Willehad Boemke’s research while affiliated with Charité Universitätsmedizin Berlin and other places

What is this page?


This page lists works of an author who doesn't have a ResearchGate profile or hasn't added the works to their profile yet. It is automatically generated from public (personal) data to further our legitimate goal of comprehensive and accurate scientific recordkeeping. If you are this author and want this page removed, please let us know.

Publications (103)


Schematic setup of the artificial lung model. Mechanical breaths from the ventilator inflate lung 1 (valve A), thereby raising a lifting bar to cause aspiration of ambient air into lung 2. Expiration into the breathing system occurs from lung 2 (valve B), while lung 1 releases to ambient air. Sites of NO injection and sampling and the distance from the ventilator or Y-piece are indicated along the breathing circuit. Injection: 20 cm distal from the ventilator, 10 cm proximal of Y-piece. Sampling: 40 and 120 cm distal from ventilator, 4 cm proximal of Y-piece, Y-piece, mid-tracheal, and artificial lung 1.
Figure adapted and modified from [13]
a Inspiratory NO concentrations achieved with “pulsed delivery”. NO was injected as a bolus during early inspiration into the breathing circuit at 20 cm after the mechanical ventilator, sampled at 120 cm thereafter, and quantified with ozone-based chemiluminescence. Ultra-low (230 ml), low (450 ml) or traditional (750 ml) tidal volumes were applied via both pressure and volume-controlled ventilation with I:E ratios of 1:1 and 1:1.9 (4 conditions). Mean NO concentrations were determined for 120 s (n = 30 respiratory cycles) per each individual ventilation condition (i.e. n = 120 per bar representing 4 ventilation conditions). Means ± SD. *p < 0.05 vs. target. b Target reliability of NO administration modes. NO was administered via “pulsed”, “flow proportional” or “continuous delivery” through injection into the breathing circuit at 20 cm after the mechanical ventilator, sampled at 120 cm thereafter, and quantified with ozone-based chemiluminescence. For each NO target concentration, mechanical ventilation was performed with ultra-low (230 ml), low (450 ml) or traditional (750 ml) tidal volumes applied via both pressure and volume-controlled ventilation with I:E ratios of 1:1 and 1:1.9 (12 conditions). NO concentrations were determined for 120 s (n = 30 respiratory cycles) per each individual ventilation condition (i.e. n = 360 per data point representing 12 ventilation conditions; n = 354 for “flow proportional delivery” at 5 ppm). The level of agreement is depicted by the mean inspiratory NO concentration expressed as a percentage of the target value ± SD. *p < 0.05 between respective modes
Sensitivity of NO quantification methods. The graph depicts the theoretical temporal resolution of ozone-based chemiluminescence (OBC, blue) versus an electrochemical sensor (ECS, black) in retracing fluctuations of the true NO concentration (red)
In vitro validation and characterization of pulsed inhaled nitric oxide administration during early inspiration
  • Article
  • Full-text available

June 2022

·

119 Reads

·

1 Citation

Journal of Clinical Monitoring and Computing

Philipp A. Pickerodt

·

Moritz B. T. Hofferberth

·

Thilo Busch

·

[...]

·

Roland C. E. Francis

Purpose Admixture of nitric oxide (NO) to the gas inspired with mechanical ventilation can be achieved through continuous, timed, or pulsed injection of NO into the inspiratory limb. The dose and timing of NO injection govern the inspired and intrapulmonary effect site concentrations achieved with different administration modes. Here we test the effectiveness and target reliability of a new mode injecting pulsed NO boluses exclusively during early inspiration. Methods An in vitro lung model was operated under various ventilator settings. Admixture of NO through injection into the inspiratory limb was timed either (i) selectively during early inspiration (“pulsed delivery”), or as customary, (ii) during inspiratory time or (iii) the entire respiratory cycle. Set NO target concentrations of 5–40 parts per million (ppm) were tested for agreement with the yield NO concentrations measured at various sites in the inspiratory limb, to assess the effectiveness of these NO administration modes. Results Pulsed delivery produced inspiratory NO concentrations comparable with those of customary modes of NO administration. At low (450 ml) and ultra-low (230 ml) tidal volumes, pulsed delivery yielded better agreement of the set target (up to 40 ppm) and inspiratory NO concentrations as compared to customary modes. Pulsed delivery with NO injection close to the artificial lung yielded higher intrapulmonary NO concentrations than with NO injection close to the ventilator. The maximum inspiratory NO concentration observed in the trachea (68 ± 30 ppm) occurred with pulsed delivery at a set target of 40 ppm. Conclusion Pulsed early inspiratory phase NO injection is as effective as continuous or non-selective admixture of NO to inspired gas and may confer improved target reliability, especially at low, lung protective tidal volumes.

Download

Introcan® Safety (IS) and Vasofix® Safety (VS), B. Braun Melsungen AG, Germany (images retrieved from https://www.bbraun.de/de/products/b0/vasofix-safety.html and https://www.bbraun.de/de/products/b/introcan-safety.html, adapted and printed with permission from B. Braun Melsungen)
A comparison of first-attempt cannulation success of peripheral venous catheter systems with and without wings and injection ports in surgical patients—a randomized trial

March 2022

·

500 Reads

·

3 Citations

BMC Anesthesiology

Background A peripheral venous catheter (PVC) is the most widely used device for obtaining vascular access, allowing the administration of fluids and medication. Up to 25% of adult patients, and 50% of pediatric patients experience a first-attempt cannulation failure. In addition to patient and clinician characteristics, device features might affect the handling and success rates. The objective of the study was to compare the first-attempt cannulation success rate between PVCs with wings and a port access (Vasofix® Safety, B. Braun, abbreviated hereon in as VS) with those without (Introcan® Safety, B. Braun, abbreviated hereon in as IS) in an anesthesiological cohort. Methods An open label, multi-center, randomized trial was performed. First-attempt cannulation success rates were examined, along with relevant patient, clinician, and device characteristics with univariate and multivariate analyses. Information on handling and adherence to use instructions was gathered, and available catheters were assessed for damage. Results Two thousand three hundred four patients were included in the intention to treat analysis. First-attempt success rate was significantly higher with winged and ported catheters (VS) than with the non-winged, non-ported design (IS) (87.5% with VS vs. 78.2% with IS; P Chi < .001). Operators rated the handling of VS as superior (rating of “good” or “very good: 86.1% VS vs. 20.8% IS, P Chi < .001). Reinsertion of the needle into the catheter after partial withdrawal—prior or during the catheterization attempt—was associated with an increased risk of cannulation failure (7.909, CI 5.989–10.443, P < .001 and 23.023, CI 10.372–51.105, P < .001, respectively) and a twofold risk of catheter damage (OR 1.999, CI 1.347–2.967, P = .001). Conclusions First-attempt cannulation success of peripheral, ported, winged catheters was higher compared to non-ported, non-winged devices. The handling of the winged and ported design was better rated by the clinicians. Needle reinsertions are related to an increase in rates of catheter damage and cannulation failure. Trial registration ClinicalTrials.gov, Identifier: NCT02213965 , Date: 12/08/2014.


Figure 1. Cannula position, blood flow, measurement of the recirculation fraction during veno-venous extracorporeal membrane oxygenation. Total extracorporeal blood flow ( Q EC ) of the extracorporeal membrane oxygenation (ECMO) is (i) pumped by the right heart into the pulmonary artery and then further into systemic circulation-resulting in an effective ECMO blood flow ( Q EFF ) and (ii) partially drained into the drainage cannula and thus recirculates through the ECMO circuit. This is called recirculation fraction (Rf). Red Arrows: oxygenated blood; deep blue arrows: deoxygenated blood; purple arrow: mixed blood; light blue arrows: saline injection. The Rf is measured with the ultrasound dilution technique requiring the injection of 0.9% saline (NaCl) into the ECMO circuit before the oxygenator. The saline bolus alters the ultrasound transit-time of blood and the transit-time is measured with two ultrasound flow probes placed around the drainage and the return tubing of the ECMO circuit. The difference in the ultrasound transit-time measurements between return and drainage cannula is used to calculate Rf.
Figure 3. Relation of oxygen delivery and oxygen consumption values to measurements of cardiac output during veno-venous extracorporeal membrane oxygenation. A: Systemic oxygen delivery (DO 2 ) values calculated using cardiac output ( ) Q measured with a pulmonary artery catheter ( Q Pa ) or measured with an ultrasound-based flow probe placed around the ascending aorta ( Q A0 ) in healthy animals at different extracorporeal membrane oxygenation (ECMO) blood flows ( Q EC ) set at defined ratios of Q EC / Q A0 . B: Calculations as in A after induction of acute lung injury (ALI). C: Systemic oxygen consumption (V̇ O 2 ) values calculated using Q Pa or Q A0 . D: Calculations as in C after induction of ALI. Values are mean ± SD; *p < 0.05 vs. ultrasound flow probe; ‡p < 0.05 vs. ECMO blood flow 25%.
Extracorporeal Membrane Oxygenation Blood Flow and Blood Recirculation Compromise Thermodilution-Based Measurements of Cardiac Output

December 2021

·

245 Reads

·

17 Citations

ASAIO Journal

The contribution of veno-venous (VV) extracorporeal membrane oxygenation (ECMO) to systemic oxygen delivery is determined by the ratio of total extracorporeal blood flow (Q˙EC) to cardiac output (Q˙). Thermodilution-based measurements of Q˙ may be compromised by blood recirculating through the ECMO (recirculation fraction; Rf). We measured the effects of Q˙EC and Rf on classic thermodilution-based measurements of Q˙ in six anesthetized pigs. An ultrasound flow probe measured total aortic blood flow (Q˙A0) at the aortic root. Rf was quantified with the ultrasound dilution technique. Q˙EC was set to 0-125% of Q˙A0 and Q˙ was measured using a pulmonary artery catheter (PAC) in healthy and lung injured animals. PAC overestimated Q˙ (Q˙Pa) at all Q˙EC settings compared to Q˙A0. The mean bias between both methods was 2.1 L/min in healthy animals and 2.7 L/min after lung injury. The difference between Q˙Pa and Q˙A0 increased with an Q˙EC of 75-125%/Q˙A0 compared to QEC <50%/Q˙A0. Overestimation of Q˙Pa was highest when Q˙EC resulted in a high Rf. Thus, thermodilution-based measurements can overestimate cardiac output during VV ECMO. The degree of overestimation of Q˙Pa depends on the Q˙EC/Q˙A0 ratio and the recirculation fraction.




Tert-butylhydroquinone augments Nrf2-dependent resilience against oxidative stress and improves survival of ventilator-induced lung injury in mice

October 2020

·

30 Reads

·

15 Citations

AJP Lung Cellular and Molecular Physiology

Background: Oxidative stress caused by mechanical ventilation contributes to the pathophysiology of ventilator-induced lung injury (VILI). A key mechanism maintaining redox balance is the up-regulation of Nrf2-dependent antioxidant gene expression. We tested whether pretreatment with a Nrf2-ARE pathway activator tert-butylhydroquinone (tBHQ) protects against VILI. Methods: Male C57BL/6J mice were pretreated with an intraperitoneal injection of tBHQ (n=10), an equivalent volume of 3% ethanol (EtOH3%, vehicle, n=13), or phosphate buffered saline (controls, n=10), and were then subjected to high tidal volume (HVT) ventilation for a maximum of 4 hours. Results: HVT ventilation severely impaired arterial oxygenation (PaO2 = 49±7 mmHg, mean±SD) and respiratory system compliance, resulting in a 100% mortality among controls. Compared with controls, tBHQ improved arterial oxygenation (PaO2 = 90±41 mmHg) and respiratory system compliance after HVT ventilation. In addition, tBHQ attenuated the HVT ventilation induced development of lung edema and pro-inflammatory response, evidenced by lower concentrations of protein and pro-inflammatory cytokines (IL-1β, TNF-α) in the bronchoalveolar lavage fluid, respectively. Moreover, tBHQ enhanced the pulmonary redox capacity, indicated by enhanced Nrf2-depentent gene expression at baseline and by the highest total glutathione concentration after HVT ventilation among all groups. Overall, tBHQ pretreatment resulted in 60% survival (p<0.001 vs. controls). Interestingly, compared with controls, EtOH3% reduced the pro-inflammatory response to HVT ventilation in the lung resulting in 38.5% survival (p=0.0054 vs. controls). Conclusions: In this murine model of VILI, tBHQ increases the pulmonary redox capacity by activating the Nrf2-ARE pathway and protects against VILI. These findings support the efficacy of pharmacological.


Fig. 1. Study flow diagram. RBCT, transfusion of RBCs.
Demographics and tumor data of the study population and the matched cohort
Surgical data and adjuvant treatment of the study population and the matched cohort
Association of perioperative transfusion of RBCs and progression-free survival according to unadjusted, multivariable and propensity score matching adjusted analyses
Transfusion of red blood cells does not impact progression‐free and overall survival after surgery for ovarian cancer

October 2019

·

103 Reads

·

13 Citations

Transfusion

Background: Allogeneic red blood cells (RBCs) have the potential to impact the immunosurveillance of the recipient and may therefore increase the risk of recurrence after cancer surgery. In this article the relationship between perioperative RBC transfusion and the risk of recurrence after ovarian cancer surgery is examined. Study design and methods: This is a retrospective cohort analysis of a prospective database of patients who underwent surgery due to primary ovarian cancer between 2006 and 2014 and who had no residual disease after surgery. Patients who did and did not receive perioperative RBC transfusion were compared. The primary endpoint was progression-free survival (PFS). Propensity score matching (PSM) and Cox proportional hazards regression (CPH) was used to control for between-group differences of prognostic determinants. Results: A total of 529 patients with a median follow-up of 51.4 months (95% CI, 46.1-56.5) were eligible for analysis. Of those, 408 patients (77.1%) received allogeneic, leukoreduced RBCs with a median of 4 units (IQR, 2-6) per patient. There was a strong selection bias of prognostic determinants between patients with and without transfusion. In unadjusted analysis, transfusion of RBCs was associated with an increased risk of cancer recurrence (hazard ratio [HR] of PFS 2.71 [95% CI, 1.94-3.77], p < 0.001). After bias reduction, transfusion of RBCs was no longer associated with an increased risk of cancer recurrence, neither in PSM-adjusted (HR 1.03 [95% CI, 0.59-1.80], p = 0.91), nor in multivariable CPH-adjusted analysis (HR 1.26 [95% CI, 0.85-1.86], p = 0.23). Conclusion: Perioperative transfusion of RBCs did not increase the risk of recurrence after ovarian cancer surgery.


Anästhesiologische Beurteilung des Patienten: Blutgasanalyse und Säure-Basen-Haushalt

April 2019

·

31 Reads

·

1 Citation

Chemische Grundlagen des Säure-Basen-Haushalts werden in diesem Kapitel ebenso erläutert, wie die unterschiedlichen Puffersysteme und Transport und Elimination von Kohlendioxid. Die Funktionen der Niere bei der Regulation des pH-Wertes sowie der Oxygenierungsstatus des Bluts werden erörtert. Zudem werden die typischen Störungen des Säure-Basen-Haushalts bei Patienten vorgestellt.


a System configuration and data flow for the automatized closed-loop control of mechanical ventilation. Blue arrows indicate signal flow to the controller (afferent signals) and red arrows indicate signal flow from the controller to steer the mechanical ventilation and norepinephrine infusion (efferent signals). Adapted and reproduced with permission from [8]. b Closed-loop mechanical ventilation using the open lung approach (OLA). A brief opening phase is used to recruit previously collapsed alveoli and is followed by identification of the critical PEEP value below which lung derecruitment occurs (closing phase). Lung recruitment is repeated (reopening) with the PEEP value set 2 cm H2O above the closing PEEP, followed by mechanical ventilation controlled by closed-loop physiological feedback algorithms (ventilation phase). PIP peak inspiratory pressure, PEEP positive end-expiratory pressure
a–c Time course after induction of lung injury (baseline lung injury) of PaO2/FIO2-ratios, PaCO2 and arterial pH measurements for animals ventilated according to the open lung algorithm (OLA). FIO2, fraction of inspired oxygen; PaO2, arterial partial pressure of oxygen; PaCO2, arterial partial pressure of carbon dioxide; pH, negative decimal logarithm of H⁺ ion concentration. Values are median and interquartile range except for pH values. pH values are depicted as the negative decimal logarithm of the mean H⁺ ion concentration. *P < 0.05 versus baseline lung injury
a–c Time course after induction of lung injury (baseline lung injury) of applied tidal volumes, end-inspiratory and end-expiratory airway pressures and lung driving pressures (ΔP) for animals ventilated according to the open lung algorithm (OLA). VT tidal volume, ∆P driving pressure. Values are median and interquartile range. *P < 0.05 versus baseline lung injury
Arterial partial pressure of oxygen/fraction of inspired oxygen (PaO2/FIO2) and ratio of ventral/dorsal (V/D) ventilation changes in the ventral- and dorsal region of interest (ROI) after lung injury induction (baseline lung injury) and at the end of the experiment. Values are medians and interquartile ranges of animals ventilated according to the open lung algorithm (OLA). FIO2 fraction of inspired oxygen, PaO2 arterial partial pressure of oxygen, V/D ratio of ventral/dorsal ventilation changes, ROI region of interest
Closed-loop mechanical ventilation for lung injury: a novel physiological-feedback mode following the principles of the open lung concept

June 2018

·

327 Reads

·

11 Citations

Journal of Clinical Monitoring and Computing

Adherence to low tidal volume (VT) ventilation and selected positive end-expiratory pressures are low during mechanical ventilation for treatment of the acute respiratory distress syndrome. Using a pig model of severe lung injury, we tested the feasibility and physiological responses to a novel fully closed-loop mechanical ventilation algorithm based on the "open lung" concept. Lung injury was induced by surfactant washout in pigs (n = 8). Animals were ventilated following the principles of the "open lung approach" (OLA) using a fully closed-loop physiological feedback algorithm for mechanical ventilation. Standard gas exchange, respiratory- and hemodynamic parameters were measured. Electrical impedance tomography was used to quantify regional ventilation distribution during mechanical ventilation. Automatized mechanical ventilation provided strict adherence to low VT-ventilation for 6 h in severely lung injured pigs. Using the "open lung" approach, tidal volume delivery required low lung distending pressures, increased recruitment and ventilation of dorsal lung regions and improved arterial blood oxygenation. Physiological feedback closed-loop mechanical ventilation according to the principles of the open lung concept is feasible and provides low tidal volume ventilation without human intervention. Of importance, the "open lung approach"-ventilation improved gas exchange and reduced lung driving pressures by opening atelectasis and shifting of ventilation to dorsal lung regions.


Arterial gas exchange parameters and ventilation
Time course of mean arterial blood partial pressure of oxygen (Pao2) (A), mean arterial blood partial pressure of carbon dioxide (P ac o2) (B), expiratory minute ventilation (V̇E) (C) and differences of arterial to alveolar partial pressure of oxygen (Pa− Ac o2) (D). Measurements were performed in 30 min intervals during 1 h of hyperoxia (FI,o2 = 0.5) and during 3 h of isobaric hypoxia (FI,o2 = 0.145–0.12) in control animals (n = 8), and pigs treated with i.v. ACZ (n = 7), inhaled NaNO2 (n = 6) and combined i.v. ACZ + iNaNO2 (n = 6). Values are means ± SD. *P < 0.05 vs. hyperoxia; †P < 0.05 vs. controls.
Pulmonary haemodynamics and lung nitric oxide release
Time course of mean pulmonary arterial pressure (MPAP) (A), mean pulmonary vascular resistance (PVR) (B), and nitric oxide release rate (NO release) (C), Measurements were performed continuously and values were obtained in 30 min intervals during 1 h of hyperoxia (FI,o2 = 0.5) and during 3 h of isobaric hypoxia (FI,o2 = 0.145–0.12) in control animals (n = 8), and pigs treated with i.v. ACZ (n = 7), inhaled NaNO2 (n = 6) and combined i.v. ACZ + iNaNO2 (n = 6). Values are means ± SD. *P < 0.05 vs. hyperoxia; †P < 0.05 vs. controls.
Carbonic anhydrase is not a relevant nitrite reductase or nitrous anhydrase in the lung

May 2018

·

47 Reads

·

11 Citations

Key points Carbonic anhydrase (CA) inhibitors such as acetazolamide inhibit hypoxic pulmonary vasoconstriction (HPV) in humans and other mammals, but the mechanism of this action remains unknown. It has been postulated that carbonic anhydrase may act as a nitrous anhydrase in vivo to generate nitric oxide (NO) from nitrite and that this formation is increased in the presence of acetazolamide. Acetazolamide reduces HPV in pigs without evidence of any NO generation, whereas nebulized sodium nitrite reduces HPV by NO formation; however; combined infusion of acetazolamide with sodium nitrite inhalation did not further increase exhaled NO concentration over inhaled nitrite alone in pigs exposed to alveolar hypoxia. We conclude that acetazolamide does not function as either a nitrous anhydrase or a nitrite reductase in the lungs of pigs, and probably other mammals, to explain its vasodilating actions in the pulmonary or systemic circulations. Abstract The carbonic anhydrase (CA) inhibitors acetazolamide and its structurally similar analogue methazolamide prevent or reduce hypoxic pulmonary vasoconstriction (HPV) in dogs and humans in vivo, by a mechanism unrelated to CA inhibition. In rodent blood and isolated blood vessels, it has been reported that inhibition of CA leads to increased generation of nitric oxide (NO) from nitrite and vascular relaxation in vitro. We tested the physiological relevance of augmented NO generation by CA from nitrite with acetazolamide in anaesthetized pigs during alveolar hypoxia in vivo. We found that acetazolamide prevents HPV in anaesthetized pigs, as in other mammalian species. A single nebulization of sodium nitrite reduces HPV, but this action wanes in the succeeding 3 h of hypoxia as nitrite is metabolized and excreted. Pulmonary artery pressure reduction and NO formation as measured by exhaled gas concentration from inhaled sodium nitrite were not increased by acetazolamide during alveolar hypoxia. Thus, our data argue against a physiological role of carbonic anhydrase as a nitrous anhydrase or nitrite reductase as a mechanism for its inhibition of HPV in the lung and blood in vivo.


Citations (70)


... After induction of anesthesia and instrumentation, hypoxia trial one was performed in healthy animals. Thereafter, lung injury was induced by saline lavage-induced surfactant depletion of the lungs [26]. The Berlin definition of severe ARDS, i.e., a aO 2 / IO 2 ≤ 100 mmHg with PEEP ≥ 5 cmH 2 O, was used to quantify the respiratory failure, followed by hypoxia trial two. ...

Reference:

Robust closed-loop control of systemic oxygenation in acute lung injury
Lavage-induced Surfactant Depletion in Pigs As a Model of the Acute Respiratory Distress Syndrome (ARDS)
  • Citing Article
  • September 2016

Journal of Visualized Experiments

... First, cardiac output, which might influence monitored lung perfusion, was not assessed in this cohort. However, considering the short study duration, the patients being deeply sedated and paralyzed, the stable circulatory conditions, and the negligible effect of VV ECMO blood flow on systemic circulation [10,13], it is assumed that the cardiac output remained constant during the trial period. Second, we did not perform repeated measurements at each blood flow rate to reduce experimental error. ...

Extracorporeal Membrane Oxygenation Blood Flow and Blood Recirculation Compromise Thermodilution-Based Measurements of Cardiac Output

ASAIO Journal

... Standard ARDS models are either animal-based (D'Alessio, 2018) or rely on transwell-insert models (Ishii et al., 2021). These models induce direct injury through mechanical ventilation or damage-inducing agents such as LPS, or indirectly through intravenous injection or shock (Russ et al., 2021). Despite advancements, more advanced models resembling human physiology are needed to better understand ARDS in humans, as animal immune systems differ (Masopust et al., 2017;Mestas and Hughes, 2004). ...

Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome (ARDS)
  • Citing Article
  • April 2021

Journal of Visualized Experiments

... In addition, this classification is valuable for understanding the substantial impact of ventilators flowby on iNO delivery. This concern has never been systematically investigated but deserves to be addressed because of the increasing variety and heterogeneity of ICU ventilators [28]. ...

In vitro validation and characterization of pulsed inhaled nitric oxide administration during early inspiration

Journal of Clinical Monitoring and Computing

... Li et al. 2020;Pérez-Rojas et al. 2011). Lilly Veskemaa et al. (2021) found that the administration of TBHQ lessened oxidative stress and improved the survival of ventilator-induced lung damage in mice (Veskemaa et al. 2021). ...

Tert-butylhydroquinone augments Nrf2-dependent resilience against oxidative stress and improves survival of ventilator-induced lung injury in mice
  • Citing Article
  • October 2020

AJP Lung Cellular and Molecular Physiology

... Surgery is therefore often extensive, enhancing intraoperative blood loss and consequently resulting in postoperative anemia [4,5]. Perioperative red blood cell (RBC) transfusions are common in patients with AOC, with a transfusion rate of 40-77% [6,7]. ...

Transfusion of red blood cells does not impact progression‐free and overall survival after surgery for ovarian cancer

Transfusion

... 12,28,29 Few studies have investigated the effect of blood transfusion therapy on improving hemoglobin. 14,15 This is the first study to evaluate the effects of the same blood transfusion bag on preterm infants. Similar to previous results, hemoglobin was most significantly elevated after two transfusions with a 1 week interval compared to other time intervals. ...

Duration of storage influences the hemoglobin rising effect of red blood cells in patients undergoing major abdominal surgery
  • Citing Article
  • April 2018

Transfusion

... Hypoxia exposure increases pulmonary vascular resistance and pulmonary artery pressure (PAP) in humans and other mammals due to hypoxic pulmonary vasoconstriction (HPV) (Swenson, 2013). Several studies in animals and humans have demonstrated that AZ attenuates HPV (Boulet et al, 2018;Höhne et al, 2007;Pickerodt et al, 2019;Pickerodt et al, 2014;Teppema et al, 2007;Teppema and Swenson, 2015;Tremblay et al, 2015). Animal studies further suggest that the reduction in HPV is independent of CA inhibition (Höhne et al, 2007;Pickerodt et al, 2014;Shimoda et al, 2007). ...

Carbonic anhydrase is not a relevant nitrite reductase or nitrous anhydrase in the lung

... A CDSS or clinical protocol can be programmed to run in a closed-loop form. Examples would be the automatic ARDSNet protocol system [21] or the automation of the open lung concept [22]. Automatic closed-loop control are often focused on either oxygenation [23,24] or ventilation (PETCO 2 ) [25][26][27]. ...

Closed-loop mechanical ventilation for lung injury: a novel physiological-feedback mode following the principles of the open lung concept

Journal of Clinical Monitoring and Computing

... For example, pulmonary lavage-induced surfactant depletion is a widely used porcine model of acute hypoxemic lung failure. However, rapid recovery of respiratory system compliance (C rs ) and blood oxygenation occur with this model after alveolar recruitment [6][7][8][9], indicating that the gas exchange abnormalities reflect collapsed alveoli with otherwise intact alveolar walls and high recruitability. In contrast, the clinical presentation of ARDS in patients typically includes alveolar injury and nonaerated lung tissue with low C rs and limited recruitability [10]. ...

Lavage-induced Surfactant Depletion in Pigs As a Model of the Acute Respiratory Distress Syndrome (ARDS)
  • Citing Article
  • September 2016

Journal of Visualized Experiments