This page lists works of an author who doesn't have a ResearchGate profile or hasn't added the works to their profile yet. It is automatically generated from public (personal) data to further our legitimate goal of comprehensive and accurate scientific recordkeeping. If you are this author and want this page removed, please let us know.

Publications (1)


Figure 1. Typical chemiluminescence analyzer.
Figure 2. 
Table 3 . CO 2 interference on NO x readings at 5 ppm NO.
Figure 4. NO x analyzer testing apparatus.
Table 4 . Estimated NO x detection limits and linearity.

+2

Evaluation of NO x Flue Gas Analyzers for Accuracy and Their Applicability for Low-Concentration Measurements
  • Article
  • Full-text available

July 2003

·

2,498 Reads

·

34 Citations

·

Chuck Glenn

·

Tim Logan

·

[...]

·

Pat Williams

The requirements of the Texas State Implementation Plan of the U.S. Clean Air Act for the Houston-Galveston Ozone Nonattainment Area stipulate large reductions in oxides of nitrogen (NO(x)) emissions. A large number of sources at Dow Chemical Co. sites within the nonattainment area may require the addition of continuous emission monitoring systems (CEMS) for online analysis of NO(x), CO, and O2. At the outset of this work, it was not known whether the analyzers could accurately measure NO(x) as low as 2 ppm. Therefore, NO(x) CEMS analyzers from five different companies were evaluated for their ability to reliably measure NO(x) in the 2-20 ppm range. Testing was performed with a laboratory apparatus that accurately simulated different mixtures of flue gas and, on a limited basis, simulated a dual-train sampling system on a gas turbine. The results indicate that this method is a reasonable approach for analyzer testing and reveal important technical performance aspects for accurate NO(x) measurements. Several commercial analyzers, if installed in a CEMS application with sampling conditioning components similar to those used in this study, can meet the U.S. Environmental Protection Agency's measurement data quality requirements for accuracy.

Download

Citations (1)


... Comparing the FTIR analyzer with the CLA at the tailpipe, the former recorded, on average, 8 mg/km higher values. CLAs have some uncertainties due to the use of NO2-to-NO converters with conversion efficiencies >95% (which need to be checked every month according to EU regulation), possible NO2 losses in their chiller, and influence exerted by quenching with water and CO2 [115]. FTIR instruments' NO2 measurement is affected by water vapor interference. ...

Reference:

Measurement of Light-Duty Vehicle Exhaust Emissions with Light Absorption Spectrometers
Evaluation of NO x Flue Gas Analyzers for Accuracy and Their Applicability for Low-Concentration Measurements