ArticlePDF Available

Comparison of Luneau SA disposable and Goldmann applanation tonometer readings

Authors:

Abstract and Figures

To test the agreement of intraocular pressure (IOP) measurements made with Luneau SA applanators and Goldmann applanator. A single-blind crossover trial. IOPs were measured in both eyes of subjects with both applanators. Type of applanator was alternated to eliminate systematic bias. Multiple observers were used. Observers were blind to the scale while performing measurements but not to the type of applanator used. The appearance of the meniscus was assessed semiquantitatively. All measurements were combined and presented in a Bland-Altman plot. A total of 140 eyes of 79 subjects were tested by seven observers. The range of measurements was 6-45 mmHg (mean 17.8 mmHg) for the Goldmann applanator. On average, the Luneau SA applanator (range of measurements 4-36 mmHg) gave a measurement of 2.35 mmHg less than the Goldmann standard. The standard deviation of these differences was 2.13 mmHg, giving an upper 95% confidence limit of 6.53 mmHg and a lower 95% confidence limit of -1.83 mmHg. The measurements agreed in only 24 out of 140 instances. In 28 eyes, the disposable tonometer end point was difficult to assess owing to excessively thick rings. Linear extrapolation suggests an increase in difference with increasing IOP. The inter-head inaccuracy, tendency to underestimate IOP, and lack of systematic inaccuracy make a corrective algorithm impossible to formulate. The range of variation between the Luneau SA disposable applanator and the Goldmann standard is sufficiently large to influence clinical management decisions. We speculate that one explanation is the interaction of the tonometer with the tear film, making end point determination difficult. Further research is being undertaken.
Content may be subject to copyright.
Comparison of
Luneau SA
disposable and
Goldmann
applanation
tonometer readings
ACJ Baddon, SF Osborne, SA Quah, M Batterbury
and D Wong
Abstract
Purpose To test the agreement of intraocular
pressure (IOP) measurements made with
Luneau SA applanators and Goldmann
applanator.
Materials and Methods A single-blind
crossover trial. IOPs were measured in both
eyes of subjects with both applanators. Type
of applanator was alternated to eliminate
systematic bias. Multiple observers were used.
Observers were blind to the scale while
performing measurements but not to the type
of applanator used. The appearance of the
meniscus was assessed semiquantitatively. All
measurements were combined and presented
in a Bland–Altman plot.
Results A total of 140 eyes of 79 subjects
were tested by seven observers. The range of
measurements was 6–45 mmHg (mean
17.8 mmHg) for the Goldmann applanator. On
average, the Luneau SA applanator (range of
measurements 4–36 mmHg) gave a
measurement of 2.35 mmHg less than the
Goldmann standard. The standard deviation
of these differences was 2.13 mmHg, giving an
upper 95% confidence limit of 6.53 mmHg and
a lower 95% confidence limit of 1.83 mmHg.
The measurements agreed in only 24 out of 140
instances. In 28 eyes, the disposable tonometer
end point was difficult to assess owing to
excessively thick rings. Linear extrapolation
suggests an increase in difference with
increasing IOP.
Conclusion The inter-head inaccuracy,
tendency to underestimate IOP, and lack of
systematic inaccuracy make a corrective
algorithm impossible to formulate. The range
of variation between the Luneau SA
disposable applanator and the Goldmann
standard is sufficiently large to influence
clinical management decisions. We speculate
that one explanation is the interaction of the
tonometer with the tear film, making end point
determination difficult. Further research is
being undertaken.
Eye (2007) 21, 789–792; doi:10.1038/sj.eye.6702329;
published online 21 April 2006
Keywords: Luneau SA; Goldmann; applanation
tonometer
Introduction
The use of disposable instruments in the
practice of ophthalmology has increased in
response to fears of iatrogenic horizontal
transmission of prion material between patients.
It has been suggested that the very common
practice of measurement of intraocular pressure
(IOP) by Goldmann prism applanation
tonometry may lead to transfer of potentially
infective material.
1–4
Because of this risk, the
Medical Devices Agency has recommended that
‘components of ophthalmic devices that touch
the surface of the eye should be restricted to
single patient use wherever practicable and
where this does not compromise clinical
outcome’.
4
There are a number of disposable
applanation tonometers in use in the NHS.
Silicone and acrylic disposable applanation
tonometers have been the subject of previous
study. Acrylic prisms (Tonosafe
s
) were found
to have good agreement with the Goldmann
applanation tonometer, but silicone
(Tonoshield
s
) prisms were found to
consistently over read by a small margin.
5
The Luneau SA disposable tonometer has
become available. It is placed in the Goldmann
tonometer device in the same way as the
Goldmann prism. The objective of this study
Received: 10 May 2005
Accepted in revised form:
1 February 2006
Published online: 21 April
2006
No proprietary interests
No research funding
Presented as Poster and
Quickfire Presentation,
Royal College of
Ophthalmologists Congress
2005
Department of
Ophthalmology, St Paul’s
Eye Unit, The Royal
Liverpool University
Hospital, Liverpool, UK
Correspondence:
ACJ Baddon,
Department of
Ophthalmology,
Walton Hospital,
Rice Lane, Walton,
Liverpool L9 1AE, UK
Tel: þ 779 860 7703
(Mobile), þ 151 424 1559
(Home);
Fax: þ 779 860 9051.
E-mail: ophthalmologist@
onetel.com
Eye (2007) 21, 789–792
& 2007 Nature Publishing Group All rights reserved 0950-222X/07 $30.00
www.nature.com/eye
CLINICAL STUDY
was to compare the IOP readings of the two devices and
to assess the end point features of each.
Materials and methods
The IOPs of a consecutive series of patients were
measured with both the Luneau SA disposable
applanator tonometer (Figure 1) and Goldmann
applanator tonometer. Each patient’s pressures were
measured by only one of the seven observers. All
observers were of consultant or specialist registrar
grades. The type of applanator and the side measured
first were alternated to eliminate systematic bias. The
tonometers’ calibration was checked before and after the
study by use of the calibration bar. The measurements
were made by setting the tonometer dial to zero and then
applanating the cornea and aligning the inner aspect of
the two half menisci. Measurements were taken with the
eye in the primary position and apart from the
applanation, there was no external pressure exerted on
the eyeball. The observers were blinded to the scale on
the tonometer dial. Observers were not blinded to the
type of applanator being used. The appearance of the
menisci was assessed semiquantitavely, without
reference to a standard photograph, by all observers as
too thin, too thick, or acceptable. To illustrate agreement,
all measurements were combined and presented in a
Bland–Altman plot
6
with calculation of 95% confidence
limits of agreement. A paired t-test was used to analyse
the difference between measurements.
Results
A total of 140 eyes of 79 subjects were tested by seven
observers. The range of measurements was 6–45 mmHg
(mean 17.7 mmHg) for the Goldmann applanator. The
range of measurements for the Luneau SA applanator
was 4–36 mmHg (mean of 16.3 mmHg). On average, the
Luneau SA applanator gave a measurement of
2.35 mmHg less than the Goldmann standard. The range
of difference in these paired readings was from 6to
þ 9 mmHg. The standard deviation of these differences
was 2.13 mmHg, giving an upper 95% confidence limit of
6.53 mmHg and a lower 95% confidence limit of
1.83 mmHg for measurements made with the
disposable prism in comparison to the Goldmann
standard. The measurements agreed in only 24 out of 140
instances (Figure 2). A paired, two-tailed t-test showed
that the difference between the means of the two data
sets was very highly significant (Po0.001) (Table 1). In 28
eyes, the disposable tonometer end point was difficult to
assess owing to excessively thick rings. In three eyes, thin
poor quality rings were seen with the disposable
tonometer. With these paired data removed, a paired
Figure 1 Luneau tonometer prism.
+9
-6
y = 0.0816x
-6.5
-4.5
-2.5
-0.5
5 101520253035404550
1.5
3.5
5.5
7.5
9.5
Goldman IOP (mmHg) ->
Difference between Goldman and Disposable
(ie Positive numbers = Goldman reading higher)
Figure 2 Bland–Altman plot.
Comparison of Luneau SA and Goldmann applanators
ACJ Baddon
790
Eye
t-test shows that the statatistical significance remains
high (Po0.001). Linear extrapolation suggests an
increase in measurement difference with increasing IOP.
Discussion
Our results show that, under conditions of normal
clinical use, the Luneau SA disposable tonometer
underestimates IOP compared with the reference
standard Goldmann nondisposable tonometer in a
nonsystematic way that, in many eyes, is clinically
significant. While only 13.6% of our readings were above
21 mmHg, accurately recording a lower pressure may, in
many cases, have clinical significance.
In this study, there was no measure made to assess
inter-observer agreement. Neither were the observers
blinded to the prism type, owing to the differing
appearance of the Luneau SA disposable prism as
compared to the Goldmann prism. Judgements on the
quality of menisci were not standardised against
photographs of a poor or satisfactory meniscus, although
all observers were ophthalmologists who were very
experienced in applanation tonometry.
Our findings are contrary to those of Maino et al,
7
who
found good agreement between disposable and
nondisposable tonometer prisms by practitioners trained
in applanation tonometry; however, each eye received
one applanation from each of the three types of prism in
every case in random order. Our readings, although
lesser in total number, were mostly within the normal
range, from a population of a catchment area adjacent to
that of Maino et al.
Our impression before setting up this study was that
for the Luneau SA prism the end point of measurement,
the alignment of the semicircular menisci, was difficult to
determine because of the occurrence of thick rings. This
impression is supported by observation in our results.
The disposable prism menisci appearance was judged to
be of acceptable quality in only 80% of cases, whereas
there was no case of the Goldmann menisci reported as
being of poor quality.
Goldmann applanation tonometry relies on the Imbert-
Fick principle. Imbert-Fick states that the force of
flattening (F) equals the pressure (P) multiplied by the
surface area flattened (A), or F ¼ P A. However, this
principle assumes that the surface is dry, perfectly
flexible, infinitely thin, and spherical. In the design of the
Goldmann applanator prism, two additional factors are
included: the surface tension of the tear film, which tends
to draw the tonometer head toward the cornea and
corneal rigidity, and the tendency of the cornea to resist
being flattened during applanation. Goldmann assumes
the corneal thickness to be exactly 520 mm and that all
corneas are equally rigid with parallel anterior and
posterior surfaces. Given these assumptions, at a
diameter of applanation of 3.06 mm, scleral rigidity and
surface tension cancel each other out.
8
We speculate that one aspect of the poor agreement
between the Luneau SA and the Goldmann is the
interaction between the prism and the tear film such that
the surface tension is altered. Not only might this change
the measurement of IOP
8–11
but also may result in thick
menisci that make end point determination inaccurate.
We also speculate that different combinations of
fluorescein and local anaesthetic agent might lead to a
range of tear film surface tensions, which might affect the
accuracy of the IOP measurement. Further studies are
being undertaken to test these hypotheses.
Design of disposable applanation prisms may
necessitate the use of different materials or the surface
modification of current materials and this in turn might
require the modification of the body of the tonometer or
its dial. This modification would of course require that
there be a systematic error. In the case of a nonconstant
error in tonometry measurement, it would seem logical
to balance the risk of horizontal transmission of
infections between patients against the risk of the over-
or underestimation of their ocular pressure and the
impact of this on treatment. The materials we received
from Luneau did not include any specification as to the
level of accuracy expected with their prism.
The inter-head inaccuracy, tendency to underestimate
IOP, and lack of systematic inaccuracy make a corrective
algorithm impossible to formulate. The range of variation
between the Luneau SA disposable applanator and the
Goldmann standard is sufficiently large to influence
clinical management decisions. This lack of agreement
means that the Luneau SA prism in its present form
cannot replace the nondisposable prism.
References
1 Lim R, Dhillon B, Kurian KM, Aspinall PA, Fernie K,
Ironside JW. Retention of corneal epithelial cells following
Goldmann tonometry: implications for CJD risk. Br J
Ophthalmol 2003; 87(5): 583–586.
Table 1 Summary of results
N ¼ 140 eyes Goldmann
(G)
Disposable
(D)
Difference
(GD)
Mean 17.7 16.34 2.35
Standard deviation 2.13
95% CI 74.17
Range 6 to 45 4 to 36 6to þ 9
For paired t-test, t ¼ 13.69, Po0.001.
All values are in mmHg.
Comparison of Luneau SA and Goldmann applanators
ACJ Baddon
791
Eye
2 Manuelidis EE, Angelo JN, Gorgacz EJ, Kim JH, Manuelidis
L. Experimental Creutzfeldt–Jakob disease transmitted via
the eye with infected cornea. N Engl J Med 1977; 296:
1334–1336.
3 Walia JS, Chronister CL. Possible iatrogenic transmission of
Creutzfeldt–Jakob disease via tonometer tips: a review of
the literature. Optometry 2001; 72: 649–652.
4 Medical Device Agency AN 1999 (03). Single patient use of
ophthalmic devices: implications for clinical practice.
Medical Devices Agency, October 1999.
5 Bhatnagar A, Gupta AK. Disposable devices for measuring
intraocular pressure: a clinical study to assess their
accuracy. Eye 2005; 19(7): 752–754.
6 Bland JM, Altman DG. Statistical methods for assessing
agreement between two methods of clinical measurement.
Lancet 1986; 1: 307–310.
7 Maino AP, Uddin HJ, Tullo AB. A comparison of clinical
performance between disposable and Goldmann
tonometers. Eye 2005 [E-pub ahead of print].
8 Viernstein LJ, Pollack IP. Validity of the Imbert-Fick law for
constant-pressure tonography. Exp Eye Res 1979; 29(6):
587–594.
9 Whitacre MM, Stein R. Sources of error with use of
Goldmann-type tonometers. Surv Ophthalmol 1993; 38: 1–30.
10 Birchall W, Kumar V. A comparative study of
proxymetacaine–fluorescein and lignocaine–fluorescein use
during applanation tonometry. Br J Ophthalmol 2001; 85(4):
477–479.
11 Goel S, Chua C, Dong B, Butcher M, AhFat A, Hindi SK et al.
Comparison between standard Goldmann applanation
prism and disposable applanation prism in tonometry. Eye
2004; 18(2): 175–178.
Comparison of Luneau SA and Goldmann applanators
ACJ Baddon
792
Eye
... Studies comparing the Luneau Tonojet to reusable tip GAT found a mean IOP difference of 0.4-2.35 mmHg [26,32,33]. There are several important considerations in the use of disposable tips. ...
... In their study, they inspected the prisms prior to use and found that 14 of 189 disposable tips had to be discarded because of the presence of surface irregularities. Baddon et al. [33] found that in 28 of the 140 eyes included in the study, the tonometer endpoint was difficult to assess with Luneau prisms because of excessively thick rings. The menisci endpoint was deemed to be of acceptable quality in 80% of the Luneau tip readings and 100% of the GAT readings. ...
Article
Full-text available
Purpose of review: Accurate and precise measurement of intraocular pressure (IOP) is a vitally important component of the ophthalmic examination. There are multiple methods of tonometry, each of which has considerations in light of the ongoing severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) pandemic. This review discusses these considerations and compares various tonometer methods with the gold standard of Goldmann applanation tonometry (GAT). Recent findings: The SARS-CoV-2 virus may spread via droplets, microaerosols, or direct contact in the ophthalmology clinic. Tonometry poses a high risk of contamination. The accuracy and reliability of various methods of tonometry with single-use disposable equipment has been compared with Goldmann applanation tonometry. Summary: Goldmann applanation tonometry with disposable applanation tips, Tono-pen, and iCare employ single use tips to decrease the risk of cross-contamination of infectious agents. Review of the literature demonstrates good correlation between these devices and GAT, although the published level of agreement between devices varies.
... [21][22][23] Literature published on the use of dGAT in glaucomatous populations, however, shows a lesser agreement between dGAT and GAT and a tendency for underestimation of IOP with dGAT, differing from the findings in our present study. [24][25][26][27] In our study we found that dGAT provided good agreement with GAT in both glaucomatous populations (ICC of 0.96) and nonglaucomatous populations (ICC of 0.93). At extreme IOP values (< 10 or > 24 mm Hg), this agreement (ICC) decreased significantly with a wider CI. ...
Article
Purpose: To compare intraocular pressure measurements by Goldmann applanation tonometer, disposable Goldmann applanation prism, ICare, and Tonopen. Materials and methods: A total of 74 patients with varying glaucoma status were examined in our outpatient clinic and intraocular pressure was measured with three tonometers. The disposable Tonojet prism, Tonopen XL, and ICare ic100 were compared with the Goldmann applanation tonometer. Results: There were good intraclass correlation coefficients between intraocular pressure measurements by Goldmann applanation tonometer and disposable Tonojet prisms (0.95), Tonopen (0.83), and ICare (0.77), all P<0.001. The intraocular pressure mean differences between disposable Tonojet prisms and Goldmann applanation tonometer were mean 0.80▒mm Hg; 95% limits of agreement: -3.35-4.96▒mm Hg. For Tonopen and Goldmann applanation tonometer: mean -1.67▒mm Hg; limits of agreement: -8.55-5.21▒mm Hg. For ICare and Goldmann applanation tonometer: mean 0.44▒mm Hg; limits of agreement: -8.18-9.06▒mm Hg. Conclusion: The most reliable modality, with good correlation with the Goldmann tonometer values, was the Goldmann applanation tonometer with disposable Tonojet prisms, followed in descending order by the Tonopen-XL and ICare. There was good interdevice agreement and consistency between all devices. On subgroup analysis, all three modalities were found to be less reliable at extreme IOP values (<10▒mm Hg and >24▒mm Hg). These disposable modalities should be avoided in extreme intraocular pressure ranges outside the normal range.
... [21][22][23] Literature published on the use of dGAT in glaucomatous populations, however, shows a lesser agreement between dGAT and GAT and a tendency for underestimation of IOP with dGAT, differing from the findings in our present study. [24][25][26][27] In our study we found that dGAT provided good agreement with GAT in both glaucomatous populations (ICC of 0.96) and nonglaucomatous populations (ICC of 0.93). At extreme IOP values (< 10 or > 24 mm Hg), this agreement (ICC) decreased significantly with a wider CI. ...
Article
To assess agreement of intraocular pressure (IOP) measured with Goldmann applanation and disposable Luneau SA applanation prisms in glaucoma patients compared with normal subjects. In a prospective comparative study, one eye per primary open-angle glaucoma (POAG) patient was measured for IOP using Goldmann and disposable Luneau SA (Tonojet) applanation tonometry prisms. Glaucoma patients found with IOP < 21 mmHg were compared to an age-matched group of normal subjects. The main outcomes measured were IOP and the difference in IOP between either prism. Bland-Altman plots were used to study the agreement between measurements. Regression analysis was applied to either group. One hundred and four POAG patients (eyes) were enrolled into this study. Seventy-one of them were found with IOPs < 21 mmHg, and were compared to 71 normal subjects (eyes). The mean difference between either prism was almost similar, 2.0 ± 2.0 and 2.1 ± 1.7 mmHg in the glaucoma and normal eye groups, respectively. Bias of two SD from the mean ranged from -1.8 to 5.8 mmHg in glaucoma group and -1.2 to 5.4 mmHg with normal subjects (p < 0.0001 in either group). The inaccuracy was non-systematic with glaucoma patients while was systematic with normal subjects. The disagreement between disposables and standard Goldmann tips followed different patterns in glaucoma patients when compared with normal subjects. The use of the Goldmann tip is advised with glaucoma patients.
Article
Full-text available
Applanation tonometry is widely performed as the most accurate method of measuring intraocular pressure in the clinical setting. Disposable tonometer heads reduce the risk of transmitting infection during applanation and eliminate tonometer head cleaning, which is time-consuming. We performed a study to investigate whether the Tonojet disposable tonometer (Luneau, France) gave comparable intraocular pressure measurements to the standard Goldmann tonometer produced by Haag Streit (Switzerland). The study was performed in a busy general secondary referral ophthalmology clinic in Beaumont Hospital over a 15-day period. We measured intraocular pressure with the Goldmann and Tonojet tonometers in 293 eyes. Patients were randomised as to which tonometer was used first. A diagnosis of glaucoma or ocular hypertension was noted. Central corneal thickness (CCT) was measured. The measurements taken with the Tonojet were on average 2.2 mmHg lower than corresponding measurements taken with the Goldmann applanation tonometer head. The bias was present over the entire range of pressures. The 95 % limits of agreement between the two tonometers ranged from -5.9 to +1.4 mmHg. In the cohort of eyes with ocular hypertension/glaucoma the difference was significantly higher (2.8 vs. 2.1 mmHg). The bias was independent of CCT. We found that the Tonojet disposable tonometer tends to underestimate intraocular pressure compared with the Goldmann. The bias is non-systematic, present over the entire range of pressures and independent of corneal thickness. We conclude that the Tonojet is not sufficiently accurate to constitute a suitable alternative to the Goldmann.
Article
The definition of glaucoma has changed over the decades from a simple ocular pressure disease to a systemic disorder of multivariate etiology. Glaucoma may be defined for the individual eye as a chronic ocular disease with various underlying pathophysiologic disorders. However, elevated intraocular pressure (IOP) is still the most important risk factor for an untreated glaucomatous eye to progress to a more severe stage of the disease. As the main risk factor within therapeutic reach, IOP and its appropriate measurement deserve our ongoing interest. Not only has our understanding of glaucoma changed but also our approach to the measurement of the IOP. In this article we focus our attention on the various developments in tonometry from the simple force-tonometers of the late 19th century to the high-technology pressure tonometers that were recently introduced for clinical use.
Article
Disposable tonometers are increasingly being adopted partly because of concerns over the transmission of variant Creutzfeldt-Jakob disease and partly for convenience. Recently, we have found one such tonometer (Tonojet by Luneau Ophthalmologie, France) underestimated the intraocular pressure (IOP). We hypothesized that this underestimation was caused by a difference in the surface property of the tonometers. A tensiometer was used to measure the suction force resulting from interfacial tension between a solution of lignocaine and fluorescein and the tonometers. The results showed that the suction force was significantly greater for the Goldmann compared to the Tonojet. The magnitude of this force was too small to account for the difference in IOP measurements. The Tonojet was less hydrophilic than the Goldmann, and the contact angle of the fluid was therefore greater. For a given tear film, less hydrophilic tonometers will tend to have thicker mires, and this may lead to underestimation of the IOP. When such disposable tonometers are used, it is recommended care should be taken to reject readings from thick mires.
Article
Full-text available
To evaluate the relative merits of proxymetacaine-fluorescein (PROX-FLU) and lignocaine-fluorescein (LIG-FLU) when used to perform applanation tonometry. This prospective, masked, double blind study assessed several aspects of the tonometry process-the duration of the stinging sensation and degree of discomfort, the extent of reflex lacrimation induced, the need for subsequent tear film manipulation to ensure an accurate tonometry reading, and total time to complete tonometry-for each preparation. PROX-FLU caused significantly less discomfort and reflex lacrimation than LIG-FLU and accurate tonometry was more rapidly completed when it was used. PROX-FLU was preferred by 98% of the study patients. PROX-FLU is a well tolerated and useful alternative to the more widely used LIG-FLU mixture.
Article
Full-text available
To assess the adequacy of current decontamination methods for the Goldmann tonometer in the context of variant Creutzfeldt-Jakob disease (vCJD). Reusable Goldmann tonometer prisms were used to perform applanation tonometry on different groups of patients. Following tonometry, retained materials were collected from the tonometer prism head and examined using cytological methods. The used tonometers were subjected to a series of conditions to evaluate their effect on the residual cell numbers found on the tonometer heads. These included wiping alone and wiping or washing followed by disinfection of the tonometer prism. The effect on cell counts of drying the prism overnight was studied, as well as drying overnight and then wiping and disinfecting. All disinfections were performed with sodium hypochlorite (0.05% w/v). The cytology specimens of 69 patients were studied. Patients using eye drops regularly desquamated significantly more corneal epithelial cells with Goldmann tonometry than patients not using regular eye drops. The mean number of cells was 156 (range 0-470) for patients using eye drops and 14 (4-57) for patients not using eye drops (p = 0.004). Wiping or washing the tonometer head reduced the cell number significantly but neither method completely eliminated cells. The two methods were not significantly different (p=0.3). Drying left a large number of cells (23-320 cells). Retained corneal epithelial cells, following the standard decontamination routine of tonometer prisms, may represent potential prion infectivity. Manual cleaning was the most important step in reducing epithelial cell retention.
Article
Full-text available
Disposable devices are increasingly becoming the preferred choice where possible in contact medical equipment. To evaluate the accuracy of the disposable applanation tonometer head as a potential substitute to the standard Goldmann applanation head. The study was prospective. The intraocular pressure recordings in 80 eyes of 42 patients were compared using the disposable and standard Goldmann applanator heads. The Bland and Altman method of assessing agreement between two methods of clinical measurement was used in the analysis. The difference in the readings between the two types of tonometer heads was highly variable (mean difference=0.78 mm Hg, range=-1 to 11 mm Hg). This was because of the distortions on the applanating surface of the disposable device. When the readings associated with the defective heads were excluded, very strong agreement was obtained (mean=0.07 mm Hg, range=-1 to 2 mm Hg). Good agreement with standard Goldmann applanation is achieved with the disposable heads except where surface distortions induce significant errors. Careful inspection to ensure well-structured disposable units is imperative in disposable applanation tonometry.
Article
In the clinical evaluation of constant-pressure tonography, charts are required by the medical technician for determining the value of aqueous outflow facility from the force readings obtained from the tonography instrument. The accuracy of the tables depends upon the validity of Imbert-Fick law in this application. As a result of earlier experiments on excised human corneas, there has been some question as to whether use of this law leads to large errors when applied to corneal applanation, and it has been suggested that use of an empirical relation would be better.In this study the accuracy of the Imbert-Fick law was tested with the use of rabbit and human cornea, and in the intact and the enucleated rabbit eye. Our results also do not conform exactly to the Imbert-Fick law, but the errors are not excessive for the practical application of the law to constant-pressure tonography. In fact, its use leads to errors of lesser magnitude than errors from other sources in the constant-pressure measurement of aqueous outflow facility.
Article
Scrapie of sheep, transmissible mink encephalopathy and Creutzfeldt-Jakob disease and kuru of man constitute the nosologic group of subacute spongiform virus encephalopathies.1 The possible transmission of Creutzfeldt-Jakob disease from person to person has been reported in a 55-year-old man in whom the disease developed 18 months after he received a corneal transplant from a donor who had died of a neurologic disorder later diagnosed as Creutzfeldt-Jakob disease.2 In 1974 infectivity was demonstrated in corneal tissues of hamsters dying of transmissible mink encephalopathy. After injection of infected corneal epithelium intracerebrally into healthy hamsters the disease developed.3 Successful transmission from a woman . . .
Article
There are many sources of error in the use of Goldmann-type applanation tonometers. In clinically normal corneas hypofluorescence of the precorneal tear film, accommodation, the Valsalva maneuver and vertical gaze are preventable causes of large tonometric errors. Repeated tonometry may induce a decline in the intraocular pressure (IOP). Variations in the corneal resistance to indentation between eyes cause significant errors. The most significant cause of error in clinically abnormal eyes is corneal epithelial edema, which causes a marked underestimation of IOP. Measurements obtained with Goldmann-type tonometers can be used with confidence to monitor changes in the IOP of an individual, but should not be relied on to determine the absolute manometric pressure within an eye or to compare the IOPs in eyes of different individuals.
Article
Tonometer tips are used by optometrists to measure intraocular pressures. The recommended procedure of soaking in bleach solution kills bacteria and certain viruses, such as human immunodeficiency virus, herpes simplex virus-1 and herpes simplex virus-2, adenovirus 8, and hepatitis B, from the tip. Conversely, recommendations made in literature to sterilize equipment that may have come in contact with virus-contaminated tissue from patients with Creutzfeldt-Jakob disease have a somewhat tougher requirement. Autoclaving for 1 hour at a temperature of at least 120 degrees C (15 psi), or a 1-hour exposure to 0.5% sodium hypochlorite (a 10-fold dilution of household bleach) should provide excellent disinfection. One-hour exposure to 1 N Sodium hydroxide has also been mentioned in the literature. Studies have shown that corneas of guinea pigs with Cruetzfeldt-Jakob disease (C-J disease) are infectious. Infected corneas have been shown to cause transmission via corneal transplants, and via experimental placement of infected guinea pig's cornea into the anterior chamber of uninfected guinea pigs. Many researchers have strongly suggested that C-J disease can be iatrogenically transmitted via applanation tonometer tips. An epidemiologic case-controlled study found statistically significant odds ratio for intraocular pressure testing in the medical history of patients with C-J disease. Even though there have not been any proven studies confirming iatrogenic transmission through tonometer tips, optometrists should be cautious if a patient has C-J disease, or manifests symptoms of C-J disease and use alternatives to Goldmann applanation tonometry.
Article
We conducted a study to compare the accuracy of two commercially available disposable tonometry devices (the acrylic biprism, Tonosafe and the silicone shield, Tonoshield) with the standard Goldmann applanation tonometer (GAT). A total of 80 eyes of 40 patients were randomly assigned to two groups. Group A (40 eyes) had the intraocular pressure IOP measured using GAT and the acrylic biprism (Tonosafe), while Group B (40 eyes) had IOP measured using GAT and the disposable silicone shield (Tonoshield). IOP was measured using the GAT both before and after the disposable device and average of the two standard Goldmann readings was compared with the IOP measured using the disposable devices. The mean difference in IOP in Group A was 0.29 mmHg (SD 0.54) (P=0.0018) while in Group B the mean difference in IOP was 2.09 mmHg (SD 1.23) (P<0.0001). The Altman-Bland method was used to assess the agreement between two methods of clinical measurement. This shows a positive bias of 0.29 (95% confidence interval (CI) 0.11-0.46) when using the disposable acrylic biprism and a positive bias of 2.09 (95% CI 1.69-2.48) with the silicone shield. The IOP recorded using the disposable acrylic biprism (Tonosafe) was in close agreement with the standard GAT. However, IOP readings obtained with the silicone shield (Tonoshield) tend to be higher than with the standard GAT. These facts should be kept in mind when recording IOP in clinical situations.
Article
Applanation tonometry as performed in routine clinical practice is a significant potential vehicle for cross-infection particularly in an emergency eye care setting. The aim of this study is to evaluate the accuracy and reliability of three single-use devices (Tonoshield, Tonosafe, Tonojet) as an alternative to standard Goldmann prisms in an emergency eye department. All patients attending the eye casualty at the Manchester Royal Eye Hospital for a period of 4 months who required intraocular pressure measurement were eligible for this prospective study. Exclusion criteria were: age below 18 years, corneal anomalies that might affect measurement, and refusal to participate. After taking informed consent, the patient was examined by one experienced nurse practitioner, who measured the intraocular pressure three times. In the first part of the study, we compared the standard Goldmann prism vsTonoshield and Tonosafe prisms, while for the second part of the study we used standard Goldmann, Tonosafe, and Tonojet prisms. Agreement and repeatability tests were carried out on separate samples. Tonosafe and Tonojet correlated well with standard Goldmann tonometry (P<0.001), while the measurements obtained with Tonoshield were higher, especially for raised intraocular pressure measurements. Tonojet and Tonosafe measurements were more reproducible than Tonoshield measurements. This study shows that Tonosafe and Tonojet are accurate and reliable alternatives to standard Goldmann tonometry.
Medical Device Agency AN 1999 (03) Single patient use of ophthalmic devices: implications for clinical practice
Medical Device Agency AN 1999 (03). Single patient use of ophthalmic devices: implications for clinical practice. Medical Devices Agency, October 1999.