ArticlePDF Available

Online Tissue Discrimination for Transcutaneous Needle Guidance Applications Using Broadband Impedance Spectroscopy

Authors:
  • Hochschule Mannheim, Germany

Abstract and Figures

This paper reports on a novel system architecture for measuring impedance spectra of a biological tissue close to the tip of a hollow needle. The measurement is performed online using fast broadband chirp signals. The time domain measurement raw data are transformed into the transfer function of the tissue in frequency domain. Correlation technique is used to analyze the characteristic shape of the derived tissue transfer function with respect to known "library functions" for different types of tissue derived in earlier experiments. Based on the resulting correlation coefficients the exact type of tissue is determined. A bipolar coaxial needle is constructed, simulated by finite element method and tested during various in vitro and in vivo experiments. The results show a good spatial resolution of approximately 1.0 mm for a needle with a diameter of 2.0 mm. The correlation coefficients for the three tested tissue types muscle, fat, and blood allow for a clear tissue classification. Best results have been obtained using the characteristic phase diagrams for each tissue. Correlated to the corresponding library transfer function the coefficients are in the range of +0.96 to +0.99 for the matching tissue. In return, the resulting coefficients for correlation with nonmatching tissues are in the range of -0.93 to +0.81.
Content may be subject to copyright.
494 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 2, FEBRUARY 2012
Online Tissue Discrimination for Transcutaneous
Needle Guidance Applications Using Broadband
Impedance Spectroscopy
Dennis Trebbels*, Felix Fellhauer, Michael Jugl, Gerd Haimerl, Mart Min, and Roland Zengerle
Abstract—This paper reports on a novel system architecture for
measuring impedance spectra of a biological tissue close to the tip
of a hollow needle. The measurement is performed online using fast
broadband chirp signals. The time domain measurement raw data
are transformed into the transfer function of the tissue in frequency
domain. Correlation technique is used to analyze the characteris-
tic shape of the derived tissue transfer function with respect to
known “library functions” for different types of tissue derived in
earlier experiments. Based on the resulting correlation coefficients
the exact type of tissue is determined. A bipolar coaxial needle is
constructed, simulated by finite element method and tested during
various in vitro and in vivo experiments. The results show a good
spatial resolution of approximately 1.0mm for a needle with a di-
ameter of 2.0 mm. The correlation coefficients for the three tested
tissue types muscle, fat, and blood allow for a clear tissue classi-
fication. Best results have been obtained using the characteristic
phase diagrams for each tissue. Correlated to the corresponding li-
brary transfer function the coefficients are in the range of +0.96 to
+0.99 for the matching tissue. In return, the resulting coefficients
for correlation with nonmatching tissues are in the range of 0.93
to +0.81.
Index Terms—Broadband impedance spectroscopy, cannula
guidance, chirp-signals, needle guidance, tissue classification.
I. INTRODUCTION
NEEDLE guidance during transcutaneous surgical inter-
ventions is a common challenge in many medical applica-
tions such as fine needle biopsies [1], regional anaesthesia [2],
drug delivery [2], catheter insertion [3], vessel puncture [4] and
Manuscript received March 8, 2011; revised June 9, 2011 and August 25,
2011; accepted October 3, 2011. Date of publication November 8, 2011; date
of current version January 20, 2012. Asterisk indicates corresponding author.
*D. Trebbels, HSG-IMIT, Institut f¨
ur Mikro- und Informationstechnik der
Hahn-Schickard-Gesellschaft e.V., 78052 Villingen-Schwenningen, Germany
(e-mail: dennis.trebbels@hsg-imit.de).
F. Fellhauer, HSG-IMIT, Institut f¨
ur Mikro- und Informationstechnik der
Hahn-Schickard-Gesellschaft e.V., 78052 Villingen-Schwenningen, Germany
(e-mail: felix.fellhauer@hsg-imit.de).
M. Jugl and G. Haimerl are with the HFU, Hochschule Furtwangen University,
78054 Villingen-Schwenningen, Germany (e-mail: michael.jugl@hsg-imit.de;
hai@hs-furtwangen.de).
M. Min is with the Tallinn University of Technology, Faculty of Informa-
tion Technology, Department of Electronics, 19086 Tallinn, Estonia (e-mail:
min@elin.ttu.ee).
R. Zengerle is with the Department of Microsystems Engineering, Labora-
tory for MEMS-Applications, University of Freiburg, 79110 Freiburg, Germany
(e-mail: roland.zengerle@gmx.de).
Color versions of one or more of the figures in this paper are available online
at http://ieeexplore.ieee.org.
Digital Object Identifier 10.1109/TBME.2011.2174990
brachytherapy [5]. In all of the listed applications it is impor-
tant to place the needle tip inside a well defined tissue type of
the human body whereas often the exact 3-D orientation of the
needle is of minor importance.
Today there are several approaches to accomplish the exact
needle placement based on 2-D or 3-D imaging systems which
allow for controlling the needle orientation and position and sup-
port the surgeon during the needle insertion procedure. Three
well known systems are magnetic resonance imaging (MRI),
computed tomography (CT) and X-Ray fluoroscopy [6]–[8].
All systems work very well and give good spatial resolution
but are too expensive for widespread use. In addition CT and
X-Ray fluoroscopy expose the patient to ionizing radiation. Fur-
thermore, fluoroscopy provides only a limited contrast for soft-
tissues, which is a crucial point for needle guidance. Alternative
imaging systems are based on ultrasound (US) [9], [10]. There
are several ultrasound systems available in the market, but ma-
jor drawbacks are the limited penetration depth and the limited
field of view. In addition, it is difficult to analyze the exact type
of soft tissue close to the needle tip by using US. The use of
force-feedback systems [11] in combination with imaging based
systems or robots [12] can improve the needle steering process
but the overall system again becomes quite complicated and
expensive. In some actual research projects microsensors are
integrated in the needle tip [13] but deal with the inherent lack
of space given by the thin needle geometry and the low-cost
requirements for disposable medical equipment.
A promising alternative approach for exact online tissue clas-
sification at the needle tip during the insertion process into the
human body is broadband impedance measurement, often re-
ferred to as impedance spectroscopy (IS). This method is based
on the continuous measurement of the characteristic impedance
spectrum of a small volume of tissue around the needle tip.
The measured spectrum is analyzed and the exact type of tissue
can be determined. Kalvoy et al. already presented prelimi-
nary results of their configuration based on a monopolar needle
electrode where the impedance is measured against a reference
electrode [14], [15]. The results look promising but a drawback
of the system is the need for the additional reference electrode,
and due to the current density distribution the achievable spatial
resolution is limited to a spherical volume having a diameter
of approximately three to four times the needle diameter. We
already proposed a coaxial needle design in combination with
time domain reflectometry (TDR) measurement technique [16].
Drawback of this technique is the poor signal-to-noise ratio,
which can be achieved for a broadband measurement.
0018-9294/$26.00 © 2011 IEEE
TREBBELS et al.: ONLINE TISSUE DISCRIMINATION FOR TRANSCUTANEOUS NEEDLE GUIDANCE APPLICATIONS 495
Fig. 1. Detailed view of the constructed coaxial hollow needle. The needle
consists of two concentric stainless steel tubes which are electrically isolated
by an inner layer of Teflon (PTFE). The given dimensions correspond to the
prototype needle, which is used throughout the simulations and experiments.
In this paper, we present a new measurement system for coax-
ial hollow needles where the impedance measurement method
is based on short broadband chirp signals. The paper discusses
the system concept, finite element method simulation results
for the expected spatial resolution, chirp signal properties and
processing, the developed laboratory measurement setup and
experimental details for the conducted in vitro and in vivo ex-
periments. The derived raw data are processed and analyzed by
correlation techniques. The correlation results are presented and
discussed particularly with regard to tissue classification. In ad-
dition, some further experiments are done in order to support the
simulation results, investigating the impact of the speed of the
needle during insertion process and evaluating the robustness of
the classification method with regard to mechanical tolerances
and imperfections of the needle.
II. SYSTEM CONCEPT AND MODEL
Goal of the developed system is to continuously analyze and
classify the exact type of tissue close to the needle tip during the
insertion process of the needle into the human body. In order
to classify the tissue close to the needle tip we continuously
measure the complex electrical impedance of the tissue close to
the needle tip at various frequencies within the range of 5 kHz
to 1 MHz. Within this frequency range the complex transfer
function of the tissue is expected to be characteristic for each
individual type of tissue [17]. The chosen range is a compromise
between overall bandwidth, signal duration, and the resulting
total amount of raw data, which must be processed in real time.
A more detailed description of the signal properties is given in
Section IV-C.
A coaxial hollow needle is constructed as illustrated in Fig. 1.
The inner stainless steel tube serves as electrical conductor for
the measurement signal whereas the outer stainless steel tube
is at ground potential. The two tubes are electrically isolated
against each other by a thin layer of Teflon (PTFE). Current can
only flow from the inner conductor to the outer conductor at
the needle tip while inserted into a conductive media (see also
Fig. 3). When the needle is inserted into biological tissue, the
tissue forms a frequency-dependent electrical load on the needle
Fig. 2. (a) Equivalent electrical schematic of biological tissue located on the
needle tip in series to a model for the electrode polarization. (b) The used
cell model from which the equivalent schematic is derived. (c) The electrode
polarization effect by a double layer structure.
Fig. 3. (a) Current density plot of the needle tip in homogeneous resistive
media. The highest current density and therefore the highest sensitivity are
close to the needle tip. (b) The simulated needle tip and the volume virtually
divided in front of the tip in thin slices with 0.1 mm thickness for further analysis
of the losses and sensitivity. The results are shown in Fig. 4.
tip as illustrated in Fig. 2(a). The presented equivalent circuit
model for the tissue is composed of Rp,R
i,and CM.Fig.3(b)
illustrates a corresponding cell model, where Rirepresents the
resistive behavior of the intracellular volume, Rprepresents the
resistive behavior of the extracellular volume, and CMrepre-
sents the capacitive behavior of the cell membrane [18]. In series
to the tissue model we place a second model, which describes the
frequency depending behavior of electrode polarization effects
(Cpol,R
pol,Z
W) [19]. Electrode polarization occurs whenever
a galvanically coupled metal electrode is in direct contact with
an aqueous electrolyte. The effect can be explained by a rigid
double layer, which is caused by metal ions on the electrode
surface and corresponding counter ions in the electrolyte. Often
this layer is referred to as Helmholtz layer. In addition, there
is a weaker diffusion layer caused by the electrostatic coulomb
forces of the charged metal surface, often referred to as Gouy-
Chapman layer [see Fig. 2(c)]. The equivalent circuit model for
the two layers consists of a common layer capacitance Cpol and
the two elements Rpol and ZW[19]. Rpol represents the resistive
behavior of the charge transfer within the Helmholtz layer and
is almost frequency independent. ZWrepresents the impedance
496 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 2, FEBRUARY 2012
of the Gouy-Chapman layer and is frequency-dependent. Often
this frequency dependency is modeled by a constant phase el-
ement. The resistive properties of the electrolyte bulk material
are already represented by Rpof the tissue model, and therefore
here not included in the double layer model.
For tissue classification,it would be ideal to measure the
broadband impedance of the tissue only and eliminating the
parasitic effects caused by the electrode polarization. However,
this is practically impossible due to the 2-electrode interface
formed by the needle tip. All measurement results will always
be a combination of the tissue transfer function and the elec-
trode polarization transfer function. Calibration and precision
correction of the polarization effects is also practically not ap-
plicable because the polarization itself depends on characteristic
tissue properties such as the conductivity of the bulk material
as well as on measurement signal parameters such as the ap-
plied voltage level and the resulting current density. In addition,
a calibration (e.g., in saline) for each hollow needle means a
significant amount of work. In the ideal case, it would be pos-
sible to measure directly without the need for any calibration
procedure.
Throughout the study presented here in this paper, we decided
to go a straight way and simply measure the resulting overall
electrode impedance as a function of frequency inside differ-
ent types of tissue including all polarization effects. Multiple
complex impedance curves are recorded for each type of tissue.
An averaged curve for each individual tissue type is calculated
out of the recorded data and serves as “library function.” Tissue
classification is done by comparing these “tissue library func-
tions” online with continuously measured impedance curves.
The comparison is made by correlation and the tissue classifica-
tion is based on the resulting correlation coefficients. We accept
that the measured impedance curves do not exactly represent
the specific tissue impedance only. Our focus is on tissue classi-
fication between several types of tissue and not on precise tissue
analysis of one specific single tissue. The results discussed later
in the paper show that a pure classification can be achieved by
this method, and there is no need for an additional calibration
procedure using a reference electrolyte.
III. SIMULATION
A low frequency finite element simulation is done in order to
evaluate the current density distribution and the resulting spatial
sensitivity for a needle with the geometry shown in Fig. 1. The
simulation is done at a frequency of 100 kHz using the software
package “Ansys Multiphysics.” The maximum mesh size is set
to 30 μm, the boundaries of the simulated volume are at ground
potential. The excitation signal amplitude is set to 1 V and the
conductivity of the bulk material is set to 1S/m. However, the
absolute values of the resistive bulk material and the excita-
tion signal amplitude are not relevant, since we calculate the
“relative losses.” This means we calculate the losses within a
small tissue volume close to the needle tip and compare them
to the total losses in the simulated system caused by the mea-
surement signal. Any absolute material properties such as the
specific conductivity cancel out. The resulting “relative losses”
Fig. 4. FEM-Simulation results: Relative losses as a function of the volume
in front of the needle tip. The volume is indicated by the distance to the needle
tip according to the slices shown in Fig. 3.
represent the sensitivity as a function of the needle tip geome-
try only. We do not simulate any frequency-dependent behavior
of the tissue since this has no effect on the spatial sensitivity.
The electrical losses are proportional to the square of the cur-
rent density within a small volume of tissue. Fig. 3(a) shows a
simulated current density pattern for the needle tip inserted into
homogeneous resistive media.
As expected, the highest current density and, therefore, the
highest sensitivity are found close to the needle tip, which serves
as two-electrode-pair. For further detailed analysis of the sensi-
tivity we virtually cut the volume in front of the needle tip into
thin slices with a thickness of 0.1 mm each, as shown in Fig. 3(b).
Fig. 4 presents the simulation result. The graph shows the accu-
mulated relative losses inside a volume in front of the needle tip,
as illustrated in Fig. 3(b). The graph shows that approximately
90% of the total losses are generated within a distance of only
1 mm to the needle tip. In addition, we investigated the effect
of the needle insertion depth. The current density on the outer
needle wall surface is very low but, in total, the needle has a
large surface compared to the needle tip. Fig. 4 contains two
graphs, one for 10 mm insertion depth and the other for 40 mm
insertion depth. Both graphs are almost identical which means
the sensitivity is almost independent of the insertion depth of
the needle. The result looks reasonable since almost all losses
(90 %) are generated directly in front of the needle tip.
The simulation result leads to the assumption that the hollow
coaxial needle design yields a very good spatial resolution of
the measurement and is ideally suited for precision needle tip
positioning. In addition to the simulation, we conducted an in
vitro experiment. The obtained results support the simulation
result very well and are described in detail within Sections VI
and VII of this paper.
IV. MEASUREMENT SIGNALS AND PROCESSING
A. Measurement Signals
As already mentioned in Section II, the measurement system
must be capable of measuring the complex impedance spec-
trum of the tissue under investigation at the needle tip. Since
needle insertion is a dynamic process, the measurement has
to be done within a short period of time. In an ideal case the
TREBBELS et al.: ONLINE TISSUE DISCRIMINATION FOR TRANSCUTANEOUS NEEDLE GUIDANCE APPLICATIONS 497
Fig. 5. Normalized linear chirp signal plot in time domain. The presented
sample signal has a duration of 30 μs and its frequency ranges from 5 kHz to
1MHz.
measurement time window is so short that the complex tissue
load on the needle tip remains stable for one complete mea-
surement cycle. However, this requirement cannot be achieved
by using conventional measurement techniques such as apply-
ing several frequencies within one slow sweep. Therefore, we
use short broadband signals. One type of broadband signals is
the so called chirp signal which has already been introduced as
a promising signal for impedance measurement [20]. The sine
wave based chirp signal has a general expression
C(t)=Asin θ(t)=Asin ω(t)dt +θ0(1)
with amplitude A, running phase θ(t), instantaneous angle fre-
quency ω(t) and initial phase angle θ0. A simple linear chirp
has an instantaneous frequency ω(t) =dθ(t)/dt, which changes
linearly during the excitation interval Texc with constant accel-
eration dω(t)/dt =d2θ(t)/dt2=kch. Expressing ω=2πf and
denoting f0as an initial, and ffin as a final frequency, also taking
θ0=0 and marking Tch as duration of the chirp pulse, we obtain
the following expression for the linear chirp excitation:
C1(t)=Asin 2πf0·t+(ffin f0)·t2
2Tch .(2)
The excitation bandwidth Bexc =ffin f0remains the same when
the excitation time Texc =Tch changes. Only the chirping rate
kch =(ffin f0)/Tch and signal energy E=(A2/2)Tch vary to-
gether with Tch.
A linear chirp signal calculated by (2) gives maximum flex-
ibility in scaling the signal to the needs of the application. In
this case amplitude, signal duration and covered bandwidth can
be controlled independently. Figs. 5 and 6 show a linear chirp
signal in time domain and frequency domain. One advantageous
property of such a chirp signal is the flat spectrum in the fre-
quency domain. The amplitude of all contained frequency bins
remains stable over the entire frequency range (Fig. 6).
This property allows for measuring the impedance over the
full bandwidth with stable signal-to-noise ratio and, therefore,
leads to precision transfer functions of the measured complex
needle tip impedance.
B. Signal Processing Concept
Goal of the measurement is to derive the complex transfer
function of the load impedance caused by the tissue under inves-
tigation in the frequency domain. Therefore, the unknown tissue
Fig. 6. Normalized amplitude plot of the power spectral density of the chirp
signal presented in Fig. 5. The signal energy equally distributed within the
bandwidth of the signal and therefore allowing broadband measurements with
a stable signal to noise ratio over the whole frequency range.
Fig. 7. Block schematic of the signal processing chain as implemented in
the developed system. The system under investigation is the load impedance
caused by the unknown tissue located at the needle tip. Stimulus and response
signals are voltage and current signals, respectively. Both signals are sampled
and converted into frequency domain for calculation of the transfer function.
is stimulated by a broadband chirp voltage signal and the result-
ing response current is recorded. Both signals are synchronously
sampled by a two-channel AD-converter and transformed into
frequency domain (Fig. 7). The quotient of both signals is the
complex transfer function.
C. Measurement Signal Properties
As already mentioned before, the chirp signal properties can
independently be adjusted to the specific requirements of the
application. In this case the bandwidth of interest is set to
5 kHz – 1.0 MHz, because here we expect a characteristic
impedance spectrum for each biological tissue [17]. Further-
more, the covered frequency range is a compromise between
the excitation signal bandwidth, excitation signal duration, and
the resulting amount of sampled raw data, which must be pro-
cessed online with a reasonable amount of computing effort.
Key consideration is the required sampling rate of the two syn-
chronous AD-converters shown in Fig. 7. According to the Shan-
non theorem, the minimum sampling rate must be at least twice
the frequency of the highest frequency component in the signal.
Therefore, the maximum frequency component fmax in the chirp
signal dictates the minimum sampling frequency fsample-min of
both AD-converters which is fsample-min =2fmax. The result-
ing minimum total number of samples Nmin required to capture
one full chirp excitation signal can be calculated and expresses
as Nmin =Tchirp fsample-min =2fmax /fmin. This equation
shows that the number of samples Nmin is an exponential func-
tion and becomes quite large for signals with a broad bandwidth
covering several decades. Therefore we choose a compromise
and use a chirp signal with a duration of 200 μs. This results in
a frequency resolution δf=1/Tchirp =1/200 μs=5 kHz and
a minimum number of samples Nmin =400. In addition we
decide to use an oversampling rate of a factor 10, resulting in
498 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 2, FEBRUARY 2012
Fig. 8. Block schematic of the laboratory setup. The unknown impedance Z
and the current measurement resistor are connected in series and the AWG ex-
cites the measurement signal. In addition, the AWG also generates an adjustable
trigger signal for the oscilloscope that allows a customized sampling rate.
4000 samples. The next matching power of 2 is 4096 samples.
The sampling frequency and thus the trigger frequency for the
AD-converter is, therefore, adjusted to 20.48MHz for capturing
a chirp with fmax =1 MHz. Our developed software running
on a laptop computer can perform a complex 4096 point FFT
within 2.4 ms and allows for fast online signal processing.
Another critical parameter is the excitation signal amplitude.
Extremely low amplitudes in the mV or even sub-mV range
cause difficulties associated with noise and the dynamic range
of AD-converters. In contrast very high amplitudes in the range
of several volts may significantly cause measurement errors due
to the resulting high current densities and nonlinear behavior
or even damage of the tissue under investigation. Based on our
experience, we choose the chirp signal amplitude to be 100 mV
as a compromise and use this value within all experiments.
V. LABORATORY MEASUREMENT SETUP
For experiments, a laboratory measurement setup is devel-
oped using state of the art equipment. Fig. 8 shows a block
schematic of the implemented setup. The chirp signal is gener-
ated by a programmable arbitrary waveform generator (AWG,
Tektronix AFG3252) which drives the coaxial cable connected
to the needle via a current measurement resistor. The stimulating
voltage signal and the resulting current signal are synchronously
sampled by a digital storage oscilloscope (DSO, LeCroy
WaveRunner 104 Xi). The AWG generates an additional trig-
ger signal with stable phase relative to the chirp measurement
signal and allows for a stable triggering and sampling. Both
oscilloscope input channels have a limited input impedance of
1 MOhm and a relatively high input capacitance of approxi-
mately 20 pF. Especially at higher frequencies the high input ca-
pacitance causes a low input impedance and leads to frequency-
dependent measurement errors. This unwanted effect is avoided
by employing two identical buffer amplifiers connected to the
oscilloscope inputs, as shown in Fig. 8. The amplifiers have an
input capacitance of approximately 1 pF.
The software calculates the chirp signal and controls the sig-
nal generation and sampling process via LAN and USB. In
addition, the software processes the measured raw data and di-
rectly calculates the transfer function. In principle the transfer
function can directly be calculated out of the measured voltage
and current signals, as illustrated in Fig. 7. Drawbacks of this
Fig. 9. Equivalent electrical schematic of the system including relevant par-
asitic components that cause measurement errors. Riis the output resistance
of the generator, Rcis the dc-resistance of the copper wires that connect the
measurement setup with the needle and Ccis the cable capacitance of the used
coaxial cable. (a) Cccan be estimated by an “open” calibration. (b) Rccan be
estimated by a “short” calibration if Ccis known.
direct method are measurement errors caused by inherent para-
sitic effects such as unwanted capacitive and resistive elements
in the setup. Fig. 9 shows an equivalent circuit including major
effects such as the cable capacitance Ccand the connector/cable
resistance Rc. In order to minimize such measurement errors, we
perform a one-time system calibration by measuring the transfer
function of the “open” system (no load at the needle tip) and the
“shortened” system (short circuit at the needle tip). The mea-
sured transfer functions of the open circuit and the shortened
circuit are used to directly correct measurement errors of the
tissue transfer functions in frequency domain. First, the “open”
calibration has to be performed. In an ideal setup, no current
is expected to flow under the “open” condition. In reality, a
frequency-dependent current can be measured. This alternating
current mainly flows through the parallel cable capacitance Cc.
The equivalent schematic for an “open” condition is presented
in Fig. 9(a).
The measured transfer function TF{open}is directly the
transfer function of the capacitive element TF{Cc}.Fig.9(b)
shows the equivalent circuit for a “short” condition on the nee-
dle tip. In parallel to the capacitance Ccthere is two times the
cable resistance Rc. The transfer function TF{Rc}can be ob-
tained by calculating TF{Rc}=TF{short}–TF{Cc}.Nowall
relevant parasitic components are known and the corresponding
transfer functions can directly be taken into account during the
calculation of any measured tissue transfer function. Fig. 10
shows two plots for a laboratory measurement result obtained
by placing the needle tip into a resistive saline bath with a mea-
sured conductivity of 16.2 mS/cm (almost isotonic). In the test
setup we used a coaxial cable of type RG174A with a length of
1.5 m to connect the needle with our measurement setup. In case
of no calibration and no error correction, a signal degradation at
high frequencies is observed. This is mainly caused by the par-
asitic coaxial cable capacitance. Applying the above described
calibration procedure virtually eliminates the parasitic effects
and shows a stable modulus and phase value as expected for a
pure resistive load within this frequency range. Any polariza-
tion effects at low frequencies could not be observed within this
experiment. According to [19] the polarization effect becomes
more visible at frequencies below 1 kHz, which is outside of our
measurement range.
TREBBELS et al.: ONLINE TISSUE DISCRIMINATION FOR TRANSCUTANEOUS NEEDLE GUIDANCE APPLICATIONS 499
Fig. 10. Influence of parasitic components and result of calibration and error
correction illustrated by the result of an experiment obtained in resistive saline
bath.
VI. MATERIALS AND METHODS
A. Overview
Within this research study, we conducted several in vivo and
in vitro experiments. Goal of the in vivo experiments was to
derive measurement data especially in order to investigate the
possibility of tissue classification. Details of the in vivo experi-
ments are presented in Section VI-B in this paper. In addition to
the in vivo experiments we conducted three in vitro experiments
in order to investigate side effects and robustness of the measure-
ment principle as well as verifying the FEM-simulation results
presented in Section III. Details about the in vitro experiments
are given in Section VI-C.
All measurements have been done with the developed labo-
ratory setup described in Section V. The measurement resistor
R(see Fig. 8) is set to 100 Ohm. The excited chirp signal had a
duration of 200 μs, an amplitude of 100 mV, and covered a band-
width from 5 kHz to 1 MHz. The signals have been captured by
a DSO with a sampling rate of 20.48 MHz and an amplitude res-
olution of 8 bits resulting in a total number of 4096 samples per
sampled chirp signal. The sampled waveforms of the stimulus
chirp signal and resulting current signal have been stored on file
for later analysis.
B. In vivo Experiments
All in vivo experiments have been conducted on anesthetized
pigs. For our study, we measured on three pigs. On each pig the
three-tissue types muscle, fat, and blood have been measured
on two locations with a prototype needle having the dimensions
presented in Fig. 1. A picture of the used prototype needle is
shown in Fig. 11(c). First measurement location was the neck of
the pig, and second measurement location was the hip of the pig.
A veterinarian carefully cut the skin of the pig and dissected the
named three-tissue samples. The samples were prepared in such
a way that the measurement needle could be inserted into either
plain muscle, fat, or arterial blood tissue without any visible
inhomogeneity. Placement of the needles was done manually.
The in vivo experiments have been done in accordance with
animal ethic standards approved by the government of Baden-
W¨
urttemberg, Germany. Institutional and national guides for the
care and use of laboratory animals were followed.
C. In vitro Experiments
Goal of the first in vitro experiment was to verify the simula-
tion result for the expected spatial resolution of the tissue clas-
sification on the needle tip. Therefore, we inserted the needle
Fig. 11. Picture of three different hollow coaxial prototype needles. Needle C
is used throughout the experiments, the dimensions are given in Fig. 1. Needles
A and B are used during additional in vitro experiments. Picture 11D illustrates
the setup used for in vitro experiments using well prepared tissue with good
visible and sharp tissue boundaries between muscle and fat.
into well prepared tissue [see Fig. 11(d)] using a programmable
linear actuator. We started with the needle tip embedded inside
homogeneous muscle tissue. In steps of 0.1 mm we moved the
needle toward the fat tissue. On each step we measured the
impedance spectrum of the tissue load. We stopped the actuator
when the needle tip was fully embedded inside fat tissue.
The second in vitro experiment was conducted with the same
setup, but the needle was moved within homogeneous muscle
tissue only. We continuously moved the needle with various
speeds from 0 to 14 mms1and continuously measured dur-
ing the movement. Goal was to evaluate if a moving needle
shows different results compared to a non-moving needle due
to disturbance effects of the double layer on the electrodes.
The third in vitro experiment was done with three different
types of prototype needles with various dimensions, as shown
in Fig. 11(a) and (b). All three needles were placed into homo-
geneous muscle and fat tissue and the impedance spectra have
been recorded. The results are compared to each other and give
a first overview about the impact of mechanical parameters on
the tissue classification.
VII. MEASUREMENT RESULTS
A. In vivo Experiments
During the in vivo experiments the three different tissue types
muscle, fat, and arterial blood have been measured. The sampled
raw data of the stimulus voltage signal and the response current
signal have been processed by a custom analysis software ac-
cording to the scheme presented in Fig. 7. The output of the
signal processing software is a Bode diagram for each measure-
ment. Since the measured impedance is complex there are two
plots for each measurement—the modulus plot and the phase
plot as a function of frequency. Below are the obtained results
for the three tissues. We conducted the measurement on three
pigs with two samples per pig resulting in a total of six measure-
ments per tissue. However, for muscle tissue we only show five
measurements since we had to exclude one extreme outlier. The
outlier was caused by accidentally scratching a blood vessel and
almost reflected the blood curve only.
B. In vitro Experiments
The first in vitro experiment was done in order to verify
the simulation result presented in Fig. 4. The experiment was
500 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 2, FEBRUARY 2012
Fig. 12. Bode diagrams for blood measured using a 200 μs chirp signal. The
left diagram shows the modulus and the right diagram shows the phase.
Fig. 13. Bode diagrams for fat measured using a 200 μs chirp signal. The left
diagram shows the modulus and the right diagram shows the phase.
Fig. 14. Bode diagrams for muscle measured using a 200 μs chirp signal. The
left diagram shows the modulus and the right diagram shows the phase.
Fig. 15. In vitro experiment for evaluating the spatial measurement resolution.
Three remarkable needle positions are shown in A, B, and C. The corresponding
positions are also marked with A, B, and C in the phase diagram on the right
side. The plotted phase curve shows the phase angle at a frequency of 100kHz.
conducted as described in Section VI-C. Measurements were
obtained for the needle transition from muscle to fat tissue.
Fig. 15(a)–(c) illustrates the needle position as also indicated
in the associated graph on the right side. The graph shows the
measured phase as a function of the penetration depth for a fixed
frequency of 100 kHz. We chose 100kHz because according to
Figs. 13 and 14, at this frequency we expect a significant change
in the signal depending on which tissue we have in front of the
needle tip.
The second in vitro experiment was done in order to verify
the impact of speed when the needle is moving inside the tissue
during the measurement cycle. The following graph shows the
measured amplitude and phase values at a frequency of 100 kHz
for muscle and fat tissue as a function of speed. The maximum
speed which could be achieved with the given linear actuator
was 14 mms1which already represents a quite fast insertion
process. During conventional surgical operations the needle in-
sertion speed is usually lower.
The results obtained in the third in vitro experiment are
presented later within Section IX because the data evaluation
is also based on the tissue classification scheme discussed in
Section IX. There the effect of the needle shape is discussed.
VIII. DISCUSSION OF MEASUREMENT RESULTS
A. In vivo Experiments
The presented diagrams in Figs. 12–14 show the measured
complex impedance formed by the tissues on the needle tip. The
data include characteristic tissue properties as well as charac-
teristic electrode polarization effects which cannot be separated
from each other. However, on the first view, the three tissues
seem to have individual characteristic shapes of the plotted
curves. Especially the phase curves show significant differences
in the shape and it looks attractive to classify the different tis-
sue types based on their individual spectra. The classification
approach is done using correlation and is described in detail
within Section IX.
It is remarkable that all measured curves for blood show
an almost perfect match to each other whereas especially the
impedance curves for fat and the phase curves for muscle show
some variations between the lines. According to our measure-
ment experience this is mainly caused by the “quality” of the
contact between the tissue and the needle. Blood is liquid and
always has an ideal contact to the needle tip whereas the con-
tact of fat and muscle tissue depends on the pressure which is
generated by the needle. A high pressure ensures a good contact
whereas a low pressure may significantly affect the measured
values. However, the pressure is given by the mechanical tis-
sue properties, the needle geometry, and the insertion speed.
In real applications, this may vary and cannot be optimized or
stabilized. For tissue classification, this circumstance should be
included in the tests and a resulting classification scheme should
be robust enough to correctly identify the different tissues de-
spite the unknown “quality” of the contact to the needle. In
addition to the variations caused by imperfections of the con-
tact, there is the possibility that small inhomogeneities in the
tested tissues caused variations as well.
B. In vitro Experiments
The measurement results for the first in vitro experiment are
presented in Fig. 15. The graph shows the measured phase
angle as a function of the insertion depth at a frequency of
100 kHz. The measurement values were picked out of the mea-
sured impedance spectrum, which has been recorded for each
measurement position, as described in Section VI-C. The graph
clearly shows that the measured phase value for muscle remains
stable until the boundary layer of muscle and fat is reached with
TREBBELS et al.: ONLINE TISSUE DISCRIMINATION FOR TRANSCUTANEOUS NEEDLE GUIDANCE APPLICATIONS 501
Fig. 16. Modulus and phase angle of the impedance as a function of needle
insertion speed for muscle and fat tissue. The diagram represents the values for
100 kHz obtained from the measured impedance spectra.
the center of the needle tip, indicated with position B. Shortly
after reaching position B, the signal drops to the value measured
for fat where it remains stable again. It is remarkable that there
is a “spatial delay” between position B and the falling signal
edge, which is approximately 0.5 to 1.0 mm. This is caused by
the higher electrical conductivity of the muscle tissue compared
to the fat tissue (according to Fig. 16 approximately by a fac-
tor of two). As long as the needle tip still has contact to the
muscle tissue with higher conductivity, the fat tissue is partially
“shortened.” This limits the steepness of the falling edge as
well. However, according to the simulation result presented in
Section III, the sensitivity of the system should be within 1mm
of the needle tip. The obtained measurement results support
the simulation results very well. The falling signal edge occurs
within 1 mm and the location of the falling edge is within 1mm
to the position B. In summary, the FEM simulation and the ex-
perimental results lead to the conclusion that the sensitivity is
approximately within 1 mm to the needle tip for a needle with
an outer diameter of 2 mm and therefore yields a very good spa-
tial resolution which is ideal for precision needle tip placement
inside a specific target tissue.
The second in vitro experiment shows the amplitude and
phase values for muscle and fat tissue at 100kHz as a func-
tion of speed of the moving needle. The results are presented in
Fig. 16. The speed was varied from 0 to 14mms1. Obviously,
the measured values remain quite stable, which indicates that
moving the needle within the tested speed regions and within
the tested tissues does not significantly affect the measurements.
IX. TISSUE CLASSIFICATION
A. Classification Method
Goal of the studies is to implement an online tissue classifi-
cation system which enables to identify the type of tissue on the
needle tip in real-time during the insertion process. Key compo-
nent of such a system is a method that allows for identification
of a specific tissue based on the characteristic properties of its
spectrum. We use traditional correlation technique for “compar-
Fig. 17. Averaged curves from n=5 measured transfer functions for muscle
and n=6 measured transfer functions for fat and blood tissue. All tissue types
show an individual characteristic shape.
Fig. 18. The table shows the average correlation coefficientsfor the impedance
modulus curves for each type of tissue correlated to each other averaged transfer
function. The standard deviation is enclosed in brackets.
Fig. 19. The table shows the average correlation coefficients for the phase
curves for each type of tissue correlated to each other averaged transfer function.
The standard deviation is enclosed in brackets.
ing” the shape of a measured spectrum with the shape of several
known tissues contained in a “tissue transfer function library.”
Within our research, we obtained the “library functions” for
each tissue by simply averaging the measured curves presented
in Figs. 12–14. The averaged result for each tissue is presented
below in Fig. 17.
The obtained measurement data (modulus and phase dia-
grams) for each single measurement within each type of tissue
are correlated to all three library functions presented in Fig. 17.
The resulting correlation coefficients for impedance are pre-
sented in Fig. 18, the results for phase in Fig. 19. The tables
show the obtained average value and the standard deviation
enclosed in brackets. A correlation factor of 1 means both cor-
related curves are identical and, therefore, indicate a perfect
match between the measured transfer function and the library
transfer function. A coefficient of 0 shows that there are no sim-
ilar properties and a coefficient of 1 indicates a perfect match
associated with a negative relationship to the original curve.
In order to derive a robust tissue classification scheme, it is
desirable to obtain a correlation coefficient close to 1 for the
502 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 59, NO. 2, FEBRUARY 2012
matching curves (e.g., measured blood curve correlated with li-
brary blood curve) and a coefficient as close to zero as possible
for non-matching curves (e.g., measured blood curve correlated
with library muscle curve). Second important factor for a ro-
bust classification scheme is a tight standard deviation of all
correlated values. This ensures proper classification even in the
case that two average correlation factors for different tissues are
relatively close to each other (e.g., blood to library blood =0.9
and blood to library muscle =0.8).
The data presented in Figs. 18 and 19 show that a tissue clas-
sification of the three tested tissues can be done without any
trouble based on the experimental dataset. Classification can ei-
ther be done on modulus or phase curves only whereas the phase
curves show slightly better correlation results for classification.
The standard deviation is at least by a factor of approximately
30 smaller than the smallest absolute difference in correlation
coefficients (e.g., fat impedance correlated with library fat ver-
sus fat impedance curve correlated with library blood yields an
average difference in the correlation coefficient of 0.992 – 0.907
=0.085 whereas the standard deviation is 0.0027 for fat).
Another interesting point that has to be investigated is the
effect of mechanical variations. Any variation in the needle
geometry, variation in size, and mechanical tolerances as well
as imperfections of the metal surface have impact on the ef-
fective electrode surface and thus affect the electrode transfer
impedance to the tissue as well as the resulting current den-
sity inside the tissue. Since it is difficult to quantitatively cover
the impact of all possible tolerances and variations, we simply
manufactured two more needles with the same outer diameter of
2 mm but different inner conducting tubes and different PTFE
insulation thicknesses. The two additional needles are shown in
Fig. 11(a) and (b). Needle A has an inner diameter of 0.4 mm
and 0.4 mm PTFE insulation, needle B has an inner diameter
of 0.8 mm and 0.2 mm PTFE insulation. All three needles are
used within the same in vitro experiment and it is measured ten
times inside muscle and fat tissue with each needle. The derived
transfer functions of all three needles have been correlated to
“library functions” obtained by only averaging the values from
the original needle C. That means we compare the measurement
results from needles A and B to needle C using the correlation
technique despite the needles have significantly different di-
mensions. The resulting correlation coefficients are presented
in Table I. Tissue classification is still possible, but the average
coefficients become closer to each other compared to the values
shown in Figs. 18 and 19. That means less “buffer” for a defi-
nite tissue classification. However, it shows that the correlation
method has good potential to be robust enough against mechan-
ical imperfections caused by manufacturing of a needle with a
well defined geometry.
B. Limitations
So far we only investigated the potential for classifying be-
tween homogeneous tissue types. Measuring a mix of tissue on
the needle tip will cause the correlation method to fail as it is
implemented right now. In further research, the classification al-
gorithm has to be optimized. By combining modulus and phase
TAB LE I
CORRELATION COEFFICIENTS FOR NEEDLES AAND B
data within the classification, there might be a way for a robust
classification even for mixed tissue volumes containing at least
two unknown tissue types. The system has to be further tested
with more different tissue types as well as with human tissues.
X. CONCLUSION
Throughout this paper, we presented a novel system approach
for online broadband impedance measurement and online tis-
sue classification on the tip of a hollow needle. The devel-
oped system architecture shows good potential for building up
a stand-alone needle guidance system or it could be integrated
into existing needle steering systems. Here it can potentially
improve the weakness of such systems that they often can-
not perfectly classify between different types of soft tissue. The
presented system architecture successfully uses short broadband
chirp signals with well chosen parameters and a following effec-
tive signal processing scheme for obtaining transfer functions
of the load on the needle tip within a very short measurement
time as required by an online system. Tissue classification is
done by simple correlation of the measured complex transfer
functions. The measured functions are compared to known li-
brary functions and based on the correlation coefficients the type
of tissue is identified. The in vivo experiments show excellent
results and allow for a clear tissue identification using either
the modulus or the phase curve of the transfer function whereas
the results obtained by analyzing the phase curve are slightly
better. Additional in vitro experiments indicate that the system
has good potential to be robust against mechanical tolerances of
the needle tip electrodes as well as to be almost immune against
dynamic needle movement within the tested speed range up to
14 mms1. An FEM simulation and a corresponding in vitro
experiment show an excellent spatial resolution of the sensitive
needle tip, approximately 90% of the signal is derived within
1 mm to the needle tip for a needle with an outer diameter of
2 mm. This is a better resolution than existing prototype systems
based on monopolar electrodes can achieve. According to [14],
a spherical shaped monopolar electrode tip measures 70% of
the total signal within a relatively large volume having a 3.3
times greater diameter than the needle. Clear advantage of the
proposed coaxial needle design is in the close proximity of the
two electrodes at the needle tip and the resulting well defined
current path.
TREBBELS et al.: ONLINE TISSUE DISCRIMINATION FOR TRANSCUTANEOUS NEEDLE GUIDANCE APPLICATIONS 503
REFERENCES
[1] D. Popovic and M. Okoniewski, “Precision open-ended coaxial probe for
dielectric spectroscopy of breast tissue,” in Proc. IEEE Antennas Propag.
Soc. Int. Symp. 2002, vol. 1, pp. 815–818.
[2] M. Vosseler, M. Jugl, and R. Zengerle, “A smart interface for reliable
intradermal injection and infusion of high and low viscosity solutions,”
Pharm. Res., vol. 28, pp. 647–661, 2011.
[3] Y. Thakur, J. S. Bax, D. W. Holdsworth, and M. Drangova, “Design and
performance evaluation of a remote catheter navigation system,” IEEE
Trans. Biomed. Eng., vol. 56, no. 7, pp. 1901–1908, Jul. 2009.
[4] H. Saito and T. Togawa, “Detection of puncturing vessel wall for automatic
blood sampling,” in Proc. 1st Joint BMES/EMBS Conf., 1999,vol.2,
Atlanta, GA, Oct.1999, p. 866.
[5] R. Alterovitz, J. Pouliot, R. Taschereau, I. Hsu, and K. Goldberg, “Sim-
ulating needle insertion and radioactive seed implantation for prostate
brachytherapy,” Stud. Health Tech. Informat., vol. 94, pp. 19–25, 2003.
[6] R. Viard, N. Betrouni, J. Rousseau, S. Mordon, O. Ernst, and S. Maouche,
“Needle positioning in interventional MRI procedure: Real time optical
localisation and accordance with the roadmap,” in Proc. 29th Annu.IEEE
Int. Conf. Eng. Med. Biol. Soc. (EMBS) 2007, Lyon, France, Aug. 22–26,,
pp. 2748–2751.
[7] G. Fichtinger, A. Deguet, K. Masamune, E. Balogh, G. S. Fischer, H.
Mathieu, R. H. Taylor, S. J. Zinreich, and L. M. Fayad, “Image overlay
guidance for needle insertion in CT scanner,” IEEE Trans. Biomed. Eng.,
vol. 52, no. 8, pp. 1415–1424, Aug. 2005.
[8] B. Bascle, N. Navab, M. Loser, B. Geiger, and R. Taylor, “Needle place-
ment under X-ray fluoroscopy using perspective invariants,” in Proc..
IEEE Workshop Math. Methods Biomed. Image Anal., 2000, Hilton Head
Island, SC, Jun. 11–12,, pp. 46–53.
[9] R. S. Singh, S. Natarajan, M. Lee, A. E. Dann, B. P. Cox, D. B. Bennett,
E. R. Brown, H. Lee, W. S. Grundfest, and M. O. Culjat, “Development
of an ultrasound imaging system for needle guidance,” in Proc. IEEE Int.
Ultrason. Symp. (IUS) 2009, Rome, Sep. 20–23,, pp. 1852–1855.
[10] S. Cochran, G. A. Corner, K. J. Kirk, D. I. A. Lines, and M. J. Watson, “P5
C-5 design and validation of an ultrasound array optimised for epidural
needle guidance,” in Proc. IEEE Ultrason. Symp., 2007, Oct. 28–31,
pp. 2255–2258.
[11] A. Haddadi, O. Goksel, S. E. Salcudean, and K. Hashtrudi-Zaad, “On the
controllability of dynamic model-based needle insertion in soft tissue,” in
Proc. 2010 IEEE Annu. Int. Conf. Eng. Med. Biol. Soc. (EMBC), Buenos
Aires, Argentina, Aug. 31–Sep. 4, pp. 2287–2291.
[12] A. Majewicz, T. R. Wedlick, K. B. Reed, and A. M. Okamura, “Evaluation
of robotic needle steering in ex vivo tissue,” in Proc. 2010 IEEE Int. Conf.
Robot. Autom. (ICRA), Anchorage, AK, May 3–7, pp. 2068–2073.
[13] W. Yang, A. Bonvilain, T. Alonso, A. Moreau-Gaudry, and S. Basrour,
“Modeling and characterization of an instrumented medical needle in sight
of new microsensor design for its insertion guidance,” in Proc. 2010 IEEE
Annu. Int. Conf. Eng. Med. Biol. Soc. (EMBC), Buenos Aires, Argentina,
Aug. 31–Sep. 4, pp. 6465–6468.
[14] H. Kalvoy, O. G. Martinsen, and S. Grimnes, “Determination of tissue
type surrounding a needle tip by electrical bioimpedance,” in Proc. 30th
IEEE Annu. Int. Conf. Eng. Med. Biol. Soc. (EMBS) 2008, Vancouver,
BC, Aug. 20–25,, pp. 2285–2286.
[15] H. Kalvoy, L. Frich, S. Grimnes, G. Martinsen, P. K. Hol, and A.
Stubbhaug, “Impedance based tissue discrimination for needle guidance,”
Physiol. Meas., vol. 30, pp. 129–140, 2009.
[16] D. Trebbels, M. Jugl, and R. Zengerle, “Real-time cannula navigation
in biological tissue with high temporal and spatial resolution based on
impedance spectroscopy,” in Proc. Annu. IEEE Int. Conf. Eng. Med. Biol.
Soc. (EMBC) 2010, Buenos Aires, Argentina, Aug. 31–Sep. 4, pp. 1886–
1889.
[17] S. Gabriel, R. W. Lau, and C. Gabriel, “The dielectric properties of biolog-
ical tissues: II. Measurements in the frequency range 10 Hz to 20 GHz,”
Phys. Med. Biol., vol. 41, pp. 2251–2269, 1996.
[18] S. Grimnes and O. G. Martinsen, Bioimpedance and Bioelectricity Basics.
Oxford, UK: Academic Press, 2000.
[19] U. Pliquett, D. Frense, M. Sch¨
onfeldt, C. Fratzer, Y. Zhang, B. Cahill,
M. Metzen, A. Barthel, T. Nacke,and D. Beckmann, “Testing miniaturized
electrodes for impedance measurements within the beta-dispersion—A
practical approach,” J. Electr. Bioimpedance, vol. 1, pp. 41–55, 2010.
[20] T. Paavle, M. Min, and T. Parve, “Using of chirp excitation for
bioimpedance estimation: Theoretical aspects and modeling,” in Proc.
11th Int. Biennial Baltic Electron. Conf. (BEC) 2008, Tallinn, Estonia,
Oct. 6–8, pp. 325–328.
Authors’ photographs and biographies not available at the time of
publication.
... The smaller the sensitive volume, the more accurate the tissue information provided. As the spatial sensitivity distribution cannot be directly determined experimentally, Finite Element Method (FEM) simulation has proven to be a useful method [10][11][12]. ...
... Bipolar setups are characterized by the utilization of two similar electrodes for both carrying the current and picking up the voltage. Various simulative investigations were conducted showing different needle geometries that minimize the sensitivity field [11,12,19,20]. Bipolar setups exist, which use at least one needle body as active electrode or even two nested needles [12,[20][21][22][23]. ...
... Various simulative investigations were conducted showing different needle geometries that minimize the sensitivity field [11,12,19,20]. Bipolar setups exist, which use at least one needle body as active electrode or even two nested needles [12,[20][21][22][23]. In other works, two active electrodes were installed at the needle tip [24][25][26][27][28], or a concentric needle electrode designed for Electromyography (EMG) was used [29]. ...
Article
Full-text available
Needle insertion is a common procedure in modern healthcare practices, such as blood sampling, tissue biopsy, and cancer treatment. Various guidance systems have been developed to reduce the risk of incorrect needle positioning. While ultrasound imaging is considered the gold standard, it has limitations such as a lack of spatial resolution and subjective interpretation of 2D images. As an alternative to conventional imaging techniques, we have developed a needle-based electrical impedance imaging system. The system involves the classification of different tissue types using impedance measurements taken with a modified needle and the visualization in a MATLAB Graphical User Interface (GUI) based on the spatial sensitivity distribution of the needle. The needle was equipped with 12 stainless steel wire electrodes, and the sensitive volumes were determined using Finite Element Method (FEM) simulation. A k-Nearest Neighbors (k-NN) algorithm was used to classify different types of tissue phantoms with an average success rate of 70.56% for individual tissue phantoms. The results showed that the classification of the fat tissue phantom was the most successful (60 out of 60 attempts correct), while the success rate decreased for layered tissue structures. The measurement can be controlled in the GUI, and the identified tissues around the needle are displayed in 3D. The average latency between measurement and visualization was 112.1 ms. This work demonstrates the feasibility of using needle-based electrical impedance imaging as an alternative to conventional imaging techniques. Further improvements to the hardware and the algorithm as well as usability testing are required to evaluate the effectiveness of the needle navigation system.
... Given the success that EIS has shown in discriminating different types of tissue, there has been much interest in integrating the electrodes used in EIS with clinical instruments: recent publications that have explored the application of EIS to surgical needles in a variety of applications including tissue recognition probes [22] [40], prostate brachytherapy needles [14] [23], and biopsy needles for prostate cancer [28] and thyroid cancer [45]. It is seen that spectroscopy may be a suitable process to compliment the current biopsy method and tools. ...
... The authors' of [22] report an overall classification accuracy of 85.4% with high classification accuracy on muscle and tendons, but poor results with blood and fat classification. Trebbels et al. used a hollow coaxial bipolar EIS needle design to measure blood, fat, and muscle [40]. The conclusion of their analysis claims that these tissues can be classified without issue due to the strong correlation coefficient values for matching tissues, despite the large coefficient values seen in some instances for non-matching tissues [40]. ...
... Trebbels et al. used a hollow coaxial bipolar EIS needle design to measure blood, fat, and muscle [40]. The conclusion of their analysis claims that these tissues can be classified without issue due to the strong correlation coefficient values for matching tissues, despite the large coefficient values seen in some instances for non-matching tissues [40]. Mishra et. ...
Article
Full-text available
Some tumours may not be detected by ultrasound during biopsy, but recent evidence has shown that different tissues can be discerned by electric impedance. This paper explores the application of vibrotactile feedback in an electrode embedded needle to help classify tissue during fine-needle aspiration biopsy from bioimpedance measurements. The process uses electric impedance spectroscopy from 10 Hz to 349 kHz to fit the double-dispersion Cole model through the Newton-Raphson algorithm. A Naive Bayes classifier is then used on the equivalent circuit parameters to estimate the tissue at the needle tip. The method is validated through a series of experiments and user trials. The results show that the vibrotactile feedback is able to help the operator in determining the tissue the needle is in, suggesting that this may prove to be a useful supplement to the standard procedure used today. This paper explores the application of vibrotactile feedback for an electrode embedded needle to help classify tissue from electric impedance measurements. The data is fit to an equivalent circuit using Newton- Raphon’s method. A Naive Bayes classifier is trained and later used in an experiment to estimate the tissue at the needle tip and provide an assigned vibrotacticle feedback to the user.
... Thus, an electrical current with 100 kHz was considered in this work as the impact of the capacitive part, which was determined by the heterogeneity of tissue structures in the region of interest to be less than 10% [5]. Moreover, as the specific electrical resistance of venous blood is several times lower than that of the surrounding tissues [6], it is in principle possible to determine the moment of needle-electrode penetration into the blood [7][8][9][10][11][12][13][14][15][16] because, in this case, the measured impedance of the needle relative to the reference electrode decreases. Figure 1 shows a simplified diagram of the proposed measuring scheme, with the sensor based on sending a small current through the needle for electrical impedance recoding in a two-electrode setup. ...
... Moreover, as the specific electrical resistance of venous blood is several times lower than that of the surrounding tissues [6], it is in principle possible to determine the moment of needle-electrode penetration into the blood [7][8][9][10][11][12][13][14][15][16] because, in this case, the measured impedance of the needle relative to the reference electrode decreases. Figure 1 shows a simplified diagram of the proposed measuring scheme, with the sensor based on sending a small current through the needle for electrical impedance recoding in a two-electrode setup. ...
Article
Full-text available
A venipuncture is the most common non-invasive medical procedure, and is frequently used with patients; however, a high probability of post-injection complications accompanies intravenous injection. The most common complication is a hematoma, which is associated with puncture of the uppermost and lowermost walls. To simplify and reduce complications of the venipuncture procedure, and as well as automation of this process, a device that can provide information of the needle tip position into patient’s tissues needs to be developed. This paper presents a peripheral vascular puncture control system based on electrical impedance measurements. A special electrode system was designed to achieve the maximum sensitivity for puncture identification using a traditional needle, which is usually used in clinical practice. An experimental study on subjects showed that the electrical impedance signal changed significantly once the standard needle entered the blood vessel. On basis of theoretical and experimental studies, a decision rule of puncture identification based on the analysis of amplitude-time parameters of experimental signals was proposed. The proposed method was tested on 15 test and 9 control samples, with the results showing that 97% accuracy was obtained.
... Research on monopolar impedance measurement setups has been conducted, showing successful identification of fat and muscle tissue [5]. To avoid the necessity of patch electrodes on the skin, bipolar setups have also been investigated [6][7][8][9]. Some of the bipolar designs consist of a hollow needle with a coated copper wire that was inserted into the needle and exposed at the needle tip. ...
... Efforts can be taken to reduce polarization effects as much as possible, e. g., tetrapolar needle electrodes have been fabricated and tested for this purpose [10]. An alternative is to measure the resulting overall electrode impedance including all polarization effects [8]. ...
Article
Full-text available
Localized impedance measurements at the needle tip identifying the present tissue type could aid clinicians in needle procedures. To assess the sensitivity field of a hollow, bipolar needle electrode, a 3D finite element approach using COMSOL Multiphysics was chosen. The simulated bipolar needle electrode consists of two hypodermic needles (17 G and 23 G) with an insulating layer of polytetrafluoroethylene (PTFE) in between. Impedance values were recorded while steadily increasing the insertion depth of the needle electrode in a layered tissue structure of skin (dermis), fat, and blood. Simulation results reveal a highly local sensitivity volume around the needle tip that can be approximated by half a tri-axial ellipsoid with elliptic radii of 0.735 mm, 2.886 mm, and 1.774 mm. A comparison with simulated and measured impedance values shows great correspondence.
... Various designs of hollow needles for blood sampling exist in which there is an active electrode in the cannula of the needle, resulting in a coaxial layout. [30][31][32][33][34][35][36] Liu et al. 23 simulated an insertion procedure through heterogeneous tissue with three tissue types and two coaxial cannulas acting as a bipolar electrode pair. Similarly to Høyum et al., 27 the voltage distribution was used to represent the sensitivity. ...
Article
Full-text available
Despite being among the most common medical procedures, needle insertions suffer from a high error rate. Impedance measurements using electrode‐equipped needles offer promise for improved tissue targeting and reduced errors. Impedance visualization usually requires an extensive pre‐measured impedance dataset for tissue differentiation and knowledge of the electric fields contributing to the resulting impedances. This work presents two finite element simulation approaches for both problems. The first approach describes the generation of a multitude of impedances with Monte Carlo simulations for both, homogeneous and inhomogeneous tissue to circumvent the need to rely on previously measured data. These datasets could be used for tissue discrimination. The second method describes the simulation of the spatial sensitivity distribution of an electrode layout. Two singularity analysis methods were employed to determine the bulk of the sensitivity within a finite volume, which in turn enables consistent 3D visualization. The modeled electrode layout consists of 12 electrodes radially placed around a hypodermic needle. Electrical excitation was simulated using two neighboring electrodes for current carriage and voltage pickup, which resulted in 12 distinct bipolar excitation states. Both, the impedance simulations and the respective singularity analysis methods were compared with each other. The results show that the statistical spread of impedances is highly dependent on the tissue type and its inhomogeneities. The bounded bulk of sensitivities of both methods are of similar extent and symmetry. Future models should incorporate more detailed tissue properties such as anisotropy or changing material properties due to tissue deformation to gain more accurate predictions.
... Also, bioimpedance needles were recently entered in tissue analyses. [30][31][32][33][34][35] It is well known that impedance recording from live tissues sheds new light through its advantages such as high sensitivity, simple measurement method, and real-time measurement capability. 23,27,30 The study is focused on evaluating an impedancebased calibration to distinct breast lesions with borderline or high-risk nature recommended for excision from low-risk benign lesions, due to the pathological guidelines, during the sonography. ...
Article
Full-text available
Purpose: Evaluating a real-time complementary bioelectrical diagnostic device based on Electrical Impedance Spectroscopy(EIS) for improving breast imaging-reporting and data system (BIRADS) scoring accuracy, especially in high-risk or borderline breast diseases. The primary purpose is to characterize breast tumors based on their dielectric properties. Early detection of high-risk lesions and increasing the accuracy of tumor sampling and pathological diagnosis are secondary objectives of the study. Methods: The tumor detection probe (TDP) was first applied to the mouse model for electrical safety evaluations by electrical current measurement, then to 138 human palpable breast lesions undergo CNB, VAB, or FNA with the surgeon's requests. Impedance phase slope(IPS) in frequency ranges of 100 kHz to 500 kHz and impedance magnitude in f = 1kHz were extracted as the classification parameters. Consistency of radiological and pathological declarations for the excisional recommendation was then compared with the IPS values. Results: Considering pathological results as the gold standard, meaningful correlations between IPS and pathophysiological status of lesions recommended for excision (such as atypical ductal hyperplasia, papillary lesions, complex sclerosing adenosis, and fibroadenoma, etc.) were observed (p<0.0001). These pathophysiological properties may include cells size, membrane permeability, packing density, adenosis, cytoplasm structure, etc. Benign breast lesions showed IPS values greater than zero, while high-risk proliferative, precancerous, or cancerous lesions had negative IPS values. Statistical analysis showed 95% sensitivity with Area Under the Curve(AUC) equal to 0.92. Conclusion: Borderline breast diseases and high-risk lesions that should be excised according to standard guidelines can be diagnosed with TDP before any sampling process. It is a precious outcome for high-risk lesions that are radiologically underestimated to BI-RADS3, specifically in younger patients with dense breast masses, challenging in mammographic and sonographic evaluations. Also, the lowest IPS value detects the most pathologic portions of the tumor for increasing sampling accuracy in large tumors. Significance: Precise detection of high-risk breast masses, which may be declared BI-RADS3 instead of BI-RADS4a. This article is protected by copyright. All rights reserved.
Article
Full-text available
Intravenous Access (IVA) is the most common invasive medical procedure. Globally, it is estimated that over one billion IVA devices (needles/catheters) are used annually. However, the overall failure rate in this procedure is unacceptably high, reaching values between 35 and 50%. This has driven a great deal of research and technological development over recent decades, including the integration of different levels of autonomy in IVA medical devices to greatly improve this process. This paper will review these recent technical developments, including methods and systems for vein imaging and localization, needle insertion, venipuncture detection, catheter placement, and complete robotic IVA platforms. Furthermore, this paper explores emerging technical aspects, current limitations, and new research directions that may enable wider clinical translation and better acceptance of robotic IVA technologies.
Article
Objective:Liver biopsy is an essential procedure in cancer diagnostics but targeting the biopsy to the actual tumor tissue is challenging. Aim of this study was to evaluate the clinical feasibility of a novel bioimpedance biopsy needle system in liver biopsy and simultaneously to gatherin vivobioimpedance data from human liver and tumor tissues.Approach:We measured human liver and tumor impedance datain vivofrom 26 patients who underwent diagnostic ultrasound-guided liver biopsy. Our novel 18G core biopsy needle tip forms a bipolar electrode that was used to measure bioimpedance during the biopsy in real-time with frequencies from 1 kHz to 349 kHz. The needle tip location was determined by ultrasound. Also, the sampled tissue type was determined histologically.Main results:The bioimpedance values showed substantial variation between individual cases, and liver and tumor data overlapped each other. However, Mann-Whitney U test showed that the median bioimpedance values of liver and tumor tissue are significantly (p<0.05) different concerning the impedance magnitude at frequencies below 25 kHz and the phase angle at frequencies below 3 kHz and above 30 kHz.Significance:This study uniquely employed a real-time bioimpedance biopsy needle in clinical liver biopsies and reported the measured humanin vivoliver and tumor impedance data. Impedance is always device-dependent and therefore not directly comparable to measurements with other devices. Although the variation in tumor types prevented coherent tumor identification, our study provides preliminary evidence that tumor tissue differs from liver tissuein vivoand this association is frequency-dependent.
Article
Full-text available
There has been a growing interest in developing electric impedance sensing surgical tools for tissue identification during surgery. A key facet of this development is identifying distinct features that can be used to identify tissues from one another. This paper explores several feature extraction techniques and classification methods applied to electric impedance data. Furthermore, a modified forward stepwise method is proposed. The method introduces a scoring metric to help select features to add to the model, that is based off of the coefficient of variation and overlapping index from the feature’s probability density functions for each of the classes. The proposed and existing methods were applied to spectral data measured at 23 frequencies, from 132 samples across 6 different tissues including ex-vivo bovine kidney, liver and muscle, poultry liver, as well as freshly excised canine testicle and ovary samples. These methods were able to successfully find impedance spectra features for the investigated biological tissues. The best predictive accuracy was with Boruta feature extraction and a Random Forest classifier but without signifigantly reducing the number of features in the classifier model. The proposed method was able to reduce the number of features in the model to an average of 5.8 features for all tested classifiers. These methods may have use in finding features to discriminate other tissue types, possibly to aid in targeting lesions in minimally invasive cancer treatment surgeries.
Article
Full-text available
Frequent use is made of lumbar epidural and combined spinal epidural anaesthesia and analgesia for pain relief during labour and childbirth. However, the conventional use of palpation to identify the position of the relevant interspinous space can be difficult in obese parturients. The safety, low cost and capability for ultrasound to be used in real time at the point of care make it a suitable clinical imaging modality to use as an alternative. With a conventional biomedical ultrasound imaging array, with or without a needle guide, the needle cannot be made to lie exactly in the imaging plane because of the physical geometry and with lengths up to more than 100 mm for lumbar epidural procedures, it is easy for the needle to leave the imaging plane altogether. The work reported here has been directed towards a practical solution to this problem via custom array design. This takes into account the need for ergonomic manipulation by the anaesthetist, insertion of the needle exactly parallel to the ultrasonic beam, and sterility. The array incorporates two separate electroacoustic components with a space for needle insertion in between. Using standard array modules has allowed this novel configuration to be connected to a conventional imaging system with only minimal software modifications. The array is operated with its body parallel to the obese parturient's back and with an optimised clinical procedure. The procedure first involves use of the array to identify the relevant interspinous space, followed by insertion of the needle through the array, which is subsequently held in place to guide the needle tip until contact is established with the spine or can be withdrawn once the needle path is well established.
Article
Full-text available
Miniaturized electrodes are introduced in life sciences in a great number and variety. They are often designed for a special purpose without the need of quantitative analysis, such as for detecting cells or water droplets in a fluid channel. Other developments aim in monitoring a single quantity in a process where all other factors held constant. To use miniaturized electrodes for quantitative measurements, their behavior should be known in detail and stable over time in order to allow a mathematical correction of the data measured. Here we show test procedures for evaluating macroscopic but also microscopic electrodes. The most important quality parameters for electrode systems used in life science are the electrode impedance, its stability, the useful frequency range as well as the limits for applied stimulus without driving the electrode system into a non-linear region of the current/voltage relation. Proper electrode design allows a bandwidth from 100 Hz up to some MHz for impedances ranging over decades from 50 Ω up to several MΩ.
Article
Full-text available
Insertion velocity, tip asymmetry, and shaft diameter may influence steerable needle insertion paths in soft tissue. In this paper we examine the effects of these variables on needle paths in ex vivo goat liver, and demonstrate practical applications of robotic needle steering for ablation, biopsy, and brachytherapy. All experiments were performed using a new portable needle steering robot that steers asymmetric-tip needles under fluoroscopic imaging. For bevel-tip needles, we found that larger diameter needles resulted in less curvature, i.e. less steerability, confirming previous experiments in artificial tissue. The needles steered with radii of curvature ranging from 3:4 cm (for the most steerable pre-bent needle) to 2:97m (for the least steerable bevel needle). Pre-bend angle significantly affected needle curvature, but bevel angle did not. We hypothesize that biological tissue characteristics such as inhomogeneity and viscoelasticity significantly increase path variability. These results underscore the need for closed-loop image guidance for needle steering in biological tissues with complex internal structure.
Article
Full-text available
We present a smart intradermal interface suitable for skin-attached drug delivery devices. Our solution enables injections or infusions that are less invasive compared to subcutaneous injections and are leakage-free at the location of penetration. The intradermal interface is based on a 31-gauge cannula embedded in a slider, movable relative to a carrier plate that can easily be fixed onto the skin. By simply pushing the slider, the cannula is inserted into the dermis. We performed injections and infusions with stained water and polyethylene glycol (PEG) solution in ex vivo pig skin. The sizes of coloured spots in the skin range from 3.5 mm(2) to 15.4 mm(2) for stained water depending on the infused volume. Infusing stained PEG solution resulted in stained tissue areas about one order of magnitude larger. One of three investigated leakage modes is unacceptable but can be reliably avoided by proper site selection. At low flow rates and at the beginning of an infusion an initial back pressure overshoot was identified. This effect was identified as the limiting parameter for the design of small programmable or intelligent devices based on micro actuators. With the proposed easy-to-use interface, intradermal injections and infusions can be performed reliably. Therefore, it is supposed to be an ideal and clinically relevant solution for self-administration of parenteral drugs in home care applications.
Conference Paper
Ultrasound transducers are commonly used to guide needle insertion procedures such as central line placement and biopsies, but these image guided procedures have steep learning curves and high user variability. An ultrasound imaging system has been developed that may significantly simplify ultrasound guidance procedures while producing high quality volumetric imagery. The system features an affixable thin flexible ultrasound transducer that conforms to curved body surfaces and a backward propagation-based image reconstruction technique. The ultrasound imaging system was used in a preliminary demonstration to image the cross-section of a needle in a cylindrical soft tissue phantom. The images compared favorably to images obtained from a commercial ultrasound system. The encouraging imaging results demonstrate the potential of the conformal ultrasound approach for ultrasound guidance procedures. The proposed conformal ultrasound approach may significantly simplify current medical procedures, while also providing high resolution volumetric imagery.
Article
Ultrasound transducers are commonly used to guide needle insertion procedures such as central line placement and biopsies, but these image guided procedures have steep learning curves and high user variability. An ultrasound imaging system has been developed that may significantly simplify ultrasound guidance procedures while producing high quality volumetric imagery. The system features an affixable thin flexible ultrasound transducer that conforms to curved body surfaces and a backward propagation-based image reconstruction technique. The ultrasound imaging system was used in a preliminary demonstration to image the cross-section of a needle in a cylindrical soft tissue phantom. The images compared favorably to images obtained from a commercial ultrasound system. The encouraging imaging results demonstrate the potential of the conformal ultrasound approach for ultrasound guidance procedures. The proposed conformal ultrasound approach may significantly simplify current medical procedures, while also providing high resolution volumetric imagery.
Article
Research suggests that there exists a significant contrast between the dielectric properties of cancerous and normal breast tissue at microwave frequencies. Open-ended coaxial probes have been extensively used and analyzed in dielectric spectroscopy. There has been an extensive numerical and experimental analysis of these probes to ensure the highest possible accuracy in the dielectric-property estimates. For simple probes made of sections of semi-rigid cable, small geometrical changes at the probe aperture, which can result from manufacturing, handling and/or usage, may have very significant effects on the observed reflection coefficient. Consequently, a precision, 3.0 mm, stainless steel open-ended coaxial probe has been designed. The probe is built out of sections of high quality coaxial line, using dielectrics whose thermal properties are similar to those of the metal. Tolerances on probe aperture dimensions and flatness are better than 10 μm. Borosilicate glass, hermetically sealed to the metal conductors, is used as the dielectric in the final section of the probe. We present preliminary experimental and numerical reflection coefficient estimates using this probe. The carefully manufactured open-ended coaxial probe is reliable and yields accurate results, and is suitable as a sensor for high fidelity in vivo tissue measurements and characterization.
Article
To achieve automatic needle puncturing for automatic blood sampling, we attempted to detect blood flowing into the needle by measuring the electric impedance between the inside of the needle and the skin. The needle with electrode was set on a syringe with a force sensor, and used to puncture a rabbit's vein. Electric impedance suddenly decreased after a final peak of puncturing force. This result shows that the changes of electric impedance of the needle can be applied to control needle insertion for automatic blood sampling
Conference Paper
In this paper, theoretical considerations and simulation results of a method for quick estimation of the bioimpedance (BI)vector over a wide range of excitation frequencies are discussed. This method uses a chirp signal with linearly modulated frequency for excitation. The response signal is processed using cross-correlation techniques together with following fast Fourier transform (FFT). As the result, the complex spectrum of the response signal is attained in the range of frequency, determined by the spread of chirp pulse. This spectrum characterizes the frequency response of the impedance under study.
Article
Soft tissue needle guidance and steering for clinical applications has been an active topic of research in the past decade. Although dynamic feedback control of needle insertion systems is expected to provide more accurate target tracking, it has received little attention due to the fact that most available models for needle-tissue interaction do not incorporate the dynamics of motions. In this paper, we study the controllability of rigid or flexible needles inside soft tissues using mechanical-based dynamic models. The results have significant implications on the design of suitable feedback controllers for different types of needle insertion systems.