Abstract

Biomechanical study. To evaluate the cervical facet loading profile of an intervertebral cervical disc prosthesis incorporating a unique saddle-shaped articulation and to determine the effect of implantation of the prosthesis on both the loading curve and area distribution of facet loading, when compared with that of an intact specimen. This is the first study that examines the effect of implantation of a cervical disc replacement on the loading of the facet joints. Fresh-frozen ovine cervical spine specimens were used. Specimens were prepared and disarticulated to yield 6 functional spinal units, which were secured and mounted in a custom fixture on a material testing apparatus. A novel pressure sensor was constructed by combining a resistive ink thin film sensor with Fuji pressure-sensitive film. The sensor was calibrated before unilateral insertion into the cervical facet joint via a small arthrotomy. Specimens were tested in both intact condition and after surgical intervention consisting of discectomy and implantation of an intervertebral cervical disc prosthesis (CerviCore, Stryker Spine, Allendale, NJ). Specimens were tested in flexion, extension, and lateral bending to 3 Nm and in torsion to 2.5 Nm. At the termination of each loading profile, the load was maintained for 30 seconds to permit full exposure of the Fuji film. For each configuration, the maximum resultant load, load rate, and contact area pressure at the endpoint of the loading profile were determined. A paired Student t test was used to determine the differences between the intact specimen and the specimen with an implanted intervertebral disc prosthesis. There were no statistically significant differences in mean or maximum pressures between the intact specimen and the specimen with an intervertebral disc in all loading configurations. Similarly, there was no significant difference in the total measured force between the groups in all loading configurations. There was no significant difference in contact areas between the groups in flexion, lateral bending, and torsion. When evaluated in extension, the intact specimen had a mean contact area of 0.8 cm compared with 0.5 cm for the intervertebral disc (P<0.02). Biomechanical testing in an ovine model demonstrated no significant difference in measured facet pressures and forces between an intact native cervical disc specimen and a cervical intervertebral disc prosthesis using a saddle-shaped articulation. Peak and mean pressures were not demonstrated to be significantly different between the implanted and intact disc conditions. Implantation of the prosthesis resulted in a significant reduction of contact area under extension moments.

No full-text available

Request Full-text Paper PDF

To read the full-text of this research,
you can request a copy directly from the authors.

... Existing experimental methods for measuring ex vivo facet joint apposition area alter the joint structures and kinematics, 12,13,23,33,36,39 and typically overestimate true apposition area, 2,14 but computational models of experimental kinematics data can provide a method of estimating joint apposition without compromising the joint. Kinematic data, combined with specimen-specific geometries of spine segments, has been used to estimate facet cartilage apposition area in the lumbar 7 and cervical spine 30 Unlike finite element (FE) modelling, where external forces are applied to a model to predict the propagation of stress and strain in the tissue(s), kinematic modelling applies measured kinematic data to the modelled (rigid) anatomic geometry to derive geometric relationships (such as facet joint articulation area) that cannot be measured directly. ...
Article
Full-text available
During cervical spine trauma, complex intervertebral motions can cause a reduction in facet joint cartilage apposition area (CAA), leading to cervical facet dislocation (CFD). Intervertebral compression and distraction likely alter the magnitude and location of CAA, and may influence the risk of facet fracture. The aim of this study was to investigate facet joint CAA resulting from intervertebral distraction (2.5 mm) or compression (50, 300 N) superimposed on shear and bending motions. Intervertebral and facet joint kinematics were applied to multi rigid-body kinematic models of twelve C6/C7 motion segments (70 ± 13 year, nine male) with specimen-specific cartilage profiles. CAA was qualitatively and quantitatively compared between distraction and compression conditions for each motion; linear mixed-effects models ( α = 0.05) were applied. Distraction significantly decreased CAA throughout all motions, compared to the compressed conditions ( p < 0.001), and shifted the apposition region towards the facet tip. These observations were consistent bilaterally for both asymmetric and symmetric motions. The results indicate that axial neck loads, which are altered by muscle activation and head loading, influences facet apposition. Investigating CAA in longer cervical spine segments subjected to quasistatic or dynamic loading may provide insight into dislocation and fracture mechanisms.
... However, most previous studies focused on the kinematics of the intervertebral segments. Fewer in vitro and computational studies have reported on the morphology of FJs (Milne, 1991;Rong et al., 2017) and uncovertebral joints (Panjabi et al., 1991), and on how intact or TDAtreated segment kinematics and load-sharing were affected by FJs (Bauman et al., 2012;Chang et al., 2007;Kang et al., 2010;Patel et al., 2017;Stieber et al., 2011) or uncovertebral joints (Kang et al., 2010;Snyder et al., 2007;Wachowski et al., 2017Wachowski et al., , 2007Wang et al., 2016). However, little is known on the in vivo kinematics and synergistic function of the IVD and the FJs in the normal human cervical spine. ...
Article
Full-text available
Functional neck motion is achieved by the cervical segments with each composed of an intervertebral disc (IVD) and two facet joints (FJs). Using biplane fluoroscopic imaging, we investigated the ranges of motion (ROMs) of the three joints in the cervical spines (from C3 to C7) of eighteen asymptomatic subjects. Three functional neck motions were examined, including flexion-extension (FE), lateral bending (LB) and axial twisting (AT). Our measurements showed that the translations of both IVD and FJs primarily occurred in the sagittal plane during all neck motions, and the anteroposterior translations of IVDs were significantly smaller than those of the corresponding FJs (p < 0.05) at all segments. For example, the ranges of IVD and FJ anteroposterior translations at C3/4 were 2.7 ± 0.7 mm vs. 3.5 ± 1.1 mm in FE, 0.9 ± 0.5 mm vs. 4.6 ± 1.1 mm in LB, and 1.0 ± 0.5 mm vs. 3.1 ± 1.0 mm in AT. Furthermore, we introduced an IVD-FJ translation ratio, which represents the ratio of the IVD to FJ translational ROMs. In FE neck motion, the IVD-FJ anteroposterior translation ratios decreased from 0.81 ± 0.18 at C3 to 0.52 ± 0.19 at C3, indicating gradually increasing resistances of IVDs compared to FJs from the proximal to distal levels. In LB neck motion, the smallest IVD-FJ translation ratios (0.14 ± 0.09 and 0.43 ± 0.30) occurred at C4/5 for both anteroposterior and left-right translations. In AT neck motion, the largest IVD-FJ anteroposterior translation ratio (0.42 ± 0.21) occurred at C3/4, and was significantly different from those at C4/5 and C5/6 (p < 0.05). These data could be used as references for improving motion-preserving cervical treatment methods that aimed to achieve the normal ranges of translational motions of both IVD and FJs.
... Compared with ACDF, our novel prosthesis and the Prestige LP could better maintain the fact joint force and delay the facet joint degeneration, and the novel prosthesis showed more similar stress with the intact group. Consistently, in vitro studies on the biomechanics of ACDR also suggested that ACDR could maintain the normal force of the facet joint [17,29]. ...
Article
An ideal implantation of artificial cervical disc replacement (ACDR) prosthesis should preserve the cervical biomechanics and maintain normal cervical biomechanics. In this study, we designed a novel prosthesis based on the physiological curvature of the cervical endplate, and conducted an in vitro test with cadaveric cervical specimens to study its effect on cervical range of motion (ROM) and facet contact force. Eighteen cadaveric cervical specimens (C2-C7) were evaluated with a 50 N follower load and a moment of ± 2.0 N-m. Testing conditions were as follows: control (group 1), C5-C6 ACDR with a novel prosthesis (group 2), C5-C6 ACDR with the Prestige LP prosthesis (group 3) and C5-C6 cervical discectomy and fusion (ACDF) with cage internal fixation (group 4). The range of motion (ROM) of all segments and facet joint contact force were measured and analyzed. The results showed that there was no significant difference among the group 1, 2 and 3 in terms of ROM (P>0.05). The ROM of C5-C6 in the group 4 showed a significant decrease compared with the group 1, 2, and 3 (P<0.05). The group 2 had a similar facet joint force with the group 1, and there is no statistical difference among the group 1, 2 and 3 (P>0.05). The facet joint force of C5-C6 in the group 4 reduced significantly compared with the other groups (P<0.05). In conclusion, both novel cervical prosthesis and Prestige LP could better preserve the normal cervical ROM and maintain facet joint force than ACDF.
... The pressure during flexion is zero, which is consistent with the finite element method (FEM) computation reported by Faizan et al. [4] This is due to the upper and lower joints being far apart during flexion. The FEM results show that the pressure of the adjacent joint increased by three times after double-segment fusion, [19] while our results show a 50-80% increase. The reasons may be that the degree of freedom of the fusion segment could be completely constrained while using FEM; however, for the in vitro test, the motion cannot be completely removed, so the operative segment can move to some extent after fusion. ...
Article
Full-text available
Background: The cause of the adjacent segment degeneration (ASD) after fusion remains unknown. It is reported that adjacent facet joint stresses increase after anterior cervical discectomy and fusion. This increase of stress rate may lead to tissue injury. Thus far, the load rate of the adjacent segment facet joint after fusion remains unclear. Methods: Six C2-C7 cadaveric spine specimens were loaded under four motion modes: Flexion, extension, rotation, and lateral bending, with a pure moment using a 6° robot arm combined with an optical motion analysis system. The Tecscan pressure test system was used for testing facet joint pressure. Results: The contact mode of the facet joints and distributions of the force center during different motions were recorded. The adjacent segment facet joint forces increased faster after fusion, compared with intact conditions. While the magnitude of pressures increased, there was no difference in distribution modes before and after fusion. No pressures were detected during flexion. The average growth velocity during extension was the fastest and was significantly faster than lateral bending. Conclusions: One of the reasons for cartilage injury was the increasing stress rate of loading. This implies that ASD after fusion may be related to habitual movement before and after fusion. More and faster extension is disadvantageous for the facet joints and should be reduced as much as possible.
... Chang et al. [69] found that following TDR, the facet loads increased at the index level in all the bending modes, with maximum in extension. Metzger et al. [70] in a cadaver study found that facet forces were sensitive to the device placement location. In contrast, a similar study conducted by Steiber et al. [71] using ovine spines demonstrated no significant increase in the facet loading following TDR. ...
Article
Full-text available
Total disc replacement (TDR) is expected to provide a more physiologic alternative to fusion. However, long-term clinical data proving the efficacy of the implants is lacking. Limited clinical data suggest somewhat of a disagreement between the in vitro biomechanical studies and in vivo assessments. This conceptual paper presents the potential biomechanical challenges affecting the TDR that should be addressed with a hope to improve the clinical outcomes and our understanding of the devices. Appropriate literature and our own research findings comparing the biomechanics of different disc designs are presented to highlight the need for additional investigations. The biomechanical effects of various surgical procedures are analyzed, reiterating the importance of parameters like preserving uncinate processes, disc placement and its orientation within the cervical spine. Moreover, the need for a 360° dynamic system for disc recipients who may experience whiplash injuries is explored. Probabilistic studies as performed already in the lumbar spine may explore high risk combinations of different parameters and explain the differences between "standard" biomechanical investigations and clinical studies. Development of a patient specific optimized finite element model that takes muscle forces into consideration may help resolve the discrepancies between biomechanics of TDR and the clinical studies. Factors affecting long-term performance such as bone remodeling, subsidence, and wear are elaborated. In vivo assessment of segmental spine motion has been, and continues to be, a challenge. In general, clinical studies while reporting the data have placed lesser emphasis on kinematics following intervertebral disc replacements. Evaluation of in vivo kinematics following TDR to analyze the quality and quantity of motion using stereoradiogrammetric technique may be needed.
... In contrast, a similar study conducted by Steiber et al. [24] using ovine spines demonstrated no significant increase in the facet loading following TDR. Our data suggests that under hybrid loading conditions TDR maintains total facet loads similar to the intact facet loads—with lower facet loads for the SPH-I and OVL-I designs. ...
Article
Full-text available
Various ball and socket-type designs of cervical artificial discs are in use or under investigation. Many artificial disc designs claim to restore the normal kinematics of the cervical spine. What differentiates one type of design from another design is currently not well understood. In this study, authors examined various clinically relevant parameters using a finite element model of C3-C7 cervical spine to study the effects of variations of ball and socket disc designs. Four variations of ball and socket-type artificial disc were placed at the C5-C6 level in an experimentally validated finite element model. Biomechanical effects of the shape (oval vs. spherical ball) and location (inferior vs. superior ball) were studied in detail. Range of motion, facet loading, implant stresses and capsule ligament strains were computed to investigate the influence of disc designs on resulting biomechanics. Motions at the implant level tended to increase following disc replacement. No major kinematic differences were observed among the disc designs tested. However, implant stresses were substantially higher in the spherical designs when compared to the oval designs. For both spherical and oval designs, the facet loads were lower for the designs with an inferior ball component. The capsule ligament strains were lower for the oval design with an inferior ball component. Overall, the oval design with an inferior ball component, produced motion, facet loads, implant stresses and capsule ligament strains closest to the intact spine, which may be key to long-term implant survival.
Article
Computational models of experimental data can provide a noninvasive method to estimate spinal facet joint biomechanics. Existing models typically consider each vertebra as one rigid-body and assume uniform facet cartilage thickness. However, facet deflection occurs during motion, and cervical facet cartilage is nonuniform. Multi rigid-body computational models were used to investigate the effect of specimen-specific cartilage profiles on facet contact area estimates. Twelve C6/C7 segments underwent non-destructive intervertebral motions. Kinematics and facet deflections were measured. Three-dimensional models of the vertebra and cartilage thickness estimates were obtained from pre-test CT data. Motion-capture data was applied to two model types (2RB: C6, C7 vertebrae each one rigid body; 3RB: left and right C6 posterior elements, and C7 vertebrae, each one rigid body) and maximum facet mesh penetration was compared. Constant thickness cartilage (CTC) and spatially-varying thickness cartilage (SVTC) profiles were applied to the facet surfaces of the 3RB model. Cartilage apposition area (CAA) was compared. Linear mixed-effects models were used for all quantitative comparisons. The 3RB model significantly reduced penetrating mesh elements by accounting for facet deflections (p = 0.001). The CTC profile resulted in incongruent facet articulation, whereas realistic congruence was observed for the SVTC profile. The SVTC profile demonstrated significantly larger CAA than the CTC model (p < 0.001).
Article
Background: Discectomy and fusion is considered the "gold standard" treatment for clinical manifestations of degenerative disc disease in the cervical spine. However, clinical and biomechanical studies suggest that fusion may lead to adjacent-segment disease. Cervical disc arthroplasty preserves the motion at the operated level and may potentially decrease the occurrence of adjacent segment degeneration. The purpose of this study was to investigate the effect of disc generation, fusion, and disc replacement on the motion, disc stresses, and facet forces on the cervical spine by using the finite element method. Methods: A validated, intact, 3-dimensional finite element model of the cervical spine (C2-T1) was modified to simulate single-level (C5-C6) and 2-level (C5-C7) degeneration. The single-level degenerative model was modified to simulate both single-level fusion and arthroplasty (total disc replacement [TDR]) using the Bryan and Prestige LP discs. The 2-level degenerative model was modified to simulate a 2-level fusion, 2-level arthroplasty, and single-level disc replacement adjacent to single-level fusion (hybrid). The intact models were loaded by applying a moment of ±2 Nm in flexion-extension, lateral bending, and axial rotation. The motion in each direction was noted and the other modified models were loaded by increasing the moment until the primary C2-T1 motion matched that of the intact (healthy) C2-T1 motion. Results: Both Bryan and Prestige discs preserved motion at the implanted level and maintained normal motions at the adjacent nonoperative levels. A fusion resulted in a decrease in motion at the fused level and an increase in motion at the unfused levels. In the hybrid construct, the TDR (both) preserved motion adjacent to the fusion, thus reducing the demand on the other levels. The disc stresses followed the same trends as motion. Facet forces increased considerably at the index level following a TDR. Conclusion: The Bryan and Prestige LP TDRs both preserved motion at the implanted level and maintained normal motion and disc stresses at the adjacent levels. The motion patterns of the spine with a TDR more closely resembled that of the intact spine than those of the degenerative or fused models.
Article
BACKGROUND: Artificial cervical disc prosthesis simulates range of motion and buffer shock function of normal intervertebral discs. Clinical experiments verify that artificial cervical disc prosthesis material has good biocompatibility and mechanical characteristics. OBJECTIVE: To evaluate artificial cervical disc replacement and zero-profile interbody fixation and fusion system for multilevel cervical disease in 2-year follow-up. METHODS: Artificial cervical disc replacement and zero-profile interbody fixation and fusion system were used to treat 42 patients with multilevel cervical disease. The patient presented typical symptoms and signs of spinal cord or nerve root compression. There were 18 cases of cervical myelopathy, 15 cases of nerve root cervical spondylosis and 10 cases of mixed type of cervical spondylosis. After treatment, mean operation time, blood loss and reoperation rate were measured. Postoperative complications, disability index of neck function, visual analog scale, function unit range of corresponding surgery segments of the cervical spine, Cobb angle of C2-C7 vertebral body, range of motion of adjacent segment of proximal and distal vertebral bodies were observed and clinical outcomes were evaluated. RESULTS AND CONCLUSION: All cases finished the operation and were scored at various time points. After treatment, radiating pain of shoulder and neck and upper extremity were remarkably lessened. Numbness and sensory loss symptoms disappeared obviously. Quality of life elevated noticeably. Visual analog scale and the disability index of neck function score were decreased in final follow-up compared with pre-treatment (P < 0.001). C2-C7 vertebrae Cobb angle, FSU angle, range of motion of proximal surgery adjacent segment and range of motion of the distal surgery adjacent segment were elevated compared with pre-treatment (P < 0.001). These data indicate that cervical spondylosis was improved after treatment. Each index of cervical spondylosis after artificial cervical disc replacement and zero-profile interbody fixation and fusion system was reconstructed to different degrees. © 2015, Journal of Clinical Rehabilitative Tissue Engineering Research. All rights reserved.
Article
Demineralized bone particle (DBP), which is widely used as a biomaterial in the field of tissue engineering, contains various bioactive molecules, such as cytokines. For this reason, in this study we investigated the effects of injectable DBP gels on cell proliferation, inflammation and maintenance of the shape of DBP gels as a scaffold able to substitute for intervertebral discs (IVDs) in vivo. DBP gels were fabricated with different percentages (5% and 10%) of DBP powder and 3% acetic acid, including 0.02% pepsin. DBP gels with 1 × 106 annulus fibrosus (AF) cells were implanted into the dorsal subcutaneous region of BALB/C-nu mice for 1, 2 and 3 weeks. Cell proliferation was measured by MTT assay. The effect of DBP gels on the inflammatory response was analysed by measuring the amount of tumour necrosis factor-alpha (TNFα) released. Also, histological methods were carried out to analyse the response of DBP gels in vivo. This study demonstrated that injectable DBP gels are able to provide physical scaffolds for growing IVD cells in vivo. Copyright © 2013 John Wiley & Sons, Ltd.
Article
Study design: A biomechanical study of facet joint pressure after total disc replacement using cadaveric human cervical spines during lateral bending and axial torsion. Objective: The goal was to measure the contact pressure in the facet joint in cadaveric human cervical spines subjected to physiologic lateral bending and axial torsion before and after implantation of a ProDisc-C implant. Summary of background data: Changes in facet biomechanics can damage the articular cartilage in the joint, potentially leading to degeneration and painful arthritis. Few cadaveric and computational studies have evaluated the changes in facet joint loading during spinal loading with an artificial disc implanted. Computational models have predicted that the design and placement of the implant influence facet joint loading, but limited cadaveric studies document changes in facet forces and pressures during nonsagittal bending after implantation of a ProDisc. As such, little is known about the local facet joint mechanics for these complicated loading scenarios in the cervical spine. Methods: Seven osteoligamentous C2-T1 cadaveric cervical spines were instrumented with a transducer to measure the C5-C6 facet pressure profiles during physiological lateral bending and axial torsion, before and after implantation of a ProDisc-C at that level. Rotations at that level and global cervical spine motions and loads were also quantified. RESULT.: Global and segmental rotations were not altered by the disc implantation. Facet contact pressure increased after implantation during ipsilateral lateral bending and contralateral torsion, but that increase was not significant compared with the intact condition. Conclusion: Implantation of a ProDisc-C does not significantly modify the kinematics and facet pressure at the index level in cadaveric specimens during lateral bending and axial torsion. However, changes in facet contact pressures after disc arthroplasty may have long-term effects on spinal loading and cartilage degeneration and should be monitored in vivo.
Article
Studies delineating the adjacent level effect of single level disc replacement systems have been reported in literature. The aim of this study was to compare the adjacent level biomechanics of bi-level disc replacement, bi-level fusion and a construct having adjoining level disc replacement and fusion system. In total, biomechanics of four models- intact, bi level disc replacement, bi level fusion and fusion plus disc replacement at adjoining levels- was studied to gain insight into the effects of various instrumentation systems on cranial and caudal adjacent levels using finite element analysis (73.6N+varying moment). The bi-level fusion models are more than twice as stiff as compared to the intact model during flexion-extension, lateral bending and axial rotation. Bi-level disc replacement model required moments lower than intact model (1.5Nm). Fusion plus disc replacement model required moment 10-25% more than intact model, except in extension. Adjacent level motions, facet loads and endplate stresses increased substantially in the bi-level fusion model. On the other hand, adjacent level motions, facet loads and endplate stresses were similar to intact for the bi-level disc replacement model. For the fusion plus disc replacement model, adjacent level motions, facet loads and endplate stresses were closer to intact model rather than the bi-level fusion model, except in extension. Based on our finite element analysis, fusion plus disc replacement procedure has less severe biomechanical effects on adjacent levels when compared to bi-level fusion procedure. Bi-level disc replacement procedure did not have any adverse mechanical effects on adjacent levels.
Article
The purpose of this study was to develop a clinical method for measuring three-dimensional motion of the cervical spine using the CA 6000 Spine Motion Analyzer (Orthopaedic Systems Inc., Hayward CA). Normal values for passive examinations of flexion-extension, lateral bending, rotation, rotation out of maximum flexion, and rotation out of maximum extension were obtained and analyzed for each gender in a group of 150 normal subjects. Gender classifications were further subdivided into age groups, with each decade containing asymptomatic volunteers. Values for each group were compared for differences with respect to age and gender differences. A detailed error analysis was also performed on the interobserver and intraobserver repeatability, differences between passive and active testing, and the use of different fixation devices. Significantly decreased motion differences were found between age groups within gender, and between gender groups in corresponding decades. Results of rotation out of maximum flexion suggest and support earlier conclusions that the rotation of the C1-C2 segment does not decrease with age, but rather increases slightly to perhaps compensate for the overall decreased motion in the lower segments.
Article
Data regarding intra-joint loads during range of motion is essential to understanding normal joint mechanics and pathology. The investigators configured and characterized the response of a 440 N range, 0.076 mm thick commercial thin film sensor to monitor joint loads through a range of motion. Following preconditioning, static and dynamic tests were performed to evaluate the sensor response under varied environmental conditions. Both tests utilized a fixture to align the sensor and applied load. Under conditions that included dry, wet, folded and kinked configurations, a static load of 100 N was applied and the sensor output monitored up to thirty minutes from the time of loading. A load of 400 N at 50 N/s was applied to the sensor to determine dynamic characteristics and calibration curves under the conditions described for static tests in addition to curved, hard and soft contact surfaces and sensor overload. Static kinked data was significantly different from the dry, wet and folded conditions. Dynamic data showed that inter-package variability was not significant but that differences between sensor packages and sensor configuration conditions were significant. To investigate applicability of these sensors to the field of orthopaedics, a cadaveric knee was instrumented with sensors to examine the role of the meniscus in load transmission and distribution across the knee. The sensors were placed bilaterally below each meniscus on the anterior, posterior and center of the tibial plateau. Sensor data were obtained at these locations during manually flexed range of motion for the intact, re-attached and lateral menisectomized conditions of the knee specimen.
Article
Through the use of pressure-sensitive color film under static loading conditions, intra-joint loads, contact area, and contact pressure have been determined. Intra-joint loads have also been elucidated through the use of computationally intensive computer simulations. We present an experimental technique and loading mechanism that can provide dynamic intra-joint loads during range of motion. The goal of the study was to establish a dynamic range-of-motion testing protocol encompassing measurement of intra-joint loads while retaining the characteristics of joint compressive loads under minimal constraints to motion. ### Materials and Methods Cadaveric knee specimens were mounted in a testing frame as shown in Figure 1. The quadriceps muscle was fastened to the hydraulic actuator of a materials testing machine (MTS Systems, Eden Prairie, Minnesota) through the proximal insertion site of the patellar tendon. The vastus lateralis and medialis and hamstrings muscles were each loaded with 45 N of tensile load at angles of 17°, 50°, and 0°, respectively, in relation to the femoral axis as described by Lieb and Perry1. A knee flexion moment was generated by a 90-N force equally distributed on both sides of the tibial fixture. All pulleys were terminated and mounted with free rotating swivel joints to reduce external bending moments (Fig. 1). Fig. 1: Apparatus for dynamic testing. MTS = materials testing machine, VL = vastus lateralis, and VMO = vastus medialis oblique. To account for the varied stiffness of the knee during range of motion, an external goniometer mounted at the approximate center of knee rotation was used by the servohydraulic controller as control feedback for the actuator motion. This feedback provided reduced actuator displacement rates at increased flexion angles, at which increased resistance to knee motion was encountered as a result of muscle forces. For all of the specimens tested, knee range of motion was from 15° to …
Article
In vitro study on the effect of increasing the height of the cervical disc space on the facet joint. To demonstrate how facet joint articulation is affected by increasing the cervical disc space height. A surgeon attempts to increase the disc space and inserts a larger artificial disc than normal in order to keep the intervertebral foramen open and the prosthesis stable. However, it is hypothesized by the current authors that this procedure could have an adverse effect on the facet joints. Computerized tomography images passing through the disc space and the center of the C4-C7 facet joints (sagittal plane) were obtained from 15 cadaveric cervical spine specimens. A 1-mm incremental increase to a total 5 mm in disc space height was performed to simulate the changes seen in disc replacement. The change in the facet joint articulation overlap and space in the sagittal plane at normal and each displacement was measured. Each 1-mm incremental increase in disc space at C4-C5 translated to a decrease in the facet joint articulation overlap in the sagittal plane by approximately 8%. The mean facet joint space increased approximately 0.8 mm. At the C5-C6 and the C6-C7 levels, the articulation overlap decreased by approximately 7% and the facet joint space increased approximately 0.8 mm. There is a significant decrease of the facet joint articulation overlap in the sagittal plane and an increase in the facet joint space following an increase in the cervical disc space. The inappropriate increase of the disc space height may result in facet joint subluxation and could lead to the accelerated failure of the artificial disc.