Evaluation of rebound tonometry in non-human primates

Dept. of Ophthalmology & Visual Sciences, University of Wisconsin-Madison, Madison, WI 53792, USA.
Experimental Eye Research (Impact Factor: 2.71). 02/2011; 92(4):268-73. DOI: 10.1016/j.exer.2011.01.013
Source: PubMed

ABSTRACT

To determine the accuracy and reproducibility of intraocular pressure (IOP) measurements obtained with the TonoVet® rebound tonometer in cynomolgus macaques and to determine the effects of corneal thickness on measurements obtained by the TonoVet®. The anterior chambers of both eyes of anesthetized monkeys were cannulated with branched 23-G needles; one branch was connected to a vertically adjustable reservoir and the other to a pressure transducer. IOP was increased by 5 mmHg increments and then decreased by 10 mmHg decrements. IOP was measured using the TonoVet® at each increment and decrement by 2 independent observers and at every other increment and every decrement by a single observer using 'minified' Goldmann applanation tonometry. Central corneal thickness was measured with a PachPen(TM) ultrasonic pachymeter. TonoVet® readings correlated well with manometric IOP (slope = 0.972, r(2) coefficient = 0.955). No significant differences were observed when comparing eyes or operators; however there was a non-significant trend for TonoVet® readings taken in right eyes to be closer to manometric IOP than those taken in left eyes. The TonoVet® had a non-significant tendency to underestimate manometric IOP. TonoVet® readings obtained during the decremental phase of the experiment were significantly closer (p < 0.004) to manometric IOP than those obtained during the incremental phase. Central corneal thickness significantly increased (p < 0.0001) over the course of the experiment. The TonoVet® rebound tonometer is a reliable and accurate tool for the measurement of IOP in cynomolgus macaques. This tonometer has several advantages, including portability, ease of use, and brief contact with the corneal surface making topical anesthetics unnecessary.

Download full-text

Full-text

Available from: Gillian J Mclellan
  • [Show abstract] [Hide abstract]
    ABSTRACT: Objective To validate intraocular pressure (IOP) readings obtained in cats with the TonoVet(®) tonometer. Animals studied IOP readings obtained with the TonoVet(®) were compared to IOP readings determined by manometry and by the Tono-Pen XL(™) in 1 normal cat and two glaucomatous cats. TonoVet(®) and Tono-Pen XL(™) readings were also compared in a further six normal and nine glaucomatous cats. Procedures The anterior chambers of both eyes of three anesthetized cats were cannulated and IOP was varied manometrically, first increasing from 5 to 70 mmHg in 5 mmHg increments, then decreasing from 70 to 10 mmHg in 10 mmHg decrements. At each point, two observers obtained three readings each from both eyes, with both the TonoVet(®) and Tono-Pen XL(™) . IOP was measured weekly for 8 weeks with both tonometers in six normal and nine glaucomatous unsedated cats. Data were analyzed by linear regression. Comparisons between tonometers and observers were made by paired student t-test. Results The TonoVet(®) was significantly more accurate than the Tono-Pen XL(™) (P = 0.001), correlating much more strongly with manometric IOP. In the clinical setting, the Tono-Pen XL(™) underestimated IOP when compared with the TonoVet(®) . Conclusions Both the TonoVet(®) and Tono-Pen XL(™) provide reproducible IOP measurements in cats; however, the TonoVet(®) provides readings much closer to the true IOP than the Tono-Pen XL(™) . The TonoVet(®) is superior in accuracy to the Tono-Pen XL(™) for the detection of ocular hypertension and/or glaucoma in cats in a clinical setting.
    No preview · Article · Jun 2012 · Veterinary Ophthalmology
  • [Show abstract] [Hide abstract]
    ABSTRACT: A large number of factors are important in conducting anti-glaucoma drug efficacy studies. It is essential to have an understanding of aqueous humor dynamics and how the tonometer, tonometrist, and animal may affect IOP estimates. Additional critical considerations in the design of an anti-glaucoma drug efficacy studies include the following: (1) selecting the most appropriate species, (2) identifying the rate of nonresponders within the study population, (3) determining whether normotensive or glaucomatous animals should be used, and deciding (4) what secondary endpoints (if any) to include, and (5) whether one eye or both should be dosed. Anti-glaucoma drug efficacy studies have an acclimation phase in which the animal becomes conditioned to the procedures, a predose phase in which baseline data is collected, a dosing phase in which the drug is administered and IOP and possibly other endpoints are monitored, and a recovery phase in which IOP returns to predose values as the drug is washed out before another predose phase is started.
    No preview · Chapter · Jan 2013
  • [Show abstract] [Hide abstract]
    ABSTRACT: Objective: To compare intraocular pressure (IOP) measurements in children 18 years of age and younger using rebound tonometry and applanation tonometry and the feasibility of using these techniques in children. Methods: Literature searches of the PubMed and the Cochrane Library databases were last conducted in June 2012 and resulted in 43 citations, including citations not in English. Of these 43 citations, 4 studies met the inclusion criterion following full text review. A level of evidence rating was assigned to each study using criteria specifically developed for this assessment. Results: No level I study was found in the literature search, and 2 level II and 2 level III studies were identified. Intraocular pressure was 2 to 3 mmHg higher using rebound tonometry compared with Goldmann applanation tonometry in the 2 level II studies performed in a clinic setting and in 1 level III study performed on children under general anesthesia. However, IOP was lower in 1 level III study in which noncontact applanation tonometry was compared with rebound tonometry. Bland-Altman plots showed that the difference in IOP for rebound versus Goldmann applanation tonometry increased as the IOP increased. The success rate for measuring IOP was markedly higher in children 3 years of age and younger using rebound tonometry compared with noncontact tonometry in 1 level III study. Conclusions: Rebound tonometry seems to be a reasonably accurate instrument that allows the IOP to be measured in many children without using general anesthesia. More data are required to assess better how the differences between instruments vary with IOP measurement.
    No preview · Article · Feb 2013 · Ophthalmology
Show more