Background: Transcranial Magnetic Stimulation (TMS) therapies use both focal and unfocal coil designs. Unfocal designs often employ bendable windings and moveable parts, making realistic simulations of their electric fields in inter-individually varying head sizes and shapes challenging. This hampers comparisons of the various coil designs and prevents sys-tematic evaluations of their dose-response relationships. Objective: Introduce and validate a novel method for optimizing the position and shape of flexible coils taking individual head anatomies into account. Evaluate the impact of realistic modeling of flexible coils on the electric field simulated in the brain. Methods: Accurate models of four coils (Brainsway H1, H4, H7; MagVenture MST-Twin) were derived from computed tomography data and mechanical measurements. A generic representation of coil deformations by concatenated linear transformations was introduced and validated. This served as basis for a principled approach to optimize the coil positions and shapes, and to optionally maximize the electric field strength in a region of interest (ROI). Results: For all four coil models, the new method achieved configurations that followed the scalp anatomy while robustly preventing coil-scalp intersections on N=1100 head models. In contrast, setting only the coil center positions without shape deformation regularly led to physically impossible configurations. This also affected the electric field calculated in the cortex, with a median peak difference of ~16%. In addition, the new method outperformed grid search-based optimization for maximizing the electric field of a standard figure 8 coil in a ROI with a comparable computational complexity. Conclusion: Our approach alleviates practical hurdles that so far hampered accurate simula-tions of bendable coils. This enables systematic comparison of dose-response relationships across the various coil designs employed in therapy.
Content may be subject to copyright.
Personalized electric field simulations of deformable large TMS
coils based on automatic position and shape optimization
Torge Worbsa,b,#, Bianka Rumib,c,d,#, Kristoffer H. Madsen b,c,#, Axel Thiel-
schera,b,#
a Section for Magnetic Resonance, DTU Health Tech, Technical University of Denmark, Kgs Lyngby,
Denmark
b Danish Research Centre for Magnetic Resonance, Department of Radiology and Nuclear Medicine,
Copenhagen University Hospital Amager and Hvidovre, Copenhagen, Denmark
c Section for Cognitive Systems, Department of Applied Mathematics and Computer Science, Tech-
nical University of Denmark, Kgs Lyngby, Denmark
d Sino-Danish College, University of Chinese Academy of Sciences, Beijing 100190, China
# All authors equally contributed to this work.
Corresponding Author
Kristoffer H. Madsen, Assoc. Prof.
Department of Applied Mathematics and Computer Science, Richard Petersens Plads, Build-
ing 321, Technical University of Denmark, 2800 Kgs. Lyngby, Denmark
Tel: + 45- 45253895
e-mail: khma@dtu.dk
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
2/24
Abstract
Background Transcranial Magnetic Stimulation (TMS) therapies use both focal and unfocal
coil designs. Unfocal designs often employ bendable windings and moveable parts, making
realistic simulations of their electric fields in inter-individually varying head sizes and shapes
challenging. This hampers comparisons of the various coil designs and prevents systematic
evaluations of their dose-response relationships.
Objective Introduce and validate a novel method for optimizing the position and shape of flex-
ible coils taking individual head anatomies into account. Evaluate the impact of realistic mod-
eling of flexible coils on the electric field simulated in the brain.
Methods Accurate models of four coils (Brainsway H1, H4, H7; MagVenture MST-Twin)
were derived from computed tomography data and mechanical measurements. A generic rep-
resentation of coil deformations by concatenated linear transformations was introduced and
validated. This served as basis for a principled approach to optimize the coil positions and
shapes, and to optionally maximize the electric field strength in a region of interest (ROI).
Results For all four coil models, the new method achieved configurations that followed the
scalp anatomy while robustly preventing coil-scalp intersections on N=1100 head models. In
contrast, setting only the coil center positions without shape deformation regularly led to phys-
ically impossible configurations. This also affected the electric field calculated in the cortex,
with a median peak difference of ~16%. In addition, the new method outperformed grid search-
based optimization for maximizing the electric field of a standard figure 8 coil in a ROI with a
comparable computational complexity.
Conclusion Our approach alleviates practical hurdles that so far hampered accurate simulations
of bendable coils. This enables systematic comparison of dose-response relationships across
the various coil designs employed in therapy.
Keywords
TMS, electric field calculations, personalized dosing, Brainsway H-coils, MagVenture MST-
twin coil
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
3/24
Highlights
automatic positioning and shape optimization of large deformable TMS coils
ensures adherence to the head anatomy and prevents coil-head intersections
enable automatic electric field maximization in target brain regions
outperforms grid search for standard flat coils
provides accurate computational models of four coils used in clinical practice
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
4/24
Introduction
Transcranial Magnetic Stimulation (TMS) has been approved by several medical agencies as
therapy against specific psychiatric diseases, including major depressive disorder and obses-
sive-compulsive disorder [1]. Interestingly, the clinically employed coil designs vary substan-
tially and range from standard rigid figure-8 geometries to large and deformable coils [2–4].
Despite being used for the same clinical indications, the different designs induce electric fields
in the brain that vary strongly in their spatial distribution and focality. The individual head and
brain anatomies additionally influence the induced electric field (E-field) [5]. It is therefore
important to understand how these two factors affect the therapeutic effects.
Personalized E-field simulations informed by structural magnetic resonance imaging (MRI)
[5,6] can be useful tools to explore these questions, for example during clinical trials. However,
this was so far practically difficult or infeasible for most of the therapeutically used large coils
with complex shapes. So far accurate computational models of these coils for use in the simu-
lations were mostly lacking. More importantly, the available simulation software does not have
automatic means to prevent intersections between the coil model and the head model which
will create physically impossible coil configurations and field distributions. Instead, time-con-
suming and practically tedious manual positioning of non-flat coil geometries on the head
model is required. So far, shape adjustments of deformable coils were not supported, prevent-
ing realistic simulations of those coils.
The SimNIBS software package is an open-source tool for personalized E-field simulations [7]
of transcranial magnetic and electric stimulation. While SimNIBS already includes many val-
idated coil models [8], so far, it lacked support for coils with deformable and movable parts. In
this paper, we introduce geometrically accurate models of the clinically used Brainsway H1,
H4, and H7 coils and the MagVenture MST-twin coil [9]. We further describe a computation-
ally efficient approximation of non-linear deformations of coil shapes by concatenated linear
transformations. We use this to establish a principled optimization approach for fitting the coil
position and shape to individual head shapes while avoiding intersections between the coil and
the head models and, in case of the MST-twin coil, also between the two individually movable
coil parts. The approach is generic and can be easily extended to further coil designs.
We assess two application scenarios, in which we evaluate the stability of our approach by
reporting the average E-field distribution induced in the brain for 1100 head models. In the first
scenario, the coil casing is fitted as close as possible to the head surface, to reach a physically
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
5/24
feasible coil configuration close to an initial position provided by the user. In the second sce-
nario, the coil position and shape are optimized to maximize the E-field in a cortical region of
interest (ROI). For further validation, we apply the approach to a flat figure-8 coil and compare
the optimized E-field with the results of a standard grid search.
Methods
Modeling of the Brainsway H1, H4 and H7 coils and the MagVenture MST-twin coil
Physical samples of the Brainsway H1, H4 and H7 coil models were scanned in a clinical com-
puterized tomography (CT) scanner, whereby the coils were carefully placed to avoid defor-
mations. The wire paths were then traced manually in the scans (Suppl. Material A). Calcula-
tions of the magnetic fields and magnetic vector potentials of the coils were then implemented
based on numerical solutions of line integrals [10] using the fast multipole method for im-
proved efficiency [11]. A wire was approximated by a single line element placed in its center
with a resolution of two line elements per millimeter. This resulted in a numerical error below
0.04% compared to reference cases with very dense sampling. This choice enabled good nu-
merical accuracy (Suppl. Material A) while maintaining computational efficiency. When posi-
tioning the H1, H4 and H7 coils physically on heads, the resulting deformations occurred as
bends of specific wire paths, which could be well approximated in the models by rotations
around “bending axes” or, more generally, affine transformations. If needed, more complex
deformations were represented by chaining affine transformations that are successively applied
to different sub-parts of the wire paths. By this procedure, coil deformations can be efficiently
expressed as a small number of affine transformations. Based on pilot tests on various head
shapes, we visually identified the deforming wire paths and defined appropriate bending axes
and physically feasible transformation ranges. The coil wires are attached to a fabric cap and
include soft padding. After fitting the coils firmly on a head using the integrated straps, we
measured the minimal gap between skin and wires caused by the cap and padding materials.
For modeling this gap, we then created surfaces around the wire paths (Fig. 1A) to account for
the padding when computationally fitting the coil model on head models. We assessed the
accuracy of the modeled deformations and gaps by performing a CT scan of the H1 coil fitted
on a ball with a 200 mm diameter, confirming that the computational coil model approximated
the physically occurring deformation well (Suppl. Material B).
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
6/24
The MagVenture MST-twin coil has two “sub-coils” that are connected to a guide rail, which
determines the range of feasible positions of the two sub-coils. Using an existing computational
model of an MST sub-coil [8] and measurements on a physical coil sample, we created a model
of the complete MST coil consisting of representations of the two sub-coils and the guide rail
(Fig. 1A). The feasible positions of the sub-coils were tested on the physical coil sample and
represented by defining suited linear transformations and parameter ranges.
A new Json-based file format (“.tcd” – TMS coil definition, Suppl. Material C) was created for
the generic representation of the four coil models in the SimNIBS simulation environment, also
substantially simplifying addition of further coil models with complex shapes.
Optimization of coil position and shape: Cost functions
We implemented automatic optimizations of the coil position and shape, supporting two dif-
ferent application scenarios. In the first scenario, the objective is to smoothly fit the coil casing
on the head surface, starting from an initial position provided by the user and then adjusting
the position and deformation of the coil. Thereby, a minimal distance to the head surface is
ensured and self-intersections (e.g. intersections of the two sub-coils of the MST-twin coil) are
avoided. This was formalized in a cost function
𝑓
 𝑑𝑝
𝑝
,
where the distance 𝑑 is determined as the average distance of predefined positions on the rele-
vant coil wires or casing parts that are close to the head (Suppl. Fig. S4). The intersection
penalty variable 𝑝 is based on the calculation of the intersection volume between the coil cas-
ing volume and the head volume, whereby deeply intersecting parts are weighted more to sup-
port faster convergence of the optimization. The self-intersection penalty variable 𝑝
is the total
intersection volume between sub-parts of a coil.
The second scenario aims at the maximization of the electric field strength in a ROI in the brain
while preventing intersections between the coil and head and coil self-intersections. The cor-
responding cost function has the form:
𝑓
||
|𝐸| 𝑝
𝑝
The variable |𝐸| is the average electric field strength in the ROI. Appropriate weighting con-
stants for the different parts of the cost functions were set in pilot tests to ensure robust con-
vergence of the optimization and avoidance of intersections. Additional information about the
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
7/24
definition and implementation of the different parts of the cost functions is stated below and in
Supplementary Material D.
Optimization framework
The optimization was performed with a combination of global optimization (DIRECT or the
faster locally-biased DIRECT-L [12]) followed by additional local optimization using a quasi-
Newton method (L-BFGS-B [13]). Unless stated otherwise, DIRECT-L and L-BFGS-B were
chosen as the default methods for the results reported in the main paper, which provided a good
tradeoff between the number of function evaluations and the optimization result. The default
method was compared to DIRECT and L-BFGS-B as reference.
Parameter bounds were set according to the physically feasible coil deformation ranges.
In addition, the coil center positions and orientations were restricted to reasonable ranges, as
determined in pilot tests. The ranges for coil translations, rotations, and deformations are listed
in Supplementary Table S4.
To evaluate the general performance of our approach, we compared our e-field optimi-
zation results for the non-flexible MagVenture Cool-B65 coil to a standard grid search ap-
proach [14]. As local optimization turned out to be sufficient to ensure adequate optimization,
the MagVenture Cool-B65 results reported in the main paper only utilized local optimization
with L-BFGS-B, while further results are shown in Supplementary Material E.
Selection of initial coil positions and orientations, and definition of target ROIs
The initial coil placements for optimization were set according to the coil model and optimiza-
tion goal. Unless stated otherwise, the coil positions and orientations were manually defined in
MNI space (Fig. 1A), in an effort to minimize the coil-head intersections, and then non-linearly
transformed to the individual subject spaces using SimNIBS functions. These coil settings are
included as the “baseline” in the results to illustrate how well non-deformable coil models op-
timized once on a template would work.
Distance optimization of the Brainsway H1 coil: The initial position of the coil center (see
Suppl. Fig. S5 for details) was chosen to be above the center of area BA46 of the MNI
template head (MNI coordinates [-44, 40, 29]), a common target in the treatment of MDD
[15]. The orientation of the coil was manually adjusted along the left-right and anterior-
posterior axes to minimize the intersection of the coil with the MNI template.
Distance optimization of the Brainsway H4 coil: The initial coil position was selected to
mimic the clinical guidelines for coil placement as provided by Brainsway. The coil center
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
8/24
was placed on the MNI template head above the group-average activation site of the FDI
muscle (MNI coordinates [41, 7, 63] [16]), then moved medially to a position above the
interhemispheric cleft and finally projected 6 cm anteriorly, resulting in the MNI position
[0, 53, 63] (Suppl. Fig. S6). The orientation of the coil along the left-right axis was manu-
ally optimized to minimize coil-head intersections. The rotations along the other two axes
were chosen to ensure symmetric placement of the coil above both sides of the head.
Distance optimization of the Brainsway H7 coil: A position between the medial parts of
the primary motor areas that contain the leg representations was visually determined and
the closest scalp position was chosen (MNI coordinates [-5, -20, 99]). The treatment posi-
tion was then found by moving 4 cm in the anterior direction (MNI coordinates [-5, 20, 87];
Suppl. Fig. S7), according to clinical guidelines of the company.
Distance and E-field optimization of the MagVenture MST-Twin coil: Clinical applications
aim to place the two coil halves above the F3/F4 electrode positions, respectively [17,18].
The coil center was thus placed at the Fz electrode position between F3/F4 (Suppl. Fig. S8),
and the coil model was additionally moved in inferior-superior direction to achieve a place-
ment of the two sub-coils approximately on the skin surface. As the MST-Twin coil is
designed for seizure induction by stimulating frontal areas, a large bi-lateral frontal ROI
was used as target for the E-field optimizations (Fig. 1B).
E-field optimization of the MagVenture Cool-B65 coil: As the Cool-B65 coil is used for
focal stimulation, we defined the hand knob area of the left precentral gyrus as an example
target ROI (Fig. 1C). The coil position was directly initialized in subject space by automat-
ically determining the skin position closest to the brain ROI (termed “auto-init” in the fol-
lowing). Alternatively, the coil position was centered over the hand knob (MNI coordinate
[-32, -26, 59]), and orthogonally projected to a 4 mm skin distance from the scalp after
transformation to subject space. The coil handle was oriented backwards and approximately
45 degrees lateral from the mid-line thereby ensuring that it pointed in the direction oppo-
site to EEG 10-10 position FCz. The results for the “auto-init” option are reported in the
main part of the paper, while the subject-specific settings serve as comparison and are re-
ported in Supplementary Figure S10C and Supplementary Table S2.
Software Implementation
The methods and coil models presented in this paper will be published as open source software
in the next SimNIBS release. Also examples on how to define additional custom coil models
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
9/24
will be included. The intersection tests are based on voxel masks (1 mm³ isotropic) of the coil
and head surfaces that are automatically voxelized during runtime [19]. Intersection volumes
between two masks are then efficiently determined by interpolating one mask to the voxel
space of the second mask (using the map_coordinates function in SciPy 1.13.1 [20]), according
to the linear transformation representing the shape deformation, and summing overlapping
voxels. The evaluation of the distance term was implemented in a similar way, but weighted
by the intersection distance (Suppl. Material D). The speed of the FEM calculations of the E-
field in the brain ROIs was optimized for computational efficiency using a similar approach as
outlined in Cao, Madsen et al [21]. The implementations of DIRECT, DIRECT-L and L-BFGS-
B in SciPy 1.13.1 [20] were used. Visualizations were created with Matplotlib 3.9.2 [22] and
PyVista 0.44.1 [23].
Evaluations of the Optimization Performance
To evaluate the stability of the optimization approach for varying head shapes, results were
assessed on the Human Connectome Project database of young healthy adults (N=1100) [24].
SimNIBS charm [25] was used for the creation of the head models from the T1- and T2-
weighted structural MRI images, incorporating reconstructions of the pial and white matter
surfaces from FreeSurfer [26] for more accurate representations of narrow sulci in the head
models. The resulting head meshes consisted of ~4.6 million tetrahedral elements representing
seven tissue types (white matter, gray matter, cerebrospinal fluid, compact bone, spongy bone,
scalp, and eye balls), and their default conductivity values were used in the simulations (0.126
S/m, 0.275 S/m, 1.654 S/m, 0.008 S/m, 0.025 S/m, 0.465 S/m, and 0.5 S/m). All simulations
were performed for a rate of change of the coil current of 1 A/µs, and the E-field strength was
evaluated on the central gray matter surfaces of two hemispheres halfway between the pial and
white matter surfaces. All head meshes were visually inspected for quality assurance. In addi-
tion to the results obtained on the large dataset, speed tests were performed on a desktop com-
puter (Ubuntu 22.04, Intel i7-11700, 16 cores, 32GB RAM) using the public SimNIBS ernie
dataset with standard conductivities to demonstrate expected run times in practical settings.
The target ROIs for the E-field optimizations were defined in the fsaverage surface space and
automatically transformed to the individual subject spaces using SimNIBS functionality.
The evaluations were based on the following criteria:
Distance optimizations: The effect of the optimization on coil position and shape was in-
dexed using the 𝑑 metric (distance, as defined above) before and after optimization. In ad-
dition, a signed distance measure was calculated between the head mesh and the coil casing,
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
10/24
whereby negative values indicated the maximal (or deepest) intersections of the coil with
the head, while positive values correspond to the minimal distance between the coil casing
and the scalp in case of no intersections.
For evaluating the effect of the optimization on the calculated electric fields, simulations
were performed for the baseline configuration and after optimization. The electric field
strengths on the central gray matter surfaces of both hemispheres (i.e. on surfaces midway
between the pial surface and gray-white matter boundary) were determined and the differ-
ences between baseline and the optimized coil configuration were calculated and scaled
relative to the 99.9-percentile of the electric fields after optimization. The results were
mapped to the fsaverage surface space to obtain group difference maps.
The computational efficiency of the optimization was determined using the average (+/-
standard deviation) number of function evaluations. In addition, the runtime on the test
desktop computer for the SimNIBS ernie dataset was evaluated.
E-field optimizations of the MagVenture MST-Twin coil: The average electric field strength
in the ROI (Fig. 2a) at baseline and after optimization was compared. In addition, signed
distance measures were calculated for both cases.
E-field optimizations of the MagVenture Cool-B65 coil: Signed distance measures were
calculated to confirm the absence of intersections after optimization. In addition, we com-
pared the average electric field strength in the ROI achieved with the new optimization to
the results of standard grid search implemented in SimNIBS [14]. The latter was feasible
for this coil as the Cool-B65 is flat and does not deform. Grid search was performed on the
full dataset (N=1100) with a grid spacing of 5 mm and 12 orientations per position with an
angular spacing of 15° (585 simulations per grid). In addition, a higher-resolution grid
search was performed on a random subset of N=48 subjects with a spacing of 2 mm and 36
orientations per position with an angular spacing of 5° (11285 simulations per grid). The
grid search only optimizes the position of the coil center on the skin surface and the orien-
tation of the coil handle, while the coil is always selected to be parallel to skin surface under
the coil center to simplify the search. However, this resulted in skin intersections for several
of the grid search results due to irregular head shapes. To enable a fair comparison, the final
coil position was moved orthogonally away from head in these cases until the intersections
were resolved, and the electric field in the ROI was reevaluated.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
11/24
Results
Distance optimization
The median electric field distributions across the 1100 head models and the interquartile ranges
(IQR) after distance optimization are shown in Figure 2 for the H1, H4, H7 and MST coils.
Table 1 lists the peak E-field strengths in the cortex (defined as the 99.9%ile), focality measures
(defined as the gray matter volume where the E-field strength exceeds 50% of the peak strength)
and depth measures (defined as the radial distance from the brain surface to the deepest point
in gray matter where the E-field strength is half of the peak strength) [27,28].
The optimization reduced the interindividual spread of the coil-scalp distances compared to the
initial positions (Fig. 3A), suggesting that it led to a more consistent fit of the coil models on
the head surfaces. It successfully removed all intersections of the coils with the heads and
achieved consistent minimal distances to the skin surfaces (Fig. 3B). The results shown for the
initial positions demonstrate that using coil positions transformed from a group template space
and ignoring deformations regularly results in impossible positions on the individual heads.
The median of the differences between the E-field strengths induced in the cortex for the opti-
mized vs initial positions reach 12% (MST coil) and 16% (H1, H4 & H7 coils) of the maximal
E-field strength (Fig. 3C). Of note, strong differences occur also in brain regions that are im-
plicated in the therapeutic effects, such as the prefrontal cortex for the H1 and H4 coils. Look-
ing at the 90-percentile (Suppl. Fig. S9) reveals differences exceeding 20% for all coils in 110
of the 1100 head models. Overall, these results suggest that distance optimization is important
to reach plausible coil configurations and that this has clear effects on the E-field distribution
on the cortex.
Comparison of the employed optimization method (DIRECT_L&L-BFGS-B) with a more ex-
tensive optimization (DIRECT with stricter convergence settings, followed by L-BFGS-B) re-
veals mostly equal performance, with the differences in the distance cost function being cen-
tered around zero (Suppl. Fig. S10A). The more extensive optimization reaches cost function
values that are better by 10% or more only in a few outlier cases. On the other hand, DI-
RECT_L&L-BFGS-B required on average far less function evaluations compared to DI-
RECT&L-BFGS-B (e.g., 2137 vs 7816 for the H1 coil; see Suppl. Table S1 for all results). The
total time for the distance optimizations on the desktop computer stayed below 3 minutes for
all four coil models and the memory requirements stayed below 4 GB (DIRECT_L&L-BFGS-
B and ernie head model; Suppl. Table S3).
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
12/24
Electric field optimization
The median electric field distributions and their IQR maps for the MST-Twin coil after opti-
mization are shown in Figure 4A&B. Table 1 lists the corresponding peak electric fields and
focality results. Figures 4E&F show that the optimization led to an increase of the average E-
field strength in the bilateral prefrontal ROI from ~0.72 V/m (median across subjects) to ~0.75
V/m while robustly avoiding coil-scalp intersections. The median of the differences between
the E-field strengths in the cortex reaches 14% of the peak E-field strength (Fig. 4C). The
medians between E-field strength from the position optimization and electric field optimiza-
tions differ with up to ±0.1 V/m across subjects (Fig. 4D).
Comparison of the DIRECT_L&L-BFGS-B optimization with the more extensive DI-
RECT&L-BFGS-B optimization again reveals similar performance, with the differences in the
achieved costs centered around zero and only a few cases where the DIRECT&L-BFGS-B
optimization is better by 10% or more (Suppl. Fig. S10B). However, DIRECT_L&L-BFGS-B
required on average substantially less function evaluations (2508 vs 4921; Suppl. Table S1).
The total time for the E-field optimization with the ernie head model was around 30 minutes
on the desktop computer and the memory requirements were below 8 GB (Suppl. Table S3).
The E-field strength induced by a Magventure Cool-B65 coil in a left precentral gyrus ROI
covering the primary motor hand area (handknob) was additionally optimized using only the
local L-BFGS-B search and compared to the results of naïve grid searches. Figure 5A shows
the differences of the achieved E-field strength in the ROI when compared to a lower-resolution
grid search for all 1100 head models. Figure 5B shows the corresponding results compared to
a high-resolution grid search for a sub-group of 48 head models. Our optimization approach
reliably prevented coil-scalp intersections (data not shown) and generally performed slightly
better than the grid search. This can be explained by the fact that the grid search enforces a
tangential coil orientation relative to the skin surface under the coil center while our approach
optimizes all 6 degrees of freedom of the coil position.
Using only local L-BFGS-B search was generally sufficient to reach stable optimization results
for the Cool-B65 coil, which did on average not improve further for more extensive search
strategies (Suppl. Fig. S10C). Using the most extensive search with DIRECT&L-BFGS-B
seemed to slightly reduce the already very low number of outliers where grid search was better
by 1% or more. Optimization with L-BFGS-B and automated initial position required on aver-
age 423 function evaluations compared to the low- and high- resolution grid searches with 585
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
13/24
and 11285 evaluations (Suppl. Table S2). The total time for the E-field optimization for the
Cool-B65 coil and the ernie head model stayed below 8 minutes on the desktop computer and
the memory requirements were below 8.5 GB (Suppl. Table S3). Overall, the results for the
Magventure Cool-B65 coil suggest that the new optimization approach can serve as a general
approach for all coil types including standard flat and rigid coils.
Discussion
Summary of findings
We introduced the first method to automatically optimize the shape of deformable TMS coils
and validated its performance on a large dataset of 1100 head models and four new models of
large TMS coils that are used in approved therapies or clinical trials. We demonstrated that it
robustly avoided coil-scalp intersections while achieving coil configurations that fitted closely
to the head surface or maximized the E-field strength in a target ROI. The reported group me-
dian and IQR maps for H1, H4, H7 and MST coils will be provided as online resources.
In contrast, we showed that the use of coil positions that were determined via transformations
from a group template space and did not include deformations regularly resulted in physically
impossible positions on the individual heads. This resulted in notable differences in the E-field
distributions between the two cases that also occurred in potential therapeutically relevant brain
areas. The number of function evaluations and total runtime on a normal desktop computer
were low enough to allow for standard practical use.
We additionally compared our approach to a previously published grid search approach, with
the aim to maximize the E-field strength induced by a flat and rigid coil (Magventure Cool-
B65) in a small target ROI. Our approach performed at least as good as the grid search and
required fewer function evaluations on average. This suggests that the new optimization
method can serve as a general-purpose approach for all coil types.
Comparison to published work
We compared the performance of our method to standard grid search. Alternative optimization
approaches are the auxiliary dipole method (ADM, [29]) that can maximize the average E-field
in a ROI and its recent extension [30] which determines the full E-field distribution in the ROI.
In principle, the FEM-based E-field calculations in our approach could be replaced by these
methods. As both methods use precalculations, it would depend on the number of function
evaluations whether this would result in a speedup of the overall time required for optimization.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
14/24
More importantly, the methods are restricted to E-field calculations and do not allow for coil-
scalp intersection testing or the handling of deformable coils.
The electric field distributions of the Brainsway H1, H4&H7 coils and the MagVenture MST
twin coils were characterized by a few prior publications [27,31–33] using a variety of ap-
proaches such as measurements in simplified saline-filled phantoms [33], simulations using a
spherical multi-layer model to coarsely mimic the head anatomy [27] or a low number of ana-
tomically detailed head models [31,32]. The coils were modelled based on geometric infor-
mation provided by the manufacturers and fitted manually to the skin surfaces. Our study com-
plements these findings by ensuring the accuracy of the coil models, systematically optimizing
the fit of the coils on the skin surfaces and assessing the E-field characteristics in a large sample
of anatomically detailed head models. Differences in the applied metrics make a direct com-
parison with these prior findings difficult. However, our focality and depth estimates for the
H1 coil were found to be comparable to a previous approximation based on a spherical head
model [27]. Another study using two anatomically detailed head models reports notably larger
depth estimates, but also included the electric field in white matter in the estimations [32].
Limitations
To the best of our knowledge, the presented method is the so far only automatic approach to
realistically place deformable coil models and to systematically avoid coil-head intersections.
Our approach well approximated the true shape of an H1 coil put on a ball, which led to stronger
coil deformations than when placing it on heads in our pilot tests. The H4 generally deformed
less in our tests. The H7 and MST-Twin coils consist of two sub-coils that do not or hardly
deform but only change their relative positions, which could be accurately modelled by linear
transformations. We thus suggest that the tested H1 deformations serve as a reasonable worst-
case scenario and are confident that our modeling approach captures practically feasible coil
configurations with good accuracy. However, coil deformations in practice also depend on, e.g.
how strongly the straps of the H-coils are tightened, so that the exact correspondence between
the modelled and real coil configurations would need additional experimental assessment.
Our approach allows for flexibility in terms of the optimization methods utilized and their pa-
rameter settings. The results in the main paper are based on settings that balance accuracy and
computational efficiency. More exhaustive search strategies gave little additional gain, except
for slightly reducing the already very low numbers of outlier cases in which the balanced
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
15/24
settings performed worse to a notable extent. However, we recommend choosing the optimi-
zation settings and by that the tradeoff between runtime and avoidance of putative outliers with
the specific research question in mind.
The reported electric field distributions might deviate from those achieved in clinical practice
due to differences in coil positioning strategies. We place the coil centers at average positions
defined in MNI space that might lead to deviations compared to positions that are defined using
features of the individual brain or head anatomy. Specifically, for the H1 coil, we chose its
center to be above area BA46 in MNI space, while it is placed relative to the motor hot spot of
the finger muscles in practice.
Future objectives
We presented a new framework for the simulation of large and deformable TMS coils that
systematically prevents coil-scalp intersections. We tested it with accurate computational mod-
els of four therapeutically employed coils and a dataset of 1100 head models. We suggest that
this will facilitate evaluations of the dose–response relationships for these coils, including es-
timations of interindividual differences of the induced E-fields and systematic comparisons of
induced E-fields across different coil models. Additionally, our approach can be used for the
principled optimization of coil positions and configurations to maximize the stimulation of a
target brain area. The method and coil models will be provided as part of SimNIBS together
with comprehensive example scripts demonstrating their usage. Moreover, examples for creat-
ing custom computational models of rigid and deformable coils will be included.
Acknowledgements
We thank BrainsWay Ltd (Jerusalem, Israel) for providing physical samples of the H1, H4 and
H7 coils and MagVenture A/S (Farum, Denmark) for providing a sample of the MST-Twin
coil.
Funding
AT was supported by Innovation Fund Denmark (Grand Solutions grant 9068-00025B “Preci-
sion-BCT”) and the Lundbeck Foundation (grants R313-2019-622 and R244-2017-196). TW
and AT received support by the National Institute of Health (grant R01MH128422).
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
16/24
Literature
[1] Cohen SL, Bikson M, Badran BW, George MS. A visual and narrative timeline of US FDA mile-
stones for Transcranial Magnetic Stimulation (TMS) devices. Brain Stimulat 2022;15:73–5.
https://doi.org/10.1016/j.brs.2021.11.010.
[2] Harmelech T, Hanlon CA, Tendler A. Transcranial Magnetic Stimulation as a Tool to Promote
Smoking Cessation and Decrease Drug and Alcohol Use. Brain Sci 2023;13:1072.
https://doi.org/10.3390/brainsci13071072.
[3] Levkovitz Y, Isserles M, Padberg F, Lisanby SH, Bystritsky A, Xia G, et al. Efficacy and safety of
deep transcranial magnetic stimulation for major depression: a prospective multicenter ran-
domized controlled trial. World Psychiatry 2015;14:64–73. https://doi.org/10.1002/wps.20199.
[4] Harmelech T, Roth Y, Tendler A. Deep TMS H7 Coil: Features, Applications & Future. Expert Rev
Med Devices 2021;18:1133–44. https://doi.org/10.1080/17434440.2021.2013803.
[5] Thielscher A, Opitz A, Windhoff M. Impact of the gyral geometry on the electric field induced
by transcranial magnetic stimulation. NeuroImage 2011;54:234–43.
https://doi.org/10.1016/j.neuroimage.2010.07.061.
[6] Opitz A, Windhoff M, Heidemann RM, Turner R, Thielscher A. How the brain tissue shapes the
electric field induced by transcranial magnetic stimulation. NeuroImage 2011;58:849–59.
https://doi.org/10.1016/j.neuroimage.2011.06.069.
[7] Thielscher A, Antunes A, Saturnino GB. Field modeling for transcranial magnetic stimulation: A
useful tool to understand the physiological effects of TMS? 2015 37th Annu. Int. Conf. IEEE
Eng. Med. Biol. Soc. EMBC, 2015, p. 222–5. https://doi.org/10.1109/EMBC.2015.7318340.
[8] Drakaki M, Mathiesen C, Siebner HR, Madsen K, Thielscher A. Database of 25 validated coil
models for electric field simulations for TMS. Brain Stimul Basic Transl Clin Res Neuromodula-
tion 2022;15:697–706. https://doi.org/10.1016/j.brs.2022.04.017.
[9] Deng Z-D, Luber B, McClintock SM, Weiner RD, Husain MM, Lisanby SH. Clinical Outcomes of
Magnetic Seizure Therapy vs Electroconvulsive Therapy for Major Depressive Episode: A Ran-
domized Clinical Trial. JAMA Psychiatry 2024;81:240. https://doi.org/10.1001/jamapsychia-
try.2023.4599.
[10] Strang G, Herman E “Jed,” Strang G, Herman E “Jed.” 6.2 Line Integrals - Calculus Volume 3 |
OpenStax 2016. https://openstax.org/books/calculus-volume-3/pages/6-2-line-integrals (ac-
cessed September 20, 2024).
[11] Cheng H, Greengard L, Rokhlin V. A Fast Adaptive Multipole Algorithm in Three Dimensions. J
Comput Phys 1999;155:468–98. https://doi.org/10.1006/jcph.1999.6355.
[12] Jones, D. R., Perttunen, C. D., Stuckman, B. E. Lipschitzian optimization without the Lipschitz
constant. J Optim Theory Appl 1993;79:157–81.
[13] Byrd RH, Lu P, Zhu C. A Limited Memory Algorithm for Bound Constrained Optimization. SIAM J
Sci Comput 1995;16.
[14] Weise K, Numssen O, Thielscher A, Hartwigsen G, Kn TR. A novel approach to localize cortical
TMS effects. NeuroImage 2020;209:116486. https://doi.org/10.1016/j.neu-
roimage.2019.116486.
[15] Fox MD, Buckner RL, White MP, Greicius MD, Pascual-Leone A. Efficacy of Transcranial Mag-
netic Stimulation Targets for Depression Is Related to Intrinsic Functional Connectivity with the
Subgenual Cingulate. Biol Psychiatry 2012;72:595–603. https://doi.org/10.1016/j.bi-
opsych.2012.04.028.
[16] Matilainen N, Kataja J, Laakso I. Predicting the hotspot location and motor threshold prior to
transcranial magnetic stimulation using electric field modelling. Phys Med Biol
2023;69:015012. https://doi.org/10.1088/1361-6560/ad0219.
[17] Tang VM, Blumberger DM, Weissman CR, Dimitrova J, Throop A, McClintock SM, et al. A pilot
study of magnetic seizure therapy for treatment-resistant obsessive–compulsive disorder. De-
press Anxiety 2021;38:161–71. https://doi.org/10.1002/da.23097.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
17/24
[18] Smith SE, Kosik EL, van Engen Q, Kohn J, Hill AT, Zomorrodi R, et al. Magnetic seizure therapy
and electroconvulsive therapy increase aperiodic activity. medRxiv 2023:2023.01.11.23284450.
https://doi.org/10.1101/2023.01.11.23284450.
[19] Cohen-Or D, Kaufman A. Fundamentals of Surface Voxelization. Graph Models Image Process
1995;57:453–61. https://doi.org/10.1006/gmip.1995.1039.
[20] Virtanen P, Gommers R, Oliphant TE, Haberland M, Reddy T, Cournapeau D, et al. SciPy 1.0:
fundamental algorithms for scientific computing in Python. Nat Methods 2020;17:261–72.
https://doi.org/10.1038/s41592-019-0686-2.
[21] Cao, Fang, Madsen, Kristoffer Hougaard, Worbs, Torge, Puonti, Oula, Siebner, Hartwig Roman,
Schmitgen, Arno, Kunz, Patrik, et al. FEM-based Electric Field Calculations for Neuronavigated
Transcranial Magnetic Stimulation. Biorxiv 10.1101/2024.12.13.628139
[22] Team TMD. Matplotlib: Visualization with Python 2024. https://doi.org/10.5281/ze-
nodo.12652732.
[23] Sullivan CB, Kaszynski AA. PyVista: 3D plotting and mesh analysis through a streamlined inter-
face for the Visualization Toolkit (VTK). J Open Source Softw 2019;4:1450.
https://doi.org/10.21105/joss.01450.
[24] Van Essen DC, Ugurbil K, Auerbach E, Barch D, Behrens TEJ, Bucholz R, et al. The Human Con-
nectome Project: a data acquisition perspective. NeuroImage 2012;62:2222–31.
https://doi.org/10.1016/j.neuroimage.2012.02.018.
[25] Puonti O, Van Leemput K, Saturnino GB, Siebner HR, Madsen KH, Thielscher A. Accurate and
robust whole-head segmentation from magnetic resonance images for individualized head
modeling. NeuroImage 2020;219:117044. https://doi.org/10.1016/j.neuroimage.2020.117044.
[26] Dale AM, Fischl B, Sereno MI. Cortical surface-based analysis - I. Segmentation and surface re-
construction. Neuroimage 1999;9:179–94. https://doi.org/DOI 10.1006/nimg.1998.0395.
[27] Deng ZD, Lisanby SH, Peterchev AV. Electric field depth-focality tradeoff in transcranial mag-
netic stimulation: Simulation comparison of 50 coil designs. Brain Stimulat 2013;6:1–13.
https://doi.org/DOI 10.1016/j.brs.2012.02.005.
[28] Antonenko D, Grittner U, Saturnino G, Nierhaus T, Thielscher A, Flöel A. Inter-individual and
age-dependent variability in simulated electric fields induced by conventional transcranial elec-
trical stimulation. NeuroImage 2021;224:117413. https://doi.org/10.1016/j.neu-
roimage.2020.117413.
[29] Gomez LJ, Dannhauer M, Peterchev AV. Fast computational optimization of TMS coil place-
ment for individualized electric field targeting. NeuroImage 2021;228:117696.
https://doi.org/10.1016/j.neuroimage.2020.117696.
[30] Wang D, Hasan NI, Dannhauer M, Yucel AC, Gomez LJ. Fast computational E-field dosimetry for
transcranial magnetic stimulation using adaptive cross approximation and auxiliary dipole
method (ACA-ADM). NeuroImage 2023;267:119850. https://doi.org/10.1016/j.neu-
roimage.2022.119850.
[31] Tzirini M, Roth Y, Harmelech T, Zibman S, Pell GS, Kimiskidis V, et al. Electrical field measure-
ments and simulations of the H7 and D-B80 coils: Non-equivalence of the TMS coils for obses-
sive compulsive disorder. Brain Stimul Basic Transl Clin Res Neuromodulation 2021;14:1525–7.
https://doi.org/10.1016/j.brs.2021.10.382.
[32] Parazzini M, Fiocchi S, Chiaramello E, Roth Y, Zangen A, Ravazzani P. Electric field estimation of
deep transcranial magnetic stimulation clinically used for the treatment of neuropsychiatric
disorders in anatomical head models. Med Eng Phys 2017;43:30–8.
https://doi.org/10.1016/j.medengphy.2017.02.003.
[33] Tendler A, Barnea Ygael N, Roth Y, Zangen A. Deep transcranial magnetic stimulation (dTMS) –
beyond depression. Expert Rev Med Devices 2016;13:987–1000.
https://doi.org/10.1080/17434440.2016.1233812.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
18/24
[34] Fischl B, Sereno MI, Tootell RBH, Dale a M. High-resolution inter-subject averaging and a sur-
face-based coordinate system. Hum Brain Mapp 1999;8:272–84.
https://doi.org/10.1002/(SICI)1097-0193(1999)8.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
19/24
FIGURES
Figure 1: A) Initial placements on the MNI head for the distance optimization of the Brainsway H1,
H4 and H7 coils, and MagVenture MST-Twin coil. B) Initial coil placements on the MNI head for
optimization of the average electric field magnitude (abbr.: mean(|E|) within a bilateral prefrontal ROI
(shown in red) for MagVenture MST-Twin coil. C) Left precentral ROI covering the handknob for use
with the MagVenture Cool-B65 coil. The masks are defined on the "fsaverage" cortical surface and
transformed to the subject space via surface-based registration [34].
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
20/24
Figure 2: E-fields of the H1, H4, H7 and MST-Twin coils after coil distance optimizations. A) Median
across N=1100 head models of the induced electric field strength. B) IQR across the 1100 head models.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
21/24
Figure 3: A) Coil-scalp distances before and after optimization. The distance is calculated as the mean
distance between the pre-defined sets of positions on the coil casings and their nearest scalp position
for each subject. B) Maximally occurring intersections of the coil with the head before and after distance
optimization. Signed distances are reported, with negative values indicating intersections, and positive
values the minimal distance between coil casing and scalp in case of no intersections. Specifically, for
each individual optimization result, the minimal value of the signed distances between the scalp and
any of the predefined positions on the coil casing is shown. This corresponds to the deepest intersection
or, in case no intersection occurred, positive values indicate the minimal distance between coil casing
and scalp. C) Median of the relative differences between the electric field strengths for the optimized
and initial coil settings.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
22/24
Figure 4: Results of the optimization of the mean electric field strength in the bilateral prefrontal ROI
for the MST-Twin coil. A) Median across N=1100 head models of the electric field strength after opti-
mization. B) IQR across 1100 head models. C) Median of the relative differences between the electric
field strengths for the optimized and initial coil settings. D) Average difference in electric field strength
between electric field strength-optimized and position-optimized MST-Twin coils (Figure 4A and 3A)
E) Mean electric field strengths in the ROI before and after optimization. F) Maximal intersections of
the coil with the head before and after optimization. Signed distances are reported, with negative values
indicating the maximal depth of the intersection, and positive values the minimal distance between coil
casing and scalp in case of no intersections.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
23/24
Figure 5: Results of the optimization of the mean electric field strength in the handknob ROI for the
MagVenture Cool-B65 coil. The differences of the mean electric field magnitudes in the ROI obtained
by our optimization approach versus a grid search are shown. A) Comparison to results of a “coarse”
grid search and N=1100 subjects. B) Comparison to results of a higher-resolution grid search (N=48).
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
24/24
Tables
Distance optimization E-field optimization
H1
DIRECT-L+
L-BFGS-B
H4
DIRECT-L+
L-BFGS-B
H7
DIRECT-L+
L-BFGS-B
MST-Twin
DIRECT-L+
L-BFGS-B
MST-Twin
DIRECT-L+
L-BFGS-B
Cool-B65
L-BFGS-B
Peak |E| (V/m) 1.9 ± 0.1 2.2 ± 0.2 1.8 ± 0.2 2.1 ± 0.2 2.1 ± 0.2 1.14 ± 0.2
Focality (cm3) 108.0 ± 18.1 89.0 ± 16.6 37.8 ± 9.0 41.5 ± 8.3 43.1 ± 8.1 12.8 ± 3.8
Depth (mm) 22.6 ± 2.1 21.5 ± 1.9 22.9 ± 2.4 21.5 ± 2.0 21.3 ± 2.0 16.7 ± 1.6
Table 1. Fields summary of the optimization results for distance optimization (H1, H4, H7,
MST-Twin) and electric field optimization (MST-Twin and Cool-B65). The values are shown
as the medians ± the interquartile range across the subject pool.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
1/18
Personalized electric field simulations of deformable
large TMS coils based on automatic position and
shape optimization
Torge Worbs, Bianka Rumi, Kristoffer H. Madsen, Axel Thielscher
Supplementary Material A: Coil implementation
based on CTs
Physical samples of the Brainsway coils (H1, H4, and H7) were provided by the manufacturer.
These coils underwent CT scanning on a Siemens SOMATOM Definition Flash using a se-
quence with 0.5x0.5x3 mm resolution, 140 keV, 600 mAs and metal artifact reduction. The
resulting CT scan data was converted into NIfTI format for further processing using custom
software written in Python 3.8. The origins and axes directions of the NIfTI images were set
manually according to the information about the coil centers provided by the manufacturer.
For each coil, a binary mask was generated by setting an appropriate intensity threshold to
isolate the metal wires and transformed into triangle surfaces using a marching cube algorithm.
The resulting surfaces represented the circumferences of the coil wires, and the central paths
of the wires were obtained by shrinking the surfaces along their local normal directions. These
central wire paths were manually tracked, yielding precise wire path coordinates. Subsequently,
these wire path coordinates were interpolated using b-splines to obtain a desired resolution (2
elements/mm). The coil center position and coil coordinate system were defined based on man-
ually identified coil landmarks in the CTs (mounting screws, brackets on the coil, see Fig. S4
below). The interpolated and aligned data points were transformed into deformable coil models,
in line with the coil generation examples included in SimNIBS 4.5. The casing thickness to
account for the fabric cap and paddings and the coil deformations were defined as outlined in
the main paper.
To estimate the induced error from representing the coil wire by a spatially discretized line
current along the central wire path, two convergence tests were performed with a round coil
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
2/18
(radius of 50 mm; wire diameter of 5 mm) placed above electrode position Cz of the standard
SimNIBS ernie head model (Suppl. Figure S1). The errors of the electric field magnitude and
the electric field vectors were calculated in relation to reference cases. The first test assessed
the error due to modeling the wire by a single line current along the central wire path. For that,
the wire was modelled by an increasing number of line currents placed in spherical layers (Fig.
S1A). The resolution along the wire paths was kept constant at 3.2 elements/mm. The reference
consisted of 8 layers with 169 lines. Using one central line already resulted in a very low error
that is practically not relevant (≈0.038% error).
The second test assessed the convergence when increasing the resolution along the wire path,
using 64 elements per mm as reference. Two elements per mm resulted in an error of 0.001%
and were selected when modelling the wire paths of the H coils.
Supplementary Figure S1. Convergence behavior of the error for the line approximation of a
circular coil. A) Layer approximation pattern for 1 layer (top), 2 layers (middle) and 5 layers
(bottom). B) Error convergence for an increasing number of layers using the relative root mean
square error (RRMSE). Number of lines representing each layer: 1: 1, 2: 7, 3: 19, 4: 37, 5: 61,
6: 91, 7: 127. C) Error convergence for an increasing number of elements per mm.
Supplementary Material B: Validation of defor-
mation modeling
We approximate coil deformations by concatenated affine transformations, resulting in a low
number of free parameters that can be efficiently optimized. To validate this approach, the
physical BrainsWay H1 coil was CT-scanned in a deformed state positioned on a football with
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
3/18
a 200 mm diameter, and a ground truth model of the deformed coil was generated from the CT
scan. The standard H1 coil model was aligned and deformed to match the ground truth model
as closely as possible by minimizing the distance between the two coil wires using the DIRECT
optimizer implemented in scipy. The electric fields were then simulated on a 200 mm diameter
sphere model, corresponding to a version of the simnibs m2m_sphere model scaled to match
the size of the football with a diameter of 200 mm. The sphere model had 5 tissue types mim-
icking the conductivity of human tissues (scalp: 200 mm, 0.465 S/m; bone: 187 mm, 0.01 S/m;
CSF: 175 mm, 1.654 S/m; GM 168 mm, 0.275 S/m; and WM 157 mm, 0.126 S/m). The pair-
wise differences in the electric field magnitudes on the GM surface nodes between the ground-
truth H1 model and the deformed and non-deformed H1 models were assessed. The error in
the electric field magnitudes between the optimized deformed coil model and ground-truth coil
model was reduced to be <2.5%, demonstrating that the approximated deformations repre-
sented the true deformed coil shape reasonably well. Supplementary Figure S2 illustrates the
differences in coil wires between the initial, ground truth and optimized coil settings (Figure
S2A and S2B), and the electric field differences between the coil settings (Figure S2C&D).
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
4/18
Supplementary Figure S2. Validation of the deformable computational model of the H1 coil.
A) Initial undeformed wire paths (red) and the ground truth (gray) derived from a CT scan in
its deformed state on the ball. B) Optimized wire paths (green) and the ground truth (gray). C)
Relative difference between the initial coil model and the ground truth. D) Relative difference
between the coil model after position and deformation optimization and the ground truth. E)
Electric field of the ground truth. F) Average relative differences between the electric fields of
the initial undeformed coil model and the ground truth model, as well as of the optimized coil
model and the ground truth model.
Supplementary Material C: Technical notes on the
TMS Coil Definition File Format (.tcd)
The TMS Coil Definition File Format (.tcd) is a JSON-based format that defines a single TMS
coil. The coil file can be loaded into SimNIBS and used to calculate the magnetic vector po-
tential (A-field) for a given rate of change of the coil current and the magnetic field (B-field)
for a given coil current. The following text gives an overview of the format. Details are pro-
vided in the SimNIBS documentation as well as in the example scripts for custom coil creation.
Supplementary Figure S3. The TMS Coil Definition File Format. Each sub-list (green) can
have an arbitrary number of items. A) Each coil can have a general coil casing (or “coilModel”)
not associated with any coil element. B) Each coil element can have a casing. C) Each coil
element must be connected to one stimulator. D) Each coil element can have an arbitrary num-
ber of deformations including no deformations. E) Each deformation uses a deformation range
that defines the valid values of the deformation. F) Each coil can have an arbitrary number of
self-intersection groups.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
5/18
The file format holds basic information about the coil, the name, the brand, and the version of
the coil file. Its main part is a list of stimulating coil elements (“coilElementList”) that is linked
to a list of deformations ("deformList") and a list of geometric models of the sub-casing
(“coilModels”). These three components allow for definitions of static coils and coils with one
or multiple movable stimulating elements. The following list gives a brief overview of the main
components:
coilElementList
Each stimulating coil element can use one of three types of stimulating elements:
o Magnetic dipoles, which are described by their location (in mm), direction, and
magnitude (in Am2).
o Line current elements, which are defined by their starting position (in mm), direc-
tion, and magnitude (in mm).
o Magnetic vector potential field (A-field), sampled on a grid, described by the A-
field value at each grid location (in Tm) and an affine transformation matrix (in
mm).
Each stimulating coil element has a name, is connected to one stimulator (from “stim-
ulatorList”) (arrow “C” in Suppl. Figure S3), has a casing (from “coilModels”; Suppl.
Figure S3 arrow B), and is associated with a list of deformations (from “deformList”)
that are applied to this coil element (Suppl. Figure S3 arrow D). The points, values and
data lists are stored as plain text or as Base64 encoded and compressed binary. The
deformations are applied in the order that they are stored in. Multiple stimulating coil
elements can share the same deformation, enabling the common rotation of several coil
elements around shared axes.
stimulatorList
A list of TMS stimulators is stored to connect different stimulating coil elements to the
same stimulator, which can be used in multi-stimulator coil settings. The TMS stimu-
lators are described by their name, brand, maximum dI/dt (in A/s), and a list of wave-
forms.
deformList
A deformation is either defined as a translation in the x, y, or z direction (in mm) or as
a rotation around an axis (in degrees) defined by two points (in mm). Its value is defined
and bound by a deformation range (Suppl. Figure S3 arrow E).
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
6/18
deformRangeList
The limits for deformations are defined as deformation ranges. Multiple deformations,
for example, rotations around different axes, can share the same deformation range.
These ranges are defined by a minimum and maximum value, and an initial value be-
tween the minimum and the maximum.
coilModels
For visualization and optimization purposes, the file format contains a list of coil casing
triangulated surfaces. Each coil casing surface has a point list (in mm) and a list of faces
which are indices into the point list. Additionally, a list of points ("minDistancePoints"
in mm) representing the parts of the coil that are supposed to be close to the head can
be stored. These are then used for evaluation of the coil-skin distance costs during op-
timization (Supplementary Figure S4 shows the “minDistancePoints” for the 4 imple-
mented coils as examples). The point and faces lists are stored as plain text or as Base64
encoded and compressed binary.
selfIntersectionTest
A list of groups of stimulating coil element indices which will be tested for self-inter-
sections in cases of optimization.
The file format also gives the possibility to store a global coil casing ("coilCasing") that is not
associated with a specific stimulating coil element (Suppl. Figure S3 arrow A). In addition, to
speed up repeated FEM evaluations for varying coil positions and deformations, the A-fields
of the stimulating coil elements can be sampled on 3D grids during preparation. They can then
be computationally efficient evaluated for varying coil configurations using simple linear trans-
formations. The file format supports this by giving the option to store information about the
axis limits ("limits" in mm) and resolution ("resolution" in mm) of the 3D grids.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
7/18
Supplementary Figure S4. The red minDistancePointsare used to calculate the 𝑑𝑖𝑠𝑡𝑎𝑛𝑐𝑒
cost during optimization and the points defining the bending axes (blue). The coil wires (H-
coils) or coil casing (MST-Twin coil) are shown in gray. The closest distance to the head is
determined for each point, and the average then used as 𝑑𝑖𝑠𝑡𝑎𝑛𝑐𝑒 cost (d . A) BrainsWay H1
coil. B) BrainsWay H4 coil. C) BrainsWay H7 coil. D) MagVenture MST-Twin coil.
Supplementary Material D: Evaluations of the differ-
ent parts of the cost functions
Distance Cost:
The cost for the distance optimization is the mean distance of pre-selected points on the coil
casing or wires (Suppl. Figure S4) to the scalp surface. The points are selected to sample the
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
8/18
parts of the coil casing that are facing towards and are close to the scalp. To efficiently calculate
the distance, a binary voxel mask of the head is created (resolution 1x1x1 mm³). Using the
Euclidean distance transform, the mask is converted into a voxel image where the voxel values
describe the signed distance from the scalp surface to the inside and outside of the scalp. The
distance at the pre-selected points is cubically interpolated based on the distance voxel image
in combination with linear extrapolation.
E-Field Magnitude Cost:
The E-field magnitude cost is the negative of the average E-field strength in a Region of Interest
(ROI). To efficiently calculate the E-field magnitude in the ROI, the A-field of each stimulating
coil element is sampled on a 3D grid during preparation. The sampled A-field is then used to
interpolate the A-field in the head model and update the right hand side of the FEM is to sim-
ulate the resulting E-field in the ROI.
Intersection Penalty:
The intersection penalty is the depth-weighted volume intersection between the casings of the
coil elements and the head mask. To efficiently calculate the intersection penalty, the scalp
surface and the surfaces of the coil element casings are used to create binary voxel masks (res-
olution 1x1x1 mm³) of the head and casings, respectively. Using the Euclidean distance trans-
form, the head mask is converted into a voxel distance image where outside voxels have a value
of 0 and the value of inside voxels represent their distance to the scalp surface. In dependence
on the tested coil configuration (i.e., the position relative to the head and the deformations), the
transformed coordinates of all voxels of a casing are used to cubically interpolate the depth
values from the scalp distance image. This is repeated for all coil elements and the sum of all
interpolated depth values is reported as the depth-weighted volume intersection.
Self-Intersection Penalty:
The self-intersection penalty is the volume intersection between selected coil element casings.
To efficiently calculate the self-intersection penalty, the surfaces of the coil element casings
are used to pre-calculate voxel mask images (resolution 1x1x1 mm³). The coordinates of one
coil casing volume are transformed into the space of the other element casing in dependence
on the tested coil configuration, and are used to cubically interpolate the volume in another coil
casing volume. The sum of these interpolated volume intersections is then used as the volume
self-intersection in mm³.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
9/18
Supplementary Material E: Additional Figures and
Tables
Supplementary Figure S5
Supplementary Figure S5. BrainsWay H1 coil on the MNI head. A) Initial coil position. The
white sphere indicates the coil center, placed above the center of BA46 (dark red sphere). The
visualized target positions were taken from [1,2]. B) Electric field and coil position after dis-
tance optimization to resolve intersections of the coil with the head. The coil center was posi-
tioned beneath the intersection of the wires running left-to-right and front-to-back. Pilot tests
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
10/18
on the MNI head confirmed that the field strength in BA46 decreased when moving the coil
center to a more medial position, suggesting that its choice was reasonable. To ensure con-
sistent orientations of the coil axes, they were aligned with the top plate of the coil and the
screw threads in the plate (both are not shown here), which are used to secure the coil in the
helmet, based on the CT image.
Supplementary Figure S6
Supplementary Figure S6. BrainsWay H4 coil on the MNI head. A) Initial coil position. B)
Electric field and coil position after distance optimization to resolve intersections of the coil
with the head. The white sphere indicates the intended coil center position at MNI coordinate
[0, 53, 63] before optimization. The coil axes were aligned with the top plate of the coil.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
11/18
Supplementary Figure S7
Supplementary Figure S7. BrainsWay H7 coil on the MNI head. A) Initial coil position. B)
Electric field and coil position after distance optimization to resolve intersections of the coil
with the head. The white sphere indicates the intended coil center position at MNI coordinate
[-5, 20, 87] before optimization. The coil axes were aligned with the top plate of the coil.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
12/18
Supplementary Figure S8
Supplementary Figure S8. MagVenture MST-Twin coil on the MNI head. A) Initial coil
position. B) Electric field and coil position after distance optimization to resolve intersections
of the coil with the head. The white sphere indicates the intended coil center position at elec-
trode position Fz before optimization. The coil axes were aligned with the guide rail.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
13/18
Supplementary Figure S9
Supplementary Figure S9. 90%iles of the differences between the E-field strength for the
initial and optimized coil configurations.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
14/18
Supplementary Figure S10
Supplementary Figure S10. Relative differences between different optimization settings for
A) distance optimization, B) E-field optimization of the MST-Twin coil and C) E-field optimi-
zation of the Cool-B65 figure-8 coil.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
15/18
Supplementary Table S1
Distance
optimization
H1
DIRECT +
L-BFGS-B
H1
DIRECT-L+
L-BFGS-B
H4
DIRECT +
L-BFGS-B
H7
DIRECT +
L-BFGS-B
H7
DIRECT-L +
L-BFGS-B
N mean ±
STD
7816 ± 3202
2137 ± 420
6544 ± 2517
5872 ± 1604
1423 ± 295
N min
2957
985
2436
2095
688
N max
19495
3447
16804
12327
2990
Distance
optimization
MST-Twin
DIRECT +
L-BFGS-B
MST-Twin
DIRECT-L +
L-BFGS-B
E-field
optimization
MST-Twin
DIRECT +
L-BFGS-B
MST-Twin
DIRECT-L +
L-BFGS-B
N mean ±
STD
7213 ±
2263
2269 ± 662
N mean ±
STD
4921 ± 917
2508 ± 722
N min
3076
980
N min
2056
808
N max
16888
4728
N max
8341
5299
Supplementary Table S1. Number of cost function evaluations for the distance and E-field
optimization with more exhaustive (DIRECT + L-BFGS-B) and standard (DIRECT-L + L-BFGS-
B) search. The results for the settings used in the main paper are marked in blue.
Supplementary Table S2
L-BFGS-B
(auto-init)
L-BFGS-B
DIRECT +
L-BFGS-B
DIRECT-L+
L-BFGS-B
DIRECT-L+
L-BFGS-B
(auto-init)
Grid search
High-res
grid search
N mean ±
STD
423± 99
489 ± 119
4750 ±
1207
940 ± 169
957 ± 151
585 ± 0
11285 ± 0
N min
210
238
2023
685
719
-
-
N max
854
1015
6734
2543
2213
-
-
Supplementary Table S2. Number of cost function evaluations for the E-field optimization of
the Cool-B65 coil. The results for the settings used in the main paper are marked in blue.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
16/18
Supplementary Table S3
Distance optimization
E-field optimization
H1
DIRECT-L+
L-BFGS-B
H4
DIRECT-L+
L-BFGS-B
H7
DIRECT-L+
L-BFGS-B
MST-Twin
DIRECT-L+
L-BFGS-B
MST-Twin
DIRECT-L+
L-BFGS-B
Cool-B65
L-BFGS-B
Time (optimization)
in sec
68
59
34
43
1886
338
Time (overhead)
in sec
83
81
66
64
116
134
Total time
in sec
151
140
100
107
2002
472
Number of function
evaluations
1390
1704
1331
1359
3089
686
Peak memory
in GB
1.7
3.6
3.5
3.8
7.5
8.2
Supplementary Table S3. Timings, number of cost function evaluations and peak memory
consumption. Tested on a desktop computer (Ubuntu 22.04, Intel i7-11700, 16 cores, 32GB
RAM) using the standard SimNIBS ernie head model. The “overhead” covers the time for a
final standard FEM simulation with the optimized coil parameters and the creation of visuali-
zations from these simulation results in addition to pre-calculations necessary for estimation of
the cost function.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
17/18
Supplementary Table S4
Translations in
mm
[xmin xmax],
[ymin ymax],
[zmin zmax]
Rotations in
degrees
(around x, y,
z)
Deformations
(if not specified otherwise, rotations (degrees)
around custom-defined axes color-coded in
Supplementary Figure 3)
Brainsway
H1
[-20, 20],
[-20, 20],
[-30, 30]
[-30, 30],
[-10,10],
[-5, 5]
[-25, 25] Rotation of left deformable wire seg-
ments (blue)
[-25, 25] Rotation of right deformable wire seg-
ments (green)
[-25, 25] Rotation of frontal deformable wire
segments (black)
Brainsway
H4
[-20, 20],
[-20, 20],
[-30, 30]
[-30, 30],
[-10,10],
[-5, 5]
[-10, 10] Rotation of left deformable wire seg-
ments around the custom-defined axis (blue)
[-10, 10] Rotation of right deformable wire seg-
ments (green)
Brainsway
H7
[-5, 5],
[-5, 5],
[-30, 30]
[-30, 30],
[-10,10],
[-5, 5]
[-10, 10] Rotation of the left coil segment (blue)
[-10, 10] Rotation of the right coil segment
(green)
Magven-
ture
MST-Twin
Pos: ([-5, 5],
[-5, 5],
[-30, 30])
|E|: (30,30,30)
Pos([-30,
30],
[-10,10],
[-5, 5])
|E|:
(30,30,30)
Rotations around the indicated axes:
C left: [-13, 13]
B left: [-13, 13]
A left: [-10, 0]
C right: [-13, 13]
B right: [-13, 13]
A right: [-10, 0]
Magven-
ture
Cool-B65
[-30, 30],
[-30, 30],
[-30, 30]
[-30, 30],
[-30, 30],
[-95, 95]
None
Supplementary Table S4. Ranges for the feasible translations and rotations around the coil
center and the allowed transformation ranges to mimic the deformation of coil parts.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
18/18
References
[1] Fox MD, Buckner RL, White MP, Greicius MD, Pascual-Leone A. Efficacy of Tran-
scranial Magnetic Stimulation Targets for Depression Is Related to Intrinsic Functional
Connectivity with the Subgenual Cingulate. Biol Psychiatry 2012;72:595603.
https://doi.org/10.1016/j.biopsych.2012.04.028.
[2] Trapp NT, Bruss J, King Johnson M, Uitermarkt BD, Garrett L, Heinzerling A, et al.
Reliability of targeting methods in TMS for depression: Beam F3 vs. 5.5 cm. Brain
Stimulat 2020;13:57881. https://doi.org/10.1016/j.brs.2020.01.010.
.CC-BY-NC 4.0 International licenseperpetuity. It is made available under a
preprint (which was not certified by peer review) is the author/funder, who has granted bioRxiv a license to display the preprint in
The copyright holder for thisthis version posted December 28, 2024. ; https://doi.org/10.1101/2024.12.27.629331doi: bioRxiv preprint
ResearchGate has not been able to resolve any citations for this publication.
Article
Full-text available
Objective To investigate whether the motor threshold (MT) and the location of the motor hotspot in transcranial magnetic stimulation (TMS) can be predicted with computational models of the induced electric field. Approach Individualized computational models were constructed from structural magnetic resonance images of ten healthy participants, and the induced electric fields were determined with the finite element method. The models were used to optimize the location and direction of the TMS coil on the scalp to produce the largest electric field at a predetermined cortical target location. The models were also used to predict how the MT changes as the magnetic coil is moved to various locations over the scalp. To validate the model predictions, the motor evoked potentials were measured from the first dorsal interosseous (FDI) muscle with TMS in the ten participants. Both computational and experimental methods were preregistered prior to the experiments. Main results Computationally optimized hotspot locations were nearly as accurate as those obtained using manual hotspot search procedures. The mean Euclidean distance between the predicted and the measured hotspot locations was approximately 1.3 cm with a 0.8 cm bias towards the anterior direction. Exploratory analyses showed that the bias could be removed by changing the cortical target location that was used for the prediction. The results also indicated a statistically significant relationship (p < 0.001) between the calculated electric field and the MT measured at several locations on the scalp. Significance The results show that the individual TMS hotspot can be located using computational analysis without stimulating the subject or patient even once. Adapting computational modelling would save time and effort in research and clinical use of TMS.
Article
Full-text available
Repetitive transcranial magnetic stimulation (rTMS) is a noninvasive, drug-free, neural-circuit-based therapeutic tool that was recently cleared by the United States Food and Drug Associate for the treatment of smoking cessation. TMS has been investigated as a tool to reduce consumption and craving for many other substance use disorders (SUDs). This review starts with a discussion of neural networks involved in the addiction process. It then provides a framework for the therapeutic efficacy of TMS describing the role of executive control circuits, default mode, and salience circuits as putative targets for neuromodulation (via targeting the DLPFC, MPFC, cingulate, and insula bilaterally). A series of the largest studies of TMS in SUDs are listed and discussed in the context of this framework. Our review concludes with an assessment of the current state of knowledge regarding the use of rTMS as a therapeutic tool in reducing drug, alcohol, and nicotine use and identifies gaps in the literature that need to be addressed in future studies. Namely, while the presumed mechanism through which TMS exerts its effects is by modulating the functional connectivity circuits involved in executive control and salience of drug-related cues, it is also possible that TMS has direct effects on subcortical dopamine, a hypothesis that could be explored in greater detail with PET imaging.
Preprint
Full-text available
Major depressive disorder (MDD) is a leading cause of disability worldwide. One of the most efficacious treatments for treatment-resistant MDD is electroconvulsive therapy (ECT). Recently, magnetic seizure therapy (MST) was developed as an alternative to ECT due to its more favorable side effect profile. While these approaches have been very successful clinically, the neural mechanisms underlying their therapeutic effects are unknown. For example, clinical slowing of the electroencephalogram has been observed in both treatment modalities. A recent longitudinal study of a small cohort of ECT patients revealed that observed clinical slowing was better explained by increases in frontal aperiodic activity, an emerging EEG signal linked to neural inhibition. Here we investigate the role of aperiodic activity in a cohort of patients who received ECT and a cohort of patients who received MST treatment. We find that across treatments, frontal aperiodic activity better explains increases in delta band power associated with clinical slowing, compared to delta oscillations. Increased aperiodic activity is also linked to therapeutic efficacy, which is suggestive of a potential shared neural mechanism of action across ECT and MST: an increase in frontal inhibitory activity.
Article
Full-text available
Introduction Transcranial magnetic stimulation(TMS) uses magnetic pulses to induce electrical current in the underlying neuronal tissue. A variety of TMS coils exists on the market, differing primarily based on the configuration, orientation, and flexibility of the wire windings of the coil. Deep TMS utilized H-coils, which are flexible coils with different configurations aimed to stimulate different brain regions implicated in different neuropsychiatric disorders. The H7-coil, designed to target primarily the medial prefrontal cortex(mPFC) and anterior cingulate cortex(ACC), is FDA-cleared for obsessive-compulsive disorder(OCD). It was chosen as the focus of this review since it has recently shown promise in various neuropsychiatric populations in addition to a growing understanding of its mechanism of action(MOA) Areas covered Here we assembled all peer-reviewed publications on the H7-coil to showcase its efficacy in: (a) various OCD patient populations (e.g., different degrees of symptom severity, treatment resistance, comorbidities) (b) other neuropsychiatric populations (e.g., addiction, major depressive disorder(MDD), and autism spectrum disorder(ASD)). Expert opinion While substantial evidence pertaining to the H7-coil’s efficacy as well as its MOA has accumulated, much work still remains. In the final section of this review, we highlight areas of ongoing and future research that will further elucidate the coil’s MOA as well as its full efficacy potential.
Article
Full-text available
Background During transcranial magnetic stimulation (TMS) a coil placed on the scalp is used to non-invasively modulate activity of targeted brain networks via a magnetically induced electric field (E-field). Ideally, the E-field induced during TMS is concentrated on a targeted cortical region of interest (ROI). Determination of the coil position and orientation that best achieve this objective presently requires a large computational effort. Objective To improve the accuracy of TMS we have developed a fast computational auxiliary dipole method (ADM) for determining the optimum coil position and orientation. The optimum coil placement maximizes the E-field along a predetermined direction or, alternatively, the overall E-field magnitude in the targeted ROI. Furthermore, ADM can assess E-field uncertainty resulting from precision limitations of TMS coil placement protocols. Method ADM leverages the electromagnetic reciprocity principle to compute rapidly the TMS induced E-field in the ROI by using the E-field generated by a virtual constant current source residing in the ROI. The framework starts by solving for the conduction currents resulting from this ROI current source. Then, it rapidly determines the average E-field induced in the ROI for each coil position by using the conduction currents and a fast-multipole method. To further speed-up the computations, the coil is approximated using auxiliary dipoles enabling it to represent all coil orientations for a given coil position with less than 600 dipoles. Results Using ADM, the E-fields generated in an MRI-derived head model when the coil is placed at 5,900 different scalp positions and 360 coil orientations per position (over 2.1 million unique configurations) can be determined in under 15 minutes on a standard laptop computer. This enables rapid extraction of the optimum coil position and orientation as well as the E-field variation resulting from coil positioning uncertainty. ADM is implemented in SimNIBS 3.2. Conclusion ADM enables the rapid determination of coil placement that maximizes E-field delivery to a specific brain target. This method can find the optimum coil placement in under 15 minutes enabling its routine use for TMS. Furthermore, it enables the fast quantification of uncertainty in the induced E-field due to limited precision of TMS coil placement protocols, enabling minimization and statistical analysis of the E-field dose variability.
Article
Full-text available
Transcranial brain stimulation (TBS) has been established as a method for modulating and mapping the function of the human brain, and as a potential treatment tool in several brain disorders. Typically, the stimulation is applied using a one-size-fits-all approach with predetermined locations for the electrodes, in electric stimulation (TES), or the coil, in magnetic stimulation (TMS), which disregards anatomical variability between individuals. However, the induced electric field distribution in the head largely depends on anatomical features implying the need for individually tailored stimulation protocols for focal dosing. This requires detailed models of the individual head anatomy, combined with electric field simulations, to find an optimal stimulation protocol for a given cortical target. Considering the anatomical and functional complexity of different brain disorders and pathologies, it is crucial to account for the anatomical variability in order to translate TBS from a research tool into a viable option for treatment. In this article we present a new method, called CHARM, for automated segmentation of fifteen different head tissues from magnetic resonance (MR) scans. The new method compares favorably to two freely available software tools on a five-tissue segmentation task, while obtaining reasonable segmentation accuracy over all fifteen tissues. The method automatically adapts to variability in the input scans and can thus be directly applied to clinical or research scans acquired with different scanners, sequences or settings. We show that an increase in automated segmentation accuracy results in a lower relative error in electric field simulations when compared to anatomical head models constructed from reference segmentations. However, also the improved segmentations and, by implication, the electric field simulations are affected by systematic artifacts in the input MR scans. As long as the artifacts are unaccounted for, this can lead to local simulation differences up to 30% of the peak field strength on reference simulations. Finally, we exemplarily demonstrate the effect of including all fifteen tissue classes in the field simulations against the standard approach of using only five tissue classes and show that for specific stimulation configurations the local differences can reach 10% of the peak field strength.
Article
Full-text available
SciPy is an open-source scientific computing library for the Python programming language. Since its initial release in 2001, SciPy has become a de facto standard for leveraging scientific algorithms in Python, with over 600 unique code contributors, thousands of dependent packages, over 100,000 dependent repositories and millions of downloads per year. In this work, we provide an overview of the capabilities and development practices of SciPy 1.0 and highlight some recent technical developments. This Perspective describes the development and capabilities of SciPy 1.0, an open source scientific computing library for the Python programming language.
Article
Background The effects of transcranial magnetic stimulation (TMS) on brain activity depend on the design of the stimulation coil. A wide range of coils from different vendors are currently used with different stimulation properties. This decreases the comparability of study results. Objective To systematically compare widely used commercial TMS coils concerning their focality, stimulation depth and efficacy. To provide validated models and data of these coils for accurate simulations of the induced electric fields. Methods We reconstructed the magnetic vector potential of 25 commercially available TMS coils of different vendors from measurements of their magnetic fields. Most coils had a figure-of-eight configuration. We employed the reconstructed magnetic vector potential in simulations of the electric field in a spherical head model. We estimated the motor thresholds of the coil-stimulator combinations using the calculated fields, the pulse waveforms and a leaky integrator model of the neural membrane. Results Our results confirm a previously reported systematic trade-off between focality and relative depth of stimulation. However, neither the peak field strength in the “cortex” of the sphere model nor the estimated motor thresholds were strongly related to the two former measures and need to be additionally determined. Conclusion Our comprehensive coil characterization facilitates objective comparisons of coils of different sizes and from different vendors. The models and auxiliary data will be made available for electric field simulations in SimNIBS. Our work will support TMS users making an informed selection of a suited coil for a specific application and will help to reduce uncertainty regarding the TMS-induced electric field in the brain target region.