ArticlePDF Available

Evaluation of the Puritan Bennett™ 980 Ventilator System Safety and Performance in the Real-World Setting

Taylor & Francis
Medical Devices: Evidence and Research
Authors:
  • LAC+USC Medical Center Keck school of Medicine of University of Southern California

Abstract and Figures

Purpose Mechanical ventilation is a life-supporting intervention but is associated with known risks and complications. To improve the efficacy and safety profile of mechanical ventilation, manufacturers have developed advanced ventilator settings, modes, and alarm strategies to optimize ventilation for patient needs while avoiding complications. However, there is little real-world data published on the deployment of ventilator technology. The main objective of this study was to assess the clinical safety and performance of the Puritan Bennett™ 980 Ventilator System (PB980) using real-world clinical data collected from a diverse, global patient population. Methods This was a multi-center, post-market registry study that included nine sites: four in the United States of America, one in Europe, and four in China. Patients were enrolled into the registry if they were intended to be treated with a PB980. Data collection began at the start of ventilation and continued until extubation off the ventilator or up to seven days of ventilation, whichever occurred first. Subjects were divided by age into three categories: infants (0–365 days), pediatric (1–17 years), and adult (18 years and older). The primary outcome was device-related complication rate. Results Two-hundred-and-eleven subjects were enrolled (41 infants, 48 pediatric, and 122 adults). Sixteen deaths, unrelated to device deficiency, occurred during the data collection timeframe (relative frequency: 7.58, 95% CI: 4.40, 12.0). Only one device-related adverse event was reported (relative frequency: 0.47% 95% CI: 0.01%, 2.61%). Conclusion Ventilation by the PB980 was delivered safely in this multi-center observational study, which included a diverse sample of patients with broad ventilatory needs.
Content may be subject to copyright.
ORIGINAL RESEARCH
Evaluation of the Puritan Bennett™ 980 Ventilator
System Safety and Performance in the Real-World
Setting
Michael Roshon
1
, Paras B Khandhar
2
, Manoj Biniwale
3
, Rangasamy Ramanathan
3
, T Patrick Frazier
4
,
Feng Xu
5
, Linlin Zhang
6
, Xiangdong Guan
7
, Dai Wenling
8
, Bernard Lambermont
9
1
Department of Emergency Medicine, Penrose-St. Francis Health Services, Colorado, Springs, CO, USA;
2
Pediatric Critical Care Medicine, Beaumont
Children’s Hospital, Royal Oak, MI, USA;
3
Division of Neonatology, University of Southern California Keck School of Medicine, Los Angeles, CA, USA;
4
Department of Medicine, University of Alabama at Birmingham, Heersink School of Medicine, Birmingham, AL, USA;
5
Department of Intensive Care,
Children’s Hospital of Chongqing Medical University, Chongqing, People’s Republic of China;
6
Department of Critical Care Medicine, Beijing Tiantan
Hospital, Capital Medical University, Beijing, People’s Republic of China;
7
Department of Critical Care Medicine, the First Afliated Hospital of Sun
Yat-Sen University, Guangzhou, People’s Republic of China;
8
Department of Critical Care Medicine, Yancheng First People’s Hospital, Yancheng,
People’s Republic of China;
9
Department of Intensive Care, University Hospital of Liege, Liege, Belgium
Correspondence: Michael Roshon, Email MichaelRoshon@Centura.Org
Purpose: Mechanical ventilation is a life-supporting intervention but is associated with known risks and complications. To improve
the efcacy and safety prole of mechanical ventilation, manufacturers have developed advanced ventilator settings, modes, and alarm
strategies to optimize ventilation for patient needs while avoiding complications. However, there is little real-world data published on
the deployment of ventilator technology. The main objective of this study was to assess the clinical safety and performance of the
Puritan Bennett™ 980 Ventilator System (PB980) using real-world clinical data collected from a diverse, global patient population.
Methods: This was a multi-center, post-market registry study that included nine sites: four in the United States of America, one in
Europe, and four in China. Patients were enrolled into the registry if they were intended to be treated with a PB980. Data collection
began at the start of ventilation and continued until extubation off the ventilator or up to seven days of ventilation, whichever occurred
rst. Subjects were divided by age into three categories: infants (0–365 days), pediatric (1–17 years), and adult (18 years and older).
The primary outcome was device-related complication rate.
Results: Two-hundred-and-eleven subjects were enrolled (41 infants, 48 pediatric, and 122 adults). Sixteen deaths, unrelated to device
deciency, occurred during the data collection timeframe (relative frequency: 7.58, 95% CI: 4.40, 12.0). Only one device-related
adverse event was reported (relative frequency: 0.47% 95% CI: 0.01%, 2.61%).
Conclusion: Ventilation by the PB980 was delivered safely in this multi-center observational study, which included a diverse sample
of patients with broad ventilatory needs.
Plain Language Summary: Mechanical ventilation is a life supporting intervention. Much progress has been made in this eld
thanks to a better knowledge of respiratory physiopathology and improved ventilation delivered by modern ventilators.
In this global, post-market registry study, ventilation by the Puritan Bennett™ 980 Ventilator was delivered safely to a diverse
sample of patients.
Keywords: ventilation, PB980, safety, complications, respiratory distress, critical care
Introduction
Mechanical ventilation is a life-supporting intervention for patients with respiratory failure or critical illness. However,
mechanical ventilation also carries the risk of signicant complications, including ventilator-associated pneumonia,
neuromuscular weakness from sedative therapies, volutrauma, and barotrauma.
1,2
Ventilation devices have advanced to
Medical Devices: Evidence and Research 2024:17 37–45 37
© 2024 Roshon et al. This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.
php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the
work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For
permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
Medical Devices: Evidence and Research Dovepress
open access to scientific and medical research
Open Access Full Text Article
Received: 8 August 2023
Accepted: 12 January 2024
Published: 23 January 2024
employ a variety of settings and modes that can be manipulated in response to patient ventilation needs to improve the
efcacy and safety prole.
3–5
Ventilator design has become more sophisticated and rened over the past century, with improvements from negative-
pressure ventilators that required the patient’s entire body to be encased in the 19th and early 20th centuries to today’s
positive-pressure ventilators that boast a plethora of settings and modes to provide responsive ventilation to a large
spectrum of patient needs.
6,7
Ventilation solutions are intended to protect the airway and support the pulmonary system
while minimizing ventilator-induced complications.
8
To optimize patient safety and comfort, ventilator advancements
aim to make mechanical ventilation synchronous with patient demands and less intermittent mandatory delivery of
positive pressure.
The Puritan Bennett™ 980 Ventilator System (PB980) is a CE-marked and FDA-cleared medical device designed to
provide continuous ventilation for patients who require respiratory support, delivered via invasive mechanical ventilation
(IMV) or noninvasive ventilation (NIV). Care can be administered to neonatal, pediatric, and adult patients weighing at
least 0.3 kg, with tidal volumes for mandatory volume-controlled breaths from 2 mL to 2500 mL. The PB980 is intended
for use in hospitals, and intra-hospital transport.
Several bench studies have been published utilizing lung models to simulate a variety of ventilation needs to support
the safety and performance of the PB980.
9–16
However, despite the advances in ventilator technology, real-world
experience and the safety performance of these devices have yet to be fully explored in the medical literature.
The PB980 Post-Approval Registry study aimed to assess the clinical safety and performance of the PB980 using
real-world clinical data collected from a diverse, global patient population. The primary objective was to obtain direct
clinical evidence to support the safety of the ventilator system.
Materials and Methods
Study Design
Data was collected from the PB980 Post-Approval Registry, a prospective, observational, multinational study conducted
from June 2020 to April 2022 in the United States of America (USA), Europe, and China (Table 1). Enrollment in the
study did not impact the care patients received before, during, or after participation. Study procedures were performed in
accordance with the Declaration of Helsinki under a protocol that was duly approved by the Institutional Review Board
or Ethics Committee (IRB/EC), as applicable, at each site. This study was sponsored and supported by Medtronic
(Minneapolis, MN).
Table 1 Participating Sites and Enrollments
Site Information Enrollments
Site Name Country Adult Pediatric Infant
Penrose-St. Francis Health Services, Colorado Springs, Colorado USA 70 0 0
Beaumont Children’s Hospital, Royal Oak, Michigan USA 4 25 10
Los Angeles General Medical Center, Los Angeles, California USA 0 9 21
University of Alabama at Birmingham, Birmingham, Alabama USA 17 0 0
Children’s Hospital of Chongqing Medical University, Chongqing China 0 6 10
Beijing Tiantan Hospital, Capital Medical University, Beijing China 15 0 0
The First Afliated Hospital of Sun Yat-Sen University, Guangzhou China 11 0 0
Yancheng First People’s Hospital, Guangzhou China 0 8 0
University Hospital of Liege, Liege Belgium 5 0 0
Total 122 48 41
https://doi.org/10.2147/MDER.S433900
DovePress
Medical Devices: Evidence and Research 2024:17
38
Roshon et al Dovepress
Powered by TCPDF (www.tcpdf.org)
Patient and Procedures
Patients receiving IMV or NIV using the PB980, as ordered by the treating physician, were eligible for study enrollment.
Enrollment took place prior to ventilation or up to 30 days following the start of ventilation. Data were collected for the
duration of mechanical ventilation to extubation or seven ventilation days, whichever occurred rst.
Subjects were divided based on age into three cohorts: infants (0–365 days), pediatric (1 year – 17 years), and adult (18
years and above). Due to the population under study, many subjects presented with trauma and/or were intubated, therefore
a variety of methods were used to obtain consent. Per local IRB/EC requirements, consent was obtained in one of the
following ways, depending on clinical situation: 1) prior to receiving care, 2) after receiving care (up to 30 days following
start of ventilation), or 3) waiver of consent. Waiver of consent was provided at the discretion of each site’s IRB/EC, with
only sites in the USA utilizing this option. For pediatric subjects who were not enrolled under waiver of consent, assent and
parental and/or guardian consent was obtained, per local IRB/EC requirements. For infant subjects, who were not enrolled
under waiver of consent, parental and/or guardian consent was obtained.
The PB980 Post-Approval Registry was designed to collect real-world clinical data on various patients using the
ventilator as intended for routine medical care. Ventilator settings were managed by the attending clinician(s), as
determined by patient clinical need and routine care. Trained research staff recorded ventilator settings every four
hours for the duration of ventilation or up to seven ventilation days, whichever occurred rst. All system-related adverse
events and device deciencies that occurred during the data collection period were promptly reported.
Data Collection
The primary endpoint was dened as the device complication rate of the PB980 during the duration of the study.
Complications were dened as any event related to the ventilator or the ventilation therapy that impacted the patient
during study participation. Site investigators assessed all complications and complication details, and investigator-
determined relatedness was reported to the study sponsor and IRB/EC, as applicable.
In addition to the primary endpoint, subject demographics, including age, sex, weight, and height were collected.
A brief medical history was obtained by reviewing the medical record to determine the history or presence of brain/
neurological disease, cardiac disease, diabetes, renal disease, respiratory disease, stroke, or musculoskeletal disease.
Admission diagnosis and reason for surgery, if applicable, were noted, along with an indication for respiratory
intervention. The type of ventilation mode, endotracheal tube use, date and time of initial ventilation, date and time of
extubation, and date and time of admission to and discharge from the intensive care unit were also recorded. Ventilation
settings were collected every four hours and included ventilation type, ventilation mode, fractional inspired oxygen
(FiO
2
), blood oxygen saturation (SpO
2
), tidal volume, peak inspiratory pressure, inspiration time, and positive end-
expiratory pressure (PEEP). Additionally, physiologic measures were collected every four hours and included respiratory
rate, heart rate, and blood gas results, if available, including blood pH, partial pressure of carbon dioxide (PaCO
2
), partial
pressure of oxygen (PaO
2
), and bicarbonate (HCO
3
). Reason for study exit, including death, was also collected.
Statistical Analysis
The data was analyzed using SAS© statistical software, Version 9.4 (TS1M7) (SAS Institute Inc., Cary, NC, USA).
Categorical variables are presented as counts and percentages. Descriptive statistics are provided for continuous variables.
As this was an observational study, no treatment groups were assigned, or statistical signicance testing completed.
A minimum sample size of 160 subjects total, with at least 30 subjects in each cohort, was determined necessary to
detect a complication rate between 5% and 10%, with a precision of 1.7% to 2.4% standard error. This sample size
allowed for a 95% chance of detecting a complication rate as low as 2%.
Results
Participants
Two-hundred-eleven subjects were enrolled in the registry across the three age groups (Table 2), exceeding the minimum
sample size calculated to achieve study goals. The most commonly reported medical history in adult subjects included
Medical Devices: Evidence and Research 2024:17 https://doi.org/10.2147/MDER.S433900
DovePress
39
Dovepress Roshon et al
Powered by TCPDF (www.tcpdf.org)
cardiac, brain/neurological, and respiratory disease (Table 3). Pediatric patients most frequently reported brain/neurolo-
gical disease and respiratory disease, while infants most commonly had a history of respiratory disease. Across all three
age cohorts, the most common indication for ventilation was related to presenting medical conditions as opposed to
trauma or surgery.
Procedural Characteristics
Only one device-related complication was reported during the study across all 211 subjects (relative frequency: 0.47%,
95% CI: 0.01%, 2.61%). The device deciency occurred in a 56-year-old female who presented with trauma due to a fall.
At baseline, the patient had an ASA status of 4 and a medical history positive for brain or neurological disease,
respiratory disease, and musculoskeletal disease. The subject received IMV using the PB980 20 days after initial
admission to the intensive care unit, after being transferred from another ventilator (no data collection occurred while
Table 2 Subject Characteristics
Variable Category Adults Pediatric Infant
Age (Years: Adult,
Pediatrics; Days: Infant)
n 122 48 41
Mean±SD 56.8±17.7 6.1±4.5 63.6±97.9
Median 59.0 4.5 21.0
95% CI of Mean [53.7, 60.0] [4.78, 7.41] [32.7, 94.5]
Range 19–87 1–17 0–365
Sex n 122 48 41
Male 68 (55.7%) 27 (56.3%) 26 (63.4%)
Female 54 (44.3%) 21 (43.8%) 15 (36.6%)
Weight (kg) n 117* 48 41
Mean±SD 79.35±23.87 24.70±18.23 3.82±2.43
Median 75.00 17.40 83.40
95% CI of Mean [75.0, 83.7] [19.4, 30.0] [3.07, 4.57]
Range 28.9–152.0 9.0–80.3 0.7–13.5
Note: *Weight is missing for ve subjects.
Abbreviations: SD, standard deviation; CI, condence interval.
Table 3 Subject Medical History and Indication for Ventilation
Medical History Adults (N=122) Pediatric (N=48) Infants (N=41)
Brain Or Neurological Disease 57 (46.7%) 20 (41.7%) 6 (14.6%)
Cardiac Disease 60 (49.2%) 2 (4.2%) 9 (22.0%)
Diabetes 24 (19.7%) 0 (0.0%) 1 (2.4%)
Musculoskeletal Disease 24 (19.7%) 13 (27.1%) 4 (9.8%)
Renal Disease 27 (22.1%) 1 (2.1%) 0 (0.0%)
Respiratory Disease 49 (40.2%) 16 (33.3%) 23 (56.1%)
Stroke 16 (13.1%) 0 (0.0%) 0 (0.0%)
Indication for Ventilation
Medical Reason 90 (73.8%) 31 (64.6%) 28 (68.3%)
Post-Operative 14 (11.5%) 7 (14.6%) 12 (29.3%)
Trauma 18 (14.8%) 10 (20.8%) 1 (2.4%)
https://doi.org/10.2147/MDER.S433900
DovePress
Medical Devices: Evidence and Research 2024:17
40
Roshon et al Dovepress
Powered by TCPDF (www.tcpdf.org)
receiving ventilation support using non-PB980 ventilator). While the subject was on the PB980, the most common
ventilation setting used was spontaneous mode set to PAV+ and ow trigger mode. The complication was an expiratory
lter blockage that occurred after the subject had been on the PB980 for approximately one day. The ventilator alarm
appropriately indicated an expiratory lter warning, and the subject was immediately transferred to another ventilator
(not PB980) to avoid further obstruction. The blockage did not appear to result in any short or long-term detriment to the
subject’s health and did not prolong hospital length of stay. Sixteen deaths (relative frequency: 7.58, 95% CI: 4.40, 12.0),
none related to device deciency, occurred during the study (Table 4). Reported reasons for death included: respiratory
failure (n=6), cardiogenic shock (n=3), sepsis (n=1), stroke (n=1), cardiac arrest (n=1), and incomplete description given
(n=4). Six patients were reported to have been removed from the ventilator as part of comfort care.
The most frequently used ventilator settings during study enrollment are presented in Table 5. All pediatric and adult
subjects, except one, received IMV, whereas approximately half of the infants received NIV. During the study period, three
adults and two infants were noted to have alternated between IMV and NIV. The most frequently used ventilator settings are
reported. Assist control was the most frequently reported breath control mode for adult patients, while pediatric and infant
patients were most frequently treated using synchronized, intermittent mandatory ventilation (SIMV). SIMV mode is
Table 5 Most Frequently Used Ventilator Settings by Age Group
Parameter Setting Adult (N=122) Pediatric (N=48) Infant (N=41)
Ventilation Type Invasive 120 (98.4%) 48 (100.0%) 22 (53.7%)
Non-Invasive 1 (0.8%) 0 (0.0%) 19 (46.3%)
Missing 1 (0.8%) 0 (0.0%) 0 (0.0%)
Ventilator Mode
Breath Control Assist Control 87 (71.3%) 12 (25.0%) 4 (9.8%)
Bi-Level 3 (2.5%) 0 (0.0%) 0 (0.0%)
CPAP 0 (0.0%) 0 (0.0%) 5 (12.2%)
SIMV 16 (13.1%) 35 (72.9%) 32 (78.0%)
SPONT 16 (13.1%) 1 (2.1%) 0 (0.0%)
Breath Type
Mandatory Type Pressure Support 19 (15.6%) 9 (18.8%) 31 (75.6%)
Volume Control 51 (41.8%) 9 (18.8%) 2 (4.9%)
Volume Control Plus 37 (30.3%) 31 (64.6%) 7 (17.1%)
Spontaneous Type Pressure Support 36 (29.5%) 42 (87.5%) 40 (97.6%)
Trigger Type Flow Triggering 96 (78.7%) 45 (93.8%) 40 (97.6%)
IE Sync Triggering 1 (0.8%) 0 (0.0%) 0 (0.0%)
Pressure Triggering 7 (5.7%) 3 (6.3%) 0 (0.0%)
(Continued)
Table 4 Summary of Deaths Reported During Study
Statistic Adult (N=122) Pediatric (N=48) Infant (N=41) Total (N=211)
Count 12 3 1 16
Relative Frequency (%) 9.84 6.25 2.44 7.58
95% CI* [5.2, 16.6] [1.3, 17.2] [0.1, 12.9] [4.40, 12.0]
Note: *Clopper-Pearson method for exact binomial condence intervals. No reported deaths were associated with device
deciency.
Abbreviation: CI, condence interval.
Medical Devices: Evidence and Research 2024:17 https://doi.org/10.2147/MDER.S433900
DovePress
41
Dovepress Roshon et al
Powered by TCPDF (www.tcpdf.org)
available for both IMV and NIV. Infants received pressure support more frequently than adult or pediatric patients. Flow
triggering was the most frequently used trigger type across all cohorts. Table 6 depicts physiologic measures collected
during study procedures and reects the range of physiologies and patient states included in this registry.
Table 5 (Continued).
Parameter Setting Adult (N=122) Pediatric (N=48) Infant (N=41)
Average FiO
2
n 122 48 41
Mean±SD 0.50±0.15 0.42±0.08 0.32±0.11
Median 0.45 0.41 0.30
95% CI of Mean [0.47, 0.52] [0.40, 0.45] [0.28, 0.35]
Range 0.21–1 0.26–0.62 0.21–0.75
Average PEEP (cm H
2
O) n 122 48 39
Mean±SD 6.5±2.3 5.6±1.7 5.4±1.4
Median 5.0 5.0 5.2
95% CI of Mean [6.09, 6.89] [5.10, 6.06] [4.97, 5.85]
Range 4.4–13.9 3.0–11.3 3.0–8.0
Note: For ventilation settings, the most frequently used ventilator setting during study enrollment is presented.
Abbreviations: SD, standard deviation; CI, condence inter val; CPAP, continuous positive airway pressure; SIMV, synchronized
intermittent mandatory ventilation; SPONT, spontaneous breathing mode; FiO
2
, fraction of inspired oxygen; PEEP, positive end-
expiratory pressure.
Table 6 Physiologic Measures by Age Group
Parameter Statistic Adult (N=122) Pediatric (N=48) Infant (N=41)
Average Respiratory Rate (breaths/min) n 122 48 41
Mean±SD 18.2±5.12 23.3±6.24 38.6±9.68
Median 17.0 23.3 39.9
95% CI of Mean [17.2, 19.1] [21.5, 25.1] [35.5, 41.6]
Range 10.0–34.0 11.3–41.4 18.8–60.0
Average SpO
2
(%) n 122 48 41
Mean±SD 97.2±2.81 98.7±1.44 96.9±2.04
Median 98.0 99.1 97.0
95% CI of Mean [96.7, 97.7] [98.3, 99.2] [96.3, 97.6]
Range 79.5–100.0 92.6–100.0 92.2–100.0
Average Heart Rate (beats/min) n 122 48 41
Mean±SD 86.1±14.55 115.4±23.06 140.6±14.95
Median 85.2 113.0 142.0
95% CI of Mean [83.4, 88.7] [108.7, 122.1] [135.9, 145.3]
Range 51.7–126.6 76.0–166.2 107.8–167.8
Average pH n 100 44 31
Mean±SD 7.39±0.106 7.35±0.130 7.41±0.075
Median 7.39 7.39 7.40
95% CI of Mean [7.37, 7.41] [7.31, 7.39] [7.38, 7.43]
Range 7.04–7.62 7.00–7.52 7.29–7.71
Average PaCO
2
(mmHg) n 100 44 31
Mean±SD 39.3±10.51 39.9±6.03 40.9±7.94
Median 37.7 39.0 40.7
95% CI of Mean [37.2, 41.4] [38.1, 41.8] [38.0, 43.8]
Range 14.3–85.0 23.3–53.6 25.0–58.0
(Continued)
https://doi.org/10.2147/MDER.S433900
DovePress
Medical Devices: Evidence and Research 2024:17
42
Roshon et al Dovepress
Powered by TCPDF (www.tcpdf.org)
Discussion
Previous studies have explored ventilator safety and performance using lung models to simulate patient scenarios.
9–16
These studies have concentrated primarily on bench studies and have demonstrated superior or comparative performance
of the PB980 to other ventilators.
9–12,14,16
There is robust clinical literature examining patient outcomes and ventilator
use. However, less is explicitly reported on ventilator safety and performance in the clinical setting. Case reports have
documented ventilator malfunction and its effects on individual patient outcomes,
17–22
with some case reports speci-
cally describing ventilator lter blockage. In previous reports, lter blockage due to patient secretions led to bilateral
tension pneumothorax, presumably secondary to defective anesthesia breathing circuit lter.
23,24
If the lter on
a ventilator is obstructed, oxygen may not ow properly to the patient. Hypoxia can have serious adverse effects,
including death. Obstruction can be detected by the ventilator with changes in pressure which would cause the ventilator
to alarm. In response to the lter blockage during the current study, the alarm went off and the ventilator was changed,
resulting in no further complications.
Three studies examining reported ventilator-related adverse events using national registries found that the most
common cause for ventilator-related adverse events was associated with user error.
25–27
A large screen displaying
respiratory curves on the PB980 allows for monitoring and storing the curves during inspiratory and expiratory pauses,
allows for the calculation and monitoring of intrinsic PEEP, total respiratory system resistance, and compliance. Those
measurements can be helpful when determining the delivery of respiratory support. Data from the UK National Reporting
and Learning System (2006–2008) found that of the 1029 incidents, 17.9% were related to the ventilator, though specic
information about ventilator type, settings, or patient condition was not reported.
27
To our knowledge, this is the rst
study to report PB980 ventilator-related safety and performance as a primary outcome in a real-world setting on a large
sample of diverse patients.
Limitations
This study was designed to collect information on ventilator safety and performance and did not collect information on
additional clinical parameters or outcomes. This limits the clinical conclusions that can be derived from the data. Any
future study or expansion of the registry should include a collection of clinically relevant outcomes. Further, subjects
were only enrolled in the study if they received ventilation with the PB980 and only up to seven days of ventilation, thus
potentially excluding large swaths of ventilatory care provided to patients. Efforts were made to create a diverse sample;
however, the data do not include all patient or disease types who receive ventilation across the globe. Additionally, data
was collected every four hours with the most frequent ventilator settings reported, obscuring any changes to ventilation
settings that may have taken place between data collection intervals.
Table 6 (Continued).
Parameter Statistic Adult (N=122) Pediatric (N=48) Infant (N=41)
Average PaO
2
(mmHg) n 100 36 29
Mean±SD 119.7±59.06 130.3±58.56 80.9±33.60
Median 101.3 129.6 70.5
95% CI of Mean [107.9, 131.4] [110.5, 150.1] [68.1, 93.6]
Range 53.8–372.0 34.0–264.4 31.5–173.0
Average HCO
3
(mEq/L) n 99 43 31
Mean±SD 23.4±5.17 24.2±4.14 24.8±4.50
Median 23.5 23.5 23.9
95% CI of Mean [22.4, 24.5] [22.9, 25.5] [23.2, 26.5]
Range 9.5–37.7 10.3–34.0 18.0–36.0
Abbreviations: SD, standard deviation; CI, condence interval; SpO
2
, oxygen saturation; PaCO
2
, partial pressure of carbon dioxide; PaO
2
,
partial pressure of oxygen; HCO
3
, bicarbonate.
Medical Devices: Evidence and Research 2024:17 https://doi.org/10.2147/MDER.S433900
DovePress
43
Dovepress Roshon et al
Powered by TCPDF (www.tcpdf.org)
Conclusion
Using real-world, global data, this study supports the safety and performance of the PB980 across a diverse group of
patients with a wide variety of ventilation settings and physiologic parameters.
Acknowledgments
Study execution support was provided by Katherine Schiller, Naomi Wang, Demarcus Williams, and Ryan Zhou of
Medtronic, Elizabeth Kring and Christine Batchelder of Beaumont Children’s, and Nathan Dureg and Kate Ramm of
University of Southern California General Medical Center. Biostatistical analysis and support was provided by Anne
Sexter, Guang Yang, Hui Xiong, and Julia Yang of Medtronic. Medical writing support was provided by Hanan Zavala of
Medtronic in accordance with Good Publication Practice (GPP 2022) guidelines.
28
An abstract of this work was accepted and presented at the Society of Critical Care Medicine 2024 Critical Care
Congress (January 21 23, Phoenix Arizona).
Disclosure
The study was sponsored and funded by Medtronic. All authors (or their institutions) received research support from
Medtronic to conduct this study.
Michael Roshon serves as a director on USACS’ National Clinical Governance Board.
References
1. Klompas M, Kleinman K, Murphy MV. Descriptive epidemiology and attributable morbidity of ventilator-associated events. Infect Control Hosp
Epidemiol. 2014;35(5):502–510. doi:10.1086/675834
2. Silva PL, Ball L, Rocco PR, Pelosi P. Physiological and Pathophysiological Consequences of Mechanical Ventilation. Thieme Medical Publishers,
Inc.; 2022.
3. Dexter AM, Clark K. Ventilator graphics: scalars, loops, & secondary measures. Respiratory Care. 2020;65(6):739–759. doi:10.4187/respcare.07805
4. Dos Santos Rocha A, Habre W, Albu G. Novel ventilation techniques in children. Pediatr Anesthesia. 2022;32(2):286–294. doi:10.1111/pan.14344
5. Pham T, Brochard LJ, Slutsky AS. Mechanical ventilation: state of the art. Elsevier. 2017;1382–1400.
6. Kacmarek RM. The mechanical ventilator: past, present, and future. Respiratory Care. 2011;56(8):1170–1180. doi:10.4187/respcare.01420
7. MacIntyre N, Rackley C, Khusid F. Fifty years of mechanical ventilation—1970s to 2020. Crit Care Med. 2021;49(4):558–574. doi:10.1097/
CCM.0000000000004894
8. Egbuta C, Easley RB. Update on ventilation management in the Pediatric Intensive Care Unit. Pediatr Anesthesia. 2022;32(2):354–362.
doi:10.1111/pan.14374
9. Beloncle F, Piquilloud L, Olivier PY, et al. Accuracy of P0.1 measurements performed by ICU ventilators: a bench study. Ann Intensive Care.
2019;9(1):104. doi:10.1186/s13613-019-0576-x
10. Itagaki T, Bennett DJ, Chenelle CT, Fisher DF, Kacmarek RM. Performance of Leak Compensation in All-Age ICU Ventilators During
Volume-Targeted Neonatal Ventilation: a Lung Model Study. Respiratory Care. 2017;62(1):10–21. doi:10.4187/respcare.05012
11. Itagaki T, Chenelle CT, Bennett DJ, Fisher DF, Kacmarek RM. Effects of Leak Compensation on Patient-Ventilator Synchrony During Premature/
Neonatal Invasive and Noninvasive Ventilation: a Lung Model Study. Respiratory Care. 2017;62(1):22–33. doi:10.4187/respcare.04825
12. Koyama Y, Uchiyama A, Yoshida J, Yoshida T, Yamashita T, Fujino Y. A Comparison of the Adjustable Ranges of Inspiratory Pressurization
During Pressure Controlled Continuous Mandatory Ventilation of 5 ICU Ventilators. Respiratory Care. 2018;63(7):849–858. doi:10.4187/
respcare.05286
13. Krieger TJ, Wald M. Volume-targeted ventilation in the neonate: benchmarking ventilators on an active lung model. Pediatr Crit Care Med.
2017;18(3):241–248. doi:10.1097/PCC.0000000000001088
14. Moon K, Takeuchi M, Tachibana K, Mizuguchi S, Hirano S. Accuracy of Reported Tidal Volume During Neonatal Ventilation With Airway Leak:
a Lung Model Study. Pediatr Crit Care Med. 2019;20(1):e37–e45. doi:10.1097/PCC.0000000000001752
15. Morita PP, Weinstein PB, Flewwelling CJ, et al. The usability of ventilators: a comparative evaluation of use safety and user experience. Crit Care.
2016;20(1):263. doi:10.1186/s13054-016-1431-1
16. Marjanovic NS, De Simone A, Jegou G, L’Her E. A new global and comprehensive model for ICU ventilator performances evaluation. Ann
Intensive Care. 2017;7(1):68. doi:10.1186/s13613-017-0285-2
17. Di Paolo M, Evangelisti L, Ambrosino N. Unexpected death of a ventilator-dependent amyotrophic lateral sclerosis patient. Revista Portuguesa de
Pneumologia. 2013;194:175–178. doi:10.1016/j.rppneu.2012.12.001
18. Kim K. Ventilator malfunction due to water condensation during low ow anaesthesia. Anaesthesia Intensive Care. 2011;39(6):1155.
19. Kumar BK, Ravi M, Dinesh K, Nanda A. Ventilator malfunction. J Anaesthesiol Clin Pharmacol. 2011;27(4):576. doi:10.4103/0970-9185.86623
20. Sripriya R, Parthasarathy S, Ravishankar M. Ventilator dysfunction: role of graphics in detection. Ain-Shams J Anaesthesiol. 2016;9(3):465.
doi:10.4103/1687-7934.189103
21. Olayode A, Mashriqi N, Liu C, et al. 882: A Case of Ventilator Malfunction-Induced Pneumothorax. Crit Care Med. 2023;51(1):434. doi:10.1097/
01.ccm.0000909256.95446.33
22. le Noble JL, Terpoorten FJ, der Snoek M, Foudraine N. Ventilator dysfunction due to unexpected salbutamol crystallisation while using a nebuliser:
a practical advice for intensivists. Intensive Care Med Exp. 2022;10(1):1–3. doi:10.1186/s40635-022-00441-y
https://doi.org/10.2147/MDER.S433900
DovePress
Medical Devices: Evidence and Research 2024:17
44
Roshon et al Dovepress
Powered by TCPDF (www.tcpdf.org)
23. McEwan AI, Dowell AL, Karis JH. Bilateral tension pneumothorax caused by a blocked bacterial lter in an anesthesia breathing circuit.
Anesthesia Analg. 1993;76(2):440–442.
24. Smith CE, Otworth JR, Kaluszyk P. Bilateral tension pneumothorax due to a defective anesthesia breathing circuit lter. J Clin Anesthesia. 1991;3
(3):229–234. doi:10.1016/0952-8180(91)90166-K
25. Kamio T, Masamune K. Mechanical ventilation-related safety incidents in general care wards and ICU settings. Respiratory Care. 2018;63
(10):1246–1252. doi:10.4187/respcare.06109
26. Pham JC, Williams TL, Sparnon EM, Cillie TK, Scharen HF, Marella WM. Ventilator-related adverse events: a taxonomy and ndings from 3
incident reporting systems. Respiratory Care. 2016;61(5):621–631. doi:10.4187/respcare.04151
27. Cassidy C, Smith A, Arnot-Smith J. Critical incident reports concerning anaesthetic equipment: analysis of the UK National Reporting and
Learning System (NRLS) data from 2006–2008. Anaesthesia. 2011;66(10):879–888. doi:10.1111/j.1365-2044.2011.06826.x
28. DeTora LM, Toroser D, Sykes A, et al. Good Publication Practice (GPP) Guidelines for Company-Sponsored Biomedical Research: 2022 Update.
Ann Intern Med. 2022;175(9):1298–1304. doi:10.7326/m22-1460
Medical Devices: Evidence and Research Dovepress
Publish your work in this journal
Medical Devices: Evidence and Research is an international, peer-reviewed, open access journal that focuses on the evidence, technology,
research, and expert opinion supporting the use and application of medical devices in the diagnosis, monitoring, treatment and management of
clinical conditions and physiological processes. The identication of novel devices and optimal use of existing devices which will lead to
improved clinical outcomes and more effective patient management and safety is a key feature of the journal. The manuscript management
system is completely online and includes a very quick and fair peer-review system. Visit http://www.dovepress.com/testimonials.php to read
real quotes from published authors.
Submit your manuscript here: https://www.dovepress.com/medical-devices-evidence-and-research-journal
Medical Devices: Evidence and Research 2024:17 DovePress
45
Dovepress Roshon et al
Powered by TCPDF (www.tcpdf.org)
ResearchGate has not been able to resolve any citations for this publication.
Article
Full-text available
Extraordinary progress has been made during the past few decades in the development of anesthesia machines and ventilation techniques. With unprecedented precision and performance, modern machines for pediatric anesthesia can deliver appropriate mechanical ventilation for children and infants of all sizes and with ongoing respiratory diseases, ensuring very small volume delivery and compensating for circuit compliance. Along with highly accurate monitoring of the delivered ventilation, modern ventilators for pediatric anesthesia also have a broad choice of ventilation modalities, including synchronized and assisted ventilation modes, which were initially conceived for ventilation weaning in the intensive care setting. Despite these technical advances, there is still room for improvement in pediatric mechanical ventilation. There is a growing effort to minimize the harm of intraoperative mechanical ventilation of children by adopting the protective ventilation strategies that were previously employed only for prolonged mechanical ventilation. More than ever, the pediatric anesthesiologist should now recognize that positive-pressure ventilation is potentially a harmful procedure, even in healthy children, as it can contribute to both ventilator-induced lung injury and ventilator-induced diaphragmatic dysfunction. Therefore, careful choice of the ventilation modality and its parameters is of paramount importance to optimize gas exchange and to protect the lungs from injury during general anesthesia. The present report reviews the novel ventilation techniques used for children, discussing the advantages and pitfalls of the ventilation modalities available in modern anesthesia machines, as well as innovative ventilation modes currently under development or research. Several innovative strategies and devices are discussed. These novel modalities are likely to become part of the armamentarium of the pediatric anesthesiologist in the near future and are particularly relevant for challenging ventilation scenarios.
Article
Full-text available
Background: Occlusion pressure at 100 ms (P0.1), defined as the negative pressure measured 100 ms after the initiation of an inspiratory effort performed against a closed respiratory circuit, has been shown to be well correlated with central respiratory drive and respiratory effort. Automated P0.1 measurement is available on modern ventilators. However, the reliability of this measurement has never been studied. This bench study aimed at assessing the accuracy of P0.1 measurements automatically performed by different ICU ventilators. Methods: Five ventilators set in pressure support mode were tested using a two-chamber test lung model simulating spontaneous breathing. P0.1 automatically displayed on the ventilator screen (P0.1vent) was recorded at three levels of simulated inspiratory effort corresponding to P0.1 of 2.5, 5 and 10 cm H2O measured directly at the test lung and considered as the reference values of P0.1 (P0.1ref). The pressure drop after 100 ms was measured offline on the airway pressure-time curves recorded during the automated P0.1 measurements (P0.1aw). P0.1vent was compared to P0.1ref and to P0.1aw. To assess the potential impact of the circuit length, P0.1 were also measured with circuits of different lengths (P0.1circuit). Results: Variations of P0.1vent correlated well with variations of P0.1ref. Overall, P0.1vent underestimated P0.1ref except for the Löwenstein® ventilator at P0.1ref 2.5 cm H2O and for the Getinge group® ventilator at P0.1ref 10 cm H2O. The agreement between P0.1vent and P0.1ref assessed with the Bland-Altman method gave a mean bias of - 1.3 cm H2O (limits of agreement: 1 and - 3.7 cm H2O). Analysis of airway pressure-time and flow-time curves showed that all the tested ventilators except the Getinge group® ventilator performed an occlusion of at least 100 ms to measure P0.1. The agreement between P0.1vent and P0.1aw assessed with the Bland-Altman method gave a mean bias of 0.5 cm H2O (limits of agreement: 2.4 and - 1.4 cm H2O). The circuit's length impacted P0.1 measurements' values. A longer circuit was associated with lower P0.1circuit values. Conclusion: P0.1vent relative changes are well correlated to P0.1ref changes in all the tested ventilators. Accuracy of absolute values of P0.1vent varies according to the ventilator model. Overall, P0.1vent underestimates P0.1ref. The length of the circuit may partially explain P0.1vent underestimation.
Article
These updated Good Publication Practice (GPP) guidelines include recommendations for publishing company-sponsored biomedical research. The GPP guidelines apply to peer-reviewed or peer-oriented biomedical publications, such as manuscripts, meeting presentations, posters, and abstracts, as well as enhanced content, such as plain-language summaries. The current GPP guidelines incorporate guidance on ethics and transparency as well as the planning, development, review, and approval of biomedical publications and policies and procedures that describe these practices. Supplemental materials lay out processes for steering committees, publication plans, publication working groups, determining authorship, and documentation. Information about new topics, such as alliances and working with patients, has been included where appropriate within these supplemental materials. Incorporating the principles and best practices presented in these GPP guidelines will result in increased transparency and a firm ethical footing. This guidance is also intended to enable the compliant incorporation of new and emerging publication tools for the ethical publication of company-sponsored research.
Article
Mechanical ventilation is a life-support system used to ensure blood gas exchange and to assist the respiratory muscles in ventilating the lung during the acute phase of lung disease or following surgery. Positive-pressure mechanical ventilation differs considerably from normal physiologic breathing. This may lead to several negative physiological consequences, both on the lungs and on peripheral organs. First, hemodynamic changes can affect cardiovascular performance, cerebral perfusion pressure (CPP), and drainage of renal veins. Second, the negative effect of mechanical ventilation (compression stress) on the alveolar-capillary membrane and extracellular matrix may cause local and systemic inflammation, promoting lung and peripheral-organ injury. Third, intra-abdominal hypertension may further impair lung and peripheral-organ function during controlled and assisted ventilation. Mechanical ventilation should be optimized and personalized in each patient according to individual clinical needs. Multiple parameters must be adjusted appropriately to minimize ventilator-induced lung injury (VILI), including: inspiratory stress (the respiratory system inspiratory plateau pressure); dynamic strain (the ratio between tidal volume and the end-expiratory lung volume, or inspiratory capacity); static strain (the end-expiratory lung volume determined by positive end-expiratory pressure [PEEP]); driving pressure (the difference between the respiratory system inspiratory plateau pressure and PEEP); and mechanical power (the amount of mechanical energy imparted as a function of respiratory rate). More recently, patient self-inflicted lung injury (P-SILI) has been proposed as a potential mechanism promoting VILI. In the present chapter, we will discuss the physiological and pathophysiological consequences of mechanical ventilation and how to personalize mechanical ventilation parameters.
Article
Ventilator graphic monitoring is common in ICUs. The graphic information provides clinicians with immediate clues regarding patient-ventilator interaction and ventilator function. These display tools are aimed at reducing complications associated with mechanical ventilation, such as patient-ventilator asynchrony. It is also useful to assess respiratory mechanics in mechanically ventilated patients using both scalar and plot displays on the ventilator. Additional information can be gained by observing secondary ventilator measures including stress index, inflection points, and work of breathing. Ventilator graphics impact mechanical ventilation management through optimizing effectiveness of patient care and enhancing promptness of clinician response. Despite being a valuable asset in providing high-quality patient care, many bedside clinicians do not have a thorough understanding of ventilator graphics. Mastery of ventilator graphics interpretation is key in managing patients who are receiving ventilatory support.
Article
Objectives: To determine the accuracy of tidal volume reported by neonatal ventilators, with and without leak compensation, in invasive and noninvasive ventilation modes in the presence of airway leak; and, to determine what factors have a significant effect on the accuracy of tidal volume reported by ventilators with leak compensation in the presence of airway leak. We hypothesized that ventilators with a leak compensation function that includes estimation of tidal volume could accurately report tidal volume in the presence of airway leak, but that the accuracy of reported tidal volume may be affected by variables such as the identity of the ventilator, lung mechanics, leak size, positive end-expiratory pressure level, and body size. Design: In vitro assessment of ventilator volume delivery was conducted for seven acute care ventilators using a passive lung simulator. Setting: Laboratory-based measurements. Interventions: The error of reported tidal volume was calculated under three ventilation modes (noninvasive-pressure-control, invasive-pressure-control, and invasive-dual-control ventilation), three models of lung mechanics (normal and restrictive and obstructive lung disease), a range of airway leak values, two positive end-expiratory pressure values, and two body weights for each ventilator. Ventilators with and without leak compensation were studied. Measurements and main results: In the absence of airway leak, all ventilators reported tidal volume accurately. In the presence of airway leak, the error of reported tidal volume increased for all ventilators without a leak compensation algorithm while ventilators with leak compensation that included estimation of tidal volume accurately reported tidal volume. In the presence of airway leak, clinically significant effects on the error of reported tidal volume by ventilators with leak compensation were associated with the choice of ventilator in all modes and with lung mechanics in invasive ventilation modes. Conclusions: Reported tidal volume is affected by the presence of airway leak, but in many ventilators a leak compensation algorithm that includes estimation of tidal volume can correct for the discrepancy between actual and reported tidal volume. However, even in ventilators with leak compensation, choice of ventilator and lung mechanics in invasive ventilation modes have a significant effect on error of reported tidal volume.
Article
Background: Although the ICU is the most appropriate place to care for mechanically ventilated patients, a considerable number are ventilated in general medical care wards all over the world. However, adverse events focusing on mechanically ventilated patients in general care have not been explored. Methods: Data from the Japan Council for Quality Health Care database were analyzed. Patient safety incidents from January 2010 to November 2017 regarding mechanical ventilation were collected, and comparisons of patient safety incidents between ICUs/high care units (HCUs) and general care wards were made. Results: We identified 261 adverse events (with at least 20 adverse events resulting in death) and 702 near-miss events related to mechanical ventilation in Japan between 2010 and 2017. Furthermore, among all adverse events, 19% (49 of 261 events) caused serious harm (residual disability or death). Human-factor issues were most frequent in both ICU/HCU and general care settings (55% and 53%, respectively), while knowledge-based errors were higher in the general care setting. Conclusions: Human-factor issues were the most frequent reasons in both settings, while knowledge-based error rates were higher in general care. Our results suggest that proper education and training is needed to minimize patient safety incidents in facilities without respiratory therapists.