Synthetic bone models are used to train surgeons as well as to test new medical devices. However, currently available models do not accurately mimic the complex structure of trabecular bone, which can provide erroneous results. This study aimed to investigate the suitability of stereolithography (SLA) to produce synthetic trabecular bone. Samples were printed based on synchrotron micro-computed tomography (micro-CT) images of human bone, with scaling factors from 1 to 4.3. Structure replicability was assessed with micro-CT, and mechanical properties were evaluated by compression and screw pull-out tests. The overall geometry was well-replicated at scale 1.8, with a volume difference to the original model of <10%. However, scaling factors below 1.8 gave major print artefacts, and a low accuracy in trabecular thickness distribution. A comparison of the model–print overlap showed printing inaccuracies of ~20% for the 1.8 scale, visible as a loss of smaller details. SLA-printed parts exhibited a higher pull-out strength compared to existing synthetic models (Sawbones ™), and a lower strength compared to cadaveric specimens and fused deposition modelling (FDM)-printed parts in poly (lactic acid). In conclusion, for the same 3D model, SLA enabled higher resolution and printing of smaller scales compared to results reported by FDM.
Content may be subject to copyright.
materials
Article
The Potential of Stereolithography for 3D Printing of Synthetic
Trabecular Bone Structures
Ana Grzeszczak 1, * , Susanne Lewin 1, Olle Eriksson 2, Johan Kreuger 2and Cecilia Persson 1


Citation: Grzeszczak, A.; Lewin, S.;
Eriksson, O.; Kreuger, J.; Persson, C.
The Potential of Stereolithography for
3D Printing of Synthetic Trabecular
Bone Structures. Materials 2021,14,
3712. https://doi.org/10.3390/
ma14133712
Academic Editor: Ana Paula Piedade
Received: 31 May 2021
Accepted: 28 June 2021
Published: 2 July 2021
Publisher’s Note: MDPI stays neutral
with regard to jurisdictional claims in
published maps and institutional affil-
iations.
Copyright: © 2021 by the authors.
Licensee MDPI, Basel, Switzerland.
This article is an open access article
distributed under the terms and
conditions of the Creative Commons
Attribution (CC BY) license (https://
creativecommons.org/licenses/by/
4.0/).
1Department of Materials Science and Engineering, Uppsala University, 751 21 Uppsala, Sweden;
susanne.lewin@angstrom.uu.se (S.L.); Cecilia.Persson@angstrom.uu.se (C.P.)
2Department of Medical Cell Biology, Uppsala University, 751 23 Uppsala, Sweden;
olle.eriksson@mcb.uu.se (O.E.); johan.kreuger@mcb.uu.se (J.K.)
*Correspondence: ana.grzeszczak@angstrom.uu.se; Tel.: +46-760-376-722
Abstract:
Synthetic bone models are used to train surgeons as well as to test new medical devices.
However, currently available models do not accurately mimic the complex structure of trabecu-
lar bone, which can provide erroneous results. This study aimed to investigate the suitability of
stereolithography (SLA) to produce synthetic trabecular bone. Samples were printed based on
synchrotron micro-computed tomography (micro-CT) images of human bone, with scaling factors
from 1 to 4.3. Structure replicability was assessed with micro-CT, and mechanical properties were
evaluated by compression and screw pull-out tests. The overall geometry was well-replicated at
scale 1.8, with a volume difference to the original model of <10%. However, scaling factors below 1.8
gave major print artefacts, and a low accuracy in trabecular thickness distribution. A comparison of
the model–print overlap showed printing inaccuracies of ~20% for the 1.8 scale, visible as a loss of
smaller details. SLA-printed parts exhibited a higher pull-out strength compared to existing synthetic
models (Sawbones
), and a lower strength compared to cadaveric specimens and fused deposition
modelling (FDM)-printed parts in poly (lactic acid). In conclusion, for the same 3D model, SLA
enabled higher resolution and printing of smaller scales compared to results reported by FDM.
Keywords:
stereolithography; trabecular bone; additive manufacturing; 3D printing; mechani-
cal properties
1. Introduction
As reported by statistics on hospital-based care in the United States, the number of
orthopaedic procedures concerning all age groups amounted to 3,399,600 in 2005 [
1
]. Bone
models are used by surgeons during preoperative planning of such procedures, and in
visual and haptic surgical simulations for training in low-stress environments [
2
,
3
]. Bone
models are also needed to test and design medical equipment, such as orthopaedic screws
and implants. However, the development of synthetic models with high resemblance
to natural bone remains challenging due to the unique properties and structure of bone
tissue [
4
]. Although cadaveric specimens may provide the most accurate representation
of the mechanical behaviour of bone, reproducible results are difficult to achieve due to
differences among individuals, anatomical sites and medical conditions [
5
]. Synthetic
models could provide more consistent morphologies, increase repeatability and be more
readily available [
6
]. With respect to ethical concerns, such models could help reduce
the number of animals used in experiments [
7
]. In addition, synthetic models could help
reduce the experimental time and cost by avoiding the need for ethical approval and
allowing for faster sample preparation. Furthermore, synthetic samples do not need to
be handled under special conditions, such as controlled temperature or immersion in
phosphate-buffered saline solution.
The currently available synthetic models do not accurately mimic the inherent highly
porous and complex structure of trabecular bone [
4
], neither in terms of internal structure
Materials 2021,14, 3712. https://doi.org/10.3390/ma14133712 https://www.mdpi.com/journal/materials
Materials 2021,14, 3712 2 of 17
nor in terms of mechanical properties (Table 1) [
5
,
6
]. Poukalova et al. [
8
] investigated the
relationship between trabecular microstructure and mechanical properties (elastic modulus,
compressive strength and suture anchor pull-out strength) in both cadaveric specimens
and in commonly used synthetic trabecular bone models (Sawbones
). Several microstruc-
tural properties differed between the cadaveric bone and the Sawbones
, e.g., trabecular
thickness. The cadaveric specimens showed a significant correlation between trabecular
thickness and elastic modulus, but the same correlation was not found for the Sawbones
.
This result was explained by the limited variability in trabecular thickness, even when using
Sawbones
with different densities. This showed that the use of commercially available
synthetic bone may not be suitable for specific applications and that the bone structure is an
important aspect to replicate as it has a significant influence on the mechanical properties.
In another study, Patel et al. [
6
] tested low-density Sawbones
mimicking osteoporotic
bone and proved these models to be unsuitable to represent osteoporotic cancellous bone
in certain aspects, e.g., energy dissipation effects. Sawbones
synthetic bone models
are produced from polyurethane foam, and the ability to influence the microstructural
properties such as trabecular thickness is limited. In response to these limitations, additive
manufacturing techniques are emerging as an attractive way of producing the desired bone
models [
9
]. Indeed, the fabrication of biomimetic bone scaffolds has been the object of
extensive research over the past two decades, with specific design considerations such as
biodegradability, mechanical similarity and control over the pores’ size [
10
]. Similarly, bio-
printing strategies have been extensively investigated to demonstrate the ability to create
vascularised cell-laden bone constructs [
10
]. However, techniques allowing for printed
structures with both geometrical accuracy of microstructural features and desired mechani-
cal properties remain to be developed. Yang et al. reported on scaffold fabrication processes
for the repair of large segmental bone defects, and especially on the current characteristics
of several typical additive manufacturing techniques, their advantages and their limita-
tions [
11
]. The latter are mainly related to insufficient densification of the printed material,
low mechanical strength, low resolution or poor surface quality [
11
].
Alsheghri et al. [12]
encountered challenges comparing mechanical tests results with numerically predicted
mechanical properties at small scales due to printing defects, and required printing scaled
up geometries. Thus, it appears necessary to investigate printing techniques that allow
for high printing resolution. Wu et al. [
13
] evaluated the possibility of using fused deposi-
tion modelling (FDM) to produce trabecular structure. The technique showed potential,
but limitations were encountered in printing resolution. Furthermore, they observed an
unsuitable effect of the layer-by-layer construction process on mechanical properties.
In a review, Szymczyk-Ziólkowska et al. [
14
] reported that recent additive manufac-
turing technologies can enable the precise manufacturing of porous scaffolds, yet this is re-
stricted by the resolution achievable with the different available materials. Ng et al. [
15
] and
Li et al. [
16
] performed reviews on the potential of vat-polymerisation printing techniques
(stereolithography (SLA), digital light processing, two-photon polymerisation) for bioprint-
ing and tissue engineering applications. They highlighted that the vat-polymerisation
printing techniques allow for a superior resolution as compared to other 3D printing meth-
ods. However, despite very promising recent studies, the highly complex bone structure
has not yet been accurately replicated with bioprinting techniques [
15
]. Chartrain et al. [
17
]
also showed that the vat-polymerisation techniques allow for repeatability and precision,
thus presenting a significant advantage over traditional scaffold fabrication techniques,
even if the resolution is often achieved at the expense of the size of the part that can be
printed. Other techniques, for example solvent casting or particulate leaching, present
drawbacks such as the inability to control pore shape and pore interconnectivity [
17
]. Vat-
polymerisation appears promising, but only simple architectures have been investigated
with the most precise printers [
17
]. Mézière et al. [
18
] used SLA, which is a 3D printing
method using a photo-reactive, resin-solidified, point-by-point by a focused laser beam, to
print horse trabecular bone models and perform ultrasound measurements. The models
were printed with reasonable accuracy, but their mechanical properties were different
Materials 2021,14, 3712 3 of 17
from those of real bone. Barak and Black [
19
] used a feeding method consisting of a spray
of droplets of photopolymerizable material in a Poly-Jet device coupled with a model
reconstructed from micro-CT scans of chimpanzee bone, to study changes in bone mor-
phology and the effect of trabecular thickness on mechanical properties. This model was
chosen because chimpanzees have thicker trabeculae compared to humans and a larger
thickness was preferable, given the limitations in printing resolution. Printing equipment
and materials allowing for a higher resolution remain to be assessed.
Table 1. Mechanical properties of cortical and trabecular bone.
Bone Type Porosity (%) Compressive
Strength (MPa)
Elastic Modulus
(GPa)
Fernandez-Cervantes et al.
[20]Milazzo et al. [21] Milazzo et al. [21]
Cortical <10 130–180 12–18
Trabecular 50–90 4–12 0.1–0.5
This study aimed to evaluate SLA as a means of producing structures replicating
human trabecular bone, including a volumetric comparison with the model for a highly
detailed assessment of the printed parts, with the following objectives:
To acquire data on the suitability of SLA technology for the intended application.
To evaluate replicability in trabecular spacing, thickness and total porosity, with
micro-computed tomography (micro-CT) analysis.
To evaluate mechanical properties with compression testing and screw pull-out tests.
2. Materials and Methods
2.1. SLA Printing
The 3D bone model used for this project was generated during a previous study, and
the process has been described in detail elsewhere [
13
]. A short summary is provided
here for clarity and context. The 3D images were obtained through synchrotron micro-CT
scanning of a human femoral head sample. The raw images had a resolution of 3.25
µ
m.
A series of image processing steps was applied for segmentation of only the bony part.
Unconnected debris resulting from bone cutting were removed. Finally, the segmented
bone was saved as a printable stl file. A 3D mesh of a cylindrical region of interest was
generated with MATLAB (Natick, MA, USA). Due to the size and large number of images,
and the associated computational burden, the processing steps resulted in a reduction in
resolution by a total factor of 8, with the voxel size being ultimately equal to 26 µm.
The 3D model of the bone was imported into the slicing software Preform (Form-
labs, Somerville, MA, USA). 3D printing was performed on Form2 and Form3 (Formlabs,
Somerville, MA, USA) SLA printers. The printed material is a resin made out of a thermoset
polymer, composed of a mixture of methacrylic acid esters, photo-initiators, proprietary
pigment and additive package. It contains 75% to 90% of methacrylated oligomer, 25% to
50% of methacrylated monomer and traces of diphenyl (2,4,6-trimethylbenzoyl) phosphine
oxide (<1%) [
22
]. Commercially available resins have different characteristics and potential,
as specified by the manufacturer, and some of the properties of Formlabs’ resins can be
found in Table 2.
Materials 2021,14, 3712 4 of 17
Table 2.
Elastic modulus for parts printed with Formlabs resin [
23
]. “Green” parts refer to parts that
have not been through any post-processing treatment. “Cured” parts refer to parts that have been
through the suggested best post-process treatment (UV-light and heat).
Formlabs Resin Elastic Modulus for “Green”
Parts (GPa)
Elastic Modulus for “Cured”
Parts (GPa)
Black 1.6 2.8
GreyPro 1.4 2.6
Durable 0.45 1.26
Once the printing process was finished, the parts were rinsed in isopropyl alcohol
(IPA), which dissolves the remaining uncured resin of the surfaces of the part. Then, the
parts were post-cured using UV-light to improve and stabilise the mechanical properties,
in accordance with the manufacturer’s recommendations (Table 3). UV-light exposure
triggers the formation of additional chemical bonds in the cured resin and makes the part
stronger and stiffer, while heat exposure accelerates the process of bond formation [
23
,
24
].
Table 3.
Recommended UV-cure for parts printed with Black, GreyPro and Durable resins (Formlabs,
Somerville, MA, USA).
Resin Recommended Cure Time (min) Recommended Cure Temperature (C)
Black 30 60
GreyPro 15 80
Durable 60 60
Available resins from Formlabs were used without changing their composition. The
Black resin, which is among the standard resins, was expected to provide the best results
in terms of structural accuracy and printed details due to its opacity preventing the laser
light from penetrating deeply into the resin, and thus preventing artefact formation due to
over-cured resin. Its fluidity also allows the liquid resin to properly flow out of the part
during the printing process. To evaluate the SLA printer capability, the three-dimensional
trabecular bone model previously mentioned was used as an input model in a Form3
printer. This model was scaled up in Preform with the following scaling factors: 1 (original
size), 1.3, 1.5, 1.8, 2, 2.3, 2.5, 2.8, 3, 4 and 4.3. The prints were configured with a resolution
of 25
µ
m and the addition of external support structures that were required due to the
overhanging parts of the model.
To assess the mechanical properties of the printed material, standard dense cylin-
ders with a diameter of 6 mm and a height of 12 mm were printed, following ISO 5833
Standard [25] and allowing for standard compression testing.
2.2. Mechanical Testing
2.2.1. Compression Tests
Compression tests were conducted on a Shimadzu AGS-X universal testing machine
(Shimadzu, Kyoto, Japan), associated with the complementing software package Trapez-
iumX (Shimadzu, Kyoto, Japan). The displacement measurement was provided by the
crosshead position. The compliance was measured as the inverse of the setup stiffness and
was corrected in the displacement data for the final results. The load cell had a capacity of
5 kN. The same setup was used in all mechanical testing.
Uniaxial compression tests were performed as preliminary tests on cylindrical dense
specimens (6 mm in diameter and 12 mm in height) printed in Black resin, at a crosshead
speed of 1 mm/min. The data collected before the force reached 2 N were removed. The
stiffness of the samples was defined as the slope of the straight line that fits the force–stroke
curve in the linear region, which was delineated independently for each curve. The ma-
chine compliance value was used to correct the stiffness, and then the elastic modulus was
calculated. These preliminary tests (details are provided in the Supplementary Material,
Materials 2021,14, 3712 5 of 17
Table S1, Figure S1) followed an experimental plan to test and choose the printing param-
eters. Each batch included 5 samples. The influence of the following parameters on the
mechanical properties was investigated: printer type, orientation, resolution, UV-curing
time and preload during the test. The parameters of the control batch were defined as the
parameters recommended by the supplier. The expected elastic modulus for the control
group was 2.8 GPa (Table 2), but the test result was significantly lower (1879
±
51 MPa).
This elastic modulus is lower than the elastic modulus range of values for cortical bone,
as described in Table 1(12–18 GPa [
21
]). The elastic modulus of the control group did
not appear to be different from the ones of the other batches, except for the batches of the
horizontally printed samples. However, in trabecular samples, the initial orientation of
the sample may be less important than in dense structures considering that the trabecular
structures are oriented in different directions. Finally, it was concluded that the mechanical
properties do not seem to be adjustable with the print settings.
Another set of preliminary tests (accessible in the Supplementary Material, Figure S2)
using the same equipment and procedure was conducted to assess the influence of post-
treatment on the SLA-printed parts. Standard dense cylinders (6 mm in diameter and
12 mm in height) were printed in GreyPro and Durable resin, and only half of the samples
underwent post-processing. UV-light and heat exposures were defined according to the
supplier’s recommendations. The results showed that the post-treatment greatly increases
the stiffness of the parts. However, the elastic moduli obtained with these compression
tests were lower than the expected elastic moduli, as provided by the supplier Formlabs
(Table 2). The manipulation of untreated samples proved difficult and implied constraints
that do not seem acceptable for the intended use of these materials. We decided to use
treated specimens as they are more stable.
Uniaxial compression tests were also performed on trabecular samples printed in
Black resin at the scales 1.8, 2, 4 and 4.3, and on samples printed at the scale 4.3 that
underwent micro-CT. Each batch included 4 samples, except for the unscanned specimens
at the scale 4.3 which included 3 samples. As for the standard cylinders, the crosshead
speed was 1 mm/min and the stiffness of the samples was defined as the slope of the
straight line that fits the force–stroke curve in the linear region, which was delineated
independently for each curve. The machine compliance value was used to correct the
stiffness, and then the elastic modulus was calculated.
2.2.2. Pull-Out Tests
The same testing machine was used for pull-out tests, with a crosshead speed of
1 mm/min and a 5 kN load cell. The 6 tested samples were printed with Black resin on
a Form3 printer, with a scale factor of 4.3, a resolution of 25 microns and the addition of
external support structures. This scale factor was chosen to allow comparison with data
from previous studies using this factor [
13
]. An orthopaedic screw with the following
characteristics was used: 6.7 mm outer diameter, 3.5 mm inner diameter, 3.6 mm thread
length, 9 mm overall length and 2.8 mm pitch distance. Prior to screw insertion, a pilot hole
of 3.5 mm diameter was drilled in the trabecular samples. This operation was performed
with machine drilling to ensure correct positioning and fixation of the samples during
drilling. The screw was then inserted manually with an insertion depth of 20 mm. The
screw was mounted on a steel holder that was attached to the load cell, which in turn was
attached to the mobile crosshead. The sample was held by another steel holder, attached to
the fixed crosshead. A diagram of the setup can be seen in Figure 1.
Materials 2021,14, 3712 6 of 17
Materials 2021, 14, x FOR PEER REVIEW 6 of 17
Figure 1. A cross-sectional illustration of the pull-out setup. The loading direction is indicated by
red arrows.
2.2.3. Statistical Analysis
Statistical analysis was conducted using SPSS Statistics (IBM, Armonk, NY, USA).
The test results are presented as mean and standard deviation. Levene’s test was used to
assess homogeneity of variances. If Levene’s test was non-significant, an analysis of vari-
ance (ANOVA) was conducted to analyse the differences among the group means. If the
ANOVA was significant, Scheffe’s post-hoc test was carried out. On the other hand, if
Levene’s test was found significant, Welch’s test was conducted to analyse the differences
among group means. Then, Tamhane’s post-hoc test was carried out, to analyse the dif-
ferences between pairs of groups. The results were considered significant if p < 0.05.
2.3. Morphological Evaluation by Micro-CT
3D evaluation of the printed samples was performed using a micro-CT scanner, Sky-
Scan 1172 (Bruker, Billerica, Massachusetts, USA). The samples were scanned with a voxel
size of 12 µm using a voltage of 60 kV and a current of 167 µA. An exposure time of 1100
ms and a rotation step of 0.4° were also used. The same settings were used for all the
printed samples, to allow for an unbiased comparison of the results between samples of
different dimensions. A different set of parameters was only used to scan the parts that
were printed at the original scale. Due to the smaller size of these samples, an increased
resolution was necessary to be able to identify internal structures. The chosen settings for
these small parts were a voxel size of 5 µm, a voltage of 60 kV, a current of 167 µA, an
exposure time of 760 ms and a rotation step of 0.4°. Samples at the original scale printed
in all different resins were compared. The purpose was to determine the accuracy that
could be reached for each resin.
Due to time limitations, only one specimen of each scale was scanned, except for
those that were to be studied in more detail. The list of the scanned samples can be found
in Table 4. Visual assessment revealed that the accuracy of the trabecular structures de-
clined for scaling factors below 2, and the small structures were most likely to exhibit
errors during printing, so it was decided to scan two samples of the scales 1.3, 1.5 and 1.8,
to have more representative results for these scales. Four samples with the scale factor 4.3
were scanned, with this scale being the one chosen during previous studies using FDM
[13], and having more representative results on this scale allowed for data comparison.
Figure 1.
A cross-sectional illustration of the pull-out setup. The loading direction is indicated by
red arrows.
2.2.3. Statistical Analysis
Statistical analysis was conducted using SPSS Statistics (IBM, Armonk, NY, USA). The
test results are presented as mean and standard deviation. Levene’s test was used to assess
homogeneity of variances. If Levene’s test was non-significant, an analysis of variance
(ANOVA) was conducted to analyse the differences among the group means. If the ANOVA
was significant, Scheffe’s post-hoc test was carried out. On the other hand, if Levene’s test
was found significant, Welch’s test was conducted to analyse the differences among group
means. Then, Tamhane’s post-hoc test was carried out, to analyse the differences between
pairs of groups. The results were considered significant if p< 0.05.
2.3. Morphological Evaluation by Micro-CT
3D evaluation of the printed samples was performed using a micro-CT scanner,
SkyScan 1172 (Bruker, Billerica, MA, USA). The samples were scanned with a voxel size
of 12
µ
m using a voltage of 60 kV and a current of 167
µ
A. An exposure time of 1100 ms
and a rotation step of 0.4
were also used. The same settings were used for all the printed
samples, to allow for an unbiased comparison of the results between samples of different
dimensions. A different set of parameters was only used to scan the parts that were printed
at the original scale. Due to the smaller size of these samples, an increased resolution was
necessary to be able to identify internal structures. The chosen settings for these small
parts were a voxel size of 5
µ
m, a voltage of 60 kV, a current of 167
µ
A, an exposure time
of 760 ms and a rotation step of 0.4
. Samples at the original scale printed in all different
resins were compared. The purpose was to determine the accuracy that could be reached
for each resin.
Due to time limitations, only one specimen of each scale was scanned, except for those
that were to be studied in more detail. The list of the scanned samples can be found in
Table 4. Visual assessment revealed that the accuracy of the trabecular structures declined
for scaling factors below 2, and the small structures were most likely to exhibit errors
during printing, so it was decided to scan two samples of the scales 1.3, 1.5 and 1.8, to have
more representative results for these scales. Four samples with the scale factor 4.3 were
scanned, with this scale being the one chosen during previous studies using FDM [
13
], and
having more representative results on this scale allowed for data comparison.
Materials 2021,14, 3712 7 of 17
Table 4.
List of samples analysed by micro-CT. The samples were printed vertically, unless other-
wise indicated.
Resin Scale Factor Number of Specimens Scanned
Black
1 1
1.3 2
1.5 2
1.8 2
2 1
2.3 1
2.5 1
2.8 1
3 1
41 horizontally printed
1 vertically printed
4.3 4
After scanning, the images were reconstructed using NRecon (Bruker, Billerica, MA,
USA) and analysed using CTAn (Bruker, Billerica, Massachusetts, USA). The images were
pre-processed by a median filter in order to remove noise. Then, the printed material
was segmented using an automated Ridler and Calvard process [
26
]. Following the seg-
mentation, parts that were not connected were removed and a morphological opening
was applied to remove noise. Quantitative analysis was conducted on the segmented
volumes in terms of the ratio bone volume/total volume (BV/TV) and trabecular thickness
distribution. Trabecular thickness was calculated with 3D analysis in CTAn, using the
“sphere-fitting” method by Hildebrand and Rüegsegger [
27
], i.e., the thickness is set as the
diameter of the largest sphere that fits a solid trabecula. The trabecular thickness values
were extracted as ranges, with mid-range values and the associated volume of trabeculae
(in µm3or in %) that are measured in this range.
A volumetric comparison of the micro-CT images was also conducted using 3D Slicer
4.10 [
28
]. The binary micro-CT image dataset which had been used to create the 3D-model
of the bone was imported, together with the binary micro-CT image dataset of the printed
samples. Datasets from three printed volumes were imported: scale 4.3, scale 1.8 and
scale 1.5. For a better comparison, the model was scaled to fit the scale of the printed
samples. A landmark-based rigid registration was made to overlap the images of the
model and the micro-CT images of the printed samples, and a cylindrical region of interest
was created through the volumes (diameter 40 mm). The total volume of printed material
was then quantified in the region of interest.
After registration, the volumes of the scaled model and print overlapped. The differ-
ences between the model and each print could be observed in more detail after registration,
and, more specifically, if material represented in the original model was missing in the
printed part (hereinafter referred to as underfilling), and if material present in the printed
part was not present in the model (hereinafter referred to as overfilling). To quantify
the over- and under-filling, the volume from the model was subtracted from each print
(overfilling), and conversely, the volume from each print was subtracted from the model
(underfilling). In addition, overlapping 3D models were created to allow for visual com-
parison of the overfilling.
Materials 2021,14, 3712 8 of 17
3. Results and Discussion
3.1. Mechanical Testing
3.1.1. Compression Tests
Compression tests were conducted on trabecular samples scaled up with different
factors, printed in Black resin, and the resulting elastic moduli were measured. Porosity
was not accounted for in the calculation of the elastic modulus from the stiffness; instead,
an apparent elastic modulus was obtained. The batches did not include enough samples to
draw conclusions on any variation of the apparent elastic modulus with the scale factors.
However, a significant difference in elastic modulus was observed between samples that
were scanned or not. The average elastic modulus for samples printed at the scale 4.3 was
54.11
±
1.92 MPa, and the average elastic modulus for samples printed at the scale 4.3
that underwent micro-CT scanning was 106.46
±
13.06 MPa. Thus, micro-CT scanning
significantly increases the elastic modulus of the parts. A typical test curve of a 4.3 scale
sample and one of a scanned 4.3 scale sample are shown in Figure 2. These curves were
chosen as representative of their respective batches. The two types of samples exhibited a
very different behaviour during the compression test. This can be explained from the X-ray
exposure during micro-CT. One hypothesis is that X-ray irradiation causes the formation
of cross-links but also breaks bonds, thus changing the properties of the material, which
becomes brittle [
29
,
30
]. This phenomenon will have to be considered for possible future
studies, in order to avoid altered results.
Materials 2021, 14, x FOR PEER REVIEW 8 of 17
was not accounted for in the calculation of the elastic modulus from the stiffness; instead,
an apparent elastic modulus was obtained. The batches did not include enough samples
to draw conclusions on any variation of the apparent elastic modulus with the scale fac-
tors. However, a significant difference in elastic modulus was observed between samples
that were scanned or not. The average elastic modulus for samples printed at the scale 4.3
was 54.11 ± 1.92 MPa, and the average elastic modulus for samples printed at the scale 4.3
that underwent micro-CT scanning was 106.46 ± 13.06 MPa. Thus, micro-CT scanning sig-
nificantly increases the elastic modulus of the parts. A typical test curve of a 4.3 scale sam-
ple and one of a scanned 4.3 scale sample are shown in Figure 2. These curves were chosen
as representative of their respective batches. The two types of samples exhibited a very
different behaviour during the compression test. This can be explained from the X-ray
exposure during micro-CT. One hypothesis is that X-ray irradiation causes the formation
of cross-links but also breaks bonds, thus changing the properties of the material, which
becomes brittle [29,30]. This phenomenon will have to be considered for possible future
studies, in order to avoid altered results.
Figure 2. Compression tests of two trabecular samples, representative of their respective batches.
One sample (red curve) went through micro-CT before compression. Both were printed in Black
resin at the scale 4.3.
Tests on standard dense cylinders showed that, for a specific resin, mechanical prop-
erties cannot be efficiently controlled by printing and post-treatment parameters. The
main resin used for this project was the Black resin from Formlabs. Its composition was
not modified, and it was used as recommended by the supplier. The control group, that
was composed of samples processed according to the supplier’s recommendations, exhib-
ited an elastic modulus of 1879 ± 51 MPa. The values exhibited by the other groups are
available in the Supplementary Material (Figure S1). In future studies, investigations on
the influence of resin composition would be of interest. Indeed, it would be advantageous
to be able to influence the mechanical properties of the samples by changing the compo-
sition of the resin, and thus obtain properties closer to the ones of bone (Table 1). Further-
more, a biocompatible resin could be investigated and evaluated with regard to cell com-
patibility, which would be a milestone towards in vivo application as bone void-filling
scaffolds.
Figure 2.
Compression tests of two trabecular samples, representative of their respective batches.
One sample (red curve) went through micro-CT before compression. Both were printed in Black resin
at the scale 4.3.
Tests on standard dense cylinders showed that, for a specific resin, mechanical proper-
ties cannot be efficiently controlled by printing and post-treatment parameters. The main
resin used for this project was the Black resin from Formlabs. Its composition was not
modified, and it was used as recommended by the supplier. The control group, that was
composed of samples processed according to the supplier’s recommendations, exhibited an
elastic modulus of 1879
±
51 MPa. The values exhibited by the other groups are available in
the Supplementary Material (Figure S1). In future studies, investigations on the influence
of resin composition would be of interest. Indeed, it would be advantageous to be able
to influence the mechanical properties of the samples by changing the composition of
the resin, and thus obtain properties closer to the ones of bone (Table 1). Furthermore, a
Materials 2021,14, 3712 9 of 17
biocompatible resin could be investigated and evaluated with regard to cell compatibility,
which would be a milestone towards in vivo application as bone void-filling scaffolds.
3.1.2. Pull-out Tests
Pull-out tests were performed on six trabecular samples, printed in Black resin at the
scale 4.3. Results from a previous study [
13
] are also displayed in Table 5. Although the
stiffness of the SLA samples seems lower than the stiffness of the samples printed by FDM,
the FDM results showed large variations within the tested batches, and no statistically
significant difference was found.
Table 5.
Pull-out test results comparison between present study (screw diameter of 6.7 mm) and
previous study (screw diameter of 6.5 mm), using the same input model, the same scale factor (4.3)
and the same type of screw, with an insertion depth of 20 mm. The same test setup was used for both
fused deposition modelling (FDM) and stereolithography (SLA) printed samples. The mean values
and standard deviations are presented.
Study Samples Stiffness (N/mm) Pull-Out Strength (N)
Present study
Printed by SLA Black resin 819.3 ±127.8 755.6 ±146.8
Wu et al. [13]
Printed by FDM
Pure poly(lactic acid)
(PLA) 931.1 ±82.3 1369.7 ±32.9
PLA + 5%
hydroxyapatite (HA) 1295.2 ±283.3 1657 ±446.9
PLA + 10% HA 1374.7 ±140.1 1891.3 ±223.8
PLA + 15% HA 1374.6 ±514.8 1555 ±613.3
A list of different studies and their respective maximum pull-out force are displayed
in Table 6. The pull-out strength values were scaled by Wu et al. [
13
] to adapt them to a
screw diameter of 6.5 mm and an insertion depth of 20 mm. A linear relation between the
pull-out load, screw diameter and insertion depth was assumed [
31
]. The screw diameter
used in the present study (6.7 mm) was very close to 6.5 mm, so it was assumed that scaling
was not required.
Table 6. Comparison of pull-out results from the literature.
Study Material Insertion
Depth (mm)
Screw Diameter
(mm)
Pull-out
Strength (N)
Scaled Pull-Out
Strength (N)
Present study SLA-printed resin 20 6.7 755.6 ±146.8 755.6 ±146.8
Shea et al.,
2015 [32]
Sawbones™, density
0.08 g/cm345 6.5 393 ±9 147 ±4
Mueller et al., 2013
[33]
Sawbones™, density
0.16–0.32 g/cm322 12.5 700–2500 305–1127
Mueller et al., 2013
[33]
Human femoral head
(screw pushed inwards
in the specimen)
22 12.5 2489–4347 1176–2055
Wu et al., 2020 [34] Human femoral head 9 4 314–635 1134–2293
Wu et al., 2020 [13]3D-printed
PLA/HA 20 6.5 1400–1900 1400–1900
Shea et al. [
32
] tested pull-out in Sawbones
with a density of 0.08 g/cm
3
. This
foam appears to exhibit a lower pull-out strength than the models tested in the present
study. As can be seen in Table 5, the pull-out strength obtained with the SLA-printed
trabecular models of the present study is lower than the results obtained for cadaveric
samples representing the same bone location (femoral head). It is important to mention
Materials 2021,14, 3712 10 of 17
that variations could be caused by the pre-mortal condition of the cadaveric specimens.
In a study, Palmer et al. [
35
] performed pull-out tests on Sawbones
samples, using
a similar setup. The pull-out was conducted with 3.2 drill bit, 6.5 mm screw diameter
and 20 mm insertion in 40 mm tall samples. The foam supposed to mimic healthy bone
(0.24 g cm
3
) was used as a control, and the maximum pull-out force achieved was 40 N.
This is significantly lower than the pull-out force obtained for SLA-printed parts in the
present study: 755.6
±
146.8 N. Wu et al. [
13
] conducted a study with models printed in
FDM and based on the same input 3D model as the one used in the present study. The
pull-out strength achieved (1400–1900 N) was found to be higher than the value found
in the current project. In another study, Wu et al. [
34
] conducted pull-out tests on human
femoral head with an average BT/TV ratio of 27.6
±
6.1, and obtained significantly higher
pull-out strength compared to the present study (BV/TV ratio estimated to be 39.5%).
To conclude, the values acquired on SLA-printed samples in the present study were higher
than previous results obtained with Sawbones
. The results were also lower than previous
results obtained with cadaveric specimens and FDM-printed samples.
In the present study, pull-out was only tested for one group of samples (scale 4.3).
Other scaling factors and screws sizes could be tested in future studies. Further pull-out
tests could be conducted for samples at different scales, as only samples at the scale 4.3
were tested in pull-out in the present study. All the trabecular samples were printed from
the same 3D model. It would be interesting to print samples from another model, to
evaluate the variability of the results in printing accuracy and mechanical tests. In addition,
the pull-out test setup is challenging in terms of screw insertion, which needs to be fully
central and vertical. Any deviation in the insertion technique can bias the results, and this
can be a contributing factor to the variability in the results.
3.2. Morphological Evaluation by Micro-CT
3.2.1. Bone Fraction Measured with Micro-CT
The ratio BV/TV was obtained with a 3D analysis in CTAn on the micro-CT images
of the samples. The results are displayed in Figure 3. BV/TV was significantly higher for
the small scales (66.1% for scale 1, 66.7% for scale 1.3) than for the bigger scales (33.1% for
scale 4, 32.4% for scale 4.3). In other terms, the small samples had a much lower porosity.
A decreasing trend is visible, i.e., the BV/TV ratio seems to decrease with the increase of
the scale factor (Figure 3).
Materials 2021, 14, x FOR PEER REVIEW 11 of 17
Figure 3. Bone fraction BV/TV (%) as provided after micro-CT and 3D analysis, for samples
printed in Black resin at different scales.
3.2.2. Trabecular Thickness Distribution
The trabecular thickness distribution is shown in Figure 4. To allow for a comparison
between the different scales, mid-range values were scaled down by the scaling factor of
their respective samples, to match the original size. Each sample’s trabecular thickness
distribution was plotted in a histogram and represented as the percent volume in each
range depending on the scaled down mid-range values. However, as the objective was to
overlay the distribution from the different samples, the histograms were transformed into
curves for better readability. The overlapped distribution curves for the different samples
can be seen in Figure 4. The trabecular thickness distribution of the input model was not
available, so the sample printed at scale 4.3 was taken as the control. For the control group,
the distribution had approximately the shape of a bell curve. This silhouette was also
found for samples with scales greater than or equal to 1.8. The curves appear similar but
shift and grow, following the decrease in scale factor. For the 1.5 scale distribution, the
curve was very shifted and became wider. For the 1.3 and 1 (original) scales, the curves
looked very different from the control group. The values were no longer grouped together
as in a bell curve but were distributed with a lot of variation. Notably, these samples had
fewer thin but also much thicker trabecular structures. This is most likely due to samples
with the scale factors below 1.8 containing printing inaccuracies. From this evaluation, it
could be concluded that scales 1.8 or 2 are candidates to be the smallest scale for printing
accurate structures with this 3D model.
Figure 3.
Bone fraction BV/TV (%) as provided after micro-CT and 3D analysis, for samples printed
in Black resin at different scales.
Materials 2021,14, 3712 11 of 17
The BV/TV ratio of the input model was estimated to be 39.5%. Prior to conducting the
measures, it was assumed that the samples with the bigger scale factor (4.3) would be the
best printed samples. Indeed, with the high resolution of the SLA printer, it was expected
that big structures should be accurately printed, while some artefacts could appear in the
printing of very small features. In fact, for the scales under 1.8, the BV/TV ratio was much
higher, and significant artefacts are likely present in the printed samples. This was further
investigated by studying the trabecular thickness distribution and conducting a volumetric
comparison, as reported in Sections 3.2.2 and 3.2.3.
It is important to mention that porosity is an overall measure that does not provide
details on the location of structures in the sample. It was therefore interesting to study
other parameters to compare the printed samples, for example the trabecular thickness
distribution, as described in the following section.
3.2.2. Trabecular Thickness Distribution
The trabecular thickness distribution is shown in Figure 4. To allow for a comparison
between the different scales, mid-range values were scaled down by the scaling factor of
their respective samples, to match the original size. Each sample’s trabecular thickness
distribution was plotted in a histogram and represented as the percent volume in each
range depending on the scaled down mid-range values. However, as the objective was to
overlay the distribution from the different samples, the histograms were transformed into
curves for better readability. The overlapped distribution curves for the different samples
can be seen in Figure 4. The trabecular thickness distribution of the input model was not
available, so the sample printed at scale 4.3 was taken as the control. For the control group,
the distribution had approximately the shape of a bell curve. This silhouette was also
found for samples with scales greater than or equal to 1.8. The curves appear similar but
shift and grow, following the decrease in scale factor. For the 1.5 scale distribution, the
curve was very shifted and became wider. For the 1.3 and 1 (original) scales, the curves
looked very different from the control group. The values were no longer grouped together
as in a bell curve but were distributed with a lot of variation. Notably, these samples had
fewer thin but also much thicker trabecular structures. This is most likely due to samples
with the scale factors below 1.8 containing printing inaccuracies. From this evaluation, it
could be concluded that scales 1.8 or 2 are candidates to be the smallest scale for printing
accurate structures with this 3D model.
Materials 2021, 14, x FOR PEER REVIEW 12 of 17
Figure 4. Trabecular thickness distributions for trabecular samples printed in Black resin. The mean for the distribution of
the trabecular thickness in the 3D model is 246 µm (marked by a red arrow) [13].
3.2.3. Volumetric Comparison
A volumetric comparison was conducted to quantify the overall volumetric differ-
ence between the input model and the printed parts (Table 7). The overall volumetric dif-
ference for the scale 1.5 is large (53.8%), but the difference for the scale 1.8 is deemed ac-
ceptable (8.9%, i.e., <20%). In fact, the volume difference for the scale 4.3 is not much
smaller than the one for scale 1.8, while the scale factor is much larger. There seems to be
a scale limit below which the printing accuracy abruptly degrades, and with this analysis
the limit appears to be 1.8.
Table 7. Quantification of the volumetric differences between the input model and printed parts.
Sample Difference to Model in Total Volume (%)
Scale 4.3 6.1%
Scale 1.8 8.9%
Scale 1.5 53.6%
The difference in volumes was afterwards assessed in more detail by overlapping a
scaled version of the model with the printed parts. First, the model volume was subtracted
from the printed volume, to obtain the volume percentage of the printed part that was not
displayed in the original model (overfilling). Then, the printed volume was subtracted
from the model volume, to obtain the volume percentage of the model that was not
printed (underfilling). The quantitative results can be seen in Table 8, and a visual repre-
sentation of the overfilling in Figure 5. Scales 1.8 and 4.3 again appeared to provide ac-
ceptable results. As expected, the quantity of underfilling and overfilling was larger for
scale 1.8 (23% and 22%) than for scale 4.3 (11% and 15%). However, both scales had a
similar amount of underfilling and overfilling, which resulted in a small difference in
overall volume compared to the model (Table 7). The comparison for scale 1.5 displayed
a large overfilling, as expected according to the results in Table 7. Nevertheless, scale 1.5
instead showed a very low value for the underfilling (5%). This can be explained by the
Figure 4.
Trabecular thickness distributions for trabecular samples printed in Black resin. The mean for the distribution of
the trabecular thickness in the 3D model is 246 µm (marked by a red arrow) [13].
Materials 2021,14, 3712 12 of 17
3.2.3. Volumetric Comparison
A volumetric comparison was conducted to quantify the overall volumetric difference
between the input model and the printed parts (Table 7). The overall volumetric difference
for the scale 1.5 is large (53.8%), but the difference for the scale 1.8 is deemed acceptable
(8.9%, i.e., <20%). In fact, the volume difference for the scale 4.3 is not much smaller than
the one for scale 1.8, while the scale factor is much larger. There seems to be a scale limit
below which the printing accuracy abruptly degrades, and with this analysis the limit
appears to be 1.8.
Table 7. Quantification of the volumetric differences between the input model and printed parts.
Sample Difference to Model in Total Volume (%)
Scale 4.3 6.1%
Scale 1.8 8.9%
Scale 1.5 53.6%
The difference in volumes was afterwards assessed in more detail by overlapping a
scaled version of the model with the printed parts. First, the model volume was subtracted
from the printed volume, to obtain the volume percentage of the printed part that was not
displayed in the original model (overfilling). Then, the printed volume was subtracted
from the model volume, to obtain the volume percentage of the model that was not printed
(underfilling). The quantitative results can be seen in Table 8, and a visual representation of
the overfilling in Figure 5. Scales 1.8 and 4.3 again appeared to provide acceptable results.
As expected, the quantity of underfilling and overfilling was larger for scale 1.8 (23%
and 22%) than for scale 4.3 (11% and 15%). However, both scales had a similar amount
of underfilling and overfilling, which resulted in a small difference in overall volume
compared to the model (Table 7). The comparison for scale 1.5 displayed a large overfilling,
as expected according to the results in Table 7. Nevertheless, scale 1.5 instead showed a
very low value for the underfilling (5%). This can be explained by the fact that if the sample
is very filled, the structures that were in the original model will probably be represented
in the printed model, but this may be due to artefacts. For example, if a comparison was
made with a sample consisting of a completely dense cylinder, the underfilling would be
0%. However, the high value for overfilling (59%) confirms that a lot of extra material was
printed for scale 1.5.
Table 8. Quantification of underfilling and overfilling for printed models of different scales (1.5, 1.8
and 4.3).
Sample
Underfilling (%)
Volume Percentage of the Model
that Was Not Printed
Overfilling (%)
Volume Percentage of the Printed
Part that Was Not in the Original
Model (Extra Volume)
Scale 4.3 11 15
Scale 1.8 23 22
Scale 1.5 5 59
Materials 2021,14, 3712 13 of 17
Materials 2021, 14, x FOR PEER REVIEW 13 of 17
fact that if the sample is very filled, the structures that were in the original model will
probably be represented in the printed model, but this may be due to artefacts. For exam-
ple, if a comparison was made with a sample consisting of a completely dense cylinder,
the underfilling would be 0%. However, the high value for overfilling (59%) confirms that
a lot of extra material was printed for scale 1.5.
Figure 5. Volumetric comparison of the original bone model and the overfilling volumes from the printed samples at the
different scales. The original model is displayed in grey, and the overfilled material in red. Note, the model has been scaled
up to the scale of the printed parts in these comparisons.
Table 8. Quantification of underfilling and overfilling for printed models of different scales (1.5,
1.8 and 4.3).
Sample
Underfilling (%)
Volume Percentage of the
Model that Was Not Printed
Overfilling (%)
Volume Percentage of the Printed Part that
Was Not in the Original Model (Extra Vol-
ume)
Scale 4.3 11 15
Scale 1.8 23 22
Scale 1.5 5 59
A more detailed visual comparison of the overfilling and underfilling could be made
in the overlapped cross-sectional micro-CT images of the 3D model and the printed sam-
ples. Overlapping images can be seen in Figure 6. The structures of the sample in scale 4.3
appear to be accurately printed, as seen in the resulting overlap images. For scale 1.8, some
trabecular structures are too thick compared to the ones of the model, and some anomalies
are visible. For scale 1.5 on the other hand, many trabecular structures appear too thick,
with the result that some pores disappear and the overall structure is quite different from
that of the model. Scale 1.8 appears to be the smallest scale at which it is possible to print
the model with acceptable quality and accuracy. However, some errors are noticeable in
the printed samples and it could be interesting to conduct further studies on this aspect,
to investigate the influence of these artefacts in more detail.
A previous study printed the same trabecular bone model, but with FDM [13]. A
comparison of similar cross-sections of the same part of the model printed by FDM and
SLA can be seen in Figure 7. These cross-sections highlight the possible underfilling issues
encountered with FDM-printing, and the better resolution that can be achieved with SLA-
printing. Already at scale 2.6, the trabecular bone model is not printed well by FDM. In-
terestingly, the difference in total volume between the model and print was less for FDM
(0.6%) than for SLA (6%) for samples of scale 4.3. However, it can be clearly seen that in a
Figure 5.
Volumetric comparison of the original bone model and the overfilling volumes from the printed samples at the
different scales. The original model is displayed in grey, and the overfilled material in red. Note, the model has been scaled
up to the scale of the printed parts in these comparisons.
A more detailed visual comparison of the overfilling and underfilling could be made in
the overlapped cross-sectional micro-CT images of the 3D model and the printed samples.
Overlapping images can be seen in Figure 6. The structures of the sample in scale 4.3
appear to be accurately printed, as seen in the resulting overlap images. For scale 1.8, some
trabecular structures are too thick compared to the ones of the model, and some anomalies
are visible. For scale 1.5 on the other hand, many trabecular structures appear too thick,
with the result that some pores disappear and the overall structure is quite different from
that of the model. Scale 1.8 appears to be the smallest scale at which it is possible to print
the model with acceptable quality and accuracy. However, some errors are noticeable in
the printed samples and it could be interesting to conduct further studies on this aspect, to
investigate the influence of these artefacts in more detail.
A previous study printed the same trabecular bone model, but with FDM [
13
]. A
comparison of similar cross-sections of the same part of the model printed by FDM and
SLA can be seen in Figure 7. These cross-sections highlight the possible underfilling is-
sues encountered with FDM-printing, and the better resolution that can be achieved with
SLA-printing. Already at scale 2.6, the trabecular bone model is not printed well by FDM.
Interestingly, the difference in total volume between the model and print was less for FDM
(0.6%) than for SLA (6%) for samples of scale 4.3. However, it can be clearly seen that in a
more detailed analysis, the FDM parts have both more underfilling and overfilling than
the SLA part. Overfilling and underfilling was not quantified for the FDM parts, but the
trabecular thickness was quantified. Here, the results for SLA (206
±
117
µ
m) are closer to
the model (246
±
70
µ
m) than the results for FDM (98
±
36
µ
m). This observation highlights
the importance of not only comparing overall volumes, but also doing more detailed obser-
vations when evaluating the accuracy of additively manufactured parts. In the evaluation
of the trabecular bone model, quantification of trabecular thickness and comparison of
over- and under-filling provided a useful representation of the print accuracy.
Materials 2021,14, 3712 14 of 17
Materials 2021, 14, x FOR PEER REVIEW 14 of 17
more detailed analysis, the FDM parts have both more underfilling and overfilling than
the SLA part. Overfilling and underfilling was not quantified for the FDM parts, but the
trabecular thickness was quantified. Here, the results for SLA (206 ± 117 µm) are closer to
the model (246 ± 70 µm) than the results for FDM (98 ± 36 µm). This observation highlights
the importance of not only comparing overall volumes, but also doing more detailed ob-
servations when evaluating the accuracy of additively manufactured parts. In the evalua-
tion of the trabecular bone model, quantification of trabecular thickness and comparison
of over- and under-filling provided a useful representation of the print accuracy.
Figure 6. Cross-sectional comparison of registered micro-CT images from the model and the
printed parts. The model is displayed in white (first row) and the printed parts of different scales
are displayed in red (second row). An overlap comparison is seen in the bottom rows, where the
printed cross-sections are superimposed over the model. For this comparison, the model (white)
was scaled up to the scale of the displayed printed part (red).
The results provided by the different evaluation methods used in this study, includ-
ing the extensive comparison between the model and the printed samples, allowed for a
detailed picture of the capabilities of SLA printers and an effective performance compar-
ison with the FDM printing technique. The Black resin permitted to reach a good printing
resolution and future studies could focus on modifying its composition to adjust the me-
chanical properties of the printed parts. Other resins could be investigated, especially bi-
ocompatible resin if in vivo applications such as bone void-filling scaffolds are considered.
The SLA printing technique proved to be a potential solution to produce anatomical bone
models and for training bone models, and in a second instance, aimed to contribute to the
overall objective to improve the production of customised medical products.
Figure 6.
Cross-sectional comparison of registered micro-CT images from the model and the printed
parts. The model is displayed in white (
first row
) and the printed parts of different scales are
displayed in red (
second row
). An overlap comparison is seen in the bottom rows, where the printed
cross-sections are superimposed over the model. For this comparison, the model (
white
) was scaled
up to the scale of the displayed printed part (red).
Figure 7.
Cross-sectional comparison of registered micro-CT images, of the same part of the 3D-
model, from the model and the printed parts produced by FDM (
upper row
) and SLA (
lower row
).
Cross-sections for the 3D-models are displayed in white, and the printed parts in blue (FDM) or red
(SLA). The images in the upper row are adapted with modifications from [13].
Materials 2021,14, 3712 15 of 17
The results provided by the different evaluation methods used in this study, including
the extensive comparison between the model and the printed samples, allowed for a
detailed picture of the capabilities of SLA printers and an effective performance comparison
with the FDM printing technique. The Black resin permitted to reach a good printing
resolution and future studies could focus on modifying its composition to adjust the
mechanical properties of the printed parts. Other resins could be investigated, especially
biocompatible resin if
in vivo
applications such as bone void-filling scaffolds are considered.
The SLA printing technique proved to be a potential solution to produce anatomical bone
models and for training bone models, and in a second instance, aimed to contribute to the
overall objective to improve the production of customised medical products.
4. Conclusions
Trabecular samples were successfully printed by SLA using a 3D trabecular bone
model scaled by factors from 1 to 4.3. Black resin (Formlabs, Somerville, MA, USA)
appeared to allow for the best printing accuracy, and the scale 1.8 appeared to be the
smallest scale with which to print with acceptable precision. Micro-CT analysis showed that
major artefacts may appear in samples with scale factors below 1.8. Trabecular thickness
analysis also showed a decreased accuracy in thickness distribution for scale factors below
1.8. A comparison with the total trabecular volume of the model found a difference of less
than 10% for samples with scales 1.8 and 4.3, and more than 50% for samples with scale 1.5.
Nevertheless, in a detailed analysis of overlapped print and model micro-CT images, scale
1.8 had a loss of accuracy in some details (~20% over- and under-filling). Overfilling was
the main problem for the 1.5 scale.
Compression tests on trabecular samples revealed that X-rays from the micro-CT
significantly increased the elastic modulus. Pull-out tests on trabecular samples showed
that SLA-printed parts exhibit a higher pull-out strength compared to Sawbones
, and
a lower pull-out strength compared to cadaveric specimens or FDM-printed poly(lactic
acid) (PLA) parts. However, for the same 3D model, smaller scales could be printed with
good geometric accuracy, and a higher resolution was achieved compared to FDM prints.
SLA printing appears promising for the intended application, but further optimisation is
required to make 3D-printed structures competitive to the commercially existing foams.
Supplementary Materials:
The following are available online at https://www.mdpi.com/article/10
.3390/ma14133712/s1, A: Preliminary tests for the choice of printing parameters, B: Compression
tests on samples with or without post-processing. Table S1. Parameters chosen for the experimental
plan performed on standard cylinders printed in Black resin. The boxes highlighted in grey indicate
the parameters that change from the control group; Figure S1. Elastic modulus of standard cylinders
printed in Black resin, tested in compression. The errors bars represent standard deviation within
the group. Statistical difference of a group is shown by *, followed by the group numbers that it is
statistically significant to; Figure S2. “Cured” and “Green” parts properties comparison. The error
bars represent standard deviation within the groups.
Author Contributions:
Conceptualisation, C.P.; methodology, A.G., S.L., O.E. and C.P.; formal
analysis, A.G. and S.L.; investigation, A.G., S.L. and O.E.; resources, C.P.; data curation, A.G. and
S.L.; writing—original draft preparation, A.G.; writing—review and editing, S.L., C.P. and J.K.;
visualisation, A.G. and S.L.; supervision, C.P.; project administration, C.P.; funding acquisition, C.P.
and J.K. All authors have read and agreed to the published version of the manuscript.
Funding:
This research was funded by Sweden’s Innovation Agency VINNOVA, grant number
2019-00029.
Institutional Review Board Statement: Not applicable.
Informed Consent Statement: Not applicable.
Acknowledgments:
3D printing was performed at U-PRINT: Uppsala University’s 3D printing
facility at the Disciplinary Domain of Medicine and Pharmacy. The authors are grateful to Adam
Materials 2021,14, 3712 16 of 17
Engberg for his support and advice on the printers, and to Alejandro López and Assoc. Caroline
Öhman Mägi for the training on other equipment.
Conflicts of Interest:
The authors declare no conflict of interest. The funders had no role in the design
of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, or
in the decision to publish the results.
References
1.
Facts and Figures. 2005. Available online: https://hcup-us.ahrq.gov/reports/factsandfigures/facts_figures_2005.jsp#ex3_6c
(accessed on 5 April 2021).
2.
Burkhard, M.; Fürnstahl, P.; Farshad, M. Three-dimensionally printed vertebrae with different bone densities for surgical training.
Eur. Spine J. 2019,28, 798–806. [CrossRef]
3.
Gilbody, J.; Prasthofer, A.; Ho, K.; Costa, M. The use and effectiveness of cadaveric workshops in higher surgical training: A
systematic review. Ann. R. Coll. Surg. Engl. 2011,93, 347–352. [CrossRef] [PubMed]
4.
Burr, D.B. Chapter 1—Bone Morphology and Organization. In Basic and Applied Bone Biology, 2nd ed.; Burr, D.B., Allen, M.R.,
Eds.; Academic Press: Cambridge, MA, USA, 2019; pp. 3–26. ISBN 978-0-12-813259-3.
5.
Cristofolini, L.; Viceconti, M. Mechanical validation of whole bone composite tibia models. J. Biomech.
2000
,33, 279–288.
[CrossRef]
6.
Patel, P.S.; Shepherd, D.E.; Hukins, D.W. Compressive properties of commercially available polyurethane foams as mechanical
models for osteoporotic human cancellous bone. BMC Musculoskelet. Disord. 2008,9, 137. [CrossRef]
7.
Ferdowsian, H.R.; Beck, N. Ethical and Scientific Considerations Regarding Animal Testing and Research. PLoS ONE
2011
,6,
e24059. [CrossRef] [PubMed]
8.
Poukalova, M.; Yakacki, C.M.; Guldberg, R.E.; Lin, A.; Saing, M.; Gillogly, S.D.; Gall, K. Pullout Strength of Suture Anchors: Effect
of Mechanical Properties of Trabecular Bone. J. Biomech. 2010,43, 1138–1145. [CrossRef] [PubMed]
9.
Bose, S.; Vahabzadeh, S.; Bandyopadhyay, A. Bone tissue engineering using 3D printing. Mater. Today
2013
,16, 496–504. [CrossRef]
10. Ng, W.L.; Chua, C.K.; Shen, Y.-F. Print Me An Organ! Why We Are Not There Yet. Prog. Polym. Sci. 2019,97, 101145. [CrossRef]
11.
Yang, Y.; Wang, G.; Liang, H.; Gao, C.; Peng, S.; Shen, L.; Shuai, C. Additive manufacturing of bone scaffolds. Int. J. Bioprinting
2018,5, 148. [CrossRef]
12.
Alsheghri, A.; Reznikov, N.; Piché, N.; McKee, M.D.; Tamimi, F.; Song, J. Optimization of 3D network topology for bioinspired
design of stiff and lightweight bone-like structures. Mater. Sci. Eng. C 2021,123, 112010. [CrossRef] [PubMed]
13.
Wu, D.; Spanou, A.; Diez-Escudero, A.; Persson, C. 3D-printed PLA/HA composite structures as synthetic trabecular bone: A
feasibility study using fused deposition modeling. J. Mech. Behav. Biomed. Mater. 2020,103, 103608. [CrossRef]
14.
Szymczyk-Ziółkowska, P.; Łabowska, M.B.; Detyna, J.; Michalak, I.; Gruber, P. A review of fabrication polymer scaffolds for
biomedical applications using additive manufacturing techniques. Biocybern. Biomed. Eng. 2020,40, 624–638. [CrossRef]
15.
Ng, W.L.; Lee, J.M.; Zhou, M.; Chen, Y.-W.; Lee, K.-X.A.; Yeong, W.Y.; Shen, Y.-F. Vat polymerization-based bioprinting—process,
materials, applications and regulatory challenges. Biofabrication 2020,12, 022001. [CrossRef] [PubMed]
16.
Li, W.; Mille, L.S.; Robledo, J.A.; Uribe, T.; Huerta, V.; Zhang, Y.S. Recent Advances in Formulating and Processing Biomaterial
Inks for Vat Polymerization-Based 3D Printing. Adv. Healthc. Mater. 2020,9, 2000156. [CrossRef]
17.
Chartrain, N.A.; Williams, C.B.; Whittington, A.R. A review on fabricating tissue scaffolds using vat photopolymerization. Acta
Biomater. 2018,74, 90–111. [CrossRef] [PubMed]
18.
Mézière, F.; Juskova, P.; Woittequand, J.; Muller, M.; Bossy, E.; Boistel, R.; Malaquin, L.; Derode, A. Experimental observation of
ultrasound fast and slow waves through three-dimensional printed trabecular bone phantoms. J. Acoust. Soc. Am.
2016
,139,
EL13–EL18. [CrossRef]
19.
Barak, M.M.; Black, M.A. A novel use of 3D printing model demonstrates the effects of deteriorated trabecular bone structure on
bone stiffness and strength. J. Mech. Behav. Biomed. Mater. 2018,78, 455–464. [CrossRef]
20.
Fernández-Cervantes, I.; Morales, M.A.; Agustín-Serrano, R.; Cardenas-García, M.; Pérez-Luna, P.V.; Arroyo-Reyes, B.L.;
Maldonado-García, A. Polylactic acid/sodium alginate/hydroxyapatite composite scaffolds with trabecular tissue morphology
designed by a bone remodeling model using 3D printing. J. Mater. Sci. 2019,54, 9478–9496. [CrossRef]
21.
Milazzo, M.; Contessi Negrini, N.; Scialla, S.; Marelli, B.; Farè, S.; Danti, S.; Buehler, M.J. Additive Manufacturing Approaches for
Hydroxyapatite-Reinforced Composites. Adv. Funct. Mater. 2019,29, 1903055. [CrossRef]
22.
Black_Resin_SDS_EU.pdf. Available online: https://archive-media.formlabs.com/upload/Black_Resin_SDS_EU.pdf (accessed
on 14 June 2020).
23.
Materials-Technical-Library FORMLABS.pdf. Available online: https://formlabs-media.formlabs.com/filer_public/ac/89/ac896
3db-f54a-4cac-8fe9-fb740a7b06f1/formlabs-materials-library.pdf (accessed on 14 June 2020).
24.
SLA 3D Printing Materials Compared. Available online: https://www.3dhubs.com/knowledge-base/sla-3d-printing-materials-
compared/ (accessed on 31 May 2020).
25.
ISO 5833:2002(en), Implants for Surgery—Acrylic Resin Cements. Available online: https://www.iso.org/obp/ui/#iso:std:iso:
5833:ed-2:v1:en (accessed on 31 May 2021).
26.
Ridler, T.W.; Calvard, S. Picture Thresholding Using an Iterative Slection Method. IEEE Trans. Syst. Man Cybern.
1978
,SMC-8,
630–632.
Materials 2021,14, 3712 17 of 17
27.
Hildebrand, T.; Rüegsegger, P. A new method for the model-independent assessment of thickness in three-dimensional images. J.
Microsc. 1997,185, 67–75. [CrossRef]
28. 3D Slicer. Available online: https://www.slicer.org/ (accessed on 26 September 2020).
29.
Coffey, T.; Urquhart, S.G.; Ade, H. Characterization of the effects of soft X-ray irradiation on polymers. J. Electron Spectrosc. Relat.
Phenom. 2002,122, 65–78. [CrossRef]
30. Alexander, P.; Fox, M. The degradation of polymethacrylic acid by X-rays. Trans. Faraday Soc. 1954,50, 605–612. [CrossRef]
31.
Chapman, J.R.; Harrington, R.M.; Lee, K.M.; Anderson, P.A.; Tencer, A.F.; Kowalski, D. Factors Affecting the Pullout Strength of
Cancellous Bone Screws. J. Biomech. Eng. 1996,118, 391–398. [CrossRef]
32.
Shea, T.M.; Doulgeris, J.J.; Gonzalez-Blohm, S.A.; Lee, W.E.; Aghayev, K.; Vrionis, F.D. Balancing Rigidity and Safety of Pedicle
Screw Fixation via a Novel Expansion Mechanism in a Severely Osteoporotic Model. BioMed Res. Int.
2015
,2015, 1–11. [CrossRef]
[PubMed]
33.
Mueller, T.L.; Basler, S.E.; Müller, R.; van Lenthe, G.H. Time-lapsed imaging of implant fixation failure in human femoral heads.
Med. Eng. Phys. 2013,35, 636–643. [CrossRef]
34.
Wu, D.; Pujari-Palmer, M.; Bojan, A.; Palmquist, A.; Procter, P.; Öhman-Mägi, C.; Ferguson, S.J.; Isaksson, P.; Persson, C. The
effect of two types of resorbable augmentation materials—A cement and an adhesive—On the screw pullout pullout resistance in
human trabecular bone. J. Mech. Behav. Biomed. Mater. 2020,110, 103897. [CrossRef] [PubMed]
35.
Pujari-Palmer, M.; Robo, C.; Persson, C.; Procter, P.; Engqvist, H. Influence of cement compressive strength and porosity on
augmentation performance in a model of orthopedic screw pull-out. J. Mech. Behav. Biomed. Mater.
2018
,77, 624–633. [CrossRef]
... The printing accuracy could be conducted by correlating the weight of the printed cubes with their volume fraction (Barak and Black, 2018;Wood et al., 2019) or using micro-CT scanning and the FEM model to validate printing accuracy and precision (Dobson et al., 2006;Grzeszczak et al., 2021;Kuhn et al., 2014b;Tellis et al., 2008), but these approaches only reveal the accuracy of structure replicability rather than the mechanical behavior. In this study, we did not verify the structure replicability since we focused on replicating the accuracy of the macromechanical behavior. ...
... Using stereolithography (compared to Multi-Jet Fusion in our printer), this study investigated the influence of scale factor on trabecular structure replicability and found an optimal scale factor of 1.8 or 2 by comparing the volume difference and trabecular thickness. A higher scale factor did not seem to improve the structure replicability (Grzeszczak et al., 2021). These results are contrary to our results. ...
Article
Full-text available
Three-dimensional (3D) printing has been used to fabricate synthetic trabeculae models and to test mechanical behavior that cannot be recognized in the actual sample, but the extent to which 3D printed trabeculae replicate the mechanical behavior of the actual trabeculae remains to be quantified. The aim of this study was to evaluate the accuracy of 3D printed trabeculae in reproducing the mechanical properties of the corresponding actual trabeculae. Twelve human trabecular cubes (5×5×5 mm) were scanned by micro-CT to form the trabecular 3D model. Each trabecular 3D model was scaled ×2-, ×3-, ×4- and ×5-fold and then printed twice at a layer thickness of 60 μm using poly (lactic acid) (PLA). The actual trabecular cubes and the 3D-printed trabecular cubes were first compressed under a loading rate of 1 mm/min; another replicated stack of 3D-printed trabecular cubes was compressed under a strain rate of 0.2/min. The results showed that the stiffness of the printed cubes tended to increase, while the strength tended to converge when the magnification increased under the two loading conditions. The strain rate effect was found in the printed cubes. The correlation coefficient (R²) of the mechanical properties between the printed and actual trabeculae can reach up to 0.94, especially under ×3-, ×4- and ×5-fold magnification. In conclusion, 3D printing could be a potential tool to evaluate the mechanical behavior of actual trabecular tissue in vitro and may help in the future to predict the risk of fracture and even personalize the treatment evaluation for osteoporosis and other trabecular bone pathologies.
... In order to train surgeons and test new medical devices, synthetic bone models will be developed. In a study, SLA printing was used to develop synthetic trabecular bone based on micro-CT images and found that SLA-printed bone parts showed a higher pull-out strength compared to existing synthetic SawbonesTM with higher resolution [44]. ...
Article
Full-text available
Three-dimensional (3D) printing is a technique where the products are printed layer by layer via a series of cross-sectional slices with the exact deposition of different cell types and biomaterials based on computer-aided design software. Three-dimensional printing can be divided into several approaches, such as extrusion-based printing, laser-induced forward transfer-based printing systems, and so on. Bio-ink is a crucial tool necessary for the fabrication of the 3D construct of living tissue in order to mimic the native tissue/cells using 3D printing technology. The formation of 3D software helps in the development of novel drug delivery systems with drug screening potential, as well as 3D constructs of tumor models. Additionally, several complex structures of inner tissues like stroma and channels of different sizes are printed through 3D printing techniques. Three-dimensional printing technology could also be used to develop therapy training simulators for educational purposes so that learners can practice complex surgical procedures. The fabrication of implantable medical devices using 3D printing technology with less risk of infections is receiving increased attention recently. A Cancer-on-a-chip is a microfluidic device that recreates tumor physiology and allows for a continuous supply of nutrients or therapeutic compounds. In this review, based on the recent literature, we have discussed various printing methods for 3D printing and types of bio-inks, and provided information on how 3D printing plays a crucial role in cancer management.
Article
Full-text available
Additive manufacturing (AM) techniques have gained interest in the tissue engineering field, thanks to their versatility and unique possibilities of producing constructs with complex macroscopic geometries and defined patterns. Recently, composite materials—namely, heterogeneous biomaterials identified as continuous phase (matrix) and reinforcement (filler)—have been proposed as inks that can be processed by AM to obtain scaffolds with improved biomimetic and bioactive properties. Significant efforts have been dedicated to hydroxyapatite (HA)‐reinforced composites, especially targeting bone tissue engineering, thanks to the chemical similarities of HA with respect to mineral components of native mineralized tissues. Herein, applications of AM techniques to process HA‐reinforced composites and biocomposites for the production of scaffolds with biological matrices, including cellular tissues, are reviewed. The primary outcomes of recent investigations in terms of morphological, structural, and in vitro and in vivo biological properties of the materials are discussed. The approaches based on the nature of the matrices employed to embed the HA reinforcements and produce the tissue substitutes are classified, and a critical discussion is provided on the presented state of the art as well as the future perspectives, to offer a comprehensive picture of the strategies investigated as well as challenges in this emerging field of materiomics. Advanced manufacturing technologies currently used to process hydroxyapatite‐reinforced composites for tissue engineering applications are reviewed. A discussion on the constructs, their features, and applications is provided through a classification based on the intimate nature of the polymeric matrices (soft and hard). A conclusive and comprehensive picture is provided to highlight future challenges and perspectives in this field.
Article
Full-text available
The article presents a new methodology that employs 3D printing technology to generate a microporous composite material of polylactic acid, sodium alginate and hydroxyapatite, whose microstructure is designed by means of the 3D numerical solution from a mathematical model. This model represents the spatio-temporal dynamics of the interaction between osteoblasts and osteoclasts in the bone remodeling. The microporosity of composite material mimics the structure of human trabecular bone. This material has density with microporosity pretty close to the one that is exhibited by the natural bone tissue. Close relationship between the material processing and its elasticity module is observed. When subjecting this composite material to a simulated body fluid treatment, the mechanical resistance to compression is increased due to induced mineralization of hydroxyapatite crystals on its surface. The methodology shows potential to generate structures that allow the control of the composite material properties. The material presents a microporosity that has morphological and chemical properties suitable for future applications in tissue engineering as bone scaffold.
Article
A truly bioinspired approach to design optimization should follow the energetically favorable natural paradigm of “minimum inventory with maximum diversity”. This study was inspired by constructive regression of trabecular bone – a natural process of network connectivity optimization occurring early in skeletal development. During trabecular network optimization, the original excessively connected network undergoes incremental pruning of redundant elements, resulting in a functional and adaptable structure operating at lowest metabolic cost. We have recapitulated this biological network topology optimization algorithm by first designing in silico an excessively connected network in which elements are dimension-independent linear connections among nodes. Based on bioinspired regression principles, least-loaded connections were iteratively pruned upon simulated loading. Evolved networks were produced along this optimization trajectory when pre-set convergence criteria were met. These biomimetic networks were compared to each other, and to the reference network derived from mature trabecular bone. Our results replicated the natural network optimization algorithm in uniaxial compressive loading. However, following triaxial loading, the optimization algorithm resulted in lattice networks that were more stretch-dominated than the reference network, and more capable of uniform load distribution. As assessed by 3D printing and mechanical testing, our heuristic network optimization procedure opens new possibilities for parametric design.
Article
Augmentation materials, such as ceramic and polymeric bone cements, have been frequently used to improve the physical engagement of screws inserted into bone. While ceramic, degradable cements may ultimately improve fixation stability, reports regarding their effect on early fixation stability have been inconsistent. On the other hand, a newly developed degradable ceramic adhesive that can bond with tissues surrounding the screw, may improve the pullout performance, ensure early stability, and subsequent bony integration. The aim of this study was to investigate failure mechanisms of screw/trabecular bone constructs by comparing non-augmented screws with screws augmented with a calcium phosphate cement or an adhesive, i.e. a phosphoserine-modified calcium phosphate. Pullout tests were performed on screws inserted into trabecular cylinders extracted from human femoral bone. Continuous and stepwise pullout loading was applied with and without real-time imaging in a synchrotron radiation micro-computed tomograph, respectively. Statistical analysis that took the bone morphology into account confirmed that augmentation with the adhesive supported significantly higher pullout loads compared to cement-augmented, or non-augmented screws. However, the adhesive also allowed for a higher injection volume compared to the cement. In-situ imaging showed cracks in the vicinity of the screw threads in all groups, and detachment of the augmentation materials from the trabecular bone in the augmented specimens. Additional cracks at the periphery of the augmentation and the bone-material interfaces were only observed in the adhesive-augmented specimen, indicating a contribution of surface bonding to the pullout resistance. An adhesive that has potential for bonding with tissues, displayed superior pullout resistance, compared to a brushite cement, and may be a promising material for cementation or augmentation of implants.
Article
3D printing and bioprinting have become a key component in precision medicine. They have been used toward the fabrication of medical devices with patient‐specific shapes, production of engineered tissues for in vivo regeneration, and preparation of in vitro tissue models used for screening therapeutics. In particular, vat polymerization‐based 3D (bio)printing as a unique strategy enables more sophisticated architectures to be rapidly built. This progress report aims to emphasize the recent advances made in vat polymerization‐based 3D printing and bioprinting, including new biomaterial ink formulations and novel vat polymerization system designs. While some of these approaches have not been utilized toward the combination with biomaterial inks, it is anticipated their rapid translation into biomedical applications.
Article
This paper presents the current state-of-the art of additive manufacturing (AM) applications in the biomedical field, especially in tissue engineering. Multiple advantages of additive manufacturing allow to precise three-dimensional objects fabrication with complex structure using various materials. Depending on the purpose of the manufactured part, different AM technologies are implemented, in which a specific material can be utilized. In the biomedical field, there are used several techniques such as: Binder Jetting, Material Extrusion, Material Jetting, Powder Bed Fusion, Sheet Lamination, Vat Polymerization. This article focuses on the utilization of polymer materials (natural and synthetic) taking into account hydrogels in scaffolds fabrication. Assessment of polymer scaffolds mechanical properties enables personalized patient care, as well as prevents damage after implantation in human body. By controlling process parameters it is possible to obtain optimised mechanical properties of manufactured parts.
Article
Additive manufacturing has significant advantages, in the biomedical field, allowing for customized medical products where complex architectures can be achieved directly. While additive manufacturing can be used to fabricate synthetic bone models, this approach is limited by the printing resolution, at the level of the trabecular bone architecture. Therefore, the aim of this study was to evaluate the possibilities of using fused deposition modeling (FDM) to this end. To better mimic real bone, both in terms of mechanical properties and biodegradability, a composite of degradable polymer, poly(lactic acid) (PLA), and hydroxyapatite (HA) was used as the filament. Three PLA/HA composite formulations with 5-10-15 wt% HA were evaluated, and scaled up human trabecular bone models were printed using these materials. Morphometric and mechanical properties of the printed models were evaluated by micro-computed tomography, compression and screw pull out tests. It was shown that the trabecular architecture could be reproduced with FDM and PLA by applying a scaling factor of 2-4. The incorporation of HA particles reduced the printing accuracy, with respect to morphology, but showed potential for enhancement of the mechanical properties. The scaled-up models displayed comparable, or slightly enhanced, strength compared to the commonly used polymeric foam synthetic bone models (i.e. Sawbones). Reproducing the trabecular morphology by 3D printed PLA/HA composites appears to be a promising strategy for synthetic bone models, when high printed resolution can be achieved.
Article
Over the years, the field of bioprinting has attracted attention for its highly automated fabrication system that enables the precise patterning of living cells and biomaterials at pre-defined positions for enhanced cell-matrix and cell-cell interactions. Notably, vat polymerization (VP)-based bioprinting is an emerging bioprinting technique for various tissue engineering applications due to its high fabrication accuracy. Particularly, different photo-initiators (PIs) are utilized during the bioprinting process to facilitate the crosslinking mechanism for fabrication of high-resolution complex tissue constructs. The advancements in VP-based printing have led to a paradigm shift in fabrication of tissue constructs from cell-seeding of tissue scaffolds (non-biocompatible fabrication process) to direct bioprinting of cell-laden tissue constructs (biocompatible fabrication process). This paper, presenting a first-time comprehensive review of the VP-based bioprinting process, provides an in-depth analysis and comparison of the various biocompatible PIs and highlights the important considerations and bioprinting requirements. This review paper reports a detailed analysis of its printing process and the influence of light-based curing modality and PIs on living cells. Lastly, this review also highlights the significance of VP-based bioprinting, the regulatory challenges and presents future directions to transform the VP-based printing technology into imperative tools in the field of tissue engineering and regenerative medicine. The readers will be informed on the current limitations and achievements of the VP-based bioprinting techniques. Notably, the readers will realize the importance and value of highly-automated platforms for tissue engineering applications and be able to develop objective viewpoints towards this field.
Article
Bioprinting offers a highly-automated and advanced manufacturing platform that facilitates the deposition of bio-inks (living cells, biomaterials and growth factors) in a scalable and reproducible manner, a process that is lacking in conventional tissue engineering approaches. There are significant improvements in the field of bioprinting over the last two decades; an in-depth analysis of current improvement in the bioprinting techniques, progress in bio-ink development, implementation of new bioprinting and tissue maturation strategies are presented. In this review, we highlight how the progress in polymer sciences complements 3D bioprinting in overcoming some of the major impediments in the field of organ printing. We provide a concise overview of the anatomy and physiology of different tissues/organs, followed by important design considerations to better facilitate the fabrication of biomimetic tissues/organs for tissue engineering and regenerative medicine (TERM). Lastly, a realistic overview of current progress in organ bioprinting by presenting the current limitations and achievements in bioprinted tissue-engineered constructs, followed by an outlook. We strongly believe that with the advances in polymer sciences, it will be an impending reality for on-demand bioprinting of patient-specific tissues/organs. Here's the Share Link – a personalized URL providing 50 days' free access to our article (until September 29, 2019). https://authors.elsevier.com/a/1ZXrF_1pxsKUci