ArticlePDF Available

Characterization of E-cigarette coil temperature and toxic metal analysis by infrared temperature sensing and scanning electron microscopy -energy- dispersive X-ray

Authors:

Abstract and Figures

Introduction: Electronic cigarettes (e-cigarettes) have rapidly evolved since their introduction to the U.S. market. The rebuildable atomizer (RBA) offers user-driven modification to the heating element (coil) and wicking systems. Different coil materials can be chosen based on user needs and preferences. However, the heating element of an e-cigarette is believed to be one-source for toxic metal exposure. Methods: E-cigarette coils from Kanthal and nichrome wires were constructed in a contact and non-contact configuration and heated at four voltages. The maximum temperatures of the coils were measured by infrared temperature sensing when dry and when saturated with 100% vegetable glycerin or 100% propylene glycol. The metal composition of each coil was analyzed with Scanning Electron Microscopy-Energy-Dispersive X-Ray (SEM-EDX) when new, and subsequently after 1, 50, and 150 heat cycles when dry. Results: The coils reached temperatures above 1000 °C when dry, but were below 300 °C in both liquid-saturated mediums. Metal analysis showed a decrease of 9-19% chromium and 39-58% iron in Kanthal wire and a decrease of 12-14% iron and 39-43% nickel in nichrome wire after 150 heat cycles. Significant metal loss was observed after one heat cycle for both coil alloys and configurations. Conclusions: The loss of metals from these heat cycles further suggests that the metals from the coils are potentially entering the aerosol of the e-cigarette, which can be inhaled by the user.
Content may be subject to copyright.
Full Terms & Conditions of access and use can be found at
https://www.tandfonline.com/action/journalInformation?journalCode=iiht20
Inhalation Toxicology
International Forum for Respiratory Research
ISSN: (Print) (Online) Journal homepage: https://www.tandfonline.com/loi/iiht20
Characterization of E-cigarette coil temperature
and toxic metal analysis by infrared temperature
sensing and scanning electron microscopy –
energy-dispersive X-ray
Haley A. Mulder , James B. Stewart , Ivy P. Blue , Rose I. Krakowiak , Jesse L.
Patterson , Kimberly N. Karin , Jasmynne M. Royals , Alexandra C. DuPont ,
Kaitlin E. Forsythe , Justin L. Poklis , Alphonse Poklis , Shelle N. Butler ,
Joseph B. McGee Turner & Michelle R. Peace
To cite this article: Haley A. Mulder , James B. Stewart , Ivy P. Blue , Rose I. Krakowiak , Jesse L.
Patterson , Kimberly N. Karin , Jasmynne M. Royals , Alexandra C. DuPont , Kaitlin E. Forsythe ,
Justin L. Poklis , Alphonse Poklis , Shelle N. Butler , Joseph B. McGee Turner & Michelle R.
Peace (2020): Characterization of E-cigarette coil temperature and toxic metal analysis by infrared
temperature sensing and scanning electron microscopy – energy-dispersive X-ray, Inhalation
Toxicology, DOI: 10.1080/08958378.2020.1840678
To link to this article: https://doi.org/10.1080/08958378.2020.1840678
Published online: 03 Nov 2020. Submit your article to this journal
Article views: 6 View related articles
View Crossmark data
RESEARCH ARTICLE
Characterization of E-cigarette coil temperature and toxic metal analysis by
infrared temperature sensing and scanning electron microscopy energy-
dispersive X-ray
Haley A. Mulder
a
, James B. Stewart
a
, Ivy P. Blue
a
, Rose I. Krakowiak
a
, Jesse L. Patterson
a
, Kimberly N. Karin
a
,
Jasmynne M. Royals
a
, Alexandra C. DuPont
a
, Kaitlin E. Forsythe
a
, Justin L. Poklis
c
, Alphonse Poklis
a,c,d
,
Shelle N. Butler
a
, Joseph B. McGee Turner
b
and Michelle R. Peace
a
a
Department of Forensic Science, Virginia Commonwealth University, Richmond, VA, USA;
b
Department of Chemistry, Virginia
Commonwealth University, Richmond, VA, USA;
c
Department of Pharmacology and Toxicology, Virginia Commonwealth University,
Richmond, VA, USA;
d
Department of Pathology, Virginia Commonwealth University, Richmond, VA, USA
ABSTRACT
Introduction: Electronic cigarettes (e-cigarettes) have rapidly evolved since their introduction to the
U.S. market. The rebuildable atomizer (RBA) offers user-driven modification to the heating element
(coil) and wicking systems. Different coil materials can be chosen based on user needs and preferen-
ces. However, the heating element of an e-cigarette is believed to be one-source for toxic
metal exposure.
Methods: E-cigarette coils from Kanthal and nichrome wires were constructed in a contact and non-
contact configuration and heated at four voltages. The maximum temperatures of the coils were
measured by infrared temperature sensing when dry and when saturated with 100% vegetable gly-
cerin or 100% propylene glycol. The metal composition of each coil was analyzed with Scanning
Electron Microscopy-Energy-Dispersive X-Ray (SEM-EDX) when new, and subsequently after 1, 50, and
150 heat cycles when dry.
Results: The coils reached temperatures above 1000 C when dry, but were below 300 Cinboth
liquid-saturated mediums. Metal analysis showed a decrease of 919% chromium and 3958% iron in
Kanthal wire and a decrease of 1214% iron and 3943% nickel in nichrome wire after 150 heat cycles.
Significant metal loss was observed after one heat cycle for both coil alloys and configurations.
Conclusions: The loss of metals from these heat cycles further suggests that the metals from the coils
are potentially entering the aerosol of the e-cigarette, which can be inhaled by the user.
ARTICLE HISTORY
Received 16 March 2020
Accepted 17 October 2020
KEYWORDS
E-cigarettes; coils; metal
analysis; metal aerosol;
SEM-EDX
Introduction
Electronic cigarettes (e-cigarettes) have rapidly evolved since
their introduction to the U.S. market in 2007. All e-cigarettes
operate under the same, basic principle. The device consists
of two main components: a battery power supply and an
atomizer that contains a coil and wicking system (Breland
et al. 2017). The coil is typically a metal filament wrapped
around a wicking material of silica or cotton, saturated in a
refill formulation liquid (e-liquid). The e-liquid consists of a
ratio of propylene glycol and vegetable glycerin, nicotine, and
usually, a flavoring agent (Peace et al. 2016). When the bat-
tery is activated, it delivers an electric charge to the coil,
which heats and aerosolizes the e-liquid. The generated aero-
sol is then inhaled by the user (Peace et al. 2016; Breland
et al. 2017).
The first e-cigarette, designed in China in 2004, resembled
a traditional cigarette and was known as the ciga-like.Future
generations of e-cigarettes consisted of the cartomizer (a cart-
ridge atomizer) and clearomizer (a transparent tank cartom-
izer), which contained replaceable tank-and-atomizer systems
(Breland et al. 2017). A popular generation of e-cigarette is
known as the rebuildable atomizer (RBA), which allows for
complete user-driven customizability. The intent of the RBA
is to enable the user to modify the device to create a nicotine
delivery system that an e-cigarette consumer would find satis-
fying (Grana et al. 2013). Users take advantage of adjustable
power supplies and coil configurations to impact heat produc-
tion with the desire to maximize drug delivery.
The use of rebuildable atomizers caused polarizing dis-
cussions about what components will create the best build
and therefore, the best vaping experience. Early e-cigarette
models used nichrome 80:20 wire for coil builds in commer-
cial atomizers (Williams et al. 2013). The metal is comprised
of nickel and chromium and has a maximum operating
temperature of 1200 C (BNC 2020). Kanthal A-1 is a resist-
ance wire that grew popular with RBA builders in the vap-
ing community. The metal is comprised of iron, chromium,
and aluminum alloys and has a maximum operating tem-
perature of 1400 C (Kanthal 2018a). With wire composition
preferences came choices in wire configurations. Generally,
the coils in the e-cigarette can be wrapped into two
CONTACT Michelle R. Peace mrpeace@vcu.edu Department of Forensic Science, Virginia Commonwealth University, Box 843079, Richmond, VA 23284, USA
ß2020 Informa UK Limited, trading as Taylor & Francis Group
INHALATION TOXICOLOGY
https://doi.org/10.1080/08958378.2020.1840678
configurations: contact or non-contact. The contact build
has the coil wraps touching together and the non-contact
build has the coil wraps separated in space. Despite the
numerous ways to create an e-cigarette coil, there is no
definitive answer on which type of coil build provides the
best user experience and most efficient drug delivery.
Potential toxic metal exposure from inhaling e-cigarette
aerosols is of concern with the use of metallic filaments as
the heating element. Chromium and nickel, two major metal
alloys present in nichrome wire are listed as carcinogens
and are linked with chronic lung infections and other
respiratory illnesses such as reduced lung infections and
bronchitis (ASTDR 2005,2012). Chromium is also present
in Kanthal wire. Some studies for iron oxides also suggest that
inhalation over time can lead to lung cancer (Kornberg et al.
2017). Early studies assessed e-cigarette aerosol for metals and
identified tin, silver, iron, nickel, aluminum, and cadmium
present in the aerosol (Williams et al. 2013; Hess et al. 2017).
Several studies highlighted a (Hess et al. 2017; Halstead et al.
2019; Williams et al. 2019; Zervas et al. 2020) variability in
metal concentrations in aerosols between e-cigarette brands,
potentially resulting from the different heating units within
the devices, device age, and access of the liquid to metal. One
study assessed the metal content present in e-liquids directly
from the manufacturer bottle verses the tank and the resulting
aerosol. The study determined that there was a higher metal
content present in the e-liquid from the tank and the aerosol,
suggesting that the coils had an effect on metal concentration
(Olmedo et al. 2018). A later study found that factors such as
e-liquid composition and boiling temperature of the e-liquids
had an effect on the transfer of metal from two commercial
coils to the e-liquid aerosolized in the device (Zervas
et al. 2020).
The purpose of this study was to characterize two types of
common metal wires used to construct coils for electronic cig-
arettes with typical user modifications. The temperature out-
put of the two coils when in a contact and non-contact
configuration under dry conditions and wet conditions when
saturated with 100% vegetable glycerin, 100% propylene glycol
e-liquids, and a 50:50 mixture was determined using an infra-
red (IR) temperature sensor. The metal profile of the two coils
before and after heating was determined using scanning elec-
tron microscopy energy-dispersive X-ray (SEM-EDX).
Experimental
Reagents and supplies
For all studies, a Kayfun Lite Styled Rebuildable Atomizer
(RBA) was purchased from FastTech (Hong Kong, China).
The Pure-Atomist Kanthal A-1 wire of 30, 32, and 34
American Wire Gauge (AWG) and ceramic tweezers were
purchased from Lightning Vapes (Brandenton, FL). The
consolidated chromium wire of 30, 32, and 34 AWG were
purchased from McMaster Carr (Elmhurst, IL). The 2 mm
high quality silica wick and Master 5-in-1 coiling kit were
purchased from Amazon (Seattle, WA). The 100% pure
vegetable glycerin (VG) and 100% pure propylene glycol
(PG) were purchased from Wizard Labs (Almonte Springs,
FL). A thermoMETER M3 and M1/M2 temperature sensors
ranging from 100600 C and 6501800 C were purchased
from Micro-Epsilon (Raleigh, NC). The Sorensen power
supply was from Amtek (San Diego, CA) and a multimeter
was purchased from Fluke (Everett, WA).
Coil building
Kanthal A-1 and nickel chromium (nichrome) wires of 30
AWG (254.6 mm), 32 AWG (201.9 mm), and 34 AWG
(160.1 mm) were used to wrap all coils in two configurations:
contact and non-contact (Figure 1). The coils in each gauge
and configuration were tested under two types of conditions:
(i) dry mediumcontaining no silica wick and no e-liquid
and (ii) wet medium,which contained the 2 mm silica wick
and e-liquid. Each coil was constructed using the 1.5 mm
diameter rod from the Coil Master 5-in-1 coiling kit and
were wrapped to 1.8 X. For non-contact configuration, the
coils were pulled apart so that the wraps were no longer in
contact. The coils were housed in a Kayfun Lite Styled RBA
and a Fluke multimeter was used to ensure that every coil
type, configuration, and gauge was wrapped to 1.8 X. In the
contact configuration, the coils were wrapped and placed in
the atomizer and then annealed to ensure even heating across
the coil before the resistance was tested. For all wet medium
conditions in the temperature studies, a 2 mm silica wick was
threaded through the coil and the atomizer was filled with
the e-liquid. E-liquid refill formulations used for the tempera-
ture studies were 100% VG, 100% PG, and a 50:50 PG:VG
made in-house with no nicotine.
Coil temperature studies
The atomizers were heated using a Sorensen power supply,
which controlled the voltage output. The coils were heated at
increasing voltages of 3.5, 3.7, 4.2, and 5.0 V (6.8, 7.6, 9.8,
13.9 W) . The trials were completed in triplicate, resulting in
three trials for each coil under each voltage, configuration,
and dry or wet medium (100% PG and 100% VG). Both coil
materials were then tested under contact and non-contact
configurations at 32 gauge in a 50:50 PG:VG solution. Two
Figure 1. Electronic cigarette coil wrapped in a contact (A) and non-contact (B)
configuration.
2 H. A. MULDER ET AL.
dual-laser infrared temperature sensors by Micro-Epsilon
were used to measure the temperature outputs of the coils. A
thermoMETER CT Laser M3 with a measuring range of
100650 C and spectral range of 2.3 mm was used for all wet
mediumstudies. A thermoMETER CT Laser M1/M2 with a
measuring range of 6501800 C and a spectral range of 1/
1.6 mm was used for all dry mediumstudies. The coils were
heated for eight seconds, and the maximum temperature at
both ends near the pole and at the center of the coil was
recorded. The temperature at the three positions was meas-
ured ten times. Compact Connect version 1.9.8.6 was used to
record the temperatures of the coils as they were heated.
Coil SEM analysis
The coils analyzed for SEM were heated from the Sorensen
power supply at their maximum operating voltage for ten
seconds under dry conditions. The coils were heated for
one, 50, and 150 times at their maximum operating temper-
atures and stored in a trace metal free microcentrifuge
tubes. The max operating temperatures were determined by
the voltage and coil combination that could withstand a ten
second burn without melting the coil. The analysis of the
coil composition pre and post-heating was conducted using
a Hitachi SU-70 Scanning Electron Microscope coupled
with Energy Dispersive X-Ray Spectroscopy (SEM-EDX).
Samples were mounted on an aluminum disk coated with
carbon tape, with the ends taped down and the wraps pulled
apart and pressed into the carbon tape. An EDX spectrum
was obtained simultaneously with the SEM images. The
SEM-EDX was used with an accelerating voltage of 20 keV
and the center of the coil was identified and a measurement
was made in that center and to the immediate left and right,
totaling 12 measurements for each gauge at each configur-
ation at 2000magnification. The EDX spectrum and
weight percent of each metal present in the coil were also
obtained. SEM-EDX data for coils at 32 gauge for both coil
types and configurations were recorded at the center of the
coil for new, one, and50 heat cycles in the presence of a wet
media of 50:50 PG to VG.
The BrownForsythe test was used to test for equality of
variances. A significance level of 0.05 was used. The data
that was determined to exhibit equal variance was analyzed
using ANOVA. Data with unequal variance were analyzed
using Tukey-HSD. Data were analyzed using JMP pro 12
statistical software.
Results
Coil temperature studies
In the dry heat studies across the length of the coil, the cen-
ter of the coils was visibly brighter than the ends in both
configurations suggesting that the coils produced the most
heat in the center of the coil, and was confirmed with
510% higher temperature measurements in the center.
Kanthal in a contact configuration with the lowest gauge
(thickest wire) had the most significant increase in
temperature output, from 1134 C to 1436 C, with increas-
ing voltage. Wire gauge only significantly impacted tempera-
ture at the lower voltages in contact configuration, with the
thicker wire exhibiting the lowest starting temperature
(1134 C) at the lowest voltage (3.5 V) (Figure 2). The thin-
nest wire, 34 gauge, in contact configuration was unable to
withstand the heat produced at 5.0 V and burnt out/melted
before the maximum temperature could be recorded.
Kanthal in non-contact configuration had lower operational
temperatures compared to the contact configuration with
increasing voltages and wire gauges. Initial temperatures of
30 and 32 gauge wires at 3.5 V, dry, non-contact configura-
tions were 975 C and 1016 C versus 1134C and 1354 C
in dry, contact configurations, respectively. Temperatures of
34 gauge at 4.2 and 5.0 V dry, non-contact configurations
could not be recorded. For nichrome wire with dry heat in
a contact configuration, the operational temperature range
of 10511234 C was lower compared to Kanthal across the
voltages and wire gauges (Figure 2). In a contact configur-
ation, the temperature profile of 34 gauge nichrome could
not be recorded even at the lowest voltage due to burnout.
In a non-contact configuration, the wire had a temperature
range of 9111234 C across the voltages and wire gauges.
Similar to non-contact Kanthal, the non-contact nichrome
wire at 34 gauge could not be recorded at 4.2 and 5.0 V.
For the wet heat studies with 100% vegetable glycerin
(VG), the temperature output was greatly reduced compared
to the dry heats. Kanthal wire in a contact configuration
had a temperature range of 266298 C with increasing volt-
age and wire gauge (Figure 2). In a non-contact configur-
ation, the Kanthal wire had a temperature range of
241275 C. Statistical analysis of the two coil configurations
determined that the temperature output in a contact verses
a non-contact configuration for Kanthal wire was not statis-
tically different (p¼0.453). For nichrome wire in a contact
configuration, the temperature of the coils ranged from
253283 C across all voltages and gauges. In a non-contact
configuration, the nichrome wire had a temperature range
of 245293 C. Statistical analysis of the two coil configura-
tions determined that the temperature output in a contact
versus non-contact configurations for nichrome wire was
not statistically different (p¼0.447). Comparing the two
coil types and the temperature output in Kanthal versus
nichrome wire was determined to not be statistically differ-
ent with varying wire gauge and voltage (p¼0.198).
For the wet heat studies with 100% propylene glycol
(PG), the coil temperature output was also greatly reduced
compared to the dry heats. Kanthal wire in a contact config-
uration had a temperature range of 182386 C with
increasing voltage and wire gauge (Figure 2). In a non-con-
tact configuration, the Kanthal wire had a temperature
range of 148186 C across all voltages and gauges.
Statistical analysis demonstrated that the differences in tem-
perature output of Kanthal in contact and non-contact con-
figurations was statistically significant (p<0.001). For
nichrome wire in a contact configuration, the temperature
of the coils ranged from 171241 C across all voltages and
wire gauges. In a non-contact configuration, the nichrome
INHALATION TOXICOLOGY 3
wire had a temperature output range of 160184 C.
Statistical analysis demonstrated that the difference in tem-
perature output of nichrome coils in contact and non-con-
tact configurations was not statistically significant
(p¼0.138). The temperature output between coil configura-
tions combined with coil metal was statistically significantly
different. Comparing the two coil configurations and the
temperature output in Kanthal versus nichrome wire was
statistically significant when varying wire gauge and volt-
age (p¼0.001).
In a 50:50 PG:VG solution, the 32-gauge Kanthal in a
contact configuration had a temperature of 197240 C, and,
in a non-contact configuration, the range was 190227 C
(Figure 3). In a 50:50 PG:VG medium, the Kanthal wire was
significantly different from coils heated in 100% vegetable
glycerin (p¼0.001), but not significantly different from coils
heated in 100% propylene glycol (p>0.05). For nichrome
wire in a contact configuration, the temperatures ranged
from 204240 C and for a non-contact configuration,
212240 C. For both coil types, the temperatures of the
nichrome coils heated in a 50:50 PG:VG medium was sig-
nificantly different than those heated in either 100% VG or
100% PG (p<0.05).
Coil SEM metal analysis
When heated dry, the metals that showed the most appre-
ciable loss following heating were iron and chromium for
Kanthal A-1 wire and iron and nickel for nichrome wire.
The oxygen content on each coil increased with each burn
cycle. Metal analysis performed on new unheated coils
yielded Kanthal baseline compositions to be 59.29-66.26%
iron, 19.7221.03% chromium, and 3.604.79% oxygen. The
baseline composition of the nichrome wire was
18.8119.69% iron, 51.5054.34% chromium, and
3.444.78% oxygen for the three gauges.
With dry heating, chromium composition in the Kanthal
wire in contact configuration decreased from 19.7221.03%
to 13.4715.53% after one heating cycle and to 5.768.54%
after 50 burn cycles across the three gauges (Figure 4). After
150 heating cycles, the final chromium composition was
4.55%, 2.54%, and 1.86% for 30, 32, and 34-gauge wire,
respectively. The iron composition decreased from
59.5966.62% to 33.5243.17% after one heating cycle and
to 14.8523.50% after 50 heating cycles. After 150 heating
cycles, the metal content for nickel was 23.50%, 14.85%, and
22.15% for the three gauges, respectively (Figure 4). The
total loss of metal in a contact configuration was between
15.2719.17% for chromium and 52.4858.26% for iron. For
each metal, a significant difference in metal composition at
each heating cycle increment was observed across the three
gauges (p<0.05). When comparing the total metal loss
between each wire gauge, for both chromium and iron,
there was a significant difference between 30-gauge and the
32 and 34-gauge wires (p<0.05). There was not a signifi-
cant difference in 32 and 34-gauge wire for total metal loss
for metal for chromium and iron (p¼0.4468, 0.5003).
When heated dry, Kanthal wire in a non-contact config-
uration, after 1 heat the metal content for chromium
dropped from 19.7221.03% to 15.8617.00% across the
three gauges and after 50 heats, to 8.9711.45% (Figure 4).
After 150 heats, the final metal content for chromium was
10.34%, 6.98%, and 5.54% for 30, 32, and 34-gauge wire,
respectively (Figure 3). For iron, the metal content dropped
from 59.5966.62% to 38.6448.33% after one heat cycle
and to 22.9832.11% after 50 heat cycles. After 150 heat
cycles, the metal content for nickel was 26.66%, 18.02%, and
16.64% for the three gauges, respectively (Figure 4). The
total loss of metal in a non-contact configuration was
between 9.3815.49% for chromium and 39.9646.10% for
Figure 2. Maximum temperature outputs of Kanthal in contact (C) and non-contact (NC) configurations in three mediums; Dry (A), 100% VG (B), and 100% PG (C).
Maximum temperature outputs of nichrome in contact and non-contact configurations in three mediums: Dry (E), 100% VG (F), and 100% PG (G).
4 H. A. MULDER ET AL.
iron. For each metal, there was a significant difference in
metal composition at each burn stage across the three
gauges (p<0.05). When comparing the total metal loss
between each wire gauge, for both chromium and iron,
there was a significant difference between 30 and 32 gauge
and 30 and 34 gauge (p<0.05). There was not a significant
difference in 32- and 34-gauge wire for total metal loss for
metal for chromium and iron (p¼0.3923, 0.8797). For
Kanthal wire, there was a significant difference between the
two configurations (contact and non-contact) at each gauge
for both chromium and iron (p<0.005).
Oxygen composition in a contact configuration for the
Kanthal wire increased from 3.604.74% to 20.1524.04%
after one heating cycle and to 30.6432.79% after 50 heat
cycles across the three gauges (Figure 4). After 150 heating
cycles, the final oxygen composition was 34.46%, 32.79%,
and 31.29% for 30, 32, and 34 G respectively. The total gain
of oxygen in the contact configuration was 30.8535.09%.
Metal composition increased significantly at each heat stage
across the three gauges (p<0.05). When comparing the
total oxygen gain between each wire gauge, there was no
significant difference (p¼0.6261). In a non-contact
configuration, oxygen content increased to 16.3923.03%
after one heat cycle and to 25.3027.79% after 50 heat
cycles. After 150 heat cycles, the final oxygen composition
was 26.17%, 28.53%, and 31.06% for 30, 32, and 34 G
respectively. The total increase of oxygen in the non-contact
configuration was 22.5726.32%. With all three gauges,
there was not a significant difference in the oxygen levels
from 50 to 150 heat cycles (p¼0.9393, 0.0819, and 0.2824
for 30, 32, and 34 G respectively). For 34 gauge, there was
also no significant difference between the oxygen levels
from one heat cycle to 50 heat cycles (p¼0.0573). When
comparing the total oxygen increase between each wire
gauge, there was no significant difference (p¼0.2848). For
Kanthal wire, there was a significant difference (p<0.05)
between the two configurations (contact and non-contact) at
each gauge at 150 heat cycles. For 30 and 32 gauge, there
was also a significant difference in oxygen accumulation at
50 heat cycles.
For nichrome wire in a contact configuration, after one -
dry heat cycle, the metal content for iron dropped from
18.8119.09% to 11.2216.91% across the three gauges and
after 50 dry heat cycles, to 5.328.19% (Figure 4). After
Figure 3. Maximum temperature outputs of 32 G Kanthal and nichrome wire in contact (C) and non-contact (NC) configurations in three mediums; 100% VG (A),
50:50 PG:VG (B), and 100% PG (C).
Figure 4. Metal loss of Kanthal (A, B, and C) and nichrome wire (D, E, and F) in contact (C) and non-contact (NC) configurations after one, 50, and 150 burns in a
dry medium.
INHALATION TOXICOLOGY 5
150 dry heat cycles, the final metal content for chromium
was 5.57%, 7.01%, and 5.75% for 30-, 32-, and 34-gauge
wire, respectively (Figure 3). For nickel, the metal content
dropped from 51.5054.30% to 21.0337.41% after 1 heat
and to 9.1614.41% after 50 heat cycles. After 150 heat
cycles, the metal content for nickel was 10.84%, 11.51%, and
9.88% for the three gauges, respectively (Figure 4). The total
loss of metal in a contact configuration was between
12.0514.12% for iron and 40.77-43.50% for nickel. For each
metal, there was a significant difference in metal compos-
ition after the first and 50th heat cycles across the three
gauges (p<0.05). However, there was not a significant dif-
ference in the loss of metal for nickel and iron from 50 to
150 heat cycles. When comparing the total metal loss
between each wire gauge, for both iron and nickel, there
was no significant difference between the three
gauges (p>0.05).
For nichrome wire in a non-contact configuration, after
1 heat cycle, the metal content for iron dropped from
18.8119.09% to 13.0415.77% across the three gauges and
after 50 burns, to 5.708.37% (Figure 2). After 150 heat
cycles, the final metal content for chromium was 12.13%,
12.87%, and 14.11% for 30-, 32-, and 34-gauge wire, respect-
ively (Figure 4). For nickel, the metal content dropped from
51.5054.30% to 21.0336.99% after one heat cycle and to
17.559.16% after 50 heat cycles (Figure 4). After 150 heat
cycles, the metal content for nickel was 14.76%, 7.59%, and
9.88% for the three gauges, respectively (Figure 3). The total
loss of metal in a non-contact configuration was between
12.1314.11% for iron and 39.5844.69% for nickel. For
each metal, there was a significant difference in metal com-
position after the first and 50th heat cycle across the three
gauges (p<0.05). However, a significant difference in the
loss of metal for nickel and iron from 50 to 150 heats was
not observed. When comparing the total metal loss between
each wire gauge, for nickel there was not a significant differ-
ence between 30- and 32-gauge wire (p¼0.5632), but a sig-
nificant difference between 30- and 34-gauge and 32- and
34-gauge was observed (p<0.05). For iron, there was a sig-
nificant difference when comparing 30-gauge wire to 32-
and 34-gauge wire (p<0.05), but not between 32- and 34-
gauge wire (p¼0.8105). When comparing contact and non-
contact configurations of nichrome wire, a significant differ-
ence in metal loss between the two configurations for both
iron and nickel was observed. However, for 32- and 34-
gauge wire, there was a significant difference for iron loss
(p<0.05), but there was no significant difference for nickel
metal loss (p>0.05).
Oxygen composition in a contact configuration for the
nichrome wire increased from 3.444.78% to 18.0427.01%
after one heat cycle and to 29.1230.14% after 50 heat cycles
(Figure 4). After 150 heating cycles, the final oxygen com-
position was 29.99%, 27.36%, and 27.95% at 30, 32, and 34
gauges, respectively. The total gain of oxygen in a contact
configuration was 23.4125.21%. With 30 and 32 gauge,
there was a significant difference in oxygen levels between
new, one heat cycle, and 50 heat cycles. (p<0.05). For 34
gauge contact, there was only a significant difference in
oxygen levels between new and one heat cycle. There was
not a significant difference in the amount of oxygen accu-
mulated between the three gauges (p¼0.5682). For a non-
contact configuration, oxygen levels increased to
20.9125.60% after one heat cycle and 24.1234.03% after
50 heat cycles. After 150 heat cycles, the oxygen levels were
22.19%, 24.48%, and 24.06% for 30, 32, and 34 gauge
respectively. The total increase of oxygen in a non-contact
configuration was 17.4120.62%. For 30 and 34 gauges,
there was no significant difference in oxygen levels after
one, 50 and 150 heat cycles (p>0.05). For 32 gauge wire,
there was a significant difference between new, one, and
fifty heat cycles, but no significant difference between oxy-
gen levels after one heat cycle and 150 heat cycles
(p¼0.6427). There was not a significant difference in the
amount of oxygen accumulated between the three gauges
(p¼0.2968). For nichrome, 30- and 34-gauge wires did not
have a significant difference between the two configurations
(contact and non-contact), but 32 gauge had a significant
difference in oxygen levels between the two configurations
at one heat cycle and 50 heat cycles (p<0.05).
The significance of the metal loss with increasing heat
cycles could be seen visually via SEM analysis. Figure 5
shows both 30-gauge Kanthal and nichrome wire at new/
pre-heated and after one, 50, and then 150 heats. Both wires
exhibit a smooth surface prior to burning and even after the
first burn, the metal begins to show signs of decay and fis-
sures. By the 150
th
heat cycle, both wires are pitted and
have rough surfaces. In some instances, charring could be
seen visually prior to SEM analysis. When all coils were
heated while with a 50:50 PG:VG e-liquid, no statistical dif-
ference was observed between contact and non-contact con-
figurations of either Kanthal or nichrome wires at either 1
or 50 heat cycles (p>0.05). Additionally, no statistical dif-
ference was observed between one and 50 heat
cycles (p>0.05).
Under a wet media of 50:50 PG:VG, the coils exhibited
less visible deterioration under SEM when compared with
the dry heat cycles. Statistical comparison of new, one, and
50 heat cycles were not statistically different in a wet media.
However, consecutive loops within coils displayed varying
degrees of deterioration (Figure 6). The difference in the
two medias (wet vs. dry) was significant after 50 heat
cycles (p<0.05).
Discussion
Coil temperature studies
Online forums for e-cigarette consumers who use rebuild-
able atomizers (RBA) have created numerous posts discus-
sing whether Kanthal A-1 and nichrome creates the best
experience. Most participants on those forums will claim
that the choice between wire type is a personal one.
However, it has been highlighted by consumers that Kanthal
wire can withstand higher temperatures and thus, lasts lon-
ger than nichrome (Reddit 2015). The other coil topic fre-
quently discussed in these forums is the configuration of the
coil. Most RBA users will say that contact configuration
6 H. A. MULDER ET AL.
coils will burn hotter than a non-contact coil, and that the
non-contact coil will create a more even burn across the
wick and coil (Contact Micro Coils Vs Spaced Coils 2014).
Consumers also say that vaping with a contact coil will lead
to a dry burnor heating the wick to the point where there
is no e-liquid left to aerosolize. Further, assessing the coil
temperature outputs while dry is critical given that a model
called the rebuildable dripper atomizer (RDA) enables users
to drip an e-liquid directly onto a coil or coil/wick system
(Poklis et al. 2017; Harrell and Eissenberg 2018).
When assessing the coil temperatures heated in the
absence of a liquid, Kanthal wire reached higher tempera-
tures than nichrome and could withstand greater voltages.
Heating coils in a dry medium and with increasing voltage
and wire gauge showed that contact coils would reach a
temperature where they would melt or burn outand cease
Figure 6. Coil surface morphology - SEM images of contact and non-contact configurations of Kanthal wire coils after one (A) 50 (B) wet heating cycles and
nichrome wire coils after one (C) and 50 (C) wet heating cycles.
Figure 5. Coil surface morphology - SEM images of Kanthal wire coils new (A & B), and after 1 (C), 50 (D), and 150 (E) dry heating cycles (2000x magnification) and
nichrome wire coils new (F & G), and after 1 (H), 50 (I), and 150 (J) dry heating cycles.
INHALATION TOXICOLOGY 7
to work above a certain voltage (Figure 1). For Kanthal
wire, 34-gauge was the only wire in a contact configuration
that could not withstand the highest voltage, 5 V. For
nichrome, only 30-gauge wire, the thickest wire, could with-
stand the temperatures produced from all four voltages.
However, in a wet medium, both coils were able to with-
stand the same amount of voltage applied in both 100% PG
and 100% VG e-liquids.
In 100% VG, no significant difference between the coil
type and configuration in regards to their observed tempera-
ture outputs was observed (p¼0.198). However, in 100%
PG, the two wire types had a statistically significant differ-
ence in temperature outputs. Kanthal also showed a differ-
ence in temperature output between the two coil
configurations. Propylene glycol is less viscous than vege-
table glycerin. As it was previously mentioned, Kanthal
reaches higher temperatures in a dry medium when com-
pared to nichrome wire and a higher temperature in its con-
tact configuration over non-contact. The less viscous
properties of propylene glycol allow Kanthal coils to reach a
higher temperature than nichrome during the eight second
heating period used in this study. At these higher tempera-
tures, a greater risk of dry burningthe coil in 100% PG e-
liquids exits. Some consumers who have recently transi-
tioned to e-cigarettes from traditional cigarette usage prefer
100% PG-based e-liquids, citing that the vapor produced
emulates a traditional smoking experience (Legostar 2013).
The 50:50 PG:VG ratio data for Kanthal wire showed a
similar trend in comparison to 100% VG vs 100% PG
(Figure 2). No significant difference in the temperature out-
put of the coils when compared to 100% VG was observed,
but when compared to the 100% PG medium, Kanthal tem-
perature outputs were statistically different. Similar 100%
VG and 100% PG, the difference in media viscosity influen-
ces the temperature outputs of the coils and their ability to
conduct heat.
Coil SEM analysis
SEM data gathered in this study showed a significant metal
loss for iron, chromium, and nickel in the coils after being
heated while dry. The metal loss from the coils may be pre-
sent in aerosol inhaled by the e-cigarette user. The metals
that exhibited composition changes in this study were con-
sistent with metals identified and quantified in previous
studies (Williams et al. 2013; Goniewicz et al 2014; Hess
et al. 2017; Olmedo et al. 2018). When reporting studies
about metals in e-cigarette aerosols, it is important to
include the coil specifications. The coil type and configur-
ation will have an effect on which metals are present and
the amount of metal present in the aerosol. Previous studies
have highlighted the dangers of nickel and chromium being
inhaled via e-cigarette aerosol. Nickel and chromium are
both carcinogenic and have been linked to illnesses such as
chronic bronchitis and lung cancer (Hess et al. 2017).
Dry heating a coil is a common technique used with
rebuildable atomizers. It allows the user to shape the coil
and ensure even heating across it when building. It is also
used to cleanthe coil of e-liquid build-up when changing
or refilling the tank. Online forums have reported that the
build-up of e-liquids primarily happens on coils that are
built in a contact configuration. This study determined that
there was a significant metal loss from newly constructed
coils in both contact and non-contact configurations after
just one dry heat cycle. Further, a significant difference in
the amount of metal loss between the contact and non-con-
tact configurations was observed. Figure 3 shows that con-
tact configurations had a greater metal loss overtime when
compared to non-contact configurations. This metal loss
and differences between wire type and configuration is sup-
ported by the temperature differences in the coil tempera-
tures when dry (Figure 2). The dry burning method for
coils has caused members of the scientific community to
caution e-cigarette users from pursuing this method, citing
that the possible metal loss from the intense heat of the dry
burn could be inhaled by the users (Farasalinos 2015).
Changes to the coil from dry heating over time showed
not only metal loss, but visual changes to the coils. SEM
images showed increased surface pitting and the formation
of a white solid, indicating oxidation. SEM images of the
coils showed that pieces of the wire were missing, possibly
due to disintegration. Oxygen levels increased with each
heat cycle and an increase of aluminum in nichrome and
Kanthal, respectively, was also observed. Kanthal and
nichrome wire in a contact configuration both showed sig-
nificant oxygen changes after the first heat cycle. Nichrome
wire, however, showed that by the 50 and 150 heat cycles,
oxygen levels did not change significantly, which was also
seen in both wire types noncontact configurations. The
apparent loss of oxygen seen in nichrome wire between 50
and 150 heat cycles may be due to the metal degrading
over time.
The increase in oxygen content is possibly due to the oxi-
dation of the coils as they are heated. Nichrome wire will
form a chromium oxide on the outer side of the coil as it is
heated and Kanthal will form an aluminum oxide coating
(Kanthal 2018b). The formation of these oxides are intended
to protect the coil from further oxidation, however, the
metal oxides can be inhaled into deep lung tissue, where
injury is likely to occur. A particle size analysis of e-cigarette
aerosol shows that in the three e-liquid compositions pre-
sented in this study (100% VG, PG, and 50:50 PG:VG),
aerosol particles were less than 5 mm in size, demonstrating
that the aerosol deposits in the deep lung tissue (Mulder
et al. 2020). Further, other studies that assessed metal aero-
sol content from e-cigarettes using Kanthal and nichrome
wire also observed aluminum and chromium, suggesting
that oxidized metals can still be inhaled by the e-cigarette
user (Olmedo et al. 2018; Williams et al. 2019).
The SEM-EDX data from wet heat cycles demonstrated
no significant difference in metal deterioration between coil
types and configurations. This supports the temperature
data which also did not change significantly between the
coil types and configuration. In the end, user preferences
are likely a reflection of subconscious behavioral effects.
8 H. A. MULDER ET AL.
Conclusions
Despite enthusiastic conversations about e-cigarette coil
types and configurations, in the presence of e-liquids, there
is no significant difference between the two types of wire
and configurations on temperature output or as observed by
SEM-EDX. Dry burning has been shown to cause significant
metal loss from the coil as early as one heat cycle and can
potentially be inhaled by the e-cigarette consumer in the
process of building or cleaning their coil. The metals identi-
fied on the coils are known to cause negative health effects
when inhaled over time. It is also important that in future
studies that report the presence and concentrations of metal
in e-cigarette aerosol include all specifications about the
atomizer, including the heating element. This is because the
metal compositions will affect the presence and abundance
of certain metals in the aerosol.
Disclosure statement
The authors have no conflicts of interest to disclose.
Funding
This project was supported by the National Institute of Justice, Office
of Justice Programs, U.S. Department of Justice [Award No. 2014-R2-
CX-K010 and 2016-DN-BX-0150] and the National Institutes of Health
[Award No. P30DA033934]. The opinions, findings, and conclusions
or recommendations expressed in this publication/program/exhibition
are those of the author(s) and do not necessarily reflect those of the
Department of Justice.
ORCID
Justin L. Poklis http://orcid.org/0000-0001-5470-5717
References
ASTDR 2005. Toxicological profile for Nickel. Atlanta (GA): U.S.
Department of Health and Human Services PHS.
ASTDR 2012. Toxicological profile for chromium. Atlanta (GA): U.S.
Department of Health and Human Services PHS.
Bare Nickel Chromium Resistance Wire (BNC). 2020. Consolidated
electronic wire and cable. http://products.conwire.com/viewitems/
bnc-bare-nickel-chromium/bare-nickel-chromium-resistance-wire-
bnc-?
Breland A, Soule E, Lopez A, Ramoa C, El-Hellani A, Eissenberg T.
2017. Electronic cigarettes: what are they and what do they do? Ann
N Y Acad Sci. 1394 :530.
Contact Micro Coils Vs Spaced Coils. 2014. Oct E-cigarette forum.
https://www.e-cigarette-forum.com/threads/contact-micro-coils-vs-
spaced-coils.61
Farasalinos KE. 2015. Dry-burning metal coils: is it a good thing? E-
Cigarette Research. http://www.ecigarette-research.org/research/
index.php/research/research-2015/212-db0692/
Goniewicz ML, Knysak J, Gawron M, Kosmider L, Sobczak A, Kurek J,
Prokopowicz A, Jablonska-Czapla M, Rosik-Dulewska C, Havel C,
et al. 2014. Levels of selected carcinogens and toxicants in vapour
from electronic cigarettes. Tob Control. 23:133139.
Grana R, Benowitz N, Glantz S. 2013. Background paper on e-cigar-
tettes (electronic nicotine delivery systems). World Health
Organization Tobacco Free Initiative.
Halstead M, Gray N, Gonzalez-Jimenez N, Fresquez M, Valentin-
Blasini L, Watson C, Pappas RS. 2019. Analysis of toxic metals in
electronic cigarette aerosols using a novel trap design. J Anal
Toxicol. 2020;44:149155.
Harrell PT, Eissenberg T. 2018. Automated dripping devices for vapers:
RDTAs, bottomfeeders, squonk mods and dripboxes. Tob Control.
27:480482.
Hess CA, Olmedo P, Navas-Acien A, Goessler W, Cohen JE, Rule AM.
2017. E-cigarettes as a source of toxic and potentially carcinogenic
metals. Environ Res. 152:221225.
Kanthal. 2018a. Kanthal A-1 resistance heating wire and resistance
wire datasheet. https://www.kanthal.com/en/products/material-
datasheets/
Kanthal. 2018b. Resistance heating alloys for electric home appliances.
www.smt.sandvik.com
Kornberg TG, Stueckle TA, Antonini JM, Rojanasakul Y, Castranova
V, Yang Y, Rojanasakul LW. 2017. Potential toxicity and underlying
mechanisms associated with pulmonary exposure to iron oxide
nanoparticles: conflicting literature and unclear risk. Nanomaterials.
7:307.
Legostar. 2013. P.G Throat Hit. E-Cigarette Forum. https://www.e-cig-
arette-forum.com/forum/threads/ph-throat-hit.410670
Mulder HA, Patterson JL, Halquist MS, Kosmider L, Turner JB, Poklis
JL, Poklis A, Peace MR. 2020. The effect of electronic cigarette user
modifications and e-liquid adulteration on the particle size profile
of an aerosolized product. Sci Rep. 9:10221.
Olmedo P, Goessler W, Tanda S, Grau-Perez M, Jarmul S, Aherrera A,
Chen R, Hilpert M, Cohen JE, Navas-Acien A, et al. 2018. Metal
concentrations in e-cigarette liquid and aerosol samples: the contri-
bution of metallic coils. Environ Health Perspect. 126:027010.
Peace MR, Baird TR, Smith N, Wolf CE, Poklis JL, Poklis A. 2016.
Concentration of nicotine and glycols in 27 electronic cigarette for-
mulations. J Anal Toxicol. 40:403407.
Poklis JL, Mulder HA, Halquist MS, Wolf CE, Poklis A, Peace MR.
2017. The blue lotus flower (Nymphea caerulea) resin used in a new
type of electronic cigarette, the re-buildable dripping atomizer. J
Psychoactive Drugs. 49:175181.
Reddit. 2015. Nichrome vs Kanthal pros and cons? https://www.reddit.
com/r/electronic_cigarette/comments/2kwx9j/nichrome_vs_kanthal_
pros_and_cons/
Williams M, Bozhilov KN, Talbot P. 2019. Analysis of the elements
and metals in multiple generations of electronic cigarette atomizers.
Environ Res. 175:156166.
Williams M, Villarreal A, Bozhilov K, Lin S, Talbot P. 2013. Metal and
silicate particles including nanoparticles are present in electronic
cigarette cartomizer fluid and aerosol. PLoS One. 8:e57987.
Zervas E, Matsouki N, Kyriakopoulos G, Poulopoulos S, Ioannides T,
Katsaounou P. 2020. Transfer of metals in the liquids of electronic
cigarettes. Inhal Toxicol. 32:240248.
INHALATION TOXICOLOGY 9
... These fi ndings by Mulder, et al. suggest that metal lost by the heating coils may enter the aerosols [29]. Hence the presence of these toxic metals in the e-liquids. ...
... These are nondestructive promising technologies for the study of metals in the device. For instance, in 2020, Mulder used SEM to study the metal loss from the heating coils in the e-cig [29]. ...
Article
Full-text available
The evolution of the cigarette industry took us from traditional to electronic cigarettes (e-cigs). These are electronic devices that produce nicotine-containing aerosols by heating a liquid (e-liquid) that contains the drug as well as other numerous compounds. Although designed to decrease the harmful effects of traditional cigarettes, little is known about the long-term health incidence of e-cigarettes. The chemical composition of e-liquids varies depending on their manufacturer, country of origin, and batches. Fundamentally, e-liquids contain glycerol, propylene glycol, water, and nicotine. However, a concerning number of toxic substances including heavy metals, nonmetals, pesticides, polycyclic aromatic hydrocarbons, volatile organic compounds, and many other molecules have been identifi ed in them. A wide number of analytical techniques have been employed in the identifi cation and quantitation of these chemicals in liquids as well as aerosols. This review will discuss the numerous chemical substances associated with the use of e-cigarettes, including the potential impact of these on human health. Additionally, this work will address the available analytical tools as well as promising technologies that can contribute to future toxicological characterization efforts. This review also covers some of the global regulatory policies including their challenges and loopholes.
... Besides coil composition, other factors that might influence the metal migration from the coil towards the e-liquid include applied battery voltage, frequency of coil replacement [22,28], rate of heat transfer within the device [29,30], device type [31], and airflow [27,31]. ...
... Unfortunately, the total metal content does not allow to determine the composition of a passivation layer in Aspire BVC coils. For example, Kanthal A-1 contains Cr, but its passivation layer is made of Al 2 O 3 [30]. The composition of the protective layer might influence the released Cr and Ni concentration towards the e-liquid. ...
Article
Full-text available
Electronic cigarettes are generally recognized as a safer alternative than conventional cigarettes. Nevertheless, previous research suggests metal (loid) leaching due to coil contact, potentially transferring to the e-liquid and its aerosolized form. In this study, Cr, Cd, Ni, and Pb levels were measured by inductively coupled plasma mass spectrometry (ICP-MS) on 17 samples of e-liquids with different chemical properties (e.g., pH, nicotine content, flavoring, free-base, and nicotine salts). Twelve e-liquids were then put in contact with 36-gauge Kanthal A-1, Nichrome 80, Stainless steel 317 L and disposable coils such as Juul, and Aspire BVC for three days at 200–250 °C for 1 h each day. Metal levels expressed as mean (standard deviation) metal concentration, were below detection (Cd) to very low in bottle samples (Ni ≤ 76 (18); Pb ≤ 16 (1.5); and Cr ≤ 386 (15.6) μg/kg). In the coil extracts, varying concentrations of the same metal (loid) were found, indicating that metal leaching capacity may differ per sample. All samples contained Ni and Cr, followed by Pb to a much lesser extent. Cd levels were mostly below detection limits. Coil + e-liquid combinations with the highest Ni, Cr, and Pb concentrations were: Aspire BVC + Melon 0 mg/mL: Ni = 1.22 E+04 (281); Aspire BVC + Hit Nicotine 40 mg/mL: Cr = 864 (116); and Nichrome 80 + Melon 0 mg/mL: Pb = 56 (5) μg/kg. Overall, results suggest that nicotine salts at 40 mg/mL enhance Cr and Ni transfer. Stainless steel 317 L released very low metal concentrations. A conservative screening level risk characterization showed that 10.5% and 3.5% of the coil extracts may exceed Ni and Cr (III) safe concentrations, respectively. In the aerosol phase, 8.8% of samples might be above Ni equivalent daily dose for chronic exposure and 1.8% for intermediate exposure. Further studies on coil metal leaching could aid in establishing coil manufacturing regulations.
... In general, if the wick is kept saturated with e-liquid, ecigarette coil temperatures stay below 300°C. Coils can reach temperatures beyond 1000°C as the wicking material dries out, which accelerates coil breakdown and makes it easier for metals to be released [23]. ...
... The operational procedure involves passing an electric current through a coil of specific resistance, generating heat in the process. The coil reaches hundreds of degrees Celsius [7], causing the e-liquid to aerosolize and allowing the generated aerosol to be inhaled. ...
Article
Full-text available
Because of the increasing popularity of e-cigarettes, monitoring the e-cigarette market has become important for national health authorities to guarantee safety and quality. In the EU, the Tobacco Products Directive requires emission studies for e-cigarette products. The absence of industry guidelines for studying these emissions and the lack of proper validation in the literature led us to develop and validate a method using the total error approach for the determination of nicotine in e-cigarette aerosols. A commercial vaping device was used to generate aerosols, which were then collected on Cambridge filter pads and measured for nicotine concentration by UHPLC-DAD after extraction. The method was successfully validated by generating accuracy profiles, which show that the β-expectation tolerance intervals remained below the acceptance limits of ±20%. Within-run repeatability and intermediate precision were considered acceptable since the highest RSD value obtained was below 5%. The method was applied to 15 commercial e-liquids. A complete validation of a method for the analysis of e-cigarette emissions is presented, including several parameters that impact the accuracy and reproducibility. Similar systematic approaches for method development and validation could be used for other e-cigarette emission analysis methods to ensure the reliability of the measurements.
... Metal aerosol in e-cigarettes are produced from vaporized fluid generated from the heating of metal filaments. These filaments are in general made from nichrome or kanthal (ferritic iron-chromiumaluminum alloy), so metals, such as silver (Ag), aluminum (Al), chromium (Cr), copper (Cu), iron (Fe), nickel (Ni), and zinc (Zn), are expected to be present in e-cigarette aerosols (5,6). A systematic review published in 2020 (7) included 24 studies. ...
Article
Full-text available
Introduction Electronic nicotine delivery systems (ENDS) are gradually becoming more popular, particularly, among today’s youth. Despite being marketed as safe by the tobacco industry, the notable absence of regulation in their composition is evident. Both the generated fluids and aerosol exhibit a wide variety of substances that are not yet fully identified. In addition to additives, the aerosol contains metals, the presence of which can be attributed to the excessive heating of metallic filaments used in vaporizing the liquid. Objective This review aimed to identify and describe studies that have assessed metal levels in biological samples obtained from electronic cigarette users and those exposed to their second-hand aerosol. This involved detailing the types and concentrations of metals identified and the biological samples in which the metals were detected. Methods Two independent researchers conducted searches in the MEDLINE and EMBASE databases to identify studies that measured the metal levels in human non-invasive biological samples from electronic cigarette users and second-hand exposure. Data were presented as a narrative review. Results In total, 18 articles were included in this review. Overall active and passive exposure to ENDS was related to higher levels of many metals, including lead and cadmium, in biological samples. ENDS users, in general, have lower metal concentrations in biological samples compared to the users of combustible cigarettes. Conclusion The exposure to primary and second-hand e-cigarette aerosol is related to higher metal concentrations in the biological samples. The adverse effects of this exposure on long-term users are yet to be determined.
... If coils exceed 1000°C [57], coils degrade rapidly and facilitate the release of metals. If 150 heat cycles were performed, chromium and iron from Kanthal coils would be lost up to 19% and 58%, respectively, and iron and nickel from nichrome coils up to 14% and 43%, respectively [58]. Studies [55] have found most metal/metalloid biosample levels in e-cigarette users were similar or even higher in comparison with tobacco cigarette users and higher in comparison with cigar users. ...
Article
Full-text available
Electronic cigarettes (e-cigarettes), as alternative nicotine delivery methods, has rapidly increased among youth and adults in recent years. However, cardiovascular safety is an important consideration regarding e-cigarettes usage. e-cigarette emissions, including nicotine, propylene glycol, flavorings, nitrosamine, and metals, might have adverse effects on cardiovascular health. A large body of epidemiological evidence has indicated that e-cigarettes are considered an independent risk factor for increased rates of cardiovascular disease occurrence and death. The incidence and mortality of various types of cardiovascular disease, such as cardiac arrhythmia, hypertension, acute coronary syndromes, and heart failure, have a modest growth in vapers (users of e-cigarettes). Although the underlying biological mechanisms have not been fully understood, studies have validated that oxidative stress, inflammation, endothelial dysfunction, atherosclerosis, hemodynamic effects, and platelet function play important roles in which e-cigarettes work in the human body. This minireview consolidates and discusses the epidemiological and biological links between e-cigarettes and various types of cardiovascular disease.
... In cigarettes where temperatures can reach 900 • C [1], Dempsey et al. [2] have provided a historical review of these flavor chemical evaluation efforts, which have generally incorporated evaluating the pyrolysisderived by-products [3,4]. As reported in the literature, electronic nicotine delivery systems (ENDSs) utilize substantially lower temperatures to generate an aerosol, typically under 300 • C [5][6][7][8][9][10]. However, non-temperature-regulated devices may exceed 300 • C, leading to the production of higher levels and different thermal degradation products [11]. ...
Article
Full-text available
Toxicological evaluations of flavor chemicals for use in inhalation products that utilize heat for aerosol generation are complicated because of the potential effect heat may have on the flavor chemical. The objective was to develop a thermal degradation technique to screen flavor chemicals as part of a toxicological testing program for their potential use in ENDS formulations. Based upon published data for acetaldehyde, acrolein, and glycidol from ENDS products (common thermal degradants of propylene glycol and glycerin), the pyrolizer temperature was adjusted until a similar ratio of acetaldehyde, acrolein, and glycidol was obtained from a 60/40 ratio (v/v) of glycerin/propylene glycol via GC/MS analysis. For each of 90 flavor chemicals, quantitative measurements of acetaldehyde, acrolein, and glycidol, in addition to semiquantitative non-targeted analysis tentatively identifying chemicals from thermal degradation, were obtained. Twenty flavor chemicals transferred at greater than 99% intact, another 26 transferred at greater than 95% intact, and another 15 flavor chemicals transferred at greater than 90% intact. Most flavor chemicals resulted in fewer than 10–12 tentatively identified thermal degradants. The practical approach to the thermal degradation of flavor chemicals provided useful information as part of the toxicological evaluation of flavor chemicals for potential use in ENDS formulations.
Article
Electronic cigarettes (ECs) emit many toxic substances, including metals, that can pose a threat to users and the environment. The toxicity of the emitted metals depends on their oxidation states. Hence, this study examines the oxidation states of metals observed in EC aerosols. X-ray photoelectron spectroscopy analysis of the filters that collected EC aerosols identified the oxidation states of five primary metals (based on surface sample analysis), including chromium(III) (close to 100%) under low power setting while a noticeable amount of chromium(VI) (15%) at higher power settings of the EC, and copper(II) (100%), zinc(II) (100%), nickel(II) (100%), lead(II) (65%), and lead(IV) (35%) regardless of power settings. This observation indicates that the increased temperature due to higher power settings could alter the oxidation states of certain metals. We noted that many metals were in their lesser toxic states; however, inhaling these metals may still pose health risks.
Article
Full-text available
Indoor particle number size distribution (0.3-10 μm), equivalent black carbon (eBC), and Ångström absorption exponent (AAE) data were collected in real conditions, over a ten-month period at a research area building, in a semirural site, to characterize indoor aerosol loading. Additionally, during the campaign, emissions from four indoor sources commonly used at the site (incense, traditional cigarettes, electronic cigarettes, and heat-not-burn products) were studied during short-term experiments with the support of ultrafine particle (UFP) monitoring. Two particle low-cost sensors (PM LCS), Sensirion SPS30 (0.3-10 μm), were evaluated in the long-term campaign and during fast emission processes, to assess their accuracy and reliability. Penetration and infiltration of both fine and coarse particles from outdoor traffic, domestic heating, and dust resuspension were inferred as the main sources of indoor aerosols on a long-term basis. Moreover, long-range transported dust aerosols were found to influence indoor coarse number concentration. Among the source events, heat-not-burn (HNB) product resulted in the lowest effect on indoor air quality, whereas the highest AAE values from incense and traditional cigarettes suggest the brown carbon (BrC) production. The highest emission of UFP was caused by electronic cigarettes (e-cig), which spanned particles from the ultrafine to the coarse fractions. This was likely due to the release of metal and silicate from the coil. Analysis of number size distributions of the four experiments revealed the emission of fine particles (0.3-1 μm) and super micron particles. SPS30s performance was satisfactory in terms of accuracy, precision, and durability, indicating that these devices are suitable for monitoring indoor air quality. Additionally, the two PM LCS were able to detect all simulated fast emission sources.
Article
Full-text available
Since 2006 the domestic popularity and sales of electronic cigarettes (i.e., electronic nicotine delivery systems or ENDS) have grown rapidly. Although the constituents of the aerosol produced by ENDS have been previously investigated, differences in puff regimens and aerosol trapping schema in published literature often complicate result comparisons and data interpretation. As the ENDS product designs continue to evolve, there is a critical need to develop and validate robust methodologies for laboratory testing, appropriate aerosol generation and trapping media required for accurate determinations of ENDS aerosol metals deliveries. A simple, high metals purity, fluoropolymer trap was developed and validated that meets standard machine puffing regimen (CORESTA Recommended Method 81) specifications and exhibits negligible acid extractable metal backgrounds. Using a standard machine puffing regimen in combination with a fluoropolymer condensation trap, aerosol was generated and collected from select ENDS devices for analysis of chromium, nickel, copper, zinc, cadmium, tin, and lead with triple quadrupole inductively coupled plasma mass spectrometry. Devices tested spanned a range of commercial products, including flavored variants of JUUL pods, refillable tank systems, rechargeable cartridges, and single-use ENDs devices. Results showed that for aerosols generated under a fixed puffing regimen (50 puffs/collection), metal concentrations ranged from below the detection limits (LOD) to 614 ng copper and 339 ng zinc per 10 puffs. Cadmium concentrations were below LOD for all devices tested. Device specific aerosol levels of Sn and Pb ranged from below LOD to low nanogram levels. Cr and Ni were transported in aerosols at levels equivalent to, or slightly higher than in mainstream cigarette smoke using a standard smoking regimen. The generally lower levels of specific metals, Cd and Pb, transmitted in ENDS aerosols compared to mainstream cigarette smoke reflect possible reduction of harm for smokers who substitute the use of ENDS as cessation devices in place of smoking cigarettes.
Article
Full-text available
Electronic cigarettes (e-cigarettes) are an alternate nicotine delivery system that generate a condensation aerosol to be inhaled by the user. The size of the droplets formed in the aerosol can vary and contributes to drug deposition and ultimate bioavailability in the lung. The growing popularity of e-cigarette products has caused an increase in internet sources promoting the use of drugs other than nicotine (DOTNs) in e-cigarettes. The purpose of this study was to determine the effect of various e-cigarette and e-liquid modifications, such as coil resistance, battery voltage, and glycol and drug formulation, on the aerosol particle size. E-liquids containing 12 mg/mL nicotine prepared in glycol compositions of 100% propylene glycol (PG), 100% vegetable glycerin (VG), or 50:50 PG:VG were aerosolized at three voltages and three coil resistances. Methamphetamine and methadone e-liquids were prepared at 60 mg/mL in 50:50 PG:VG and all e-liquids were aerosolized onto a 10 stage Micro-Orifice Uniform Deposit Impactor. Glycol deposition correlated with drug deposition, and the majority of particles centered between 0.172–0.5 μm in diameter, representing pulmonary deposition. The 100% PG e-liquid produced the largest aerosol particles and the 100% VG and 50:50 PG:VG e-liquids produced ultra-fine particles <0.3 μm. The presence of ultrafine particles indicates that drugs can be aerosolized and reach the pulmonary alveolar regions, highlighting a potential for abuse and risk of overdose with DOTNs aerosolized in an e-cigarette system.
Article
Full-text available
Background: Electronic cigarettes (e-cigarettes) generate an aerosol by heating a solution (e-liquid) with a metallic coil. Whether metals are transferred from the coil to the aerosol is unknown. Objective: Our goal was to investigate the transfer of metals from the heating coil to the e-liquid in the e-cigarette tank and the generated aerosol. Methods: We sampled 56 e-cigarette devices from daily e-cigarette users and obtained samples from the refilling dispenser, aerosol, and remaining e-liquid in the tank. Aerosol liquid was collected via deposition of aerosol droplets in a series of conical pipette tips. Metals were reported as mass fractions (μg/kg) in liquids and converted to mass concentrations (mg/m3) for aerosols. Results: Median metal concentrations (μg/kg) were higher in samples from the aerosol and tank vs. the dispenser (all p<0.001): 16.3 and 31.2 vs. 10.9 for Al; 8.38 and 55.4 vs. <0.5 for Cr; 68.4 and 233 vs. 2.03 for Ni; 14.8 and 40.2 vs. 0.476 for Pb; and 515 and 426 vs. 13.1 for Zn. Mn, Fe, Cu, Sb, and Sn were detectable in most samples. Cd was detected in 0.0, 30.4, and 55.1% of the dispenser, aerosol, and tank samples respectively. Arsenic was detected in 10.7% of dispenser samples (median 26.7μg/kg) and these concentrations were similar in aerosol and tank samples. Aerosol mass concentrations (mg/m3) for the detected metals spanned several orders of magnitude and exceeded current health-based limits in close to 50% or more of the samples for Cr, Mn, Ni, and Pb. Conclusions: Our findings indicate that e-cigarettes are a potential source of exposure to toxic metals (Cr, Ni, and Pb), and to metals that are toxic when inhaled (Mn and Zn). Markedly higher concentrations in the aerosol and tank samples versus the dispenser demonstrate that coil contact induced e-liquid contamination. https://doi.org/10.1289/EHP2175.
Article
Full-text available
Fine/micron-sized iron oxide particulates are incidentally released from a number of industrial processes, including iron ore mining, steel processing, welding, and pyrite production. Some research suggests that occupational exposure to these particulates is linked to an increased risk of adverse respiratory outcomes, whereas other studies suggest that iron oxide is biologically benign. Iron oxide nanoparticles (IONPs), which are less than 100 nm in diameter, have recently surged in use as components of novel drug delivery systems, unique imaging protocols, as environmental catalysts, and for incorporation into thermoplastics. However, the adverse outcomes associated with occupational exposure to IONPs remain relatively unknown. Relevant in vivo studies suggest that pulmonary exposure to IONPs may induce inflammation, pulmonary fibrosis, genotoxicity, and extra-pulmonary effects. This correlates well with in vitro studies that utilize relevant dose, cell type(s), and meaningful end points. A majority of these adverse outcomes are attributed to increased oxidative stress, most likely caused by particle internalization, dissolution, release of free iron ions, and disruption of iron homeostasis. However, because the overall toxicity profile of IONPs is not well understood, it is difficult to set safe exposure limit recommendations that would be adequate for the protection of at-risk workers. This review article will focus on known risks following IONPs exposure supported by human, animal, and cell culture-based studies, the potential challenges intrinsic to IONPs toxicity assessment, and how these may contribute to the poorly characterized IONPs toxicity profile.
Article
Full-text available
The blue lotus flower (Nymphea caerulea) is an Egyptian water lily containing apomorphine and nuciferine. Apomorphine has been described as a psychoactive alkaloid and is a non-selective dopamine agonist primarily used to treat Parkinson’s disease as it stimulates dopamine receptors and improves motor function. Nuciferine is an alkaloid associated with dopamine receptor blockade. Today, blue lotus flower is used as a sleep aid and anxiety reliever. The rebuildable dripping atomizer (RDA) is an electronic cigarette that allows direct application of an e-liquid onto the coil in the atomizer for aerosolization, compared to a typical electronic cigarette where the e-liquid is wicked from a storage vessel to the coil. Our laboratory received a dark-brown resin material from a concerned parent. The resin had been confiscated from an adolescent who had a reported history of marijuana use. The resin was later identified as blue lotus flower (N. caerulea). This resin, together with four commercially available blue lotus products, was analyzed for content. Apomorphine was detected in two samples, and nuciferine was detected in all five samples. The confiscated resin was determined to contain no apomorphine and 4300 ng/g of nuciferine. The nuciferine resin was shown to aerosolize using aRDA electric cigarette.
Article
Full-text available
Personal battery-powered vaporizers or electronic cigarettes were developed to deliver a nicotine vapor such that smokers could simulate smoking tobacco without the inherent pathology of inhaled tobacco smoke. Electronic cigarettes and their e-cigarette liquid formulations are virtually unregulated. These formulations are typically composed of propylene glycol and/or glycerin, flavoring components and an active drug, such as nicotine. Twenty-seven e-cigarette liquid formulations that contain nicotine between 6 and 22 mg/L were acquired within the USA and analyzed by various methods to determine their contents. They were screened by Direct Analysis in Real Time™ Mass Spectrometry (DART-MS). Nicotine was confirmed and quantitated by high-performance liquid chromatography–tandem mass spectrometry, and the glycol composition was confirmed and quantitated by gas chromatography–mass spectrometry. The DART-MS screening method was able to consistently identify the exact mass peaks resulting from the protonated molecular ion of nicotine, glycol and a number of flavor additives within 5 mmu. Nicotine concentrations were determined to range from 45 to 131% of the stated label concentration, with 18 of the 27 have >10% variance. Glycol composition was generally accurate to the product description, with only one exception where the propylene glycol to glycerin percentage ratio was stated as 50:50 and the determined concentration of propylene glycol to glycerin was 81:19 (% v/v). No unlabeled glycols were detected in these formulations. LINK TO FREE ARTICLE ACCESS: http://jat.oxfordjournals.org/cgi/reprint/bkw037? ijkey=8BDu5z9k19hiLJZ&keytype=ref
Article
ABSTRACT Objectives: E-cigarettes are electronic devices containing a liquid that usually consists of a mixture of glycerol, propylene glycol and nicotine, with or without flavorings, in various concentrations. A vapor or aerosol is produced, and inhaled from the user, when this liquid is heated by a heating coil. This work examines the impact of three parameters (e-liquid composition, nicotine content and air flow) on the transfer of metals’ from the heating coils to the e-liquids. Materials and methods: A distillation unit was used, where 20ml of an e-liquid were boiled with two commercial heating elements. Four e-liquids: 100% Propylene Glycol, 100% Glycerol, 50/50% Propylene Glycol/Glycerol, 33.3/33.3/33.3% Propylene Glycol/Glycerol/Water, three nicotine contents: 0, 0.4 and 0.8% per volume and three air flows: 0, 0.5 and 1.0 L/min, were used. The liquids were analyzed by Total Reflection X-Ray Fluorescence spectrometry to determine the final content of metals. Results and discussion: Five metals, Fe, Ni, Cu, Zn, and Pb, were found to be transferred from the heating coils to the e-liquids. The transfer of those metals increases with air flow and nicotine concentration, while e-liquid composition also has a significant impact. Glycerol enhances the transfer of metals compared to propylene glycol and their mixtures. The boiling temperature of the e-liquids increases significantly the transfer of metals in the e-liquids. Conclusions: There is a transfer of metals from the heating coils to the e-liquids. This transfer depends on the e-liquid composition and on the boiling temperature.
Article
Background: Since their release in 2004, electronic cigarettes (ECs) and their atomizers have undergone significant evolution. Objective: The purpose of this study was to evaluate and compare the elemental/metal composition of atomizers in cartomizer and tank style ECs produced over a 5-year period. Methods: Popular cartomizer and tank models of ECs were dissected and photographed using a stereoscopic microscope, and elemental analysis of EC atomizers was done using scanning electron microscopy coupled with energy dispersive x-ray spectroscopy. Results: Eight elements/metals were found in most products across and within brands purchased at different times. These included chromium, nickel, copper, silver, tin, silicon, aluminum, and zinc. Iron and lead were found in some but not all products, while manganese, cobalt, molybdenum, titanium, and tungsten were only found in a few of the products. The metals used in various components were often similar in cartomizer and tank models. Filaments were usually chromium and nickel (nichrome), although in some newer products, the filament also contained iron, copper, and manganese. The thick wire in earlier products was usually copper coated with silver, while in some newer products, the thick wire was predominantly nickel. In all products, the wick was silica, and sheaths, when present, were fiberglass (silicon, oxygen, calcium, aluminum, magnesium). Wire-to-wire joints were either brazed or clamped with brass (copper and zinc), and air-tube-to-thick wire joints, when present, were usually soldered with tin. Tank style products generally lacked a thick wire and sheaths. Conclusion: In general, atomizer components in ECs were remarkably similar over time and between brands. Certain elements/metals were consistently found in most models from all generations, and these should be studied carefully to determine if their transfer to aerosols affects user's health and if their accumulation in trash affects the environment.