Every year, 60,000-100,000 Americans die of complications arising from deep vein thrombosis (DVT), blood clots that form in deep lower extremity veins. DVT is the leading cause of preventable hospital death for trauma patients, and despite 0.1% of the population receiving a new diagnosis annually, DVT is underdiagnosed due to nonspecific symptoms, diagnostic difficulty, and overburdened radiologists, especially when examining trauma patients with other urgent conditions. Developing computer-aided detection (CAD) systems would alleviate these concerns; however, current systems classify computed tomography (CT) slices individually and yield many false positives from DVT-like dark spots in poorly contrasted areas. Recent advances in deep neural networks (DNNs) allow us to leverage spatial dependencies between slices in imaging studies to identify false positives and ultimately deploy DNN systems that lighten physicians workloads while not exacerbating alarm fatigue. To train our DNN, we have acquired 119 lower-body CT imaging studies labeled by radiol-ogists for DVT at the pixel level. Using these studies, we have developed a DNN-based CAD system that will (1) segment targeted deep veins in a CT slice, (2) classify whether a DVT is present within multiple slices given segmentations of deep veins, and (3) evaluate different deep learning approaches for handling 3D datasets for DVT detection. For segmen-tation, we use a 2D U-Net, 2D VGG encoder-decoder, and 3D U-Net and find that VGG performs best (Dice: 0.07, IoU: 0.48, AUROC: 0.78). For classification, we use a 2D ResNet, CNN-RNN, and 3D Inception model with and without segmentation masks. We find that our CNN-RNN without masks performs best in AUROC (Average Precision: 0.31, AUROC 0.64) and 3D Inception with masks performs best in average precision (Average Precision: 0.33, AUROC: 0.62). By developing more effective detection algorithms, we hope to ensure more frequent and accurate diagnosis of DVT, thereby reducing its high mortality rate.
Content may be subject to copyright.
Deep Vein Thrombosis Screening with Three-Dimensional Deep Learning on
Lower Extremity Computed Tomography Studies
Brianna Kozemzak, Anoop Manjunath, Trevor Tsue, and Jonathan X Wang
Department of Biomedical Data Science, Stanford University,
Stanford, CA, 94305, USA
kozemzak@stanford.edu, amanjuna@stanford.edu, ttsue@stanford.edu, jonxwang@stanford.edu
www.stanford.edu
Equal Contributors
Every year, 60,000-100,000 Americans die of complications arising from deep vein thrombo-
sis (DVT), blood clots that form in deep lower extremity veins. DVT is the leading cause of
preventable hospital death for trauma patients, and despite 0.1% of the population receiving
a new diagnosis annually, DVT is underdiagnosed due to nonspecific symptoms, diagnostic
difficulty, and overburdened radiologists, especially when examining trauma patients with
other urgent conditions. Developing computer-aided detection (CAD) systems would alle-
viate these concerns; however, current systems classify computed tomography (CT) slices
individually and yield many false positives from DVT-like dark spots in poorly contrasted
areas. Recent advances in deep neural networks (DNNs) allow us to leverage spatial depen-
dencies between slices in imaging studies to identify false positives and ultimately deploy
DNN systems that lighten physicians workloads while not exacerbating alarm fatigue. To
train our DNN, we have acquired 119 lower-body CT imaging studies labeled by radiol-
ogists for DVT at the pixel level. Using these studies, we have developed a DNN-based
CAD system that will (1) segment targeted deep veins in a CT slice, (2) classify whether a
DVT is present within multiple slices given segmentations of deep veins, and (3) evaluate
different deep learning approaches for handling 3D datasets for DVT detection. For segmen-
tation, we use a 2D U-Net, 2D VGG encoder-decoder, and 3D U-Net and find that VGG
performs best (Dice: 0.07, IoU: 0.48, AUROC: 0.78). For classification, we use a 2D ResNet,
CNN-RNN, and 3D Inception model with and without segmentation masks. We find that
our CNN-RNN without masks performs best in AUROC (Average Precision: 0.31, AUROC
0.64) and 3D Inception with masks performs best in average precision (Average Precision:
0.33, AUROC: 0.62). By developing more effective detection algorithms, we hope to ensure
more frequent and accurate diagnosis of DVT, thereby reducing its high mortality rate.
Keywords: Deep Vein Thrombosis; Computer Vision; Deep Learning
1. Introduction
Deep vein thrombosis (DVT) is a blood clot commonly found in deep veins of the lower
extremities. Approximately 2.5-5% of individuals in the United States will be affected by DVT
during their lifetime.1Pulmonary Embolism (PE) is a fatal complication that can arise when
DVT goes undetected or is left untreated, and is caused by clots breaking off and traveling into
c
2016 The Authors. Open Access chapter published by World Scientific Publishing Company and
distributed under the terms of the Creative Commons Attribution Non-Commercial (CC BY-NC)
4.0 License.
the lungs.2According to some studies, it is the leading cause of preventable hospital death for
trauma patients3,4 and the leading cause of maternal mortality,5resulting in 60,000-100,000
American deaths annually.6
Despite its high prevalence, DVT continues to be under-diagnosed in the clinical setting.6
One reason for this is the nonspecific physical symptoms of DVT, including swelling, pain,
tenderness, and redness in the area of the thrombosis.7Another reason arises due to the
diagnostic difficulty. Common risk factors include injury and trauma, where DVT can be
overshadowed by the primary diagnoses.4A study of severe trauma patients found that three
cases of DVT and three cases of PE were missed when 205 CT scans were reviewed by an
independent radiologist.4Finally, physician burnout has been on the rise, and even though
half of blood clots form following a clinical visit, the sheer number of cases and immediate risks
commonly associated with DVT (e.g. obesity, pregnancy, trauma) overburden physicians.8,9
One potential solution to these problems are computer-aided detection (CAD) systems
whereby patients at risk are automatically screened for blood clots (and potentially other
diagnoses). Although Ultrasound (US) is most commonly used for the diagnosis of DVT, the
results of recent studies suggest that Computed Tomography (CT) venography studies can
have equal or greater diagnostic power and may be preferable for patients who are also being
examined for PE, in which a CT angiography is required, rather than performing two separate
diagnostic tests.10,11
Fig. 1: Inputs, outputs, and models used to ac-
complish the two primary tasks in this study:
segmentation and classification.
A major challenge in developing such
systems is that the quality of a CT venog-
raphy study relies on proper diffusion and
timing of a venous contrast injection prior
to the imaging study being performed. In
areas of confluence and turbulent blood
flow, hypodense streaks may occur on the
resulting CT images.12 These streaks may
resemble a DVT when visualizing a single
slice of a CT study and can therefore lead
to false positives when analyzing the slices
individually. Fortunately, the false positive
streaks tend to span more slices than a true
DVT, and therefore can be distinguished
from DVT by leveraging spatial informa-
tion between adjacent slices, just as radi-
ologists do when they analyze CT studies
for DVT.
Methods for incorporating 3D spatial
context have only recently become avail-
able. Increased computational power and
larger datasets have led to a resurgence of
deep neural networks (DNNs) and progress
in the field of computer vision. Many networks have been trained on millions of images and
are easily adapted for new tasks. We are optimistic that this new technology, coupled with
innovative architectures to process 3D spatial information, will resolve many of the diagnostic
difficulties arising from the low contrast areas.
In this study, we aim to (1) segment targeted deep veins in a CT slice, (2) classify whether
a DVT is present within multiple slices given segmentations of deep veins, and (3) evaluate
different deep learning approaches for handling 3D datasets for DVT detection. In our segmen-
tation task, a CT slice or set of slices is passed as input to our model and a binary segmentation
mask that highlights deep veins of interest is outputted from our model. For classification,
we input a CT slice or set of slices, and we output a single binary value indicating whether
the center slice contains a DVT. For both tasks, we employ 2D models as a baseline, and
compare their performance to various 3D approaches. An overview of our workflow is depicted
in Figure 1.
2. Related Work
To the best of our knowledge, no computer-aided detection (CAD) systems have been
developed for the automated detection of deep vein thrombosis (DVT) from computed to-
mography (CT) studies. However, 3D convolutional approaches have demonstrated superior
performance to 2D approaches for the related task of detecting pulmonary embolism in chest
CT studies, achieving a sensitivity of 83% with 2 false positives per volume. This is the current
state of the art for computer-aided pulmonary embolism detection.13
In 2016, two major approaches for leveraging 3D spatial information in computer vision
tasks emerged: (1) 3D convolutional neural networks (CNNs) such as 3D U-Net,14 and (2) 2D
CNN and recurrent neural network (RNN) hybrids such as the FCN-RNN hybrid developed
by Chen et al.15 3D U-Net has already been demonstrated to outperform U-Net,16 its 2D
counterpart, in many segmentation tasks.17–19 Likewise, the RNN-CNN hybrid outperformed
U-Net in segmenting 3D neuronal and fungal structures.15 However, no studies have compared
the effectiveness of the two approaches, so we utilize architectures from both approaches in
an effort to compare their performance.
3. Data
We were granted access to a dataset consisting of 119 lower extremity computed tomogra-
phy (CT) studies from 97 Stanford patients, 22 of whom had two CT studies performed. Each
study consisted of 29 to 1,519 grayscale slices with pixel values represented by 16-bit unsigned
integers. Anywhere from 0 to 1,031 slices were annotated by a radiologist. See Figure 2 for
distributions of slices per study and annotated slices per study. Slices without an annotation
were assumed to be negative for DVT in our classification task based on our knowledge of
how the data was annotated. Our dataset was split randomly by patient ID into training,
validation, and testing subsets at a ratio of 60% - 20% - 20%.
(a) Slices (b) Annotated Slices
Fig. 2: Distribution of slices per CT study and of annotated slices per CT study.
4. Methods
4.1. Data Pre-processing
For both the segmentation and classification tasks, CT slices were resized to 224 ×224
pixels from 512 ×512 pixels in an effort to make their dimensions compatible with weights pre-
trained on ImageNet for our models. The pixel values were mean-centered and standardized
according to the mean and standard deviation of the pixel values of images in our training
set. For the classification task, pixel values were clipped to the interval [0,4095].
Depending on the dimension and task, our definition of an example from our dataset
changed. For our 2D baselines, individual slices were treated as separate examples for input
to the models with their corresponding mask/label the targeted output. For the 3D segmen-
tation task, chunks of 8 consecutive CT slices were used as input to the models, with the
corresponding 8 consecutive binary deep vein masks used as the targeted output. Thus, our
input and output dimensions for 3D segmentation were 8 ×224 ×224 pixels. For the 3D
classification task, the input consisted of chunks of 11 consecutive CT slices, and the targeted
output was a single binary indicator of whether DVT was present in the center slice of the
chunk. The number of consecutive slices to include was selected based on its ability to capture
the entirety of a true DVT, allowing the model to identify false positives that typically span
more slices than a true DVT. Images on the edges of the input volume were included in our
examples by padding the chunks with empty images.
4.2. Data Augmentation
The training set of reference images and masks was augmented in order to combat over-
fitting. A generator was developed for the segmentation task that applied simple transforma-
tions to both 2D and 3D training examples. Rotations up to 72, shifts up to 5% of the width
and height, reflections over the x-axis, and scaling up to 5% were permitted. The parameters
for image augmentation were selected based on a brief hyperparameter search according to
segmentation performance on the validation set. For classification, a training generator that
permitted reflections over both the x-axis and y-axis with a probability of 0.5 was used.
4.3. Segmentation
4.3.1. Models
We implement three convolutional neural network (CNN) architectures for the deep vein
segmentation task. All of these networks have architectures consisting of an encoder, which
takes an image and generates a high dimensional feature matrix, and a decoder, which takes
a high dimensional feature matrix and generates a segmentation mask.
The first model is a 2D U-Net. Convolutional layers in the encoder are separated by max
pooling operations while the second half are separated by upsampling operations.16 U-Net is
known for its increased precision in segmentation that results from concatenating high res-
olution features to outputs in the decoder. It yields state of the art performance on image
segmentation tasks in biological contexts with fast train and test speeds. This is especially
important for the potential application of real-time DVT image segmentation. Our implemen-
tation of U-Net involves 10 convolutional layers, selected according to a brief hyperparameter
search (infrastructure especially inspired by20). The inputs to this model are individual 224 ×
224 pixel images, and the outputs are equivalently shaped masks.
A modified VGG16 architecture was implemented for our second segmentation model. This
model uses the encoder from the original VGG16 model,21 but adds a decoder that mirrors
the encoder’s max pooling layers with upsampling. The major advantage of this architecture
is that it allows for transfer learning in the encoder by using weights that were pre-trained on
ImageNet. By using weights that were trained on a larger data set, we decrease the likelihood
of over-fitting to the training set. Like our 2D U-Net, this model takes 224 ×224 pixel images
as input and outputs an equivalently shaped mask.
The third CNN architecture employed for the segmentation task was a 3D implementation
of U-Net. This model is capable of performing volumetric segmentation.22 The model archi-
tecture is very similar to that of U-Net, with 2D convolutions and max pooling operations
replaced by 3D variants. The input to this model consists of a volume of 8 adjacent images,
each 224 ×224 pixels. Its outputted masks maintain this shape at 8 ×224 ×224 pixels.
4.3.2. Loss Function
The choice of loss function is especially important with the presence of class imbalance,
as is the case in in this study. The ratio of pixels representing deep veins to those that do
not is approximately 1:536 in our training set. We experimented with dice loss and weighted
binary cross-entropy loss. Ultimately, we found that dice loss had the best performance on
our validation set:
L=2
N
N
P
i=1
tipi+ 1
N
P
i=1
ti+
N
P
i=1
pi+ 1!
where Nis the number of pixels in the image, tiis the true label for a pixel, and piis the
probability output by the model. The 1 is added for smoothing in case the denominator would
ever become zero.
4.3.3. Evaluation
Model performance was evaluated using traditional metrics of segmentation performance,
including intersection over union (IoU),23 and dice coefficient.24 We perform an additional
pixel-wise analysis where we treat each pixel as a binary classification task of containing deep
vein tissue (1) or not (0). We then plot a receiver operating characteristic (ROC) curve, which
demonstrates the trade-off between sensitivity and specificity across a range of probability
thresholds. Hyperparameters including learning rate, number of layers, and layer activation
function were determined according to model performance on these metrics on the validation
set.
4.4. Classification
4.4.1. Models
To approach the DVT classification task, we utilized both 2D and 3D images. Our 2D
images used a single CT slice in the study. For the 2D images, we utilized a standard 2D
CNN. The 3D images utilized 11 slices in the study: 5 of the slices before, the center slice,
and 5 slices after. Our 3D models were trained to predict DVT for the center slice. For the
3D volumes, we utilized a CNN-RNN and a 3D CNN. In all cases, we also experimented with
providing the labeled segmentation masks of the veins as an extra channel.
Our 2D CNN utilized the ResNet18 architecture25 pretrained on ImageNet. This archi-
tecture presents residual connections that act as shortcuts, skipping one or more layers in
the network. Consequently, this network becomes easier to optimize, addresses the vanishing
gradient problem, and also has groundbreaking performance on the ImageNet dataset. The
final layers are a global average pooling and a fully connected layer with a softmax activation
to output a probability over the two possible classes.
The 3D CNN-RNN also utilized the ResNet18 architecture. For the 11 slices, we passed
them individually through the CNN, creating a feature vector for each one of the slices. Then,
we pass each of the feature vectores through a bidirectional LSTM and a final fully connected
layer, predicting the output for the middle slice.
The Kinetics 3D Inception model26 utilizes a 3D version of Google’s Inception v3 network.
This network utilizes different convolutional layers in parallel and concatenates the results
together. Unlike the CNN-RNN, this model uses a 3D kernel to run over multiple slices,
incorporating this spatial information in the convolution calculations. In recent work, this has
been shown to produce state of the art results for lung cancer screening on CT scans.27
4.4.2. Loss Function
For our classification problem, we used the Cross Entropy Loss on the softmax probability
outputs:
L=1
M
M
X
i=1
C
X
c=1
y(i)
clog(ˆy(i)
c)
where Mis the batch size, Cis the number of classes (i.e., 2), yis the ground truth label, and ˆy
is the probability output of the model. Additionally, we added weight decay regularization to
our loss function to penalize the model for having large weights, thereby reducing over-fitting.
4.4.3. Evaluation
To analyze our classification performance, we examined area under the receiver operator
characteristic (AUROC) and average precision (AP). Because we have a unbalanced dataset
with more negative examples, accuracy is not the best metric, as a null classifier could have
very high accuracy (e.g., if 80% of the data were negative, then the null classifier would classify
80% correctly by always guessing the majority class).
AUROC shows the true positive rate (T P
P) as a function of false positive rate (F P
N), where
the different values are a result of changing the classification threshold of the model. Essen-
tially, this metric allows us to choose a false positive rate (what percent of negatives we miss
classify as positives) that we accept and tells us how correctly we label the positives.
AP shows precision ( T P
T P +F P ) as a function of recall (T P
P). Essentially, we can choose the
percent of true case that we want to catch (recall) and then show how many true positives
compare to false positives. Therefore, for our problem, the average precision is the strongest
metric, as we want to minimize the number of false positives resulting from poor CT contrast.
5. Results
5.1. Segmentation
Fig. 3: Segmentation test ROC curve.
Fig. 4: Example deep vein segmentations from
a random test set image on all four models.
Table 1 illustrates the performance of
each deep vein segmentation model accord-
ing to the putative metrics of segmenta-
tion performance: dice coefficient and in-
tersection over union. The 2D VGG with
encoder pre-trained on ImageNet had the
best performance according to both dice
coefficient and intersection over union. The
performance of the 2D VGG architecture
without weights closely followed in perfor-
mance. The 2D U-Net model had the worst
performance on the validation and test set
despite middling performance on the train-
ing set indicating over-fitting. This over-
fitting problem is especially present for the
3D U-Net architecture, which despite its
superior performance on the training set,
performed worse than the 2D VGG archi-
tecture with and without weights on the
validation and test sets.
Figure 3 depicts the performance of
each model according to pixel-wise binary
classification metrics of specificity and sen-
sitivity. The ranking of the area under the
receiver operating characteristic (AUROC)
recapitulates the results presented by the segmentation metrics. In particular, the VGG net-
work with a pre-trained encoder has the best performance followed closely by the VGG network
without pre-training. This model’s performance is followed by that of the 3D U-Net, and fi-
nally by the 2D U-Net. With an AUROC of 0.48, the 2D U-Net has almost no prediction
power. One interesting feature is that the pre-trained encoder appears to regularize the ROC
curve of the VGG network. In particular, transfer learning improves performance when the
false positive rate is set to low values, whereas the curve for non-pre-trained encoder hugs the
diagonal at these low false positive values.
Table 1: Segmentation performance on inter-
section over union and dice coefficient metrics.
Model Metric Training Validation Test
2D U-Net Dice 0.12 0.01 0.01
IoU 0.23 0.05 0.04
2D VGG Dice 0.31 0.05 0.05
IoU 0.56 0.43 0.46
2D VGG Pre-trained Dice 0.38 0.12 0.07
IoU 0.67 0.55 0.48
3D U-Net Dice 0.46 0.09 0.04
IoU 0.72 0.41 0.42
Finally, Figure 4 presents a qualitative
analysis of the relative ability of all four
models to segment the deep veins from a
randomly selected image from the test set
(ground truth shown on the left). The 2D
U-Net in this case predicts almost all the
pixels in the image to belong to a deep vein.
The VGG with pre-trained encoder is able
to segment slightly large circles around
each of the deep veins, whereas the VGG
model without pre-trained weights appears
to have a general sense of the area of the
CT scan. Finally, the 3D U-Net appears to
have functioned as essentially a null classi-
fier.
5.2. Classification
Table 2: Classification performance for our
models on AUROC and average precision
evaluation metrics.
Model Metric Validation Test
2D ResNet AUROC 0.80 0.60
Avg. Precision 0.74 0.29
2D ResNet, Masks AUROC 0.82 0.59
Avg. Precision 0.77 0.27
CNN-RNN AUROC 0.81 0.64
Avg. Precision 0.77 0.31
CNN-RNN, Masks AUROC 0.79 0.59
Avg. Precision 0.70 0.26
3D Inception AUROC 0.81 0.63
Avg. Precision 0.70 0.30
3D Inception, Masks AUROC 0.73 0.62
Avg. Precision 0.70 0.33
Table 2 shows the classification results for
both the validation and test sets on all mod-
els. Figure 5 shows the ROC and precision-
recall curves for our models, which demon-
strate the performance of the model across a
range of probability thresholds. In the valida-
tion dataset, the 2D ResNet with masks per-
forms best in both the AUROC and average
precision. In the test set, the CNN-RNN has the
best AUROC and the 3D inception with masks
has the best average precision. Overall, we do
see a slight increase in performance on the test
set when we use the 3D models: CNN-RNN and
3D Inception. However, adding the segmenta-
tion masks to the models does not necessarily
help the model.
(a) Receiver Operating Characteristic (ROC) (b) Precision-Recall Curve
Fig. 5: Performance of DVT classification models on the test set.
6. Discussion
Fig. 6: Example of ground
truth annotation missing vein
in red circle.
In segmentation, we find our models perform rather
poorly on the task, with a highest dice coefficient of 0.07
and intersection over union of 0.48 for the VGG model with
pre-trained weights. One factor that may contribute to this
poor performance is missing deep veins in the segmentation
masks provided by the labeling radiologist (Figure 6). Addi-
tionally, these missing masks seem to be disproportionately
impacting the 3D model as a particular vein may not be
traced fully through the volume, thus the model learns to
be more careful in mask prediction. We believe one potential
solution to mitigate this problem would be to ensure that
masks are symmetric during data pre-preprocessing. We are
also interested in restructuring the 3D model, such that the
segmentations predicted from a volume is the single center
slice rather than a volume of slices. We believe this may
help the 3D model become more robust to the label issue as well.
Interestingly, our VGG outperforms U-Net, which traditionally has been shown to generate
state of the art results on biological applications. The major functional difference between the
models is that U-Net concatenates layers from the encoder in the decoder. We hypothesize
that this concatenation is contributing to the over-fitting problem, as it increases the variance
of the model. Additionally, our results validate the importance of pre-training, as the VGG
model had better performance with pre-trained weights than without pre-trained weights.
For classification, our main issue seems to arise from our train, validation, and test split
(Table 2). With limited tuning, our validation set results are disproportionately greater than
Fig. 7: Distribution of deep veins in the train, validation, and test set images. Each pixel
represents the proportion of slices that had a deep vein annotation that included the pixel.
our test sets, with an AUROC of 0.82 and average precision of 0.77 in comparison to an
AUROC of 0.64 and average precision of 0.33 on the test. We believe our small dataset size is
leading to significant differences in the characteristics of the data splits. We randomly split our
data into train, validation, and test sets based on patient ID. However, if the distribution of
the test set randomly happens to be different from that of our validation set, our model would
not be able to generalize to this different distribution, especially since we have limited labeled
data. For example, in the validation set, 15% of slices were positive for DVT, while in the
test only 11% were positive for DVT. Without these issues, it does seem that the algorithms
are performing well based on the validation metrics. Interestingly, the dataset splitting issue
seems to be less of a problem in the segmentation task than in the classification task, since
the difference between validation and test performance is less. However, the locations of the
ground truth deep veins appear fairly different between all dataset splits (Figure 7).
As with most medical imaging studies, our work is limited to data from a single institution.
Thus, it is likely that we are over-fitting to certain institutional practices, such as imaging
instrument quality and patient admission diagnoses. The dataset also comes from patients
seeing an interventional radiologist, rather than general trauma patients, which is the intended
use case. Unfortunately, our models currently do not perform well enough to be used as part
of a computer aided detection system in clinical practice.
Our main interest is further iteration on our models to get interpretable and clinically
relevant performance. Of primary importance is fixing the data issues with missing labels, as
well as generating a more balanced data split through some form of matching based on CT
study statistics. Additionally, we had limited time for hyperparameter tuning, and strongly
believe that a more extensive hyperparameter search could yield much better model perfor-
mance. Finally, we would be interested in performing ensembling, as this has also been shown
to improve performance, especially in cases of over-fitting.28
Acknowledgments
We would like to thank Dr. Lawrence Hofmann for granting us access to the dataset used
in this project and for his guidance throughout. We would also like to thank Andre Souffrant,
David Cohn, and Jared Dunnmon for introducing us to the dataset and for providing high-level
algorithmic advising.
References
1. K. Burnand, A. Irvine, N. Wilson et al., Deep vein thrombosis: pathology, in Diseases of the
veins, (Arnold London, 1999) pp. 249–289.
2. H. Decousus, A. Leizorovicz, F. Parent, Y. Page, B. Tardy, P. Girard, S. Laporte, R. Faivre,
B. Charbonnier, F.-G. Barral et al., A clinical trial of vena caval filters in the prevention of
pulmonary embolism in patients with proximal deep-vein thrombosis, New England Journal of
Medicine 338, 409 (1998).
3. G. A. Maynard, Preventing hospital-associated venous thromboembolism: a guide for effective
quality improvement (Agency for Healthcare Research and Quality, US Department of Health
and , 2016).
4. S. Hamada, C. Espina, T. Guedj, R. Buaron, A. Harrois, S. Figueiredo and J. Duranteau, High
level of venous thromboembolism in critically ill trauma patients despite early and well-driven
thromboprophylaxis protocol, Annals of intensive care 7, p. 97 (2017).
5. C. J. Berg, J. Chang, L. Elam-Evans, L. Flowers, J. Herndon, K. A. Seed and C. J. Syverson,
Pregnancy-related mortality surveillance–united states, 1991–1999 (2003).
6. M. G. Beckman, W. C. Hooper, S. E. Critchley and T. L. Ortel, Venous thromboembolism: a
public health concern, American journal of preventive medicine 38, S495 (2010).
7. F. A. Anderson, H. B. Wheeler, R. J. Goldberg, D. W. Hosmer, N. A. Patwardhan, B. Jovanovic,
A. Forcier and J. E. Dalen, A population-based perspective of the hospital incidence and case-
fatality rates of deep vein thrombosis and pulmonary embolism: the worcester dvt study, Archives
of internal medicine 151, 933 (1991).
8. N. McLeod and G. Montane, The radiologist assistant: the solution to radiology workforce needs,
Emergency radiology 17, 253 (2010).
9. A. P. Kiraly, C. L. Novak, D. P. Naidich, I. Vlahos, J. P. Ko and G. T. Brusca-Augello, A
comparison of 2d and 3d evaluation methods for pulmonary embolism detection in ct images, in
Medical Imaging 2006: Image Perception, Observer Performance, and Technology Assessment,
2006.
10. S. Thomas, S. Goodacre, F. Sampson and E. Van Beek, Diagnostic value of ct for deep vein
thrombosis: results of a systematic review and meta-analysis, Clinical radiology 63, 299 (2008).
11. M. D. Cham, D. F. Yankelevitz, D. Shaham, A. A. Shah, L. Sherman, A. Lewis, J. Rademaker,
G. Pearson, J. Choi, W. Wolff et al., Deep venous thrombosis: detection by using indirect ct
venography, Radiology 216, 744 (2000).
12. W.-Y. Shi, L.-W. Wang, S.-J. Wang, X.-D. Yin and J.-P. Gu, Combined direct and indirect
ct venography (combined ctv) in detecting lower extremity deep vein thrombosis, Medicine 95
(2016).
13. N. Tajbakhsh, M. B. Gotway and J. Liang, Computer-aided pulmonary embolism detection using
a novel vessel-aligned multi-planar image representation and convolutional neural networks, in
International Conference on Medical Image Computing and Computer-Assisted Intervention,
2015.
14. F. Milletari, N. Navab and S.-A. Ahmadi, V-net: Fully convolutional neural networks for vol-
umetric medical image segmentation, in 2016 Fourth International Conference on 3D Vision
(3DV), 2016.
15. J. Chen, L. Yang, Y. Zhang, M. Alber and D. Z. Chen, Combining fully convolutional and recur-
rent neural networks for 3d biomedical image segmentation, in Advances in neural information
processing systems, 2016.
16. O. Ronneberger, P. Fischer and T. Brox, U-net: Convolutional networks for biomedical image
segmentation, in International Conference on Medical image computing and computer-assisted
intervention, 2015.
17. J. Patravali, S. Jain and S. Chilamkurthy, 2d-3d fully convolutional neural networks for cardiac
mr segmentation, in International Workshop on Statistical Atlases and Computational Models
of the Heart, 2017.
18. X. Zhou, K. Yamada, T. Kojima, R. Takayama, S. Wang, X. Zhou, T. Hara and H. Fujita,
Performance evaluation of 2d and 3d deep learning approaches for automatic segmentation of
multiple organs on ct images, in Medical Imaging 2018: Computer-Aided Diagnosis, 2018.
19. ¨
O. C¸cek, A. Abdulkadir, S. S. Lienkamp, T. Brox and O. Ronneberger, 3d u-net: learning dense
volumetric segmentation from sparse annotation, in International conference on medical image
computing and computer-assisted intervention, 2016.
20. Z. Hao, Implementation of deep learning framework – unet, using keras (Github.
21. K. Simonyan and A. Zisserman, Very deep convolutional networks for large-scale image recogni-
tion, arXiv preprint arXiv:1409.1556 (2014).
22. ¨
O. iek, A. Abdulkadir, S. S. Lienkamp, T. Brox and O. Ronneberger, 3d u-net: Learning dense
volumetric segmentation from sparse annotation, in MICCAI , 2016.
23. M. A. Rahman and Y. Wang, Optimizing intersection-over-union in deep neural networks for
image segmentation, in International symposium on visual computing, 2016.
24. C. H. Sudre, W. Li, T. Vercauteren, S. Ourselin and M. J. Cardoso, Generalised dice overlap as
a deep learning loss function for highly unbalanced segmentations, in Deep learning in medical
image analysis and multimodal learning for clinical decision support, (Springer, 2017) pp. 240–
248.
25. K. He, X. Zhang, S. Ren and J. Sun, Deep residual learning for image recognition, CoRR
abs/1512.03385 (2015).
26. J. Carreira and A. Zisserman, Quo vadis, action recognition? A new model and the kinetics
dataset, CoRR abs/1705.07750 (2017).
27. D. Ardila, A. P. Kiraly, S. Bharadwaj, B. Choi, J. J. Reicher, L. Peng, D. Tse, M. Etemadi,
W. Ye, G. Corrado et al., End-to-end lung cancer screening with three-dimensional deep learning
on low-dose chest computed tomography, Nature medicine , p. 1 (2019).
28. Y. Wu and J. I. Arribas, Fusing output information in neural networks: Ensemble performs
better, in Proceedings of the 25th Annual International Conference of the IEEE Engineering in
Medicine and Biology Society (IEEE Cat. No. 03CH37439), 2003.
ResearchGate has not been able to resolve any citations for this publication.
Article
Full-text available
With an estimated 160,000 deaths in 2018, lung cancer is the most common cause of cancer death in the United States¹. Lung cancer screening using low-dose computed tomography has been shown to reduce mortality by 20–43% and is now included in US screening guidelines1–6. Existing challenges include inter-grader variability and high false-positive and false-negative rates7–10. We propose a deep learning algorithm that uses a patient’s current and prior computed tomography volumes to predict the risk of lung cancer. Our model achieves a state-of-the-art performance (94.4% area under the curve) on 6,716 National Lung Cancer Screening Trial cases, and performs similarly on an independent clinical validation set of 1,139 cases. We conducted two reader studies. When prior computed tomography imaging was not available, our model outperformed all six radiologists with absolute reductions of 11% in false positives and 5% in false negatives. Where prior computed tomography imaging was available, the model performance was on-par with the same radiologists. This creates an opportunity to optimize the screening process via computer assistance and automation. While the vast majority of patients remain unscreened, we show the potential for deep learning models to increase the accuracy, consistency and adoption of lung cancer screening worldwide.
Article
Full-text available
Background: Venous thromboembolism (VTE) is one of the most common preventable causes of in-hospital death in trauma patients surviving their injuries. We assessed the prevalence, incidence and risk factors for deep venous thrombosis (DVT) and pulmonary embolism (PE) in critically ill trauma patients, in the setting of a mature and early mechanical and pharmacological thromboprophylaxis protocol. Methods: This was a prospective observational study on a cohort of patients from a surgical intensive care unit of a university level 1 trauma centre. We enrolled consecutive primary trauma patients expected to be in intensive care for ≥48 h. Thromboprophylaxis was protocol driven. DVT screening was performed by duplex ultrasound of upper and lower extremities within the first 48 h, between 5 and 7 days and then weekly until discharge. We recorded VTE risk factors at baseline and on each examination day. Independent risk factors were analysed using a multivariate logistic regression. Results: In 153 patients with a mean Injury Severity Score of 23 ± 12, the prevalence of VTE was 30.7%, 95 CI [23.7-38.8] (29.4% DVT and 4.6% PE). The incidence was 18%, 95 CI [14-24] patients-week. The median time of apparition of DVT was 6 days [1; 4]. The global protocol compliance was 77.8% with a median time of introduction of the pharmacological prophylaxis of 1 day [1; 2]. We identified four independent risk factors for VTE: central venous catheter (OR 4.39, 95 CI [1.1-29]), medullar injury (OR 5.59, 95 CI [1.7-12.9]), initial systolic arterial pressure <80 mmHg (OR 3.64, 95 CI [1.3-10.8]), and pelvic fracture (OR 3.04, 95 CI [1.2-7.9]). Conclusion: Despite a rigorous, protocol-driven thromboprophylaxis, critically ill trauma patients showed a high incidence of VTE. Further research is needed to tailor pharmacological prophylaxis and balance the risks and benefits.
Article
Full-text available
Segmentation of 3D images is a fundamental problem in biomedical image analysis. Deep learning (DL) approaches have achieved the state-of-the-art segmentation performance. To exploit the 3D contexts using neural networks, known DL segmen- tation methods, including 3D convolution, 2D convolution on the planes orthogonal to 2D slices, and LSTM in multiple directions, all suffer incompatibility with the highly anisotropic dimensions in common 3D biomedical images. In this paper, we propose a new DL framework for 3D image segmentation, based on a com- bination of a fully convolutional network (FCN) and a recurrent neural network (RNN), which are responsible for exploiting the intra-slice and inter-slice contexts, respectively. To our best knowledge, this is the first DL framework for 3D image segmentation that explicitly leverages 3D image anisotropism. Evaluating using a dataset from the ISBI Neuronal Structure Segmentation Challenge and in-house image stacks for 3D fungus segmentation, our approach achieves promising results, comparing to the known DL-based 3D segmentation approaches.
Conference Paper
Full-text available
Computer-aided detection (CAD) can play a major role in diagnosing pulmonary embolism (PE) at CT pulmonary angiography (CTPA). However, despite their demonstrated utility, to achieve a clinically acceptable sensitivity, existing PE CAD systems generate a high number of false positives, imposing extra burdens on radiologists to adjudicate these superfluous CAD findings. In this study, we investigate the feasibility of convolutional neural networks (CNNs) as an effective mechanism for eliminating false positives. A critical issue in successfully utilizing CNNs for detecting an object in 3D images is to develop a “right” image representation for the object. Toward this end, we have developed a vessel-aligned multi-planar image representation of emboli. Our image representation offers three advantages: (1) efficiency and compactness—concisely summarizing the 3D contextual information around an embolus in only 2 image channels, (2) consistency—automatically aligning the embolus in the 2-channel images according to the orientation of the affected vessel, and (3) expandability—naturally supporting data augmentation for training CNNs. We have evaluated our CAD approach using 121 CTPA datasets with a total of 326 emboli, achieving a sensitivity of 83% at 2 false positives per volume. This performance is superior to the best performing CAD system in the literature, which achieves a sensitivity of 71% at the same level of false positives. We have further evaluated our system using the entire 20 CTPA test datasets from the PE challenge. Our system outperforms the winning system from the challenge at 0mm localization error but is outperformed by it at 2mm and 5mm localization errors. In our view, the performance at 0mm localization error is more important than those at 2mm and 5mm localization errors.
Chapter
The significance of deep vein thrombosis (DVT) lies in its potential to cause pulmonary embolism and post-thrombotic calf pump failure. Virtually all venous thrombi arise in the deep veins of the legs or pelvis. The incidence of DVT in the general population is approximately 0.5%. Where no preventative measures are employed, the incidence of DVT in general surgical patients over 40 years undergoing major surgery is 30% and 6080% in patients undergoing hip or knee replacement or surgery for hip fracture(b). In patients recovering from myocardial infarction or cerebrovascular accident, the incidence is 20-60%(b).
Article
Deep-learning has proved in recent years to be a powerful tool for image analysis and is now widely used to segment both 2D and 3D medical images. Deep-learning segmentation frameworks rely not only on the choice of network architecture but also on the choice of loss function. When the segmentation process targets rare observations, a severe class imbalance is likely to occur between candidate labels, thus resulting in sub-optimal performance. In order to mitigate this issue, strategies such as the weighted cross-entropy function, the sensitivity function or the Dice loss function, have been proposed. In this work, we investigate the behavior of these loss functions and their sensitivity to learning rate tuning in the presence of different rates of label imbalance across 2D and 3D segmentation tasks. We also propose to use the class re-balancing properties of the Generalized Dice overlap, a known metric for segmentation assessment, as a robust and accurate deep-learning loss function for unbalanced tasks.
Conference Paper
We consider the problem of learning deep neural networks (DNNs) for object category segmentation, where the goal is to label each pixel in an image as being part of a given object (foreground) or not (background). Deep neural networks are usually trained with simple loss functions (e.g., softmax loss). These loss functions are appropriate for standard classification problems where the performance is measured by the overall classification accuracy. For object category segmentation, the two classes (foreground and background) are very imbalanced. The intersection-over-union (IoU) is usually used to measure the performance of any object category segmentation method. In this paper, we propose an approach for directly optimizing this IoU measure in deep neural networks. Our experimental results on two object category segmentation datasets demonstrate that our approach outperforms DNNs trained with standard softmax loss.
Conference Paper
There is large consent that successful training of deep networks requires many thousand annotated training samples. In this paper, we present a network and training strategy that relies on the strong use of data augmentation to use the available annotated samples more efficiently. The architecture consists of a contracting path to capture context and a symmetric expanding path that enables precise localization. We show that such a network can be trained end-to-end from very few images and outperforms the prior best method (a sliding-window convolutional network) on the ISBI challenge for segmentation of neuronal structures in electron microscopic stacks. Using the same network trained on transmitted light microscopy images (phase contrast and DIC) we won the ISBI cell tracking challenge 2015 in these categories by a large margin. Moreover, the network is fast. Segmentation of a 512x512 image takes less than a second on a recent GPU. The full implementation (based on Caffe) and the trained networks are available at http://lmb.informatik.uni-freiburg.de/people/ronneber/u-net .