A targeted development of safe medical products can be supported by design methods. This paper analyses which design methods are applied in the development of medical devices and whether they are adapted for considering medical devices’ special features (legal, human and technical issues). In particular, variety management, risk assessment and user-centered design for medical devices are examined. Typically, interdisciplinary risk assessment is methodically supported. Additionally, user-centered design methods for requirements assessment, design verification and design validation are applied.
Content may be subject to copyright.
SOCIO-TECHNICAL ISSUES IN DESIGN 1969
INTERNATIONAL DESIGN CONFERENCE DESIGN 2020
https://doi.org/10.1017/dsd.2020.95
INVESTIGATION ON METHODS AND CHARACTERISTICS IN MEDICAL
DEVICE DEVELOPMENT
J. Kuhl , O. Sankowski and D. Krause
Hamburg University of Technology, Germany
juliane.kuhl@tuhh.de
Abstract
A targeted development of safe medical products can be supported by design methods. This paper
analyses which design methods are applied in the development of medical devices and whether
they are adapted for considering medical devices’ special features (legal, human and technical
issues). In particular, variety management, risk assessment and user-centered design for medical
devices are examined. Typically, interdisciplinary risk assessment is methodically supported.
Additionally, user-centered design methods for requirements assessment, design verification and
design validation are applied.
Keywords: biomedical design, design methods, risk management, user-centred design, medical
device development
1. Introduction
The increasing demand for patient safety and security of supply within the medical sector is
challenging medical device companies to further optimize and personalize their products. In order to
develop high-quality products with low error rates and low risk, different design methods can be
applied to support a goal-oriented product development process (Pahl et al., 2007). As safety issues
are particularly important for medical devices, the application of design methods within the healthcare
sector should be given appropriate importance. However, medical devices have special issues
(Lantada and Morgado, 2013; Santos et al., 2013) that may need to be considered when applying
design methods in the development process.
The investigation on how these special issues and characteristics of medical devices influence the
application of design methods is the subject of this paper. Since the consideration of individual, patient
specific requirements within the design of medical devices is increasingly focused (e.g. Maier et al.,
2017), special attention is paid to the application of variety management and modularization methods,
through which the management of variety-induced complexity can be supported (Simpson et al., 2014;
Krause and Gebhardt, 2018).
After the presentation of known special issues of medical devices in section 2, the research method is
introduced (section 3). Afterwards, the application of variety management methods for medical
devices (section 4.1) as well as the use of general product development methods for medical devices
(see section 4.2) is investigated. In Section 4.3 and 4.4, the application of risk assessment as well as
user-centered methods for medical devices is emphasised in detail. In doing so, special attention is
paid to the extent to which design methods were adapted to make them applicable to medical devices.
1970 SOCIO-TECHNICAL ISSUES IN DESIGN
Finally, the results are summarized (section 4.5 and 4.6) and discussed (section 5), before a conclusion
is given in section 6.
2. Background
As a background for this paper, the state of the art in the development of medical devices and its
special issues and characteristics as well as variety management and user-centered design is presented
in the following.
2.1. Special issues in medical device development
According to the Regulation (EU) 2017/745 of the European Parliament (2017), the European
definition for medical devices is as follows: “‘medical device’ means any instrument, apparatus,
appliance, software, implant, reagent, material or other article intended by the manufacturer to be
used, alone or in combination, for human beings for one or more of the following specific medical
purposes:
 diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease,
 diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability,
 investigation, replacement or modification of the anatomy or of a physiological or
pathological process or state,
 providing information by means of in vitro examination of specimens derived from the human
body, including organ, blood and tissue donations,
and which does not achieve its principal intended action by pharmacological, immunological or
metabolic means, in or on the human body, but which may be assisted in its function by such means.
This definition covers a broad range of products, which differ from each other in aspects like risk
potential, development time or production volumes (Santos et al., 2013). To classify products in
Europe, medical devices are divided into four different risk classes I, IIa, IIb and III (European
Parliament, 2017). In a nutshell, the risk class depends on the duration and location of use of the
medical device, i.e. whether it is used externally or invasively. The longer and the more contact the
product has with tissues in the body, the higher the risk class. Depending on the risk class, medical
devices are subject to strict regulations that must be fulfilled in order to be admitted to the market as a
medical device. In Europe the conformity of the developed product must be proven by the
manufacturer himself. Notified bodies check the declaration of conformity (except for certain products
in risk class I) (BVMed, 2016). Both technical as well as clinical factors must be examined within the
scope of the conformity assessment. During design verification it is checked, if all technical
requirements were considered during the design phase while design validation confirms that the final
product fulfils the intended use (Alexander and Clarkson, 2000). If conformity is approved, the
medical device can be provided with a CE mark and offered on the market (BVMed, 2016). Taking
into account different methodologies for the development of medical devices (e.g. US Food and Drug
Administration (FDA), 1997; Santos et al., 2013) and combining it with the general engineering
development process of VDI 2221 (Pahl et al., 2007), a medical device development process can be
summarized as shown in Figure 1, paying special emphasis to design verification and validation.
Figure 1. Development process for medical devices
Phase I
Clarification
Definition of
the intended
use
Design
Input
Design
Output
Design Verification
Design Validation
Phase II
Conceptual Design
Phase III
Embodiment
Phase IV
Detailed Design
Standards and Guidelines
Risk
Class ification
Medical Device Development Process
Product
Class ification:
Medical Device
General
Produ ct
Development
Process
Real
Human
Need
SOCIO-TECHNICAL ISSUES IN DESIGN 1971
Legal issues, which include the points just described, also cover ethical issues including beneficence,
non-maleficence, justice, and respect for autonomy, which must be taken into account when
developing medical devices (Santos et al., 2013). Lantada and Morgado (2013) further add technical
and human issues. Technical issues refer to geometric and material details, which have certain
characteristics due to the direct contact of the medical device with the human body. These include
aspects such as biocompatibility of materials or the cleanable and sterilizable design of the devices.
Human issues include the coordination of different experts, who all have an interest in the medical
device and can provide different inputs during development (Lantada and Morgado, 2013). The
heterogeneity of the stakeholders is also emphasized by Santos et al. (2013). In many cases medical
data and databases can be used to specify the use case and clarify the anatomical requirements. On the
other side, according to Santos et al. (2013), economic issues are less important. Although, in order to
establish itself, a medical device must be competitive when introducing it to the market, it is more
important to offer a product with good quality and low risk. Figure 2 summarizes special issues, which
have to be considered in medical device development.
Figure 2. Special issues in medical device development
Despite the many special issues that must be taken into account in the development of a medical
device, there is no uniform procedure for the development of medical devices (Azman et al., 2017).
Therefore, at this point the question arises why there are no methods and methodologies for the
development of medical devices. For answering this question, it is necessary to check whether there
are specific methods for the development of medical devices and how they differ.
2.2. Design methods
Design methodologies support the development of products and technical systems by proposing a sequence
of different design phases through which a solution can be systematically developed (Pahl et al., 2007).
One example for a methodology is contained in VDI 2221 (Pahl et al., 2007). Within each phase, different
methods can support the systematic solution finding (Pahl et al., 2007). A method is a description of a rule-
based and scheduled procedure for obtaining knowledge or results (Krause and Gebhardt, 2018).
2.2.1. Personalization strategies
For a better adaption of the medical device to the patient specific anatomies, personalized medical
devices become more and more important (Federal Ministry of Education and Research, 2018; Maier
et al., 2017). Thus, an optimal functionality of the device for every patient can be reached. Therefore,
the consideration of individual, patient specific requirements within the design of medical devices is
increasingly focused. However, considering individual requirements within the product development
process results in an increase of the variety-induced complexity emerging of the variety of
requirements, parts, components, products as well as processes within a company (Simpson et al.,
2014). To manage this variety-induced complexity, design for variety methods and modularization
approaches can be applied (Simpson et al, 2014; Krause and Gebhardt, 2018). It includes mass
customization strategies, which differ in their achievable level of customization (da Silveira et al.,
2001). Thereby, personalized modules are combined with standard and variant modules to create
technical
issues
legal issues
human
issues
Medical Device
ethics standards
risk
classification
interdisziplinarity
biocompatibility
usability
safety
quality
real
human
need
sterilization
1972 SOCIO-TECHNICAL ISSUES IN DESIGN
specific product variants (e.g. Lindemann et al., 2006; Koren et al., 2013). Gräßler (2004) introduces a
methodology for the Development of Mass Customized Products.
Beside the product, variety-induced process complexity needs to be managed by adequate methods
(Lindemann et al., 2006; Spallek and Krause 2016).
Usually, personalization methods focus on how to manage variety-induced complexity. A systematic
analyse on which product properties can be personalized while bringing benefit for the customer and
being manageable concerning internal variety at the same time, is rarely supported (Kuhl and Krause,
2018). But especially for personalized medical devices, such a method would be very helpful, to
balance risk, benefit and complexity at the same time.
2.2.2. User-centered design
For the design of (personalized) medical devices, collecting and analysing requirements of various
stakeholders, especially the individual and special requirements of patients, as well their implementation
into the design process is of high importance. These goals can be achieved by choosing and applying
user-centered approaches, methods or tools. Here, user-centered design (UCD) stands for the whole
research field of user-centrism (Sankowski and Krause, 2018). Apart of UCD, further terms exist to
describe this research area, e.g. Human-Centered Design, Participatory Design, user research or, as
stated above, Design for Usability. Herein a plethora of different methods, approaches and techniques
exist to elicit user requirements, e.g. interviews, prototyping, usability tests. UCD methods and tools can
be applied in all design phases, generally spoken these are planning and task clarification phase,
ideation or concept phase and the testing and evaluation phase. Furthermore, methods and tools can be
applied for or with primary users (the target or main users that interact with the product to perform its
main function), secondary users (those users that also interact with the product on a regular and intended
basis to perform auxiliary or subordinate functions), and tertiary users (those persons that do not directly
or unintentionally interact with the product and could be negatively affected by it).
Several researchers proposed categorizations of UCD methods and approaches to enhance goal-
oriented method selection. Sanders (2006) presents a map of design research, where different methods
and approaches are clustered and allocated according to their underlying mindset of user-involvement;
depending on whether users are seen as subjects, i.e. reactive informers (expert mindset), such as in
Human Factors, Ergonomics or Usability Testing, or as partners in design, i.e. active co-creators
(participatory mindset), such as in Co-Creation or Storytelling. A broader overview of UCD methods
and tools is described by Sankowski and Krause (2018). Here, every kind of method and tool that
supports the elicitation and definition of user and usage requirements is allocated to the UCD research,
ranging from standards and guidelines with a weak focus on the user, to virtual simulation techniques
and emphatical methods (moderate focus) up to methods and approaches with user involvement, e.g.
co-design or usability tests, with a strong focus on individual user needs.
User-centered approaches and especially methods with intensive user participation of different user
groups throughout the entire product life-cycle are proven to be helpful in creating innovative,
satisfactory, efficient and effective products by involving and addressing diverse users needs. This is
also reported in context of medical devices (e.g. Lettl, 2007; Luna et al., 2016). However, ill-targeted or
wrongly applied user centrism may also delay or threaten the outcome of a product development process.
It is therefore necessary to understand, which kind of UCD methods, at what point of the process and
with participation of which stakeholders are necessary for a targeted product development.
3. Research method
In order to support the development of safe and high-quality medical devices, this paper will examine
the extent to which design methods exist specifically for medical devices and how these differ from
other not specially adapted or created methods. Therefore, a large literature review, using the Scopus
database, is conducted which includes the investigation on certain keyword combinations. The search
results are limited to English publications within the engineering area. Quotation marks () around
search terms symbolize that the words have to appear together, a star (*) represents that any characters
can be inserted instead, e.g. *medical also includes terms like biomedical.
SOCIO-TECHNICAL ISSUES IN DESIGN 1973
First of all, the existing methods in the field of variety management for medical devices are examined
(keyword combination: *medical and variety/ complexity/ modulari*) and compared with the number
of publications in the field of personalization/individualization/ customization (see Figure 3).
The search is then extended to general design methods that are used for medical devices. For this
purpose, it was examined which topics are addressed in the product development of medical devices.
Topic clusters are created using the VOSviewer software. All keywords, which are named at least
three times by the author, are included. All too basic terms, such as design, children or
entrepreneurship, are excluded manually. Within the defined clusters, which are shown in Figure 3,
publications were examined with regard to the application of design methods.
Figure 3. Overview of emphasised keyword combinations and clusters including the number of
found publications
The most frequently addressed methods and topics were then analysed (keyword combinations on risk
and user*, see Figure 3). Publications that use or present methods to support risk assessment are
analysed to identify method adaptations that are associated with the application of methods to medical
devices. Since the search for *medical and product risk or device risk resulted in 1.320 hits, the search
is narrowed down by adding further keywords including risk&evaluat*, risk&analys*,
validation, design method (see Figure 3, total of 147 hits remain when excluding duplications). In an
iterative loop, the keyword FMEA (short form for Failure Mode and Effects Analysis) is added. By
first analysing the abstract and then having a closer look on publications, which apply a certain design
method, general conclusions can be drawn.
Additionally, publications dealing with any kind of UCD and common synonyms (Human-Centered
Design, User Involvement, etc.) are emphasised; without restricting the search to any specific method
or approach, such as Design Thinking, Co-Design or User Experience Design. Hits are checked to see
if they describe a medical device development case study and whether they are using any type of UCD
method or tool. All paper fulfilling both criteria are selected for description and analysation of the
state of the art in UCD for medical device development.
Finally, the findings are summarized and their significance for the design of personalized medical
devices is addressed. Furthermore, limitations are discussed before the conclusion is reached.
4. Design of medical devices
The results of the research procedure described in section 3 are presented below.
4.1. Design for variety for medical devices
The large number of publications on keyword combinations dealing with *medical in personalization,
customization and individualization prove the close relation of those topics as earlier presented in
Compatibility
Conformity
Reliablity and
Safety
Usability
design process
management
„design for variety
product
product
product
internal
product complexity
process
external
14
1
1
2
776
1.200
41
1.320
3.187
84
5
2
6
0
variety
1
0
individual*
personal*
20
84
5
6
process
149
method* 26
approach
11
modulari*
risk&evaluat*“
4
22
validation
47
risk&analys*“
86
5.002
user&cent*“ 179
user&involv*“
18
user&integrat*“
2
user&orient*“ 16
product
product
customi*
„internal complexity
0
complexity
FMEA
9
product/
device risk
design method
25
578
*medical
engineering unterstan ding,
publications in english
human&cent*“
user*
18
22
user
particip*
8
product
development
1974 SOCIO-TECHNICAL ISSUES IN DESIGN
section 1. Although the importance of variety management for an increasing number of variants as in
case of personalized product design is known within the engineering design community, the analysis
on the keywords shows that the variety management itself is not focused at all in related literature.
There is no publication in Scopus database dealing with medical devices and Design for Variety or
variety management. Also, the search for keywords closely linked to variety like complexity,
modularity or product family, product structure or product architecture show no relevant matches on
how to define and manage the number of variants within a medical device product family. Since even
the search for variety management named in the title or abstract offers no results, no typical
characteristics for medical devices in variety management methods can be summarized.
Therefore, at this point, the question arises why there is no variety management for medical devices.
Can standard methods be applied directly for medical devices without changing or adapting them? Or
is there an unresolved need for adaptation? In order to answer these questions, it is investigated which
methods specified for the use in medical device development exist so far and which adaptations have
been made to them.
4.2. Design methods for medical device development
The investigation on medical device development topics emphasise, that design validation and
verification are of high importance and part of every topic focused in medical device development.
This makes clear that one of the main tasks during development is to ensure the safety and usability of
the medical device. This can also be proved by having a closer look on methods for medical device
development. Different methods can be found that support the design of safe and high-quality products
in the medical device sector. Within methods for medical devices, two aspects are nearly always
included, both contributing to a high quality and good usability of medical devices:
 risk assessment as part of quality management
 some kind of user involvement method or tool, i.e. UCD approach
Legal and product-related issues of medical devices are implemented in the development process by
the methodical evaluation and assessment of risks. For this purpose, different methods and tools are
used, which are explained below. Additionally, human issues, that need to be taken into account for
medical devices, are also supported by methods and tools. Here, the user-centered design approach is
applied, which is also examined in more detail below.
4.3. Risk assessment for medical devices
The found publications are further analysed concerning their method application for risk assessment.
Special attention is paid to method adaptions and the target of the methods, which are applied for
executing the risk assessment within the development of medical devices. It is not possible at this
point to name an exact number of publications that use a particular method of risk assessment. Rather,
the analysis of the publications reveals generally valid aspects that are sometimes more and sometimes
less focused. In many cases the method is hardly emphasized, but rather the application example is
focused. Common features that occur particularly frequently are explained below:
Within the found methods, risk management for medical devices is most likely supported by Failure
Mode and Effects Analysis (FMEA) (e.g. Liu et al., 2012; Wang et al., 2019). Hereby, possible
product defects are tried to be estimated and possible consequences are assessed by the company.
Searching for *medical in combination with FMEA, further published and applied methods can be
analysed concerning their method adaptions. For medical products, the general approach can be
applied as usual, including the definition of occurrence, impact and prevention factors by the engineer.
For supporting the assessment of the exact factors, some publications provide different tables and
checklists with indicators directly related to the clinical situation (e.g. assessment of urgency or
stress) (e.g. Liu et al., 2012; Catelani et al., 2014).
Besides using tables resp. checklists, the rating of the occurrence, impact and prevention factors
should not be done only by the company, but by different experts with medical and technical
background. Thus, the risk analysis becomes more realistic since expertise of both disciplines is
integrated. For expressing the risk factors, it is suggested to not use crisp numbers but linguistic terms,
SOCIO-TECHNICAL ISSUES IN DESIGN 1975
like very low, medium or high, to express the failure modes, which help to integrate the mental
behaviour of the rating person.
With interdisciplinary FMEA supported by different checklists, the company pays special attention not
only to product failures but also to human factors and human failures occurring during the use phase.
But not only the use phase should be evaluated in risk management. Possible failures of the whole
product life cycle should be included and checked within the risk management to evaluate an overall
performance of a product. Within product development, at each milestone, risk management should be
integrated and requirements verified repeatedly.
Among other things, material checklists can for example support the detailed design and avoid failures
concerning biocompatibility and sustainability. Simulations can help with identifying failure modes
for the load and design of the medical device. Nevertheless, design validations should be performed in
practical tests which are carried out in an appropriate validation environment with the device user, as
errors can be best identified in this way.
To sum up, particular for the risk assessment for medical devices is the high degree of user integration
and the intensive assessment of human failures.
4.4. User-centered design for medical devices
It is clear from the methods applied in risk assessment that interdisciplinary work is of great
importance in the field of medical device development. In particular, the user of the medical device is
involved. User-centered design methods can be applied here. Therefore, UCD for medical devices will
be analysed in more detail in the following. One of the main points of interest is to what extent which
user is involved in which UCD method.
Publications found through the Scopus search are further analysed for the actual state of the art of
UCD methods, their point of application and the users involved in medical device development. The
results shall help to conclude recommendations for the design of personalized medical devices.
Excluding the duplications from the hit list, the remaining number of relevant papers from the UCD cluster
are 142. Herein, in 114 papers a UCD method or tool is applied. In 48 papers a case study of a medical
device development is described. Reducing the list of papers to case studies is necessary since only already
applied and proven methods shall be included in the analysis. Keeping all publications dealing with UCD
method application as well as a case study, 39 publications remain. However, none of these covers actually
a personalized medical device. Figure 4 combines both in section 2.2 presented UCD categorisations in
one, while collecting and counting all methods, tools and approaches within the remaining 39 papers. For
instance, in 8 of the 39 papers an interview technique is applied. It shall be explained here that in most of
the case studies more than one method, tool or approach is named; they are counted individually.
Think aloud (1)
Focus groups
(5)
User-driven
innovation (1)
User-centered
design (14)
Cognitive
walkthrough (1)
Designer empathy &
simulations
III IIIa
Strong focus
on (individual) user
Weak focus
on the (individual) user
User involvement with
expert mindset
Guidelines & standards
Universal user data & best
practices, publically available
Contextual user and usage data
and information of fictitious users
Contextual and individual user and usage data and information of real
users taking part in the design and/ or development process ...
Moderate focus
on the (individual) user
IIIb User involvement with
participatory mindset
... as research/ test subjects and a
source of data elicitation
as design partners and
collaborators
ISO 9241/10 (1)
ISO 9126 (1)
Activity theory (1)
Ergonomics (2)
Heuristics (2)
Nielsen´s expert heuristic (1)
Stakeholder analysis (1)
Users´needs analysis (1)
UTAUT2 model (1)
Empathic design (1)
Kano method (1)
Storyboard (1)
Simulation (1)
Recording (1)
Survey (2)
Questionnaire (4)
Interview (8)
Observation (5)
Contextual investigation (1)
Feedback (4)
Human factors (1)
Usability test (12)
Simulation/ scenario test (2)
Prototypes (3)
Sketch (1)
CAD model (1)
Mockups (2)
Digital measurement tool (1)
Ethnografic fieldwork (1) Co-elaboration (1)
Prototyping (1)
Contextual design (1)
Generative sessions (1)
Participatory
design (3) Workshop (1)
Methods
Tools
Approaches/ strategies
Figure 4. Overview of UCD approaches, methods and tools in medical device development
1976 SOCIO-TECHNICAL ISSUES IN DESIGN
Some of the approaches, methods and tools collected cannot be easily allocated into the sub-categories
of expert and participatory mindset and are therefore placed in between of both ends. This applies for
example for the user-centered design approach, which in theory can be applied with both mindsets, but
in most of the papers it is used with an expert mindset.
By collecting, allocating and counting the methods mentioned, a strong focus on the individual user
with an expert mindset can be seen in most case studies. Indeed, only very few papers do not mention
any kind of user involvement at all, i.e. weak or moderate user focus. In most of the cases, primary
users, i.e. medical and clinical staff (15) as well as patients (10), are involved. In a few cases also
secondary users, i.e. relatives and informal caregivers (3) are taking part in the design process, but
only on behalf of primary users. There is no mention of tertiary users. In almost all of the case studies
users are involved for the testing and evaluation of prototypes. This is further confirmed by the
number of UCD methods and tools related to testing, such as usability or scenario tests, feedback,
survey, questionnaires, measurement tools and all kinds of prototypes. Also, users are often involved
in the fuzzy front end of development, i.e. the planning and task clarification phase. Methods and
tools related to this phase are: interviews, observations, investigations, fieldwork, thinking aloud and
cognitive walkthrough. Only in a few case studies (5) users were involved in the concept phase.
Looking from the literature results, an already strong focus on individual users in design and
development of (non-personalized) medical devices can be seen. No especially for the development of
medical devices designed or adapted UCD method or tool was found inside of the results. In context
of medical devices, where various users, such as patients, relatives, caregivers, prosthetists and
physicians, may interact with the device on a regular basis, the involvement of secondary and maybe
also tertiary users needs become increasingly important.
4.5. Conclusion on method adaptions for medical device development
Many of the methods applied for medical devices deal with the risk assessment of medical devices. What
is repeatedly taken up in the methods is the fact that, in addition to theoretical considerations, practical
validation tests must be carried out to detect errors. The same result can be seen in the analysis of UCD
methods, where testing and evaluation of prototypes with user involvement is often focused.
Interdisciplinary work is of great importance. Otherwise checklists should at least be provided to help an
engineer to better understand the clinical routine. In summary, the study has shown that, when applying
design methods for medical devices, special issues of a medical device are considered as follows:
 legal and technical issues: continuous assessment of risks including product failures as well as
human failures in all development phases, especially by practical tests during design
validation (product validation and usability validation) and FMEA with user involvement
 human issues: method application within an interdisciplinary team of engineers and user of the
medical device; involvement of the primary user in particular during the early stages of the
development process including requirements assessment, but also for verification and especially
for design validation; at least assistance (e.g. checklists) for engineers when clinical/ medical
situations need to be assessed
4.6. Application of the results for personalized medical devices
As section 4.1 has shown, there is no variety management for medical devices or no application of
variety management for medical devices that has been found through the structured literature review.
A personalized medical device means the development or adaptation of the product specifically for the
individual patient, so that each patient receives his own specific product variant. As presented in the
introduction, an adequate approach to control variety-induced complexity during the development of
personalized products is needed.
The literature review shows, that safety issues are of particular importance. Hence, one of the biggest
challenges for variety management for medical devices is to ensure the safety of each patient specific
product variant. At the same time, in order to meet individual requirements, the degree and amount of
user involvement in the development of personalized medical devices has to be further increased.
Practically spoken, this means that the design process has to shift to a holistic user involvement with a
SOCIO-TECHNICAL ISSUES IN DESIGN 1977
participatory mindset. This includes that primary users are involved as partners over the whole design
process, especially in the concept phase, to discover unmet user needs and further increase usability
and user experience. Both, the necessity of risk assessment for each product variant and the
involvement of each specific user increases the variety-induced complexity. Therefore, with the help
of a new method, the search for a compromise between variety-induced complexity, emerging
especially of risk assessment and user involvement, and the benefits of an individual adaptation should
be supported.
5. Discussion and limitations
Overall, it was very difficult to find methods specifically for the development of medical devices. This
could be due to the fact that very different products are grouped under the definition. All medical devices
have in common that legal issues must be met and high-quality products with proven conformity and low
risk must be developed. The results of the paper underline this once again.
Only publications from the Scopus database were used for the research. Although Scopus contains a very
large number of peer-reviewed publications, it is possible that some relevant publications are not
included. With the help of other databases, further publications not listed in Scopus can be found and
analysed to further strengthen the results of the paper. The same limitation results from the fact that
maybe not all relevant combinations of keywords have been used and therefore relevant results remained
hidden. Particularly in case of the UCD research field, to which numerous approaches and strategies as
well as synonyms can be assigned, the search carried out here may not show the true circumstances,
since not all researchers, who use a UCD method or approach may also name it that way. In this case
too, the search could be extended so that further relevant publications may be found.
Also, many publications presented the application of a method in the development process of a medical
device, but did not detail the method adaptations that resulted. However, it may be that small and
unobtrusive method adaptations indeed took place in context of medical product development, but were
not reported in that detail. Therefore, in many places it was difficult to assess whether an adaptation was
due solely to the fact that it was a medical device. Often the medical device is only a use case for one
method, which can also be used for other products. This fact can also be found in the methods specified
for the application in design of medical devices. Aspects such as comprehensive risk management or the
involvement of the user in the development also apply to non-medical products. However, these are
explicitly emphasized as an indispensable necessity for medical devices. This is presumably due to the
severity of the effects that a defective medical device that has not been developed with the intended
purpose in mind will have. Since medical devices serve nothing less than human health, special attention
should be paid during development to intensive risk management and sufficient user involvement. These
particularities should already be considered in the methods and methodologies that support the
development of medical devices.
6. Conclusion
In this paper, methods supporting the development of medical devices are examined as well as the
extend, to which legal, human and technical issues are addressed. For this purpose, an extensive
literature review was conducted, in which different design methods used in medical device
development were examined. Taking into account the limitations of the paper, it was shown that in
particular risk management and user involvement are methodically supported. Methods of user-
centered design show a strong focus on primary users with involved under an expert mindset. These
users are particularly involved in the early development phases as well as in testing and evaluation.
Methods of risk management show a strong focus on interdisciplinary and test-based assessment of
risk factors, usually by means of FMEA. In this way, both product and human failures can be detected.
In summary, it can be seen that safety issues are of paramount importance for medical devices. For
this reason, both product and usability risks are examined in the risk assessment with user
involvement. For personalizing medical devices, the risk and the real benefit of a personalized device
needs to be assessed in addition to the management of variety-induced complexity.
1978 SOCIO-TECHNICAL ISSUES IN DESIGN
Acknowledgment
The authors would like to thank the German Federal Ministry of Education and Research for founding this work
within the project BELUCCI (13GW0274D) belonging to the program “Individualisierte Medizintechnik 2”.
References
Alexander, K. and Clarkson, P.J. (2000), Good design practice for medical devices and equipment, Part I: a
review of current literature”, Journal of Medical Engineering & Technology, Vol. 24 No. 1, pp. 5-13.
https://doi.org/10.1080/030919000293987
Azman, N. et al. (2017), “Review on Design for Medical Device”, MATEC Web Conference, Vol. 135 No. 5, p.
20. https://doi.org/10.1051/matecconf/201713500020
Bundesverband Medizintechnologie (BVMed) (2016): Der lange Weg eines Medizinprodukts von der Idee bis
zur Anwendung am Patienten”, Article, 20.01.2016
Catelani, M., Ciani, L. and Risaliti, C. (2014): Risk assessment in the use of medical devices: A proposal to
evaluate the impact of the human factor, 2014 IEEE International Symposium on Medical Measurements
and Applications (MeMeA), Lisboa, pp. 1-6. https://doi.org/10.1109/MeMeA.2014.6860088
da Silveira, G.J.C., Borenstein, D. and Fogliatto, F.S. (2001), “Mass customization: Literature review and
research directions”, Internatioanl Journal of Production Economics, Vol. 72 No. 1, pp. 1-13.
https://doi.org/10.1016/S0925-5273(00)00079-7
European Parliament (2017), Regulation (EU) 2017/745, Council of 5 April 2017
Gräßler, I. (2004), Kundenindividuelle Massenproduktion. Entwicklung, Vorbereitung der Herstellung,
Veränderungs-management, Springer, Heidelberg. https://doi.org/10.1007/978-3-642-18681-3
Kuhl, J. and Krause, D. (2018), Strategies for Customer Satisfaction and Customer Requirement Fulfillment
within the Trend of Individualization”, 29th CIRP Design Conference, 8. - 10. Mai 2019, Procedia CIRP,
Póvoa de Varzim, https://doi.org/10.1016/j.procir.2019.04.278
Koren, Y. et al. (2013), Open-architecture products”, In: CIRP Annals Vol. 62 No. 2, pp. 719-729.
https://doi.org/10.1016/j.cirp.2013.06.001
Lantada, A.D. and Morgado, P.L. (2013), General Considerations for the Development of Biomedical Devices,
In: Lantada, A.L. (Eds.): Handbook on Advanced Design and Manufacturing Technologies for Biomedical
Devices, Springer, Boston, pp. 19-47. https://doi.org/10.1007/978-1-4614-6789-2
Lindemann, U., Reichwald, R. and Zäh, M.F. (2006), Individualisierte Produkte - Komplexität beherrschen in
Entwicklung und Produktion, Springer, Berlin. https://doi.org/10.1007/3-540-34274-5
Liu, L. et al. (2012), Use-related risk analysis for medical devices based on improved FMEA, WORK: A Journal
of Prevention, Assessment & Rehabilitation, https://doi.org/10.3233/WOR-2012-0976-5860
Luna, D. et al. (2016), “Impact of participatory design for drug-drug interaction alerts. A comparison study
between two interfaces”, MIE, pp. 68-72.
Lettl, C. (2007), “User involvement competence for radical innovation”, Journal of engineering and technology
management, Vol. 24 No. 1-2, pp. 53-75. https://doi.org/10.1016/j.jengtecman.2007.01.004
Maier, A. et al. (2017), “Sensing behaviour in healthcare design”, In: Proceedings of the 21st International
Conference on Engineering Design (ICED 17), Vancouver, Canada, 21-25.08.2017.
Pahl, G. et al. (Eds.) (2007), Engineering Design. A Systematic Approach, Vol. 3, Springer, Berlin.
https://doi.org/10.1007/978-1-84628-319-2
Sanders, E.B.N. (2006), Design research in 2006. Design research quarterly, Vol. 1, No.1, pp.1-8.
Sankowski, O. and Krause, D. (2018), “Using Multi-Channel Human-System Interaction for User-Centered
Product Design”. Proceedings of ASME 2018 International Mechanical Engineering Congress and
Exposition, November 9-15, Pittsburgh, Pennsylvania, USA. https://doi.org/10.1115/IMECE2018-88091
Santos, I.C.T (2013), Product development methodologies: the case of medical devices, Thesis, Faculdade de
Engenharia da Universidade do Porto, Portugal
Simpson, T.W. et al. (2014), Advances in Product Family and Product Platform Design”, Methods &
Applications, Springer, New York. https://doi.org/10.1007/978-1-4614-7937-6
Spallek, J. and Krause, D. (2016), Process Types of Customisation and Personalisation in Design for Additive
Manufacturing Applied to Vascular Models”, Procedia CIRP, Vol. 50, pp. 281-286. https://doi.org/10.1016/
j.procir.2016.05.022
US Food and Drug Administration (FDA) (1997), Design control guidance for medical device manufacturers,
Food and Drug Administration, Rockville.
Wang, L. et al. (2019), A linguistic risk prioritization approach for failure mode and effects analysis: A case
study of medical product development. Quality and Reliability Engineering International, pp. 1735-1752.
https://doi.org/10.1002/qre.2472
... To effectively design solutions that meet the specific requirements of a given medical task, it is essential to conduct a thorough and comprehensive characterization of the treatment process (Coulentianos et al., 2020;Kuhl et al., 2020;Surma-aho et al., 2021). ...
... Specifically, the initial definition of the Tool Center Point (TCP) and robotic base three-dimensional (3D) position is critical since they directly affect the kinematic control of the robotic arm as well as their 3D positioning in the task space. Indeed, technology solution errors are normally related to poor usability/task requirement specification rather than direct device failure (Coulentianos et al., 2020;Kuhl et al., 2020;Surma-aho et al., 2021). Traditionally, the treatment characterization can be performed through contact with clinical experts, interviews, observation of medical treatment sessions, and treatment investigation (Saeedi-Hosseiny et al., 2022;Surma-aho et al., 2021). ...
... An important topic in the development of medical devices is the increasing demand for international standards for patient safety and security (Gualtieri et al., 2022;Kuhl et al., 2020). ...
Article
Full-text available
While designing an end‐effector tool for a specific task using an already available robotic arm is a common strategy for developing robot‐based solutions, it is subject to the designer's subjective biases, especially in the medical robotics field. Consequently, this approach can result in suboptimal solutions that may limit the performance of the robotic arm and the overall system. This work introduces a novel framework for developing a virtual scenario that enables the robust optimization of design guidelines for robotics‐based medical solutions. The proposed framework aims to address the issue of subjectivity involved in the design process, improving the reliability and accuracy of the robotic system. Specifically, it aims to find the optimized 3D positioning of the Tool Center Point and the robot base that takes full advantage of the robotic arm kinematics. The feasibility of the proposed methodology is validated for designing a system for robotic‐guided vascular laser treatments. First, a simulated treatment environment is constructed to compute the optimal trajectories to treat a wide population of synthetic human models. Next, a design optimization methodology based on robot task performance metrics is developed. Overall, the proposed design guidelines allowed the performance of 94.1% of the treatment trajectories free of collisions with median precision of around 0.5 mm while guaranteeing high manipulability. Thus, proving the added value of the proposed strategy and showing the potential for promoting the development of optimized robotics‐based solutions.
... In some cases, identifying the "main" end user can be challenging, on top of the divergence of concerns and the difficulty to align them in between patients and practitioners (Andersen, 2019;Money et al., 2011;Oliveira et al., 2020). Among the consequences, "ill-targeted or wrongly applied user centrism may also delay or threaten the outcome of a product development process" (Kuhl et al., 2020). Finally, an intrinsic limitation of Design thinking could be its lack of emphasis on regulatory requirements, leading Arandia et al. (2023) to question its use for the development of MedTech. ...
Conference Paper
Smart medical technologies arise from the collaboration of multiple engineering disciplines, resulting in increased technical and organizational complexity. Compounded by the intrinsic complexities of the medical sector, their development can prove challenging. In this sense, this article focuses on their development from a methodological standpoint. Leveraging the review of six distinct product development approaches and processes, the proposed research direction explores Agile Systems engineering and especially a V-model-Scrum process model for the development of smart medical technologies.
... Finally, as the BreathSpec devices were strategically designed and manufactured for this project, we thought it was important to consider a patient's perspective on providing a breath sample and on the diagnostic device itself. To achieve this, a user-centerd design was implemented, which has shown to lead to better quality products with lower error rates, while establishing efficient and effective devices that address the diverse needs of participants, 28,29 . This study reported a 49% RR for feedback associated to delivering a breath sample and a 29% RR on the device itself. ...
Article
Full-text available
Diagnosis of respiratory tract infections, especially in primary care, is typically made on clinical features and in the absence of quick and reliable diagnostic tests. Even in secondary care, where diagnostic microbiology facilities are available, these tests take 24-48 hours to provide an indication of the aetiology. This multicentre study used a portable Gas Chromatography-Ion Mobility Spectrometer for the diagnosis of bacterial RTIs. Breath samples taken from 570 participants with 149 clinically validated bacterial and 421 non-bacterial respiratory tract infections were analysed to distinguish bacterial from non-bacterial RTIs. Through the integration of a sparse logistic regression model, we identified a moderate diagnostic accuracy of 0.73 (95% CI 0·69, 0·77) alongside a sensitivity of 0·85 (95% CI 0·79, 0·91) and a specificity of 0·55 (95% CI 0·50, 0·60). The GC-IMS diagnostic device provides a promising outlook in distinguishing bacterial from non-bacterial respiratory tract infections and was also favourably viewed by participants.
... For a better adaptation of the medical device to the patient specific anatomies, the consideration of user's specific requirements in the product development process is essential but results in an increase of the variety-induced complexity caused by the variety of requirements, parts, components, products as well as processes within a company (Simpson et al., 2014). To manage this variety-induced complexity, design for variety methods and modularization approaches can be applied (Simpson et al, 2014;Rennpferdt et al., 2020), but a lack in application to medical devices has been found in literature review (Kuhl, J. et at., 2020) except for the analysis made by Kuhl et al. (2021). Several works focused on the application of user-centred design for the development of personalised wearable devices (Perego et al., 2022), therapies (Imbesi et al., 2022) or assistive devices (Santos et al., 2020) are available, but none of them report the design and implementation of a methodology to develop an automated tool for product customisation based on user anthropometry as reported in this paper. ...
Conference Paper
IntroductionThe objective of this study was to define a new approach for product design and customization. The approach focused on the development of several technical components able to provide product personalisation as an integrated solution to simplify the interactions between the end-user and the producer. The solution was developed as a web platform for an easy configuration of the product by the user at home and applied to a wheelchair. The finalization of a new order and the assembly procedure of the product is fully digital. This research activity is funded by the EU Project n. 872570 “KYKLOS 4.0 - An Advanced Circular and Agile Manufacturing Ecosystem based on rapid reconfigurable manufacturing process and individualized consumer preferences” (H2020-DT-2018-2020 - Digitising and transforming European industry and services: digital innovation hubs and platforms) which aims at simplifying mass customization while improving circularity and sustainability.Methods Our Product Parametric Design Methodology was developed through a specific ergonomic analysis of products related to human factors and anthropometry. The use of virtual manikins and simulation of different percentiles, allowed to test and better define correct inputs and parameters for product customization. The solution has been developed and applied to a medical use case and validated and integrated in the KYKLOS 4.0 main platform. The aim was to provide a coherent user experience for the wheelchair ordering and customization, so the user can easily select wheelchair frame and mandatory parts, optional accessories and customize additional elements such as the cushion and footrest. Several Technical Component were developed in order to guarantee the final workflow. The methodology was applied into our “Recommendation Engine” and integrated with the “Users' Behaviour Model”, which use patient’s information, such as anthropometry and behaviour to better define best fitting products and accessories. Additionally, the workflow integrated a specifically designed “3D Modelling Component” which allows direct parts customization.ResultsThe Product Personalization Solution was intended to support the customer in the selection of parts and accessories to fully customize the product and, at the same time, also to support product designer to convert customers’ requests efficiently and smoothly into personalized product specifications, thus providing a complete customer-oriented design for manufacturing with time and cost reduction. The methodology can be considered part of what is called "ECO-DESIGN," which is the design of products based on the sustainable and minimal use of resources and which enables high-quality recycling of materials at the end of the product life cycle. The use of the circular economy concept adopted in the specific pilot case is certainly relevant to the production of customized components for the wheelchair provided by our end-user, as the production of customized components allows a reduction in the resources required to produce parts. Moreover, since these are products that are perfectly compatible with the end user's needs, the possibility of having returned products (when components does not meet patient's needs) or unused products is considerably reduced, resulting in waste of materials used. Within this specific Use Case the customization of the wheelchair frame and accessories, based on the user's anthropometry, allows the reduction of unnecessary machining such as, for example, unused holes for ergonomic adaptation. It also turns out to be reduced waste production of aluminium tubulars resulting in reduced processing time and harmful emissions released into the environment.Conclusion Although the solution has been applied for demonstration purposes to the personalization of a wheelchair, it is exploitable to other products where personalization based on anthropometric measurements is essential for maximizing patient comfort and reduce production errors and waste. The addressed medical case study highlighted some very interesting aspects concerning the replication and scale up potential in the medical domain. The methodology based on anthropometry to define product requirements can also be applied very effectively to a variety of other products. Some application cases already exist in fashion for clothes and accessories size selection. Furniture products such as chairs, desks and wardrobes can also be customized using the same approach. We also believe that sport equipment has a high potential for such application. The automation of this process and the direct relationship with the user's dimensions could open potential markets currently reserved only to very expensive bicycle frames. Also, the production of accessories and tool, made with additive manufacturing technologies, could lead to disruptive innovation in those sectors providing true mass customization capabilities.
... While their outputs constitute essential elements of an RM framework, yet they did not establish the aspired process and the level of details aimed by this research. Furthermore, according to a recent study, 28 legal, human, and technical issues were found essential while implementing RM in MDD which respectively include for example safety, interdisciplinarity and biocompatibility risks. Hence, if the studies of Pietzsch et al 18 and Kuhl et al 28 were to be investigated and explored further, the types of risk to be explored throughout the early MDD phases can be researched. ...
Article
Full-text available
Introduction Whilst risk management has become fundamental in the development of medical devices, enforced by regulations and international standards, there is still no comprehensive model that explains how risk management in medical devices’ development should be tackled, especially with regard to the type of risks that should be addressed. Risk management in the medical devices’ development field is currently focused on technical risks, comprising product, usability, and development process risks, in alignment with standards’ requirements and regulations, without giving enough attention to non-technical risks, which include business and project risks. Start-ups within this heavily regulated domain have a key role in the innovation process, yet they suffer a structural lack of tangible, such as financial capacity, and intangible resources such as development, risk management, and regulations’ compliance. Nonetheless, they can still optimize their risk identification coverage beyond the enforced requirements to increase their products’ chances of success. Methods A set of qualitative interviews, serving the adopted grounded theory building research method, with seven start-ups who are involved in the development, commercialization, and quality control of medical devices was accomplished. The purpose was to determine the applied risk management practices and most importantly identify the risk types covered by them. Since every start-up is a project by itself, a sample of project risks, as identified by the project management institute, was utilized to scope the risk coverage and flag missing non-technical risks by the participating start-ups. Results Un-identified risk types, lack of involvement of the right teams, and other related loopholes were presented. Discussion A list of requirements was developed and sketched in a user-friendly risk management framework, which is believed to be crucial in helping start-ups attain successful, safe, and regulatory compliant medical devices production, is shared in the discussion and proposed framework section of this paper.
Article
Full-text available
The risk-based approach is one of the keys used in the implementation of management systems using requirements from the management standards. If the management systems are implemented separately, the risks are evaluated from only one perspective, but when two or more systems are integrated, the approach must take into account the synergetic effects of the risks due to system integration. The paper aims to highlight the specific risks encountered when implementing management systems in organizations, underlining the risks associated with systems management and integrated management systems. In this work, a synthesis was presented including an analysis of the risks that occurred during the implementation of management systems and those that occurred during their integration, the need and opportunity of risk management, the advantages, and barriers in the approach of risk management in the industrial sector. The review of the literature comprises the assessment of 190 papers, 11 books, and 5 standards. The content of the paper highlights a synthesis of risk-based thinking and the risks that can manifest in the implementation of quality, environmental, and occupational safety and health management systems in industrial organizations, as well as those which can manifest in the implementation of integrated management systems. In addition, the benefits and barriers in risk management and risk management strategies, with an emphasis on the presentation of the Deming approach to risk management, were evidenced. Some recommendations and proposals for future research were made.
Chapter
The central role of hospitals in today’s healthcare organisation provided its ineffectiveness in building a resilient healthcare service, especially during the COVID-19 pandemic. The desire and effort to move healthcare services to the territory, with a perspective of de-hospitalisation, have offered new healthcare places in a capillary network whose home becomes the focal point. The shift of some medical and care practices towards the household, and the role of technologies in this transition, impose a rethinking of the medical device design used by patients, family members, and caregivers. The presented work fits into the debate on designing new systems and products for homecare services with a human-centred telemedicine vision. Indeed, to ensure the success of telemedicine products and homecare devices, it is critical to understand the potential of emerging technologies and the design requirements needed to support their use. Therefore, the contribution starts with three questions: (i) what characteristics should a homecare medical device have? (ii) what method and process should be used to design medical devices? (iii) what skills must the designer need to deal with a medical project? and would define the knowledge necessary for the designer to operate in the medical world, the desirable characteristics of homecare medical products and through which strategies it is possible to achieve them. Thus, combining academic research and educational experience outlined some design requirements that might innovate and improve telemedicine and homecare products to draft a preliminary guide for designers intending to operate in the healthcare sector.
Article
Full-text available
The increase of the aging population brings numerous challenges to health and aesthetic segments. Here, the use of laser therapy for dermatology is expected to increase since it allows for non-invasive and infection-free treatments. However, existing laser devices require doctors’ manually handling and visually inspecting the skin. As such, the treatment outcome is dependent on the user’s expertise, which frequently results in ineffective treatments and side effects. This study aims to determine the workspace and limits of operation of laser treatments for vascular lesions of the lower limbs. The results of this study can be used to develop a robotic-guided technology to help address the aforementioned problems. Specifically, workspace and limits of operation were studied in eight vascular laser treatments. For it, an electromagnetic tracking system was used to collect the real-time positioning of the laser during the treatments. The computed average workspace length, height, and width were 0.84 ± 0.15, 0.41 ± 0.06, and 0.78 ± 0.16 m, respectively. This corresponds to an average volume of treatment of 0.277 ± 0.093 m³. The average treatment time was 23.2 ± 10.2 min, with an average laser orientation of 40.6 ± 5.6 degrees. Additionally, the average velocities of 0.124 ± 0.103 m/s and 31.5 + 25.4 deg/s were measured. This knowledge characterizes the vascular laser treatment workspace and limits of operation, which may ease the understanding for future robotic system development.
Conference Paper
Full-text available
Individualization is one of the current megatrends in the product development. The literature offers approaches for customization of which individualization emerged. Nevertheless, it is not emphasized if the approaches are also applicable for focusing each individual customer and his requirements within individualization. Also, the achievable customer’s requirements fulfillment compared to the induced internal complexity is disregarded. Therefore, this paper clusters different individualization strategies for fulfilling customer-individual requirements and emphasizes the induced complexity, solution approaches, the resulting customer benefit and limitations for each strategy. Afterwards, decision-making models for finding the right strategy for the own business are reviewed and discussed.
Article
Full-text available
Failure mode and effects analysis (FMEA) is a widely used risk management technique for identifying the potential failures from a system, design, or process and determining the most serious ones for risk reduction. Nonetheless, the traditional FMEA method has been criticized for having many deficiencies. Further, in the real world, FMEA team members are usually bounded rationality, and thus, their psychological behaviors should be considered. In response, this study presents a novel risk priority model for FMEA by using interval two‐tuple linguistic variables and an integrated multicriteria decision‐making (MCDM) method. The interval two‐tuple linguistic variables are used to capture FMEA team members' diverse assessments on the risk of failure modes and the weights of risk factors. An integrated MCDM method based on regret theory and TODIM (an acronym in Portuguese for interactive MCDM) is developed to prioritize failure modes taking experts' psychological behaviors into account. Finally, an illustrative example regarding medical product development is included to verify the feasibility and effectiveness of the proposed FMEA. By comparing with other existing methods, the proposed linguistic FMEA approach is shown to be more advantageous in ranking failure modes under the uncertain and complex environment.
Article
Full-text available
Medical devices are one of the uprising star in product development due to high demand among users. Developing safe, effective and usable design is indeed a challenging work which faced by designer nowadays. There are comparative methods in designing medical device to achieve desired outcome. Weather designing a “User-friendly”, “Safe to use” or “Eco-friendly” product requires an exclusive method. This review paper might illuminate some light to the designers and researchers who choose to integrate those aspect in their design work in healthcare sector. A review on current method used on user-friendly approach, ensuring safety, regulations on safety of medical device and suggested standards to ensure safety in less complicated way. Thus, there is no solid method or standard yet to establish on safety for medical device.
Conference Paper
Full-text available
We are entering an era of distributed healthcare that should fit and respond to individual needs, behaviour and lifestyles. Designing such systems is a challenging task that requires continuous information about human behaviour on a large scale, for which pervasive sensing (e.g. using smartphones and wearables) presents exciting opportunities. While mobile sensing approaches are fuelling research in many areas, their use in engineering design remains limited. In this work, we present a collection of common behavioural measures from literature that can be used for a broad range of applications. We focus specifically on activity and location data that can easily be obtained from smartphones or wearables. We further demonstrate how these are applied in healthcare design using an example from dementia care. Comparing a current and proposed scenario exemplifies how integrating sensor-derived information about user behaviour can support the healthcare design goals of personalisation, adaptability and scalability, while emphasising patient quality of life.
Article
Full-text available
Manufacturing companies face high demand for products that fulfil individual customer desires. Recent improvements in additive manufacturing (AM) enable the fabrication of customer-specific components of a product.
Article
Full-text available
In order to effectively analyze and control use-related risk of medical devices, quantitative methodologies must be applied. Failure Mode and Effects Analysis (FMEA) is a proactive technique for error detection and risk reduction. In this article, an improved FMEA based on Fuzzy Mathematics and Grey Relational Theory is developed to better carry out use-related risk analysis for medical devices. As an example, the analysis process using this improved FMEA method for a certain medical device (C-arm X-ray machine) is described.
Conference Paper
User-centered product design (UCPD) and especially its methods and tools offer a lot of benefits to product development. By using specific data of the user group or by including them into the design process, systems with better functionality and usability arise. However, including the users in an optimum manner means to include them over the whole product development process, which is costly and often too time-consuming regarding the ever shorter product life cycles. An extensive application of UCPD methods is therefore usually not practical for industry. In order to (1) support the user-centered development process in general and (2) support the selection of appropriate UCPD methods, a multi-channel human-system interaction framework is proposed. It is derived from existing human-computer and human-machine interaction models and further includes additional factors influencing the human-system interaction. However, the framework itself needs further and more detailed elaboration and discussion and currently lacks an allocation of UCPD methods.
Article
Decision support systems for alert drug-drug interactions have been shown as valid strategy to reduce medical error. Even so the use of these systems has not been as expected, probably due to the lack of a suitable design. This study compares two interfaces, one of them developed using participatory design techniques (based on user centered design processes). This work showed that the use of these techniques improves satisfaction, effectiveness and efficiency in an alert system for drug-drug interactions, a fact that was evident in specific situations such as the decrease of errors to meet the specified task, the time, the workload optimization and users overall satisfaction with the system.
Chapter
The development process of medical devices and of any products oriented to interacting with biological systems (biodevices) involves several special features deriving from the typical multidisciplinary characteristics of such devices and of their surrounding environment. Therefore, the systematic development methodologies previously described have to integrate several additional special considerations and indeed very specific recommendations, for adequately helping to face up with the development of novel biodevices. Aspects such as the existence of a relevant medical need; the effects of biological conditions; the selection of adequate biomaterials, sometimes with unusual mechanical and chemical properties; the consequences of corrosion; or the sterilisation methods available have to be considered, almost from the beginning, when developing a new biodevice. Development teams also integrate normally engineers, physicians, biologists and personnel from different disciplines, and sometimes, communication problems, together with project delays and even cost mismatches, arise. All this has to be taken into account in these projects. The regulatory framework is also especially noteworthy in the medical device field, due to their potential harm when interacting with tissues and organs, and different directives have to be followed carefully. The most relevant EU directives of application for medical devices, together with advices included in important ISO standards, as well as some discussion and comparison with approaches from other countries (United States' FDA, Asian market...), are commented at the end of the chapter.