ChapterPDF Available

On the Applicability of Low-cost Sensors for Measurements of Aerosol Concentrations

Authors:

Abstract

One of the most important parameters of air pollution, concentration of particulate matter suspended in the air (aerosols), is measured by means of low-cost sensors of latest generation. Such measurements have high resolution in both space and time, but there is a big question about applicability of low-cost sensors for measurements of aerosol concentrations that arise from the complexity of measurement technique, and inevitable errors associated. In this paper we analyze all aspects of applicability of low-cost sensors for long-term measurement campaigns.
30TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
DOI: 10.2507/30th.daaam.proceedings.060
ON THE APPLICABILITY OF LOW-COST SENSORS FOR
MEASUREMENTS OF AEROSOL CONCENTRATIONS
Adnan Masic, Dzevad Bibic & Boran Pikula
This Publication has to be referred as: Masic, A[dnan]; Bibic, D[zevad] & Pikula, B[oran] (2019). On the Applicability
of Low-cost Sensors for Measurements of Aerosol Concentrations, Proceedings of the 30th DAAAM International
Symposium, pp.0452-0456, B. Katalinic (Ed.), Published by DAAAM International, ISBN 978-3-902734-22-8, ISSN
1726-9679, Vienna, Austria
DOI: 10.2507/30th.daaam.proceedings.060
Abstract
One of the most important parameters of air pollution, concentration of particulate matter suspended in the air (aerosols),
is measured by means of low-cost sensors of latest generation. Such measurements have high resolution in both space
and time, but there is a big question about applicability of low-cost sensors for measurements of aerosol concentrations
that arise from the complexity of measurement technique, and inevitable errors associated. In this paper we analyze all
aspects of applicability of low-cost sensors for long-term measurement campaigns.
Keywords: Low-cost sensor; Air pollution; Aerosol concentration; Calibration.
1. Introduction
Air pollution is a major issue nowadays: it affects many aspects of our lives, from the health (directly) to the economy
(indirectly). There are various components of air pollution, and among all, particles suspended in the air (aerosols) are
primary pollutant in most cases [1], [2]. In order to assess the level of air pollution by aerosols, we need to measure their
concentrations in the air. But not all aerosols are equally harmful. Particles are classified by their size, and generally,
smaller particles are more dangerous. There are many standards and recommendations for particle concentrations,
proposed from different organizations. All of them are based on the concentrations of particles smaller than 10 m (PM10)
or 2.5m (PM2.5). In this study, we will use PM2.5 measurements, since they appear in most studies (PM10 is slowly
becoming deprecated for this purpose). The problem is the fact that proper measurement of aerosol concentrations is
complex and expensive. Thus, for most polluted areas (poor and developing countries), we don’t have much (if any) data.
Recently, new technology has emerged in this field: the low-cost sensors. Table 1 shows main classification of devices
for PM2.5 concentration measurements. Most accurate methods are based on gravimetric measurements, but they are
expensive and the resolution is very limited in both space and time. Equivalent methods (usually based on the attenuation
of beta radiation, BAM) are not as accurate as referent methods, but they are cheaper to run and offer better time
resolution. Research grade portable devices, usually based on optical measurement techniques, have even lower cost,
better resolution, but they may give inaccurate results, due to the principle of their work. Low-cost sensors are very
popular because of their price, size and time resolution (typically 1 s) but the question is: can they be used for
measurements of PM concentrations at all? We will try to answer this question in the present paper.
- 0452 -
30TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
Method
Reference
Research grade calibrated
Low-cost
Device
example
Cost
Very high
Moderate
Very low
Accuracy
Best
Moderate/good, depending
on conditions
Questionable
Typ. time
resolution
1 day
1 second
1 second
Portability
Very low
High
Very High
Table 1. Classification of devices for measurement of aerosol concentrations
There are many papers in which performance of low-cost aerosol sensors is evaluated, for example [3], [4], [5]. However,
articles with long campaigns in realistic conditions are very rare. In this paper results of successful long-term comparison
of low-cost sensor with professional BAM monitor nearby are presented.
2. Low-cost sensor of choice
Taking into account conclusions from our previous studies [5], [6] among all commercially available low-cost sensors,
we have selected Plantower PMS5003 sensor. Main reasons for such choice are reliability of the sensors, consistence
between units from different batches of sensors, well implemented digital interface and good mean time before failure
(MTBF). It is optical based sensor which uses Mie scattering theory for laser scattering on spherical particles to calculate
size and concentration of particles in the unit volume of air. This approach, however, inherits several shortcomings for
accurate measurement of mass concentrations of aerosols, including:
- assumption that particles are spherical,
- inability to detect particles smaller than 0.3 m in diameter,
- assumption for index of refraction of particles,
- assumption of particles’ density.
These parameters may change in the atmosphere over time, which means that there is no correct and permanent factory
calibration. Furthermore, high relative humidity in the air may lead to the effect of hygroscopic growth of aerosols [7],
[8], which leads to overshoots in measurements at such conditions.
Low-cost sensors are not weather proof, which means that we need to provide proper enclosure. That is not an easy
task, because the enclosure must not obstruct sampling of aerosols. Putting PMS5003 sensor into simple box with some
openings for the air is not solution, because the sensor has both inlet and outlet on the same side, and such enclosure
would lead to inaccurate results due to the cyclone effect. Data acquisition must be performed flawlessly because
measurement loses are not acceptable here. All these challenges were previously solved within the framework of our in-
house developed MAQS sensor ecosystem and data acquisition system [9].
3. Case study
The city of Sarajevo is affected by aerosol air pollution very much, and it was selected as our case study. Sarajevo is
situated in the natural valley, with strong episodes of temperature inversions during winter [10], [11] and very low number
of windy days. Results of this study may be relevant for other cities with similar topology of terrain as well. Figure 1
shows the locations of devices for the measurement campaign. We set up two MAQS sensors: MAQS@MEF was located
at the Faculty of Mechanical Engineering while MAQS@ETF was installed at the Faculty of Electrical Engineering,
University of Sarajevo. The referent device was professional BAM monitor (equivalent method in Table 1) located inside
of the complex of US Embassy in Sarajevo (BAM@Embassy). Data collected from the BAM@Embassy are publicly
available, but access to the Embassy device was not possible. Strictly speaking, two MAQS sensors were not collocated
with BAM, but they were close to each other (200 m from ETF to Embassy and 500 m from MEF to Embassy), so that it
is unlikely that atmosphere can differ significantly at such short distances.
- 0453 -
30TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
Fig. 1. Location of measurement devices
4. Results
Data from Embassy monitor were collected from 23/MAY/2018, MAQS@MEF started data acquisition on 7/17/2018
and MAQS@ETF started on 2/7/2019. All devices are running continuously as of today, 10/18/2019, which is the last
day included in this paper. During this period, many different scenarios were recorded.
Figure 2 shows comparisons of daily average PM2.5 values between two MAQS sensors and BAM@Embassy. The
coefficient of determination between MAQS and BAM was sensational: 𝑅2= 0.96 for MAQS@MEF and 𝑅2= 0.93 for
MAQS@ETF (please note that MAQS@MEF collected more data than MAQS@ETF, which started after the peak of air
pollution during winter 2018/19). The absolute values, however, didn’t agree well: MAQS@MEF measured 37% higher
value (on average) than BAM@Embassy, while MAQS@ETF recorded 31% higher value than BAM (on average).
Fig. 2. Daily average values of PM2.5 - comparisons between MAQS sensors and BAM
Figure 3 shows comparisons of hourly average values. Different factors associated to the principle of work of optical
aerosol sensor appear as noise in measurements on such timescale. The fact that sensors are not strictly collocated may
also play some role here. BAM monitor is not perfectly accurate either, so it is natural, that correlation is weaker in case
of hourly average values. But the results are still good: correlation between MAQS sensors and BAM device is high: 𝑅2=
0.86 for MAQS@MEF and 𝑅2= 0.79 for MAQS@ETF. Absolute values, again, are overestimated by similar factors as
in the case of hourly averages.
y = 0.7295x - 2.0544
R² = 0.9567
0
50
100
150
200
250
300
350
400
450
0100 200 300 400 500
BAM@Embassy
MAQS@MEF
PM2.5 daily avg from 7/17/18 [g/m3]
y = 0.7368x + 0.1248
R² = 0.9312
0
20
40
60
80
100
120
050 100 150
BAM@Embassy
MAQS@ETF
PM2.5 daily avg from 2/7/19 [g/m3]
- 0454 -
30TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
Fig. 3. Hourly average values of PM2.5 - comparisons between MAQS sensors and BAM
Consistency between two MAQS sensors is also important. Figure 4 shows daily and hourly average values for two
MAQS sensors. Since the sensors are not collocated (distance between them was about 300 m) we don’t expect perfect
match, but they should be well correlated. Indeed, daily averages had correlation coefficient 𝑅2= 0.98 while hourly
averages had 𝑅2= 0.96, which is excellent. Their absolute values were close (well within 10%).
Fig. 4. Daily and hourly average values of PM2.5 - comparisons between MAQS sensors at MEF and ETF
5. Conclusion
Low-cost sensors have potential for monitoring of aerosol concentrations: they are affordable and robust (at least those
used in this study), they have modern digital interface and can be adapted for outdoor use with proper enclosure. However,
low-cost sensors can’t be used out-of-the-box for direct measurements of mass concentrations. They must be calibrated,
which may be difficult and expensive. If there is any reference or equivalent public station nearby, that is excellent
opportunity for comparison and (some kind of) calibration of low-cost sensors. In such cases low-cost sensors could fill
the gaps between professional stations, taking into account their time and space resolution.
As suggestion for future work we propose the following activities:
- investigation of humidity influence on low-cost sensors and development of model of correction,
- measurement of concentrations of nanoparticles, which can’t be detected with low-cost sensors,
- strict colocation between low-cost sensors and reference monitor,
- using reference (gravimetric) methods for calibration of low-cost sensors under various conditions.
y = 0.6963x - 0.7673
R² = 0.8642
-100
0
100
200
300
400
500
600
0200 400 600
BAM@Embassy
MAQS@MEF
PM2.5 hourly avg from 7/17/18 [g/m3]
y = 0.6756x + 1.2625
R² = 0.793
0
20
40
60
80
100
120
140
160
180
200
050 100 150 200 250
BAM@Embassy
MAQS@ETF
PM2.5 hourly avg from 2/7/19 [g/m3]
y = 0.944x + 0.7886
R² = 0.9826
0
20
40
60
80
100
120
050 100 150
MAQS@ETF
MAQS@MEF
PM2.5 daily avg from 2/7/19 [g/m3]
y = 0.9047x + 2.6842
R² = 0.9577
0
50
100
150
200
250
300
350
400
0100 200 300 400
MAQS@ETF
MAQS@MEF
PM2.5 hourly avg from 2/7/19 [g/m3]
- 0455 -
30TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
5. Acknowledgments
We would like to thank to Mr. Damir Muslić for his generous contribution to our project by writing core part of the in-
house developed MAQS software.
6. References
[1] https://www.epa.gov/pm-pollution/health-and-environmental-effects-particulate-matter-pm, (2019). United States
Environmental Protection Agency, Accessed on: 2019-10-18
[2] https://www.who.int/airpollution/data/cities/en, (2018). WHO Global Ambient Air Quality Database (update 2018),
Accessed on: 2019-10-18
[3] Crilley, L. R.; Shaw, M.; Pound, R.; Kramer, L. J.; Price, R.; Young, S.; Lewis, A. C. & Pope F. D.(2018).
Evaluation of a Low-cost Optical Particle Counter (Alphasense OPC-N2) for Ambient Air Monitoring. Atmospheric
Measurement Techniques, 11, 709720, 2018, DOI: https://doi.org/10.5194/amt-11-709-2018
[4] Chatzidiakou, L. et al. Characterising low-cost sensors in highly portable platforms to quantify personal exposure
in diverse environments. Atmospheric Measurement Techniques, 12, 46434657, 2019, DOI:
https://doi.org/10.5194/amt-12-4643-2019
[5] Masic, A.; Pikula, B. & Bibic, D. (2017). Mobile Measurements of Particulate Matter Concentrations in Urban Area,
Proceedings of the 28th DAAAM International Symposium, pp.0452-0456, B. Katalinic (Ed.), Published by
DAAAM International, ISBN 978-3-902734-11-2, ISSN 1726-9679, Vienna, Austria DOI:
10.2507/28th.daaam.proceedings.063
[6] Masic, A.; Pikula, B.; Bibic, Dz.; Musemic, R. & Halac, A. (2018). Calibration and Assessment of Low-cost Dust
sensors, Proceedings of the 29th DAAAM International Symposium, pp.0523-0528, B. Katalinic (Ed.), Published
by DAAAM International, ISBN 978-3-902734-20-4, ISSN 1726-9679, Vienna, Austria, DOI:
10.2507/29th.daaam.proceedings.075
[7] Chen, J. et al. Aerosol hygroscopic growth, contributing factors, and impact on haze events in a severely polluted
region in northern China. Atmospheric Chemistry and Physics, 19, 13271342, 2019, DOI:
https://doi.org/10.5194/acp-19-1327-2019
[8] Jayaratne, R.; Liu, X.; Thai, P.; Dunbabin, M. & Morawska, L. (2018). The Influence of Humidity on the
Performance of a Low-cost Air Particle Mass Sensor and the Effect of Atmospheric Fog. Atmospheric Measurement
Techniques, Atmospheric Measurement Techniques, 11, 48834890, DOI: https://doi.org/10.5194/amt-11-4883-
2018
[9] Masic, A. (2015). Unmanned Aerial Vehicle as Data Acquisition System. Journal of Trends in the Development of
Machinery and Associated Technology. Vol. 19, No. 1, 2015, ISSN 2303-4009 (online), p.p. 181-184.
[10] Masic, A.; Musemic, R. & Dzaferovic-Masic, E. (2016). Temperature Inversion Measurements in Sarajevo Valley
Using Unmanned Aerial Vehicles, Proceedings of the 27th DAAAM International Symposium, pp.0423-0427, B.
Katalinic (Ed.), Published by DAAAM International, ISBN 978-3-902734-08-2, ISSN 1726-9679, Vienna, Austria,
DOI: 10.2507/27th.daaam.proceedings.062
[11] Masic, A.; Bibic, Dz.; Pikula, B.; Dzaferovic-Masic, E. & Musemic, R. Experimental study of temperature
inversions above urban area using unmanned aerial vehicle. Thermal Science, 2018, DOI:
10.2298/TSCI180227250M
- 0456 -
... This platform has a modular design and can carry various sensors and equipment, including electrochemical sensors for toxic gases [17], [18], and the abovementioned PM sensors. In our previous work we discussed quality assurance (QA) procedures that were applied to ensure reliable data [19], [20]. For this research, LiDAR with 100 meters range, LW20/C (LightWare, USA) was configured and installed on the drone (Figure 1). ...
Chapter
Full-text available
A novel method for terrain mapping is presented in this paper. It is based on low-cost open source components, which brings this innovative technique to a much wider group of researchers than ever before. The in-house developed drone was used together with a commercially available LiDAR sensor to produce a terrain mapping platform. This system was tested in two field locations with good results. In some aspects of performance, the drone surpassed commercially available solutions that are much more expensive and closed. Keywords: LiDAR; drone; terrain mapping; inertial measurement unit; global navigation satellite system.
... Mobile aethalometer MA200 from AethLabs (San Francisco, USA) with an inhouse developed enclosure was used. This component was designed to complement existing systems for mobile measurements of particulate matter [6], [7], toxic gases [8], [9] and vertical measurements using unmanned aerial vehicles [10]. Basic specifications of MA200 are: If the aethalometer uses single sampling spot at a time, some non-linear effects may appear as a consequence of filter loading effect. ...
Article
Full-text available
A new method for mobile measurements of black carbon (BC) concentration measurements is presented in this paper. An advanced pocket-sized aethalometer with 5 wavelengths and dual-spot technology was used. Different measurement locations were selected to evaluate the contribution of car emissions and biomass burning emissions during the winter campaign in heavily polluted urban environments. A model based on the Angstrom exponent is proposed for source apportionment studies. Obtained results confirmed the system's capability to distinguish BC sources with high temporal and spatial resolution.
... Two important features of our cloud system are: quality of service (QoS), which prevents data loss if there are temporary problems with internet connection and advanced calibration algorithm. The calibration procedure for particulate matter (PM) is explained in our previous work [6], [7], for toxic gases and CO2 in [8], [9], and for temperature and humidity in [10]. Fig. 1. (right) shows the batch of 10 smart sensors undergoing the calibration procedure in our laboratory. ...
Chapter
Full-text available
A novel method for indoor air quality monitoring is presented in this paper. It is based on the network of smart sensors permanently connected to the cloud. The prototype system, consisting of 10 smart sensors is evaluated in laboratory and real use. Each smart sensor was able to measure air temperature, relative humidity, particulate matter concentration, and carbon dioxide concentration. The cloud-based architecture of the system is explained, followed by the calibration method and real scenario results. The system proved to be suitable for real-time monitoring of indoor air quality parameters for large buildings.
Chapter
Full-text available
Nanoparticles with diameters from 10 nm to 1000 nm were analyzed using the scanning mobility particle sizer (SMPS) continuously from October 2020 till May 2021. Mass spectrometry was performed in 129 channels. Measurements of some other air pollution parameters were used for comparison and analysis. A large number of nanoparticles was detected in the period of strong air pollution, as expected. However, a surprisingly large number of nanoparticles was detected in spring and summer, where other parameters of air pollution were good. Keywords: SMPS; nanoparticles; air pollution; mass spectrometry.
Chapter
Full-text available
To provide the reference dataset for particulate matter concentrations, a long-term campaign of continuous gravimetric measurements of PM10 and PM2.5 concentrations was performed. Strict implementation of standard EN12341:2014 was fulfilled. Daily and monthly concentrations of both PM10 and PM2.5 were measured with the highest possible accuracy. Ratio PM2.5/PM10 was determined and analyzed. Different distribution of particles was observed during the winter period, where most of the particulate matter originates from burning fossil fuel, and during an episode of Sahara dust, occasionally observed in spring and summer. Keywords: PM10; PM2.5; gravimetric measurement; Sahara dust.
Chapter
Full-text available
A new approach for measurements of vertical profiles of temperature and particulate matter (PM) concentrations is developed and tested. Special purpose unmanned aerial vehicle with data acquisition system has been developed in-house. The temperature sensor is based on a fast response thermistor, while the PM sensor uses optical detection of particles. All sensors were tested in the laboratory and calibrated against reference instruments. Two urban locations were selected for field tests. This research produced a very useful asset that characterizes the relation between temperature inversion and air pollution. Keywords: temperature inversion; particulate matter; drone; boundary layer.
Chapter
Full-text available
Continuous measurement of black carbon (BC) concentrations in ambient air was performed using the optical-based instrument called aethalometer. The attenuation of light by BC was measured at 7 different wavelengths. This gives not only the pure concentration of BC in the ambient air but also dependence of attenuation on the wavelength of light, from which the Angstrom exponent can be derived. Based on the Angstrom exponent, we can estimate the percentage of BC that originates from solid fuel (wood and coal burning) or liquid fuel (diesel). Thus, a simple source apportionment analysis is presented. Keywords: black carbon; source apportionment; fossil fuels; biomass burning.
Article
Full-text available
PM2.5 hourly average measurements from 33 outdoor PurpleAir particle monitors were compared with hourly measurements from 27 nearby US EPA Air Quality System (AQS) stations employing Federal Equivalent Method (FEM) monitors in California over an 18-month (77-week) period. A transparent and reproducible alternative method (ALT) of calculating PM2.5 from the particle numbers in three size categories was used in place of the estimates provided by Plantower, the manufacturer of the sensors used in PurpleAir monitors. The ALT method was superior in several ways (better precision, lower limit of detection, improved size distribution) compared to Plantower’s CF1 or ATM data series. PurpleAir monitors were strongly correlated with the nearby US EPA Air Quality System AQS stations. A calibration factor (CF) ranging between 2.9 and 3.1 was empirically derived for the PurpleAir estimates using the ALT method. This value was based on comparing the average value of 177,329 PurpleAir measurements to the value calculated from the FEM stations. The monitoring period included about 13 weeks showing very high outdoor values due to several major fires covering several hundred thousand acres. The CF during these 13 weeks averaged 2.39, whereas the CF for the remaining 64 weeks averaged about 3.21, suggesting a different response to the smoke from wildfires compared with normal ambient fine particulate matter (PM2.5). The standard Plantower CF1 data series overestimated the FEM values by about 40%, in agreement with several other studies.
Article
Full-text available
The inaccurate quantification of personal exposure to air pollution introduces error and bias in health estimations, severely limiting causal inference in epidemiological research worldwide. Rapid advancements in affordable, miniaturised air pollution sensor technologies offer the potential to address this limitation by capturing the high variability of personal exposure during daily life in large-scale studies with unprecedented spatial and temporal resolution. However, concerns remain regarding the suitability of novel sensing technologies for scientific and policy purposes. In this paper we characterise the performance of a portable personal air quality monitor (PAM) that integrates multiple miniaturised sensors for nitrogen oxides (NOx), carbon monoxide (CO), ozone (O3) and particulate matter (PM) measurements along with temperature, relative humidity, acceleration, noise and GPS sensors. Overall, the air pollution sensors showed high reproducibility (mean R‾2=0.93, min–max: 0.80–1.00) and excellent agreement with standard instrumentation (mean R‾2=0.82, min–max: 0.54–0.99) in outdoor, indoor and commuting microenvironments across seasons and different geographical settings. An important outcome of this study is that the error of the PAM is significantly smaller than the error introduced when estimating personal exposure based on sparsely distributed outdoor fixed monitoring stations. Hence, novel sensing technologies such as the ones demonstrated here can revolutionise health studies by providing highly resolved reliable exposure metrics at a large scale to investigate the underlying mechanisms of the effects of air pollution on health.
Article
Full-text available
This study investigates the impact of the aerosol hygroscopic growth effect on haze events in Xingtai, a heavily polluted city in the central part of the North China Plain (NCP), using a large array of instruments measuring aerosol optical, physical, and chemical properties. Key instruments used and measurements made include the Raman lidar for atmospheric water vapor content and aerosol optical profiles, the PC-3016A GrayWolf six-channel handheld particle and mass meter for atmospheric total particulate matter (PM) that has diameters less than 1 and 2.5 µm (PM1 and PM2.5, respectively), the aerosol chemical speciation monitor (ACSM) for chemical components in PM1, and the hygroscopic tandem differential mobility analyzer (H-TDMA) for aerosol hygroscopicity. The changes in PM1 and PM2.5 agreed well with that of the water vapor content due to the aerosol hygroscopic growth effect. Two cases were selected to further analyze the effects of aerosol hygroscopic growth on haze events. The lidar-estimated hygroscopic enhancement factor for the aerosol backscattering coefficient during a relatively clean period (Case I) was lower than that during a pollution event (Case II) with similar relative humidity (RH) levels of 80 %–91 %. The Kasten model was used to fit the aerosol optical hygroscopic growth factor (GF) whose parameter b differed considerably between the two cases, i.e., 0.1000 (Case I) versus 0.9346 (Case II). The aerosol acidity value calculated from ACSM data for Case I (1.35) was less than that for Case II (1.50) due to different amounts of inorganics such as NH4NO3, NH4HSO4, and (NH4)2SO4. Model results based on H-TDMA data showed that aerosol hygroscopic growth factors in each size category (40, 80, 110, 150, and 200 nm) at different RH levels (80 %–91 %) for Case I were lower than those for Case II. For similar ambient RH levels, the high content of nitrate facilitates the hygroscopic growth of aerosols, which may be a major factor contributing to heavy haze episodes in Xingtai.
Chapter
Full-text available
Air pollution with particulate matter (PM) represents major risk for humans’ health. Important step towards the solution of the aforementioned problem is the correct measurement of PM concentrations in the air. Commodity (low-cost) sensors are very popular nowadays, since they are affordable and they fit well in the trend called Internet of Things (IoT). However, measurement of PM concentrations in the air is difficult task and low-cost sensors are unable to provide accurate readings out of the box. Thus, advanced calibration is necessary for any application of low-cost dust sensors. In this paper we propose calibration technique and discuss initial results. Keywords: low-cost sensors; particulate matter; air pollution; calibration; humidity corrections
Article
Full-text available
Vertical temperature profiles represent a very important factor for various analytical and numerical studies, such as weather forecasts, air pollution models and computational fluid dynamics simulations. These temperature profiles are especially important during the winter periods, when temperature inversions occur. The cities in the natural valleys, such as the city of Sarajevo, are strongly affected by this phenomenon. In this paper, a method for quantitative characterization of vertical temperature profiles, which is based on the in-house developed data acquisition system and the unmanned aerial vehicle, is presented. Comprehensive calibration and verification procedure was performed and explained in details. Field measurements were focused on the winter period and extreme temperature inversion scenarios. The correlation with the air pollution in the city, for the same period, was discussed as well.
Article
Full-text available
While low-cost particle sensors are increasingly being used in numerous applications, most of them have no heater or dryer at the inlet to remove water from the sample before measurement. Deliquescent growth of particles and the formation of fog droplets in the atmosphere can lead to significant increases in particle number concentration (PNC) and mass concentrations reported by such sensors. We carried out a detailed study using a Plantower PMS1003 low-cost particle sensor, both in the laboratory and under actual ambient field conditions, to investigate its response to increasing humidity and the presence of fog in the air. We found significant increases in particle number and mass concentrations at relative humidity above about 75 %. During a period of fog, the total PNC increased by 28 %, while the PNC larger than 2.5 µm increased by over 50 %. The PM10 concentration reported by the PMS1003 was 46 % greater than that on the standard monitor with a charcoal dryer at the inlet. While there is a causal link between particle pollution and adverse health effects, the presence of water on the particles is not harmful to humans. Therefore, air quality standards for particles are specifically limited to solid particles and standard particle monitoring instruments are fitted with a heater or dryer at the inlet to remove all liquid material from the sample before the concentrations are measured. This study shows that it is important to understand that the results provided by low-cost particle sensors, such as the PMS1003, cannot be used to ascertain if air quality standards are being met.
Article
Full-text available
A fast-growing area of research is the development of low-cost sensors for measuring air pollutants. The affordability and size of low-cost particle sensors makes them an attractive option for use in experiments requiring a number of instruments such as high-density spatial mapping. However, for these low-cost sensors to be useful for these types of studies their accuracy and precision need to be quantified. We evaluated the Alphasense OPC-N2, a promising low-cost miniature optical particle counter, for monitoring ambient airborne particles at typical urban background sites in the UK. The precision of the OPC-N2 was assessed by co-locating 14 instruments at a site to investigate the variation in measured concentrations. Comparison to two different reference optical particle counters as well as a TEOM-FDMS enabled the accuracy of the OPC-N2 to be evaluated. Comparison of the OPC-N2 to the reference optical instruments shows some limitations for measuring mass concentrations of PM1, PM2.5 and PM10. The OPC-N2 demonstrated a significant positive artefact in measured particle mass during times of high ambient RH (> 85 %) and a calibration factor was developed based upon κ-Köhler theory, using average bulk particle aerosol hygroscopicity. Application of this RH correction factor resulted in the OPC-N2 measurements being within 33 % of the TEOM-FDMS, comparable to the agreement between a reference optical particle counter and the TEOM-FDMS (20 %). Inter-unit precision for the 14 OPC-N2 sensors of 22 ± 13 % for PM10 mass concentrations was observed. Overall, the OPC-N2 was found to accurately measure ambient airborne particle mass concentration provided they are (i) correctly calibrated and (ii) corrected for ambient RH. The level of precision demonstrated between multiple OPC-N2s suggests that they would be suitable devices for applications where the spatial variability in particle concentration was to be determined.
Article
Full-text available
A domestic unmanned aerial vehicle (UAV) was built with the data acquisition capabilities. It is completely based on open-source projects which enables great level of customization. Two different data acquisition systems (DAS) were tested: a proprietary DAS and an in-house developed open platform DAS. Some measurements were presented and possible applications are discussed.