materials
Article
Miniaturized Near-Infrared (MicroNIR) Spectrometer
in Plastic Waste Sorting
Monika Rani 1, 2, †, Claudio Marchesi 1, 2, †, Stefania Federici 1, 2, * , Gianluca Rovelli 3,
Ivano Alessandri 2,4,5 , Irene Vassalini 2,4,5, Serena Ducoli 2,4, Laura Borgese 1,2,
Annalisa Zacco 1,2, Fabjola Bilo 1,2, Elza Bontempi 1,2 and Laura E. Depero 1,2
1Department of Mechanical and Industrial Engineering, University of Brescia, via Branze,
38-25123 Brescia, Italy
2Consorzio Interuniversitario Nazionale per la Scienza e Tecnologia dei Materiali (INSTM),
50121 Firenze, Italy
3ABCS Srl-Scientific Instruments & Materials Science, viale Regina Giovanna, 37-20129 Milan, Italy
4Department of Information Engineering, University of Brescia, via Branze, 38-25123 Brescia, Italy
5CNR-INO, Unit of Brescia, 25123 Brescia, Italy
*Correspondence: stefania.federici@unibs.it
†These authors are equally contributed.
Received: 24 July 2019; Accepted: 23 August 2019; Published: 27 August 2019
Abstract:
Valorisation of the urban plastic waste in high-quality recyclates is an imperative challenge
in the new paradigm of the circular economy. In this scenario, a key role in the improvement of
the recycling process is exerted by the optimization of waste sorting. In spite of the enormous
developments achieved in the field of automated sorting systems, the quest for the reduction
of cross-contamination of incompatible polymers as well as a rapid and punctual sorting of the
unmatched polymers has not been sufficiently developed. In this paper, we demonstrate that a
miniaturized handheld near-infrared (NIR) spectrometer can be used to successfully fingerprint and
classify different plastic polymers. The investigated urban plastic waste comprised polyethylene (PE),
polypropylene (PP), poly(vinyl chloride) (PVC), poly(ethylene terephthalate) (PET), and poly(styrene)
(PS), collected directly in a recycling plastic waste plant, without any kind of sample washing or
treatment. The application of unsupervised and supervised chemometric tools such as principal
component analysis (PCA) and partial least squares-discriminant analysis (PLS-DA) on the NIR
dataset resulted in a complete classification of the polymer classes. In addition, several kinds of PET
(clear, blue, coloured, opaque, and boxes) were correctly classified as PET class, and PE samples with
different branching degrees were properly separated.
Keywords:
urban plastic waste; near-infrared (NIR) spectroscopy; chemometrics; principal
component analysis (PCA); partial least squares-discriminant analysis (PLS-DA); circular economy
1. Introduction
The huge amount of urban plastic waste and the continuous growth in human plastic consumption
require a high valorisation of the collected waste in the direction of a whole-system economic
sustainability. By the numbers, European plastic production reached almost 60 million tonnes in 2016,
and 335 million tonnes globally (Source: PlasticsEurope Market Research Group (PEMRG)/Conversio
Market & Strategy GmbH). In this framework, over 8.4 million tonnes of plastic waste were collected
in order to be recycled inside and outside the EU [
1
]. The recycling of polymer waste has valuable
environmental benefits due to the substitution of primary production, and a key role in the improvement
of the recycling process is exerted by the optimization of waste sorting [2,3].
Materials 2019,12, 2740; doi:10.3390/ma12172740 www.mdpi.com/journal/materials
Materials 2019,12, 2740 2 of 13
Generally, automated sorting processes are widespread in plastic recycling facilities [
4
].
These systems are usually based on vibrational spectroscopy techniques for polymer identification [
5
–
9
]
and camera systems for optical recognition of clear and coloured products [
10
,
11
]. Other sorting
technologies include UV–Vis spectroscopy [
12
] or mass spectroscopy (Py-GC/MS) for identification [
13
],
hyper-spectral imaging methods [
14
], X-ray detection for the separation of specific containers [
15
],
fluorescence spectroscopy for identifying halogens and heavy metals [
16
], and laser-induced plasma
spectroscopy to detect additives [
17
]. This approach has increased the purity of the output plastic
over the years, reaching a high percentage of recyclates in the production of secondary materials.
Nevertheless, there are severe limitations of these techniques in handling mixed plastic, which requires
additional sorting elsewhere and can affect the quality of the recyclate, if not properly assigned.
The positive cost/benefit balance can take place only if the separated fractions of polymers match a
high purity grade, fulfilling the market requirement of high recyclate quality. Therefore, a key step in
post-consumer recycling is reducing the cross-contamination of incompatible polymers [10].
Besides the automated sorting of the plastic stream, manual sorting is still necessary for the full
separation of the collected waste. However, manual sorting based only on plastic code recognition is
inefficient and time consuming. In some cases, plastic goods can be badly crushed or have missing or
unreadable codes. Manufacturers can also use different polymers for the same product, depending on
marketing choices. In addition, the high turnover rate can be affected by an improperly trained or
unskilled workforce.
In this scenario, the possibility of exploiting a well-established technique for polymer identification
combined with a miniaturized, portable, low-cost, and real-time spectrometer for local and punctual
semi-automated sorting is highly desirable. Urban plastic waste normally contains plastics such as
polyethylene (PE), polypropylene (PP), poly(vinyl chloride) (PVC), poly(ethylene terephthalate) (PET),
and poly(styrene) (PS). These polymers have characteristic near-infrared (NIR) spectral fingerprints that
can be used for accurate sorting and separation [5]. The progress in miniaturization technologies has
generated specific instrumental features that have been reported in polymer identification applications,
but results have been reported only for a selection of polymers from commercial libraries, and are not
fully representative of real plastic waste [18].
In the present work, we challenged a compacted and miniaturized NIR spectrometer for a rapid
identification of urban plastic waste collected directly from a selection and recycling plant, without
any sample treatment. We collected a large amount of plastic samples directly from the plant and we
created a robust database of NIR spectra. Different chemometric tools were applied for data evaluation
of all the recorded samples. Principal component analysis (PCA) for preliminary data exploration and
partial least squares-discriminant analysis (PLS-DA) as supervised pattern recognition were applied to
fingerprint and classify the different plastic polymers.
2. Materials and Methods
2.1. Polymer Samples Collection
Plastic samples were collected at the selection division of the recovery and recycling plant Montello
SpA (Bergamo, Italy), which receives post-consumer plastic in the form of urban waste for recycling.
In total, 250 samples belonging to various classes of polymer were used (see Supplementary Materials
Figure S1). The polymers considered were: several kinds of PET (clear, blue, coloured, opaque,
and boxes), PE with different branching degrees (low-density polyethylene, LDPE, and high-density
polyethylene, HDPE), PVC, PP, and PS. The collection included bottles, containers, and packets of
different shapes, sizes, and colours. The samples were flattened or partially flattened at the sampling
site, and were not washed prior to testing. Black samples were not considered, due to the very low
reflectance in the NIR spectral region and the low signal-to-noise ratio of NIR sensors, which hamper
successful black polymer spectral acquisition [19].
Materials 2019,12, 2740 3 of 13
2.2. NIR Measurements
Spectra acquisition was realized using MicroNIR On-site (Viavi Solutions Inc., CA, United States
in reflectance mode without any sample preparation. The instrument is a miniaturized palm-sized
portable spectrometer of about 250 g in weight, less than 200 mm in length, and 50 mm in diameter.
This handheld spectrometer employs a linear variable filter (LVF) as dispersing element, in contrast
to traditional diffraction-based spectrometers (see Figure 1for an instrument operating scheme).
The controlling parameters for spectral data collection were set at 10 ms integration time and 50 scans,
resulting in a short measurement time of 0.25 s. Not less than five replicates in different positions were
acquired by a point-and-shoot technique, in order to minimize effects driven by the non-uniformity of
samples. A total of 1303 NIR spectra were collected. A dark and a reference scan were carried out
approximately every 10 min. Data acquisition was realized through MicroNIR
TM
Pro v3.0 software
(Viavi Solutions Inc., CA, United States).
Materials 2019, 12, x FOR PEER REVIEW 3 of 13
Spectra acquisition was realized using MicroNIR On-site (Viavi Solutions Inc., CA, United
States in reflectance mode without any sample preparation. The instrument is a miniaturized
palm-sized portable spectrometer of about 250 g in weight, less than 200 mm in length, and 50 mm in
diameter. This handheld spectrometer employs a linear variable filter (LVF) as dispersing element,
in contrast to traditional diffraction-based spectrometers (see Figure 1 for an instrument operating
scheme). The controlling parameters for spectral data collection were set at 10 ms integration time
and 50 scans, resulting in a short measurement time of 0.25 s. Not less than five replicates in different
positions were acquired by a point-and-shoot technique, in order to minimize effects driven by the
non-uniformity of samples. A total of 1303 NIR spectra were collected. A dark and a reference scan
were carried out approximately every 10 min. Data acquisition was realized through MicroNIRTM
Pro v3.0 software (Viavi Solutions Inc., CA, United States).
Figure 1. (a) Miniaturized near-infrared (microNIR) device and (b) instrument operating scheme.
2.3. Signal Pretreatment
Signal correction methods are mathematical tools employed to remove additive and/or
multiplicative effects on spectroscopic data, which often affect the performance of chemometric
analysis [20]. These phenomena are typical of light scattering effects, which induce a photon loss
(addictive effect) and an increased path length (multiplicative effect), among others. These methods
are “row-wise”, and therefore the correction is made sample-by-sample. This is in contrast to mean
centring and autoscaling, which are “column-wise” treatments. In most cases, the combination of the
two types of method is required in order to make a satisfactory data analysis [21]. In this study,
spectra were collected into a single 1307 × 125 matrix (samples × scans), and several pre-processing
tools were applied to the raw data. The best result was obtained by applying Savitzky-Golay
second-derivative procedure of seven data points and a second-order polynomial followed by
standard normal variate (SNV). Moreover, after the preprocessing, spectra were normalized by
mean centring.
2.4. Data Analysis
Different chemometric tools were applied in order to have a correct data evaluation of all the
analysed samples. The first step consisted of an exploratory analysis by PCA aiming to explore the
data structure. PCA was carried out with a total of 1303 NIR spectra for all five polymer classes.
Subsequently, PLS-DA was used as supervised pattern recognition with the aim of separating the
different commodities, and it was performed on a total of 1287 NIR spectra (16 spectra were
Figure 1. (a) Miniaturized near-infrared (microNIR) device and (b) instrument operating scheme.
2.3. Signal Pretreatment
Signal correction methods are mathematical tools employed to remove additive and/or
multiplicative effects on spectroscopic data, which often affect the performance of chemometric
analysis [
20
]. These phenomena are typical of light scattering effects, which induce a photon loss
(addictive effect) and an increased path length (multiplicative effect), among others. These methods
are “row-wise”, and therefore the correction is made sample-by-sample. This is in contrast to mean
centring and autoscaling, which are “column-wise” treatments. In most cases, the combination of the
two types of method is required in order to make a satisfactory data analysis [
21
]. In this study, spectra
were collected into a single 1307
×
125 matrix (samples
×
scans), and several pre-processing tools were
applied to the raw data. The best result was obtained by applying Savitzky-Golay second-derivative
procedure of seven data points and a second-order polynomial followed by standard normal variate
(SNV). Moreover, after the preprocessing, spectra were normalized by mean centring.
2.4. Data Analysis
Different chemometric tools were applied in order to have a correct data evaluation of all the
analysed samples. The first step consisted of an exploratory analysis by PCA aiming to explore the
data structure. PCA was carried out with a total of 1303 NIR spectra for all five polymer classes.
Subsequently, PLS-DA was used as supervised pattern recognition with the aim of separating the
Materials 2019,12, 2740 4 of 13
different commodities, and it was performed on a total of 1287 NIR spectra (16 spectra were removed as
outliers). All calculations were performed using MATLAB (R2018a) with chemometric toolboxes [
22
,
23
].
3. Results and Discussion
NIR spectroscopy has the key advantage of being a rapid-response analytical tool, recording
spectra with no prior manipulation and predicting physical and chemical parameters from a single
spectrum [
24
]. In the NIR region, the absorption bands occur due to overtones or combination bands
of mainly carbon–hydrogen vibrations and oxygen-hydrogen vibrations. A correct band assignment is
challenging because it may be due to different combinations of fundamental vibrations. Moreover,
overtone vibrations are highly overlapped [25].
The NIR reflectance spectra of the five polymers (i.e., PET, PP, PVC, PE, and PS) in the range
900–1700 nm, obtained by the MicroNIR On-site, are shown in Figure 2.
Materials 2019, 12, x FOR PEER REVIEW 4 of 13
removed as outliers). All calculations were performed using MATLAB (R2018a) with chemometric
toolboxes [22,23].
3. Results and Discussion
NIR spectroscopy has the key advantage of being a rapid-response analytical tool, recording
spectra with no prior manipulation and predicting physical and chemical parameters from a single
spectrum [24]. In the NIR region, the absorption bands occur due to overtones or combination bands
of mainly carbon–hydrogen vibrations and oxygen-hydrogen vibrations. A correct band assignment
is challenging because it may be due to different combinations of fundamental vibrations. Moreover,
overtone vibrations are highly overlapped [25].
The NIR reflectance spectra of the five polymers (i.e., PET, PP, PVC, PE, and PS) in the range
900–1700 nm, obtained by the MicroNIR On-site, are shown in Figure 2.
Figure 2. Near-infrared (NIR) spectra of five classes of plastics: (a) representative raw spectra of the
five classes; (b) corresponding spectra after pre-treatment by second derivative and standard normal
variate (SNV). PE: polyethylene; PET: poly(ethylene terephthalate); PP: polypropylene; PS:
poly(styrene); PVC: poly(vinyl chloride).
In the absorbance spectra (Figure 2a), the principal absorbance band for PET was found at 1660
nm due to the first overtone of C–H stretching [26], with two other small peaks around 1130 nm and
1415 nm. In the case of PVC, this main peak was shifted and was not recorded in the instrumental
operating window, and only the two small peaks around 1190 nm and 1420 nm were seen [26]. The
same behaviour appeared for PE and PP, assigning two bands around 1210 nm and 1430 nm for PE
and around 1190 nm and 1400 nm for PP. PS exhibited a dominant band around 1675 nm and two
small peaks around 1140 nm and 1205 nm [5]. Figure 2b refers to the corresponding spectra shown in
Figure 2a after pre-treatment by second derivative and SNV. In general, by comparing the spectra of
samples made of one type of polymer, a variation in the absorbance values was noted and this was
ascribed to variations in sample thickness [5]. However, the overall form of the spectra was
preserved, concluding that dust or liquid contaminations randomly characterizing each sample did
not affect the quality of NIR spectra (see Appendix A, Figure A1) [27]. In addition, NIR spectra of
samples of different colours did not display considerable differences (Figure A1), meaning that the
colorants present in the plastic resin matrix did not significantly influence the NIR spectra.
Moreover, no differences in band positions were found when studying the spectra of several kinds
of PET (clear, coloured, opaque, box, and blue) (see Appendix A, Figure A2), implying that the
spectra can be successfully exploited to discriminate PET samples, despite their differences. For this
reason, hereinafter, only one PET class is considered.
3.1. Principal Component Analysis
Principal component analysis is a very powerful chemometric tool for analysing data structure
[28]. The aim of PCA is to extract the information encoded in a certain number of variables into a
Figure 2.
Near-infrared (NIR) spectra of five classes of plastics: (
a
) representative raw spectra of the five
classes; (
b
) corresponding spectra after pre-treatment by second derivative and standard normal variate
(SNV). PE: polyethylene; PET: poly(ethylene terephthalate); PP: polypropylene; PS: poly(styrene); PVC:
poly(vinyl chloride).
In the absorbance spectra (Figure 2a), the principal absorbance band for PET was found at 1660 nm
due to the first overtone of C–H stretching [
26
], with two other small peaks around 1130 nm and
1415 nm. In the case of PVC, this main peak was shifted and was not recorded in the instrumental
operating window, and only the two small peaks around 1190 nm and 1420 nm were seen [
26
].
The same behaviour appeared for PE and PP, assigning two bands around 1210 nm and 1430 nm for PE
and around 1190 nm and 1400 nm for PP. PS exhibited a dominant band around 1675 nm and two
small peaks around 1140 nm and 1205 nm [
5
]. Figure 2b refers to the corresponding spectra shown in
Figure 2a after pre-treatment by second derivative and SNV. In general, by comparing the spectra of
samples made of one type of polymer, a variation in the absorbance values was noted and this was
ascribed to variations in sample thickness [
5
]. However, the overall form of the spectra was preserved,
concluding that dust or liquid contaminations randomly characterizing each sample did not affect the
quality of NIR spectra (see Appendix A, Figure A1) [
27
]. In addition, NIR spectra of samples of different
colours did not display considerable differences (Figure A1), meaning that the colorants present in
the plastic resin matrix did not significantly influence the NIR spectra. Moreover, no differences in
band positions were found when studying the spectra of several kinds of PET (clear, coloured, opaque,
box, and blue) (see Appendix A, Figure A2), implying that the spectra can be successfully exploited to
discriminate PET samples, despite their differences. For this reason, hereinafter, only one PET class
is considered.
Materials 2019,12, 2740 5 of 13
3.1. Principal Component Analysis
Principal component analysis is a very powerful chemometric tool for analysing data structure [
28
].
The aim of PCA is to extract the information encoded in a certain number of variables into a smaller
set of new orthogonal variables called principal components. PCA calculation was performed over
the entire spectral region and using all collected spectra. The first three components accounted for
87.11% of the total variability (PC1 56.75%, PC2 15.63%, and PC3 14.73%). Figure 3displays the score
plots of PC1 vs. PC2 (Figure 3a) and PC1 vs. PC3 (Figure 3b). In the score plot displayed in Figure 3a,
a clear separation between the five commodities can be seen. Indeed, in the new space determined
by the first two components, polymers formed very tight and homogeneous clusters. Along the first
component, a clear differentiation between PET, PVC, PP, and PE was highlighted, while the second
component allowed the differentiation of PS from the other polymers, and PE from PP. On the other
hand, Figure 3b shows the score plot of PC1 vs. PC3. In this case PET was differentiated from the other
polymers along PC1, while PC3 distinguished between the remaining sets of polymers.
Materials 2019, 12, x FOR PEER REVIEW 5 of 13
smaller set of new orthogonal variables called principal components. PCA calculation was
performed over the entire spectral region and using all collected spectra. The first three components
accounted for 87.11% of the total variability (PC1 56.75%, PC2 15.63%, and PC3 14.73%). Figure 3
displays the score plots of PC1 vs. PC2 (Figure 3a) and PC1 vs. PC3 (Figure 3b). In the score plot
displayed in Figure 3a, a clear separation between the five commodities can be seen. Indeed, in the
new space determined by the first two components, polymers formed very tight and homogeneous
clusters. Along the first component, a clear differentiation between PET, PVC, PP, and PE was
highlighted, while the second component allowed the differentiation of PS from the other polymers,
and PE from PP. On the other hand, Figure 3b shows the score plot of PC1 vs. PC3. In this case PET
was differentiated from the other polymers along PC1, while PC3 distinguished between the
remaining sets of polymers.
Figure 3. Results of principal components analysis carried out with spectral data of the different
commodities. (a) The score plot of the first two components is shown, as well as (b) the score plot of
PC1 vs. PC3 EV: explained variance.
Figure 3.
Results of principal components analysis carried out with spectral data of the different
commodities. (
a
) The score plot of the first two components is shown, as well as (
b
) the score plot of
PC1 vs. PC3 EV: explained variance.
Materials 2019,12, 2740 6 of 13
Furthermore, the results of PCA applied to the restricted dataset of PET samples showed no
separation among the five kinds of PET samples (Figure 4), confirming the interpretation of the band
assignment of NIR spectra.
Materials 2019, 12, x FOR PEER REVIEW 6 of 13
Furthermore, the results of PCA applied to the restricted dataset of PET samples showed no
separation among the five kinds of PET samples (Figure 4), confirming the interpretation of the band
assignment of NIR spectra.
Figure 4. Principal component analysis (PCA) analysis of several kinds of poly(ethylene
terephthalate) (PET) samples for a total of 659 spectra considered. EV: explained variance.
3.2. HDPE versus LDPE
The discrimination between HDPE and LDPE is an important issue in terms of plastic recycling
[3]. PE is largely used in the food packaging sector, thanks to its remarkable properties in terms of
mechanical and optical performances, water-vapor-resistance, and heat-sealing attitude. The
amount of side branches that are attached to the main PE polymer chain determines the
morphological properties, which in turn affects optical, physical, and thermal properties [29,30]. NIR
spectroscopy successfully discriminates between HDPE and LDPE [31,32], even if severe limitations
can occur depending on the characteristics of the measuring instruments [27]. In the sampling
campaign, we collected 25 samples of HDPE and 25 samples of LDPE. Raw NIR spectra did not
show any significant feature to discriminate the two plastics, nor did a clear separation in the score
plots resulting from the previous PCA analysis (Figure 3). However, by limiting the PCA analysis to
the PE samples, a separation between HDPE and LDPE was obtained; the first two components
explicated 61.22% of the total variability (PC1 43.22%, PC2 18.00%), confirming the successful
application of NIR spectroscopy in discriminating these two polymers. Figure 5 displays the score
plots of PC1 vs. PC2.
Figure 4.
Principal component analysis (PCA) analysis of several kinds of poly(ethylene terephthalate)
(PET) samples for a total of 659 spectra considered. EV: explained variance.
3.2. HDPE versus LDPE
The discrimination between HDPE and LDPE is an important issue in terms of plastic recycling [
3
].
PE is largely used in the food packaging sector, thanks to its remarkable properties in terms of
mechanical and optical performances, water-vapor-resistance, and heat-sealing attitude. The amount of
side branches that are attached to the main PE polymer chain determines the morphological properties,
which in turn affects optical, physical, and thermal properties [
29
,
30
]. NIR spectroscopy successfully
discriminates between HDPE and LDPE [
31
,
32
], even if severe limitations can occur depending on the
characteristics of the measuring instruments [
27
]. In the sampling campaign, we collected 25 samples of
HDPE and 25 samples of LDPE. Raw NIR spectra did not show any significant feature to discriminate
the two plastics, nor did a clear separation in the score plots resulting from the previous PCA analysis
(Figure 3). However, by limiting the PCA analysis to the PE samples, a separation between HDPE and
LDPE was obtained; the first two components explicated 61.22% of the total variability (PC1 43.22%,
PC2 18.00%), confirming the successful application of NIR spectroscopy in discriminating these two
polymers. Figure 5displays the score plots of PC1 vs. PC2.
Materials 2019,12, 2740 7 of 13
Materials 2019, 12, x FOR PEER REVIEW 7 of 13
Figure 5. Results of principal components analysis carried out with spectral data of polyethylene (PE)
class polymers. Score plot of the first two components is shown.
3.3. Partial Least Squares-Discriminant Analysis
After the exploratory PCA analysis, a supervised classification tool was applied in order to
distinguish the different plastic classes. In PLS-DA, the well-known PLS regression algorithm is
modified with a classification goal. The response variable is categorical, reflecting the belonging
class of the statistical units. PLS-DA returns the prediction in a vector of size equal to the number of
classes in the predictor variables, with values ranging from 0 to 1 [33].
Validation tools are very useful methods in chemometrics, and they are used to verify the
capability of the model prediction. Before the model calibration, data were split into training and test
sets containing 901 and 386 samples, respectively. The training set was used to calibrate the model
both in fitting and in cross-validation, while the test set was employed only at the end of the
procedure in order to evaluate the true predictive capability of the calibrated model [22].
During the calibration, the selection of the optimal number of latent variables (LVs) is a crucial
point, which is performed based on the cross-validation procedure [22]. This should allow the
optimization of the complexity of the multivariate model according to the predictive capacity of the
model itself. Cross-validation is generally carried out by dividing the training set into different
cross-validation groups, and during each round one group is removed from the training set. Over
the rounds, the model is calibrated on the remaining training samples and then used to predict
samples of the cross-validation group. The cross-validation procedure was based on a Venetian
blinds approach with 10 groups. In the Venetian blinds procedure, the calibration group and the
cross-validation group are selected by choosing every n-th sample from the dataset starting from the
first one [22]. The selection of the number of cross-validation groups was set to 10, as a reasonable
number in order to avoid the so-called “overfitting” (i.e., the overestimation of the predictive
capability of the model). Figure 6 shows the error rate (Figure 6a) and the degree of not assigned
samples as a function of the number of latent variables (Figure 6b). The optimal number of LVs was
set to 6 (explaining 94% of the total variability and 0.0099% of not assigned samples), which was
associated with the minimum error rate and, simultaneously, with the minimum percentage of not
assigned samples. Four LVs reached the same error rate, but 0.019% of not assigned samples.
Figure 5.
Results of principal components analysis carried out with spectral data of polyethylene (PE)
class polymers. Score plot of the first two components is shown.
3.3. Partial Least Squares-Discriminant Analysis
After the exploratory PCA analysis, a supervised classification tool was applied in order to
distinguish the different plastic classes. In PLS-DA, the well-known PLS regression algorithm is
modified with a classification goal. The response variable is categorical, reflecting the belonging class
of the statistical units. PLS-DA returns the prediction in a vector of size equal to the number of classes
in the predictor variables, with values ranging from 0 to 1 [33].
Validation tools are very useful methods in chemometrics, and they are used to verify the capability
of the model prediction. Before the model calibration, data were split into training and test sets
containing 901 and 386 samples, respectively. The training set was used to calibrate the model both in
fitting and in cross-validation, while the test set was employed only at the end of the procedure in
order to evaluate the true predictive capability of the calibrated model [22].
During the calibration, the selection of the optimal number of latent variables (LVs) is a crucial
point, which is performed based on the cross-validation procedure [
22
]. This should allow the
optimization of the complexity of the multivariate model according to the predictive capacity of
the model itself. Cross-validation is generally carried out by dividing the training set into different
cross-validation groups, and during each round one group is removed from the training set. Over the
rounds, the model is calibrated on the remaining training samples and then used to predict samples of
the cross-validation group. The cross-validation procedure was based on a Venetian blinds approach
with 10 groups. In the Venetian blinds procedure, the calibration group and the cross-validation
group are selected by choosing every n-th sample from the dataset starting from the first one [
22
].
The selection of the number of cross-validation groups was set to 10, as a reasonable number in order
to avoid the so-called “overfitting” (i.e., the overestimation of the predictive capability of the model).
Figure 6shows the error rate (Figure 6a) and the degree of not assigned samples as a function of the
number of latent variables (Figure 6b). The optimal number of LVs was set to 6 (explaining 94% of the
total variability and 0.0099% of not assigned samples), which was associated with the minimum error
rate and, simultaneously, with the minimum percentage of not assigned samples. Four LVs reached
the same error rate, but 0.019% of not assigned samples.
Materials 2019,12, 2740 8 of 13
Materials 2019, 12, x FOR PEER REVIEW 8 of 13
Figure 6. Error rate (a) and not assigned samples (b) as a function of latent variables calculated in the
partial least squares-discriminant analysis (PLS-DA) model. Six latent variables (LVs) was the
optimal number, marked in red, with 0.0099% of not assigned samples.
After the determination of the optimal number of LVs, the calculation of the PLS-DA model was
performed selecting six latent variables and 10 cross-validation groups for internal validation. In
Table 1, the two confusion matrices achieved in fitting and in cross-validation were presented. Both
in fitting and in cross-validation, the performance of the model was very high, proved by the
correlation between observed and predicted classes. All the investigated samples were correctly
designated, and only four samples in fitting and five in cross-validation were not assigned.
Table 1. Confusion matrices obtained from the PLS-DA model, both in fitting and in cross-validation
(based on Venetian blinds with 10 groups). The “Not Assigned” column contains samples which
were not assigned to any of the considered classes.
Experimental
Class
Calculated Class
PET PE PP PVC PS Not
Assigned
Fitting - - - - - -
PET 459 0 0 0 0 2
PE 0 181 0 0 0 1
PP 0 0 82 0 0 0
PVC 0 0 0 84 0 1
PS 0 0 0 0 91 0
Cross-Validation - - - - - -
PET 460 0 0 0 0 1
PE 0 181 0 0 0 1
PP 0 0 82 0 0 0
PVC 0 0 0 82 0 3
PS 0 0 0 0 91 0
The performance of the classification model could be also evaluated by considering the
classification parameters derived from the confusion matrix: sensitivity (Sn), specificity (Sp), and
Figure 6.
Error rate (
a
) and not assigned samples (
b
) as a function of latent variables calculated in the
partial least squares-discriminant analysis (PLS-DA) model. Six latent variables (LVs) was the optimal
number, marked in red, with 0.0099% of not assigned samples.
After the determination of the optimal number of LVs, the calculation of the PLS-DA model
was performed selecting six latent variables and 10 cross-validation groups for internal validation.
In Table 1, the two confusion matrices achieved in fitting and in cross-validation were presented.
Both in fitting and in cross-validation, the performance of the model was very high, proved by the
correlation between observed and predicted classes. All the investigated samples were correctly
designated, and only four samples in fitting and five in cross-validation were not assigned.
Table 1.
Confusion matrices obtained from the PLS-DA model, both in fitting and in cross-validation
(based on Venetian blinds with 10 groups). The “Not Assigned” column contains samples which were
not assigned to any of the considered classes.
Experimental Class Calculated Class
PET PE PP PVC PS Not Assigned
Fitting - - - - - -
PET 459 0 0 0 0 2
PE 0 181 0 0 0 1
PP 0 0 82 0 0 0
PVC 0 0 0 84 0 1
PS 0 0 0 0
91
0
Cross-Validation - - - - - -
PET 460 0 0 0 0 1
PE 0 181 0 0 0 1
PP 0 0 82 0 0 0
PVC 0 0 0 82 0 3
PS 0 0 0 0
91
0
The performance of the classification model could be also evaluated by considering the classification
parameters derived from the confusion matrix: sensitivity (Sn), specificity (Sp), and non-error rate
(NER). These parameters represent the ability of the model to correctly identify the samples of the
Materials 2019,12, 2740 9 of 13
given class, the ability of a classifier to reject the samples of other classes, and the average of the class
sensitivities, respectively [
34
]. Table 2collects the parameters of the PLS-DA model. In both fitting
and cross-validation, NER, Sn, and Sp were equal to 1, meaning that 100% of the polymers were
correctly classified.
Table 2.
Classification parameters (non-error rate (NER), class sensitivity (Sn), and specificity (Sp))
calculated in fitting and in cross-validation.
-- PET PE PP PVC PS
NER Sn Sp Sn Sp Sn Sp Sn Sp Sn Sp
Fitting 1 1 1 1 1 1 1 1 1 1 1
Cross-validation 1 1 1 1 1 1 1 1 1 1 1
Furthermore, we analysed the confusion matrix and the classification parameters for the test set,
used as an additional trial to validate the model. Table 3shows the confusion matrix derived from
the test set; again, 100% of the samples were classified, and only four samples were not assigned.
Table 4summarizes the classification parameters obtained from the confusion matrix, with the NER,
Sn, and Sp of each class equal to 1.
Table 3.
Confusion matrix obtained from the PLS-DA model fitted on the validation set. The “Not
Assigned” column contains samples which were not assigned to any of the considered classes.
Experimental Class Calculated Class
PET PE PP PVC PS Not Assigned
PET 197 0 0 0 0 1
PE 0 78 0 0 0 1
PP 0 0 33 0 0 1
PVC 0 0 0 35 0 1
PS 0 0 0 0
39
0
Table 4.
Classification parameters (non-error rate (NER), class sensitivity (Sn), and specificity (Sp))
calculated on the validation set.
-- PET PE PP PVC PS
NER Sn Sp Sn Sp Sn Sp Sn Sp Sn Sp
Test 1 1 1 1 1 1 1 1 1 1 1
In the case of the restricted data set composed of HDPE and LDPE spectra, the optimal number
of LVs was set to six, assigning 90% of total variability (see Supplementary Materials, Figure S2).
With respect to the analysis of the polymer classes, the degree of not assigned samples as a function
of the number of latent variables was equal to zero, meaning that all the samples were assigned to a
class. After the determination of the optimal number of LVs, the calculation of the PLS-DA model
was performed, selecting six latent variables and 10 cross-validation groups for internal validation.
Both in fitting and in cross-validation, all the investigated samples were correctly designated (see
Table S1 in the Supplementary Materials for the two confusion matrices achieved in fitting and in
cross-validation).
4. Conclusions
In the present study, the use of a handheld NIR spectrometer combined with robust chemometrics
tools to fingerprint different urban plastic waste collected directly from a recycling plant was presented.
Plastic samples included clear, blue, coloured, opaque, and boxes of PET, PE with different branching
Materials 2019,12, 2740 10 of 13
degrees (HDPE and LDPE), PP, PVC, and PS. In a first step, PCA was applied to NIR spectra of all
polymer classes, and polymers were correctly separated in clusters. Moreover, by limiting the analysis
on PE samples, HDPE and LDPE were correctly discriminated.
In addition, by comparing samples of different colours, no differences in band positions were
found, suggesting that colorants present in the plastic resin matrix do not significantly influence NIR
spectra. Moreover, even if samples were not washed or treated prior to the analysis, NIR spectra were
not affected by dust contamination or liquid residuals, meaning a successful application directly to the
urban plastic waste.
After the exploratory PCA analysis, PLS-DA was applied. Before the model calibration, data were
split into a training and test set, using the training set to calibrate the model. In both fitting and
cross-validation, the performance of the model was very high, proved by the correlation between
observed and predicted classes. All the analysed samples were also correctly designated in the case of
the restricted data set composed of HDPE and LDPE spectra.
The results show promising outcomes in order to improve the reliability and efficiency of the
manual sorting of plastic waste, increasing the volume of recycled plastic and the management of
multi-component polymer types.
Supplementary Materials: The following are available online at http://www.mdpi.com/1996-1944/12/17/2740/s1.
Figure S1: Systematic organization of the collected urban plastic waste for laboratory measurements, Figure S2:
Error rate as a function of latent variables calculated in the PLS-DA model for HDPE and LDPE classification. Six
LVs was the optimal number, Table S1: Confusion matrices obtained from PLS-DA model, both in fitting and in
cross-validation (based on Venetian blinds with 10 groups) restricted to the PE class.
Author Contributions:
Conceptualization, S.F. and L.E.D.; Funding acquisition, G.R., E.B., and L.E.D.;
Investigation, M.R., C.M., I.V., S.D., A.Z., and F.B.; Resources, E.B. and L.E.D.; Supervision, S.F. and L.E.D.;
Writing—original draft, M.R., C.M., and S.F.; Writing—review and editing, S.F., I.A., L.B., and L.E.D.
Funding:
This research was funded by PON “R&I” 2014-2020: SIRIMAP—SIstemi di Rilevamento
dell’Inquinamento MArino da Plastiche e successivo recupero-riciclo (No. ARS01_01183) CUP D86C18000520008.
Conflicts of Interest: The authors declare no conflicts of interest.
Appendix A
Appendix A.1 Evaluation of Sample Contaminations and Colour Influence
Figure A1 shows a representative collection of NIR spectra of coloured PET samples. The coloured
PET batch was chosen as a demonstrator set to compare samples with random contaminations and
to evaluate the presence of colorants. A total of 136 spectra were collected (25 samples with at least
5 replicates in different positions for each sample). The samples obtained from the recycling plant
were flattened or partially flattened directly on site and were not washed prior the testing. Liquid
residuals, pieces of labels, and environmental dust were largely present on the surface and inside the
samples. Nevertheless, NIR spectra showed only a variation in the total absorbance values, attributable
to variations in sample thickness, according to Lambert–Beer law [
35
]. Moreover, no differences in
band positions were found despite the different colours of the commodities.
Materials 2019,12, 2740 11 of 13
Materials 2019, 12, x FOR PEER REVIEW 11 of 13
Figure A1. Representative NIR spectra of coloured PET samples.
Appendix A.2. PET Samples
Several kinds of PET were considered. In particular, clear, coloured, opaque, blue, and box
samples of PET were collected at the recycling plant (25 samples for each kind). In order to
investigate the NIR contribution of the different kinds, we acquired at least five replicates for each
sample (for a total of 659 spectra) to make spectra comparison and multivariate analysis. Figure A2
shows representative NIR spectra of a restriction of two random samples for each kind, in order to
better visualize the spectra in the plot. The overall form of the spectra was preserved and no
significant differences are noted.
Figure A2. Representative NIR spectra of two random samples for each kind of PET commodity.
References
1. Plastics Europe. Available online: http://www.plasticseurope.org (accessed on 24 August 2019).
2. Al-Salem, S.M.; Lettieri, P.; Baeyens, J. Recycling and recovery routes of plastic solid waste (PSW): A
review. Waste Manag. 2009, 29, 2625–2643.
Figure A1. Representative NIR spectra of coloured PET samples.
Appendix A.2 PET Samples
Several kinds of PET were considered. In particular, clear, coloured, opaque, blue, and box samples
of PET were collected at the recycling plant (25 samples for each kind). In order to investigate the NIR
contribution of the different kinds, we acquired at least five replicates for each sample (for a total of 659
spectra) to make spectra comparison and multivariate analysis. Figure A2 shows representative NIR
spectra of a restriction of two random samples for each kind, in order to better visualize the spectra in
the plot. The overall form of the spectra was preserved and no significant differences are noted.
Materials 2019, 12, x FOR PEER REVIEW 11 of 13
Figure A1. Representative NIR spectra of coloured PET samples.
Appendix A.2. PET Samples
Several kinds of PET were considered. In particular, clear, coloured, opaque, blue, and box
samples of PET were collected at the recycling plant (25 samples for each kind). In order to
investigate the NIR contribution of the different kinds, we acquired at least five replicates for each
sample (for a total of 659 spectra) to make spectra comparison and multivariate analysis. Figure A2
shows representative NIR spectra of a restriction of two random samples for each kind, in order to
better visualize the spectra in the plot. The overall form of the spectra was preserved and no
significant differences are noted.
Figure A2. Representative NIR spectra of two random samples for each kind of PET commodity.
References
1. Plastics Europe. Available online: http://www.plasticseurope.org (accessed on 24 August 2019).
2. Al-Salem, S.M.; Lettieri, P.; Baeyens, J. Recycling and recovery routes of plastic solid waste (PSW): A
review. Waste Manag. 2009, 29, 2625–2643.
Figure A2. Representative NIR spectra of two random samples for each kind of PET commodity.
Materials 2019,12, 2740 12 of 13
References
1. Plastics Europe. Available online: http://www.plasticseurope.org (accessed on 19 June 2019).
2.
Al-Salem, S.M.; Lettieri, P.; Baeyens, J. Recycling and recovery routes of plastic solid waste (PSW): A review.
Waste Manag. 2009,29, 2625–2643. [CrossRef] [PubMed]
3.
Siddique, R.; Khatib, J.; Kaur, I. Use of recycled plastic in concrete: A review. Waste Manag.
2008
,28,
1835–1852. [CrossRef] [PubMed]
4.
Gundupalli, S.P.; Hait, S.; Thakur, A. A review on automated sorting of source-separated municipal solid
waste for recycling. Waste Manag. 2017,60, 56–74. [CrossRef] [PubMed]
5.
Masoumi, H.; Safavi, S.M.; Khani, Z. Identification and classification of plastic resins using near infrared
reflectance spectroscopy. Int. J. Mech. Mechatron. Eng. 2012,6, 877–884.
6.
Masson, J.; Pelletier, L.; Collins, P. Rapid FTIR method for quantification of styrene-butadiene type copolymers
in bitumen. J. Appl. Polym. Sci. 2001,79, 1034–1041. [CrossRef]
7.
Allen, V.; Kalivas, J.H.; Rodriguez, R.G. Post-consumer plastic identification using raman spectroscopy. Appl.
Spectrosc. 1999,53, 672–681. [CrossRef]
8.
Chalmers, J.M.; Everall, N.J.; Measurement, I.C.I. Handbook of Vibrational Spectroscopy: Qualitative and
Quantitative Analysis of Plastics, Polymers and Rubbers by Vibrational Spectroscopy; John Wiley & Sons, Ltd.:
Hoboken, NJ, USA, 2007; ISBN 9780470027325.
9.
Florestan, J.; Lachambre, A.; Mermilliod, N.; Boulou, J.C.; Marfisi, C. Recycling of plastics: Automatic
identification of polymers by spectroscopic methods. Resour. Conserv. Recycl. 1994,10, 67–74. [CrossRef]
10.
Hopewell, J.; Dvorak, R.; Kosior, E. Plastics recycling: Challenges and opportunities. Philos. Trans. R. Soc. B
Biol. Sci. 2009,364, 2115–2126. [CrossRef]
11.
Tachwali, Y.; Al-Assaf, Y.; Al-Ali, A.R. Automatic multistage classification system for plastic bottles recycling.
Resour. Conserv. Recycl. 2007,52, 266–285. [CrossRef]
12.
Safavi, S.M.; Masoumi, H.; Mirian, S.S.; Tabrizchi, M. Sorting of polypropylene resins by color in MSW using
visible reflectance spectroscopy. Waste Manag. 2010,30, 2216–2222. [CrossRef]
13.
Herrera, M.; Matuschek, G.; Kettrup, A. Fast identification of polymer additives by pyrolysis-gas
chromatography/mass spectrometry. J. Anal. Appl. Pyrol. 2003,70, 35–42. [CrossRef]
14.
Ulrici, A.; Serranti, S.; Ferrari, C.; Cesare, D.; Foca, G.; Bonifazi, G. Efficient chemometric strategies for
PET-PLA discrimination in recycling plants using hyperspectral imaging. Chemometr. Intell. Lab.
2013
,122,
31–39. [CrossRef]
15.
Brunner, S.; Fomin, P.; Kargel, C. Automated sorting of polymer flakes: Fluorescence labeling and development
of a measurement system prototype. Waste Manag. 2015,38, 49–60. [CrossRef]
16.
Ernst, T.; Popp, R.; Van Eldik, R. Quantification of heavy metals for the recycling of waste plastics from
electrotechnical applications. Talanta 2000,53, 347–357. [CrossRef]
17.
Anzano, J.M.; Gornushkin, I.B.; Smith, B.W.; Winefofdner, J.D. Laser-Induced Plasma Spectroscopy for Plastic
Identification. Polym. Eng. Sci. 2000,40, 2423–2429. [CrossRef]
18.
Yan, H.; Siesler, H.W. Identification performance of different types of handheld near-infrared (NIR)
spectrometers for the recycling of polymer commodities. Appl. Spectrosc. 2018,72, 1362–1370. [CrossRef]
19.
Rozenstein, O.; Puckrin, E.; Adamowski, J. Development of a new approach based on midwave infrared
spectroscopy for post-consumer black plastic waste sorting in the recycling industry. Waste Manag.
2017
,68,
38–44. [CrossRef]
20.
Jansen, J.J.; Downey,G.; Blanchet, L.; Engel, J.; Gerretzen, J.; Szyman, E. Breaking with trends in pre-processing?
Trend. Anal. Chem. 2013,50, 96–106.
21.
Rinnan, Å.; van den Berg, F.; Engelsen, S.B. Review of the most common pre-processing techniques for
near-infrared spectra. TrAC-Trend. Anal. Chem. 2009,28, 1201–1222. [CrossRef]
22.
Ballabio, D.; Consonni, V. Classification tools in chemistry. Part 1: Linear models. PLS-DA. Anal. Methods
2013,5, 3790. [CrossRef]
23.
Ballabio, D. A Matlab toolbox for principal component analysis and unsupervised exploration of data
structure. Chemometr. Intell. Lab. 2015,149, 1–9. [CrossRef]
24.
Blanco, M.; Villarroya, I. NIR spectroscopy: A rapid-response analytical tool. TrAC-Trend. Anal. Chem.
2002
,
21, 240–250. [CrossRef]
Materials 2019,12, 2740 13 of 13
25.
Weyer, L.; Workman, J., Jr. Practical Guide and Spectral Atlas for Interpretive Near-Infrared Spectroscopy;
Workman, J., Jr., Weyer, L., Eds.; CRC Press: Boca Raton, FL, USA, 2012.
26.
Crandall, E.W.; Jagtap, A.N. The near-infrared spectra of polymers. J. Appl. Polym. Sci.
1977
,21, 449–454.
[CrossRef]
27.
Feldhoff, R.; Wienke, D.; Cammann, K.; Fuchs, H. On-line post consumer package identification by NIR
spectroscopy combined with a FuzzyARTMAP classifier in an industrial environment. Appl. Spectrosc.
1997
,
51, 362–368. [CrossRef]
28. Bro, R.; Smilde, A.K. Principal component analysis. Anal. Methods 2014,6, 2812–2831. [CrossRef]
29.
Martinez-Salazar, J.; S
á
nchez Cuesta, M.; Balt
á
Calleja, F.J. Equilibrium and non-equilibrium melting of
branched polyethylene relating to defect incorporation within crystals. Colloid Polym. Sci.
1987
,265, 239–245.
[CrossRef]
30.
Mart
í
nez-Salazar, J.; Cuesta, M.S.; Plans, J. On phase separation in high- and low-density polyethylene
blends: 1. Melting-point depression analysis. Polymer (Guildf) 1991,32, 2984–2988. [CrossRef]
31.
Miller, C.E. Use of near-infrared spectroscopy to determine the composition of high-density/low-density
polyethylene blend films. Appl. Spectrosc. 1993,47, 222–228. [CrossRef]
32.
De Biasio, M.; Arnold, T.; McGunnigle, G.; Leitner, R.; Balthasar, D.; Rehrmann, V. Detecting and
discriminating PE and PP polymers for plastics recycling using NIR imaging spectroscopy. In Proceedings of
the Thermosense XXXII, Orlando, FL, USA, 3 May 2010; Volume 7661, p. 76610V.
33. Barker, M.; Rayens, W. Partial least squares for discrimination. J. Chemom. 2003,17, 166–173. [CrossRef]
34.
Ballabio, D.; Grisoni, F.; Todeschini, R. Multivariate comparison of classification performance measures.
Chemometr. Intell. Lab. 2018,174, 33–44. [CrossRef]
35.
Willard, H.H.; Merritt, L.L.J.; Dean, J.A.; Settle, F.A.J. Instrumental Methods of Analysis, 7th ed.; Wadsworth
Pub Co.: Belmont, CA, USA, 1988.
©
2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access
article distributed under the terms and conditions of the Creative Commons Attribution
(CC BY) license (http://creativecommons.org/licenses/by/4.0/).