ThesisPDF Available

Evaluation of Light Curing Units and Dentists’ Knowledge About Photo Polymerization Techniques in Sulaimani Governmental Dental Clinics

Authors:

Abstract

Abstract Objective: To measure the output intensity of light curing units (LCU) in governmental dental clinics of the Sulaimani governorate. To evaluate practitioners’ knowledge on light cure application. Methods: Ninety-four dental units and eighty-eight dentists in the Sulaimani governorate were included in this research. Output intensity and diameter of tips of the dental LCU devices were measured. Two survey questionnaires were filled, one for the LCU and its maintenance, the other for evaluating the knowledge, attitude, and practice of the dentists. Results: Output power intensity results ranged from 50-2000W, with an average of 993W. The output intensity of 78% of the tested devices was found to be acceptable regarding the manufacturer instruction of the radiometers used (every intensity below the given range that given in the Result section is regarded as unusable or non-acceptable). The range of age in service of the devices was 1-12 years with a mean of 4.7 years. The average restorations performed in a week by a device was 17.8. Dentists had poor knowledge, attitude, and practice towards the maintenance of the LCU devices. 51% of the dentists were quite satisfied with the devices. 48.9% of the dentists reported that they don’t have an idea of LCU intensity. Conclusions: Devices should be checked regularly for output intensity, light bulb efficiency, and composite build up. Dentists express poor knowledge, maintenance, and attitude toward LCUs. Educational programs are advisable and recommended.
 
Vol 6(1) Mohammed and Faraj


33
Cite this article as: Mahmood MA, Faraj BM. Evaluation of Light Curing Units and Dentists’ Knowledge About
Photo Polymerization Techniques in Sulaimani Governmental Dental Clinics. Sulaimani Dent J. 2019;6(1):33-39.
Mohammed A. Mahmood1, Bestoon M. Faraj2
Abstract
Objective: To measure the output intensity of light curing units (LCU) in governmental dental clinics of the Sulaimani governorate.
To evaluate practitioners’ knowledge on light cure application.
Methods: Ninety-four dental units and eighty-eight dentists in the Sulaimani governorate were included in this research. Output
intensity and diameter of tips of the dental LCU devices were measured. Two survey questionnaires were filled, one for the LCU and
its maintenance, the other for evaluating the knowledge, attitude, and practice of the dentists.
Results: Output power intensity results ranged from 50-2000W, with an average of 993W. The output intensity of 78% of the tested
devices was found to be acceptable regarding the manufacturer instruction of the radiometers used (every intensity below the given
range that given in the Result section is regarded as unusable or non-acceptable). The range of age in service of the devices was 1-12
years with a mean of 4.7 years. The average restorations performed in a week by a device was 17.8. Dentists had poor knowledge,
attitude, and practice towards the maintenance of the LCU devices. 51% of the dentists were quite satisfied with the devices. 48.9%
of the dentists reported that they don’t have an idea of LCU intensity.
Conclusions: Devices should be checked regularly for output intensity, light bulb efficiency, and composite build up. Dentists
express poor knowledge, maintenance, and attitude toward LCUs. Educational programs are advisable and recommended.
Keywords: Light-cure intensity, Dentists knowledge.
Submitted: March 17, 2019, Accepted: May 30, 2019, Published: June 1, 2019
DOI: https://doi.org/10.17656/sdj.10088
1. Kurdistan Board for Medical Specialties, Iraq.
2. Kurdistan Board for Medical Specialties and Department of Conservative Dentistry, College of Dentistry, University of
Sulaimani, Sulaimani, Iraq.
* Corresponding author: mhamadabm@gmail.com.
Published by College of Dentistry, University of Sulaimani
Evaluation of Light Curing Units and Dentists’
Knowledge About Photo Polymerization
Techniques in Sulaimani Governmental Dental
Clinics
Research Article
 
Light curing intensity and dentists’ knowledge Sulaimani Dent J. June 2019
34
Introduction
The development of the light-curing device came with
the development of light-activated composite materials
in the 1970s heralded a period of rapid progress in the
field of tooth-colored restorations. There is an
increasing demand for aesthetic restorative dentistry,
using primarily direct or indirect composites, or
composite bonded porcelain veneers(14).
Adequate depth of cure of a light-activated composite
depends on the intensity of the light source and the
exposure time, as well as on the material and other
factors(1). Many factors influence the clinical
performance of composites, of which the utmost
importance is the functioning of the light curing unit
(LCU), Duration of application is relatively easy to
control, but the intensity is not so readily monitored(2,
3). Three essential components are required for
adequate polymerization, namely, sufficient radiant
intensity, the correct wavelength of the light, and
ample curing time(4,5).
Visible light photo-activating LCU usually emit
wavelength spectra from 400 and 515 nm. Some units
may have small to moderate amounts of radiation
outside this range. Even though the dentist may
increase the irradiation time, there is a law of
diminishing returns, and the dentist must possess an
effective light source(6).
According to the results of a recent survey by Berry et
al., many current light activating units do not produce
adequate energy and increased irradiation time may not
compensate for this(7).
Light activation units are the standard items of
equipment in dental practice. This necessitates every
practitioner to understand various factors relating to the
maintenance of an LCU and their effect on clinical
performance and longevity of a restoration. For any
failure of a restoration, practitioners usually blame the
material used rather than technique or method of
placement(4,8).
Several studies have been conducted about the
measurement of output intensity and evaluation of
dentists’ knowledge on the type of light cure, exposure
time, maintenance of the device, and mode of
polymerization. Measured output intensity differs
among various researches, for example, acceptable
output intensity (every intensity below the given range
that given in the Result section is regarded as unusable
or non-acceptable) was found to be 68% in Malaysia(9)
and %45 in Israel(10), 25% in India(4). Researches from
England(11), Saudi Arabia(12) and Turkey(13) agree on
the necessity of further education, training, and
guidance of dentists about using light curing units(12).
This study has two aims: 1. to measure the output
intensity of light curing units in governmental dental
clinics of the Sulaimani governorate. 2. To evaluate
practitioners’ knowledge on light cure application.
Patients and methods
Ethical approval was taken from the Kurdistan Board
of Medical Specialties in Erbil. All of the
governmental dental clinics of Sulaimani Governorate
were selected. Two survey questionnaires were
prepared and validated by using the literature review
and previous researches with slight modification. One
of the surveys targeted light cure evaluation and
covered the following items: type of curing unit,
measurement of output intensity, manufacturer brand
or company name, age of curing unit, number of
restorations per week, maintenance of curing unit for
intensity, frequency of changing the bulb, diameter of
the light tip guides and presence or absence of
composite build-ups on the curing tips. The second
survey covered different aspects of light cure
application aiming evaluation of the dentists’
knowledge, such as knowledge of the light cure type,
used the mode of curing, type of restoration material,
application time, and regular maintenance of the device
and level of satisfaction.
The output intensity of the light cure devices was
measured by two manual radiometers (COXO) (Serial
No: P07237, CO04026) from China (Figure.1). Each
LCU was tested twice by two different radiometers for
the measurement of output intensity. The diameter of
the LCU tips was measured by a metal gauge and
checked for the presence or absence of composite build
up.
The first questionnaire was filled out by the author.
The second questionnaire was filled by the participants
(dentists) themselves in a one-to-one meeting.
Participants had the chance of asking about any part of
the survey that not understood enough. Participants
were the users of the LCUs measured in this research.
 
Vol 6(1) Mohammed and Faraj


35
Table 1: Dentists’ knowledge on photopolymerization techniques.
Results
Light-cure units
Ninety-four dental units were included in the research.
The majority of LCUs were Light Emission Diode
(LED) (91.5%), while the rest were Quartz Tungsten
Halogen (QTH) (8.5%). Regarding the output intensity,
the results ranged from 50-2000W with an average of
993W. According to the manufacturer of the used
radiometer, two independent criteria were used: 500-
800 mW/cm2 (recommended) and 300-500 mWcm2
(the composite should be tested to be sure).
Percentages of the unusable LCUs were 28.7% and
11.7% respectively. The majority (71.27%) of the
tested LCUs were above 500 mW/cm2. (every intensity
below the given range that given in the previous
sentence is regarded as unusable or non-acceptable).
Most of the units enrolled were of Satelec® device
made by action company from France (45.7%,
n=43)29.8% of the devices were of the unknown made
since they were without labels. The rest of the light-
cures were mostly Chinese (OSAKA, KASO,
Woodpecker, DTE, YDL-Hangzhou, Quayle, LY-
CZ40 China). The range of age in service of the
devices was 1-12 years with a mean of 4.7 years
(Figure 2). Besides, the tiredness of the devices was
measured by the number of restorations per week, with
an average of 17.8 restorations per week per each LCU
(Figure 3). All of the dental LCU lacked light intensity
maintenance, and none of them had given the chance of
bulb changing.
Variables No. %
Type of your light cure:
1. QTH.
2. LED.
3.Other
4. Not sure.
3
74
2
9
3.4
84.1
2.3
10.2
Mode of curing:
1. Soft Start
2. Continuous
3. Ramping
4. Not sure
21
62
3
2
23.9
70.5
3.4
2.3
For which restorations do you use the light cure?
1. Composite.
2. Glass Ionomer.
3. Composite + Calcium Hydroxide.
4. All.
42
2
30
14
47.7
2.3
34.1
15.9
Difference between the types regarding the efficiency of curing:
1. Yes
2. No
3. Not sure
55
9
24
62.5
10.2
27.3
How many seconds do you perform curing for composite restoration?
1. 10-15 sec.
2. 20-30 sec.
3. 30-60 sec.
9
61
18
10.2
69.3
20.5
Did you have any idea about light cure intensity previously?
1. Yes
2. No
45
43
51.1
48.9
Do you check your light cure regularly for maintenance?
Yes
N
o
25
63
28.4
71.6
Level of satisfaction with your light cure.
1. Very satisfied.
2. Quite satisfied.
3. Neither.
4. Quite dissatisfied.
5. Very dissatisfied.
27
45
8
2
6
30.7
51.1
9.1
2.3
6.8
 
Light curing intensity and dentists’ knowledge Sulaimani Dent J. June 2019
36
Figure 2: Linear correlation between output intensity and age of the device.
Figure 3: linear correlation between output intensity and number of
restorations/weeks.
Figure 1: Front and back side of the used radiometers.
 
Vol 6(1) Mohammed and Faraj


37
The diameter of the tip of the device, 52 of them (55%)
were 8 mm, followed by 7.5 mm in 36% of the devices.
As for the presence of the composite build up on the
device, most of them had composite buildups on their
tip of the light guide (75%). Sixty-seven percent of
67% of the devices were wired, while 33% were
portable. Summary of the results is shown in Table 1.
Dentists’ knowledge and the attitude most of the
participants correctly predicted the type of LCU (90%).
Regarding the common type of curing mode preferred
by the dentists, the majority (70%) used continuous
mode, followed by 23.9% for the soft start mode.
Composite was the main filling material for which
light-cure was used for; as 47.7% of the respondents
used light only for the composite, while 34% used in
combination with resin-modified glass ionomer
cement. 16% used light curing for all types of
materials. 62.5% of the participants believed that the
type of polymerization technique affects the efficiency
of the curing and the subsequent quality of the filling.
However, 10% said that there is no link between the
two, and, 27% were not sure. Most of the participating
dentists (69%) preferred 20-30 seconds for the curing,
while 10% selected 10-15 seconds and 20% selected
30-60 seconds. Nearly half of the respondents (48.9%)
reported that they don’t have an idea of light-cure
intensity. Besides, 71.6% they don’t check the device
for regular maintenance. Finally, the dentists asked
about the level of satisfaction with the light-cure they
used and, only 51% were quite satisfied.
Discussion
Light curing unit
Over a period, the intensity of LCUs will reduce due to
many factors, like composite resin build‑up on the
curing tip, condition of the bulb in units, orientation
position of the curing tip to the radiometer, voltage
regulation, and handling of the LCU(4,14).
The output of dental light-curing units is usually
evaluated indirectly by subjective tactile examination
of the upper surface of restorations. However, it has
been shown that it is not possible to completely
differentiate between adequately cured and under-cured
composite and because even an inferior curing unit can
polymerize the surface as well as an effective unit(15).
The recently marketed hand-held radiometers should
serve as a means to objectively measure the
performance of these units, instead of depending on
unreliable subjective parameters. These are simplified,
less accurate, chair‑side versions of sophisticated
laboratory equipment used to measure the output
intensity of curing lights(10).
According to the manufacturer of the radiometer used
in this research, two independent criteria were
conducted: 500-800 mWcm2 (recommended) and 300-
500 mWcm2 (the composite should be tested to be
sure) below 300mW considered as unusable.
Percentages of the unusable lighting cures (less than
300mW) were 28.7% and 11.7% respectively
(according to the two independent criteria mentioned
respectively in the previous sentence) with the average
of (20.2%). This finding is by Yogesh et al.(16) in which
unusable LCUs were estimated to be 22%.
As the age advances, the output intensity of these LCU
gradually diminishes(2). The present survey also
showed a significant reduction in light intensity with
older units. As the measured devices relatively old with
an average of 4.7 years. This agrees with the result of
similar researches(4).
Resin‑based composite build‑up on the light curing tip
was evident on 76% of 100 LCUs. This build‑up may
have a significant negative effect on intensity because
the resin‑based composite material partially blocks the
light output(8,17) since it is likely that intensity from the
LCU can be improved by removing any resin
contamination on the tip(17). The influence of the
number of composite build‑ups on the intensity of
curing light was not measured in the present survey;
hence, the effect of the cleaning of curing tips on the
outcome of this study is unknown.
About the monitoring of output intensity of curing
units, the majority of the participants reported that they
never checked the output intensity of their curing
lights. Findings from the neighbor country, Saudi
Arabia reflect the same problem(12). (the problem is
lack of maintenance).
A number of the restorations conducted with a device
in a given period could give a clue on the tiredness of
the device(17). In this study, the average number of
restorations per week per each light-cure unit is 17.8.
This result agrees with the findings of an Indian article
in which 49% of the tested LCU devices were
performing <10 restorations in a week(4).
Dentists’ knowledge
Adequate polymerization is a crucial factor in
obtaining optimal physical properties and satisfactory
 
Light curing intensity and dentists’ knowledge Sulaimani Dent J. June 2019
38
clinical performance of composite materials.
Inadequate polymerization reduces the strength,
hardness, water sorption, and color stability of
composites(18,19). Several factors affect resin
polymerization, including a selection of the appropriate
LCU; light intensity, wavelength and exposure time;
and size, location, and orientation of the LCU tip(20,21).
In addition to the type of LCU used, the clinician’s
knowledge and skill regarding the use of LCUs plays
an important role in the outcome of polymerization and
thus the outcome of resin-based restorations(13). The
findings of this study indicated that dental clinicians
had low levels of knowledge regarding LCUs and
materials science.
Knowledge and understanding of the LCU output
intensity are important to ensure sufficient curing of
the resin and to prevent any damage or harm(12). Most
of the respondents did not know the output intensity of
the LCU they were using, which increases their
chances of having insufficient curing and
polymerization by the LCU. This agrees with the
results of Alsuliman et al.(12) (In which 88.7% of the
participants were unaware of the intensity value of
their LCUs).
Conclusions
Based on the study results, it’s concluded that about
78% of the tested LCUs were found to be acceptable
(more than 300mW). Dentists express poor knowledge,
maintenance, and attitude toward LCUs. Educational
programs are advisable and recommended.
Acknowledgment
The authors are giving their thanks and showing their
appreciation to all effective Members of Kurdistan
Board for Medical Specialties, Restorative Dentistry.
Very special thanks to Professor Ali Al-Zubaidi for his
great help and support during the years of our study in
Kurdistan Board.
References
1. Sakaguchi R, Douglas W, Peters M. Curing light
performance and polymerization of composite
restorative materials. J Dent. 1992;20(3):183-8.
2. Mitton B, Wilson N. Practice maintenance: The
use and maintenance of visible light activating
units in general practice. Br Dent J.
2001;191(2):82.
3. Shortall A, Harrington E, Wilson H. Light curing
unit effectiveness assessed by dental radiometers. J
Dent. 1995;23(4):227-32.
4. Madhusudhana K, Swathi T, Suneelkumar C,
Lavanya A. A clinical survey of the output
intensity of light curing units in dental offices
across the Nellore urban area. SRM J Res Dent
Sci. 2016;7(2):64.
5. Santini A. Current status of visible light activation
units and the curing of light-activated resin-based
composite materials. Dent. update.
2010;37(4):214-27.
6. Rueggeberg F, Caughman WF, Curtis J. Effect of
light intensity and exposure duration on cure of
resin composite. Oper. Dent. 1994;19:26-.
7. Berry T, Barghi N, Godwin J, Hunter K.
Measurement of intensity of curing light units in
dental offices. J Dent Res. 1992;71:161.
8. Hegde V, Jadhav S, Aher GB. A clinical survey of
the output intensity of 200 light curing units in
dental offices across Maharashtra. J conserv dent.
2009;12(3):105.
9. Ab Rahman A, Husein A, Ahmed HMA,
Mohamad D, Bakar WZW, Farea M, et al. A
survey on light intensity outputs of QTH, cabled
and cordless LED light curing units. Arch Orofac
Sci. 2014;9:85-90.
10. Pilo R, Oelgiesser D, Cardash H. A survey of
output intensity and potential for depth of cure
among light-curing units in clinical use. J Dent.
1999;27(3):235-41.
11. Santini A, Turner S. General dental practitioners'
knowledge of polymerization of resin-based
composite restorations and light curing unit
technology. Br Dent J. 2011;211(6):E13.
12. Alsuliman SA. Knowledge of Dental Practitioners
about Light Curing Units in Saudi Arabia. Int J
Dent & Oral Heal. 2018;4(9):127-9.
13. Nuray Tüloğlu, Sezin Özer, Emine Şen Tunç,
Sevgi Canbaz, Şule Bayrak. Knowledge And
Attitudes of Dental Clinicians Regarding Light-
Curing Units In Northern Turkey. cdr.
2016;40(1):26-34.
 
Vol 6(1) Mohammed and Faraj


39
B
14. Burgess JO, Walker RS, Porche CJ, Rappold AJ.
Light curing--an update. Compendium of
continuing education in dentistry (Jamesburg, NJ:
1995). 2002;23(10):889-92.
15. Hansen EK, Asmussen E. Correlation between
depth of cure and surface hardness of a light‐
activated resin. Eur J Oral Sci. 1993;101(1):62-4.
16. Yogesh Sahu VK, Aditi Jain, Pooja Deshmukh,
Deepti Jain, Shruti Sharma. To evaluate the
functional efficacy of various light curing units in
dental clinics across Chhattisgarh state. IJCE.
2017;2(1):1-5.
17. Maghaireh GA, Alzraikat H, Taha NA. Assessing
the irradiance delivered from light-curing units in
private dental offices in Jordan. J Am Dent Assoc.
2013;144(8):922-7.
18. Pires JAF, Cvitko E, Denehy GE, Swift Jr EJ.
Effects of curing tip distance on light intensity and
composite resin microhardness. Quintessence Int.
1993;24.
19. Oberländer H, Friedl K-H, Schmalz G, Hiller K-A,
Kopp A. Clinical performance of polyacid-
modified resin restorations using” softstart-
polymerization”. Clin Oral Investig. 1999;3(2):55-
61.
20. Kwon YH, Kwon T-Y, Ong JL, Kim K-H. Light-
polymerized compomers: coefficient of thermal
expansion and microhardness. J Prosthet Dent.
2002;88(4):396-401.
21. Nomoto R, Uchida K, Hirasawa T. Effect of light
intensity on polymerization of light-cured
composite resins. Dent Mater J. 1994;13(2):198-
205.
... The majority of results documented that the LCUs used in their studies recorded an acceptable, adequate, and as being within the required range. 18,[20][21][22][23][24][25][26] Other studies concluded an inverse relationship between the clinical age of LCUs, 12,18,27 the presence of uncleaned with remaining adhesive and resin materials stick on the LCU tip and the radiant exitance of LCUs. 22,28 No previous study was conducted in different dental colleges in Sanaa city. ...
... For radiant exitance measurements, each measurement of each LCU device was recorded three times as the 1 st , 2 nd , 3 rd reading, then the averages were taken as done in many earlier studies. 18,[20][21][22][23][24]28 They divided into three groups and graded as (insufficient intensity that cannot be compensated even by increasing the light time), (marginal intensity at which additional curing was needed), and (sufficient intensity where further curing time would not be necessary) and represented as <400 mW/cm², between 400 and 850 mW/cm², and >850 mW/cm². ...
Article
Full-text available
Background To evaluate the radiant power of the light cure units (LCUs) in relation to their type, radiant exitance, number of years in clinical use, and condition of LCUs tips in governmental and public clinics in Dental Faculties in Sana’a City. Materials and Methods LCUs were collected from different colleges at Sanaa City, Yemen, then LCU data as type, clinical age (<1 year, between 1–5 and ˃ 5-years), tip condition was visually inspected for damage and adhering debris, and the radiant exitance values of the tested LCUs. Radiant exitance values were subcategorized into three groups: <400, 400–850, and >850 mW/cm², labeled as inadequate, marginal, and adequate radiant exitances, respectively. A Woodpecker radiometer was used with a mode lasting of 20 seconds was used with each LCU. Descriptive statistics of the different parameters were evaluated with SPSS version 25. One-way ANOVA and Mann–Whitney tests were performed to determine the mean difference between the groups with a significance value of ˂ 0.05 was considered. Results Two hundred twenty-three LCUs were surveyed, and the majority were Light–emitting diode (LED). Forty-nine (21.9%), 117 (52.4%), 57 (25.6%) recorded lesser than, 400–850, and more than 850 mW/cm², respectively. Radiant exitances of < year-old units were found to be higher than those of units used for ˃ 5 years with significant differences (p=0.001). The ANOVA test showed significant differences between the radiant exitance with clinical age and LCU tip conditions and a strong correlation p ˃ 0.050. Conclusion LED curing lights were the most used in the tested Dental Faculties. More than half of the used LCU offered sufficient radiant exitance. Clinical age, the presence or absence of composite buildups, and damage to curing tips showed significantly affect radiant exitance values.
... Donde el rendimiento de estas restauraciones depende directamente de la polimerización del material restaurador (6,9,18).La integridad mecánica del material correctamente fotocurado, indica la capacidad de resistir a las fuerzas compresivas. Dichas fuerzas se ven afectadas por la intensidad lumínica con que fue fotocurado el material (19).La intensidad de las lámparas de fotocurado, pueden llegar a una intensidad lumínica muy alta, cuyo rango varía desde los 900 a 1600 mW/cm2(20,21). Hay que tener en cuenta que para lograr activar al fotoiniciador que está contenido en los materiales fotopolimerizables, se requerirá un rango de intensidad lumínica de 420 -580 nm. ...
... cuando la intensidad lumínica oscilaba entre 500 -800 mW/cm2 y si los valores se encontrasen entre 300 -500 mW/cm2, los odontólogos deberían asegurarse de que el material esté fotocurado correctamente(19).Otro de los resultados que concordaron con el estudio fue el de Madhusudhana y cols; encontraron que el 57 % de las lámparas de fotocurado mantenían una intensidad marginal y que de ellas solo el 60,7 % eran tipo LED. Para ellos una intensidad lumínica inadecuada era cuando tenía menos de 400 mW/cm2. ...
Article
Full-text available
A eficácia de um tratamento restaurador com compostos resinosos pode ser afetada pela intensidade luminosa que o fotopolimerizador é capaz de emitir. O objetivo deste estudo foi determinar a prevalência de intensidade de luz de lâmpadas de fotopolimerização em clínicas odontológicas de centros de saúde na cidade de Cuenca, Equador. A intensidade luminosa de 38 lâmpadas fotopolimerizadoras correspondentes a 23 centros de saúde foi avaliada e classificada em baixa potência (<300mw/cm2), potência regular (300 - 800 mw/cm2) e alta potência (>800 mw/cm2). A intensidade da potência luminosa foi avaliada com o radiômetro CK12024 por três observadores antes da autorização dos profissionais responsáveis pela análise de seus fotopolimerizadores. Verificou-se que, das 38 lâmpadas fotopolimerizáveis avaliadas, 5,26% (n=8) apresentaram potência baixa, 21,05% (n=8) potência regular e 73,68% (28) potência alta. De acordo com os dados obtidos, concluímos que a maioria das lâmpadas polimerizadoras nos centros de saúde da cidade de Cuenca tem um ótimo desempenho, além de enfatizar a avaliação periódica das lâmpadas polimerizadoras para controlar o desempenho ideal.
Article
Full-text available
Introducción: La lámpara de fotocurado, que utiliza diodos emisores de luz (LED), se emplea en odontología para la conversión polimérica de los materiales de restauración dental. Se ha comunicado que una intensidad lumínica inadecuada de la lámpara no aseguraría la correcta polimerización del material de restauración. Objetivo: Determinar la intensidad lumínica de las lámparas de fotocurado LED en consultorios odontológicos de la ciudad de Piura, Perú, 2020. Métodos: Estudio observacional, descriptivo. Se midió la intensidad lumínica en 70 lámparas de fotocurado LED, usando un radiómetro con una longitud de onda de 400-500 nm, con capacidad de medida de la intensidad lumínica de hasta 3500 mw/cm². Por debajo de los 400 mw/cm² indica intensidad baja, de 400 a 800 mw/cm² intensidad media, de 800 a 1200 mw/cm² intensidad alta y por encima de los 1200 mw/cm² indica intensidad muy alta. Resultados: El 48,5 % de las lámparas analizadas presentaban intensidad media, el 22,86 % intensidad alta, mientras que el 15,71 % intensidad baja y finalmente el 12,86 % de las lámparas presentaban intensidad muy alta. Se reportó menor frecuencia de lámparas con mayor uso clínico. Conclusiones: Las lámparas de fotocurado LED, utilizadas en los consultorios dentales de la provincia de Piura durante el 2020, emiten una intensidad lumínica promedio de 778,14 mW/cm², equivalente a la intensidad media.
Article
Full-text available
Background and Aim: Different types of light-curing units have been used for the polymerization of resin-based materials, however, dental clinicians’ knowledge and attitudes regarding light-curing units and materials science have been the focus of limited studies so far. This study aimed to evaluate the knowledge and attitudes regarding light-curing units and materials science among dental clinicians in Northern Turkey. Materials and Methods: A survey questionnaire was sent by email to 448 dental clinicians, who were asked to provide demographic information and respond to questions on their knowledge and attitudes regarding current LCUs and materials science. Chi-square tests were used to identify the differences in responses for different variables. Results: A total of 184 questionnaires were responded, which corresponds to a rate of 41%. All of the respondents were the ones who had already used resin-based composite materials. 41% (n=75) of the respondents preferred LED units. Although 102 (55%) of the respondents had access to LCUs that offered various options for irradiation, 82 (53%) of them used the conventional mode. Neither of the variables examined (specialization, years of professional experience) were found to have a statistically significant effect on the number of correct responses regarding LCUs and materials science (p> 0.05). Conclusion: The results of this study demonstrated that the knowledge of dentists on materials science and LCUs was poor. Therefore, materials science and LCUs educational programs for dentists are needed. � 2016, Hacettepe Universitesi Yerbilmleri. All rights reserved.
Article
Full-text available
Aim: The aim of the survey was to examine the output intensity of curing units and other related factors in private dental offices across Nellore urban area. Materials and Methods: A questionnaire was prepared about the type of curing unit, number of restorations performed in a week, maintenance of curing unit, frequency of changing bulb, measurement of output intensity, presence, or absence of composite build--ups on curing tips. The questionnaire was submitted to 100 private dental offices located in Nellore urban area. Each variable in the questionnaire had an impact on quality of the composite restoration. Each curing unit light tip was also observed for the presence or absence of composite build--up. The output intensity of the curing light was measured using a digital radiometer (Ivoclar). The average of the three readings of the output intensity was obtained for each curing light. The average output intensity was divided into three categories ( Statistical Analysis: The results were statistically analyzed using linear regression analysis. Results: Among the 100 curing units examined, 84 were light emitting diode (LED) units, and 16 were quartz tungsten halogen (QTH) units. Only 22% LED machines and 3% QTH curing units had adequate intensities (>850 mW/cm2). A significant reduction in output intensity is seen with both types of older light curing units. Nearly 50% of practitioners had never checked the light output of their unit. Conclusion: It was concluded that there is a general lack of awareness among dentists of the need for maintenance of these units.
Article
Full-text available
Light intensity output is one of the determinants for adequate curing of visible light-cured materials. The aim of this survey was to evaluate the light intensity outputs (LIOs) of light curing units (LCUs) in dental clinics of Hospital Universiti Sains Malaysia (HUSM) and School of Dental Sciences, Universiti Sains Malaysia (USM). The respective LIOs of all functioning Quartz Tungsten Halogen (QTH) and Light Emitting Diode (LED) LCUs were tested using two light radiometers. For cordless LED LCUs, the testing procedure was done in situ and after being fully charged. Statistical analysis using Kruskal Wallis and Wilcoxon signed ranks tests were performed to compare the LIOs between groups and between the LIOs of in situ and post-charged cordless LED LCUs, respectively. The level of significance was set at 0.05 (P<0.05). The results reveal that 72.72%, 42.47% and 92% of QTH, cabled LED and cordless LED LCUs exhibited acceptable LIOs, respectively. Data analysis using Kruskal Wallis test shows a statistically significant difference between groups (P<0.05). The intergroup comparison using multiple Mann Whitney test with Bonferroni correction reveals a significant difference between the LIOs of cordless LED and both QTH and cabled LED (P<0.017). The difference between the LIOs of in situ and post charged cordless LED LCUs is also significant (P<0.05). In conclusion, both QTH and cordless LED LCUs performed better in term of LIOs than cabled LED LCUs. Periodic testing of LCUs is essential to ensure optimal performance.
Article
Full-text available
The authors conducted a study to examine the irradiance from light-curing units (LCUs) used in dental offices in Jordan. Two of the authors visited 295 private dental offices (15 percent) in Jordan and collected the following information about the LCUs: age, type (quartz-tungsten-halogen or light-emitting diode), date of last maintenance, type of maintenance, last date of use, number of times used during the day, availability of a radiometer, exposure time for each resin-based composite increment, size of light-curing tips and presence of resin-based composite on the tips. The authors used a radiometer to measure the irradiance from the LCUs. They used linear regression with stepwise correlation for the statistical analysis. The authors set the minimum acceptable irradiance at 300 milliwatts/square centimeter. The mean irradiance of the 295 LCUs examined was 361 mW/cm(2), and 136 LCUs (46.1 percent) delivered an irradiance of less than 300 mW/cm(2). The unit's age, type and presence of resin-based composite on the light-curing tips had a significant effect on the irradiance (P ≤ .001). Only 37 of the 141 quartz-tungsten-halogen units (26.2 percent) and 122 of the 154 light-emitting diode units (79.2 percent) delivered at least 300 mW/cm(2). Resin contamination on the light-curing tips had a significant effect on the irradiance delivered. The irradiance from the LCUs decreased with use. Practical Implications. The irradiance from many of the units in this study was less than 300 mW/cm(2), which may affect the quality of resin-based composite restorations. Dentists should monitor the performance of the LCUs in their offices weekly.
Article
Full-text available
Clinical successful use of resin-based composite restorations (RBCs) depends on knowledge of material and light curing unit (LCU) related factors. The purpose of this study was to evaluate general dental practitioners' knowledge of polymerisation of RBCs and LCU technology. Members of the Active Research Group of the Faculty of General Dental Practice (UK) in England, Scotland and Wales engaged in primary dental care were sent a letter introducing the study and asking for their cooperation, followed by an email containing a link to the online survey questionnaire, hosted on Surveymonkey.com. The questionnaire enquired about current LCUs, and asked a series of questions on material science. Sixty-six percent of the 274 members contacted responded. Fifty-seven percent used LED units, 25% quartz tungsten halogen (QTH), and 1% plasma arc (missing: 17%). Thirty percent reported having access to a radiometer. Appropriate responses regarding the degree of conversion of composite and adhesive materials were given by 32% and 23% respectively, and 22% agreed that LED and QTH LCUs had comparable efficiency in polymerising composites. Thirty-three percent were aware that RBCs eluted substances that may have adverse local or systemic consequences. Fifty-eight percent stated that if polymerisation of RBC is slowed down, polymerisation stress will be lower, and 43% said that polymerisation shrinkage will be reduced if the degree of conversion is reduced. Knowledge (measured by appropriate responses to these questions) was not related to years since qualification (r=-0.05, n=168, p=0.53). The study suggests that dentists' knowledge of curing RBC restorations and LCUs is poor. This indicates that there is a need for training and guidance in this aspect of primary dental care.
Article
Full-text available
Unlabelled: Light activation units are standard items of equipment in dental practice. It is essential to understand the many factors which affect the polymerization of light-activated resin composite materials and the choice of a light curing unit. In this respect, the development of high-intensity halogen and light-emitting diode (LED) light curing units (LCUs), many with multiple curing modes, has revolutionized light curing techniques. This article reviews visible light activation unit design and development. Factors influencing the effective use of LCUs and polymerization of resin-based composite materials are discussed, as are the steps which should be taken to maintain the efficiency of units in clinical use. Clinical relevance: Many LCUs produce lower output intensities than stated by the manufacturer. Newer high power LEDs may present as much of a heat problem as high power quartz tungsten halogen lamps (QTHs).The manufacturer's data should be followed to ensure that the emission spectra of the unit is compatible with the photo-initiator in the resin-based composite material.
Conference Paper
Objective: A wide gulf between the knowledge about curing lights (LCUs) held by researchers and by practitioners has been reported in Scotland (Santini & Turner, BDJ, 2011). This study examined if a similar situation exists in Canada. Method: An on-line survey of was conducted of 344 Canadian dentists to discover what these practitioners knew about their LCU. Result: The survey showed that Canadian dentists have low level of technical knowledge about the LCU they were using on their patients. Amongst the respondents, 73% were using LED, 13% were using halogen, 5% were using plasma arc curing lights, and the remainder (9%) did not know. 50% did not know the intensity of their LCU, and 58% checked the output from their curing light once a year or never, and 16% checked the output from their LCU at least once a week. 60% did not watch what they are doing when light curing. 72% did not know the wavelengths produced by their LCU, 73% did not know which photoinitiators were used in the resin restorations and what resins their light was most compatible with. While 98% of the respondents were familiar with and could explain the term ‘Depth of Cure’, only 42% were familiar with the term irradiance. The majority of the dentists (241) followed the resin composite manufacturers' instructions to determine the light exposure time they used. Conclusion: Canadian dentists exhibited a similar low level of knowledge about the LCUs and resins they were using. They did not understand the importance of proper light curing when placing light cured restorations. The majority did not routinely test the light output from their LCU and did not know if the light output matched the requirements of the resin they were using. Several key terms were not well understood and this may lead to inadequately cured restorations and premature restoration failure.
Article
The purpose of this study is to examine the intensity of light curing units and factors affecting it in dental offices. The output intensity of 200 light curing units in dental offices across Maharashtra were examined. The collection of related information (thenumber of months of use of curing unit, the approximate number of times used in a day, and presence or absence of composite build-ups) and measurement of the intensity was performed by two operators. L.E.D Radiometer (Kerr) was used for measuring the output intensity. The average output intensity was divided into three categories (<200 mW/cm(2), 200-400 mW/ cm(2)and >400 mW/cm(2)). Among the 200 curing units examined, 81 were LED units and 119 were QTH units. Only 10% LED machines and 2% QTH curing units had good intensities (>400 mW/cm( 2)). Most of the examined curing lights had low output intensity ranging from 200 to 400 mW/cm(2), and most of the curing units had composite build-ups on them.
Article
Aim The present study to investigate the use, care and maintenance of light units in everyday clinical practice was undertaken to complement light unit emission surveys, with a view to developing a protocol for light unit use and care in everyday clinical practice.
Article
The majority of modern composite restorative materials require light activation for polymerization. Variables affecting light energy absorption by the composite have been examined for their effect on the polymerization contraction. Since the polymerization contraction is closely associated in a complex way to the degree of cure of the restoration, this parameter served as an empirical indicator for the extent of polymerization. Variables included the composite shade, distance between the light source and composite sample, and light intensity. Three resin composites are evaluated. Post-gel polymerization contraction was evaluated using a strain gauge method. Curing light intensity diminished rapidly for distances greater than 2 mm between the tip of the light guide and material surface. A linear relationship was demonstrated between polymerization contraction and light intensity. The polymerization contraction of a microfilled composite and posterior composite, using a constant curing time and light intensity, decreased linearly with increasing sample thickness. Less than optimal light output of the curing light source can be compensated by increasing application time within reasonable limits.