Conference PaperPDF Available

Sunram 5: An MR Safe Robotic System for Breast Biopsy

Authors:
Sunram 5: An MR Safe Robotic System for Breast Biopsy
V. Groenhuis1, F.J. Siepel1, M.K. Welleweerd1, J. Veltman2, S. Stramigioli1,3
1Robotics and Mechatronics, University of Twente, The Netherlands
2Ziekenhuisgroep Twente, Almelo, The Netherlands
3ITMO, Saint Peterburg, Russia
v.groenhuis@utwente.nl
INTRODUCTION
The conventional MRI-guided breast biopsy procedure
is inaccurate due to manual needle insertions and the
necessity of rescanning. Therefore, the procedure would
benefit from an MR safe robotic system to manipulate
the needle precisely towards suspicious lesions under
near-realtime MRI guidance. The main challenge is to
design a system consisting of non-metallic, non-
magnetic and non-conductive materials in order to
eliminate safety hazards and imaging artifacts due to
interference with the MRI’s magnetic field.
Pneumatic stepper motors have been proved an effective
actuation method in several state-of-art robotic
systems[1, 2]. These systems are limited in bandwidth as
5 m long pneumatic tubes are needed between the valve
manifold and the robotic system. This results in a
stepper motor frequency of approximately 10 Hz when
maximum force is necessary. The Stormram 4 needs
approximately one minute to move from one position to
another considering an axis of 640 steps. Higher
velocities are desirable for practical applications[2].
Workspace is important for the breast biopsy robot: a
suspicious lesion may be identified at the medial side or
close to the chest wall. These locations are difficult to
reach in state-of-art systems. Safety is also a concern: in
case of emergency (including loss of pressure or mains
power), the system should be able to eject the needle
from the patient within one second.
This research aims to develop a new robotic system
addressing the aforementioned challenges. The specific
novelties are enhanced speed and accuracy, enlarged
workspace and improved safety.
MATERIALS AND METHODS
Fig. 1 Photo of Sunram 5.
The Sunram 5 robot is shown in Fig. 1. The five degree
of freedom manipulator is actuated by six pneumatic
stepper motors, measures 107 x 72 x 56 mm and has a
mass of 260 g. Two of the motors have a relatively large
step size of 1.7 mm, allowing for high-speed lateral and
needle insertion movements. These motors are coupled
with small-step motors (step size 0.3 mm) to maintain
sub-millimeter accuracy in the same direction.
Each stepper motor consists of two or three pneumatic
cylinders which alternatingly press against a straight or
curved rack to make discrete steps[2]. Besides these six
stepper motors three individual cylinders are also
present to fire the biopsy gun and activate the needle
ejection safety mechanism.
Fig. 2 Needle insertion actuator with safety ejection
mechanism.
Fig. 2 shows a schematic of the three-cylinder needle
insertion stepper motor with integrated safety ejection
mechanism. In normal operation pistons A, B and C are
alternatingly pushed down and up, nudging the rack
(with needle) to the left or right in discrete steps. In case
of emergency all three pistons A, B and C are lifted up
and piston E is pushed to the right by a separate
cylinder. As piston E is connected to the rack holding
the needle, the needle is retracted out of the breast.
The robot is attached to a breast fixation system inspired
by Machnet’s device to immobilize the breast[3]. It can
be manually positioned at a range of orientations
relative to the breast allowing to adapt the compression
direction and biopsy access window to the specific
location of the lesion to be targeted. Fig. 3 shows a
photograph of the entire system during the biopsy
process.
Hamlyn Symposium on Medical Robotics 2018
82
Fig. 2 Sunram 5 (left) targeting a breast phantom inside the
fixation system (right).
The majority of the Sunram 5 parts was printed using a
high-resolution polyjet printer. Acrylic rods support the
hinges and laser-cut silicone rubber seals make the
cylinders airtight. The removable 14G (2.1 mm) MR
conditional titanium needle is the only metallic
component of the system.
Three aspects of the robotic system were tested
including speed/accuracy, workspace and safety. The
highest manipulation time from one configuration to
another is based on the range of motion of the different
motors expressed as the number of steps, while the step
sizes determine the theoretical positioning accuracy.
The workspace was evaluated by estimating the volume
percentile of the breast phantom which can be reached
by the biopsy system, taking both the positioning
freedom of the fixation system and the range of motion
of the robot into account. The safety aspect was
evaluated by first inserting a needle in the phantom to a
depth of 50 mm and then activating the safety
mechanism at pressures ranging from 0.1 to 0.4 MPa
while recording the needle movement with a high-speed
camera to measure the needle ejection time.
RESULTS
Speed and accuracy
The six stepper motors have a range of motion varying
from 61 to 167 steps. At a stepping frequency of 10 Hz
the upper bound of the manipulation time from one
position to another is 16.7 seconds. When starting in a
mid-way position and/or taking into account that two
motors can work together in certain directions the
effective maximum manipulation time is reduced to 8.4
seconds.
In the lateral and needle-insertion directions the small-
step actuators yield a step size of 0.3 mm. In vertical
direction two rotational joints have a step size of 0.3°,
which translates to a 0.52 mm displacement at 100 mm
from the rotation axis. Based on these step sizes, sub-
millimeter accuracy can be achieved. Further enhanced
accuracy is possible by coordinating movements of two
motors: in collinear motor pairs, the smallest achievable
step is 0.1 mm which is the greatest common divisor of
its individual step sizes.
Workspace
The orientable fixation system significantly enhances
workspace when the approximate location of the target
lesion is already known. In combination with the high
dexterity of the robotic system, any lesion that is
situated below the fixation system’s top surface is
reachable by the robotic system.
Safety
A minimum pressure of 0.2 MPa is required to
successfully activate the safety mechanism, ejecting the
needle out of the phantom in 1.8 s. At 0.3 MPa, the time
is reduced to 0.53 s, while at 0.4 MPa the needle is
ejected in 0.31 s.
DISCUSSION
Pneumatic stepper motor technology has shown to be an
effective actuation method for the Sunram 5. The six
degrees of freedom combined with a versatile breast
fixation system has potential to target lesions in the
whole breast. Thanks to the dual-speed motors the
manipulation time is under ten seconds, making it an
order of magnitude faster than state-of-art robotic
systems without compromising on accuracy. In case of
emergency the safety mechanism is able to eject the
needle out of the breast within one second at a pressure
of 0.3 MPa or higher. A limitation might be that no
direct position feedback is employed, so a calibration
routine based on MRI or visual feedback may be
necessary prior to the procedure.
The next steps in this research involve quantitative
evaluation of targeting accuracy when performing
biopsies in phantoms under MRI guidance. Needle-
tissue interaction may contribute to deformations of the
needle and/or the breast, potentially introducing
localization errors which need to be assessed and
mitigated. While the Sunram 5 has shown to be a major
advancement over its predecessors, additional work is
required concerning the placement of RF coils inside the
breast fixation system and in sterilization of the device.
REFERENCES
[1] D. Stoianovici et al., Multi-Imager Compatible, MR
Safe, Remote Center of Motion Needle-Guide Robot, in
IEEE Transactions on Biomedical Engineering, vol. 65,
no. 1, pp. 165-177, Jan. 2018.
[2] V. Groenhuis et al., Design and characterization of
Stormram 4: An MRI-compatible robotic system for
breast biopsy, 2017 IEEE/RSJ International Conference
on Intelligent Robots and Systems (IROS), Vancouver,
BC, 2017, pp. 928-933.
[3] J. Veltman et al., “Magnetic Resonance-Guided Biopsies
and Localizations of the Breast: Initial Experiences Using
an Open Breast Coil and Compatible Intervention
Device,” Investigative Radiology, 2005, vol. 40, no. 1,
pp. 379-384
Hamlyn Symposium on Medical Robotics 2018
83
... MR safe surgical robots powered by pneumatic stepper motors have also been demonstrated. The Sunram 5 and previous iterations from the authors of this paper are driven by combinations of linear and curved stepper motors [54,55], while pneumatic robotic systems by other authors are typically driven by rotational or helical stepper motors [15,121]. art motors in blue. ...
... In the next year the Sunram 5 was developed (Figure 10.4, right) which utilizes dual-speed motors on certain axes. The Sunram 5 has five degrees of freedom driven by a total of six stepper motors and also includes a breast fixation system, a pneumatic biopsy gun and a safety needle ejection mechanism [54]. All stepper motors used in Stormram 1-4 and Sunram 5 consist of two or three double-acting cylinders that engage on a straight or curved rack by means of a wedge mechanism [47,53]. ...
... These changes are not visible due to the low sampling frequency. This has led to the launch of research projects focused on the introduction of surgical robotics to overcome the existing limitations and improve the accuracy of biopsies of MRI-visible lesions [54]. ...
... In the lateral direction, the actuators yield a step size of 0.3 mm, and in the vertical direction, the rotational joints have a step size of 0.3 • , which translates to 0.52-mm displacement. Submillimeter accuracy was achieved based on the step sizes [194]. ...
... Although ultrasound is free of ionizing radiation, it is userdependent, and good nerve visualization can be technically difficult, especially for nerves in and around the bony an integrated imaging coil [49]. (b) Sunram 5 targeted a breast phantom inside the fixation system [194]. ...
Article
Full-text available
Magnetic resonance imaging (MRI) can provide high-quality 3-D visualization of target anatomy, surrounding tissue, and instrumentation, but there are significant challenges in harnessing it for effectively guiding interventional procedures. Challenges include the strong static magnetic field, rapidly switching magnetic field gradients, high-power radio frequency pulses, sensitivity to electrical noise, and constrained space to operate within the bore of the scanner. MRI has a number of advantages over other medical imaging modalities, including no ionizing radiation, excellent soft-tissue contrast that allows for visualization of tumors and other features that are not readily visible by other modalities, true 3-D imaging capabilities, including the ability to image arbitrary scan plane geometry or perform volumetric imaging, and capability for multimodality sensing, including diffusion, dynamic contrast, blood flow, blood oxygenation, temperature, and tracking of biomarkers. The use of robotic assistants within the MRI bore, alongside the patient during imaging, enables intraoperative MR imaging (iMRI) to guide a surgical intervention in a closed-loop fashion that can include tracking of tissue deformation and target motion, localization of instrumentation, and monitoring of therapy delivery. With the ever-expanding clinical use of MRI, MRI-compatible robotic systems have been heralded as a new approach to assist interventional procedures to allow physicians to treat patients more accurately and effectively. Deploying robotic systems inside the bore synergizes the visual capability of MRI and the manipulation capability of robotic assistance, resulting in a closed-loop surgery architecture. This article details the challenges and history of robotic systems intended to operate in an MRI environment and outlines promising clinical applications and associated state-of-the-art MRI-compatible robotic systems and technology for making this possible.
... These changes are often not detectable due to the low sampling frequency. This has led to the launch of research projects focused on the introduction of surgical robotics to overcome the existing limitations and improve the accuracy of biopsies of MRIvisible lesions [1], [2]. ...
Article
Biopsy of breast lesions only visible on MRI may benefit from patient-specific finite-element modeling of the breast. This requires acquisition of the heterogeneous elasticity distribution. An acoustically transparent pad has been developed which enables strain elastography imaging using ultrasound without need for additional force or pressure sensors. The elasticity of the pad is known a priori while the change in thickness is detectable on ultrasound by edge detection. Based on the calculated pressure distribution across the pad, the pressure in inner tissue can be approximated. The strain is measured using a cross-correlation block matching algorithm and from these measurements the heterogeneous elasticity distribution can be estimated. The ultrasound probe with pad is moved over the breast surface by a robot arm. At sixteen sites on the surface of one phantom the probe with pad performed an inward push movement perpendicular to the breast surface. The ultrasound scans were post-processed to reconstruct a heterogeneous elasticity map in the local imaging plane. While it is challenging to accurately estimate the stress and strain in deeper tissue, certain features such as the skin can be effectively distinguished from the underlying soft tissue which demonstrates the potential of the system.
... Several MR safe surgical robots powered by pneumatic stepper motors have also been demonstrated. The Sunram 5 and previous iterations from the authors of this paper are driven by combinations of linear and curved stepper motors [11], [12], while pneumatic robotic systems by different authors are generally driven by rotational or helical stepper motors [13], [14]. Fig. 1 graphically shows the build volumes and step sizes of the state-of-art motors in blue. ...
Chapter
This work presents a soft robotic system that supports interventionalists in tumor diagnostics through biopsy and ablation in the magnetic resonance imaging (MRI) environment. The system is designed to indicate the location of the puncture, assist in holding the biopsy and ablation needles, and support the fine adjustment of the puncture angle. The robot consists of a six-chamber soft actuator that can be extended with additional actuators or end effectors using a connector system. Two end effector designs are presented, a needle holder, and a system for indicating and, if necessary, marking the puncture site. The soft robot components are prototyped and characterized in an experimental set-up regarding the bending capabilities and evaluated with the help of an experienced interventionalist. A minimum bending angle of 23.5123.51^\circ was achieved, which satisfies the predefined requirements.Keywordssoft roboticsinterventional radiologytumor ablation
Article
Full-text available
Computerized Tomography Angiography (CTA) based follow-up of Abdominal Aortic Aneurysms (AAA) treated with Endovascular Aneurysm Repair (EVAR) is essential to evaluate the progress of the patient and detect complications. In this context, accurate quantification of post-operative thrombus volume is required. However, a proper evaluation is hindered by the lack of automatic, robust and reproducible thrombus segmentation algorithms. We propose a new fully automatic approach based on Deep Convolutional Neural Networks (DCNN) for robust and reproducible thrombus region of interest detection and subsequent fine thrombus segmentation. The DetecNet detection network is adapted to perform region of interest extraction from a complete CTA and a new segmentation network architecture, based on Fully Convolutional Networks and a Holistically-Nested Edge Detection Network, is presented. These networks are trained, validated and tested in 13 post-operative CTA volumes of different patients using a 4-fold cross-validation approach to provide more robustness to the results. Our pipeline achieves a Dice score of more than 82% for post-operative thrombus segmentation and provides a mean relative volume difference between ground truth and automatic segmentation that lays within the experienced human observer variance without the need of human intervention in most common cases.
Article
Full-text available
In laparoscopic surgery, image quality can be severely degraded by surgical smoke, which not only introduces error for the image processing (used in image guided surgery), but also reduces the visibility of the surgeons. In this paper, we propose to enhance the laparoscopic images by decomposing them into unwanted smoke part and enhanced part using a variational approach. The proposed method relies on the observation that smoke has low contrast and low inter-channel differences. A cost function is defined based on this prior knowledge and is solved using an augmented Lagrangian method. The obtained unwanted smoke component is then subtracted from the original degraded image, resulting in the enhanced image. The obtained quantitative scores in terms of FADE, JNBM and RE metrics show that our proposed method performs rather well. Furthermore, the qualitative visual inspection of the results show that it removes smoke effectively from the laparoscopic images.
Article
Full-text available
Age-related macular degeneration (AMD) remains a major cause of blindness, with dysfunction and loss of retinal pigment epithelium (RPE) central to disease progression. We engineered an RPE patch comprising a fully differentiated, human embryonic stem cell (hESC)–derived RPE monolayer on a coated, synthetic basement membrane. We delivered the patch, using a purpose-designed microsurgical tool, into the subretinal space of one eye in each of two patients with severe exudative AMD. Primary endpoints were incidence and severity of adverse events and proportion of subjects with improved best-corrected visual acuity of 15 letters or more. We report successful delivery and survival of the RPE patch by biomicroscopy and optical coherence tomography, and a visual acuity gain of 29 and 21 letters in the two patients, respectively, over 12 months. Only local immunosuppression was used long-term. We also present the preclinical surgical, cell safety and tumorigenicity studies leading to trial approval. This work supports the feasibility and safety of hESC-RPE patch transplantation as a regenerative strategy for AMD.
Article
Retinal Vein Occlusion (RVO) is a blinding disease caused by one or more occluded retinal veins. Current treatment methods only focus on symptom mitigation rather than targeting a solution for the root cause of the disorder. Retinal vein cannulation is an experimental eye surgical procedure which could potentially cure RVO. Its goal is to dissolve the occlusion by injecting an anticoagulant directly into the blocked vein. Given the scale and the fragility of retinal veins on one end and surgeons’ limited positioning precision on the other, performing this procedure manually is considered to be too risky. The authors have been developing robotic devices and instruments to assist surgeons in performing this therapy in a safe and successful manner. This work reports on the clinical translation of the technology, resulting in the world-first in-human robot-assisted retinal vein cannulation. Four RVO patients have been treated with the technology in the context of a phase I clinical trial. The results show that it is technically feasible to safely inject an anticoagulant into a 100μm100\,{\mu} {\rm m}-thick retinal vein of an RVO patient for a period of 10 min with the aid of the presented robotic technology and instrumentation. Full text available at https://rdcu.be/bVKgk
Article
Pose estimation methods for robotically guided magnetic actuation of capsule endoscopes have recently enabled trajectory following and automation of repetitive endoscopic maneuvers. However, these methods face significant challenges in their path to clinical adoption including the presence of regions of magnetic field singularity, where the accuracy of the system degrades, and the need for accurate initialization of the capsule's pose. In particular, the singularity problem exists for any pose estimation method that utilizes a single source of magnetic field if the method does not rely on the motion of the magnet to obtain multiple measurements from different vantage points. We analyze the workspace of such pose estimation methods with the use of the point-dipole magnetic field model and show that singular regions exist in areas where the capsule is nominally located during magnetic actuation. Since the dipole model can approximate most magnetic field sources, the problem discussed herein pertains to a wider set of pose estimation techniques. We then propose a novel hybrid approach employing static and time-varying magnetic field sources and show that this system has no regions of singularity. The proposed system was experimentally validated for accuracy, workspace size, update rate and performance in regions of magnetic singularity. The system performed as well or better than prior pose estimation methods without requiring accurate initialization and was robust to magnetic singularity. Experimental demonstration of closed-loop control of a tethered magnetic device utilizing the developed pose estimation technique is provided to ascertain its suitability for robotically guided capsule endoscopy. Hence, advances in closed-loop control and intelligent automation of magnetically actuated capsule endoscopes can be further pursued toward clinical realization by employing this pose estimation system.
Article
Background Malignant pleural effusion affects more than 750,000 persons each year across Europe and the United States. Pleurodesis with the administration of talc in hospitalized patients is the most common treatment, but indwelling pleural catheters placed for drainage offer an ambulatory alternative. We examined whether talc administered through an indwelling pleural catheter was more effective at inducing pleurodesis than the use of an indwelling pleural catheter alone. Methods Over a period of 4 years, we recruited patients with malignant pleural effusion at 18 centers in the United Kingdom. After the insertion of an indwelling pleural catheter, patients underwent drainage regularly on an outpatient basis. If there was no evidence of substantial lung entrapment (nonexpandable lung, in which lung expansion and pleural apposition are not possible because of visceral fibrosis or bronchial obstruction) at 10 days, patients were randomly assigned to receive either 4 g of talc slurry or placebo through the indwelling pleural catheter on an outpatient basis. Talc or placebo was administered on a single-blind basis. Follow-up lasted for 70 days. The primary outcome was successful pleurodesis at day 35 after randomization. Results The target of 154 patients undergoing randomization was reached after 584 patients were approached. At day 35, a total of 30 of 69 patients (43%) in the talc group had successful pleurodesis, as compared with 16 of 70 (23%) in the placebo group (hazard ratio, 2.20; 95% confidence interval, 1.23 to 3.92; P=0.008). No significant between-group differences in effusion size and complexity, number of inpatient days, mortality, or number of adverse events were identified. No significant excess of blockages of the indwelling pleural catheter was noted in the talc group. Conclusions Among patients without substantial lung entrapment, the outpatient administration of talc through an indwelling pleural catheter for the treatment of malignant pleural effusion resulted in a significantly higher chance of pleurodesis at 35 days than an indwelling catheter alone, with no deleterious effects. (Funded by Becton Dickinson; EudraCT number, 2012-000599-40.)