Electrical Structure of Biological Cells and Tissues:
impedance spectroscopy, stereology, and singular perturbation theory
Robert Eisenberg
Department of Molecular Biophysics
Rush University
Chicago IL
October 15, 2015
A chapter in
Impedance Spectroscopy
Theory, Experiment, and Applications:
Solid State, Corrosion, Power sources,
3rd Edition
Evgenij Barsoukov (ed.), J. Ross Macdonald (ed.),
Wiley-Interscience, 2016
Electrical Structure of Biological Cells and Tissues 1
Impedance Spectroscopy is a technique with exquisite resolution since it resolves linear
electrical properties into uncorrelated variables, as a function of frequency. This separation is robust
and most useful when the system being studied is linear (If the system is significantly nonlinear, the
meaning of the parameters is not clear, and they are likely to vary substantially with conditions, making
models badly posed.) Impedance spectroscopy can be combined with appropriate structural knowledge
(qualitative and quantitative) based on the actual anatomy of the system being studied. The
combination provides biologically useful insight into the pathways for current flow in a number of cells
and tissues, with more success than other methods.
Biological applications of impedance spectroscopy are often unable to take advantage of the
strengths of impedance spectroscopy since so much of biology is strongly nonlinear in its essential
features, and impedance spectroscopy is fundamentally a linear analysis. There is an important special
case, however, present in all cells and tissues in plants and animals, the cell membrane and its
capacitance, where bioelectric properties are both linear and important to cell function. Here we take
advantage of the ideal properties of membrane capacitance to make impedance spectroscopy a most
useful tool for determining the electrical structure of cells and tissues [18].
The capacitance of cell membranes is close to perfect, with 90 degree phase angle as close as
can be measured in the entire biologically relevant frequency range of DC to (say) 104 or even 105 Hz [21,
23, 26-28]. Impedance spectroscopy is a useful tool for studying cell membranes then, because linear
analysis is appropriate. Cell membranes define every cell in animals and plants and so impedance
spectroscopy is a tool that can be used throughout biology, exploiting the linear electrical properties of
this structure. Cell membranes determine a wide range of biological function and the organization of
these membranes into folded and more complex structures is a general motif in biology. Indeed, a
glance at the structure of a cell in any cell biology or histology text book shows that membranes inside
cells are extensively folded and organized as well. Impedance spectroscopy can help provide insight into
membrane structure and function throughout biology.
Cells are not simple spheres or cylinders. Rather, they include complex membrane structures
that channel electrical current to the regions of interest. These structures are important in epithelia that
form the kidney and gall bladder and secretory cells [6]. These structures are crucial in muscle fibers,
skeletal and cardiac, where electrical signals across membranes coordinate contraction. Without
coordination, skeletal muscle cannot provide useful force [61] [24]. Without coordination the heart
cannot pump blood [5]. Thus, the arrangement of membranes in skeletal muscle, and some cardiac
muscle, include tubules that are invaginations of the outer cell membrane [62]. The outer membrane is
a cylinder (roughly speaking) that sprouts tubular invaginations that bring current flow and electrical
signals into the interior of the muscle fiber. These signals, like ALL biological signals are voltages ACROSS
MEMBRANES. They allow communication of the propagating action potential of the surface membrane
to the depths of the muscle fiber [17].
The tubular system of muscle have an interior accessible to the solutions outside of cells.
Markers diffuse quickly into the T-system as it is called. Extracellular solutions are derived from salt
water and have simple linear electrical properties. Thus, the electrical properties of the T-system are
roughly those of a resistor (of the extracellular solution, more or less) in series with a membrane
capacitance, both actually resembling the linear electrical circuit elements used to represent them. The
capacitance is of course distributed along the series resistance of the resistance. The T-system
capacitance is distributed along the T-system luminal resistance.
The resistive properties of membranes in biology is hardly ever linear [29-32]. Even worse, it is
the nonlinearities that are interesting. Specialized proteins called ion channels, with a hole down their
middle, carry current [48, 63]. The channels open and close in response to stimuli, and so the currents
through the ensemble of channels in a cell membrane performs complex biological functions and is
anything but linear. Thus, studies of linear electrical properties of cells and tissues, including muscle and
Electrical Structure of Biological Cells and Tissues 2
its T-system, choose experimental conditions in which the number of channels open remains fixed, and
the resistive properties of cell membranes are linear. Fortunately, such conditions are easy to find and
enforce. If perturbing potentials are chosen that make the interior of the cell more negative than its
normal resting potential (of say -80 mv, inside minus outside), and not more than say 10 mv in
amplitude, the resistance of membranes is constant and the membrane system (of resistor and
capacitor in parallel) is a linear circuit element that can be well analyzed by impedance spectroscopy
[13-15, 18, 19, 39].
The linear properties of cells of course depend on the layout of the membranes and how they
are distributed along resistances [18]. Thus, to understand linear properties in physical terms one must
know the anatomy of the membranes, and one must know it quantitatively. How much membrane is
there? What are the dimensions of the T-tubules? How much outer membrane is there? How much
membrane is in the T-system? and so on. Without quantitative anatomy, one can only measure effective
phenomenological parameters of equivalent circuits. Equivalent circuits are equivalent to useful
representations only in the minds of physical scientists. Biologists want to know actual circuits with
parameters of definite structures and amounts of membrane. Neglect of anatomy, particularly
anatomical measurements, is a sad characteristic of many impedance measurements of biological
systems.
Anatomical measurements cannot be used in an electrical circuit however without a theory. To
put it baldly, the units are wrong! One must have a theory to connect the structures and their amounts
and sizes to the electrical measurements made by impedance spectroscopy. A theory is needed to
describe current flow and the role of structure. We see then that
Impedance spectroscopy of biological structures is a platform resting on three pillars.
The pillars of the platform are
(1) Impedance measurements themselves (with their technical requirements discussed below)
(2) Anatomical observations and measurements that provide quantitative information describing
the amount of membranes and their connections to each other.
(3) A theory (field theory or circuit theory or both) linking the electrical properties of tiny pieces of
membrane (differential elements, to be precise) and extra and intra cellular space with the
anatomy of the system to the impedance measurements themselves.
(4) Estimation procedures that allow the impedance measurements, anatomical observations, and
theories to evaluate properties of individual membrane systems.
This paper discusses those pillars in the context of work done long ago. The paper is meant to explain
and illustrate this use of impedance spectroscopy. The paper is a guide (I hope) to how the impedance
spectroscopy has been successfully applied to determine the electrical structure of several biological
systems of importance [18]. I apologize to the many more recent workers whom I have slighted out of
ignorance and hope that they nonetheless can find this story useful.
Anatomical measurements. The qualitative understanding of anatomical structures underlies any
quantitative measurements of those structures, and any electrical models and theories tying those
structures to impedance measurements. Humans are visual animals so the unconscious visualization of
structure is natural for all of us, particularly for the subset who are visually gifted. Anatomy is an ancient
subject for that reason, and only rarely is our (unconscious) visualization qualitatively incorrect, and
then usually when dealing with sectioned (i.e., sliced) material. When a three dimensional structure is
sliced into thin sections, reconstruction into the three dimensional structure can (and has) occasionally
confused spheres and cylinders because they are both circles in slices. But in general we begin with the
presumption that the qualitative anatomy of the system is known through the centuries of effort of
anatomists and histologists, who are anatomists on a smaller scale, and cell and structural biologists,
Electrical Structure of Biological Cells and Tissues 3
who are anatomists on a still smaller scale reaching nowadays to the atoms that make up the structure
of membranes, channels, and proteins in general.
The obvious way to measure an anatomical structure has been used ‘forever’. Slice it into
sections, trace out the structure, and measure the tracing. This approach is fraught with danger. As is
obvious, the tissue must be fixed and embedded in a hard sliceable material before it can be sectioned
and those steps must be controlled to have minimal artifact. All this is the stuff of classical histology and
is more or less under control [10]. All of this can be improved or replaced by more modern three
dimensional imaging techniques, if those are used to make QUANTITATIVE calibrated measurements of
membrane area and so on.
What is often overlooked is that the statistical sampling methods called stereology provide far
less biased estimates of crucial membrane parameters (e.g., membrane area) than tracing methods.
These stereological methods are of course not the subject of this article, but their existence is so often
overlooked over the years that I must point them out [8, 9]. It is easy to show in experiments on
computer generated images that stereological methods are fast and accurate, whereas tracing methods
are slow and badly biased. Slow because errors abound. Biased because humans and machines trace in
ways that seem natural but are in fact deterministically skewed.
Stereology depends on the projection of an image onto a grid (usually rectangular) of LOW
resolution with spacing between lines much larger than one might guess, and very much larger than that
necessary to resolve the details of the membranes, or even their location. Resolution has to be enough
only---I repeat ONLY---to allow identification of the grid point: is it inside or is it outside the structure?
The loss of information in any one image is very large. BUT A LARGE NUMBER (say 500) of images of
sections are needed IN ANY CASE to sample the three dimensional structure successfully. When the low
resolution sampling is done of these number of images, the resulting ensemble estimates in fact have
(more than) acceptable variance. And most importantly they have much less bias than tracing methods.
It might be thought that making this number of observations is unacceptably tedious, but in fact the
experience of many investigators show the tedium and work involved is much less than in say typical
experiments of biophysics or molecular biology.
There are many modern visualization methods that allow direct measurement of tiny membrane
structures, including the tubular systems of muscle, in unfixed material. These measurements are very
much more reliable than measurements made in fixed and sliced material, because the fixation process
introduces large, unknown, and irreproducible changes in volume
However, the stereological methods developed long ago remain very good, and probably the
best method to measure membrane area. Membrane area is not disturbed by fixation; membranes and
their area cannot be directly visualized by high resolution microscopy. Of course, methods that tag
membranes uniformly and observe the density of the tag in high resolution microscopy would be better
yet.
Impedance Measurements. Impedance measurements in biology are most helpful when the impedance
being measured is of a system of direct biological importance. When current is applied outside cells, to a
whole tissue or to a suspension of cells, the impedance measured is mostly that of the extracellular
space. Properties of the extracellular space are not central to most biological function although of
course they can be useful assays of malfunction (in clinical medicine) and useful adjuncts to more
meaningful measurements.
Impedance measurements discussed here force the applied current to flow across membranes,
through membrane capacitance and ion channels that form a conductive pathway in parallel with the
membrane capacitance [11, 22, 41, 64, 67-69]. Current is forced to flow in these paths because current
is applied to a tiny probe inserted into a cell, in classical experiments, or in current applied inside cells by
the patch clamp method [25]. The current flows out of the probe through the cytoplasm of the cell,
Electrical Structure of Biological Cells and Tissues 4
across the cell membrane, which is by far the largest impedance to flow, through specialized structures
just outside the membrane, and then into the extracellular space and the electrode in the extracellular
space, the bath electrode, which forms the other side of the current injection circuit.
With care (and some luck) probes can be inserted into most cells larger than say 10-5 m in
diameter, and that includes most cardiac and skeletal muscle, although many cells do not fall into this
category. The damage and perturbation produced by one probe can be assessed by inserting one probe
for recording and assessment and then inserted another probe. These errors are kept well below 2% in
the successful experiments discussed here.
Measurements of impedance are made in the biologically relevant frequency range of say 0.1 Hz
to say 10,000 Hz where most biological function occurs. Originally, measurements were made using
sinusoids and indeed much impedance spectroscopy is still done that way. Reading the engineering
literature [4, 49] showed clearly, however, that measurements could be made much more quickly if the
sum of sinusoids was used as the perturbing signal [39]. Since biological cells impaled with probes are
dying preparations (typically lasting only an hour or so), it is important to make measurements as rapidly
as possible. The rule of thumb in the engineering literature (which is easy to verify with simulations or
breadboard experiments) is that the time to take a broadband measurement with the optimal sum of
sinusoids is determined by the lowest frequency sinusoid. The higher frequency measurements come for
free, in the sense that they do not take (significant) extra time, although of course the equipment to
perform such measurements is anything but free!
This is not the place to derive or justify these statements. They have been textbook material for
more than fifty years [4, 49], even if not widely known outside electrical engineering. The sum of
sinusoids mentioned above is made with random phase (to keep the amplitude of the sum under
control) and so the signals are stochastic requiring some knowledge of stochastic signal processing. This
is knowledge that is not wide spread or trivial to master, so it is understandable that stochastic signals
have not been widely used. They have great advantages nonetheless [39].
One should point out a particular artifact that is easy to make in stochastic analysis. It is
tempting, even seductive to estimate a transfer function by taking a numerical estimate of a Fourier
transform of the output of a system and dividing it (in the complex domain) by a numerical estimate of
the Fourier transform of the input. This estimate however is so gravely flawed that is not usable, in the
presence of essentially any nonlinearity or contaminating noise, as explained in great detail in the books
of Otnes and Enochson [4, 14, 49]. If this estimation procedure is tried on a mock circuit, with added
noise, one will find that the mean value of the estimate changes with the noise level, but the variance of
the estimate does not. The mock circuit of course does not change with the level of added noise, so the
qualitative properties of the estimate are wrong. If one fails to check the procedure on a mock circuit,
the error can be overlooked because most workers look for variance as a sign of noise contamination,
and in this case the sign of noise contamination is a bias in the mean value withOUT an increase in noise.
Measurement Difficulties. Probes small enough to insert into cells with acceptable damage and
reproducibility are small and have high impedance even if they are micropipettes filled with nearly
saturated salt (e.g., 3M KCl). At frequencies of 104 Hz, stray capacitances of even 0.1 pF are important.
Thus great care must be taken in the construction of setups, analysis and treatment of errors, and
calibration. This process was begun in [11, 22, 64], and extended dramatically in [41, 67-69], where
attention to detail and use of a new recording amplifier allowed a calibrated bandwidth of at least 105
Hz, as checked in an actual setup making biological recording. The key was the use of an inverting
amplifier circuit in which the current through the micropipette is measured as a replica of the potential
within the cell. The resistance of the micropipette is quite pure (because it is lumped in a tiny part of the
micropipette, just a few micrometers long). This circuit has optimal noise performance as well, which
Electrical Structure of Biological Cells and Tissues 5
was later recognized and exploited brilliantly by Neher and Sigworth [65, 66], in the electronics which
Neher used in his patch clamp work.
Calibration was performed by injection of current through an AIR capacitor: a physical capacitor
proved unreliable because even finger prints seriously changed the phase angle of current through it
(because of the parallel conductance of finger grease). These realities proved to be the key to the later
development of the integrating head state which has provided the optimal electronics for recording the
picoamps of current through single channels in the AxoPatch amplifier [34-36, 56, 59] still the leading
amplifier in 2015 after nearly 30 years on the market, now sold by Molecular Devices Inc.
Future Measurements. The much lower impedance (by a factor of at least 5 times, without further
fiddling) of patch pipettes should allow much wider bandwidth recording which might, or might not
prove to lead to more biological insight. One cannot be sure until one addresses biological questions
specific to a particular cell, tissues, and preparation.
Interpreting Impedance Spectroscopy. The results just mentioned all depend on the interpretation of
the impedance spectrum. This interpretation requires a specific theory linking the measurement to the
structure of the system in question, including the experimental setup and artifacts in the theory, and
measurements of the anatomical parameters of the structures involved.
These requirements are often ‘swept under the rug’ because they require extensive work to
fulfill by using an ‘equivalent circuit’ to replace the theory. The equivalent circuit is an essential part of
the analysis to be sure, and the insight required to understand its simplifications prove to be most
useful, but it is not enough to allow the full use of the wonderful resolution of impedance
measurements.
The first step in extending the equivalent circuit analysis is the understanding of how the
equivalent circuit approximates the electrodynamics of the biological structure and recording apparatus.
The artifacts of the recording apparatus have been discussed above and enumerated and analyzed
definitively (it seems to me in my prejudiced opinion) in the literature, and so will not be discussed
further here [12, 68].
The relation of the equivalent circuit to the electrodynamics of the system was studied by
solving Poisson’s equation by several authors, but the simplification to the usual equivalent circuit only
became apparent when series expansions of the solution of the Poisson equation, and the Poisson
equation itself were developed[3, 12, 16, 33, 51-54].
These expansions provided wonderful insight [51]: the usual equivalent circuit (of a uniform
potential for a finite cell, like a sphere, or of a transmission line for a long cylinder) arose as a separate
term from the point source effects (i.e., the terms that vary with angular coordinates). Indeed, the
beauty and power of the analysis (by matched asymptotic expansions for example) led applied
mathematicians to use this work as a teaching example of the power of singular perturbation theory.
Fitting Data Once the impedance spectroscopy data is available (as plots of amplitude and phase vs
frequency, for example) it is necessary to extract the parameters of the model that fits the data. The
least squares fitting procedure proved robust and fast, even with ancient computers of the 1970’s, and
able to do the job. In fact, fitting to biological data was one of the first applications of the Brown Denis
version of the Levenberg Marquardt minimization technique, now widely used and enshrined in
MATLAB, thanks to the help of many coworkers.
Just as important as fitting the data, is determining the significance of the parameter estimates.
The issue is not just what is the statistical variance of a parameter estimate. Here there is an additional
problem because of the correlation between parameter estimates. Perhaps changing two parameters
(or more) might produce a fit nearly as good as the one chosen. This issue can be addressed by fixing
one parameter at a different value from its optimal, and then comparing the quality of fit, using the F-
Electrical Structure of Biological Cells and Tissues 6
variance ratio test (which is quite robust) in (for example) the form of the R-test used widely in
crystallography.
Results. Measurements proved straightforward in skeletal muscle, cardiac muscle, tendon regions of
skeletal muscle, and lens of the eye [6, 7, 11, 18, 22, 37, 38, 41-43, 46, 57, 64, 67-69]. In each case the
measurements of impedance provided useful information concerning the tissue. In skeletal muscle,
measurements provided the best estimates of the predominant (cell) membrane system that dominates
electrical properties. In cardiac muscle, measurements showed definitively that classical microelectrode
voltage clamp could not control the potential of the predominant membranes, that were in the tubular
system separated from the extracellular space by substantial distributed resistance. In the tendon
regions of skeletal muscle, electrical properties needed to interpret extensive voltage clamp analysis of
the Cambridge (UK) group were measured [1]. In the lens of the eye, impedance spectroscopy changed
the basis of all recording and interpretation of electrical measurements and laid the basis for Rae and
Mathias’ extensive later experimental work [2, 19, 40, 42-45, 50, 55, 57-60].
Many tissues are riddled with extracellular space as clefts and tubules, for example, cardiac
muscle, the lens of the eye, most epithelia, and of course frog muscle. These tissues are best analyzed
with a bi-domain theory [20, 42, 43] that arose from the work on electrical structure described here.
There has been a great deal of work since then on the bi-domain model admirably reviewed in [47] and
this represents the most important contribution to biology of the analysis of electrical structure in my
view.
Future Perspectives: Science is subject to fashions like all human behavior. The electrical structure of
cells and tissues is out of fashion as biologists move to understand the ion channels so important for
biological function. These channels function in a biological environment determined in large measure by
the electrical structure of the tissues in which they are found. Computational tools are available to
compute the properties of channels in the electrical structures in which they are found. Impedance
spectroscopy will come back into fashion as scientists realize there are few other techniques with the
resolution necessary to determine the electrical structure and how it modulates the current flows driven
by ion channels.
Electrical Structure of Biological Cells and Tissues 7
References
1. Adrian, R.H., W.K. Chandler, and A.L. Hodgkin (1966) Voltage clamp experiments in skeletal
muscle fibres. J Physiol 186(2): p. 51P-52P.
2. Baldo, G.J. and R.T. Mathias (1992) Spatial variations in membrane properties in the intact rat
lens. Biophys J 63(2): p. 518-529.
3. Barcilon, V., J. Cole, and R.S. Eisenberg (1971) A singular perturbation analysis of induced electric
fields in nerve cells. SIAM J. Appl. Math. 21(2): p. 339-354.
4. Bendat, J.S. and A.G. Piersol, (2011) Random data: analysis and measurement procedures. Vol.
729. 2011: John Wiley & Sons.
5. Bers, D.M. (2002) Cardiac excitation-contraction coupling. Nature 415(6868): p. 198-205.
6. Clausen, C., S.A. Lewis, and J.M. Diamond (1979) Impedance analysis of a tight epithelium using
a distributed resistance model. Biophysical Journal 26(2): p. 291-317.
7. Ebihara, L. and R.T. Mathias (1985) Linear impedance studies of voltage-dependent
conductances in tissue cultured chick heart cells. Biophys J 48(3): p. 449-460.
8. Eisenberg, B.R., (1979) Skeletal muscle fibers: stereology applied to anisotropic and periodic
structures, in Stereological Methods: Practical methods for biological morphometry, E. Weibel,
Editor. p. 274-284.
9. Eisenberg, B.R., (1983) Quantitative ultrastructure of mammalian skeletal muscle, in Handbook
of Physiology, L.D. Peachey and R.H. Adrian, Editors. American Physiological Society Bethesda
MD. p. 73-112.
10. Eisenberg, B.R. and B.A. Mobley (1975) Size changes in single muscle fibers during fixation and
embedding. Tissue and Cell 7: p. 383-387.
11. Eisenberg, R.S. (1967) The equivalent circuit of single crab muscle fibers as determined by
impedance measurements with intracellular electrodes. J Gen Physiol 50(6): p. 1785-1806.
12. Eisenberg, R.S. (1967) Equivalent circuit of single crab muscle fibers as determined by
impedance measurement with intracellular electrodes. J. Gen. Physiol. 50: p. 1785-1806.
13. Eisenberg, R.S. (1980) Structural complexity, circuit models, and ion accumulation. Fed Proc
39(5): p. 1540-1543.
14. Eisenberg, R.S., (1983) Impedance Measurement of the Electrical Structure of Skeletal Muscle, in
Handbook of Physiology, Section 10: Skeletal Muscle, L.D. Peachey, Editor. American
Physiological Society: Washington DC. p. 301-323.
15. Eisenberg, R.S. (1987) Impedance measurements as estimators of the properties of the
extracellular space. Ann. NY Acad. Sci. 481: p. 116-122.
16. Eisenberg, R.S. and E.A. Johnson (1970) Three dimensional electrical field problem in physiology.
Prog. Biophys. Mol. Biol. 20: p. 1-65
17. Eisenberg, R.S. and L.L. Costantin (1971) The radial variation of potential in the transverse
tubular system of skeletal muscle. J Gen Physiol 58(6): p. 700-701.
18. Eisenberg, R.S. and R.T. Mathias (1980) Structural analysis of electrical properties. Critical
Reviews in Bioengineering 4: p. 203-232.
19. Eisenberg, R.S., R.T. Mathias, and J.S. Rae (1977) Measurement, modeling, and analysis of the
linear electrical properties of cells. Ann N Y Acad Sci 303: p. 342-354.
20. Eisenberg, R.S., V. Barcilon, and R.T. Mathias (1979) Electrical properties of spherical syncytia.
Biophys J 25(1): p. 151-180.
21. Everitt, C.T. and D.A. Haydon (1968) Electrical capacitance of a lipid membrane separating two
aqueous phases. J Theor Biol 18(3): p. 371-379.
Electrical Structure of Biological Cells and Tissues 8
22. Falk, G. and P. Fatt (1964) Linear Electrical Properties of Striated Muscle Fibres Observed with
Intracellular Electrodes. Proc R Soc Lond B Biol Sci 160: p. 69-123.
23. Fettiplace, R., D.M. Andrews, and D.A. Haydon (1971) The thickness, composition and structure
of some lipid bilayers and natural membranes. J Membr Biol 5(3): p. 277-296.
24. Fill, M. and J.A. Copello (2002) Ryanodine Receptor Calcium Release Channels. Physiol Rev
82(4): p. 893-922.
25. Hamill, O.P., A. Marty, E. Neher, B. Sakmann, and F.J. Sigworth (1981) Improved patch-clamp
techniques for high-resolution current recording from cells and cell-free membrane patches.
Pflugers Arch 391(2): p. 85-100.
26. Hanai, T., D.A. Haydon, and J. Taylor (1965) The variation of capacitance and conductance of
bimolecular lipid membranes with area. J Theor Biol 9(3): p. 433-443.
27. Hanai, T., D.A. Haydon, and J. Taylor (1965) The influence of lipid composition and of some
adsorbed proteins on the capacitance of black hydrocarbon membranes. J Theor Biol 9(3): p.
422-432.
28. Haydon, D.A., J. Requena, and B.W. Urban (1980) Some effects of aliphatic hydrocarbons on the
electrical capacity and ionic currents of the squid giant axon membrane. J Physiol 309: p. 229-
245.
29. Hodgkin, A.L., (1992) Chance and Design. 1992, New York: Cambridge University Press. 401.
30. Huxley, A. (1999) Hodgkin Obituary. The Independent (newspaper) January 4, 1999
http://www.independent.co.uk/arts-entertainment/obituaries-professor-sir-alan-hodgkin-
1044924.html.
31. Huxley, A.F. (1992) Kenneth Stewart Cole. Biographical Memoirs of Fellows of the Royal Society
38: p. 98-110 , see http://books.nap.edu/html/biomems/kcole.pdf
32. Huxley, A.F. (2002) From overshoot to voltage clamp. Trends Neurosci 25 (11): p. 553-558.
33. Kevorkian, J. and J.D. Cole, (1996) Multiple Scale and Singular Perturbation Methods. 1996, New
York: Springer-Verlag. pp. 1-632.
34. Levis, R.A. and J.L. Rae, (1992) Constructing a patch clamp setup., in Methods Enzymol, L.
Iverson and B. Rudy, Editors. Academic Press: NY. p. 66-92.
35. Levis, R.A. and J.L. Rae, (1995) Technology of patch clamp recording electrodes, in Patch-clamp
Applications and Protocols, W. Walz, A. Boulton, and G. Baker, Editors. Humana Press.: Totowa,
NJ.
36. Levis, R.A. and J.L. Rae (1998) Low-noise patch-clamp techniques. Methods Enzymol 293: p.
218-266.
37. Levis, R.A., R.T. Mathias, and R.S. Eisenberg (1983) Electrical properties of sheep Purkinje
strands. Electrical and chemical potentials in the clefts. Biophys J 44(2): p. 225-248.
38. Mathias, R.T. (1983) Effect of tortuous extracellular pathways on resistance measurements.
Biophys J 42(1): p. 55-59.
39. Mathias, R.T., (1984) Analysis of Membrane Properties Using Extrinsic Noise, in Membranes,
Channels, and Noise, R. Eisenberg, M. Frank, and C. Stevens, Editors. Springer US. p. 49-116.
40. Mathias, R.T. and J.L. Rae (1985) Transport properties of the lens. Am J Physiol 249(3 Pt 1): p.
C181-190.
41. Mathias, R.T., R.S. Eisenberg, and R. Valdiosera (1977) Electrical properties of frog skeletal
muscle fibers interpreted with a mesh model of the tubular system. Biophys J 17(1): p. 57-93.
42. Mathias, R.T., J.L. Rae, and R.S. Eisenberg (1979) Electrical properties of structural components
of the crystalline lens. Biophys J 25(1): p. 181-201.
43. Mathias, R.T., J.L. Rae, and R.S. Eisenberg (1981) The lens as a nonuniform spherical syncytium.
Biophys J 34(1): p. 61-83.
Electrical Structure of Biological Cells and Tissues 9
44. Mathias, R.T., J.L. Rae, and G.J. Baldo (1997) Physiological properties of the normal lens. Physiol
Rev 77(1): p. 21-50.
45. Mathias, R.T., J.L. Rae, L. Ebihara, and R.T. McCarthy (1985) The localization of transport
properties in the frog lens. Biophys J 48(3): p. 423-434.
46. Milton, R.L., R.T. Mathias, and R.S. Eisenberg (1985) Electrical properties of the myotendon
region of frog twitch muscle fibers measured in the frequency domain. Biophys J 48(2): p. 253-
267.
47. Mori, Y. (2015) A multidomain model for ionic electrodiffusion and osmosis with an application
to cortical spreading depression. Physica D: Nonlinear Phenomena 308: p. 94-108.
48. Neher, E., (1997) Ion channels for communication between and within cells Nobel Lecture,
December 9, 1991, in Nobel Lectures, Physiology or Medicine 1991-1995, N. Ringertz, Editor.
World Scientific Publishing Co: Singapore. p. 10-25.
49. Otnes, R.K. and L. Enochson, (1972) Digital Time Series Analysis. Robert K. Otnes (Author), Loren
Enochson (1972, New York: John Wiley. 480.
50. Pasquale, L.R., R.T. Mathias, L.R. Austin, P.R. Brink, and M. Ciunga (1990) Electrostatic properties
of fiber cell membranes from the frog lens. Biophys J 58(4): p. 939-945.
51. Peskoff, A. and R.S. Eisenberg (1973) Interpretation of some microelectrode measurements of
electrical properties of cells. Annu Rev Biophys Bioeng 2: p. 65-79.
52. Peskoff, A. and D.M. Ramirez (1975) Potential induced in a spherical cell by an intracellular point
source and an extracellular point sink. Journal of Mathematical Biology 2(4): p. 301-316.
53. Peskoff, A. and R.S. Eisenberg (1975) The time-dependent potential in a spherical cell using
matched asymptotic expansions. Journal of Mathematical Biology 2(The time-dependent
potential in a spherical cell using matched asymptotic expansions): p. 277-300.
54. Peskoff, A., R.S. Eisenberg, and J.D. Cole (1976) Matched asymptotic expansions of the Green's
function for the electric potential in an infinite cylindrical cell. SIAM J. Appl. Math. 30(2): p. 222-
239.
55. Rae, J.L. (1979) The electrophysiology of the crystalline lens. Curr Top Eye Res 1: p. 37-90.
56. Rae, J.L. and R.A. Levis (1992) A method for exceptionally low noise single channel recordings.
Pflugers Arch 420(5-6): p. 618-620.
57. Rae, J.L., R.T. Mathias, and R.S. Eisenberg (1982) Physiological role of the membranes and
extracellular space with the ocular lens. Exp Eye Res 35(5): p. 471-489.
58. Rae, J.L., R.D. Thomson, and R.S. Eisenberg (1982) The effect of 2-4 dinitrophenol on cell to cell
communication in the frog lens. Exp Eye Res 35(6): p. 597-609.
59. Rae, J.L., R.A. Levis, and R.S. Eisenberg, (1988) Ionic channels in ocular epithelia, in Ion Channels,
T. Narahashi, Editor. Plenum: New York. p. 283-327.
60. Rae, J.L., C. Bartling, J. Rae, and R.T. Mathias (1996) Dye transfer between cells of the lens. J
Membr Biol 150(1): p. 89-103.
61. Ríos, E. and G. Pizarro (1991) Voltage sensor of excitation-contraction coupling in skeletal
muscle. Physiol Rev 71(3): p. 849-908.
62. Russell, B., (1995) Striated Muscle, in Gray's Anatomy. Churchill Livinstone London.
63. Sakmann, B. and E. Neher, (1995) Single Channel Recording. Second ed. 1995, New York:
Plenum. 700.
64. Schneider, M.F. (1970) Linear electrical properties of the transverse tubules and surface
membrane of skeletal muscle fibers. J Gen Physiol 56(5): p. 640-671.
65. Sigworth, F., (1995) Electronic Design of Patch Clamp, in Single Channel Recording, B. Sakmann
and E. Neher, Editors. Plenum: New York. p. 95-128.
66. Sigworth, F.J. and E. Neher (1980) Single Na+ channel currents observed in cultured rat muscle
cells. Nature 287(5781): p. 447-449.
Electrical Structure of Biological Cells and Tissues 10
67. Valdiosera, R., C. Clausen, and R.S. Eisenberg (1974) Circuit models of the passive electrical
properties of frog skeletal muscle fibers. J. Gen. Physiol. 63: p. 432-459.
68. Valdiosera, R., C. Clausen, and R.S. Eisenberg (1974) Measurement of the impedance of frog
skeletal muscle fibers. Biophys. J. 14: p. 295-314.
69. Valdiosera, R., C. Clausen, and R.S. Eisenberg (1974) Impedance of frog skeletal muscle fibers in
various solutions. J Gen Physiol 63(4): p. 460-491.