ChapterPDF Available

New Approach of Measuring Toxic Gases Concentrations: Principle of Operation

Authors:

Abstract

Measurement of the concentrations of toxic gases (such as CO, SO2, NOX and O3) in the air represent important but challenging task. Traditionally, these concentrations are measured by means of heavy and expensive toxic gas monitors, usually operated by the governmental institutions. In this paper we present a novel method based on commodity (lowcost) electrochemical sensors and in-house developed data acquisition system. Principle of operation is explained, together with the calibration procedure and initial evaluation of sensors’ performance. Possible applications are discussed as well. Keywords: air pollution; toxic gas concentration; electrochemical sensor; temperature drift; amperometric cell
29TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
DOI: 10.2507/29th.daaam.proceedings.127
NEW APPROACH OF MEASURING TOXIC GASES
CONCENTRATIONS: PRINCIPLE OF OPERATION
Adnan Masic, Dzevad Bibic, Boran Pikula & Faruk Razic
This Publication has to be referred as: Masic, A[dnan]; Bibic, D[zevad]; Pikula, B[oran] & Razic, F[aruk] (2018). New
Approach of Measuring Toxic Gases Concentrations: Principle of Operation, Proceedings of the 29th DAAAM
International Symposium, pp.0882-0887, B. Katalinic (Ed.), Published by DAAAM International, ISBN 978-3-902734-
20-4, ISSN 1726-9679, Vienna, Austria
DOI: 10.2507/29th.daaam.proceedings.127
Abstract
Measurement of the concentrations of toxic gases (such as CO, SO2, NOX and O3) in the air represent important but
challenging task. Traditionally, these concentrations are measured by means of heavy and expensive toxic gas monitors,
usually operated by the governmental institutions. In this paper we present a novel method based on commodity (low-
cost) electrochemical sensors and in-house developed data acquisition system. Principle of operation is explained,
together with the calibration procedure and initial evaluation of sensors’ performance. Possible applications are discussed
as well.
Keywords: air pollution; toxic gas concentration; electrochemical sensor; temperature drift; amperometric cell
1. Introduction
According to the World Health Organization (WHO), 4.2 million people die every year due to the ambient air pollution
and 3.8 million more as a result of household exposure to smoke from dirty cookstoves and fuels, while 91% of the
world’s population lives in places where air quality exceeds WHO guideline limits [1]. An important step towards the
solution of the problem of air pollution is monitoring of the concentrations of toxic gases in the air. However, vast majority
of the population affected by the excessive air pollution live in developing countries [2]. Hence, affordable and reliable
methods for monitoring of toxic gases are urgently needed. A very promising approach is to use commodity
electrochemical sensors [3], recently developed by Alphasense Ltd (UK).
Alphasense toxic gas sensors are electrochemical cells that operate in the amperometric mode. They generate a current
that is linearly proportional to the fractional volume of the toxic gas [4]. Electrochemical amperometric gas sensors have
a background current in addition to the current from oxidation or reduction of the sampled gas. This background current
is commonly called the zero current. These zero currents can be significant in the measurements at low gas concentrations.
Sources of these zero currents can include anodisation or cathodic reduction of the working electrode (WE),
electrochemical oxidation or reduction of the sensor electrolyte or electrolyte contaminants by the WE and reduction of
oxygen in the sampled air [5].
- 0882 -
29TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
In order to correct for zero currents, the Alphasense sensors include an additional electrode, an auxiliary electrode
(AE), buried within the sensor. The AE has the same catalyst structure as the working electrode (WE) so, in principle,
since it is not in contact with the sampled gas, any background current arising from solid electrode processes or from
electrochemistry involving the electrolyte will be measured on both the WE and the AE. This AE current could then be
subtracted from the WE total current to give a corrected WE current corresponding solely to the electrochemical reaction
of the sampled gas [5].
2. The test setup
We have acquired and analysed the following Alphasense sensors:
- SO2-B4, range (0-100) ppm,
- CO-A4, range (0-500) ppm,
- NO-A4, range (0-20) ppm,
- NO2-A4, range (0-20) ppm,
- OX-A4, range (0-20) ppm.
Sensor SO2-B4 was installed on Alphasense ISB (Individual Sensor Board), while A-type sensors were combined on
one 4-way AFE (Analogue Front End) board from Alphasense. ISB and AFE boards contain electronic components which
convert small currents from electrochemical cells to the buffered voltage (0-5) V, which is more suitable for interfacing
and integration into DAS (Data Acquisition System). Our DAS is in-house developed and based on previous work
described in [6], [7] and [8].
Taking into account voltage range and typical sensor sensitivities, we arrive at conclusion that at least 16-bit ADC
(Analogue to Digital Converter) is necessary. We have chosen ADS1115 from Texas Instruments (USA) as suitable ADC
for Alphasense sensor boards. This is 4-channel 16-bit ADC with I2C interface, which can be easily connected to various
microcontrollers, such as Arduino. The Figure 1 shows simplified schematics of our platform. Since we have 5
electrochemical sensors, we need at least 10 ADC channels (working and auxiliary electrode voltage for each sensor).
Thus, three ADS1115 devices are required. Data storage is flash memory card, which communicates with the Arduino
over SPI bus. The OX sensor reacts to ozone (O3) and nitrogen dioxide (NO2); together with NO2 sensor, we can calculate
results for O3 alone.
Fig. 1. Simplified schematics
3. Results
Five electrochemical sensors, together with electronic boards and DAS as described above, were installed at the
Federal Institute for Hydrometeorology in Sarajevo, Bosnia-Herzegovina, in order to assess sensors’ performance and
key characteristics. System was continuously running from 18-05-2018 until 22-06-2018. One minute average values
from the sensors were logged to the flash memory. Primary goal of the experiment was investigation of temperature drifts
of both working and auxiliary electrode and noise levels. The Figure 2 shows auxiliary electrode voltage (one minute
average) vs. temperature for all sensors. We can see that SO2 and NO boards show parabolic-like temperature dependence
in the interval (15-35) °C (second-order polynomial was fitted in the corresponding graphs), while other sensors do not
show clear trend. Assessing the working electrode was much more difficult, since it reacts to target gas, whose
concentration is not known. However, reference monitors were available at the Federal Institute for Hydrometeorology,
so we could detect periods of clean air, with near-zero concentration of the target gas. Figure 3 shows working electrode
zero values vs. temperature for all sensors.
- 0883 -
29TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
a)
b)
c)
d)
e)
Fig. 2. Temperature drift of auxiliary electrodes of used sensors for SO2, NO, NO2, CO, OX
This time SO2 sensor follows linear trend of dependence of working electrode zero vs. temperature, while NO sensor
follows polynomial fit of second order. This statement is not firm (as in the case of auxiliary electrode), since the zero
concentration of target gas couldn’t be verified. Other sensors do not show clear trends, again.
To conclude this part, it seems that for room temperatures, (20-25) °C, the working electrode zero and the auxiliary
electrode zero may be taken as constants for each sensor (with board). In the wider temperature range, corrections for SO2
and NO may be necessary, though. By looking at the schematics in the Figure 1 (and knowing the nature of amperometric
cells), it is clear that the noise would be a great problem. Before we present new graphs, let’s clarify how concentration c
of the target gas is actually calculated. We measure two voltages from each sensor/board: working electrode voltage (UWE)
and auxiliary electrode voltage (UAE). We must subtract zero values, UWE0 and UAE0 and divide the difference by
sensitivity of the sensor s:
- 0884 -
29TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
(1)
Using equation (1), we get volume concentrations (measured in ppb). If we want the mass concentration cm (in g/m3),
we can use the following conversion:
𝑐𝑚= 𝑐 ∙ 𝑝𝑀
𝑅𝑇
(2)
where p is the pressure, M is the molar mass, R is the universal gas constant and T is the temperature.
a)
b)
c)
d)
e)
Fig. 3. Temperature drift of working electrode of used sensors for SO2, NO, NO2, CO, OX
- 0885 -
29TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
According to the manufacturers recommendation, we have soldered capacitors of 10 nF and 100 nF close to the power
supply connectors of both ISB and AFE boards, in order to minimize noise injection (Figure 4).
Fig. 4. Decoupling capacitors on AFE and ISB boards with sensors
Despite this effort, noise appears on all measurements, to some extent. The Figure 5 shows almost instantaneous
readings (sampling period was 3 seconds) from one typical run of SO2 sensor. Noise of few mV in the signal from
electrodes exists, which translates into few tens of ppb in volume concentration. Black line in the Figure 5 is smoother
(moving average trend line with 15 samples per move). Noise may be minimized using combination of hardware (filters)
and software (processing of the signal), but it turns out that these sensors are not suitable for very low concentrations of
target gas.
Fig. 5. Noise on SO2 sensor
4. Conclusion
Electrochemical amperometric sensors, such as Alphasense 4-electrode sensor family, together with appropriate
interface and data acquisition system, represent promising, new technology for measuring the concentrations of toxic
gases in the air. However, their use is not straight-forward. Each sensor must be (individually) calibrated. Signals from
the electrodes should be corrected for (non-trivial) temperature drifts, sensitivities of individual sensors are also
temperature dependant and decay over time. There is also cross-sensitivity problem [9], which is difficult to deal with.
The noise in the signals from electrodes (amplified using circuit boards) are especially problematic in case of low
concentrations of toxic gases.
As a suggestion for further research we propose:
- Application of electrochemical sensors to real-life problems, such as air pollution in underground garages or real-
time street measurements,
- Investigation of sensitivity of the sensor over longer periods,
- Investigation and modelling of the cross-sensitivity.
0
10
20
30
40
50
60
70
80
0 200 400 600 800 1000
SO2, ppb
Time, s
- 0886 -
29TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION
5. Acknowledgments
This research was supported by the Ministry of Education of Sarajevo Canton and the University of Sarajevo - Faculty
of Mechanical Engineering. We would like to express gratitude to the Federal Institute for Hydrometeorology as well.
The Last but not the least, we would like to thank to Mr. Damir Muslic for his generous contribution to our project by
writing core part of the in-house developed software.
6. References
[1] http://www.who.int/airpollution/en/, (2018). World Health Organization, Accessed on: 2018-09-23
[2] Kumar, R. (2018). Five steps to improve air-quality forecasts. Nature, 561 (2018) 27-29
[3] Mead, M. I.; Popoola, O. A. M.; Stewart, G. B.; Landshoff, P.; Calleja, M.; Hayes, M.; Baldovi, J. J.; McLeod M.
W.; Hodgson T. F.; Dicks J.; Lewis A.; Cohen J.; Baron R.; Saffell J. R. & Jones, R. L. (2013). The use of
electrochemical sensors for monitoring urban air quality in low-cost, high-density networks. Atmospheric
Environment, 70 (2013) 186-203
[4] How electrochemical gas sensors work (2013). Alphasense Application Note AAN 104
[5] Correcting for background currents in four electrode toxic gas sensors (2017). Alphasense Application Note AAN
803-04
[6] Masic, A. (2015).Unmanned Aerial Vehicle as Data Acquisition System. Journal of Trends in the Development of
Machinery and Associated Technology. Vol. 19, No. 1, 2015, ISSN 2303-4009 (online), 181-184
[7] Masic, A.; Musemic, R. & Dzaferovic-Masic, E. (2016). Temperature Inversion Measurements in Sarajevo Valley
Using Unmanned Aerial Vehicles, Proceedings of the 27th DAAAM International Symposium, pp.0423-0427,
B.Katalinic (Ed.), Published by DAAAM International, ISBN 978-3-902734-08-2, ISSN 1726-9679, Vienna,
Austria, DOI: 10.2507/27th.daaam.proceedings.062
[8] Masic, A.; Pikula, B. & Bibic, Dz. (2017). Mobile Measurements of Particulate Matter Concentrations in Urban
Area, Proceedings of the 28th DAAAM International Symposium, pp.0452-0456, B. Katalinic (Ed.), Published by
DAAAM International, ISBN 978-3-902734-11-2, ISSN 1726-9679, Vienna, Austria, DOI:
10.2507/28th.daaam.proceedings.063
[9] Cross E. S.; Williams, L. R.; David K.; Lewis, D. K.; Magoon G. R.; Onasch T. B.; Kaminsky, M. L.; Worsnop, D.
R. & Jayne, J. T. (2017). Use of electrochemical sensors for measurement of air pollution: correcting interference
response and validating measurements, Atmospheric Measurement Techniques, 10 (2017) 3575358, DOI:
10.5194/amt-10-3575-2017
- 0887 -
... This leads to different interpretations of the formulas in the literature. Some assume a correction factor n T equal to 1 [38][39][40]. Others consider n T a constant provided in tabular or in graphic form by Alphasense [41,42]. ...
Article
Full-text available
In the Western world, the SO2 concentration in ambient air dropped to low levels, but some emission sources (e.g., merchant ships) and some regions (e.g., low-income countries) still emit substantial amounts of SO2. At those locations, SO2 monitoring is critical. However, low-income countries do not have much access to expensive reference instruments. Low-cost gas sensors might be an alternative, but it is unclear how reliable such measurements are. To evaluate the performance of the low-cost alternative, the same SO2 gas sensor has been subjected to three different calibration methods: (1) low-cost calibration performed in the tropical climate of Cuba; (2) high-end calibration performed in Belgium; (3) a field calibration at an air quality measuring station in Belgium. The first two methods showed similar trends, suggesting that the gas sensor can be calibrated with a low-cost method. The field calibration was hampered by the low SO2 concentrations. For the monitoring campaign in Cienfuegos, Cuba, the low-cost SO2 sensor calibrated by the low-cost method appeared to be sufficiently reliable. The reliability of the sensor increases with the increase in SO2 concentration, so it can be used in Cuba instead of Belgium.
... By taking into account the results of our previous campaigns [9] and [13] we were looking for representative test locations for vertical measurements up to 500m above ground level. The following locations were selected: 19.01.2021. with strong temperature inversion starting from the ground and very low air temperature. ...
Chapter
Full-text available
A new approach for measurements of vertical profiles of temperature and particulate matter (PM) concentrations is developed and tested. Special purpose unmanned aerial vehicle with data acquisition system has been developed in-house. The temperature sensor is based on a fast response thermistor, while the PM sensor uses optical detection of particles. All sensors were tested in the laboratory and calibrated against reference instruments. Two urban locations were selected for field tests. This research produced a very useful asset that characterizes the relation between temperature inversion and air pollution. Keywords: temperature inversion; particulate matter; drone; boundary layer.
... Our drone platform was developed in-house, based on Ardupilot open-source autopilot and commercially available electronic components [16]. This platform has a modular design and can carry various sensors and equipment, including electrochemical sensors for toxic gases [17], [18], and the abovementioned PM sensors. In our previous work we discussed quality assurance (QA) procedures that were applied to ensure reliable data [19], [20]. ...
Chapter
Full-text available
A novel method for terrain mapping is presented in this paper. It is based on low-cost open source components, which brings this innovative technique to a much wider group of researchers than ever before. The in-house developed drone was used together with a commercially available LiDAR sensor to produce a terrain mapping platform. This system was tested in two field locations with good results. In some aspects of performance, the drone surpassed commercially available solutions that are much more expensive and closed. Keywords: LiDAR; drone; terrain mapping; inertial measurement unit; global navigation satellite system.
Chapter
Full-text available
Nanoparticles with diameters from 10 nm to 1000 nm were analyzed using the scanning mobility particle sizer (SMPS) continuously from October 2020 till May 2021. Mass spectrometry was performed in 129 channels. Measurements of some other air pollution parameters were used for comparison and analysis. A large number of nanoparticles was detected in the period of strong air pollution, as expected. However, a surprisingly large number of nanoparticles was detected in spring and summer, where other parameters of air pollution were good. Keywords: SMPS; nanoparticles; air pollution; mass spectrometry.
Chapter
Full-text available
To provide the reference dataset for particulate matter concentrations, a long-term campaign of continuous gravimetric measurements of PM10 and PM2.5 concentrations was performed. Strict implementation of standard EN12341:2014 was fulfilled. Daily and monthly concentrations of both PM10 and PM2.5 were measured with the highest possible accuracy. Ratio PM2.5/PM10 was determined and analyzed. Different distribution of particles was observed during the winter period, where most of the particulate matter originates from burning fossil fuel, and during an episode of Sahara dust, occasionally observed in spring and summer. Keywords: PM10; PM2.5; gravimetric measurement; Sahara dust.
Article
Full-text available
A new method for mobile measurements of black carbon (BC) concentration measurements is presented in this paper. An advanced pocket-sized aethalometer with 5 wavelengths and dual-spot technology was used. Different measurement locations were selected to evaluate the contribution of car emissions and biomass burning emissions during the winter campaign in heavily polluted urban environments. A model based on the Angstrom exponent is proposed for source apportionment studies. Obtained results confirmed the system's capability to distinguish BC sources with high temporal and spatial resolution.
Chapter
Full-text available
A novel method for indoor air quality monitoring is presented in this paper. It is based on the network of smart sensors permanently connected to the cloud. The prototype system, consisting of 10 smart sensors is evaluated in laboratory and real use. Each smart sensor was able to measure air temperature, relative humidity, particulate matter concentration, and carbon dioxide concentration. The cloud-based architecture of the system is explained, followed by the calibration method and real scenario results. The system proved to be suitable for real-time monitoring of indoor air quality parameters for large buildings.
Article
Full-text available
The environments in which we live, work, breathe, and play are subject to enormous variability in air pollutant concentrations. To adequately characterize air quality, measurements must be fast (real-time), scalable, and reliable (with known accuracy, precision, and stability over time). Low-cost AQ sensor technologies offer new opportunities for fast and distributed measurements, but a persistent characterization gap remains when it comes to evaluating sensor performance under realistic environmental sampling conditions. This limits our ability to inform stakeholders about pollution sources and inspire policy makers to address environmental justice air quality issues. In this paper, initial results obtained with a recently developed low-cost air quality sensor system are reported. In this project, data were acquired with the ARISense integrated sensor package over a 4-month time interval during which the sensor system was co-located with a state-operated (Massachusetts, USA) air quality monitoring station equipped with reference instrumentation measuring the same pollutant species. This paper focuses on validating electrochemical sensor measurements of CO, NO, NO2, and O3. Through the use of High Dimensional Model Representation (HDMR), we show that interference effects derived from changing environmental conditions and the ambient-gas concentration mix encountered at an urban neighborhood site can be effectively modelled for the Alphasense CO-B4, NO-B4, NO2-B43F, and Ox-B421 sensors, improving the credibility of air pollutant measurements made with these sensors.
Article
Full-text available
A domestic unmanned aerial vehicle (UAV) was built with the data acquisition capabilities. It is completely based on open-source projects which enables great level of customization. Two different data acquisition systems (DAS) were tested: a proprietary DAS and an in-house developed open platform DAS. Some measurements were presented and possible applications are discussed.
Article
A worldwide monitoring and modelling network would reduce the dramatic toll of air pollution on health and food production, urge Rajesh Kumar and colleagues. A worldwide monitoring and modelling network would reduce the dramatic toll of air pollution on health and food production, urge Rajesh Kumar and colleagues.
World Health Organization
http://www.who.int/airpollution/en/, (2018). World Health Organization, Accessed on: 2018-09-23