Access to this full-text is provided by MDPI.
Content available from Sensors
This content is subject to copyright.
sensors
Article
Occupational Risk Prevention through Smartwatches:
Precision and Uncertainty Effects of the
Built-In Accelerometer
Luis Sigcha 1, Ignacio Pavón1, * , Pedro Arezes 2, Nélson Costa 2, Guillermo De Arcas 1and
Juan Manuel López 1
1
Instrumentation and Applied Acoustics Research Group (I2A2), ETSI Industriales, Universidad Politécnica
de Madrid, Campus Sur UPM, Ctra. Valencia, Km 7., 28031 Madrid, Spain;
luisfrancisco.sigcha@upm.es (L.S.); g.dearcas@upm.es (G.D.A.); juanmanuel.lopez@upm.es (J.M.L.)
2ALGORITMI Research Center, School of Engineering, University of Minho, 4800-058 Guimaraes, Portugal;
parezes@dps.uminho.pt (P.A.); ncosta@dps.uminho.pt (N.C.)
*Correspondence: ignacio.pavon@upm.es; Tel.: +34-91-336-4696
Received: 11 October 2018; Accepted: 1 November 2018; Published: 6 November 2018
Abstract:
Wearable technology has had a significant growth in the last years; this is particularly
true of smartwatches, due to their potential advantages and ease of use. These smart devices
integrate sensors that can be potentially used within industrial settings and for several applications,
such as safety, monitoring, and the identification of occupational risks. The accelerometer is one
of the main sensors integrated into these devices. However, several studies have identified that
sensors integrated into smart devices may present inaccuracies during data acquisition, which may
influence the performance of their potential applications. This article presents an analysis from the
metrological point of view to characterize the amplitude and frequency response of the integrated
accelerometers in three currently available commercial smartwatches, and it also includes an analysis
of the uncertainties associated with these measurements by adapting the procedures described in
several International Organization for Standardization (ISO) standards. The results show that despite
the technical limitations produced by the factory configuration, these devices can be used in various
applications related to occupational risk assessment. Opportunities for improvement have also been
identified, which will allow us to take advantage of this technology in several innovative applications
within industrial settings and, in particular, for occupational health purposes.
Keywords:
wearable; risk assessment; smartwatches; microelectronic mechanical systems (MEMS)
accelerometers
1. Introduction
Wearables are electronic devices (or sensors) that can be used discreetly and comfortably on
the body and, by using the embedded sensors, can be used to measure physical parameters such as
movement, body temperature, and heart rate, among others. The sensors provide functionalities that
are used in biometric and security applications. In recent years, there has been a large growth in this
technological area along with the advances in telecommunications and mobile technology (mainly due
to consumer electronics and miniaturization). Due to the diversity of sensors that can be integrated
into smartwatches, they are currently used in many areas and applications such as health, wellness,
safety, sports, fitness, communication, and business [1–4].
One of the most prominent wearable devices is the smartwatch. In recent years, these watches
have been turned from portable mechanical devices used to measure time to electronic devices with
features and functionalities similar to those available in smartphones, allowing the development of
Sensors 2018,18, 3805; doi:10.3390/s18113805 www.mdpi.com/journal/sensors
Sensors 2018,18, 3805 2 of 20
innovative applications in diverse areas. Smartwatches have advantages over smartphones, such as
the location on the body and the direct contact with users’ skin, which facilitates the use of specific
sensors such as a heart rate monitor, skin conductance and others [1–6].
Although the adoption of wearables is at an early stage, smartwatches have become the most
popular wearables due to some of their advantages, such as their size, portability, and energy autonomy.
Several studies have determined that the technological improvements, the development of innovative
applications, and the reduction of costs can cause these devices to be massively adopted by the
market [3–11].
Software applications for wearables devices are being constantly developed which take advantage
of the increasing computer-processing capabilities, wireless communication, and different sensors.
The sensors that are usually integrated into smart devices are accelerometers, gyroscopes, ambient
light and magnetic sensors [1,2,12–14].
Currently, some researchers use specific wearable devices [
15
–
25
], smartphones with body
adapters [
26
–
35
] and smartwatches [
36
–
38
] for developing occupational risk assessments. Wearables
can be used for continuous monitoring and measurement over time, as well as warning or training
devices for the worker; some of these applications take advantage of the wide range of available sensors
and the discreet use of wearables, which makes them an excellent option for collecting information in
an economic and discreet way [5,16].
In [
39
], certain applications of wearables have been identified for occupational risk assessment,
among which are motion detection and physical activity, recognition of musculoskeletal injuries, fall
detection, evaluation of exposure to physical agents, evaluation of exposure to chemical agents and
the location of potential hazards (moving machinery). This work identifies that there is a growing
number of solutions that are being proposed that are basically supported by the use of wearables. It is
also mentioned that, in the future, this kind of system will take advantage of wireless communication
capabilities to facilitate the exchange of information in real-time and to improve occupational safety
management systems by using emerging technologies such as big data analysis.
With the current technical characteristics (processing speed, data storage, and wireless
communication), wearable devices have the potential to be used as part of integrated risk management
systems. The first experiences of integrated management systems based on wearable technology and
the Internet of Things (IoT) for the industry are described in [
22
,
40
,
41
]. In [
40
], a framework has
been established for the creation of intelligent work environments that allow risk assessment in a
personalized manner in real-time using personal protective equipment with IoT technology, whereas
in [
22
,
41
] industrial security management systems have been implemented that use wearables and
IoT devices to supervise, raise awareness and send emergency alerts to workers to promote better
labor practices and safe work environments. These studies show that the IoT approach can be used in
sensors, tools, and workers using wearables for constant monitoring and real-time alerting. All this
technology will allow the improvement of the management of occupational safety.
Based on the literature review, several applications have been identified in which the wearables’
vibration sensors are used (directly or indirectly) for occupational risk prevention (ORP), health, and
safety. The identified applications are the evaluation of physical activity (PA), fall detection and
vibration monitoring [42].
One of the main applications of wearables devices related to users’ health is the evaluation
of PA. In [
43
], a bibliography referring to the development of the systems dedicated to movement
detection is compiled. In these systems, triaxial accelerometers and movement classification methods
based on threshold or statistical classification schemes are mainly used (supervised machine learning
techniques). In this study, the typical applications of PA detection are also identified, including posture
and movement classification, the estimation of the expended energy, and automatic fall detection.
In [
44
], a compilation of the techniques used to perform the monitoring of occupational physical
activity has been made, in which sensors such as pedometers, accelerometers and global positioning
systems (GPS) can be used simultaneously as integrated systems for the determination of physical
Sensors 2018,18, 3805 3 of 20
activity in the workspace. This article concludes that the addition of sensors allows us to obtain
physiological variables that can be used in other areas of occupational health.
Meanwhile, in [
45
], the effects produced by the location of accelerometers on the wrist or ankle for
the estimation of PA in activities such as walking or cycling has been analyzed. The results show high
precision in identification using accelerometers with a sample a rate of 90 Hz and configured to a range
of
±
4 g, using the support vector machine (SVM) classification algorithm. This study also mentions
that it is possible to detect activities by using characteristics of the signal in the time or frequency
domain, and it is feasible to decrease the sampling rate of the signal at 40 Hz without decreasing the
accuracy of the classification algorithm too much.
Recent solutions have started to use commercial smartwatches to identify human activities.
In [
46
], it is identified that a smartwatch’s accelerometer can provide accurate information to perform
a satisfactory recognition of activities. This article shows that the technical requirements for the
sensors depend on the application for which they are intended (e.g., monitoring of activity or inactivity,
recognition of gestures or industrial monitoring related to exposure to vibrations), whereas in [
47
] it
has been shown that it is possible to identify postures in an accurate way using low sampling rates in
the accelerometer of a smartwatch, which can help to improve the battery life of the devices.
The different efficiencies in the recognition of activities between a smartwatch and a smartphone
were compared in [
48
]: the results suggest that smartwatches can identify manual activities with a
higher accuracy than smartphones, giving them the potential to be used in the biomedical and health
domains. However, in [
49
], a smartphone and a commercial smartwatch were simultaneously used,
showing that the combination of devices can increase the accuracy of the identification of activities.
In addition, in [
36
], a smartwatch was used to identify gestures and activities, in addition to presenting
information in a discreet way to the workers, through the screen of the device and haptic feedback.
The developed system allows permanent assistance to the user during the performance of their work
tasks. This article also presents new forms of interaction between the devices and the users, using
novel applications in areas such as industry (mechanical and construction), the consumer market at
home, and ambient assisted living.
Automatic fall detection is another application in which the wearables’ accelerometers have been
extensively used. In [
50
], a compilatory study was made of fall detection and prevention systems using
environmental and wearable sensors. The study determined that wearables have advantages over
environmental sensor systems, such as better mobility, extended spatial coverage, easier installation
and use. In [
51
], a complete systematic analysis of fall detection systems implemented in the Android
platform was carried out. The results suggest that these devices can be a viable solution for fall
detection due to their excellent cost-effectiveness ratio. Some open issues such as the lack of precision
in the sensors and the reduced range in which the sensors are programmed by default were also
identified; these issues reduce the attractiveness of this technology.
For work-related fall detection, a system is presented in [
52
] that uses various accelerometers
coupled with safety equipment (belt and vest) placed on the chest, hip, arm, and hand. These sensors
communicate with a smartphone and a server to predict potential falls through the automatic detection
of fall precursor events with different classifying algorithms. The obtained results show a considerable
accuracy in fall prediction by detecting precursor events. The authors have also concluded that
it is possible to improve the accuracy of the prediction by increasing, for example, the number of
used sensors.
The authors of [
53
] showed that it is possible to identify daily activities and detect falls with a
smartwatch placed on the left hand using different pattern classification algorithms. In this article,
it is highlighted that the position of the device on the wrist can decrease system performance due to
random movements of the hand. In [
54
] a commercial smartphone and smartwatch with algorithms
based on threshold analysis have been used in combination, demonstrating that the performance of the
system to discriminate daily activities and falls can be increased with a small decrease in the efficiency
in the detection of falls when performing simultaneous analyzes.
Sensors 2018,18, 3805 4 of 20
Solutions that use commercial smartwatches for the evaluation of risks associated with vibrations
that affect the hand–arm system (HAV) have been also developed. An approach based on machine
learning was presented in [
37
]. In this study, a system was developed which uses the built-in
microphone and accelerometer of a commercial (not modified) smartwatch to recognize the kind
of tool used and the time of use to estimate the value of the daily dose of vibration, A(8). The results
suggest that it is feasible to implement recognition systems for work tasks and use this information to
prevent risks. In [
38
], a vibration measurement system was developed using the built-in accelerometer
from a smartwatch to measure the value of A(8), and it was concluded that this system can be used in
specific HAV evaluation tasks due to the technical limitations imposed by the operating system on the
sensors to increase the battery’s autonomy. However, the opportunities for improvement in the future
and the smartwatches’ advantages, such as being a low-cost solution for the continuous monitoring
of workers, its portability and comfort of use were highlighted, as well as the compliance with the
principle of not increasing or causing new risks.
Accelerometers
One of the primary sensors included in smart devices is the accelerometer. At present, this type of
sensor is widely used as a stand-alone device, in smart devices or within wearable devices for specific
applications such as the monitoring of physical and sports activities, fall detection, and the analysis of
structural or machine vibrations, as well as vibrations that affect the human body [
16
–
20
,
22
–
32
,
36
–
38
].
Accelerometers are a type of inertial sensor that can measure acceleration in their sensitive axes.
The principle of operation of the accelerometers is based on a mechanical sensor element, composed of
a test mass attached to a reference frame with a mechanical suspension system. Based on this principle,
the acceleration can be measured electrically, using the physical changes in the displacement of the test
mass. There are three types of accelerometers: piezoresistive, piezoelectric and capacitive [44,55].
The accelerometers used in smartwatches are called MEMS (microelectronic mechanical systems):
these sensors have a reduced dimension and can be used to capture different physical variables [
1
].
The technical characteristics and performance of MEMS accelerometers have been analyzed in several
studies, in most cases comparing them with piezoelectric accelerometers (PCB). The results of these
studies concluded that these sensors are a viable and low-cost solution that can be used in different
applications [56–64].
In the work of Tarabini [
65
], the technical characteristics of the MEMS accelerometers and their
possibilities for use in the evaluation of hand–arm vibrations, full body and other types of mechanical
vibration measurement have been evaluated. This article shows that this kind of transducer complies
with most of the technical requirements of ISO 8041:2005 [
66
]. The limitations found in the frequency
response can be improved using equalization. Therefore, these sensors present advantages over PCB,
such as their reduced cost and size, which allow the sensor to be used directly on the handles of tools
or fixed on the skin.
Although there is an extensive bibliography focused on the analysis of the different characteristics of
MEMS accelerometers, there are fewer studies that analyze the behavior of sensors within the wearable
device (specific device or smartphone). To the best of our knowledge, no study has been identified that
specifically analyzes the performance of accelerometers within commercial smartwatches.
The MEMS accelerometers included in the device circuitry have enough technical characteristics
for optimal operation in specific applications [
39
]. However, when they are used in intelligent devices,
their performance is reduced by the operating systems to reduce the energy consumption and increase
the energy autonomy. Even with this reduction, intelligent devices can be used in many applications.
Knowing the behavior and limitations of the accelerometer (which is one of the most commonly used
sensors) can be useful in future research in which the use (direct or indirect) of the accelerometers in a
commercial smartwatch is considered.
The main objective of the current paper is to determine the accuracy and uncertainty associated
with the acceleration measurements made by the vibration sensors of a few current commercial
Sensors 2018,18, 3805 5 of 20
smartwatches. The results are used to analyze the current and future potential applications related to
the prevention and management of occupational risks. It is also discussed whether these devices can
be used as precision measuring instruments, control instruments or as training and teaching devices.
Specifically, this paper presents the methodology used for the characterization of the frequency
response and variability of the amplitude of the acceleration magnitude captured by the MEMS
accelerometers of three commercial smartwatches from different manufacturers. The smartwatches
used in the experiments use the Android Wear operating system (version 1.5, Google), which facilitates
the preparation of trials and the comparison of results.
2. Materials and Methods
The amplitude and frequency response of three commercial smartwatches was evaluated in order
to analyze the accuracy of the accelerometer measurements. In this analysis, the repeatability of the
measurements has been considered, and the associated uncertainties have been calculated.
In this kind of intelligent device, it is possible to obtain the digital representation of the acceleration
signal in the three axial axes from the sensor (triaxial MEMS type accelerometer) to analyze its
amplitude and frequency. The amplitude can be analyzed by using the maximum amplitude (peak
value) of the signal or by analyzing the root mean square value (RMS), while the frequency tests can be
carried out using discrete signals such as sinusoidal waves at different frequencies. Also, the amplitude
and frequency test can be performed in the frequency domain by using the Fourier transform with test
signals such as pink noise or frequency sweeps of sine waves [67].
In the current study, the amplitude and frequency analysis is performed by analyzing the RMS
values of sinusoidal vibration signals at discrete frequencies. The mechanical signals used in the
measurements were generated by a vibration shaker.
Also, the estimation of the maximum sampling rate of each of the smartwatches’ accelerometers
was verified at first by using time-stamps created during the capture of signals. It has been identified
that the sampling time-intervals of the accelerometer are not constant, which generates jitter errors.
The average sampling rates of the smartwatches configured in Android Studio 3.2 (Google, California,
USA) at the fastest sensor sampling rate are 250 Hz for smartwatch 1, 200 Hz for smartwatch 2,
and 50 Hz for smartwatch 3.
The analysis was made at different amplitudes to verify the amplitude and frequency behavior
of the built-in accelerometers. The analysis methodology was chosen according to the existing
bibliography and ISO standards regarding the measurement instruments used for the analysis of
the human response to vibrations [66] and secondary accelerometer calibration methods [68].
This section has been divided into six subsections. Section 2.1 describes the technical
characteristics of the built-in accelerometers from the smartwatches used in the tests. Section 2.2
describes the test bench used in the calibration and measurement tests. Section 2.3 details the
development of the vibration measurement system for the Android test devices. Section 2.4 presents
the process of calibration and adjustment of the implemented measurement system in the devices,
before performing frequency and amplitude tests. Section 2.5 describes the frequency response
measuring process of the test devices at different amplitudes. Lastly, Section 2.6 presents the procedure
used to perform the analysis of uncertainties associated with calibration and measurements.
2.1. Test Devices
Three commercial smartwatches with the Android Wear system (version 1.5, Google) were used
to perform the amplitude and frequency analysis. For the choices of the test smartwatches, several
high-end commercial smartwatches available on the market in 2017 were identified.
From these devices, two smartwatches were chosen which had the highest factory-set sampling
rate in the Android Wear system for the accelerometer. In addition, a third smartwatch was chosen
with a standard sample rate factory-set for movement analysis to verify its behavior.
Sensors 2018,18, 3805 6 of 20
The three smartwatches have similar technical characteristics, and their sensors are programmable.
However, the factory settings (maximum sampling frequency and amplitude range) for the Android
system are lower than that at which accelerometers can work.
The technical characteristics of the smartwatches and the built-in accelerometers are specified in
Table 1.
Table 1. Technical characteristics of the smartwatches and the accelerometers.
Smartwatch 1 Smartwatch 2 Smartwatch 3
Dimension (mm) 36 ×10 ×51 45.5 ×52.2 ×10.9 45 ×45 ×11.5
Weight (g) 38 66.5 54
Accelerometer amplitude range
(standard acceleration due to gravity: g)±2, ±4, ±8, ±16 ±2, ±4, ±8, ±16 ±2, ±4, ±8, ±16
Accelerometer maximum sample rate (Hz) 2000 4000 1000
Accelerometer maximum sample rate set in Android (Hz)
250 200 50
Analog-to-Digital Converter (ADC) (bit) 12 16 16
2.2. Test Bench
The calibration processes and the amplitude and frequency measurements were made by using
a vibration generation and analysis system composed of a PULSE 7537 system (Bruel & Kjaer,
Copenhagen, Denmark), connected to an LDS PA 100E power amplifier (Bruel & Kjaer), which feeds to
an LDS V406 CE M4 vibration shaker (Bruel & Kjaer). A SV106 vibrometer (Svantek, Warsaw, Poland)
connected to a Dytran 3023M3 accelerometer (Dytran Instruments, California, USA) was used as a
reference. The equipment was connected according to Figure 1.
Figure 1.
Vibration analysis and generation system: (
a
) block diagram of the connections; (
b
) devices
of the test bench.
2.3. Development of the Vibration Measurement System for the Android Wear Devices
A software application for Android Wear devices was developed [
69
] and installed in the
smartwatches to determine the RMS acceleration value. The application captures and processes the triaxial
accelerometer signals at the maximum frequency sample rate allowed by the device. The application
shows, as a result, the RMS acceleration values without the gravity offset. The scheme in Figure 2was
used to implement a measurement system following the guidelines of ISO 8041:2005 [66].
The acceleration analysis for each of the three axial axes was made in the frequency domain
using the fast Fourier transform (FFT) [
70
]. The digital signals captured from the sensor must be
subjected to the windowing process before the transformation to the frequency domain; after that,
the signal must be filtered to avoid the effects of gravity using a high-pass filter that eliminates
low-frequency components.
Sensors 2018,18, 3805 7 of 20
Figure 2.
Vibration measurement system (frequency domain analysis). FFT, fast Fourier transform;
RMS, root mean square value.
The RMS (total band power) value for each analysis window is obtained using the linear
integration method, calculated using the Parseval theorem described in the following equation:
RMS =r2∑f=SR/2
f=0|X(f)|2, (1)
where |X(f)| is the module of the FFT components in each spectral line.
The values displayed by the smartwatch measurement system are obtained from the moving
average of the RMS acceleration, which is calculated using the instantaneous values of the acceleration
of each analysis window.
2.4. Calibration and Adjustment through the Comparison Method
Although the smartwatches have a factory-calibrated triaxial accelerometer, calibration and
adjustment tests must be performed before performing frequency and amplitude measurement tests
on the devices. The calibration tests must verify the degree of deviation of the measurements obtained
with the smartwatch to later perform the adjustment process and compensate this deviation to avoid
mistakes in data collection in the rest of the experiments.
The calibration process was made by adapting the guidelines of ISO 16063-21:2003 [
68
] regarding
the calibration of a vibration transducer by comparison to a reference transducer. From the different
methods of secondary calibration that exist, the method of comparison with a reference accelerometer
is the most used due to its ease of implementation and reliability. The method consists of coupling the
accelerometer to be calibrated on another accelerometer which is pattern calibrated and then driving
the coupled pair with vibrations with a determined frequency and amplitude.
For this study, the calibration of the vibration measurement system in smartwatches consists in
making a comparison between the smartwatch acceleration measurement value and the reference to
get the acceleration bias and its associated uncertainty. For this process, the reference accelerometer
and the smartwatches were coupled using adhesive. For smartwatches 2 and 3, it was necessary to use
a thin plate between the devices due to the size and the straps of the smartwatches.
The calibration of the devices can be performed at any frequency and amplitude within the
operational limits of the device. The ISO 16063-21:2003 [
68
] standard recommends using the
standardized third-octave frequencies indicated in ISO 266:1997 [
71
] and amplitudes of 1, 2, 5 or
multiples of 10 m/s2.
In the case of digital devices, since it is not possible to access the voltage output of the
accelerometer directly to get the sensitivity value, the calibration can be performed by verifying
the measurement error, which is the difference between the acceleration value obtained with the test
device and the value of the reference device. By getting the value of this error, a correction factor can
be applied to each accelerometer axis to get the RMS value. For the calibration of the smartwatches, a
mechanical sinusoidal vibration signal with an amplitude of 10 m/s
2
at a frequency of 12.5 Hz was
used, considering the limitations of the devices in both amplitude and frequency.
2.5. Test of the Frequency Response at Different Amplitudes
The present test was made to verify the smartwatches’ frequency linearity at three different
amplitudes. The test was made following the guidelines of the ISO 8041:2005 [
66
] standard and the
ISO 16063-21 [68] criteria for coupling the reference and the test transducers.
Sensors 2018,18, 3805 8 of 20
According to ISO 8041:2005 [
66
], when performing measurements at different amplitudes and
frequencies, the indication error (Ex) can be expressed in absolute terms with the difference between
the value of the analyzed device and the value shown by reference analyzer (Equation (2)). The ISO
standard also explains that the indication error can be shown in relative terms, expressed in decibels or
the percentage of deviation from the reference value. The deviation percentage (
ε
) for the reference can
be calculated using Equation (3) [66].
Ex =atest −are f , (2)
ε= atest −are f
are f !∗100%. (3)
The frequency response tests were performed by adapting the method of comparison with
a reference accelerometer described in ISO 16063-21 [
68
] (in a similar way to the calibration test).
The transducers were coupled according to the procedures described in the method of calibration of
accelerometers by comparison (back-to-back calibration method). The results of these measurements
are the RMS acceleration values of the deviation percentage to the reference vibrometer and its
associated uncertainties at different frequencies and amplitudes.
The mechanical test signals were generated with the vibration shaker fed by discrete sinusoidal
waves at third-octave intervals, with central frequencies in the range of 6.3–100 Hz. The frequencies
were selected based on the maximum frequency analysis of each device according to Nyquist’s
theorem [67]. The vibration amplitudes used in the test were 10, 5 and 2 m/s2RMS.
The reference accelerometer and each smartwatch were simultaneously coupled to the shaker’s
vibrating surface; then, six RMS acceleration determinations were performed at each of the
frequencies, keeping the acceleration amplitude constant, then repeating the process for the other
amplitudes mentioned.
2.6. Calibration and Measurement Uncertainties
Due to the sources that can produce variability in the measurements of the magnitude of vibration,
such as the physical equipment conditions, experimental variability, and environmental factors such as
humidity, temperature or atmospheric pressure, it is necessary to perform the analysis of uncertainties
associated to the measurements and the calibration to be able to characterize the dispersion of the
values that could reasonably be attributed to the measurand.
The estimation and expression of the uncertainties of the calibration and measurement results
were made following the criteria and guidelines of ISO/IEC Guide to the Expression of Uncertainty
in Measurement (GUM) [
72
] and adapting ISO 16063-21:2003 [
68
]. Significant contributions to the
final uncertainty are considered as each of the corrections of the devices used in the calibration and
measurement processes:
•The uncertainty associated to the reference device.
•The temporary drift of the reference device.
•The scale division of the reference device and the device to be calibrated.
•The repeatability of the test during the calibration process.
•The alignment of the device to be calibrated against the reference.
•Environmental effects (temperature and humidity variation).
•Variations produced by the magnetic field of the vibration shaker.
•The uncertainty of the power amplifier and the vibration shaker.
•The uncertainty produced by the sampling rate variation of the smartwatches.
2.6.1. Model
According to ISO-GUM, for the determination of a measurand, other quantities that are related
through a functional relationship are normally used. The function contains all quantities, including
Sensors 2018,18, 3805 9 of 20
the corrections and correction factors for systematic effects that can bring a significant component of
uncertainty to the measurement result. The corrections and correction factors are applied to the model
to compensate for the significant influences that affect the measurement results due to the effects of
systematic and random errors [68,72].
For the determination of uncertainty associated with calibration and measurements, it is necessary
to adapt the procedure described in ISO 16063-21:2003 [
68
], because the output voltage of the
accelerometer to be calibrated cannot be directly accessed to use the sensitivity as a function to
calculate the uncertainty.
In the current case, to facilitate the uncertainty calculations of the calibration and measurements,
the indication error (Ex) shown in Equation (2) has been used as a function of the model (Equation (4)).
In this model, the corrections produced by systematic effects are considered.
Ex =atest +ctemp test +cres test +ccalib test +ccross test +cjitt test
−are f +cte mp re f +cres re f +ccal ib ref +ccross re f ,(4)
where:
•ctemp is the correction due to the temperature;
•cres is the correction due to the finite resolution of the meter;
•ccalib is the correction due to the calibration of the meter;
•ccross is the correction due to the cross sensitivity;
•cjitt is the correction due the sampling rate variation in the smartwatch (jitter error).
The components of uncertainty have been expressed as relative standard uncertainties to simplify
the calculations, as mentioned in [
73
]. The relative uncertainty values have been calculated using
Equation (3).
Although in some cases the value of the corrections can be a null value when calculating the
indication error, their uncertainties will not be null. The coefficients of sensitivity for each contribution
are equal to one because the input quantities can be considered as uncorrelated for the calculation of
the relative expanded uncertainty, as indicated in [68,73].
2.6.2. Characterization and Estimation of the Value of Contributions Type A and Type B
According to GUM [
72
], the uncertainty components can be grouped into two categories according
to the evaluation method: “type A” and “type B”. The classification into type A and type B consists of
two different ways of evaluating uncertainty components, based on probability distributions.
The characterization of the sources of uncertainty used in the analysis are described in Table 2.
Some sources of uncertainty indicated in ISO 16063-21:2003 [
68
] have been ignored because they are
not components of the proposed model, or their information is not available.
Table 2. Characterization of the uncertainty sources for calibration.
Variable Estimator Density Function Type Description
Ex Y Normal A Indication error
u accelref X1 Normal A Reference vibrometer RMS acceleration
u accelsw X2 Normal A Smartwatch RMS acceleration
u tempref X3 Rectangular B Reference vibrometer variation due to the temperature
u tempsw X4 Rectangular B Smartwatch based vibrometer variation due to the temperature
u resref X5 Rectangular B Reference vibrometer finite resolution
u ressw X6 Rectangular B Smartwatch vibrometer finite resolution
u calibref X7 Normal B Reference vibrometer uncertainty
u calibsw X8 Normal B Smartwatch calibration uncertainty
u crosssw X9 Rectangular B Smartwatch accelerometer cross sensitivity
u crossref X10 Rectangular B Reference accelerometer cross sensitivity
u swjitt X11 Normal B Smartwatch RMS uncertainty due to the sampling rate variation
Sensors 2018,18, 3805 10 of 20
For the estimation of the combined uncertainty, all the sources of uncertainty from the proposed
model have been analyzed. The uncertainty associated to the acceleration of the meters has
been analyzed as a normal distribution because this variable is determined by performing several
quantity determinations.
The uncertainty components of the temperature, the finite resolution of the meter and the
cross-axis sensitivity have been analyzed as rectangular distributions because only the range in
which the quantity can be found is known.
The uncertainty contribution of the devices’ calibration is analyzed as a normal distribution. Also,
the uncertainty contribution of the sampling rate of the smartwatch has been analyzed by using the
Monte Carlo method [74] with the model indicated in [75].
The combined relative standard uncertainty associated to the calibration test is estimated using
Equation (5), according to the law of propagation of the variances. For uncorrelated elements, the
relative combined uncertainty can be estimated by Equation (6), adapted from ISO 16063-21:2003 [
68
].
uc,rel (y)=v
u
u
t∑
i=0∂f
∂xi2
urel i 2(xi)+2
N−1
∑
i=0
N
∑
i=0
∂f
∂xi
∂f
∂xj
urel xi,xj, (5)
where:
•uc,rel is the combined relative standard uncertainty;
•yis the measurand determinate for Ninput quantities x1,x2, . . . , xn;
•∂f/∂xiare partial derivatives evaluated at Xi=xi;
•xiare estimated input values;
•urel (xi,xj) are estimated covariance.
u2
c,rel (y)=
n
∑
i=0urel i ci
xi2
, (6)
where ciare sensitivity coefficients.
As indicated in [
68
], the results of the measurement must be reported as the expanded uncertainty.
In this case, it is necessary to multiply the value of the combined uncertainty by a coverage factor kto
get the expanded uncertainty. As mentioned in the ISO 16063-21:2003 standard, it has been considered
that the results of the measurements follow a normal distribution so that the coverage factor k= 2 can
be used for coverage of approximately 95%. The expanded relative uncertainty can be estimated by
Equation (7), adapted from [68].
Urel (y)=kuc,rel(y). (7)
The same model and method were used to calculate each of the uncertainties associated with the
frequency response measurement tests, obtaining results for each of the analysis points. The calculation
of the associated uncertainties was performed using the relative uncertainties expressed as a percentage
to facilitate the calculation and expression of results.
3. Results
The following section shows the results achieved in the experiments of the frequency amplitude
response of the three test smartwatches. In Section 3.1, the results of frequency linearity are illustrated
graphically at different amplitudes. Section 3.2 details the smartwatches’ measurements of deviation
as compared to reference vibrometer measurements, expressed in percentages, together with their
respective associated uncertainty.
3.1. Frequency Response Test at Different Amplitudes
The results of the tests were obtained from the measurements of the RMS magnitude of the three
smartwatches (average value of six determinations at each point) in the z-axis according to the Android
Sensors 2018,18, 3805 11 of 20
sensor coordinate system [
76
]. The reference vibrometer is used to adjust the amplitude value of the
mechanical test signal to the different test frequencies and amplitudes. The frequency analysis is
performed only for the first third-octave bands, due to the maximum sampling frequency of each
smartwatch. The frequency response curves in three different amplitudes (10, 5 and 2 m/s
2
) are shown
in Figures 3–5, respectively. An error band has been defined for amplitude and frequency linearity
according to the specifications of ISO 8041:2005 [
66
], where values with a deviation greater than 6%
are considered as amplitude linearity errors. Deviation values greater than 6% with respect to the
reference have been highlighted in the figures and tables.
3.1.1. Smartwatch 1
The frequency response curves of smartwatch 1 (see Figure 3) show a relatively flat frequency
response up to the frequency of 40 Hz for the three amplitudes. For frequencies higher than 50 Hz,
the amplitude gradually decreases; this fall in the amplitude could be produced by the slope of the
anti-aliasing filter applied by the sensor before the digitalization and the fixing method. A deviation of
more than 6% (error band) has been identified in the bands of 80 and 100 Hz.
Figure 3. Smartwatch 1 frequency response for three amplitudes.
3.1.2. Smartwatch 2
For smartwatch 2 (see Figure 4), the frequency response curves present a relatively flat response
up to the frequency of 50 Hz for the three vibration amplitudes used in the tests. For frequencies
higher than 63 Hz, the error increases up to a maximum of 0.84 m/s
2
. This increase in amplitude at
high frequency could be produced by the physical location of the accelerometer inside the device,
the accelerometer and smartwatch fixing method (different from smartwatch 1, due to its size) and the
vibration modes of the device itself. In this smartwatch, the deviation induced by the anti-aliasing
filter is not noticeable; this may be due to the cutoff frequency of the low pass filter, which can be
greater than 100 Hz. A deviation of more than 6% has been identified in the band of 80 Hz.
Figure 4. Smartwatch 2 frequency response for several amplitudes.
Sensors 2018,18, 3805 12 of 20
3.1.3. Smartwatch 3
For smartwatch 3 (see Figure 5), the frequency response curves have a relative flat response up
to the frequency of 20 Hz for the three amplitudes analyzed. For the maximum frequency analyzed
of 25 Hz, the error increases up to 1.19 m/s
2
; this increase in the amplitude can be produced by
the sampling error (jitter), the accelerometer and smartwatch fixing method (different from the fist
smartwatch, due the difficulty of removing the smartwatch’s straps) and the location of the sensor and
the vibration modes of the device itself. The frequency analysis in this device has only been possible in
the first seven third-octave bands because the maximum sampling frequency set at the factory is 50 Hz.
As in smartwatch 2, there is no decay in the level at frequencies of 20 or 25 Hz induced by the
anti-aliasing filter. A deviation of more than 6% has been identified in the band of 25 Hz.
Figure 5. Smartwatch 3 frequency response for several amplitudes.
3.2. Results of the Expanded Relative Uncertainties and Relative Deviation of the Smartwatches Measured from
the Reference at Different Amplitudes and Frequencies
In order to compare the deviation of the measurements at different amplitudes of each smartwatch,
Tables 3–5show the relative deviations expressed as the deviation percentage (
ε
) and the expanded
relative uncertainty (U
rel
) (expressed in percentage) of each smartwatch with respect to the reference
vibrometer at each frequency and amplitude analyzed in the previous test. The deviation percentage
(ε) values are determined using Equation (3).
The uncertainty values from each of the frequencies and amplitudes were calculated through
the calibration model by using the comparison method for all the frequencies and amplitudes in the
three devices.
3.2.1. Smartwatch 1
Although the absolute error of the device at high frequencies is different according to the magnitude
of the vibration signal (see Figure 3), the relative error expressed in percentage for each frequency band
at different amplitudes is similar, reaching a maximum of
−
20.82% at the frequency 100 Hz (see Table 3).
These results show a linear behavior with respect to the amplitude in the entire analysis range of the
device. The deviation in the error band of 6% is exceeded in the bands of 80 and 100 Hz.
3.2.2. Smartwatch 2
For smartwatch 2, the relative error expressed in percentage remains at low values (see Table 4).
For frequencies higher than 63 Hz, the error increases up to a maximum of 8.41%. This increase
in the frequency response at high frequencies could be produced by the physical location of the
accelerometer inside the device and the vibration modes of the device itself. In this smartwatch,
the errors induced by the anti-aliasing filter are not noticeable. The deviation in the error band is
exceeded in the band of 80 Hz.
Sensors 2018,18, 3805 13 of 20
Table 3.
Deviation percentage and relative uncertainty of smartwatch 1 as compared to the reference
vibrometer.
Frequency (Hz)
Amplitude 6.3 8 10 12.5 16 20 25 31.5 40 50 63 80 100
10 m/s2ε(%) −1.14 1.28 0.63 0.00 0.81 −0.70 −1.15 −1.56 −1.31 −2.81 −4.83 −9.32 −20.64
Urel ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.1
5 m/s2ε(%) 1.10 −0.37 −0.20 −0.29 −0.30 −0.65 −1.28 −1.65 −1.12 −2.45 −4.54 −9.13 −20.68
Urel ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.2
1 m/s2ε(%) 1.41 0.49 0.25 0.35 −0.45 −0.63 −1.54 −2.51 −1.61 −3.03 −4.84 −9.44 −20.82
Urel ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.8
Values in red: deviation values greater than 6% with respect to the reference.
Table 4.
Deviation percentage and relative uncertainty of smartwatch 2 as compared to the reference
vibrometer.
Frequency (Hz)
Amplitude 6.3 8 10 12.5 16 20 25 31.5 40 50 63 80
10 m/s2ε(%) −0.74 1.21 0.73 0.02 0.50 0.24 −0.21 −0.14 1.35 1.56 3.89 8.41
Urel ±3.1 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.1
5 m/s2ε(%) 0.89 0.23 0.31 −0.49 0.15 −0.10 −0.59 −0.26 1.20 1.29 3.63 7.71
Urel ±3.1 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0 ±3.0
1 m/s2ε(%) 2.18 −0.74 −0.57 −0.08 0.17 0.06 −0.85 −0.38 1.50 0.94 3.51 7.06
Urel ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1 ±3.1
Values in red: deviation values greater than 6% with respect to the reference.
3.2.3. Smartwatch 3
For smartwatch 3, the relative error expressed in percentage remains at low values until 20 Hz
(see Table 5). For frequencies higher than 25 Hz, the error increases up to a maximum of 11.9%. As with
smartwatch 2, the increase in amplitude at high frequencies could be produced by the physical location
of the accelerometer inside the device and the vibration modes of the device itself. Errors induced by
the anti-aliasing filter are not noticeable. The deviation in the error band is exceeded only in the band
of 25 Hz.
In the three devices, a gradual increase in the uncertainty can be observed when increasing the
frequency of analysis. This behavior is produced by the contribution of uncertainty produced by
the variation in the sampling rate (jitter effect) that affects the high frequencies that are close to the
Nyquist frequency.
Table 5.
Deviation percentage and relative uncertainty of smartwatch 3 as compared to the reference
vibrometer.
Frequency (Hz)
Amplitude 6.3 8 10 12.5 16 20 25
10 m/s2ε(%) −0.88 0.71 −0.17 0.03 −0.12 −0.29 11.90
Urel ±3.1 ±3.1 ±3.1 ±3.2 ±3.3 ±3.4 ±3.6
5 m/s2ε(%) 0.74 −0.11 0.06 0.05 −0.04 −1.13 11.68
Urel ±3.2 ±3.1 ±3.1 ±3.2 ±3.3 ±3.4 ±3.6
1 m/s2ε(%) −0.42 −1.50 −1.53 −1.84 −1.43 −1.41 7.41
Urel ±3.1 ±3.1 ±3.2 ±3.2 ±3.3 ±3.4 ±3.7
Values in red: deviation values greater than 6% with respect to the reference.
4. Discussion
The frequency analysis of the three devices can only be performed for the first third-octave
bands, depending on the maximum sampling rate of each smartwatch. The limitation identified in
Sensors 2018,18, 3805 14 of 20
the bandwidth restricts the use of these devices to specific evaluation tasks related to occupational
risk prevention.
The value of 10 m/s
2
RMS was used as the maximum amplitude of analysis, because higher
amplitudes vibrations can saturate the accelerometer depending on its orientation with respect to the
ground due to the gravity offset. The acceleration magnitude added to the gravity offset can exceed
the factory set amplitude limit of ±2 g (13.85 m/s2RMS) easily.
4.1. Analysis of the Frequency Response at Different Amplitudes
Regarding the frequency response analysis, smartwatch 1 presents the higher analysis bandwidth
(125 Hz), and presents a flat frequency response up to the frequency of 40 Hz with a maximum
amplitude range of
±
2 g. This characteristic makes this device an excellent option for vibration
analysis tasks that require a range up to the band of 80 Hz.
For smartwatch 2, the flat frequency response makes this device an excellent option to perform
vibration measurements. The primary considerations for this device are the maximum range of
frequency analysis, due to the sampling rate (200 Hz) and the amplitude range (
±
2 g) configured at
the factory.
For smartwatch 3, the results present a flat frequency response at low frequencies. However, due
to its reduced sampling rate (50 Hz) and its amplitude range (
±
2 g), this device can underestimate
the magnitude of vibrations when performing vibration measurement tests with signals that have
frequencies higher than 25 Hz.
4.2. Measurement Uncertainties Analysis
The values obtained in the test of amplitude and frequency are in the range from
±
3.0% to
±
3.8%. According to ISO 8041:2005, the expanded uncertainties for mechanical frequency response
tests should be a maximum of 4.5% for pattern device verification and 5% in periodic verification for
users. These maximum values have not been exceeded in any of the tests, even at very high or low
frequencies where it is possible to get variability in the test signal due to the technical characteristics of
the equipment.
In this analysis, the uncertainty associated with the calibration of the reference vibrometer has
been identified as the primary source of contribution to the combined uncertainty, followed by the
contribution to the uncertainty produced by the finite resolution of the meters, especially when
analyzing low acceleration amplitudes (5 and 1 m/s2).
A gradual increase in the uncertainty has been noticed in the three devices when increasing
the frequency of analysis. The increase in the uncertainty is produced by the jitter effect due to the
variation in the sampling rate [75].
The other uncertainties sources analyzed have been characterized by having a low contribution to
uncertainty because the tests were conducted in a laboratory with a controlled environment.
4.3. Possible Uses with the Current Configuration and Limitations
Although the smartwatches that use the Android Wear system have certain limitations in
bandwidth and the maximum amplitude analyzed with the accelerometer, several applications for
occupational risk assessment have been identified in which the vibration sensor can be used with the
current factory configuration.
In addition, in the bibliography review, two main limitations for the identification of activities
and the analysis of hand–arm vibrations have been identified which are related with the location of
the device and the sensor in the body. The first limitation is that the smartwatch is located on the wrist
and there can be a relative movement between the sensor itself and the bone, due to the presence of
soft tissues that can make the analysis of the vibration signals difficult.
The second limitation regards hand–arm vibration assessment. In this case, the sensor must be
placed between the part of the body exposed to vibrations and the vibrating surface; therefore, when
Sensors 2018,18, 3805 15 of 20
evaluating the magnitude of vibrations with smartwatches, changes in the acceleration response can
occur because the transmissibility between the palm and the wrist is not in unity [77].
4.3.1. Smartwatches for Physical Activity
For monitoring occupational physical activity, it has been identified that the technical
characteristics of the smartwatches analyzed in this article can meet the sampling rate requirements
in activities such as sleep detection, heart rate [
47
], ambulatory movement (frequencies of analysis
in the range of 1–3 Hz) [
44
], identification of postures (sampling rate of 10 Hz) [
48
], identification of
specialized manual activities (sampling rate 20 Hz) [
49
], and for environmental interaction in the work
environment using gestures (25 Hz sampling rate) [
36
]. In these applications, low requirements in
the sampling rate are identified, while to identify the PA with sensors placed on the wrist in [
46
], it is
mentioned that it is possible to use a sampling frequency of 40 Hz without degrading the accuracy of
the classifier.
4.3.2. Smartwatches for Fall Detection
For fall detection, different results are identified depending on the configuration of the devices.
In [
78
], it is mentioned that for fall detection using smartphones, the acceptable amplitude ranges of
the devices are
±
4 or
±
16 g. This range is higher than the one currently configured in the factory
(±2 g)
. In [
79
], a 100 Hz sampling frequency has been used in a smartphone, getting satisfying results
in the fall detection.
In [
53
], the authors suggest that the current factory configuration of Android devices may not
have the necessary precision due to the built-in accelerometer sample rate. The operating system
allows the configuration of the sample rate in a range from 7–200 Hz depending on the accelerometer
model. Also, the sample rate presents some variation over time. Besides this, it is mentioned by other
authors such as Abbate and Fudicar [
80
,
81
] that using low sampling frequencies (50 Hz) is a good
compromise between performance and energy consumption in the fall detection.
For fall detection with wearables placed on the wrists, in [
54
], fall datasets have been recorded and
evaluated using commercial devices with inertial sensors. The maximum sample rate of this devices
was 50 Hz; with this configuration, a high percentage of precision in fall detection has been achieved.
These results suggest that is possible to get reliable performance in fall detection using the current
accelerometer factory configuration in smartwatches, also maintaining a useful energy autonomy of
the device to be used continuously over time.
4.3.3. Smartwatches for Hand–Arm Vibration
For the measurement of the magnitude of the vibrations in applications related to exposure to
hand–arm vibrations, the results show that the current configuration of the Android smartwatches
does not meet the requirements for accurate measurement in the complete range of frequency and
amplitudes [38].
When comparing the results of the smartwatches’ measurements with the results achieved by
Tarabini (direct analysis of the MEMS accelerometers), it is confirmed that the main limitations are the
reduced bandwidth and the lack of linearity found in the high frequencies, although, depending on
the application, the linearity problems can be overcome by sensor equalization.
The analyzed accelerometer factory configuration of the smartwatches (in amplitude and
frequency) is lower than the described by Tarabini [
65
]. However, it is feasible to use some smartwatch
models for the evaluation of vibrations that affect the whole body. Because the analysis bandwidth
proposed by ISO 8041:2005 is from 0.5–80 Hz for health and well-being. On the other hand,
for hand–arm vibrations, the current configuration of the smartwatch can make mistakes in the
measurement of the vibration magnitude when analyzing vibrations with high magnitudes and
extended frequency content.
Sensors 2018,18, 3805 16 of 20
As described by Moschioni [
82
], the vibration direct measurement is the best procedure for
assessing the hand–arm vibration-related risk with minimum uncertainty; nevertheless, to perform the
hand–arm vibration risk assessment with the actual sensor configuration, it could be convenient to
use accelerometers for the detection of a pattern of activities and tools and then estimate the vibration
exposure instead of directly measuring the magnitude, as proposed in [37].
The sensor configuration problem can be solved quickly with improvements in the energy
consumption management of the devices and sensors, and, when the wearables APIs allow, a more
detailed configuration of the characteristics of the accelerometers.
In the work of Matthies [
37
] referring to the estimation of the exposure to vibrations using an
Android smartwatch with an accelerometer configured with a sampling frequency of 50 Hz, it has been
identified that smartwatches have a sufficient precision to analyze and detect several kinds of tools.
However, with the simultaneous use of a smartwatch’s microphone, 85% of the overall detection rate
has been achieved in several tools. Thus, with the results achieved in this study for the smartwatches,
it is feasible to get an improvement in the percentage of accuracy in the detection of tools in at least
two of the smartwatches analyzed, because they provide higher sampling rates that can be used by
classification algorithms.
In the current work, the frequency response of three smartwatches at different amplitudes has
been characterized. The result achieved can be used as a reference for future research work related
to the use of accelerometers of wearables devices in different applications such as the prevention of
occupational hazards, fall detection or activity or posture detection [39].
The methodology used for the evaluation of the uncertainties for calibration has been presented.
The model can be adapted for the evaluation of the uncertainties associated with the measurements
of the magnitude of the vibration, which may allow us in the future to use these devices in
metrological areas.
It has been identified that the maximum sampling rate and dynamic range to which accelerometers
are adjusted within the Android Wear operating system are lower than those at which they could
work. Although the design and construction of these sensors allow the use of higher sample rates
and dynamic ranges, it is not currently possible to modify these characteristics with the traditional
programming methods on the Android platform [
83
], restricting the operation of the devices of this
platform to the factory settings.
Although this kind of wearable device has a reduced cost when compared with the specific
measurement equipment, the results show that the sensors have a high potential to be used in different
tasks in which it is possible to analyze the acceleration through a built-in accelerometer. With the
current configuration, the smartwatches can be used as an option for providing general information
about hand–arm vibrations and physical activities related to ORP, but it is necessary to make technical
improvements for their use in specific applications such as the assessment of hand–arm vibration, in
which limitations in the frequency and amplitude range might be relevant.
Despite the limitations currently imposed by the operating system and the application
programming interface, it is possible that solutions will be developed shortly to improve the open
issues, such as the battery autonomy of these devices, to avoid reducing the technical characteristics of
the sensors. Moreover, if the frequency response curve is known, it is possible to use a reverse filter to
improve the frequency linearity in the accelerometer if it is required [34].
Author Contributions:
L.S., I.P., G.D.A. and J.M.L. conceived, designed and performed the experiments and
wrote the manuscript. P.A. and N.C. contributed to the early versions and reviewed the manuscript. All authors
have contributed to the analysis, interpretation and discussion of the results, and have read and approved the
final version of the manuscript.
Funding:
This research was partially funded by Instrumentation and Applied Acoustics Research Group
(I2A2) ETSI Industriales. Universidad Politécnica de Madrid. And was also supported by COMPETE:
POCI-01-0145-FEDER-007043 and FCT—Fundação para a Ciência e Tecnologia within the Project Scope:
UID/CEC/00319/2013, and Fundación Prevent (X Convocatoria de becas I + D en prevención de riesgos laborales).
Conflicts of Interest: The authors declare no conflict of interest.
Sensors 2018,18, 3805 17 of 20
References
1.
Sazonov, E.; Neuman, M.R. Wearable Sensors: Fundamentals, Implementation and Applications; Elsevier:
Amsterdam, The Netherlands, 2014.
2.
Park, S.; Chung, K.; Jayaraman, S. Chapter 1.1—Wearables: Fundamentals, Advancements, and a Roadmap
for the Future. Wearable Sens. 2014, 1–23. [CrossRef]
3.
Page, T. A Forecast of the Adoption of Wearable Technology. Int. J. Technol. Diffus.
2015
,6, 12–29. [CrossRef]
4.
Wright, R.; Keith, L. Wearable Technology: If the Tech Fits, Wear It. J. Electron. Resour. Med. Libr.
2014
,
11, 204–216. [CrossRef]
5.
Rawassizadeh, R.; Price, B.; Petre, M. Wearables: Has the age of smartwatches finally arrived? Commun. ACM
2015,58, 45–47. [CrossRef]
6.
Cecchinato, M.; Cox, A.; Bird, J. Smartwatches: The Good, the Bad and the Ugly? ACM: New York, NY, USA,
2015; pp. 2133–2138.
7.
Choi, J.; Kim, S. Is the smartwatch an IT product or a fashion product? A study on factors affecting the
intention to use smartwatches. Comput. Hum. Behav. 2016,63, 777–786. [CrossRef]
8.
Kim, K.; Shin, D. An acceptance model for smart watches Implications for the adoption of future wearable
technology. Internet Res. 2015,25, 527–541. [CrossRef]
9.
Page, T. Barriers to the Adoption of Wearable Technology. i-Manager’s J. Inform. Technol.
2015
,4, 1. [CrossRef]
10.
Chuah, S.; Rauschnabel, P.; Krey, N.; Nguyen, B.; Ramayah, T.; Lade, S. Wearable technologies: The role of
usefulness and visibility in smartwatch adoption. Comput. Hum. Behav. 2016,65, 276–284. [CrossRef]
11.
Wu, L.; Wu, L.; Chang, S. Exploring consumers’ intention to accept smartwatch. Comput. Hum. Behav.
2016
,
64, 383–392. [CrossRef]
12. Jackson, W. Pro Android Wearables: Building Apps for Smartwatches, 1st ed.; Apress: Berkeley, CA, USA, 2015.
13.
Hameed, S. Mastering Android Wear Application Development, 1st ed.; Packt Publishing: Mumbai, India;
Birmingham, UK, 2016.
14.
Mishra, S. Wearable Android: Android Wear & Google Fit App Development, 1st ed.; Wiley: Hoboken, NJ,
USA, 2015.
15.
Negi, I.; Tsow, F.; Tanwar, K.; Zhang, L.; Iglesias, R.A.; Chen, C.; Rai, A.; Forzani, E.S.; Tao, N. Novel monitor
paradigm for real-time exposure assessment. J. Expo. Sci. Environ. Epidemiol.
2011
,21, 419–426. [CrossRef]
[PubMed]
16.
Chan, M.; Estève, D.; Fourniols, J.; Escriba, C.; Campo, E. Smart wearable systems: Current status and future
challenges. Artif. Intell. Med. 2012,56, 137–156. [CrossRef] [PubMed]
17.
Kim, S.; Nussbaum, M. Performance evaluation of a wearable inertial motion capture system for capturing
physical exposures during manual material handling tasks. Ergonomics
2013
,56, 314–326. [CrossRef]
[PubMed]
18.
Peppoloni, L.; Filippeschi, A.; Ruffaldi, E.; Avizzano, C. A novel wearable system for the online assessment
of risk for biomechanical load in repetitive efforts. Int. J. Ind. Ergon. 2016,52, 1–11. [CrossRef]
19.
Aiello, G.; La Scalia, G.; Vallone, M.; Catania, P.; Venticinque, M. Real time assessment of hand-arm vibration
system based on capacitive MEMS accelerometers. Comput. Electron. Agric. 2012,85, 45–52. [CrossRef]
20.
Morello, R.; De Capua, C.; Meduri, A. A Wireless Measurement System for Estimation of Human Exposure to
Vibration During the Use of Handheld Percussion Machines. IEEE Trans. Instrum. Meas.
2010
,59, 2513–2521.
[CrossRef]
21.
Ferreira, B.; Serejo, G.; Ferreira, M.; Ferreira, D.; Cardoso, L.; Yoshidome, E.; Arruda, H.; Lira, W.; Ferreira, J.;
Carvalho, E.; et al. Wearable computing for railway environments: Proposal and evaluation of a safety
solution. IET Intell. Transp. Syst. 2017,11, 319–325. [CrossRef]
22.
Bernal, G.; Colombo, S.; Al Ai Baky, M.; Casalegno, F. Safety: Designing IoT and Wearable Systems for
Industrial Safety through a User Centered Design Approach. In Proceedings of the 10th International
Conference on Pervasive Technologies Related to Assistive Environments, Island of Rhodes, Greece,
21–23 June 2017; ACM: New York, NY, USA, 2017; pp. 163–170.
23.
Lee, W.; Lin, K.; Seto, E.; Migliaccio, G. Wearable sensors for monitoring on-duty and off-duty worker
physiological status and activities in construction. Autom. Constr. 2017,83, 341–353. [CrossRef]
24.
Valero, E.; Sivanathan, A.; Bosche, F.; Abdel-Wahab, M. Analysis of construction trade worker body motions
using a wearable and wireless motion sensor network. Autom. Constr. 2017,83, 48–55. [CrossRef]
Sensors 2018,18, 3805 18 of 20
25.
Yan, X.; Li, H.; Li, A.; Zhang, H. Wearable IMU-based real-time motion warning system for construction
workers’ musculoskeletal disorders prevention. Autom. Constr. 2017,74, 2–11. [CrossRef]
26.
Daponte, P.; De Vito, L.; Picariello, F.; Riccio, M. State of the art and future developments of measurement
applications on smartphones. Meas. J. Int. Meas. Confed. 2013,46, 3291–3307. [CrossRef]
27.
Liu, B.; Koc, A.B. Hand-Arm Vibration Measurements and Analysis Using Smartphones. In Proceedings
of the ASABE Annual International Meeting, Kansas City, MO, USA, 21–24 July 2013; American Society of
Agricultural and Biological Engineers: St. Joseph, MI, USA, 2013.
28.
Cutini, M.; Bisaglia, C. Whole body vibration monitoring using a smartphone. In Proceedings of the
International Conference of Agricultural Engineering (AgEng), Zurich, Switzerland, 6–10 July 2014.
29.
Pancardo, P.; Acosta, F.; Hernandez-Nolasco, J.; Wister, M.; Lopez-de-Ipina, D. Real-Time Personalized
Monitoring to Estimate Occupational Heat Stress in Ambient Assisted Working. Sensors
2015
,15, 16956–16980.
[CrossRef] [PubMed]
30.
Nath, N.; Akhavian, R.; Behzadan, A. Ergonomic analysis of construction worker’s body postures using
wearable mobile sensors. Appl. Ergon. 2017,62, 107–117. [CrossRef] [PubMed]
31.
Wolfgang, R.; Burgess-Limerick, R. Using Consumer Electronic Devices to Estimate Whole-Body Vibration
Exposure. J. Occup. Environ. Hyg. 2014,11, D77–D81. [CrossRef] [PubMed]
32.
Wolfgang, R.; Di Corleto, L.; Burgess-Limerick, R. Can an iPod Touch Be Used to Assess Whole-Body
Vibration Associated with Mining Equipment? Ann. Occup. Hyg.
2014
,58, 1200–1204. [CrossRef] [PubMed]
33.
Kardous, C.A.; Celestina, M. Use of smartphone sound measurement apps for occupational noise assessments.
J. Acoust. Soc. Am. 2015,137, 2292. [CrossRef]
34.
Robinson, D.; Tingay, J. Comparative study of the performance of smartphone-based sound level meter apps,
with and without the application of a
1
2
” IEC-61094-4 working standard microphone, to IEC-61672 standard
metering equipment in the detection of various problematic workplace noise environments. In Proceedings of
the 43rd International Congress on Noise Control Engineering, Melbourne, Australia,
16–19 November 2014
.
35.
Cerqueira, D.; Carvalho, F.; Melo, R.B. Is It Smart to Use Smartphones to Measure Illuminance for
Occupational Health and Safety Purposes? In Proceedings of the International Conference on Applied
Human Factors and Ergonomics, Los Angeles, CA, USA, 17–21 July 2017; Springer: Cham, Switzerland, 2017;
pp. 258–268.
36.
Bieber, G.; Kirste, T.; Urban, B. Ambient Interaction by Smart Watches; ACM: New York, NY, USA, 2012; pp. 1–6.
37.
Matthies, D.; Bieber, G.; Kaulbars, U. AGIS: Automated Tool Detection & Hand-Arm Vibration Estimation Using
an Unmodified Smartwatch; ACM: New York, NY, USA, 2016; pp. 1–4.
38.
Pavón, I.; Sigcha, L.; López, J.; De Arcas, G. Wearable technology usefulness for occupational risk prevention:
Smartwatches for hand-arm vibration exposure assessment. In Proceedings of the Occupational Safety and
Hygiene V: Selected Papers from the International Symposium on Occupational Safety and Hygiene (SHO
2017), Guimarães, Portugal, 10–11 April 2017; CRC Press: Boca Raton, FL, USA, 2017; p. 65.
39.
Pavón, I.; Sigcha, L.; Arezes, P.; Costa, N.; Arcas, G.; López, J. Wearable technology for occupational risk
assessment: Potential avenues for applications. In Occupational Safety and Hygiene VI, Proceedings of the 6th
International Symposium on Occupation Safety and Hygiene (SHO 2018), Guimarães, Portugal, 26–27 March 2018;
CRC Press: Boca Raton, FL, USA, 2018; p. 447.
40.
Podgorski, D.; Majchrzycka, K.; Dabrowska, A.; Gralewicz, G.; Okrasa, M. Towards a conceptual framework
of OSH risk management in smart working environments based on smart PPE, Ambient intelligence and the
Internet of Things technologies. Int. J. Occup. Saf. Ergon. 2017,23, 1–20. [CrossRef] [PubMed]
41.
Li, R.Y.M. Smart Working Environments Using the Internet of Things and Construction Site Safety.
In An Economic Analysis on Automated Construction Safety; Springer: Berlin, Germany, 2018; pp. 137–153.
42.
Knight, J.F.; Bristow, H.W.; Anastopoulou, S.; Baber, C.; Schwirtz, A.; Arvanitis, T.N. Uses of accelerometer
data collected from a wearable system. Pers. Ubiquitous Comput. 2007,11, 117–132. [CrossRef]
43.
Yang, C.; Hsu, Y. A review of accelerometry-based wearable motion detectors for physical activity monitoring.
Sensors 2010,10, 7772–7788. [CrossRef] [PubMed]
44.
Ohtaki, Y.; Susumago, M.; Suzuki, A.; Sagawa, K.; Nagatomi, R.; Inooka, H. Automatic classification of
ambulatory movements and evaluation of energy consumptions utilizing accelerometers and a barometer.
Microsyst. Technol. 2005,11, 1034–1040. [CrossRef]
Sensors 2018,18, 3805 19 of 20
45.
Scott, K.; Browning, R. Occupational physical activity assessment for chronic disease prevention and
management: A review of methods for both occupational health practitioners and researchers. J. Occup.
Environ. Hyg. 2016,13, 451–463. [CrossRef] [PubMed]
46.
Mannini, A.; Intille, S.S.; Rosenberger, M.; Sabatini, A.M.; Haskell, W. Activity Recognition Using a Single
Accelerometer Placed at the Wrist or Ankle. Med. Sci. Sports Exerc.
2013
,45, 2193–2203. [CrossRef] [PubMed]
47.
Bieber, G.; Haescher, M.; Vahl, M. Sensor Requirements for Activity Recognition on Smart Watches; ACM: New
York, NY, USA, 2013; pp. 1–6.
48.
Mortazavi, B.; Nemati, E.; VanderWall, K.; Flores-Rodriguez, H.; Cai, J.; Lucier, J.; Naeim, A.; Sarrafzadeh, M.
Can Smartwatches Replace Smartphones for Posture Tracking? Sensors
2015
,15, 26783–26800. [CrossRef]
[PubMed]
49.
Weiss, G.M.; Timko, J.L.; Gallagher, C.M.; Yoneda, K.; Schreiber, A.J. Smartwatch-based activity recognition:
A machine learning approach. In Proceedings of the 2016 IEEE EMBS International Conference on Biomedical
and Health Informatics (BHI), Las Vegas, NV, USA, 24–27 February 2016; pp. 426–429.
50.
Ramos, F.; Moreira, A.; Costa, A.; Rolim, R.; Almeida, H.; Perkusich, A. Combining Smartphone and
Smartwatch Sensor Data in Activity Recognition Approaches: An Experimental Evaluation. In Proceedings of
the International Conference on Software Engineering & Knowledge Engineering; SEKE: San Francisco Bay, CA,
USA, 2016; pp. 267–272.
51.
Pannurat, N.; Thiemjarus, S.; Nantajeewarawat, E. Automatic Fall Monitoring: A Review. Sensors
2014
,
14, 12900–12936. [CrossRef] [PubMed]
52.
Casilari, E.; Luque, R.; Moron, M. Analysis of Android Device-Based Solutions for Fall Detection. Sensors
2015,15, 17827–17894. [CrossRef] [PubMed]
53.
Fang, Y.; Dzeng, R. Accelerometer-based fall-portent detection algorithm for construction tiling operation.
Autom. Constr. 2017,84, 214–230. [CrossRef]
54.
Gjoreski, M.; Gjoreski, H.; Lustrek, M.; Gams, M. How Accurately Can Your Wrist Device Recognize Daily
Activities and Detect Falls? Sensors 2016,16, 800. [CrossRef] [PubMed]
55.
Casilari, E.; Oviedo-Jiménez, M.A. Automatic Fall Detection System Based on the Combined Use of a
Smartphone and a Smartwatch. PLoS ONE 2015, 10. [CrossRef] [PubMed]
56. Westbrook, M.H.; Turner, J.D. Automotive Sensors; CRC Press: Bristol, UK, 1994.
57.
Jean-Michel, S. Market opportunities for advanced MEMS accelerometers and overview of actual capabilities
vs. required specifications. In Proceedings of the PLANS 2004, Position Location and Navigation Symposium
(IEEE Cat. No.04CH37556), Monterey, CA, USA, 26–29 April 2004; pp. 78–82.
58.
Mohd-Yasin, F.; Korman, C.E.; Nagel, D.J. Measurement of noise characteristics of MEMS accelerometers.
Solid State Electron. 2003,47, 357–360. [CrossRef]
59.
Albarbar, A.; Badri, A.; Sinha, J.K.; Starr, A. Performance evaluation of MEMS accelerometers. Measurement
2009,42, 790–795. [CrossRef]
60.
Ghisi, A.; Kalicinski, S.; Mariani, S.; De Wolf, I.; Corigliano, A. Polysilicon MEMS accelerometers exposed to
shocks: Numerical–experimental investigation. J. Micromech. Microeng. 2009,19, 035023. [CrossRef]
61.
Mariani, S.; Ghisi, A.; Corigliano, A.; Zerbini, S. Multi-scale analysis of MEMS sensors subject to drop
impacts. Sensors 2007,7, 1817–1833. [CrossRef] [PubMed]
62.
Ratcliffe, C.; Heider, D.; Crane, R.; Krauthauser, C.; Yoon, M.K.; Gillespie, J.W. Investigation into the use
of low cost MEMS accelerometers for vibration based damage detection. Compos. Struct.
2008
,82, 61–70.
[CrossRef]
63.
Xiong, X.; Wu, Y.; Jone, W.B. Material Fatigue and Reliability of MEMS Accelerometers. In Proceedings of
the 2008 IEEE International Symposium on Defect and Fault Tolerance of VLSI Systems, Boston, MA, USA,
1–3 Octber 2008; pp. 314–322.
64. Mohd-Yasin, F.; Nagel, D.J.; Korman, C.E. Noise in MEMS. Meas. Sci. Technol. 2010,21, 012001. [CrossRef]
65.
Tarabini, M.; Saggin, B.; Scaccabarozzi, D.; Moschioni, G. The potential of micro-electro-mechanical
accelerometers in human vibration measurements. J. Sound Vib. 2012,331, 487–499. [CrossRef]
66. ISO 8041:2005. Human Response to Vibration—Measuring Instrumentation; ISO: Geneva, Switzerland, 2005.
67.
Brandt, A. Noise and Vibration Analysis: Signal Analysis and Experimental Procedures; Wiley: Chichester,
UK, 2011.
68.
ISO 16063-21:2003. Methods for the Calibration of Vibration and Shock Transducers–Part 21: Vibration Calibration
by Comparison to a Reference Transducer; ISO: Geneva, Switzerland, 2003.
Sensors 2018,18, 3805 20 of 20
69.
Sigcha, L. Soluciones Basadas en Tecnología Wearable Para la Valoración de Vibraciones en el Puesto de
Trabajo Que Afectan Al Sistema Mano Brazo. Master ’s Thesis, Universidad Politécnica de Madrid, Madrid,
Spain, 2017.
70.
Smith, S.W. Digital Signal Processing: A Practical Guide for Engineers and Scientists; Newnes: Amsterdam,
The Netherlands, 2003.
71. ISO 266:1997. Acoustics—Preferred Frequencies; ISO: Geneva, Switzerland, 1997.
72.
ISO/IEC Guide 98-3:2008. Uncertainty of Measurement—Part 3: Guide to the Expression of Uncertainty in
Measurement (GUM:1995); ISO: Geneva, Switzerland, 2008.
73.
Centro Español de Metrología AC-001 Procedimiento Para la Calibración de Acelerómetros 1999. Available
online: http://www.cem.es/divulgacion/publicaciones/ac-001-procedimiento-para-la-calibración-de-
acelerómetros (accessed on 1 July 2018).
74.
JCGM. Evaluation of Measurement Data—Supplement 1 to the “Guide to the Expression of Uncertainty in
Measurement”—Propagation of Distributions Using a Monte Carlo Method. 2008. Available online: https:
//www.bipm.org/utils/common/documents/jcgm/JCGM_101_2008_E.pdf (accessed on 25 October 2018).
75.
Brannon, B.; Barlow, A. Aperture Uncertainty and ADC System Performance. Application Note AN 2006,
501. Available online: http://www.analog.com/media/en/technical- documentation/application-notes/an
-501.pdf (accessed on 11 June 2018).
76.
Sensors Overview Sensor Coordinate System. Available online: https://developer.android.com/guide/to
pics/sensors/sensors_overview.html (accessed on 9 April 2018).
77.
Concettoni, E.; Griffin, M. The apparent mass and mechanical impedance of the hand and the transmission
of vibration to the fingers, hand, and arm. J. Sound Vib. 2009,325, 664–678. [CrossRef]
78.
Habib, M.; Mohktar, M.; Kamaruzzaman, S.; Lim, K.; Pin, T.; Ibrahim, F. Smartphone-Based Solutions for Fall
Detection and Prevention: Challenges and Open Issues. Sensors 2014,14, 7181–7208. [CrossRef] [PubMed]
79.
Mellone, S.; Tacconi, C.; Schwickert, L.; Klenk, J.; Becker, C.; Chiari, L. Smartphone-based solutions for
fall detection and prevention: The FARSEEING approach. Zeitschrift für Gerontologie und Geriatrie
2012
,
45, 722–727. [CrossRef] [PubMed]
80.
Abbate, S.; Avvenuti, M.; Bonatesta, F.; Cola, G.; Corsini, P.; Vecchio, A. A smartphone-based fall detection
system. Pervasive Mob. Comput. 2012,8, 883–899. [CrossRef]
81.
Fudickar, S.; Lindemann, A.; Schnor, B. Threshold-Based Fall Detection on Smart Phones. In Proceedings of
the International Conference on Health Informatics, Angers, France, 3–6 March 2014; pp. 303–309.
82.
Moschioni, G.; Saggin, B.; Tarabini, M. Prediction of data variability in hand-arm vibration measurements.
Measurement 2011,44, 1679–1690. [CrossRef]
83.
Laput, G.; Xiao, R.; Harrison, C. ViBand: High-Fidelity Bio-Acoustic Sensing Using Commodity Smartwatch
Accelerometers; ACM: New York, NY, USA, 2016; pp. 321–333.
©
2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access
article distributed under the terms and conditions of the Creative Commons Attribution
(CC BY) license (http://creativecommons.org/licenses/by/4.0/).
Available via license: CC BY
Content may be subject to copyright.
Content uploaded by Nelson Costa
Author content
All content in this area was uploaded by Nelson Costa on Nov 09, 2018
Content may be subject to copyright.