Available via license: CC BY
Content may be subject to copyright.
sensors
Article
Validation of the Apple Watch for Heart Rate
Variability Measurements during Relax and Mental
Stress in Healthy Subjects
David Hernando 1, 2, *ID , Surya Roca 3, Jorge Sancho 3ID ,Álvaro Alesanco 3ID
and Raquel Bailón1,2
1Biomedical Signal Interpretation & Computational Simulation (BSICoS) Group, Aragón Institute of
Engineering Research (I3A), IIS Aragón, University of Zaragoza, 50018 Zaragoza, Spain; rbailon@unizar.es
2Centro de Investigación Biomédica en Red en Bioingeniería, Biomateriales y Nanomedicina (CIBER-BBN),
28029 Madrid, Spain
3
Communications Networks and Information Technologies (CeNIT) Group, Aragón Institute of Engineering
Research (I3A), University of Zaragoza, 50018 Zaragoza, Spain; surya@unizar.es (S.R.);
jslarraz@unizar.es (J.S.); alesanco@unizar.es (Á.A.)
*Correspondence: dhernand@unizar.es; Tel.: +34-876-555462
Received: 4 July 2018; Accepted: 8 August 2018; Published: 10 August 2018
Abstract:
Heart rate variability (HRV) analysis is a noninvasive tool widely used to assess autonomic
nervous system state. The market for wearable devices that measure the heart rate has grown
exponentially, as well as their potential use for healthcare and wellbeing applications. Still, there is a
lack of validation of these devices. In particular, this work aims to validate the Apple Watch in terms
of HRV derived from the RR interval series provided by the device, both in temporal (HRM (mean
heart rate), SDNN, RMSSD and pNN50) and frequency (low and high frequency powers, LF and HF)
domain. For this purpose, a database of 20 healthy volunteers subjected to relax and a mild cognitive
stress was used. First, RR interval series provided by Apple Watch were validated using as reference
the RR interval series provided by a Polar H7 using Bland-Altman plots and reliability and agreement
coefficients. Then, HRV parameters derived from both RR interval series were compared and their
ability to identify autonomic nervous system (ANS) response to mild cognitive stress was studied.
Apple Watch measurements presented very good reliability and agreement (>0.9). RR interval series
provided by Apple Watch contain gaps due to missing RR interval values (on average, 5 gaps per
recording, lasting 6.5 s per gap). Temporal HRV indices were not significantly affected by the gaps.
However, they produced a significant decrease in the LF and HF power. Despite these differences,
HRV indices derived from the Apple Watch RR interval series were able to reflect changes induced
by a mild mental stress, showing a significant decrease of HF power as well as RMSSD in stress with
respect to relax, suggesting the potential use of HRV measurements derived from Apple Watch for
stress monitoring.
Keywords:
heart rate variability; wearable device; Apple Watch; RR series; validation; ANS assessment
1. Introduction
Heart rate variability (HRV) analysis is a widely accepted tool for the noninvasive assessment of
autonomic nervous system (ANS) [
1
]. Its analysis and use are becoming increasingly common, since it
is sensitive to both physiological and psychological changes [
2
]. Altered HRV measurements have
been reported in several diseases related to ANS dysregulation, including cardiovascular diseases,
such as ischemia, myocardial infarction and heart failure [
1
,
3
], metabolic diseases, such as diabetes and
Sensors 2018,18, 2619; doi:10.3390/s18082619 www.mdpi.com/journal/sensors
Sensors 2018,18, 2619 2 of 11
obesity [
3
], and mental disorders, such as anxiety and depression [
4
,
5
]. Moreover, HRV measurements
have also been used to monitor sleep [6], stress [7,8], drowsiness [9] and exercise training [10–12].
Over the past 10 years, the market for health related wearable devices has grown exponentially,
as well as their potential use for healthcare and wellbeing applications. The most popular and
accepted ones, especially among the non-patient population, are the watch-like devices, which
usually provide average heart rate (HR) estimates, derived from the pulse signal recorded at the
wrist. They are based on photophethysmography (PPG), including a light source for illuminating
the skin and a photodetector, which measures the intensity of the reflected light. HR estimates are
based on the pulsatile changes in reflected light induced by fluctuations in blood flow every heartbeat.
Although technically feasible, only a small subset of these devices allows HRV measures. As pointed
out in [
13
], these devices usually lack from proper validation, limiting its applicability for enhancing
clinical care.
Different wearable devices have been evaluated and compared for HR estimation. In [
13
] the
Apple Watch achieved the best performance estimating HR at a one minute granularity, with median
error below 3% during sitting, walking, running, and cycling. However, neither of these devices has
been validated for HRV estimation. Although pulse rate variability (PRV) derived from PPG has been
validated as a surrogate of HRV derived from the electrocardiogram (ECG) [
14
], it is not the case for
PRV derived from the wrist. Only a few studies have addressed this issue, such as [
15
], where wrist
PPG-based HRV was evaluated for Emphatica E4 (Empatica Srl, Milano, Italy) and PulseOn (PulseOn
Oy, Espoo, Finland) devices, yielding reliable results only during sitting condition.
The objective of this study is to validate the Apple Watch for HRV measurement. Prior studies have
only validated aggregate HR estimates [
13
,
16
–
19
], but its outperformance over other devices [
13
,
20
]
makes it a potential candidate among wrist-worn devices for HRV measurement. Moreover, its spread
and acceptance are beating the wearable market [
21
]. The present study sets out the comparison of RR
intervals derived from Apple Watch using as reference the RR intervals derived from the Polar H7 chest
belt heart rate monitor (Polar Electro Oy, Kempele, Finland), which has been already validated [
22
,
23
].
Additionally, this study includes the validation of HRV measurements, derived from the RR intervals
from both devices, in two scenarios with different ANS balance: relax and metal stress.
2. Materials and Methods
2.1. Experimental Setup
A total of 20 volunteers agreed to participate in the study. All of them were apparently healthy
subjects. Written informed consent was obtained from each subject. The study protocol was approved
by the institutional ethics committee and was in accordance with the Declaration of Helsinki for
Human Research of 1974 (last modified in 2008).
Participants were asked to abstain from caffeine and smoking prior to the test and to only consume
a light meal 2 h prior to testing. The Polar H7 chest belt was placed tightly just below the chest muscles
and the Apple Watch was placed tightly as well in the left wrist. Both devices were cleaned up between
uses to avoid possible misreading effects. Measures were taken while participants were sitting in front
of a 21” computer screen, completing relax and mental stress test protocols. Instructions were given to
minimize left arm movements, where the Apple Watch was placed.
The experimental protocol consisted of two consecutive 5-min stages, namely relax and stress
stages, separated by a 1-min break. During the relax stage volunteers watched a relaxing video,
consisting in relaxing music and pleasant images, which was previously used in a study on emotion
recognition [
24
]. During the stress stage, participants performed an online version of the Stroop test
(www.onlinestrooptest.com), which is an attentional test in which participants see words (names of
different colors) written in colored ink and they are asked to name the color of the ink while ignoring
the meaning of the word.
Sensors 2018,18, 2619 3 of 11
Raw RR data was obtained from Polar H7 (
RRH7
) using Elite HRV app installed in an Android
device. This app is able to export a text file containing the raw RR interval series with a precision
of milliseconds. On the other hand, the Breath app was used to extract RR interval series from
Apple Watch (
RRAW
). Breath app, developed by Apple, is the only way at this moment (watchOS
4.2 operating system) to obtain RR raw values since Apple does not include any programing method
for developers to directly access the values. This app stores the raw RR values, with a precision of
centiseconds, in the user’s Personal Health Record, accessible to be exported in XML format using
Apple’s Health App.
2.2. Synchronization and RR Matching
Figure 1shows an example of both RR series,
RRH7
and
RRAW
. It can be seen that the RR series
derived from the Apple Watch has some gaps, probably corresponding to time instants where Apple
Watch proprietary algorithms have not been able to reliably detect the PPG pulses. These gaps were
present in almost all Apple Watch recordings, with no apparent relation with the subject or the test
stage (relax or stress). Synchronization between
RRH7
and
RRAW
was performed using the delay
that maximized their cross correlation using the first 20 RR intervals, where no gaps appeared in
Apple Watch recordings. All RR intervals need to be matched from both devices: we create
RRH7g
by
intentionally creating the same gaps in
RRH7
than those in
RRAW
. This way, both
RRAW
and
RRH7g
have the same number of samples and all RR intervals are matched. For each recording, the percentage
of missing RR intervals missed with respect to the total number of RR intervals were noted.
Sensors 2018, 18, x 3 of 11
Raw RR data was obtained from Polar H7 () using Elite HRV app installed in an Android
device. This app is able to export a text file containing the raw RR interval series with a precision of
milliseconds. On the other hand, the Breath app was used to extract RR interval series from Apple
Watch ( ). Breath app, developed by Apple, is the only way at this moment (watchOS 4.2
operating system) to obtain RR raw values since Apple does not include any programing method for
developers to directly access the values. This app stores the raw RR values, with a precision of
centiseconds, in the user’s Personal Health Record, accessible to be exported in XML format using
Apple’s Health App.
2.2. Synchronization and RR Matching
Figure 1 shows an example of both RR series, and . It can be seen that the RR series
derived from the Apple Watch has some gaps, probably corresponding to time instants where
Apple Watch proprietary algorithms have not been able to reliably detect the PPG pulses. These
gaps were present in almost all Apple Watch recordings, with no apparent relation with the subject
or the test stage (relax or stress). Synchronization between and was performed using
the delay that maximized their cross correlation using the first 20 RR intervals, where no gaps
appeared in Apple Watch recordings. All RR intervals need to be matched from both devices: we
create by intentionally creating the same gaps in than those in . This way, both
and have the same number of samples and all RR intervals are matched. For each
recording, the percentage of missing RR intervals missed with respect to the total number of RR
intervals were noted.
Figure 1. Example of the RR series for a subject in both stages: Polar H7 (blue) and Apple Watch (red).
In the stress stage, there are gaps in the Apple Watch recording where no beats are detected.
2.3. Validation of RR Series
A Bland–Altman plot [25] was used to study the validity of the Apple Watch RR series, by
representing both and series. The bias, the limits of agreement (LOA, ±2*std values)
and the percentage of paired RR measurements out of the LOA were also obtained.
To measure the interchangeability between both RR series, two reliability indices were used. First,
Lin’s concordance correlation coefficient (CCC) determines how far the observed data deviate from the
line of perfect concordance line at 45° on a square axis scatter plot [26]. Second, intraclass correlation
coefficient (ICC) represents the ratio of between-sample variance and the total variance (between- and
within-sample) to measure precision under the model of equal marginal distributions [27]. Lastly, the
Figure 1.
Example of the RR series for a subject in both stages: Polar H7 (blue) and Apple Watch (red).
In the stress stage, there are gaps in the Apple Watch recording where no beats are detected.
2.3. Validation of RR Series
A Bland–Altman plot [
25
] was used to study the validity of the Apple Watch RR series, by
representing both
RRH7g
and
RRAW
series. The bias, the limits of agreement (LOA,
±
2*std values)
and the percentage of paired RR measurements out of the LOA were also obtained.
To measure the interchangeability between both RR series, two reliability indices were
used. First, Lin’s concordance correlation coefficient (CCC) determines how far the observed
data deviate from the line of perfect concordance line at 45
◦
on a square axis scatter plot [
26
].
Second, intraclass correlation coefficient (ICC) represents the ratio of between-sample variance and
the total variance (between- and within-sample) to measure precision under the model of equal
Sensors 2018,18, 2619 4 of 11
marginal distributions [
27
]. Lastly, the agreement was measured by an information-based measure
of disagreement (IBMD), which is based on Shannon’s entropy [
28
]. This index equals 0 when the
observers agree (no disagreement); i.e., there is no information in the differences between both methods.
The agreement (A) can be quantified as A = 1−IBMD.
2.4. Heart Rate Variability Parameters
Heart rate variability parameters were divided in two classes: time and frequency domain indices.
Since both are calculated in different ways, the gaps also are addressed differently. For the temporal
domain analysis, we will derive HRV parameters from
RRAW
,
RRH7
and
RRH7g
, the latter created
with simulated gaps as defined in Section 2.2. For the frequency domain analysis, most of the methods
used for power spectral density estimation require evenly sampled series, and interpolation is usually
employed to overcome the intrinsic unevenly sampled RR interval series. However, since gaps in
RRAW
and
RRH7g
are too large (from 3.3 to 10.4 s), we first filled the gaps with artificial RR intervals,
created by linear interpolation. Note that, for simplicity, we use the same nomenclature,
RRAW
and
RRH7g
, for temporal and frequency domain analysis, although they are slightly different: for temporal
HRV parameters, the gaps are omitted and the RR intervals are concatenated (
RRAW
and
RRH7g
are
shorter than
RRH7
), while for frequency HRV parameters the gaps are filled with artificially created
RR intervals (RRAW and RRH7gare the same length than RRH7).
For the temporal parameters, four parameters were proposed [
1
]: HRM, SDNN, RMSSD and
pNN50. HRM is the mean heart rate, which is obtained as the inverse of the mean heart period (mean
of the RR intervals). SDNN is the standard deviation of the NN intervals (or normal RR intervals),
which is a measure of the total power in the analyzed segment. A dispersion measure can be obtained
by calculating the root mean-square of successive differences of adjacent intervals (RMSSD). The last
parameter, pNN50, represents the percentage of pairs of adjacent intervals differing by more than
50 ms. RMSSD and pNN50 describe short-time variability. Note that whenever a gap appears in
RRAW
, the corresponding intervals in
RRH7
are also removed, and the remaining RR intervals are
concatenated. The difference of the intervals just before and after the gaps are not included in the
calculation of RMSSD and pNN50.
For the frequency domain analysis, the heart rate variability signal was obtained following a
method based on the integral pulse frequency modulation (IPFM) model described in [
29
]: from the
RR interval series, we derived the instantaneous heart rate signal,
dHR (n)
, which is a continuous
and evenly sampled signal (sampled at 4 Hz). Then, we obtained the mean heart rate,
dHR M(n)
,
using a low pass filter up to 0.03 Hz. The variability signal,
dHRV (n)
, was obtained as
dHRV (n)=
dHR (n)−dHRM (n)
. Finally, the modulating signal, which is assumed to represent the ANS modulation
on the sinoatrial node, was obtained as ˆ
m(n)=dHRV (n)/dHR M(n)[30].
The Welch periodogram was applied to estimate the spectral properties of the HRV signals
ˆ
mAW(n)
,
ˆ
mH7(n)
and
ˆ
mH7g(n)
(derived from
RRAW
,
RRH7
and
RRH7g
, respectively), using a
Hamming window with a length of 150 s with an overlap of 120 s. Low frequency (LF) and high
frequency (HF) powers,
PLF
and
PHF
respectively, were obtained integrating the power in their bands
(LF: 0.04–0.15 Hz, HF: 0.15–0.4 Hz) [1].
2.5. Statistical Analysis
None of all HRV parameters followed a normal distribution (tested with a Kolmogorov test).
To compare HRV parameters, a paired Wilcoxon test was applied. Then we performed 3 different
comparisons: (1) we analyzed the influence of the gaps by comparing HRV parameters derived from
RRH7
and
RRH7g
; (2) we compared HRV parameters derived from
RRAW
and
RRH7g
; and (3) HRV
parameters derived from
RRAW
were compared in relax vs. stress stage to evaluate the ability of those
parameters to reflect ANS response to stress, using changes in HRV parameters derived from
RRH7
as the reference ANS response. The difference was considered to be significantly different from zero
when p< 0.05.
Sensors 2018,18, 2619 5 of 11
Distribution of HRV parameters are shown using boxplots: on each box, the central mark indicates
the median, and the bottom and top edges of the box indicate the 25th and 75th percentiles, respectively.
The whiskers extend to the most extreme data points not considered outliers, and the outliers are
plotted individually using the ‘+’ symbol. An outlier is defined as a point that falls more than 1.5
times the interquartile range (difference between 75th and 25th percentiles) above the 75th percentile
or below the 25th percentile.
3. Results
3.1. Validity of RR Series
A total of 206 gaps were found in the Apple Watch recordings, equivalent to 1321 missing RR
intervals (around 10% of total intervals). On average, each recording presents 5 gaps, with a length of
6 s per gap. The minimum length found was 3.3 s and the maximum length was 10.4 s. No differences
in the number or length of these gaps were found between relax and stress recordings.
Figure 2shows the Bland–Altman plot which evaluates the RR interval discrepancies between
Polar H7 and Apple Watch measurements and the stability across the different values of the intervals.
A total of 12,109 paired RR intervals were used from both relax and stress stages. The bias (central
horizontal line) was 0.06 ms. The LOA (upper and lower lines) contained 96.21% of the intervals.
No visual differences were found in the discrepancies between lower and higher RR intervals.
Sensors 2018, 18, x 5 of 11
respectively. The whiskers extend to the most extreme data points not considered outliers, and the
outliers are plotted individually using the ‘+’ symbol. An outlier is defined as a point that falls more
than 1.5 times the interquartile range (difference between 75th and 25th percentiles) above the 75th
percentile or below the 25th percentile.
3. Results
3.1. Validity of RR Series
A total of 206 gaps were found in the Apple Watch recordings, equivalent to 1321 missing RR
intervals (around 10% of total intervals). On average, each recording presents 5 gaps, with a length
of 6 s per gap. The minimum length found was 3.3 s and the maximum length was 10.4 s. No
differences in the number or length of these gaps were found between relax and stress recordings.
Figure 2 shows the Bland–Altman plot which evaluates the RR interval discrepancies between
Polar H7 and Apple Watch measurements and the stability across the different values of the intervals.
A total of 12,109 paired RR intervals were used from both relax and stress stages. The bias (central
horizontal line) was 0.06 ms. The LOA (upper and lower lines) contained 96.21% of the intervals. No
visual differences were found in the discrepancies between lower and higher RR intervals.
Figure 2. Bland-Altman plot: vs. . Mean of the difference of the RR series ±2*std values
(limits of agreement, LOA).
Table 1 shows the mean values of the RR intervals for each stage, as well as the reliability and
agreement indices. While slightly lower in the stress stage, these indices show excellent reliability
and agreement between the measurements in both stages. Bias and LOA are also shown for relax and
stress stages, which are very similar to those obtained with the combined data. The last row shows
the mean of the percentage of missing RR intervals in .
Table 1. Validity of RR series from Apple Watch. CI = Confidence interval.
RELAX STRESS
Mean (SD) H7 RR intervals (ms) 869.28 (114.01) 834.78 (97.43)
Mean (SD) AW RR intervals (ms) 869.23 (114.39) 834.70 (97.84)
CCC (90% CI) 0.989 (0.981, 0.998) 0.977 (0.970, 0.985)
ICC (90% CI) 0.989 (0.984, 0.996) 0.982 (0.977, 0.987)
A (90% CI) 0.993 (0.987, 0.999) 0.983 (0.975, 0.991)
Bias (Out LOA) 0.049 (3.29%) 0.078 (3.93%)
Mean (SD)% missed RR intervals 10.98 (9.78) 9.45 (7.30)
Figure 2.
Bland-Altman plot:
RRH7gvs
.
RRAW
. Mean of the difference of the RR series
±
2*std values
(limits of agreement, LOA).
Table 1shows the mean values of the RR intervals for each stage, as well as the reliability and
agreement indices. While slightly lower in the stress stage, these indices show excellent reliability and
agreement between the measurements in both stages. Bias and LOA are also shown for relax and stress
stages, which are very similar to those obtained with the combined data. The last row shows the mean
of the percentage of missing RR intervals in RRAW.
Table 1. Validity of RR series from Apple Watch. CI = Confidence interval.
RELAX STRESS
Mean (SD) H7 RR intervals (ms) 869.28 (114.01) 834.78 (97.43)
Mean (SD) AW RR intervals (ms) 869.23 (114.39) 834.70 (97.84)
CCC (90% CI) 0.989 (0.981, 0.998) 0.977 (0.970, 0.985)
ICC (90% CI) 0.989 (0.984, 0.996) 0.982 (0.977, 0.987)
A (90% CI) 0.993 (0.987, 0.999) 0.983 (0.975, 0.991)
Bias (Out LOA) 0.049 (3.29%) 0.078 (3.93%)
Mean (SD)% missed RR intervals 10.98 (9.78) 9.45 (7.30)
Sensors 2018,18, 2619 6 of 11
3.2. HRV Parameters: Temporal Domain
Figure 3shows the temporal parameters for both relax and stress stages. No significant differences
were found in any parameter neither between RRH7and RRH7g, nor between RRH7gand RRAW.
Sensors 2018, 18, x 6 of 11
3.2. HRV Parameters: Temporal Domain
Figure 3 shows the temporal parameters for both relax and stress stages. No significant
differences were found in any parameter neither between and , nor between and
.
Figure 3. Heart rate variability (HRV) parameters: time domain.
3.3. HRV Parameters: Frequency Domain
Figure 4 shows the frequency HRV parameters derived from , and in relax
and stress stages. Both and show significant differences (marked with an asterisk between
the boxplots) in when comparing to , being lower in the presence of the gaps. When
comparing HRV parameters derived from and , neither nor presented
significant differences.
Figure 4. HRV parameters: frequency domain, derived from , and . * denotes
significant differences (p < 0.05) between adjacent boxplots. Adim refers to adimensional units.
Figure 3. Heart rate variability (HRV) parameters: time domain.
3.3. HRV Parameters: Frequency Domain
Figure 4shows the frequency HRV parameters derived from
RRH7
,
RRH7g
and
RRAW
in
relax and stress stages. Both
PLF
and
PHF
show significant differences (marked with an asterisk
between the boxplots) in
RRH7g
when comparing to
RRH7
, being lower in the presence of the gaps.
When comparing HRV parameters derived from
RRAW
and
RRH7g
, neither
PLF
nor
PHF
presented
significant differences.
Sensors 2018, 18, x 6 of 11
3.2. HRV Parameters: Temporal Domain
Figure 3 shows the temporal parameters for both relax and stress stages. No significant
differences were found in any parameter neither between and , nor between and
.
Figure 3. Heart rate variability (HRV) parameters: time domain.
3.3. HRV Parameters: Frequency Domain
Figure 4 shows the frequency HRV parameters derived from , and in relax
and stress stages. Both and show significant differences (marked with an asterisk between
the boxplots) in when comparing to , being lower in the presence of the gaps. When
comparing HRV parameters derived from and , neither nor presented
significant differences.
Figure 4. HRV parameters: frequency domain, derived from , and . * denotes
significant differences (p < 0.05) between adjacent boxplots. Adim refers to adimensional units.
Figure 4.
HRV parameters: frequency domain, derived from
RRH7
,
RRH7g
and
RRAW
. * denotes
significant differences (p< 0.05) between adjacent boxplots. Adim refers to adimensional units.
Sensors 2018,18, 2619 7 of 11
3.4. HRV Parameters: Relax vs. Stress
Figures 5and 6display HRV parameters derived from
RRAW
and
RRH7
in relax vs. stress stages.
All temporal HRV parameters (HRM, SDNN, RMSSD and pNN50) presented changes from relax to
stress in both Apple Watch and Polar H7 recordings: an increase in HRM and a decrease in the rest of
parameters. We observed a decrease in
PLF
in stress with respect to relax stage, being only statistically
significant in Polar H7 recordings (p= 0.030 for Polar H7, p= 0.065 for Apple Watch). A significant
decrease was found in both devices when comparing PHF.
Sensors 2018, 18, x 7 of 11
3.4. HRV Parameters: Relax vs. Stress
Figures 5 and 6 display HRV parameters derived from and in relax vs. stress
stages. All temporal HRV parameters (HRM, SDNN, RMSSD and pNN50) presented changes from
relax to stress in both Apple Watch and Polar H7 recordings: an increase in HRM and a decrease in
the rest of parameters. We observed a decrease in in stress with respect to relax stage, being only
statistically significant in Polar H7 recordings (p = 0.030 for Polar H7, p = 0.065 for Apple Watch). A
significant decrease was found in both devices when comparing .
Figure 5. HRV parameters: temporal domain (Relax vs. Stress). * denotes significant differences
(p < 0.05) between adjacent boxplots.
Figure 6. HRV parameters: frequency domain (Relax vs. Stress). * denotes significant differences (p <
0.05) between adjacent boxplots. Adim refers to adimensional units.
Figure 5.
HRV parameters: temporal domain (Relax vs. Stress). * denotes significant differences (p<
0.05) between adjacent boxplots.
Sensors 2018, 18, x 7 of 11
3.4. HRV Parameters: Relax vs. Stress
Figures 5 and 6 display HRV parameters derived from and in relax vs. stress
stages. All temporal HRV parameters (HRM, SDNN, RMSSD and pNN50) presented changes from
relax to stress in both Apple Watch and Polar H7 recordings: an increase in HRM and a decrease in
the rest of parameters. We observed a decrease in in stress with respect to relax stage, being only
statistically significant in Polar H7 recordings (p = 0.030 for Polar H7, p = 0.065 for Apple Watch). A
significant decrease was found in both devices when comparing .
Figure 5. HRV parameters: temporal domain (Relax vs. Stress). * denotes significant differences
(p < 0.05) between adjacent boxplots.
Figure 6. HRV parameters: frequency domain (Relax vs. Stress). * denotes significant differences (p <
0.05) between adjacent boxplots. Adim refers to adimensional units.
Figure 6.
HRV parameters: frequency domain (Relax vs. Stress). * denotes significant differences
(p< 0.05) between adjacent boxplots. Adim refers to adimensional units.
Sensors 2018,18, 2619 8 of 11
4. Discussion
This is the first study to validate HRV measurements using an Apple Watch to the authors’
knowledge. Many of the wrist-worn devices aimed at assessing well-being, stress, and fitness provide
average HR values, derived from a pulse signal, instead of HRV. Fitbit, Jawbone, TomTom, Garming,
Withing, Samsumg, among others, commercialize such devices. However, although changes in
average HR are mainly induced by ANS and can be significantly different in some physio-pathological
situations, it cannot be considered a measure of autonomic function [
3
,
31
]. Moreover, there are
situations where altered ANS function is manifested in HRV but not in HR changes, such as in
depressed patients with respect to controls [32] or in exercise contexts [33,34].
In order to validate HRV measurements derived from an Apple Watch, in this work an
experimental study was conducted on young healthy volunteers subjected to mild relax and stress.
HRV measurements derived from Apple Watch were compared to those obtained from a validated
heart rate monitor, in particular Polar H7 chest band. Several works in the literature have already
validated, using as reference the RR interval series derived from a synchronous ECG, the RR interval
series given by different models of Polar devices: S810 [
35
,
36
], RS800 [
37
,
38
], and V800 with a H7
chest belt [22,23].
Regarding the RR intervals series, which are the basis for any HRV analysis, the bias was almost 0
and most pairs of RR intervals (around 96%) were within the limits of agreement in the Bland Altman
plot. These results are consistent when studying both stages separately. Moreover, the discrepancy
between Apple Watch and Polar H7 is similar with no dependency of the heart rate. The RR series
present excellent reliability and agreement when analyzing matched pairs of RR intervals. These results
agree with other studies for Apple Watch [17,18] and other heart rate watches [13,16,19].
However, it is important to note the existence in Apple Watch measurements of missing beats.
Around 10% of the beats could not be detected by the Apple Watch in both the relax and stress stages.
This could be not problematic if they were isolated beats, but they usually were consecutive beats,
which lead to gaps in the RR series. The shorter gaps were longer than 3 s, being easily identifiable as
abnormally large RR intervals. The origin of these gaps could be varied, from bad skin contact to fast
arm movement, but ultimately, we cannot conclude anything about these gaps because we do not have
direct access to the algorithms used by the Apple Watch. In [
18
], they also reported missing values
in the Apple Watch, with the proportion of heart rate values actually measured by the Apple Watch
decreasing with increasing exercise intensity. They also reported that these missing heart rate values
were higher in the first minute of exercise (between 20 and 40% of RR intervals), particularly at higher
intensities of exercise. In our data, however, we did not find more missing beats in the stress stage
compared to the relax stage, and the total missing RR intervals was about 10% of total intervals.
These missing beats are important when extracting HRV parameters in the frequency domain,
since the RR interval series needs to be interpolated in the gaps. This might influence HRV frequency
parameters. To study the influence of these gaps in frequency parameters
PLF
and
PHF
, we created
RRH7g
by intentionally creating gaps in the
RRH7
series, removing the missing RR intervals found
in
RRAW
. This analysis showed that both
PLF
and
PHF
were significantly lower in the presence of
simulated gaps. However, when comparing HRV frequency parameters derived from
RRAW
and from
RRH7g
, no significant difference was found, suggesting that differences between HRV parameters
derived from RR series provided by Apple Watch and Polar H7 are indeed due to the presence of gaps
in the Apple Watch series, rather than to the different signal from which they are derived or to the
different time resolution. The effect of gaps on HRV temporal parameters was also investigated, but no
significant differences were found, suggesting that these parameters are less sensitive to the presence
of gaps.
Despite these differences, HRV measurements derived from Apple Watch were still able to
reflect the ANS response induced by the stress stage, replicating most of the trends observed in HRV
measurements derived from the Polar H7. Induced mental stress resulted in a significant decrease
in
PHF
with respect to relax (in both the Apple Watch and Polar H7 measures, note that there were
Sensors 2018,18, 2619 9 of 11
no gaps in the Polar H7 recordings), suggesting the inhibition of parasympathetic stimulation, as
already reported in numerous studies [
8
,
39
]. A similar conclusion can be extracted with the decrease in
RMSSD parameter in the stress phase, being this parameter associated to the parasympathetic activity.
Although not shown in the results, the ratio between LF and HF powers (R), as well as the normalized
LF power (
PLFn
), were also obtained. Comparing relax and stress, these parameters have shown in
other studies a significant increase during stress [
8
], which means a predominance of the sympathetic
activity. In this work, however, the increase found in Rand
PLFn
was not significant even in the Polar
H7 measures, possibly due to the fact that the level of stress was moderate: see the moderately low
maximum HR achieved in the stress stage (around 100 bpm). Still, the significant decrease in
PHF
was
evident using both the Polar H7 and the Apple Watch, supporting the potential use of the Apple Watch
for stress monitoring.
However, one of the main limitations of using the Apple Watch to obtain and use HRV measures
is that, at this moment, the RR series is only available through the use of the Apple Breath App
in an indirect way (exporting the measured values using the Health App, where all Health-related
data from the user is stored). Apple does not provide within the watch OS 4.2 SDK any function to
access raw RR measures. Thus, developers are not able to develop an App that takes advantage of
the potential shown by Apple Watch in HRV monitoring. Besides, Breath App stops when repeated
motion is detected. This does not allow us to validate HRV measurements in low or moderate exercise.
Nevertheless, this SDK limitation can be removed by Apple in a further SDK release. This inclusion
would leverage the development of Apps able to monitoring stress conditions, among other things, by
the developer’s community.
5. Conclusions
Heart rate variability parameters extracted from an Apple Watch device have been validated using
a Polar H7 band as a reference during relax and mental stress in 20 healthy volunteers. Reliability and
agreement coefficients were computed for the RR interval series provided by both devices in relax and
stress stages, achieving very good results (reliability and agreement > 0.9). No significant differences
were found when comparing temporal HRV parameters (SDNN, RMSSD, pNN50 and HRM) derived
from the RR interval series provided by both devices. However, frequency HRV parameters LF and
HF powers were significantly different when derived from the Apple Watch, due to the appearance of
gaps in the RR series, both in relax and stress stages. Nonetheless, a decrease in the HF power (and in
the RMSSD parameter) was observed in stress with respect to relax stage when using both devices,
which supports the potential use of the Apple Watch for stress monitoring.
Author Contributions:
Conceptualization, D.H., Á.A. and R.B.; Methodology, D.H. and R.B.; Formal Analysis,
D.H.; Data Curation, S.R., J.S. and Á.A.; Writing-Original Draft Preparation, D.H., S.R., J.S., Á.A. and R.B.;
Writing-Review & Editing, D.H., Á.A. and R.B.; Supervision, Á.A. and R.B.
Funding:
This work was supported by Centro de Investigación Biomédica en Red–Bioingeniería, Biomateriales y
Nanomedicina (CIBER-BBN) through Instituto de Salud Carlos III, by the Ministerio de Economía, Industria y
Competitividad, Gobierno de España, European Regional Development Fund (TIN2016-76770-R) and FEDER
“Construyendo Europa desde Aragón” (Grupo de investigación T31_17R), by Aragón Government through Grupo
de Referencia BSICoS (T39_17R), by Aragón Institute of Engineering Research (I3A), IIS Aragón and European
Social Fund (EU).
Acknowledgments:
The computation was performed by the ICTS “NANBIOSIS”, more specifically by the
High Performance Computing Unit of the CIBER-BBN at the University of Zaragoza. The authors also want to
acknowledge the support from Ogilvy Paris.
Conflicts of Interest:
The authors declare no conflict of interest. The funders had no role in the design of the
study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, and in the decision to
publish the results.
Sensors 2018,18, 2619 10 of 11
References
1.
Camm, A.J.; Malik, M.; Bigger, J.T.; Brethardt, G.; Cerutti, S.; Cohen, R.J.; Coumel, P.; Fallen, E.L.;
Kennedy, H.L.; Kleiger, R.E.; et al. Heart rate variability: Standards of measurement, physiological
interpretation, and clinical use. Circulation 1996,93, 1043–1065.
2.
Thayer, J.; Ahs, F.; Fredrikson, M.; Sollers, J.; Wager, T. A meta-analysis of heart rate variability
and neuroimaging studies: Implications for heart rate variability as a marker of stress and health.
Neurosci. Biobehav. Rev. 2012,36, 747–756. [CrossRef] [PubMed]
3.
Billman, G. Heart rate variability—A historical perspective. Front. Physiol.
2011
,2, 86. [CrossRef] [PubMed]
4.
Beauchaine, T.; Thayer, J. Heart rate variability as a transdiagnostic biomarker of psychopathology.
Int. J. Psychophysiol. 2015,98, 338–350. [CrossRef] [PubMed]
5.
Koenig, J.; Kemp, A.; Beauchaine, T.; Thayer, J.; Kaess, M. Depression and resting state heart rate variability
in children and adolescents—A systematic review and meta-analysis. Clin. Psychol. Rev.
2016
,46, 136–150.
[CrossRef] [PubMed]
6.
Tobaldini, E.; Nobili, L.; Strada, S.; Casali, K.; Braghiroli, A.; Montano, N. Heart rate variability in normal
and pathological sleep. Front. Physiol. 2013,4, 294. [CrossRef] [PubMed]
7.
Kim, H.; Cheon, E.; Bai, D.; Lee, Y.; Koo, B. Stress and heart rate variability: A meta-analysis and review of
the literature. Psychiatry Investig. 2018,15, 235–245. [CrossRef] [PubMed]
8.
Hernando, A.; Lázaro, J.; Gil, E.; Arza, A.; Garzón, J.; López-Antón, R.; de la Cámara, C.; Laguna, P.;
Aguiló, J.; Bailón, R. Inclusion of respiratory frequency information in heart rate variability analysis for
stress assessment. IEEE J. Biomed. Health Inform. 2016,20, 1016–1025. [CrossRef] [PubMed]
9.
Vicente, J.; Laguna, P.; Bartra, A.; Bailón, R. Drowsiness detection using heart rate variability. Med. Biol.
Eng. Comput. 2016,54, 927–937. [CrossRef] [PubMed]
10.
Kiviniemi, A.; Hautala, A.; Kinnunen, H.; Tulppo, M. Endurance training guided individually by daily heart
rate variability measurements. Eur. J. Appl. Physiol. 2007,101, 743–751. [CrossRef] [PubMed]
11.
Plews, D.; Laursen, P.; Stanley, J.; Kilding, A.; Buchheit, M. Training adaptation and heart rate variability in
elite endurance athletes: Opening the door to effective monitoring. Sports Med.
2013
,43, 773–781. [CrossRef]
[PubMed]
12.
Tian, Y.; He, Z.; Zhao, J.; Tao, D.; Xu, K.; Earnest, C.; McNaughton, L. Heart rate variability threshold values
for early-warning nonfunctional overreaching in elite female wrestlers. J. Strength Cond. Res.
2013
,27,
1511–1519. [CrossRef] [PubMed]
13.
Shcherbina, A.; Mattsson, C.; Waggott, D.; Salisbury, H.; Christle, J.; Hastie, T.; Wheeler, M.; Ashley, E.
Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse
cohort. J. Pers. Med. 2017,7, 3. [CrossRef] [PubMed]
14.
Gil, E.; Orini, M.; Bailón, R.; Vergara, J.; Mainardi, L.; Laguna, P. Photoplethysmography pulse rate variability
as a surrogate measurement of heart rate variability during non-stationary conditions. Physiol. Meas.
2010
,
31, 1271–1290. [CrossRef] [PubMed]
15.
Eskola, H.; Väisänen, O.; Viik, J.; Hyttinen, J. Evaluation of the accuracy and reliability for
photoplethysmography based heart rate and beat-to-beat detection during daily activities. In Proceedings
of the Joint Conference of the European Medical and Biological Engineering Conference (EMBEC) and the
Nordic-Baltic Conference on Biomedical Engineering and Medical Physics (NBC) 2017, Tampere, Finland,
11–15 June 2017.
16.
Wallen, M.; Gomersall, S.; Keating, S.; Wisløff, U.; Coombes, J. Accuracy of heart rate watches: Implications
for weight management. PLoS ONE 2016,11, e0154420. [CrossRef] [PubMed]
17.
Abt, G.; Bray, J.; Benson, A. The validity and inter-device variability of the Apple Watch for measuring
maximal heart rate. J. Sports Sci. 2017,36, 1447–1452. [CrossRef] [PubMed]
18.
Khushhal, A.; Nichols, S.; Evans, W.; Gleadall-Siddall, D.; Page, R.; O’Doherty, A.; Carroll, S.; Ingle, L.;
Abt, G. Validity and reliability of the AppleWatch for measuring heart rate during exercise. Sports Med.
Int. Open 2017,1, 206–211.
19.
Wang, R.; Blackburn, G.; Desai, M.; Phelan, D.; Gillinov, L.; Houghtaling, P.; Gillinov, M. Accuracy of
wrist-worn heart rate monitors. JAMA Cardiol. 2017,2, 104–106. [CrossRef] [PubMed]
20.
Bai, Y.; Hibbing, P.; Mantis, C.; Welk, G. Comparative evaluation of heart rate-based monitors: Apple Watch
vs Fitbit Charge HR. J. Sports Sci. 2017,36, 1734–1741. [CrossRef] [PubMed]
Sensors 2018,18, 2619 11 of 11
21.
Canalys. 18 Million Apple Watches Ship in 2017, up 54% on 2016. 2018. Available online: https://www.
canalys.com/newsroom/18-million-apple-watches-ship-2017-54-2016 (accessed on 4 July 2018).
22.
Giles, D.; Draper, N.; Neil, W. Validity of the Polar V800 heart rate monitor to measure RR intervals at rest.
Eur. J. Appl. Physiol. 2016,116, 563–571. [CrossRef] [PubMed]
23.
Caminal, P.; Sola, F.; Gomis, P.; Guasch, E.; Perera, A.; Soriano, N.; Mont, L. Validity of the Polar V800
monitor for measuring heart rate variability in mountain running route conditions. Eur. J. Appl. Physiol.
2018,118, 669–677. [CrossRef] [PubMed]
24.
Valderas, M.T.; Bolea, J.; Laguna, P.; Vallverdú, M.; Bailón, R. Human Emotion Recognition Using Heart
Rate Variability Analysis with Spectral Bands Based on Respiration. In Proceedings of the 37nd Annual
International Conference of the IEEE EMBS, Milan, Italy, 25–29 August 2015; pp. 6134–6137.
25.
Bland, J.; Altman, D. Statistical methods for assessing agreement between two methods of clinical
measurement. Lancet 1986,8, 307–310. [CrossRef]
26.
Lin, L. A concordance correlation coefficient to evaluate reproducibility. Biometrics
1989
,45, 255–268.
[CrossRef] [PubMed]
27. Fisher, R. Statistical Methods for Research Workers; Oliver & Boyd: Edinburgh, UK, 1925.
28.
Costa-Santos, C.; Antunes, L.; Souto, A.; Bernardes, J. Assessment of disagreement: A new information-based
approach. Ann. Epidemiol. 2010,20, 555–561. [CrossRef] [PubMed]
29.
Mateo, J.; Laguna, P. Improved heart rate variability signal analysis from the beat occurrence times according
to the IPFM model. IEEE Trans. Biomed. Eng. 2000,47, 985–996. [CrossRef] [PubMed]
30.
Bailón, R.; Laouini, G.; Grao, C.; Orini, M.; Laguna, P.; Meste, O. The integral pulse frequency modulation
with time-varying threshold: Application to heart rate variability analysis during exercise stress testing.
IEEE Trans. Biomed. Eng. 2011,58, 642–652. [CrossRef] [PubMed]
31.
Sacha, J. Interplay between heart rate and its variability: A. prognostic game. Front. Physiol.
2014
,5, 347.
[CrossRef] [PubMed]
32.
Valenza, G.; Garcia, R.; Citi, L.; Scilingo, E.; Tomaz, C.; Barbieri, R. Nonlinear digital signal processing in
mental health: Characterization of major depression using instantaneous entropy measures of heartbeat
dynamics. Front. Physiol. 2015,6, 74. [CrossRef] [PubMed]
33.
Weippert, M.; Behrens, K.; Rieger, A.; Stoll, R.; Kreuzfeld, S. Heart rate variability and blood pressure during
dynamic and static exercise at similar heart rate levels. PLoS ONE 2013,8, e83690. [CrossRef] [PubMed]
34.
Buchheit, M. Monitoring training status with HR measures: Do all roads lead to Rome? Front. Physiol.
2014
,
5, 73. [CrossRef] [PubMed]
35.
Gamelin, F.X.; Baquet, G.; Berthoin, S.; Bosquet, L. Validity of the polar S810 to measure R-R intervals in
children. Int. J. Sports Med. 2008,29, 134–138. [CrossRef] [PubMed]
36.
Nunan, D.; Jakovljevic, D.G.; Donovan, G.; Hodges, L.D.; Sandercock, G.R.; Brodie, D.A. Levels of agreement
for RR intervals and short-term heart rate variability obtained from the Polar S810 and an alternative system.
Eur. J. Appl. Physiol. 2008,103, 529–537. [CrossRef] [PubMed]
37.
Barbosa, M.P.; da Silva, N.T.; de Azevedo, F.M.; Pastre, C.M.; Vanderlei, L.C. Comparison of Polar
®
RS800G3
™
heart rate monitor with Polar
®
S810i
™
and electrocardiogram to obtain the series of RR intervals and analysis
of heart rate variability at rest. Clin. Physiol. Funct. Imaging 2016,36, 112–117. [CrossRef] [PubMed]
38.
Hernando, D.; Garatachea, N.; Almeida, R.; Casajús, J.A.; Bailón, R. Validation of Heart Rate Monitor Polar
RS800 for Heart Rate Variability Analysis During Exercise. J. Strength Cond. Res.
2018
,32, 716–725. [CrossRef]
[PubMed]
39.
Porges, S. Cardiac vagal tone: A physiological index of stress. Neurosci. Biobehav. Rev.
1995
,19, 225–233.
[CrossRef]
©
2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access
article distributed under the terms and conditions of the Creative Commons Attribution
(CC BY) license (http://creativecommons.org/licenses/by/4.0/).