sensors
Article
On the Beat Detection Performance in Long-Term
ECG Monitoring Scenarios
Francisco-Manuel Melgarejo-Meseguer 1,*ID , Estrella Everss-Villalba 1ID ,
Francisco-Javier Gimeno-Blanes 2,*ID , Manuel Blanco-Velasco 3ID , Zaida Molins-Bordallo 1ID ,
José-Antonio Flores-Yepes 2ID , José-Luis Rojo-Álvarez 4,5,*ID and Arcadi García-Alberola 1ID
1Cardiology Service, Arrhythmia Unit, Hospital General Universitario Virgen de la Arrixaca, El Palmar,
30120 Murcia, Spain; estrella.everss@urjc.es (E.E.-V.); zaidamb23@gmail.com (Z.M.-B.);
arcadi@secardiologia.es (A.G.-A.)
2Department of Signal Theory and Communications, Miguel Hernández University, Elche,
03202 Alicante, Spain; ja.flores@umh.es
3Department of Signal Theory and Communications, University of Alcalá, Alcalá de Henares,
28805 Madrid, Spain; manuel.blanco@uah.es
4Center for Computational Simulation, Universidad Politécnica de Madrid, Boadilla, 28223 Madrid, Spain
5Department of Signal Theory and Communications, Rey Juan Carlos University, Fuenlabrada,
28943 Madrid, Spain
*Correspondence: francisco.melgarejo@goumh.umh.es (F.-M.M.-M.); javier.gimeno@umh.es (F.-J.G.-B.);
joseluis.rojo@urjc.es (J.L.R.-Á.); Tel.: +34-966-658-881 (F.-J.G.-B.)
Received: 26 March 2018; Accepted: 28 April 2018; Published: 1 May 2018
Abstract:
Despite the wide literature on R-wave detection algorithms for ECG Holter recordings,
the long-term monitoring applications are bringing new requirements, and it is not clear that the
existing methods can be straightforwardly used in those scenarios. Our aim in this work was twofold:
First, we scrutinized the scope and limitations of existing methods for Holter monitoring when
moving to long-term monitoring; Second, we proposed and benchmarked a beat detection method
with adequate accuracy and usefulness in long-term scenarios. A longitudinal study was made with
the most widely used waveform analysis algorithms, which allowed us to tune the free parameters of
the required blocks, and a transversal study analyzed how these parameters change when moving
to different databases. With all the above, the extension to long-term monitoring in a database of
7-day Holter monitoring was proposed and analyzed, by using an optimized simultaneous-multilead
processing. We considered both own and public databases. In this new scenario, the noise-avoid
mechanisms are more important due to the amount of noise that exists in these recordings, moreover,
the computational efficiency is a key parameter in order to export the algorithm to the clinical
practice. The method based on a Polling function outperformed the others in terms of accuracy and
computational efficiency, yielding 99.48% sensitivity, 99.54% specificity, 99.69% positive predictive
value, 99.46% accuracy, and 0.85% error for MIT-BIH arrhythmia database. We conclude that the
method can be used in long-term Holter monitoring systems.
Keywords: QRS detection; ECG; long-term monitoring; Holter; 7-day
1. Introduction
The main purpose of cardiac monitoring is the detection of cardiac diseases with transient
manifestations or symptoms. The most frequently used device to continuously record the heart’s
rhythm during daily activities is the 24-h Holter, in which patients are continuously monitored during
one day through a series of electrodes attached to the skin and while performing their daily activities.
This procedure can be also carried out during longer periods, such as seven days, 15 days, or 21 days,
Sensors 2018,18, 1387; doi:10.3390/s18051387 www.mdpi.com/journal/sensors
Sensors 2018,18, 1387 2 of 24
thus recording what we refer to as Long-term Monitoring (LTM). In previous research, LTM has
been proven to be an important diagnostic tool, since it can detect several diagnostic indicators that
are not always present in shorter monitoring periods (specially for transient and very infrequent
electrocardiographic abnormalities). For this reason, recent research is increasingly focusing on LTM
protocols and systems [
1
–
4
]. Inside this new scenario, one of the most important challenges is the
extraction of different parameters that can be used to improve the diagnostic accuracy. Using traditional
protocols and signal processing techniques, the massive amount of data that the clinicians must
analyze and validate for each patient represents a time-consuming challenge. To solve this problem,
new protocols and new more advanced signal processing techniques are required. Beat detection is
one of the most relevant electrocardiogram (ECG) processing techniques because finding the exact
location of QRS complexes (i.e., the signal deflections corresponding to ventricular depolarization)
enables the segmentation of the ECG into its sub–waves, which is often a basic step for multiple and
more complex algorithms [
5
]. For this reason, R-detection is usually done in the initial stage before
beat-feature extraction and beat classification [
6
]. An example of the importance of the correct heartbeat
detection is its application to the study of changes in the heart rate, which is a key parameter to analyze
many heart diseases, such as arrhythmias, and arrhythmic risk assessment of patients, among others [
7
].
Although the detection of the QRS complex has been extensively studied over the years, it is still
a field under development. The reason is that the detection of the R-wave in the ECG is an apparently
very simple problem, which can be tackled from many different viewpoints. Initially, the development
of accurate methods was the point of interest [
8
,
9
], but, currently, research mainly focuses on the
design of novel methods usually based on modern processing techniques. Despite the application of
advanced data processing methods, the performance of the new approaches seldom overcome the
former ones. In fact, in many cases, some recent techniques are applied in combination with traditional
methods such as the Pan and Tompkins algorithm (PT) or frequency transforms [
9
–
12
]. Accordingly,
huge amounts of contributions have been reported, the literature in this area is large and complicated,
and the task of determining which is the best suited or the more efficient algorithm among the available
ones has become difficult.
Beyond the accuracy of the methods, not all of them can be straightforwardly applied to LTM
recordings. The initial constraint is given by the limited computational capacity of current equipment
to process the big amount of data resulting from 7- to 21-day ECG recordings. Additionally, the high
performance exhibited by current methods is difficult to reach in LTM applications. One of the reasons
is the testing methodology, and, despite the algorithms being checked by means of comparison tests
against state-of-the-art methods and over public databases, the set size often is limited, which makes
some analysis be scarcely representative. Another, and perhaps more important reason, is that the
algorithms are sensitive to the dataset, so that the parameter tuning needs to be readjusted depending
on the working conditions. Therefore, an R-wave detector for LTM recordings must be only be accurate,
but also reliable, robust, and efficient from a computational cost viewpoint, in order to be able to deal
with very long sets in the presence of noise and artifacts [13].
Therefore, and taking into account the current situation of the R-peak detection field, the purpose
of this work was twofold. On the one hand, we aimed to determine the scope and limitations of the
R-wave detection algorithms currently available when they are applied to LTM records. On the other
hand, we proposed an accurate and reliable method to be specifically designed for LTM. In order to
achieve these targets, we drove both a qualitative and quantitative study by means of a longitudinal
and a transversal analysis. The longitudinal analysis aimed to identify and analyze the most commonly
used processing blocks in this field, and the results derived from this part of the study enabled the
design of an efficient detection system with the most suitable processing blocks. The transversal
analysis was carried out to overcome the effect of using different databases and given the dependency
of the detector with the dataset used in its original design. This part of the study provided the clues to
correctly refine the parameters of the different elements of the detection algorithm. The transversal
analysis was tackled through comparisons carried out over several public and own large databases.
Sensors 2018,18, 1387 3 of 24
As a result, we present here a reliable, robust, and efficient R-wave detector for LTM, outpacing
the existing ones, and benchmarking it simultaneously over a wide range of databases. To do so,
we deeply stress-tested the performance of the most popular published methods, not only over the
traditional public (and well known) databases, but also over two trusted and solid new databases of
24-h and 7-day LTM clinical recordings, from Hospital Virgen de la Arrixaca de Murcia. These private
databases contain a variety of patient illnesses, as well as different signal noises, lengths, or recording
characteristics, such as sampling frequency and number of bits per sample. The developed optimal
detector consisted of a new–and–modified PT algorithm based on simultaneous-multilead processing,
which outperformed all the tested methods in terms of efficiency and accuracy.
This paper is structured as follows. Section 2describes the appliances and tools used for signal
registering and processing, and then it elaborates on details for all the algorithms and databases used
for signal processing. Section 3includes the different selected methods, while, in Section 4, experiments
are applied over our database and the corresponding results are reported. Finally, Section 5contains
the conclusions of all the work carried out.
2. Databases
For testing the algorithms and methods analyzed and designed in this paper, several private and
public databases have been used. More precisely, a private 24-h Holter database that was manually
labeled by experts was used to analyze the optimization of classical algorithms in a high–quality
annotated and long-duration set of recordings. After that, the most popular public labeled databases
in the scientific literature were used to compare our results with those of others, aiming to account
here for possible deviations in the algorithm accuracy due to overfitting to the database used for
algorithm tuning. Finally, we worked with another set of labeled 7-day Holter recordings in order to
adapt the algorithms to a larger time-scale database. To our best knowledge, there is not a public LTM
database manually and beat-by-beat labeled with respect to arrhythmic events, hence we decided to
use again the available recordings from precedent studies, which had been carefully revised because
their research use. Table 1shows a summary of the names of the databases and their characteristics.
Table 1. Duration and number of ECGs used from different databases in this study.
Experiment Database Type #ECGs Duration
Longitudinal vPREDICT private 16 24 h
Transversal
MITBIH
public
48 30 m
NSTDB 15 30 m
LTSTDB 86 21–24 h
LTAFDB 84 24–25 h
LTMITDB 7 14–22 h
INCART 75 30 m
LTM PREMARIS private 5 7 days
(1) vPREDICT Private Database. The vPREDICT project [
14
,
15
] was designed to establish data
relationships between electric cardiac signals and sudden cardiac death events using signal processing
based indices. Given that these recording were assembled for research purposes, they had been
carefully annotated, which made them suitable for quality validation of ECG signal-processing
algorithms. This database contains 24-h Holter recordings from 122 heart failure patients in 5 different
hospitals. For testing the algorithms and methods in this work, only 16 patients were selected among
them, including diverse clinical conditions, namely, two subjects with normal heart function, five with
an implanted defibrillator, five with atrial fibrillation, and the last five with frequent premature
ventricular contractions. Two different Holter recorders were used, namely, SpiderView former ELA
Medical, and now part of Livanova , and SpiderView Plus (former Sorin Group, and now also part
of Livanova). The acquisition was made with 3 or 5 sensors, yielding the usual two or three leads in
Sensors 2018,18, 1387 4 of 24
Holter, and the sampling rate was 200 Hz. The electrodes were placed according to SpiderView manual,
i.e., in the case of 3-electrode configuration, the electrodes were placed on the sternal manubrium,
the xyphoid process and the area of the apex generating a pseudo-II and V5 leads.
(2) Public Databases. For decades, Physionet has offered online large physiological collections
of signals that have powered the clinical and technical research in many fields. In this work, and in
order to evaluate the generalization of the optimization of classical algorithms previously used in the
vPREDICT database, six open–access labeled ECG signal databases from Physionet were used:
•
MIT–BIH. The MIT–BIH Arrhythmia database contains a representative collection of independently
annotated ECG signals, consisting of 48 half-hour excerpts of two–lead 24-h ambulatory ECGs with
clinically significant arrhythmias sampled at 360 Hz [
16
,
17
]. We chose this classical database due to
its widespread and current use [18].
•
NSTDB. The MIT–BIH Noise Stress database includes 12 half–hour ECG recordings and three
half-hour recordings of three types of noise that are more commonly present in ambulatory
monitorization, namely, baseline wander, muscle artifact, and motion artifact. Noise data were
made using physically active volunteers with standard ECG recorders, leads, and electrodes,
and then adding it to the recordings from the MIT–BIH [
19
,
20
]. We selected these data for the
wide use of its noise components (e.g., see [21]).
•
LTSTDB. The Long-term ST database includes 86 Holter recordings of two or three leads with a duration
between 21 and 24 h. This database shows a variety of events of ST segment changes [20,22].
•
LTAFDB. The Long Term AF database includes 84 two-lead Holter recordings sampled at 128 Hz
with a duration between 24 and 25 h [20,23].
•
LTMITBIHDB. The Long-term MIT–BIH database includes 7 Holter recordings between 14 and 22 h [
20
].
•
INCART. St. Petersburg Institute of Cardiological Technics 12–lead Arrhythmia database consists
of 75 annotated recordings extracted from 32–half–hour Holter recordings of coronary artery
disease patients sampled at 257 Hz [
20
]. In order to compare our Holter recordings with other
Holter segments, we chose this data following other current researchers (e.g., see [24]).
(3) PREMARIS Private Database. This database was manually labeled by experts, and it has
been used in previous research by our group [
25
,
26
]. It includes 84 real cases of 7-day- Holters,
in patients suffering from heart failure (HF) and/or left ventricular dysfunction. In order to evaluate
the performance of the previously developed algorithms over real LTM recordings, five patients
with sinus rhythm were selected from the database. Four out of these five were males. It should be
mentioned here that 7-day Holter recordings are registered during the normal daily life of patients,
and so these records included not only the usual amount of artifacts and noise in standard ECG, but also
strong presence of sensors decoupling and signal losses due to the long exposition. The sampling
frequency was 128 Hz.
3. R-Wave Detector System
Several types of QRS detector methods have been proposed to date in the literature and in
the commercial systems. The most usual family of algorithms used for 24-h Holter monitoring is
based on waveform analysis principles. These algorithms use signal processing techniques to reduce
noise as well as to emphasize relevant signal elements through a number of computational strategies.
In this work, we scrutinize the performance of the most commonly used R-wave detector systems
applied to 24-h Holter recordings, and specially how they behave when they are moved towards LTM
scenarios. Towards that end, we present in this section a compiled architecture of the best knowledge
in the literature. Waveform algorithms applied in literature are generally categorized into two subsets.
The first one incorporates all those methods using multiple leads that are recorded through parallel
acquisition. Examples of these multilead processing techniques are those using Principal Component
Analysis (PCA), Independent Component Analysis (ICA), or Linear Discriminant Analysis (LDA) [
27
].
The second subset of algorithms treats each lead individually, and they apply filters in order to reduce
Sensors 2018,18, 1387 5 of 24
the noise and enhance the relevant signal components. Examples of this second type include the
very well-known PT, as well as the Hamilton–Tompkins (HT) and the Chouhan methods [
9
,
28
–
30
].
Table 2shows the accuracy rates benchmarked over the MIT–BIH of relevant methods on the literature,
such as Continuous Wavelet Transform (CWT) [
31
], Short Time Fourier Transform (STFT) [
32
], Hilbert
Transform (HiT) [
33
], or Hilbert–Huang Transform (HHT). These reported values should be considered
as the standard target accuracy also for LTM scenarios, but the methods should also be competitive in
terms of computational cost, taking into account the dramatically longer signal duration.
Table 2. Key QRS detection methods and statistical comparison of their performance over MIT–BIH.
Method Statistical Measures [%] Ref.
SEN SPE PPV
PT 99.75 - 99.53 [9]
HT 99.69 - 99.77 [28]
CWT 99.85 - 99.48 [34]
STFT 99.1 99.6 - [32]
HiT 99.81 - 99.83 [33]
HHT 98.88 99.04 - [35]
Our literature review highlighted the common stages for R-wave detection in commercial
algorithms [
9
,
10
,
15
,
28
,
29
,
36
]. Roughly speaking, three steps are generally applied (see Figure 1).
First, Signal Preprocessing is usually performed, in which the signal is adjusted to fit a certain common
standard by removing noise, baseline wander, and undesirable artifacts. To do so, different types of
filters are applied, such as low–pass, high pass, and notch filters. In a second stage, usually known as
Feature Signal Extraction, a much stronger and sometimes uncompromisingly processing is performed in
order to isolate the QRS complexes for their better detection. At this stage, the focal point is to detect the
presence of the QRS complexes by the creation of a new signal, a so-called feature signal, where the QRS
complexes are enhanced in order to ease their detection in the next stage. In the final R-wave Detection
stage, the R-waves are situated by using an adaptive–threshold based on statistical and temporal
properties of the feature signal. Many algorithms are proposed in the literature for this final and critical
steps, although the main differences in terms of accuracy and computational cost are found in the
Feature Extraction stage. Here follows a deeper detailed description of each of the stages including
a short description of the methods present in literature, as well as new adapted methods developed in
this work to enhance the performance. The reader should notice that no exhaustive description or full
enumeration of all the reported methods are included in this work, as over 500 papers are released
yearly in this topic, but rather the most relevant and effective methods are described, implemented,
benchmarked, and summarized hereafter.
3.1. Signal Preprocessing Stage
Preprocessing is the first stage of the algorithm, and decisions taken at this point will impact
all the subsequent blocks. In our specific application for LTM, a prior segmentation process is
applied for computational convenience. Segmentation is the process of dividing the ECG record into
smaller segments to be processed efficiently [
18
,
37
,
38
], according to standard computational resources.
Although segmentation is not always included or mentioned in the literature, as it has no impact
on standard clinical ECG records of few seconds, it is of relevant impact in LTM implementations.
As a token, the reader should be aware that a 3-lead signal, sampled at 200 Hz with 15–bit resolution
and during 48 consecutive hours, creates a single file of about 100 MB. This file size becomes intractable
when it has to be handled with intensive processing techniques, and commercial devices do not always
have enough resources.
Sensors 2018,18, 1387 6 of 24
Figure 1.
Main stages of the QRS detection algorithms proposed in this work. The yellow boxes are
inputs/outputs, the blue boxes represent the preprocessing stage, the green boxes represent the feature
signal extraction stage, and the red boxes represent the QRS detection stage.
As a part of the standard preprocessing, denoising, baseline wander, and signal detrending
are very well described in literature, and frequently built-in in almost any commercial or research
system. The removal of these undesirable components is generally referred to as preprocessing,
although as we mentioned before it could also include some other signal accommodation. The most
common types of noise present in the ECG methodology are power line interference, other device
electromagnetic interference, muscle artifacts, breathing induction, and artifacts related to decoupling
of sensor interfaces. The relevance of all these possible elements have a capital impact on LTM due to
the unfeasibility to keep the recording device in a noise–free environment and connected to ground,
as the device is portable and the patient should be running his normal life during the recording period.
Our implementation incorporated the following algorithmic stages: cubic spline detrend baseline,
power line and harmonic notch filtering, a low–pass filter to remove muscle artifact and other high
frequency interferences, and, finally, a filter based on the time–local variance, which is used to remove
the patient–electrode motion artifacts [
13
]. An example of the result of this stage is shown in Figure 2.
Sensors 2018,18, 1387 7 of 24
2.7666 2.7668 2.767 2.7672 2.7674
104
0
0.2
0.4
0.6
0.8 Raw signal from lead 1
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
-0.3
-0.2
-0.1
0
0.1
Amplitude [mV]
Filtered signal from lead 1
2.7666 2.7668 2.767 2.7672 2.7674
104
-1
-0.5
0
Raw signal from lead 2
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
-0.4
-0.2
0
0.2
0.4
0.6
0.8
Amplitude [mV]
Filtered signal from lead 2
2.7666 2.7668 2.767 2.7672 2.7674
104
-1.5
-1
-0.5
0
Raw signal from lead 3
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
-0.5
0
0.5
1
Amplitude [mV]
Filtered signal from lead 3
Figure 2.
Example of the signal preprocessing stage over a segment of 3-lead signal from
vPREDICT database.
3.2. Feature Signal Extraction Stage
The next step after preprocessing is Feature Signal Extraction. At this stage, the focus is not to
retain the ECG morphology for diagnostic purposes, but rather to undoubtedly identify the region
where the QRS complexes are. A number of algorithms have been proposed that compute the feature
signal in many ways. We can classify them all in two families attending to the way it is being processed.
The first family could be called Matrix Methods. These methods leverage on the multiple available
leads to enhance the QRS by applying very well-known multichannel processing techniques, such as
PCA and ICA [
39
–
44
]. The second family could be generally named as Digital Filtering methods, and it
includes PT, Root Mean Square (RMS), and combinational methods. All the methods in this family
share the use of filtering techniques to enhance the QRS against the rest of the elements present in the
ECG signal.
PCA is one of the most relevant Matrix Methods since it was first introduced [
45
], and it consists
of a multivariate linear conversion that transforms the original signals into a new coordinate system
where signals are orthogonal. This transformation may eventually highlight relevant elements, as the
QRS, in one of the output signals if the input signals incorporated different projections of these
elements. As a result, this processing could reduce both noise and data dimensionality, selecting only
the output signals with relevant information. PCA is often used in a number of denoising processes,
and it has also shown to be appropriate for multilead Holter recording analysis, where different noise
sources distort the measured signal. To perform this processing, the covariance matrix of available
leads is calculated, as well as its eigenvectors. As a second step, a transformation is applied ensuring
orthogonality, and the output new signals are sorted according to their eigenvalues, which ensures
to retain the largest variance components. The eigenvector associated with the largest eigenvalue
corresponds to the largest explained variance from the decomposition [
46
], which usually makes the
projection of all the leads onto that direction the most suitable to be used as a feature signal for this
stage. Once it has been obtained, it is often squared to highlight the R-peak magnitude, which usually
enables an enhancement on its presence. The second technique applied under the Matrix Methods is
the ICA, which also allows us to represent a set of signals as a combination of statistically independent
components. Its main purpose is the extraction of signals from a signal mixture model created under
unknown conditions. This situation mirrors and matches the real fact of the different projections
of the QRS in multiple lead signals. In order to apply this technique, the signals must meet certain
Sensors 2018,18, 1387 8 of 24
conditions, namely, the original signals are assumed to be independent among them, and they follow
a non-Gaussian distribution. Different from PCA, it is not clear which component is more relevant
than another in ICA [
47
], and in this work we used the heuristically accepted approach of sorting the
relevance of the projection directions with the descending kurtosis of the projected output.
Entering now in the Digital Filtering methods, the PT algorithm [
9
] designed to provide QRS
complex detection in real time is widely used, and it may be the most benchmarked method of all
the ones reported in the literature. This method works on a single lead, and in the case of having
more than one lead, the one with lowest noise content is selected. The signal is then filtered, first
with a narrow-band-pass filter, and then with a derivative, which emphasizes the QRS segment.
The resulting signal is squared to boost the amplitude of R-peaks, and a mean filter is applied with
a sliding window in order to integrate the signal. Finally, the R-peak detection is done with the
application of a threshold over the amplitude of the before computed signal. Two main drawbacks
of this PT algorithm method should be mentioned. First, the initial selection of the less–noisy lead is
crucial, having a relevant impact on final results, and, second, the threshold selection is not always
direct or evident, and it has noticeable effects on the final results.
To improve the observed disadvantages involving the PT detector, especially the first one,
we proposed and developed a set of new algorithms based on this very well known method.
We describe and present here only that provided better results. One of them is the RMS method,
which creates a ready–for–thresholding feature signal from a multi–lead Holter recording by just
combining all the leads as follows:
yc[n] = v
u
u
t
L
∑
l=1
x2
f[n,li], (1)
where
yc[n]
is the feature signal,
xf[n
,
li]
is the
i
th lead of the pre–processed signal filtered with the PT
derivative–filter, and Lis the number of leads.
Another implemented variation consisted on applying PT derivative–filter to each lead and the
result is mixed using the following combinational lead approach:
yc[n,li] = xf[n,li]v
u
u
t
L
∏
j=1,i6=j
xf[n,lj]. (2)
This last implementation of a combinational method consists of the evaluation of the feature signal
for R-peak detection on each lead and to jointly assess the detection based on a certain algorithm
applied to all of them simultaneously. Four different possibilities were calculated and benchmarked.
The first one is called the And method, in which all the detection vectors are combined by using the
and logical function. The second one is called the Polling (Poll) combination, and, in this method,
if a peak is detected in the majority of the leads, then it is tagged as a valid peak. The third one is called
Simple Coupling (SC), and it is based on the And method, but to report an R-peak as valid, it must be
present in at least two leads. The last one is known as Or, and all the detected R-peaks are combined
by the use of the or logical function. As the physical location of the R-peak may differ slightly among
leads, a tolerance window is implemented around the peaks as far these combinational methods are
concerned. Maximum tolerance was set to 10 ms based on the clinical knowledge and empirical review.
Additionally, a refractory period for R-peak detection was set in order to avoid artifact detection and
other prominent false positive detections, such as prominent T–peaks.
Figure 3shows the different feature signals computed for each method presented in this work
using the signal shown in Figure 2.
Sensors 2018,18, 1387 9 of 24
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
2
4
6
8
Amplitude
ICA
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
0.1
0.2
0.3
0.4
Amplitude
PT
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
0.05
0.1
0.15
0.2
0.25
Amplitude
RMS
Feature signal
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
0.5
1
1.5
2
Amplitude
10-4 PCA
2.7665 2.7666 2.7667 2.7668 2.7669 2.767 2.7671 2.7672 2.7673 2.7674
104
2
4
6
10-3 Lead 1
2.7665 2.7666 2.7667 2.7668 2.7669 2.767 2.7671 2.7672 2.7673 2.7674
104
5
10
15 10-3 Lead 2
Feature signals from combinational methods
2.7665 2.7666 2.7667 2.7668 2.7669 2.767 2.7671 2.7672 2.7673 2.7674
Time [s] 104
5
10
15
Amplitude [mV]
10-3 Lead 3
Figure 3.
Computed feature signals for each proposed method. The upper four panels show the feature
signal for the next methods: ICA, PCA, PT, and RMS. In the lower panel, the feature signals from the
combinational methods are presented.
3.3. R-Wave Detection Stage
This subsection describes the method followed to finally locate the R-peak in the ECG previously
determined region. We implemented a thresholding method as usual in the literature [
10
,
15
,
29
,
36
].
The threshold is defined as
α·y+β·std(y)
, where
y
is the feature signal that was calculated according
to the methods seen before, and it is applied to a widen by 75 ms of the preselected QRS segment.
Once all the segments that contain an R-wave are determined, the detection is carried out using
an algorithm with two steps. First, we implemented a coarse detection stage where the algorithm
searches the maximums of each of the previous regions candidates. This allows for a specific detection
of the R-waves using the following equation:
Prepeak(r) = max(|(xf[r
,
l]) |)
, where
r
refers to the
previously calculated segments and
xf[r
,
li]
stands for the
i
th lead of featured signal computed before.
Second, we implemented a fine detection stage, where the algorithm looks for the maximum value in
a window centered in each pre–detected peak over a softer filtered version of the initial input signal.
In Figure 4, an example of peak detection for each proposed method is shown.
Sensors 2018,18, 1387 10 of 24
2.7666 2.7668 2.767 2.7672 2.7674
104
2
4
6
8
Feature signal from ICA
2.7666 2.7668 2.767 2.7672 2.7674
104
0.2
0.4
0.6
0.8
Detected QRS usign ICA
2.7666 2.7668 2.767 2.7672 2.7674
104
0.2
0.4
Feature signal from PT
2.7666 2.7668 2.767 2.7672 2.7674
104
0.2
0.4
0.6
0.8
Detected QRS usign PT
2.7666 2.7668 2.767 2.7672 2.7674
104
0.05
0.1
0.15
0.2
0.25
Feature signal from RMS
2.7666 2.7668 2.767 2.7672 2.7674
104
0.2
0.4
0.6
0.8
Detected QRS usign RMS
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
0.5
1
1.5
2
Amplitude
10-4 Feature signal from PCA
2.7666 2.7668 2.767 2.7672 2.7674
Time [s] 104
0.2
0.4
0.6
0.8
Amplitude [mV]
Detected QRS usign PCA
2.7665 2.7666 2.7667 2.7668 2.7669 2.767 2.7671 2.7672 2.7673 2.7674
104
2
4
6
Amplitude
10-3 Feature signal lead 1
2.7665 2.7666 2.7667 2.7668 2.7669 2.767 2.7671 2.7672 2.7673 2.7674
104
2
4
6
10-3 Feature signal lead 2
2.7665 2.7666 2.7667 2.7668 2.7669 2.767 2.7671 2.7672 2.7673 2.7674
104
2
4
6
10-3 Feature signal lead 3
2.7666 2.767 2.7674
Time [s] 104
0.2
0.4
0.6
0.8
Amplitude [mV]
Poll
2.7666 2.767 2.7674
Time [s] 104
0.2
0.4
0.6
0.8
Amplitude [mV]
Or
2.7666 2.767 2.7674
Time [s] 104
0.2
0.4
0.6
0.8
Amplitude [mV]
And
2.7666 2.767 2.7674
Time [s] 104
0.2
0.4
0.6
0.8
Amplitude [mV]
SC
Figure 4.
Thresholding process and peak detection for each proposed method. Panel one shows the
feature signal, the threshold, and the peaks detected for the next methods, ICA, PCA, PT, and RMS.
In panel two, the feature signal, the threshold, and the peaks detected for the combinational methods
are presented. The yellow crosses shown the real peak presented in the signal, and the red circles
shown the peak detected by each method
4. Experiments and Results
The algorithms presented in the previous section need to be tuned by setting different parameters
in order to maximize their performance. This section is divided into three subsections. The first
subsection correspond to the so-called Longitudinal Optimization, and includes the free-parameter
optimization on the private database vPREDICT. The results of this step allows the best algorithm preset
to be used here and after. In the second subsection, so-called Transversal Optimization, the knowledge
acquired in the previous section is applied to different public databases from the Physionet repository.
This application provides a three-fold outcome. First, it allows us to check the performance versus
different databases and to make the model independent from the used database. Additionally, it allows
us to benchmark versus well known databases, and improve the performance over a second parameter
adjustment. The last subsection, described as LTM Application, extends the resulting algorithm over our
Sensors 2018,18, 1387 11 of 24
LTM database and evaluates the different inaccuracies arising in the significant different environment
of 24-h to 7-day Holter monitoring.
The throughput analysis of the system relies on the statistics of detections achieved by the
algorithm, which can be either correct or incorrect. The study of detections only provide information
about the ability of detecting R waves, but it tells nothing about the failures to detect QRS complexes.
For this reason, the statistic of the non-detected events must also be incorporated. Regarding detections,
a correct match of the algorithm is a detection that falls within the range of a short interval around an
annotated QRS, which is referred to as a True Positive (TP). Conversely, a wrong detection is a detected
QRS complex that is not corresponding to an actual one; in this case, the detected event is labeled to be
a False Positive (FP) detection. Regarding the non-detected events, the failure to detect a QRS stands
for a wrong non-detection that is labeled as a False Negative (FN). However, the non-detection of any
event between consecutive R waves, i.e., no FPs in between two QRS complexes, stands for a correct
match of the algorithm, and it is then classified to as a True Negative (TN). The benchmark comparison
is then performed using the classical figures of merit, namely, Sensitivity (Sen),Specificity (Spe),Positive
Predictive Value (PPV),Accuracy (Acc), and Error, described as follows:
Sen =TP
TP +F N , (3)
PPV =T P
TP +F P , (4)
Spe =TN
TN +FP , (5)
Acc =TP +T N
B, (6)
Error =FP +F N
B. (7)
In summary, we should mention at this point that standard definition of SPE requires the existence of True
Negatives, and, for benchmarking purposes, we have slightly re-defined this very well-known statistic as
mentioned for the R-wave detection setting. Thus, the results of SPE should be considered with caution as
this new parameter does not reflect exactly the standard definition in other environments.
4.1. Longitudinal Optimization
LTM presents two main differences compared with standard ECG records. One of them is its
higher relevance as a consequence of the uncontrolled environment in contrast to ambulatory ECG
recordings. Another difference is the time length of LTM recordings compared with usual ambulatory
recordings. This can have a twofold impact because it is possible to implement techniques that take
advantage of the record length in order to improve the processing, for instance by statistical denoising
through averaging, but also digital processing may collapse the existing commercial devices. In order
to approach this second aspect, the first part of this subsection evaluates the required Segmentation
Optimization for processing feasibility. Additionally, a Threshold Optimization process is included to
set the minimum size of the most accurate level of peak detection that should be considered in the
subsequent processing steps.
Segmentation Optimization
. As described before, segmentation is a relevant part of any R-wave
detection algorithm, as a suboptimal election of the segment length can cause a heavy computational
load and make the algorithm unsuitable for LTM scenarios. For that reason, the aim of this first
experiment was to determine the adequate segment length that achieves the minimum processing
time. For this objective, a subset of the vPREDICT database was selected and an intensive analysis
from 1-min to 24-h segments on 5-min steps was preformed. As shown in the upper panel of Figure 5,
all the algorithms exhibited similar behavior in terms of processing time, except for ICA being instable
in short segments and divergent above 8-h segments. The errors shown in the lower panel showed
a stable performance over the segment lengths, except for PT and ICA, where errors were much higher
Sensors 2018,18, 1387 12 of 24
than in the other methods. In ICA, the performance enhancement was directly related to the segment
length, while PT showed the opposite behavior. The best segment lengths and the results for each
method are shown in Table 3. Our segmentation approach allowed us to reduce the computational
complexity of the problem. Our algorithm is based on the conventional PT algorithms, and it uses the
squared-sample operation to build the feature signal. For this reason, the PT complexity has an order
of magnitude about
n2
, where
n
is the number of samples of the whole signal. The use of segmentation
reduces the complexity to about
S·n2
seg
, where
S
is the number of segments, and
nseg
is the number of
samples in a single segment. Figure 6shows the computational complexity in terms of the number of
operations carried out. We compared the following scenarios: (i) our modified PT algorithm applied
to one single lead without any segmentation processing, meaning that the application was applied
to the whole signal; (ii) in all of the other analysis, our modified PT was applied to various leads
simultaneously and including also segmentation. It was proven earlier that segmentation enhanced
the computational performance.
12345678
Segment length (s) 104
40
60
80
100
Processing time (s)
Mean of processing time
Poll
Or
And
SC
ICA
PT
RMS
PCA
12345678
Segment length (s) 104
10
20
30
40
50
60
Error (%)
Mean of detection error
Figure 5.
Mean detection errors and processing times for different segment lengths in the
proposed algorithms.
5 10 15 20 25 30
Time [days]
1014
1015
1016
1017
1018
Number of operations
Non-segmented signal, 1 lead
Segmented signal, 2 leads
Segmented signal, 3 leads
Segmented signal, 5 leads
Segmented signal, 7 leads
Segmented signal, 12 leads
Figure 6.
Estimated computational complexity in terms of the number of operations per recording and
of its duration (from 1 to 31 days), and for different number of leads. The dark blue line represents the
number of operations of a complete single-lead signal, and the others show different lead configurations
when including the segmentation stage.
Threshold Optimization
. The aim of the next experiment was to tune the R-wave detection
threshold. In order to obtain the best threshold for each proposed algorithm, we implemented a
K
–fold
cross–validation with
K=
4. The results of this experiment are shown in Table 4for each proposed
Sensors 2018,18, 1387 13 of 24
method, showing that the algorithms with lower detection errors are Polling and And. We selected the
Polling method because it was the best algorithm in terms of error and processing time.
Table 3.
Optimal segment lengths, processing time per segment, and detection error for each proposed
method in vPREDICT database.
Method Processing Time (s) Error (%) Segment Length (s)
Poll 0.0637 4.05 300
Or 0.0645 4.63 300
And 0.0639 4.05 300
SC 0.0641 3.92 300
ICA 0.1459 26.45 7200
PCA 0.0594 7.27 300
PT 0.0597 12.02 300
RMS 0.0617 4.98 300
Table 4.
Merit figures using the optimal threshold for each method proposed in the vPREDICT database.
Method Alpha Beta Error (%)
Poll 2.21 0.50 3.92
Or 3.93 0.50 4.51
And 2.21 0.50 3.92
SC 2.79 0.50 8.36
ICA 0.50 1.00 47.14
PCA 2.21 0.50 5.93
PT 1.64 0.50 27.75
RMS 0.50 2.00 4.67
Optimal R-peak Detection Method
. According to previous experiments, the Polling method
outperformed all the other implemented ones. It standouts in terms of computational time, as well as
in detection accuracy. The selected algorithm, over vPREDICT and MIT–BIH, presented quite similar
results in both cases and in line with algorithms in the literature, as it can be observed in Table 5.
Table 5. R-wave results on 24-h Holter recordings from vPREDICT and MIT–BIH databases.
vPREDICT Sen Spe PPV Acc Error
Mean ±std 99.32 ±0.96 95.83 ±3.83 96.99 ±2.85 97.70 ±2.08 3.84 ±3.56
95% CI [96.49, 99.99] [88.33, 99.97] [91.31, 99.98] [93.45, 99.97] [0.04, 10.63]
MIT–BIH
Mean ±std 99.01 ±3.01 99.59 ±1.53 99.71 ±1.32 99.19 ±2.14 1.29 ±3.40
95% CI [82.35, 99.96] [89.87, 100.00] [90.91, 100.00] [88.91, 99.98] [0.04, 12.28]
4.2. Transversal Optimization
In this subsection, we checked the method independence from the database used during the tuning
process. To do so, we first tested on the MIT–BIH, and after that on the rest of available databases.
For better performance and model optimization, some empirically found improvements were
applied, which are summarized next: (1) the low-pass-filter cut-off frequency was set to 25 Hz;
Sensors 2018,18, 1387 14 of 24
(2) the refractory period was adjusted according to empirical findings; and (3) the amplitude threshold
was set to be updated every three seconds for a closer signal follow–up. These updates required a
new
K
-fold cross-validation process, which was applied to MIT–BIH, and it was established
K=
12.
The results were now α= 3 and β= 0.5, and the averaged error was 0.85%.
In Figure 7, the improvements can be scrutinized on several examples. The first two plots depict
leads one and two for record 100. In these panels, the reader can see clearly how the new processing
is able to closer follow the signal evolution, so that the detection is possible even in the second lead,
where it is found a relevant difference on the level of the signal form one beat to another. The following
four panels also show the same concept, but now overprinted at two intermediate feature signals,
without applying the last mentioned improvements (left) and after applying it (right). Table 6details
these improved results on a per-patient basis.
1.073 1.074 1.075 1.076 1.077 1.078 1.079
Samples 105
-0.5
0
0.5
1
1.5
Amplitude in mV
Lead 1 record 100 from MIT-BIH
Signal
Real points
Detection non optimized
Detection optimized
1.073 1.074 1.075 1.076 1.077 1.078 1.079
Samples 105
-0.4
-0.2
0
0.2
Amplitude in mV
Lead 2 record 100 from MIT-BIH
Signal
Real points
Detection non optimized
Detection optimized
1.074 1.076 1.078
Samples 105
0.01
0.02
0.03
0.04
0.05
Amplitude in mV3/2
First feature signal record 100 from MIT-BIH
Feat sig
Non-optimized th
Detected
1.074 1.076 1.078
Samples 105
0.005
0.01
0.015
0.02
Amplitude in mV3/2
Second feature signal record 100 from MIT-BIH
Feat sig
Non-optimized th
Detected
1.074 1.076 1.078
Samples 105
0.01
0.02
0.03
0.04
0.05
Amplitude in mV3/2
First feature signal record 100 from MIT-BIH
Feat sig
Optimized th
Detected
1.074 1.076 1.078
Samples 105
0.005
0.01
0.015
0.02
Amplitude in mV3/2
Second Feature signal record 100 from MIT-BIH
Feat sig
Optimized th
Detected
Figure 7.
Comparison between the non-optimized and the optimized algorithm in record 100. The two
upper panels depict leads 1 and 2. The four lower panels depict the feature signals for both leads.
Sensors 2018,18, 1387 15 of 24
Table 6. Results of the transversal-optimized algorithm using the complete MIT–BIH.
MITBIH
Record Sen Spe PPV Acc Error
100 99.91 100.00 100.00 99.96 0.09
101 99.89 99.87 99.89 99.88 0.21
102 99.95 100.00 100.00 99.98 0.05
103 98.61 99.68 99.90 98.85 1.49
104 99.78 99.79 99.82 99.78 0.40
105 98.64 97.05 98.95 98.16 2.41
106 99.56 99.75 99.85 99.62 0.59
107 99.91 99.95 99.95 99.93 0.14
108 99.49 98.76 99.21 99.05 1.30
109 99.76 100.00 100.00 99.88 0.24
111 99.86 100.00 100.00 99.93 0.14
112 100.00 100.00 100.00 100.00 0.00
113 99.94 100.00 100.00 99.97 0.06
114 99.84 100.00 100.00 99.92 0.16
115 99.69 99.46 99.95 99.67 0.36
116 99.59 99.90 99.92 99.72 0.50
117 99.93 100.00 100.00 99.97 0.07
118 99.96 100.00 100.00 99.98 0.04
119 99.95 99.30 99.95 99.91 0.10
121 100.00 100.00 100.00 100.00 0.00
122 99.96 100.00 100.00 99.98 0.04
123 99.93 100.00 100.00 99.97 0.07
124 99.94 100.00 100.00 99.97 0.06
200 95.42 98.94 99.20 96.88 5.34
201 99.49 100.00 100.00 99.75 0.51
202 99.95 100.00 100.00 99.98 0.05
203 99.26 99.66 99.80 99.41 0.94
205 99.32 100.00 100.00 99.66 0.68
207 99.68 88.89 90.26 94.19 11.08
208 90.66 99.37 99.74 92.96 9.58
209 99.70 99.32 99.93 99.67 0.37
210 99.62 99.92 99.96 99.72 0.42
212 99.78 99.89 99.96 99.81 0.25
213 99.88 100.00 100.00 99.94 0.12
214 99.82 99.94 99.96 99.87 0.22
215 99.97 100.00 100.00 99.99 0.03
217 99.82 100.00 100.00 99.91 0.18
219 99.91 100.00 100.00 99.95 0.09
220 99.95 100.00 100.00 99.98 0.05
221 99.96 100.00 100.00 99.98 0.04
222 99.88 100.00 100.00 99.94 0.12
223 99.58 100.00 100.00 99.79 0.42
228 99.81 99.83 99.95 99.81 0.24
230 99.96 100.00 100.00 99.98 0.04
231 99.94 100.00 100.00 99.97 0.06
232 99.78 98.55 99.16 99.28 1.07
233 99.68 100.00 100.00 99.84 0.32
234 99.93 100.00 100.00 99.96 0.07
Mean [CI low–CI high] 99.48 [90.66–100] 99.54 [88.89–100] 99.69 [90.26–100] 99.46 [92.96–100] 0.85 [0–9.58]
Sensors 2018,18, 1387 16 of 24
For testing purposes, the algorithm was also applied to the other databases described earlier,
namely, NSTDB, LTSTDB, LTAFDB, LTMITDB, INCART, and vPREDICT. As an example of the resulting
performance, Table 7shows the outcome for the NSTDB. The high performance values obtained show
that the algorithm is robust against strong noise. Table 8additionally summarizes the behavior on all
the non-LTM databases used in this paper. It is remarkable that the algorithm settings cope well with
all the databases simultaneously, making it a valid algorithm to use in clinical environments, since it
has proven its efficacy in this wide variety of cases.
Table 7. Results of the transversal-optimized algorithm using the NSTDB.
Record Sen Spe PPV Acc Error
118e00 94.56 82.89 89.75 89.61 16.24
118e6 98.29 95.88 97.77 97.29 3.95
118e12 99.74 99.78 99.87 99.72 0.40
118e18 99.96 100.00 100.00 99.98 0.04
118e24 99.96 100.00 100.00 99.98 0.04
118e-6 84.64 75.62 83.14 80.19 32.53
119e00 96.02 81.86 88.70 89.92 16.21
119e06 99.45 95.38 97.34 97.82 3.27
119e12 99.95 98.88 99.90 99.86 0.15
119e18 99.95 99.30 99.95 99.91 0.10
119e24 99.95 99.30 99.95 99.91 0.10
119e-6 89.28 71.09 80.20 80.76 32.76
Table 8. Results of the transversal-optimized algorithm using non-LTM databases.
Database Sen Spe PPV Acc Error
vPREDICT 99.55 [98.06–99.99] 95.64 [87.53–99.97] 96.87 [90.63–99.98] 97.77 [93.58–99.98] 3.75 [0.03–10.56]
LTAFDB 98.89 [91.08–99.99] 96.38 [80.96–99.95] 97.56 [88.71–99.96] 97.69 [91.01–99.92] 3.66 [0.07–12.92]
LTSTDB 99.88 [99.19–100] 99.80 [97.49–100] 99.87 [98.74–100] 99.84 [98.54–100] 0.25 [0.00–1.87]
LTMITDB 99.68 [98.86–99.99] 98.52 [96.57–99.97] 99.02 [97.62–99.98] 99.16 [98.25–99.98] 1.32 [0.03–2.86]
INCART 99.36 [100–95.33] 99.63 [96.64–100] 99.78 [98.00–100] 99.46 [96.92–100] 0.86 [0–4.89]
4.3. Application to LTM Recordings
After the algorithm has been designed, tuned, and proven accurate over a number of short-term
databases, we finally evaluated its behavior in LTM recordings. To do so, the PREMARIS database
was used. PREMARIS contains a set of 84 archives of 7-day recordings from patients suffering of HF
with left ventricular dysfunction. The details of this database have been introduced previously in the
literature [
25
]. The computational cost of computing over half a million beats on each record suggested
to evaluate a subsample of five records of this database. Table 9summarizes the results over these
LTM records. The first finding is a higher number of errors compared with the 24-h recording signals
in absolute terms. The reason for this could be found in the difficulties for the user to keep the same
care level about the system over the seven days, hence inducing higher noise and probably worst
electrodes contact.
Sensors 2018,18, 1387 17 of 24
Table 9. Results of the optimized algorithm on 7-day Holter database PREMARIS.
Patient SEN SPE PPV ACC ERR
1 99.32 91.00 93.46 95.46 7.70
2 99.44 88.32 91.04 94.09 11.05
3 99.87 95.32 96.61 97.77 3.72
4 99.90 96.30 97.22 98.26 2.97
5 99.98 98.67 99.09 99.43 0.94
Mean [CI low–CI high] 99.70 [99.98–53.43] 98.67 [93.92–60.52] 99.09 [95.48–57.61] 99.43 [97–55.94] 86.89 [5.28–0.94]
In order to explore this speculation, the detection error evolution was plotted as a function of time
for all the cases, which allowed a visual representation of the possible errors produced throughout the
days of the test. As seen in Figure 8, upper panel, the evolution of the detection errors shows a clear
increase in many cases about the middle of the test (days 3 to 5), and a reduction in less proportion
when approaching to its end (days 6 to 7). Partial or complete disconnection of the electrodes were the
main causes of error, which is related either to incorrect positioning after disconnection, or to daily
physical activity of the patient. Any of these situations debuts in an increase of noise levels in the
signal. An example of this is shown in the same figure, in the middle and lower panels.
1234567
Time in days
0
5
10
15
20
25
30
%
Error
Patient 1
Patient 2
Patient 3
Patient 4
Patient 5
9.754 9.7545 9.755 9.7555 9.756
Samples 106
5
10
15
20
25
30
35
40
Amplitude in mV
Signal quality LTM record 1 day 4
5.4546 5.4548 5.455 5.4552 5.4554 5.4556 5.4558 5.456
Samples 106
5
10
15
20
25
30
35
40
Amplitude in mV
Signal quality LTM record 2 day 5
Figure 8.
Evolution of the detection error for each LTM patient across the test duration (
up
),
and examples of noise types present in 7-day Holter recordings (middle and down).
As previously stated, one of the main purposes of this study was to scrutinize what are the main
issues in R-wave detector when applied to LTM. Hence, we manually analyzed the first 500 errors
from the worst day in terms of detection errors for each patient, and we classified them as shown in
Table 10. On the one hand, those beats that were not detected by the commercial algorithm, but were
Sensors 2018,18, 1387 18 of 24
correctly detected by the proposed algorithm, are referred to as Holter miss-detection. Inside this group,
we split them into two separate groups, namely, those corresponding to a Single beat or those included
in a Burst of detection errors. On the other hand, those detection errors that were generated by
our final algorithm are categorized under the name Algorithm errors, regardless they are FP or FN.
As an informative type of identified errors, we included those ones that were not detected either by the
commercial software or by our algorithm, and they were called Shared errors. The reader can check in
Table 10 that about 90% of error detections did not correspond to real errors of the presented algorithm.
In addition, within the last 10% of detection errors, up to 5% corresponded to effective errors of the
introduced algorithm, and the other 5% corresponded to beats that none of the methods were able to
detect. Figure 9illustrates some examples of error detections in the commercial software, and Figure 10
shows additional examples of error detections by the newly introduced algorithm.
Table 10.
Results of analyzing the first 500 errors in the worst day of the LTM recordings in
PREMARIS database.
Holter Miss-Detections Algorithm Errors Shared Errors
Patient Analyzed Segments (h) Burst Single Beat FP FN
1 4.23 217 233 25 15 10
2 3.5 449 17 10 10 14
3 2.54 59 335 19 18 69
4 5.32 118 342 2 13 25
5 7.52 37 462 0 1 0
Figure 9.
Example of single beat miss-detection in conventional Holter LTM. The panels in the first
row show leads 1 and 2 for an example of Holter miss-detection, the second row shows the feature
signal for the leads plotted before. The last four panels follow the same scheme for other Holter
miss-detection examples.
Sensors 2018,18, 1387 19 of 24
Figure 10.
Example of beat miss-detection in proposed algorithm on 7-day Holter recording. The panels
in the first row show leads 1 and 2 for an example of algorithm miss-detection, the second row shows
the feature signal for the leads plotted before. The last four panels follow the same scheme for other
algorithm miss-detection examples.
5. Discussion and Conclusions
A number of methods and algorithms have been reported in literature for R-wave detection in
ECG recordings. A subset including the most relevant R-wave detectors in the literature, as well as
new ones based on multilead processing and specially devoted to LTM records, have been proposed,
developed, and benchmarked in the present work. LTM algorithms have specific requirements when
compared to standard ECG for two reasons. First, the signal processing needs to be extremely efficient
to cope with extremely long records (over half a million beats in 7-day recordings), and also it should
be simple enough to run on commercial hardware. Second, a much deeper and precise analysis is
necessary to extract a clean R-wave, as the signals recorded in LTM incorporate a significantly higher
content on noise and other artifacts. A reason for this much worse signal quality is that the recording
is not performed in an ambulatory environment. However, even more, the extension over one week or
Sensors 2018,18, 1387 20 of 24
more in LTM, it unfailingly does not facilitate the attention that the patient can make of the system and
the electrodes, this becoming an uncomfortable element as the duration extends.
In order to work under these circumstances, an innovative algorithm has been proposed in this
paper to handle this new singular environment of LTM, which was compared with the most promising
published methods in literature, both over public and private databases. To check its performance in
LTM environments, a gold standard database has been labeled by experts and clinicians who manually
evaluated two sets of 24-h and 7-day Holter recordings. As a result, not only the designed algorithm
improves the detection, but also it incorporates relevant efficient noise removal that can be considered
for other Holter and ECG recording types, and not only to LTM.
In order to justify the good behavior of the polling strategy, we can argue that, as a voting
strategy itself, and from the statistical point of view, it can be assimilated to median modeling in
statistical biological processes. Thus, in our problem, and as in many biological processes where
results are better characterized by using a figure of merit that is resilient to single-errors or outliers,
the pooling method offers a solid tool to better detect QRS in multilead ECG recordings, as it is ready
to overcome singularities, impulsive noise, and other effects that may apply due to the more than
frequent uncoupling of electrodes in LTM.
In general, the proposed scheme yields better results than other counterpart detectors in such
a way that some of the improvements proposed in this work are not limited to LTM recordings, but they
also can be very well applied to conventional Holter. The proposed methods to increase robustness
against noise are among the most reliable procedures to hold the performance. Nonetheless, one of
the main drawbacks of the current application is the long patient monitoring period itself. Having to
deal with the acquisition system during such long time periods becomes uncomfortable and highly
tiring for any individual, which thus inevitably reduces the proper care into the system during some
of the periods of the seven recording days. This natural disregard causes, therefore and unavoidably,
disconnection failures of the electrodes, which are in fact the main cause of errors, and also the reason
for which the error rate increases in LTM. Improving the recording systems to be technically more
efficient would partially alleviate this situation. However, from an analytical standpoint, a better
understanding and characterization of noise is needed to enable a correct delineation of the ECG
according to its clinical validity. The elimination of invalid segments from a clinical viewpoint would
partially and realistically reduce the error rate in QRS detection. Some works, such as [
13
], are already
facing this problem based on the belief that approaching this issue would provide with room for
improvement.
Not all of the categories of proposed ECG analysis algorithms are presented in this work. We only
considered those with higher prevalence in the literature, and specially those included on articles
dealing with R-wave detection. Other families have not been considered here because they seem
not to match with the specific LTM criteria, specially the computational efficiency. As an example,
time–frequency analysis methods offer a versatile representation in both the frequency and time
domains, by varying the scaled parameters according to global and local signal features, as in the case
of detectors based on CWT [
31
], Discrete Wavelet Transformation [
34
], STFT [
32
], HiT [
33
], or HHT
and Empirical Mode Decomposition [
35
]. The adaptation ability of these methods demand high
computational requirements, which currently precludes their use in LTM scenarios. Methods based
on advanced statistical and multivariate analysis, like artificial neural networks (ANN), offer high
adaptability, but they also demand extended processing time for training. In the field of ECG signal
processing, ANN have been proposed as adaptive preprocessing stages [
38
], R-peak detectors [
48
],
and commonly as classifiers of the detected and segmented QRS complexes [
49
]. ANN–based detectors
deal with ECG signal prediction and examination error; therefore, their extension to LTM scenarios
remains to be efficiently formulated.
In the future, we want to scrutinize the improvements of these lately mentioned families of
methods and to check what can be expected from them in terms of the preprocessing, feature extraction,
and detection logic in LTM scenarios, as a natural continuation of this work. Very recent alternatives are
Sensors 2018,18, 1387 21 of 24
being proposed in the very last generation of works in the literature, which point towards an inflection
point in this field. Works in short-term monitoring (duration of some minutes) and new public
databases [
50
] can be found, as the intelligent heart-monitoring public cryptosystem, which involves
a patient-worn ECG sensor and a remote monitoring station, using PT and classification trees as
heartbeat detection and classification algorithms, respectively. More advanced methods, such as
sample entropy, fast Fourier transform, and ANN, were recently selected for the integrated detection
algorithms [
21
], in order to validate an arrhythmia detection method for supra and ventricular
ectopic beats and atrial/ventricular fibrillation, when using the ANSI/AAMI EC57:2012 standard.
Lossy methods, which traditionally were avoided in this setting, have been successfully revisited [
18
],
and different compression methods based on the compression ratio and percentage root-mean-squared
difference were proposed and advantageously benchmarked, including in terms of the accuracy of
R-wave detection. Many other works are being developed in the very last years accounting for wavelet
multiresolution analysis with promising results [51].
Special attention is being paid to deep learning approaches in ECG monitoring scenarios. In [
24
],
active classification of ECG signals is achieved using a suitable unsupervised feature learning from the
raw ECG data, using a softmax regression layer on the top of the resulting hidden representation layer
and fine-tuning it with active learning. Their results showed significant accuracy improvements with
less expert interaction and faster online retraining compared to existing methods. Other groups [
52
]
showed a deep learning framework to implement a technique for the classification of three selected
different cardiac conditions. It was proven to be an efficient automatic arrhythmia detection
method while eliminating the burden of training a deep convolutional neural network from scratch.
Despite some of these solutions seeming to obviate the need of R-wave detection stages, we should keep
in mind that many diagnostic parameters still have to be made on morphological measurements within
the delineated ECG, even (specially) in LTM scenarios. Many studies have tested algorithm detection
with 24-h Holter recordings, but only some segments of a few seconds or minutes are considered.
Two major issues require attention from researchers when analyzing R-wave detection algorithms.
First, the never-ending and growing list of proposed methods in literature almost discourages any
attempt to check them all. Second, the difficulty to be able to compare the results as the coding is not
described deeply enough in papers, and databases are not always available or shared. This situation is
even more complex, as when it comes to LTM cardiac recordings, we are still lacking from a standard
and shared framework for comparison. For that reason, in this paper, we first addressed a Longitudinal
Optimization, which allowed to benchmark, tune, and improve the best available waveform algorithms
in the literature for non-LTM. After that, a modified and multilead R-wave detection algorithm based on
PT was proposed with a suitable segmentation and using the complete, and Polling-function combined,
content present in all of the available leads. As a consequence, significant improvements were found
in terms of accuracy and computational efficiency. In order to get a Transversal Optimization, further
analysis and comparisons over public databases were performed to validate and ensure the results’
independence from the tuning databases, sampling frequency, or any other over-fitting effect. Hence,
we proved that it is possible to implement an accurate and efficient method, with low computational
cost, valid for LTM and suitable for its clinical use. Our 7-day Holter recordings on previous research
have been found to be useful in HF patients for the detection of non-sustained atrial or ventricular
arrhythmias [
26
], for the evaluation of heart rate turbulence [
53
], and for the assessment of autonomic
parameters obtained from heart rate variability analysis [25].
Author Contributions:
F.-J.G.-B., J.L.R.-A. and A.G.-A. proposed the clinical and technical definitions of QRS
detector. F.M.M.-M. developed and tested the algorithm. E.E.-V. collected the databases. A.G.-A., J.L.R.-A.,
F.-J.G.-B., Z.M.-B. and M.B.-V. reviewed the manuscript and managed the process. All authors contributed to
write the manuscript.
Funding:
This research was funded by Ministerio de Economía, Industria y Competitividad, Gobierno de España,
Agencia Estatal de Investigación, and co-funded by European Regional Development Fund, under the following
project codes: TEC2013-48439-C4-1-R, TEC2013-48439-C4-2-R, TEC2016-75161-C2-1-R, TEC2016-75161-C2-2-R,
TEC2016-81900-REDT, TEC2015-64835-C3-1-R.
Sensors 2018,18, 1387 22 of 24
Conflicts of Interest: The authors declare no conflict of interest.
References
1.
Dagres, N.; Kottkamp, H.; Piorkowski, C.; Weis, S.; Arya, A.; Sommer, P.; Bode, K.; Gerds-Li, J.;
Kremastinos, D.T.; Hindricks, G. Influence of the duration of Holter monitoring on the detection of
arrhythmia recurrences after catheter ablation of atrial fibrillation Implications for patient follow-up.
Int. J. Cardiol. 2010,139, 305–306. [CrossRef] [PubMed]
2.
Steven, D.; Rostock, T.; Lutomsky, B.; Klemm, H.; Servatius, H.; Drewitz, I.; Friedrichs, K.; Ventura, R.;
Meinertz, T.; Willems, S. What is the real atrial fibrillation burden after catheter ablation of atrial fibrillation?
A prospective rhythm analysis in pacemaker patients with continuous atrial monitoring. Eur. Heart J.
2008
,
29, 1037–1042. [CrossRef] [PubMed]
3.
Jabaudon, D.; Sztajzel, J.; Sievert, K.; Landis, T.; Sztajzel, R. Usefulness of ambulatory 7-day ECG monitoring
for the detection of atrial fibrillation and flutter after acute stroke and transient ischemic attack. Stroke
2004
,
35, 1647–1651. [CrossRef] [PubMed]
4.
Pastor-Pérez, F.; Manzano-Fernández, S.; Goya-Esteban, R.; Pascual-Figal, D.; Barquero-Pérez, O.;
Rojo-Álvarez, J.; Everss-Villalba, E.; Martínez Martínez-Espejo, M.; Valdes Chavarri, M.; García-Alberola, A.
Prognostic significance of long-period heart rate rhythms in chronic heart failure. Circ. J.
2012
,76, 2124–2129.
[CrossRef] [PubMed]
5.
Kohler, B.U.; Hennig, C.; Orglmeister, R. The principles of software QRS detection. IEEE Eng. Med. Biol. Mag.
2002,21, 42–57. [CrossRef] [PubMed]
6. Mehmet, E. ECG beat classification using neuro-fuzzy network. Pattern Recognit. Lett. 2004,25, 1715–1722.
7. Walraven, G. Basic Arrhythmias; Addison Wesley: Boston, MA, USA, 2010.
8.
Sornmo, L.; Paklm, O.; Nygards, M.E. Adaptive QRS Detection: A Study of Performance. IEEE Trans.
Biomed. Eng. 1985,BME-32, 392–401. [CrossRef]
9.
Pan, J.; Tompkins, W. A Real-Time QRS Detection Algorithm. IEEE Trans. Biomed. Eng.
1985
,32, 230–236.
[CrossRef] [PubMed]
10.
Gutiérrez-Rivas, R.; García, J.J.; Marnane, W.P.; Hernández, A. Novel Real-Time Low-Complexity QRS
Complex Detector Based on Adaptive Thresholding. IEEE Sens. J. 2015,15, 6036–6043. [CrossRef]
11.
Zidelmal, Z.; Amirou, A.; Ould-Abdeslam, D.; Moukadem, A.; Dieterlen, A. QRS detection using S-Transform
and Shannon energy. Comput. Methods Prog. Biomed. 2014,116, 1–9. [CrossRef] [PubMed]
12.
Merah, M.; Abdelmalik, T.; Larbi, B. R-peaks detection based on stationary wavelet transform.
Comput. Methods Prog. Biomed. 2015,121, 149–160. [CrossRef] [PubMed]
13.
Everss-Villalba, E.; Melgarejo-Meseguer, F.; Blanco-Velasco, M.; Gimeno-Blanes, F.; Sala-Pla, S.; Rojo-Álvarez, J.;
García-Alberola, A. Noise Maps for Quantitative and Clinical Severity Towards Long-Term ECG Monitoring.
Sensors 2017,17, 2448. [CrossRef] [PubMed]
14.
Barquero-Pérez, O.; Quintanilla, T.; García-Muñoz, J.; Soguero-Ruiz, C.; Wilby, M.; de la Rosa, M.; Cabanas, M.;
González, I.; Bravo, R.; García-Alberola, A.; et al. eLab: A web-based platform to perform HRV and HRT analysis
and store cardiac signals. In Proceedings of the Computing in Cardiology Conference (CinC), Zaragoza, Spain,
22–25 September 2013; pp. 21–24.
15.
Casañez-Ventura, A.; Gimeno-Blanes, F.; Rojo-Álvarez, J.; Flores-Yepes, J.; Gimeno-Blanes, J.; López-Ayala, J.M.;
García-Alberola, A. QRS delineation algorithms comparison and model fine tuning for automatic clinical
classification. In Proceedings of the Computing in Cardiology Conference (CinC), Zaragoza, Spain,
22–25 September 2013; pp. 1163–1166.
16.
Mark, R.; Schluter, P.; Moody, G.; Devlin, P.; Chernoff, D. An annotated ECG database for evaluating
arrhythmia detectors. In Proceedings of the Annual Conference IEEE Engineering in Medicine and Biology
Society Frontiers of Engineering in Health Care, Philadelphia, PA, USA, 20–21 September 1982; pp. 205–210.
17.
Moody, G.; Mark, R. The impact of the MIT-BIH arrhythmia database. IEEE Eng. Med. Biol. Mag.
2001
,
20, 45–50. [CrossRef] [PubMed]
18.
Elgendi, M.; Mohamed, A.; Ward, R. Efficient ECG Compression and QRS Detection for E-Health
Applications. Sci. Rep. 2017,7, 459. [CrossRef] [PubMed]
19.
Moody, G.; Muldrow, W.W.; Mark, R. A noise stress test for arrhythmia detectors. Comput. Cardiol.
1984
,
11, 381–384.
Sensors 2018,18, 1387 23 of 24
20.
Goldberger, A.; Amaral, L.; Glass, L.; Hausdorff, J.; Ivanov, P.; Mark, R.; Mietus, J.; Moody, G.; Peng, C.K.;
Stanley, H. PhysioBank, PhysioToolkit, and PhysioNet Components of a New Research Resource for Complex
Physiologic Signals. Circulation 2000,101, 215–220. [CrossRef]
21.
Sadrawi, M.; Lin, C.; Lin, Y.; Hsieh, Y.; Kuo, C.; Chien, J.; Haraikawa, K.; Abbod, M.; Shieh, J. Arrhythmia
Evaluation in Wearable ECG Devices. Sensors 2017,17, 2445. [CrossRef] [PubMed]
22.
Jager, F.; Taddei, A.; Moody, G.B.; Emdin, M.; Antoliˇc, G.; Dorn, R.; Smrdel, A.; Marchesi, C.; Mark, R.G.
Long-term ST database: A reference for the development and evaluation of automated ischaemia detectors
and for the study of the dynamics of myocardial ischaemia. Med. Biol. Eng. Comput.
2003
,41, 172–182.
[CrossRef] [PubMed]
23.
Petrutiu, S.; Sahakian, A.V.; Swiryn, S. Abrupt changes in fibrillatory wave characteristics at the termination
of paroxysmal atrial fibrillation in humans. Europace 2007,9, 466–470. [CrossRef] [PubMed]
24.
Rahhal Al, M.; Bazi, Y.; AlHichri, H.; Alajlan, N.; Melgani, F.; Yager, R. Deep learning approach for active
classification of electrocardiogram signals. Inf. Sci. 2016,345, 340–354. [CrossRef]
25.
Goya-Esteban, R.; Barquero-Pérez, O.; Caamaño-Fernández, A.; Rojo-Álvarez, J.; Pastor-Pérez, F.;
Manzano-Fernández, S.; García-Alberola, A. Usefulness of 7-day Holter monitoring for heart rate variability
nonlinear dynamics evaluation. In Proceedings of the Computing in Cardiology (CinC), Hangzhou, China,
18–21 September 2011; Volume 38, pp. 105–108.
26.
Pastor-Pérez, F.; Manzano-Fernández, S.; Goya-Esteban, R.; Pascual-Figal, D.; Barquero-Pérez, O.; Rojo-Álvarez,
J.; Martinez-Espejo, M.; Chavarri, M.V.; García-Alberola, A. Comparison of Detection of Arrhythmias in Patients
With Chronic Heart Failure Secondary to Non-Ischemic Versus Ischemic Cardiomyopathy by 1 Versus 7-Day
Holter Monitoring. Am. J. Cardiol. 2010,106, 677–681. [CrossRef] [PubMed]
27.
Yeh, Y.; Wang, W.; Chiou, C. Cardiac arrhythmia diagnosis method using linear discriminant analysis on
ECG signals. Measurement 2009,42, 778–789. [CrossRef]
28.
Hamilton, P.; Tompkins, W. Quantitative Investigation of QRS Detection Rules Using the MIT/BIH
Arrhythmia Database. IEEE Trans. Biomed. Eng. 1986,33, 1157–1165. [CrossRef] [PubMed]
29.
Chouhan, V.; Mehta, S. Detection of QRS Complexes in 12-lead ECG using Adaptive Quantized Threshold.
Int. J. Comput. Sci. Netw. Secur. 2008,8, 155–163.
30.
Arzeno, N.; Deng, Z.; Poon, C. Analysis of first-derivative based QRS detection algorithms. IEEE Trans.
Biomed. Eng. 2008,55, 478–484. [CrossRef] [PubMed]
31.
Lin, C.; Yang, C. Heartbeat classification using normalized RR intervals and wavelet features. In Proceedings
of the International Symposium on Computer, Consumer and Control (IS3C), Taichung, Taiwan,
10–12 June 2014; pp. 650–653.
32.
Uchaipichat, N.; Inban, S. Development of QRS Detection using Short-time Fourier Transform based
Technique. Int. J. Comput. Appl. CASCT 2010,8, 7–10. [CrossRef]
33.
Benítez, D.; Gaydecki, P.; Zaidi, A.; Fitzpatrick, A. A new QRS detection algorithm based on the Hilbert
transform. In Proceedings of the Computers in Cardiology, Cambridge, MA, USA, 24–27 September 2000;
pp. 379–382.
34.
Yochum, M.; Renaud, C.; Jacquir, S. Automatic detection of P, QRS and T patterns in 12 leads ECG signal
based on CWT. Biomed. Signal Process. Control 2016,25, 46–52. [CrossRef]
35.
Pal, S.; Mitra, M. QRS Complex detection using Empirical Mode Decomposition based windowing
technique. In Proceedings of the International Conference on Signal Processing and Communications
(SPCOM), Bangalore, India, 18–21 July 2010; Volume 20, pp. 1–5.
36.
Elgendi, M.; Mahalingam, S.; Jonkman, M.; Boer, F. A Robust QRS Complex Detection Algorithm Using
Dynamic Thresholds. In Proceedings of the International Symposium on Computer Science and its
Applications, Hobart, Australia, 13–15 October 2008; Volume 1, pp. 153–158.
37.
Chen, C.; Chuang, C. A QRS Detection and R Point Recognition Method for Wearable Single-Lead ECG
Devices. Sensors 2017,17, 1969. [CrossRef] [PubMed]
38.
Szilágyi, L.; Szilágyi, S.; Fördos, G.; Benyó, Z. Quick ECG analysis for on-line holter monitoring systems.
In Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology
Society, New York, NY, USA, 30 August–3 September 2006; Volume 1, pp. 1678–1681.
39.
Khawaja, A.; Dössel, O. Predicting the QRS complex and detecting small changes using principal component
analysis. Biomed. Technik 2007,52, 11–17. [CrossRef] [PubMed]
Sensors 2018,18, 1387 24 of 24
40.
Yu, S.N.; Chou, K.T. A switchable scheme for ECG beat classification based on independent component
analysis. Expert Syst. Appl. 2007,33, 824–829. [CrossRef]
41.
Yu, S.N.; Chou, K.T. Selection of significant independent components for ECG beat classification.
Expert Syst. Appl. 2009,36, 2088–2096. [CrossRef]
42.
Kužílek, J.; Lhotská, L.; Hanuliak, M. Processing holter ECG signal corrupted with noise: Using ICA for QRS
complex detection. In Proceedings of the 3rd International Symposium on Applied Sciences in Biomedical
and Communication Technologies, Roma, Italy, 7–10 November 2010; pp. 1–4.
43.
Kuzilek, J.; Lhotska, L. Electrocardiogram beat detection enhancement using Independent Component
Analysis. Med. Eng. Phys. 2013,35, 704–711. [CrossRef] [PubMed]
44.
Rodríguez, R.; Mexicano, A.; Bila, J.; Cervantes, S.; Ponce, R. Feature Extraction of Electrocardiogram Signals
by Applying Adaptive Threshold and Principal Component Analysis. J. Appl. Res. Technol.
2015
,13, 261–269.
[CrossRef]
45.
Pearson, K. On lines and planes of closest fit to systems of points in space. Lond. Edinb. Dublin Philos. Mag.
J. Sci. 1901,2, 559–572. [CrossRef]
46.
Haddadi, R.; Abdelmounim, E.; El Hanine, M.; Belaguid, A. Discrete Wavelet Transform based algorithm for
recognition of QRS complexes. In Proceedings of the International Conference on Multimedia Computing
and Systems (ICMCS), Marrakech, Morocco, 14–16 April 2014; pp. 375–379.
47.
Hyvärinen, A.; Karhunenen, J.; Oja, E. Independent Component Analysis. Neural Comput.
2001
,13, 504.
[CrossRef] [PubMed]
48.
Suzuki, Y. Self-organizing QRS-wave recognition in ECG using neural networks. IEEE Trans. Neural Netw.
1995,6, 1469–1477. [CrossRef] [PubMed]
49.
De Lannoy, G.; François, D.; Delbeke, J.; Verleysen, M. Weighted Conditional Random Fields for Supervised
Interpatient Heartbeat Classification. IEEE Trans. Biomed. Eng. 2012,59, 241–247. [CrossRef] [PubMed]
50.
Son, J.; Park, J.; Oh, H.; Bhuiyan, M.Z.; Hur, J.; Kang, K. Privacy-Preserving Electrocardiogram Monitoring
for Intelligent Arrhythmia Detection. Sensors 2017,17, 1360. [CrossRef] [PubMed]
51.
Qin, Q.; Li, J.; Yue, Y.; Liu, C. An Adaptive and Time-Efficient ECG R-Peak Detection Algorithm.
J. Healthc. Eng. 2017,7, 5980541. [CrossRef] [PubMed]
52.
Isin, A.; Ozdalili, S. Cardiac arrhythmia detection using deep learning. Procedia Comput. Sci.
2017
,
120, 268–275. [CrossRef]
53.
Fernández-Fernández, S.; Manzano-Fernández, S.; Pastor-Pérez, F.; Barquero-Pérez, O.; Goya-Esteban, R.;
Salar, M.; Rojo-Álvarez, J.; García-Alberola, A. Usefulness of 7-day Holter monitoring for the evaluation of
heart rate turbulence in chronic heart failure. Europace 2011,13, 484.
c
2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access
article distributed under the terms and conditions of the Creative Commons Attribution
(CC BY) license (http://creativecommons.org/licenses/by/4.0/).