Continuous ECG monitoring in hospital: part 2, practical issues

Authors:

Abstract and Figures

Continuous monitoring of the electrocardiogram (ECG) is a common intervention performed by cardiac nurses. In the first article of this two-part series, we evaluated the principal indications for ECG monitoring; these are the detection of arrhythmias, monitoring for myocardial ischaemia, and measurement of the QT interval. In this second article, the practical issues surrounding ECG monitoring are evaluated. The use of 3- and 5-electrode systems is discussed, and related to ECG theory and normal coronary artery anatomy. Recommendations are made for lead selection, correct electrode placement, and the recording of 12-lead ECGs. The current article concludes with a consideration of safe practice with respect to alarm management, and considers both individual and organisational approaches to reducing inappropriate alarms.
Content may be subject to copyright.
1
Michael Sampson
Continuous ECG monitoring in hospital: part 2, practical issues
British Journal of Cardiac Nursing, March 2018, Vol 13, No 3, p. 128-134.
This is the accepted manuscript version of a published work that appeared in final form in the
British Journal of Cardiac Nursing, copyright (c) MA Healthcare, after technical editing by the
publisher. To access the final edited and published work see
https://www.magonlinelibrary.com/doi/10.12968/bjca.2018.13.3.128
Originally published March 2018.
Abstract
Continuous monitoring of the electrocardiogram (ECG) is a common intervention performed by
cardiac nurses. In the first article of this two-part series, we evaluated the principal indications for
ECG monitoring. These are the detection of arrhythmias, monitoring for myocardial ischaemia, and
measurement of the QT interval. In this second article, the practical issues surrounding ECG
monitoring are evaluated. The use of 3- and 5-electrode systems is discussed, and is related to ECG
theory and normal coronary artery anatomy. Recommendations are made for lead selection, correct
electrode placement, and the recording of 12-lead ECGs. The article concludes with a consideration
of safe practice in respect of alarm management, and considers both individual and organisational
approaches to reducing inappropriate alarms.
Key words
ECG monitoring, arrhythmia, myocardial ischaemia, QT interval, ECG lead
Introduction
Continuous monitoring of the electrocardiogram (ECG) is a common intervention performed by
cardiac nurses. Last month, we evaluated the principal indications for ECG monitoring. These are
the detection of arrhythmias, monitoring for myocardial ischaemia, and measurement of the QT
interval. In this second article we conclude our review of the topic by examining the practical issues
that nurses face in relation to ECG monitoring. These include correct electrode positioning,
adequate skin preparation, and the appropriate selection of ECG leads. We evaluate the theory that
underpins ECG monitoring, and provide practical advice on how to set up 3- and 5-lead monitoring
systems. The article concludes with an evaluation of alarm management, and considers how to
reduce inappropriate alarms and ensure patient safety.
2
Leads and electrodes
Before discussing specific issues relating to ECG monitoring, we need to define two key terms. In
clinical practice, we often talk about a 3-lead ECG or a 5-lead ECG”. Unfortunately, this
terminology is somewhat misleading. In ECG terms, a “lead” is a recording of the heart’s electrical
activity, taken from a specific anatomical location, and not an electrical cable (Garcia, 2015). For
example, lead I views the heart from the left-hand side of the body, while lead aVF views it from
below. These different leads are obtained from skin electrodes placed on the body. It would be
more correct to describe monitoring systems as “3-electrode” or “5-electrode”, and that’s the
convention we’ll use in this article.
To understand the relationship between leads and electrodes, its useful to think about the 12-lead
ECG. A 12-lead ECG records 12 different electrical views of the heart, using 10 skin electrodes
(Hampton, 2013). The leads can be divided into two groups (figure 1).
Figure 1. The layout of the 12-lead ECG. The rhythm strip is a longer printout of lead II.
The limb leads are I, II, III, aVR, aVL, and aVF. They are recorded using four electrodes; one
attached to each limb. The standard position for these electrodes is on the wrists and ankles
(Campbell et al, 2017).
The chest (or precordial) leads are V1, V2, V3, V4, V5 and V6. They are recorded using the six
electrodes placed across the chest. Unlike the limb leads, each electrode corresponds with a
single ECG lead (Hampton, 2013).
3
The purpose of recording so many leads is to build up a three-dimensional picture of the heart’s
electrical activity (Garcia, 2015). The limb leads view the heart in a frontal plane, and see electrical
activity from a number of different angles (figure 2). The chest leads view the heart in a transverse
plane (i.e. through the chest wall), again from multiple angles (figure 3).
Figure 2. Anatomical perspective of the six limb leads
Figure 3. Anatomical perspective of the six chest leads
4
This detailed picture of cardiac electrical activity is extremely useful when diagnosing cardiac
conditions such as ischaemia, which are often localised to one area of the heart. For this purpose,
the ECG leads can be considered in three broad groups which correspond to anatomical regions of
the heart, and the coronary arteries that supply them (Houghton and Gray, 2014): -
Inferior leads: II, III and aVF. These leads view the heart from below, and get a good view of the
diaphragmatic surface of the left ventricle. This area of the heart is supplied by the right
coronary artery (RCA) in most people (Cademartiri et al, 2008).
Lateral leads: I, aVL, V5 and V6. These leads view the left-hand side of the heart, and “see” the
lateral wall of the left ventricle. The predominant artery supplying this region of the heart is the
left circumflex artery (LCx) (Rinta-Kiikka et al, 2014).
Anteroseptal leads: V1, V2, V3, V4. These leads sit over the front of the left ventricle, and view
its anterior, apical and septal regions. These areas are supplied primarily by the left anterior
descending (LAD) artery (Tortora and Nielsen, 2014).
Although continuous monitoring of all 12 ECG leads is possible, and is performed on some units, in
many clinical areas only one or two leads are constantly evaluated. This reduces the number of
electrodes that must be placed on the patient, as well as the complexity of the monitoring
equipment required. It also reduces the number of electrical views obtained; careful lead selection
is therefore required, especially when ischaemia monitoring is initiated (Sandau and Smith, 2009).
The lead(s) selected will depend on the monitor being used, the number of electrodes, and the
indication for monitoring. Let’s examine the two common monitoring setups, 3-electrode and 5-
electrode, and think about which ECG lead(s) might be selected for monitoring.
3-electrode ECG monitoring
3-electrode is the simplest form of continuous ECG monitoring, and is commonly employed in a
range of clinical settings including cardiac and intensive care units, operating theatres, and
resuscitation situations (Jevon, 2010). The system uses three limb electrodes, which are colour
coded: -
Right arm (RA red)
Left arm (LA yellow)
Left leg (LL - green)
Note that these colours refer to European equipment, and are different in the USA.
Rather than placing the electrodes on the limbs, they are positioned on the chest wall (Hatchett,
2017). This modification was first described by Mason and Likar (1966), and was developed for
exercise testing. Moving the limb leads away from the extremities allows greater patient movement,
and reduces artefact, making it suitable for both exercise testing and continuous ECG monitoring
(Pahlm and Wagner, 2008). Although suitable for these applications, it should be noted that Mason-
Likar electrode positions alter the amplitude of the QRS voltage in the inferior leads, and shift the
cardiac axis to the right (Jowett et al, 2005). Diagnostic 12-lead ECGs should, therefore, be
performed using limb leads placed at the limb extremities, and not on the chest wall (Campbell et al,
2017).
5
For continuous ECG monitoring, the positions of the three electrodes are as follows (figure 4): -
RA (red) - below the right clavicle
LA (yellow) - below the left clavicle
LL (green) - on the lower part of the ribcage, on the left-hand side
Figure 4. 3-electrode monitoring
Placement of ECG electrodes is a relatively simple task; however, care must be taken to ensure good
adhesion and correct location (Hatchett, 2017). Skin must be dry and free from oil, loose skin flakes
or excessive hair (Campbell et al, 2017). If necessary, skin should be cleaned with soap and water, or
an alcohol wipe, prior to electrode placement. Loose skin can be removed by gentle exfoliation
using a dry gauze swab, and chest hair removed by shaving or clipping in those areas where
electrodes will be placed (American Association of Critical Care Nurses [AACCN], 2013). Although
the placement of the arm electrodes below the clavicles is standard practice (Hatchett, 2017; Jevon,
2010), it should be noted that the right arm electrode is positioned where a defibrillator pad would
be placed. In patients at high risk of cardiac arrest, or who may need transcutaneous pacing,
placement of arm electrodes on the shoulders is recommended, keeping the chest clear for
defibrillation pads (Resuscitation Council UK, 2016). For the same reason, the left leg electrode
should be placed low down on the rib cage, and not immediately below the axilla.
3-electrode systems typically allow a single ECG lead to be displayed and monitored. The available
leads with this setup are I, II and III. These are referred to as the standard limb leads, and are the
three leads used in the early years of electrocardiography (AlGhatrif and Lindsay, 2012). For
arrhythmia monitoring, lead II is typically selected as P-waves and QRS complexes are usually upright
in this view, facilitating rhythm interpretation (Sandau et al, 2017). If lead II does not give a clear
ECG trace, the monitor can be adjusted to display lead I or III (Hatchett, 2017). If the ECG trace is
poor in all leads, electrode position and adherence should be checked. Although 3-electrode
6
systems are ideal for rhythm monitoring, the limited range of leads available make them a poor
choice for monitoring ischaemia or the QT interval. For these applications, a 5-electrode system is
better.
5-electrode ECG monitoring
In 5-electrode monitoring, two additional electrodes are added (figure 5): -
A right leg electrode (RL - black)
A chest electrode (V - white)
Figure 5. 5-electrode monitoring
The right leg electrode is placed on the lower ribcage, to the right-hand side of the body. Placement
of this electrode makes the remaining limb leads available; the augmented limb leads aVR, aVL
and aVF (Hampton, 2013). The chest electrode is typically placed in the V1 position the fourth
intercostal space at the right sternal margin (Hatchett, 2017). This lead gets a good view of atrial
electrical activity, so is useful in rhythm interpretation (Bennett, 2013). It is also useful in the
diagnosis of bundle branch blocks (BBB), as characteristic QRS changes are seen in V1 (Garcia, 2015).
If a different chest lead is required, the white electrode must physically be moved to that position.
For example, to monitor V4, the white electrode must be placed in the fifth intercostal space, in the
mid-clavicular line (Campbell et al, 2017). As 5-electrode monitors often allow two ECG leads to be
displayed and monitored, leads II and V1 are recommended for arrhythmia monitoring (Sandau and
Smith, 2009).
7
If a change in rhythm is detected during monitoring, and time allows (i.e. the patient is not in cardiac
arrest or close to it), a 12-lead ECG should be recorded using standard electrode positions (Pitcher
and Nolan, 2015). This ensures documentation of the arrhythmia, and may be helpful for rhythm
interpretation. Information available on the 12-lead may also have prognostic importance, or guide
future treatment such as catheter ablation (Bennett, 2013).
ST-segment monitoring
As discussed above, lead selection is one of the most important considerations in ST-segment
monitoring (Sandau et al, 2017). Ischaemia is often localised to one area of the heart; the leads that
“see” that area must be monitored if ischaemia is to be detected. If patients present with an
ischaemic ECG, this should form the basis for initial lead selection by identifying the ”ST fingerprint”
(Collins, 2010). This refers to the pattern of ST elevation or depression seen at admission; the leads
showing the most important changes are selected for subsequent monitoring.
When no prior ischaemic changes have been recorded, the leads that are most likely to detect
ischaemia should be selected (Sandau and Smith, 2009). Myocardial infarction most often affects
the anteroseptal and inferior regions of the heart (Hreybe and Saba, 2009). In suspected ACS,
selection of leads monitoring these regions is therefore suggested; leads III and V3 are common
choices (Collins, 2010). In a study using experimental coronary artery occlusion, ST elevation was
greatest in these two leads, although aVF had the greatest sensitivity and specificity for right
coronary artery occlusion (Haeberlin et al, 2014).
Following cardiac surgery, or PCI, lead selection should be guided by the vessels that have been
revascularized; for example, V3 is a good choice following PCI to the LAD (Sandau et al, 2017). Given
the limitations imposed by continuous monitoring (e.g. limited lead choice, modified electrode
positions), a standard 12-lead ECG should be recorded at least once per shift, and whenever
significant changes in ST-segments are detected (Sandau and Smith, 2009).
The main limitation to ST-segment monitoring, aside from those discussed above, is the large
variation in the ST-segment itself. There are multiple causes of ST-segment elevation or depression,
many of which are benign or unrelated to cardiac ischaemia (Garcia, 2015). The ST-segment is
deviated in bundle branch blocks, ventricular pacing, ventricular hypertrophy, and chronic digoxin
use (Houghton and Gray, 2014). Evaluation may be difficult or impossible in patients with these
conditions. ST-segment monitoring is also difficult in restless patients, because of artefact, and
should not be attempted if patients are agitated (Sangkachand et al, 2011).
Monitoring the QT interval
There are three methods of monitoring the QT interval; continuous fully automated, semi-
automated, and manual (Sandau et al, 2017). The method chosen will depend to some degree on
the monitoring equipment available, as well as usual practice within the individual unit.
Continuous, fully automated QT interval monitoring is a feature available on some, but not all,
modern patient monitors (Chenoweth et al, 2017). Special software allows the monitor to sample all
available ECG leads, and calculate representative QT and RR intervals. These are used to calculate
the QTc automatically every five minutes, with the result displayed on the monitor (Zhou et al,
2009). This has the advantage that changes in the QTc can be mapped over time; significant
8
prolongation can be picked up quickly and acted on. A study by Chenworth et al (2017)
demonstrated moderate correlation between measurements using an automated system and 12-
lead ECGs. The authors concluded that while the automated system was useful for trend evaluation,
12-lead ECG remains the standard for evaluation, and should be used to confirm the QT interval
before changes to treatment are made.
In semi-automated systems, electronic callipers are accessed via the monitor, and are used to
measure the QT interval and RR interval on the displayed rhythm strip. The monitor uses these
values to calculate the QTc (Sandau et al, 2017). Although this allows more flexibility than an
automated system, and greater human oversight, it also means that nurses must remember to take
measurements. Unless this is incorporated into standard observation charts and clinical routines, it
may be forgotten (Pickham and Drew, 2008). As with fully automated systems, significant changes
must be corroborated by 12-lead ECG.
Manual evaluation implies measurement of the QT interval using manual callipers and a 12-lead
ECG. This remains the gold standard for evaluation of the QT interval, and is probably the most
widely used method in clinical practice (Chenworth et al, 2017; Pickham and Drew, 2008; Sandau et
al, 2017). In practice, many clinicians use the QTc calculated by the ECG machine, which is printed at
the top of the recording. While this is often the most accurate calculation of the QT interval, ECG
machines are not infallible and manual verification is recommended (Bennett, 2013). Most clinicians
use Bazett’s formula to calculate QTc from the QT and R-R intervals (Rautaharju et al, 2009) (figure
6). This is fairly straightforward: -
The QT interval appears longer in some leads than others, and may be difficult to measure if the
onset and offset of waveforms are unclear. Choose a lead in which the QT appears the longest,
and which also has a clear beginning to the QRS complex, and a clear end to the T-wave
(Pickham and Drew, 2008). Rautaharju et al (2009) suggest that leads V2 and V3 often have the
longest QT intervals, while Yaldren and Richley (2014) suggest lead II as the clearest for
measurement. Other suggested leads are I, III and V5 (Yaldren and Richley, 2014).
Measure from the beginning of the QRS complex to the end of the T-wave. Do not include U-
waves; if they merge with the T-wave, choose a lead in which they do not appear if possible
(Rautaharju et al, 2009). Multiply each millimetre by 0.04 seconds to calculate the QT interval.
If you don’t have callipers, count small squares; each small square is 1 millimetre.
Next, measure the preceding R-R interval; this is the time interval between the two previous QRS
complexes. This is best measured by placing the tips of the callipers on the peaks of the R-waves
(Garcia, 2015). If the ECG shows atrial fibrillation, a mean R-R interval should be calculated by
either adding the shortest and longest R-R intervals, and dividing by two, or by averaging
multiple R-R intervals (Sandau et al, 2017). Multiply the number of millimetres by 0.04 seconds
to give the R-R interval, and then work out the square root using a calculator.
Divide the square root of the R-R interval into the QT interval to give the QTc in seconds (figure
6). Multiply by 1000 for the same value in milliseconds.
A number of online calculators exist to make this easier. All you need to use them is the heart
rate, and measured QT interval. One example can be found at
https://www.mdcalc.com/corrected-qt-interval-qtc
9
Figure 6. Using Bazett’s formula to calculate the QTc from the QT and R-R intervals
Note that because 12-lead ECG machines measure the QT interval in all 12 leads, and calculate using
the earliest QRS and latest T-wave, the QTc will often be slightly longer than that calculated
manually using a single lead (Rautaharju et al, 2009). For manual calculations, use the same lead for
every measurement; document the QTc in the patients notes (Pickham and Drew, 2008).
The suggested frequency of QT monitoring is a minimum of eight hourly, however frequency should
be increased for patients who are clinically unstable, or whose QTc is already prolonged (Sandau et
al, 2017). Progressive increases in QTc should be reported to the patient’s medical team, and urgent
advice sought if the QTc exceeds 500ms as a change in treatment, for example drug discontinuation,
may be required (Bennett, 2013). Bear in mind that increases in QRS width will also increase the
QTc; if the patient has a bundle branch block or is ventricularly paced, the effect of increased QRS
width on the QT interval should be factored in (Rautaharju et al, 2009). Close monitoring for TdP is
essential in any patient with significant QT prolongation.
Alarm management
The final consideration in continuous ECG monitoring is alarm management. Alarms are an essential
element of any monitoring system; their purpose is to alert the clinician to a change in physiological
parameters that requires action (Spratt, 2016). Unfortunately, most alerts are false alarms; in other
words, they do not require intervention on the part of the nurse. Drew et al (2014) studied audible
alarms in an intensive care unit over a 31-day period. On average, there were 187 alarms per bed
per day, of which less than 12% required action.
This high burden of false alarm is widely reported in the literature, and is undesirable for several
reasons (Graham and Cvach, 2010). Firstly, false alarms add to the clinical workload; dealing with
them reduces the time available for more important patient care. Secondly, regular false alarms
desensitize staff, and may lead to slow or inadequate responses when genuine emergencies occur.
Pelter and Drew (2015) describe the case of a man admitted for observation with chest pain and a
non-diagnostic 12-lead ECG. During the night shift, the monitor alarmed repeatedly because of
“bradycardia” and “asystole”. Each time, the patient was checked, and the rate and rhythm were
normal. After consulting with the on-call doctor, the nurse switched off the alarms. When the
patient was checked several hours later, he was cold and pulseless; the presumed cause of death
was ventricular arrhythmia secondary to myocardial ischaemia. This was not detected because the
alarms had been disabled. Similar case studies have been reported elsewhere; a US report on the
10
problem documented 98 adverse events in four years, 80 of which resulted in death (Joint
Commission, 2013).
The level of harm reported makes appropriate alarm management a priority for any nurse caring for
monitored patients. Alarms needs to be sensitive enough to detect important changes in patient
physiology, but not so sensitive that insignificant changes or artefact trigger alarms (Spratt, 2016).
Alarms can be divided into three categories: -
Parameter violation - upper and lower limits for physiological measurements are set by the
user, and the alarm sounds if the parameter is exceeded. In ECG monitoring, parameter alarms
would include upper and lower heart rate limits, ST-segment deviation, and QTc prolongation.
Arrhythmia - ECG rhythm is constantly evaluated by software that analyses waveform
morphology and rhythm. An alarm sounds if an abnormal rhythm is detected.
Technical for example poor signal, lead disconnection, and power failure.
Parameter violations are the most common cause of false alarms, and often result from limits that
are too narrow (Spratt, 2016). Typical default settings for heart rate are a low of 60bpm, and a high
of 120bpm. If the patient has a naturally low resting heart rate, or is beta-blocked, the alarm will
constantly sound when the rate falls below 60. Given that no action is taken in this case, a lower
limit would be more appropriate. Similarly, a patient with atrial fibrillation, and a heart rate
averaging 110bpm, will frequently trip the upper alarm limit due to fluctuation in heart rate. A
higher upper limit is appropriate if the problem is known, and treatment already underway (Pelter
and Drew, 2015). This underlines the importance of patient assessment; alarms should be checked
and adjusted to individual circumstances when taking over the care of any monitored patient, and
should only sound when an actionable change in physiology occurs (AACCN, 2013). Changes to
default settings within organisations can also be helpful. Altering default heart rate limits to a low of
50bpm, and a high of 150bpm, reduced alarm frequency by 43% in a study by Graham and Cvach
(2010).
Preventing false arrhythmia and technical alarms relies largely on correct electrode placement, and
appropriate lead selection (Sendelbach and Funk, 2013). Poor skin preparation can result in lead
detachment, and unnecessary alarms, whilst poorly sited electrodes can suffer from poor signal or
interference from patient movement (Hatchett, 2017). Selecting the best lead for monitoring
rhythm, ST-segment or QTc has been discussed already, however lead selection must also include
assessment of the size of the QRS complex and T-wave (Pelter and Drew, 2015). A low amplitude
QRS complex can result in under-detection by the monitor, and result in false bradycardia or asystole
alarms. This was the cause of the false alarms in the case study discussed earlier. A lead with a T-
wave that is very tall can result in overcounting the monitor thinks the T-wave is another QRS, and
double counts the heart rate as a result. When false alarms occur, electrode placement, lead
selection, and alarm parameters should all be reviewed with the aim of preventing further false
alarms. Alarms should never be permanently silenced (Pelter and drew, 2015). Training is
recommended to ensure that clinicians are familiar with monitoring equipment, and know how to
adjust and troubleshoot alarms (AACCN, 2013).
11
Conclusion
ECG monitoring can be a lifesaving intervention when used appropriately, however patients must be
carefully selected, and the right type of monitoring instigated. This monitoring compliments, but
does not replace, the need for regular evaluation using a 12-lead ECG, which remains the gold
standard for assessment of rhythm, ischaemia, and repolarisation.
Key areas issues in ECG monitoring include adequate skin preparation, correct siting of electrodes,
and well-judged lead selection. Good practice in these areas ensures effective monitoring, and helps
to prevent false alarms. Practice must be underpinned by a knowledge of the patient’s history and
previous ECGs, the reason for monitoring, and an understanding of ECG theory. Nurses must be
familiar with the equipment used, and know how to adjust and troubleshoot alarms. Most
importantly, all clinicians need to recognise the limitations of continuous ECG monitoring, and
ensure that action or omission does not result in monitoring failure.
Key points
Common monitoring setups include 3- and 5-electrode systems. 3-electrode systems can
monitor ECG leads I, II or III, and typically display a single lead. 5-electrode systems can monitor
these leads plus aVR, aVL, aVF, and a single chest lead. 5-electrode systems can often display
two ECG leads.
For arrhythmia monitoring, lead II is usually selected on 3-electrode systems. On 5-electrode
systems, V1 is usually added.
For ST-segment monitoring, lead selection is determined by the circumstances and prior ECG. In
known or suspected ACS, lead selection is based on ST-segment changes on the presenting ECG;
if these are absent, leads III and V3 are commonly monitored. Following cardiac surgery or PCI,
lead selection should be based on the arteries that have been revascularized.
In QT interval monitoring, the QTc may be measured automatically, semi-automatically, or
manually. Measurement by 12-lead ECG is considered the gold standard; many clinicians use the
machine calculated QTc, although manual verification is recommended. When evaluating the
QTc manually, select a lead in which the QT appears the longest, but in which QRS onset and T-
wave offset are clear. Use Bazett’s formula to convert QT to QTc.
Alarms are essential, and alert clinicians to important changes in physiology. They are often
poorly used, resulting in multiple false alarms, and alarm fatigue. Individualising alarm
parameters, correct electrode placement, and well-judged lead selection reduces the number of
false alarms. The cause of false alarms should always be investigated; alarms should never be
permanently disabled.
12
CPD questions
1. Find out what type of monitors are in use on your ward or unit. Make sure that you know how
to switch between ECG leads, and alter alarm parameters.
2. Would you know how to monitor a patient for ST segment changes, or QT prolongation? If not,
find out how this is usually done on your ward or unit, and familiarise yourself with the changes
that occur on the ECG, and how to measure them.
3. Think about a busy shift that you have worked recently. How much time did you spend dealing
with unnecessary alarms? In hindsight, how could you have reduced this?
References
AlGhatrif M, Lindsay J (2012) A brief review: history to understand fundamentals of
electrocardiography. Journal of community hospital internal medicine perspectives. 2(1):14383.
American Association of Critical Care Nurses (2013) AACCN practice alert: Alarm management,
available at https://www.aacn.org/clinical-resources/practice-alerts/alarm-management accessed
20/12/2017.
Bennett DH (2013) Bennett’s Cardiac Arrhythmias: Practical notes on interpretation and treatment,
8th edition, London: Hodder Arnold.
Cademartiri F, La Grutta L, Malagò L, Alberghina F, Meijboom WB, Pugliese F, Maffei E, Palumbo AA,
Aldrovandi A, Fusaro M, Brambilla V, Coruzzi P, Midiri M, Mollet NRA, Krestin GP (2008) Prevalence
of anatomical variants and coronary anomalies in 543 consecutive patients studied with 64-slice CT
coronary angiography, Eur Radiol, 18: 781791.
Campbell B, Richley D, Ross C, Eggett CJ (2017) Clinical Guidelines by Consensus: Recording a
standard 12-lead electrocardiogram. An approved method by the Society for Cardiological Science
and Technology (SCST). Available at:
http://www.scst.org.uk/resources/SCST_ECG_Recording_Guidelines_2017 (accessed 30/11/2017)
Chenoweth JA, Hougham AM, Colby DK, Ford JB, Sandhu J, Albertson TE, Sutter ME (2017)
Monitoring the corrected QT in the acute care setting: A comparison of the 12-lead
electrocardiogram and bedside monitor. The American Journal of Emergency Medicine. In press.
Collins ML (2010) Using continuous ST-segment monitoring, Nursing2016, 40:11-3.
Drew BJ, Harris P, Zègre-Hemsey JK, Mammone T, Schindler D, Salas-Boni R, Bai Y, Tinoco A, Ding Q,
Hu X (2014) Insights into the problem of alarm fatigue with physiologic monitor devices: a
comprehensive observational study of consecutive intensive care unit patients. PloS one.
9(10):e110274.
Garcia TB (2015) 12-lead ECG: The Art of Interpretation. 2nd Edition. Burlington, Ma : Jones and
Bartlett
Graham KC, Cvach M (2010) Monitor alarm fatigue: standardizing use of physiological monitors and
decreasing nuisance alarms, American Journal of Critical Care, 19(1):28-34.
13
Haeberlin A, Studer E, Niederhauser T, Stoller M, Marisa T, Goette J, Jacomet M, Traupe T, Seiler C,
Vogel R (2014) Electrocardiographic ST-segment monitoring during controlled occlusion of coronary
arteries. Journal of Electrocardiology. 47(1):29-37.
Hampton JR (2013) The ECG made easy, 8th edition, London: Churchill Livingstone.
Hatchett R (2017) Cardiac monitoring and the use of a systematic approach in interpreting
electrocardiogram rhythms, Nursing Standard, 32, 11, 51-62.
Houghton AR & Gray D (2014) Making sense of the ECG: A hands-on guide, 4th edition, Boca Raton:
CRC Press.
Hreybe H & Saba S (2009) Location of Acute Myocardial Infarction and Associated Arrhythmias and
Outcome, Clinical Cardiology, 32(5), 274-277.
Jevon P (2010) An introduction to electrocardiogram monitoring, Nursing in Critical Care, 15(1), 34-
38.
Joint Commission (2013) Sentinel Event Alert Issue 50: Medical device alarm safety in hospitals,
available at https://www.jointcommission.org/sea_issue_50/ accessed 20/12/2017.
Jowett NI, Turner AM, Cole A, Jones PA. (2005) Modified electrode placement must be recorded
when performing 12-lead electrocardiograms. Postgraduate Medical Journal. 81(952):122-5
Mason RE, Likar I. (1966) A new system of multiple-lead exercise electrocardiography. Am Heart J
71:196.
Pahlm O, Wagner GS. (2008) Potential solutions for providing standard electrocardiogram recordings
from nonstandard recording sites. Journal of Electrocardiology. 41(3):207-10.
Pelter MM, Drew BJ (2015) Harm from alarm fatigue, Agency for Healthcare Research and Quality,
available at https://psnet.ahrq.gov/webmm/case/362/harm-from-alarm-fatigue (accessed
06/12/2017)
Pickham D, Drew BJ (2008) QT/QTc interval monitoring in the emergency department. Journal of
Emergency Nursing. 34(5):428-34.
Pitcher D, Nolan J (2015) Peri-arrest arrhythmias. Resuscitation Council UK, London. http://tinyurl.
com/ogeh2jt (accessed 30/11/2017)
Rautaharju PM, Surawicz B & Gettes LS (2009) AHA/ACCF/HRS Recommendations for the
Standardization and Interpretation of the Electrocardiogram. Part IV: The ST Segment, T and U
Waves, and the QT Interval. A Scientific Statement from the American Heart Association
Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American
College of Cardiology Foundation; and the Heart Rhythm Society, Circulation, 119, e241-e250.
Resuscitation Council (UK) (2016) Advanced Life Support. Seventh edition. RCUK, London.
Rinta-Kiikka I, Tuohinen S, Ryymin P, Kosonen P, Huhtala H, Gorgels A, Bayes de Luna A, Nikus K
(2014) Correlation of Electrocardiogram and Regional Cardiac Magnetic Resonance Imaging Findings
in ST-Elevation Myocardial Infarction: A Literature Review, Ann Noninvasive Electrocardiol, 19(6),
509523.
Sangkachand P, Sarosario B, Funk M (2011) Continuous ST-segment monitoring: nurses’ attitudes,
practices, and quality of patient care. American Journal of Critical Care. 20(3):226-38.
14
Sandau KE, Funk M, Auerbach A et al (2017) Update to Practice Standards for Electrocardiographic
Monitoring in Hospital Settings: A Scientific Statement From the American Heart Association,
Circulation, 136:e273e344. DOI: 10.1161/CIR.0000000000000527
Sandau KE, Smith M (2009) Continuous ST-segment monitoring: 3 case studies in progressive care.
Crit Care Nurse. 29(5):18-30.
Sendelbach S, Funk M (2013) Alarm Fatigue; A Patient Safety Concern. AACN advanced critical care.
24(4):378-86.
Spratt G (2016) Three steps to reduce alarm fatigue and improve patient safety, AARC Times, August
2016, 13-16.
Tortora GJ & Nielsen MT (2014) Principles of human anatomy, 13th edition, Hoboken, NJ : Wiley.
Yaldren J & Richley D (2014) Accurate measurement and assessment of the QT interval, British
Journal of Cardiac Nursing, 9(3), 137-141.
Yaldren J & Richley D (2014) Accurate measurement and assessment of the QT interval, British
Journal of Cardiac Nursing, 9(3), 137-141.
Zhou SH, Helfenbein ED, Lindauer JM, Gregg RE, Feild DQ.(2009) Philips QT interval measurement
algorithms for diagnostic, ambulatory, and patient monitoring ECG applications. Ann Noninvasive
Electrocardiol. 14(suppl):S3S8
... Αν και οι επικίνδυνες για τη ζωή αρρυθμίες είναι συχνές σε ασθενείς με οξέα στεφανιαία σύνδρομα, εντούτοις, μια μεγάλη ποικιλία φαρμάκων όπως αντιαρρυθμικά, αντιβιοτικά, αντιεμετικά, αντιψυχωτικά, αντικαταθλιπτικά, μπορούν να προκαλέσουν καθυστέρηση της επαναπόλωσης στα κύτταρα του μυοκαρδίου και να εμφανιστεί στο ΗΚΓ ως παράταση του διαστήματος QT, με αυξημένο κίνδυνο πρόκλησης πολύμορφης κοιλιακής ταχυκαρδίας, γνωστή και ως torsades de pointes (TdP), καθώς ακόμη και αιφνίδιο καρδιακό θάνατο. [4][5][6] Με αφορμή την πρόσφατη δημοσιευμένη βιβλιογραφία για τους καρδιολογικούς κινδύνους που ελλοχεύει η χορήγηση των πρώτων φαρμάκων που προτάθηκαν για τη θεραπεία της νόσου COVID-19, αναδεικνύεται ξανά η σημασία της ΗΚΓ παρακολούθησης κατά την έναρξη και τη συνέχιση της θεραπείας με ορισμένα φάρμακα. 7-9 Ο συνδυασμός των φαρμάκων που προτάθηκε αρχικά (υδροξυχλωροκίνη και αζυθρομυκίνη), μπορούσαν δυνητικά να προκαλέσουν παράταση του διαστήματος QT και ως αποτέλεσμα την εμφάνιση TdP. 8 Αν θεωρηθεί ότι η αξιολόγηση του πόνου αποτελεί επιπρόσθετο μέρος των ζωτικών σημείων για τη νοσηλευτική, 1 η μέτρηση του διαστήματος QT στο ΗΓΚ μπορεί να θεωρηθεί ως ένα επιπλέον ζωτικό σημείο, αφού η έγκαιρη ανίχνευση μιας παράτασης, αποτελεί υψίστης σημασίας γεγονός για την έναρξη ή και συνέχιση της φαρμακευτικής αγωγής, βελτιώνοντας έτσι την ασφάλεια του ασθενούς. ...
... Συστήνεται η παρακολούθηση του διαστήματος QTc να παρακολουθείται κάθε 8 ώρες, ενώ σε κλινικά ασταθείς ασθενείς ή σε ασθενείς που ήδη έχουν παράταση του QTc, η παρακολούθηση να γίνεται πιο συχνά. 6 Οποιαδήποτε παράταση του QTc, μετά την έναρξη της φαρμακευτικής αγωγής, πέραν των 60 msec από την αρχική μέτρηση, θα πρέπει να αξιολογείται το όφελος συνέχισης της θεραπείας. 8,14 Η γνώση όμως ότι ο κίνδυνος εμφάνισης TdP είναι μεγαλύτερος όταν το διάστημα QTc > 500 ms και η υιοθέτηση μεθόδων παρακολούθησης του QTc, μπο-Πίνακας 2. Φάρμακα που είναι γνωστό ότι προκαλούν παράταση του QT διαστήματος και χρησιμοποιούνται συχνά στην κλινική πρακτική. ...
Article
Full-text available
Monitoring the clinical condition of the patients and identifying signs of deterioration is one of the traditional roles of nurses. Many of drugs that that are administered everyday by nurses such as macrolides, fluoroquinolones, Imipenem/ cilastatin, citalopram, haloperidol and Piperacillin/Tazobactam are related with QT prolongation on the electrocardiogram, which cause an increased risk of life-threatening arrhythmias. In light of recent reports on the cardiac risks involved in administering the first drugs proposed for the treatment of COVID-19, nurses can play an important role in measuring the QT interval in each shift. Nurses should be familiar with QT interval measurement and monitoring methods, especially in patients with other risk factors such as age over 68, cardiology history, potassium and magnesium electrolyte disturbances. Calculating and monitoring the QT interval as another vital sign is crucial for patient safety and can help reduce the risk of life-threatening arrhythmias
... By identifying these specific ECG changes, healthcare providers can initiate appropriate treatment strategies to correct the hypercalcemia and prevent potential cardiac complications. Regular monitoring of the ECG in patients with hypercalcemia is essential to assess the effectiveness of treatment and ensure the restoration of normal cardiac function [18]. ...
Article
Electrolyte imbalances are a common occurrence in patients with various medical conditions, and their impact on cardiovascular health cannot be underestimated. The electrocardiogram (ECG) is a valuable tool for monitoring cardiac function, and changes in its morphology can provide important insights into electrolyte abnormalities. Electrolyte abnormalities can be life-threatening, and their management depends on the speed of onset of the electrolyte's derangement and the patient's existing comorbidities. Electrolyte imbalances can affect the heart's electrical activity and can be seen on an ECG. The specific ECG changes will depend on the specific electrolyte imbalance. This article explores the relationship between electrolyte imbalances and morphological alterations in the ECG. By understanding these changes, healthcare professionals can improve the diagnosis, treatment, and management of patients with electrolyte disturbances, ultimately enhancing patient outcomes and reducing the risk of cardiac events. Background One of the ways to assess the effects of electrolyte imbalances on the heart is through electrocardiogram (ECG) morphology changes. An ECG records the heart's electrical activity and can detect abnormalities in its rhythm and conduction system. Electrolyte imbalances can cause distinct changes in the ECG, which can serve as important diagnostic clues. A common electrolyte abnormality that affects the ECG is hyperkalemia, which refers to elevated levels of potassium in the blood. In patients with hyperkalemia, the ECG typically shows tall, peaked T-waves, prolonged PR intervals, and widened QRS complexes. These abnormalities reflect a delay in ventricular repolarization and can lead to potentially life-threatening arrhythmias, such as ventricular fibrillation.
Chapter
With the number of surgical procedures requiring sedation increasing every year, most hospitals and other facilities now have a procedural (moderate/deep) sedation program. The third edition of this popular handbook provides concise, practical, and evidence-based guidance on safe and effective procedural sedation. Featuring contributions from national experts, chapters cover the description/definition of sedation levels, patient evaluation, pharmacology, legal and quality assurance issues, as well as sedation for specific populations (ambulatory/office settings, elderly, pediatric, ICU, emergency room, endoscopy, reproductive technologies). The book also reviews the specific clinical and administrative considerations for the nursing and PA staff often involved in administering sedation. Comprehensively updated to incorporate the most current, evidence-based information including updates to existing guidelines, patient outcomes data from the most recently published papers and expanded sedation-related content in sub-specialties. An essential manual for a wide array of healthcare providers to develop safer techniques, policies, and procedures for moderate and sedation.
Preprint
Full-text available
This research paper presents a comprehensive investigation into the acquisition, processing, and transmission of Electrocardiogram (ECG) data. The study comprises two distinct but interrelated sections. The first part focuses on the intricacies of ECG data collection, utilizing an Arduino UNO micro controller and a heart rate sensor (AD8232) connected to MATLAB on laptop. The collected data undergoes digital signal processing in MATLAB, where various filters are used to cancel out noise and drift present in the data and to extract useful information like heartbeat rate of the patient. The second part of the research delves into the domain of data transmission, encompassing Phase Modulation (PM) and Quadrature Phase Shift Keying (QPSK) modulation techniques. Transmission and reception are facilitated through simulated Additive White Gaussian Noise (AWGN) channel and ADALM-PLUTO Software-Defined Radio (SDR) technology, allowing for efficient communication and analysis.
Article
Objective Chest pain is common, and current risk-stratification methods, requiring 12-lead electrocardiograms (ECGs) and serial biomarker assays, are static and restricted to highly resourced settings. Our objective was to predict myocardial injury using continuous single-lead ECG waveforms similar to those obtained from wearable devices and to evaluate the potential of transfer learning from labeled 12-lead ECGs to improve these predictions. Methods We studied 10 874 Emergency Department (ED) patients who received continuous ECG monitoring and troponin testing from 2020 to 2021. We defined myocardial injury as newly elevated troponin in patients with chest pain or shortness of breath. We developed deep learning models of myocardial injury using continuous lead II ECG from bedside monitors as well as conventional 12-lead ECGs from triage. We pretrained single-lead models on a pre-existing corpus of labeled 12-lead ECGs. We compared model predictions to those of ED physicians. Results A transfer learning strategy, whereby models for continuous single-lead ECGs were first pretrained on 12-lead ECGs from a separate cohort, predicted myocardial injury as accurately as models using patients’ own 12-lead ECGs: area under the receiver operating characteristic curve 0.760 (95% confidence interval [CI], 0.721–0.799) and area under the precision–recall curve 0.321 (95% CI, 0.251–0.397). Models demonstrated a high negative predictive value for myocardial injury among patients with chest pain or shortness of breath, exceeding the predictive performance of ED physicians, while attending to known stigmata of myocardial injury. Conclusions Deep learning models pretrained on labeled 12-lead ECGs can predict myocardial injury from noisy, continuous monitor data early in a patient’s presentation. The utility of continuous single-lead ECG in the risk stratification of chest pain has implications for wearable devices and preclinical settings, where external validation of the approach is needed.
Article
Full-text available
Purpose: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing "alarm fatigue" which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO(2), and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. Results: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. Discussion: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue.
Article
Full-text available
Background Patients with acute ST-elevation myocardial infarction (STEMI) benefit substantially from emergent coronary reperfusion. The principal mechanism is to open the occluded coronary artery to minimize myocardial injury. Thus the size of the area at risk is a critical determinant of the patient outcome, although other factors, such as reperfusion injury, have major impact on the final infarct size. Acute coronary occlusion almost immediately induces metabolic changes within the myocardium, which can be assessed with both the electrocardiogram (ECG) and cardiac magnetic resonance (CMR) imaging.Methods The 12-lead ECG is the principal diagnostic method to detect and risk-stratify acute STEMI. However, to achieve a correct diagnosis, it is paramount to compare different ECG parameters with golden standards in imaging, such as CMR. In this review, we discuss aspects of ECG and CMR in the assessment of acute regional ischemic changes in the myocardium using the 17 segment model of the left ventricle presented by American Heart Association (AHA), and their relation to coronary artery anatomy.ResultsUsing the 17 segment model of AHA, the segments 12 and 16 remain controversial. There is an important overlap in myocardial blood supply at the antero-lateral region between LAD and LCx territories concerning these two segments.Conclusion No all-encompassing correlation can be found between ECG and CMR findings in acute ischemia with respect to coronary anatomy.
Article
Attaching a patient to a cardiac monitor and obtaining a clear electrocardiogram (ECG) trace may now be considered basic nursing skills. In line with the UK professional standards and code of conduct, healthcare practitioners are required to practise effectively and preserve patients' safety. Therefore, healthcare practitioners undertaking cardiac monitoring are required to have a basic understanding of normal sinus rhythms and some of the common types of cardiac arrhythmia. This will enable prompt recognition of early warning signs of potential and actual clinical conditions, and the timely initiation of treatment. This article reviews the clinical skill of attaching a patient to a three and five-lead cardiac monitor, discussing appropriate skin preparation and lead selection. It also outlines the identification of several of the common types of cardiac arrhythmia on an ECG rhythm strip using a systematic approach.
Article
Introduction Prolongation of the QT interval is a well-recognized complication associated with many commonly used medications. Emergency Department monitoring of the corrected QT (QTc) both before and after medication administration is typically performed using the 12‑lead electrocardiogram (ECG). The purpose of this study is to compare the QTc reported on the 12‑lead ECG to that reported by single brand of bedside monitor. Methods A convenience sample of emergency department patients over the age of 18 undergoing bedside monitoring and who had an ECG ordered by their treating physician were enrolled. These patients underwent simultaneous ECG and monitor QTc calculation. The primary outcome of interest was the correlation between the monitor and ECG QTc. Secondary outcomes included ability of each method to identify patients with a QTc > 500 ms and the ability of each method to identify patients with a QTc < 450 ms. Results A total of 125 patients had simultaneous ECG and monitor QTc measurements recorded. There was moderate correlation between the monitor and ECG QTc (Pearson's correlation coefficient = 0.55). The median difference between the ECG QTc and the monitor QTc (ECG QTc minus monitor QTc) was − 7 ms (IQR − 23 to 11 ms). Conclusion We found that there was moderate correlation between the QTc reported on the 12 lead ECG and that reported by the bedside monitor. This correlation is not strong enough to support the use of the bedside monitor as a substitute for the 12‑lead ECG when evaluating a patient's QTc.
Article
Background and purpose: This scientific statement provides an interprofessional, comprehensive review of evidence and recommendations for indications, duration, and implementation of continuous electro cardiographic monitoring of hospitalized patients. Since the original practice standards were published in 2004, new issues have emerged that need to be addressed: overuse of arrhythmia monitoring among a variety of patient populations, appropriate use of ischemia and QT-interval monitoring among select populations, alarm management, and documentation in electronic health records. Methods: Authors were commissioned by the American Heart Association and included experts from general cardiology, electrophysiology (adult and pediatric), and interventional cardiology, as well as a hospitalist and experts in alarm management. Strict adherence to the American Heart Association conflict of interest policy was maintained throughout the consensus process. Authors were assigned topics relevant to their areas of expertise, reviewed the literature with an emphasis on publications since the prior practice standards, and drafted recommendations on indications and duration for electrocardiographic monitoring in accordance with the American Heart Association Level of Evidence grading algorithm that was in place at the time of commissioning. Results: The comprehensive document is grouped into 5 sections: (1) Overview of Arrhythmia, Ischemia, and QTc Monitoring; (2) Recommendations for Indication and Duration of Electrocardiographic Monitoring presented by patient population; (3) Organizational Aspects: Alarm Management, Education of Staff, and Documentation; (4) Implementation of Practice Standards; and (5) Call for Research. Conclusions: Many of the recommendations are based on limited data, so authors conclude with specific questions for further research.
Article
QT measurement should be included as part of the systematic interpretation of the electrocardiogram (ECG). Accurate measurement and interpretation of the QT interval will help to identify potentially fatal long QT syndrome (LQTS). The assessment of the QT interval, correction of the QT and the guidelines relating to it are complex. This article discusses the importance of this element of the ECG and offers practical advice on QT assessment based on a review of the relevant literature.