ArticlePDF Available

A Swab Limit-Derived Scale For Assessing The Detectability Of Total Organic Carbon Analysis

Authors:

Abstract and Figures

This article discusses how the detection limit for analytical methods can be combined with cleaning validation swab limits to create a detectability scale similar to that described in our earlier article on visual inspection. This new detectability scale can assist in determining whether an analytical method is acceptable for use in a cleaning validation or verification. Combined with the HBELderived toxicity scale and Cpu (process capability)-derived probability scale, it can also provide for a total measure of risk in cleaning.
Content may be subject to copyright.
A Swab Limit-Derived Scale For Assessing The Detectability Of Total Organic Carbon
Analysis
By Andrew Walsh; Thomas Altmann; Alfredo Canhoto, Ph.D.; Ester Lovsin Barle, Ph.D.; David G. Dolan, Ph.D.; Andreas Flueckiger, M.D.; Igor Gorsky;
Robert Kowal; Mariann Neverovitch; Mohammad Ovais; Osamu Shirokizawa; and Kelly Waldron
Part of the Cleaning Validation for the 21st Century series
This article will discuss how the detection limit for analytical methods can be combined with
cleaning validation swab limits to create a detectability scale similar to that described in our article
on visual inspection. This new detectability scale can assist in determining whether an analytical
method is acceptable for use in a cleaning validation or verification. Combined with the HBEL-
derived toxicity scale and Cpu (process capability)-derived probability scale, it can also provide
for a total measure of risk in cleaning.
Note: This article uses the term health-based exposure limit (HBEL), which is synonymous with
the terms acceptable daily exposure (ADE) and permitted daily exposure (PDE).
Selection Of Analytical Methods In Cleaning
Analytical methods typically used in cleaning validation fall into the two broad categories of specific methods and nonspecific methods, and the decision
for using one or the other should be science-based and risk-based. Figure 1 presents a hierarchy for selecting analytical methods in reference to the
HBEL-derived toxicity scale. For low-risk situations, visual inspection may be the only method needed, supported by nonspecific methods or by specific
methods as necessitated by the increasing hazard level. As the level of the hazard increases, the rigor required of the analytical method should also
increase. However, as indicated by the question marks, the transitions from using only a visual inspection to needing a nonspecific analysis such as total
organic carbon (TOC) and then to needing a specific analysis are not obvious. The use of the scale discussed in this article may provide a tool to help
resolve these questions for analytical methods as was shown for visual inspection. This article is focused on TOC as an example for nonspecific methods,
as compounds containing organic carbon are the most common; however, this scale could be applied to other analytical methods.
Figure 1 - Risk hierarchy of analytical methods (Note: Toxicity scale is based on –log(HBEL) where HBEL is the acceptable daily exposure in
grams)
Determination Of Analytical Detection Limits
Detection limits (DLs) and how they are determined are fundamental to this discussion. It is fairly well known that DLs for HPLC are normally
determined by evaluating the signal-to-noise ratio. As stated in ICH Q2(R1):
"Determination of the signal-to-noise ratio is performed by comparing measured signals from samples with known low concentrations of analyte
with those of blank samples and establishing the minimum concentration at which the analyte can be reliably detected. A signal-to-noise ratio
between 3 or 2:1 is generally considered acceptable for estimating the detection limit."
Guest Column | January 24, 2018
1
2 3
1
2
1
4
For methods where there is no specific background noise to measure, such as TOC, other techniques may be employed, such as the standard deviation of
the blank. ICH Q2 states:
Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and
calculating the standard deviation of these responses.
Similarly, a multiple of 3 is applied to the standard deviation of the blank and set as the DL. For example, for a blank with a mean of 100 and a standard
deviation of 30, the DL would be set at 190 (100 + 30 x 3 = 190). This type of approach has been used for TOC.
Detection Limits For TOC
Acceptance of TOC for use in cleaning validation has grown over the past 20 years, with a number of articles being published on its application to APIs
and cleaning agents. Of the 15 articles cited, only eight addressed the DL for TOC in one way or another. Since the methods for calculating DLs are a
matter of debate among analytical chemists, this may account for the lack of information on DLs for TOC in past articles. However, for the purposes of
this article, the DL of TOC is very important.
An early article on using TOC for cleaning validation by R. Baffi, et. al. examined its use for biologic compounds. The authors mention in their abstract
that "...a limit of detection of approximately 0.1 ppm," but the text offered no details on how this DL was derived.
Gavlik, et. al. in 1995 published an article on the potential use of TOC for cleaning agents, but focused on recovery and did not report any other method
parameters.
In 1996, Jenkins, et. al. published a comprehensive review of swab and rinse recoveries for a variety of swab and filter materials and briefly discussed
the DL. In their article, DL was defined as "...the absolute value of the intercept plus three times the estimated standard deviation." The authors reported
DLs that ranged from 1 ppm to 14 ppm.
Strege, et. al. in 1996 discussed the DL and stated, "A limit of detection and limit of quantitation were established at 9.2 µg/swab and 12.1 µg/swab." The
authors did not provide details on how the DL was arrived at but wrote that "...a set of 10 swab blanks were prepared and analyzed." Data or calculations
for the DL were not provided, but the authors included a glossary from the USP XXII that mentioned "...analyzing a number of blank samples and
calculating the standard deviation of this response. The standard deviation multiplied by a factor, usually 3, provides an estimate of the limit of
detection." Since the rest of this article will present DLs in ppb (parts per billion), these results need to be converted. Based on their description of the
handling of other swab samples, it appears that the swab dilution volume was 5 mL, and this would translate to a DL of 1,840 ppb (9.2 µg/5 mL = 1.84
µg/mL = 1,840 ppb).
Holmes, Alison J. and A. J. Vanderwielen in 1997 reported using TOC for analysis of aspirin residues on several materials of construction and reported
DLs of 3 to 15 ppm. These investigators included swabbing an unspiked coupon surface as part of the swab blank, which other investigators have not
mentioned.
Guazzaroni, et. al. in 1998 discussed the use of TOC for a number of compounds (cleaning agents, endotoxin, biologic media, and PEG) and reported
the DL as 50 ppb "…as per the manufacturer's specification," but stated that the TOC background, including the swab and filter material, was about 2.5
ppm.
Kirsch in 1998 discussed the parameters important to the validation of methods used for cleaning and mentioned the applicability of TOC. Kirsch stated
that the DL is "… most practically defined as approximately three times the standard deviation of the baseline noise level around the analyte peak." While
this is applicable to HPLC and some TOC analyzers, it is not applicable to all TOC analyzers.
In 2000, Karen Clark analyzed "swab blanks" as a means to calculate a DL for TOC. In her study, a swab blank is defined as a vial containing low TOC
water (<25 ppb) along with the head of one swab. Four replicate analyses were performed on each swab blank and the mean and standard deviation were
calculated. Using a Student t-test analysis of 10 swab blanks, Clark found the DL for TOC to be 50 ppb.
Table 1 - Calculated TOC Averages from 10 Blank Vials
5-19
20
5
6
7
8
9
10
11
12 5
12
In 2004, Wallace, et. al. of Teledyne Instruments published a brief review of some factors to consider for implementing TOC, such as detergent
selection and acceptance criteria, and discussed the choice of TOC technologies, but did not discuss analytical method parameters. They did provide a
table comparing the two major technologies used to oxidize the carbon in the sample to CO : high temperature combustion (HTC) and UV/persulfate
(UV/P). This table contained data on the reagent water used as a blank. From this data we can determine that the DL for HTC in this study is 51 ppb (12
ppb + 13 ppb x 3) and UV/P is 10 ppb (7 ppb + 1 ppb x 3). Both of these values are lower than they should be, as these were not "swab blanks" and did not
contain any swab material. It is well known that swab material can contribute significant carbon background to the swab blank.
In 2006, Chris Glover performed a study of TOC using albumin and included the DL. Glover used a different approach by evaluating the accuracy data.
The DL was designated as the lowest albumin weight with acceptable accuracy results (no less than 50 percent recovery). Glover set the DL at 49 ppb, but
the water blank was subtracted from this value. Glover provided a table containing the raw data for 15 swab blanks. From this data, a DL can be
calculated as above to be 237 ppb (186 ppb + 17 ppb x 3).
Nieves and Strege reported a study of the development of a test method for polysorbate 20, which was being used as a cleaning agent for vial closures.
These authors used the ICH Q2 3sigma/slope approach for calculating the DL. Their calculation yielded a DL value of 660 ppb for this study.
In their article on using TOC for cleaning validation of nutraceuticals, Frey, et. al. mentioned the importance of method parameters including DLs, but
did not provide any values obtained.
Bader, et. al., in their study of the use of online TOC, state that the instruments examined met the "instrumental limit of detection of 50 ppb TOC
required by USP," but did not state how that was determined or what the actual results were.
In 2012 Clifford and Tanaka published a study on six water soluble and water-insoluble compounds and compared the results for recovery by rinse
sampling, swab sampling with a water extraction, and swab sampling using direct combustion. No analytical method parameters were discussed other
than recovery.
Most recently, Xue Li, et. al., in a study on cleaning agents, reported a QL of 114 ppb based on the linearity data and defined the DL as QL/3, or 38 ppb.
Table 2 summarizes the DLs reported in the literature or calculated from the data provided in the articles.
Table 2 - DLs Reported or Calculated from the Literature
While there were significant differences in the methods used to determine the DLs in these articles, it is more important to note the wide disparity in the
DL values reported/calculated, which range over three orders of magnitude. As described in the visual inspection article, if the DL of TOC is known,
then it can be compared to the TOC limit for a compound to justify the use of TOC for that compound. It should be immediately obvious that the higher
the DL of TOC, the harder it will be to justify its use for compounds with lower limits. Clearly, obtaining a low DL is a very important task for the analyst
developing the method, and this is something that the analyst should be aware of and address during the TOC method development.
Impact Of 1/1,000th Dose And 10 ppm Limits On The Use Of TOC
As described in the visual inspection article, to demonstrate the undesirable impact that retaining the 1/1,000 or 10 ppm limits would have on the use
of TOC, swab limits were calculated for 304 drug compounds based on their HBELs and their corresponding 1/1,000th or 10 ppm limits using the
assumptions in Table 3.
Table 3 - Parameter Assumptions for TOC Limit Calculations
13
2
14
15
16
17
18
19
1
1 th
21
The data obtained was plotted using R statistical software on a log scale in order to visualize it all on one graph (Figure 2).
Figure 2: Comparison of HBEL, 1/1,000th and 10 ppm to TOC Detection Limits (100 ppb)
A line has been drawn in Figure 2 at a 100-ppb level (0.1 pbm) as the TOC DL for an example. Many of these compounds cannot be considered for
evaluation by TOC as the TOC swab limits based on the 1/1,000th or 10 ppm are below this 100 ppb DL or too close to it to be justified as a safe method
of analysis. If some of the DLs in Table 2 were shown, even fewer compounds could be justified.
However, if only the HBELs are used to calculate the TOC swab limits, many of these compounds could easily be considered for evaluation by TOC
(Figure 3). It would seem that the compounds on the right side of the curve in the rectangle could easily be justified for evaluation by TOC. But as we
move to the left and the TOC swab limits get closer and closer to the 100 ppb DL it would seem harder to justify using TOC. To appropriately evaluate the
acceptable use of TOC, it would be helpful to have some way of judging how close the TOC swab limit is to the TOC detection limit.
Figure 3: Drugs where HBEL can meet TOC detection limits (100 ppb)
Using The Detection Limit Of TOC As A Measure Of Detectability
In the same way as with visual inspection, we are trying to measure how close the TOC swab limit is to the detection limit of TOC so we can make a
decision on whether we can use TOC. Again, a simple method to measure the relative distance is to look at the ratio of the two values. If we then take the
log of this ratio we can obtain a logarithmic scale that equals zero when the values of the TOC swab limit and detection limit of TOC are equal and
becomes negative when the detection limit of TOC is lower than the TOC swab limit and becomes positive when it is higher. This calculation can provide
us with a carbon detection index that can be applied in all manufacturing cleaning situations (Equation 1).
where:
CDI = Carbon Detection Index
DL = TOC Detection Limit
TOC
SL = TOC Swab Limit
TOC detection limits can also be converted into a scale by simply taking the logarithm of the ratio of TOC detection limit to the TOC swab limit as derived
as described above in Equation 1 (Table 4).
Table 4 - Detection Limit-based Scales for Detectability of Residues by TOC
In this example, any CDIs above zero are unacceptable and CDIs below -1.0 are acceptable. As can be seen comparing the three columns, as the DL
increases, fewer and fewer swab limits can be met. Each company can select how close to a CDI of zero it is comfortable with. For example, one company
may require its CDIs to be < -2.0, or at least 2 logarithms below the zero point.
Discussion
As stated in the article on visual inspection, the selection of methods for assessing cleaning should be science-based and risk-based. Key considerations
for the risk assessment may include the hazard or risk of the process residue to be analyzed (toxicity score), level of detection required, applicability of
existing methods, other quality and compliance risks, as well as risks to the business such as difficulty of implementation and the possible long-term
maintenance of the method for ongoing monitoring programs. A reasoned and logical approach needs to be taken, as some methods may be
unnecessarily expensive or difficult to implement for the process residues under consideration. Conversely, a simple inexpensive method may not be
appropriate for all process residues. In general, the simplest techniques should be examined first and used if determined to be appropriate through an
assessment based on science and risk. Ultimately, the goal should be to use the simplest technique that is appropriate and can be justified. TOC has
proved to be one of the easiest analytical methods to implement, and is becoming a method of choice, for cleaning validation.
The scale reveals the two aspects of method development for cleaning that work in concert with each other. One is the HBEL, which drives the swab or
rinse limits, which must be determined judiciously. Undue conservatism in calculating the HBEL through the excessive application of adjustment factors
can easily result in swab or rinse limits that are so low as to be unachievable, which is in conflict with the longstanding guidance on cleaning from the
FDA that limits should be "practical, achievable and verifiable." This can lead to excessive cleaning efforts or unnecessary dedication, which conflicts
with not only business goals but the intended purposes of Risk-MaPP. The second aspect, the method DL, decides whether the method can be used
based on the first aspect. Inattention to the DL when developing a TOC method can lead to high DLs (see Table 2) and result in TOC being restricted
from use with many compounds. Conversely, lower DLs would allow the TOC method to be used more widely. If the TOC limits are set too low using
arbitrary or non-health-based limits (as discussed above), then, again, TOC could not be justified. Regulators should ask to see scientifically justified
swab limits (i.e., based on the HBEL), along with the corresponding DL when TOC or any other analytical methods are used for cleaning validation.
This article is intended to specifically address the use of TOC, but this scale is appropriate for any analytical method being developed for swab (or rinse)
sample testing and can be applied to large biological molecules as well as small molecules. The principle simply informs the user whether a method for a
given compound can be considered acceptable based on their swab (or rinse) limit. If the HBEL is very low ("Green Zone"), the corresponding swab limit
will be very low, too, and will probably surpass the method's DL, and the scale will give a measure of how good that is. The user can simply clean and
measure how well they have achieved that using the process capability scale. If they are in the "Red Zone" and the DL is equal to or above the swab limit,
they cannot detect the compound at a level that assures meeting the HBEL-based limit. Such methods should not be considered completely useless; they
can still be used to demonstrate that residues have been removed close to the limit. However, the user would need to pursue additional steps to provide
assurance that the residues are at safe levels, such as demonstration of inactivation, degradation, or decontamination.
It should be obvious that the DL for TOC is very important, and one of the main goals in swab method development should be to reduce the DL as much
as possible. Previously, the limits on the applicability of TOC have been unclear, and this tool may be helpful in such assessments. Prior to this, careful
consideration should be given to how DLs are experimentally determined and a standard procedure for determining the DL for TOC should probably be
established.
A subsequent article will discuss how these new detectability scales for TOC (or any other analytical method) and for visual inspection can be used in
conjunction with the HBEL-derived toxicity scale and the Cpu-derived probability scale as tools to evaluate the level of risk in cleaning.
Peer Review
TOC
3
22
23
24
3
1 2
The authors wish to thank our peer reviewers, Bharat Agrawal; Sarra Boujelben; Gabriela Cruz, Ph.D.; Parth Desai; Ioanna-Maria Gerostathi; Jessica
Graham, Ph.D., DABT; Miguel Romero Obon; Laurence O'Leary; John Leahy; and Ersa Yuliza, for reviewing this article and for their insightful
comments and helpful suggestions.
References
1. Walsh, Andrew, Thomas Altmann, Alfredo Canhoto, Ester Lovsin Barle, David G. Dolan, Mariann Neverovitch, Mohammad Ovais, Osamu
Shirokizawa, and Kelly Waldron. "An MSSR-derived Scale for Assessing the Detectability of Compound-Carryover in Shared Facilities,"
Pharmaceuticalonline.com, December 2017
2. Walsh, Andrew, Ester Lovsin Barle, Michel Crevoisier, David G. Dolan, Andreas Flueckiger, Mohammad Ovais, Osamu Shirokizawa, and Kelly
Waldron. "An ADE-Derived Scale For Assessing Product Cross-Contamination Risk In Shared Facilities," Pharmaceuticalonline.com, May 2017
3. Walsh, Andrew, Ester Lovsin Barle, David G. Dolan, Andreas Flueckiger, Igor Gorsky, Robert Kowal, Mohammad Ovais, Osamu Shirokizawa, and
Kelly Waldron. "A Process Capability-Derived Scale For Assessing Product Cross-Contamination Risk In Shared Facilities,"
Pharmaceuticalonline.com, August 2017
4. ICH Q2 (R1): Validation of Analytical Procedures: Text and Methodology International Conference on Harmonisation of Technical Requirements
for Registration of Pharmaceuticals for Human Use [ICH], London, 2005.
5. Baffi, R. G. Dolch, R. Garnicka, Y. F. Huang, B. Mar, D. Matsuhiro, B. Niepelt, C. Parra, and M. Stephan "A Total Organic Carbon Analysis Method
for Validating Cleaning Between Products in Biopharmaceutieal Manufacturing," PDA J Pharm Sci and Tech 1991, 45 13-19
6. Gavlik, Walter K., Lane A. Ohlemeir and Herbert Kaiser Analytical Strategies for Cleaning Agent Residue Determination, Pharmaceutical
Technology March 1995
7. Jenkins, K.M., A.j. Vanderweilen, J.A Armstronmg, L.M. Leonard, G.P. Murphy and N.A Piros "Application of Total Organic Carbon Analysis to
Cleaning Validation," PDA Journal of Pharmaceutical Science and Technology Vol. 50, No. 1 January/February 1996
8. Strege, Mark A., Tery L Stinger, Brett T. Farrell and Avinash L. Lagu "Total Organic Carbon Analysis for Cleaning Validation of Bioprocess
Fermentation Equipment," Biopharm International April 1996
9. Holmes, Alison J. and A. J. Vanderwielen "Total Organic Carbon Method for Aspirin Cleaning Validation," PDA J Pharm Sci and Tech 1997, 51 149-
152
10. Guazzaroni, Maria, Boudin Yiin and Julia L. Yu "Application of total organic carbon analysis for cleaning validation in pharmaceutical
manufacturing," American Biotechnology Laboratory September 1998
11. Kirsch, Robert Validation of Analytical Methods Used in Pharmaceutical Cleaning Assessment and Validation, Pharmaceutical Technology
Analytical Validation 1998
12. Clark, Karen A., “Total Organic Carbon Analysis for Cleaning Validation in Pharmaceutical Manufacturing” October 2000
13. Wallace, Brian and Stephen Lawson "Detection Selection for TOC Analysis: Analytical Considerations for Cleaning Validation and Purified Water
Testing," Teledyne Tekmar White Paper, www.teledynetekmar.com
14. Glover, Chris "Validation of the Total Organic Carbon (TOC) Swab Sampling and Test Method," PDA Journal of Pharmaceutical Science and
Technology Vol. 60, No. 5, September-October 2006
15. Juarbe, Nieves and Mark Strege "Validation of a Method for the Determination of Polysorbate 20 Residues for the Support of the Cleaning of
Pharmaceutical Vuial Closures," Journal of Validation Technology February 2007, Vol. 13, No. 2
16. Frey, Benjamin, David Koczan, and Kenneth Shrout "Implementing a Compliant Cleaning Program in the Dietary Supplement Industry," Contract
Pharma/Nutraceuticals World Manufacturing Supplement, March 2007
17. Bader, Keith, John Hyde, Peter Watler and Amber Lane "Online Total Organic Carbon (TOC) as a Process Analytical Technology for Cleaning
Validation Risk Assessment" Pharmaceutical Engineering January/February 2009
18. Clifford, Robert and Minako Tanaka "Carbon Measurement Methods for Cleaning Validation: Comparing Direct Combustion with Rinse and Swab
Sampling Methods," Pharmaceutical Technology Europe August 2012
19. Xue Li, Imad A. Haidar Ahmad, James Tam, Yan Wang, Gina Dao and Andrei Blasko " Cleaning verification: A five parameter study of a Total
Organic Carbon method development and validation for the cleaning assessment of residual detergents in manufacturing equipment," Journal of
Pharmaceutical and Biomedical Analysis 149 (2018) 33–39
20. Vander Heyden, Yvan and Ricard Boqué “The Limit of Detection,” LCGC Europe 2009 Volume 22, Issue 2, pg 82–85
21. Walsh, Andrew, Michel Crevoisier, Ester Lovsin Barle, Andreas Flueckiger, David G. Dolan and Mohammad Ovais "Cleaning Limits—Why the 10-
ppm and 0.001-Dose Criteria Should be Abandoned, Part II," Pharmaceutical Technology 40 (8), 45-55 (2016).
22. FDA Guide to Inspections: Validation of Cleaning Processes, July 1993, U.S. Food and Drug Administration (FDA), www.fda.gov.
23. ISPE, Baseline® Pharmaceutical Engineering Guide: Risk-Based Manufacture of Pharmaceutical Products: A Guide to Managing Risks Associated
with Cross-Contamination. (ISPE, Tampa, FL, ed. First, 2010), vol. 7, pp. 186.
24. Walsh, Andrew "Cleaning Validation for the 21st Century: Can Alternative Approaches to the PDE-ADE be Justified?" Biopharm International
October 2015
... As both the HBEL-based toxicity scale for severity of hazard and the Cpu-based process capability scale for probability of exposure (occurrence) are not arbitrary values, they consequently have real significance. The toxicity and probability of exposure may be evaluated first, and then detectability can be considered for prioritization when the toxicity and 3,4 probability of exposure of two hazards are equal. Table 3 shows the toxicity and process capability scales side by side from the highest to the lowest possible values. ...
Article
Full-text available
This article discusses the concept and measurement of risk as it applies to the cleaning of pharmaceutical products. Four previous articles discussed how science-based data-derived scales could be created using compound HBELs (health-based exposure limits), from the process capability (Cpu) of the products’ cleaning processes and from the detection limits for visual inspection or for total organic carbon (TOC) analyses of these compounds. This article continues the discussion about the potential use and application of these new scales in cleaning failure modes and affects analysis (cleaning FMEA) to assist in measuring the risk of cleaning process failures as well as how these scales can be applied to develop a cleaning risk dashboard. The article will also discuss how these new scales can be utilized to accelerate new product introductions.
Article
Full-text available
This article will provide a detailed discussion of the science-, risk-, and statistics-based approaches in the American Society for Testing and Material (ASTM) E3106 "Standard Guide for Science Based and Risk Based Cleaning Process Development and Validation".
Article
Full-text available
The International Congress on Harmonization Quality Risk Management Guidance (ICH Q9) lists both cleaning (in Annex II.4) and validation (in Annex II.6) as potential areas for the application of quality risk management. This clearly implies that the ICH Q9 principle for adjusting the level of "effort, formality, and documentation" based on the level of risk could be applied to cleaning and its validation. Previous articles discussed how science-based and data-derived scales could be created from HBELs (health-based exposure limits), from the process capability (Cpu) of cleaning processes, from the detection limits for total organic carbon (TOC) analyses of these compounds, or from visual inspection. Another article discussed how these scales could be used to measure the level of risk in cleaning validation. This article builds on these discussions and shows how these HBEL-based and process capability-based scales can be combined into a matrix that provides a clear visual guide for adjusting the level of effort, formality, and documentation for cleaning validation based on the level of risk.
Article
Full-text available
This article is the first in a series of articles that will explore some of the typical issues that may be encountered during recovery studies and show how the use of statistical tools for assessing the recovery data can provide greater insight into the results and enable data-driven decisions concerning recovery studies. In this installment, we will look at three case studies: 1. Case Study No. 1 will examine how to establish a relationship between the level of compound spiked and the percent recovery through regression analysis. It will also show how regression analysis can be used to provide an indication of how the recovery method works at higher or lower concentrations than those examined in the study. 2. Case Study No. 2 will examine how the 2-sample t-Test might be used to evaluate changes to a swab method to see if the differences in recovery results are significant. 3. Case Study No. 3 will examine how the 2-sample t-Test might be used to evaluate differences in recovery results between two analysts performing swab sampling.
Article
Full-text available
This article reviews the European Medicines Agency (EMA) April 16, 2018 Questions and answers on implementation of risk-based prevention of cross-contamination in production and Guideline on setting health-based exposure limits for use in risk identification in the manufacture of different medicinal products in shared facilities. The article discusses the significance of these changes to cleaning validation programs and how they compare to the new ASTM E3106-18 Standard Guide for Science-Based and Risk-Based Cleaning Process Development and Validation and the forthcoming ASTM Standard Guide for the Derivation of Health-Based Exposure Limits (HBELs).
Article
Full-text available
This article discusses regulatory views on the use of VI as a sole criterion in cleaning validation, presents a case study on how inspectors can be qualified for VI, recommends the use of statistical techniques, and suggests how VI could be implemented as part of a control strategy in a cleaning validation program based on the level of risk.
Article
Full-text available
This article discusses how to derive a Toxicity Scale from the ADE (with scores form 1-10) and how the ADE of any compound can be directly converted into a score on this scale. This Toxicity Scale can be used to quickly assess the potential for the risk of cross contamination in manufacturing situations through an easy to use visual scale. This Toxicity Scale can be used as the "Severity" Score in FMEAs of Cleaning Processes.
Article
Full-text available
This article discusses how the 10ppm limit, along with the 0.001 dose limit, are not truly risk-based approaches and are also unsound from an operational standpoint as they have caused unnecessary difficulties for many companies. A statistical process-based approach to demonstrate that acceptable cleaning has been achieved compared to a health-based limit is proposed.
Article
Full-text available
Many pharmaceutical companies have adopted the permitted daily exposure (PDE), or acceptable daily exposure (ADE) to set cleaning validation limits. Some regulatory agencies now require its use; but some companies continue to argue against it, particularly biologics manufacturers. Critics of the approach argue that protein molecules are denatured or degraded by their cleaning procedures, and that alternate approaches should be used. This article explores whether alternative approaches to using the PDE or ADE for determining cleaning validation limits for biologics can be justified and, if not, what approaches could be used.
Article
A Total Organic Carbon (TOC) based analytical method to quantitate trace residues of clean-in-place (CIP) detergents CIP100(®) and CIP200(®) on the surfaces of pharmaceutical manufacturing equipment was developed and validated. Five factors affecting the development and validation of the method were identified: diluent composition, diluent volume, extraction method, location for TOC sample preparation, and oxidant flow rate. Key experimental parameters were optimized to minimize contamination and to improve the sensitivity, recovery, and reliability of the method. The optimized concentration of the phosphoric acid in the swabbing solution was 0.05M, and the optimal volume of the sample solution was 30mL. The swab extraction method was 1min sonication. The use of a clean room, as compared to an isolated lab environment, was not required for method validation. The method was demonstrated to be linear with a correlation coefficient (R) of 0.9999. The average recoveries from stainless steel surfaces at multiple spike levels were >90%. The repeatability and intermediate precision results were ≤5% across the 2.2-6.6ppm range (50-150% of the target maximum carry over, MACO, limit). The method was also shown to be sensitive with a detection limit (DL) of 38ppb and a quantitation limit (QL) of 114ppb. The method validation demonstrated that the developed method is suitable for its intended use. The methodology developed in this study is generally applicable to the cleaning verification of any organic detergents used for the cleaning of pharmaceutical manufacturing equipment made of electropolished stainless steel material.
Article
Validated cleaning procedures are needed to ensure the absence of contaminants from bioprocessing equipment, and these procedures must be supported by appropriate analytical methodology. This article describes the development of a quantitative total organic carbon (TOC) assay for residual carbon-containing materials on stainless steel surfaces using E. coli cells as a model substances.
Article
Sophisticated measurement and control strategies have been successfully applied to CIP systems and operations for many years. The utilization of online TOC measurement represents a significant step forward in assurance of product quality and safety through more effective real-time monitoring and control of the cleaning processes. With enhanced quality assurance and reduced cost of goods as driving forces, pharmaceutical manufacturers are automating manufacturing operations to accommodate more complex processes, including more complicated cleaning sequences commensurate with increasingly complex manufacturing equipment configurations and production methodologies. More robust automated systems will provide higher levels of assurance of removal of potential contaminants to acceptable levels. CIP systems can be automated to the point that risk from manual operator actions are eliminated from the process stream, except for manual set-up activities, such as the loading and un-loading of a glass-washer or the starting of a unit operation from a control point.
Article
The concept of the limit of detection (LOD) has been, and still is, one of the most controversial in analytical chemistry. The multiple definitions and calculation methods proposed have contributed to this situation. Although in the last years, several international organizations, such as ISO or IUPAC, have tried to reach a consensus in their definitions and have issued guidelines for the estimation of this important parameter in chemical analysis, the subject is still a matter of scientific debate. In this article, we try to clarify the definition and provide guidelines to estimate LOD in chromatographic methods of analysis.
Article
Validation of cleaning processes following production of pharmaceutical products has received much industry and regulatory attention. Because companies are increasingly using multipurpose process equipment and automated clean-in-place procedures, it has become even more important to establish evidence that the product is not contaminated. This article discusses requirements for validating a chromatographic procedure for assessing cleaning efficiency and for cleaning validation purposes. Recovery experiments are particularly important because firms must demonstrate removal of analytes at low levels from potentially active stainless steel and glass surfaces as well as from polyester swabs to ensure the accurate evaluation of the cleaning process.