Recent advances in ubiquitous sensing technologies have exploited various approaches to monitoring vital signs. One of the vital signs is human respiration which typically requires reliable monitoring with low error rate in practice. Previous works in respiration monitoring however either incur high cost or suffer from poor error rate. In this paper, we propose a correlation based ranging method C-FMCW which is able to achieve high ranging resolution. Based on C-FMCW, we present the design and implementation of an audio-based highly-accurate system for human respiration monitoring, leveraging on commodity speaker and microphone widely available in home environments. The basic idea behind the audio-based method is that when a user is close to a pair of speaker and microphone, body movement during respiration causes periodic audio signal changes, which can be extracted to obtain the respiration rate. However, several technical challenges exist when applying C-FMCW to detect respiration with commodity acoustic devices. First, the sampling frequency offset between speakers and microphones if not being corrected properly would cause high ranging errors. Second, the uncertain starting time difference between the speaker and microphone varies over time. Moreover, due to multipath effect, weak periodic components due to respiration can easily be overwhelmed by strong static components in practice. To address those challenges, we 1) propose an algorithm to compensate dynamically acoustic signal and counteract the offset between speaker and microphone; 2) co-locate speaker and microphone and use the received signal without reflection (self-interference) as a reference to eliminate the starting time difference; and 3) leverage the periodicity of respiration to extract weak periodic components with autocorrelation. Extensive experimental results show that our system accurately detects respiration in real environment with the median error lower than 0.35 breaths/min, outperforming the state-of-the-arts.
Content may be subject to copyright.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
C-FMCW Based Contactless Respiration Detection Using Acoustic
Signal
TIANBEN WANG, Northwestern Polytechnical University
DAQING ZHANG, Institut Mines-Télécom/Télécom SudParis
YUANQING ZHENG, The Hong Kong Polytechnic University
TAO GU, RMIT University, Australia
XINGSHE ZHOU, Northwestern Polytechnical University
BERNADETTE DORIZZI, Institut Mines-Télécom/Télécom SudParis
Recent advances in ubiquitous sensing technologies have exploited various approaches to monitoring vital signs. One of the
vital signs is human respiration which typically requires reliable monitoring with low error rate in practice. Previous works
in respiration monitoring however either incur high cost or suffer from poor error rate. In this paper, we propose a
Correlation based Frequency Modulated Continuous Wave method (C-FMCW) which is able to achieve high ranging
resolution. Based on C-FMCW, we present the design and implementation of an audio -based highly-accurate system for
human respiration monitoring, leveraging on commodity speaker and microphone widely available in home environments.
The basic idea behind the audio-based method is that when a user is close to a pair of speaker and microphone, body
movement during respiration causes periodic audio signal changes, which can be extracted to obtain the respiration rate.
However, several technical challenges exist when applying C-FMCW to detect respiration with commodity acoustic devices.
First, the sampling frequency offset between speakers and microphones if not being corrected properly would cause high
ranging errors. Second, the uncertain starting time difference between the speaker and microphone varies over time.
Moreover, due to multipath effect, weak periodic components due to respiration can easily be overwhelmed by strong static
components in practice. To address those challenges, we 1) propose an algorithm to compensate dynamically acoustic signal
and counteract the offset between speaker and microphone; 2) co-locate speaker and microphone and use the received signal
without reflection (self-interference) as a reference to eliminate the starting time difference; and 3) leverage the periodicity
of respiration to extract weak periodic components with autocorrelation. Extensive experimental results show that our system
detects respiration in real environments with the median error lower than 0.35 breaths/min, outperforming the state-of-the-
arts.
CCS Concepts: Human-centered computing → Ubiquitous and mobile computing systems and tools
KEYWORDS
Health Monitoring, Respiration Detection, Contactless Sensing, Acoustic sensing, Device-free Ranging,
ACM Reference format:
Tianben Wang, Daqing Zhang, Yuanqing Zheng, Tao Gu, Xingshe Zhou, and Bernadett Dorizzi. 2017. C-FMCW Based
Contactless Respiration Detection Using Acoustic Signal. PACM Interact. Mob. Wearable Ubiquitous Technol. 1, 4, Article
170 (December 2017), 20 pages. https://doi.org/10.1145/31611881
This work was supported by National Natural Science Foundation of China (No. 61332013), Chinese Scholarship Council Program.
Author’s addresses: Tianben Wang, Xingshe Zhou, School of Computer Science, Northwestern Polytechnical University, Xi’an Shaanxi,
China; Daqing Zhang, Berbadette Dorizzi, Institut Mines-Técom/Técom SudParis 9, rue Charles Fourier 91011 Evry Cedex, France;
Yuanqing Zheng, The Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong; Tao Gu, School of Computer Science and IT,
RMIT University, Australia. Daqing Zhang is the corresponding author (daqing.zhang@telecom-sudparis.eu).
Permission to make digital or hard copies of all or part of this work for personal or classroom use is granted without fee provided that copies
are not made or distributed for profit or commercial advantage and that copies bear this notice and the full citation on the first page.
Copyrights for components of this work owned by others than ACM must be honored. Abstracting with credit is permitted. To copy
otherwise, or republish, to post on servers or to redistribute to lists, requires prior specific permission and/or a fee. Request permissions
from permissions@acm.org.
© 2017 Association for Computing Machinery.
2474-9567/2017/12-ART170 $15.00
https://doi.org/10.1145/3161188
170
170:2 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
1 INTRODUCTION
1.1 Motivation
Recent years have witnessed the surge of ubiquitous sensing technologies and their applications in vital sign
monitoring [1, 2]. Respiration rate is one of the important vital signs, often used to assess and track human
health condition. For instance, research studies show that a deep breath helps reduce blood pressure while
irregular respiration patterns may indicate chronic stress even progression of illness and decline in health [3, 4,
5]. Respiration monitoring in bed also helps track sleep quality and allow timely intervention in case of
respiration abnormality such as obstructive or central sleep apnea-hypopnea which have been found quite
common among the elderly [6, 7, 8]. In particular, chronic obstructive pulmonary disease (COPD) is the third
most common causes of death for people aged 65 and above [9]. As such, respiration monitoring is essential to
enable various ubiquitous healthcare applications. Since these systems often target patients or the elderly with
chronic disease, a convenient and non-intrusive system is highly desirable.
1.2 Limitations of Prior Work
Several respiration monitoring systems have been proposed in clinical setting in early days such as thoracic
impedance pneumography [10] and capnography [11], however these systems are inconvenient and intrusive.
For instance, these systems typically require users to attach/wear special devices such as nasal probes or chest
bands thus restricting body movements and causing inconvenience in their daily lives. Moreover, clinical
respiration monitoring systems typically require well-trained professionals to set up the systems and assist users
to attach/wear respiration monitoring devices properly. To minimize the discomfort brought by the invasive
respiration systems, several attempts have been made for long-term respiration monitoring leveraging on
wearable devices [12, 13]. While these systems are made more tolerable for the elderly, additional issues may
arise such as user acceptance and usability [14].
Contactless sensing technologies explore a new possibility of monitoring vital signs without direct and
physical contact to users. For instance, Penne et al. use a Time-of-Flight camera and apply advanced image
processing algorithms to estimate human respiration rate in a home [15]. Unfortunately, such camera-based
approaches require a subject to face the camera closely besides having privacy concerns and being affected by
the lighting conditions. Kondo et al. deploy a laser sensor to measure the chest wall motion during respiration
[16]. RF based methods have been also widely studied, ranging from Doppler radar [17], UWB radar [18],
FMCW radar [19] to USRP (Universal Software Radio Peripheral) based solutions [20, 21]. The basic idea of
these methods is to detect the chest movement displacement during breathing. Though these methods are able to
accurately detect human respiration, they incur prohibitive cost, preventing these systems from large-scale
deployment in typical home settings. To discover cost-effective solutions, researchers have recently turned their
attentions to widely available commodity WiFi or Zigbee devices. These works can be further classified into two
sub-categories: channel state information (CSI) based method [22-26] and received signal strength (RSS) based
method [27-30]. Limited by the granularity of RSS, the RSS based approach suffers from low resolution, and it
is subjected to interference in the environment. The CSI based method typically yields a higher resolution due to
fine-grained channel information used. However, the study in [22] shows that the minute body movement may
not generate detectable CSI changes. As such, the CSI based approach typically calls for careful selection of
device locations and subcarriers to achieve robustness.
In recent years, acoustic signal has been used to enable accurate affordable contactless reparation monitoring
in home environments. The study in [31] uses 40KHz ultrasonic signal to detect human respiration by sensing
the exhaled airflow. The Doppler effect introduced by the velocity of exhaled airflow can be used to detect
respiration. While the idea of using acoustic signal to detect respiration is promising, these systems are
vulnerable to ambient airflow, such as wind or even the airflow introduced by body movement around it.
Frequency Modulated Continuous Wave (FMCW) has been first used to detect human respiration using acoustic
signal by measuring the chest movement displacement during breathing [32]. It achieves about 0.7 cm distance
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:3
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
estimation resolution and can detect the respiration with chest movement displacement of about 2 cm. However,
the average respiration amplitude, i.e., the chest movement displacement, is only about 1cm. Especially for
Asian people whose body size is relatively thin, new-borns or the elderly, the chest movement displacement is
usually within 1cm [22], and this is beyond the detection capability of the FMCW based systems [32].
1.3 Proposed Approach
Due to the fact that the distance estimation resolution of FMCW is restricted by the sweep band width, it is
difficult to achieve very high distance estimation resolution (i.e., detecting human respiration with chest
movement displacement of less than 1cm.) with narrow-band commodity acoustic device. In this paper, we
propose a Correlation based FMCW (C-FMCW) approach to reliably monitor respiration with commodity audio
devices. Unlike traditional FMCW, C-FMCW estimates the round-trip propagation time of acoustic signals by
discovering the maximum correlation between transmitted signal and received signal. For digital FMCW signal,
the round-trip propagation time can be measured by detecting the number of samples corresponding to the
maximum correlation. As such, the distance estimation resolution of C-FMCW is limited only by the acoustic
signal sampling rate. With current prevailing audio systems of 48KHz, our approach can achieve a ranging
resolution of around 0.4 cm. The ranging resolution can be further improved if higher sampling rates (e.g.,
96Khz) are supported by audio systems. In C-FMCW, linear chirps facilitate the correlation detection, and also
tolerate the frequency selective fading of audio signals due to multipath effect.
Implementing such an audio based respiration system however entails several practical challenges. First, the
sampling frequency offset between commodity speakers and microphones can seriously affect the correlation
based detection. In practice, the sampling frequency offset squeezes or stretches audio signals such that the
received signals and the transmitted signals cannot be aligned during the correlation process. Such an offset if
not corrected properly may incur high ranging error. Second, in practice, even though we send a starting
command to speaker and microphone simultaneously, the actual start time of the speaker is essentially different
from that of the microphone. What’s worse, the starting time difference varies every time we send a starting
command to them, which will introduce intolerable absolute distance estimation errors. Third, due to multipath
effect, weak reflected signals capturing user’s respiration can be overwhelmed by stronger signals reflected from
the static body parts of the user. It still remains elusive to extract the reflected signals capturing respiration.
We hence propose three novel techniques to address the above challenges. First, we propose an algorithm to
compensate dynamically to the transmitted signal and counteract the signal shift caused by sampling frequency
offset between speaker and microphone. Second, we co-locate speaker and microphone, thus, part of the
transmitted signal will be directly received by the microphone without reflection (self-interference). We regard
this part of received signal as a reference to estimate the absolute distance. In this way, we can counteract the
starting time difference between speaker and microphone since the starting time difference remains the same for
the signal without reflection and the signal reflected from abdomen. Third, based on the fact that respiration is
almost periodic in nature, we extract periodic components by using autocorrelation to capture respiration while
filtering out static signals.
1.4 Contribution
The main contributions of this paper can be summarized as follows:
1) We propose a novel C-FMCW method for human respiration monitoring, which is able to achieve sub-
centimeter ranging accuracy. C-FMCW adopts the correlation based detection method to accurately measure the
propagation time of audio signals with high resolution audio samples.
2) We propose several novel ideas to address practical issues, such as developing a compensation algorithm
to dynamically add artificial shift to the transmitted signal to counteract the signal shift caused by sampling
frequency offset between speaker and microphone; leveraging the received signal without reflection to eliminate
the absolute distance estimation error introduced by the starting time difference between speaker and
microphone; leveraging the prior knowledge of respiration periodicity to enhance and extract weak periodic
170:4 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
signals which can be otherwise overwhelmed by stronger reflection signals.
3) We conduct extensive experiments to evaluate our system with 22 subjects. The experimental settings vary
in different sleep postures, different rooms, and different scenarios. We also verify the system robustness with
respect to different sensing distance, different scenarios, different respiration rates, different kinds of simulated
apnea, and body movement. The results show that our system achieves a median error of 0.35 breaths/min.
2 Related Work
In this section, we briefly review the existing contactless respiration detection work, which can be roughly
grouped into two categories:
2.1 Customized Device based Methods
Recently, customized device based technologies ranging from laser [16], infrared [33], Doppler radar [17],
UWB radar [18], FMCW radar [19], USRP [20, 21], ultrasonic sensor [31] have been proposed to detect human
respiration. The basic idea is to measure the chest movement displacement during respiration. Although these
methods achieve accurate respiration detection, they typically require costly signal transceivers or customized
radio frontends. As such, these methods are less appealing than the commodity device based methods for large-
scale deployment in home settings.
2.2 Commodity Device based Methods
In recent years, researchers have turned their attentions to commodity devices since they are readily available in
home environments. The most widely studied methods are WiFi based and audio based.
2.2.1 WiFi based methods. Ubibreathe [27] measures the received signal strength (RSS) to extract the
respiration rate of a person in vicinity. RSS-based methods, however, require the subject in the line-of-sight path.
WiSleep [24] and the studies in [22, 23, 25, 26] use the fine-grained channel state information (CSI) in
commodity WiFi cards to detect respiration rate. Wang et al. [22] apply the Fresnel zone model to characterize
the radio propagation in indoor environments and reveal the principle and limitations of CSI based human
respiration sensing with WiFi signals. According to their findings, CSI based approaches cannot always reliably
track human respiration and require careful selection of relative distance between human and WiFi devices.
2.2.2 Audio based methods. ApneaApp [32] is a mobile application that allows smartphone users to monitor
respiration rate. ApneaApp sends audio chirps and measures frequency shifts to monitor the breathing patterns
so as to detect Apnea. The ranging resolution of ApneaApp is fundamentally limited by the narrow sweeping
band of the device. As the ranging resolution of ApneaApp is only 0.7 cm, it can detect a deep breath with chest
displacement of around 2 cm. Our system offers a more fine-grained ranging resolution of 0.4 cm within the
80cm range. Moreover, the maximum ranging error is less than 0.2 cm when the distance between the acoustic
devices and subject is from 40 cm to 60 cm, which can reliably detect a normal breath of young kids or elders
who typically have a chest displacement of less than 1 cm. iSleep [34] uses microphone on a smartphone to
detect the events that are closely related to sleep quality such as body movement and snore. However, iSleep
cannot detect respiration. BeepBeep [35] detects the roundtrip time of flight at the resolution of audio sampling
rate and achieves high ranging accuracy between two smartphones. BeepBeep requires both smartphones to
actively generate audio chirps, while our work detects the time of flight of reflected audio chirps. RunBuddy [36]
measures the breathing pattern of a runner with Bluetooth headset and suggests a proper running rhythm. The
headset is not intrusive for a runner but may cause inconvenience to a user during sleep.
In our work, we propose a novel acoustic respiration detection method leveraging commodity acoustic
devices, which achieves a ranging resolution of around 0.4 cm. Based on the high ranging estimation resolution,
we manage to detect human respiration with the chest movement displacement of less than 1cm.
3 C-FMCW: A HIGH RESOLUTION DISTANCE ESTIMATION METHOD
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:5
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
3.1 FMCW and its Limitation
3.1.1 FMCW. The naïve way to estimate distance to an object using wireless signal is to transmit and receive
a sharp pulse and compute the time delay , then estimate the distanceusing their relationship , where
denotes the speed of wireless signal. However, sending a sharp pulse requires a large bandwidth. Instead,
FMCW [37] indirectly estimates the propagation delay based on the frequency shift of the chirp signal as
follows.
As shown in Fig. 1, the red line denotes the transmitted signal, whose frequency linearly increases over time.
The frequency of the transmitted signal at time is given by: 
, where , and denote carrier
frequency, sweep bandwidth and sweep period, respectively. The phase of the transmitted signal is calculated by
integrating over time: 


the transmitted signal is then presented as . For simplicity, the amplitude is assumed to be 1.
The transmitted signal is reflected by the target with the time delay . If amplitude attenuation is ignored, the
received signal at time can be represented as:  (the blue line in Fig. 1). Consider the
transmitter and receiver are co-located. is calculated as:


where denotes the distance between transceiver and target, denotes the moving speed of the target, and
denotes the propagation speed. The receiver multiplies the transmitted signal with the received signal as
 . By using  
and filtering out the high frequency
component , is then simplified as 
  (the green line in Fig. 1).
Plugging Eq. (1), the frequency of can be represented as:

 

 


. The terms with respect to  
is too small and can be ignored, thus,  can be then simplified as:


 
 . For the static or very slowly moving target, i.e., is close to 0, we can further simplify
the frequency of mixed signal, and thus obtain the distance between transceiver and target as follows.

   
3.1.2 Theoretical Range Resolution Upper Bound of FMCW. The transmitted signal  has the
Fig 1. Chirp signal in FMCW
Time
Time
Frequency
Transmitted signal Receive signal
170:6 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
fundamental frequencies at all multiples of the frequency  
[38, 39]. If ignoring the amplitude attenuation and
assuming that the target is static or moving slowly, the received signal  is a time shifted version of
transmitted signal , so  and  have the same fundamental frequencies. Thus,   has the
fundamental frequencies at all multiples of the frequency  
, since the periodicity is preserved when two same
period signals are multiplied [38]. It means that in the frequency domain, has the spectral points only at  
Hz. In other words, the frequency resolution of mixed signal,  
.
Based on Eq. (2), we can infer the theoretical range resolution of FMCW as:

 
Eq. (3) indicates that the range resolution, i.e., fixed error,  depends on frequency resolution of the mixed
signal . However, as mentioned above,  is bounded by the chirp frequency  
, i.e.,  
. which
means that in order to distinguish targets in two different distances, the frequency of the mixed signal cannot be
smaller than the chirp frequency [39]. Thus, the theoretical range resolution upper bound of FMCW is computed
as follows: 

 
Eq. (4) indicates that the range resolution of linear modulated FMCW depends on sweep bandwidth .
Narrower bandwidth results in lower range resolution.
3.1.3 Highest Range Resolution of FMCW Using Commodity Audio Devices. There are two practical
frequency bounds that restrict the available sweeping bandwidth (as shown in Fig. 1). On one hand, the upper
bound of human audible range is about 18KHz. To be inaudible to users, the starting frequency of chirp signal
(as shown in Fig. 1) should be set higher than 18KHz. On the other hand, for commodity audio devices, the
frequency response starts to decrease rapidly in the band higher than 23KHz. Thus, the available sweeping
bandwidth = 23KHz-18KHz= 5KHz. According to Eq. (4), the highest range resolution of FMCW using
commodity audio devices is represented as follows:



in which 3.43 cm is much larger than the typical chest displacement of 1 cm during breathing. It implies that
directly implementing FMCW using commodity audio devices cannot reliably detect human respiration.
3.2 Cross-correlation function based FMCW (C-FMCW)
3.2.1 Basic Idea of C-FMCW. The objective of distance estimation is to measure delay . For transmitted
signal with a modulation period , the delay within  
reflected from target can be measured using the main
peak of cross-correlation of transmitted signal and received signal. In this paper, we use FMCW wave as
transmitted signal, and exploit cross-correlation to measure the delay. We call this method Cross-correlation
function based FMCW (C-FMCW).
The cross-correlation function is defined as:
 


 



where is the number of samples of transmitted signal in one modulation period , 
. measures the similarity of  and -sample shifted version of . Suppose reaches its
peak  with the -sample shift. Since  is a time shifted version of  with amplitude attenuation, it
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:7
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
means that  overlaps  with the -sample shift. Thus, we can measure the time of flight by counting the
number of audio samples. For example, Fig. 2(a) and Fig.2(b) show the transmitted signal and received signal
with 3 modulation periods (sampling frequency  KHz,  KHz, KHz, 0.08 s),
respectively. The received signal is set to  samples shifted version of transmitted signal. Fig. 2(c)
shows the cross-correlation of transmitted signal and received signal. We can see the peak in .
According to the definition of , the peak should be . In Fig 2(d), we can see that peak
appears indeed at . Thus, the delay can be calculated as:


where is the sampling frequency. Combining with Eq. (1), we measure the distance between transceiver and
target as:
 



 
For the static or very slowly moving target, i.e., is close to 0, we can simplify as:


3.2.2 Theoretical Range Resolution Upper Bound of C-FMCW. Based on Eq. (8), the theoretical range
resolution of C-FMCW can be represented as: 

According to the definition of cross-correlation function in Eq. (5), the theoretical resolution of  is
, since the cross-correlation is calculated at the resolution of each audio sample. Thus, the theoretical
range resolution of C-FMCW can be calculated as:



From Eq. (10) we can see that the range resolution of C-FMCW is limited only by the sampling frequency
and independent of sweeping bandwidth . Higher sampling frequency results in higher ranging resolution. In
other words, even if available sweeping bandwidth is narrow, C-FMCW can still achieve very high ranging
resolution.
3.2.3 Range Distance Resolution of C-FMCW Using Commodity Audio Devices. Generally, commodity
audio systems support 22.05KHz, 44.1KHz and 48KHz sampling frequencies. In order to achieve high ranging
resolution, we set the sampling frequency to , and the ranging resolution of C-FMCW is calculated
as:
(a) (b) (c) (d)
Fig. 2. Cross-correlation function of transmitted and received chirp signals with delay. (a) transmitted signal. (b)received
signal with delay. (c) cross-correlation function of transmitted and received signals. (d) enlarged view round the peak of (c)
170:8 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.



Such a ranging resolution suffices to reliably detect human respiration. In fact, by combining appropriate
interpolation techniques, the resolution can be further improved. In this work, we have adopted 2x interpolation
method to improve the resolution to



Note that excessive interpolation has a limit in improving the resolution.
3.3 C-FMCW Verification
In this section, we conduct real experiments to verify Eq. (8) and Eq. (10) in C-FMCW. The experimental
settings and results are reported as follows.
We bind a general speaker (JBL Jembe, 6 Watt, 80 dB) and microphone (SAMSON MeteorMic, 16 bit, 48
KHz) together (as shown in Fig. 3(a)), they work as a simple acoustic radar (transceiver). The transceiver
illuminates target directly (as shown in Fig. 3(b)). The speaker transmits FMCW signal (,
, , ) continuously, which is beyond human audibility range. Meanwhile, the
microphone receives the signal reflected by the target at the frequency of 48 KHz with 16 bits.
A ruler is placed on the desk as the ground truth. The position, where the reading of the ruler is 0, is regarded
as the reference position (highlighted with blue dashed line in Fig. 3(b)). The distance between transceiver and
reference position is set to 0. To verify the ranging resolution of C-FMCW, we move the target to 13 different
positions ranging from 0.1 m to 1.2 m. The distance estimation results are shown in Fig. 3(c). We observe that
within 80 cm, the distance estimated error can be controlled below 0.4 cm. Moreover, the maximum ranging
error is less than 0.2 cm when the distance between the acoustic devices and subject is from 40 cm to 60 cm,
meaning that chest displacement which is less than 0.5 cm can be detected using C-FMCW. As distance
increases, we observe that error also increases due to degradation of reflected acoustic signal. For more details,
readers are encouraged to watch the video at this link: http://dwz.cn/6EE2Dx
4 CONTACTLESS RESPIRATION DETECTION USING C-FMCW WITH COMMODITY
ACOUSTIC DEVICES
In this section, we present the detailed design of C-FMCW based contactless respiration detection system using
commodity acoustic devices. We first identify the technical challenges and our solutions. We then give an
overview of the system architecture and describe each major function component in detail.
(a) (b) (c)
Fig. 3. C-FMCW verification. (a) transceiver, (b) full view of experiment, (c) distance estimation error as ground truth is
varied
Microphone
Speaker
Reference position
Transceiver
Ground Truth: ruler Target
R
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:9
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
4.1 Practical Challenges and Solutions using C-FMCW to Detect Respiration with Commodity
Acoustic Devices
Even though C-FMCW is promising to achieve high ranging resolution, a number of practical challenges exist
when applied to detect respiration with commodity acoustic devices such as sampling frequency offset between
transmitter and receiver, the starting time offset between transmitter and receiver, as well as major reflection
path selection problems. In this section, we present these challenges and our solutions.
4.1.1 Challenge 1: Sampling Frequency Offset Between Transmitter and Receiver. Existing audio based
ranging approaches typically assume that the sampling frequencies of transmitter and receiver are exactly the
same. Unfortunately, due to imperfect clocks, the frequency offset between transmitter and receiver introduces
error in calculating  (in Eq. (8)). Moreover, the error will accumulate over time. In Fig. 4(a), the red line
highlights the  shift caused by the sampling frequency offset between transmitter and receiver. We find that
within 30 seconds (about 10 breathes), the  shift accumulates up to 87 samples, which leads to a distance
measurement error of 31.1cm according to Eq. (8). In other words, during each breathing, the ranging error
caused by the sampling frequency offset can exceed 3 cm, which is much larger than the chest displacement of 1
cm during breathing. In order to reliably detect respiration, we must eliminate the errors caused by the sampling
frequency offset between transmitter and receiver.
4.1.2 Solution 1: Sampling Frequency Offset Calibration Algorithm. As shown in Fig. 5, the basic idea of
our solution is to add a shift into transmitted signal to compensate the  shift caused by sampling frequency
offset between transmitter and receiver. Specifically, given the  shift threshold  and time duration , our
compensation algorithm will shift audio samples so that the  shift in duration is smaller than  by
adapting two parameters compensation rate and compensation unit . Compensation rate  
( and both of them are positive integers) means that in every iterations (as shown in Fig. 5), we will
add times shifts into transmitted signal. Compensation unit  means that we add shift in unit 
samples. Our compensation algorithm is summarized in ALGORITHM 1. The algorithm firstly finds shift unit
∆Shift to make the symbols of system shift when r = 0 and r = 1 are different. Then, the algorithm uses the
binary search method to find appropriate compensation rate r that makes system shift during period D smaller
(a) (b) (c) (d)
Fig. 4. Practical issues using C-FMCW to detect respiration with commodity acoustic devices. (a) lag shift over time caused
by sampling frequency offset between transmitter and receiver. The red line indicates the delay shift over time before
correction, while, the blue line indicates the delay shift over time after correction. (b) starting time difference between
speaker and microphone. (c) multiple reflection path from abdomen, and (d) cross-correlation function peak amplitude.
Fig. 5. Solution for sampling frequency offset between transmitter and receiver
87 samples
Reference point
Multiple paths reflected
from abdomen

Transmit Acoustic
Signal  Receive Acoustic
Signal  C-FMCW(, )
etc.
Shift to

Compensation
Algorithm
170:10 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
than shift threshold OT. Note that during the execution of the ALGORITHM 1, the target should be static. The
target movement will introduce shift and influence compensation performance. In practice, we only need to use
a static object to get the compensation parameters for once during system calibration before starting respiration
detection. The time cost incurred by ALGORITHM 1 depends on the input parameters and . Larger and
smaller  will result in longer running time but more accurate compensation rate and compensation unit
.
The blue line in Fig. 4(a) highlights the system shift after applying our compensation algorithm. We observe
that there is almost no  shift after applying the compensation. In summary, with this compensation algorithm,
the effect caused by sampling frequency offset between transmitter and receiver can be well controlled.
4.1.3 Challenge 2: Starting Time Offset Between Transmitter and Receiver. Ideally, if speaker and
microphone start to work simultaneously, we can directly estimate the absolute distance from transceiver to
abdomen. Unfortunately, even though we send the starting commands to speaker and microphone
simultaneously, the starting time of speaker and microphone can be different in practice. Moreover, the starting
time difference varies over time. Fig. 4(b) shows 30 times starting time difference between speaker and
microphone in our system. The varied starting time difference will introduce a large error in initial distance
estimation. In the example shown in Fig. 4(b) the starting time difference (average 0.034s) will introduce
1632 samples  shift. According to C-FMCW distance estimation equation Eq. (8), it will
introduce an absolute distance error of  .
4.1.4 Solution 2: Cancelling Starting Time Offset Using Self-interference Effect. As shown in Fig. 3(a), in
our system, speaker and microphone are co-located. As a result, part of the transmitted signal will be received by
the microphone without reflection (i.e. self-interference). This part of received signal will generate a special
peak in the cross-correlation function of transmitted signal and received signal (the reference point shown in Fig.
4(c) and Fig. 4(d)). Due to no reflection, this part of received signal will not be affected by the target. Thus, both
the amplitude and location of this special peak will keep relatively static. It is easy to use variation of amplitude
and location to distinguish this special peak from others. We name the special peak reference point. As shown in
Fig. 4(c) and Fig. 4(d), with the reference point, we can cancel the starting time offset between transmitter and
receiver by using the location difference between reference point and the peak generated by the receive signal
reflected from the target. Then, by using C-FMCW distance estimation equation Eq. (8), we can accurately get
the distance. This solution can completely avoid the interference from the starting time difference between
speaker and microphone, since the  shift caused by the starting time difference is the same for the reference
point and the peak generated by the receive signal reflected from the target. Thus, when we calculate the  by
ALGORITHM 1 Compensation Algorithm
Input: Duration D, and offset threshold OT
Output: compensation rate r and compensation unit ∆Shift
 0;  1;
find the smallest , which makes getShift() and getShift () with different symbol;
% getShift (r, ∆Shift, D) gets the Lag shift in duration D with the compensation rate r and compensation unit ∆Shift
while |shift|  do % Binary search method to find appropriate
shift getShift (
, );
if shift < 0 % update the upper bound or lower bound according to the symbol of current system shift do
 ;
else
 ;
end
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:11
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
using the location difference between the reference point and the peak generated by the receive signal reflected
from abdomen, the  shift caused by the starting time difference can be cancelled.
4.1.5 Challenge 3: Multiple Reflection Paths from Different Parts of Body. In practice, there are multiple
reflection paths from different parts of abdomen. As shown in both Fig. 4(c) and Fig. 4(d), there are two peaks
(in the blue circle) corresponding to two different reflection paths (note that in practice, there may be more than
2 peaks). Different parts of abdomen have different movement displacements during breathing. The reflection
paths from static parts cannot be used for respiration detection. Only the reflection paths from the moving parts
during breathing can contribute to respiration detection. Moreover, the power of received signal from different
reflection path fluctuates dynamically, which will result in amplitude fluctuation of cross-correlation peaks. As
shown in Fig. 4(c) and Fig. 4(d), the two peaks in blue circle alternatively become the largest peak. It is
challenging to identify the representative peak that can effectively reflect human respiration among the multiple
peaks.
4.1.6 Solution 3: Selecting Major Reflection Path Using the Periodicity of Respiration. Even though there
may exist multiple reflection paths from different parts of abdomen, we know that human respiration is almost
periodical in nature. It means that the distances from the moving parts of the target during breathing will vary
periodically, while other distances from static reflection paths are not periodical. Thus, we can select the
effective reflection path by measuring the periodicity. The periodicity can be easily measured by using an
autocorrelation function. For the sequence , autocorrelation function is defined as follows:
 
 
Fig. 6(a) shows two different distance estimations from two different reflection paths over time. Fig. 6(b)
shows the autocorrelation functions of the two distance estimations. In Fig. 6(b), we see that the peaks of the
autocorrelation function is higher if the distance estimation is more cyclical.
(a) (b)
Fig. 6. Autocorrelations of two different distance estimation results. (a) two distance estimation sequences. (b) the
autocorrelation function of the two distance estimation sequences in (a)
Fig. 7. Overall framework of C-FMCW based contactless respiration detection system using acoustic signal
Acoustic Signal Pre-processing
Acoustic signal t ransmitting
and receiving
Band-pass filtering
Frequency offset
compensation
Distance Estimation
Calculating cross-
correlation function
Finding reference point
Estimating the distances
from abdomen
Respiration Detection
Measuring the periodicity of
the distance estimations
Selecting one distance
estimation corresponding to
major reflection path
Respiration detection and
visualization
170:12 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
4.2 C-FMCW Based Respiration Detection System Framework
In this section, we present the key workflow of C-FMCW based contactless respiration detection system using
acoustic signal. We first give an overview of the system architecture, and then describe each key functional
module in detail.
4.2.1 System Architecture. As shown in Fig. 7, our system is composed of three major components, namely
acoustic signal pre-processing, distance estimation, and respiration detection. The acoustic signal pre-processing
module filters out noise and compensates the Lag shift. The distance estimation module estimates the distance of
each reflection path from the target. The respiration detection module detects the respiration from the distance
estimation results. In the following, we summarize the key functional modules of C-FMCW.
4.2.2 Acoustic Signal Pre-processing. The objectives of this module are to: 1) ensure real-time continuous
transmission and reception of acoustic signals; 2) denoise received signals; and 3) compensate the Lag shift
caused by sampling frequency offset between transmitter and receiver.
In order to filter background noise, we apply a band-pass filter (passband [, ]), where , denotes
carrier frequency and denotes sweep band width.
To compensate the Lag shift caused by sampling frequency offset between transmitter and receiver, we firstly
get the compensation rate and compensation unit  by using our composed compensation algorithm in
ALGORITHM 1. Then, we distribute times compensation (each time ∆Shift samples) to transmitted signal
(as shown in Fig. 5) during iterations uniformly to counteract the shift caused by the sampling frequency
offset between transmitter and receiver. and have the relation  
.
4.2.3 Distance Estimation. The key function of this module is to estimate the distance of each reflection
path from abdomen using our proposed C-FMCW. Specifically, we first calculate the cross-correlation function
of compensated transmitted signal and received signal using Eq. (5), and find all the peaks of cross-correlation
function. Then, we identify the reference point by finding the peak whose variation of the amplitude and
location are smallest. After the reference point is located, we estimate the absolute distances (here the distance
means half of the length of reflection path) of all reflection path by using the solution presented in Section 4.2.2.
We calculate  by using the location difference between reference point and the peak corresponding to
current reflection path. Then, we estimate the distance by using C-FMCW distance estimation equation Eq. (8).
The distance estimation result of each reflection path over time is saved as a distance sequence. These distance
sequences of all the reflection paths will be regarded as the input of next module.
4.2.4 Respiration Detection. The major work of this module is to detect respiration based on the distance
estimation results. As mentioned in Section 4.1.3, not all the distance estimation results can be used for
respiration detection. We first calculate the periodicity of all the distance sequences by using autocorrelation
function. Then, we select the distance sequence corresponding to the major reflection path which can reflect the
respiration well if the periodicities of all the distance sequences are lower than a given threshold. Thereby, the
interference factors (e.g., body movement, sleep posture change) can be detected. Fig. 8 shows the respiration
detection result while the user changes sleep posture. The system will stop identifying breathing when the user
changes sleep posture, since during this period the distance estimation fluctuation cannot reliably capture the
chest movement. The system will recover to identify breathing once at least one distance sequence’s periodicity
Fig. 8. Respiration identification with body movement
Body movement
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:13
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
becomes higher than a threshold.
5 EVALUATION
In this section, we conduct comprehensive experiments to evaluate the proposed system. First, we describe our
experiment settings. We then briefly describe the baseline approach. We conduct comprehensive experiments to
evaluate our system in four different rooms with 22 subjects, who have three different sleep postures and
compare the performance with that of the baseline approach. We also evaluate the system robustness with
different scenarios, different sensing distance, different respiration rates, different kinds of simulated apnea, and
body movement. Finally, we discuss the limitations of the system.
5.1 Experiment Setting and Set-ups
To set up our respiration monitoring system, we use two laptops (Thinkpad T450 with Intel Core i5-5200 CPU,
8G RAM; Dell Latitude E6540 with Intel Core i7-4800MQ, 4GB RAM) and connect them with a pair of
commodity speaker (JBL Jembe, 6 Watt, 80 dB) and microphone (SAMSON MeteorMic, 16 bit, 48 KHz) which
forms an audio transceiver in our system, as shown in Fig 3(a). In practice, we use two transceivers to increase
the coverage of monitoring area. The proposed respiration detection algorithm implemented in Matlab keeps
monitoring user’s respiration in real-time. The speakers are programmed to transmit FMCW signal continuously.
Meanwhile, the microphones receive the echo with 48 KHz sampling rate and send the received audio signal to
the connected laptop for data processing and respiration detection. In order to avoid frequency band interference,
the speaker in the first transceiver is programmed to transmit FMCW signal with the parameters:
 , while the speaker in the second
transceiver transmits FMCW signal with the parameters: 
. , and denote the carrier frequency, sweep band width and sweep period,
respectively. is FMCW signal length of time, i.e., the system processes second signal for one time. In other
words, our system processes  
sweep periods FMCW signal in one system iteration. is sampling
frequency of speakers. The microphones in two transceivers receive the reflected signal with 48KHz and filter
the received signal with band-pass filters. In particular, the pass bands of the two transceivers are set to [18KHz,
20.3KHz] and [20.7KHz, 23KHz], respectively.
Due to clothes absorption and attenuation in the air, the received signal reflected from abdomen is very weak.
Thus, we place the transceiver toward subject’s abdomen. However, one transceiver will fail to detect
respiration if the subject changes sleep posture. We notice that subjects may change their sleeping posture from
lying on one’s back to lying on right side or left side. For system configuration, we firstly detect the effective
range, covering area and max slant angle of the transceiver. To this end, we ask each subject to lie on a bed and
place a transceiver above the subject’s abdomen for respiration detection. As shown in Fig. 9(a), the acoustic
beam width is about 25° and the effective detection distance is about 80cm, while the effective detection
(a) (b) (c) (d)
Fig. 9. The detectable area of transceiver and its layout facing in different sleep postures. (a) the range, covering area and
max slant angle of the transceiver. (b) ~ (d) the device layout and detectable area facing in different sleep postures.
Transceiver
Acoustic beam
Mattress
25°
50cm
Human abdomen
110cm
50cm
50cm
Undetectable area
Detectable area
170:14 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
distance is about 70cm when the slant angle is 45°. In practice, if the slant angle is set as larger than 45°, the
transceiver cannot receive the effective reflection signal.
Based on the above results, we deploy two transceivers as shown in Fig. 9(b) -Fig. 9(d). The transceivers are
placed at the upper-left and upper-right of the subject’s abdomen, respectively with the angle about 45° pointing
to the user’s abdomen. The perpendicular distance between mattress and device is about 50 cm and the distance
between two transceivers is about 110 cm. We can see that, theoretically this setup covers all the facings of a
subject, no matter where the subject faces during sleep.
5.2 Baseline Approach
We compare C-FMCW with two representative audio based approaches [31, 32]. The theoretical ranging
resolution of App [32] is around 0.7 cm which can be used to detect respiration with chest movement
displacement of about 2 cm. Different from the subjects selected in [32] who are mainly western adults, people
in Asia, young kids and elders have relatively mild chest movement during normal breathing, i.e. with
displacement of less than 1 cm [22], which is beyond the detection capability of the ApneaApp system. Based
on this fact and extensive verification experiments, we find that [31] is the only existing work that can detect
human respiration with small chest movement using acoustic signal. Instead of directly monitoring the chest
movement, the work [31] measures the airflow of respiration and monitor respiration. Specifically, the velocity
of exhaled airflow will introduce the Doppler effect to the reflected acoustic signal. Thus, we choose it as the
baseline to verify whether our system is also able to detect respiration accurately.
Yet, the work [31] needs a customized device to generate 40 KHz acoustic signals. As the off-the-shelf
speaker can only generate acoustic signals with maximum frequency of 23 KHz, their proposed algorithm
Fig. 10. System settings in test rooms. The red circles mark the transceivers.
(a) (b) (c)
Fig. 11. Respiration detection results of 22 subjects with three different sleep postures. (a) sleep on one’s back, (b) facing
left, (c) facing right
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:15
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
cannot be applied directly. We thus implement it with a 23 KHz acoustic signal, setting the key parameters
accordingly to make the system work well. The key difference between two systems is that the transceiver faces
different part of human body. Specifically, the system in [31] monitors the area around nose, while our system
monitors the area around chest and abdomen.
5.3 System Performance Evaluation
We conduct comprehensive experiments to evaluate our system in four different rooms with 22 subjects, who
have three different sleep postures. We compare the performance with baseline approaches in four different
practical scenarios. We also evaluate the system robustness with different scenarios, different sensing distance,
different respiration rates, different kinds of simulated apnea, and body movement.
5.3.1 Evaluation with Different Subjects. We recruit 22 participants (7 elders, 2 young kids and 13 adults,
including 10 females and 12 males) to evaluate the effectiveness of our system. The participants are asked to set
up the system and properly adjust facing direction of audio transceivers according to the requirements specified
in Section 5.1. The experiment environment is shown in Fig. 10. We set aside 15 minutes for the participants to
lie on his/her back so that the participants really fall asleep before measurements. We then detect the respiration
with each posture for about 200 seconds. During the measurements, two subjects watch the video stream to
record the ground truth manually. Fig. 11(a) shows the respiration detection error CDF. We can see that the
median respiration detection error of our system is lower than 0.35 breaths/min. In most cases (p > 0.95), our
system achieves an error lower than 0.33 breaths/min. The median error of the baseline approach is around 1
breaths/min. In addition, the max error of our system is 0.6 breaths/min, while the maximum error of the
baseline reaches 3.7 breaths/min. Note that, based on our measurements, even the highest and fattest subject
(male, height 1.85m, weight 95 kg), his chest movement displacement is about 1cm when he breathes naturally.
This experimental result indicates: 1) the proposed C-FMCW achieves very high distance estimation resolution;
2) our system is able to accurately detect human respiration when the participants sleep on his/her back; 3) the
system is easy to set up in practice.
5.3.2 Evaluation with Different Sleep Postures. Except for sleeping on ones back, lying on one side is also
(a) (b)
(c) (d)
Fig. 12 Respiration detection error in four different rooms.
170:16 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
one common sleep posture. With the same experiment settings, all the subjects are recruited to evaluate the
performance of our system when subjects lying on one side. Fig. 11(b) and Fig. 11(c) show the respiration
detection error CDF when subjects sleeping on the left and right side, respectively. The result shows that the
median error of our system is 0 breaths/min, while the median error of the baseline is 0.35 breaths/min. In
addition, the max error of our system can be controlled smaller than 0.35 breaths/min, while the maximum error
of the baseline is 1 breaths/min.
The experimental results indicate that our system is able to accurately detect human respiration for all three
common sleep postures in real environment and achieve better performance.
In addition, we observe an increased error of the baseline when a user lies on his/her back. With this sleeping
posture, the angle between the acoustic signal and the exhaled airflow is closed to 90°, which will result in
relatively small effective velocity component of exhaled airflow. Thus, the Doppler effect will decrease and
result in larger detection error.
5.3.3 Evaluation in Different Environments. We conduct the experiments in four different rooms with
different size and layout. Fig.12(a) ~ (d) show the average respiration detection error CDF in the four test rooms,
respectively. The median detection errors of our system for all three sleep postures in four test rooms are 0
breaths/min, while the median errors of the baseline are 0.35 breaths/min. There is no obvious difference in the
four test rooms for both our method and the baseline. It indicates that our system is not sensitive to the
experiment environment. The real demo video is provided at link below: http://dwz.cn/6EDZnK
5.3.4 Evaluation with Different Scenarios. In order to test the applicability in real bed rooms, we recruit two
subjects to evaluate our system and compare with the baseline in the following four practical scenarios.
1) A subject wears thin cloth or is covered by thin quilt with windows opened or air condition turned on (let
breeze blow into the room, e.g., in summer scenario)
2) A subject wears thick cloth or is covered with quilt (e.g., in winter scenario)
3) A subject wears thin cloth or is covered by thin quilt with windows closed and air condition turned off (e.g.,
in spring and autumn scenarios)
4) A subject sleeps with quilt covering his/her head or sleep prostrate.
Fig. 13 (a) shows the respiration detection error in the above four scenarios. We can see that in the first
scenario, our system works well and the error is smaller than 0.5 breaths/min, while the baseline fails to detect
respiration. This is due to the fact that the baseline is to detect respiration by sensing exhaled airflow which can
be easily disturbed by ambient airflow such as wind or even people passing by. In contrast, our system still
works since our distance measurement between transceiver and the chest of subject will not be affected by
airflow. In the second scenario, the baseline works well while our system fails to detect respiration. This is due
to the fact that in this scenario, the chest movement displacement during breathing will be buried since the
displacement of thick cloth or quilt is very small. The sensing target of baseline is the airflow around nose, so it
will not be affected by thick cloth or quilt. In the third scenario, both systems work well, since there is no
interference for both systems. In the fourth scenario, both systems fail to detect respiration, since the exhaled
(a) (b)
Fig. 13 System roobustness to (a) different test scenarios and (b) different sensing distance
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:17
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
airflow around the nose are blocked and displacement reflected on her/his back is too small to be detected
directly.
5.3.5 Evaluation with Different Sensing Distance. We vary the distance between the transceiver and the body
of subject ranging from 0.3 m to 1.1 m with an interval of 0.1 m. Two subjects are recruited to evaluate the
performance of our system. At each position, we test for 6 minutes. Fig. 13 (b) shows the average respiration
detect error as distance is varied when subjects sleeping with three different postures. We can see that within 0.7
m, the system achieves respiration detection error smaller than 0.5 breaths/min. Beyond 0.7 m, the error will
increase with the distance between the transceiver and the subject mainly because of the signal attenuation. The
device with higher output power and higher frequency response in effective frequency band (18KHz~23KHz)
will result in larger effective sensing range.
5.3.6 Evaluation with Respiration Rate Change
.
As our system detects the breathing rate by measuring the
periodicity of distance variation, respiration rate change will weaken the periodicity and may influence our
system. In this work, we use a sliding window method which can discard the obsolete signals and adapt to the
change of respiration rate.
Two participants are asked to do high-intensity exercises like push-ups, which will increase the respiration
(a) (b)
Fig. 14. Respiration detection results of two participants while respiration rate is changing over time.
(a) (b)
Fig. 15. Partial apnea identification results of two subjects.
Fig. 16. Respiration detection with body movement
ApneaApnea Apnea
Hypopnea Hypopnea
Body movement Body movement
Body movement Body movement
170:18 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
rate significantly. Then, we let the participants lie on the bed and detect their respiration rates until the
respiration rates gradually fall back to a normal level (10 ~ 15 breathing counts per minute). Fig. 14 shows the
respiration detection results of the two participants after high-intensity exercises. The red lines show the
measured displacements and the green lines track the instantaneous respiration rates. We can observe that the
respiration rates of two participants change from about 20 breathing counts per minute to about 10 breathing
counts per minute. Our system provides sufficient ranging resolutions to accurately track both rapid shallow
breathing and slow deep breathing throughout the process.
5.3.7 Detection of Apnea
.
Detecting apnea is an important objective of monitoring respiration during sleep.
When apnea happens, the periodicity of distance variation will be absent. Benefitting from the usage of sliding
window to detect the periodicity of distance variation, after the subject recovers to breathe normally, the system
will recover to detect breathing.
Limited by legal constraints, we could not test our system with real apnea patients in hospitals for now.
Instead, we simulate Central Apnea (CA) and Obstructive Apnea (OA) following the clinical symptom
described as “hold breath for a while” [32, 40], and simulate Hypopnea event following the clinical symptom
described as “breathing becomes shallow” [32]. We recruit two participants (one male and one female) to test
apnea detection performance. We repeat the experiment for 10 times each of which lasts for 2 to3 minutes. Fig.
15 (a) shows that our system accurately detects CA and OA. Fig. 15(b) shows that our system also accurately
detects the Hypopnea event.
5.3.8 Evaluation with body movement
.
Body movement (e.g., sleep posture change) generates strong but
un-rhythmical distance displacement variation. The displacement caused by chest movement would be
submerged. Under this condition, it is difficult to detect breathing. Our system is designed to suspend respiration
detection once body movement is detected. Two subjects are recruited to test whether our system can actually
suspend when a body movement occurs. We detect subjects’ respiration for about 15 minutes. During respiration
detection process, the subjects change sleep postures or move limbs several times. Fig. 16 shows that our system
suspends respiration detection while body movement occurs and recovers for detecting respiration after the
movement stops.
5.4 Discussion
5.4.1 Limitations. Even though the experimental results demonstrate that the proposed system can detect
human respiration with three common sleep postures in real rooms and can adapt to different real scenarios,
different sensing distances, changing respiration rate, different kinds of simulated apnea and body movement,
the current implementation has the following two limitations. First, the transceivers must face the abdomen
directly to ensure that the transceivers can receive reflected acoustic signals. As mentioned in Section 5.3.4,
wearing thick cloth or covering with quilt and sleeping prostrate may influence accuracy. This is due to the fact
that when sleeping on one’s chest, the displacement reflected on her/his back is too small to be detected. Second,
so far, we have not tested our system with subjects suffering sleep apnea yet due to users’ privacy and legal
issues. We note that the apnea detection study is based on simulated apnea, where volunteers try their best to
simulate the apnea events according to the clinical symptoms. Yet, we believe the experiments can at least allow
us to evaluate whether our system can detect some types of apnea events that would lead to breath stopping or
other irregular breathing patterns.
5.4.2 Potential applications. C-FMCW achieves 0.4 cm ranging resolution within 80 cm using commodity
audio devices. The ranging resolution can be further improved if higher sampling rates (e.g., 96Khz) are
supported by audio systems. Beneficial from the high ranging resolution, C-FMCW can be potentially applied to
detect other types of subtle movements like heartbeat, finger gestures and even mechanical oscillations if the
sampling rate is further increased.
6 CONCLUSION
In this work, we study contactless human respiration detection using acoustic signal. By exploring the traditional
C-FMCW Based Contactless Respiration Detection Using Acoustic Signal 170:19
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
FMCW principle, we proposed a novel correlation based distance estimation method C-FMCW. Based on C-
FMCW, we develop a contactless respiration detection system using commodity audio devices. We develop and
integrate several elaborated acoustic signal processing methods to address some practical challenges. We
conduct comprehensive experiments to evaluate the proposed respiration detection system. We implement the
proposed C-FMCW with commodity audio devices and demonstrate that our system can achieve a ranging
resolution of less than 0.4 cm. We evaluate the performance of respiration monitoring in four different rooms
with 22 subjects who have three different sleep postures. Experimental results show that our system accurately
detects subjects’ respiration with the median error lower than 0.35 breaths/min in different experiment settings.
ACKNOWLEDGMENTS
The authors would like to express their special appreciation to the 22 volunteers for participating in our
experiments. The study was approved by the Medical Ethics Committee (MEC) of NPU. This work was
supported in part by a grant from National Natural Science Foundation of China (No. 61332013), Chinese
Scholarship Council Program, National Natural Science Foundation of China (No. 61702437), Hong Kong ECS
under Grant PolyU 252053/15E, and Hong Kong PolyU under Grant G-YBMT.
REFERENCES
[1]
Adib Fadel, Hongzi Mao, Zachary Kabelac, Dina Katabi, and Robert C Miller. 2015. Smart homes that monitor breathing and heart
rate. In Proceedings of the 33rd Annual ACM Conference on Human Factors in Computing Systems. ACM, 837846.
[2]
Healthcare. 2012. F. M. Market for embedded health monitoring-gadgets to hit 170M devices by 2017. http://www.fiercemobilehealth
care.com/story/market-embedded-health-monitoring-gadgets-hit-170m\-devices-2017/2012-08-03. (2012).
[3]
Michelle A Cretikos, Rinaldo Bellomo, Ken Hillman, Jack Chen, Simon Finfer and Arthas Flabouris. 2008. Respiratory rate: the
neglected vital sign. Medical Journal of Australia 188, 11 (2008): 657.
[4]
J. Brian North and Sheila Jennett. Abnormal breathing patterns associated with acute brain damage. 1974. Archives of neurology 31, 5
(1974), 338-344.
[5]
Yuan, George, Nicole A. Drost, and R. Andrew McIvor. 2013. Respiratory rate and breathing pattern. McMaster University Medical
Journal 10, 1 (2013), 23-25.
[6]
Cooke, Jana R., and Sonia Ancoli-Israel. Normal and abnormal sleep in the elderly. 2011. Handbook of clinical neurology/edited by PJ
Vinken and GW Bruyn 98 (2011), 653.
[7]
Norman, Daniel, and José S. Loredo. Obstructive sleep apnea in older adults. 2008. Clinics in geriatric medicine 24.1 (2008), 151-165.
[8]
Lee-Chiong, L. Teofilo. 2003. Monitoring respiration during sleep. Clinics in chest medicine 24, 2 (2003), 297-306.
[9]
R. Madeline, MPH. Vann. 2015. The 15 Most Common Health Concerns for Seniors. URL: http://goo.gl/EQn2fn, 2015
[10]
J. N. Wilkinson, and V. U. Thanawala. 2009. Thoracic impedance monitoring of respiratory rate during sedation – is it safe?.
Anaesthesia, 64 (2009), 455–456.
[11]
Jaffe, B. Michael. 2008. Infrared measurement of carbon dioxide in the human breath:“breathe-through devices from Tyndall to the
present day. Anesthesia & Analgesia 107, 3 (2008), 890-904.
[12]
Rita Paradiso. 2003. Wearable health care system for vital signs monitoring. In Information Technology Applications in Biomedicine,
2003. 4th International IEEE EMBS Special Topic Conference on. IEEE, 283–286.
[13]
Shoko Nukaya, Toshihiro Shino, Yosuke Kurihara, Kajiro Watanabe, Hiroshi Tanaka. 2012. Noninvasive bed sensing of human
biosignals via piezoceramic devices sandwiched between the floor and bed. IEEE Sensors journal 12, 3 (2012), 431-438.
[14]
Hulya Gokalp and Malcolm Clarke. 2013. Monitoring activities of daily living of the elderly and the potential for its use in telecare and
telehealth: a review. TELEMEDICINE and e-HEALTH 19, 12 (2013), 910–923.
[15]
Jochen Penne, Christian Schaller, Joachim Hornegger, and Torsten Kuwert. 2008. Robust real-time 3D respiratory motion detection
using time-of-flight cameras. International Journal of Computer Assisted Radiology and Surgery 3, 5 (2008), 427–431.
[16]
T Kondo, T Uhlig, P Pemberton, and PD Sly. 1997. Laser monitoring of chest wall displacement. European Respiratory Journal 10, 8
(1997), 1865–1869.
[17]
M Nowogrodzki, DD Mawhinney, and HF Milgazo. 1984. Non-invasive microwave instruments for the measurement of respiration
and heart rates. NAECON 1984 (1984), 958–960.
[18]
Svetha Venkatesh, Christopher R Anderson, Natalia V Rivera, and R Michael Buehrer. 2005. Implementation and analysis of
respiration-rate estimation using impulse-based UWB. In Military Communications Conference, 2005. MILCOM 2005. IEEE. IEEE,
3314–3320.
[19]
Fadel Adib, Hongzi Mao, Zachary Kabelac, Dina Katabi, and Robert C Miller. 2015. Smart homes that monitor breathing and heart
rate. In Proceedings of the 33rd Annual ACM Conference on Human Factors in Computing Systems. ACM, 837–846.
[20]
Ruth Ravichandran, Elliot Saba, Ke-Yu Chen, Mayank Goel, Sidhant Gupta, and Shwetak N Patel. 2015. WiBreathe: Estimating
respiration rate using wireless signals in natural settings in the home. In Pervasive Computing and Communications (PerCom), 2015
IEEE International Conference on. IEEE, 131–139.
[21]
Fadel Adib, Zachary Kabelac, Dina Katabi, Robert C. Miller. 2013. 3D Tracking via Body Radio Reflections. In Proceedings of the
170:20 T. Wang et al.
Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies, Vol. 1, No. 4, Article 170. Publication date:
December 2017.
11th USENIX Conference on Networked Systems Design and Implementation. USENIX Association, 317-329.
[22]
Hao Wang, Daqing Zhang, Junyi Ma, Yasha Wang, Yuxiang Wang, Dan WU, Tao Gu, Bing Xie. 2016. Human respiration detection
with commodity wifi devices: do user location and body orientation matter?. In Proceedings of the 2016 ACM International Joint
Conference on Pervasive and Ubiquitous Computing. ACM, 25-36.
[23]
Jian Liu, Yan Wang, Yingying Chen, Jie Yang, Xu Chen, and Jerry Cheng. 2015. Tracking Vital Signs During Sleep Leveraging Off-
the-shelf WiFi. In Proceedings of the 16th ACM International Symposium on Mobile Ad Hoc Networking and Computing. ACM, 267–
276.
[24]
Xuefeng Liu, Jiannong Cao, Shaojie Tang, and Jiaqi Wen. 2014. Wi-Sleep: Contactless sleep monitoring via WiFi signals. In Real-
Time Systems Symposium (RTSS), 2014 IEEE. IEEE, 346–355.
[25]
Xuefeng Liu, Jiannong Cao, Shaojie Tang, Jiaqi Wen, and Peng Guo. 2016a. Contactless Respiration Monitoring via WiFi Signals.
Mobile Computing, IEEE Transactions on (2016).
[26]
Chenshu Wu, Zheng Yang, Zimu Zhou, Xuefeng Liu, Yunhao Liu, and Jiannong Cao. 2015. Non-Invasive Detection of Moving and
Stationary Human With WiFi. Selected Areas in Communications, IEEE Journal on 33, 11 (2015), 2329–2342.
[27]
Heba Abdelnasser, Khaled A Harras, and Moustafa Youssef. 2015. Ubibreathe: A ubiquitous non-invasive wifi-based breathing
estimator. arXiv preprint arXiv:1505.02388 (2015).
[28]
Ossi Kaltiokallio, Huseyin Yigitler, Riku Jantti, and Neal Patwari. 2014. Non-invasive respiration rate monitoring using a single COTS
TX-RX pair. In Information Processing in Sensor Networks, IPSN-14 Proceedings of the 13th International Symposium on. IEEE, 59
69.
[29]
Neal Patwari, Lara Brewer, Quinn Tate, Ossi Kaltiokallio, and Maurizio Bocca. 2014a. Breathfinding: A wireless network that
monitors and locates breathing in a home. Selected Topics in Signal Processing, IEEE Journal of 8, 1 (2014), 30–42.
[30]
Neal Patwari, James Wilson, Sundaram Ananthanarayanan, Sneha Kumar Kasera, and Dwayne R Westenskow. 2014b. Monitoring
breathing via signal strength in wireless networks. Mobile Computing, IEEE Transactions on 13, 8 (2014), 1774–1786.
[31]
Philippe Arlotto, Michel Grimaldi, Roomila Naeck and Jean-Marc Ginoux. 2014. An ultrasonic contactless sensor for breathing
monitoring. Sensors 14.8 (2014), 15371-86.
[32]
Rajalakshmi Nandakumar, Shyamnath Gollakota, Nathaniel Watson M.D.. 2015. Contactless Sleep Apnea Detection on Smartphones.
In Proceedings of the 13th Annual International Conference on Mobile Systems, Applications, and Services. ACM, 45-57.
[33]
Carina Barbosa Pereira, Xinchi Yu, Michael Czaplik, Rolf Rossaint, Vladimir Blazek, and Steffen Leonhardt. 2015. Remote
monitoring of breathing dynamics using infrared thermography. Biomedical optics express, 6, 11 (2015), 4378-4394.
[34]
Hao Tian, Guoliang Xing, and Gang Zhou. 2013. iSleep: unobtrusive sleep quality monitoring using smartphones. In Proceedings of
the 11th ACM Conference on Embedded Networked Sensor Systems. ACM, 1-14.
[35]
Chunyi Peng, Guobin Shen, Yongguang Zhang, Yanlin Li, Kun Tan (2007). BeepBeep:a high accuracy acoustic ranging system using
COTS mobile devices. In Proceedings of the 5th international conference on Embedded networked sensor systems. ACM,1-14
[36]
Tian Hao, Guoliang Xing, Gang Zhou. 2015. RunBuddy: A Smartphone system for running rhythm monitoring. In Proceedings of the
2015 ACM International Joint Conference on Pervasive and Ubiquitous Computing. ACM, 133-144
[37]
A. G. Stove. (1992). Linear FMCW radar techniques. Radar & Signal Processing Iee Proceedings F, 139, 5(1992):343-350.
[38]
Oppenheim A V, Willsky A S, Nawab S H. 1996. Signals & systems (2nd ed.) Prentice-Hall, Inc. 1996.
[39]
S. Suleymanov. (2016). Design and Implementation of an FMCW Radar Signal Processing Module for Automotive Applications
(Master's thesis, University of Twente).
[40] WebMD. http://www.webmd.com/sleep-disorders/guide/sleep-disorders-symptoms-types.
Received: May 2017; revised: September 2017; accepted: October 2017.
... Wireless sensing has attracted significant amounts of attentions in recent years and various wireless signals have been explored for contactless sensing, including WiFi [12,53], RFID [54,56], FMCW radar [11,13], ultrasound [52,55], 60GHz radar [25,39], UWB [51], and visible light [30,49]. Many new applications have been enabled, such as localization [26], contactless activity sensing [44], respiration monitoring [13], and keystroke detection [14]. ...
... Fullbreath [62] TagBreath [56] Vital-Radio [13] C-FMCW [55] Doppler Radar [ In this work, we find that chirp-based LoRa signal presents us with great opportunities for long-range sensing. We first develop the LoRa signal propagation model to capture the relationship between target movement and signal variation. ...
... We compare the proposed LoRa sensing system with the state-of-the-art methods based on Wi-Fi, FMCW and Ultrasound. We implement three stateof-the-art sensing systems based on Wi-Fi [62], FMCW radar [5] and Ultrasound [55] and compare them with the proposed LoRa sensing system in three aspects: 1) sensing range; 2) through-wall sensing capability; and 3) accuracy. Experiment setup: As shown in Fig. 26a, we implement the WiFi-based system proposed in [62] with one Wi-Fi transceiver pair. ...
Article
Full-text available
Wireless signals have been extensively utilized for contactless sensing in the past few years. Due to the intrinsic nature of employing the weak target-reflected signal for sensing, the sensing range is limited. For instance, WiFi and RFID can achieve 3-6 meter sensing range while acoustic-based sensing is limited to less than one meter. In this work, we identify exciting sensing opportunities with LoRa, which is the new long-range communication technology designed for IoT communication. We explore the sensing capability of LoRa, both theoretically and experimentally. We develop the sensing model to characterize the relationship between target movement and signal variation, and propose novel techniques to increase LoRa sensing range to over 25 meters for human respiration sensing. We further build a prototype system which is capable of sensing both coarse-grained and fine-grained human activities. Experimental results show that (1) human respiration can still be sensed when the target is 25 meters away from the LoRa devices, and 15 meters away with a wall in between; and (2) human walking (both displacement and direction) can be tracked accurately even when the target is 30 meters away from the LoRa transceiver pair.
... Wireless sensing has attracted significant amounts of attentions in recent years and various wireless signals have been explored for contactless sensing, including WiFi [12,53], RFID [54,56], FMCW radar [11,13], ultrasound [52,55], 60GHz radar [25,39], UWB [51], and visible light [30,49]. Many new applications have been enabled, such as localization [26], contactless activity sensing [44], respiration monitoring [13], and keystroke detection [14]. ...
... Fullbreath [62] TagBreath [56] Vital-Radio [13] C-FMCW [55] Doppler Radar [ In this work, we find that chirp-based LoRa signal presents us with great opportunities for long-range sensing. We first develop the LoRa signal propagation model to capture the relationship between target movement and signal variation. ...
... We compare the proposed LoRa sensing system with the state-of-the-art methods based on Wi-Fi, FMCW and Ultrasound. We implement three stateof-the-art sensing systems based on Wi-Fi [62], FMCW radar [5] and Ultrasound [55] and compare them with the proposed LoRa sensing system in three aspects: 1) sensing range; 2) through-wall sensing capability; and 3) accuracy. Experiment setup: As shown in Fig. 26a, we implement the WiFi-based system proposed in [62] with one Wi-Fi transceiver pair. ...
Article
Full-text available
Wireless sensing received a great amount of attention in recent years and various wireless technologies have been exploited for sensing, including WiFi [1], RFID [2], ultrasound [3], 60 GHz mmWave [4] and visible light [5]. The key advantage of wireless sensing over traditional sensing is that the target does not need to be equipped with any sensor(s) and the wireless signal itself is being used for sensing. Exciting new applications have been enabled, such as passive localization [6] and contactless human activity sensing [7]. While promising in many aspects, one key limitation of current wireless sensing techniques is the very small sensing range. This is because while both direct path and reflection path signals are used for communication, only the weak target-reflection signals can be used for sensing. Take Wi-Fi as an example: the communication range can reach 20 to 50 meters indoors but its sensing range is merely 4 to 8 meters. This small range further limits the through-wall sensing capability of Wi-Fi. On the other hand, many applications do require long-range and through-wall sensing capability. In a fire rescue scenario, the sensing device cannot be placed close to the building, and the long-range through-wall sensing capabilities are critical for detecting people deep inside the building. Table I summarizes the sensing range of existing wireless technologies. We can see that long-range through-wall sensing is still missing with wireless sensing.
... Doppler radar based [18,19], WiFi based [20,21]) and the micro-signals of mechanical waves (e.g. acoustic or sonar based [22,23]) have been used for contactless respiration measurement. Although these are different from optical based measurement and mostly discussed in the wireless sensing community, we foresee a potential of fusing different contactless sensing modalities and a cross-fertilization between two communities (i.e. computer vision and wireless sensing) leading towards multi-modal contactless health monitoring. ...
Article
Full-text available
Objective: Measuring the respiratory signal from a video based on body motion has been proposed and recently matured in products for contactless health monitoring. The core algorithm for this application is the measurement of tiny chest/abdominal motions induced by respiration (i.e. capturing sub-pixel displacement caused by subtle motion between subsequent video frames), and the fundamental challenge is motion sensitivity. Though prior art reported on the validation with real human subjects, there is no thorough or rigorous benchmark to quantify the sensitivities and boundary conditions of motion-based core respiratory algorithms. Approach: A set-up was designed with a fully-controllable physical phantom to investigate the essence of core algorithms, together with a mathematical model incorporating two motion estimation strategies and three spatial representations, leading to six algorithmic combinations for respiratory signal extraction. Their promises and limitations are discussed and clarified through the phantom benchmark. Main results: With the variation of phantom motion intensity between 0.5 mm and 8 mm, the recommended approach obtains an average precision, recall, coverage and MAE of 88.1%, 91.8%, 95.5% and 2.1bpm in the day-light condition, and 81.7%, 90.0%, 93.9% and 4.4 bpm in the night condition. Significance: The insights gained in this paper are intended to improve the understanding and applications of camera-based respiration measurement in health monitoring. The limitations of this study stem from the used physical phantom that does not consider human factors like body shape, sleeping posture, respiratory diseases, etc., and the investigated scenario is focused on sleep monitoring, not including scenarios with a sitting or standing patient like in clinical ward and triage.
... Many of these efforts are geared towards utilizing the microphone sensor but for sensing human respiration [62][63][64]. For example, Wang et al. developed a respiration monitoring system by sensing audio signal changes during respiration [65]. Nam et al. detected nasal breath sound recordings from a microphone built-in smartphones to estimate respiration rate [66]. ...
Preprint
Full-text available
Proper indoor ventilation through buildings' heating, ventilation, and air conditioning (HVAC) systems has become an increasing public health concern that significantly impacts individuals' health and safety at home, work, and school. While much work has progressed in providing energy-efficient and user comfort for HVAC systems through IoT devices and mobile-sensing approaches, ventilation is an aspect that has received lesser attention despite its importance. With a motivation to monitor airflow from building ventilation systems through commodity sensing devices, we present FlowSense, a machine learning-based algorithm to predict airflow rate from sensed audio data in indoor spaces. Our ML technique can predict the state of an air vent-whether it is on or off-as well as the rate of air flowing through active vents. By exploiting a low-pass filter to obtain low-frequency audio signals, we put together a privacy-preserving pipeline that leverages a silence detection algorithm to only sense for sounds of air from HVAC air vent when no human speech is detected. We also propose the Minimum Persistent Sensing (MPS) as a post-processing algorithm to reduce interference from ambient noise, including ongoing human conversation, office machines, and traffic noises. Together, these techniques ensure user privacy and improve the robustness of FlowSense. We validate our approach yielding over 90% accuracy in predicting vent status and 0.96 MSE in predicting airflow rate when the device is placed within 2.25 meters away from an air vent. Additionally, we demonstrate how our approach as a mobile audio-sensing platform is robust to smartphone models, distance, and orientation. Finally, we evaluate FlowSense privacy-preserving pipeline through a user study and a Google Speech Recognition service, confirming that the audio signals we used as input data are inaudible and inconstructible.
... Acoustic sensing has been increasingly adopted for a wide range of tasks, from gesture recognition to respiratory monitoring, from lip-reading to authentication [5,12,19,35,48,52,60,63]. As speakers and microphones have become prevalent and accessible in people's everyday lives, acoustic sensing has also found traction in being utilized as a ubiquitous human-computer interface. ...
Article
We propose SonicASL, a real-time gesture recognition system that can recognize sign language gestures on the fly, leveraging front-facing microphones and speakers added to commodity earphones worn by someone facing the person making the gestures. In a user study (N=8), we evaluate the recognition performance of various sign language gestures at both the word and sentence levels. Given 42 frequently used individual words and 30 meaningful sentences, SonicASL can achieve an accuracy of 93.8% and 90.6% for word-level and sentence-level recognition, respectively. The proposed system is tested in two real-world scenarios: indoor (apartment, office, and corridor) and outdoor (sidewalk) environments with pedestrians walking nearby. The results show that our system can provide users with an effective gesture recognition tool with high reliability against environmental factors such as ambient noises and nearby pedestrians.
... Acoustic sensing has been widely explored in the wearable and mobile computing community, mostly as a function of geometry approximation (e.g., floor plans [43,49], obstacle estimation [57]), and motion detection including hand gestures [6,8], respiration [56], heart beat [44], mouth movement [31], and object angular speed [30]. ...
Article
With the rapid growth of artificial intelligence and mobile computing, intelligent speech interface has recently become one of the prevalent trends and has already presented huge potentials to the public. To address the privacy leakage issue during the speech interaction or accommodate some special demands, silent speech interfaces have been proposed to enable people's communication without vocalizing their sound (e.g., lip reading, tongue tracking). However, most existing silent speech mechanisms require either background illuminations or additional wearable devices. In this study, we propose the EchoWhisper as a novel user-friendly, smartphone-based silent speech interface. The proposed technique takes advantage of the micro-Doppler effect of the acoustic wave resulting from mouth and tongue movements and assesses the acoustic features of beamformed reflected echoes captured by the dual microphones in the smartphone. Using human subjects who perform a daily conversation task with over 45 different words, our system can achieve a WER (word error rate) of 8.33%, which shows the effectiveness of inferring silent speech content. Moreover, EchoWhisper has also demonstrated its reliability and robustness to a variety of configuration settings and environmental factors, such as smartphone orientations and distances, ambient noises, body motions, and so on.
... Other solutions [24] leverage cameras, but they do not work in non-line-of-sight (NLoS) scenario and also raise privacy concerns. The acoustic-based solutions [34] can achieve a high respiration rate accuracy, however, they are vulnerable against environmental noise and have a limited sensing range (around 1 m). RF-based solutions such as RFID system [18,46] and FMCW radar [2,49] need dedicated hardware, which are relatively expensive. ...
Article
Full-text available
In recent years, we have seen efforts made to simultaneously monitor the respiration of multiple persons based on the channel state information (CSI) retrieved from commodity WiFi devices. Existing approaches mainly rely on spectral analysis of the CSI amplitude to obtain respiration rate information, leading to multiple limitations: (1) spectral analysis works when multiple persons exhibit dramatically different respiration rates, however, it fails to resolve similar rates; (2) spectral analysis can only obtain the average respiration rate over a period of time, and it is unable to capture the detailed rate change over time; (3) they fail to sense the respiration when a target is located at the ``blind spots'' even the target is close to the sensing devices. To overcome these limitations, we propose MultiSense, the first WiFi-based system that can robustly and continuously sense the detailed respiration patterns of multiple persons even they have very similar respiration rates and are physically closely located. The key insight of our solution is that the commodity WiFi hardware nowadays is usually equipped with multiple antennas. Thus, each individual antenna can receive a different mix copy of signals reflected from multiple persons. We successfully prove that the reflected signals are linearly mixed at each antenna and propose to model the multi-person respiration sensing as a blind source separation (BSS) problem. Then, we solve it using independent component analysis (ICA) to separate the mixed signal and obtain the reparation information of each person. Extensive experiments show that with only one pair of transceivers, each equipped with three antennas, MultiSense is able to accurately monitor respiration even in the presence of four persons, with the mean absolute respiration rate error of 0.73 bpm (breaths per minute).
... We use a standard FMCW radio similar to the one used in past work [2,40,49,51]. The radio sweeps the frequencies from 5.4 GHz to 7.2 GHz and transmits at sub-milliwatt power in accordance with the FCC regulations. ...
Article
Full-text available
Monitoring sleep posture is important for avoiding bedsores after surgery, reducing apnea events, tracking the progression of Parkinson's disease, and even alerting epilepsy patients to potentially fatal sleep postures. Today, there is no easy way to track sleep postures. Past work has proposed installing cameras in the bedroom, mounting accelerometers on the subject's chest, or embedding pressure sensors in their bedsheets. Unfortunately, such solutions jeopardize either the privacy of the user or their sleep comfort. In this paper, we introduce BodyCompass, the first RF-based system that provides accurate sleep posture monitoring overnight in the user's own home. BodyCompass works by studying the RF reflections in the environment. It disentangles RF signals that bounced off the subject's body from other multipath signals. It then analyzes those signals via a custom machine learning algorithm to infer the subject's sleep posture. BodyCompass is easily transferable and can apply to new homes and users with minimal effort. We empirically evaluate BodyCompass using over 200 nights of sleep data from 26 subjects in their own homes. Our results show that, given one week, one night, or 16 minutes of labeled data from the subject, BodyCompass's corresponding accuracy is 94%, 87%, and 84%, respectively.
Article
Full-text available
In the era of pervasively connected and sensed Internet of Things, many of our interactions with machines have been shifted from conventional computer keyboards and mouses to hand gestures and writing in the air. While gesture recognition and handwriting recognition have been well studied, many new methods are being investigated to enable pervasive handwriting tracking. Most of the existing handwriting tracking systems either require cameras, handheld sensors or involve dedicated hardware restricting user-convenience and the scale of usage. In this paper, we present mmWrite, the first high-precision passive handwriting tracking system using a single commodity millimeter wave (mmWave) radio. Leveraging the short wavelength and large bandwidth of 60 GHz signals and the radar-like capabilities enabled by the large phased array, mmWrite transforms any flat region into an interactive writing surface that supports handwriting tracking at millimeter accuracy. MmWrite employs an end-to-end pipeline of signal processing to enhance the range and spatial resolution limited by the hardware, boost the coverage, and suppress interference from backgrounds and irrelevant objects. We implement and evaluate mmWrite on a commodity 60 GHz device. Experimental results show that mmWrite can track a finger/pen with a median error of 2.8 mm and thus can reproduce handwritten characters as small as 1 cm × 1 cm, with a coverage of up to 8 m2 supported. With minimal infrastructure needed, mmWrite promises ubiquitous handwriting tracking for new applications in the field of human-computer interactions.
Article
Full-text available
Vital sign monitoring is a common practice amongst medical professionals, and plays a key role in patient care and clinical diagnosis. Traditionally, dedicated equipment is employed to monitor these vital signs. For example, electrocardiograms (ECG) with 3-12 electrodes are attached to the target chest for heartbeat monitoring. In the last few years, wireless sensing becomes a hot research topic and wireless signal itself is utilized for sensing purposes without requiring the target to wear any sensors. The contact-free nature of wireless sensing makes it particularly appealing in current COVID-19 pandemic. Recently, promising progress has been achieved and the sensing granularity has been pushed to millimeter level, fine enough to monitor respiration which causes a chest displacement of 5 mm. While a great success with respiration monitoring, it is still very challenging to monitor heartbeat due to the extremely subtle chest displacement (0.1 - 0.5 mm) - smaller than 10% of that caused by respiration. What makes it worse is that the tiny heartbeat-caused chest displacement is buried inside the respiration-caused displacement. In this paper, we show the feasibility of employing the popular smart speakers (e.g., Amazon Echo) to monitor an individual's heartbeats in a contact-free manner. To extract the submillimeter heartbeat motion in the presence of other interference movements, a series of novel signal processing schemes are employed. We successfully prototype the first real-time heartbeat monitoring system using a commodity smart speaker. Experiment results show that the proposed system can monitor a target's heartbeat accurately, achieving a median heart rate estimation error of 0.75 beat per minute (bpm), and a median heartbeat interval estimation error of 13.28 ms (less than 1.8%), outperforming even some popular commodity products available on the market.
Conference Paper
Full-text available
Tracking human vital signs of breathing and heart rates during sleep is important as it can help to assess the general physical health of a person and provide useful clues for diagnosing possible diseases. Traditional approaches (e.g., Polysomnography (PSG)) are limited to clinic usage. Recent radio frequency (RF) based approaches require specialized devices or dedicated wireless sensors and are only able to track breathing rate. In this work, we propose to track the vital signs of both breathing rate and heart rate during sleep by using off-the-shelf WiFi without any wearable or dedicated devices. Our system re-uses existing WiFi network and exploits the fine-grained channel information to capture the minute movements caused by breathing and heart beats. Our system thus has the potential to be widely deployed and perform continuous long-term monitoring. The developed algorithm makes use of the channel information in both time and frequency domain to estimate breathing and heart rates, and it works well when either individual or two persons are in bed. Our extensive experiments demonstrate that our system can accurately capture vital signs during sleep under realistic settings, and achieve comparable or even better performance comparing to traditional and existing approaches, which is a strong indication of providing non-invasive, continuous fine-grained vital signs monitoring without any additional cost.
Article
Full-text available
Non-invasive human sensing based on radio signals has attracted numerous research interests and fostered a broad range of innovative applications of localization, gesture recognition, smart health-care, etc, for which a primary primitive is to detect human presence. Previous works have studied to detect moving humans via signal variations caused by human movements. For stationary people, however, existing approaches often employ a prerequisite scenario-tailored calibration of channel profile in human-free environments. Based on in-depth understanding of human motion induced signal attenuation reflected by PHY layer channel state information (CSI), we propose DeMan, a unified scheme for non-invasive detection of moving and stationary human on commodity WiFi devices. DeMan takes advantages of both amplitude and phase information of CSI to detect moving targets. In addition, DeMan considers human breathing as an intrinsic indicator of stationary human presence and adopts sophisticated mechanisms to detect particular signal patterns caused by minute chest motions, which could be destroyed by significant whole-body motion or hidden by environmental noises. By doing this, DeMan is capable of simultaneously detecting moving and stationary people with only a small number of prior measurements for model parameter determination, yet without the cumbersome scenario-specific calibration. Extensive experimental evaluation in typical indoor environments validates the great performance of DeMan in case of various human poses and locations and diverse channel conditions. Particularly, DeMan provides detection rate of around 95% for both moving and stationary people, while identifies human-free scenarios by 96%, all of which outperforms existing methods by about 30%.
Article
Full-text available
The monitoring of breathing activity of a human during a long period has multiple fundamental applications in medicine. In breathing sleep disorders such as apnea, the diagnosis is based on events during which the person stops breathing for several periods during sleep. In polysomnography, the standard of sleep disordered breathing analysis, chest movement and airflow are used to monitor the respiratory activity. However, this method has serious drawbacks. Indeed, as the subject should sleep overnight in a laboratory and because of sensors being in direct contact with him, artifacts modifying sleep quality are often observed. This work investigates an analysis of the viability of an ultrasonic device to quantify the breathing activity, without contact and without any perception by the subject. Based on a low power ultrasonic active source and transducer, the device measures the frequency shift produced by the velocity difference between the exhaled air flow and the ambient environment, i.e. the Doppler effect. After acquisition and digitization, a specific signal processing is applied to separate the effects of breath from those due to subject movements from the Doppler signal. The distance between the source and the sensor, about 50 cm, and the use of ultrasound frequency well above audible frequencies, 40 kHz, allow monitoring the breathing activity without any perception by the subject, and therefore without any modification of the sleep quality which is very important for sleep disorders diagnostic applications. This work is patented (patent pending 2013-7-31 number FR.13/57569).
Conference Paper
Recent research has demonstrated the feasibility of detecting human respiration rate non-intrusively leveraging commodity WiFi devices. However, is it always possible to sense human respiration no matter where the subject stays and faces? What affects human respiration sensing and what's the theory behind? In this paper, we first introduce the Fresnel model in free space, then verify the Fresnel model for WiFi radio propagation in indoor environment. Leveraging the Fresnel model and WiFi radio propagation properties derived, we investigate the impact of human respiration on the receiving RF signals and develop the theory to relate one's breathing depth, location and orientation to the detectability of respiration. With the developed theory, not only when and why human respiration is detectable using WiFi devices become clear, it also sheds lights on understanding the physical limit and foundation of WiFi-based sensing systems. Intensive evaluations validate the developed theory and case studies demonstrate how to apply the theory to the respiration monitoring system design.
Conference Paper
As one of the most popular exercises, running is accomplished through a tight cooperation between the respiratory and locomotor systems. Research has suggested that a proper running rhythm -- the coordination between breathing and strides -- helps improve exercise efficiency and postpone fatigue. This paper presents RunBuddy -- the first smartphone-based system for continuous running rhythm monitoring. RunBuddy is designed to be a convenient and unobtrusive exercise feedback system, and only utilizes commodity devices including smartphone and Bluetooth headset. A key challenge in designing RunBuddy is that the sound of breathing typically has very low intensity and is susceptible to interference. To reliably measure running rhythm, we propose a novel approach that integrates ambient sensing based on accelerometer and microphone, and a physiological model called Locomotor Respiratory Coupling (LRC), which indicates possible ratios between the stride and breathing frequencies. We evaluate RunBuddy through experiments involving 13 subjects and 39 runs. Our results show that, by leveraging the LRC model, RunBuddy correctly measures the running rhythm for indoor/outdoor running 92:7% of the time. Moreover, RunBuddy also provides detailed physiological profile of running that can help users better understand their running process and improve exercise self-efficacy.
Conference Paper
We present a contactless solution for detecting sleep apnea events on smartphones. To achieve this, we introduce a novel system that monitors the minute chest and abdomen movements caused by breathing on smartphones. Our system works with the phone away from the subject and can simultaneously identify and track the fine-grained breathing movements from multiple subjects. We do this by transforming the phone into an active sonar system that emits frequency-modulated sound signals and listens to their reflections; our design monitors the minute changes to these reflections to extract the chest movements. Results from a home bedroom environment shows that our design operates efficiently at distances of up to a meter and works even with the subject under a blanket. Building on the above system, we develop algorithms that identify various sleep apnea events including obstructive apnea, central apnea, and hypopnea from the sonar reflections. We deploy our system at the UW Medicine Sleep Center at Harborview and perform a clinical study with 37 patients for a total of 296 hours. Our study demonstrates that the number of respiratory events identified by our system is highly correlated with the ground truth and has a correlation coefficient of 0.9957, 0.9860, and 0.9533 for central apnea, obstructive apnea and hypopnea respectively. Furthermore, the average error in computing of rate of apnea and hypopnea events is as low as 1.9 events/hr.
Conference Paper
The evolution of ubiquitous sensing technologies has led to intelligent environments that can monitor and react to our daily activities, such as adapting our heating and cooling systems, responding to our gestures, and monitoring our elderly. In this paper, we ask whether it is possible for smart environments to monitor our vital signs remotely, without instrumenting our bodies. We introduce Vital-Radio, a wireless sensing technology that monitors breathing and heart rate without body contact. Vital-Radio exploits the fact that wireless signals are affected by motion in the environment, including chest movements due to inhaling and exhaling and skin vibrations due to heartbeats. We describe the operation of Vital-Radio and demonstrate through a user study that it can track users' breathing and heart rates with a median accuracy of 99%, even when users are 8~meters away from the device, or in a different room. Furthermore, it can monitor the vital signs of multiple people simultaneously. We envision that Vital-Radio can enable smart homes that monitor people's vital signs without body instrumentation, and actively contribute to their inhabitants' well-being.
Article
An atypical or irregular respiratory frequency is considered to be one of the earliest markers of physiological distress. In addition, monitoring of this vital parameter plays a major role in diagnosis of respiratory disorders, as well as in early detection of sudden infant death syndrome. Nevertheless, the current measurement modalities require attachment of sensors to the patient's body, leading to discomfort and stress. The current paper presents a new robust algorithm to remotely monitor breathing rate (BR) by using thermal imaging. This approach permits to detect and to track the region of interest (nose) as well as to estimate BR. In order to study the performance of the algorithm, and its robustness against motion and breathing disorders, three different thermal recordings of 11 healthy volunteers were acquired (sequence 1: normal breathing; sequence 2: normal breathing plus arbitrary head movements; and sequence 3: sequence of specific breathing patterns). Thoracic effort (piezoplethysmography) served as "gold standard" for validation of our results. An excellent agreement between estimated BR and ground truth was achieved. Whereas the mean correlation for sequence 1-3 were 0.968, 0.940 and 0.974, the mean absolute BR errors reached 0.33, 0.55 and 0.96 bpm (breaths per minute), respectively. In brief, this work demonstrates that infrared thermography is a promising, clinically relevant alternative for the currently available measuring modalities due to its performance and diverse remarkable advantages.
Article
Is it possible to leverage WiFi signals collected in bedrooms to monitor a person's sleep? In this paper, we show that with off-the-shelf WiFi devices, fine-grained sleep information like a person's respiration, sleeping postures and rollovers can be successfully extracted. We do this by introducing Wi-Sleep, the first sleep monitoring system based on WiFi signals. Wi-Sleep adopts off-the-shelf WiFi devices to continuously collect the fine-grained wireless channel state information (CSI) around a person. From the CSI, Wi-Sleep extracts rhythmic patterns associated with respiration and abrupt changes due to the body movement. Compared to existing sleep monitoring systems that usually require special devices attached to human body (i.e. Probes, head belt, and wrist band), Wi-Sleep is completely contact less. In addition, different from many vision-based sleep monitoring systems, Wi-Sleep is robust to low-light environments and does not raise privacy concerns. Preliminary testing results show that the Wi-Sleep can reliably track a person's respiration and sleeping postures in different conditions.