ArticlePDF Available

Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses

Authors:

Abstract

Purpose To compare the effect of spherical aberration on optical quality in eyes with two different aspherical intraocular lenses. Methods 120 eyes of 60 patients underwent phacoemulsification. In patients' eyes, an aberration-free IOL (Aspira-aA; Human Optics) or an aberration-correcting aspherical IOL (Tecnis ZCB00; Abott Medical Optics) was randomly implanted. After surgery, contrast sensitivity and wavefront measurements as well as tilt and decentration measurements were performed. Results Contrast sensitivity was significantly higher in eyes with Aspira lens under mesopic conditions with 12 cycles per degree (CPD) and under photopic conditions with 18 CPD (p = 0.02). Wavefront measurements showed a higher total spherical aberration with a minimal pupil size of 4 mm in the Aspira group (0.05 ± 0.03) than in the Tecnis group (0.03 ± 0.02) (p = 0.001). Strehl ratio was higher in eyes with Tecnis (0.28 ± 0.17) with a minimal pupil size larger than 5 mm than that with Aspira (0.16 ± 0.14) (p = 0.04). In pupils with a minimum diameter of 4 mm spherical aberration had a significant effect on Strehl ratio, but not in pupils with a diameter less than 4 mm. Conclusions Optical quality was better in eyes with the aberration-correcting Tecnis IOL when pupils were large. In contrast, this could not be shown in eyes with pupils under 4 mm or larger. This trial is registered with Clinicaltrials.gov NCT03224728.
Clinical Study
Effect of Spherical Aberration on the Optical Quality after
Implantation of Two Different Aspherical Intraocular Lenses
Michael Lasta,
1,2
Kata Miháltz,
1,2
Illés Kovács,
3
and Pia Veronika Vécsei-Marlovits
1,2
1
Department of Ophthalmology, Hospital Hietzing, Vienna, Austria
2
Karl Landsteiner Institute of Process Optimization and QM in Cataract Surgery, Vienna, Austria
3
Department of Ophthalmology, Semmelweis University, Budapest, Hungary
Correspondence should be addressed to Kata Miháltz; kata.mihaltz@wienkav.at
Received 26 February 2017; Revised 21 May 2017; Accepted 12 June 2017; Published 16 August 2017
Academic Editor: Tamer A. Macky
Copyright © 2017 Michael Lasta et al. This is an open access article distributed under the Creative Commons Attribution License,
which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Purpose. To compare the eect of spherical aberration on optical quality in eyes with two dierent aspherical intraocular lenses.
Methods. 120 eyes of 60 patients underwent phacoemulsication. In patientseyes, an aberration-free IOL (Aspira-aA; Human
Optics) or an aberration-correcting aspherical IOL (Tecnis ZCB00; Abott Medical Optics) was randomly implanted. After
surgery, contrast sensitivity and wavefront measurements as well as tilt and decentration measurements were performed. Results.
Contrast sensitivity was signicantly higher in eyes with Aspira lens under mesopic conditions with 12 cycles per degree (CPD)
and under photopic conditions with 18 CPD (p=002). Wavefront measurements showed a higher total spherical aberration
with a minimal pupil size of 4 mm in the Aspira group (0.05 ±0.03) than in the Tecnis group (0.03 ±0.02) (p=0001). Strehl
ratio was higher in eyes with Tecnis (0.28 ±0.17) with a minimal pupil size larger than 5 mm than that with Aspira (0.16
±0.14) (p=004). In pupils with a minimum diameter of 4 mm spherical aberration had a signicant eect on Strehl ratio, but
not in pupils with a diameter less than 4 mm. Conclusions. Optical quality was better in eyes with the aberration-correcting
Tecnis IOL when pupils were large. In contrast, this could not be shown in eyes with pupils under 4 mm or larger. This trial is
registered with Clinicaltrials.gov NCT03224728.
1. Introduction
In recent years, cataract surgery has become a highly precise
and safe surgical procedure due to phacoemulsication tech-
nique and advanced intraocular lens (IOL) designs. To face
increasing demands on visual and refractive outcome, eorts
have been made to further improve IOL designs. Today, it is
widely recognized that implantation of spherical IOLs leads
to increased spherical aberrations (SA), therefore decreasing
retinal image quality [1]. Aspherical IOLs have been devel-
oped in order to overcome this issue. Aberration-correcting
aspherical IOLs are aiming to compensate for the corneal
spherical aberration of the eye. It is known that aspherical
lenses signicantly decrease higher-order aberrations (HOA)
postoperatively [1]. A number of studies have shown that
aspherical IOLs perform better compared to conventional
spherical ones in terms of contrast sensitivity (CS), particu-
larly under mesopic conditions [2, 3]. Furthermore it is
known that a precise position of an aspherical IOL is essen-
tial for perfect optical quality [4]. IOL displacement such as
tilt and decentration could not only impair its aberration-
correcting eect but may even induce additional optical
aberrations [4, 5].
The question whether correction of the total SA of the eye
or some residual SA leads to better visual outcome and better
optical quality remains to be elucidated. It is known that
healthy young eyes with good visual acuity have some posi-
tive ocular SA [6]. However, results of other studies indicate
that a postoperative SA of zero should be aimed in order to
optimize visual outcome [3, 7].
Aberration-free aspherical IOLs follow another concept
of aspherical IOL design. These implants have no inherent
Hindawi
Journal of Ophthalmology
Volume 2017, Article ID 8039719, 6 pages
https://doi.org/10.1155/2017/8039719
SA; thus, they do not introduce additional SA into the
optical system of the eye. Some studies indicate that
aberration-free aspheric IOLs may result in better optical
quality compared to aberration-correcting aspheric IOL-
seven in case of decentration and tilt of the IOL [8].
Our study aimed to evaluate the optical quality and poten-
tial benets on visual outcome of an aberration-free
aspherical IOL compared to those of an aspherical IOL
with negative SA.
Optical quality is a subjective construct that can only be
described indirectly by objective metrics such as wavefront
error measurements and visual quality metrics or functional
data like visual acuity and contrast sensitivity. Wavefront
analysis isolates the eect of lower-order aberrations (defo-
cus and astigmatism) and higher-order aberrations as well
as the contribution of individual aberrations on optical qual-
ity. Modulation transfer function (MTF) and Strehl ratio are
well related to the image quality of an optical system, includ-
ing the human eye [9].
In this study, an aberration-free aspherical IOL (Aspira-
aA, Human Optics AG; Erlangen, Germany) was compared
to an aberration-correcting aspherical IOL with negative SA
(Tecnis ZCB00, Abbott Medical Optics; Illinois, USA)
regarding the eect of spherical aberration on visual outcome
and optical quality.
2. Methods and Subjects
This prospective single-center study was conducted at the
Department of Ophthalmology, Hospital Hietzing, Vienna,
Austria, in accordance with the Declaration of Helsinki.
The study received approval from the Ethics Committee of
the City of Vienna, and each subject gave written informed
consent before taking part.
Before inclusion, all patients underwent a full ophthalmic
assessment including slit lamp examination and funduscopy.
Intraocular eye pressure was measured with the Goldmann
applanation tonometry. Optical biometry was performed
using the IOLMaster 500 (Carl Zeiss Meditec AG; Jena,
Germany) for axial length and corneal radii of curvature
measurements. For IOL calculation, SRK®/T, Holladay I,
and Hoer® Q were used, depending on the length of the
patientseyes.
Patients with bilateral age-related cataract and an age
range between 50 and 80 years were included. Eyes with
relevant other ocular pathologies or previous surgeries, a
potential postoperative corrected distance visual acuity
worse than 20/20, and corneal astigmatism of 1.5 diopters
and more were excluded from the study. For patientssubjec-
tive refraction, a plus cylinder design was used. To be able to
compare IOL function, patients had to manifest similar
corneal asphericity.
37 female and 23 male subjects were enrolled in this trial.
They were randomly assigned to implantation of the
aberration-free aspherical Aspira-aA in one eye and the
aberration-correcting aspherical Tecnis ZCB00 in the fellow
eye. Both IOLs are monofocal acrylic one-piece lenses with
C-loop haptic design.
The eye with the worse visual acuity was operated rst,
followed by the second eye one week afterwards. All of our
patients have been corrected for emmetropy after surgery.
2.1. Surgical Technique. Surgery was performed in topical
anesthesia using oxybuprocaine eye drops. The self-sealing
2.4 mm incision, capsulorrhexis, phacoemulsication, irriga-
tion and aspiration of cortical material, and injection of
viscoelastic substance into the capsular bag were performed
as standard procedures. The IOL was implanted via an injector
into the capsular bag followed by thorough aspiration of the
viscoelastic substance from the eye.
2.2. Postoperative Assessments. Three months postopera-
tively, the following measurements were carried out in
each subject:
Uncorrected distance visual acuity (UDVA) and cor-
rected distance visual acuity (CDVA) were measured using
the ETDRS chart at a viewing distance of four meters. In
addition, monocular defocus curve measurements were per-
formed with the best distance correction adding glasses in
0.50-diopter increments from +0.5 D to 3.00 diopters.
Photopic and mesopic contrast sensitivity was evaluated
with best refraction using the Optec 6500 contrast sensitivity
tester (Stereo Optical Co.; Chicago, Illinois, USA) under
photopic (85 cd/m
2
) and mesopic (3 cd/m
2
) conditions. This
item uses functional acuity contrast charts (FACT) testing
ve spatial frequencies and nine levels of contrast. The
patient determines the last grating seen for each row (A, B,
C, D, and E) and reports the orientation of the grating: right,
up, or left. The last correct grating seen for each spatial
frequency is plotted on a contrast sensitivity curve.
Thereafter, pupils were dilated using one eye drop
containing 0.5% tropicamide and one eye drop containing
10% phenylephrine.
To assess optical quality characteristics including higher-
order aberrations, wavefront measurements were obtained
using the HOYA iTraceSurgical Workstation (HOYA
Surgical Optics GmbH; Frankfurt, Germany).
The HOYA iTrace Surgical Workstation provides an
objective clinical evaluation of the eyes optical quality.
Performing these measurements with maximum-dilated
pupils, we calculated all relevant parameters with a pupil size
of two, three, four, and ve millimeters.
The following parameters have been calculated:
Wavefront aberrometry describes the optical properties
of the eye in individual Zernike polynomials. Optical quality
can be described by metrics of image quality for point objects
(point spread function (PSF)) and for grating objects (modu-
lation transfer function (MTF)). The iTrace is able to
measure the following image quality metrics: PSF, Strehl
ratio, and MTF.
Strehl ratio is the ratio of the peak height of the PSF
divided by the maximum intensity of PSF in the diraction-
limited perfect eye. Strehl ratio ranges from 0 to 1; the greater
Strehl ratio, the better the quality of vision [10]. By the
iTrace, Strehl ratio is calculated from the retinal PSF. The
MTF is the modulus of the Fourier transform of the PSF. It
2 Journal of Ophthalmology
characterizes the degradation of the image for every spatial
frequency of the object.
Root mean square (RMS) error was calculated by the
iTrace software from Zernike coecients. The total HO
(higher order) is the RMS of higher-order terms (Z3 to Z6).
IOL tilt and decentration were measured with the Visante
anterior segment OCT (Carl Zeiss Meditec AG; Jena,
Germany). Images were obtained in mesopic conditions for
each examination. Anterior single-scan mode was used to
assess pictures. These were obtained in four axes (180 to 0
degrees, 45 to 225 degrees, 315 to 135 degrees, and 270 to
90 degrees). Tilt and decentration of IOLs were calculated
accordingly as previously described by Rosales et al. [11].
2.3. Statistical Analysis. Statistical analyses were performed
with Statistica 12.0 (Statsoft Inc.; Tulsa, OK, USA). The
Shapiro-Wilk Wtest was used to conrm normal distribu-
tion of the variables. Paired-samples t-test was used to
compare means between the eyes of the same subject. This
test allows for comparing within-subject parameters (visual
acuity, keratometric, OCT, and intraocular optical quality
parameters) in the two study groups by taking into
account between-eye correlations by treating data from
the two eyes of the same patient in statistical analyses as
repeated measures.
To determine the eect of intraocular spherical aberra-
tion on visual quality in relation to the pupillary diameter,
multivariable regression analysis using the generalized esti-
mating equation (GEE) model was performed where data
from patients who had measurements in both eyes were sta-
tistically analyzed as repeated measures. The type of the IOL
and corneal spherical aberration were included as con-
founders in these multivariable regression models to adjust
for their eect on total ocular Strehl ratio. Model t was
assessed using the value of the corrected quasilikelihood
under independence model criterion (QICC) with a lower
QICC values indicating a better t to data. In all analyses, a
pvalue less than 0.05 was considered statistically signicant.
3. Results
The age range of the included patients was between 47 and 79
years (mean 69 ±7.8 years). For the three-month follow-up,
data of 47 persons were eligible for evaluation.
Preoperative ndings of the patients are shown in Table 1.
Three months after phacoemulsication, mean corrected
distance visual acuity (CDVA) increased in all patients
(logMAR 0.073 ±0.092 in eyes with Aspira and 0.040 ±0.098
in eyes with Tecnis. There was no signicant dierence
between both IOL groups (p=0 22).
In all patients, postoperative refraction showed a slight
myopic shift. The dierence between planned and manifest
postoperative sphere was found in a majority of the eyes;
however, we could not nd a signicant dierence between
both lenses.
In addition, we measured no signicant dierence in
depth of focus between Aspira and Tecnis IOLs.
Postoperative characteristics are shown in Table 2.
3.1. Contrast Sensitivity. In eyes with the aberration-free
Aspira, higher contrast sensitivity under photopic as well as
under mesopic conditions was measured compared to that
in eyes with Tecnis IOL. However, only one out of 5 measure-
ments under photopic conditions with 18 cycles per degree
(CPD) as well as one measurement under mesopic condi-
tions with 12 CPD presented a signicant dierence between
both IOLs (Table 3).
3.2. Tilt and Decentration. Three months after implantation,
both IOLs showed a slight tendency to decentration in tem-
poral direction; however, no signicant dierence between
lenses was found (Table 4).
Vertical tilt was higher in Aspira than in Tecnis IOL.
However, this nding did not reach the level of signicance.
Table 1: Patientspreoperative characteristics (mean ±SD; n=60).
Aspira-aA Tecnis ZCB00 pvalue
CDVA (logMAR) 0.38 ±0.13 0.39 ±0.14 0.85
Sphere (D) 0.50 ±2.32 0.42 ±2.63 0.43
Cylinder (D) 0.68 ±0.61 0.74 ±0.5 0.40
Axial eye length (mm) 23.40 ±0.99 23.45 ±1.05 0.87
CDVA: corrected distance visual acuity.
Table 2: Patientspostoperative characteristics (mean ±SD; n=60).
Aspira-aA Tecnis ZCB00 pvalue
UDVA (logMAR) 0.14 ±0.13 0.12 ±0.15 0.25
CDVA (logMAR) 0.07 ±0.09 0.04 ±0.09 0.22
Sphere Δ(D) 0.22 ±0.67 0.50 ±0.52 0.29
Depth of focus (D) 0.61 ±0.33 0.65 ±0.29 0.74
Corneal asphericity with
a pupil size of 4 mm 0.19 ±0.24 0.26 ±0.16 0.19
UDVA: uncorrected distance visual acuity; CDVA: corrected distance
visual acuity; sphere Δ:dierence between scheduled and manifest
postoperative sphere.
Table 3: Contrast sensitivity values in the eyes, implanted with
Aspira or Tecnis IOLs. p: Studentst-test on dependent samples
(log; mean ±SD; n=60).
Aspira-aA Tecnis ZCB00 pvalue
Photopic
CPD 1.5 1.54 ±0.20 1.52 ±0.17 0.62
CPD 3 1.78 ±0.20 1.76 ±0.18 0.60
CPD 6 1.72 ±0.27 1.74 ±0.20 0.77
CPD 12 1.27 ±0.49 1.29 ±0.39 0.85
CPD 18 0.52 ±0.51 0.36 ±0.41 0.02
Mesopic
CPD 1.5 1.60 ±0.22 1.60 ±0.20 0.92
CPD 3 1.73 ±0.23 1.69 ±0.24 0.40
CPD 6 1.36 ±0.64 1.49 ±0.33 0.22
CPD 12 0.71 ±0.61 0.48 ±0.58 0.02
CPD 18 0.09 ±0.13 0.16 ±0.09 0.19
CDP: cycles per degree.
3Journal of Ophthalmology
In contrast, horizontal tilt was signicantly lower in Aspira
compared to Tecnis (0.35 ±1.65
°
and 0.67 ±1.42
°
, resp.;
p=002; Table 4).
3.3. Wavefront Measurements. All wavefront measurements
were performed with a pupil diameter of 5 mm. Data with
a pupil size of 2, 3, and 4 mm were calculated with the
software program.
Higher-order root mean square (HORMS) showed no
signicant dierence between the two lenses. From 0.054
±0.042 in Aspira and 0.050 ±0.036 in Tecnis with a pupil size
of 2 mm, values increased to 0.52 ±0.25 and 0.56 ±0.52 with a
maximum-dilated pupil.
In contrast, total spherical aberration (TSA) was signif-
icantly higher in Aspira compared to Tecnis. However,
this was only shown with pupil sizes of minimum 4 mm
(p=0001; Table 5).
Total Strehl ratio was comparable in both lenses with a
pupil size of 2, 3, and 4 mm. However, with a pupil size of
5 mm, we calculated signicantly higher values in eyes with
Tecnis IOL than in eyes with Aspira IOL (0.16 ±0.14 versus
0.08 ±0.17, resp.; p=004). This eect was shown in modula-
tion transfer function as well (0.46 ±0.13 versus 0.39 ±0.15;
p=004). However, these dierences did not reach the level
of signicance after a Bonferroni correction.
Multivariate regression models showed that total spheri-
cal aberration had a signicant negative eect (beta: 1.54,
95% CL: 2.56 to 0.51; p=0004) on total ocular Strehl ratio
when the pupillary diameter was 4 mm, but not when pupil-
lary diameter was less than 4 mm after adjustment for the
eect of IOL type and corneal spherical aberration.
4. Discussion
The surgical technique of phacoemulsication has been
improved continuously. The measure of successful surgery
is more than just an increased visual quality. Patients expect
high optical quality without disturbing aberrations as well as
high contrast sensitivity during night and day.
A various number of spherical and aspherical IOL
designs have been developed aiming to achieve perfect
optical quality.
In this trial we compared the aberration-free aspher-
ical Aspira-aA with the aberration-correcting aspherical
Tecnis ZCB00.
Several authors mentioned that there is increased opti-
cal quality and subjective satisfaction in patients with
aspherical lenses compared to those with spherical ones
[12, 13]. In the clinical trial of Santhiago et al. [12], the
aspherical aberration-free Akreos AO was compared with
spherical Akreos Fit. The authors described signicantly
higher contrast sensitivity and decreased aberrations in the
aspherical lens.
Apart from comparisons between spherical and aspheri-
cal IOLs, a few authors compared dierent aspherical lenses
with each other.
In the work of Rajabi et al. [14], the Aspira-aA was
compared with Akreos AO. In contrast to our ndings, they
mentioned that there was no signicant dierence in higher-
order aberrations between the two lenses. Due to the fact
that they compared two IOLs with similar SA, the major
question (what is the ideal SA regarding optical quality)
could not be answered.
Johansson et al. [15] evaluated Akreos AO and Tecnis
Z9000. Although the Tecnis showed less aberration than
the Akreos, more patients reported better subjective quality
with Akreos IOL. The same IOLs were evaluated in the work
of Baghi et al. [16]. In this trial, Akreos AO causes signi-
cantly more spherical aberration in pupil diameter of 4 and
6 mm. This seems to be in accordance with our ndings,
where Aspira lens causes higher spherical aberrations in
pupil diameters of 4 and 5 mm.
Nochez and coworkers [17] stated that an IOL with a
spherical aberration of zero leads to better optical quality
and even better MTF contrast than aberration-correcting
intraocular lens. However, the best compromise of subjective
depth of contrast and objective contrast sensitivity is reached
with a total spherical aberration between 0.07 μm and
0.10 μm. We found a total spherical aberration between
0.006 μm with a pupil size of 2 mm and 0.11 μm with a pupil
size of 5 mm.
In general a higher spherical aberration decreases Strehl
ratio resulting in lower optical quality [10]. We can reassure
this phenomenon as our multivariable regression models
showed that intraocular spherical aberration had signicant
negative eect on total ocular Strehl ratio. Nevertheless, it
has to be mentioned that this eect applied only when the
pupillary diameter was 4 mm and more. With pupil sizes of
less than 4 mm, adjustment for the eect of IOL type and cor-
neal spherical aberration did not show this eect. We can
assume that spherical aberration inuences Strehl ratio and
optical quality less when pupil sizes are small due to the fact
that small pupils present smaller values of spherical aberra-
tion. In contrast, higher spherical aberration signicantly
inuences optical quality when the pupils are small.
Results of evaluating contrast sensitivity between aspher-
ical lenses dier in various studies. Shentu et al. [18] could
not nd any dierence in contrast sensitivity between the
Table 4: Results of Visante OCT measurements (mean ±SD; n=60).
Aspira-aA Tecnis ZCB00 pvalue
Horizontal tilt (
°
) 0.35 ±1.65 0.67 ±1.42 0.02
Vertical tilt (
°
) 0.63 ±1.76 0.19 ±1.24 0.35
Horizontal decentration (mm) 0.03 ±0.16 0.04 ±0.27 0.94
Vertical decentration (mm) 0.08 ±0.15 0.07 ±0.23 0.96
Table 5: Total ocular higher-order spherical aberrations measured
at dierent pupillary diameters in eyes implanted with Aspira or
with Tecnis IOL (μm; mean ±SD; n=60).
Aspira-aA Tecnis ZCB00 pvalue
2 mm 0.006 ±0.004 0.006 ±0.005 0.88
3 mm 0.02 ±0.01 0.02 ±0.01 0.9
4 mm 0.05 ±0.03 0.03 ±0.02 0.001
5 mm 0.11 ±0.05 0.07 ±0.05 0.001
4 Journal of Ophthalmology
two aspherical lenses. Our results show signicant dierences
in contrast sensitivity, although this was found at 18 CPD
under photopic conditions and at 12 CPD under mesopic
conditions. Hence, it can be assumed that patients might
not notice any distinction in their daily life.
Jia and Li [19] mentioned that a presurgical measure-
ment of corneal spherical aberration should be performed
to determine which type of IOL would lead to optimal optical
quality. A possible solution for this problem would be to
choose the SA of the implanted IOL individually as suggested
by Jia and Li.
Tilt and decentration of IOLs may lead to higher spheri-
cal aberration resulting in decreased optical quality [20]. In
our study, three months postoperatively, a signicantly
higher horizontal tilt of Tecnis lens was measured. In
contrast, both lenses did not dier in vertical tilt as well as
decentration, although a horizontal temporal decentration
were found in both eyes. These ndings comply with the work
of Rosales et al. [11]. They described that IOL decentration
tends to be mirror symmetric between the right and left eye.
Study limitations are that only total aberrations of study
eyes were calculated. Further, in order to obtain precise
wavefront measurements, detailed knowledge about internal
aberrations would be essential, which will be subject of future
investigations. Nevertheless, we aimed to constitute the total
eciency of pseudophakic eyes. Furthermore, we included
patients with similar corneal asphericity in order to compare
both IOLs. Regarding statistics, Bonferroni calculations may
have been more suitable concerning data calculations, which
is considered as a drawback of the current study.
4.1. Conclusion. We can conclude that aberration-free as well
as aberration-correcting aspherical intraocular lenses feature
high optical quality conditions. When pupils are large, eyes
with aberration-correcting Tecnis IOL showed higher modu-
lation transfer function and Strehl ratio values than eyes with
aberration-free Aspira lens. Nevertheless, it has to be claried
that this signicance is not shown after a Bonferroni correc-
tion. Moreover, this benet could not be veried in eyes with
pupils under 4 mm.
Disclosure
This work has been presented as a poster at DOG annual
meeting, Berlin, October 2017.
Conflicts of Interest
The authors declare that they have no conict of interest.
References
[1] R. Yagci, F. Uzun, S. Acer, and I. F. Hepsen, Comparison of
visual quality between aspheric and spherical IOLs,European
Journal of Ophthalmology, vol. 24, pp. 688692, 2014.
[2] A. Assaf and A. Kotb, Ocular aberrations and visual perfor-
mance with an aspheric single-piece intraocular lens: contra-
lateral comparative study,Journal of Cataract and Refractive
Surgery, vol. 36, pp. 15361542, 2010.
[3] S. Ohtani, K. Miyata, T. Samejima, M. Honbou, and T. Oshika,
Intraindividual comparison of aspherical and spherical intra-
ocular lenses of same material and platform,Ophthalmology,
vol. 116, pp. 896901, 2000.
[4] T. Eppig, K. Scholz, A. er, A. Messner, and
A. Langenbucher, Eect of decentration and tilt on the image
quality of aspheric intraocular lens designs in a model eye,
Journal of Cataract and Refractive Surgery, vol. 35, pp. 1091
1100, 2009.
[5] S. K. Choi, J. H. Kim, D. Lee, S. H. Park, N. Maeda, and
K. J. Ma, IOL tilt and decentration,Ophthalmology, vol. 117,
pp. 18611864, 2010.
[6] L. N. Thibos, X. Hong, A. Bradley, and X. Cheng, Statistical
variation of aberration structure and image quality in a normal
population of healthy eyes,Journal of the Optical Society of
America A: Optics, Image Science, and Vision, vol. 19,
pp. 23292349, 2002.
[7] E. A. Tuzcu, K. Erkilic, B. Bulut, and N. Ilhan, Comparing the
eect of two dierent intraocular lenses on optical aberrations
in bilaterally operated eyes for cataract,Journal of Medical
Sciences, vol. 29, pp. 982985, 2013.
[8] S. Pieh, W. Fiala, A. Malz, and W. Stork, In vitro strehl ratios
with spherical, aberration-free, average, and customized
spherical aberration-correcting intraocular lenses,Investiga-
tive Ophthalmology & Visual Science, vol. 50, pp. 12641270,
2009.
[9] L. J. Moreno, D. P. Piñero, J. L. Alío, A. Fimia, and A. B. Plaza,
Double-pass system analysis on the visual outcomes and
optical performance of an apodized diractive multifocal
intraocular lens,Journal of Cataract and Refractive Surgery,
vol. 35, pp. 663671, 2009.
[10] M. Lombardo and G. Lombardo, Wave aberration of human
eyes and new descriptors of image optical quality and visual
performance,Journal of Cataract and Refractive Surgery,
vol. 36, pp. 313331, 2010.
[11] P. Rosales, A. De Castro, I. Jiménez-Alfaro, and S. Marcos,
Intraocular lens alignment from purkinje and Scheimpug
imaging,Clinical & Experimental Optometry, vol. 93,
pp. 400408, 2010.
[12] M. R. Santhiago, M. V. Netto, J. Barreto Jr. et al., Wavefront
analysis, contrast sensitivity, and depth of focus after cataract
surgery with aspherical intraocular lens implantation,
American Journal of Ophthalmology, vol. 149, pp. 383389,
2010.
[13] C. Pérez-Vives, T. Ferrer-Blasco, S. García-Lázaro, C. Albarrán-
Diego, and R. Montés-Micó, Optical quality comparison
between spherical and aspheric toric intraocular lenses,
European Journal of Ophthalmology, vol. 24, pp. 699706, 2014.
[14] M. T. Rajabi, S. Korouji, M. Farjadnia et al., Higher order
aberration comparison between two aspherical intraocular
lenses: MC6125AS and Akreos advanced optics,International
Journal of Ophthalmology, vol. 8, pp. 565568, 2015.
[15] B. Johansson, S. Sundelin, A. Wikberg-Matsson, P. Unsbo, and
A. Behndig, Visual and optical performance of the Akreos
adapt advanced optics and Tecnis Z9000 intraocular lenses:
Swedish multicenter study,Journal of Cataract and Refractive
Surgery, vol. 33, pp. 15651572, 2007.
[16] A. R. Baghi, M. R. Jafarinasab, H. Ziaei, and Z. Rahmani,
Visual outcomes of two aspheric PCIOLs: Tecnis Z9000
versus Akreos AO,Journal of Ophthalmic & Vision Research,
vol. 23, pp. 3236, 2008.
5Journal of Ophthalmology
[17] Y. Nochez, S. Majzoub, and P. J. Pisella, Eect of residual ocu-
lar spherical aberration on objective and subjective quality of
vision in pseudophakic eyes,Journal of Cataract and Refrac-
tive Surgery, vol. 37, pp. 10761081, 2011.
[18] X. Shentu, X. Tang, and K. Yao, Spherical aberration, visual
performance and pseudoaccommodation of eyes implanted
with dierent aspheric intraocular lens,Clinical & Experi-
mental Ophthalmology, vol. 36, pp. 620624, 2008.
[19] L. X. Jia and Z. H. Li, Clinical study of customized aspherical
intraocular lens implants,International Journal of Ophthal-
mology, vol. 7, pp. 816821, 2014.
[20] J. McKelvie, B. McArdle, and C. McGhee, The inuence of
tilt, decentration, and pupil size on the higher-order aberration
prole of aspheric intraocular lenses,Ophthalmology, vol. 118,
pp. 17241731, 2011.
6 Journal of Ophthalmology
... Strengths of our study include the use of real-world data and no patients being excluded due to their comorbidities. Together, these factors provide an authentic prevalence of concomitant ocular pathologies and a preoperative mean CDVA that is more representative of population averages than in other studies [23][24][25]. Our study was also strengthened by the involvement of multiple surgeons and the excellent success rate of ensuring follow-up. ...
Article
Full-text available
Background This study was designed to evaluate visual, refractive and safety outcomes in eyes after they underwent phacoemulsification and implantation of a preloaded monofocal hydrophobic acrylic intraocular lens. Methods This was a single center observational study conducted at Ashford and St Peter’s Hospitals NHS Foundation Trust, United Kingdom. Patients were included if they had cataract extraction with in-the-bag implantation of the EyeCee® One preloaded intraocular lens from August to October 2019. Pre-operative, surgery-related and 2 weeks and 3 months post-operative data was collected. Surgeons at this trust were then asked to complete a feedback form to evaluate their experience of implanting the EyeCee® One. Results One hundred fifty-two eyes were included in the study. Ninety-four (62%) of these eyes had cataract but no concomitant ocular pathology that could potentially affect visual acuity. Three months post-operatively, 98.7% of all eyes had monocular CDVA ≤0.3 logMAR. 100% of the eyes without concomitant ocular pathology achieved this target. The mean CDVA of all eyes in this study improved from 0.43 ± 0.43 logMAR pre-operatively, to 0.05 ± 0.11 logMAR post-operatively (p < 0.05). The mean sphere and spherical equivalent values showed significant improvements (p < 0.05) and (p < 0.05). There were no intraoperative complications and 1.3% of patients reported complications 2 weeks post-operatively. All of the participating surgeons said they would use the EyeCee® One again with 64% providing an overall rating of ‘excellent’ for their experience of implanting this intraocular lens. Conclusions This study indicates excellent post-operative visual acuity and refractive outcomes in eyes after EyeCee® One implantation. This is accompanied with very little risk of intraoperative and post-operative complications.
... However, when the aperture size increased, the AT LARA outperformed the Proming lens (Fig. 2, Table 2), which results from differences in the amount of spherical aberration induced by each model. This impact of spherical aberration on image and visual quality has been reported by many researchers [22][23][24]. Note that the AT LARA features an aberration-neutral design. ...
Article
Full-text available
Background To assess the optical behavior of a new diffractive intraocular lens (IOL) and compare its performance to that of an established extended-depth-of-focus (EDOF) IOL. Methods This study assessed the Proming EDOF Multifocal AM2UX [Eyebright Medical Technology (Beijing) Co., Ltd., China] and the AT LARA 829MP [Carl Zeiss Meditec, Germany]. An experimental set-up with 0.01% fluorescein solution and monochromatic light (532 nm) was used to visualize the IOLs’ ray propagation. In addition, the optical quality of the IOLs was assessed by measuring the modulation transfer function (MTF) values at 50lp/mm and 3.0 and 4.5 mm apertures on the optical bench OptiSpheric® IOL PRO II [Trioptics GmbH, Germany]. Results The ray propagation of the two IOLs showed two distinct foci. Light intensity assessment revealed that both IOLs allocate more energy to primary than secondary focus. At 3.0 mm pupil, the MTF values at 50lp/mm for the primary focus were 0.39 and 0.37, and for the secondary focus, 0.29 and 0.26 for the AT LARA and Proming IOLs, respectively. At 4.5 mm pupil, the single-frequency MTF for the primary focus was 0.51 and 0.24 and for the secondary focus 0.21 and 0.15 for the AT LARA and Proming IOLs, respectively. Conclusions When tested with an aberration-free model cornea under monochromatic conditions, the Proming behaved as a low-add bifocal lens; however, its properties did not differ much from the well-established AT LARA EDOF IOL. The AT LARA outperformed the Proming at low defocus (up to 2D), while the latter demonstrated better image quality in the 2-3D range.
... However, in a study that compared outcomes Strengths of our study include the use of real-world data and no patients being excluded due to their comorbidities. Together, these factors provide an authentic prevalence of concomitant ocular pathologies and a pre-operative mean CDVA that is more representative of population averages than in other studies [19][20][21] ...
Preprint
Full-text available
Background: This study was designed to evaluate visual, refractive and safety outcomes in eyes after they underwent phacoemulsification and implantation of a preloaded monofocal hydrophobic acrylic intraocular lens. Methods: This was a single center prospective study conducted at Ashford and St Peter’s Hospitals NHS Foundation Trust, United Kingdom. Patients were included if they had cataract extraction with in-the-bag implantation of the EyeCee® One preloaded intraocular lens from August to October 2019. Pre-operative, surgery-related and 2 weeks and 3 months post-operative data was collected. Surgeons at this trust were then asked to complete a feedback form to evaluate their experience of implanting the EyeCee® One. Results: 152 eyes were included in the study. 94 (62%) of these eyes had cataract but no concomitant ocular pathology that could potentially affect visual acuity. Three months post-operatively, 98.7% of all eyes had monocular CDVA ≤ 0.3 logMAR. 100% of the eyes without concomitant ocular pathology achieved this target. The mean CDVA of all eyes in this study improved from 0.43 ± 0.43 logMAR pre-operatively, to 0.05 ± 0.11 logMAR post-operatively (p < 0.05). The mean sphere and spherical equivalent values showed significant improvements (p <0.05) and (p < 0.05). There were no intraoperative complications and 1.3% of patients reported complications 2 weeks post-operatively. All of the participating surgeons said they would use the EyeCee® One again with 64% providing an overall rating of ‘excellent’ for their experience of implanting this intraocular lens. Conclusions: This study indicates excellent post-operative visual acuity and refractive outcomes in eyes after EyeCee® One implantation. This is accompanied with very little risk of intraoperative and post-operative complications.
... However, when the aperture size increased, the AT LARA outperformed the Proming lens (Fig. 2, Table 2), which results from differences in the amount of spherical aberration induced by each model. This impact of spherical aberration on image and visual quality has been reported by many researchers [23][24][25] . Note that the AT LARA features an aberration-neutral design. ...
Preprint
Full-text available
Background To assess the optical behavior of a new diffractive intraocular lens (IOL) and compare its performance to that of an established extended-depth-of-focus (EDOF) IOL. Methods This study assessed the Proming EDOF Multifocal AM2UX [Eyebright Medical Technology (Beijing) Co., Ltd., China] and the AT LARA 829MP [Carl Zeiss Meditec, Germany]. An experimental set-up with 0.01% fluorescein solution and monochromatic light (532nm) was used to visualize the IOLs’ ray propagation. In addition, the optical quality of the IOLs was assessed by measuring the modulation transfer function (MTF) values at 50lp/mm and 3.0 and 4.5mm apertures on the optical bench OptiSpheric® IOL PRO II [Trioptics GmbH, Germany]. Results The ray propagation of the two IOLs showed two distinct foci. Light intensity assessment revealed that both IOLs allocate more energy to primary than secondary focus. At 3.0mm pupil, the MTF values at 50lp/mm for the primary focus were 0.39 and 0.37, and for the secondary focus, 0.29 and 0.26 for the AT LARA and Proming IOLs, respectively. At 4.5mm pupil, the single-frequency MTF for the primary focus was 0.51 and 0.24 and for the secondary focus 0.21 and 0.15 for the AT LARA and Proming IOLs, respectively. Conclusions From the optical point of view, the Proming behaved as a low-add bifocal lens; however, its properties did not differ much from the well-established AT LARA EDOF IOL. The AT LARA outperformed the Proming at low defocus (up to 2D), while the latter demonstrated better image quality in the 2-3D range.
... Several studies have evaluated the effect of aspheric monofocal intraocular lenses on the optical quality of the eye [8,9], but there are relatively few studies with multifocal IOLs. Eppig et al. have studied the impact of corneal spherical aberration on the optical quality of the eye with spheric and aspheric MIOLs on model eyes [10]. ...
Article
Full-text available
PurposeTo evaluate the effect of visual axis positioning on the optical performance of the Tecnis MIOL and the Diff-aA MIOL.Methods In this prospective, randomized comparative study, 70 eyes of 35 subjects with senile cataract were implanted with the spherical aberration–correcting diffractive, bifocal Tecnis ZLB00 IOL and 60 eyes of 30 age-matched subjects with the spherical aberration neutral, diffractive, bifocal Diffractiva IOL. Observation procedure was performed 1, 3, and 6 months postoperatively. Main outcome measures included uncorrected and corrected distance and near visual acuity, manifest refraction, ocular aberrations, and visual quality metrics with 2 mm and 4 mm pupil and the position of visual axis.ResultsAt the 6-month visit, no significant difference was found in monocular and binocular uncorrected (UDVA) and corrected (CDVA) distance and near (UNVA, CNVA) visual acuity between the groups. Spherical and coma-like aberrations were similar measured with a 2-mm pupil, but with a 4-mm pupil, the SA was significantly larger (in negative direction) in the Diffractiva group. The higher-order Strehl ratio and MTF was significantly larger in the Diffractiva group measured at 2 mm entrance pupil; however, this difference disappeared by the 4-mm pupil measurements. Postoperative angle alpha distance had a significant influence on HO Strehl value.Conclusions The size of angle alpha is a predictive factor of image quality by multifocal IOL patients.Trial RegistrationTrial registration number and date of registration: NCT04274088, 14.02.2020.
... The result was comparable to previous study [3] . Lasta et al [4] suggested that the closer the ocular SA was to zero, the better the retinal image quality was, whereas the focal depth would be reduced. Vázquez-Villa et al [5] suggested that the best focal depth and CS could be obtained with a target ocular SA value of 0.10 μm in pseudophakic eye. ...
Article
Full-text available
Aim: To compare the visual performance of pseudophakic eyes implanted with A1-UV and SN60WF aspheric intraocular lens (IOL), and to investigate the correlations between visual quality parameters and pupil size. Methods: This prospective comparative study included 105 eyes of 90 patients with age-related cataract who underwent uneventful phacoemulsification. The subjects were divided into two groups according to the implanted IOL type. Three months postoperatively, visual acuity and contrast sensitivity were measured, wave-front aberrations were assessed using a KR-1W aberrometer (Topcon), and objective optical quality parameters were performed using an optical quality analysis system-OQAS II (Visiometrics). Independent sample t-test and Spearman correlation analysis were used for data analysis. Results: There were no significant differences found in visual acuity, contrast sensitivity and visual quality parameters between the two groups (P>0.05). The measured intraocular spherical aberration (SA) in A1-UV IOL eyes of -0.19±0.05 µm was close to the designed SA value of -0.20 µm. The modulation transfer function cutoff, Strehl ratio and OQAS values were negatively correlated with pupil size in both groups (P<0.01). Conclusion: The subjective and objective visual quality in pseudophakic eyes with A1-UV and SN60WF IOLs are comparable. For aspheric IOL eyes, visual quality decreases with increasing pupil size.
... The biomechanical stability inside the capsular bag is the key feature leading to a successful surgical procedure with this type of lenses. Misalignment or tilt can affect the optical performance inducing wavefront aberrations that may result in significant visual disturbances [14][15][16]. On the one hand, there are some clinical studies [17][18][19] that measure the IOL misalignment, tilt or rotation once it has been implanted inside a pseudophakic eye using different measurement methods such as Scheimplug imaging, Purkinje reflections [20], optical coherence tomography [19], and slit lamp assessment [17]. ...
Article
Full-text available
Purpose To evaluate and compare the effect of misalignment and tilt on the optical performance of different aspheric intraocular lens (IOL) designs. Methods Three aspheric IOLs with a different quantity of spherical aberration (SA) have been designed and the effect of IOL misalignment and tilt on the imaging quality of an eye model has been numerically assessed using a commercial optical design software. The prototypes have been manufactured by lathe turning and tested in vitro using the same optical bench (PMTF, Lambda-X) that complies with International Organization for Standardization standard 11979–2 requirements. Image quality was evaluated from the modulation transfer functions (MTFs), through-focus modulation transfer functions (TF-MTFs), root mean square (RMS) values of defocus, astigmatism and coma, and images of the United States Air Force (USAF) target were taken. A comparison with the optical performance of spherical IOLs has also been performed. Results Intraocular lens misalignment and tilt increased wavefront aberrations; the effect of misalignment on root mean square (RMS) astigmatism and coma was positively correlated with the spherical aberration of the IOL. Aberration-free IOLs showed the highest MTF for all misalignment values and for IOLs with negative SA correction the MTF decays below 0.43 when they are decentered 0.50 mm. Conclusions Aspherical IOLs are more sensitive than spherical IOLs to misalignment or tilt, depending on their SA correction. The optical degradation caused by IOL misalignment had a greater effect on IOL designs with a higher amount of negative spherical aberration. In contrast, the effect of tilt on the optical performance was less sensitive to the IOL design.
Article
Purpose: To determine the median spherical aberration (SA) of the cataractous population, how it relates to biometry, and the theoretical effect of different intraocular lens (IOL) platforms. Methods: A retrospective chart review of patients undergoing cataract surgery evaluation with a high quality Pentacam (Oculus Optikgeräte GmbH) were included. Age, gender, Q-value, mean total SA, higher order aberration root mean square wavefront error, and equivalent keratometry were collected from the Holladay report and axial length and anterior chamber depth (ACD) from the IOLMaster 700 (Carl Zeiss Meditec AG). Results: Data from 1,725 eyes of 999 patients were collected. SA had a median of 0.37 µm (95% confidence interval: 0.36 to 0.38. Age (r = .136, P < .001), Q-factor (r = .743, P < .001), and higher order aberration root mean square wavefront error (r = .307, P < .001) were positively correlated with SA. Average equivalent keratometry (r = -.310, P < .001) was negatively correlated with SA. Axial length (r = -0.037, P = .120) and ACD (r = .004, P = .856) had no association with SA. Up to 1,499 (86.9%) theoretically had SA moved closer to zero with IOLs that had negative SA. Up to 102 (5.9%) had SA theoretically worsened. Conclusions: SA is not normally distributed, suggesting that there may be no "average" SA that IOLs should aim to correct. Patients might benefit from tailoring IOL choice to individual SA. Without access to SA data, eyes with steeper average keratometry or younger patients may have less SA, which could influence IOL choice. [J Refract Surg. 2023;39(2):89-94.].
Article
Full-text available
Purpose: To investigate the impact of different sizes of steep meridian clear corneal incisions for phacoemul sification on anterior corneal higher-order aberrations. Methods: Medical records of patients who underwent 2.2-mm coaxial micro-incision cataract surgery or 2.75-mm coaxial small-incision cataract surgery were retrospectively reviewed. Only patients with preexisting anterior corneal astigmatism <2.00 diopters (D) and ≥0.50 D who underwent a steep meridian clear corneal incision were included. Primary outcomes were 3rd- to 6th-order anterior corneal higher-order aberrations with an 8-mm pupil. Anterior corneal astigmatism and effective phaco time were evaluated as secondary outcomes. Preoperative and 3-month postoperative outcomes were evaluated. Results: Anterior corneal astigmatism significantly decreased after both procedures; however, there was no significant difference found in surgically induced anterior corneal astigmatism between the two procedures (p=0.146). Although the total higher-order aberrations did not significantly change after both procedures, the group comparison showed a significant difference in surgically induced total higher-order aberrations (a decrease of 0.337 ± 1.156 mm in 2.2-mm coaxial micro-incision cataract surgery and an increase of 0.106 ± 0.521 mm in 2.75-mm coaxial small-incision cataract surgery, p=0.046). Spherical aberrations significantly decreased after 2.2-mm coaxial micro-incision cataract surgery (p=0.001), whereas they did not change significantly after 2.75-mm coaxial small-incision cataract surgery (p=0.564). Coma did not significantly change after either of the procedures. Trefoil did not significantly change after 2.2-mm coaxial micro-incision cataract surgery (p=0.361), whereas it significantly increased after 2.75-mm coaxial small-incision cataract surgery (p<0.001). There was no significant difference shown in effective phaco time between the procedures. A significantly positive correlation was shown between surgically induced anterior corneal astigmatism and coma in 2.75-mm coaxial small-incision cataract surgery (r=0.387, p=0.006). There was no significant correlation found between any surgically induced higher-order aberration changes and effective phaco time. Conclusions: The results showed that 2.2-mm coaxial micro-incision cataract surgery and 2.75-mm coaxial small-incision cataract surgery did not significantly degrade the total higher-order aberrations of the anterior cornea. However, the surgically induced changes in total higher-order aberration showed a significant difference between the two procedures, with a slight reduction after 2.2-mm coaxial micro-incision cataract surgery and a slight increase after 2.75-mm coaxial small-incision cataract surgery. Phaco time and power used during surgery had no impact on corneal aberrations.
Article
Purpose: To compare the visual performance after implantation of three aberration-correcting aspherical intraocular lens (IOL). Materials and methods: Seventy-seven eyes of 77 cataract patients were divided into three groups: 26 eyes implanted with a non-constant aberration IOL (LUCIA 601P IOL, Zeiss Company, Germany); 26 eyes implanted with a spherical aberration -0.18μm IOL (CT ASPHINA 509M, Zeiss Company, Germany) and 27 eyes implanted with a spherical aberration -0.27μm IOL (AMO Tecnis ZCB00, Johnson & Johnson Surgical Vision, USA). Three months after operation, the distance visual acuity, wavefront aberrometry, contrast sensitivity, intraocular stray light, IOL decentration, and tilt were evaluated. Results: Three months postoperatively, no statistically significant differences were found in uncorrected distance visual acuity and corrected distance visual acuity (p≥.83). The RMS for total ocular coma was statistically significantly lower in the Lucia group (p=.03) and spherical aberration was statistically significantly lower in the Tecnis group (p<.01). No statistically significant differences were observed among the three lenses in higher order aberration (p=.85) and in contrast sensitivity under both photopic and mesopic lighting conditions (p≥.05). The intraocular stray light was statistically significantly better in the Lucia group (p=.04). No statistically significant differences were observed with respect to IOL decentration (p=.75) and tilt (p=.89). Conclusions: Cataract surgery with non-constant aberration IOL resulted in lower coma and better intraocular stray light than with the spherical aberration -0.18μm and -0.27μm IOLs despite equivalent postoperative levels of visual acuity and contrast sensitivity.
Article
Full-text available
To compare higher order aberrations in two aspherical intraocular lenses (IOLs): Akreos advanced optics (AO) and Dr. Schmidt Microcrystalline 6125 aspheric anterior surface (MC6125AS) with each other. Forty eyes of 39 patients underwent phacoemulsification and Akreos AO and MC6125AS were implanted in their eyes in a random manner. Three months post-operatively, higher order aberrations including spherical aberration, coma aberration, and total aberrations were measured and compared. The total aberration was 0.24±0.17 in eyes with Dr. Schmidt and 0.20±0.01 in eyes with Akreos AO (P=0.361). The mean of coma aberration was 0.17±0.21 and 0.09±0.86 in Dr. Schmidt and Akreos lenses, respectively (P=0.825). Total spherical aberration was almost the same in both groups (Mean: 0.05, P=0.933). Best corrected visual acuity in Akreos AO (0.10±0.68) and Dr. Schmidt (0.09±0.67) did not differ significantly (P=0.700). There is no statistically significant difference in the higher order aberrations between these two aspherical lenses.
Article
Full-text available
Aim: To compare if there is an improvement in visual functions with age-related cataracts between patients receiving a aspherical intraocular lens (IOL) based on corneal wavefront aberration and patients randomly assigned lenses. Methods: A total of 124 eyes of 124 patients with age-related cataracts were placed in experimental group and a group receiving randomly assigned (RA) lenses. The experimental group was undergone Pentacam corneal spherical aberration measurement before surgery; the targeted range for residual total spherical aberration after surgery was set to 0-0.3 µm. Patients with a corneal spherical aberration <0.3 µm were implanted with a zero-spherical aberration advanced optics (AO) aspherical IOL and patients with an aberration ≥0.3 µm received a Tecnis Z9003 aspherical lens in experimental group. RA patients were randomly implanted with an AO lens or a Tecnis Z9003 lens. Three months after surgery total spherical aberration, photopic/mesopic contrast sensitivities, photopic/mesopic with glare contrast sensitivities, and logMAR vision were measured. Results: Statistical analysis on logMAR vision showed no significant difference between two groups (P=0.413). The post-surgical total spherical aberration was 0.126±0.097 µm and 0.152±0.151 µm in the experimental and RA groups, respectively (P=0.12). The mesopic contrast sensitivities at spatial frequencies of 6, 12 and 18 c/d in the experimental group were significantly higher than of the RA group (P=0.00; P=0.04; P=0.02). The mesopic with glare contrast sensitivity in the experimental group at a spatial frequency of 18 c/d was also significantly higher vs the RA group (P=0.01). Conclusion: Pre-surgical corneal spherical aberration measurement in cataract patients followed by customized selection of aspherical IOL implants improved mesopic contrast sensitivities at high spatial frequencies, and thus is a superior strategy compared to the random selection of aspherical IOL implants.
Article
Full-text available
Objective: To assess high order and spherical aberrations results of hydrophobic acrylic AMO Sensar AR40E and hydrophobic acrylic Alcon AcrySof SA60AT intraocular lenses after implantation in cases with bilateral cataract. Methods: Cases diagnosed as bilateral cataract were included in the study and preoperative aberration measurements were recorded by using Nidek OPD SCAN-ARK 1000. Groups were created by implanting AMO Sensar AR40E to one eye of the patients, while Alcon AcrySof SA60AT into the other in a prospective and randomized manner. Aberration measurements were recorded after one and two months of surgery. Results: Overall, 40 eyes in 20 patients (11 women and 9 men) were included in the study. All patients underwent bilateral phacoemulsification surgery due to cataract. There were 20 eyes in both groups. Mean age was 62.4 (range: 31-82) years. There was no significant difference in aberrations recorded before surgery and one and two months after surgery in both groups. (p<0.05). Conclusion: There was no difference among spherical intraocular lenses used in this study.
Article
Purpose: To determine if aspheric intraocular lens (IOL) implantation produces the same degree of postoperative ocular aberration and contrast sensitivity as spherical IOL implantation. Methods: In this randomized prospective comparative study, 60 eyes of 30 cataract surgery patients were randomly assigned to receive a spherical IOL (Rayner 620H) in one eye and an aspheric IOL (Rayner 920H) in the contralateral eye. All patients were examined at 1 month postoperatively. Primary outcomes of contrast sensitivity and ocular wavefront higher order aberrations (HOAs) were assessed. Results: Aspheric IOLs (median total HOAs 0.26 root mean square [RMS]; range 0.13-0.82 RMS) produced significantly lower total HOAs than spherical IOLs (median total HOAs 0.34 RMS; range 0.18-1.08 RMS; p<0.05). Contrast sensitivity was significantly better with aspheric IOLs (median contrast sensitivity 1.8 log units; range 1.35-1.8 log units) than with spherical IOLs (median contrast sensitivity 1.65 log units; range 1.35-1.8 log units; p<0.05). Conclusions: When compared with a structurally (platform and material) similar spherical IOL (Rayner 620H), aspheric IOLs (Rayner 920H) appear to significantly reduce HOAs and yield better levels of contrast sensitivity under photopic conditions.
Article
Purpose: To measure and compare the optical quality of spherical and aspheric toric intraocular lenses (IOLs). Methods: Wavefront aberrations of AcrySof Toric and IQ Toric IOLs (Alcon Laboratories) for different powers were measured at 3- and 5-mm pupils by Nimo TR0805 instrument. The Zernike coefficients of trefoil, coma, tetrafoil, secondary astigmatism, and spherical aberration were evaluated. The point spread functions (PSFs) of each IOL evaluated were calculated from the wavefront aberrations. The PSF images also were calculated from the IOL wavefront aberrations, adding the cornea's aberrations to simulate the optical quality after their implantation. Results: Spherical toric IOLs showed positive and aspheric toric IOLs negative spherical aberrations. Statistically significant differences were found in spherical aberration root mean square (RMS) values between spherical and aspheric IOLs for both pupil sizes (p<0.05). Aspheric toric IOLs showed higher spherical aberration RMS values. We found differences in PSF images between both IOL designs at 5-mm pupil. The PSFs corresponding to the aspheric toric IOLs showed more spread out than the PSFs corresponding to the spherical toric IOLs. However, when corneal aberrations were added, aspheric toric IOLs showed better optical quality than spherical toric IOLs. Conclusions: At 3-mm pupil, the optical quality between the IOL designs was similar, but at 5-mm pupil the optical quality was higher for spherical IOLs than aspheric IOLs. However, when theoretical corneal aberrations were added, aspheric toric IOLs showed better optical quality than spherical toric IOLs, due to the compensation between aspheric toric IOL negative spherical aberration and corneal positive spherical aberration.
Article
The improved designs of intraocular lenses (IOLs) implanted during cataract surgery demand understanding of the possible effects of lens misalignment on optical performance. In this review, we describe the implementation, set-up and validation of two methods to measure in vivo tilt and decentration of IOLs, one based on Purkinje imaging and the other on Scheimpflug imaging. The Purkinje system images the reflections of an oblique collimated light source on the anterior cornea and anterior and posterior IOL surfaces and relies on the well supported assumption of the linearity of the Purkinje images with respect to IOL tilt and decentration. Scheimpflug imaging requires geometrical distortion correction and image processing techniques to retrieve the pupillary axis, IOL axis and pupil centre from the three-dimensional anterior segment image of the eye. Validation of the techniques using a physical eye model indicates that IOL tilt is estimated within an accuracy of 0.261 degree and decentration within 0.161 mm. Measurements on patients implanted with aspheric IOLs indicate that IOL tilt and decentration tend to be mirror symmetric between left and right eyes. The average tilt was 1.54 degrees and the average decentration was 0.21 mm. Simulated aberration patterns using custom models of the patients eyes, built using anatomical data of the anterior cornea and foveal position, the IOL geometry and the measured IOL tilt and decentration predict the experimental wave aberrations measured using laser ray tracing aberrometry on the same eyes. This reveals a relatively minor contribution of IOL tilt and decentration on the higher-order aberrations of the normal pseudophakic eye.
Article
To characterize the influence of tilt angle, decentration, and pupil size on the higher-order aberration (HOA) profile of 3 aspheric intraocular lenses (IOLs) using a physical model eye. A 4-factor (model, pupil, angle, decentration) in vitro experimental design comparing 3 aspheric IOLs using a physical model eye. Measurements of HOA were obtained using the Zywave aberrometer (Bausch & Lomb, Rochester, NY) and a purpose-built physical model eye. The following IOLs were assessed with various levels and combinations of pupil diameter, decentration, and tilt angle: the AcrySof IQ SN60WF aspheric (Alcon, Hünenberg, Switzerland), Technis ZA9003 aspheric (Advanced Medical Optics, Santa Ana, CA), and Adapt Advanced Optics (Bausch & Lomb). Fifteen Zernike modes were compared using multivariate analysis of variance, canonical discrimination, and regression analysis. Four identical IOLs of each IOL model were assessed at all possible combinations of 3 pupil sizes, 4 levels of decentration, and 4 tilt angles. Individual HOA from z200 to z550. Pupil size, decentration, model, and tilt angle all had statically significant effects on the HOA profile. Pupil size contributed most to observed total variability (54.9%), followed by decentration (22.7%), then model (16.6%), and finally tilt angle (5.7%). All factors demonstrated significant interaction terms with respect to HOA. Intraocular lenses with increased aspheric properties inherent in the design of the optic were more sensitive to decentration and change in pupil size. The 3 IOL models demonstrated significant differences in HOAs in response to changes in pupil size, decentration, and tilt angle. All IOL models assessed in this study demonstrated minimal HOA at small pupil diameters. The IOL models with lower, or an absence of, negative spherical aberration were most robust to displacement with increased decentration and tilt angle.
Article
To determine the level of residual spherical aberration that gives the best objective and subjective quality of image after cataract surgery with intraocular lens (IOL) implantation. Department of Ophthalmology, CHU Bretonneau, Tours, France. Cohort study. Six months after microincision (1.8 mm) cataract surgery with aspheric IOL implantation, total aberrations were computed using a Wavescan aberrometer. The modulation transfer function (MTF), Strehl ratio, and objective index of scattering were measured using the Objective Quality Analysis System. Objective depth of focus was computed as the focus range at which the Strehl ratio did not fall below 50% of maximum. Subjective depth of focus was calculated as the difference between the vergence of the punctum remotum and that of the punctum proximum. Thirty patients (54 eyes) were evaluated. The MTF cutoff values were higher with decreasing total ocular spherical aberration (r = 0.56; P < .05). Objective and subjective depth of focus were positively correlated with total spherical aberration (r = 0.26 and r = 0.46, respectively; P < .05). A final spherical aberration of zero obtained by compensation of IOL asphericity gave the greatest improvement in objective quality of vision and better MTF contrast. However, a final target ocular spherical aberration between 0.07 μm and 0.10 μm should be considered to be the best compromise between subjective depth of focus and objective contrast sensitivity. No author has a financial or proprietary interest in any material or method mentioned.
Article
To evaluate the postoperative ocular optical quality using a double-pass method and assess visual outcomes in eyes with an apodized multifocal intraocular lens (IOL) and to correlate the findings with IOL power. Vissum Corporation, Alicante, Spain. Case series. This study evaluated eyes that had cataract surgery with implantation of an AcrySof ReSTOR SN6AD3 multifocal IOL. Near and far visual acuities were measured postoperatively at 1 month and 6 months. Ocular optical quality was assessed at 1 month using a double-pass system (Optical Quality Analysis System), and the point-spread function (PSF) and modulation transfer function (MTF) were characterized. The study included 38 eyes of 19 patients with a mean age of 63.4 years ± 8.4 (SD). The improvements in uncorrected distance visual acuity (from 0.57 ± 0.31 logMAR preoperatively to 0.11 ± 0.13 logMAR 1 month postoperatively and 0.11 ± 0.12 logMAR at 6 months) were statistically significant (P<.00001), as were the improvements in uncorrected near visual acuity (0.76 ± 0.40 logMAR preoperatively to 0.21 ± 0.11 logMAR and 0.12 ± 0.09 logMAR, respectively) (P<.00001). The IOL power correlated significantly with PSF width (r = 0.57, P = .017) and the MTF cutoff frequency (r = -0.36, P = .05). Furthermore, the lower the IOL power, the better the near and distance visual acuities, especially 1 month postoperatively. The apodized multifocal IOL restored distance and near visual function, although IOL power was a limiting factor to the final visual outcomes and optical quality.