Mechanical Vibration Measurement of Solidly
Mounted Resonator in Fluid by Atomic
Force Microscopy
Fei Xu †ID , Xinyi Guo †, Linyan Xu *, Xuexin Duan ID , Hao Zhang, Wei Pang and Xing Fu
State Key Lab of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China;
feifeixu@tju.edu.cn (F.X.); xinyiguo@tju.edu.cn (X.G.); xduan@tju.edu.cn (X.D.); haozhang@tju.edu.cn (H.Z.);
weipang@tju.edu.cn (W.P.); xingfu@tju.edu.cn (X.F.)
*Correspondence: xulinyan@tju.edu.cn; Tel.: +86-186-9806-3909
† These authors contributed equally to this work.
Received: 11 July 2017; Accepted: 4 August 2017; Published: 7 August 2017
Abstract:
The very small vibration of a solidly-mounted resonator (SMR) in fluid may trigger a
relatively large motion of the covering fluid, which was implied by our protein-related experimental
results. Therefore, a series of experimental methods for characterizing the mechanical longitudinal
vibration of the SMR and the corresponding out-of-plane dynamic response of the fluid above the
SMR surface is described in this paper. A SMR device with theoretical resonance frequency of 2.5 GHz
was driven by an amplitude-modulated (AM) signal, in which the amplitude is modulated by a
signal of the second resonance frequency of the atomic force microscope (AFM) cantilever. A lock-in
amplifier is used to demodulate the vibration response of the AFM cantilever, which is proportional to
the amplitude of the sample vibration in contact mode and tapping mode. The amplitude-frequency
curve of the SMR surface is obtained in contact mode with a relatively stronger interaction force
between the AFM tip and the SMR surface. The amplitude-frequency curve of the motion of the
liquid above the SMR device and the peak amplitude of the fluid at different distances above the
SMR surface are measured in tapping mode with a relatively weak interaction force between the
AFM tip and the fluid sample.
Keywords:
solidly mounted resonator; mechanical vibration; atomic force microscopy; amplitude
modulation; lock-in amplifier
1. Introduction
In the last several years, film bulk acoustic resonator (FBAR) devices [
1
–
9
], especially the
solidly-mounted resonator (SMR), has been increasingly studied as ultrasensitive biosensors [
10
],
such as biosensors for human IgE detection [
11
], sensors for DNA hybridization detection, sensors for
special protein detection, and immune sensors [
12
]. However, it is of vital importance to solve some
issues concerning mechanical vibration characteristics of SMRs and the hydrodynamic responses of
the fluid when the resonator is vibrating among these versatile applications.
There is a considerable amount of literature investigating the vibration of various microresonators
based on AFM. Ryder et al. presented a high-speed and accessible detection technique for an annular
microresonator with contact-mode atomic force microscope (AFM) in air. The fundamental resonance
mode was found to be 2.75 MHz with a quality factor of 6 and the peak amplitude is ~230 nm [
13
]. Two
years later, a simple and direct method based on AFM operating in tapping mode was demonstrated to
characterize the vibration amplitude of electrostatically-actuated MEMS resonators with the resonant
frequency of 6.35 MHz and the corresponding amplitude of 90 nm [
14
]. The above two methods are
suitable for a relatively stronger vibration, i.e., generally larger than 20 nm. The vibrations of the
Micromachines 2017,8, 244; doi:10.3390/mi8080244 www.mdpi.com/journal/micromachines
Micromachines 2017,8, 244 2 of 10
first three bending modes of carbon nanotube (CNT) resonators were mechanically detected in air
via AFM. Especially for the third mode of CNT, the measured peak amplitude is only 0.02 nm at the
resonant frequency 1.078 GHz. The CNT resonator is driven by an amplitude-modulated (AM) signal
superposing on its eigenfrequency carrier signal, and the vibration response component of the AFM
cantilever at an AM frequency, which is proportional to the vibration amplitude of the CNT resonator,
could be extracted by a lock-in amplifier [
15
]. Similar experiments were carried out on suspended
graphene resonators by Garcia-Sanchez [
16
]. In the same year, San Paulo et al. successively utilized
atomic force microscopy in contact mode and dynamic force microscopy to characterize the mechanical
vibrations of RF resonators including FBAR [17] in air.
In our previous research, it was confusing that the proteins behaved with a much stronger
dynamic response monitored by optical microscopy when motivated by the SMR device in fluid [18].
Consequently, the dynamic behavior of the bare SMR device without proteins was characterized using
Garcia-Sanchez’s method [
15
]. It was found that even for the same SMR device, the resonant amplitude
measured in fluid had a remarkable discrepancy with that measured in air. Thus, Garcia-Sanchez’s
method has been comprehensively improved in this paper, which can obtain the SMR’s mechanical
vibration properties, as well as the dynamic motion of liquid which is the direct surrounding
environment of the proteins. The stiffness of SMR device, similar to the FBAR device [
19
], is much
larger than CNT and graphene resonators. Here, contact mode, with a relatively stronger interaction
force between the AFM tip and the SMR surface, and tapping mode, with a weaker interaction force,
have been both applied to study the dynamic behavior of the SMR surface and the hydrodynamic
motion of water induced by the vibration of the SMR. The measured amplitudes at different heights
above the resonator surface have also been acquired to characterize the intensity of the movement
of water.
2. Experimental Methods
2.1. Structure of the Solidly Mounted Resonator
Figure 1a,b illustrates the structure of the SMR used in this work. The resonant part of the SMR
consists of a piezoelectric film of aluminum nitride (AlN) sandwiched between two molybdenum (Mo)
electrodes (top and bottom). The overlapping region of the two electrodes defines a pentagonally-active
area (0.02 mm
2
) of the SMR. The resonant part, which is more reliable than an air cavity in fluid,
is isolated from the silicon substrate with a Bragg mirror consisting of silicon dioxide (SiO
2
) and
molybdenum (Mo). Furthermore, a passivation layer composed of AlN is covered on the top electrode
to protect the SMR from oxidation and corrosion. This device is designed to have a main thickness
vibration mode at 2.5 GHz.
Micromachines 2017, 8, 244 2 of 10
vibration, i.e., generally larger than 20 nm. The vibrations of the first three bending modes of carbon
nanotube (CNT) resonators were mechanically detected in air via AFM. Especially for the third
mode of CNT, the measured peak amplitude is only 0.02 nm at the resonant frequency 1.078 GHz.
The CNT resonator is driven by an amplitude-modulated (AM) signal superposing on its
eigenfrequency carrier signal, and the vibration response component of the AFM cantilever at an
AM frequency, which is proportional to the vibration amplitude of the CNT resonator, could be
extracted by a lock-in amplifier [15]. Similar experiments were carried out on suspended graphene
resonators by Garcia-Sanchez [16]. In the same year, San Paulo et al. successively utilized atomic
force microscopy in contact mode and dynamic force microscopy to characterize the mechanical
vibrations of RF resonators including FBAR [17] in air.
In our previous research, it was confusing that the proteins behaved with a much stronger
dynamic response monitored by optical microscopy when motivated by the SMR device in fluid [18].
Consequently, the dynamic behavior of the bare SMR device without proteins was characterized using
Garcia-Sanchez’s method [15]. It was found that even for the same SMR device, the resonant
amplitude measured in fluid had a remarkable discrepancy with that measured in air. Thus,
Garcia-Sanchez’s method has been comprehensively improved in this paper, which can obtain the
SMR’s mechanical vibration properties, as well as the dynamic motion of liquid which is the direct
surrounding environment of the proteins. The stiffness of SMR device, similar to the FBAR device [19],
is much larger than CNT and graphene resonators. Here, contact mode, with a relatively stronger
interaction force between the AFM tip and the SMR surface, and tapping mode, with a weaker
interaction force, have been both applied to study the dynamic behavior of the SMR surface and the
hydrodynamic motion of water induced by the vibration of the SMR. The measured amplitudes at
different heights above the resonator surface have also been acquired to characterize the intensity of
the movement of water.
2. Experimental Methods
2.1. Structure of the Solidly Mounted Resonator
Figure 1a,b illustrates the structure of the SMR used in this work. The resonant part of the SMR
consists of a piezoelectric film of aluminum nitride (AlN) sandwiched between two molybdenum
(Mo) electrodes (top and bottom). The overlapping region of the two electrodes defines a
pentagonally-active area (0.02 mm2) of the SMR. The resonant part, which is more reliable than an air
cavity in fluid, is isolated from the silicon substrate with a Bragg mirror consisting of silicon dioxide
(SiO2) and molybdenum (Mo). Furthermore, a passivation layer composed of AlN is covered on the
top electrode to protect the SMR from oxidation and corrosion. This device is designed to have a
main thickness vibration mode at 2.5 GHz.
(a) (b)
Figure 1. The schematic of SMR: (a) optical photo of SMR; and (b) cross-section of the pentagon
region in (a).
Figure 1.
The schematic of SMR: (
a
) optical photo of SMR; and (
b
) cross-section of the pentagon region
in (a).
Micromachines 2017,8, 244 3 of 10
2.2. Experimental Setup and Principle
Figure 2shows an integrated experimental setup used in this work. A commercial AFM
(Dimension Icon, Bruker Corp., Bremen, Germany) can be operated in both contact mode and
tapping mode. SNL-10 probes made of silicon nitride were purchased from Bruker Corp., and one
of the cantilevers with a nominal spring constant of 0.35 N/m was selected. The AFM cantilever
used is a triangular-shaped one with two arms. This cantilever has two clean peaks of resonant
frequency. The first eigenmode frequency f
1
and second eigenmode frequency f
2
are ~18 kHz and
~110 kHz, respectively.
Micromachines 2017, 8, 244 3 of 10
2.2. Experimental Setup and Principle
Figure 2 shows an integrated experimental setup used in this work. A commercial AFM
(Dimension Icon, Bruker Corp., Bremen, Germany) can be operated in both contact mode and
tapping mode. SNL-10 probes made of silicon nitride were purchased from Bruker Corp., and one of
the cantilevers with a nominal spring constant of 0.35 N/m was selected. The AFM cantilever used is
a triangular-shaped one with two arms. This cantilever has two clean peaks of resonant frequency.
The first eigenmode frequency f1 and second eigenmode frequency f2 are ~18 kHz and ~110 kHz,
respectively.
In Figure 2, the liquid covering the surface of the SMR is actually several drops of water placed
on top of the resonator. The resonator is driven by a 100% amplitude-modulated signal with an RF
power of 0 dBm provided by an RF excitation source, and the modulation frequency fmod is equal to
the second resonance frequency f2 of the AFM cantilever bending mode. As the AFM tip is scanning
along the surface of the resonator, it follows the envelope of the resonator vibration and shows a
response at the modulation frequency fmod with an amplitude which is proportional to the resonator
amplitude. The response of the AFM tip is transformed into electric signal through the detector.
Afterwards, the magnified electric signal can be extracted by “signal access module (SAM)” and it
then can be demodulated by the lock-in amplifier tuned at fmod, which is the same as the frequency of
the reference signal set by the Zi control software. The demodulated signal is actually a voltage
signal that can be displayed in the AFM software. In order to record the value of the signal
automatically instead of an artificial reading, the AFM controller gives a “synchronic” signal to the
trigger of the RF excitation source and the automatic frequency sweeps simultaneously. Finally, an
input signal of the AFM laser photodetector in vertical direction is transferred into a voltage signal
which corresponds to a resonator amplitude stored in AFM software. In tapping mode, the first
eigenmode frequency f1 of the AFM cantilever should be used for topography imaging.
Simultaneously, the second eigenmode frequency f2 is employed for the detection of the vibration
amplitude with a lock-in amplifier tuned at fmod. Hence, it could be recognized as a “dual-frequency
method”. Here, the AFM cantilever is sensitive to the longitudinal vibration, but is not sensitive to
the shear mode vibration.
Figure 2. Automatic and fast experimental detection setup for contact/tapping mode of atomic force
microscopy used in this work.
In contact mode, the value of deflection setpoint reflects different interaction forces between the
AFM tip and the resonator surface. The larger the value is, the greater the interaction force increases.
In tapping mode, tip offset is a parameter describing the distance between the AFM tip and the
resonator within several micrometers. The value of the deflection setpoint and tip offset can be
regulated at any time after the AFM tip is engaged. In order to demonstrate the influence of the
water motion, a mapping of the amplitude versus deflection setpoint in contact mode and the
amplitude versus tip offset above SMR surface in tapping mode are applied.
Figure 2.
Automatic and fast experimental detection setup for contact/tapping mode of atomic force
microscopy used in this work.
In Figure 2, the liquid covering the surface of the SMR is actually several drops of water placed on
top of the resonator. The resonator is driven by a 100% amplitude-modulated signal with an RF power
of 0 dBm provided by an RF excitation source, and the modulation frequency f
mod
is equal to the
second resonance frequency f
2
of the AFM cantilever bending mode. As the AFM tip is scanning along
the surface of the resonator, it follows the envelope of the resonator vibration and shows a response at
the modulation frequency f
mod
with an amplitude which is proportional to the resonator amplitude.
The response of the AFM tip is transformed into electric signal through the detector. Afterwards,
the magnified electric signal can be extracted by “signal access module (SAM)” and it then can be
demodulated by the lock-in amplifier tuned at f
mod
, which is the same as the frequency of the reference
signal set by the Zi control software. The demodulated signal is actually a voltage signal that can be
displayed in the AFM software. In order to record the value of the signal automatically instead of an
artificial reading, the AFM controller gives a “synchronic” signal to the trigger of the RF excitation
source and the automatic frequency sweeps simultaneously. Finally, an input signal of the AFM laser
photodetector in vertical direction is transferred into a voltage signal which corresponds to a resonator
amplitude stored in AFM software. In tapping mode, the first eigenmode frequency f
1
of the AFM
cantilever should be used for topography imaging. Simultaneously, the second eigenmode frequency
f
2
is employed for the detection of the vibration amplitude with a lock-in amplifier tuned at f
mod
.
Hence, it could be recognized as a “dual-frequency method”. Here, the AFM cantilever is sensitive to
the longitudinal vibration, but is not sensitive to the shear mode vibration.
In contact mode, the value of deflection setpoint reflects different interaction forces between
the AFM tip and the resonator surface. The larger the value is, the greater the interaction force
increases. In tapping mode, tip offset is a parameter describing the distance between the AFM tip and
the resonator within several micrometers. The value of the deflection setpoint and tip offset can be
regulated at any time after the AFM tip is engaged. In order to demonstrate the influence of the water
motion, a mapping of the amplitude versus deflection setpoint in contact mode and the amplitude
versus tip offset above SMR surface in tapping mode are applied.
Micromachines 2017,8, 244 4 of 10
3. Experimental Results and Analysis
3.1. Amplitude-Frequency Curve of SMR in Air
Previously, our group used an SMR for driving biomolecules in fluid and found that the dynamic
behavior of the particles was more intense than expected [
20
]. It was guessed that the phenomenon
may be related to the fluid working environment. Therefore, we planned to measure the dynamic
characteristics of the SMR operated in air and in fluid, respectively. First, the amplitude-frequency
curve of the SMR working in air has been measured. According to Garcia-Sanchez’s method in
tapping mode and contact mode in air, amplitude-frequency curves of the SMR driven by a 100%
amplitude-modulated signal with an RF power of 0 dBm are shown in Figure 3a,b. It is easily seen
from the curve that there is only one resonant peak at 2.533 GHz, and the amplitude is 3.44 nm in
tapping mode and 0.794 nm in contact mode. The force exerted in tapping mode has little effect on the
resonator vibration and the results obtained in tapping mode is closer to the real vibration amplitude
of the SMR device [17,19].
Micromachines 2017, 8, 244 4 of 10
3. Experimental Results and Analysis
3.1. Amplitude-Frequency Curve of SMR in Air
Previously, our group used an SMR for driving biomolecules in fluid and found that the
dynamic behavior of the particles was more intense than expected [20]. It was guessed that the
phenomenon may be related to the fluid working environment. Therefore, we planned to measure
the dynamic characteristics of the SMR operated in air and in fluid, respectively. First, the
amplitude-frequency curve of the SMR working in air has been measured. According to
Garcia-Sanchez’s method in tapping mode and contact mode in air, amplitude-frequency curves of
the SMR driven by a 100% amplitude-modulated signal with an RF power of 0 dBm are shown in
Figure 3a,b. It is easily seen from the curve that there is only one resonant peak at 2.533 GHz, and the
amplitude is 3.44 nm in tapping mode and 0.794 nm in contact mode. The force exerted in tapping
mode has little effect on the resonator vibration and the results obtained in tapping mode is closer to
the real vibration amplitude of the SMR device [17,19].
Figure 3. Amplitude-frequency curve measured in: (a) tapping mode in air; (b) contact mode in air;
(c) tapping mode in fluid; and (d) contact mode in fluid.
3.2. Amplitude-Frequency Curve of SMR in Fluid
Utilizing identical measurement conditions and same radio-frequency power, the comparison
test on the same resonator in Section 3.1 was carried on in fluid. Figure 3c shows an
amplitude-frequency curve obtained in tapping mode in fluid. The free vibration amplitude of the
AFM cantilever is set to be 300 mV, and the amplitude setpoint is set to be 150 mV, which is 50% of
the free vibration amplitude. The range of sweep frequency is from 2.328 GHz to 2.583 GHz and the
measured resonance frequency is 2.456 GHz. The resonance frequency of the resonator and Q-factors of
amplitude-frequency curves are different in air and in fluid. The deviation of the resonance frequency
reaches 77 MHz and the Q-factor differs from 63 in air (Figure 3a) to 58 in fluid (Figure 3c). The
interaction of the SMR with liquid due to viscosity and the larger emission of acoustic power due to
lower acoustic impedance mismatch are responsible for the frequency shift and the lower quality factor.
In Figure 3c, the peak amplitude is 8.79 nm, which is ~2.5 times larger than the measurement
result obtained in air in Figure 3a. Theoretically, the energy dissipation of the SMR in fluid is larger
than that in air. It is reasonable to estimate that under the same power the vibration amplitude of
SMR in fluid should be smaller than 3.44 nm (Figure 3a). The vibration amplitude of SMR in fluid
should not be this large.
How can we verify that the amplitude detected is due to a large movement of the SMR surface
or a dynamic influence of the liquid? A more straightforward experiment is conducted with a
“single-frequency method” [14,21] in tapping mode in fluid. The idea of vibration amplitude
measurement is shown in Figure 4a. When the SMR is not excited, let the probe tap on a single point
of the SMR surface in imaging mode. The AFM cantilever will give out the position of the static
Figure 3.
Amplitude-frequency curve measured in: (
a
) tapping mode in air; (
b
) contact mode in air;
(c) tapping mode in fluid; and (d) contact mode in fluid.
3.2. Amplitude-Frequency Curve of SMR in Fluid
Utilizing identical measurement conditions and same radio-frequency power, the comparison test
on the same resonator in Section 3.1 was carried on in fluid. Figure 3c shows an amplitude-frequency
curve obtained in tapping mode in fluid. The free vibration amplitude of the AFM cantilever is set to
be 300 mV, and the amplitude setpoint is set to be 150 mV, which is 50% of the free vibration amplitude.
The range of sweep frequency is from 2.328 GHz to 2.583 GHz and the measured resonance frequency
is 2.456 GHz. The resonance frequency of the resonator and Q-factors of amplitude-frequency curves
are different in air and in fluid. The deviation of the resonance frequency reaches 77 MHz and the
Q-factor differs from 63 in air (Figure 3a) to 58 in fluid (Figure 3c). The interaction of the SMR with
liquid due to viscosity and the larger emission of acoustic power due to lower acoustic impedance
mismatch are responsible for the frequency shift and the lower quality factor.
In Figure 3c, the peak amplitude is 8.79 nm, which is ~2.5 times larger than the measurement
result obtained in air in Figure 3a. Theoretically, the energy dissipation of the SMR in fluid is larger
than that in air. It is reasonable to estimate that under the same power the vibration amplitude of SMR
in fluid should be smaller than 3.44 nm (Figure 3a). The vibration amplitude of SMR in fluid should
not be this large.
How can we verify that the amplitude detected is due to a large movement of the SMR surface
or a dynamic influence of the liquid? A more straightforward experiment is conducted with a
“single-frequency method” [
14
,
21
] in tapping mode in fluid. The idea of vibration amplitude
measurement is shown in Figure 4a. When the SMR is not excited, let the probe tap on a single
Micromachines 2017,8, 244 5 of 10
point of the SMR surface in imaging mode. The AFM cantilever will give out the position of the static
sample surface as a reference. However when the SMR is excited by the RF signal without modulation,
the AFM cantilever will step up to the envelope of the vibration peaks of the sample. The AFM
cantilever acts as a mechanical low-pass filter as it encounters the RF high-frequency mechanical
vibration of the SMR. Thu, it is easy to understand that the vibration amplitude of the SMR could be
read from the step height in the topography of the scanning image in tapping mode.
Micromachines 2017, 8, 244 5 of 10
sample surface as a reference. However when the SMR is excited by the RF signal without
modulation, the AFM cantilever will step up to the envelope of the vibration peaks of the sample.
The AFM cantilever acts as a mechanical low-pass filter as it encounters the RF high-frequency
mechanical vibration of the SMR. Thu, it is easy to understand that the vibration amplitude of the
SMR could be read from the step height in the topography of the scanning image in tapping mode.
Figure 5a shows the “step-type” topography of the static and vibrating SMR in fluid obtained
from bottom to top. In Figure 5a, the upper half is recorded while the SMR is driven by a RF signal of
0 dBm. Figure 5b display the cross-section where the white line in Figure 5a is located. As is known
from Figure 3a in Section 3.1, the vibration amplitude of the SMR with a RF signal of 0 dBm should
be no larger than 3.44 nm. However, in Figure 4b, there is a ~10 nm step that the AFM cantilever lifts
up sharply to a high position beyond the SMR surface. Figure 4b gives an illustration of the relative
position of the cantilever and the sample. The vibration valleys of the cantilever do not reach the
vibration peaks of the SMR. Thus, Figures 4 and 5 indicate that the measured amplitude in Figure 3c
in fluid in tapping mode is due to the hydrodynamic motion of the liquid above the SMR surface.
Hence, the explanation for larger amplitude detected in tapping mode is that the mechanical
vibration at the SMR surface can trigger much more intense movement of water in the vicinity.
Consequently, the response of the AFM cantilever is not directly transmitted from the surface of
SMR. The much stronger hydrodynamic motion of water hinders the contact of the tip and the SMR
surface. As observed previously, the strong dynamic behavior of the proteins in fluid is reasonable
to be considered the excitation of vibrating water [18].
(a) (b)
Figure 4. An illustration of the relative position of the cantilever and the sample: (a) in air; and (b) in
water.
(a) (b)
Figure 5. Height information of the AFM cantilever: (a) image of the height of the AFM cantilever
with a half of excitation and another half without excitation on the SMR; and (b) values of the height
sensor in the area where the white line in (a) lies.
Figure 4.
An illustration of the relative position of the cantilever and the sample: (
a
) in air; and (
b
) in water.
Figure 5a shows the “step-type” topography of the static and vibrating SMR in fluid obtained
from bottom to top. In Figure 5a, the upper half is recorded while the SMR is driven by a RF signal of
0 dBm. Figure 5b display the cross-section where the white line in Figure 5a is located. As is known
from Figure 3a in Section 3.1, the vibration amplitude of the SMR with a RF signal of 0 dBm should
be no larger than 3.44 nm. However, in Figure 4b, there is a ~10 nm step that the AFM cantilever
lifts up sharply to a high position beyond the SMR surface. Figure 4b gives an illustration of the
relative position of the cantilever and the sample. The vibration valleys of the cantilever do not
reach the vibration peaks of the SMR. Thus, Figures 4and 5indicate that the measured amplitude
in Figure 3c in fluid in tapping mode is due to the hydrodynamic motion of the liquid above the
SMR surface. Hence, the explanation for larger amplitude detected in tapping mode is that the
mechanical vibration at the SMR surface can trigger much more intense movement of water in the
vicinity. Consequently, the response of the AFM cantilever is not directly transmitted from the surface
of SMR. The much stronger hydrodynamic motion of water hinders the contact of the tip and the SMR
surface. As observed previously, the strong dynamic behavior of the proteins in fluid is reasonable to
be considered the excitation of vibrating water [18].
Micromachines 2017, 8, 244 5 of 10
sample surface as a reference. However when the SMR is excited by the RF signal without
modulation, the AFM cantilever will step up to the envelope of the vibration peaks of the sample.
The AFM cantilever acts as a mechanical low-pass filter as it encounters the RF high-frequency
mechanical vibration of the SMR. Thu, it is easy to understand that the vibration amplitude of the
SMR could be read from the step height in the topography of the scanning image in tapping mode.
Figure 5a shows the “step-type” topography of the static and vibrating SMR in fluid obtained
from bottom to top. In Figure 5a, the upper half is recorded while the SMR is driven by a RF signal of
0 dBm. Figure 5b display the cross-section where the white line in Figure 5a is located. As is known
from Figure 3a in Section 3.1, the vibration amplitude of the SMR with a RF signal of 0 dBm should
be no larger than 3.44 nm. However, in Figure 4b, there is a ~10 nm step that the AFM cantilever lifts
up sharply to a high position beyond the SMR surface. Figure 4b gives an illustration of the relative
position of the cantilever and the sample. The vibration valleys of the cantilever do not reach the
vibration peaks of the SMR. Thus, Figures 4 and 5 indicate that the measured amplitude in Figure 3c
in fluid in tapping mode is due to the hydrodynamic motion of the liquid above the SMR surface.
Hence, the explanation for larger amplitude detected in tapping mode is that the mechanical
vibration at the SMR surface can trigger much more intense movement of water in the vicinity.
Consequently, the response of the AFM cantilever is not directly transmitted from the surface of
SMR. The much stronger hydrodynamic motion of water hinders the contact of the tip and the SMR
surface. As observed previously, the strong dynamic behavior of the proteins in fluid is reasonable
to be considered the excitation of vibrating water [18].
(a) (b)
Figure 4. An illustration of the relative position of the cantilever and the sample: (a) in air; and (b) in
water.
(a) (b)
Figure 5. Height information of the AFM cantilever: (a) image of the height of the AFM cantilever
with a half of excitation and another half without excitation on the SMR; and (b) values of the height
sensor in the area where the white line in (a) lies.
Figure 5.
Height information of the AFM cantilever: (
a
) image of the height of the AFM cantilever with
a half of excitation and another half without excitation on the SMR; and (
b
) values of the height sensor
in the area where the white line in (a) lies.
Micromachines 2017,8, 244 6 of 10
In order to obtain accurate and stable vibration amplitude of SMR surface, the idea of increasing
the contact force between the AFM tip and the resonator surface has been attempted through changing
the value of the amplitude setpoint in tapping mode. The relationship of the peak amplitude of
SMR and the amplitude setpoint is shown in Figure 6, in which the measured vibration amplitude is
commonly larger than that in air. The measured amplitude is larger than 8 nm when the amplitude
setpoint is 33–50% (100–150 mV) of the free amplitude of AFM cantilever (300 mV). The measured
amplitude is sharply damping to 0 when the amplitude setpoint is less than 7% (~20 mV). It is inferred
that the AFM tip is easier to be influenced in fluid that the measurement method based on tapping
mode is not suitable to detect the real vibration amplitude of the SMR device in fluid. Moreover,
this kind of magnified effect was also found in other SMR devices working in fluid, and the ratio may
sometimes be up to five times. Nevertheless, the unique application for the method is that it can be
used to characterize the movement of water, which is just the working environment of our protein
samples. In Section 3.3, the dynamic response of water at different height above the SMR surface will
be further measured.
Micromachines 2017, 8, 244 6 of 10
In order to obtain accurate and stable vibration amplitude of SMR surface, the idea of increasing
the contact force between the AFM tip and the resonator surface has been attempted through
changing the value of the amplitude setpoint in tapping mode. The relationship of the peak
amplitude of SMR and the amplitude setpoint is shown in Figure 6, in which the measured vibration
amplitude is commonly larger than that in air. The measured amplitude is larger than 8 nm when the
amplitude setpoint is 33–50% (100–150 mV) of the free amplitude of AFM cantilever (300 mV). The
measured amplitude is sharply damping to 0 when the amplitude setpoint is less than 7% (~20 mV). It
is inferred that the AFM tip is easier to be influenced in fluid that the measurement method based on
tapping mode is not suitable to detect the real vibration amplitude of the SMR device in fluid.
Moreover, this kind of magnified effect was also found in other SMR devices working in fluid, and
the ratio may sometimes be up to five times. Nevertheless, the unique application for the method is
that it can be used to characterize the movement of water, which is just the working environment of
our protein samples. In Section 3.3, the dynamic response of water at different height above the SMR
surface will be further measured.
Figure 6. Amplitude response measured at different amplitude setpoints in tapping mode in fluid.
3.3. Measured Amplitudes at Different Heights above the Resonator in Tapping Mode
According to Section 3.2, the motion of water has a significant effect on the measured
amplitude. It is supposed that the magnitude of water motion differs with the height above the SMR
surface. Through changing the values of tip offset, the distance between the AFM tip and the SMR
surface can be adjusted. The corresponding measured amplitudes are shown in Figure 7. The
distance ranges from 0 nm to 3.91 µm, and the reference position is the central position of the AFM
tip working in Section 3.2. The corresponding measured amplitude gradually changes from 8.79 nm
to 4.65 nm. It is inferred that the vibration intensity of water is negatively correlated with the
distance above the SMR and the AFM tip experiences a relatively large impact force induced by the
intense motion of the water. The method utilizing tapping mode is beneficial to investigate the
movement of the water, and the results will help us significantly to understand the behavior of
biomolecules.
0 20 40 60 80 100 120 140 160 180 200 220
-1
0
1
2
3
4
5
6
7
8
9
10
Amplitude (nm)
setpoint (mV)
Figure 6. Amplitude response measured at different amplitude setpoints in tapping mode in fluid.
3.3. Measured Amplitudes at Different Heights above the Resonator in Tapping Mode
According to Section 3.2, the motion of water has a significant effect on the measured amplitude.
It is supposed that the magnitude of water motion differs with the height above the SMR surface.
Through changing the values of tip offset, the distance between the AFM tip and the SMR surface can
be adjusted. The corresponding measured amplitudes are shown in Figure 7. The distance ranges
from 0 nm to 3.91
µ
m, and the reference position is the central position of the AFM tip working in
Section 3.2. The corresponding measured amplitude gradually changes from 8.79 nm to 4.65 nm. It is
inferred that the vibration intensity of water is negatively correlated with the distance above the SMR
and the AFM tip experiences a relatively large impact force induced by the intense motion of the water.
The method utilizing tapping mode is beneficial to investigate the movement of the water, and the
results will help us significantly to understand the behavior of biomolecules.
Micromachines 2017,8, 244 7 of 10
Micromachines 2017, 8, 244 7 of 10
Figure 7. Amplitude versus different values of distance between the AFM tip and SMR.
3.4. Amplitude-Frequency Curve of SMR in Contact Mode in Fluid
Taking the weak interaction force in tapping mode into account, contact mode is better to
measure the real vibration amplitude of the SMR regardless of the effect of water through setting a
larger value of the deflection setpoint for a larger interaction force. Figure 3d shows an
amplitude-frequency curve of the SMR in contact mode in fluid. The deflection setpoint is set to be 1.2 V.
Considering that the spring constant of the AFM cantilever is 0.35 N/m and the sensitivity is 70 nm/V, the
deflection setpoint value of 1 V amounts to a contact force of 29.4 nN between the AFM tip and the SMR.
To compare with the result from tapping mode, the curve is obtained at the same position of the
resonator with the same range of sweep frequency. By chance, the resonance frequency is also 2.456
GHz. The peak amplitude is 2.90 nm, which is slightly smaller than the data obtained in air. Here,
caution must be taken while obtaining the curve. Only under the condition that the deflection
setpoint is set to be large enough, the AFM tip can penetrate the water layer between them and sense
the accurate vibration amplitude of the device. We assume that the choice of the interaction force is
reasonable within a certain range and the confirmation of the specific parameters need further
experiments, as described in Section 3.5.
Additionally, in order to study the relationship of the mechanical resonance frequency of the
SMR and electrical resonance frequency of the series-circuit model, the electrical characterization of
the SMR is measured by the network analyzer. The electrical results are shown in Figure 8. It can be
seen that the resonance peak is 2.483 GHz in fluid, which has only ~27 MHz deviation according to
the result introduced above and the resonance peak increases to 2.511 GHz in air that has ~22 MHz
less than that measured by AFM. The AFM method based on contact mode with proper parameters
cannot only detect amplitude-frequency curve of SMR surface vibration, but also indicates the inner
consistency of the electrical resonance frequency and the mechanical resonance frequency.
(a)
0 1000 2000 3000 4000
4
5
6
7
8
9
Amplitude (nm)
Distance (nm)
Figure 7. Amplitude versus different values of distance between the AFM tip and SMR.
3.4. Amplitude-Frequency Curve of SMR in Contact Mode in Fluid
Taking the weak interaction force in tapping mode into account, contact mode is better to measure
the real vibration amplitude of the SMR regardless of the effect of water through setting a larger value of
the deflection setpoint for a larger interaction force. Figure 3d shows an amplitude-frequency curve of
the SMR in contact mode in fluid. The deflection setpoint is set to be 1.2 V. Considering that the spring
constant of the AFM cantilever is 0.35 N/m and the sensitivity is 70 nm/V, the deflection setpoint
value of 1 V amounts to a contact force of 29.4 nN between the AFM tip and the SMR. To compare with
the result from tapping mode, the curve is obtained at the same position of the resonator with the same
range of sweep frequency. By chance, the resonance frequency is also 2.456 GHz. The peak amplitude
is 2.90 nm, which is slightly smaller than the data obtained in air. Here, caution must be taken while
obtaining the curve. Only under the condition that the deflection setpoint is set to be large enough,
the AFM tip can penetrate the water layer between them and sense the accurate vibration amplitude
of the device. We assume that the choice of the interaction force is reasonable within a certain range
and the confirmation of the specific parameters need further experiments, as described in Section 3.5.
Additionally, in order to study the relationship of the mechanical resonance frequency of the SMR
and electrical resonance frequency of the series-circuit model, the electrical characterization of the SMR
is measured by the network analyzer. The electrical results are shown in Figure 8. It can be seen that
the resonance peak is 2.483 GHz in fluid, which has only ~27 MHz deviation according to the result
introduced above and the resonance peak increases to 2.511 GHz in air that has ~22 MHz less than
that measured by AFM. The AFM method based on contact mode with proper parameters cannot only
detect amplitude-frequency curve of SMR surface vibration, but also indicates the inner consistency of
the electrical resonance frequency and the mechanical resonance frequency.
Micromachines 2017, 8, 244 7 of 10
Figure 7. Amplitude versus different values of distance between the AFM tip and SMR.
3.4. Amplitude-Frequency Curve of SMR in Contact Mode in Fluid
Taking the weak interaction force in tapping mode into account, contact mode is better to
measure the real vibration amplitude of the SMR regardless of the effect of water through setting a
larger value of the deflection setpoint for a larger interaction force. Figure 3d shows an
amplitude-frequency curve of the SMR in contact mode in fluid. The deflection setpoint is set to be 1.2 V.
Considering that the spring constant of the AFM cantilever is 0.35 N/m and the sensitivity is 70 nm/V, the
deflection setpoint value of 1 V amounts to a contact force of 29.4 nN between the AFM tip and the SMR.
To compare with the result from tapping mode, the curve is obtained at the same position of the
resonator with the same range of sweep frequency. By chance, the resonance frequency is also 2.456
GHz. The peak amplitude is 2.90 nm, which is slightly smaller than the data obtained in air. Here,
caution must be taken while obtaining the curve. Only under the condition that the deflection
setpoint is set to be large enough, the AFM tip can penetrate the water layer between them and sense
the accurate vibration amplitude of the device. We assume that the choice of the interaction force is
reasonable within a certain range and the confirmation of the specific parameters need further
experiments, as described in Section 3.5.
Additionally, in order to study the relationship of the mechanical resonance frequency of the
SMR and electrical resonance frequency of the series-circuit model, the electrical characterization of
the SMR is measured by the network analyzer. The electrical results are shown in Figure 8. It can be
seen that the resonance peak is 2.483 GHz in fluid, which has only ~27 MHz deviation according to
the result introduced above and the resonance peak increases to 2.511 GHz in air that has ~22 MHz
less than that measured by AFM. The AFM method based on contact mode with proper parameters
cannot only detect amplitude-frequency curve of SMR surface vibration, but also indicates the inner
consistency of the electrical resonance frequency and the mechanical resonance frequency.
(a)
0 1000 2000 3000 4000
4
5
6
7
8
9
Amplitude (nm)
Distance (nm)
Figure 8. Cont.
Micromachines 2017,8, 244 8 of 10
Micromachines 2017, 8, 244 8 of 10
(b)
Figure 8. Frequency response curve of the SMR measured by a network analyzer: (a) in fluid; and
(b) in air.
3.5. Measured Amplitude with Different Interaction Force between the AFM Tip and the Resonator in Contact
Mode
If the value of the deflection setpoint is set to be smaller, the interaction force between the AFM
tip and the resonator will also become smaller. As a result, the measured amplitude will increase.
The relationship between the measured amplitude of the SMR and the deflection setpoint is shown
in Figure 9, in which the setpoint increases from 0.8 V to 1.0 V with 0.02 V intervals and from 1.0 V to
2.0 V with 0.1 V intervals in order to obtain a relatively smooth and continuous curve. As the value
of the deflection setpoint increases from 0.8 V to 2.0 V, the vibration amplitude decreases from 5.27
nm to 2.90 nm. It can be seen that with the increasing deflection setpoint larger than 1.2 V, the
amplitude tends to be a constant value of 2.90 nm. It is hypothesized that the interaction force
between the AFM tip and the SMR surface contributes to this result. As the value of the deflection
setpoint varies from 0.8 V to 1.2 V, the tip is more easily affected by water motion as a result of not
contacting the resonator closely. Here, when the value of the deflection setpoint is set to be 1.2 V to
2.0 V, the measured amplitude will not be influenced by water. The SMR device used in this paper
has a larger stiffness, which is determined by its structure. Thus, the larger force does not seriously
affect its amplitude attenuation.
Figure 9. Amplitude response measured at different deflection setpoints in contact mode.
0.8 1.0 1.2 1.4 1.6 1.8 2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
Amplitude (nm)
Deflection setpoint (V)
Figure 8.
Frequency response curve of the SMR measured by a network analyzer: (
a
) in fluid; and (
b
) in air.
3.5. Measured Amplitude with Different Interaction Force between the AFM Tip and the Resonator in
Contact Mode
If the value of the deflection setpoint is set to be smaller, the interaction force between the AFM
tip and the resonator will also become smaller. As a result, the measured amplitude will increase.
The relationship between the measured amplitude of the SMR and the deflection setpoint is shown in
Figure 9, in which the setpoint increases from 0.8 V to 1.0 V with 0.02 V intervals and from 1.0 V to
2.0 V with 0.1 V intervals in order to obtain a relatively smooth and continuous curve. As the value of
the deflection setpoint increases from 0.8 V to 2.0 V, the vibration amplitude decreases from 5.27 nm to
2.90 nm. It can be seen that with the increasing deflection setpoint larger than 1.2 V, the amplitude tends
to be a constant value of 2.90 nm. It is hypothesized that the interaction force between the AFM tip and
the SMR surface contributes to this result. As the value of the deflection setpoint varies from 0.8 V to
1.2 V, the tip is more easily affected by water motion as a result of not contacting the resonator closely.
Here, when the value of the deflection setpoint is set to be 1.2 V to 2.0 V, the measured amplitude
will not be influenced by water. The SMR device used in this paper has a larger stiffness, which is
determined by its structure. Thus, the larger force does not seriously affect its amplitude attenuation.
Micromachines 2017, 8, 244 8 of 10
(b)
Figure 8. Frequency response curve of the SMR measured by a network analyzer: (a) in fluid; and
(b) in air.
3.5. Measured Amplitude with Different Interaction Force between the AFM Tip and the Resonator in Contact
Mode
If the value of the deflection setpoint is set to be smaller, the interaction force between the AFM
tip and the resonator will also become smaller. As a result, the measured amplitude will increase.
The relationship between the measured amplitude of the SMR and the deflection setpoint is shown
in Figure 9, in which the setpoint increases from 0.8 V to 1.0 V with 0.02 V intervals and from 1.0 V to
2.0 V with 0.1 V intervals in order to obtain a relatively smooth and continuous curve. As the value
of the deflection setpoint increases from 0.8 V to 2.0 V, the vibration amplitude decreases from 5.27
nm to 2.90 nm. It can be seen that with the increasing deflection setpoint larger than 1.2 V, the
amplitude tends to be a constant value of 2.90 nm. It is hypothesized that the interaction force
between the AFM tip and the SMR surface contributes to this result. As the value of the deflection
setpoint varies from 0.8 V to 1.2 V, the tip is more easily affected by water motion as a result of not
contacting the resonator closely. Here, when the value of the deflection setpoint is set to be 1.2 V to
2.0 V, the measured amplitude will not be influenced by water. The SMR device used in this paper
has a larger stiffness, which is determined by its structure. Thus, the larger force does not seriously
affect its amplitude attenuation.
Figure 9. Amplitude response measured at different deflection setpoints in contact mode.
0.8 1.0 1.2 1.4 1.6 1.8 2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
Amplitude (nm)
Deflection setpoint (V)
Figure 9. Amplitude response measured at different deflection setpoints in contact mode.
4. Conclusions
In summary, a comprehensive detection method based on atomic force microscopy in contact
mode should be employed for the sake of obtaining the real longitudinal vibration amplitude of
a solidly-mounted resonator in fluid. It is noted that there should be a relatively large interaction
Micromachines 2017,8, 244 9 of 10
force between the AFM tip and the resonator. In addition, a series of improved methods in tapping
mode utilizing a frequency sweep and tip offset change are suitable to measure the out-of-plane
vibration amplitude caused by the motion of water. Differently-measured amplitudes can reflect
different intensities of the movement of water. With respect to the SMR, the vibration amplitude of
its surface in fluid is slightly smaller than that in air. In a fluid environment, there exists a relatively
intense longitudinal vibration of water covering the surface of the SMR. The frequency response
characteristic of the water above the SMR surface is similar to that of the SMR itself, and the maximum
value of the amplitude occurs close to the surface of the SMR. The vibration amplitude decreases
with the increased distance away from the SMR surface. According to incomplete statistics, the peak
longitudinal vibration amplitude of water is 2–5 times than that of the SMR. Consequently, the method
also provides a new thought to explain the dynamic behavior of protein above the working SMR
device in fluid.
Acknowledgments:
This project was supported by National Science Foundation of China (grant No. 51305298) and
the Tianjin Research Program of Application Foundation and Advanced Technology (grant No. 13JCQNJC04700).
Author Contributions:
Fei Xu, Linyan Xu, and Xinyi Guo conceived and designed the experiments, performed
the experiments, and analyzed the data; Fei Xu and Linyan Xu wrote the paper; Xuexin Duan, Hao Zhang,
and Wei Pang afforded the experimental devices; and Xing Fu contributed to the supervision and gave valuable
advice to this work.
Conflicts of Interest: The authors declare no conflict of interest.
References
1.
Gao, J.; Liu, G.; Li, J.; Li, G. Recent developments of film bulk acoustic resonators (FBAR). Funct. Mater. Lett.
2016,9, 1630002. [CrossRef]
2.
Moreira, M.; Bjurström, J.; Katardjev, I.; Yantchev, V. Aluminum scandium nitride thin-film bulk acoustic
resonators for wide band applications. Vacuum 2011,86, 23–26. [CrossRef]
3.
Mastromatteo, U.; Villa, F.F. High sensitivity acoustic wave AlN/Si mass detectors arrays for artificial
olfactory and biosensing applications: A review. Sens. Actuators B Chem. 2013,179, 319–327. [CrossRef]
4.
Dong, S.R.; Bian, X.L.; Jin, H.; Hu, N.N.; Zhou, J.; Wong, H.; Deen, M.J. Electrically tunable film bulk acoustic
resonator based on Au/ZnO/Al structure. Appl. Phys. Lett. 2013,103, 062904. [CrossRef]
5.
Wang, Q.; Wu, X.; Ren, W.; Li, X.; Wang, G.; Shi, P.; Ye, Z. Design and micro-machining fabrication of
piezoelectric diaphragm chambers for biomaterial detection. Ceram. Int. 2015,41, S612–S617. [CrossRef]
6.
Xu, W.; Zhang, X.; Yu, H.; Abbaspour-Tamijani, A.; Chae, J. In-liquid quality factor improvement for film
bulk acoustic resonators by integration of microfluidic channels. IEEE Electron Device Lett.
2009
,30, 647–649.
[CrossRef]
7.
Wang, J.; Chen, D.; Xu, Y.; Liu, W. Label-free immunosensor based on micromachined bulk acoustic resonator
for the detection of trace pesticide residues. Sens. Actuators B Chem. 2014,190, 378–383. [CrossRef]
8.
Fattinger, G.G.; Kaitila, J.; Aigner, R.; Nessler, W. Thin film bulk acoustic wave devices for applications at
5.2 GHz. In Proceedings of the 2003 IEEE Symposium on Ultrasonics, Honolulu, HI, USA, 5–8 October 2003;
Volume 1, pp. 174–177.
9.
Sorokin, B.P.; Kvashnin, G.M.; Volkov, A.P.; Bormashov, V.S.; Aksenenkov, V.V.; Kuznetsov, M.S.; Gordeev, G.I.;
Telichko, A.V. AlN/single crystalline diamond piezoelectric structure as a high overtone bulk acoustic
resonator. Appl. Phys. Lett. 2013,102, 113507. [CrossRef]
10.
Flewitt, A.J.; Luo, J.K.; Fu, Y.Q.; Garcia-Gancedo, L.; Du, X.Y.; Lu, J.R.; Zhao, X.B.; lobrra, E.; Ramos, M.;
Milne, W.I. ZnO based SAW and FBAR devices for bio-sensing applications. J. Non-Newton. Fluid Mech.
2015
,
222, 209–216. [CrossRef]
11.
Chen, Y.C.; Shih, W.C.; Chang, W.T.; Yang, C.H.; Kao, K.S.; Cheng, C.C. Biosensor for human IgE detection
using shear-mode FBAR devices. Nanoscale Res. Lett. 2015,10, 69. [CrossRef] [PubMed]
12.
Pang, W.; Zhao, H.; Kim, E.S.; Zhang, H.; Yu, H.; Hu, X. Piezoelectric microelectromechanical resonant
sensors for chemical and biological detection. Lab Chip 2012,12, 29–44. [CrossRef] [PubMed]
Micromachines 2017,8, 244 10 of 10
13.
Ryder, S.; Lee, K.B.; Meng, X.; Lin, L. Characterization of out-of-plane high frequency microresonators by
AFM. In Proceedings of the 12th International Conference on TRANSDUCERS, Solid-State Sensors, Actuators
and Microsystems, Boston, MA, USA, 8–12 June 2003; Volume 1, pp. 424–427.
14.
Agache, V.; Legrand, B.; Nakamura, K.; Kawakatsu, H.; Buchaillot, L.; Toshiyoshi, H.; Collard, D.; Fujita, H.
Characterization of vertical vibration of electrostatically actuated resonators using atomic force microscope
in non-contact mode. In Proceedings of the 13th International Conference on Solid-State Sensors, Actuators
and Microsystems, Digest of Technical Papers 2005, Seoul, Korea, 5–9 June 2005; Volume 2, pp. 2023–2026.
15.
Garcia-Sanchez, D.; San-Paulo, A.; Esplandiu, M.J.; Perez-Murano, F.; Forró, L.; Aguasca, A.; Bachtold, A.
Mechanical detection of carbon nanotube resonator vibrations. Phys. Rev. Lett. 2007,99, 088501. [CrossRef]
[PubMed]
16.
Garcia-Sanchez, D.; van-der-Zande, A.M.; San-Paulo, A.; Lassagne, B.; McEuen, P.L.; Bachtold, A. Imaging
mechanical vibrations in suspended graphene sheets. Nano Lett. 2008,8, 1399–1403. [CrossRef] [PubMed]
17.
San-Paulo, A.; Black, J.P.; White, R.M.; Bokor, J. Detection of nanomechanical vibrations by dynamic force
microscopy in higher cantilever eigenmodes. Appl. Phys. Lett. 2007,91, 053116. [CrossRef]
18.
Pan, S.; Zhang, H.; Liu, W.; Wang, Y.; Pang, W.; Duan, X. Biofouling Removal and Protein Detection Using a
Hypersonic Resonator. ACS Sens. 2017. [CrossRef] [PubMed]
19.
San-Paulo, A.; Quevy, E.; Black, J.; Howe, R.T.; White, R.; Bokor, J. Mode shape imaging of out-of-plane and
in-plane vibrating RF micromechanical resonators by atomic force microscopy. Microelectron. Eng.
2007
,84,
1354–1357. [CrossRef]
20.
Zhang, Z.; Wang, Y.; Zhang, H.; Tang, Z.; Liu, W.; Lu, Y.; Wang, Z.; Yang, H.; Pang, W.; Zhang, H.;
et al. Hypersonic Poration: A New Versatile Cell Poration Method to Enhance Cellular Uptake Using a
Piezoelectric Nano-Electromechanical Device. Small 2017,13. [CrossRef] [PubMed]
21.
Xu, L.Y.; Li, D.C.; Fu, X.; Hu, X.T. Characterization of the vibration behavior of nanobeam-resonator using
AFM dynamic mode. J. Harbin Inst. Technol. 2011,43, 393–397.
©
2017 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access
article distributed under the terms and conditions of the Creative Commons Attribution
(CC BY) license (http://creativecommons.org/licenses/by/4.0/).