Conference PaperPDF Available

Performance of Systolic Blood Pressure estimation from radial Pulse Arrival Time (PAT) in anesthetized patients

Authors:

Abstract and Figures

The performance of estimating Systolic Blood Pressure (SBP) in anesthetized patients via Pulse Arrival Time (PAT) techniques was studied with respect to the minimum required time in between two recalibration procedures. Materials: a clinical trial [NCT02651558] involving 14 patients was conducted measuring PAT from an ECG and an arterial line inserted into the radial artery. Methods: comparison of BP estimates from PAT measurements against invasive BP values was performed in terms of mean error and standard deviation of the error (AAMI/ANSI/ISO 81060-2), cumulative percentage of readings falling within 5, 10 and 15 mmHg (BHS criteria), and MAD-Mean Absolute Difference (IEEE Std 1708). Two calibration strategies were explored, involving time between recalibration periods ranging from 10 seconds to 8 minutes. Results: assuming an affine calibration function between PAT and SBP, different slope (Mean Slope:-1.45, CI:-1.64 to-1.27 mmHg/ms) and offset values (Mean Offset: 575, CI: 517 to 633 mmHg) were found in between patients. In addition, given a patient, affine calibration functions at different anesthesia phases also showed to be variable. When assessing agreement in terms of existing international standards it was found that PAT-based SBP estimates complied with requirements when time between two calibrations was smaller than 60 seconds. Conclusions: the use of anesthetic agents compromises the implementation of PAT-based techniques to estimate SBP.
Content may be subject to copyright.
Solà et al, 2017 - Performance of Systolic Blood Pressure estimation from radial Pulse Arrival Time (PAT) in anesthetized patients 1
Presented at EMBEC2017 - European Medical and Biological Engineering Conference 2017, Tampere (FI), 11-15 June 2017
Performance of Systolic Blood Pressure estimation
from radial Pulse Arrival Time (PAT) in anesthetized patients
J. Solà1, A. Vybornova1, F. Braun1, M. Proença1, R. Delgado-Gonzalo1, D. Ferrario1, C. Verjus1,
M. Bertschi1, N. Pierrel2 and P. Schoettker2
1 Systems Division, CSEM, Neuchâtel, Switzerland
2 Anesthesia Department, CHUV, Lausanne, Switzerland
AbstractThe performance of estimating Systolic Blood
Pressure (SBP) in anesthetized patients via Pulse Arrival Time
(PAT) techniques was studied with respect to the minimum re-
quired time in between two recalibration procedures.
Materials: a clinical trial [NCT02651558] involving 14 pa-
tients was conducted measuring PAT from an ECG and an ar-
terial line inserted into the radial artery.
Methods: comparison of BP estimates from PAT measure-
ments against invasive BP values was performed in terms of
mean error and standard deviation of the error
(AAMI/ANSI/ISO 81060-2), cumulative percentage of readings
falling within 5, 10 and 15 mmHg (BHS criteria), and MAD -
Mean Absolute Difference (IEEE Std 1708). Two calibration
strategies were explored, involving time between recalibration
periods ranging from 10 seconds to 8 minutes.
Results: assuming an affine calibration function between
PAT and SBP, different slope (Mean Slope: -1.45, CI: -1.64 to
-1.27 mmHg/ms) and offset values (Mean Offset: 575, CI: 517 to
633 mmHg) were found in between patients. In addition, given
a patient, affine calibration functions at different anesthesia
phases also showed to be variable. When assessing agreement in
terms of existing international standards it was found that PAT-
based SBP estimates complied with requirements when time be-
tween two calibrations was smaller than 60 seconds.
Conclusions: the use of anesthetic agents compromises the
implementation of PAT-based techniques to estimate SBP.
KeywordsBlood Pressure (BP), Pulse Arrival Time (PAT),
calibration function, anesthetized patients.
I. INTRODUCTION
There is large demand for new techniques to monitor Sys-
tolic Blood Pressure (SBP) in anesthetized patients in a con-
tinuous and non-invasive way. Current anesthesia protocols
require either the use of invasive arterial line monitoring
(providing continuous SBP measurements), or the inflation
of oscillometric cuffs (providing intermittent SBP measure-
ments every 60 seconds). Although controversial, Pulse Ar-
rival Time (PAT) has been suggested in the past to monitor
SBP [1]. The goal of this study is to provide figures of merit
of PAT-based techniques under the effects of anesthesia.
II. MATERIALS
Preliminary data from an on-going clinical trial [Clinical-
Trials.gov NCT02651558] were analysed. Patients were
placed on the operating table and monitoring was carried out
via the Philips monitor IntelliVue MX800. Monitoring con-
sisted of a 3-lead electrocardiogram, pulse oximetry at the
finger of the right hand, and an invasive recording of pressure
signals via a dedicated catheter inserted into the right radial
artery (BD Arterial Cannula 20G/1.1mmx45mm, Becton
Dickinson Infusion Therapy Syst. Inc., Utah USA).
Recording of raw data was performed using iXTrend Ex-
press (Ixellence GmbH, Wildaum, Germany) during 15
minutes of general anesthesia induction phase with target-
control infusion of Propofol as induction agent, and Remifen-
tanyl as opioid. Rocuronium was administered as a paralys-
ing agent before tracheal intubation. Decreases in BP were
managed with boluses of Ephedrin or Neosynephrin to coun-
teract the hemodynamic effect of anesthetic agents, and
adapted to patients physio-pathology and specifics. Data
from first 14 enrolled patients were used for the analysis.
Fig. 1 Correlation plots between PAT and Systolic BP for the 14 re-
cruited subjects: colored dots depict measurement points, and dashed lines
depict their associated affine calibration functions.
2 Solà et al, 2017 - Performance of Systolic Blood Pressure estimation from radial Pulse Arrival Time (PAT) in anesthetized patients
Presented at EMBEC2017 - European Medical and Biological Engineering Conference 2017, Tampere (FI), 11-15 June 2017
III. METHODS
The recorded ECG and arterial line signals were processed
every 10 seconds to calculate two time series. First, a time
series of systolic BP values: average systolic BP value within
each 10 seconds window. Second, a time series of PAT val-
ues: time delay between the ECG R-wave and the onset of
the pressure pulse waveform, calculated via Chiu’s method
[2] on the ensemble averaged arterial pulse (as for [3]).
For each patient, the calculated SBP and PAT time series
were post-processed in order to estimate an affine calibration
function relating SBP to PAT during time windows of differ-
ent lengths (ranging from 20 seconds to the entire recording).
Affine calibration functions of all patients were then com-
bined into a general-population calibration function averag-
ing the coefficients of each patient-dependent calibration.
Finally, for each patient, the PAT time series were trans-
formed into predicted SBP time series (pSBP) by applying
two calibration strategies. A first calibration strategy as-
sumed pSBP to be a constant value during a “time in-between
recalibration” window. The constant value was a cuff meas-
ured SBP value at the beginning of the window. This strategy
mimics a constant calibration strategy, as for a standard cuff-
based monitoring of a patient. A second calibration strategy
calculated pSBP within a window by applying an affine cal-
ibration function to the PAT time series. The calibration
function was assumed to be the general-population calibra-
tion function, with a corrected offset in order to match the
measured SBP value at the beginning of each window. This
strategy mimics a PAT-based estimation of SBP with recali-
brated offset values at every new performed SBP measure-
ment via an oscillometric cuff.
IV. RESULTS
Figure 1 provides a visual example of patient-dependent
calibration functions estimate over entire recordings.
A general-population calibration function was then calcu-
lated, relating PAT and SBP time series in the form of SBP =
Slope PAT +Offset, with average coefficients:
Slope: -1.45, 95% CI: -1.64 to -1.28 mmHg/ms
Offset: 575, 95% CI: 517 to 633 mmHg
Figure 2 provides examples of the temporal evolution of
slope estimates calculated from the PAT and the SBP series
within the analysis windows of 120 s. Dashed lines depict the
patient-dependent overall calibration function estimated over
the entire recording, bold black lines depict a reliable calibra-
tion function estimated at a given analysis window, and red
lines depict an unreliable local calibration function (i.e. asso-
ciated to non-significant correlation between SBP and PAT
time series). Temporal evolution of the slope for these illus-
trative patients demonstrates the large variability of local af-
fine calibration functions during anesthesia, associated to re-
current use of anesthetic agents acting on both arterial
muscular tonus and cardiac contractibility.
Figure 3 provides a performance assessment of the two
calibration strategies according to the figures of merit sug-
gested by the following existing international protocols:
IEEE Std 1708 [4]: the Mean Absolute Difference
score (MAD).
AAMI/ANSI/ISO 81060-2 [5]: the mean error and
the standard deviation of the error.
BHS Criteria [6]: the cumulative percentage of read-
ings falling within 5, 10 and 15 mmHg.
Fig. 2 Temporal evolution of local affine calibration functions between PAT and Systolic BP, with their associated slope coefficients. For a particular
patient, calibration functions were fitted for the entire recording (dashed lines), and within analysis windows of 120 seconds (solid lines).
Solà et al, 2017 - Performance of Systolic Blood Pressure estimation from radial Pulse Arrival Time (PAT) in anesthetized patients 3
Presented at EMBEC2017 - European Medical and Biological Engineering Conference 2017, Tampere (FI), 11-15 June 2017
.
Fig. 3 Performance assessment according to IEEE Std 1708, ISO 81060-2:2013 and BHS Criteria at predicting SBP following two strategies.
Upper panel depicts a standard cuff-based monitoring of a patient. Lower panel depicts PAT-based estimation of SBP with recalibrated offset val-
ues at every performed oscillometric cuff measurement. Mean and 95% CI values are calculated at different time in-between calibration windows.
4 Solà et al, 2017 - Performance of Systolic Blood Pressure estimation from radial Pulse Arrival Time (PAT) in anesthetized patients
Presented at EMBEC2017 - European Medical and Biological Engineering Conference 2017, Tampere (FI), 11-15 June 2017
For the standard cuff-based BP monitoring strategy of a
patient, it is found that requirements of ISO, BHS, and IEEE
are only met if the time between consecutive measurements
is lowered to 60 seconds. After this one-minute threshold, the
performance continuously decreases as time in between
measurements is increased.
For the PAT-based estimation of SBP with recalibrated
offset values, it is found that the same one-minute threshold
applies in order to meet the requirements of ISO, BHS and
IEEE. However, after this one-minute threshold, the perfor-
mance degrades until reaching a plateau around the two-mi-
nute time in between calibrations. This plateau is reached at
approximately:
IEEE Std 1708, MAD of 11 mmHg.
ISO 81060-2, mean error of -2.5 mmHg, and stand-
ard deviation of the error of 11 mmHg.
BHS Criteria, cumulative percentage of readings
falling within 5, 10 and 15 mmHg of respectively of
40%, 60% and 70%.
Figure 4 provides particular examples of SBP prediction
for the two studied strategies, when applying a two-minute
“time in-between recalibration” window. The illustrated pa-
tients correspond to those depicted in Figure 2.
V. DISCUSSION AND CONCLUSIONS
The presented results point at the fact that the use of anes-
thetic agents compromises the implementation of PAT-based
techniques to estimate SBP during anesthesia.
At least during anesthesia induction, the standard use of
anesthetic agents on a patient creates large variability of local
calibration functions. In order to fulfill the requirement of in-
ternational standard for BP measurements, such variability
would force recalibration procedures to be performed at an
interval of 60 seconds. Furthermore, within this 60 s win-
dows, performances of a PAT-based SBP monitoring did not
outperform the current SBP monitoring implemented anes-
thesia induction protocols, which may require, depending on
patients' physiologic status, an oscillometric cuff measure-
ment to be performed every 60 seconds.
The analyzed data provides valuable statistical analysis of
the coefficients of an affine calibration function relating SBP
and PAT time series in anesthetized patients. These results
might be used in the future to support the design of new clin-
ical trials in the field of cuffless blood pressure monitoring.
REFERENCES
1. Mukkamala et al, Toward Ubiquitous Blood Pressure Monitoring
via Pulse Transit Time: Theory and Practice. IEEE Trans Biomed
Eng. 2015 Aug;62(8):1879-901
2. Chiu et al, Determination of pulse wave velocities with comput-
erized algorithms. Am Heart J. 1991 May;121(5):1460-70
3. Sola, Continuous non-invasive blood pressure estimation. ETHZ
Thesis No 20093, 2011.
4. IEEE Std. 1708-2014 - IEEE Standard for Wearable Cuffless
Blood Pressure Measuring Devices.
5. ISO, ANSI/AAMI/ISO 81060-2:2013 Non-invasive sphygmoma-
nometers Part 2: Clinical investigation of automated measure-
ment type
6. ESH, Working Group on Blood Pressure Monitoring of the Euro-
pean Society of Hypertension International Protocol for valida-
tion of blood pressure measuring devices in adults, Blood Press
Monit. 2002 Feb;7(1):3-17
Author: Josep Solà
Institute: CSEM
Street: Jacquet-Droz, 1
City: CH-2002 Neuchâtel
Country: Switzerland
Email: Josep.Sola@csem.ch
Fig. 4 Three examples of SBP prediction during anesthesia when
applying strategy 1, i.e. standard cuff-based monitoring strategy, and
strategy 2, i.e. PAT-based estimation of SBP with recalibrated offset
values. Time in-between recalibration was set to two minutes.
... However, the changes in BP induced by these techniques are much lower than for exercise-induced changes 9 . There have been a few studies that have changed BP by medication infusion 14,15 . This has the advantage of achieving a wide range of BP values with minimal motion artefacts. ...
Article
Full-text available
Various models have been proposed for the estimation of blood pressure (BP) from pulse transit time (PTT). PTT is defined as the time delay of the pressure wave, produced by left ventricular contraction, measured between a proximal and a distal site along the arterial tree. Most researchers, when they measure the time difference between the peak of the R-wave in the electrocardiogram signal (corresponding to left ventricular depolarisation) and a fiducial point in the photoplethysmogram waveform (as measured by a pulse oximeter attached to the fingertip), describe this erroneously as the PTT. In fact, this is the pulse arrival time (PAT), which includes not only PTT, but also the time delay between the electrical depolarisation of the heart’s left ventricle and the opening of the aortic valve, known as pre-ejection period (PEP). PEP has been suggested to present a significant limitation to BP estimation using PAT. This work investigates the impact of PEP on PAT, leading to a discussion on the best models for BP estimation using PAT or PTT. We conducted a clinical study involving 30 healthy volunteers (53.3% female, 30.9 ± 9.35 years old, with a body mass index of 22.7 ± 3.2 kg/m2). Each session lasted on average 27.9 ± 0.6 min and BP was varied by an infusion of phenylephrine (a medication that causes venous and arterial vasoconstriction). We introduced new processing steps for the analysis of PAT and PEP signals. Various population-based models (Poon, Gesche and Fung) and a posteriori models (inverse linear, inverse squared and logarithm) for estimation of BP from PTT or PAT were evaluated. Across the cohort, PEP was found to increase by 5.5 ms ± 4.5 ms from its baseline value. Variations in PTT were significantly larger in amplitude, − 16.8 ms ± 7.5 ms. We suggest, therefore, that for infusions of phenylephrine, the contribution of PEP on PAT can be neglected. All population-based models produced large BP estimation errors, suggesting that they are insufficient for modelling the complex pathways relating changes in PTT or PAT to changes in BP. Although PAT is inversely correlated with systolic blood pressure (SBP), the gradient of this relationship varies significantly from individual to individual, from − 2946 to − 470.64 mmHg/s in our dataset. For the a posteriori inverse squared model, the root mean squared errors (RMSE) for systolic and diastolic blood pressure (DBP) estimation from PAT were 5.49 mmHg and 3.82 mmHg, respectively. The RMSEs for SBP and DBP estimation by PTT were 4.51 mmHg and 3.53 mmHg, respectively. These models take into account individual calibration curves required for accurate blood pressure estimation. The best performing population-based model (Poon) reported error values around double that of the a posteriori inverse squared model, and so the use of population-based models is not justified.
Article
Full-text available
Objective We aimed to examine the accuracy of noninvasively-derived peripheral arterial BP by Caretaker device (CT) against invasively measured arterial BP. We also examined the fidelity of heart rate variability by CT compared to ECG derived data. Design Prospective cohort study. Participants Adult surgical and trauma patients admitted to the ICU. Setting Academic tertiary care medical center. Interventions In a prospective manner, beat-by-beat BP by CT was recorded simultaneously with invasive arterial BP measured in patients in the intensive care unit. Invasive arterial BPs were compared with those obtained by the CT system. All comparisons between the CT data, arterial catheter data, and ECG data were post-processed. Measurements and Main Results From 37 enrolled patients, 34 were included with satisfactory data that overlapped between arterial catheter and CT. A total of 87,757 comparative data points were obtained for the 40 minute time window comparisons of the 34 patients, spanning approximately 22.5 hours in total. Systolic BP and diastolic BP correlations (Pearson's coefficient), as well as the mean difference (standard deviation) were, 0.92 and -0.36 (7.57) mmHg and 0.83 and -2.11 (6.00) mmHg, respectively. The overall inter-beat correlation was 0.99 with the mean difference between inter-beats obtained with the arterial BP and the CT was -0.056ms (6.0). Conclusions This study validates the non-invasive tracking of BP using the CT device and the pulse decomposition analysis approach is possible within the guidelines of the standard.
Thesis
Full-text available
Elevated Blood Pressure (BP) is a human-specific illness affecting a quarter of the worldwide population. Clinically known as hypertension, elevated BP is considered the major risk factor for cardiovascular disease: the most common cause of death in developed countries. Detecting, treating and controlling hypertension are major goals of modern medicine. For more than one century, the non-invasive measurement of BP has relied on the inflation of pneumatic cuffs around a limb, typically the upper arm. In addition to being occlusive, and thus cumbersome, clinical cuff-based methods provide intermittent BP readings, i.e. every twenty minutes, hence impeding the suitable monitoring of short-term BP regulation mechanisms. In addition, cuff-based methods may not yield representative BP during sleep as repeated inflations induce arousal reactions, leading to non-representative overestimated BP values. Therefore, the development of novel technologies that reduce the recurrent use of pneumatic cuffs is clearly justified. The goal of this thesis is to investigate novel non-invasive technologies for the continuous measurement of BP. Particular emphasis is given to non-occlusive technologies that can be used in ambulatory scenarios, during daily life activities and not only within hospitals or physicians’ offices. This thesis addresses thus the challenge of ambulatory BP monitoring from four different perspectives: cardiovascular physiology, clinical applicability, system integration, and signal/information processing. The thesis starts by reviewing basic concepts of cardiovascular physiology related to the control of BP in humans: this analysis aims at setting the background knowledge for the understanding of the challenges faced by the BP monitoring field. Currently existing approaches for the non-invasive monitoring of BP are then reviewed systematically: their principles of work and their respective advantages/limitations are identified from both, clinical and ambulatory perspectives. The studied approaches are the auscultatory, oscillometric, tonometric, volume-clamp and pulse wave velocity techniques. The introductory part of this thesis is completed with a comprehensive review of the metrological means for the non-invasive and non-occlusive monitoring of cardiovascular parameters that have been used for this research, namely: electro-cardiography (ECG), photo-plethysmography (PPG), phono-cardiography (PCG), impedance-cardiography (ICG) and electrical impedance tomography (EIT). The thorough appraisal of the state of the art identifies the Pulse Wave Velocity (PWV) technique as the most promising track to follow, since it provides the best trade-off between clinical and ambulatory compliances. PWV-based techniques rely on the fact that the velocity at which arterial pressure pulses propagate along the arterial tree depends on the underlying BP. Therefore, by continuously measuring PWV along the arterial tree one obtains beat-by-beat surrogate values of mean BP. However, this principle can only be exploited reliably if PWV is measured along central elastic arteries, where no vasomotion phenomenon exists. State-of-the-art metrological techniques to measure PWV are either unable to assess central PWVs, or not adapted for ambulatory monitoring. Therefore, to date the potential of PWV has not been fully exploited. The major contribution of this thesis is thus the development and testing of new PWV-based techniques for the continuous, non-invasive and non-occlusive measurement of BP. In particular this thesis explores two new strategies to assess the Pulse Transit Time (PTT) of pressure pulses along central segments of the arterial tree. These techniques are based on 1) the use of the EIT technology, and 2) a chest sensor implementing mature sensing technologies such as ECG, PPG, ICG and PCG. Electrical Impedance Tomography (EIT) is a non-invasive monitoring technology based on the analysis of bioimpedance signals. This thesis provides first proof that EIT applied at the chest skin is capable of providing information on the pulsatility of the aortic arterial wall, as if a virtual catheter was placed into the descending aorta. After constructing a novel algorithm for the identification of EIT pixels providing functional information on pulsatility of the descending aorta, and describing a method to estimate aortic Pulse Transit Times (PTT), experimental data on animal models is provided. Accordingly, EIT-derived PTT estimates show to highly correlate with invasive BP measurements (r=-0.967, p<0.00001) for a wide range of mean BP values (from 60 to 150 mmHg). A chest sensor for the measurement of central PWV values integrating multiple non-occlusive technologies is introduced by this thesis as well. The novel sensing approach relies on the detection of the opening of the aortic valve (genesis of pressure pulses) by the joint analysis of ECG, ICG and PCG time series, and the detection of the arrival time of the pulses at the subcutaneous vessels of the chest by the joint analysis of ECG and multi-channel PPG time series. This thesis describes the integration of the depicted sensing technologies in a single chest sensor, introduces new dedicated multi-parametric signal processing routines, and provides experimental data on humans. Accordingly, chest sensor-derived PWV estimates show to positively correlate with reference carotid-to-femoral PWV measurements (r=0.88, p<10-9) for a population of 31 normo- and hypertensive male subjects. The second contribution of this thesis is the introduction of a novel signal processing tool enabling the reliable determination of the arrival time of arterial pressure pulses even under noisy measurement conditions. Called parametric estimation of Pulse Arrival Times (PAT), this technique is a key element for the deployment of novel technologies aiming at measuring PWV at central arterial locations by means of non-invasive measuring means. This thesis introduces thus the new concept of PAT estimation via the fitting of parametric models to non-invasive arterial time series, and tests its agreement with state-of-the-art approaches. Accordingly, when evaluated on 200 hours of intensive care unit PPG signals, PAT values determined by the novel technique show to particularly correlate with PAT values determined by the state-of-the-art first derivative technique (r=0.99, p<0.001), while increasing its robustness to real motion noise and simulated multiplicative colored Gaussian noise. In conclusion, this thesis introduces a collection of novel technological and algorithmic strategies paving the way towards the deployment of devices for the ambulatory, continuous, non-invasive and non-occlusive measurement of BP.
Article
Ubiquitous blood pressure (BP) monitoring is needed to improve hypertension detection and control and is becoming feasible due to recent technological advances such as in wearable sensing. Pulse transit time (PTT) represents a wellknown, potential approach for ubiquitous BP monitoring. The goal of this review is to facilitate the achievement of reliable, ubiquitous BP monitoring via PTT. We explain the conventional BP measurement methods and their limitations; present models to summarize the theory of the PTT-BP relationship; outline the approach while pinpointing the key challenges; overview the previous work towards putting the theory to practice; make suggestions for best practice and future research; and discuss realistic expectations for the approach.
Article
Careful determination of pulse wave velocity is important in the study of arterial viscoelastic properties, wave reflections, and ventricular-arterial interactions. In spite of its increasingly widespread use, there is as yet no standardized method for its determination. Most studies have manually identified the transit time of the pressure wave front as it travels over a known distance in the arterial system, but the issues of accuracy and reproducibility have not been addressed. This study was designed to investigate the efficacy of four computerized algorithms in the determination of pulse wave velocities in invasive as well as in noninvasive pressure determinations. The four methods were the identification of: (1) the point of minimum diastolic pressure, (2) the point at which the first derivative of pressure is maximum, (3) the point at which the second derivative of pressure is maximum, and (4) the point yielded by the intersection of a line tangent to the initial systolic upstroke of the pressure tracing and a horizontal line through the minimum point. High-fidelity aortic pressure recordings were obtained in 26 patients with a multi-sensor micromanometer catheter. Noninvasive brachial and radial pressure waveforms were recorded in 11 volunteers with external piezoelectric transducers. The results show that the first derivative method consistently provided results that were different from the other methods for both the invasive and noninvasive methods because of changes in the structure of the upstroke as the arterial pulse propagates distally. Although the minimum method worked well for the invasive determinations, it was erratic with the noninvasive determinations, probably because of the higher amount of noise and reflection in the latter. Among the four algorithms, the second derivative and the intersecting tangents methods worked well with both invasive and noninvasive determinations with mean variation coefficients of less than 7% and correlation coefficients between the methods of greater than 0.90 for all data. In conclusion, computerized algorithms allow accurate determination of pulse wave velocity in invasively and noninvasively measured arterial pressure waveforms.
Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults
  • Working Esh
ESH, Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults, Blood Press Monit. 2002 Feb;7(1):3-17
Josep Solà Institute: CSEM Street: Jacquet-Droz
  • Author
Author: Josep Solà Institute: CSEM Street: Jacquet-Droz, 1
1708-2014 - IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices
  • Ieee Std
IEEE Std. 1708-2014 -IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices.