Water-analysis kits are useful for practical aquaculture only if they provide equivalent decision-making as compared to standard water-analysis methods. This study used weighted Cohen's kappa (κ) statistics to compare management decisions made by farmers who used water-analysis kits (e.g., Seneye slide kit, Tetra EasyStrips, API test strips, Seachem Ammonia Alert, Salifert Profi test kit, and Hach dissolved oxygen (DO) and alkalinity kit) and decisions made by those who used standard methods. The decisions made by farmers were similar for water-analysis kits and standard methods, except for Tetra and API test strips, when measuring nitrate concentrations. The highest conformity between the two methods (κ-value = 1.0, P < 0.0001) was obtained with the Hach and Salifert Profi test kits (for measuring DO) and the API test strip (for measuring total hardness). The rapid, simple measurements by the kits appear suitable for use by farmers if they are properly maintained and manufacturer's instructions are followed.
Content may be subject to copyright.
A preview of the PDF is not available
... WQS have been used to analyze various species of free chlorine, hydrogen sulfide and formaldehyde (Arsawiset and Teepoo, 2020), tetracycline antibiotics (Li et al., 2019), Hg (II) ions (Lan et al., 2018), Zn (II) ions (Takahashi, 2014), microcystins (Humpage et al., 2012), hypochlorite (Ballesta Claver et al., 2004), water hardness (Capitán-Vallvey et al., 2003) and microorganisms (Martins et al., 1997). The reliability of WQS for WQM was assessed against laboratory-based standard methods (Naigaga et al., 2016;Mosley and Donald, 2005;Bischoff et al., 1996). However, test strips lack selectivity of detection due to the random diffusion of chromogenic reagents in the detection zone in the presence of common interfering agents (Evans et al., 2014). ...
... The limited performance of the test strips for total alkalinity analyses may have resulted from the fact that the range of concentration of this parameter in Medjerda watershed has been often not high enough for assessing the reliability over their entire analytical range. Earlier comparison of the reliability of the WTS revealed that test strips generally gave a substantial agreement with the standard methods for total alkalinity (Naigaga et al., 2016). Nitrates test strips provide slight reliability in the 0-10 ppm NO 3 concentration range where one method indicated zero nitrates while the standard revealed measurable concentrations. ...
... The inferior reliability of test strips for nitrates compared to standard methods may be due to the presence of common interfering colloids agents (Forest et al., 2004). Other strips, such as API TM five-in-one aquarium and Tetra Easy TM six-in-one aquarium test strips, have been found not to agree with the standard methods for nitrates (Naigaga et al., 2016). The results show also that sodium chloridesensitive test strips are the most reliable for measuring a wide range of concentrations (0-7510 ppm) with the highest correlation parameters (PCC (0.962), R 2 (0.926), and α c (0.982)). ...
Chapter
Full-text available
This paper explores deeper dimensions of fluorosis and seeks to develop a clear communitydriven policy framework around fluorosis. It aims at redistributive policies that will lead to resource benefits to the local residents. The work will provide the foundation necessary to lobby for the implementation of policies that favor rainwater harvesting at household level. Exploitation of water for energy production needs to be redefined and readdressed properly. Naivasha geothermal exploration and fluorosis is used as the case study. Mount Longonot’s previous volcanic events were a blessing and a curse to the geographical area in that, it is the reason for geothermal potential and is also the generator of observed high fluoride aquifers of the area. Geothermal energy production has always required water in the form of steam to produce energy and therefore, legislation should be able to induce good quality and affordable consumption water as well as wastewater treatment in Naivasha.
... WQS have been used to analyze various species of free chlorine, hydrogen sulfide and formaldehyde (Arsawiset and Teepoo, 2020), tetracycline antibiotics (Li et al., 2019), Hg (II) ions (Lan et al., 2018), Zn (II) ions (Takahashi, 2014), microcystins (Humpage et al., 2012), hypochlorite (Ballesta Claver et al., 2004), water hardness (Capitán-Vallvey et al., 2003) and microorganisms (Martins et al., 1997). The reliability of WQS for WQM was assessed against laboratory-based standard methods (Naigaga et al., 2016;Mosley and Donald, 2005;Bischoff et al., 1996). However, test strips lack selectivity of detection due to the random diffusion of chromogenic reagents in the detection zone in the presence of common interfering agents (Evans et al., 2014). ...
... The limited performance of the test strips for total alkalinity analyses may have resulted from the fact that the range of concentration of this parameter in Medjerda watershed has been often not high enough for assessing the reliability over their entire analytical range. Earlier comparison of the reliability of the WTS revealed that test strips generally gave a substantial agreement with the standard methods for total alkalinity (Naigaga et al., 2016). Nitrates test strips provide slight reliability in the 0-10 ppm NO 3 concentration range where one method indicated zero nitrates while the standard revealed measurable concentrations. ...
... The inferior reliability of test strips for nitrates compared to standard methods may be due to the presence of common interfering colloids agents (Forest et al., 2004). Other strips, such as API TM five-in-one aquarium and Tetra Easy TM six-in-one aquarium test strips, have been found not to agree with the standard methods for nitrates (Naigaga et al., 2016). The results show also that sodium chloridesensitive test strips are the most reliable for measuring a wide range of concentrations (0-7510 ppm) with the highest correlation parameters (PCC (0.962), R 2 (0.926), and α c (0.982)). ...
Article
Full-text available
Citizen Science (CS) that is applicable to Water Quality Monitoring (WQM) is considered as an innovative approach to enhancement of the efficiency of hydrosystems’ monitoring. The main objective of this study was to assess the quality of a CS-based WQM program based on water test strips for the Medjerda River, the main hydrosystem of Tunisia. The study focused on total alkalinity, nitrates and sodium chloride content. Three citizens participated in WQM with test strips, for wet and dry periods, using 96 samples collected and monitored in 12 sampling sites located in the Medjerda watershed. Regression analysis, based on Pearson Correlation Coefficient (PCC), determination coefficient (R²), F test, and Cronbach’s alpha (αc) test were performed to statistically assess the difference between the citizens’ datasets and the standard values. The results show that sodium chloride-sensitive test strips are the most reliable for measuring sodium chloride within wide range concentrations (0-7510 ppm). Nitrates-sensitive test strips are moderately reliable in the 0-25 ppm range. The total alkalinity-sensitive test strips are less reliable in the 3-20°F range concentration. Results also demonstrate a good agreement between the citizen-based outcomes and the analytical methods for the three citizens. However, the reliability depends on the citizen and the parameter to be tested. The test strips appear to be a suitable tool for WQM of the Medjerda watershed by citizens if the instructions of use are strictly followed. The latter can be achieved by a more consistent citizen training.
... Sudden changes in three water quality parameters taken in aggregate characterized a decline in water quality that was recognized as needing immediate remediation to minimize impending impacts on coral health (Delbeek and Sprung 1994;Borneman 2001). The accuracy of colorimetric rapid test kits relative to standard validated bench chemistry methods varies substantially by test, but they generally provide sufficient precision to support water management decisions in aquaculture, and their use is a common practice in small-to medium-sized systems to support water quality management decisions made by biologists and farmers (Naigaga et al. 2016). To optimize accuracy and precision, it is important to use kits that are in date and properly stored, and to follow the directions carefully (Naigaga et al. 2016), as was done for the system in this study. ...
... The accuracy of colorimetric rapid test kits relative to standard validated bench chemistry methods varies substantially by test, but they generally provide sufficient precision to support water management decisions in aquaculture, and their use is a common practice in small-to medium-sized systems to support water quality management decisions made by biologists and farmers (Naigaga et al. 2016). To optimize accuracy and precision, it is important to use kits that are in date and properly stored, and to follow the directions carefully (Naigaga et al. 2016), as was done for the system in this study. Quality control of API commercial test kits to assure agreement with laboratory prepared standards is conducted by lot prior to distribution to consumers by the manufacturer, Mars Fishcare (GJ. ...
Article
A reliable marker of early coral response to environmental stressors can help guide decision-making to mitigate global coral reef decline by detecting problems before the development of clinically observable disease. We document the accumulation of acrylic acid in two divergent coral taxa, stony small polyp coral ( Acropora sp.) and soft coral ( Lobophytum sp.), in response to deteriorating water quality characterized by moderately increased ammonia (0.25 ppm) and phosphate (0.15 ppm) concentrations and decreased calcium (360 ppm) concentration, using nuclear magnetic resonance spectroscopy (NMR)-based metabolomic techniques. Changes in acrylic acid concentration in polyp tissues free of zooxanthellae suggest that acrylic acid could be a product of animal metabolism and not exclusively a metabolic by-product of the osmolyte dimethylsulfoniopropionate (DMSP) in marine algae or bacteria. Our findings build on previously documented depletions of acrylic acid in wild coral potentially correlated to temperature stress and provide additional insight into approaches to further characterize the nature of the metabolic accumulation of acrylic acid under controlled experimental conditions.
... Most instances of fish kills, disease outbreaks, poor growth, and poor feed-conversion efficiency are directly or indirectly related to poor water quality. The most important water-quality variables in aquaculture usually are dissolved oxygen (DO), unionized ammonia (NH3), nitrite, nitrate, total alkalinity, total hardness, and pH (Naigaga, Boyd, Gaillard, Abdelrahman, & Molnar, 2017). The most important treatment of water in recirculation aquaculture systems (RAS) is the biological process that uses the nitrification process to convert toxic ammonia to less toxic nitrate. ...
... The use of RAS has contributed immensely to the development of aquaculture due to its various advantages, including reduced water use through partial re-use of cultured water, reduced environmental impacts from fish farming systems through improved waste management and nutrient recycling, improved sanitation, reduced disease outbreaks, and limited biological control (Dauda, Natrah, Karim, Kamarudin, & Bichi, 2018). The Water-analysis kit (Hach DO kit) can be used to make acceptable decisions related to water-quality management in aquaculture since this test kit gives results in κ value ≥ 0.6 (preferred result) (Naigaga et al., 2017). ...
Article
Full-text available
Fish farming is considered an important agricultural activity that is capable of ending nutritional deficiencies of the world and contributing to poverty reduction. Its proponents argue that aquaculture will meet the food security needs of millions of people in developing countries who will benefit from relatively inexpensive protein, while depleted capture fisheries are allowed to replenish. Egypt is Africa's leading aquaculture country, followed by Nigeria with similar production systems. The dominant species of fish cultured in Egypt and Nigeria are Tilapia and African catfish, respectively. The aquaculture industries in these two counties are very promising, as there is a presence of water bodies, some institutional commitment, and high demand for fish among others. Despite some gains made by the countries and the huge potential of the aquaculture sector, it is however bedeviled with constraints such as low technology adoption, inadequate supply of fingerlings, high cost of fish feed, etc. This work is an overview of the aquaculture sectors of Egypt and Nigeria, which includes the production systems, prospects, and potentials of the sector and the constraints that affect aquaculture.
... Colorimetric tests require a colorimeter (about $1500 in 2022) and are easy to use because they use preformulated reagents (available from several manufacturers, e.g., LaMotte and Hach). These test kits provide accurate measurements of all the macronutrient ions, and for most of the micronutrient ions [79][80][81], but the kits are expensive, and adding the colorimetric reagents to each sample is labor intensive. ...
Article
Full-text available
Mass balance principles are a cornerstone of efficient fertilizer use and can be utilized to optimize plant nutrition without discarding or leaching solution. Here, we describe the maintenance of closed hydroponic and soilless substrate systems based on mass balance. Water removed by transpiration is restored with solution that replaces the nutrients that were taken up with the water. The concentration of nutrients in this refill/irrigation solution is determined by multiplying the optimal concentration of each nutrient in plant tissue by the water-use efficiency (WUE; ratio of dry mass to water transpired). Optimal leaf nutrient concentrations are well established, but WUE in controlled environments varies widely and is less well characterized. Elevated CO2 increases photosynthesis and demand for nutrients, but partially closes stomata and reduces transpiration; so high CO2 dramatically increases WUE. The concentration of the refill/irrigation solution must be adjusted to account for a two-fold range of WUE, from 3 g L−1 in ambient CO2 in lower humidity, to 6 g L−1 in elevated CO2 in higher humidity. WUE and nutrient requirements vary during the vegetative and reproductive stages of growth, and adjustment of the solution over the lifecycle can be beneficial. Measurement of solution electrical conductivity (EC) is helpful, but if the solution is appropriate, low EC usually means healthy plants and active nutrient uptake. The ammonium to nitrate ratio is critical to pH management. We have applied these principles across multiple species and environments to achieve long-term, steady-state nutrient concentrations with no discharge or leaching of solution.
... For example, mobile phones with GPS, photography, and social media applications have proven effective in collecting data on the color and transparency of surface waters by citizens, as has been implemented in the River Chief Mechanism (Zheng et al., 2017). Simple and inexpensive water test kits, available from many pet stores or online retailers, are useful to small farmers when preventing water quality degradation in fish ponds (Naigaga et al., 2017). High-throughput sequencing of DNA extracted from water samples serves as a quick, cost-effective and standardized method for scientists and conservation agencies when evaluating rare and threatened species across a wide range of taxonomic groups (Thomsen et al., 2012). ...
Article
Farm ponds, which are sometimes numerous and widely distributed in agricultural regions, have faced widespread degradation in recent decades. Although conservation strategies for these biodiversity hotspots have gradually increased, appropriate approaches for developing country contexts are lacking. Farm ponds provided hydrologic, biogeochemical, and socioeconomic benefits to southern China for thousands of years, but they are facing contemporary threats and management challenges, including (1) inadequate planning in terms of construction and conservation regulations; (2) rural nonpoint source and mini-point source pollution; (3) climate change induced abnormalities in the hydroperiod and disturbance to wildlife; (4) invasive species; and (5) inadequate social and political capacity to consider ecological conservation. Because farm ponds function as wetland complexes that are embedded within or integral to larger ecosystems, their conservation requires collaborative efforts over scales ranging from within-pond to regional. We highlight approaches that build public awareness and involve inventory maps as a basis. Policies that integrate top-down regulation and bottom-up engagement and emphasize sustainable management and utilization are recommended to ensure the effectiveness and continuous improvement of conservation programs. Techniques that involve interconnected smart sensors, volunteering and citizen science, and integrated process-based modeling are preferred when conducting comprehensive descriptions of the pond landscape, numerical assessments on their ecosystem services, and associated conservation cost analyses. Nature-based solutions are increasingly recognized as an important opportunity for coping with water-related crises. This paper presents the first synthetic perspective on the ecological roles of farm ponds in agriculturally dominated developing countries. The analytical framework and conservation suggestions are referential to sustainable rural development and the management of other small, scattered wetlands.
Article
Full-text available
A Hach DR-EL/2 Direct Reading Engineer's Laboratory Kit (Hach Chemical Co., Ames, Iowa) was evaluated by comparison to standard analytical methods. Results of Hach kit methods were highly correlated with concentrations of standard solutions. The y-intercepts (a) did not differ from 0 for most methods, but for 11 methods regression coefficients (b) differed from 1.0. For each of the 20 parameters, three natural water samples were analyzed seven times by Hach kit and by the standard method. Mean values obtained by the two analytical systems were usually significantly different and the standard method sometimes gave higher precision than the Hach kit. Nevertheless, the Hach kit usually gave close approximations of values obtained by standard methods. Comparisons also included analyses of a large series of natural water samples. Correlation coefficients between Hach kit and standard methods were often > 0.99. Although a = 0 for 14 methods, b = 1.0 for only 8 methods. Reliability of the Hach kit was adequate for general surveys of water quality, fisheries management decisions, and research requiring only approximate data on water quality. Please view the pdf by using the Full Text (PDF) link under 'View' to the left. Copyright © . .
Article
Full-text available
Water analysis kits (Bausch and Lomb SpectroKits TM, Ecologic test kits, and CHEMetrics test kits) were evaluated by comparison with standard analytical methods. Results of analyses with water analysis kits were highly correlated with concentrations of standard solutions. In replicate analyses of pond water, kits gave results that often differed from, and usually were less precise than, standard methods. When large series of water samples were analyzed, values from kits and standard methods were highly correlated (r > 0.85; P < 0.01), but regression intercepts seldom were 0 and slopes seldom were 1.0. Thus, water analysis kits are not suitable when high accuracy and precison are necessary, but kits are sufficiently reliable in many applications.
Article
An assessment was conducted to compare the precision and accuracy of Nessler, phenate, salicylate, and ammonia electrode procedures for total ammonia nitrogen (TAN) concentration determination in waters of aquaculture. Salicylate method was selected as a standard method for its high precision and accuracy. In replicate analyses of water samples for precision estimate, Nessler method usually gave higher mean concentration of TAN than other methods and it is not recommended for use for its high coefficient of variance. Results from Nessler's with Rochelle salt method were more accurate than Nessler without Rochelle salt method. Electrodes for sensing NH3 and NH4+ usually gave higher mean concentration of TAN than salicylate method. NH4+ electrode, although with high accuracy, was strongly affected by sodium and potassium and not recommended to use especially in waters of high salinity. The finding also suggested that the salicylate kit by using YSI photometer also was a preferable alternative to salicylate method while Nessler kit was not. TAN concentrations measured in 27 water samples by all methods were highly correlated (R2=0.919 to 0.996) with salicylate method. All of the slopes except phenate I and II method were different from 1.0 (P<0.05); all of the intercepts except phenate I and salicylate kit were different from 0.0 (P<0.05). Statement of relevance: In this study, we evaluated the accuracy and precision of different possible ways of measuring TAN concentration at aquaculture facilities. These findings will be useful because producers need reliable measurement of TAN concentration to effectively manage ponds and other culture systems. The conclusions made in this manuscript provide valuable information when researchers and farmers need to measure TAN concentration.
Chapter
There is a consensus among aquaculturists that water circulation in ponds is beneficial. Water circulation prevents thermal and chemical stratification. This makes the entire pond volume habitable for aquatic animals, and it eliminates the danger of thermal overturns in deep ponds. Water circulation devices create surface turbulence and this causes a small degree of aeration. Air-lift pumps use air bubbles to move water, and some oxygenation is affected by the rising bubbles. Water circulators should not be considered aerators in the usual sense. The greatest influence of water circulators on dissolved oxygen concentration is the blending of surface water with subsurface water. During daylight hours, surface water in ponds often is supersaturated with dissolved oxygen, and water at greater depths may have a low dissolved oxygen concentration. By mixing pond water, a uniform dissolved oxygen profile can be established. Oxygen produced by phytoplankton is partially conserved by water mixing, because the high degree of dissolved oxygen supersaturation normally found at pond surfaces during daylight is eliminated. Circulation of pond water also may stimulate phytoplankton growth (Sanares et al. 1986), and this could possibly increase dissolved oxygen production by photosynthesis.
Article
Methods of analysis used in the comparison of two methods of measurement are reviewed. The use of correlation, regression and the difference between means is criticized. A simple parametric approach is proposed based on analysis of variance and simple graphical methods.