An open access database for the evaluation of heart sound algorithms

Authors:

Abstract and Figures

In the past few decades, analysis of heart sound signals (i.e. the phonocardiogram or PCG), especially for automated heart sound segmentation and classification, has been widely studied and has been reported to have the potential value to detect pathology accurately in clinical applications. However, comparative analyses of algorithms in the literature have been hindered by the lack of high-quality, rigorously validated, and standardized open databases of heart sound recordings. This paper describes a public heart sound database, assembled for an international competition, the PhysioNet/Computing in Cardiology (CinC) Challenge 2016. The archive comprises nine different heart sound databases sourced from multiple research groups around the world. It includes 2435 heart sound recordings in total collected from 1297 healthy subjects and patients with a variety of conditions, including heart valve disease and coronary artery disease. The recordings were collected from a variety of clinical or nonclinical (such as in-home visits) environments and equipment. The length of recording varied from several seconds to several minutes. This article reports detailed information about the subjects/patients including demographics (number, age, gender), recordings (number, location, state and time length), associated synchronously recorded signals, sampling frequency and sensor type used. We also provide a brief summary of the commonly used heart sound segmentation and classification methods, including open source code provided concurrently for the Challenge. A description of the PhysioNet/CinC Challenge 2016, including the main aims, the training and test sets, the hand corrected annotations for different heart sound states, the scoring mechanism, and associated open source code are provided. In addition, several potential benefits from the public heart sound database are discussed.
Content may be subject to copyright.
1
An Open Access Database for the Evaluation of
Heart Sound Algorithms
Chengyu Liu1, David Springer2, Qiao Li1, Benjamin Moody3, Ricardo Abad Juan4,5, Francisco J Chorro6,
Francisco Castells5, José Millet Roig5, Ikaro Silva3, Alistair E.W. Johnson3, Zeeshan Syed7, Samuel E.
Schmidt8, Chrysa D. Papadaniil9, Leontios Hadjileontiadis9, Hosein Naseri10, Ali Moukadem11, Alain
Dieterlen11, Christian Brandt12, Hong Tang13, Maryam Samieinasab14, Mohammad Reza Samieinasab15,
Reza Sameni14, Roger G. Mark3, Gari D. Clifford1,4*
Affiliations:
1 Department of Biomedical Informatics, Emory University, USA
2 Institute of Biomedical Engineering, Department of Engineering Science, University of Oxford, UK
3 Institute for Medical Engineering & Science, Massachusetts Institute of Technology, USA
4 Department of Biomedical Engineering, Georgia Institute of Technology, USA
5 ITACA Institute, Universitat Politecnica de Valencia, Spain
6 Service of Cardiology, Valencia University Clinic Hospital, INCLIVA, Spain
7 Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI, USA
8 Department of Health Science and Technology, Aalborg University, Denmark
9 Department of Electrical and Computer Engineering, Aristotle University of Thessaloniki, Greece
10 Department of Mechanical Engineering, K. N. Toosi University of Technology, Iran
11 MIPS Laboratory, University of Haute Alsace, France
12 Hospital University of Strasbourg, France
13 Faculty of Electronic and Electrical Engineering, Dalian University of Technology, China
14 School of Electrical & Computer Engineering, Shiraz University, Shiraz, Iran
15 Department of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran
* Corresponding author: Gari D. Clifford, E-mail: gari@gatech.edu
2
Abstract: In the past few decades, analysis of heart sound signals (i.e., the phonocardiogram or PCG),
especially for automated heart sound segmentation and classification, has been widely studied and has
been reported to have the potential value to detect pathology accurately in clinical applications. However,
comparative analyses of algorithms in the literature have been hindered by the lack of high-quality,
rigorously validated, and standardized open databases of heart sound recordings. This paper describes a
public heart sound database, assembled for an international competition, the PhysioNet/Computing in
Cardiology (CinC) Challenge 2016. The archive comprises nine different heart sound databases sourced
from multiple research groups around the world. It includes 2,435 heart sound recordings in total
collected from 1,297 healthy subjects and patients with a variety of conditions, including heart valve
disease and coronary artery disease. The recordings were collected from a variety of clinical or
nonclinical (such as in-home visits) environments and equipment. The length of recording varied from
several seconds to several minutes. This article reports detailed information about the subjects/patients
including demographics (number, age, gender), recordings (number, location, state and time length),
associated synchronously recorded signals, sampling frequency and sensor type used. We also provide a
brief summary of the commonly used heart sound segmentation and classification methods, including
open source code provided concurrently for the Challenge. A description of the PhysioNet/CinC
Challenge 2016, including the main aims, the training and test sets, the hand corrected annotations for
different heart sound states, the scoring mechanism, and associated open source code are provided. In
addition, several potential benefits from the public heart sound database are discussed.
Key words: heart sound; phonocardiogram (PCG); database; heart sound classification; heart sound
segmentation; PhysioNet/CinC Challenge
1. Introduction
Cardiovascular diseases (CVDs) continue to be the leading cause of morbidity and mortality worldwide.
An estimated 17.5 million people died from CVDs in 2012, representing 31% of all global deaths (WHO,
2015). One of the first steps in evaluating the cardiovascular system in clinical practice is physical
examination. Auscultation of the heart sounds is an essential part of the physical examination and may
reveal many pathologic cardiac conditions such as arrhythmias, valve disease, heart failure, and more.
Heart sounds provide important initial clues in disease evaluation, serve as a guide for further diagnostic
examination, and thus play an important role in the early detection for CVDs.
During the cardiac cycle, the heart first experiences electrical activation, which then leads to mechanical
activity in the form of atrial and ventricular contractions. This in turn forces blood between the chambers
of the heart and around the body, as a result of the opening and closure of the heart valves. This
mechanical activity, and the sudden start or stop of the flow of blood within the heart, gives rise to
vibrations of the entire cardiac structure (Leatham, 1975). These vibrations are audible on the chest wall,
and listening for specific heart sounds can give an indication of the health of the heart. An audio
recording (or graphical) time series representation of the resultant sounds, transduced at the chest surface
is known as a heart sound recording or phonocardiogram (PCG).
3
Four locations are most often used to listen to and transduce the heart sounds, which are named according
to the positions in which the valves can be best heard (Springer, 2015):
Aortic area - centred at the second right intercostal space.
Pulmonic area - in the second intercostal space along the left sternal border.
Tricuspid area - in the fourth intercostal space along the left sternal edge.
Mitral area - at the cardiac apex, in the fifth intercostal space on the midclavicular line.
Fundamental heart sounds (FHSs) usually include the first (S1) and second (S2) heart sounds (Leatham,
1975). S1 occurs at the beginning of isovolumetric ventricular contraction, when already closed mitral
and tricuspid valves suddenly reach their elastic limit due to the rapid increase in pressure within the
ventricles. S2 occurs at the beginning of diastole with the closure of the aortic and pulmonic valves (See
Figure 1.) While the FHSs are the most recognizable sounds of the heart cycle, the mechanical activity of
the heart may also cause other audible sounds, such as the third heart sound (S3), the fourth heart sound
(S4), systolic ejection click (EC), mid-systolic click (MC), the diastolic sound or opening snap (OS), as
well as heart murmurs caused by turbulent, high-velocity flow of blood.
Figure 1. Phonocardiograms (above) from normal and abnormal heart sounds with pressure diagrams
(below). Red indicates aortic pressure, green ventricular pressure and blue atrial pressure. Reproduced
under the CC BY-SA 3.0 license and adapted from (Madhero, 2010).
4
The spectral properties of heart sounds and PCG recording artifacts have been well described (Leatham,
1975). The upper panel of Figure 2 shows the frequency distribution examples of different components in
heart sound (A from a normal heart sound and B from a heart sound with S3 component, both recorded at
the tricuspid area). As shown, the S1, S2, S3 and S4 components overlap with each other in the frequency
domain. Similarly, murmurs and artifacts from respiration and other non-physiological events also overlap
significantly. Arrows indicate (theoretical) typical frequency regions for each type of heart sound: S1 for
10-140 Hz (energy concentration usually in low frequencies of 25-45 Hz), S2 for 10-200 Hz (energy
concentration usually in low frequencies of 55-75) and S3 & S4 for 20-70 Hz. Murmurs tend to manifest
diverse frequency ranges and depending on their nature they can be as high as 600 Hz. Respiration
usually has a frequency range of 200-700 Hz (Tilkian and Conover, 2001). This makes the separation of
heart sounds from each other, and from abnormal sounds or artifacts, impossible in the frequency domain.
The morphological similarity of the noise to normal and abnormal heart sounds makes identification of
the latter also extremely difficult in the time domain. The lower panel of Figure 2 shows the sound
pressure levels for different frequency ranges.
5
Figure 2. General spectral regions for different heart sounds, and other physiological sounds during heart
sound recordings. Adapted from (Springer, 2015; Leatham, 1975).
Heart sound
recordings Step 1
Pre-processing Step 2
Segmentation Step 3
Classification Clinical
applications
Ø To assess the signal
quality
Ø To filter out baseline
changes and high
frequency noises
Ø To extract relevant
features
Ø To delineate the start
and end of each phase
of the heart beat (S1,
systolic, S2, diastolic,
etc.)
Ø To map the features for
each segmented phase of
the beat to a known phase
or sound, or the entire
recording to a pathology.
Sometimes quality
classifications are made on
sections or entire
recordings.
Figure 3. Typical three steps for automated analysis of heart sound in clinical applications.
Automated analysis of the heart sound in clinical applications usually consists of three steps shown in
Figure 3; pre-processing, segmentation and classification. Over the past few decades, methods for
automated segmentation and classification of heart sounds have been widely studied. Many methods have
demonstrated potential to accurately detect pathologies in clinical applications. Unfortunately,
comparisons between techniques have been hindered by the lack of high-quality, rigorously validated, and
standardized databases of heart sound signals obtained from a variety of healthy and pathological
conditions. In many cases, both experimental and clinical data are collected at considerable expense, but
6
only analyzed once by their collectors and then filed away indefinitely, because funding climates change,
and collaborators move on. Moreover, the activation energy needed to document data for external use,
store and share data in a semi-permanent manner is rarely available at the end of a research project.
The PhysioNet/Computing in Cardiology Challenge 2016 (PhysioNet/CinC Challenge 2016) attempts to
address some of these issues by assembling the research community to contribute multiple promising
databases (2016; Clifford et al., 2016). Prior to the PhysioNet/CinC Challenge 2016 there were only three
public heart sound databases available: i) The Michigan Heart Sound and Murmur database (UMHS), ii)
The PASCAL database (Bentley et al., 2011) and iii) The Cardiac Auscultation of Heart Murmurs
database (eGeneralMedical). These three databases can be summarized as follows:
The Michigan Heart Sound and Murmur database (MHSDB) was provided by the University of Michigan
Health System. It includes only 23 heart sound recordings with a total of time length of 1496.8 s and is
available from http://www.med.umich.edu/lrc/psb/heartsounds/index.htm
The PASCAL database comprises 176 recordings for heart sound segmentation and 656 recordings for
heart sound classification. Although the number of the recordings is relatively large, the recordings have
the limited time length from 1 s to 30 s. They also have a limited frequency range below 195 Hz due to
the applied low-pass filter, which removes many of the useful heart sound components for clinical
diagnosis. It is available from http://www.peterjbentley.com/heartchallenge
The Cardiac Auscultation of Heart Murmurs database is provided by eGeneral Medical Inc., includes 64
recordings. It is not open and requires payment for access from:
http://www.egeneralmedical.com/listohearmur.html
It is important to note that these three databases are limited by the recording number, length or signal
frequency range. In addition, two of these databases are intended to teach medical students auscultation,
and therefore comprise high-quality recordings of very pronounced murmurs, not often seen in real-world
recordings. In the PhysioNet/CinC Challenge 2016, a large collection of heart sound recordings was
obtained from different real-world clinical and nonclinical environments (such as in-home visits). The
data include not only clean heart sounds but also very noisy recordings, providing authenticity to the
challenge. The data were recorded from both normal subjects and pathological patients, and from both
children and adults. The data were also recorded from different locations, depending on the individual
protocols used for each data set. However, they were generally recorded at the four common recording
locations of aortic area, pulmonic area, tricuspid area and mitral area. Although a limited portion of the
data has been held back for test purposes (Challenge scoring), much of the hidden test data will be
released on PhysioNet after the conclusion of the Challenge and subsequent special issue in the Journal
Physiological Measurement. The purpose of this paper is to provide a detailed description of the heart
sound data that comprise the training and test sets for the PhysioNet/CinC Challenge 2016, and to help
researchers improve their algorithms in the Official Phase of the Challenge.
2. Description of the assembled heart sound databases
Table 1 details the composition of the assembled heart sound database. There are a total of nine heart
sound databases collected independently by seven different research teams from seven countries and three
7
continents, over a period of more than a decade. As a result, the hardware, recording locations, data
quality and patient types differ substantially, and the methods for identifying gold standard diagnoses also
vary. A description of each composite database is now given. The acoustic data were saved in either the
text format or the .wav format.
2.1. MIT heart sounds database
The Massachusetts Institute of Technology heart sounds database (hereafter referred to as MITHSDB)
was contributed by Prof John Guttag, Dr Zeeshan Syed and colleagues. An extensive description of the
data can be found in Syed et al. (Syed, 2003; Syed et al., 2007). Heart sounds were recorded
simultaneously with an electrocardiogram (ECG) using a Welch Allyn Meditron electronic stethoscope
(Skaneateles Falls, New York, USA), with a frequency response of 20 Hz to 20 kHz. Both PCG and ECG
signals were sampled at 44,100 Hz with 16-bit quantization. A total of 409 PCG recordings were made at
nine different recording positions and orientations from 121 subjects. Each subject contributed several
recordings. The subjects were divided into 5 groups: 1) normal control group: 117 recordings from 38
subjects, 2) murmurs relating to mitral valve prolapse (MVP): 134 recordings from 37 patients, 3)
innocent or benign murmurs group (Benign): 118 recordings from 34 patients, 4) aortic disease (AD): 17
recordings from 5 patients, and 5) other miscellaneous pathological conditions (MPC): 23 recordings
from 7 patients. The diagnosis for each patient was verified through echocardiographic examination at the
Massachusetts General Hospital, Boston, MA, USA. These recordings were either performed during in-
home visits or in the hospital, and were performed in an uncontrolled environment, resulting in many of
the recordings being corrupted by various sources of noise, such as talking, dogs barking and children
playing. Other noise sources included stethoscope motion, breathing and intestinal sounds. The recording
length varied from 9 s to 37 s, with mean and standard deviation (SD) of 33±5 s. For the purposes of the
competition, the ECGs were extracted and stored in a WFDB-compliant format.
2.2. AAD heart sounds database
The Aalborg University heart sounds database (AADHSDB) was contributed by Dr. Samuel E. Schmidt
and colleagues (Schmidt et al., 2010a; Schmidt et al., 2015; Schmidt et al., 2010b). Heart sound
recordings were made from the 4th intercostal space at the left sternal border on the chest of subjects
using a Littmann E4000 electronic stethoscope (3M, Maplewood, Minnesota). The frequency response of
the stethoscope was 20 to 1,000 Hz. The sample rate was 4,000 Hz with 16-bit quantization. A total of
151 subjects were recorded from patients were referred for coronary angiography at the Cardiology
Department at Aalborg Hospital, Denmark. The aim of the study was diagnosis of coronary artery disease
(CAD) from heart sound, however in the current database normal and abnormal are defined base on if the
patient has a heart valve defect either identified in the patient record or identified by a clear systolic or
diastolic murmur. A total of 30 subjects had heart valve defect and where defined as abnormal. Patients
were asked to breathe normally during the heart sound acquisition and between one and six PCG
recordings were collected from each subject, resulting in a total of 695 recordings. Most of the recordings
have a fixed time length of 8 s while a few recordings have a time length less than 8 s.
8
2.3. AUTH heart sounds database
The Aristotle University of Thessaloniki heart sounds database (AUTHHSDB) was contributed by Dr.
Chrysa D. Papadaniil and colleagues (Papadaniil and Hadjileontiadis, 2014). Heart sounds were recorded
in the first Cardiac Clinic of Papanikolaou General Hospital in Thessaloniki, Greece, using
AUDIOSCOPE, a custom-made electronic stethoscope that records signals amplified and unfiltered. The
sample rate was 4,000 Hz with 16-bit quantization. Forty-five subjects were enrolled within an age range
of 18-90 years; in particular, 11 normal subjects, 17 patients with aortic stenosis (AS) and 17 patients with
mitral regurgitation (MR). The diagnosis and the severity of the heart valve diseases were determined by
the doctors, based on the echocardiogram of the patient. The recordings were recorded from the
auscultation position of the chest where the murmur is best heard for each valve dysfunction, while the
normal heart sounds were recorded from the apex. Each subject gave one PCG recording (total 45
recordings) and the recordings had varied time length from 10 s to 122 s (mean±SD: 50±26 s).
2.4. TUT heart sounds database
The K. N. Toosi University of Technology heart sounds database (TUTHSDB) was contributed by Dr.
Hosein Naseri (Naseri and Homaeinezhad, 2013; Naseri et al., 2013). It includes a total of 28 healthy
volunteers and 16 patients with different types of valve diseases. The actual diagnoses were determined
by echocardiography prior to recording of PCG signals. PCG signals were recorded by using an electronic
stethoscope (3M Littmanns 3200) at four different locations (not simultaneously): pulmonic, aortic,
tricuspid and apex at a sampling rate of 4,000 Hz with 16-bit amplitude resolution for exactly 15 s each.
Two subjects only had 3 PCG recordings, resulting in a total of 174 PCG recordings.
2.5. UHA heart sounds database
The University of Haute Alsace heart sounds database (UHAHSDB) was contributed by Dr. Ali
Moukadema (Moukadem et al., 2013; Moukadem et al., 2011). Heart sound signals were recorded using
prototype stethoscopes produced by Infral Corporation (Strasbourg, France). The sample rate was 8,000
Hz with 16-bit quantization. The dataset contains total 79 PCG recordings, including 39 normal sounds
and 40 pathological cardiac sounds. The normal sound recordings were separated into two sub-files:
‘NHC’ (19 recordings) and ‘MARS500’ (20 recordings). ‘NHC’ recordings were collected from 19
normal subjects, aged from 18 to 40 years. The recording length varied from 7 s to 29 s (mean±SD: 14±5
s). ‘MARS500’ recordings were collected from 6 volunteers (astronauts), dedicating to the Cardio-Psy
experience as a part of the MARS500 project (IBMP–Russia) promoted by European Spatial Agency. The
recording length varied from 7 s to 17 s (mean±SD: 10±3 s). The pathologic recordings were from 30
patients (10 female and 20 male), who were recruited during hospitalization in the Hospital of Strasbourg.
They were aged from 44 to 90 years. Ten of them were recorded twice generally before and after valvular
surgery. The diagnoses of the pathologic patients were made by an experienced cardiologist using
additional information from the ECG and echocardiography-Doppler. Among 30 patients, 9 patients had
prosthetic valves with 1 bioprosthesis, 4 patients had double prostheses (in aortic and mitral positions),
and the other patients presented rhythm disturbances (ventricular extra systoles, AV block and
9
tachyarrhythmia) in the context of ischemic cardiomyopathy. The recordings varied in length from 6 s to
49 s (mean±SD: 16±9 s).
2.6. DLUT heart sounds database
The Dalian University of Technology heart sounds database (DLUTHSDB) was contributed by Dr. Hong
Tang (Li et al., 2011; Tang et al., 2010a; Tang et al., 2010b; Tang et al., 2012). Subjects included 174
healthy volunteers (2 female and 172 male, aged from 4 to 35 years, mean±SD: 25±3 years) and 335
CAD patients (227 female and 108 male, aged from 10 to 88 years, mean±SD: 60±12 years). Heart
sounds from the CAD patients were recorded in the Second Hospital of Dalian Medical University using
an electronic stethoscope (3M Littmann). CAD patients were confirmed based on the cardiologist’s
diagnosis. Only PCG signals were available and all of them were collected from the mitral position at the
chest. Data were saved in the .wav format using a sampling rate of 8,000 Hz with 16-bit quantization.
Each patient provided one PCG recording and there were a total of 335 recordings. The recording length
varied from about 3 s to 98 s (mean±SD: 17±12 s). Heart sound signals from the healthy volunteers were
recorded using a microphone sensor (MLT201, ADinstrument, Australia) or a piezoelectric sensor
(Xinhangxingye Technology Co. Lt., China) at the Biomedical Engineering Lab in DLUT, China. Each
subject contributed one or several recordings and a total of 338 recordings were collected. Recordings
included either a single channel (PCG) or several channels (PCG combined with ECG,
photoplethysmogram or respiratory signals). ECG signals were the standard lead-II ECG.
Photoplethysmogram signals were recorded from the carotid artery or finger. Respiratory signals were
collected using a MLT1132 belt transducer (ADinstrument, Australia) to record chest movement. The
recording lengths varied from about 27.5 s to 312.5 s (mean±SD: 209±78 s). Various sampling rates were
used (800 Hz, 1,000 Hz, 2,000 Hz, 3,000Hz, 4,000 Hz, 8,000 Hz or 22,050 Hz) depending on different
research aims. All 338 recordings from the healthy volunteers could be separated into two sub-types:
recordings during rest (218 recordings) where the subjects were in peaceful calm states, and recordings
during non-resting states (120 recordings). Non-resting recordings were collected immediately after step
climbing (116 recordings), during cycles of breath holding (3 recordings), and after the bike cycling (1
recording).
2.7. SUA heart sounds database
The Shiraz University adult heart sounds database (SUAHSDB) was contributed by Dr. Reza Sameni and
colleagues (Samieinasab and Sameni, 2015). This database was constructed using recordings made from
79 healthy subjects and 33 patients (total 69 female and 43 male, aged from 16 to 88 years, mean±SD:
56±16 years). The JABES digital electronic stethoscope (GS Technology Co. Ltd., South Korea) was
used, placed on the chest, commonly above the apex region of the heart. The Audacity cross-platform
audio software was used for recording and editing the signals on a PC. The subjects were asked to relax
and breathe normally during the recording session. The database consists of 114 recordings (each
subject/patient had one heart sound signal but one healthy subject had three), resulting in 81 normal
recordings and 33 pathological recordings. The recording length varied from approximately 30 s to 60 s
(mean±SD: 33±5 s). The sampling rate was 8,000 Hz with 16-bit quantization except for three recordings
10
at 44,100 Hz and one at 384,000 Hz. The data were recorded in wideband mode of the digital stethoscope,
with a frequency response of 20 Hz to 1 kHz.
2.8. SSH heart sounds database
The Skejby Sygehus Hospital heart sounds database (SSHHSDB) was assembled from patients referred to
Skejby Sygehus Hospital, Denmark. It comprises 35 recordings from 12 normal subjects and 23
pathological patients with heart valve defect. All recordings are obtained from the 2nd intercostal room
just right to sternum. The recording length varied from approximately 15 s to 69 s (mean±SD: 36±12 s)
and the sampling rate was 8,000 Hz.
2.9. SUF heart sounds database (not used for challenge)
The Shiraz University fetal heart sounds database (SUFHSDB) was also contributed by Dr. Reza Sameni
and colleagues (Samieinasab and Sameni, 2015). This database was constructed using recordings made
from 109 pregnant women (mothers aged from 16 to 47 years, mean±SD: 29±6 years with BMI from 19.5
to 38.9, mean±SD: 29.2±4.0). The JABES digital electronic stethoscope (GS Technology Co. Ltd., South
Korea) was used, and placed on the lower maternal abdomen as described in (Samieinasab and Sameni,
2015). In the case of twins (seven cases) the data were collected twice according to the locations advised
by the expert gynecologist. The Audacity cross-platform audio software was used for recording and
editing the signals on a PC. In total, 99 subjects had one signal recorded, three subjects had two and seven
cases of twins were recorded individually, resulting in 119 total recordings. The average duration of each
record was about 90 seconds. The sampling rate was generally 8,000 Hz with 16-bit quantization and a
few recordings were sampled at 44,100 Hz. The data were recorded in wideband mode of the digital
stethoscope, with a frequency response of 20 Hz to 1 kHz. In most cases (91 subjects), the heart sounds of
the mothers were also recorded before each fetal PCG recording session. As a result, a total number of 92
maternal heart sounds data (90 subjects had one heart sound signal but one had two signals recorded) are
also available in the dataset.
Note that since the PhysioNet/CinC Challenge 2016 was focused on adult heart sounds, this SUFHSDB
dataset was excluded only from the challenge; but has been included in the online database. The inclusion
of this dataset in the open-access database was provided to enable researchers to test single channel fetal,
maternal, and environmental noise separation algorithms, although it is not part of the Challenge
described in this article.
11
Table 1. Detailed profiles for the assembled heart sound databases for the 2016 PhysioNet/CinC Challenge.
Database
Subject type
#
subject
Age
Gender
(F/M)
Recording
position
Recording
state
#
recording
Recording
length (s)
Simultaneous
signal
Sample
rate
Sensor
Sensor
frequency
bandwidth
MITHSDB
Normal
38
unknown
unknown
Nine different
recording
positions
Recorded in-home
visits or in
hospital,
uncontrolled
recording
environment
117
33±5
One PCG
One ECG
44,100
Hz
Welch Allyn
Meditron
electronic
stethoscope
20 Hz –
20 kHz
MVP
37
134
Benign
34
118
AD
5
17
MPC
7
23
AADHSDB
Normal
121
unknown
58/93
Tricuspid area
Rest
544
8
One PCG
4,000
Hz
3M Littmann
E4000
20 Hz –
1,000 Hz
CAD
30
151
AUTHHSDB
Normal
11
29±8
5/6
Apex
Rest
11
47±25
Two PCGs
4,000
Hz
Welch Allyn
Meditron
electronic
stethoscope
unknown
MR
17
75±7
12/5
Auscultation
positions
17
60±30
AS
17
76±10
11/6
17
43±21
TUTHSDB
Normal
28
unknown
unknown
Four typical
auscultation
positions
Rest
174
15
One PCG
4,000
Hz
unknown
unknown
Pathologic
16
UHAHSDB
Normal:
NHC
19
18-40
unknown
unknown
Rest
19
14±5
One PCG
8,000
Hz
Prototype
(Infral
Corporation)
unknown
Normal:
MARS500
6
unknown
unknown
20
10±3
Pathologic
30
44-90
10/20
40
16±9
DLUTHSDB
Normal
174
25±3
2/172
Multi-position
at chest
Rest or after
exercise
338
209±78
PCG, ECG,
PPG and
RESP
800 Hz
22,050
Hz
MLT201 or
piezoelectric
sensor
unknown
CAD
335
60±12
227/108
Mitral
Rest
335
17±12
One PCG
8,000
Hz
3M Littmann
1-1,000 Hz
SUAHSDB
Normal
79
56±16
69/43
Apex
Rest
81
33±5
One PCG
8,000
Hz
JABES
electronic
stethoscope
20-1000 Hz
Pathologic
33
33
SSHHSDB
Normal
12
unknown
unknown
2th
intercostal
Rest
12
36±12
One PCG
8,000
Hz
unknown
unknown
Pathological
23
23
SUFHSDB
Fetal
116
--
--
Maternal
abdomen
Rest
119
90
One PCG
8,000
Hz,
44,100
JABES
electronic
stethoscope
20-1000 Hz
Maternal
109
29±6
109/0
unknown
Rest
92
90
One PCG
12
Hz
Total
--
1,297
--
--
--
--
2,435
--
--
--
--
--
Note: MIT: Massachusetts Institute of Technology, AAD: Aalborg University, AUTH: Aristotle University of Thessaloniki, TUT: K.N. Toosi University of Technology, UHA:
University of Haute Alsace, DLUT: Dalian University of Technology, SU: Shiraz University, SSH: Skejby Sygehus Hospital, MVP: mitral valve prolapse, Benign: innocent
or benign murmurs, AD: aortic disease, MPC: miscellaneous pathological conditions, CAD: coronary artery disease, MR: mitral regurgitation, AS: aortic stenosis, PCG:
phonocardiogram, ECG: electrocardiogram, PPG: photoplethysmogram, RESP: respiratory.
13
3. Brief review on heart sound segmentation methods
The segmentation of the FHSs is a first step in the automatic analysis of heart sounds. The accurate
localization of the FHSs is a prerequisite for the identification of the systolic or diastolic regions, allowing
the subsequent classification of pathological situations in these regions (Liang et al., 1997b; Springer,
2015; Springer et al., 2014). S1 is initiated by the closure of the atrioventricular valves at the beginning of
the systole and occurs immediately after the R-peak (ventricular depolarization) of the ECG. S2 is
initiated by the closure of the semilunar valves at the beginning of the diastole and occurs approximately
at the end-T-wave of the ECG (the end of ventricular depolarization). The time order of these features in
ECG and PCG is shown in Figure 4 (Springer, 2015). In clinical practice, the criteria adopted by the
cardiologist to annotate the beginning and the ending of S1 and S2 sounds was defined as follows: the
beginning of S1 is the start of the high frequency vibration due to mitral closure, the beginning of S2 is
the start of the high frequency vibration due to aortic closure, and the endings of S1 and S2 are annotated
by the end of the high frequency vibrations (Moukadem et al., 2013).
Figure 4. Example of an ECG-labelled PCG, with the ECG, PCG and four states of the heart cycle (S1,
systole, S2 and diastole) shown. The R-peak and end-T-wave are labelled as references for defining the
approximate positions of S1 and S2 respectively. Mid-systolic clicks, typical of mitral valve prolapse, can
be seen in the systole states. Adapted from (Springer, 2015).
14
Many methods of heart sound segmentation have been studied over the past few decades. The typical
methods can be classified into four types: the first type is envelope-based method, i.e., using a variety of
techniques to construct the envelopes of heart sound and thus to perform the heart sound segmentation;
the second one is feature-based method, i.e., by calculating the features of heart sounds to segment the
signal; the third one is the machine learning method and the last one, also as the state-of-the-art method, is
hidden Markov model (HMM) method. We will give a brief summary for the aforementioned four types
of heart sound segmentation methods. The size of the database of subjects and recordings used in each
study, as well as the numerical results, will be also presented (see Table 2).
3.1. Envelope-based methods
Shannon energy envelope is the most used envelope for PCG envelope extraction. Liang et al proposed a
normalized average Shannon energy envelope (Liang et al., 1997a), which emphasized the medium-
intensity sounds while attenuating the low-intensity components. The performance of this method was
evaluated using 515 PCG cycles from 37 recordings acquired from children with murmurs and achieved
93% accuracy for PCG segmentation. Another study from Liang et al employed wavelet decomposition
before estimation of the Shannon envelope and segmented heart sound into four parts: S1, systole, S2 and
diastole (Liang et al., 1997b). This method was evaluated using 1,165 cardiac cycles and resulted in an
improved accuracy from 84% (without wavelet decomposition) to 93% (with wavelet decomposition) on
a set of 77 noisy recordings including both normal and abnormal heart sounds. Moukadem et al proposed
a method to calculate the Shannon energy envelope of the local spectrum calculated by the S-transform
for each sample of heart sound signal. This method was evaluated on 40 normal and 40 pathological heart
sound recordings. The sensitivity and positive predictivity were both higher than 95% for normal and
pathological heart sound segmentation (Moukadem et al., 2013).
Envelope extraction based on Hilbert transform can be divided into two aspects: 1) the envelope is the
decimated signal of the real part of a complex analytic signal, and 2) the instantaneous frequency is the
derivative of the imaginary part of complex analytic signal. Sun et al proposed an automatic segmentation
method based on Hilbert transform (Sun et al., 2014). This method considered the characteristics of
envelopes near the peaks of S1, the peaks of S2, the transmission points T12 from S1 to S2, and the
transmission points T21 from S2 to S1. It was validated using 7,730 s of heart sounds from pathological
patients, 600 s from normal subjects, and 1496.8 s from Michigan MHSDB database. For the sounds
where S1 cannot be separated from S2, an average accuracy of 96.69% was achieved. For the sounds
where S1 can be separated from S2, an average accuracy of 97.37% was achieved.
Jiang and Choi proposed an envelope extraction method named cardiac sound characteristic waveform
(CSCW) (Jiang and Choi, 2006). However, they only reported the example figures without reporting any
quantitative results. In their following study, they compared this CSCW method with other two popular
envelope-based methods: Shannon energy and Hilbert transform envelopes, and found the CSCW method
to be superior to both of these, concluding that their method led to more accurate segmentation results:
100% and 88.2% on normal and pathological patients respectively, as compared to 78.2% and 89.4% for
the Shannon energy envelope and 51.4% and 47.3% for the Hilbert transform envelope (Choi and Jiang,
2008). However, these results were only evaluated on 500 selected cardiac cycles without a split between
15
their training and test sets. Yan et al also used a similar characteristic moment waveform envelope method
for segmenting heart sound (Yan et al., 2010). This method was only evaluated on a small dataset of 9
recordings and reported an accuracy of 99.0%, again without a train-test split.
A simple squared-energy envelope was proposed by Ari et al (Ari et al., 2008). It is primarily based on
the use of frequency content present in the signal, calculation of energy in time windows and timing
relations of signal components. It was shown to have a better performance than Shannon energy envelope
when employing a threshold-based detection method. Testing on a total of 357 cycles from 71 recordings
showed the segmentation accuracy is 97.47% (without a train-test split).
3.2. Feature-based methods
Naseri and Homaeinezhad used frequency- and amplitude-based features, and then employed a synthetic
decision making algorithm for heart sound segmentation (Naseri and Homaeinezhad, 2013). The proposed
method was applied to 52 PCG signals gathered from patients with different valve diseases and achieved
an average sensitivity of 99.00% and positive predictivity of 98.60%. Kumar et al proposed a detection
method based on a high frequency feature, which is extracted from the heart sound using the fast wavelet
decomposition (Kumar et al., 2006). This feature is physiologically motivated by the accentuated pressure
differences found across heart valves, both in native and prosthetic valves. The method was validated on
patients with mechanical and bioprosthetic heart valve implants in different locations, as well as with
patients with native valves, and achieved an averaged sensitivity of 97.95% and positive predictivity of
98.20%.
Varghees and Ramachandran used an instantaneous phase feature from the analytical signal after
calculating the Shannon entropy (Varghees and Ramachandran, 2014). This method is a quite
straightforward approach that does not use any search-back steps. It was tested using both clean and noisy
PCG signals with both normal and pathological heart sounds (701 cycles), and achieved an average
sensitivity of 99.43% and positive predictivity of 93.56% without a train-test split. Pedrosa et al used
periodic component features from the analysis signal of the autocorrelation function to segment heart
sound signal (Pedrosa et al., 2014). Their method was tested on 72 recordings and had sensitivity and
positive predictivity of 89.2% and 98.6% respectively.
Unlike using the absolute amplitude or frequency characteristics of heart sounds, Nigam and Priemer used
complexity-based features by utilizing the underlying complexity of the dynamical heart sound for PCG
segmentation and this method showed good performance on the synthetic data (Nigam and Priemer,
2005). However, this study did not provide any quantitative results for evaluation. Vepa et al also used
complexity-based features for heart sound segmentation, which combined energy-based and simplicity-
based features computed from multi-level wavelet decomposition coefficients (Vepa et al., 2008). The
method was evaluated on only 166 cycles and achieved an accuracy of 84.0%.
Papadaniil and Hadjileontiadis employed kurtosis-based features alongside ensemble empirical mode
decomposition to select non-Gaussian intrinsic mode functions (IMFs), and then detected the start and
end positions of heart sounds within the selected IMFs (Papadaniil and Hadjileontiadis, 2014). The
method was tested on 11 normal subjects and 32 pathological patients, and achieved an accuracy of
16
83.05%. In addition, an ECG-referred pediatric heart sound segmentation method was proposed in
(Gharehbaghi et al., 2011). This algorithm was applied on 120 recordings of normal and pathological
children, totally containing 1,976 cardiac cycles, and achieved accuracy of 97% for S1 and 94% for S2.
3.3. Machine learning methods
Neural network technology is widely used as a typical machine learning method for heart sound
segmentation. Oskiper and Watrous proposed a time-delay neural network method for detecting the S1
sound (Oskiper and Watrous, 2002). The method consists of a single hidden layer network, with time-
delay links connecting the hidden units to the time-frequency energy coefficients of Morlet wavelet
decomposition. The results tested on 30 healthy subjects (without a train-test split) showed an accuracy of
96.2%. Sepehri et al used a multi-layer perceptron neural network classifier for heart sound segmentation,
which paid special attention to the physiological effects of respiration on pediatric heart sounds (Sepehri
et al., 2010). A total of 823 cycles from 40 recordings of normal children and 80 recordings of children
with congenital heart diseases were tested and an accuracy of 93.6% was achieved when splitting the
recordings equally between training and test datasets.
K-means clustering is another widely used method. Chen et al used a K-means clustering and a threshold
method to identify the heart sounds, achieving 92.1% sensitivity and 88.4% positive predictivity tested on
27 recordings from healthy subjects (Chen et al., 2009). Gupta et al also used K-means clustering
combined with homomorphic filtering for segmenting heart sounds into single cardiac cycle (S1-systole-
S2-diastole) (Gupta et al., 2007). This method was tested on 340 cycles and achieved an accuracy of
90.29%. Tang et al employed dynamic clustering for segmenting heart sounds (Tang et al., 2012). In this
method, the heart sound signal was first separated into cardiac cycles based on the instantaneous cycle
frequency and then was decomposed into time–frequency atoms, and finally the atoms of heart sounds
were clustered in time–frequency plane allowing the classification of S1 and S2. The results tested on 25
subjects showed an accuracy of 94.9% for S1 and 95.9% for S2.
Rajan et al developed an unsupervised segmentation method by first using Morlet wavelet decomposition
to obtain a time-scale representation of the heart sounds and then using an energy profile of the time-scale
representation and a singular value decomposition technique to identify heart sound segments (Rajan et
al., 2006). This method was tested on a dataset of 42 adult patients and achieved an accuracy of 90.5%.
3.4. Hidden Markov Model (HMM) methods
Gamero and Watrous proposed an HMM-based methodology, which employed a probabilistic finite state-
machine to model systolic and diastolic interval duration (Gamero and Watrous, 2003). The detection of
S1 and S2 was performed using a network of two HMM with grammar constraints to parse the sequence
of systolic and diastolic intervals. Results were evaluated on 80 subjects and a sensitivity of 95% and a
positive predictivity of 97% were achieved (without a train-test split). Ricke et al also used an HMM
method for segmenting heart sounds into four components (S1-systole-S2-diastole), and achieved an
accuracy of 98% when using eight-fold cross-validation (Ricke et al., 2005). However, this study was
only performed on a relative small subject size of 9.
17
Gill et al were the first researchers to incorporate timing durations within the HMM method for heart
sound segmentation (Gill et al., 2005). In their method, homomorphic filtering was first performed and
then sequences of features were extracted to be used as observations within the HMM. Evaluation on 44
PCG recordings taken from 17 subjects showed that for S1 detection, sensitivity and positive predictivity
were 98.6% and 96.9% respectively, and for S2 detection, they were 98.3% and 96.5% respectively.
Sedighian et al (Sedighian et al., 2014) also used homomorphic filtering and an HMM method on the
PASCAL database (Bentley et al., 2011) and obtained an average accuracy of 92.4% for S1 segmentation
and 93.5% for S2 segmentation. By comparison, Castro et al (Castro et al., 2013) used the wavelet
analysis on the same database and achieved an average accuracy of 90.9% for S1 segmentation and 93.3%
for S2 segmentation.
Schmidt et al were the first researchers to explicitly model the expected duration of heart sounds within
the HMM using a hidden semi-Markov model (HSMM) (Schmidt et al., 2010a). They first hand-labelled
the positions of the S1 and S2 sounds in 113 recordings, and then used the average duration of these
sounds and autocorrelation analysis of systolic and diastolic durations to derive Gaussian distributions for
the expected duration of each of the four states, i.e., S1, systole, S2 and diastole. The employed features
were the homomorphic envelope and three frequency band features (25-50, 50-100 and 100-150 Hz).
These features, along with the hand-labelled positions of the states, were used to derive Gaussian
distribution-based emission probabilities for the HMM. The duration distributions were then incorporated
into the forward and backward paths of the Viterbi algorithm. The results on the separate test set were
98.8% sensitivity and 98.6% positive predictivity.
Based on Schmidt et als work (Schmidt et al., 2010a), Springer et al used the HSMM method and
extended it with the use of logistic regression for emission probability estimation, to address the problem
of accurate segmentation of noisy, real-world heart sound recordings (Springer et al., 2015). Meanwhile,
a modified Viterbi algorithm for decoding the most-likely sequence of states was also implemented. It
was evaluated on a large dataset of 10,172 s of heart sounds recorded from 112 patients and achieved an
average F1 score of 95.63% on a separate test dataset, significantly improving upon the highest score of
86.28% achieved by the other reported methods in the literature when evaluated on the same test data.
Therefore, this method is regarded as the state-of-the-art method in heart sound segmentation studies.
18
Table 2. Summary of the major heart sound segmentation works. Se: sensitivity, P+: positive predictivity and Acc: accuracy.
Author
Subject type
#
subject
# recording
Recording
length
Cycle
number
Sample rate
(Hz)
Segmentation results
Se (%)
P+ (%)
Acc (%)
Envelope-based method
(Liang et al., 1997a)
Normal and pathological
children
--
37
Each 7-12 s
515
11,025
--
--
93
(Liang et al., 1997b)
Normal and pathological
children
--
77
Each 6-13 s
1,165
11,025
--
--
93
(Moukadem et al., 2013)
Normal
--
80
Each 6-12 s
--
8,000
96
95
--
Pathological
97
95
--
(Sun et al., 2014)
Normal
45
--
Total 600 s
--
44,100
--
--
96.69
Pathological
76
Total 7,730 s
MHSDB
--
23
Total 1,497 s
(Choi and Jiang, 2008)
Normal
--
--
--
500
--
--
--
100
Pathological
--
--
88.2
(Yan et al., 2010)
Normal and pathological
--
9
Each < 5 s
--
--
--
--
99.0
(Ari et al., 2008)
Normal and pathological
71
71
--
357
Varied
--
--
97.47
Feature-based method
(Naseri and Homaeinezhad,
2013)
Pathological
--
--
Total 42 min
--
4,000
99.00
98.60
--
(Kumar et al., 2006)
Pathological
55
55
Each < 120 s
7,530
44,100
97.95
98.20
--
(Varghees and Ramachandran,
2014)
Normal and pathological
--
64
Each < 10 s
701
Varied
99.43
93.56
--
(Pedrosa et al., 2014)
Pathological adults and PASCAL
database
72
72
Each 60 s
--
--
89.2
98.6
--
(Vepa et al., 2008)
Normal and pathological
--
--
--
166
--
--
--
84.0
(Papadaniil and
Hadjileontiadis, 2014)
Normal and pathological
43
43
--
2,602
44,100
--
--
83.05
(Gharehbaghi et al., 2011)
Normal and pathological
children
120
120
Each 10 s
1,976
44,100
--
--
S1: 97
S2: 94
Machine learning method
(Oskiper and Watrous, 2002)
Normal
30
--
Each 20 s
--
--
--
--
S1 96.2
(Sepehri et al., 2010)
Normal and pathological
60
120
Total 1,200 s
--
--
--
--
93.6
19
children
(Chen et al., 2009)
Normal
--
27
Each 30 s
997
--
92.1
88.4
--
(Gupta et al., 2007)
Normal and pathological
--
41
--
340
8,000
--
--
90.29
(Tang et al., 2012)
Normal
3
3
--
565
2,000
--
--
S1 94.9
S2 95.9
Pathological
23
23
(Rajan et al., 2006)
Normal and pathological
42
42
Each 13 s
--
--
--
--
90.5
Hidden Markov Model (HMM) methods
(Gamero and Watrous, 2003)
Normal
80
80
Each 20 s
--
11,000
95
97
--
(Ricke et al., 2005)
--
9
9
--
--
997
--
--
98
(Gill et al., 2005)
Normal
17
44
Each 30-60 s
--
4,000
S1:98.6
S2:98.3
S1:96.9
S2:96.5
--
(Sedighian et al., 2014)
PASCAL database
--
84
Total 416 s
S1: 639
S2: 626
4,000
--
--
S1: 92.4
S2: 93.5
(Castro et al., 2013)
PASCAL database
--
84
Total 416 s
S1: 639
S2: 630
4,000
--
--
S1: 90.9
S2: 933
(Schmidt et al., 2010a)
Normal and pathological
--
113
Each 8 s
--
4,000
98.8
98.6
--
(Springer et al., 2015)
Normal and pathological
112
--
Total 10,172 s
S1: 12,181
S2: 11,627
Varied
--
--
F1 score
95.63
20
4. Brief review on heart sound classification methods
The automated classification of pathology in heart sounds has been described in the literature for over 50
years, but accurate diagnosis remains a significant challenge. Gerbarg et al (Gerbarg et al., 1963) were the
first to publish on the automatic classification of pathology in heart sounds, (specifically to aid the
identification of children with rheumatic heart disease) and used a threshold-based method. The typical
methods for heart sound classification can be grouped into four categories: 1) artificial neural network-
based classification; 2) support vector machine-based classification; 3) hidden Markov model-based
classification and 4) clustering-based classification. The current prominent works in this field are
summarized in Table 3. The important notes about the evaluation of the method, such as whether the data
was split into training and test sets, are also reported. For relative brevity, only the notable studies with
sizeable datasets are summarized in detail below.
4.1. Artificial neural network-based classification
The artificial neural network (ANN) is the most widely used machine learning-based approach for heart
sound classification. Unless auto-associative in nature, ANN classifiers require discriminative signal
features as inputs. Relatively little work has been performed on optimizing network architectures in this
context. Typical signal features include: wavelet features, time, frequency and complexity-based features
and time-frequency features.
Wavelet-based features are most widely employed in ANN approaches to classification of heart sounds.
Akay et al combined wavelet features with an ANN for the automatic detection of CAD patients (Akay et
al., 1994). They computed four features (mean, variance, skewness and kurtosis) of the extracted
coefficients of wavelet transform from the diastolic period of heart cycles. These features, alongside
physical characteristics (sex, age, weight, blood pressure), were fed into a fuzzy neural network, and a
sensitivity of 85% and a specificity of 89% on a separate test set of 82 recordings were reported. Liang et
al (Liang and Hartimo, 1998) employed wavelet packet decomposition with the aim of differentiating
between pathological and innocent murmurs in children when using ANN classification. Eight nodes of
the wavelet packet tree were selected automatically using on an information-based cost function. The cost
function values then served as the feature vector. With a 65/20 patient train/test split they achieved 80%
sensitivity and 90% specificity on the test data. Uguz (Uguz, 2012a) employed an ANN with the features
from a discrete wavelet transform and a fuzzy logic approach to perform three-class classification:
normal, pulmonary stenosis, and mitral stenosis. With a 50/50 train/test split of a dataset of 120 subjects,
they reported 100% sensitivity, 95.24% specificity, and 98.33% average accuracy for the three-classes.
Bhatikar et al (Bhatikar et al., 2005) used the fast Fourier transform (FFT) to extract the energy spectrum
features in frequency domain, and then used these as inputs to an ANN. Using a separate test set of 53
patients they reported 83% sensitivity and 90% specificity when differentiating between innocent and
pathological murmurs. Sepehri et al (Sepehri et al., 2008) identified the five frequency bands that led to
the greatest difference in spectral energy between normal and pathological recordings and used the
spectral energy in these bands as the input features for the ANN. Reported results on 50 test records were
95% sensitivity and 93.33% specificity for a binary classification. Ahlstrom et al (Ahlstrom et al., 2006)
21
assessed a range of non-linear complexity-based features that had not previously been used for murmur
classification. They included up to 207 features and finally selected 14 features to present to an ANN.
They reported 86% classification accuracy for a three-class problem: normal, aortic stenosis and mitral
regurgitation.
De Vos et al (De Vos and Blanckenberg, 2007) used time-frequency features and extracted the energy in
12 frequency bins at 10 equally-spaced time intervals over each heart cycle to presents to an ANN. They
reported a sensitivity and specificity of 90% and 96.5% respectively on 163 test patients (aged between 2
months and 16 years). Uguz (Uguz, 2012b) also used time-frequency as an input to an ANN. A total of
120 heart sound recordings, split 50/50 into train/test, and reported 90.48% sensitivity, 97.44% specificity
and 95% accuracy for a three-class classification problem (normal, pulmonary and mitral stenosis heart
valve diseases).
4.2. Support vector machine-based classification
A number of researchers have applied a support vector machine (SVM) approach to the heart sound
classification in recent years. Since SVMs are another form of supervised machine learning, the features
chosen are rather similar to those based on ANN approaches.
Wavelet-based features are therefore widely employed in SVM-based methods. Ari et al (Ari et al., 2010)
used a least square SVM (LSSVM) method for classification of normal and abnormal heart sounds based
on the wavelet features. The performance of the proposed method was evaluated on 64 recordings
comprising of normal and pathological cases. The LSSVM was trained and tested on a 50/50 split (32
patients in each set) and the authors reported an 86.72% accuracy on their test dataset. Zheng et al (Zheng
et al., 2015) decomposed heart sounds using wavelet packets and then extracted the energy fraction and
sample entropy as features for the SVM input. Tested on 40 normal and 67 pathological patients, they
reported a 97.17% accuracy, 93.48% sensitivity and 98.55% specificity. Patidar et al (Patidar et al., 2015)
investigated the use of the tunable-Q wavelet transform as an input to LSSVM with varying kernel
functions. Testing on a dataset of 4,628 cycles from 163 heart sound recordings (and an unknown number
of patients) they reported a 98.8% sensitivity and 99.3% specificity, but without stratifying patients
(having mutually exclusive patients in testing and training sets), and therefore overfitting to their data.
Maglogiannis et al (Maglogiannis et al., 2009) used Shannon energy and frequency features from four
frequency bands (50-250, 100-300, 150-350, 200-400 Hz) to develop an automated diagnosis system for
the identification of heart valve diseases based on an SVM classifier. Testing on 38 normal and 160 heart
valve disease patients they reported an 87.5% sensitivity, 94.74% specificity and 91.43% accuracy.
Gharehbaghi et al (Gharehbaghi et al., 2015) used frequency band power over varying length frames
during systole as input features and used a growing-time SVM (GTSVM) to classify pathological and
innocent murmurs. When using a 50/50 train/test split (from a total of 30 patients with aortic stenosis, 26
with innocent murmurs and 30 normals), they reported 86.4% sensitivity and 89.3% specificity.
4.3. Hidden Markov model-based classification
HMM methods are not only widely employed for heart sound segmentation, but are also used for
pathology classification of heart sounds. In the case of classifying pathology, the posterior probability of
22
the heart sound signal or the extracted features given a trained HMM can be used to differentiate between
healthy and pathological recordings.
Wang et al (Wang et al., 2007) used a combination of HMM and mel-frequency cepstral coefficients
(MFCCs) to classify heart sound signals. The feature extraction was performed using three methods:
time-domain feature, short-time Fourier transforms (STFT) and MFCCs. Testing on 20 normal and 21
abnormal patients with murmurs they reported a sensitivity of 95.2% and a specificity of 95.3%. In a
subsequent study, they also used MFCCs to extract representative features and developed a HMM-based
method for heart sound classification (Chauhan et al., 2008). The method was applied to 1,381 cycles of
real and simulated, normal and abnormal heart sounds and they reported an accuracy of 99.21%.
However, both studies failed to make use of a separate test set when evaluating their classification
methods and the methods are likely to be highly over-trained. Saracoglu (Saracoglu, 2012) applied a
HMM in an unconventional manner, by fitting an HMM to the frequency spectrum extracted from entire
heart cycles. The exact classification procedure of using the HMMs is unclear, but it is thought that they
trained four HMMs, and then evaluated the posterior probability of the features given each model to
classify the recordings. They optimized the HMM parameters and PCA-based feature selection on a
training set and reported 95% sensitivity, 98.8% specificity and 97.5% accuracy on a test dataset of 60
recordings.
In summary, although HMM-based approaches are regarded as the state-of-the-art heart sound
segmentation method, their potential to classify heart sounds has not yet been adequately demonstrated.
4.4. Clustering-based classification
A number of researchers have made use of the unsupervised k-nearest neighbours (kNN) algorithm to
classify pathology in heart sounds. Bentley et al (Bentley et al., 1998) showed that discrete wavelet
transform features outperformed morphological features (time and frequency features from S1 and S2)
when performing heart sound classification using such a method. They used a binary kNN classifier and
reported 100% and 87% accuracy when detecting pathology in patients with heart valve disease and
prosthetic heart valves respectively on an unspecified sized database. Quiceno-Manrique et al (Quiceno-
Manrique et al., 2010) used a simple kNN classifier with features from various time-frequency
representations on a subset of 16 normal and 6 pathological patients. They reported 98% accuracy for
discriminating between normal and pathologic beats. However, the kNN classifier parameters were
optimized on the test set, indicating a likelihood of over-training. Avendano-Valencia et al (Avendano-
Valencia et al., 2010) also employed time-frequency features and kNN approach for classifying normal
and murmur patients. In order to extract the most relevant time-frequency features, two specific
approaches for dimensionality reduction were presented in their method: feature extraction by linear
decomposition, and tiling partition of the time-frequency plane. The experiments were carried out using
26 normal and 19 pathological recordings and they reported an average accuracy of 99.0% when using
11-fold cross-validation with grid-based dimensionality reduction.
23
Table 3. Summary of the previous heart sound classification works. Se: sensitivity, Sp: specificity and
Acc: accuracy.
Author
Database
Recording
length
Classification
method
Features
Se
(%)
Sp
(%)
Acc (%)
Notes on
database
(Akay et al.,
1994)
42 normal and 72
CAD patients
Each 10
cycles
ANN
Wavelet
85
89
86
30 training, 82
test
(Liang and
Hartimo, 1998)
40 normal and 45
pathological children
Each 7-12
s
ANN
Wavelet
80
90
85
65 training, 20
test
(Uguz, 2012a)
40 normal, 40
pulmonary and 40
mitral stenosis
--
ANN
Wavelet
100
95.24
98.33
50-50 train-test
split
(Bhatikar et al.,
2005)
88 innocent murmurs
and 153 pathological
murmurs
Each 10-
15 s
ANN
Frequency
83
90
--
188 training, 53
test
(Sepehri et al.,
2008)
36 normal and 54
pathological
Each 10 s
ANN
Frequency
95
93.33
--
40 training, 50
test
(Ahlstrom et
al., 2006)
7 normal, 23 aortic
stenosis and 6 mitral
regurgitation
Each 12
cycles
ANN
Complexity
--
--
86
Cross-validation
(De Vos and
Blanckenberg,
2007)
113 normal and 50
pathological
Each 6
cycles
ANN
Time-
frequency
90
96.5
--
Cross-validation
(Uguz, 2012b)
40 normal, 40
pulmonary and 40
mitral stenosis
--
ANN
Time-
frequency
90.4
8
97.44
95
50-50 train-test
split
(Ari et al.,
2010)
64 patients (normal
and pathological)
Each 8
cycles
SVM
Wavelet
--
--
86.72
50-50 train-test
split
(Zheng et al.,
2015)
40 normal and 67
pathological
--
SVM
Wavelet
93.4
8
98.55
97.17
Cross-validation
(Patidar et al.,
2015)
Total 4,628 heart
cycles, 626 normal
and 4,002
pathological
--
SVM
Wavelet
98.8
99.3
98.9
80% training,
20% test
(Maglogiannis
et al., 2009)
38 normal and 160
heart valve disease
patients
--
SVM
Frequency
87.5
94.74
91.43
Cross-validation
(Gharehbaghi et
al., 2015)
30 normal, 26
innocent and 30 aortic
stenosis
Each 10 s
SVM
Frequency
86.4
89.3
--
50-50 train-test
split
(Wang et al.,
2007)
20 normal and 21
murmurs patients
--
HMM
Signal
amplitude,
STFT and
MFCC
≥95.
2
≥95.3
--
No separate
training and test
(Chauhan et al.,
2008)
20 normal and 21
murmurs patients
--
HMM
Signal
amplitude,
STFT and
MFCC
--
--
99.21
No separate
training and test
(Saracoglu,
40 normal, 40
HMM
DFT and
95
98.8
97.5
50-50 train-test
24
2012)
pulmonary and 40
mitral stenosis
PCA
split
(Bentley et al.,
1998)
Unspecified size:
native and prosthetic
heart valves patients
--
kNN
Wavelet
--
--
100 for
native and
87 for
prosthetic
Cross-validation
(Quiceno-
Manrique et al.,
2010)
16 normal and 6
pathological
--
kNN
Time-
frequency
--
--
98
Cross-validation
(Avendano-
Valencia et al.,
2010)
26 normal and 19
pathological
--
kNN
Time-
frequency
99.5
6
98.45
99.0
Cross-validation
5. Description of the 2016 PhysioNet/CinC Challenge
5.1. Main aim
The 2016 PhysioNet/CinC Challenge aims to encourage the development of algorithms to classify heart
sound recordings collected from a variety of clinical or nonclinical environments (such as in-home visits).
The practical aim is to identify, from a single short recording (10-60s) from a single precordial location,
whether the subject of the recording should be referred on for an expert diagnosis (2016; Clifford et al.,
2016).
As pointed out in the above reviews, a number of studies have investigated the performances of different
methods for heart sound segmentation and classification. However, many of these investigations are
flawed because: 1) the studies were marred by poor methodology, often without the use of a separate test
set or by allowing data from the same patient to appear in both the training and test sets, almost certainly
resulting in over-fitting of the model and inflated statistics; 2) the studies did not clearly describe the
database used (type of patient, size, etc.) and did not report the method/location for heart sound recording;
3) the studies tended to use hand-picked clean data in their database, used manual labels, and excluded
noisy data, which leads to an algorithm that is of little use in the real world; 4) failure to use enough or a
variety of heart sound recordings; and 5) failure to post the data (and any code to process the data)
publicly so others may compare their results directly. The latter issue is often due to lack of time and
resources, and therefore this challenge is an attempt to address both this and the aforementioned issues.
In this Challenge, we focused only on the accurate classification of normal and abnormal heart sound
recordings, particularly in the context of real world (extremely noisy) recordings with low signal quality.
By providing the largest public collection of heart sound recordings from a variety of clinical and
nonclinical environments, the Challenge permits the challengers to develop accurate and robust
algorithms. In addition, due to the uncontrolled environment of the recordings, many recordings provided
in this Challenge are corrupted by various noise sources, such as speech, stethoscope motion, breathing
and intestinal activity. Some recordings were difficult or even impossible to classify as normal or
abnormal. Figure 5 shows an example of a section of a heart sound recording with good (upper plot) and
poor (lower plot) signal quality respectively. Therefore the challengers were given the choice to classify
25
some recordings as ‘unsure’ and the Challenge penalizes this in a different manner (see section 5.3:
Scoring Mechanism). Classifications for the heart sound recordings were therefore three-level: normal (do
not refer), abnormal (refer for further diagnostics) and unsure (too noisy to make a decision; retake the
recording). In this way, any algorithm developed could be employed in an expert-free environment and
used as decision support.
Figure 5. Example of a heart sound recording segment with good signal quality (A) and poor signal
quality (B).
5.2. Challenge data
Heart sound recordings (from nine independent databases) sourced from seven contributing research
groups described in section 2 (with the exception of the SUFHSDB since it was from fetal and maternal
heart sounds), were used in the Challenge, resulting in eight independent heart sound databases. Four of
the databases were divided into both training and test sets with a 70-30 training-test split. The other four
databases were exclusively assigned to either training or test set with the consideration of balancing the
data as much as possible between categories. The Challenge training set includes data from six databases
(with file names prefixed alphabetically, a through f, training sets a through e were provided before the
official phase and training set f was added after the beginning of the official phase) containing a total of
3,153 heart sound recordings from 764 subjects/patients, lasting from 5 s to just over 120 s. The
Challenge test set also included data from six databases (b through e, plus g and i) containing a total of
1,277 heart sound recordings from 308 subjects/patients, lasting from 6 s to 104 s. The total number of
26
recordings created for the Challenge was 4,430 and is different from the reported number of 2,435 in
Table 1. This is because the 338 recordings from normal subjects in the DLUTHSDB are generally longer
than 100 s and each recording was segmented into several relatively short recordings. All recordings were
resampled to 2,000 Hz using an anti-alias filter and provided as .wav format. Each recording contains
only one PCG lead, except for training set a, which also contains a simultaneously recorded ECG (2016).
In each of the databases, each recording begins with the same letter followed by a sequential, but random
number. Files from the same patient are unlikely to be numerically adjacent. The training and test sets
have each been divided so that they are two sets of mutually exclusive populations (i.e., no recordings
from the same subject/patient were in both training and test sets). Moreover, there are four collected
databases that have been semi-randomly placed exclusively in either the training or test sets (to ensure
there are ‘novel’ recording types and to reduce over-fitting on the recording methods). Databases a and f
are found exclusively in the training set and g and i are exclusively found in the test set. The test set is
unavailable to the public and will remain private for the purpose of scoring. (In the future, as more data
are added, we may release all the data to the public.) Participants may note the existence of a validation
dataset in the data folder. This data is a copy of 301 recordings from the training set, and is used to
validate uploaded entries before their evaluation on the test set.
In both training and test sets, heart sound recordings were divided into two types: normal and abnormal
recordings. The normal recordings were from healthy subjects and the abnormal ones were from patients
with a confirmed cardiac diagnosis. The patients were noted to suffer from a variety of illnesses (which is
not provided here on a case-by-case basis but is detailed in an online appendix to this article for the
training set data), but typically they are heart valve defects and CAD patients. Heart valve defects include
mitral valve prolapse, mitral regurgitation, aortic regurgitation, aortic stenosis and valvular surgery. All
the recordings from the patients were generally labeled as abnormal. We do not provide more specific
classification for these abnormal recordings. Please note that both training and test sets are unbalanced,
i.e., the number of normal recordings does not equal that of abnormal ones. Challengers will have to
consider this when they train and test their algorithms.
In addition, to facilitate the challengers in training their algorithms to identify low signal quality
recordings, we provided the labels for ‘unsure’ recordings with poor signal quality in all training data. We
also provided reference annotations for the four heart sound states (S1, systole, S2 and diastole) for each
beat for all recordings that were not belong to ‘unsure’ type. The reference annotations were obtained by
using Springer’s segmentation algorithm (Springer et al., 2015) and subsequently manually reviewing and
correcting each beat labels, resulting in a total of 84,425 beats in training set and 32,440 beats in test set
after hand correction. Figure 6 illustrates an example where the automatic segmentation algorithm outputs
the wrong annotation and the corresponding correct annotation from hand-correction. Table 4 summarizes
the number of patients and recordings, the recording percentages and time lengths, the percentages of
hand corrected recordings and heart beats, as well as the corresponding number of hand corrected
recordings/beats for each database, for both training and test sets. As shown in Table 4, 20.7% of the
recordings in the training set and 15.3% of the recordings in the test set required hand correction, with
corresponding percentages of hand corrected heart beats at 11.7% and 10.9% respectively.
27
Figure 6. (A) An example of the state labels of a heart sound segment with automatically generated
annotations (using Springer’s segmentation algorithm) and (B) the same data and annotations after hand-
correction.
28
Table 4. Summary of the training and test sets used in 2016 PhysioNet/CinC Challenge.
Challenge
set
Sub-set
Data source
#
patients
#
recordings
Proportion of recordings (%)
Recording length (s)
Hand
corrected
recordings
(%)
Hand
corrected
beats (%)
# beats (after hand corrected)
Abnormal
Normal
Unsure
Min
Median
Max
Min
Median
Max
Total
Training
training-a
MITHSDB
121
409
67.5
28.4
4.2
9.3
35.6
36.5
28.9
11.6
12
37
78
14,559
training-b
AADHSDB
106
490
14.9
60.2
24.9
5.3
8
8
32.9
25.9
4
9
15
3,353
training-c
AUTHHSDB
31
31
64.5
22.6
12.9
9.6
44.4
122.0
67.7
31.5
15
67
143
1,808
training-d
UHAHSDB
38
55
47. 3
47. 3
5. 5
6.6
12.3
48.5
56.4
19.5
6
14
72
853
training-e
DLUTHSDB
356
2,054
7.1
86.7
6.2
8.1
21.1
101.7
13.7
9.7
4
27
174
59,593
training-f
SUAHSDB
112
114
27.2
68.4
4.4
29.4
31.7
59.6
35.1
16.9
7
39
75
4,259
Total/
Average
764
3,153
18.1
73.0
8.8
5.3
20.8
122.0
20.7
11.7
4
26
174
84,425
Test
test-b
AADHSDB
45
205
15.6
48.8
35.6
6.3
8
8
35.6
33.7
6
9
16
1,269
test-c
AUTHHSDB
14
14
64.3
28.6
7.1
19.3
54.4
86.9
28.6
24.7
32
57
107
853
test-d
UHAHSDB
17
24
45.8
45.8
8.3
6.1
11.4
17.1
37.5
19.7
7
11
24
260
test-e
DLUTHSDB
153
883
6.7
86.4
6.9
8.1
21.8
103.6
11.4
8.8
3
28
169
26,724
test-g
TUTHSDB
44
116
18.1
81.9
0
15
15
15
0
0
9
18
29
2,048
test-i
SSHHSDB
35
35
60
34.3
5.7
15.0
31.7
68.8
22.9
26.4
18
36
59
1,286
Total/
Average
308
1,277
12.0
77.1
10.9
6.1
17.7
103.6
15.3
10.9
3
24
169
32,440
29
5.3. Scoring mechanism
The overall score is computed based on the number of recordings classified as normal, abnormal or
unsure, in each of the two reference categories. Table 5 shows the rules for determining the classification
result of current recording from Challenger’s algorithm (2016; Clifford et al., 2016).
Table 5: Rules for determining the classification result of current recording from Challenger’s algorithm.
Diagnosis
Signal
quality
Percentages of
recordings
Challenger report result
Abnormal
Unsure
Normal
Reference
label
Abnormal (1)
good (1)
wa1
Aa1
Aq1
An1
poor (0)
wa2
Aa2
Aq2
An2
Normal (-1)
good (1)
wn1
Na1
Nq1
Nn1
poor (0)
wn2
Na2
Nq2
Nn2
The modified sensitivity () and specificity () are defined as:
 

 (1)
 

 (2)
where wa1 and wa2 are the percentages of good signal quality and poor signal quality recordings in all
abnormal recordings respectively, and are used as weights for calculating  , wn1 and wn2 are the
percentages of good signal quality and poor signal quality recordings in all normal recordings
respectively, and are used as weights for calculating .
For all 3,153 training set recordings, the weight parameters of wa1, wa2, wn1 and wn2 are 0.8602, 0.1398,
0.8948 and 0.1052 respectively. For all 301 validation set recordings, the weight parameters of wa1, wa2,
wn1 and wn2 are 0.78881, 0.2119, 0.9467 and 0.0533 respectively.
The overall Challenge ‘Score is then given by    , i.e. the average of the values of the
 and .
6. A simple benchmark classifier for the 2016 PhysioNet/CinC Challenge
As a basic starting point for the Challenge we provided a benchmark classifier that relied on relatively
obvious parameters extracted from the heart sound segmentation code. For the pending competition
results in the 2016 PhysioNet/CinC Challenge, challengers can refer to (Clifford et al., 2016). Here we
briefly describe the approach for training and testing the code on the Challenge training data only.
6.1. Selected balanced database from training set
Since both training and test sets are unbalanced, first, a balanced heart sound database from training set
was selected. (Otherwise, without prior probabilities on the illness, a prevalence bias would be created.)
30
Table 6 summarizes the numbers of the raw heart sound recordings in training set, and the numbers of the
selected recordings for each training database.
Table 6. Numbers of raw and selected recordings for each database in the training set.
Database
name
# raw recordings
# recordings after balanced
Abnormal
Normal
Total
Abnormal
Normal
Total
training-a
292
117
409
117
117
234
training-b
104
386
490
104
104
208
training-c
24
7
31
7
7
14
training-d
28
27
55
27
27
54
training-e
183
1,958
2,141
183
183
366
training-f
34
80
114
34
34
68
Total
665
2,575
3,240
472
472
944
6.2. Definition for features
Springer’s segmentation code (Springer et al., 2015) was used to segment each selected heart sound
recording to generate the time durations for the four states: S1, systole, S2 and diastole. Twenty features
were extracted from the position information of the four states as follows:
1. m_RR: mean value of RR intervals
2. sd_RR: standard deviation (SD) of RR intervals
3. m_IntS1: mean value of S1 intervals
4. sd_IntS1: SD of S1 intervals
5. m_IntS2: mean value of S2 intervals
6. sd_IntS2: SD of S2 intervals
7. m_IntSys: mean of systolic intervals
8. sd_IntSys: SD of systolic intervals
9. m_IntDia: mean of diastolic intervals
10. sd_IntDia: SD of diastolic intervals
11. m_Ratio_SysRR: mean of the ratio of systolic interval to RR of each heart beat
12. sd_Ratio_SysRR: SD of the ratio of systolic interval to RR of each heart beat
13. m_Ratio_DiaRR: mean of ratio of diastolic interval to RR of each heart beat
14. sd_Ratio_DiaRR: SD of ratio of diastolic interval to RR of each heart beat
15. m_Ratio_SysDia: mean of the ratio of systolic to diastolic interval of each heart beat
16. sd_Ratio_SysDia: SD of the ratio of systolic to diastolic interval of each heart beat
17. m_Amp_SysS1: mean of the ratio of the mean absolute amplitude during systole to that during the S1
period in each heart beat
31
18. sd_Amp_SysS1: SD of the ratio of the mean absolute amplitude during systole to that during the S1
period in each heart beat
19. m_Amp_DiaS2: mean of the ratio of the mean absolute amplitude during diastole to that during the S2
period in each heart beat
20. sd_Amp_DiaS2: SD of the ratio of the mean absolute amplitude during diastole to that during the S2
period in each heart beat
6.3. Logistic regression for feature selection
Logistic regression (LR) allows the identification of the impact of multiple independent variables in
predicting the membership of one of the multiple dependent categories. Binary logistic regression (BLR)
is an extension of linear regression, to address the fact that the latter struggles with dichotomous
problems. This difficulty is overcome by applying a mathematical transformation of the output of the
classifier, transforming it into a bounded value between 0 and 1 more appropriate for binary predictions.
In the current study, the output variable Y is a positive (1, abnormal) or negative (-1, normal)
classification for heart sound recording.
All 20 features were tested and a forward likelihood ratio selection was used, in order of likelihood. If the
accuracy of the model exhibited a statistical difference with the model prior to the addition of a feature,
the newly added feature is included in the model. The forward selection is terminated if the newly added
feature did not significantly improve the normal/abnormal classification results. In this way, correlated
predictors are unlikely to be included in the model, but it does not guarantee an optimal combination of
features. Moreover, we note that the features we have chosen are by no means likely to include the most
useful features.
6.4. Feature results comparison between the selected balanced data from training set
Table 7 shows the average values of all 20 features for normal and abnormal heart sound recordings on
the selected balanced data from training set. The Kolmogorov-Smirnov test for verifying the normal
distribution of all features was applied using the SPSS Statistics 19 software package (SPSS Inc., USA).
The results showed that only the sd_Ratio_DiaRR feature exhibited Gaussian distributions in both normal
and abnormal groups. Therefore, the group t test was performed for the sd_Ratio_DiaRR feature and a
Wilcoxon rank sum test was performed for other 19 features to test the statistical differences between the
two groups. The results showed that 13 features exhibited statistical differences between the two groups
whereas 7 features did not exhibit statistically significant differences.
Table 7. Statistical results for comparison between normal and abnormal heart sound recordings on all
selected balanced data from training set. Statistically significant differences (p<0.01) are marked with *.
Feature
Abnormal
Normal
P-value
m_RR (ms)
875 ± 279
863 ± 232
0.1
sd_RR (ms)
42 ± 40
35 ± 30
<0.01 *
m_IntS1 (ms)
131 ± 9
129 ± 8
<0.01 *
sd_IntS1 (ms)
17 ± 6
14 ± 5
<0.01 *
32
m_IntS2 (ms)
104 ± 10
105 ± 10
0.4
sd_IntS2 (ms)
15 ± 6
12 ± 5
<0.01 *
m_IntSys (ms)
200 ± 112
197 ± 82
0.1
sd_IntSys (ms)
18 ± 9
13 ± 7
<0.01 *
m_IntDia (ms)
433 ± 216
428 ± 176
0.3
sd_IntDia (ms)
31 ± 29
28 ± 22
<0.01 *
m_Ratio_SysRR (%)
23 ± 3
23 ± 3
0.4
sd_Ratio_SysRR (%)
3.6 ± 1.7
3.2 ± 1.6
<0.01 *
m_Ratio_DiaRR (%)
44 ± 5
44 ± 5
0.2
sd_Ratio_DiaRR (%)a
6.1 ± 2.9
5.5 ± 2.6
<0.01 *
m_Ratio_SysDia (%)
54 ± 12
53 ± 12
0.1
sd_Ratio_SysDia (%)
15 ± 6
12 ± 6
<0.01 *
m_Amp_SysS1 (%)
42 ± 22
36 ± 22
<0.01 *
sd_Amp_SysS1 (%)
27 ± 16
19 ± 19
<0.01 *
m_Amp_DiaS2 (%)
60 ± 24
50 ± 27
<0.01 *
sd_Amp_DiaS2 (%)
31 ± 23
28 ± 19
<0.01 *
Note: data are presented as median ± standard deviation (SD). a: only feature with the normal distribution.
6.5. Classification results using Logistic Regression
Equation (3) shows the derived BLR prediction formula with the corresponding regression coefficients
for normal/abnormal heart sound recordings classification on all selected balanced data from training set.
Seven features were identified as the predictable features, including: sd_RR, sd_IntS1, m_IntS2,
sd_IntS2, sd_IntSys, m_IntDia and sd_Ratio_SysDia.
0.062-0.013 sd_RR+0.067 sd_IntS1-0.032 m_IntS2
0.041 sd_IntS2+0.058 sd_IntSys
+0.002 m_IntDia+0.035 sd_Ratio_SysDia
T
z w X 
 

(3)
Table 8 provides the results of Aa, An, Na and Nn numbers and the three evaluation metrics (Se, Sp and
Score) defined in section 5.3. Using equation (3), the normal/abnormal classification results were 0.62 for
Se, 0.70 for Sp and a Challenge Score of 0.66 on the training data.
Table 8. BLR results (equation (3)) of the Aa, An, Na and Nn numbers and the three indices (Se, Sp and
Score) for all selected balanced training database: 472 abnormal and 472 normal recordings.
Aa
An
Na
Nn
Se
Sp
Score
293
179
141
331
0.62
0.70
0.66
We also use both a K=10-fold cross validation, stratifying by patient, and a leave-one-out (database) cross
validation, stratifying by database to test the performances of BLR model on all selected balanced training
data. This is important to note, since including patients in the training data and reporting on test data that
includes the same data will give a falsely inflated accuracy. Similarly, using a leave-one-out approach to
each database, provides a deeper understanding of which databases can result in heavy biases, and may
help provide a more accurate estimate of the out of sample accuracy of the algorithm. Tables 9 and 10
show the corresponding results from 10-fold cross validation and leave-one-out cross validation. Note that
33
the results are subject to statistical variation because of the subsampling. We also note that the average
running time on the training set used 5.26% of quota and 5.22% of quota on the hidden test set using
Matlab 2016a. We note that this classification algorithm is not intended to provide a sensible way to
classify the recordings, but rather to illustrate how a simple algorithm can achieve basic results, but that
the results will also vary highly based on which databases are used to train and test the classifiers. We also
note that improving the segmentation algorithm may be key to improving the results of any given
classifier. Finally, we note that our classifier did not attempt to label any recordings as unknown or
unreadable. Any useful algorithm must endeavor to do so, since the intention is for this algorithm to be
used at the source of recording, where a re-recording can be triggered in the event that an automated
algorithm is likely to fail. Differentiating abnormality from noise is often a difficult but critical issue in
biomedical signal analysis, as we have noted in previous competitions (Clifford and Moody, 2012).
Table 9. K=10-fold cross validation results for all selected balanced training database: 472 abnormal and
472 normal recordings.
Fold
iterate
K-fold (10-fold) cross validation on the selected balanced training set
Aa
An
Na
Nn
Se
Sp
Score
1
30
17
13
34
0.64
0.72
0.68
2
25
22
18
29
0.53
0.62
0.57
3
30
17
16
32
0.64
0.67
0.65
4
31
17
14
33
0.65
0.70
0.67
5
31
16
11
36
0.66
0.77
0.71
6
30
17
16
31
0.64
0.66
0.65
7
21
26
18
29
0.45
0.62
0.53
8
29
18
16
31
0.62
0.66
0.64
9
30
18
10
38
0.63
0.79
0.71
10
30
17
14
33
0.64
0.70
0.67
Mean
29
19
15
33
0.61
0.69
0.65
SD
3
3
3
3
0.07
0.06
0.06
Note: SD, standard deviation.
Table 10. Balanced leave-one-out cross validation results for all training databases: 472 abnormal and
472 normal recordings.
Excluded
database
Leave-one-out cross validation on the balanced training set
Aa
An
Na
Nn
Se
Sp
Score
training-a
28
89
23
94
0.24
0.80
0.52
training-b
84
20
91
13
0.81
0.13
0.47
training-c
6
1
1
6
0.86
0.86
0.86
training-d
4
23
5
22
0.15
0.81
0.48
training-e
134
49
69
114
0.73
0.62
0.68
training-f
33
1
30
4
0.97
0.12
0.54
Mean
--
--
--
--
0.63
0.56
0.59
SD
--
--
--
--
0.34
0.34
0.15
Note: SD, standard deviation.
34
7. Potential benefits from the public heart sound data
The public release of the heart sound database has many potential benefits to a wide range of users. First,
those who lack access to well-characterized real clinical signals may benefit from access to these data for
developing prototype algorithms. The availability of these data can encourage researchers from a variety
of backgrounds to develop innovative methods to tackle problems in heart sound signal processing that
they might not otherwise have attempted.
An additional benefit is that the data can be re-evaluated with new advances in machine learning and
signal processing as they become available. The public data are also essential resources for developers
and evaluators who need to test their algorithms with realistic data and to perform these tests repeatedly
and reproducibly on a public platform.
In addition, these databases have value in medical and biomedical engineering education by providing
well-documented heart sound recordings from both healthy subjects and patients with a variety of
clinically significant diseases. By making well-characterized clinical data available to educational
institutions, these databases will make it possible to answer numerous physiological or pathological
questions without the need to develop a new set of reference data.
The availability of open source state of the art signal processing algorithms for heart sound segmentation
provided for the competition, and the subsequent open source classification algorithms provided by
competitors is likely to provide an impulse into the field and raise the benchmark for FDA approval and
diagnostic performance of industrial systems (Goldberger et al., 2000). We hope that this new heart sound
database will help realize these benefits and their often-unanticipated rewards to those with an interest in
heart sound signal processing.
Acknowledgments
We wish to thank the providers of the heart sound databases described in this paper and made available
for the competition:
The MITHSDB was provided by Prof. John Guttag and Dr. Zeeshan Syed from MIT.
The AADHSDB was provided by Dr. Samuel E. Schmidt from Aalborg University.
The AUTHHSDB was provided by Dr. Chrysa D. Papadaniil from Aristotle University of Thessaloniki.
The TUTHSDB was provided by Dr. Hosein Naseri from K. N. Toosi University of Technology.
The UHAHSDB was provided by Dr. Ali Moukadema from University of Haute Alsace.
The DLUTHSDB was provided by Dr. Hong Tang from Dalian University of Technology.
The SUAHSDB and SUFHSDB were provided by Dr. Reza Sameni from Shiraz University and annotated
by Dr. Mohammad Reza Samieinasab from Isfahan University of Medical Sciences. The two datasets
were recorded as part of the MS thesis of Ms. Maryam Samieinasab at Shiraz University. The authors
would like to thank Dr. M. Hosseiniasl and Ms. Nasihatkon from Shiraz Hafez Hospital, for their valuable
assistance during fetal PCG recordings.
35
The SSHHSDB was provided by the company of Medicom Innovation Partner and Mr. Bjørn Knud
Andersen at http://www.medicomip.com/home/.
This work was supported by the National Institutes of Health (NIH) grant R01- EB001659 from the
National Institute of Biomedical Imaging and Bioengineering (NIBIB) and R01GM104987 from the
National Institute of General Medical Sciences.
References:
2016 Classification of normal/abnormal heart sound recordings: the PhysioNet/Computing in Cardiology
Challenge 2016. http://physionet.org/challenge/2016/
Ahlstrom C, Hult P, Rask P, Karlsson J E, Nylander E, Dahlstrom U and Ask P 2006 Feature extraction
for systolic heart murmur classification Ann Biomed Eng 34 1666-77
Akay Y M, Akay M, Welkowitz W and Kostis J 1994 Noninvasive detection of coronary artery disease
IEEE Eng Med Biol 13 761-4
Ari S, Hembram K and Saha G 2010 Detection of cardiac abnormality from PCG signal using LMS based
least square SVM classier, Expert Syst Appl 37 8019-26
Ari S, Kumar P and Saha G 2008 A robust heart sound segmentation algorithm for commonly occurring
heart valve diseases J Med Eng Technol 32 456-65
Avendano-Valencia L D, Godino-Llorente J I, Blanco-Velasco M and Castellanos-Dominguez G 2010
Feature extraction from parametric time-frequency representations for heart murmur detection
Ann Biomed Eng 38 2716-32
Bentley P, Nordehn G, Coimbra M, Mannor S and Getz R 2011 The PASCAL classifying heart sounds
challenge 2011 (CHSC2011). http://www.peterjbentley.com/heartchallenge/index.html
Bentley P M, Nokia R D, Camberley U K, Grant P M and McDonnell J T E 1998 Time-frequency and
time-scale techniques for the classification of native and bioprosthetic heart valve sounds IEEE
Trans Biomed Eng 45 125-8
Bhatikar S R, DeGroff C and Mahajan R L 2005 A classifier based on the artificial neural network
approach for cardiologic auscultation in pediatrics Artif Intell Med 33 251-60
Castro A, Vinhoza T T V, Mattos S S and Coimbra M T 2013 Heart sound segmentation of pediatric
auscultations using wavelet analysis. In: Annual International Conference of the IEEE
Engineering in Medicine and Biology Society, (Osaka: IEEE) pp 3909-12
Chauhan S, Wang P, Sing Lim C and Anantharaman V 2008 A computer-aided MFCC-based HMM
system for automatic auscultation Comput Biol Med 38 221-33
Chen T, Kuan K, Celi L and Clifford G D 2009 Intelligent heartsound diagnostics on a cellphone using a
hands-free kit. In: AAAI Spring Symposium on Articial Intelligence for Development, (Stanford
University pp 26-31
Choi S and Jiang Z 2008 Comparison of envelope extraction algorithms for cardiac sound signal
segmentation Expert Syst Appl 34 1056-69
Clifford G D, Liu C Y, Springer D, Moody B, Li Q, Juan R A, Millet J, Silva I, Johnson A and Mark R G
2016 Classification of normal/abnormal heart sound recordings: the PhysioNet/Computing in
Cardiology Challenge 2016. In: Computing in Cardiology, (Vancouver: IEEE), in press
Clifford G D and Moody G B 2012 Signal quality in cardiorespiratory monitoring Physiol Meas 33 E01
De Vos J P and Blanckenberg M M 2007 Automated pediatric cardiac auscultation IEEE Trans Biomed
Eng 54 244-52
eGeneralMedical Cardiac Auscultation of Heart Murmurs.
http://www.egeneralmedical.com/listohearmur.html
Gamero L G and Watrous R 2003 Detection of the first and second heart sound using probabilistic
models. In: Annual International Conference of the IEEE Engineering in Medicine and Biology
Society, (Cancun: IEEE) pp 2877-80
36
Gerbarg D S, Taranta A, Spagnuolo M and Hofler J J 1963 Computer analysis of phonocardiograms Prog
Cardiovasc Dis 5 393-405
Gharehbaghi A, Dutoir T, Sepehri A, Hult P and Ask P 2011 An automatic tool for pediatric heart sounds
segmentation. In: Computing in Cardiology, (Hangzhou: IEEE) pp 37-40
Gharehbaghi A, Ekman I, Ask P, Nylander E and Janerot-Sjoberg B 2015 Assessment of aortic valve
stenosis severity using intelligent phonocardiography Int J Cardiol 198 58-60
Gill D, Gavrieli N and Intrator N 2005 Detection and identification of heart sounds using homomorphic
envelogram and self-organizing probabilistic model. In: Computers in Cardiology, (Lyon: IEEE)
pp 957-60
Goldberger A L, Amaral L A N, Glass L, Hausdorff J M, Ivanov P C, Mark R G, Mietus J E, Moody G B,
Peng C K and Stanley H E 2000 PhysioBank, PhysioToolkit, and PhysioNet: Components of a
new research resource for complex physiologic signals Circulation 101 e215-e20
Gupta C, Palaniappan R, Swaminathan S and Krishnan S 2007 Neural network classification of
homomorphic segmented heart sounds Appl Soft Comput 7 286-97
Jiang Z and Choi S 2006 A cardiac sound characteristic waveform method for in-home heart disorder
monitoring with electric stethoscope Expert Syst Appl 31 286-98
Kumar D, Carvalho P, Antunes M and Henriques J 2006 Detection of S1 and S2 heart sounds by high
frequency signatures. In: Annual International Conference of the IEEE Engineering in Medicine
and Biology Society, (New York: IEEE) pp 1410-6
Leatham A 1975 Auscultation of the heart and phonocardiography: Churchill Livingstone)
Li T, Tang H, Qiu T S and Park Y 2011 Best subsequence selection of heart sound recording based on
degree of sound periodicity Electron Lett 47 841-3
Liang H and Hartimo I 1998 A feature extraction algorithm based on wavelet packet decomposition for
heart sound signals. In: Proceedings of the IEEE-SP International Symposium on Time-Frequency
and Time-Scale Analysis, (Pittsburgh, PA: IEEE) pp 93-6
Liang H, Lukkarinen S and Hartimo I 1997a Heart sound segmentation algorithm based on heart sound
envelolgram. In: Computing in Cardiology, (Lund: IEEE) pp 105-8
Liang H Y, Sakari L and Iiro H 1997b A heart sound segmentation algorithm using wavelet decomposition
and reconstruction. In: Proceedings of the 19th Annual International Conference of the IEEE
Engineering in Medicine and Biology Society, (Chicago, IL: IEEE) pp 1630-3
Madhero 2010
https://commons.wikimedia.org/wiki/File:Phonocardiograms_from_normal_and_abnormal_heart
_sounds.png.
Maglogiannis I, Loukis E, Zafiropoulos E and Stasis A 2009 Support Vectors Machine-based
identification of heart valve diseases using heart sounds Comput Methods Programs Biomed 95
47-61
Moukadem A, Dieterlen A, Hueber N and Brandt C 2011 Localization of heart sounds based on S-
Transform and radial basis function neural network. In: 15th Nordic-Baltic Conference on
Biomedical Engineering and Medical Physics, (Aalborg: IFMBE Proceedings) pp 168-71
Moukadem A, Dieterlena A, Hueberb N and Brandtc C 2013 A robust heart sounds segmentation module
based on S-transform Biomed Signal Process Control 8 273–81
Naseri H and Homaeinezhad M R 2013 Detection and boundary identification of phonocardiogram
sounds using an expert frequency-energy based metric Ann Biomed Eng 41 279-92
Naseri H, Homaeinezhad M R and Pourkhajeh H 2013 Noise/spike detection in phonocardiogram signal
as a cyclic random process with non-stationary period interval Comput Biol Med 43 1205-13
Nigam V and Priemer R 2005 Accessing heart dynamics to estimate durations of heart sounds Physiol
Meas 26 1005-18
Oskiper T and Watrous R 2002 Detection of the first heart sound using a time-delay neural network. In:
Computing in Cardiology, (Memphis: IEEE) pp 537-40
37
Papadaniil C D and Hadjileontiadis L J 2014 Efficient heart sound segmentation and extraction using
ensemble empirical mode decomposition and kurtosis features IEEE J Biomed Health Inform 18
1138-52
Patidar S, Pachori R B and Garg N 2015 Automatic diagnosis of septal defects based on tunable-Q
wavelet transform of cardiac sound signals Expert Syst Appl 42 3315-26
Pedrosa J, Castro A and Vinhoza T T V 2014 Automatic heart sound segmentation and murmur detection
in pediatric phonocardiograms. In: Annual International Conference of the IEEE Engineering in
Medicine and Biology Society, (Chicago: IEEE) pp 2294-7
Quiceno-Manrique A F, Godino-Llorente J I, Blanco-Velasco M and Castellanos-Dominguez G 2010
Selection of dynamic features based on time-frequency representations for heart murmur
detection from phonocardiographic signals Ann Biomed Eng 38 118-37
Rajan S, Budd E, Stevenson M and Doraiswami R 2006 Unsupervised and uncued segmentation of the
fundamental heart sounds in phonocardiograms using a time-scale representation. In:
International Conference of the IEEE Engineering in Medicine and Biology Society, (New York:
IEEE) pp 3732-5
Ricke A D, Povinelli R J and Johnson M T 2005 Automatic segmentation of heart sound signals using
hidden markov models. In: Computers in Cardiology, (Lyon: IEEE) pp 953-6
Samieinasab M and Sameni R 2015 Fetal phonocardiogram extraction using single channel blind source
separation. In: 23rd Iranian Conference on Electrical Engineering: IEEE) pp 78-83
Saracoglu R 2012 Hidden Markov model-based classification of heart valve disease with PCA for
dimension reduction Eng Appl Artif Intell 25 1523-8
Schmidt S E, Holst-Hansen C, Graff C, Toft E and Struijk J J 2010a Segmentation of heart sound
recordings by a duration-dependent hidden Markov model Physiol Meas 31 513-29
Schmidt S E, Holst-Hansen C, Hansen J, Toft E and Struijk J J 2015 Acoustic features for the
identification of coronary artery disease IEEE Trans Biomed Eng 62 2611-9
Schmidt S E, Toft E, Holst-Hansen C and Struijk J J 2010b Noise and the detection of coronary artery
disease with an electronic stethoscope. In: 2010 5th Cairo International Biomedical Engineering
Conference (CIBEC), (Cairo: IEEE) pp 53-6
Sedighian P, Subudhi A W, Scalzo F and Asgari S 2014 Pediatric heart sound segmentation using Hidden
Markov Model. In: Annual International Conference of the IEEE Engineering in Medicine and
Biology Society, (Chicago: IEEE) pp 5490-3
Sepehri A A, Gharehbaghi A, Dutoit T, Kocharian A and Kiani A 2010 A novel method for pediatric heart
sound segmentation without using the ECG Comput Methods Programs Biomed 99 43-8
Sepehri A A, Hancq J, Dutoit T, Gharehbaghi A, Kocharian A and Kiani A 2008 Computerized screening
of children congenital heart diseases Comput Methods Programs Biomed 92 186-92
Springer D B 2015 Mobile phone-based rheumatic heart disease detection. In: Department of Engineering
Science: University of Oxford)
Springer D B, Tarassenko L and Clifford G D 2014 Support vector machine hidden semi-Markov model-
based heart sound segmentation. In: Computing in Cardiology, (Cambridge, MA: IEEE) pp 625-8
Springer D B, Tarassenko L and Clifford G D 2015 Logistic regression-HSMM-based heart sound
segmentation IEEE Trans Biomed Eng In press
Sun S, Jiang Z, Wang H and Fang Y 2014 Automatic moment segmentation and peak detection analysis of
heart sound pattern via short-time modified Hilbert transform Comput Methods Programs Biomed
114 219-30
Syed Z 2003 MIT automated auscultation system. In: Department of Electrical Engineering and
Computer Science, (Boston: Massachusetts Institute of Technology) pp 73-4
Syed Z, Leeds D, Curtis D, Nesta F, Levine R A and Guttag J 2007 A framework for the analysis of
acoustical cardiac signals IEEE Trans Biomed Eng 54 651-62
Tang H, Li T, Park Y and Qiu T S 2010a Separation of heart sound signal from noise in joint cycle
frequency-time-frequency domains based on fuzzy detection IEEE Trans Biomed Eng 57 2438-47
38
Tang H, Li T and Qiu T S 2010b Noise and disturbance reduction for heart sounds in the cycle frequency
domain based on non-linear time scaling IEEE Trans Biomed Eng 57 325-33
Tang H, Li T, Qiu T S and Park Y 2012 Segmentation of heart sounds based on dynamic clustering
Biomed Signal Process Control 7 509-16
Tilkian A G and Conover M B 2001 Understanding heart sounds and murmurs with an introduction to
lung sounds: Elsevier Health Sciences)
Uguz H 2012a Adaptive neuro-fuzzy inference system for diagnosis of the heart valve diseases using
wavelet transform with entropy Neural Comput Appl 21 1617-28
Uguz H 2012b A biomedical system based on artificial neural network and principal component analysis
for diagnosis of the heart valve diseases J Med Syst 36 61-72
UMHS Michigan Heart Sound and Murmur Library.
http://www.med.umich.edu/lrc/psb_open/html/repo/primer_heartsound/primer_heartsound.html
Varghees V N and Ramachandran K 2014 A novel heart sound activity detection framework for
automated heart sound analysis Biomed Signal Process Control 13 174-88
Vepa J, Tolay P and Jain A 2008 Segmentation of heart sounds using simplicity features and timing
information. In: IEEE International Conference on Acoustics, Speech and Signal Processing, (Las
Vegas, NV: IEEE) pp 469-72
Wang P, Lim C S, Chauhan S, Foo J Y and Anantharaman V 2007 Phonocardiographic signal analysis
method using a modified hidden Markov model Ann Biomed Eng 35 367-74
WHO 2015 World statistics on cardiovascular disease
http://www.who.int/mediacentre/factsheets/fs317/en/
Yan Z, Jiang Z, Miyamoto A and Wei Y 2010 The moment segmentation analysis of heart sound pattern
Comput Methods Programs Biomed 98 140-50
Zheng Y N, Guo X M and Ding X R 2015 A novel hybrid energy fraction and entropy-based approach for
systolic heart murmurs identification Expert Syst Appl 42 2710-21
... Heart Beat Cycle Fig. 3. The PCG recording of a normal heart sound recording from the PhysioNet/CinC Database [58]. Frames in the middle with four states labelled (i. ...
... The PhysioNet/CinC Database [58], used in the Phy-sioNet/CinC Challenge 2016 [143], consists of multiple databases that were recorded from different data collectors. Especially, the publicly available training set has five databases collected from both healthy individuals and patients. ...
... We discuss publicly available codes in the papers discussed in Section IV-B as well as other papers on openaccess repositories during 2017-2022. There are only a few publicly available codes in the past years, while there are many more released codes in 2016 benefiting from the PhysioNet challenge 2016 [58]. Notably, codes in 2016 are not listed, as we want to list the most state-of-the-art studies in the past six years. ...
Preprint
Full-text available
Heart sound auscultation has been demonstrated to be beneficial in clinical usage for early screening of cardiovascular diseases. Due to the high requirement of well-trained professionals for auscultation, automatic auscultation benefiting from signal processing and machine learning can help auxiliary diagnosis and reduce the burdens of training professional clinicians. Nevertheless, classic machine learning is limited to performance improvement in the era of big data. Deep learning has achieved better performance than classic machine learning in many research fields, as it employs more complex model architectures with stronger capability of extracting effective representations. Deep learning has been successfully applied to heart sound analysis in the past years. As most review works about heart sound analysis were given before 2017, the present survey is the first to work on a comprehensive overview to summarise papers on heart sound analysis with deep learning in the past six years 2017--2022. We introduce both classic machine learning and deep learning for comparison, and further offer insights about the advances and future research directions in deep learning for heart sound analysis.
... Early diagnosis and treatment can be helpful in alleviating the adverse outcomes of CVDs. Heart sounds, generated by the mechanical activity of the heart valves during blood flow, reveal many diagnostically important information regarding the physiological condition of the heart [3]. Thus, the phonocardiogram (PCG) is of paramount importance, being an effective, non-invasive, and comparatively inexpensive way for preliminary screening of CVDs [4]. ...
... However, reliable and objective assessment of CVDs through these computerized methods is still a challenge. High susceptibility to motion artifact during signal acquisition, intrinsic spectral overlap among heart sound and lung sound (breathing sound), noise due to power supply interference, ambient noise in a hospital (door opening/closing/knocking, phone ringing, movement and speech of other patients in the outpatient department (OPD)), distortions introduced by the variation in device functionality (diaphragm, sensor, amplifier) affect the quality of PCGs and may even mask the presence of abnormalities in the perceived signal [3]. These factors can further degrade the performance of the AI-based automated diagnostic methods in real-life scenario [5]. ...
... In the case of heart sounds, one of the primary causes of signal deterioration is the impact of noise from several sources, such as lung sounds caused by breathing, hospital ambient noise, human conversation, intestinal activity, stethoscope movements, sensor variability, and so forth [3]. Susceptibility to noise is a matter of concern in the automated evaluation of cardiac disorders, especially in lowresource settings. ...
Preprint
p>Objective: The heart sound signals captured via a digital stethoscope are often distorted by environmental and physiological noise, altering their salient and critical properties. The problem is exacerbated in crowded low-resource hospital settings with high noise levels which degrades the diagnostic performance. In this study, we present a novel deep encoder-decoder based denoising architecture (LU-Net) to suppress ambient and internal lung sound noises. Methods: Training is done using a large benchmark PCG dataset mixed with physiological noise, i.e., breathing sounds. Two different noisy datasets were prepared for experimental evaluation by mixing unseen lung sounds and hospital ambient noises with the clean heart sound recordings. We also use the inherently noisy portion of the PASCAL heart sound dataset for evaluation. Results: The proposed framework showed effective suppression of background noises in both un?seen real-world data and synthetically generated noisy heart sound recordings, improving the signal-to-noise ratio (SNR) level by 5.575 dB on an average using only 1.32 M parameters. The proposed model outperforms the current state-of-the-art U-Net model with an average SNR improvement of 5.613 dB and 5.537 dB in the presence of lung sound and unseen hospital noise, respectively. LU-Net also outperformed the state-of-the-art Fully Convolutional Network (FCN) by 1.750 dB and 1.748 dB for lung sound and unseen hospital noise conditions, respectively. In addition, the proposed denoising method model improves classification accuracy by 38.93% in the noisy portion of the PASCAL heart sound dataset. Conclusion: The results presented in the paper indicate that our proposed architecture demonstrated a robust denoising performance on different datasets with diverse levels and characteristics of noise. Significance: The proposed deep learning-based PCG denoising approach is a pioneering study that can significantly improve the accuracy of computer-aided auscultation systems for detecting cardiac diseases in noisy, low-resource hospitals and underserved communities. </p
... Stethoscopes with bells and diaphragms were employed to gather samples. B) The PhysioNet/CinC 2016 Challenge (Liu et al. 2016) intends to create algorithms for categorizing sound captured from the collected heart in a variety of clinical and non-clinical settings. Sound recordings of the heart were taken from various areas of the body. ...
... edu/ lrc/ psb_ open/ html/ repo/ primer_ heart sound/ primer_ heart sound. html", and The PhysioNet/CinC 2016 Challenge(Liu et al. 2016), accessible from "https:// physi onet. org/ conte nt/ chall enge-2016/1. ...
Article
Full-text available
Among all causes of death, cardiovascular disease (CVD) is the leading one. In the medical field, the phonocardiogram (PCG) signal is widely used due to its cost-effectiveness and simplicity. We propose a hybrid classification approach for automatically classifying different cardiac auscultation classes using a PCG signal using handcrafted and deep learned features. Firstly, the PCG signal is split into equal segments using the windowing procedure. The power spectrum of a signal is constructed using l-spectrograms and then the novel modified dilated ResNet structure generates the deep features from the spectrogram images. Simultaneously, handcrafted frequency domain features are extracted from correspond windowed signals. After features are extracted, radial basis function (RBF) kernel in support vector machines (SVM) is used to classify various patterns. Moreover, the optimized SVM is performed using a modified drone squadron optimization (mDSO) algorithm to prevent overfitting. The proposed approach is 94.16% accurate for Michigan Heart Sound and Murmur Database (MHSDB) and 99.38% accurate for PhysioNet/CinC 2016 Challenge. The introduced hybrid model is more computationally efficient and accurate than earlier state-of-the-art strategies, based on the evaluation measures. Using the designed model, underserved and diverse communities can implement automated cardiac screening systems.
... The users, especially the elderly patients, can detect their own heart health status anytime and anywhere through smart stethoscope. The PhysioNet/CinC Challenge 2016 database [72] is currently the largest publicly available database of heart sounds. The database labels all pathologies, i. e., Murmurs relating to mitral Valve Prolapse (MVP), Innocent or Benign Murmurs (IBM), Aortic Disease (AD), Coronary Artery Disease (CAD), Mitral Regurgitation (MR), Aortic Stenosis (AS), Pathologic (PA), and other Miscellaneous Pathological Conditions (MPC), as abnormal. ...
... For contrastive analysis of six time-frequency representations and performance evaluation of the proposed method, we use sensitivity Se, specificity Sp, and mean accuracy MAcc as figures-of-merit according to the official scoring mechanism of the PhysioNet/CinC Challenge 2016 [72]. Here, Se is defined as the proportion of the detected abnormal heart sound samples in the actual total abnormal heart sound samples, which reflects the ability of the diagnostic method to determine the abnormal heart sound. ...
Article
Full-text available
The advantages of non-invasive, real-time and convenient, computer audition-based heart sound abnormality detection methods have increasingly attracted efforts among the community of cardiovascular diseases. Time-frequency analyses are crucial for computer audition-based applications. However, a comprehensive investigation on discovering an optimised way for extracting time-frequency representations from heart sounds is lacking until now. To this end, we propose a comprehensive investigation on time-frequency methods for analysing the heart sound, i. e., short-time Fourier transformation, Log-Mel transformation, Hilbert-Huang transformation, wavelet transformation , Mel transformation, and Stockwell transformation. The time-frequency representations are automatically learnt via pre-trained deep convolutional neural networks. Considering the urgent need of smart stethoscopes for high robust detection algorithms in real environment, the training, verification, and testing sets employed in the extensive evaluation are subject-independent. Finally, to further understand the heart sound-based digital phenotype for cardiovascular diseases, explainable artificial intelligence approaches are used to reveal the reasons for the performance differences of six time-frequency representations in heart sound abnormality detection. Experimental results show thatStockwell transformation can beat other methods by reaching the highest overall score of 65.2 %. The interpretable results demonstrate that Stockwell transformation does not only present more information for heart sounds, but also provides a certain noise robustness. Besides, the considered fine-tuned deep model brings an improvement of the mean accuracy over the previous state-of-the-art results by 9.0 % in subject-independent testing.
... Due to the lack of publicly available PCG datasets considering mental stress as a response variable, this work will use the Physionet CinC 2016 challenge PCG heart disease database [16] to develop an automated binary classification system for predicting two cardiovascular classes from PCG signals, namely 'normal' and 'abnormal'. How the resulting model will behave when trained on a non-heart disease-related response variable will have to be tested in later work when a sizable mental stress database is collected. ...
... The PhysioNet/Computing in cardiography 2016 challenge PCG dataset [16] has been used to assess the status of heart abnormality. The dataset was made to encourage development in segmentation and classification algorithms for heart sound data. ...
Thesis
Full-text available
Phonocardiography (PCG) is a noninvasive biomedical measurement technique conventionally used to examine heart health and to study audible phenomena that are perceivable at the chest wall in the cardiac region. Recent research shows impressive results using machine learning methods on PCG data, e.g. heart disease classification and heart sound segmentation with performance more than adequate for clinical use. Because of recent breakthroughs, the applications of machine learning with PCG data for non-conventional applications have been put on the radar. This specialization project summarizes, based on recent literature and testing, different preprocessing and feature extraction techniques for PCG data, and classifiers applicable to these features. Based on the findings, a generalized binary classification model that can be modified for PCG datasets with unconventional labeling is presented. In the presented work, a binary classification system has been made for the PhysioNet CinC 2016 challenge database consisting of unsegmented PCG recordings labeled with heart disease ('normal/healthy' or 'abnormal/unhealthy' heart function). The database is unbalanced with an over-representation of healthy recordings. 1D adaptive ternary patterns (1D-ALTP), Mel frequency cepstral coefficient (MFCC), and wavelet scattering coefficient (WSC) features have been tested with K-nearest neighbor (KNN) and support vector machine (SVM) classifiers. Empirical mode decomposition (EMD) as a signal-denoising technique was evaluated, however, the results indicated limited to no performance improvement. A test set accuracy, sensitivity, and specificity of 97,4%, 94,7%, and 98,3% respectively have been achieved with a model based on serialized features with a KNN classifier (SF+KNN1-MAN), and another simpler model using only WSC features and a KNN classifier (WSC+KNN1-MAN) has achieved a test set accuracy, sensitivity, and specificity of 96,7%, 93.5%, and 97.9% respectively. Both models have shown strong sensitivity scores which is important when training on a database with healthy heart function bias. As part of the project, a mental health dataset has been collected of students experiencing mental stress in the context of examination or project deliveries before end of year 2022. The dataset will be finished in January 2023 as part of a master thesis, extending the work in this project. PCG recordings were done with EKO DUO digital stethoscope and Lab Streaming Layer. This dataset will be used for detecting stress/no stress in the master thesis and the algorithm developed in this specialization project will be adapted for the classification of the stress/no stress classes. 2
... In this study we utilize the popular and public PhysioNet/CinC Challenge 2016 heart sound database [22,34], which is available at the following website: https://physionet.org/content/ challenge-2016. ...
... The datasets used and/or analyzed during in current study are available from the public PhysioNet/CinC Challenge heart sound database [22,34] and heart valve disease (HVD) database [53]. ...
Article
Full-text available
Phonocardiogram (PCG) is commonly used as a diagnostic tool in ambulatory monitoring in order to evaluate cardiac abnormalities and detect cardiovascular diseases. Although cardiac auscultation is widely used for evaluation of cardiac function, the analysis of heart sound signals mostly depends on the clinician’s experience and skills. There is growing demand for automatic and objective heart sound interpretation techniques. The objective of this study is to develop an automatic classification method for anomaly (binary and multi-class) detection of PCG recordings without any segmentation. A deep neural network (DNN) model is used on the raw data during the extraction of the features of the PCG inputs. Deep feature maps obtained from hierarchically placed layers in DNN are fed to various shallow classifiers for the anomaly detection, including support vector classifier (SVC), k-nearest neighbors (KNN), random forest (RF), gradient boosting (GB) classifier, decision tree (DT) classifier, quadratic discriminant analysis (QDA), and multi-layer perception (MLP). Principal component analysis (PCA) technique is used to reduce the high dimensions of feature maps.Finally, two famous heart sound databases, namely PhysioNet/Computing in Cardiology (CinC) Challenge heart sound database and heart valve disease (HVD) database, are used for evaluation. The databases are significantly different in terms of the tools used for data acquisition, clinical protocols, digital storages and signal qualities, making it challenging to process and analyze. By using the 10-fold cross-validation style, experimental results demonstrate that the proposed deep features with shallow classifiers yield highest performance with accuracy of 99.61% and 99.44% for binary and multi-class classification on the two databases, respectively. The results indicate that our method is effective for the detection of abnormal heart sound signals and outperforms other state-of-the-art methods.
... More specifically, they were either performed during in-home visits or in the hospital. Furthermore, they also included stethoscope movements, breathing and intestinal sounds [23]. ...
... Data Availability Statement: Please refer to suggested Data Availability Statements at Ref. [23]. ...
Article
Full-text available
The accurate localization of S1 and S2 is essential for heart sound segmentation and classification. However, current direct heart sound segmentation algorithms have poor noise immunity and low accuracy. Therefore, this paper proposes a new optimal heart sound segmentation algorithm based on K-means clustering and Haar wavelet transform. The algorithm includes three parts. Firstly, this method uses the Viola integral method and Shannon’s energy-based algorithm to extract the function of the envelope of the heart sound energy. Secondly, the time–frequency domain features of the acquired envelope are extracted from different dimensions and the optimal peak is searched adaptively based on a dynamic segmentation threshold. Finally, K-means clustering and Haar wavelet transform are implemented to localize S1 and S2 of heart sounds in the time domain. After validation, the recognition rate of S1 reached 98.02% and that of S2 reached 96.76%. The model outperforms other effective methods that have been implemented. The algorithm has high robustness and noise immunity. Therefore, it can provide a new method for feature extraction and analysis of heart sound signals collected in clinical settings.