Content uploaded by Mattia Stagnaro
Author content
All content in this area was uploaded by Mattia Stagnaro on Nov 03, 2016
Content may be subject to copyright.
Available via license: CC BY 3.0
Content may be subject to copyright.
Performance of post-processing algorithms for rainfall intensity
measurements of tipping-bucket rain gauges
Mattia Stagnaro1,2, Matteo Colli1,2 , Luca Giovanni Lanza1,2, and Pak Wai Chan3
1University of Genova, Department of Civil, Chemical and Environmental Engineering, Via Montallegro 1, 16145 Genova,
Italy
2WMO/CIMO Lead Centre “Benedetto Castelli” on Precipitation Intensity, Italy
3Hong Kong Observatory, 134A Nathan Road, Honk Kong, China
Correspondence to: Mattia Stagnaro (Mattia.Stagnaro@unige.it)
Abstract. A number of rain events recorded from May to September 2013 at the Hong Kong International Airport (HKIA)
have been selected to investigate the performance of post-processing algorithms used to calculate the Rainfall Intensity (RI)
from Tipping-Bucket Rain Gauges (TBRGs). We assumed a drop counter catching-type gauge as a working reference and
compared rainfall intensity measurements with two calibrated TBRGs operated at a time resolution of 1 min. The two TBRGs
differ in their internal mechanics, one being a traditional single-layer dual-bucket assembly, while the other has two layers of5
buckets. The drop counter gauge operates at a time resolution of 10 s, while the time of tipping is recorded for the two TBRGs.
The post-processing algorithms employed for the two TBRGs are based on the assumption that the tip volume is uniformly
distributed over the inter-tip period. A series of data of an ideal TBRG is reconstructed using the virtual time of tipping derived
from the drop counter data. From the comparison between the ideal gauge and the measurements from the two real TBRGs
the performance of different post-processing and correction algorithms are statistically evaluated over the set of recorded rain10
events. The improvement obtained by adopting the inter-tip time algorithm in the calculation of the RI is confirmed. However,
by comparing the performance of the real and ideal TBRGs, the beneficial effect of the inter-tip algorithm is shown to be
relevant for the mid-low range of rainfall intensity values (where the sampling errors prevail), while its role vanishes with
increasing the RI, in the range where the mechanical errors prevail.
1 Introduction15
Application-driven requirements of rainfall data (see e.g. Lanza and Stagi, 2008), the recommendations of international bodies
such as the World Meteorological Organization (WMO, 2008), and new measurement quality standards issued at the national
(UNI 11452:2012; BS 7843-3:2012) and international (CEN/TR 16469:2013) level provide an increasingly demanding frame-
work in terms of proven instrumental accuracy and reliability.
Following the effort leaded in the last decade by WMO and aimed at quantifying the achievable accuracy of rainfall intensity20
measurements (Lanza and Vuerich, 2009), both users and manufacturers of precipitation gauges are developing strategies to
reduce the uncertainty and to provide suitably documented performance evaluation of rainfall measurements.
Sound metrological procedures for the assessment of the uncertainty of meteorological measurements have been recently
1
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
introduced within the framework of Europe-wide collaborative projects (Merlone et al., 2015), and therein extended to the
measurement of liquid precipitation (see Santana et al., 2015).
Beside the inherent instrumental factors (e.g. the systematic mechanical bias of tipping-bucket rain gauges, or the dynamic
response bias of weighing gauges), post-processing of the raw data to obtain accurate rain intensity records at a pre-determined
temporal resolution is common practice in rain gauge measurements. In the case of tipping-bucket rain gauges (TBRGs), ded-5
icated post-processing algorithms must be employed to achieve sufficient accuracy and to minimize the impact of sampling
errors and the discrete nature of the measurement.
Various algorithms have been proposed to this aim, and discussed in the literature (Costello and Williams, 1991; Habib
et al., 2001; Colli et al., 2013b, a). Yet, the operational practice of most users, including national weather services, still relies
on the trivial counting of the number of tips occurring in the desired time frame, and the rain depth per one minute (the WMO10
recommended time frame for rain intensity measurements) is obtained as the product of that number by the nominal volume of
the bucket.
This method (as already observed by Costello and Williams, 1991) results in a general underestimation of rain intensity figures
and in a high level of uncertainty, due to the random nature of the number of tips per minute within any real-world, highly
variable rainfall event. Moreover, the correction of systematic mechanical biases can not be optimized with this method since15
it would be applied to the averaged values only, and most tipping-bucket rain gauges show a nonlinear correction curve after
laboratory calibration (Lanza and Stagi, 2009).
In this paper, the performance of different post-processing algorithms employed in the calculation of the rainfall intensity
from tipping-bucked rain gauges are compared and discussed. Data recorded at a field test site by two TBRGs using different
mechanical solution are used, and a catching-type drop-counting gauge is assumed as the working reference. The comparison20
aims to highlight the benefits of employing smart algorithms in post-processing of the raw data, and their capability to improve
the accuracy of rain intensity measurements obtained from TBRGs.
2 Field site and instrumentation
The Hong Kong Observatory performs rainfall measurements at the weather station of the Hong Kong International Airport
(HKIA). An Ogawa drop-counting rain gauge, model Osaka PC1122 (OSK), is available at the field site, providing rainfall25
measurements at the time resolution of 10 s.
Based on the calibrated drop size volume (calculated in October 2013) of 63.93 mm3, the OSK drop counter rain gauge is able
to measure rainfall rates up to 200 mm h−1with a resolution of 5.21·10−3mm (Chan and Yeung, 2004; Colli et al., 2013b),
therefore fulfilling the WMO (2008) accuracy requirements.
Due to the high accuracy and time resolution we adopted the OSK drop counter gauge as a working reference, to be compared30
with co-located observations performed by two TBRGs manufactured by Logotronic MRF-C (LGO) and Shanghai SL3-1
(SL3). The main characteristics of the instruments employed in this study are summarized in Table 1. Note, in particular, that
the two TBRGs have the same nominal sensitivity (equal to 0.1 mm), but employ different mechanical solutions. The LGO
2
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
is a traditional TBRG with a dual-compartment single-bucket assembly, while the SL3 gauge has two consecutive layers of
dual-compartment buckets (see Figure 1).
This is not a common solution for TBRGs, and the objective of the two layers of buckets employed seems to reside in the
attempt to reduce the systematic mechanical bias, typical of traditional TBRGs. In a sense, this is a hardware-type of correction
similar to the use of a syphon or other mechanical solutions.5
In order to correct the systematic mechanical errors of the two TBRGs, both of them were subjected to appropriate dynamic
calibration in the laboratory. The dynamic calibration consists of providing the gauge with a sufficient number of equivalent
rainfall intensities, using calibrated constant flow rates. By comparing the reference values with those measured by the rain
gauge under test, the parameters of a suitable correction curve (usually a power law) are derived. The measurements from the
two TBRGs were corrected before performing any comparison with the drop counter time series and/or with the ideal TBRG10
data obtained from the reference.
In this work, the computation of statistical estimators and deviations between paired observations was performed with no
reference to any ancillary data (wind speed and direction, air temperature and absolute pressure, etc.) although it is known that
some of them (especially the wind) may actually affect the accuracy of the measurement.
The field data available for this study cover a five-month period of observations from May to September 2013. Eight significant15
events in this period were selected on the basis of the total rainfall depth, after checking that the reference rainfall intensity
values were lower than the given factory limits for the instruments under test.
Table 2 reports a short description of the selected events, in terms of total rainfall depth (htot), maximum rainfall intensity in
one minute (Imax) and event duration (d).
Figure 2 shows a sample hyetograph of the raw data for the rain event occurred on 22 May 2013. The one-minute reference20
rainfall intensity is depicted (shaded gray background) as calculated from the 10 shigh-resolution data of the OSK drop
counter, as well as the cumulated rainfall for the OSK (solid red line) and the SL3 and LGO gauges (dashed and dotter line).
The underestimation of the cumulative rain depth by the two TBRGs is evident from Figure 2. The relative underestimation of
the total depth, for this particular event, is equal to 13.6% and 12.5% for the SL3 and LGO rain gauge respectively.
Figure 3 reports a non-parametric statistical description of the set of events in terms of one-minute rainfall intensity distri-25
bution from the reference DC (upper part), together with the total accumulation for the DC and the two TBRGs (lower part).
Note that the two TBRGs show a larger underestimation of the total rain amount for events characterized by the highest values
of the rainfall rate, in terms of both the mean and the extreme values, while the difference decreases for lower RI events.
3 Method
We adopted the catching-type drop counter gauge as the working reference for this work due to the high sensitivity of the30
measurement. Indeed, the instrument provides the number of generated drops with a time resolution of 10 s. We first aggregated
this information to obtain the one-minute reference rainfall intensity values (RIref), for use in the overall assessment of the
accuracy of the two involved TBRGs.
3
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
Both the SL3 and LGO rain gauges provide records of the time stamp of each tip. This feature allows using various algorithms
to calculate the one-minute rainfall intensity values for the two investigated TBRGs.
The first, traditional and widely applied method to derive the one-minute rainfall intensity (RIraw) simply relies on the
counting of the number of tips within each minute. The product of this number by the nominal volume of the bucket provides
the rainfall amount in any single minute, and therefore the average rainfall intensity at such and any higher time resolution.5
Using a suitable correction curve derived from laboratory calibration allows accounting for systematic mechanical errors as a
function of the rainfall intensity. The traditional method assigns the whole volume of each bucket to the minute in which the tip
occurs, even when part of the bucket is actually filled already in the previous minute, introducing significant counting errors in
the calculation of the one-minute rainfall intensity RIraw. The uncertainty introduced by the tip counting method also affects
the efficacy of the calibration, since the correction applied to the volume of the bucket in each minute does not precisely derive10
from the actual rain intensity occurring in that minute.
The second method used to obtain the one-minute rainfall intensity values (RIT tip) employs the inter-tip time algorithm (see
e.g., Costello and Williams, 1991; Colli et al., 2013b), which is based on the assumption that the nominal volume of each
bucket is equally distributed over the inter-tip period. The calculation of the RIT tip for each minute accounts for the portion of
the inter-tip period actually falling into that minute. In this way, also the calibration is the most effective since the correction15
applied to the volume of the bucket at the variable inter-tip scale is precisely the one corresponding to the measured rainfall
intensity.
For both the TBRGs, the two values of one-minute RI derived from the two post-processing algorithms described above
(generally indicated here as the measured rain intensity RIm) were employed to calculate the accuracy of rainfall intensity
measurements in terms of deviations from the DC reference value as follow:20
erel(%) = RIm−RIr ef
RIref
·100 (1)
In addition, in order to compare the performance of post-processing and correction algorithms for the TBRG measurements,
we derived a virtual sequence of tips of an ideal TBRG from the high-resolution DC data. This simulates the behavior of a best
performing TBRG, representing the maximum accuracy to be expected when using a tipping-bucket mechanics. The inter-tip
algorithm was employed to derive the one-minute ideal rainfall intensity values (RIideal).25
In order to assess the capability of the employed algorithm to describe the inner variability of the considered events and to
capture their finer details, we calculated the correlation coefficients between all the derived time series and the reference ones.
In particular, for each event, we calculated the RMS error of paired deviations between the measured and the ideal/reference
RI signal.
In an effort to homogenize the dataset, we considered the normalized value of the generic rainfall intensity value RI obtained30
as follow:
RIn(%) = (RI −µref )
σref
(2)
4
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
where the mean value µref and the standard deviation σref employed to obtain the normalized values of RI for all the TBRGs,
are those derived from the drop counter reference rainfall rates. In this way, we obtained comparable results in terms of standard
deviation of the normalized RI time series for all the investigated rainfall events.
4 Results
We first evaluated the accuracy of the investigated TBRGs by comparing their performance with the working reference. Figure5
4 shows the relative deviations (erel) from the reference DC for the three TBRGs (including the ideal virtual gauge). The
reported boxplots provide a synthesis of the results obtained adopting the inter-tip algorithm and are classified according to
different ranges of RIref .
Two different regions of this graph show different behavior of the relative deviations: for low values of the RI the relative
deviations of all the TBRGs exhibit a large variability, whereas this scatter suddenly decreases just above the RI value of 610
mm h−1(highlighted in black) and then continues to reduce with increasing the RI. This limit coincides with the sensitivity
of the TBRGs (both the real and virtual ones); in fact, values of RI higher than 6 mm/h generate at least one tip per minute for
TBRGs with a sensitivity of 0.1 mm.
It emerges from the graph in Figure 4 that the average deviation of both the LGO and SL3 gauges is always negative. In
particular, the SL3 gauge shows an average underestimation that is larger than the LGO (except for one bin). This is coherent15
with the daily amounts shown in Figure 3, where the cumulative rain depth of the SL3 gauge is lower than or equal to the value
recorded by the LGO. Despite the SL3 rain gauge underestimates rainfall more than the LGO on average, the variability of the
deviations from the reference for different values of RI is slightly reduced with respect to the LGO.
The behavior of the ideal TBRG is clearly different, since it is not affected by instrumental mechanical errors (ideal mechanics),
and the average value of erel becomes close to zero immediately after the threshold value of the instrument sensitivity.20
Note that the ideal TBRG shows a large variability and an average value of the relative deviations that is comparable to the
real TBRGs when the RI values are below 6 mm h−1. This means that the error caused by the aggregation time is relevant in
this region when compared to the mechanical one.
As the rainfall rate increases, the variability of erel considerably decreases, even if does not vanish due to the sampling time of
the DC (10 seconds), whereas the average values are close to zero.25
Figure 5 describes the relative deviation (erel) from the DC reference of the measured value of RI from the three TBRGs
adopting the rough tip counting algorithm. Below the instrument sensitivity limit of 6 mm h−1, where less than one tip per
minute occurs, the relative errors of all TBRGs are similar and show a very high scatter. Above this threshold value, the
variability gradually decreases. Also in this case the ideal TBRG shows an average value of deviation which is close to zero,
while the SL3 and LGO continue to maintain a mean value of erel negative for all the RI classes . Close to the threshold value30
of 6 mm h−1, the variability of the relative error (erel) is drastically reduced for all three gauges, and the average value is the
closest to zero.
5
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
In Figure 6 the Taylor diagram is depicted (Taylor, 2001) to show the effect of the inter-tip algorithm in terms of the standard
deviation of the normalized RI signal, the correlation coefficient between the TBRGs and the reference and its deviations from
the reference. Considering the standard deviation of the RI signal for the real TBRGs, note that the two algorithms used to
compute the RI values provide comparable results (approximately equal to 0.9). Since the ideal TBRG directly derives from
the reference, the normalized standard deviation is the closest to unity.5
In the same figure, the benefit of using the inter-tip time algorithm instead of the tip counting is evident: the correlation
coefficients of the two real TBRGs increase using the former one for both TBRGs. Therefore, the beneficial effect in terms
of deviations of the measured RI from the reference is highlighted. In fact, note the reduction of the RMS difference from the
reference, approaching the value of the ideal TBRG. This reduction can be quantified in about 1 mm h−1.
Also, the RI time series of the ideal TBRG show a normalized standard deviation approximately equal to 1, that is the same10
as the reference, and a mean correlation coefficient greater than 0.99. However, the average value of the RMS between the
synthetic TBRG and the reference continues to show a relevant value slightly below 2 mm h−1, which does not decreases
under the value of 1 mm h−1in all the considered events.
In order to evaluate the effectiveness of the post-processing algorithms on the accuracy of the measurements over different
ranges of rainfall intensity, we plotted the standard deviation of the relative error (erel) for different classes of RI. Figure 715
reports the results of this analysis.
It is evident from the graph that the raw counting of the number of tips results, for both the investigated TBRG, in a
continuous trend of linear (in a log-log scale) reduction of the error variance with increasing the RI. This reflects the fact that
the random attribution of one tip to the wrong minute does not affect much the derived RI since the number of tips per minute
is relatively high.20
At very low values of RI, there is little difference between the results obtained by employing the simple counting of tips or the
inter-tip algorithm, with respect to the ideal TBR. By increasing the RI, but still below the threshold intensity corresponding
to the sensitivity of the gauges, the effectiveness of the inter-tip time algorithm is relevant and results are very near to the
ideal gauge. This effectiveness decreases beyond the sensitivity value and, with increasing the RI beyond 50-60 mm h−1, the
difference with respect to the counting of tips becomes negligible. In this range, the LGO performs slightly better than the SL325
when the inter-tip time algorithm is used.
5 Conclusions
The raw data recorded during a dedicated monitoring campaign have been analysed using two different post-processing algo-
rithms to calculate the one-minute RI series. The results allow comparing the performance of the inter-tip time algorithm with
the more common tip counting method, when using two different types of TBRGs. The field reference chosen for this compar-30
ison is a catching type, optical drop counter that, although calibrated in the laboratory, is still subject to unknown uncertainties
in field operation. Notwithstanding this residual uncertainty, comparison of the two gauges with a virtual TBRG obtained from
6
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
the reference measurements was able to show relevant differences in the calculated rainfall intensity, and the relation of such
differences with the rainfall rate itself.
In particular, the main benefit of adopting the inter-tip time method as a post-processing algorithm to calculate rainfall
intensity from the raw data resides in a better representation of the inner variability of rainfall events. The measured RI series
shows an improved correlation coefficient and a lower RMS with respect to the reference, closely approaching the performance5
of an ideal TBRG, which is not affected by mechanical biases.
In terms of accuracy, the inter-tip time algorithm contributes its greater beneficial effects in the range of low to mid RI values.
In the very low RI range, below the threshold value of 6 mm h−1(corresponding to the sensitivity of the involved instrument
and a typical value for operational TBRG) the performance of the inter-tip time method in terms of the statistical amplitude of
the deviations from the reference of the calculated RI are comparable to an ideal TBRG. In this range, indeed, the time period10
between consecutive tips exceeds the time resolution of the measurement, and the sampling error represents the main source
of uncertainty.
Beyond that threshold value, a step change is observed since at least one tip per minute is recorded when RI > 6 mm h−1.
The inter-tip time algorithm is still better than the tip counting method in this range, up to about 50 mm h−1, although the
performance are no more comparable to the ideal gauge. Mechanical errors become prevalent here, so that deviations from15
the ideal gauge result from the residual uncertainty of the calibration process. At the highest RI values the benefits of the
inter-tip time algorithm vanish due to the high number of tips per minutes recorded in this range, and the performance of the
two post-processing algorithms become comparable, though both of them perform worse than the ideal TBRG.
Data availability: the data presented in this paper are available on request from the corresponding author.
Acknowledgements. This research is supported through funding from European Metrology Research Programme MeteoMet 2 (ENV58-20
REG3).
7
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
References
BS 7843-3:2012: Acquisition and management of meteorological precipitation data from a gauge network., Standard, British Standards
Institution, 2012.
CEN/TR 16469:2013: Hydrometry - Measurement of the rainfall intensity (liquid precipitation): requirements, calibration methods and field
measurements, Standard, European Committee for Standardization, 2013.5
Chan, P. W. and Yeung, C. L.: A Study of Drop-counting Rain Gauges, Hong Kong Observatory, 2004.
Colli, M., Lanza, L., and La Barbera, P.: Performance of a weighing rain gauge under laboratory simulated time-varying reference rainfall
rates, Atmospheric Research, 131, 3–12, 2013a.
Colli, M., Lanza, L. G., and Chan, P. W.: Co-located tipping-bucket and optical drop counter RI measurements and a simulated correction
algorithm, Atmospheric Research, 119, 3–12, 2013b.10
Costello, T. A. and Williams, H. J.: Short duration rainfall intensity measured using calibrated time-of-tip data from a tipping bucket raingage,
Agricultural and Forest Meteorology, 57, 147 – 155, 1991.
Habib, E., Krajewski, W. F., and Kruger, A.: Sampling Errors of Tipping-Bucket Rain Gauge Measurements, Journal of Hydrologic Engi-
neering, 6, 159–166, 2001.
Lanza, L. and Stagi, L.: Certified accuracy of rainfall data as a standard requirement in scientific investigations, Advances in geosciences,15
16, 43–48, 2008.
Lanza, L. G. and Stagi, L.: High resolution performance of catching type rain gauges from the laboratory phase of the WMO Field Intercom-
parison of Rain Intensity Gauges, Atmospheric Research, 94, 555–563, 2009.
Lanza, L. G. and Vuerich, E.: The WMO field intercomparison of rain intensity gauges, Atmospheric Research, 94, 534–543, 2009.
Merlone, A., Lopardo, G., Sanna, F., Bell, S., Benyon, R., Bergerud, R., Bertiglia, F., Bojkovski, J., Böse, N., Brunet, M., et al.: The MeteoMet20
project–metrology for meteorology: challenges and results, Meteorological Applications, 22, 820–829, 2015.
Santana, M. A., Guimarães, P. L., Lanza, L. G., and Vuerich, E.: Metrological analysis of a gravimetric calibration system for tipping-bucket
rain gauges, Meteorological Applications, 22, 879–885, 2015.
Taylor, K. E.: Summarizing multiple aspects of model performance in a single diagram, Journal of Geophysical Research, 106, 7183–7192,
2001.25
UNI 11452:2012: Hydrometry - Measurement of rainfall intensity (liquid precipitation) - Metrological requirements and test methods for
catching type gauges., Standard, Ente Nazionale Italiano di Unificazione, Milano, IT, 2012.
WMO: Guide to Meteorological Instruments and Methods of Observation-No. 8, World Meteorological Organization, 7th edn., 2008.
8
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
Figure 1. Shanghai SL3-1 (SL3): internal mechanism with double layer of tipping buckets (a) and external case (b).
Figure 2. Reference rainfall intensity measured by the Ogawa drop counter (shaded gray background) and comparison of the cumulated
reference with the cumulated Logotronic (LGO) and Shanghai (SL3) measurements during the sample event of 22 May 2013.
9
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
Figure 3. Non-parametric distribution of the reference rainfall intensity for each event and corresponding daily rain amount for the Drop
Counter (reference) and the two TBRGs (SL3 and LGO).
Figure 4. One-minute relative deviations (erel) between the three TBRGs (including the ideal one) and the reference (DC) when adopting
the inter-tip post-processing algorithm.
10
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
Figure 5. One-minute relative deviations (erel) between the three TBRGs (including the ideal one) and the reference (DC) when adopting
the tip counting algorithm.
Figure 6. Taylor diagram representation of pattern statistics of the various RI series. The radial distance from the origin is proportional to
the normalized standard deviation of the RI signal; the blue contour lines highlight the RMS difference from the reference (blue dot) for
each recorded event; the azimuthal position indicates the correlation coefficient between the RI signal and the reference. Crosses indicate the
statistics of each single event, while the dots indicate the average values of the whole campaign (colors according to the legend).
11
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
Figure 7. Standard deviation of the relative error for the ideal and the investigated TBRGs when adopting the inter-tip time algorithm and
the tip counting method with respect to the ideal gauge for both the SL3 (a) ad LGO (b) instruments.
12
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
Table 1. Type of rain gauges employed for this comparison and their principal characteristics.
Maximum
Rain gauge Measuring principles Resolution rainfall intensity* Funnel diameter
(mm) (mm h−1) (mm)
Ogawa OSK PC1122 Drop counter 5.21 ·10−3200 127.0
Logotronic (LGO) Tipping-bucket 0.1 200 252.3
Shanghai (SL3) 2-layers tipping-bucket 0.1 240 200.0
*Maximum measured rainfall intensity as declared by the manufacturer
13
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.
Table 2. Total rainfall depth (htot), maximum one-minute rainfall rate (Imax ) and duration (d) of selected events recorded by the Ogawa
drop counter during the observation period.
Date htot Imax d
(mm) (mm h−1)
22 May 2013 205 125 11h 27’
25 May 2013 125 102 06h 56’
24 June 2013 142 114 06h 19’
24 July 2013 51 95 02h 54’
25 July 2013 48 96 03h 10’
04 September 2013 66 72 06h 44’
05 September 2013 65 73 04h 53’
23 September 2013 89 64 06h 09’
14
Atmos. Meas. Tech. Discuss., doi:10.5194/amt-2016-257, 2016
Manuscript under review for journal Atmos. Meas. Tech.
This is just a preview and not the published paper.
c
Author(s) 2016. CC-BY 3.0 License.