ArticlePDF Available

In Vivo Arthroscopic Temperatures: A Comparison Between 2 Types of Radiofrequency Ablation Systems in Arthroscopic Anterior Cruciate Ligament Reconstruction—A Randomized Controlled Trial

Authors:
  • Orthopaedic Research Institute of Queensland

Abstract and Figures

Purpose: To compare a plasma ablation device with a standard ablation device in anterior cruciate ligament (ACL) reconstruction to determine which system is superior in terms of intra-articular heat generation and diathermy efficiency. Methods: This was a prospective, randomized controlled trial. The inclusion criteria were adult patients undergoing primary ACL reconstruction. Patients were randomized preoperatively to the standard ablation group or the plasma ablation group. A thermometer was inserted into the inferior suprapatellar pouch, and the temperature, time, and duration of radiofrequency ablation were measured continually. Results: No significant differences were found between the standard ablation system and the plasma ablation system for maximum temperature (29.77°C and 29.34°C, respectively; P = .95), mean temperature (26.16°C and 26.99°C, respectively; P = .44), minimum temperature (22.66°C and 23.94°C, respectively; P = .54), and baseline temperature (26.80°C and 27.93°C, respectively; P = .35). Similarly, no significant differences were found for operative time (82.90 minutes and 80.50 minutes, respectively; P = .72) and mean diathermy activation times (2.6 minutes for both systems; P = .90). The between-system coefficient of variation for the measured parameters ranged from 0.12% to 3.69%. No intra-articular readings above the temperature likely to damage chondrocytes were recorded. The mean irrigation fluid temperature had a significant correlation with the maximum temperature reached during the procedure (Spearman rank correlation, r = 0.87; P < .01). Conclusions: No difference in temperature was observed between the standard ablation and plasma ablation probes during ACL reconstruction. Temperatures did not exceed critical temperatures associated with chondrocyte death. Level of evidence: Level I, randomized controlled trial.
Content may be subject to copyright.
In Vivo Arthroscopic Temperatures: A Comparison
Between 2 Types of Radiofrequency Ablation
Systems in Arthroscopic Anterior Cruciate Ligament
ReconstructiondA Randomized Controlled Trial
Brent Matthews, M.B.B.S., Matthew Wilkinson, M.B.B.S., F.R.A.C.S.,
Peter McEwen, M.B.B.S., F.R.A.C.S., Kaushik Hazratwala, M.B.B.S., F.R.A.C.S.,
Kenji Doma, Ph.D., B.Sp.Exc.Sci.(Hons), Varaguna Manoharan, M.B.B.S.,
Zaid Bahho, M.B.B.S., and Shannon McEwen
Purpose: To compare a plasma ablation device with a standard ablation device in anterior cruciate ligament (ACL)
reconstruction to determine which system is superior in terms of intra-articular heat generation and diathermy efciency.
Methods: This was a prospective, randomized controlled trial. The inclusion criteria were adult patients undergoing
primary ACL reconstruction. Patients were randomized preoperatively to the standard ablation group or the plasma
ablation group. A thermometer was inserted into the inferior suprapatellar pouch, and the temperature, time, and
duration of radiofrequency ablation were measured continually. Results: No signicant differences were found between
the standard ablation system and the plasma ablation system for maximum temperature (29.77C and 29.34C, respec-
tively; P¼.95), mean temperature (26.16C and 26.99C, respectively; P¼.44), minimum temperature (22.66C and
23.94C, respectively; P¼.54), and baseline temperature (26.80C and 27.93C, respectively; P¼.35). Similarly, no
signicant differences were found for operative time (82.90 minutes and 80.50 minutes, respectively; P¼.72) and mean
diathermy activation times (2.6 minutes for both systems; P¼.90). The between-system coefcient of variation for the
measured parameters ranged from 0.12% to 3.69%. No intra-articular readings above the temperature likely to damage
chondrocytes were recorded. The mean irrigation uid temperature had a signicant correlation with the maximum
temperature reached during the procedure (Spearman rank correlation, r¼0.87; P<.01). Conclusions: No difference in
temperature was observed between the standard ablation and plasma ablation probes during ACL reconstruction. Tem-
peratures did not exceed critical temperatures associated with chondrocyte death. Level of Evidence: Level I, ran-
domized controlled trial.
Arthroscopic anterior cruciate ligament (ACL)
reconstruction is a frequently performed surgical
treatment for ACL rupture. During the procedure,
radiofrequency (RF) ablation is used to debride the ACL
remnant and to dene anatomic landmarks. These
landmarks are used to dene the entry point for the
tibia and femoral tunnel positions for subsequent
placement of a hamstring tendon graft.
As a byproduct of RF ablation, heat is produced, with
the potential within the enclosed space of the knee joint
for temperatures to increase to levels at which damage
to sensitive tissues occurs. A temperature of 45C can
From the Orthopaedic Research Institute of Queensland (B.M., M.W., P.M.,
K.H., V.M., Z.B., S.M.); and James Cook University (K.D., S.M.), Townsville,
Australia.
The authors report the following potential conict of interest or source of
funding: B.M., M.W., P.E., K.H., and V.M. receive support from Argo, Device
Technologies (Arthrex). S.M. receives support from Dr. Peter McEwen. This
project received no external funding grants. The project was funded entirely by
the Orthopaedic Research Institute of Queensland (ORIQL). ORIQL receives
annual funding from Argo and Device Technologies for its Arthroplasty
Fellowship Program. These fellowships support the research and contribute to
salary for the Research Assistant and Research Fellows.
Received December 8, 2015; accepted May 17, 2016.
Address correspondence to Brent Matthews, M.B.B.S., Orthopaedic
Research Institute of Queensland, 7 Turner St, Pimlico, Townsville 4817,
Australia. E-mail: drbrentmatthews@gmail.com
Ó2016 The Authors. Published by Elsevier Inc. on behalf of the Arthros-
copy Association of North America. This is an open access article under the CC
BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
0749-8063/151133
http://dx.doi.org/10.1016/j.arthro.2016.05.026
Arthroscopy: The Journal of Arthroscopic and Related Surgery, Vol -,No-(Month), 2016: pp 1-8 1
damage muscle and nerve cells, whereas a temperature
of approximately 55C has been shown to cause
destruction of chondrocytes in vitro.
1
Both extra-articular RF ablation and intra-articular
RF ablation are generally considered safe; however,
case reports linking incidences of glenohumeral chon-
drolysis to RF ablation exist.
2
RF devices are developed
and tested in simulated clinical environments, often
using animal or cadaveric tissues. Although these mo-
dalities are reasonable surrogates, the characteristics of
these environments differ from the environment of
true, human intraoperative use. Recently, RF ablation
instruments have incorporated the use of a plasma
bubble to ablate tissue with the aim of more efcient
diathermy and reduction of temperature transmission
to surrounding tissue. We conducted a literature review
using PubMed, Ovid, and Medline. No studies involving
in vivo monitoring of joint uid temperature during
ACL reconstruction were identied. Furthermore, no
robust data comparing intra-articular temperature dif-
ferences between standard ablation and plasma abla-
tion could be found.
The purpose of this study was to compare a plasma
ablation device with a standard ablation device in ACL
reconstruction to determine which system is superior in
terms of intra-articular heat generation and diathermy
efciency. We hypothesized that the plasma ablation
group would operate at a lower temperature and more
efciently, resulting in lower intra-articular tempera-
tures and reduced diathermy times intraoperatively.
We further hypothesized that by adhering to safe
diathermy practices as dened by Zoric et al.,
3
intra-
articular temperatures would be maintained below
45C for both systems.
Methods
Ethics committee approval was obtained from the
health region ethics committee (MHS20130806-02)
before commencement of the study, and the trial
was registered with the country of origins clinical
trials registry (Australian Clinical Trials Registry,
ACTRN12613000569707).
Trial Design
This trial was conducted in a prospective, randomized
manner, with 47 participants over the age of 18 years
and of both male and female gender enrolled.
Recruiting occurred between March 2014 and June
2015. The inclusion criteria were adult patients un-
dergoing primary arthroscopic ACL reconstruction by a
single bundle secured with double adjustable suspen-
sory loop xation to an anatomically referenced ACL
footprint. Revision ACL reconstructions and/or ACL
reconstructions using boneepatellar tendonebone graft
were not eligible for inclusion. The exclusion criteria
were patients who were unable to provide informed
consent or unable to attend a 6-month follow-up visit.
All surgical procedures were performed by 1 of 3 or-
thopaedic consultants (P.M., K.H., M.W.), in private
practice, within a large regional hospital. All of the
surgeons are current members of the Australian Or-
thopaedic Association and hold an F.R.A.C.S.(Ortho).
They have 16, 9, and 6 yearsexperience since attaining
fellowship. The 2 most experienced surgeons (P.M.,
K.H.) specialize in lower limb surgery, with the most
senior (P.M.) specializing in surgery of the knee only.
At the time of recruitment, patients were randomly
assigned to either the standard ablation group, treated
with the Stryker SERFAS probe (Kalamazoo, MI), or
the plasma ablation group, treated with the ArthroCare
RF Wand Super TurboVac 90 IFS system (Austin, TX),
with a computer program used to generate a random
number list. The device to be used was concealed from
the surgeon until the time of surgery, when it was
revealed by the nursing staff under the direction of the
research assistant.
Intraoperatively, 1 or 2 medial portals and 1 lateral
arthroscopic portal were placed, and a Luxtron 812
thermometer (LumaSense Technologies, Santa Clara,
CA) connected to a LumaSense Technologies beroptic
temperature probe (STB Probe [part No. SQ10939L])
was inserted into the inferior aspect of the suprapatellar
pouch, percutaneously, independent of the ports. The
probe was positioned so that it was as close to the
working area of the RF ablation probe as possible without
interfering with the arthroscopic instrumentation. The
thermometer was connected to a computer running
Luxtron TrueTemp (version 2.0.0; LumaSense Tech-
nologies), and the temperature was recorded in real time
every 0.5 seconds. Subsequent analysis was performed
with SPSS software (version 20; IBM, Armonk, NY).
The surgical technique consisted of ACL reconstruc-
tion with a single-bundle semitendinosus graft in a
double adjustable suspensory conguration placed in
the anatomic ACL footprint. Normal saline solution
stored at room temperature was used for irrigation of
the joint, with the operating room temperature set at
21C. The uid was delivered either by a gravity-fed
system that used two 3-L bags elevated to 1.85 m and
a hand pump or by a pressure-regulated electric pump.
The pressure of the gravity-fed irrigation system was
calculated by connecting the system to the arterial
blood gas manometer and was zeroed at the level of the
patients knee. This process was repeated 3 times and
returned an average reading of 80 mm Hg with both
bags full and open to the circuit. This gure has been
adopted in statistical calculations using this variable.
The pump pressure setting varied according to the
surgeons preference and ranged from 35 to 90 mm Hg.
An independent suprapatellar outow portal was
created and connected to free drainage in cases in
which it was the surgeons preference to do so.
2B. MATTHEWS ET AL.
The RF ablation probes were used intermittently.
Suction was attached to the RF ablation wand when
requested by the surgeon. When suction was not
requested, free drainage was used.
We set a temperature of 45C as the maximum
acceptable temperature for the joint. The worthwhile
differences in the maximum temperature were deter-
mined according to a nomogram for the estimation of
the measurement repeatability error from the coef-
cient of variation (CV).
4
The worthwhile differences for
the current sample size (20 per group) and a hypo-
thetical sample size of 10 per group were determined
using linear regression equations (y¼0.7955xþ0.1284
and y¼1.1096x0.1132, respectively).
5
On the basis
of these calculations, the worthwhile differences of the
current sample size of 20 per group and a hypothetical
sample size of 10 per group were 0.93% and 1.32%,
respectively.
The plasma ablation device has an inbuilt thermom-
eter that was set to alarm at 45C. On occasion, the
inbuilt alarm would be triggered; however, there was
no change to the intra-articular temperature as
measured by the independent beroptic probe. When
the RF ablation tool was activated by the surgeon, the
time and duration of activation were measured with a
stopwatch operated by the research assistant. Times
were documented in Microsoft Excel. At the completion
of the case, the probe was removed and inspected for
damage before being sent to the institutions central
sterilizing department for reprocessing.
The primary outcome measure was the intra-articular
temperature recorded every 0.5 seconds for the dura-
tion of the operation that involved RF ablation. The
secondary outcome measures were the duration of
application of RF ablation for each system and the total
operating time for each system.
As reported in the Results section, the total diathermy
time is dened as the total duration of the active RF
ablation within the intra-articular space during the
procedure, calculated by the sum of all individual
diathermy applications. The total temperature time is
dened as the total duration for which the temperature
probe was actively recording temperature during the
procedure. The temperature probe was removed during
the case once no further RF ablation was required to
mitigate the risk of probe damage or entanglement
during the passage of the graft. The total surgery time is
dened as the time from knife to skin until completion
of wound closure. A port is dened as a stab incision to
the skin for the passage of instruments, an accessory
outow portal, or the arthroscope; the probe was
inserted using a cannulated needle and is not classied
as a port. Pressure, as reported in the Results section, is
as measured by a manometer at the end of the irriga-
tion tubing at knee-high level for the manual
pressureefed system or the setting for pressure control
pumps. The baseline mean temperature is dened as
the average temperature for the rst 5 seconds after
probe insertion into the knee. The mean temperature is
dened as the average temperature of all readings
recorded during the procedure, and the maximum
temperature is the highest single reading recorded
during the case.
Statistics
All statistical analyses were conducted with SPSS
software (version 20). An a priori calculation was
conducted with an anticipated effect size of 0.8, an
a
level of .05, and a power of 0.8 (G*Power 3.1.9.2,
Heinrich-Heine-Universität, Düsseldorf, Germany) for
differences between 2 independent measures. The
sample size was based on 2 groups (standard ablation
and plasma ablation). According to these effects, a total
sample size of 40 was required.
On the basis of the Shapiro-Wilk test, the dependent
variables were not normally distributed and thus all
measures were log transformed (log
10
[x]) before ana-
lyses. The measures of central tendency and dispersion
are reported as mean standard deviation. By use of
the log10[x] data, the intersystem CV (% CV ¼Stan-
dard deviation 100/Mean) of the dependent variables
was calculated. Furthermore, differences in the
dependent variables between the Stryker and Arthro-
Care systems were determined using analysis of
covariance with 3 separate surgeons treated as a co-
variate. The correlations between dependent variables
were examined by use of the Pearson correlation co-
efcient for pooled data, as well as the correlations for
the Stryker and ArthroCare systems. The level of sig-
nicance was established at .05.
Results
A total of 47 patients were recruited into the study; 7
of these were excluded. Two participants withdrew
consent to have their temperatures recorded after
randomization but before surgery, and 1 patient was
deemed medically unt for the procedure in the anes-
thetic holding bay and the procedure was cancelled.
The remaining 4 were withdrawn because of equip-
ment issues (laptop hardware or software failure) at the
time of data collection that resulted in incomplete data
collection during the procedure.
Of the remaining 40 patients, 19 were randomized to
the plasma ablation group and 21 to the standard
ablation group. ACL reconstruction was performed in
18 right and 22 left knees. Eight patients had an
accessory suprapatellar portal under free drainage; of
these, 4 were in the standard ablation group and 4 were
in the plasma ablation group. Twenty-three patients
had 2 medial portals and no suction attached to the
diathermy device; of these, 12 were in the standard RF
group and 11 were in the plasma ablation group.
IN VIVO ARTHROSCOPIC TEMPERATURES 3
During the course of the study, 3 temperature probes
required replacement as a result of the beroptic la-
ment being kinked intraoperatively; this occurred at the
time the probe was removed through the skin. The
beroptic probe was noted to be particularly prone to
damage when not withdrawn directly perpendicular to
the skin surface.
The patients irrigation pressure was determined
intraoperatively by the surgeon. The pressure
was 35 mm Hg in 5 patients, 40 mm Hg in 4 patients,
45 mm Hg in 2 patients, 50 mm Hg in 1 patient, 55 mm Hg
in 1 patient, 60 mm Hg in 1 patient, 70 mm Hg in twelve
patients, 75 mm Hg in 1 patient, 80 mm Hg in 11 patients,
and 90 mm Hg in 2 patients.
For comparisons between the standard RF ablation
andplasmaRFablationsystems,nosignicant differ-
ences were found for any of the variables examined
(P>.05, Figs 1-3). The mean values (log
10
[x]) for the
maximum temperature in the standard ablation and
plasma ablation groups were 29.77C and 29.34C,
respectively (CV, 1.01%); the mean temperature was
26.16C and 26.99C, respectively (CV, 0.96%). The
operative times for the standard and plasma ablation
groups were 82.90 minutes and 80.50 minutes,
Fig 1. Log-transformed
maximum temperature,
minimum temperature, and
mean temperature for stan-
dard radiofrequency (RF)
ablationand plasma ablation.
Fig 2. Log-transformed to-
tal diathermy time, total
temperature time, and total
surgery time for standard
radiofrequency (RF) abla-
tion and plasma ablation.
4B. MATTHEWS ET AL.
respectively (CV, 2.08%). The mean diathermy acti-
vation time was 2.6 minutes for each group (CV,
1.91%).
No interaction effect was found between any of the
dependent variables with different surgeons as a co-
variate (P>.05). The intersystem CVs ranged from
1.2% to 3.6% (Table 1). For the correlation analyses of
the pooled data (i.e., standard RF and plasma RF sys-
tems combined), signicant correlations were found
between a number of dependent variables (P<.05,
Table 2). The minimum temperature had a strong cor-
relation (r¼0.74, P<.001) with the maximum tem-
perature recorded for both systems combined. This
correlation was also seen when the 2 systems were
analyzed independently: r¼0.63 and P¼.005 for
standard ablation and r¼0.86 and P<.001 for plasma
ablation (Tables 3 and 4). The mean temperature for
both systems correlated with the maximum tempera-
ture (r¼0.86, P<.001); again, this was also seen when
both systems were analyzed independently: r¼0.85
and P<.001 for standard ablation and r¼0.91 and
P<.001 for plasma ablation.
At no time when the plasma RF probes inbuilt
thermometer alarmed at 45C was a temperature
change detected in the joint. All patients attended
6-month follow-up visits. Although not a dened
outcome measure, there were no presentations
consistent with chondrolysis, dened as complete
cartilage loss in 1 or more compartments, in any of the
patients. Because no patients had clinical symptoms
Fig 3. CONSORT (Consoli-
dated Standards of Report-
ing Trials) ow diagram.
(ACL, anterior cruciate
ligament.)
Table 1. Coefcients of Variation and Percentage Differences of Log-Transformed (log
10
[x]) Means of Dependent Variables
Parameter
Mean (log
10
[x])
CV, % PD, % PValue
*
Standard RF Ablation Plasma Ablation
No. of ports 2.95 (0.47) 2.95 (0.47) 0.12 0 .97
Pressure, mm Hg 65.71 (1.82) 62.37 (1.79) 3.69 1.25 .34
Surgery, min 82.90 (1.92) 80.50 (1.91) 2.08 0.67 .72
Temperature, min 47.9 (1.68) 47.6 (1.68) 0.44 0.16 .74
Diathermy, min 2.60 (0.41) 2.60 (0.41) 1.91 2.81 .90
Mean temperature, C 26.16 (1.42) 26.99 (1.43) 2.22 0.96 .44
Minimum temperature, C 22.66 (1.36) 23.94 (1.38) 2.89 1.73 .54
Maximum temperature, C 29.77 (1.47) 29.34 (1.47) 1.01 0.42 .95
Baseline temperature, C 26.80 (1.42) 27.93 (1.44) 2.93 4.1 .35
CV, coefcient of variation; PD, percentage difference; RF, radiofrequency.
*The level of signicance was set at P<.05.
IN VIVO ARTHROSCOPIC TEMPERATURES 5
that warranted further investigation, no further radio-
graphs or magnetic resonance imaging scans were
performed to conrm this.
Discussion
Our study conrmed the safety of the 2 aforemen-
tioned types of RF ablation systems and showed no
difference in operative time or temperature generation
between standard RF ablation and plasma ablation in
ACL reconstruction. Furthermore, we found that the
mean intra-articular temperature correlated closely
with the maximum temperature reached in the joint, a
similar nding to that observed by Barker et al.
6
The use of appropriate-temperature irrigation uid is
a key factor because it closely correlates with mean
temperature and maximum temperature intra-
articularly. Cheng et al.
7
showed in a rat model that
the intra-articular temperature changes according to
the temperature of the irrigation uid and that the
magnitude of this change is related to the initial irri-
gation uid temperature.
At no point in our study did the use of the standard
RF ablation device or the plasma RF ablation device
result in a temperature that exceeded the safe working
limit of 45C. These ndings replicate those of Barker
et al.,
6
who examined intra-articular temperatures
during subacromial decompression. Barker et al.
conrmed that the heat generated by the probe escapes
through its drainage tube; at no time when the
ArthroCare systems temperature probe alarmed at
Table 2. Spearman
r
Correlation Between Selected Parameters for Pooled Data (i.e., Standard and Plasma Ablation Systems
Combined)
Parameter Pressure Surgery Temperature Diathermy
Mean
Temperature
Minimum
Temperature
Maximum
Temperature
Baseline
Temperature
Ports r¼0.08
P¼.61
r¼0.40
P¼.01
r¼0.38
P¼.02
r¼0.01
P¼.95
r¼0.24
P¼.14
r¼0.12
P¼.45
r¼0.14
P¼.40
r¼0.21
P¼.21
Pressure dr¼0.34
P¼.03
r¼0.21
P¼.19
r¼0.02
P¼.90
r¼0.32
P¼.046
r¼0.47
P¼.006
r¼0.25
P¼.12
r¼0.06
P¼.72
Surgery ddr¼0.75
P<.001
r¼0.009
P¼.96
r¼0.25
P¼.13
r¼0.30
P¼.06
r¼0.16
P¼.33
r¼0.13
P¼.45
Temperature dd dr¼0.13
P¼.43
r¼0.30
P¼.06
r¼0.32
P¼.047
r¼0.28
P¼.08
r¼0.19
P¼.27
Diathermy dd d dr¼0.19
P¼.26
r¼0.05
P¼.76
r¼0.02
P¼.93
r¼0.18
P¼.31
Mean
temperature
dddddr¼0.88
P<.0001
r¼0.86
P<.001
r¼0.04
P¼.83
Minimum
temperature
ddddd dr¼0.74
P<.001
r¼0.03
P¼.92
Maximum
temperature
ddddd d dr¼0.02
P¼.92
NOTE. Pressure was measured in millimeters of mercury; surgery, temperature, and diathermy were measured in seconds; and mean, mini-
mum, maximum, and baseline temperatures were measured in degrees Celsius.
Table 3. Spearman
r
Correlation Between Selected Parameters for Standard Radiofrequency Ablation System
Parameter Pressure Surgery Temperature Diathermy
Mean
Temperature
Minimum
Temperature
Maximum
Temperature
Baseline
Temperature
Ports r¼0.313
P¼.17
r¼0.19
P¼.42
r¼0.23
P¼.33
r¼0.30
P¼.22
r¼0.18
P¼.43
r¼0.01
P¼.98
r¼0.09
P¼.70
r¼0.20
P¼.39
Pressure dr¼0.23
P¼.31
r¼0.09
P¼.70
r¼0.01
P¼.96
r¼0.05
P¼.84
r¼0.23
P¼.31
r¼0.01
P¼.96
r¼0.14
P¼.55
Surgery ddr¼0.71
P<.001
r¼0.04
P¼.87
r¼0.24
P¼.29
r¼0.37
P¼.10
r¼0.22
P¼.34
r¼0.04
P¼.22
Temperature dd dr¼0.12
P¼.63
r¼0.47
P¼.03
r¼0.48
P¼.03
r¼0.39
P¼.08
r¼0.21
P¼.35
Diathermy dd d dr¼0.39
P¼.10
r¼0.29
P¼.25
r¼0.05
P¼.84
r¼0.10
P¼.71
Mean temperature dd d d dr¼0.85
P<.001
r¼0.82
P<.001
r¼0.15
P¼.53
Minimum temperature dd d d d dr¼0.63
P¼.005
r¼0.08
P¼.75
Maximum temperature dddddddr¼0.13
P¼.57
NOTE. Pressure was measured in millimeters of mercury; surgery, temperature, and diathermy were measured in seconds; and mean, mini-
mum, maximum, and baseline temperatures were measured in degrees Celsius.
6B. MATTHEWS ET AL.
45C did we detect a change in the temperature of the
joint uid being measured independently.
A cadaveric study by Zoric et al.
3
identied 3 primary
factors that inuence the quantity and effect of heat
generated during RF ablation in arthroscopic shoulder
surgery. These are dened as follows: (1) the ow rate
of the irrigation uid, (2) the duration of application of
RF ablation, and (3) the distance between the probe tip
and target tissue.
3
In their discussion Zoric et al. rec-
ommended that suction be applied to the RF ablation
device, that adequate irrigation uid ow be main-
tained, that RF ablation be activated only intermit-
tently, and that the arthroscope be focused on the RF
probe tip at the time of application.
Flow is proportional to pressure and can be calculated
through the Hagen-Poiseuille relation. Variable pres-
sure was used in our study, ranging from a minimum of
35 mm Hg up to a maximum of 90 mm Hg. No signif-
icant correlation was found between inow pressure
and intra-articular temperature; we assume that this is
because inow pressure of 35 mm Hg is above the
threshold required to prevent heat build-up when
ablation is used.
Good et al.
2
were able to show temperatures of
greater than 45C in all arms of their cadaveric trial, in
which temperatures were recorded by a probe posi-
tioned 1 cm from the active diathermy area. High
temperatures in vicinity to the probe were, however,
not replicated in regions of the glenoid distant to the
site of diathermy activation. By design, RF ablation,
whether plasma based or noneplasma based, is the
controlled application of heat; it is therefore reasonable
to expect a transient temperature increase in proximity
to the target eld. The theoretical advantage of plasma
ablation technology is that surrounding tissue is pro-
tected from heat, which is contained by the plasma
eld. The amount of heat generated by RF ablation is a
function of the power output of the device, the design
of the electrode, and the duration of application; this
has been extensively examined in the eld of tumor
ablation.
8
Nonetheless, thermal injury and, by association, the
use of RF frequency or intra-articular electrocautery
devices have the potential to cause intra-articular
temperatures that may lead to chondrolysis. A strong
association of chondrolysis of the knee with RF abla-
tion has not been established in the literature; it has
only been implicated in knee chondrolysis in 3 of the
24 international studies examined by Provencher
et al.
9
Limitations
Our study examined only 2 different manufacturers
RF ablation systems. Given that we examined the RF
probes only in the knee joint, which is a larger joint
with higher uid capacity than other joints, our nd-
ings cannot be extrapolated to other joints. We concede
that we did not control the irrigation pressure, appli-
cation of suction, and number of arthroscopic ports
between the 2 groups. The potential bias from these
variables is minimized by randomization, and the study
design reects the variability seen in orthopaedic
practice. In addition, the study was not designed or
powered to determine the effects of the different vari-
ables (use of suction, differences in inow temperature,
or number of portals) on intra-articular temperature.
We did not measure the irrigation uid temperature
before its infusion into the knee.
Conclusions
No difference in temperature was observed between
the standard ablation and plasma ablation probes
Table 4. Spearman
r
Correlation Between Selected Parameters for Plasma Radiofrequency Ablation System
Parameter Pressure Surgery Temperature Diathermy
Mean
Temperature
Minimum
Temperature
Maximum
Temperature
Baseline
Temperature
Ports r¼0.13
P¼.59
r¼0.71
P<.001
r¼0.58
P¼.01
r¼0.29
P¼.22
r¼0.29
P¼.21
r¼0.23
P¼.34
r¼0.18
P¼.46
r¼0.23
P¼.41
Pressure dr¼0.49
P¼.04
r¼0.30
P¼.22
r¼0.01
P¼.96
r¼0.56
P¼.01
r¼0.63
P¼.008
r¼0.50
P¼.03
r¼0.10
P¼.72
Surgery ddr¼0.76
P<.001
r¼0.04
P¼.89
r¼0.21
P¼.39
r¼0.18
P¼.45
r¼0.13
P¼.59
r¼0.32
P¼.25
Temperature dd d r¼0.14
P¼.57
r¼0.13
P¼.61
r¼0.19
P¼.44
r¼0.13
P¼.61
r¼0.61
P¼.02
Diathermy dd d dr¼0.05
P¼.83
r¼0.12
P¼.64
r¼0.11
P¼.64
r¼0.27
P¼.34
Mean temperature dd d d d r¼0.91
P<.001
r¼0.95
P<.01
r¼0.14
P¼.61
Minimum temperature dd d d d d r¼0.86
P<.001
r¼0.04
P¼.90
Maximum temperature dddddddr¼0.28
P¼.30
NOTE. Pressure was measured in millimeters of mercury; surgery, temperature, and diathermy were measured in seconds; and mean, mini-
mum, maximum, and baseline temperatures were measured in degrees Celsius.
IN VIVO ARTHROSCOPIC TEMPERATURES 7
during ACL reconstruction. Temperatures did not
exceed critical temperatures associated with chon-
drocyte death.
Acknowledgment
The authors thank A. Grant and L. Murphy for
manuscript proofreading assistance.
References
1. Kaplan LD, Ionescu D, Ernsthausen JM, Bradley JP, Fu FH,
Farkas DL. Temperature requirements for altering the
morphology of osteoarthritic and nonarthritic articular
cartilage: In vitro thermal alteration of articular cartilage.
Am J Sports Med 2004;32:688-692.
2. Good CR, Shindle MK, Kelly BT, Wanich T,
Warren RF. Glenohumeral chondrolysis after shoulder
arthroscopy with thermal capsulorrhaphy. Arthroscopy
2007;23:1-5.
3. Zoric BB, Horn N, Braun S, Millett PJ. Factors inuen-
cing intra-articular uid temperature proles with
radiofrequency ablation. J Bone Joint Surg Am 2009;91:
2448-2454.
4. Atkinson G, Nevill AM. Method agreement and measure-
ment error in the physiology of exercise. In: Winter EM,
Jones AM, Davison RCR, Bromley PD, Mercer TH, eds.
Sport and exercise physiology testing guidelines: Volume 1dsport
testing. London: Routledge, 2006;41-48.
5. Doma K, Deakin GB, Sealey RM. The reliability of lower
extremity and thoracic kinematics at various running
speeds. Int J Sports Med 2012;33:364-369.
6. Barker SL, Johnstone AJ, Kumar K. In vivo temperature
measurement in the subacromial bursa during arthroscopic
subacromial decompression. J Shoulder Elbow Surg 2012;21:
804-807.
7. Cheng S-C, Jou I-M, Chern T-C, Wang P-H, Chen W-C.
The effect of normal saline irrigation at different tempera-
tures on the surface of articular cartilage: An experimental
study in the rat. Arthroscopy 2004;20:55-61.
8. Ahmed M, Goldberg SN. Basic science research in thermal
ablation. Surg Oncol Clin N Am 2011;20:237-258.
9. Provencher MT, Navaie M, Solomon DJ, Smith JC,
Romeo AA, Cole BJ. Joint chondrolysis. J Bone Joint Surg
Am 2011;93:2033-2044.
8B. MATTHEWS ET AL.
... Currently, some studies have tried to assess the variations in joint temperature during arthroscopic procedures on the shoulder and knee, both in human and animal models [2,7,10,23,24]. However, there is still a lack of evidence determining variations in joint and tissue temperature during radiofrequency ablation treatments in the hip joint. ...
... For these reasons, it is important to consider all the possible variables that may influence intra-articular temperature during arthroscopic procedures. According to Zoric et al., the three primary factors that affect the quantity and effect of heat produced during arthroscopic surgery are: the flow rate of the irrigation fluid, the duration of application of radiofrequency ablation, and the distance between the probe tip and target tissue [23]. The most significant effect came from the flow, and it was crucial to understand that in situations when the flow was restricted or stopped entirely, temperatures might exceed 50 °C after only 5 s of ablation [38]. ...
... Regarding knee joint arthroscopy, no difference were found in temperature generation between standard radiofrequency ablation and plasma ablation [23]. ...
Article
Full-text available
Purpose Radiofrequency ablation is an increasingly used surgical option for ablation, resection and coagulation of soft tissues in joint arthroscopy. One of the major issues of thermal ablation is the temperature monitoring across the target areas, as cellular mortality is a direct consequence of thermal dosimetry. Temperatures from 45 °C to 50 °C are at risk of damage to chondrocytes. One of the most reliable tools for temperature monitoring is represented by fiber optic sensors, as they allow accurate and real-time temperature measurement via a minimally invasive approach. The aim of this study was to determine, by fiber Bragg grating sensors (FBGs), the safety of radiofrequency ablation in tissue heating applied to ex-vivo bovine hip joints. Methods Ex vivo bovine hips were subjected to radiofrequency ablation, specifically in the acetabular labrum, for a total of two experiments. The WEREWOLF System (Smith + Nephew, Watford, UK) was employed in high operating mode and in a controlled ablation way. One optical fiber embedding seven FBGs was used to record multipoint temperature variations. Each sensor was 1 mm in length with a distance from edge to edge with each other of 2 mm. Results The maximum variation was recorded in both the tests by the FBG1 (i.e., the closest one to the electrode tip) and was lower than to 2.8 °C. The other sensors (from FBG2 to FBG7) did not record a significant temperature change throughout the duration of the experiment (maximum up to 0.7 °C for FBG7). Conclusions No significant increase in temperature was observed at any of the seven sites. The sensor nearest to the radiofrequency source exhibited the highest temperature rise, but the variation was only 3 °C. The minimal temperature increase registered at the measurement sites, according to existing literature, is not expected to be cytotoxic. FBGs demonstrate the potential to fulfil the strict requirements for temperature measurements during arthroscopic surgery.
... A previous in vivo study found no difference between the 2 ablation techniques in the knee joint during arthroscopic anterior cruciate ligament reconstruction (ACLR). 11 The shoulder, however, has a smaller articular volume than the knee, and the effect of RF in the context of RCR may not be the same as in ACLR. ...
... Interval data-type definitions previously published by Matthews et al 11 were used and are as follows. The total diathermy time was defined as the total duration of active RF ablation within the intra-articular space during the procedure, calculated by the sum of all individual diathermy applications. ...
... This is consistent with previously published comparisons in arthroscopic surgery. 6,11 Of concern, we detected temperatures greater than 45 C; this had not previously occurred in a similar study on ACLR. These high temperatures were transient, lasting less than a few seconds. ...
Article
Full-text available
Background Radiofrequency ablation is commonly used in arthroscopic rotator cuff repair (RCR). New technology devices incorporating a plasma bubble may generate lower intra-articular temperatures and be more efficient. Purpose To compare a plasma ablation device with a standard ablation device in arthroscopic RCR to determine which system is superior in terms of intra-articular heat generation and diathermy efficiency. Study Design Randomized controlled trial; Level of evidence, 1. Methods This was a single-center randomized controlled trial. The inclusion criteria were adult patients undergoing primary RCR. Patients were randomized preoperatively to the standard ablation group (n = 20) or plasma ablation group (n = 20). A thermometer was inserted into the shoulder joint during surgery, and the temperature, surgery, and diathermy times of radiofrequency ablation were measured continually. Results No significant differences were found between the standard ablation group and plasma ablation group for maximum temperature (38.20°C and 39.38°C, respectively; P = .433), mean temperature (31.66°C and 30.64°C, respectively; P = .757), minimum temperature (21.83°C and 23.45°C, respectively; P = .584), and baseline temperature (28.49°C and 29.94°C, respectively; P = .379). Similarly, no significant differences were found for surgery time (74 and 75 minutes, respectively; P = .866) and diathermy time (10 minutes for both; P = .678). Seven patients registered transient high temperatures greater than 45°C. Conclusion There was no difference between plasma ablation and standard ablation in terms of intra-articular temperature in the joint and diathermy efficiency. Transient high intra-articular temperatures occurred in both groups. Registration ACTRN1261300056970 (Australian New Zealand Clinical Trials Registry).
... The low-temperature plasma radio frequency generator produces a lower temperature (40-70 °C) than that of conventional radio frequency ablation or laser equipment; therefore, the thermal injury to surrounding tissues is minimized, the generated water mist is more likely to dissipate, and the impact on the surgical field is also small. At present, plasma generator has been widely used in surgeries that need to strictly control the area of thermal injury in otolaryngology, urology, and joint surgery [14][15][16][17][18]. ...
Article
Full-text available
Background To evaluate the efficacy and safety of novel plasma radio frequency generator and its single-use polypectomy snares for endoscopic mucosal resection (EMR) of gastrointestinal (GI) polyps. Methods A total of 217 patients with 413 GI polyps were recruited from four centers in China. Patients were assigned to experimental or control groups using a central randomization method. The experimental group used the novel plasma radio frequency generator and its matched single-use polypectomy snares (Neowing, Shanghai), while the control group used the high-frequency electrosurgical unit (Erbe, Germany) and disposable electrosurgical snares (Olympus, Japan). The primary endpoint was the en bloc resection rate, and the non-inferiority margin was set at 10%. Secondary endpoint included operation time, coagulation success rate, intraoperative and postoperative bleeding rate, and perforation rate. Results The en bloc resection rate was 97.20% (104/107) in the experimental group and 95.45% (105/110) in the control group (P = 0.496). The operation time was 29.14 ± 20.21 min in the experimental group and 30.26 ± 18.74 min in the control group (P = 0.671). The average removal time of a single polyp in the experimental group was 7.52 ± 4.45 min, which was slightly shorter than that in the control group 8.90 ± 6.67 min, with no statistical difference (P = 0.076). The intraoperative bleeding rates of the experimental group and control group were 8.41% (9/107) and 10.00% (11/110), respectively (P = 0.686). No intraoperative perforation occurred in either group. The postoperative bleeding rates of the experimental group and the control group were 1.87% (2/107) and 4.55% (5/110), respectively (P = 0.465). No postoperative perforation occurred in the experimental group (0/107), while one case of delayed perforation occurred in the control group (1/110, 0.91%). There was no statistical difference between the two groups. Conclusions Endoscopic mucosal resection of GI polyps with the novel plasma radio frequency generator is safe and effective, and non-inferior to the conventional high-frequency electrosurgical system. Graphical abstract
... 10 However, LTPRA inevitably damages the healthy tissue while removing the diseased tissue, and it is difficult for LTPRA to realize the cooperation of visualization system and water supply system in a narrow joint space, which is still not a method to solve RA joint lesions. 11 Recent studies have provided important clues concerning jointtargeted therapy of RA. RA-FLS are the core effector cells of knee joint synovium. ...
Article
Full-text available
Rheumatoid arthritis (RA) has plagued physicians and patients for years due to the lack of targeted treatment. In this study, inspired by the commonality between rheumatoid arthritis fibroblast‐like synoviocytes (RA‐FLS) and cancer cells, the therapeutic effects of cold air plasma (CAP) on RA are studied systematically and thoroughly. In/ex vivo results show that CAP with the proper dosage significantly relieves symptoms including synovial hyperplasia, inflammatory infiltration, and angiogenesis and eliminates the root cause by triggering the self‐antioxidant capability of the surrounding tissue. The mechanism on the molecular and cellular level is also revealed that the spontaneous reactive oxygen species (ROS) cascade induces the mitochondrial apoptosis pathway on RA‐FLS. This study reveals a new strategy for targeted treatment of RA and the mechanistic study provides the theoretical foundation for future development of plasma medicine.
... The inclusion criteria for all of AG included undergone ACLR at least 12 months prior and obtained a score of 80 or higher on the knee injury and osteoarthritis outcome score (KOOS), indicating complete medical recovery [41,42]. Required sample size was estimated base on the previous kinematic study conducted on ACLR patients [43,44] with the minimum statistical power being 80% (effect size = 0.80). G-Power software (version 3.1.9.2, Kiel University, Germany) was used in the power and sample size calculation. ...
Article
Full-text available
Patients experience various biomechanical changes following reconstruction for anterior cruciate ligament (ACL) injury. However, previous studies have focused on lower extremity joints as a single joint rather than simultaneous lower extremity movements. Therefore, this study aimed to determine the movement changes in the lower limb coordination patterns according to movement type following ACL reconstruction. Twenty-one post ACL reconstruction patients (AG) and an equal number of healthy adults (CG) participated in this study. They were asked to perform walking, running, and cutting maneuvers. The continuous relative phase and variability were calculated to examine the coordination pattern. During running and cutting at 30 and 60°, the AG demonstrated a lower in-phase hip–knee coordination pattern in the sagittal plane. The AG demonstrated low hip–knee variability in the sagittal plane during cutting at 60°. The low in-phase coordination pattern can burden the knee by generating unnatural movements following muscle contraction in the opposite direction. Based on the results, it would be useful to identify the problem and provide the fundamental evidence for the optimal timing of return-to-sport after ACL reconstruction (ACLR) rehabilitation, if the coordination variable is measured with various sensors promptly in the sports field to evaluate the coordination of human movement.
... We have reported that low-temperature plasma radiofrequency ablation (LTPRA) obtained better clinical outcomes than RFT in treatment of thoracic neuropathic pain [11], phantom limb pain [12] and trigeminal neuralgia [13]. The mechanisms of LTPRA are based on molecular dissociation rather than heat damage as in RFT [10,14]. Nevertheless, the efficacy and safety of SPG LTPRA for CH are still unclear. ...
Article
Full-text available
Introduction: Although the sphenopalatine ganglion (SPG) has been considered a site of therapeutic potential for cluster headache (CH), the optimal technique of SPG is still to be determined. Low-temperature plasma radiofrequency ablation (LTPRA) has been proposed as an alternative treatment for several neuropathic pain diseases. Aim: To evaluate the effect of LTPRA of SPG in treating chronic and episodic CH. Material and methods: The patients with CH, who achieved temporary pain relief following SPG block, treated using LTPRA between January 2015 and October 2017 were reviewed. Seventy-six patients were included: 50 patients suffered from episodic CH and the remaining 26 patients from chronic CH. The primary outcomes were clinical improvement rate, defined as the percentage of partial and complete pain relief results at 1 day, 12 months, and 24 months of follow-up after the operation. Results: Clinical improvement rates were 92.3%, 92.3% and 73.1% in chronic CH and 73.1%, 84% and 68% in episodic CH at each follow-up time point, respectively. 3 chronic CH patients and 7 episodic CH patients showed no pain relief after the operation. Drooping eyelids were found in 2 cases, one recovered at the 3-month follow-up but another one did not in the 24-month follow-up. No serious complications occurred intraoperatively or postoperatively. Conclusions: LTPRA can be considered an effective and alternative surgical modality in treating patients with chronic and episodic CH, based on SPG block.
... inter-day test-retest reliability (i.e., reliability between the PAP conditions from Weeks 2-4 and the PAP condition in Week 6). The ICC estimate of 0.75, between 0.74 and 0.51 and less than 0.5 were interpreted as good, moderate and poor, respectively [26]. Measurement errors between these reliability time points were also determined by calculating coefficient of variation (CV) with associated 95% CI, with values less than 10% considered acceptable [27]. ...
Article
Full-text available
This study examined the postactivation potentiation effects of combining squat and deadlift exercises on subsequent repeated jump performance. Fifteen, resistance-trained youth wrestlers were randomly allocated to either undertake back squats (BSq), deadlift (DL) or BSq and DL as supersets (BSq+DL), with a repeated jump protocol performed 8-minutes post-exercise in each session. Thereafter, a control condition (CON) was completed involving a general warm-up, followed by the repeated jump protocols. Power outputs, flight time, contact time and reactive strength index were recorded from each repeated jump protocol. Measures were compared between the BSq, DL and BSq+DL sessions and between sessions that generated the best power output (BEST) with CON via inferential statistics and effect size (ES) calculations. The BSq condition exhibited significantly greater power output compared to the CON condition (p
... The heat effect is the mainly therapeutic mechanism of ablation and cutting function. Its instruments incorporated with the use of a plasma bubble to ablate tissue with the aim of more efficient diathermy and reduction of temperature transmission to surrounding tissue [12]. In the authors' opinion, building surgical approaches close to the RF muscle origin would be more propitious to expose the band or muscle and more beneficial to debridement and hemostasis. ...
Article
Full-text available
Background: Isolated rectus femoris (RF) contracture is encountered very rarely in orthopaedic practices. There are few reports on its imaging manifestations and no cases reported to be treated with arthroscopy. Case presentation: A 11-year-old girl with a more than 7 years history of restricted left knee flexion was presented. The clinical assessment and magnetic resonance imaging (MRI) findings were detailed here. A strip-like induration was palpated in the left thigh, which tends to be more obvious with knee flexion. MRI demonstrated a hypointensity band connected the anterior inferior iliac spine with the patella, and marked atrophy of the left RF muscle. Fibrosis contracture band was confirmed with arthroscope, then divided by radiofrequency ablation (RFA) under arthroscopic observation. Followed by debridement of the fibrillar connective tissue and hemostasis around the broken ends. The movement of left knee joint significantly improved after the operation, and the patient recovered nearly full range of motion of this joint after 6 months. Conclusion: The specific MRI findings could assist in confirming clinical early diagnosis of isolated RF contracture. Arthroscopic RFA treatment is an effective technique to treat this disorder with minimally incision.
Article
Background: Plyometric and resistance exercises are known to cause exercise-induced muscle damage (EIMD). Thus, this study examined the impact of EIMD on various field-specific performance measures following a training session that combined plyometric and resistance exercises. Methods: Nine competitive Ultimate Frisbee players undertook a training session consisting of several modes of plyometric and resistance exercises. Indirect markers of muscle damage (i.e., creatine kinase [CK] and delayed onset of muscle soreness [DOMS]) and field-specific performance measures (i.e., run-up vertical jump, standing broad jump, linear sprint and repeated agility) were measured prior to, 24 hours (T24) and 48 hours (T48) post training. Results: The combined plyometric and resistance training session significantly increased muscle damage markers at T24 (CK: 326.5±210.4% and DOMS: 343.3±181.6%) and T48 (CK: 969.2±1262.3% and DOMS: 371.1±179.3%). The jump performance measures were significantly reduced at T24 (run-up vertical jump -5.5±6.3% and standing broad jump -4.7±3.7%) and T48 (run-up vertical jump -4.2±5.1% and standing broad jump -5.0±4.4%). Furthermore, completion times for linear sprint performance was significantly increased at T24 (4.5±3.4%) and T48 (7.2±4.2%), whilst the average completion time for the repeated agility protocol was significantly increased at T24 (1.4±1.4%). Conclusions: Competitive team sport athletes may require at least 48 hours of recovery when implementing field-based conditioning sessions after a training session that combines plyometrics and resistance exercises, particular if unfamiliar with such training modalities.
Article
Nakata, H, Nakanishi, Y, Otsuki, S, Mizuno, M, Connor, J, and Doma, K. Six weeks of hip joint training using a novel multihip joint board improves sprint performance in competitive collegiate male sprinters. J Strength Cond Res XX(X): 000-000, 2021-In a previous study, we identified the possibility that hip joint training using a multihip joint board (MHJB) may increase the cross-sectional area (CSA) of the psoas major (PM) muscle and improve sprint performance. However, the preliminary study reported descriptive findings because of a limited sample size. Therefore, we aimed to investigate and statistically infer the effects of the MHJB training protocol with a larger sample of male collegiate sprinters. The sprinters were randomly assigned to either the MHJB group (n = 7) or the control group (n = 7). The MHJB protocol consisted of 7 separate exercises targeting the development of the hip musculature, all using the MHJB device. The MHJB group undertook the MHJB protocol 3 times per week for 6 weeks, after their normal daily workout, whereas the control group completed their normal daily workout without the MHJB protocol. Sprint performance was recorded from official 100-m competitions before, and after, the 6-week training period. At the end of the training period, significantly greater improvement in PM CSA and 100-m sprint was noted in the MHJB group compared with the control group. These findings confirm the efficacy of MHJB training for increasing PM CSA and improve sprint performance in competitive male sprinters.
Article
Purpose: To determine, with an electron microscope, the effects of irrigation fluid at different temperatures on the surface of articular cartilage. Type of Study: Experimental in vivo study. Methods: With a scanning electron microscope, we investigated the effect of 2 hours of irrigation with normal saline at 3 different temperatures (4degreesC, 25degreesC, and 37degreesC) on articular hyaline cartilage in an in vivo rat knee model study. The contralateral knees were used as controls. We then compared the results. Results: Specimens irrigated with cold (4degreesC) fluid seemed to show the most uneven surface and apparent fibril exposure. Those irrigated with fluid close to body temperature (37degreesC) showed the most even surface without fibril exposure. Conclusions: Although further work is needed to evaluate the long-term effects, using warm irrigation fluid to maintain the intra-articular environment at a more physiologically normal temperature may help reduce damage to cartilage during irrigation.
Article
Whilst various studies have examined lower extremity joint kinematics during running, there is limited investigation on joint kinematics at steady-state running and at intensities close to exhaustion. Subsequently, the purpose of this study was to determine whether the reliability of kinematics in the lower extremity and thorax is affected by varying the running speeds during a running economy test. 14 trained and moderately trained runners undertook 2 running economy tests with each test incorporating 3 intensity stages: 70-, 90- and 110% of the second ventilatory threshold, respectively. The participants ran for 10 min during each of the first 2 stages and to exhaustion during the last stage. Kinematics of the ankle, knee, hip, pelvis and thorax were recorded using a 3-dimensional motion analysis system. Intra-class correlation coefficient (ICC), limits of agreement (LOA) and coefficient of variation (CV) were used to calculate reliability. The ICC, LOA and CV of the lower extremity and thoracic kinematic variables ranged from 0.33-0.97, 1.03-1.39 and 2.0-18.6, respectively. Whilst the reliability did vary between the kinematic variables, the majority of results showed minimal within-subject variation and moderate to high reliability. In conclusion, examining thoracic and lower extremity kinematics is useful in determining whether running kinematics is altered with varying running intensities.
Article
The purpose of the study was to evaluate whether use of a bi-polar radiofrequency (RF) ablation wand would cause excess heating, which may lead to collateral damage to the surrounding tissues during arthroscopic subacromial decompression. Cadaveric studies have shown that high temperatures can potentially be reached when using RF ablation wands in arthroscopic shoulder surgery. Only 1 other published study assesses these temperature rises in the clinical setting. Fifteen patients were recruited to participate in the study. A standard arthroscopic subacromial decompression was performed using continuous flow irrigation, with intermittent use of the RF ablation wand for soft tissue debridement. The temperature of the irrigation fluid within the subacromial bursa and the outflow fluid from the suction port of the wand were measured during the procedure using fiber-optic thermometers. The mean peak temperature recorded in the subacromial bursa was 32.0°C (29.3-43.1°C), with a mean rise from baseline of 9.8°C. The mean peak temperature recorded from the outflow fluid from the wand was 71.6°C (65.6-77.6°C), with a mean rise from baseline of 49.4°C. High temperatures were noted in the outflow fluid from the wand; however, this was not evident in the subacromial bursa itself. Use of room temperature inflow fluid, maintenance of flow through the bursa, and avoidance of prolonged uninterrupted use of the wand all appear to ensure that safe temperatures are maintained in the subacromial bursa not only in the laboratory but also in a clinical setting.
Article
Although the disease was first described in the hip, reports of chondrolysis in nearly all diarthrodial joints have since emerged with considerable variations in the literature.Despite speculation among clinicians and researchers about the implicit causal pathways and etiologic contributors associated with chondrolysis, definitive answers remain elusive.The term chondrolysis has been applied to varied levels of joint cartilage destruction from focal chondral defects to diffuse cartilage loss, revealing a lack of consistency in the application of diagnostic criteria to guide differential disease classification.Differentiating between the various potential etiologies associated with chondrolysis provides opportunities for the prevention of the disease.
Article
Thermal ablation, which induces irreversible cellular injury from focal high-temperature tissue heating that is generated from a focal energy source, has become an accepted treatment option for focal primary and secondary malignancies in a wide range of organs including the liver, lung, kidney, bone, and adrenal glands. Given the rising complexity of treatment types and paradigms in oncology, and the wider application of thermal ablation techniques and adjuvant therapy reviewed in this article, a thorough understanding of the basic principles and recent advances in thermal ablation is a necessary prerequisite for their effective clinical use.
Article
Radiofrequency ablation devices are being used increasingly in arthroscopic surgery. However, there are concerns that excessive temperatures may damage the articular cartilage. The purpose of this study was to investigate the temperature profiles that occur within the glenohumeral space with the use of one commercially available radiofrequency ablation probe. Ten fresh-frozen human cadaver shoulder specimens were used. Intra-articular temperatures were measured at different time intervals over a two-minute period at a distance of 1, 3, 5, and 10 mm away from the probe. The radiofrequency probe was activated throughout the range of machine power settings, and irrigation fluid flow was varied (no flow, a flow at 60 mm Hg without suction, and a flow at 60 mm Hg with suction). Temperatures deleterious to articular cartilage chondrocytes (i.e., those in excess of 50 degrees C) were seen with an increased duration of application, a decreased distance between the thermometer and the probe, and a decreased irrigation fluid flow rate. The highest recorded irrigation fluid temperature reached >80 degrees C after two minutes in a no-flow setting. The flow rate was found to be the most significant predictor of intra-articular temperature profiles. The various machine power settings had no apparent influence on temperature, meaning that higher probe settings are not necessarily associated with higher temperature profiles. These results demonstrate the importance of the management of the irrigation fluid flow rate across the joint during arthroscopic procedures that involve radiofrequency ablation. Even short intervals of limited flow could lead to supraphysiological temperature profiles and potentially to cartilage damage.
Article
To determine, with an electron microscope, the effects of irrigation fluid at different temperatures on the surface of articular cartilage. TYPE OD STUDY: Experimental in vivo study. With a scanning electron microscope, we investigated the effect of 2 hours of irrigation with normal saline at 3 different temperatures (4 degrees C, 25 degrees C, and 37 degrees C) on articular hyaline cartilage in an in vivo rat knee model study. The contralateral knees were used as controls. We then compared the results. Specimens irrigated with cold (4 degrees C) fluid seemed to show the most uneven surface and apparent fibril exposure. Those irrigated with fluid close to body temperature (37 degrees C) showed the most even surface without fibril exposure. Although further work is needed to evaluate the long-term effects, using warm irrigation fluid to maintain the intra-articular environment at a more physiologically normal temperature may help reduce damage to cartilage during irrigation.
Article
Radiofrequency and laser thermal chondroplasty procedures are performed to debride and smooth fibrillated, articular cartilage. Temperature requirements necessary to achieve morphological change will be lower in fibrillated arthritic cartilage as compared with nonarthritic articular cartilage. Controlled laboratory study. A thermal cell-culture chamber was mounted on a stereoscopic microscope and coordinated with a custom temperature-control program. Nonarthritic and osteoarthritic articular cartilage specimens were sectioned into full-thickness slices. The articular sections were exposed to temperatures incrementally from 37 masculine C to 75 masculine C. Real-time, digital capture microscopy was used to visualize and analyze the morphological changes undergone by the articular cartilage specimens. Arthritic articular cartilage displayed morphological change at 56.5 +/- 1.7 masculine C. Loss of fibrillation was the initial morphological change visualized. Continued thermal exposure caused a shrinkage effect of the entire tissue section that was similar to the change seen in nonarthritic sections. Nonarthritic cartilage displayed morphological change at 60.9 +/- 1.9 masculine C. Consistent characteristic morphological changes were found at distinct temperatures in osteoarthritic and nonarthritic articular cartilage. This information begins to establish the thermal parameters required for morphological change of osteoarthritic articular cartilage.
Article
Glenohumeral chondrolysis is a rare but devastating complication that can occur after shoulder arthroscopy and thermal capsulorrhaphy. We retrospectively reviewed the medical records and imaging studies of 8 patients in whom glenohumeral chondrolysis developed after shoulder arthroscopy in which thermal energy was used. Of the 8 patients, 5 had previous thermal capsulorrhaphy for the diagnosis of instability. Two patients were diagnosed with instability with associated labral tears and underwent labral repair with thermal capsular shrinkage. One patient was diagnosed with a labral tear and underwent labral debridement with an extensive glenohumeral synovectomy via a thermal probe. No patients had evidence of chondral damage at their index arthroscopy, and none received postoperative pain pumps. In all patients, radiographic evidence of chondrolysis developed and repeat arthroscopy was performed to confirm the diagnosis. Open surgical stabilization has not been known to have this complication, and it is speculated that heating of the joint fluid at the time of arthroscopy from any source plays a role in cartilage death. Further studies are warranted to determine whether adequate outflow during shoulder arthroscopy where the fluid volume is relatively small will aid in avoiding complications associated with the use of heat sources.
Method agreement and measurement error in the physiology of exercise.
  • Atkinson G.
  • Nevill A.M.
Atkinson G, Nevill AM. Method agreement and measurement error in the physiology of exercise. In: Winter EM, Jones AM, Davison RCR, Bromley PD, Mercer TH, eds. Sport and exercise physiology testing guidelines: Volume 1dsport testing. London: Routledge, 2006;41-48.