Available via license: CC BY 4.0
Content may be subject to copyright.
177
2006; Quinney et al., 2008; Rahimi, 2006; Sanchez,
Sanz, & Zabala, 2007) in medicine, it helps us assess
the effect of the applied treatment (such as diet). Very
often monitored parameter is therefore body fat (BF).
There are many methods applied to the evaluation
of body composition that can be classified as reference
and non-reference. According to Heyward and Wagner
(2004) the reference methods (“gold standard”) are
underwater weighting, air-displacement plethysmogra-
phy and dual-energy X-ray absorptiometry (DEXA).
These methods are used for the evaluation of the valid-
ity of other (non-reference) methods. Since the refer-
ence methods are demanding for equipment as well
as implementation of measurement, they are mainly
used in medicine. In practice, non-reference methods
are used most frequently: these are field methods
(standardized anthropometry, methods based on bio-
electrical impedance analysis) that allow examining
larger sample groups in the field, are less demanding
for instrumentation and also are affordable. There have
already been many studies that compare the final values
Introduction
Nowadays, the evaluation of body composition is com-
monly used for the assessment of the medical condi-
tion of an individual, the level of nutrition and physical
fitness (McArdle, Katch, & Katch, 2007). In medicine,
it is used as a part of diagnostics in diabetics, nephrol-
ogy, obesity science and osteology (Parikh et al., 2004;
Pluijm et al., 2001; Pravn et al., 1999). In sports, it
facilitates efficient management of the training pro-
cess. On the basis of the body composition evaluation,
we can to some extent determine the level of readiness
of the athlete’s organism for strain. By monitoring
changes in the body composition, we can also evaluate
the effect of physical exercise on the athlete’s organism
and asses its adequacy (Bauer, Pivarnik, Fornetti, Jallo,
& Nassar, 2005; Green, Pivarnik, Carrier, & Womack,
* Address for correspondence: Petr Kutáč, Human Motion Di-
agnostics Center, University of Ostrava, Varenska 40a, 702 00
Ostrava 1, Czech Republic. E-mail: petr.kutac@osu.cz
Comparison of body fat using various bioelectrical impedance analyzers
in university students
Petr Kutáč
1,
*
and Miroslav Kopecký
2
1
Human Motion Diagnostics Center, University of Ostrava, Ostrava, Czech Republic; and
2
Faculty of Health Sciences,
Palacký University Olomouc, Olomouc, Czech Republic
Copyright: © 2015 P. Kutáč and M. Kopecký. This is an open access article licensed under the Creative Commons Attribution License
(http://creativecommons.org/licenses/by/4.0/).
Background: At present, the portfolio of devices using the bioelectrical impedance (BIA) method is continuously
expanding as a result of the wide use of this method in the field as measurements by this method are fast and staff
training is simple and reasonably priced. Nonetheless, the problem is that despite using the same method, bioimped-
ance analyzers can differ in many parameters. They use different electric current frequencies, a different number
of electrodes and the electric current may be conducted through different parts of the body. Objective: The main
objective of the study is to compare and evaluate the differences of values of the analysis of the body fat of university
students measured by BIA analyzers that differ in the applied electric current frequency, number of electrodes and
flow of the electric current through the individual body parts. Methods: The research included 125 participants
(70 male and 55 female). The measurements were taken by the following analyzers: Tanita 418 MA, InBody 720,
InBody R20 and Omron BF 300. Results: The differences in the mean values of the body fat representation between
the used analyzers in men ranged from 0.1 to 3.4% and from 0.0 to 2.4 kg, in women from 0.5 to 6.5% and from 0.4
to 3.8 kg in relation to the used analyzer. Conclusions: In men with regular physical activity, the values measured by
InBody R20 were statistically and practically different. The analyzer measured higher values that other analyzers. In
women, there were statistically and practically significant differences in the values measured by Omron BF 300. This
analyzer measured lower values than other analyzers.
Keywords: adipose tissue, young adult, single frequency analyzer, multi frequency analyzer, Bland-Altman analysis
Acta Gymnica, vol. 45, no. 4, 2015, 177–186
doi: 10.5507/ag.2015.021
178
P. Kutáč and M. Kopecký
of the body composition parameters acquired by vari-
ous methods. There have been comparisons of results
measured by the DEXA method (and other laboratory
methods), anthropometric methods as well as the bio-
electrical impedance method (BIA). The studies state
both the found differences in the final values of the
measured parameters and the validity of the applied
methods to laboratory methods (Beeson et al., 2010;
Dolezal, Lau, Abrazado, Storer, & Cooper, 2013; Gába,
Kapuš, Cuberek, & Botek, 2015; Gupta, Balasekaran,
Victor, Hwa, & Shun, 2011; Leahy, O’Neill, Sohun, &
Jakeman, 2012; Mojtahedi, Valentine, & Evans 2009;
Rutherford, Diemer, & Scott 2011).
At present, however, the portfolio of devices using
the BIA method is continuously expanding as a result
of the wide use of this method in the field as measure-
ments by this method are fast and staff training is sim-
ple and reasonably priced. Nonetheless, the problem
is that despite using the same method, bioimpedance
analyzers can differ in many parameters. They use dif-
ferent electric current frequencies, a different number
of electrodes and the electric current may be conducted
through different parts of the body. Another issue is
the unavailability of the used equations in the analyzer
software and the lack of information about the proband
groups from which reference data were taken for the
calculation of the final values. It is not possible to cal-
culate any potential differences in the final values when
more sophisticated instrumentation or an analyzer by a
different producer is acquired. A similar problem may
occur when the individual is measured in a different
work station. The only way how to encompass such
differences is to take such measurements in practice
and check any potential differences. The knowledge of
any potential differences is essential in case of repeated
measurements with the aim to understand changes in
body composition that could be caused by ontogenetic
changes or external interventions.
The main objective of the study is to compare and
evaluate the differences of values of the analysis of the
body fat of university students measured by bioimped-
ance analyzers that differ in the applied electric current
frequency, number of electrodes and flow of the elec-
tric current through the individual body parts.
Methods
Participants
The research group included 130 individuals in total
(73 males and 57 females). Three male and two female
were removed from the group after outliers. Thus, the
final number of monitored individuals was 125 (70
males and 55 females). The basic characteristic of the
study sample is presented in the Results part in Table 1.
None of the participants had any medical issues; they
did not take any medicine or food supplements. They
participated in the research voluntarily and they were
informed about the process of the research in advance.
Also, they signed informed consent with the participa-
tion in the research. The research was approved by the
Ethical Board of the University of Ostrava and it is in
compliance with the Helsinki Declaration. The males
were university students studying physical education
and sports. Therefore, we can call them a specific pop-
ulation group and the acquired results may be applied
to athletes of a whole range of sports disciplines that
will show similar values of the monitored parameters
(e.g. body fat representation which is a very frequently
monitored parameter in sports). The females were uni-
versity students of fields that did not focus on sports.
Therefore, we can apply their results on the general
population of women without any medical problems.
Procedures
The participants of the measurements were informed
of the conditions they had to observe prior to measure-
ment in advance (no alcohol consumption for 24 hours
prior to measurement, no vigorous exercise less than
12 hours prior to measurement, no food and beverages
3 hours prior to measurement, urination immediately
before measurement; only women that did not have
their menstrual period were measured).
Measurements took place in the morning (7.30
a.m.–9.00 a.m.) on the same day in the week. All
principles of measurement defined in the operating
instructions for the individual analyzers were met. The
participants attended all measurements wearing under-
wear. The measurement was executed standing, always
by the same team of researchers who have several years
of experience with such measurements. The body fat
of each participant was successively measured on all
applied BIA analyzers in the following order: Tanita
418 MA (SFBIA
4
), InBody 720 (MFBIA
4-720
), InBody
R20 (MFBIA
4-R20
), Omron BF 300 (SFBIA
2
). To
exclude any potential influence of the final measured
value due to delays between measurements (e.g. by con-
sumption of food or liquids), the individual measure-
ments were executed in immediate succession and the
participants were under continuous supervision. The
total body water (TBW), which is a primarily measured
parameter when the BIA method is used, was also
measured. TBW values are not stated for the SFBIA
2
analyzer because the output of this analyzer does not
specify the value. The body height, which is an input
parameter for measurements by the used analyzers,
was measured using the A-226 anthropometer (Trys-
tom, Olomouc, Czech Republic). The body weight as
179
Comparison of body fat using various BIA analyzers
Results
The basic characteristics of the monitored group and
the BF value measured by the individual analyzers are
presented in Table 1 and 2.
The differences in the values of the measured BF
and TBW representation by the used analyzers and
the results of their statistical analyses are presented in
Table 3 and 4.
an input parameter for the Omron BF 300 analyzer,
which is not a scale, was taken by the Tanita BC 587
digital scale (Tanita Corporation, Tokyo, Japan).
The used analyzers for the body composition diag-
nostics and their basic characteristics:
• Tanita 418 MA (Tanita Corporation, Tokyo, Japan)
is a tetrapolar single-frequency BIA analyzer that
uses the electric current frequency of 50 kHz for
measurement. Eight point touch electrodes are
used for measurement. The analyzer is also a digital
scale.
• InBody 720 (Biospace, Seoul, Korea) is a tetrapolar
multi-frequency BIA analyzer that uses the gradual
electric current frequency of 1, 5, 50, 250, 500 and
1000 kHz for measurement. Eight point touch elec-
trodes are used for measurement. The analyzer is
also a digital scale.
• InBody R20 (Biospace, Seoul, Korea) is a tetrapo-
lar multi-frequency BIA analyzer that uses the elec-
tric current frequency of 20 and 100 kHz for mea-
surement. Eight point touch electrodes are used for
measurement. The analyzer is also a digital scale.
• Omron BF 300 (Omron Corporation, Tokyo,
Japan) is a bipolar single-frequency BIA analyzer
(hand-hand) that uses the electric current frequency
of 50 kHz for measurement.
Statistical processing
The results were statistically processed using the IBM
SPSS Statistics (Version 21; IBM, Armonk, NY, USA).
Remote observations were identified by box plots and
the normality of distribution was verified by the Shap-
iro-Wilk test. With regard to the normal distribution of
values, we used the paired samples t-test to verify the
statistical significance of the differences of the results
between the individual devices. The statistical signifi-
cance level was determined to be α = .05 for all tests
used.
In values where statistically significant differences
were found, we used the effect size to assess practical
significance (Cohen, 1988). Recommendations for
Cohen’s d: 0.2 = minor change, 0.5 = medium change,
0.8 = major change. The value of Cohen’s d ≥ 0.5 was
considered to be a practically significant difference.
To express the level of correlation between the
results of measurement by the individual analyzers for
the body fat, we used Pearson correlation coefficient
(Westgard,
2008). To evaluate the homogeneity of the
results between two analyzers, we also used the Bland-
Altman’s analysis (Bland & Altman, 2010).
Table 1
Characteristics of the monitored group
Males (n = 70) Females (n = 55)
M SD M SD
Age (years) 20.2 1.1 19.8 1.2
Height (cm) 180.8 5.9 166.6 6.0
Weight (kg)
SFBIA
4
75.1 7.4 59.2 5.9
MFBIA
4-720
75.1 7.7 59.4 5.8
MFBIA
4-R20
75.2 7.4 59.3 5.9
BMI (kg/m
2
) 23.0 1.6 21.4 1.8
Note. BMI=body mass index, SFBIA
4
=Tanita BC 418 MA,
MFBIA
4-720
=InBody 720, MFBIA
4-R20
=InBody R20
Table 2
Values of the body fat and total body water
Males (n = 70) Females (n = 55)
M SD M SD
BF (%)
SFBIA
4
10.6 4.0 24.2 4.1
MFBIA
4-720
10.6 4.0 23.6 5.1
MFBIA
4-R20
13.2 4.0 25.2 5.0
SFBIA
2
9.6 3.3 19.0 3.9
BF (kg)
SFBIA
4
8.0 3.3 14.3 3.4
MFBIA
4-720
8.0 3.4 14.0 3.5
MFBIA
4-R20
9.9 3.4 15.0 3.3
SFBIA
2
7.2 3.0 11.3 3.0
TBW (%)
SFBIA
4
65.4 2.9 55.2 2.9
MFBIA
4-720
65.6 3.0 55.7 3.9
MFBIA
4-R20
64.0 2.9 55.1 3.7
TBW (kg)
SFBIA
4
49.1 4.5 32.7 2.6
MFBIA
4-720
49.3 4.8 33.1 3.7
MFBIA
4-R20
48.1 4.7 32.7 3.7
Note. BF=body fat, TBW=total body water, SFBIA
4
=Tanita
BC 418 MA, MFBIA
4-720
= InBody 720, MFBIA
4-R20
=InBody
R20, SFBIA
2
=Omron BF 300.
180
P. Kutáč and M. Kopecký
There was no significant difference in the mea-
sured results only between the analyzers SFBIA
4
and
MFBIA
4-720
. In other cases, the final significance val-
ues ranged from p < .001 to p < .0001. As for results
with significant differences, medium practical signifi-
cance was found between the results of SFBIA
4
and
MFBIA
4-R20
(BF %, kg and TBW %), MFBIA
4-720
and
MFBIA
4-R20
(BF %, kg and TBW %) and MFBIA
4-R20
and SFBIA
2
(BF %, kg) (d = 0.5–0.7) and high practi-
cal significance only between MFBIA
4-R20
and SFBIA
2
in the values expressed in percentage (d = 0.9). Prac-
tical significance between the analyzers SFBIA
4
and
MFBIA
4-R20
(TBW kg), MFBIA
4-720
vs. MFBIA
4-R20
(TBW kg), SFBIA
4
– SFBIA
2
and MFBIA
4-720
and
SFBIA
2
was not shown in spite of significant differ-
ences (d = 0.1–0.3). The closeness of results between
the individual analyzers expressed in kilograms can
be considered to be high, up to very high (Westgard,
2008). The values of Pearson’s correlation coefficient
r explain 51–90% of variability. In values expressed in
percentage, there is a high closeness of results only
between the values of the analyzers MFBIA
4-720
and
MFBIA
4-R20
and SFBIA
4
and SFBIA
2.
The r values
explain 59–65% of variability. The closeness of results
between other analyzers is considerable (Westgard,
2008). The r values explain 41–46% of variability.
Similarly to the men, there was no significant dif-
ference in the measured values between SFBIA
4
and
MFBIA
4-720
; moreover, the female group also did not
show any significant difference in the values between
SFBIA
4
and MFBIA
4-R20
(except BF %). In other cases,
the significance ranged from p < .05 to p = .0001.
In spite of the significant difference in the results of
SFBIA
4
and MFBIA
4-R20
(BF %), MFBIA
4-720
and
MFBIA
4-R20
(BF %, kg and TBW %, kg), practical sig-
nificance was not shown (d = 0.1–0.3). The practical
significance in other cases was always high (d ≥ 0.8).
The values of the Pearson’s correlation coefficient r
show higher correlations between the results measured
by the used analyzers than in men. The closeness of
Table 3
Differences in measured values – males (n=70)
Diff r d 95% LoA
SFBIA
4
vs. MFBIA
4-720
BF (%) 0.0 .68 – (–6.3, 6.3)
BF (kg) 0.0 .75 – (–4.8, 4.8)
TBW (%) –0.2 .65 – (–5.0, 4.6)
TBW (kg) –0.2 .93 – (–3.6, 3.2)
SFBIA
4
vs. MFBIA
4-R20
BF (%) –2.6*** .67 0.6 (–9.1, 3.9)
BF (kg) –1.9*** .75 0.5 (–6.7, 2.9)
TBW (%) 1.4*** .63 0.5 (–3.5, 6.3)
TBW (kg) 1.0*** .93 0.2 (–2.4, 4.4)
MFBIA
4-720
vs. MFBIA
4-R20
BF (%) –2.6*** .81 0.6 (–7.4, 2.2)
BF (kg) –1.9*** .84 0.5 (–5.7, 1.9)
TBW (%) 1.6*** .79 0.5 (–2.1, 5.3)
TBW (kg) 1.2*** .95 0.3 (–4.1, 1.7)
SFBIA
4
vs. SFBIA
2
BF (%) 1.0** .77 0.2 (–4.0, 6.0)
BF (kg) 0.8** .82 0.1 (–3.1, 4.7)
MFBIA
4-720
vs. SFBIA
2
BF (%) 1.0** .66 0.2 (–5.0, 7.0)
BF (kg) 0.8** .72 0.1 (–3.9, 5.5)
MFBIA
4-R20
vs. SFBIA
2
BF (%) 3.6*** .64 0.9 (–2.6, 9.8)
BF (kg) 2.7*** .72 0.7 (–2.2, 7.4)
Note. Diff=difference, r=Pearson correlation coefficient, d=effect size, 95%LoA=95% limits
of agreement, SFBIA
4
=Tanita BC 418 MA, MFBIA
4-720
=InBody 720, MFBIA
4-R20
=InBody
R20, SFBIA
2
=Omron BF 300. **p<.001, ***p<.0001.
181
Comparison of body fat using various BIA analyzers
results is high, up to very high (Westgard, 2008). The
r values between analyzers SFBIA
4
and MFBIA
4-720
(TBW kg), SFBIA
4
and MFBIA
4-R20
(TBW kg), and
MFBIA
4-720
and MFBIA
4-R20
(BF, TBW % and kg)
explain 81–94% of variability. In other cases, the values
range from 50 to 79%.
The results of the Bland-Altman’s analysis of BF rep-
resentation (a primarily monitored parameter in this
study) are illustrated in the form of Bland-Altman’s
plots (Figures 1 and 2). The plots present the differ-
ences found in BF values expressed in percentage,
measured by two different analyzers. The plots show
that for the male group, the smallest differences in
the mean are between analyzers MFBIA
4-720
and
SFBIA
4
where the mean is almost zero (mean = 0.6).
In women, there were differences between MFBIA
4-720
and SFBIA
4
and MFBIA
4-R20
and SFBIA
4
(mean = –3.5
and 3.6). However, according to the 95% interval of
agreement, there are large differences in the individual
persons among the analyzers, which are manifested
by the wide interval of agreement. To assess the size
of the values measured by the individual analyzers,
we can use the assessment of mean displacement (the
Mean axis) in the plots. The displacement of the mean
axis downwards means that the results measured by the
second analyzer are higher than the results measured
by the first analyzers. The displacement of the mean
axis upwards means that the results measured by the
first analyzer are higher than the results measured by
the second analyzer. As an example of the mean dis-
placement downwards, in the male group we provide
a comparison between MFBA
4-720
and MFBIA
4-R20
(mean = –24.4) where MFBIA
4-R20
measures consider-
ably higher values. As an example of the mean displace-
ment upwards, in the male group we provide a compari-
son between MFBIA
4-R20
and SFBIA
2
(mean = 32.6)
where MFBIA
4-R20
also measured considerably higher
values. The relative differences in the measured values
in the individual participants are mostly concentrated
around the mean relative difference in the values of
Table 4
Differences in measured values – females (n = 55)
Diff r d 95% LoA
SFBIA
4
vs. MFBIA
4-720
BF (%) 0.6 .75 – (–6.4, 7.6)
BF (kg) 0.3 .82 – (–4.6, 4.0)
TBW (%) –0.5 .71 – (–6.0, 5.0)
TBW (kg) –0.4 .94 – (–3.6, 2.8)
SFBIA
4
vs. MFBIA
4-R20
BF (%) –1.0* .71 0.2 (–6.0, 8.0)
BF (kg) –0.7 .81 – (–4.9, 3.5)
TBW (%) 0.1 .72 – (–4.8, 5.0)
TBW (kg) 0.0 .95 – (–3.0, 3.0)
MFBIA
4-720
vs. MFBIA
4-R20
BF (%) –1.6*** .94 0.3 (–5.1, 1.9)
BF (kg) –1.0*** .95 0.2 (–3.1, 1.1)
TBW (%) 0.6** .90 0.2 (–2.8, 4.0)
TBW (kg) 0.4** .97 0.1 (–1.4, 2.2)
SFBIA
4
vs. SFBIA
2
BF (%) 5.2*** .82 1.3 (–9.9, –0.5)
BF (kg) 3.0*** .89 1.0 (0.0, 6.0)
MFBIA
4-720
vs. SFBIA
2
BF (%) 4.6*** .74 1.0 (–2.2, 11.4)
BF (kg) 2.7*** .82 0.8 (–1.3, 6.7)
MFBIA
4-R20
vs. SFBIA
2
BF (%) 6.2*** .74 1.4 (–0.1, 12.5)
BF (kg) 3.7*** .84 1.1 (0.2, 7.2)
Note. Diff=difference, r=Pearson correlation coefficient, d=effect size, 95%LoA=95% limits
of agreement, SFBIA
4
=Tanita BC 418 MA, MFBIA
4-720
=InBody 720, MFBIA
4-R20
=InBody
R20, SFBIA
2
=Omron BF 300. *p<.05, **p<.001, ***p<.0001.
182
P. Kutáč and M. Kopecký
the two analyzers MFBIA
4-720
and MFBIA
4-R20
which
is also reflected in the values of the Pearson’s correla-
tion coefficient r (Tables 3 and 4). It is thus obvious that
these two analyzers provide the most predictable results.
The analyzers have one manufacturer and therefore they
should have the same software for the calculation of BF.
Discussion
The study used BIA analyzers that use different fre-
quencies for the measurement, with electric current
going through different body parts. The objective of
the study was not to evaluate their validity against the
reference method as many studies dealing with this
issue have already been published. For the InBody
analyzers, the correlation with the DEXA reference
method was determined to be at the level of .94–.96;
the study included healthy men and women by the age
of 18 (Karelis, Chamberland, Aubertin-Leheudre, &
Duval, 2013). Even though they used a different ana-
lyzer than we did in this study, we can assume that the
InBody analyzers we used will have similar correlations
as they are made by the same manufacturer and use
the same frequencies, number of electrodes as well as
the method of conducting current through the human
Figure 1. Bland-Altman plots with 95% limits of agreement and correlation analysis of the differ-
ences between the body fat values measured by the used analyzers in percentage – males
183
Comparison of body fat using various BIA analyzers
body for the measurement. As for the single-frequency
analyzer SFBIA
4
, the value of correlation to the DEXA
method found in physical education students was
.82–.84 in relation to the used measuring mode (Kutáč,
Gajda, Přidalová, & Šmajstrla, 2008). As for single fre-
quency bipolar hand-to-hand analyzers, the values of
correlation with the DEXA method for the verification
of validity in sporting young men and women ranged
from .82 to .88 in relation to the used analyzers (Esco,
Olson, Williford, Lizana, & Russell, 2011; Loenneke
et al., 2013; Wang et al., 2013). Even though InBody
seems to be the most accurate analyzer, other analyz-
ers can also be considered sufficiently accurate as the
values of correlation in all the aforementioned studies
exceeded .8 which is a high closeness of results (West-
gard, 2008). Analyzers with such closeness of results
to the reference method may be considered to be suffi-
ciently accurate for the needs of the process of physical
education and diagnostics of athletes.
In the males, the lowest differences between the
mean values of BF representation were found between
the analyzers SFBIA
4
and MFBIA
4-720.
The overall anal-
ysis of the differences in the mean values measured by
the individual BIA analyzers did not show any depen-
dence that would be related to the used BIA analyzer.
The difference in the values measured by the analyzer
Figure 2. Bland-Altman plots with 95% limits of agreement and correlation analysis of the differ-
ences between the body fat values measured by the used analyzers in percentage – females
184
P. Kutáč and M. Kopecký
of the same manufacturer, MFBIA
4-720
and MFBIA
4-R20
was greater than the difference between the single and
multi-frequency analyzers MFBIA
4-720
and SFBIA
4
or
MFBIA
4-720
and SFBIA
2
. On the other hand, the dif-
ference between the analyzers MFBIA
4-R20
and SFBIA
2
was the greatest of all comparisons in the male.
The diagnostic practice also needs to respond to the
question of the impact of the differences in the results
on the interpretation of the measured values. The
monitored men were students of physical education.
The mean values of the BF representation percentage
measured by the BIA method in physical education
students that are presented in some professional stud-
ies do not exceed 13% (Kutáč, 2012; Kutáč, Gajda, &
Přidalová, 2009; Kutáč, Přidalová, & Riegerová, 2008).
Therefore, the detailed analysis of the values measured
by the individual analyzers also focused on how many
participants would correspond with the mean values of
physical education students. The limit value was 13%
of body fat. For the SFBIA
4
analyzer it was found that
the body fat representation of 18 participants (25.71%)
did not correspond with the values of physical edu-
cation students (exceeded 13%); the number of 20
(28.57%) for MFBIA
4-720
, 31 (44.28%) for MFBIA
4-R20
and 13 (18.57%) for SFBIA
2
. The results show that
there would be different evaluation of several partici-
pants in case of interpretation of the acquired results.
There would be a significant difference especially when
using MFBIA
4-R20.
In the females, similarly to males, the lowest dif-
ferences in the mean values were found between the
analyzers SFBIA
4
and MFBIA
4-720
. The greatest dif-
ferences were found between the single frequency
BIA hand-to-hand analyzer (SFBIA
2
) and the other
analyzers. The SFBIA
2
analyzer measures the lowest
values. The found differences were even higher than
in the comparison of the measured values by the BIA
and DEXA methods (Gupta et al., 2011; Mojtahedi et
al., 2009; Kutáč et al., 2008; Trutschnigg et al., 2008).
In these studies, the differences did not exceed 2.7%
BF. To assess the impact of the found differences on
the interpretation of results in the diagnostic practice,
we will use the BF representation at the level of 25%,
which is the value stated for young female (Görner,
Boraczyński, & Štihec, 2009; Nazmi, Irfan, Osman,
& Serdar, 2011; Rutherford, Diemer, & Scott, 2011).
The value of 25% BF representation was exceeded in
25 female (45.45%) measured by SFBIA
4
, in 16 women
(29.09%) measured by MFBIA
4-720
, in 27 (49.09%) mea-
sured by MFBIA
4-R20
and only in 5 (9.09%) measured by
SFBIA
2
. There is a noticeable difference that became
apparent when the analyzer was changed. The greatest
difference would occur with the use of SFBIA
2
.
The differences in the mean values we found that
were measured by the used BIA analyzers are lower
than differences stated in other studies. The differences
in studies that dealt with the comparison of whole-
body analyzers with leg-to-leg analyzers reached the
mean value of 7.4% BF, and the value of 6.2% BF when
compared with hand-to-hand analyzers (Chin, Kiew, &
Girandola, 2006; Trutschnigg et al., 2008).
TBW is the primarily measured parameter in the
BIA method, BF values are calculated additionally.
The TBW value predicates the status of organism
hydration. Professional studies state that when hydra-
tion decreases by 2 to 3%, there is a substantial reduc-
tion of performance in physical activities (García-
Jiménez, Lucas, & García-Pellicer, 2011; Hamouti, Del
Coso, Estevez, & Mora-Rodriguez, 2010; Maughan
& Shirreffs, 2010). A decrease of hydration by 3–5%
causes digestive issues during training and muscle
spasms (Beachle & Earle, 2008; Burke, 2007; Montain,
2008; Oppliger & Bartok, 2002). From this point of
view, we can state that the differences we found in the
mean values measured by the individual analyzers are
negligible. Even though some differences were statisti-
cally significant and the value of Cohen’s d showed the
medium value of the effect of size, none of the differ-
ences exceeded the level of 2% TBW; the differences
ranged from 0 to 1.6% TBW. As for BF, the differences
found were within the difference intentions. However,
a more detailed analysis of the differences in the indi-
vidual participants showed that the difference in the
range of 2–3% TBW was found in 13 (18.6%) males
between SFBIA
4
and MFBIA
4-720
, in 16 (22.3%) males
between SFBIA
4
and MFBIA
4-R20
, and in 11 (15.7%)
males between MFBIA
4-720
and MFBIA
4-R20
. As for
females, there were 11 (20%) participants between
SFBIA
4
and MFBIA
4-720
, 15 (27.3%) between SFBIA
4
and MFBIA
4-R20
, and 6 (10.1%) between MFBIA
4-720
and MFBIA
4-R20
. The difference in the range of 3 to 5%
TBW was found in 10 (14.3%) males between SFBIA
4
and MFBIA
4-720
, 19 (27.1%) males between SFBIA
4
and MFBIA
4-R20
, and 16 (22.9%) males between
MFBIA
4-720
and MFBIA
4-R20
. In females, the differ-
ence was found in 13 (23.6%) participants between
SFBIA
4
and MFBIA
4-720
, 11 (20%) between SFBIA
4
and MFBIA
4-R20
, and 3 (5.5%) between MFBIA
4-720
and
MFBIA
4-R20
. As the detailed analysis implies, the evalu-
ation of the final values of several participants could be
misinterpreted if the analyzers were changed.
Study limitations
We are aware of the fact that the results we obtained
might be influenced by the selected groups. The moni-
tored males are individuals with regular physical activ-
ity which they perform in their field of study. These
185
Comparison of body fat using various BIA analyzers
individuals are also active athletes at the performance
level. Therefore, their results may only apply to the
sporting population.
The validity of the results is also limited by the
used BIA analyzers. Since there is a wide range of BIA
analyzers on the market, the submitted study could be
considered a base for including other BIA analyzers, or
other population groups in the research.
Conclusions
Even though the differences between the mean values
measured by the used analyzers were low in majority
of the cases and ranged at the level of the errors of
measurement, a detailed analysis showed substantially
higher differences in several participants. Replacing an
analyzer with a different one could lead to misinterpre-
tation of the measured values in diagnostics. The differ-
ences found during repeated measurements would not
need to be a result of an external intervention or the
ontogenetic development of the individual; they could
be caused by different measuring of the analyzers.
The results also showed significant (statistically and
practically) differences between analyzers by the same
manufacturer, but a different series. It is thus obvious
that a high correlation of measured values does not
guarantee conformity of results and therefore, it is not
recommended to even use different types of analyzers
by one producer in practice.
Acknowledgment
This study was financed by the project no. SGS 6136/
PdF/2013.
References
Bauer, P. W., Pivarnik, J. M., Fornetti, W. C., Jallo, J. J., &
Nassar, L. (2005). Cross validation of fat free mass predic-
tion models for elite female gymnasts. Pediatric Exercise
Science, 17, 337–344.
Beachle, T. R., & Earle, R. W. (2008). Essentials of strength
training and conditioning. Champaign, IL: Human Kinetics.
Beeson, W. L., Batech, M., Schultz, E., Salto, L., Firek, A.,
Deleon, M., … Cordero-MacIntyre, Z. (2010). Compari-
son of body composition by bioelectrical impedance analy-
sis and dual-energy X-ray absorptiometry in Hispanic dia-
betics. International Journal of Body Composition Research,
8, 45–50.
Bland, J. M., & Altman, D. G. (2010). Statistical methods
for assessing agreement between two methods of clinical
measurement. International Journal of Nursing Studies, 47,
931–936.
Burke, L. (2007). Practical sport nutrition. Champaign, IL:
Human Kinetics.
Chin, M. K., Kiew, O. F., & Girandola, R. N. (2006). A com-
parison of body fat measurement by BodPod, skinfolds,
and three bioelectrical impedance analysis techniques in
Chinese college student. International Journal of Physical
Education, 43(2), 77–85.
Cohen, J. (1988). Statistical power analysis for the behav-
ioral sciences (2nd ed.). Hillsdale, NJ: Lawrence Erlbaum
Associates.
Dolezal, B. A., Lau, M. J., Abrazado, M., Storer, T. W., &
Cooper, C. B. (2013). Validity of two commercial grade
bioelectrical impedance analyzers for measurement of
body fat percentage. Journal of Exercise Physiology Online,
16(4), 74–83.
Esco, M. R., Olson, M. S., Williford, H. N., Lizana, S. N., &
Russell, A. R. (2011). The accuracy of hand-to-hand bio-
electrical impedance analysis in predicting body composi-
tion in college-age female athletes. Journal of Strength and
Conditioning Research, 25, 1040–1044.
Gába, A., Kapuš, O., Cuberek, R., & Botek, M. (2015).
Comparison of multi- and single-frequency bioelectrical
impedance analysis with dual-energy X-ray absorptiometry
for assessment of body composition in post-menopausal
women: Effects of body mass index and accelerometer-
determined physical activity. Journal of Human Nutrition
and Dietetics, 28, 390–400.
García-Jiménez, J. V., Lucas, J. L. Y., & García-Pellicer, J. J.
(2011). Fluid balance and dehydration in futsal players:
Goalkeepers vs. field players. International Journal of Sport
Science, 7(22), 3–13.
Görner, K., Boraczyński, T., & Štihec, J. (2009). Physical
activity, body mass, body composition and the level of aer-
obic capacity among young, adult women and men. Sport
Scientific & Practical Aspects, 6(2), 7–14.
Green, M. R., Pivarnik, J. M., Carrier, D. P., & Womack, C.
J. (2006). Relationship between physiological profiles and
on-ice performance of a national collegiate athletic asso-
ciation division I Hockey Team. Journal of Strength and
Conditioning Research, 20, 43–48.
Gupta, N., Balasekaran, G., Victor, G. V., Hwa, C. Y., &
Shun, L. M. (2011). Comparison of body composition
with bioelectric impedance (BIA) and dual energy X-ray
absorptiometry (DEXA) among Singapore Chinese. Jour-
nal of Science and Medicine in Sport, 14, 33–35.
Hamouti, N., Del Coso, J., Estevez, E., & Mora-Rodriguez,
R. (2010). Dehydration and sodium deficit during indoor
practice in elite European male team players. European
Journal of Sport Science, 10, 329–336.
Heyward, V. H., & Wagner, D. R. (2004). Applied body com-
position assessment. Champaign, IL: Human Kinetics.
Karelis, A. D., Chamberland, G., Aubertin-Leheudre, M.,
& Duval, C. (2013). Validation of a portable bioelectri-
cal impedance analyzer for the assessment of body com-
position. Applied Physiology, Nutrition and Metabolism, 38,
27–32.
Kutáč, P. (2012). Application of typical error of measure-
ment for accuracy of measurement of body composition
in athletes using the BIA method. Medicina Sportiva, 16,
150–154.
186
P. Kutáč and M. Kopecký
Kutáč, P., Gajda, V., & Přidalová, M. (2009). The body com-
position of PE teacher. New Educational Review, 19(3–4),
263–272.
Kutáč, P., Gajda, V., Přidalová, M., & Šmajstrla, V. (2008).
Validity of measuring body composition by means of the
BIA method. New Medicine, 12, 89–93.
Kutáč, P., Přidalová, M., & Riegerová, J. (2008). Somatic
characteristics of present male and female university stu-
dents of physical education at various types of universi-
ties in the Czech Republic. Slovenská Antropológia, 11(1) ,
46–56.
Leahy, S., O’Neill, C., Sohun, R., & Jakeman, P. (2012). A
comparison of dual energy X-ray absorptiometry and bio-
electrical impedance analysis to measure total and segmen-
tal body composition in healthy young adults. European
Journal of Applied Physiology, 112, 589–595.
Loenneke, J. P., Wray, M. E., Wilson, J. M., Barnes, J. T.,
Kearney, M. L., & Pujol, T. J. (2013). Accuracy of field
methods in assessing body fat in collegiate baseball play-
ers. Research in Sports Medicine, 21, 286–291.
Maughan, R. J., & Shirreffs, S. M. (2010). Development of
hydration strategies to optimize performance for athletes
in high-intensity sports and in sports with repeated intense
efforts. Scandinavian Journal of Medicine & Science in
Sports, 20(Suppl. 2), 59–69.
McArdle, W. D., Katch, F. I., & Katch, V. L. (2007). Exer-
cise physiology. Energy, nutrition, and human performance.
Philadelphia, PA: Lippincott Williams & Wilkins.
Mojtahedi, M. C., Valentine, R. J., & Evans, E. M. (2009).
Body composition assessment in athletes with spinal cord
injury: Comparison of field methods with dual-energy
X-ray absorptiometry. Spinal Cord, 47, 698–704.
Montain, S. J. (2008). Hydration recommendations for sport
2008. Current Sports Medicine Reports, 7, 187–192.
Nazmi, S., Irfan, Ö., Osman, P., & Serdar, B. (2011). Evalu-
ation of body fat percentage of female university students
according to three different methods. Ovidius University
Annals, Series Physical Education & Sport / Science, Move-
ment & Health, 11, 244–249.
Oppliger, R. A., & Bartok, C. (2002). Hydration testing of
athletes. Sports Medicine, 32, 959–971.
Parikh, S. J., Edelman, M., Uwaifo, G. I., Freedman, R. J.,
Semega-Janneh, M., & Reynolds, J. (2004). The relation-
ship between obesity and serum 1,25-dihydroxy vitamin D
concentrations in healthy adults. Journal of Clinical Endo-
crinology & Metabolism, 89, 1196–1199.
Pluijm, S. M., Visser, M., Smit, J. H., Popp-Snijders, C.,
Roos, J. C., & Lips, P. (2001). Determinants of bone min-
eral density in older men and women: Body composition
as mediator. Journal of Bone and Mineral Research, 16,
2142–2151.
Pravn, P., Cizza, G., Bjarnason, N. H., Thompson, D., Daley,
M., Wasnich, R. D., … Christiansen, C. (1999). Low body
mass is an important risk factor for low bone mass and
increased bone loss in early postmenopausal women. Jour-
nal of Bone and Mineral Research, 14, 1622–1627.
Quinney, H. A., Dewart, R., Game, A., Snydmiller, G., War-
burton, D., & Bell, G. (2008). A 26 year physiological
description of a National Hockey League team. Applied
Physiology, Nutrition, and Metabolism, 33, 753–760.
Rahimi, R. (2006). Effect of moderate and high intensity
weight training on the body composition of overweight
men. Facta Universitatis: Series Physical Education and
Sport, 4, 93–101.
Rutherford, W. J., Diemer, G. A., & Scott, E. D. (2011).
Comparison of bioelectrical impedance and skinfolds with
hydrodensitometry in the assessment of body composi-
tion in healthy young adults. Journal of Research in Health,
Physical Education, Recreation, Sport, and Dance, 6, 56–60.
Sanchez, M. C., Sanz, D., & Zabala, M. (2007). Anthropo-
metric characteristics, body composition and somatotype
of elite junior tennis players. British Journal of Sports Medi-
cine, 41, 793–799.
Trutschnigg, B., Kilgour, R. D., Reinglas, J., Rosenthall, L.,
Hornby, L., Morais, J. A., & Vigano, A. (2008). Precision
and reliability of strength (Jamar vs. Biodex handgrip) and
body composition (dual-energy X-ray absorptiometry vs.
bioimpedance analysis) measurements in advanced cancer
patients. Applied Physiology, Nutrition, and Metabolism, 33,
132–140.
Wang, J.-G., Zhang, Y., Chen, H.-E., Li, Y., Cheng, X.-G.,
Xu, L., … Li, B. (2013). Comparison of two bioelectri-
cal impedance analysis devices with dual energy X-ray
absorptiometry and magnetic resonance imaging in the
estimation of body composition. Journal of Strength and
Conditioning Research, 27, 236–243.
Westgard, J. O. (2008). Basic method validation (3rd ed.).
Madison, WI: Westgard QC.