ArticlePDF Available

Development and validation of a correction equation for Corvis tonometry

Authors:

Abstract and Figures

Primary objective: This study uses numerical analysis and validation against clinical data to develop a method to correct intraocular pressure (IOP) measurements obtained using the Corvis Tonometer for the effects of central corneal thickness (CCT), and age. Materials and methods: Finite element analysis was conducted to simulate the effect of tonometric air pressure on the intact eye globe. The analyses considered eyes with wide variations in IOP (10-30 mm Hg), CCT (445-645 microns), R (7.2-8.4 mm), shape factor, P (0.6-1) and age (30-90 years). In each case, corneal deformation was predicted and used to estimate the IOP measurement by Corvis (CVS-IOP). Analysis of the results led to an algorithm relating estimates of true IOP as a function of CVS-IOP, CCT and age. All other parameters had negligible effect on CVS-IOP and have therefore been omitted from the algorithm. Predictions of corrected CVS-IOP, as obtained by applying the algorithm to a clinical data-set involving 634 eyes, were assessed for their association with the cornea stiffness parameters; CCT and age. Results: Analysis of CVS-IOP measurements within the 634-large clinical data-set showed strong correlation with CCT (3.06 mm Hg/100 microns, r(2) = 0.204) and weaker correlation with age (0.24 mm Hg/decade, r(2) = 0.009). Applying the algorithm to IOP measurements resulted in IOP estimations that became less correlated with both CCT (0.04 mm Hg/100 microns, r(2) = 0.005) and age (0.09 mm Hg/decade, r(2) = 0.002). Conclusions: The IOP correction process developed in this study was successful in reducing reliance of IOP measurements on both corneal thickness and age in a healthy European population.
Content may be subject to copyright.
1
Development and Validation of a Correction Equation for Corvis Tonometry
Akram Joda1, Mir Mohi Sefat Shervin2, Daniel Kook3, Ahmed Elshiekh1,4*
1School of Engineering, University of Liverpool, Liverpool L69 3GH, UK
2Smile Eyes Clinic, Munich, Germany
3Department of Ophthalmology, Ludwig-Maximilians-University, Munich, Germany
4NIHR Biomedical Research Centre for Ophthalmology, Moorfields Eye Hospital NHS Foundation
Trust and UCL Institute of Ophthalmology, UK
* Corresponding author
Proprietary interest statement - nil
Financial Disclosures
This work was partly funded by Oculus Optikgeräte GmbH, Wetzlar, Germany
Keywords: tonometry; cornea; ocular biomechanics; intraocular pressure
Author for correspondence:
Ahmed Elsheikh, School of Engineering, University of Liverpool, Liverpool L69 3GH, UK
elsheikh@liv.ac.uk, Tel: +44-151-7944848
Number of words: 3295
2
Abstract:
Purpose:
This study uses numerical analysis and validation against clinical data to develop a method to
correct intraocular pressure (IOP) measurements obtained using the Corvis ST Tonometer (CVS)
for the effects of central corneal thickness (CCT), central radius of corneal curvature (R) and age.
Materials and Methods:
Numerical analysis based on the finite element method was conducted to simulate the effect of
tonometric air pressure on the intact eye globe. The analyses considered eyes with wide variations
in IOP (10 to 30 mmHg), CCT (445 to 645 microns), R (7.2 to 8.4 mm), shape factor, P (0.6 to 1)
and age (30 to 90 years). In each case, corneal deformation was predicted and used to estimate the
IOP measurement by Corvis (CVS-IOP). Analysis of the results led to an algorithm relating
estimates of true IOP as a function of CVS-IOP, CCT and age. All other parameters had negligible
effect on eye deformation under air pressure and have therefore been omitted from the algorithm.
The models have been validated in two steps. First, the output of four models representing 4 eyes
with wide variations in IOP, CCT and age was compared to the eye deformation measured with the
CVS. Second, predictions of corrected IOP, as obtained by applying the algorithm to a clinical
dataset involving 634 patients, were assessed for their association with the cornea stiffness
parameters; CCT and age.
Results:
In four cases with wide variations in IOP, CCT and age, model predictions of the maximum apical
deformation under air pressure and the time to first applanation were within ±8.0% and ±1.5% of the
Corvis data. Analysis of CVS-IOP measurements within the 634-large clinical dataset showed
strong correlation with CCT (3.06 mmHg/100 microns, r2 = 0.204) and weaker correlation with age
(0.24 mmHg/decade, r2 = 0.009). Applying the algorithm to IOP measurements resulted in IOP
estimations that became less correlated with both CCT (0.04 mmHg/100 micros, r2 = 0.005) and age
(0.09 mmHg/decade, r2 = 0.002).
Conclusions:
CCT accounted for the majority of variance in CVS-IOP, while age and R had a much smaller effect.
The IOP correction process developed in this study was successful in reducing reliance of IOP
measurements on both corneal thickness and curvature in a healthy European population.
3
Introduction
Glaucoma is a group of diseases that can lead to optic nerve damage and irreversible loss of vision.
60 million people worldwide are affected by glaucoma; the second most-common cause of
blindness [1]. The diseases are associated with an elevated intraocular pressure (IOP), the accurate
determination of which is important for the effective management of glaucoma. The most-commonly
used method to measure IOP, and the reference standard in tonometry, is the Goldmann
applanation tonometer (GAT) [2]. The method, which determines IOP by measuring the force
required to applanate a certain area of the central cornea, has been found to be affected by corneal
stiffness parameters including the central corneal thickness (CCT), the mechanical properties of
corneal tissue and corneal curvature [37]. As a result, several correction equations have been
developed to compensate for the effect of stiffness and hence obtain a more accurate estimate of
the true IOP [5,810].
Over the past five decades several other tonometers have been developed including those that still
rely on contact techniques (most notably the Rebound Tonometer and the Dynamic Contour
Tonometer) and non-contact techniques that use an air-puff to indent the cornea. The advantages
of non-contact tonometers over contact tonometers include their relative ease of use and less-
invasive operation. However, non-contact tonometers, which are similar to contact tonometers in
that they apply a mechanical force and correlate the resulting deformation to the value of IOP, have
also been found to be influenced by corneal stiffness parameters, and in particular corneal
thickness, curvature and mechanical properties [1113]. Additionally, as non-contact tonometers
have traditionally been known to be less reliable than contact methods, their use has been mainly in
clinics, leaving hospital applications to be dominated by contact tonometers.
However, this trend is changing with the emergence of reliable non-contact tonometers such as the
Ocular Response Analyzer, which has been shown to provide close results to GAT and other
contact devices such as the Dynamic Contour Tonometer. More recently, a non-contact tonometer,
the Corvis ST (Corneal Visualization Scheimpflug Technology), has been developed by OCULUS
4
Optikgeräte, Inc. (Wetzlar, Germany) [14]. The Corvis relies on high-precision, ultra-high-speed,
Scheimpflug technology to monitor corneal deformation under air puff and produce a wide range of
tomography and deformation parameters, which have the potential to enable accurate estimates of
both corneal stiffness and IOP.
This paper presents a parametric study of the Corvis procedure to determine the effect of the main
stiffness parameters; corneal thickness, curvature, shape factor and the tissue’s material properties,
on IOP measurements. The study uses nonlinear finite element simulations of the air pressure
application on the eye as applied by the Corvis. Analysis of the results allowed developmed of a
closed-form algorithm providing estimates of IOP with significantly reduced correlation with the
stiffness parameters. Successful validation of the equation has been carried out using a clinical
dataset of 634 healthy eyes.
Methods
The finite element (FE) software ABAQUS 6.13 (Dassault Systèmes Simulia Corp.,Rhode Island,
USA) was used to model the Corvis ST testing procedure. In order to ensure accurate
representation of in-vivo conditions, the FE models adopted the following features from previous
work [1517]:
Full representation of the human eye’s outer tunic with consideration of cornea’s and sclera’s
thickness variation;
Representation of the eye’s internal fluids; the aqueous and the vitreous;
Stress-free form of the eye globe (under zero IOP);
Regional variation of sclera’s mechanical properties; and
Dynamic representation of the Corvis air pressure.
5
The models employed 10952 fifteen-noded elements organised in 25 element rings in the cornea,
124 element rings in the sclera and 1 element layer (Figure 1). This high mesh density allowed
smooth representation of ocular topography and thickness variation.
Third-order, hyperelastic Ogden models were used to represent the ocular tissues mechanical
behaviour and its variation with age [18,19]. Scleral regional variation in stiffness and its gradual
reduction from the limbus towards the optic nerve was incorporated in the models [15].
To prevent the models from rigid-body motion, all nodes along the equator were restrained in the
anterior-posterior direction (z-direction), and corneal apex and posterior pole nodes were restrained
in both the superior-inferior and temporal-nasal directions. To account for the aqueous and vitreous
incompressible behaviour, the ocular globe models were filled with an incompressible fluid with a
density of 1000 kg/m3 [20].
Before conducting the study, the stress-free configuration for each model was obtained while
following an iterative procedure explained in [16]. Two subsequent steps were then adopted in the
simulations. First, the models started from their stress-free configurations and the IOP was applied
gradually as a pressure increase in the internal incompressible fluid up to the desired level. In the
second step, space- and time-varying external air pressure was applied on the anterior surface of
the cornea. The spatial distribution of the air pressure (Figure 2a) was obtained from [12] and the
time variation was obtained from data acquired from the device manufacturers (Figure 2b). The
maximum air pressure that Corvis produces is about 180 mmHg and that was found by the
manufacturer to be reduced by approximately 50% as the air puff reached the cornea’s anterior
surface.
In the Corvis device, successive images are taken by the device’s Scheimpflug camera during the
30 ms duration of the air-puff. The images are analysed by an integrated computer to determine
IOP and several other parameters including corneal pachymetry, apical deformation, first and
second applanation time (A1, A2-time), first and second applanation length (A1, A2 length), velocity
of corneal apex at first and second applanation (A1, A2 velocity), highest concavity time (HC time),
6
and the distance between the two peaks at the point of highest concavity (Figure 3). The IOP is
measured in Corvis (CVS-IOP) as a function of the time to the first applanation event (A1-time), or
when the cornea starts to change its shape from convex to concave. Once the A1-time is known,
the external pressure acting on the cornea at that time (AP1) is measured and the IOP estimate is
calculated as a function of AP1. This process was replicated in the analysis of the FE model results
to determine AP1 and hence estimate CVS-IOP.
Parametric Study
The numerical models were used in a parametric study to quantify the effect of parameters with
potential considerable influence on CVS-IOP measurements. The parameters included the true IOP
in addition to the main stiffness parameters of the cornea, namely the thickness, curvature and
shape factor. Age was introduced for its known effect on the stress-strain behaviour of the tissue,
and it was therefore used as a parameter controlling the mechanical stiffness of both the cornea
and sclera [19,21,22]. In the study, IOP was varied from 10 to 30 mmHg in steps of 5 mmHg, central
corneal thickness (CCT) from 445 to 645 μm in steps of 50 μm, age from 30 to 90 years in steps of
10 years, central radius of anterior curvature (R) from 7.2 to 8.4 mm in steps of 0.3 mm and corneal
anterior shape factors (P) of 0.6, 0.71, 0.82 and 1. These values were compatible with the ranges of
variation reported in earlier clinical studies [2327].
The total number of models in the parametric study was 1575. In each model specific values of
CCT, R, P, age and IOP were used. The analysis step of the air puff application was dynamic and
consisted of 300 pressure increments (time step = 0.0001s) covering the 0.03 s of the Corvis
procedure. The coordinates of corneal anterior nodes were extracted at each time step using a
Python code, and a MATLAB code (MathWorks, MA) was used to determine the point of
applanation (A1-time), the external pressure at this point (AP1) and hence IOP estimate as a
product of AP1 and a calibration factor provided by Oculus.
The results of the parametric study were used to analyse the effect of CCT, R, P and age on the
CVS-IOP estimates, and to develop an algorithm relating estimates of true IOP to both the
7
measured CVS-IOP and the cornea’s stiffness parameters. Following development, the algorithm
was validated in a clinical dataset by testing its effect on the strength of association between IOP
estimates and the stiffness parameters considered. This validation exercise was preceeded by a
short comparative study where the match between the output of 4 models of four randomly-selected
eyes with wide variations in IOP, CCT and age was assessed in detail against the Corvis output for
the same eyes.
Validation clinical dataset
A clinical dataset was collected at Smile Eyes Clinics in Munich, Germany, and used in an exercise
to assess the success of the IOP algorithm developed in this study in reducing association between
IOP measurments and the cornea’s stiffness parameters. The dataset involved 634 eyes of 317
healthy participants with no pathological conditions. All patients signed a written informed consent
form. The study was approved by the local institutional review board and adhered to the tenets of
the Declaration of Helsinki. For each participant, CCT, IOP, apical deformation, A1 time and AP1
were measured by the Corvis. All measurements were performed by the same investigator (SM).
Mean, standard deviation and range of measurements are presented in Table 1.
Results
Validation of numerical results
In order to validate the numerical simulations of the Corvis procedure, the numerical results of four
models representing four randomly-selected eyes with wide variations in IOP, CCT, R and age were
considered in detail. Table 2 shows part of the Corvis output for the four eyes where the mean
values of three measurements are presented.
An eye-specific model was generated for each eye based on the CCT and R values, and the
material properties for the cornea and sclera were assumed to follow the association identified in
earlier work between stress-strain behaviour and age [15,18,19]. Constant values of the shape
8
factor, axial length and sclera diameter of 0.82, 23.7 mm and 23.0 mm, respectively, were assumed
since they were not measured clinically and were found numerically to have a negligible effect on
IOP estimations.
The four models were analysed and their output compared to Corvis parameters. Figure 4 shows a
selection of the comparisons held, which concentrate on two parameters with good repeatability and
direct relevance to corneal stiffness [28]; namely the maximum apical deformation and the first
applanation time (A1-time). The comparisons demonstrated a close match between the numerical
predictions and the Corvis output with the differences remaining below ±8% in all cases.
Parametric Study
The numerical results illustrate a clear effect of increased CCT (from 445 μm to 645 μm) in
decreasing maximum apical displacement by 37% and increasing A1-time by 14% on average,
Figure 5a. Similarly, an increase in age from 30 to 90 years (and hence increased material stiffness)
was associated with an average decrease in corneal displacement of 27% and a slight increase in
A1-time of 4%, Figure 5b. Moreover, an increase in true IOP from 10 to 30 mmHg led to an average
reduction in apical displacement of 47% and an average increase in A1-time of 48% (Figure 5c).
Changes in corneal curvature and shape factor within the considered range led to only slight
changes in corneal deformation and A1-time that were <3% as shown in Figures 5d & e. The results
show that the apical deformation and applanation time are associated with changes in CCT, IOP
and age, while variations in corneal curvature parameters (R and P) have only negligible effects on
corneal deformation behaviour.
Further, the influence of true IOP, CCT, age, R and P on estimated CVS-IOP is presented in Figure
6 (a-d). The results demonstrate that CVS-IOP is strongly associated with (or strongly influenced by)
CCT, correlated with age but with weaker association, while it is almost independent of variations in
R and P. These results illustrate that for the IOP to be estimated with reduced influence of corneal
stiffness, consideration must be made of variations in CCT and age.
IOP Correction algorithm
9
The parametric study predictions of CVS-IOP and the input parameters of true IOP, CCT and age
were used to develop equation (1) linking the four parameters together and providing estimates of
IOP that were less affected by the stiffness parameters than CVS-IOP. Values of the equations
parameters were obtained using the least squares method by minimising the sum of squared errors
between predicted and corrected IOP (Σ(IOPc CVS-IOP)2). The resulting equation has the form:
IOPc = (CCCT1 x CCVS-IOP + CCCT2) x Cage + C (1)
where CCCT1, CCCT2 are parameters representing the effect of variation in CCT (mm);
CCCT1 = 4.67 x 10-7 x CCT2 7.8 x 10-4 x CCT + 0.63
CCCT2 = -1.73 x 10-5 x CCT2 + 2.02 x 10-3 x CCT 0.97
CCVS-IOP represents effect of variation in measured CVS-IOP (mmHg) = 10 + (CVS-IOP + 1.16) /
0.389
Cage denotes effect of variation in age (years) = -2.01 x 10-5 x age2 + 1.3 x 10-3 x age + 1.00
C = 1.50 mmHg
Figure 7a shows the difference between the corrected IOP and CVS-IOP increasing mainly with
CCT but also with CVS-IOP and age. Without compensating for CCT and age variation, CVS-IOP
had a predicted measurement error as high as 10 mmHg when CCT = 645 μm and age > 60 years.
After IOP correction, the error in IOP reduced in most cases to below 1 mmHg (Figure 7b).
Correction Equation Assessment using Clinical Data
The clinical dataset described above was used to evaluate the effectiveness of the correction
algorithm in reducing reliance of IOP on the cornea’s stiffness parameters. Figure 8a presents
uncorrected CVS-IOP versus CCT, where strong association was evident from the regression and
gradient of the trend line (r2 = 0.204, slope = 0.0306 mmHg/µm). Figure 8b shows the results after
applying equation (1), leading to a reduction in r2 to 0.004 and the gradient to -0.0035 mmHg/µm.
Meanwhile, the mean CVS-IOP increased slightly from 14.45±2.83 mmHg before correction to
14.92±2.40 mmHg after correction. Similar to the numerical results, CVS-IOP was found to be
10
correlated weakly with age (Figure 9a). Using Equation (1), the coefficient of determination was
reduced from 0.009 to 0.0005 and the gradient from 0.024 to 0.0043 mmHg/year.
Discussion
The study evaluated the effect of major corneal stiffness parameters on the IOP measurements by
the Corvis. The Corvis has a number of unique features over other tonometers. First, it is able to
measure corneal thickness directly without a need for a separate device, making it possible to
directly correct for the effect of CCT on IOP. CCT measurements by the Corvis were found to have
good repeatability and accuracy compared to ultrasound pachymetry [28,30]. Second, the several
deformation parameters the device collects may make it possible to quantify corneal material
behaviour, which could then be considered in the further correction of IOP measurements.
In this paper, the effect on CVS-IOP measurements of both corneal geometric stiffness parameters
(CCT, R, P) and material stiffness (while assuming correlation with age [19,21,22]) has been
quantified. The results demonstrated clear effect of CCT on CVS-IOP, a relatively smaller effect of
material behaviour (as it varies with age) and almost no influence of R or P. Similar results were
obtained for GAT-IOP which, while being different in the nature of the force applied on the cornea,
still applies a mechanical force and correlates the resulting corneal deformation to the value of IOP
[9,3234].
The development of a correction algorithm for CVS-IOP relied initially on numerical simulation that is
representative of the eye’s geometric and material characteristics and the Corvis procedure.
Numerical simulation was found to be a reliable tool in modelling the cornea’s response to
mechanical loads such as those applied by tonometers. Similar earlier work has led to a number of
correction equations for GAT and ORA, which were later successfully validated clinically [12,35,36].
The numerical simulations of the Corvis procedure were first validated against clinical results
obtained in-vivo for four randomly-selected eyes with wide variations in CVS-IOP, age and CCT.
11
The close match between the numerical and clinical results, including the values of central
displacement and A1-time, demonstrated the reliability of the simulations and their ability to
accurately model the Corvis procedure. Subsequently, a parametric study considering wide ranges
of variation in CCT, R, P, age and true IOP was conducted. The study provided confirmation that
the A1-time is strongly correlated with IOP, CCT and age. Using the least squares method, an
algorithm quantifying the correlation of CVS-IOP with CCT and age was developed and proposed
as a means to provide estimates of IOP that were less affected by variations in corneal mechanical
stiffness.
The correction algorithm was tested against a clinical dataset of 634 healthy eyes. Uncorrected
CVS-IOP measurements were significantly correlated with CCT (r2 = 0.204, slope = 0.0306
mmHg/µm) and less correlated with age (r2 = 0.009, slope = 0.024 mmHg/year). Introducing the
correction algorithm reduced the dependency of CVS-IOP on both CCT (r2=0.004, slope = -0.0035
mmHg/µm) and age (r2 = 0.0005, slope = 0.0043 mmHg/year) considerably.
The correction algorithm presented in this paper offers a novel, simple, yet effective, method to
obtain IOP estimates that are less affected by the main corneal stiffness parameters, removing
dependency on a major error source and producing more reliable IOP estimates for glaucoma
management.
Acknowledgements:
The research was partially supported by the National Institute for Health Research (NIHR)
Biomedical Research Centre based at Moorfields Eye Hospital NHS Foundation Trust and UCL
Institute of Ophthalmology (AE). The views expressed are those of the author(s) and not necessarily
those of the NHS, the NIHR or the Department of Health of the United Kingdom.
12
References:
[1] Quigley H. A, and Broman A T., 2006, “The number of people with glaucoma worldwide in
2010 and 2020.,” Br. J. Ophthalmol., 90(3), pp. 2627.
[2] 2001, “International Organization for Standardization. Ophthalmic Instruments
Tonometers:,” ISO86122001. Geneva, Switz. Int. Organ. Stand.
[3] Imbert-Fick T., Fick I.-, and Pi I., 1887, “{ I I ),” pp. 155–162.
[4] Ehlers N., Bramsen T., and Sperling S., 1975, “< 0.001).,” 5(512), pp. 3443.
[5] Kotecha A, White E. T., Shewry J. M., and Garway-Heath D. F., 2005, “The relative effects of
corneal thickness and age on Goldmann applanation tonometry and dynamic contour
tonometry.,” Br. J. Ophthalmol., 89(12), pp. 15725.
[6] Shimmyo M., Ross A. J., Moy A., and Mostafavi R., 2003, “Intraocular pressure, Goldmann
applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians,
Hispanics, and African Americans,” Am. J. Ophthalmol., 136(4), pp. 603613.
[7] Liu J., and Roberts C. J., 2005, “Influence of corneal biomechanical properties on intraocular
pressure measurement: quantitative analysis.,” J. Cataract Refract. Surg., 31(1), pp. 14655.
[8] Elsheikh A., Alhasso D., Gunvant P., and Garway-Heath D., 2011, “Multiparameter correction
equation for Goldmann applanation tonometry,” Optom. Vis. …, 88(1), pp. 102112.
[9] Stodtmeister R., 1998, “Applanation tonometry and correction according to corneal
thickness.,” Acta Ophthalmol. Scand., 76(3), pp. 31924.
[10] Ko Y.-C., Liu C. J.-L., and Hsu W.-M., 2005, “Varying effects of corneal thickness on
intraocular pressure measurements with different tonometers.,” Eye (Lond)., 19(3), pp. 327
32.
13
[11] Kotecha A., Elsheikh A., Roberts C. R., Zhu H., and Garway-Heath D. F., 2006, “Corneal
thickness- and age-related biomechanical properties of the cornea measured with the ocular
response analyzer.,” Invest. Ophthalmol. Vis. Sci., 47(12), pp. 533747.
[12] Elsheikh A., Alhasso D., Kotecha A., and Garway-Heath D., 2009, “Assessment of the ocular
response analyzer as a tool for intraocular pressure measurement.,” J. Biomech. Eng.,
131(8), p. 081010.
[13] Tonnu P., Ho T., Newson T., El Sheikh a, Sharma K., White E., Bunce C., and Garway-Heath
D., 2005, “The influence of central corneal thickness and age on intraocular pressure
measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann
applanation tonometry.,” Br. J. Ophthalmol., 89(7), pp. 8514.
[14] Renato A., Nogueira L. P., Caldas D. L., Fontes B. M., Luz A., and Belin M. W., “Evaluation of
Corneal Shape and Biomechanics Before LASIK,” 51(2), pp. 1139.
[15] Elsheikh A., Geraghty B., Alhasso D., Knappett J., Campanelli M., and Rama P., 2010,
“Regional variation in the biomechanical properties of the human sclera.,” Exp. Eye Res.,
90(5), pp. 62433.
[16] Elsheikh A., Whitford C., Hamarashid R., Kassem W., Joda A., and Büchler P., 2013, “Stress
free configuration of the human eye.,” Med. Eng. Phys., 35(2), pp. 2116.
[17] Elsheikh A., 2010, “Finite element modeling of corneal biomechanical behavior.,” J. Refract.
Surg., 26(4), pp. 289300.
[18] Elsheikh A., Geraghty B., Rama P., Campanelli M., and Meek K. M., 2010, “Characterization
of age-related variation in corneal biomechanical properties.,” J. R. Soc. Interface, 7(51), pp.
147585.
14
[19] Geraghty B., Jones S., and Rama P., 2012, “Age-related variations in the biomechanical
properties of human sclera,” J. Mech. ….
[20] Villamarin A., Roy S., Hasballa R., Vardoulis O., Reymond P., and Stergiopulos N., 2012, “3D
simulation of the aqueous flow in the human eye.,” Med. Eng. Phys., 34(10), pp. 146270.
[21] Pallikaris I. G., Kymionis G. D., Ginis H. S., Kounis G. a, and Tsilimbaris M. K., 2005, “Ocular
rigidity in living human eyes.,” Invest. Ophthalmol. Vis. Sci., 46(2), pp. 40914.
[22] Elsheikh A., Wang D., Brown M., Rama P., Campanelli M., and Pye D., 2007, “Assessment of
corneal biomechanical properties and their variation with age.,” Curr. Eye Res., 32(1), pp.
119.
[23] Douthwaite W. a, Hough T., Edwards K., and Notay H., 1999, “The EyeSys
videokeratoscopic assessment of apical radius and p-value in the normal human cornea.,”
Ophthalmic Physiol. Opt., 19(6), pp. 46774.
[24] Douthwaite W. A, 2003, “The asphericity, curvature and tilt of the human cornea measured
using a videokeratoscope.,” Ophthalmic Physiol. Opt., 23(2), pp. 14150.
[25] Yildirim N., Sahin A., Basmak H., and Bal C., 2007, “Effect of central corneal thickness and
radius of the corneal curvature on intraocular pressure measured with the Tono-Pen and
noncontact tonometer in healthy schoolchildren.,” J. Pediatr. Ophthalmol. Strabismus, 44(4),
pp. 21622.
[26] Shimmyo M., and Orloff P. N., 2005, “Corneal thickness and axial length.,” Am. J.
Ophthalmol., 139(3), pp. 5534.
[27] Gatinel D., Hoang-Xuan T., and Azar D. T., 2001, “Determination of corneal asphericity after
myopia surgery with the excimer laser: a mathematical model.,” Invest. Ophthalmol. Vis. Sci.,
42(8), pp. 173642.
15
[28] Nemeth G., Hassan Z., Csutak A., Szalai E., Berta A., and Modis L., 2013, “Repeatability of
ocular biomechanical data measurements with a Scheimpflug-based noncontact device on
normal corneas.,” J. Refract. Surg., 29(8), pp. 55863.
[29] Smedowski A., Weglarz B., Tarnawska D., Kaarniranta K., and Wylegala E., 2014,
“Comparison of three intraocular pressure measurement methods including biomechanical
properties of the cornea.,” Invest. Ophthalmol. Vis. Sci., 55(2), pp. 66673.
[30] Reznicek L., Muth D., Kampik A., Neubauer A. S., and Hirneiss C., 2013, “Evaluation of a
novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular
hypertension and glaucoma,” pp. 1410–1414.
[31] Hong J., Xu J., Wei A., Deng S. X., Cui X., Yu X., and Sun X., 2013, “A new tonometer--the
Corvis ST tonometer: clinical comparison with noncontact and Goldmann applanation
tonometers.,” Invest. Ophthalmol. Vis. Sci., 54(1), pp. 65965.
[32] Browning A C., Bhan A, Rotchford A P., Shah S., and Dua H. S., 2004, “The effect of corneal
thickness on intraocular pressure measurement in patients with corneal pathology.,” Br. J.
Ophthalmol., 88(11), pp. 13959.
[33] Eysteinsson T., Jonasson F., Sasaki H., Arnarsson A., Sverrisson T., Sasaki K., and
Stefánsson E., 2002, “Central corneal thickness, radius of the corneal curvature and
intraocular pressure in normal subjects using non-contact techniques: Reykjavik Eye Study.,”
Acta Ophthalmol. Scand., 80(1), pp. 115.
[34] Kohlhaas M., Boehm A. G., Spoerl E., Pürsten A., Grein H. J., and Pillunat L. E., 2006,
“Effect of central corneal thickness, corneal curvature, and axial length on applanation
tonometry.,” Arch. Ophthalmol., 124(4), pp. 4716.
16
[35] Davey P. G., Elsheikh A, and Garway-Heath D. F., 2013, “Clinical evaluation of
multiparameter correction equations for Goldmann applanation tonometry.,” Eye (Lond).,
(January), pp. 19.
[36] Kwon T. H., Ghaboussi J., Pecknold D. a, and Hashash Y. M. a, 2008, “Effect of cornea
material stiffness on measured intraocular pressure.,” J. Biomech., 41(8), pp. 170713.
17
Figure captions
Figure 1 Computational mesh of the whole eye model (a) and Von mises stress distribution at the
highest concavity (b).
Figure 2 Spatial distribution (a) and time variation (b) of air pressure on the cornea’s surface. In (b)
thick black line represents air-puff produced at the device piston and grey line represents the
pressure acting on the cornea’s surface.
Figure 3 Example of a Corvis measurement showing the deformed cornea at the highest concavity.
Figure 4 Comparison of numerical predictions with clinical measurements of (a) the maximum apical
deformation and (b) the first applanation time (A1-time).
Figure 5 Relationships between maximum apical deformation and A1-time and (a) age, (b) CCT, (c)
true IOP, (d) radius of curvature and (e) shape factor.
Figure 6 CVS-IOP as a function (a) age, (b) CCT, (c) radius of curvature, and (d) shape factor.
Figure 7 Difference between the (a) true IOP and CVS-IOP and (b) true IOP and IOPc for different
true IOP levels, CCT values and ages.
Figure 8 Association between CVS-IOP measurement and CCT, (a) before correction and (b) after
correction.
Figure 9 Association between IOP measurements and age, (a) before correction and (b) after
correction.
18
Table 1 Details of the clinical dataset
Parameter
CCT (µm)
CVS-IOP (mm Hg)
Age (years)
Mean ± SD
537.3±41.8
14.5±2.8
40.0±11.6
Range
404 650
6.5 35.5
21 83
Table 2 Mean Corvis output for four cases considered in a validation study of numerical results
Case #
CCT(µm)
Age (year)
R (mm)
Case 1
17.3
581
68
7.82
Case 2
529
58
7.29
Case 3
537
31
7.55
Case 4
12.3
554
46
7.28
19
(a)
(b)
Figure 1 Computational mesh of the whole eye model (a) and Von mises stress distribution at the
highest concavity
(a)
(b)
Figure 2 Spatial distribution (a) and time variation (b) of air pressure on the cornea’s surface. In (b)
thick black line represents air-puff produced at the device piston and grey line represents the
pressure acting on the cornea’s surface
20
Figure 3 Example of a Corvis measurement showing the deformed cornea at the highest concavity
(a)
(b)
Figure 4 Comparison of numerical predictions with clinical measurements of (a) the
maximum apical deformation and (b) the first applanation time (A1-time)
21
(a)
(b)
(c)
(d)
(e)
Figure 5 Relationships between maximum apical deformation and A1-time and (a) age, (b) CCT, (c)
true IOP, (d) radius of curvature and (e) shape factor
22
(a)
(b)
(c)
(d)
Figure 6 CVS-IOP as a function (a) age, (b) CCT, (c) radius of curvature, and (d) shape factor
23
(a)
(b)
Figure 7 Difference between the (a) true IOP and uncorrected CVS-IOP and (b) true IOP
and corrected CVS-IOP for different true IOP levels, CCT values and ages
24
(a)
(b)
Figure 8 Association between CVS-IOP measurement and CCT, (a) before correction and (b) after
correction.
(a)
(b)
Figure 9 Association between IOP measurements and age, (a) before correction and (b) after correction
... Corvis ST provides a corneal-corrected IOP measurement, the bIOP, designed to exclude the influence of central corneal thickness and age (Joda et al., 2016;Eliasy et al., 2022). There is still a need to reduce its dependence on corneal mechanical properties. ...
... Previous numerical works attempted to estimate the IOP removing the influence of corneal structural parameters. Joda et al. (Joda et al., 2016) and more recently Eliasy et al. (Eliasy et al., 2022) proposed an algorithm to correct the IOP estimated by Corvis ST trying to exclude the influence of CCT and age. However, their algorithm is based on structural numerical simulations in which the air pressure on the cornea does not vary depending on the mechanical properties and thickness of the patient. ...
Article
Full-text available
Introduction: Glaucoma, a leading cause of blindness worldwide, is primarily caused by elevated intraocular pressure (IOP). Accurate and reliable IOP measurements are the key to diagnose the pathology in time and to provide for effective treatment strategies. The currently available methods for measuring IOP include contact and non contact tonometers (NCT), which estimate IOP based on the corneal deformation caused by an external load, that in the case of NCT is an air pulse. The deformation of the cornea during the tonometry is the result of the coupling between the IOP, the mechanical properties of the corneal tissue, the corneal thickness, and the external force applied. Therefore, there is the need to decouple the four contributions to estimate the IOP more reliably. Methods: This paper aims to propose a new methodology to estimate the IOP based on the analysis of the mechanical work performed by the air jet and by the IOP during the NCT test. A numerical eye model is presented, initially deformed by the action of a falling mass to study the energy balance. Subsequently, Fluid-Structure Interaction (FSI) simulations are conducted to simulate the action of Corvis ST. Results and discussion: The new IOP estimation procedure is proposed based on the results of the simulations. The methodology is centred on the analysis of the time of maximum apex velocity rather than the point of first applanation leading to a new IOP estimation not influenced by the geometrical and mechanical corneal factors.
... Полноценная диагностика кератоконуса в настоящее время невозможна без анализа жесткости или вязко-эластических свойств роговицы. Внедрение в клиническую практику бесконтактных тонометров, измеряющих внутриглазное давление с учетом биомеханических свойств роговицы, открыло новые возможности для изучения кератэктатического процесса [11,12,13,14,15]. Однако деформации под воздействием воздушной струи при проведении вышеуказанного исследования подвергается центральная зона роговицы диаметром 2,0 мм, вследствие чего оценить биомеханические показатели парацентральных участков не представляется возможным. ...
... С помощью Pentacam AXL оценивали среднее и максимальное значения кератометрии (Кm и Кmax), а также толщины роговицы в центре и на вершине кератоконуса (ЦТР и ТВ). С помощью Corvis ST определяли относительную толщину роговицы по Амброзио (Ambrosio Relational Thickness -ARTh), которая описывает соотношение между толщиной роговицы в самой тонкой точке и индексом прогрессии пахиметрии [14], индекс напряжения-деформации (Stress-Strain Index или SSI), жесткость роговицы (Stiffness Parameter -SPA1), рассчитываемую как разность между силой воздушного импульса на поверхности роговицы и биомеханически компенсированным внутриглазным давлением (bIOP) [16]. ...
Article
Full-text available
256 subjects (256 eyes) participated in the study: 174 healthy individuals (174 eyes) with normal corneas with different refraction and 82 patients (82 eyes) with keratoconus at stages 1, 2 and 3. Based on mathematical modeling, the biomechanical corneal properties with different patterns of keratoconus at stages 1, 2, and 3 were evaluated. Mathematical modeling made it possible to determine the dimensions and degrees of local corneal biomechanical properties reduction, which depended on the disease stage and practically did not change from the shape and keratopographic pattern localization.
... This approach has been used to estimate the effect of the corneal elastic modulus, intraocular pressure (IOP), and central corneal thickness (CCT) on the corneal deformation under air-puff (Kling et al., 2014;Ariza-Gracia et al., 2015). The insights obtained from such models can be used to estimate material properties based on easily-quantifiable APDI outputs (Joda et al., 2016;Chen et al., 2019;Eliasy et al., 2019). FE models can also be used to estimate material parameters by matching the output of a simulation with experimental results, a procedure known as inverse analysis. ...
... The thickness could be estimated from the image obtained in scleral APDI provided that this is a visible region, as is done in corneal APDI. The IOP could be estimated from tonometry measurements of the cornea, where work has been done on estimating IOP without the influence of material and geometric properties (Joda et al., 2016). ...
Article
Full-text available
Introduction: The mechanical properties of the sclera are related to its structural function, and changes to these properties are believed to contribute to pathologies such as myopia. Air-puff deformation imaging is a tool that uses an imaging system coupled with an air-puff excitation source to induce and measure deformation in a tissue in vivo. Typically used for the study of the cornea’s mechanical properties and IOP, this tool has been proposed as a method to evaluate scleral stiffness. Methods: In this work, we present a computational model of the rabbit eye to assess scleral deformation under air-puff. Parametric studies were conducted to evaluate the effects of material properties, intraocular pressure, and other parameters on the deformation response. Output from the model was also compared to experimental measurements of air-puff deformation in rabbit eyes under varying IOP. Results: Central deformation response was found to be most influenced by material properties of the sclera (at site of air-puff and posterior), thickness, and IOP, whereas deformation profile was most influenced by material properties. Experimental and simulated IOP dependence were found to be similar (RMSE = 0.13 mm). Discussion: Scleral APDI could be a useful tool for quick in vivo assessment of scleral stiffness.
... Corvis ST, apart from the standard GAT-correlated IOP (CVS-IOP) provides information on a biomechanically compensated IOP (bIOP). The bIOP algorithm incorporates factors such as CCT and age in addition to deformation response parameters to adjust for the effect of stiffness on IOP measurements [25][26][27]. Compared to manometric measurements of the interior of ex vivo human eyes, bIOP provided estimates very close to the true IOP without correlating to the CCT [28]. ...
Article
Full-text available
Background/Objectives: Intraocular pressure (IOP) readings using three different methods (Goldmann applanation tonometry (GAT), Corvis ST, and iCare) were compared in patients who underwent penetrating keratoplasty (PK). Methods: An observational cross-sectional study with prospective recruitment of patients was conducted. IOP measurements were acquired using GAT, iCare, and Corvis (including both uncorrected IOP (CVS-IOP) and biomechanical IOP (bIOP)), and the agreement among methods was analyzed using Bland–Altman plots. Secondary outcomes included the influence of CCT, the number of sutures, the size of the corneal donor button, and the use of antiglaucoma topical medications on the IOP readings using the three methods. Results: Twenty-five eyes from 25 patients were included. The Bland–Altman analysis showed the narrowest limits of agreement (LoA) between GAT and bIOP (7.5 mmHg). The difference between iCare and GAT IOP showed a bias of 1.26 ± 3.8 mmHg, with increased variability in cases with more remaining sutures (p = 0.0079). A higher CCT was moderately associated with lower bIOP readings (p = 0.0067), but no significant impact of CCT on the difference in the IOP measurements between GAT and other tonometers was found. Additionally, there were no significant differences in tonometer readings based on the use of antiglaucoma medications or the corneal donor button size. Conclusions: Good agreement was found between iCare, CVS-IOP, bIOP, and GAT-IOP readings with the comparison between GAT-IOP and bIOP resulting in the narrowest 95% LoA. The difference between the GAT-IOP and iCare readings tended to be influenced by the number of sutures at the graft–host interface. Higher CCT values were associated with lower bIOP readings; however, the differences in tonometer readings compared to GAT-IOP were not found to be influenced by CCT.
... The SSI best describes the material stiffness of the cornea due to its independence from the intraocular pressure and the corneal thickness. 22 In addition, the biomechanical corrected intraocular pressure (bIOP) 23 and the pachymetry values (CVS-CT) were gathered. ...
Article
Full-text available
Purpose The purpose of this study was to investigate corneal stiffening after epithelium-off accelerated corneal cross-linking (CXL; 9 mW/cm²) in progressive keratoconus (KC) with different methods of epithelial debridement. Methods This was a retrospective, interventional, and non-randomized study. In group 1, the epithelium was removed using a hockey knife (N = 45). In group 2 (N = 39) and group 3 (N = 22), the epithelial thickness was measured by optical coherence tomography (OCT) and the epithelium was ablated by excimer laser, but, in group 3, stromal ablation was performed additionally to correct high order aberrations (HOAs). Corneal biomechanics (integrated invers radius [IIR], stress-strain index [SSI]) and corneal tomography (thinnest corneal thickness [TCT]) were assessed with Corvis ST and Pentacam prior to and 1 month after CXL. Results Corneal tomography did not differ among the groups preoperatively (P > 0.05). TCT decreased significantly in all groups after surgery (all P < 0.05). Nonetheless, corneal biomechanical stiffening was found in all three groups indicated by a decreased IIR and an increased SSI (all P < 0.05). For group 3, the HOA improved significantly (P < 0.001). Among the groups, there were no significant differences in changes of biomechanical parameters, but TCT was significantly reduced after laser ablation. Conclusions Corneal stiffening after CXL is independent from epithelial removal. In particular, despite the removal of stromal tissue to correct HOA, a stiffening effect was achieved in keratoconic corneas, even it was less pronounced compared to mechanical epithelial removal. The reduction in HOA indicates a possible improvement in visual acuity. Translation Relevance Cross-linking stiffens the keratoconus independent of epithelial debridement technique and may compensate minor stromal laser ablation.
... The bIOP was derived by nite element simulation taking into account the in uence of central corneal thickness, age, and DCR parameters. This value was validated both experimentally and clinically 17 . CBI includes several dynamic corneal response parameters in addition to ARTh. ...
Preprint
Full-text available
Objective To assess the performance of a hybrid Transformer-based convolutional neural network (CNN) model for automated detection of keratoconus in stand-alone Scheimpflug-based dynamic corneal deformation videos (DCDV). Design Retrospective cohort study. Methods We used transfer learning for feature extraction from DCDVs. These feature maps were augmented by self-attention to model long-range dependencies before classification to directly identify keratoconus. Model performance was evaluated by objective accuracy metrics based on DCDVs from two independent cohorts with 275 and 546 subjects. Main outcome measures: Area under the receiver operating characteristics curve (AUC), accuracy, specificity, sensitivity, and F1 score. Results The sensitivity and specificity of the model in detecting keratoconus were 93% and 84%, respectively. The AUC of the keratoconus probability score based on the external validation database was 0.97. Conclusions The hybrid Transformer-based model was highly sensitive and specific in discriminating normal from keratoconic eyes using DCDV(s) at levels that may prove useful in clinical practice. Translational Relevance The hybrid Transformer-based model can detect keratoconus from non-invasive corneal videos directly without requiring corneal topography or tomography exhibiting potential application in corneal research and clinical practice.
... It has been reported that CST is more sensitive to corneal biomechanical changes after cataract surgery than the ORA (18,19). Recently, biomechanically-corrected intraocular pressure (bIOP), a newly released CST parameter, was corrected for corneal stiffness, CCT and age (20). ...
Article
Full-text available
Purpose To compare corneal biomechanical properties and intraocular pressure (IOP) measurements in patients who underwent Descemet’s stripping with endothelial keratoplasty (DSEK) with those of the follow healthy eyes. Methods In this retrospective comparative study, a total of 35 eyes of 35 patients who underwent DSEK by a single surgeon from 2015.02 to 2019.12 were enrolled along with their fellow healthy eyes. Corneal biomechanical parameters were assessed at least 3 months post-DSEK using Corneal Visualization Scheimpflug Technology (CST). IOP was measured by CST, Goldmann applanation tonometry (GAT), and MacKay-Marg tonometer. Results Central corneal thickness (CCT) and stiffness parameter at first applanation (SP-A1) were significantly increased after DSEK when compared to the fellow eyes. In DSEK eyes, biomechanically-corrected intraocular pressure (bIOP) and MacKay-Marg IOP correlated significantly with GAT IOP measurements, with bIOP showed the lowest IOP values. All the IOP values did not correlate with CCT. However, GAT-IOP and MacKay-Marg IOP showed a positive correlation with SP-A1. Conclusion The corneal stiffness increased after DSEK. Central corneal thickness may have less influence than corneal biomechanics on IOP measurements in eyes after DSEK. Biomechanically-corrected IOP obtained by CST seemed to be lower than other tonometry techniques in DSEK eyes, perhaps because of correction for corneal stiffness, CCT and age.
Article
Purpose Chronic hyperglycemia causes changes in corneal biomechanics that can be measured with the Scheimpflug Analyzer Corvis ST. The diagnostic reliability of the new diabetes mellitus (DM) index developed based on this should be evaluated. Methods In a prospective cross-sectional study, the index was initially developed using data from 81 patients with DM and 75 healthy subjects based on logistic regression analysis. The reliability of the DM index was subsequently assessed using data from another 61 patients and 37 healthy individuals. In addition, the dependence of the DM index on indicators of disease severity was analyzed. Results The index initially achieved a sensitivity of 79% and specificity of 80% with a cutoff value of 0.58. The evaluation showed a sensitivity of 67% and specificity of 76% with an optimized cutoff of 0.51 (area under the curve = 0.737, P < 0.001). The DM index correlated weakly with the severity of diabetic retinopathy (r = 0.209, P = 0.014). It was increased in the presence of diabetic maculopathy ( P = 0.037) and in type 1 DM compared with patients with type 2 disease ( P = 0.039). Conclusions In this first evaluation, the new DM index achieved sufficiently good sensitivity and specificity and was weakly associated with disease-specific factors. With further improvements, it could complement the diagnostic options in DM with a simple, rapid, and noninvasive assessment method.
Article
Full-text available
To evaluate the agreement of intraocular pressure (IOP) and central corneal thickness (CCT) measurements obtained with the non-contact tonometer Corvis Scheimpflug Technology (Corvis ST, OCULUS, Wetzlar, Germany) versus Goldmann applanation tonometry (GAT) and ultrasound-based corneal pachymetry (US-CCT). Eye healthy participants, patients with ocular hypertension (OHT) and patients with open-angle glaucoma were included in this prospective study. In each participant, GAT, US-CCT and measurements with Corvis ST were obtained (Corvis-IOP and Corvis-CCT). Accuracy and repeatability were tested by correlation and regression analyses, Bland-Altman plots and assessment of intraclass correlation coefficients. A consecutive series of 188 right study eyes of 188 participants (142 eyes with glaucoma, 10 eyes with OHT and 36 control eyes) were included in this prospective study. The mean GAT of all included was 14.5±4.8 mm Hg compared with mean Corvis-IOP of 15.4±5.6 mm Hg (Spearman's r=0.75, p<0.0001). Mean US-CCT was 544.56±40.0 µm compared with Corvis-CCT of 545.2±46.5 µm (Pearson's r=0.78, p<0.0001). Bland-Altman plots of all included eyes as well as subgroup analyses revealed good agreement of the IOP and CCT measurement techniques. High intraclass correlation coefficient values in 17 patients with repeated measurements revealed very good repeatability (0.942 and 0.937 for Corvis-IOP and Corvis-CCT, respectively). Corvis-IOP but not GAT showed a trend of dependence on CCT. Obtaining CCT and measuring IOP with the Corvis ST reveals very good repeatability and good accuracy in healthy subjects and patients with OHT and glaucoma when compared with standardised US pachymetry or GAT.
Article
Full-text available
To analyze the repeatability of a new device measuring ocular biomechanical properties, central corneal thickness (CCT), and intraocular pressure (IOP) and to investigate these parameters and their correlations in healthy eyes. Three consecutive measurements were performed on each eye using the CorVis ST device (Oculus Optikgeräte, Inc., Wetzler, Germany). Ten specific parameters, CCT, and IOP were measured. Biometric data were recorded with IOLMaster (Carl Zeiss Meditec, Jena, Germany). This study comprised 75 eyes of 75 healthy volunteers (mean age: 61.24 ± 15.72 years). Mean IOP was 15.02 ± 2.90 mm Hg and mean CCT was 556.33 ± 33.13 μm. Intraclass correlation coefficient (ICC) was 0.865 for IOP and 0.970 for CCT, and coefficient of variation was 0.069 for IOP and 0.008 for CCT. ICC was 0.758 for maximum amplitude at highest concavity and 0.784 for first applanation time, and less than 0.6 for all other parameters. The device-specific data showed no significant relationship with age and axial length. Flattest and steepest keratometric values and IOP showed a significant correlation with the 10 device-specific parameters. The CorVis ST showed high repeatability for only IOP and pachymetric values. Single measurements are not reliable for the 10 device-specific parameters. The device allows for conducting clinical examinations and screening for surgeries altering ocular biomechanical properties with some form of averaging of multiple measurements. [J Refract Surg. 2013;29(8):558-563.].
Article
Full-text available
PurposeTo evaluate and compare the multiparameter equations in correcting intraocular pressure (IOP) measurements obtained using the Goldmann applanation tonometer (IOPG) for the effects of central corneal thickness (CCT), corneal curvature (R), and age in different ethnic populations.Methods Data of IOPG, CCT, R, and age were collected from three clinical centers. The sample size consisted of 945 eyes of 945 glaucoma patients or suspects (669 Europeans, 127 African Americans, and 149 Indians). The 'corrected IOP' was calculated using five multiparameter equations to decrease the association of CCT, R, and age with measured IOP. Regression analyses were performed to calculate variance (r(2)) and determine the association of CCT, R, and age with IOPG and corrected IOP (residual association).ResultsOverall, CCT accounted for the majority of variance in IOPG, while R and age had a much smaller effect, with the combined effect on IOPG ranging from 4.7 to 7.5% in the three data sets. The residual association of CCT, R, and age with corrected IOP in the three groups ranged from 0.2 to 1.3% and 0.5 to 1.8% with the application of the Elsheikh and the Chihara equations, respectively. The residual association of CCT, R, and age with corrected IOP calculated using the Ehlers, Orssengo and Pye, and Shimmoyo equations were 7-11.5, 1.8-11.7, and 4.6-8.3%, respectively.Conclusion The Elsheikh and the Chihara equations better decreased the association of CCT, R, and age with measured IOP than the Ehlers, Orssengo and Pye, and Shimmoyo equations.Eye advance online publication, 15 March 2013; doi:10.1038/eye.2013.23.
Article
Full-text available
Purpose: To compare intraocular pressure (IOP) measurements obtained using the Topocon noncontact tonometer (NCT), the Goldmann applanation tonometer (GAT), and the Corvis ST (CST), a newly developed tonometer with features of visualization and measurement of the corneal deformation response to an air impulse. A secondary objective was to assess the agreement among the devices. Methods: Fifty-nine participants, including glaucoma patients (36 cases) and control volunteers (23 cases), were enrolled. One eye was selected randomly for further study. IOP measurements were obtained with the CST, NCT, and GAT by two experienced clinicians. IOP values were compared. Intraobserver variability and interobserver variability were assessed by the coefficient of variation and intraclass correlation coefficient. Device agreement was calculated by Bland-Altman analysis. Results: Mean IOP for all examined eyes was 18.9 ± 5.8 mm Hg for CST, 21.3 ± 6.8 mm Hg for NCT, and 20.3 ± 5.7 mm Hg for GAT. There was no statistically significant difference in IOP measurements among the tonometers except between the CST and NCT. Correlation analysis showed a high correlation between each pair of devices (all P < 0.001). The CST displayed the best intraobserver variability and interobserver variability. Bland-Altman analysis revealed a bias between CST and GAT, CST and NCT, and GAT and NCT of -1.3, -2.4, and -1.1 mm Hg, with 95% limits of agreement of -6.2 to 3.5 mm Hg, -10.1 to 5.2 mm Hg, and -8.3 to 6.2 mm Hg, respectively. Conclusions: The CST offers an alternative method for measuring IOP. IOP measurements taken with these devices may not be interchangeable.
Article
The clinical usefulness of an automatic tonometer, the NIDEK NT-1000 (NT-1000) was evaluated by comparing results obtained with the non-contact tonometer (NCT) to results of Goldmann applanation tonometry (GAT) in 229 eyes of 127 apparently healthy persons enrolled in a health screening program. The mean±SD for the NT-1000 and GAT readings were 13.9±3.5 mmHg and 15.2±3.5 mmHg. The regression equation was NT-1000=0.7 x (GAT) +3.2 and the coefficient of correlation was 0.70. Readings greater than 20 mmHg were obtained for 22 eyes using GAT but for only 7 eyes using NT-1000. Furthermore, NT-1000 readings greater than 18.7 mmHg, which corresponded in our study to GAT readings of 20.0 mmHg, were obtained in only 9 eyes. The coefficient of correlation between NT-1000 and GAT in this study was lower than the coefficient measured using manually operated NCTs in previous studies, leading us to conclude that NT-1000 readings must be evaluated carefully.
Article
Purpose: The aim of this study was to show the usefulness of three methods for measuring IOP: Goldmann applanation tonometry, rebound tonometry, and Ultra-High-Speed Scheimpflug technology. Methods: The examined group consisted of 96 patients (192 eyes), including 63 women and 33 men with a mean age of 59.3 ± 19.9 years. Together, 152 healthy eyes and 40 eyes with different pathologies were examined. Intraocular pressure was measured using the Goldmann applanation tonometer (GAT), the Icare Pro rebound tonometer (RT), and Ultra-High-Speed Scheimpflug technology (UHS ST; Corvis ST with pachymetry). Additionally, corneal pachymetry was conducted with a Scheimpflug camera (Pentacam) and an Ultrasound Pachymeter (A-scan Plus) as a comparison for Corvis ST pachymetry. Results: The mean IOPs were 15.6 ± 3.75 mm Hg, 15.6 ± 3.5 mm Hg, and 16.1 ± 4.0 mm Hg when measured with the GAT, the RT, and the UHS ST, respectively. The mean central corneal thickness (CCT) was 543.7 ± 52.7 μm, 547.9 ± 54.0 μm, and 556.25 ± 38.8 μm as measured with the UHS ST, the Pentacam, and the Ultrasound Pachymeter, respectively. In comparison between devices, there was a significant difference between IOP values measured with the GAT and the RT versus the UHS ST (P < 0.001), and there was no significant difference between GAT and RT (P = 0.5). No significant differences were observed in CCT measured with the UHS ST, Pentacam, and Ultrasound Pachymeter. Conclusions: We showed that the RT Icare Pro ensures IOP measurements that are more comparable with the measurements obtained with the GAT than the measurements that are provided by UHS ST.
Article
This study examined age-related changes in biomechanical behaviour in the anterior, equatorial and posterior regions of the human sclera (white of the eye). Circumferential strip specimens were extracted from areas close to the limbus, equator and posterior pole in 45 donor scleras ranging in age between 51 and 84 years. The strips were subjected to cycles of uniaxial tension loading at a strain rate of 8% per minute while monitoring their load-deformation behaviour. All specimens demonstrated nonlinear behaviour with an initially low tangent modulus (a measure of material stiffness) increasing under higher stresses. The average ratios between the tangent modulus at a high stress of 1MPa and that at a low stress of 0.05MPa were 11.2±1.7, 12.0±1.7 and 12.4±1.5 for anterior, equatorial and posterior specimens, respectively. Stiffening was observed with age in all regions, but it was statistically significant only in the anterior region (P<0.01). Anterior specimens showed the largest stiffness growth with advancing age in both the initial, matrix regulated phase of behaviour (0.32MPa/decade), and the final, collagen regulated phase (3.97MPa/decade), followed by equatorial (0.27 and 2.15MPa/decade) then posterior specimens (0.14 and 0.26MPa/decade). The stress-strain behaviour of scleral tissue exhibits increasing stiffness with higher age. In addition to a regional variation of material stiffness, the rate of stiffness growth with age also varies between regions.
Article
Numerical simulations of eye globes often rely on topographies that have been measured in vivo using devices such as the Pentacam or OCT. The topographies, which represent the form of the already stressed eye under the existing intraocular pressure, introduce approximations in the analysis. The accuracy of the simulations could be improved if either the stress state of the eye under the effect of intraocular pressure is determined, or the stress-free form of the eye estimated prior to conducting the analysis. This study reviews earlier attempts to address this problem and assesses the performance of an iterative technique proposed by Pandolfi and Holzapfel [1], which is both simple to implement and promises high accuracy in estimating the eye's stress-free form. A parametric study has been conducted and demonstrated reliance of the error level on the level of flexibility of the eye model, especially in the cornea region. However, in all cases considered 3-4 analysis iterations were sufficient to produce a stress-free form with average errors in node location <10(-6)mm and a maximal error <10(-4)mm. This error level, which is similar to what has been achieved with other methods and orders of magnitude lower than the accuracy of current clinical topography systems, justifies the use of the technique as a pre-processing step in ocular numerical simulations.
Article
Glaucoma results in an increase in the resistance of the aqueous humor outflow, which in turn leads to an increase of the intraocular pressure (IOP). Several treatments are proposed to reduce and stabilize the IOP that include medications, filtering surgery and glaucoma drainage devices (GDD). So far computational fluid dynamics (CFD) modeling of the eye drainage system has not yet been well studied. Therefore our goal was to provide a 3D CFD model of the eye based on the anatomy of a real human eye. Such a tool would serve for future evaluation of new glaucoma surgical techniques involving, for example, GDD. The model was based on stacks of microphotographs from human eye slides from which digital processing of the images of the eye structure and 3D reconstruction of the model were performed. Simulations of the distribution of pressure and flow velocity in the model of a healthy eye gave results comparable to physiology references. Mimicking glaucoma conditions led to an increase of the IOP from normal range, which went down to lower values after a filtering procedure. Further refinements in the boundary conditions for the filtering procedure shall improve the accuracy of this innovative tool for modeling glaucoma surgery.