ArticlePDF Available

Can we Replace Arterial Blood Gas Analysis by Pulse Oximetry in Neonates with Respiratory Distress Syndrome, who are Treated According to INSURE Protocol?

Authors:

Abstract

Neonates with respiratory distress syndrome (RDS), who are treated according to INSURE protocol; require arterial blood gas (ABG) analysis to decide on appropriate management. We conducted this study to investigate the validity of pulse oximetry instead of frequent ABG analysis in the evaluation of these patients. From a total of 193 blood samples obtained from 30 neonates <1500 grams with RDS, 7.2% were found to have one or more of the followings: acidosis, hypercapnia, or hypoxemia. We found that pulse oximetry in the detection of hyperoxemia had a good validity to appropriately manage patients without blood gas analysis. However, the validity of pulse oximetry was not good enough to detect acidosis, hypercapnia, and hypoxemia.
Iran J Med Sci May 2015; Vol 40 No 3
IJMS
Vol 40, No 3, May 2015
264
Can we Replace Arterial Blood Gas Analysis by
Pulse Oximetry in Neonates with Respiratory
Distress Syndrome, who are Treated According
to INSURE Protocol?
Pedram Niknafs1, MD; Elahe Norouzi1,
MD; Bahareh Bahman Bijari1, MD;
Mohammad Reza Baneshi2, PhD
1Division of Neonatology, Afzalipour
Medical Center, Kerman University of
Medical Sciences, Kerman, Iran;
2Research Center for Modeling in Health,
Institute for Future Studies, Kerman
University of Medical Sciences, Kerman,
Iran
Correspondence:
Pedram Niknafs, MD;
Division of Neonatology,
Afzalipour Medical Center,
Imam Khomeini Highway,
Kerman, Iran
Tel / Fa x : +98 34 33222763
Email: pniknafs@yahoo.com
Received: 24 August 2013
Revised: 1 December 2013
Accepted: 6 December 2013
Abstract
Neonates with respiratory distress syndrome (RDS), who are
treated according to INSURE protocol; require arterial blood
gas (ABG) analysis to decide on appropriate management. We
conducted this study to investigate the validity of pulse oximetry
instead of frequent ABG analysis in the evaluation of these
patients. From a total of 193 blood samples obtained from 30
neonates <1500 grams with RDS, 7.2% were found to have one
or more of the followings: acidosis, hypercapnia, or hypoxemia.
We found that pulse oximetry in the detection of hyperoxemia
had a good validity to appropriately manage patients without
blood gas analysis. However, the validity of pulse oximetry was
not good enough to detect acidosis, hypercapnia, and hypoxemia.
Please cite this article as: Niknafs P, Norouzi E, Bahman Bijari B, Baneshi MR.
Can we Replace Arterial Blood Gas Analysis by Pulse Oximetry in Neonates with
Respirator y Distress Syndrome, who are Treated According to INSURE Protocol?
Iran J Med Sci. 2015;40(3):264-267.
Keywords Infant Respiratory distress syndrome Oximetry
Blood gaz analysis
Brief Report
Introduction
In neonates with birth weight <1500 grams, when the clinical
diagnosis of respiratory distress syndrome (RDS) is made based on
the respiratory symptoms and conrmatory studies (e.g. blood gas
analysis, chest x-ray, etc.), the appropriate management is to place
the patient under nasal continuous positive airway pressure (NCPAP).
If respiratory distress increases, surfactant is administered according
to INSURE protocol1 (transient intubation, surfactant administration,
rapid extubation to NCPAP2). These infants require an umbilical
artery catheter to obtain arterial blood gas (ABG) every 30 minutes
to 4 hours for accurate monitoring of gas exchange.3
Arterial blood gas analysis is the gold standard by which the
adequacy of oxygenation and ventilation are assessed.1 Although
umbilical catheterization is safe and well tolerated in most infants,
it is associated with serious complications which could be life
threatening.3-5
Pulse oximetry is a technique that indirectly determines
oxygenation in a continuous noninvasive manner. The percentage
of saturated hemoglobin is calculated from the difference between
light frequencies of the pulsatile ow as it passes beneath the
sensor at distal extremity of the infant.3,4
We did not nd any study regarding the validity of pulse oximetry as
a bedside monitoring in the assessment of neonates with birth weight
265
Pulse oximetr y versus ABG in RDS
Iran J Med Sci May 2015; Vol 40 No 3
lower than 1500 grams with RDS. Therefore, we
conducted this study to investigate the diagnostic
value of pulse oximetry in detecting acidosis,
hypercapnia, hypoxemia, and hyperoxemia.
Patien ts and M ethods
This was a prospective diagnostic test study (carried
out at the Afzalipour Medical Center in Kerman-Iran,
from October 2011 to March 2012) on 30 preterm
infants with moderately severe RDS weighing less
than 1500 grams at birth.
6
Neonates in whom
surfactant had been introduced as INSURE protocol
met the criteria for the study. These infants were
placed under bubble NCPAP with FiO2 of 40-50%
and CPAP of 5-6 cmH2O after INSURE procedure.
Umbilical or peripheral artery catheter was
inserted for all patients and ABG according to
the patient status was obtained every 30 minutes
to 4 hours (OPTI CCA-TS blood gas analyzer,
OPTI Medical Company, USA). All patients were
monitored continuously, physical examination
was done repeatedly, and pulse oximetry was
performed by a probe at distal extremity (SpO
2
Masimo set, NOVIN S1800 patient monitor, Saadat
Co., Ltd., Iran). In this study, blood gas analysis
was the gold standard. pH higher than 7.20, PaO2
equal to 50-80 mmHg, PaCO
2
less than 60 mmHg,
and oxygen saturation in the range of 85 -95%
were considered as normal values.7
Variables including pH, PaO
2
, PaCO
2
, oxygen
saturation, as well as gestational age, birth
weight, and gender of the infant were recorded.
SPSS 19 was used to analyze the data. To
assess the degree of dependency among
observations of the same patients, we estimated
the ICC value by the applied multilevel analysis.
Sensitivity, specicity, positive predictive and
negative predictive values were extracted from
2×2 contingency tables of results.8
In this study, any abnormality in blood gas
analysis including low pH, high PaCO2, low
PaO2, and high PaO2 was considered as “disease/
condition”, and SpO
2
value by pulse oximetry was
considered as “test”.
The study protocol received approval from the
Ethics Committee of Kerman University of Medical
Sciences with the code number 323/90/K, and all
patients provided written informed consent prior
to participation.
Results
Among 30 preterm infants who were studied, 14
(46.7%) were female and 16 (53.3%) were male.
Mean gestational age at birth was 311/7-week, with
minimum of 27-week and maximum of 351/7-week.
Mean birth weight was 1340 grams, with the range
of 900 to 1500 grams.
193 ABG specimens were obtained from the
infants through umbilical or peripheral arterial
catheters. Of the 193 blood samples, only 14
specimens (7.2%) were found to have one or
more of acidosis, hypercapnia, and hypoxemia.
Acidosis (pH<7.20) was found in 10 specimens
from a total of 193. Sensitivity of pulse oximeter in
the estimation of acidosis was 40% and its specicity
was 100%. Positive predictive value and negative
predictive value of pulse oximetry in the prediction
of acidosis were 100% and 96.8%, respectively.
From 193 ABG specimens, hypercapnia
(PaCO2>60) was found in six specimens.
Sensitivity of pulse oximetry with a cut-off point
of 85% in the detection of hypercapnia was 50%.
Specicity of pulse oximetry in the prediction of
normocapnia was 99.6%. Positive predictive value
and negative predictive value of pulse oximetry
for PaCO2 were 75% and 98.4%, respectively.
From 193 specimens, four showed hypoxemia
(PaO2<50) by blood gas analysis. Sensitivity of
pulse oximetry in the detection of hypoxemia
was 75% and its specicity was 99.5%. Positive
predictive value and negative predictive value
of pulse oximetry for hypoxemia were 75% and
98.5%, respectively.
In the evaluation of hyperoxemia (PaO2>80),
from 193 ABG specimens, 61 had PaO
2
higher than
80 (hyperoxemia). Sensitivity of pulse oximetry in
the detection of hyperoxemia was 83% and its
specicity was 92.4%. Positive predictive value
and negative predictive value for hyperoxemia
were 83% and 92.4%, respectively (table 1).
Discussion
In a study performed by Witting et al., sensitivity
Table 1: Prediction of ABG abnormalities by pulse oximetry in neonates with RDS
Acidosis
(pH<7.20)
Hypercapnia
(PaCO2>60)
Hypoxemia
(PaO2<50)
Hyperoxemia
(PaO2>80)
Cut-off point SpO2 (%) 85 85 85 95
Sensitivity (%) 40 50 75 83
Specicity (%) 100 99.6 99.5 92.4
PPV (%) 100 75 75 83
NPV (%) 96.8 98.4 98.5 92.4
NPV: Negative predictive value; PPV: Positive predictive value
266
Niknafs P, Norouzi E, Bahman Bijari B, Baneshi MR
Iran J Med Sci May 2015; Vol 40 No 3
and specicity of room-air pulse oximetry (with
SpO2≥96%) in detecting moderate hypercapnia
(PaCO2>50) in patients admitted to emergency ward
were 96% and 39%, respectively. They concluded
that room-air oxygen saturation below 95% could
alert the physician to the onset of hypoventilation.9
In another study, Ritonga et al. assessed
the validity of pulse oximetry in the estimation
of hypoxemia and hyperoxemia in neonates
and children. Pulse oximetry test (with cut-off
point 91%) for detecting hypoxemia in neonates
(PaO2<35 mmHg) had a sensitivity of 81%, and
specicity of 79%. Pulse oximetry test with cut-off
point 95% for detecting hyperoxemia (PaO2>50)
had a sensitivity of 78% and specicity of 66%.
They concluded that the validity of pulse oximetry
in the detection of hypoxemia in neonates was
fairly good, but, it was not good enough to be used
in the estimation of hyperoxemia in neonates.10
In a study performed by Bakr et al., the
diagnostic value of fetal pulse oximetry in
comparison with fetal scalp blood gas in predicting
neonatal outcome was assessed and it was found
that these two tests were favorably comparable.
11
In another study, Carruthers et al. compared
arterial blood gas analysis with oxygen saturation
by pulse oximetry in the assessment of acute
asthma. They concluded that, in SpO2>92%,
respiratory failure is not probable and ABG
analysis is not necessary.12
In our study, the diagnostic value of pulse
oximetry in the detection of hyperoxemia was high.
None of the infants with SpO2 value greater than
95% were acidotic or hypercapnic. So, one can
rely on SpO
2
values for further management of
patients with high SpO
2
(>95%), which is sensitive
for hyperoxemia, and decrease FiO2 o r CPA P.
The validity of pulse oximetry in predicting
hypoxemia in our study was not good. In cases
that were hypoxemic according to blood gas,
the sensitivity of pulse oximetry was 75%. This
means that hypoxemia detection by contenting
oneself with pulse oximetry alone; one would
miss hypoxemia in 25% of cases.
The sensitivity of pulse oximetry in the
detection of acidosis in our study was only 40%.
In other words, the prediction of acidosis by pulse
oximetry alone will result in missing acidosis in
60% of cases.
In the present study, the sensitivity of pulse
oximetry in detecting hypercapnia was 50%. In
other words, based on SpO2 by simultaneous
pulse oximetry alone, half of hypercapnic patients
would be missed.
Therefore, in patients with SpO2 values lower
than 95%, arterial blood gas measurements
should be performed, and according to ABG
results, increasing FiO2 or CPAP, surfactant
administration, or initiation of mechanical
ventilation should be done.
Conclusion
Pulse oximetry in the detection of hyperoxemia in
neonates <1500 grams with RDS, who are treated
according to INSURE protocol, has a good validity
to appropriately manage the patient without blood
gas analysis. However, the validity of pulse oximetry
is not good enough to be used to detect acidosis,
hypercapnia, and hypoxemia.
Acknowledgment
We are grateful to the nursing staff of NICU at the
Afzalipour Medical Center for their help to carry out
this research. In addition, we appreciate the support
of the Deputy for Research Affairs at Kerman
University of Medical Sciences.
Conict of Interest: None declared.
References
1
Spitzer AR, Clark RH. Positive- Pressure
ventilation in the treatment of neonatal lung
disease. In: Goldsmith JP, Karotkin EH,
editors. Assisted ventilation of the neonate.
5th ed. USA: Saunders; 2011. p. 178-9.
2
Spitzer AR, Clark RH. Special ventilation
techniques II: Lung protective strategies. In:
Goldsmith JP, Karotkin EH, editors. Assisted
ventilation of the neonate. 5th ed. USA:
Saunders; 2011. p. 236-7.
3
Hamvas A. Pathophysiology and management
of RDS. In: Martin RJ, Fanaroff AA, Walsh
MC, editors. Fanaroff and Martin’s neonatal-
perinatal medicine. 9th ed. USA: Elsevier;
2011. p. 110 6 -15 .
4
Durand DJ, Mickas NA. Blood gases: Technical
aspects and interpretation. In: Goldsmith JP,
Karotkin EH, editors. Assisted ventilation of
the neonate. 5th ed. USA: Saunders; 2011.
p. 294-305.
5
Cardenas G, Finelli M, Harris C. Online Central
Access: Umbilical artery & vein cannulation
[Internet]. c2008 [cited 2012 May 1]. Available
from: http://www.crto.on.ca.
6
Solimano A, Littleford J, Ling E, et al. The
ACoRN, Acute Care of at-Risk Newborns.
3rd ed. Vancouver: McCallum Printing Group;
2010. p. 3-8.
7 Ambalavanan N, Schelonka RL, Carlo WA.
Ventilation strategies. In: Goldsmith JP,
Karotkin EH, editors. Assisted ventilation of
the neonate. 5th ed. USA: Saunders; 2011.
p. 265-76.
267
Pulse oximetr y versus ABG in RDS
Iran J Med Sci May 2015; Vol 40 No 3
8
Parikh R, Mathai A, Parikh S, Chandra
Sekhar G, Thomas R. Understanding and
using sensitivity, specicity and predictive
values. Indian J Ophtalmol. 2008;56:45-50.
doi: 10.4103/0301-4738.37595. PubMed
PMID: 18158403; PubMed Central PMCID:
PMC2636062.
9
Witting MD, Hsu S, Granja CA. The sensitivity
of room air pulse oximetry in detection of
hypercapnea. Am J Emerg Med. 2005;23:497-
500. Doi: 10.1016/j.ajem.20 04.12.006.
PubMed PMID: 16032619.
10
Ritonga SMA, Haksari EL, Suryantoro P.
Detection of hypoxemia and hyperoxemia
by pulse oximetry in neonates and children.
Pediatr Indones. 2008;48:346-9.
11
Bakr AF, Al-Abd M, Karkour T. Fetal Pulse
oximetry and neonatal outcome: A study in a
developing country. J perinatol. 2005;25:759-
62. doi: 10.1038/sj.jp.7211406. PubMed PMID:
16281048.
12 Carruthers DM, Harrison BD. Arterial blood
gas or oxygen saturation in the assessment
of acute asthma? Thorax. 1995;50:186-
8. doi: 10.1136/thx.50.2.186. PubMed
PMID: 7701461; PubMed Central PMCID:
PMC77014 61.
... Pulse oximeters are small devices that are fixed on fingers or on ear lobes. In this way, the so-called peripheral oxygenation (SpO 2 ) is measured [5,6]. ...
... Pulse oximetry signals (SpO 2 and SaO 2 ) given in [5,6] as time sequences are no exceptions to this pattern. Research [5] shows the possibility of determining hyperoxemia using pulse oximetry. ...
... Pulse oximetry signals (SpO 2 and SaO 2 ) given in [5,6] as time sequences are no exceptions to this pattern. Research [5] shows the possibility of determining hyperoxemia using pulse oximetry. The variability of pulse oximetry signals is mentioned but is not studied in detail. ...
Article
Full-text available
Monitoring of arterial blood saturation with oxygen (oxygenation) has gained special significance as a result of the COVID-19 pandemic. A new method for computer processing of saturation records (so-called SaO2 signals), based on the study of differentials (increments) from signals, was proposed. Finding a differential for a time series involves calculating the difference between the pairs of its adjacent elements. The differential is non-zero only if the elements in a pair are different. The study of differentials together with primary signals for a set of records (20 subjects) shows that the spectrum of observed levels of blood saturation is discrete and limited (from 2 to 10 levels). In addition, changes in saturation levels (switches) occur only between the nearest levels. New indicators of the variability of blood saturation were proposed. These are the frequencies of saturation level switches (event intensities) and the intervals between them. It was established that these indicators are described by statistical distributions of Poisson and Erlang, respectively. Comparison of new variability indicators with the most reliable statistical – inter-quartile range – indicates that the new indicators also provide for the division of the data set into three subgroups according to the magnitude of variability. This division is statistically significant at a confidence level of 0.99 in both approaches, however, the division into sub-groups is slightly different in these methods. It was shown that the proposed indicators of the variability of SaO2 signals are scale-invariant, that is, they do not depend on the length of observation interval. This is a consequence of the fractality of the positions of differentials in the observation interval. The established switch frequencies for subgroups in order of increasing variability are (0.06, 0.11, and 0.20) Hz. These frequencies are manifested on Fourier spectra of differentials of SaO2
... Arterial Blood Gas (ABG or SaO2) analysis is one of two used in clinical practice methods of blood oxygen saturation measuring [1]. It is direct but invasive, more laboring, and pricey than another oneperipheral Pulse Oximetry (SpO2). ...
Preprint
Full-text available
This report deals with arterial oxygen saturation (SaO2) for healthy adults. A comparably small data set (20 persons) holds 3-minute records of SaO2. The sample rate was 200 Hz. The charts have the looks of a "devil's stairs." A few (from 1 to 10) detectable oxygenation levels form the stair's treads, more or less long. "The risers" have two types (up and down), and all have virtually the same height, about 1 %. The inter-level shifts ( 0 to 42 switches per record) turned out a rare event at the actual sample rate. The number of switchings meets the Poisson distribution. There were found three visibly varied intensities for the switch-overs within the data set. Histograms also show the co-existing of no fewer than three subsets into the data set. The subsets differ by the intensity of switch-overs, amounts of possible levels, relative frequencies of most probable levels (modes), etcetera. In short, those all are diverse variability quantifiers. The higher variability subset has about 25 %, the lower one - 45%.
... Pulse oximetry is a simple and non-invasive method used to examine oxygen saturation (SpO2) in various parts of body [1]. Using pulse oximetry is effective in accelerating the weaning from mechanical ventilation and extubation and reduces the frequency of bleeding for analysis of arterial blood gases (ABG), because for the patients who just need checking for the O2 saturation, pulse oximetry could be a proper alternative [2,3]. Convenient use, speed and high accuracy in detection of hypoxia and continuous monitoring of patients are other features of pulse oximetry [3][4][5]. ...
Article
Full-text available
Background Heart surgery patients are more at risk of poor peripheral perfusion, and peripheral capillary oxygen saturation (SpO2) measurement is regular care for continuous analysis of blood oxygen saturation in these patients. With regard to controversial studies on accuracy of the current pulse oximetry probes and lack of data related to patients undergoing heart surgery, the present study was conducted to determine accuracy of pulse oximetry probes of finger, toe, forehead and earlobe in detection of oxygen saturation in patients admitted to intensive care units for coronary artery bypass surgery. Methods In this clinical trial, 67 patients were recruited based on convenience sampling method among those admitted to intensive care units for coronary artery bypass surgery. The SpO2 value was measured using finger, toe, forehead and earlobe probes and then compared with the standard value of arterial oxygen saturation (SaO2). Data were entered into STATA-11 software and analyzed using descriptive, inferential and Bland-Altman statistical analyses. Results Highest and lowest correlational mean values of SpO2 and SaO2 were related to finger and earlobe probes, respectively. The highest and lowest agreement of SpO2 and SaO2 were related to forehead and earlobe probes. Conclusion The SpO2 of earlobe probes due to lesser mean difference, more limited confidence level and higher agreement ration with SaO2 resulted by arterial blood gas (ABG) analysis had higher accuracy. Thus, it is suggested to use earlobe probes in patients admitted to the intensive care unit for coronary artery bypass surgery. Trial registration Registration of this trial protocol has been approved in Iranian Registry of Clinical Trials at 2018–03-19 with reference IRCT20100913004736N22. “Retrospectively registered.”
... Thus, continuous monitoring of the adequacy of breathing and oxygenation is necessary. Although pulse oximetry is widely used as a noninvasive method for continuous monitoring [3], oxygen saturation may be normal even if there is inadequate ventilation [4]. Previous studies have indicated that both low and high partial pressures of arterial carbon dioxide (PaCO 2) are associated with long-term morbidity in preterm and term infants [5]. ...
Article
Full-text available
Objective There are conflicting data regarding the use of end-tidal carbon dioxide (PetCO2) measurement in preterm infants. The aim of this study was to evaluate the effects of different dead space to tidal volume ratios (VD/VT) on the correlation between PetCO2 and arterial carbon dioxide pressure (PaCO2) in ventilated preterm infants with respiratory distress syndrome (RDS). Methods We enrolled ventilated preterm infants (with assist control mode or synchronous intermittent mandatory mode) with RDS who were treated with surfactant in this prospective study. Simultaneous PetCO2 and PaCO2 data pairs were obtained from ventilated neonates monitored using mainstream capnography. Data obtained before and after surfactant treatment were also analyzed. Results One-hundred and one PetCO2 and PaCO2 pairs from 34 neonates were analyzed. There was a moderate correlation between PetCO2 and PaCO2 values (r = 0.603, P < 0.01). The correlation was higher in the post-surfactant treatment group (r = 0.786, P < 0.01) than the pre-surfactant treatment group (r = 0.235). The values of PaCO2 and PetCO2 obtained based on the treatment stage of surfactant therapy were 42.4 ± 8.6 mmHg and 32.6 ± 7.2 mmHg, respectively, in pre-surfactant treatment group, and 37.8 ± 10.3 mmHg and 33.7 ± 9.3 mmHg, respectively, in the post-surfactant treatment group. Furthermore, we found a significant decrease in VD/VT in the post-surfactant treatment group when compared to the pre-surfactant treatment group (P = 0.003). Conclusions VD/VT decreased significantly after surfactant therapy and the correlation between PetCO2 and PaCO2 was higher after surfactant therapy in preterm infants with RDS.
Article
Full-text available
Background: Pediatric patients undergone mechanical ventilation after cardiac surgeries need rapid and reliable evaluation of their respiratory status. Monitoring of pulse oximetery and capnography as a surrogate, noninvasive measurement of arterial oxygen saturation (SaO2) and arterial carbon dioxide tension (PaCO2) becameimportant respiratory monitoring. Aim of the study: Was to compare pulse oximetery and capnography monitoring values with arterial blood gas parameters among postoperative pediatric cardiac surgeries. Subjects and Methods: A descriptive correlational design was adopted utilizing a convenience sample of 88 pediatric patients with cardiac anomalies who undergoing total repair surgeries, intubated for mechanical ventilation and had an indwelling arterial catheter as part of their evaluation in postoperative period were included in the study. Patient’s demographic, clinical relevant data as well as invasive and noninvasive monitoring of arterial oxygen saturation and arterial carbon dioxide partial pressure were utilized in data collection Results: Statistical analysis demonstrated mean differences between SaO2 and SpO2 (98.2 ±1.6% and 97.5±2.4% respectively; p=0.024)) as well as the mean difference between PaCO2 and PetCO2 (36.9±6.9mmHg and 33.1±7.7mmHg respectively; p <0.001). As well, a significant positive correlation was found between the mean SaO2 and SpO2 (r= 0.935, p<0.001) as well as, between the mean PaCO2 and PetCO2 (r= 0.930, p<0.001). Conclusion: It can be concluded that SpO2 and EtCO2 values are positively correlated with SaO2 and PaCO2 in mechanically ventilated pediatric patients after cardiac surgeries. Therefore SpO2 and EtCO2 monitoring can be used as a continuous, non-invasive predictor for SaO2 and PaCO2.
Article
Full-text available
Objective To assess the validity of pulse oximetry for detecting hypoxemia and hyperoxemia in neonates and children. Methods This was a diagnostic test study conducted in Neonatal Intensive Care Unit (NICU). The subjects of the study were neonates of 0-28 days and children aged one month to 14 years in Pediatric Intensive Care Unit (PICU) of Sardjito Hospital. Hypoxemia and hyperoxemia were obtained by using NellcorR pulse oximetry, at the same time as blood gas analysis was obtained as a gold standard. Results There were 240 neonates and 268 children enrolled in this study. Pulse oximetry test with cut-off point 91% (neonates) and 90% (children) for detecting hypoxemia had sensitivity of 81% and 80%, specificity of 79% and 95% positive predictive value of 54% and 82% negative predictive value of93% and 93% positive likelihood ratio of 3. 79 and 14.20 and negative likelihood ratio of0.25 and 0.24. Pulse oximetry test with cut-off point 95% (neonates) and 97% (children) for detecting hyperoxemia had sensitivity of 7 8% and 81% specificity of 66% and 79%, positive predictive value of77% and 76% negative predictive value of 66% and 83% positive likelihood ratio of 2.26 and 3.84 and negative likelihood ratio of 0.34 and 0.24. Conclusion Pulse oximetry has a fairly good validity in detecting hypoxemia in neonates and children and in detecting hyperoxemia in children to however, it is not good enough to be used to detect hyperoxemia in neonates.
Article
Full-text available
In this article, we have discussed the basic knowledge to calculate sensitivity, specificity, positive predictive value and negative predictive value. We have discussed the advantage and limitations of these measures and have provided how we should use these measures in our day-to-day clinical practice. We also have illustrated how to calculate sensitivity and specificity while combining two tests and how to use these results for our patients in day-to-day practice.
Article
A study was undertaken to determine if arterial blood gas estimation is always necessary in the assessment of patients presenting to hospital with acute severe asthma, or whether oxygen saturation as measured by pulse oximetry is a reliable screening test for predicting those in respiratory failure. A prospective study was conducted in a specialist respiratory medical unit. Arterial blood gas tensions and pulse oximetry were measured in 89 consecutive patients admitted with acute severe asthma. Respiratory failure was defined as PaO2 < 8.0 kPa or PaCO2 > 6 kPa. When oxygen saturation was 92% or higher (72 patients) respiratory failure was found in three (4.2%) cases. In the 82 patients with a saturation of 90% or higher six patients (7.3%) had respiratory failure. In the initial assessment of acute severe asthma an oxygen saturation of > 92% suggests that respiratory failure is unlikely and therefore arterial blood gas measurement is unnecessary. This study is only relevant to the assessment of asthmatic patients at presentation. Other parameters of severity must be continually assessed in all asthmatic patients admitted to hospital irrespective of the initial SaO2, and blood gases measured when clinically indicated.
Article
To estimate the sensitivity of room-air pulse oximetry in the detection of moderate hypercapnia. In this retrospective case-control study, charts were reviewed from patients with and without moderate hypercapnia (Pa co 2 >50 mm Hg), as determined by analysis of arterial blood gas samples obtained in the ED. Test characteristics (sensitivity, specificity, and likelihood ratios [LR) for room-air pulse oximetry < or = 96% to detect hypercapnia were calculated, as were confidence intervals. A total of 349 charts were eligible for abstraction-92 cases and 257 controls. A room-air pulse oximetry reading < or = 96% detected 88 of 92 cases of hypercapnia. Test characteristics were as follows (with 95% confidence interval): sensitivity, 0.96 (0.89-0.99); specificity, 0.39 (0.33-0.45), LR of a room-air pulse oximetry value >96%, 0.1 (0.04-0.3); and LR of a room-air pulse oximetry value < or = 96%, 1.6 (1.4-1.7). Room-air pulse oximetry detects moderate hypercapnia with high sensitivity.
Article
The aim of this cohort, prospective study was to compare the diagnostic value of intrapartum fetal pulse oximetry (FPO) with that of fetal scalp blood gas (FSBG) for an abnormal neonatal outcome in cases with abnormal fetal heart rate (FHR) tracings. Fetal oxygen saturation was continuously monitored with Nellcor N-400 FPO during labor. Simultaneous FSBG determinations were obtained. The results were analyzed in relation to umbilical arterial cord blood pH and neonatal outcome. Studied FPO cutoff levels were 30 and 40% hemoglobin saturation and that of FSBG pH was 7.2. During the study, there were 9825 deliveries; 415 had abnormal FHR. Only 150 fulfilled the whole screening panel. When the outcome variable was umbilical arterial pH, the positive predictive values of the three methods (FPO30, FPO40, FSBG) were 57, 61 and 65% and the negative predictive values were 43, 39 and 35% respectively. The sensitivity of FPO30 was highest (75%). Considering abnormal neonatal outcome, again the sensitivity was also highest for FPO30 (89%). The sensitivity of FSBG was 82%. The specificity of the three methods were 53, 49 and 38% respectively. The diagnostic value of intrapartum FPO compares favorably with FSBG. FPO seems to be a reliable and less invasive tool and may decrease unnecessary interventions and unnecessary fetal scalp blood sampling in cases of suspected fetal distress. The FPO cutoff of 30% saturation defined by previous studies appears to be appropriate.
Online Central Access: Umbilical artery & vein cannulation
  • G Cardenas
  • M Finelli
  • C Harris
Cardenas G, Finelli M, Harris C. Online Central Access: Umbilical artery & vein cannulation [Internet]. c2008 [cited 2012 May 1]. Available from: http://www.crto.on.ca.