In view of in vivo surgical smoke studies a difference-frequency-generation (DFG) laser spectrometer (spectral range 2900-3144 cm(-1)) and a Fourier-transform infrared (FTIR) spectrometer were employed for infrared absorption spectroscopy. The chemical composition of smoke produced in vitro with an electroknife by cauterization of different animal tissues in different atmospheres was investigated. Average concentrations derived are: water vapor (0.87%), methane (20 ppm), ethane (4.8 ppm), ethene (17 ppm), carbon monoxide (190 ppm), nitric oxide (25 ppm), nitrous oxide (40 ppm), ethyne (50 ppm) and hydrogen cyanide (25 ppm). No correlation between smoke composition and the atmosphere or the kind of cauterized tissue was found.
Content may be subject to copyright.
Sensors 2010, 10, 2694-2708; doi:10.3390/s100402694
sensors
ISSN 1424-8220
www.mdpi.com/journal/sensors
Article
Infrared Spectroscopy on Smoke Produced by Cauterization of
Animal Tissue
Michele Gianella and Markus W. Sigrist *
Laser Spectroscopy and Sensing Lab, Institute for Quantum Electronics, ETH Zurich, Schafmattstr. 16,
8093 Zurich, Switzerland; E-Mail: michele.gianella@phys.ethz.ch
* Author to whom correspondence should be addressed; E-Mail: sigristm@phys.ethz.ch;
Tel.: +41-44-633-2289; Fax: +41-44-633-1230.
Received: 24 January 2010; in revised form: 20 February 2010 / Accepted: 19 March 2010 /
Published: 26 March 2010
Abstract: In view of in vivo surgical smoke studies a difference-frequency-generation
(DFG) laser spectrometer (spectral range 2900–3144 cm-1) and a Fourier-transform infrared
(FTIR) spectrometer were employed for infrared absorption spectroscopy. The chemical
composition of smoke produced in vitro with an electroknife by cauterization of different
animal tissues in different atmospheres was investigated. Average concentrations derived
are: water vapor (0.87%), methane (20 ppm), ethane (4.8 ppm), ethene (17 ppm), carbon
monoxide (190 ppm), nitric oxide (25 ppm), nitrous oxide (40 ppm), ethyne (50 ppm) and
hydrogen cyanide (25 ppm). No correlation between smoke composition and the atmosphere
or the kind of cauterized tissue was found.
Keywords: infrared laser spectroscopy; surgical smoke; in vitro; difference frequency
generation
Classification: PACS; 42.62.Fi; 33.20.Ea; 87.64.km
1. Introduction
Surgical smoke is a generic term that describes gases, vapors and aerosols produced in surgery with
lasers, high frequency electric knives and vessel sealing devices, as well as ultrasonic scalpels. These
tools are employed to cut, coagulate, ablate, dissect or fulgurate biological tissues. Heat is usually
OPEN ACCESS
Sensors 2010, 10
2695
produced and (surgical) smoke develops. There is concern about the health risks posed by surgical
smoke, both to the patient and to the medical staff who is exposed on a day-to-day basis [1–4]. A
related issue concerns smoke produced by cooking meat, to which, e.g., kitchen personnel in
restaurants is exposed daily.
So far, the biological [5,6], particulate [7–9] and chemical [9–14] compositions of surgical smoke
have been studied; an excellent review is given in Ref. [15]. With respect to the chemical composition
there is a general lack of quantitative data. Also of interest are possible correlations between the smoke
composition and the employed tool, the kind of tissue and the insufflant gas.
As a preliminary step towards the investigation of surgical smoke produced in vivo, we present
results on the chemical gaseous composition of smoke produced in vitro in our lab with animal meat in
a specially designed cell and analyzed with a difference-frequency-generation (DFG) based mid-
infrared laser spectrometer. Additionally, Fourier-transform infrared (FTIR) spectra were recorded for
a few samples. For all samples a quantitative analysis was carried out with a specially developed
algorithm [16]. Possible correlations between composition and insufflant as well as between
composition and tissue were investigated.
2. Experimental Section
2.1. Laser Spectrometer and FTIR Spectrometer
The spectrometer used in this study (Figure 1) is a difference-frequency-generation (DFG) based
laser spectrometer. It has been used for several studies [16–18] but since some modifications were
made it is briefly described here.
Figure 1. Difference-frequency-generation (DFG) based laser spectrometer used in this
study. Signal beam: fiber-coupled ECDL (external cavity diode laser); PM fiber:
polarization maintaining fiber; pump beam: Nd:YAG laser. !/2 and !/4 plate: half-wave
and quarter-wave plate.










!

"
#
!
$ %
%
The pump laser is a diode-pumped, passively Q-switched Nd:YAG laser (InnoLight, Model M800,
Germany), with a repetition rate of about 8 kHz, 6 ns pulse duration, 300 mW average power, >5 kW
Sensors 2010, 10
2696
peak power and 1064.555 nm wavelength. The signal laser is a cw external cavity diode laser (ECDL,
Santec Corp., Model TSL-210, Japan) tunable between 1520 and 1600 nm with a minimum of 5 mW
of power over the entire range. The non-linear material is a periodically poled lithium niobate (PPLN)
crystal, 0.5 mm thick, 10 mm wide and 50 mm long with 8 different poling periods of which only the
29.9 !m period was used in this study. It is placed in a crystal oven (Super Optronics, USA) within a
Teflon housing. Quasi phase matching (QPM) is achieved by temperature tuning. The full mid-IR
range of the idler beam attainable with the current signal laser (and by using both the 29.9 and 29.5 !m
poling periods of the PPLN) spans from 2815 to 3144 cm-1 (329 cm-1) and could be extended by using
a signal laser at wavelengths below 1520 nm. The average power of the (pulsed) idler depends on the
wavelength and is typically 100-150 !W. The line width of 155 MHz is given by the pulsed Nd:YAG
laser. The homemade multipass cell has been described in detail elsewhere [19]. It is vacuum-tight and
can be operated up to high temperatures (T " 720 K), although in this study we never exceeded 120 °C
(393 K). The idler leaves the cell through the entry window after a number of passes that can be altered
by changing the distance between the two mirrors and is then reflected into the transmission detector
(Vigo Systems SA, Model PDI-2TE-4/VPDC-0.1i, Poland). The total optical path length inside the cell
is 34.5 m.
The preamps of the two detectors are connected to the two input channels of an A/D card (GaGe
CS14100, USA) via BNC cables. The quantity of interest is the detector signal ratio (DSR) Q,
This quantity is not yet the transmittance of the sample because the denominator only represents a
part of the incident power, and because the signal of the transmission detector includes losses due to
the multiple reflections in the multipass cell. Additionally, fringes caused by the uncoated BaF2
windows of the detectors can be observed in the DSR. To obtain the transmittance of a sample the
DSR Q0 of a non-absorbing gas or of the buffer gas is needed. The transmittance T can then be
computed via
Spectra are recorded by setting a temperature ramp on the PPLN crystal (2 °C/min) and measuring
the DSR at predefined times after an initial PPLN temperature has been crossed. Typically we record
1.5 data points per second and the temperature ramp runs from 40 °C to 173 °C. At 2 °C/min a
spectrum (from 2900 to 3144 cm-1) takes 67 minutes to complete and consists of about 6000 points.
Our data shows that this method provides better reproducibility in a considerably shorter time
compared to tuning the wavelength to a desired value and then setting an appropriate PPLN temperature.
In addition to the laser spectrometer measurements, transmission spectra of a few samples in a
White-type cell with 4 m of total optical path length were recorded with an FTIR spectrometer (Bruker
Optics Inc., Model IFS 66v, USA).
(1)
(2)
Sensors 2010, 10
2697
2.2. Preparation of the Smoke Samples
The investigation of in vitro samples offers some advantages over in vivo samples such as:
production is less time-consuming, freedom to alter any variable (e.g., change of atmosphere, type of
tissue), possibility to increase concentrations by generating more smoke and so on. We designed and
built a small cell that allows smoke production in a controlled atmosphere (Figure 2). The cell consists
of two aluminum plates (14 " 14 " 0.5 cm), two neoprene rings (i.d. = 11 cm, o.d. = 12 cm, thickness
3 mm) and a plexi-glass cylinder (i.d. = 11 cm, o.d. = 12 cm, height 5.5 cm). A 0.5-mm thick neoprene
layer is compressed between the upper aluminum plate and an aluminum ring screwed onto the plate.
Two concentric holes (a 2-cm hole in the aluminum plate and a 5-mm hole in the neoprene layer)
allow the insertion of a medical monopolar electroknife (Coagulasem, Ets. Dolley SA, France). All the
smoke samples were produced by slowly moving the tip of the electroknife across the surface of pieces
of fresh animal meat. The meat was weighed before it was put into the smoke production cell and
again after smoke production to determine the loss of mass caused by the cauterization.
Figure 2. Photograph of the setup for generating and collecting in vitro smoke samples.
The samples were prepared by cauterizing fresh animal meat with an electroknife inside a
cell that allows smoke production in a specific atmosphere (e.g., CO2, N2, synthetic air).
A gas of choice (CO2, N2, synthetic air) was pumped into the cell through the gas inlet, while the
produced smoke was evacuated through the gas outlet. A micro-glass-fiber particle filter (Infiltec
GmbH, housing: SL 215.401, filter elements: 25-64-30, Germany) was connected through a stainless
steel tube to the outlet of the cell on one side and to a glass bottle on the other side. The housing of the
particle filter and all the tubings were heated to 150 °C to prevent condensation of the smoke. The
glass bottle was filled with smoke by either creating a vacuum and then aspirating the generated smoke
by opening a valve, or by letting the smoke flow through the bottle for some time and then closing the
valves mounted on the cap of the bottle.
Sensors 2010, 10
2698
Once the smoke had been stored inside the glass bottle it had to be transferred into the multipass
cell. The bottle was placed into a water bath and connected to the evacuated multipass cell. The water
was then heated until it boiled while small amounts of gas were transferred into the multipass cell
whenever the pressure inside the bottle exceeded 1.2 bar. Once the boiling point was reached the valve
between the bottle and the multipass cell was opened completely and the pressure equalized at
typically 350-400 mbar. For measurements carried out at higher pressure (e.g., 930 mbar) the same gas
used for the production of the smoke was added into the multipass cell directly from the gas bottle.
We investigated a total of 15 in vitro smoke samples all produced in a carbon dioxide (purity 2.3)
atmosphere except one which was produced in nitrogen (purity 5.0), and one in synthetic air (80%
nitrogen and 20% oxygen, purity unknown). The samples were produced by cauterizing fresh animal
meat with a monopolar high frequency electroknife (a type of electroknife also used in human
surgery). The selection of a CO2 atmosphere occurred in view of in vivo studies in laparoscopy where
CO2 is the common insufflant gas [20]. Spectra of all samples were recorded with our DFG laser
spectrometer; two of the samples were also measured with an FTIR spectrometer.
3. Results and Discussion
3.1. Quantitative Analysis
The spectrum of a typical sample is shown in Figure 3a. Spectra of the four substances that were
found in all samples are shown in Figures 3b-e. All the narrow absorption features in every
measurement can be accounted for by water vapor (H2O), methane (CH4), ethane (C2H6) and ethene
(C2H4) (ethylene), as can be seen in Figure 3f. The spectra in Figure 3b-e were retrieved from the
PNNL database [21]. Because of the different resolution and pressure of the database spectra (0.1 cm-1
and 1013 mbar) and of our measurements (0.05 cm-1 and 900–960 mbar) some artifacts appear in
Figure 3f–especially in places where several absorption lines lie closely together–due to discrepancies
in the shape of the absorption lines. The concentrations for methane, ethane and ethene lie in the ppm
range and water vapor is in the percent range. There is a very broad (>100 cm-1) absorption from
2,900 cm-1 to above 3,000 cm-1 that could not be identified unequivocally and which could be the
result of the cumulative absorption of several substances or of scattering on particles smaller than
0.1 !m, if at all (since a particle filter was used). Acyclic alkanes such as pentane, hexane, etc., and
alcohols like pentanol, hexanol, etc. manifest broad absorptions between 2900 and 3000 cm-1 that
resemble the observed absorption feature. In the region above 3000 cm-1 there is a weaker and
approximately constant absorption that is not an instrumental offset. Again, there is not enough
information to associate this absorption to one or to a set of substances.
Table 1 lists for each measured sample which tissue and atmosphere were used for the smoke
production, the loss of mass following cauterization, the pressure and temperature at which the
spectrum was recorded, whether or not a particle filter was used, the concentrations of the four
substances found in all samples, the algorithm used for the quantitative analysis (see below), and
whether a Fourier-transform infrared (FTIR) spectrum was also recorded or not.
Sensors 2010, 10
2699
Figure 3. (a) Typical spectrum of an in vitro sample and its four main components:
(b) water, (c) methane, (d) ethane and (e) ethene. (f) When the four spectra (b)-(e) are
subtracted from (a) a residual spectrum remains.










     












 

 !
"#$
%
&'()**+*,
Concentrations are given in ppm = ppmV = !mol/mol. Production rates such as mol/s or mol/g
would be more useful but require the total gas volume to be measured; this was not done as in many
cases the smoke was produced in a continuous gas flow, of which only a fraction was sampled and
measured.
The quantitative analysis of the recorded spectra was performed in most cases with the improved
mix-match algorithm [16], which is based on the quantitative spectral database PNNL [21] and
includes an iterative rating procedure (Alg. 1 in Table 1). The cited database consists of 360 FTIR
spectra measured at ambient pressure and 25 °C: it therefore cannot be used for spectra recorded at
much lower pressure (pressure broadening). For samples measured at low pressure (p " 900 mbar) a
simpler alternative to the mix-match algorithm was used (Alg. 2). It should be noted, however, that
although the mix-match algorithm did not provide accurate concentrations for those samples, it still
identified methane, ethane, ethene and water vapor correctly. The spectra of water, methane, ethane
and ethene (at 1 ppm and 34.5 m path length) were computed with the Molspec software (Molspec III,
Laser Components GmbH) using the HITRAN 2004 [22] database for p = 200 mbar, T = 100°C (for
the sample A12) and p = 300 mbar, T = 120 °C (for the samples A13 and A14). The spectra were
interpolated at the wavelengths at which each of the three samples was measured. The interpolated
spectra were inserted into a matrix X, one per row, plus an additional row of zeros at the end, yielding
a 5 " k matrix, where k is the number of wavelength points (typically about 6000). The principal
components V and the scores U = X*V of X* were computed, where X* denotes X without the last row.
The purpose of the row of zeros in X is to ensure that the zero spectrum belongs to the space spanned
by the four principal component vectors (columns of V). Assuming the linear relationship C = UR
between the scores U and the concentrations C (cij = concentration of substance j in spectrum i), the
matrix of regression coefficients R was computed in a least-square way with
(3)
Sensors 2010, 10
2700
Table 1. Overview of the 15 smoke samples (A05-A20) investigated with our DFG laser
spectrometer. The atmosphere (atm.) is either carbon dioxide, nitrogen or synthetic air
(s.a.). #m: computed loss of mass. The spectra of the smoke samples were recorded at the
pressure p and temperature T. The concentrations of methane, ethane, ethene and water
vapor were computed by using one of two algorithms (Alg., 1 = improved mix-match [16],
2 = PCR (see text)).
Concentrations
ID
Tissue
Atm.
#m
Filt.a
p
T
CH4
C2H6
C2H4
H2O
Alg.
FTb
mg
mbar
°C
ppm
ppm
ppm
%
A05
liver, rabbit
CO2
no
957
25
15
3.5
18
0.66
1
A06
liver, rabbit
CO2
yes
956
25
14
2.8
13
0.69
1
A07
liver, rabbit
s.a.
316.7
yes
960
25
19
4.0
20
0.99
1
A08
heart, beef
CO2
yes
960
25
6.4
0.7
7.5
0.37
1
A09
heart, veal
CO2
394.7
yes
960
25
11
2.1
7.1
0.74
1
A10
heart, beef
CO2
769.4
yes
967
100
4.2
0.8
9.3
1.5
1
A11
heart, pig
CO2
yes
900
100
13
3.0
19
1.3
1
A12
loin, beef
CO2
292.6
no
200
100
12
5.5
~15
0.35
2
A13
heart, veal
CO2
269.1
yes
300
120
14
3.8
"10
0.24
2
A14
liver, beef
CO2
556.1
yes
300
120
17
6.3
"10
1.2
2
A15
liver, beef
CO2
352.8
yes
930
25
27
6.5
25
1.7
1
A16
liver, pig
N2
669.0
yes
930
25
36
9.5
37
2.3
1
A17
pig
CO2
yes
930
25
29
6.1
17
1.1
1
*
A19
kidney, pig
CO2
686.3
yes
930
25
41
11
32
0.27
1
*
A20
kidney, pig
CO2
yes
930
25
34
6.5
11
0.49
1
Average
20
4.8
17
0.87
Min.-Max.
4.2-41
0.7-11
7.1-37
0.15-2.3
a Smoke filtered with a particle filter retaining particles #0.1 !m.
b A Fourier-transform infrared (FTIR) spectrum was measured for entries marked with *.
The last equality in (3) follows from the fact that since the four spectra in X* (and therefore U) are
of pure substances, the matrix C is the 4 " 4 identity matrix (the units of C are ppm). The
concentrations of water, methane, ethane and ethene of samples A12, A13 and A14 were then
computed through
where s = A12, A13 or A14, and Ms is the measured spectrum. The concentrations of samples A12,
A13 and A14 have a lower accuracy–especially for ethene which has very weak absorption lines that
are often in coincidence with stronger water or methane lines–compared to the samples measured at
higher pressure. This is because the width of the absorption lines is smaller (typical line width at
300 mbar: 0.1 cm-1) but the resolution of the spectrometer was the same (0.05 cm-1) as for the
measurements carried out at higher pressure. The purpose of the measurements carried out at higher
temperature (samples A10-A14) was to increase the sensitivity to low vapor pressure species that
might condensate on the inner walls of the multipass cell or on the mirrors. Every stage between the
(4)
Sensors 2010, 10
2701
smoke production cell (Figure 2) and the multipass cell was heated in every measurement even though
for only five of them the multipass cell was heated as well. No additional substance could be identified
in the spectra of samples A10-A14.
Uncertainties of concentrations in Table 1 can be estimated by repeatedly measuring the same
sample. The uncertainties (1 standard deviation) computed for one specific sample are: for methane,
ethane and ethene about 0.5 ppm, for water about 0.025%. These values are representative for all the
samples due to the similarity of their spectra and their concentrations.
3.2. Correlation Smoke-Tissue and Smoke-Atmosphere
Water vapor, methane, ethane and ethene are the only substances that could be identified
unequivocally in each measurement and they were found in all 15 samples. Their concentrations
depend on the amount of cauterized tissue (mass loss), on the volume of gas within which the smoke
was diluted (dilution effect) and possibly on the atmosphere and tissue. Unfortunately, mass loss and
gas volume data is not available for all measurements. We can, however, normalize the concentrations
of each measurement with respect to the computed methane concentration for that spectrum and
compare samples that were produced with the same tissue and atmosphere. This procedure cancels the
dilution effect and might cancel the dependence of the concentrations on the mass loss, assuming all
concentrations manifest the same dependence on the latter.
Figure 4a shows a comparison between the normalized concentrations of four pairs of smoke
samples all produced in carbon dioxide. Notice that the concentration of each substance is plotted on a
different scale, given by kx, for better visibility. To obtain the normalized concentration of substance x
multiply the value $ plotted on the ordinate with kx = 0.2 for x = ethane, 1 for x = ethene, 800 for x = water.
A correlation between tissue type and smoke composition cannot be established: for example, sample
A13 and A19 have very similar concentrations even though different tissue was cauterized, but sample
A8 and A10 have very different concentrations although the same tissue was used. The variance of
concentrations of samples produced with the same tissue is similar to the variance of concentrations of
samples produced with different tissue, so that no significant tissue-smoke composition dependence
can be inferred. In Figure 4b the normalized concentrations of six samples are compared, four
produced in a carbon dioxide atmosphere, one in nitrogen and one in synthetic air. The atmosphere
used does not seem to have any effect on the normalized concentrations of water, ethane and ethene;
this is in agreement with a previous study conducted in vitro on animal meat [23].
The absolute concentrations of sample A16 (nitrogen atmosphere) are comparable to those of
sample A6 and A15 (carbon dioxide atmosphere), see Figure 4b and Table 1. This implies that the
carbon atoms in methane, ethane and ethene are not provided by the carbon dioxide atmosphere but by
the tissue itself. Figure 4c shows the normalized concentrations of a sample that was filtered with a
particle filter (A6) and one that was unfiltered (A5). The difference between the two samples is not
significant; the filter was therefore used for most smoke samples to prevent the multipass cell and the
mirrors inside it from becoming contaminated with soot.
As indicated in Table 1 FTIR spectra of samples A17 and A19 were recorded in addition. Aside
from water, methane, ethane and ethene some additional substances that do not absorb at all or not
strongly enough in the range 2900 to 3144 cm-1 of the DFG laser spectrometer could be identified
Sensors 2010, 10
2702
(Figure 5): nitric oxide (NO) 25 ppm, carbon monoxide (CO) 200 ppm, nitrous oxide (N2O) 50 ppm,
ethyne (acetylene, C2H2) 45 ppm, and hydrogen cyanide (HCN) 30 ppm. Carbon dioxide was used as
atmosphere for the production of both samples, its presence is therefore obvious. Figure 5a shows the
measured FTIR spectrum of sample A17 between 900 and 4000 cm-1. Water vapor and carbon dioxide
absorptions are strong in the intervals 1320–1910 cm-1 (water), 2240–2380 cm-1 (carbon dioxide) and
3530–3960 cm-1 (both). The inset displays the region accessible with our DFG laser spectrometer and
shows a comparison between measurements of the same sample obtained with the laser spectrometer
and the FTIR spectrometer, respectively. Many water absorption lines are missing in the FTIR
spectrum because of its lower resolution (0.125 cm-1 versus 0.05 cm-1 of the laser spectrometer).
Figures 5b-d show magnified regions of the measured FTIR and spectra from PNNL [21] of identified
components with concentrations computed with the improved mix-match algorithm [16]. The analysis
of the FTIR spectrum of sample A19 yielded very similar results.
Figure 4. Concentrations of ethane, ethene and water vapor normalized with respect to the
methane concentration in the corresponding sample. (a) Comparison of normalized
concentrations for samples produced in carbon dioxide by cauterizing different tissues. (b)
Normalized concentrations for samples produced in different atmospheres. (c) Normalized
concentrations for filtered/unfiltered smoke samples. Each substance's normalized
concentration is plotted on a different scale for better visibility. To obtain the actual
normalized concentration of a substance multiply $ with the corresponding kx.
  

  











 
 
!""
#$% 
&
   $
 &  "$


"&

'(
#


 )
$"



*

& +
& +
+
+
&
&*
&
&
&
&*
&
&
Sensors 2010, 10
2703
Figure 5. (a) Fourier-transform infrared (FTIR) spectrum of in vitro smoke sample A17
and magnification of the spectral region accessible with our DFG laser spectrometer (inset).
Strong water and carbon dioxide absorptions limit the sensitivity of the spectrometer in the
spectral ranges 1320–1910 cm-1, 2240–2380 cm-1 and 3530–3960 cm-1. (b) Nitric oxide
(25 ppm) is visible at 1800–2000 cm-1. (c) Carbon monoxide (200 ppm) and nitrous oxide
(50 ppm) can be seen at 20502300 cm-1. (d) Ethyne (45 ppm) and hydrogen cyanide
(30 ppm) have overlapping absorption branches at 3200-3500 cm-1. Nitrous oxide is also
visible.
 







    
 










      







     
!"#
$ %
&&
"#
$ %
&&
'()*
+#&!
      





  


,'-
&!!
'()*
'()*
'()*

+#&!
!"#
$ %
&&

!"#
$ %
&&
./#0/#
$ %
&&
1!2/!/
$ %
&&
'()*
+#&!
 
  

 

Sensors 2010, 10
2704
3.3. Detection Thresholds for Undetected Species
Across several previous studies a few hundred substances have been reported to be present in
surgical smoke [9-11,15]. It is somewhat surprising that apart from the substances mentioned in the
previous sections the presence of no other compound could be confirmed. In Table 2 we give the
detection thresholds for some selected components that were identified in one or more previous studies
and, for comparison, the detection thresholds for our DFG and FTIR spectrometer for the substances
found in the present study. The presence of one or more interfering species can severely increase the
detection thresholds; substances with several narrow absorption lines are only slightly affected (e.g.,
methane), but if large and unstructured absorption features are present the detection threshold can
considerably increase. An example is toluene, which has roughly one tenth of the absorption cross-
section of methane but a 330 times higher detection limit. The minimum detectable concentration (cmin)
for a total path length of L = 34.5 m (L = 4 m for the FTIR spectra) for a given substance x was
determined as follows: (i) the absorbance of substance x at a concentration cmin is greater than or equal
to Amin = 0.02 at one wavelength (at least) in the range 2900–3144 cm-1 (9004000 cm-1 for the FTIR
spectra), and (ii) the (absolute value of the) derivative of the absorbance spectrum (i.e., its slope) is
greater than or equal to Dmin = 0.1/cm-1 at one wavelength (at least) in the range 2900–3144 cm-1 (900-
4000 cm-1 for the FTIR spectra). Amin is the smallest measurable absorbance: it depends on the noise
and the reproducibility of the measurements. By measuring a non-absorbing gas (e.g., nitrogen) several
times we obtain slightly different spectra because of noise and drifts. From the variance of these
spectra one can estimate Amin. Since some unidentified absorption features were present in the spectra
of the measured smoke samples, additional broad and unstructured absorptions are hard to notice
especially if they lie in the range 2900–3000 cm-1. However, if the slope of the spectrum is sufficiently
large, i.e., the absorption is sufficiently "sharp", they are visible. For small molecules with narrow
absorption lines condition (ii) is generally automatically fulfilled if condition (i) is true; hence, the
minimum measurable absorbance Amin determines the detection threshold cmin. The detection
thresholds of larger molecules with broad absorptions (e.g., toluene, benzene) are quite high. The 8-
hour time-averaged recommended exposure limits (REL) in Switzerland [24] are also given in Table 2.
A rule of thumb for a sensor is that it should be at least a factor 10 more sensitive than the REL of the
substance(s) under investigation. This condition is fulfilled for the harmless methane, ethane and
ethene but not for the potentially toxic other compounds listed in Table 2.
More sensitive detection schemes or preconcentration techniques like cooling traps and
thermal/solvent desorption tubes are necessary to bring the detection limits for some of the species to
useful values (ppm range). Appropriate methods are currently under evaluation.
4. Conclusions
We present results about several in vitro samples of surgical smoke. Mid-infrared absorption spectra
were recorded with a difference-frequency-generation (DFG) based laser spectrometer and, in a few
cases, with a Fourier-transform infrared (FTIR) spectrometer. Quantitative information about the
chemical composition of the samples was obtained from the absorption spectra with the improved mix-
match algorithm [16]. The qualitative composition of all the samples we studied was very similar:
Sensors 2010, 10
2705
water, methane, ethane and ethene were always found, plus additional broad absorptions that could not
be identified because they are not characteristic enough. Measurements at 100 °C and 120 °C did not
reveal any additional substances that were not visible at room temperature.
Table 2. Minimum measurable concentration (cmin) with our DFG laser spectrometer and
with the FTIR spectrometer (value in parentheses) for a few selected compounds that have
been reported to be present in surgical smoke and compounds that were detected in at least
one sample in this study. A dash in the minimum concentration column means that the
substance has no absorption between 2900 and 3144 cm-1 and can therefore not be detected
with our DFG laser spectrometer. Recommended exposure limits (REL, 8-hour
time-average) in Switzerland [24] are also indicated.
cmin
REL
cmin
REL
Substance
ppm
ppm
Substance
ppm
ppm
Not detected
Toluene
100
50
Benzene
20
0.5
p-Xylene
140
100
o-Xylene
25
100
Styrene
70
20
Ethyl benzene
40
100
Benzaldehyde
50
N/A
Benzonitrile
25
N/A
Detected
Carbon monoxide
— (2.5)
30
Hydrogen
cyanide
— (5.4)
1.9
Ethyne
— (4.7)
1000
Nitric oxide
— (7.5)
25
Nitrous oxide
2300 (1.2)
100
Water
120
Methane
0.3
10000
Ethane
0.3
10000
Ethene
2
10000
We investigated the relationships between smoke composition and kind of cauterized tissue on the
one hand, and the atmosphere within which the smoke was produced on the other hand. In general, the
variances of the concentrations in samples produced under the same conditions are comparable or
larger than the variances of the concentrations of samples produced under different conditions
(Figure 4). Hence, there appeared to be neither a correlation between smoke composition and
atmosphere, nor between smoke composition and kind of tissue. The presence of methane, ethane and
ethene in smoke produced in a nitrogen atmosphere proves that the carbon atoms originate from the
tissue and not from the atmosphere.
The two FTIR spectra that were measured provided additional information about the composition of
the smoke samples: ethyne, nitric oxide, nitrous oxide and hydrogen cyanide were detected with
concentrations of a few tens of ppm; carbon monoxide was measured at approximately 200 ppm. Since
the two samples for which the FTIR spectra were recorded were produced in a carbon dioxide
atmosphere, the nitrogen atoms of nitrous and nitric oxide and of hydrogen cyanide are probably
provided by the cauterized tissue (a nitrogen contamination of the carbon dioxide (purity 2.3) bottle
cannot be excluded).
Our study is the first to employ exclusively infrared (laser and FTIR) spectroscopy to determine the
quantitative chemical composition of surgical smoke produced in vitro. While the FTIR spectrometer
Sensors 2010, 10
2706
yields a broad wavelength range, the DFG laser spectrometer enables a better spectral resolution and
hence higher sensitivity yet only for a limited spectral range.
In most previous studies gas chromatography (GC), sometimes coupled with mass spectrometry
(GC-MS), was the analytical tool of choice. With such techniques a few hundred chemical species
have been identified (though only qualitatively) in surgical smoke (mostly in vitro and some in vivo)
[10,11,13,15]. Most of them could not be confirmed here. For many, the currently used spectral range
of the DFG laser spectrometer (2900-3144 cm-1) is not the most appropriate: although most substances
of interest have some absorption in the given range (fundamental C–H stretch), the detection
thresholds can be too high.
The sensitivity could be increased with a preconcentration technique like a cooling trap or with
thermal or solvent desorption tubes. Without preconcentrating the samples, greater sensitivity could be
achieved elsewhere in the infrared spectrum, but would limit the detectable substances to certain
chemical groups (e.g., aldehydes, ketones, alkynes and so on).
Another option is the increase of power by using either a high-power cw-OPO or DFG with
waveguides. Both schemes would enable the employment of more sensitive photoacoustic or cavity
ringdown detection schemes, however at the cost of a large, mode-hop free tuning range which proves
to be essential.
We intend to extend our study to surgical smoke produced in vivo in laparoscopic surgery by
performing measurements in the range 2900-3144 cm-1 and increase sensitivity by preconcentrating the
smoke samples. We are also considering the possibility of investigating surgical smoke samples
around 1000 cm-1 (10 !m) with an external cavity quantum cascade laser (ECQCL) to address the
recording of additional species.
Acknowledgements
The financial support by the Swiss National Science Foundation and ETH Zurich is gratefully
acknowledged. We thank G. Seyfang (ETH) for the fruitful collaboration.
References and Notes
1. Freitag, L.; Chapman, G.A.; Sielczak, M.; Ahmed A.; Russin D. Laser smoke effect on the
bronchial system. Lasers Surg. Med. 1987, 7 283–288.
2. Gloster, H.M; Roenigk, R.K. Risk of acquiring human papillomavirus from the plume produced
by the carbon dioxide laser in the treatment of warts. J. Am. Acad. Dermatol. 1995, 32 436–441.
3. Alp, E.; Bijl, D.; Bleichrodt, R.P.; Hansson, B.;Voss, A. Surgical smoke and infection control. J.
Hosp. Infect. 2006, 62, 1–5.
4. Champault, G.; Taffinder, N.; Ziol, M.; Riskalla, H.; Catheline, J.M.C. Cells are present in the
smoke created during laparoscopic surgery. Br. J. Surg. 1997, 84, 993–995.
5. Baggish, M.; Poiesz, B.; Joret, D.; Williamson, P.; Rebai, A. Presence of human
immunodeficiency virus DNA in laser smoke. Lasers Surg. Med. 1991, 11, 197–203.
6. Hallmo, P.; Naess, O. Laryngeal papillomatosis with papilloma virus DNA contracted by a laser
surgeon. Em. Arch. Otorhinolaryngol. 1992, 248, 425–427.
Sensors 2010, 10
2707
7. Weld, K.J.; Dryer, S.; Ames, C.D.; Cho, K.; Hogan, C.; Lee, M.; Biswas, P.; Landman, J.
Analysis of surgical smoke produced by various energy-based instruments and effect on
laparoscopic visibility. J. Endourol. 2007, 21, 347–251.
8. Nezhat, C.; Winer, W.K.; Nezhat, F.; Nezhat, C.; Forrest, D.; Reeves, W.G. Smoke from laser
surgery-is there a health-hazard. Lasers Surg. Med. 1987, 7, 376–382.
9. Waesche, W.; Albrecht, H. Investigation of the distribution of aerosoles and VOC in plume
produced during laser treatment under OR conditions. Proc. SPIE 1996, 2624, 270–275.
10. Spleiss, M.; Weber, L.; Meier, T.; Treffler, B. Identification and quantification of selected
chemicals in laser pyrolysis products of mammalian tissues. Proc. SPIE 1995, 2323, 409–422.
11. Francke, W.; Fleck, O.; Mihalache, D.L.; Woellmer, W. Identification of volatile compounds
released from biological tissue during CO2 laser treatment. Proc. SPIE 1995, 2323, 423–431.
12. Krones, C.J.; Conze, J.; Hoelzl, F.; Stumpf, M.; Klinge, U.; Moeller, M.; Dott, W.; Schumpelick,
V.; Hollender, J. Chemical composition of surgical smoke produced by electrocautery; harmonic
scalpel and argon beaming-a short study. Eur. Sur. 2007, 39, 118–121.
13. Al Sahaf, O.S.; Vega-Carrascal, I.; Cunningham, F.O.; McGrath, J.P.; Bloomfield, F.J. Chemical
composition of smoke produced by high-frequency electrosurgery. Irish J. Med. Sci. 2007, 176,
229–232.
14. Rey, J.M.; Schramm. D.; Hahnloser D.; Marinov D.; Sigrist, M.W. Spectroscopic investigation of
volatile compounds produced during thermal and radiofrequency bipolar cautery on porcine liver.
Meas. Sci. Technol. 2008, 19, 075602.
15. Barrett, W.L.; Garber, S.M. Surgical smoke-a review of the literature-Is this just a lot of hot air?
Surg. Endosc. 2003, 17, 979–987.
16. Gianella, M.; Sigrist, M.W. Improved algorithm for quantitative analyses of infrared spectra of
multicomponent gas mixtures with unknown compositions. Appl. Spectrosc. 2009, 63, 338–343.
17. Bartlome, R.; Rey, J.M.; Sigrist, M.W. Vapor phase infrared laser spectroscopy: From gas sensing
to forensic urinalysis. Anal. Chem. 2008, 80, 5334–5341.
18. Bartlome, R.; Sigrist, M.W. Laser-based human breath analysis: D/H isotope ratio increase
following heavy water intake. Opt. Lett. 2009, 34, 866–868.
19. Bartlome, R.; Baer, M.; Sigrist, M.W. High-temperature multipass cell for infrared spectroscopy
of heated gases and vapors. Rev. Sci. Instrum. 2007, 78, 013110.
20. Menes, T.; Spivak, H. Laparoscopy-Searching for the proper insufflation gas. Surg. Endosc. 2000,
14, 1050–1056.
21. Sharpe, S.W.; Johnson, T.J.; Sams, R.L.; Chu, P.M.; Rhoderick, G.C.; Johnson, P.A. Gas-phase
databases for quantitative infrared spectroscopy. Appl. Spectrosc. 2004, 58, 1452–1461.
22. Rothman, L.S.; Jacquemart, D.; Barbe, A; Benner, D.C.; Birk, M; Brown, L.R.; Carleer, M.R.;
Chackerian, C.; Chance, K.; Coudert, L.H.; Dana, V.; Devi, V.M.; Flaud, J.M.; Gamache, R.R.;
Goldman, A.; Hartmann, J.M.; Jucks, K.W.; Maki, A.G.; Mandin, J.Y.; Massie, S.T.; Orphal, J.;
Perrin, A.; Rinsland, C.P.; Smith, M.A.H; Tennyson, J.; Tolchenov, R.N.; Toth, R.A.; Vander
Auwera, J.; Varanasi, P.; Wagner, G. The HITRAN 2004 molecular spectroscopic database. J.
Quant. Spectrosc. Radiat. Transfer 2005, 96, 139–204.
Sensors 2010, 10
2708
23. Hensman, C.; Baty, D.; Willis, R.G.; Cuschieri, A. Chemical composition of smoke produced by
high-frequency electrosurgery in a closed gaseous environment-An in vitro study. Surg. Endosc.
1998, 12, 1017–1019.
24. Suva. Grenzwerte am Arbeitsplatz; SUVA: Luzern, Switzerland, 2009.
© 2010 by the authors; licensee Molecular Diversity Preservation International, Basel, Switzerland.
This article is an open-access article distributed under the terms and conditions of the Creative
Commons Attribution license (http://creativecommons.org/licenses/by/3.0/).
... A literature search in the PubMed database (May 2021) comprised the following search terms: 'chemical composition' AND 'surgical smoke', 'volatile organic compounds' AND 'surgical smoke', 'health hazard' AND 'surgical smoke', 'liver' AND 'surgical smoke', 'porcine liver' AND 'smoke', 'pig liver' AND 'smoke', 'air contaminants' AND 'medical application', 'toxic byproducts' AND 'surgery'. These search terms yielded 68 results and thereof 16 original research articles [3][4][5][6][7][8][9][10][11][12][13][14][15][16][17][18] contained VOC analyses of laser-generated or electrosurgical smoke and were included as references to accumulate a list of candidate VOCs contained in the surgical smoke from porcine liver. In addition, 1 review article [19] and 2 relevant articles [20,21], not listed in the PubMed database, were included. ...
... In addition, 1 review article [19] and 2 relevant articles [20,21], not listed in the PubMed database, were included. Several of these articles explicitly refer to electrosurgical smoke generated from porcine liver tissue [4,16,18,20], which was also used in this study. These publications were given preferential consideration. ...
Article
Full-text available
OBJECTIVES Evaluation of smoke capture efficiency of different mobile smoke evacuation devices with respect to volatile organic compounds and their noise emission. METHODS Electrosurgical incisions were performed on fresh porcine liver in an operating room with vertical laminar flow. The generated surgical smoke was analysed with proton-transfer-reaction mass spectrometry with and without the use of a mobile smoke evacuation system consisting of a smoke evacuator machine, a suction hose and a handpiece. The inlet of the mass spectrometer was positioned 40 cm above the specimen. Various devices were compared: a hard plastic funnel, a flexible foam funnel, an on-tip integrated aspirator of an electrosurgical knife and a standard secretion suction (Yankauer). Also, sound levels were measured at a distance of 40 cm from the handpieces’ inlet. RESULTS The smoke capture efficiency of the secretion suction was only 53%, while foam funnel, plastic funnel and integrated aspirator were all significantly more effective with a clearance of 95%, 91% and 91%, respectively. The mean sound levels were 68 and 59 A-weighted decibels with the plastic and foam funnel, respectively, 66 A-weighted decibels with the integrated aspirator and 63 A-weighted decibels with the secretion suction. CONCLUSIONS Carcinogenic, mutagenic and reprotoxic volatile organic compounds in surgical smoke can be efficiently reduced by mobile smoke evacuation system, providing improved protection for medical personnel. Devices specifically designed for smoke evacuation are more efficient than standard suction tools. Noise exposure for the surgeon was lowest with the flexible foam funnel and higher with the other handpieces tested.
... The optical multi-pass cell (MPC) is regarded as an important part of laser absorption spectroscopy for an effective long-optical path length. Besides, combined with modulation spectroscopy of the laser absorption spectroscopy, the optical MPC is widely used in the fields of atmosphere, environmental pollution and industrial process, industrial emissions for trace gas chemical composition analysis and measurement [1][2][3][4][5][6][7][8][9][10][11][12][13][14][15] due to its real-time on-line, high sensitivity, high selectivity non-contacted and non-intrusive advantages. Early White [16] and Herriot [17] cells are still used in a laser absorption spectroscopy for greenhouse gas. ...
Article
A novel augmented 4*4 matrix model for the slightly misaligned optical multi-pass cell whose spherical mirrors are not coaxial due to the presence of small misaligned angle and linear displacement of the spherical mirrors is reported. The augmented 4*4 matrix model is proposed to describe the behavior of light rays in the misaligned optical multi-pass cell. By augmented 4*4 matrix model, a series of numerical calculation are performed to validate that the incident light ray remains unchanged, and then retrace the same spot pattern. It is found that the set of spot patterns from numerical analysis displays a displacement on the surface of the mirror and the misaligned optical elements increase the loss and reduce pass counts. More importantly, the augmented 4*4 matrix model can be used to analyze the misalignment sensitivity of the optical multi-pass cells.
... The multipass cell (MPC) has increasingly played an important role not only in physics and chemistry but also in environmental monitoring, analyses of combustion processes, and performance of medical diagnostics. [1][2][3][4][5][6] The importance of MPC stems from its ability to detect trace gases, such as CH 4 , CO 2 , HCHO, and NH 3 . Early spherical MPCs developed by White and Herriot have been used in laser-based spectroscopic trace gas monitoring 7 due to their simplicity and operability. ...
... These results may indicate possible chemical reaction between the ammonia and oxygen and/or other compound released from the resin, hindering formation of CO 2 . Since, at the same time, high amounts of H 2 O were registered, catalytic oxidation of ammonia could have occurred, resulting in formation of nitrous oxide (same absorption bands as water) or nitrogen (not infrared-active) [33,34]. All of above mentioned processes are exothermic and most likely led to an increased exothermic peak on the DSC curve. ...
Article
Increased application of organic compounds, mainly in form of synthetic resins, used as binders for moulding and core sands in metal casting, may have an adverse effect on the environment and work conditions in foundry plants. In this article we focused on the identification of the degradation products formed during laboratory scale simulation of complex thermal degradation of commercial binder. For our investigation we have chosen the phenol-formaldehyde based resin hardened by a mixture of organic esters (PFRE) as it is widely used in the core and mould sand technology. The identified degradation products obtained during the experiments varied depending on mechanism used for pyrolysis. In the “slow” evaluation, simple compounds such as: CO, CO2, NH3, H2O, phenol and CH4 were generated. Meanwhile, during the “flash” pyrolysis (500, 700, 900 and 1100 °C), mainly phenol and its methyl and ethyl derivatives as well as benzene were released. It was determined that the pyrolysis products and their ratio depended on the temperature during the degradation process.
... Several important advances are made in this area beginning in the 1930s, and the research into a wide range of applications continues to the present days [1][2][3][4][5][6]. Investigations show that MPCs have been widely used in varies areas, including environmental monitoring [4][5][6], combustion processes [7], medical diagnostics [8], and fundamental atomic and molecular physics [9,10]. ...
Article
Full-text available
Multi-pass cells (MPCs) are commonly used in trace-gas detection and weak spectrum measurement. It is essential to accomplish a high-precision measurement of MPCs' effective optical path length (EOPL). A direct high-precision measuring method of MPCs' EOPL with optical frequency domain reflectometer (OFDR) was reported and demonstrated in this paper. Several important parameters of a MPC, such as EOPL and base length, were derived with high-precision by identifying the complicated signal of OFDR. The MPC's EOPL was also verified with the prevailing absorbance method. The results showed that the MPC's EOPL measured by each of these two methods is highly consistent. However, the relative uncertainty with the OFDR dramatically decreased 2 orders of magnitude (about 0.0085%) than that with the absorbance method. It demonstrated that the OFDR method with fewer measurement links is more conducive to a direct measurement. The performances of beam spread and stray light in the White-cell were also evaluated with the method. (C) 2016 Optical Society of America
... Surgical smoke and aerosols are known to be byproducts of the interaction between biological tissues and laser beams, electro surgical units, drills, ultrasonic scalpels. These energy based instruments are dedicated to cut, coagulate, ablate, order to establish the risks for OR personnel when they are exposed to surgical smoke [14,15,16,17]. Gas components from the surgical smoke were detected with different methods like ion flow tube-mass spectroscopy (SIFT-MS) [14], gas chromatography-mass spectroscopy (GC-MS) [18,19] and mid-infrared differencefrequency-generation laser spectroscopy (MIR-DFG) [16], but despite a very good selectivity and sensitivity, the calibrations are elaborated processes (for the first two methods) or the detection limits remain in the ppm range. ...
Article
Full-text available
Energy-based surgery has greatly expanded in recent years because of the many areas that can be applied in the field of human medicine. These procedures are known to produce a toxic smoke in operating room that can pose a health risk both for surgeons and patients. Compared to abundant studies on the qualitative composition on surgical smoke, the quantitative data on personnel exposure are incomplete. More information regarding smoke concentrations are required to discuss the potential health impact of the resulting exposures. Therefore, the quantitative composition of surgical smoke is of great interest. Using Laser Photoacoustic Spectroscopy (LPAS), smoke samples produced in vitro by CO2 laser vaporization of fresh animal tissues were investigated. Traces of acetonitrile, acrolein, ammonia, benzene, ethylene, and toluene were detected in surgical smoke in the ppb to ppm range. Half of the identified gases were found exceeding in quantities the occupational exposure limit. The measured concentrations of these compounds indicated a strong relation of the potential hazards with irradiation conditions (laser power and exposure time). The sensitive CO2 laser photoacoustic spectroscopy measurements are well suited to provide quantitative information on gas products emitted during laser-tissue interaction. © 2015 Quantitative analysis of laser surgical smoke Allrights received.
... The spectra of surgical smoke samples measured with the MIR DFG spectrometer were similar in composition to smoke samples produced in the laboratory on animal tissue in an earlier study [56], where water vapor, methane, ethane and ethylene were the only detected compounds. Methane, ethane and ethylene are relatively harmless, with recommended exposure limits (REL) of 1% [57] . ...
Article
The chemical composition of surgical smoke, a gaseous by-product of some surgical devices—lasers, drills, vessel sealing devices—is of great interest due to the many toxic components that have been found to date. For the first time, surgical smoke samples collected during routine keyhole surgery were analyzed with infrared laser spectroscopy. Traces (ppm range) of methane, ethane, ethylene, carbon monoxide and sevoflurane were detected in the samples which consisted mostly of carbon dioxide and water vapor. Except for the anaesthetic sevoflurane, none of the compounds were present at dangerous concentrations. Negative effects on the health of operation room personnel can be excluded for many toxic compounds found in earlier studies, since their concentrations are below recommended exposure limits.
Article
Importance Smoke generated during surgical procedures has long been thought to be hazardous to hospital personnel; however, the degree of danger has yet to be determined. Observations The dangers of surgical smoke are associated with the composition of the plume. Small-particulate matter is found in the smoke that is easily inhaled. Particulates deposit in the lungs, circulatory system, and other organs, which may cause numerous health problems. The smoke also contains many gaseous compounds known to cause cancer. The type of tissue and the form of cautery have been shown to alter the composition of the surgical smoke. Well-cited articles have asserted that daily inhaled surgical smoke could be equivalent to smoking dozens of cigarettes. Unsafe levels of cancer-causing compounds have been found in the smoke. However, there is no firm evidence that operating room personnel show increased cancer rates compared with the general population. Data implicating other forms of passively inhaled smoke, such as fumes associated with cooking and wood burning, are likely comparable to the smoke issued during surgical procedures. Most operating rooms do not require smoke evacuation, and the elimination techniques are variable. Most systems rely on dispersion and, possibly, mechanical evacuation and filtration. Newer electric filters show promise but require more development. Conclusions and Relevance Surgical smoke is dangerous, but the severity of the risk has yet to be determined. Therefore, no safe level is known at this point. Efforts should be made to reduce and possibly eliminate smoke from the operating room. Research into cost-effective forms of smoke evacuation is necessary. Studies of respiratory and cancer sequelae of exposure to operating room smoke in personnel who have had long-term exposure to surgical smoke is also needed.
Article
Introduction A cell salvage device is used in cardiac surgery with the aim of reducing allogeneic blood transfusion. Suction of blood from the operating field used for the device is often accompanied by diathermy smoke. There is limited published research to know if this blood is then contaminated with clinically significant levels of harmful chemicals from this smoke. Postoperative cardiac surgery patients are already physiologically vulnerable, making the optimization of salvaged blood worth considering. Methods Ten patients who had cardiac surgery using a cell salvage device from a single institution had samples taken from the processed blood just prior to transfusion. Samples were tested for carbon monoxide (CO), cyanide and benzene. Results were compared to preoperative co-oximetry results and normal adult laboratory reference ranges. Demographic data about the patients was collected, including the type of operation, gender, age, body mass index, smoking status and amount of salvaged blood collected and processed. Results Primary surgery was coronary artery bypass grafting (CABG) in three patients, mitral valve repair or replacement in three patients and one each of aortic valve replacement (AVR)/CABG, AVR/plication of the aorta, CABG/external wrapping of the aorta and valve-sparing root replacement. None were smokers. Neither the blood CO level prior to surgery nor in the processed salvaged blood was above the normal limit of 2% for non-smokers. There was no processed blood benzene found. Cyanide levels in the processed blood ranged from 1.8 to 44.1 μmol/l (where <8 μmol/l is considered within the normal adult laboratory limit). Conclusion Despite the obvious limitations of the current study, it shows that cyanide levels can be found many times the normal level in processed salvaged blood. Whilst the total dose of cyanide is small, the potential impact may be clinically significant due to cyanide’s effect on mitochondrial metabolism in the heart and brain.
Article
Full-text available
This paper describes the status of the 2004 edition of the HITRAN molecular spectroscopic database. The HITRAN compilation consists of several components that serve as input for radiative transfer calculation codes: individual line parameters for the microwave through visible spectra of molecules in the gas phase; absorption cross-sections for molecules having dense spectral features, i.e., spectra in which the individual lines are unresolvable; individual line parameters and absorption cross-sections for bands in the ultra-violet; refractive indices of aerosols; tables and files of general properties associated with the database; and database management software. The line-by-line portion of the database contains spectroscopic parameters for 39 molecules including many of their isotopologues.The format of the section of the database on individual line parameters of HITRAN has undergone the most extensive enhancement in almost two decades. It now lists the Einstein A-coefficients, statistical weights of the upper and lower levels of the transitions, a better system for the representation of quantum identifications, and enhanced referencing and uncertainty codes. In addition, there is a provision for making corrections to the broadening of line transitions due to line mixing.
Article
Information on pyrolysis products such as particle size distribution, concentration of plume aerosols and volatile organic compounds produced by laser treatment is required to evaluate the potential health hazards during treatment of biological tissue. In-vitro measurements and `in vivo' collection of samples were carried out for comparison. The concentration of pyrolysis products were determined for ENT and gynecological surgical laser treatments using a Nd:YAG laser and CO2 laser. In consideration of the ventilation systems, room volume and suction efficiency the concentration of volatile organic compounds were measured by means of a multisorbent sampler and a gas chromatography system coupled with a FTIR spectrometer.
Article
A total of 148 volatile compounds were identified from the laser plume produced by a 10 W CO2 laser upon irradiation of pig liver. Structure elucidation was carried out by coupled gas chromatography mass spectrometry using reference samples. The identified compounds include hydrocarbons, alcohols, aldehydes, acids, esters, amides nitriles, ketones, furans, phenols, pyrazines, pyridines, and sulphur-compounds. Toluene, styrene, methylpyrazine, benzaldehyde, benzylcyanide, 4-methylphenol, indol and scatol for the major components of the pyrolysis products; derivatives may also reach higher concentrations. An interesting group of O/N-compounds with closely related chemical structures remained unknown.
Article
Liver and muscle tissue have been irradiated with a surgical CO2-laser. The prefiltered fumes were adsorbed on different sorbents (activated charcoal type NIOSH and Carbotrap) and desorbed with different solvents (carbondisulphide and acetone). Analysis was done by gas chromatography/mass spectrometry. An updated list of identified substances is shown. Typical Maillard reaction products as found in warmed over flavour as aldehydes, aromatics, heterocyclic and sulphur compounds were detected. Quantification of some toxicological relevant substances is presented. The amounts of these substances are given in relation to the laser parameters and different tissues for further toxicological assessment.
Article
The potential risks associated with the emission of by-products emitted by surgical cautery are of concern. Various investigations—mostly based on gas chromatography—have been performed to analyse the so-called surgical smoke but controversies remain in terms of composition and concentrations of compounds present in the smoke and hence the associated risk to human health. This quantitative model study uses for the first time CO2-laser-based photoacoustic spectroscopy and focuses on the analysis of volatile organic compounds produced during thermal- and radiofrequency bipolar cautery on porcine liver. The latter instrument is employed in actual human surgery. Concentrations in the ppm to sub-ppm range and molar fractions could be determined for carbon dioxide, water vapour, ammonia, ethanol and methanol. Distinct differences particularly in the methanol and ethanol contents were found between the two cautery devices employed.
Article
Background: The inhalation of aerosols during electrosurgery is part of daily surgical life. This study analyses the toxic and cancerogenic components of the aerosols. Methods: Electrocautery, harmonic scalpel and argon beaming were utilised in a pig model. The collected aerosols were analysed by gas chromatography coupled with mass spectrometry for acrylamide, aldehydes and ketones, volatile and semivolatile organic compounds, and polycyclic aromatic hydrocarbons. Results: Surgical plume of all instruments comprehends toxic components including e.g. acrylamide, acetaldehyde, formaldehyde, and benzene. Composition and toxicity varies from applied technique, energy, and processed tissue. High temperature induces more toxic aerosols. Offtake devices reduce toxic inhalations. Conclusions: Surgical smoke contains toxic, partly cancerogenic compounds. Concentrations estimated for daily routine are probably below relevant health risk. Nevertheless, the exposition to surgical "smog" should be minimised at any time using offtake devices. Further examinations during clinical practice are necessary.
Article
Following the ingestion of only 5.1 mL of D2O, a mid-infrared laser spectrometer determines the D/H isotope ratio increase in exhaled water vapor for the first time, to the best of our knowledge. This increase is still detectable several weeks after the heavy water intake. Collected breath samples are directly transferred into a high-temperature multipass cell operated at 373 K. No breath sample preparation is required. Aside from the capability to hinder unwanted condensation, measurements at elevated temperatures offer other advantages such as a lower temperature dependence of the delta value or the possibility to vary the intensity of absorption lines. We lay the foundation for many laser-based clinical applications. As an example, we measure a total body water weight of 55.2%+/-1.8% with respect to the total body weight, in agreement with the normal value of the male population.
Article
We present a major improvement of an algorithm based on a spectral library search for the quantitative analysis of multicomponent gas samples with unknown compositions. A quantitative spectral database of infrared spectra is used as a training set to compute regression coefficients. Concentrations are computed in the principal component space via principal component regression (PCR). In addition to previous algorithms, we introduce a rating for each candidate substance depending on the concentration found with PCR and a filter that removes candidates that are predicted with negative concentrations if their rating is below a certain threshold. Negative concentrations arise when the measured spectrum contains components that are not contained in the database. The PCR is recomputed until all candidates have a rating above the threshold. Then an adaptive filter "subtracts" the substance with the highest rating from both the measured spectrum and the library and appends it to a hit list. The iteration of this procedure directly produces a list of substances in order of descending importance (i.e., contribution to the measured absorbance) with their corresponding concentrations. The algorithm is tested on spectra of multicomponent surgical smoke samples. The four main components (water, methane, ethane, and ethene) are identified correctly (within the top 5 of the hit list) for an appropriate choice of the rating threshold. The algorithm describes the composition of the smoke sample correctly despite the presence of features in the spectrum that cannot be explained by the spectrum of any single substance present in the database.
Article
A 44-year-old laser surgeon presented with laryngeal papillomatosis. In situ DNA hybridization of tissue from these tumors revealed human papillomavirus DNA types 6 and 11. Past history revealed that the surgeon had given laser therapy to patients with anogenital condylomas, which are known to harbor the same viral types. These findings suggest that the papillomas in our patient may have been caused by inhaled virus particles present in the laser plume.
Article
Concentrated tissue culture pellets infected with human immunodeficiency virus (HIV) containing 1 x 10(7) cells/ml were vaporized by means of a carbon dioxide laser. The vaporous debris resulting from the laser's impact were evacuated through sterile silastic tubing, then bubbled through sterile culture medium (RPMI) positioned in series with a commercial smoke evacuator. No HIV DNA was detected in the culture medium flask. Tissue culture studies of the silastic collection tubing revealed p24 HIV gag antigen in 3 of 12 tube segments at the end of 1 week and in 1 of 12 tube segments at 2 weeks. No sustained infection of HIV cultured cells was observed at the 28th day. Polymerase chain reaction (PCR) analysis of particulate debris obtained from the silastic collection tubing was positive from proviral HIV DNA in both immediately sampled and day 14 cultured material.