Figure 2 - Assessment of Electromagnetic Interference with Active Cardiovascular Implantable Electronic Devices (CIEDs) Caused by the Qi A13 Design Wireless Charging Board

Figure 2. Results of the induced voltage for the homogenous exposure for (a) the bipolar lead, and (b) the unipolar lead. The linear fit to the measured values is indicated by the solid lines. The dashed lines indicate the extrapolation of the linear fit to the ranges where no induced voltages were measured due to the operation range of the used differential amplifier of the measurement system.  
Results of the induced voltage for the homogenous exposure for (a) the bipolar lead, and (b) the unipolar lead. The linear fit to the measured values is indicated by the solid lines. The dashed lines indicate the extrapolation of the linear fit to the ranges where no induced voltages were measured due to the operation range of the used differential amplifier of the measurement system.  