Willi A. Kalender

Universitätsklinikum Erlangen, Erlangen, Bavaria, Germany

Are you Willi A. Kalender?

Claim your profile

Publications (512)866.04 Total impact

  • C Steiding, D Kolditz, W Kalender
    [Show abstract] [Hide abstract]
    ABSTRACT: Purpose: The aim of this work was to implement, validate, and compare two procedures for routine image quality (IQ) assurance in dental cone-beam computed tomography (CBCT): 1. the German standard DIN 6868 - 161 introduced in 2013 and 2. the established standard IEC 61 223 - 3-5 for clinical CT x-ray equipment referenced as "DIN" and "IEC" below. Materials and Methods: The approximated in-plane modulation transfer function (MTF), the contrast-to-noise indicator (CNI), and the uniformity indicator (UI*) were determined in accordance with DIN. Image noise, the uniformity index (UI), the contrast-to-noise ratio (CNR), and the 3 D MTF were measured according to IEC 61 223 - 3-5 using a previously proposed quality assurance (QA) framework. For this, a modular phantom was used. All experiments were performed on a clinical dental CBCT unit. The severity of image artefacts was measured at different z-positions. A dedicated computer program was implemented to allow for automated QA procedure. Results: The position and orientation of the phantoms were detected automatically in all of the measurements providing a reproducible placement of the evaluation regions and volumes. 50 % and 10 % in-plane MTF values of the approximated and the exact MTF calculation procedure were in agreement to within 5 %. With increasing axial distance from the isocentre, UI* and CNI dropped by 30 % and 19 %, respectively. Conventional IQ parameters showed higher sensitivity to image artefacts; i. e., UI and CNR were reduced by about 197 % and 37 %. Conclusion: The implemented automated QA routines are compatible with both the DIN and the IEC approach and offer reliable and quantitative tracking of imaging performance in dental CBCT for clinical practice. However, there is no equivalence between the DIN and the IEC metrics. In addition, direct measurements of physical IQ parameters such as image contrast and noise, uniformity, and axial resolution are not supported by the new concept according to DIN. Key points: • The new DIN 6868 - 161 is not equivalent to the established IEC 61 223 - 3-5.• Noise, uniformity, and contrast are well-suited to assess image artefacts.• The implemented automated quality assurance program fits clinical routine. Citation Format: • Steiding C, Kolditz D, Kalender W. Comparison of Methods for Acceptance and Constancy Testing in Dental Cone-beam Computed Tomography. Fortschr Röntgenstr 2014; DOI: 10.1055/s-0034-1385333.
    RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 11/2014;
  • [Show abstract] [Hide abstract]
    ABSTRACT: The estimation of patient dose using Monte Carlo (MC) simulations based on the available patient CT images is limited to the length of the scan. Software tools for dose estimation based on standard computational phantoms overcome this problem; however, they are limited with respect to taking individual patient anatomy into account. The purpose of this study was to generate whole-body patient models in order to take scattered radiation and over-scanning effects into account. Thorax examinations were performed on three physical anthropomorphic phantoms at tube voltages of 80 kV and 120 kV; absorbed dose was measured using thermoluminescence dosimeters (TLD). Whole-body voxel models were built as a combination of the acquired CT images appended by data taken from widely used anthropomorphic voxel phantoms. MC simulations were performed both for the CT image volumes alone and for the whole-body models. Measured and calculated dose distributions were compared for each TLD chip position; additionally, organ doses were determined. MC simulations based only on CT data underestimated dose by 8%-15% on average depending on patient size with highest underestimation values of 37% for the adult phantom at the caudal border of the image volume. The use of whole-body models substantially reduced these errors; measured and simulated results consistently agreed to better than 10%. This study demonstrates that combined whole-body models can provide three-dimensional dose distributions with improved accuracy. Using the presented concept should be of high interest for research studies which demand high accuracy, e.g. for dose optimization efforts.
    Physica Medica 10/2014; · 1.17 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Purpose: The presented study aimed at optimizing a patient table design for breast CT (BCT) systems with respect to breast tissue coverage and patient comfort. Additionally, the benefits and acceptance of an immobilization device for BCT using underpressure were evaluated. Materials and Methods: Three different study parts were carried out. In a positioning study women were investigated on an MRI tabletop with exchangeable inserts (flat and cone-shaped with different opening diameters) to evaluate their influence on breast coverage and patient comfort in various positioning alternatives. Breast length and volume were calculated to compare positioning modalities including various opening diameters and forms. In the second study part, an underpressure system was tested for its functionality and comfort on a stereotactic biopsy table mimicking a future CT scanner table. In the last study part, this system was tested regarding breast tissue coverage. Results: Best results for breast tissue coverage were shown for cone-shaped table inserts with an opening of 180 mm. Flat inserts did not provide complete coverage of breast tissue. The underpressure system showed robust function and tended to pull more breast tissue into the field of view. Patient comfort was rated good for all table inserts, with highest ratings for cone-shaped inserts. Conclusion: Cone-shaped tabletops appeared to be adequate for BCT systems and to allow imaging of almost the complete breast. An underpressure system proved promising for the fixation of the breast during imaging and increased coverage. Patient comfort appears to be adequate. Key points: • Tissue coverage in breast CT is highly dependent on patient table design.• An underpressure fixation system shows potential to increase breast coverage.• The proposed breast CT patient table design combines good coverage and patient comfort. Citation Format: • Rößler AC, Wenkel E, Althoff F et al. The Influence of Patient Positioning in Breast CT on Breast Tissue Coverage and Patient Comfort. Fortschr Röntgenstr 2014; DOI: 10.1055/s-0034-1385208.
    RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 09/2014;
  • [Show abstract] [Hide abstract]
    ABSTRACT: The purpose of this study was to validate the use of a single shaped filter (SF) for computed tomography (CT) using variable source-to-filter distance (SFD) for the examination of different object diameters.A SF was designed by performing simulations with the purpose of achieving noise homogeneity in the reconstructed volume and dose reduction for arbitrary phantom diameters. This was accomplished by using a filter design method thats target is to achieve a homogeneous detector noise, but also uses a correction factor for the filtered back projection process. According to simulation results, a single SF designed for one of the largest phantom diameters meets the requirements for all diameters when SFD can be adjusted. To validate these results, a SF made of aluminium alloy was manufactured. Measurements were performed on a CT scanner with polymethyl methacrylate (PMMA) phantoms of diameters from 40-100 mm. The filter was positioned at SFDs ranging from 97-168 mm depending on the phantom diameter. Image quality was evaluated for the reconstructed volume by assessing CT value accuracy, noise homogeneity, contrast-to-noise ratio weighted by dose (CNRD) and spatial resolution. Furthermore, scatter distribution was determined with the use of a beam-stop phantom. Dose was measured for a PMMA phantom with a diameter of 100 mm using a calibrated ionization chamber.The application of a single SF at variable SFD led to improved noise uniformity and dose reduction: noise homogeneity was improved from 15% down to about 0%, and dose was reduced by about 37%. Furthermore, scatter dropped by about 32%, which led to reduced cupping artifacts and improved CT value accuracy. Spatial resolution and CNRD was not affected by the SF.By means of a single SF with variable SFD designed for CT, significant dose reduction can be achieved and image quality can be improved by reducing noise inhomogeneity as well as scatter-induced artifacts.
    Physics in Medicine and Biology 09/2014; 59(19):5691-5706. · 2.70 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Purpose To determine the dose-length product (DLP)-effective dose (ED) (DLP/ED) conversion coefficient (k) tables for the lower extremities that can be used for calculating ED. Materials and Methods Dose calculations were performed on standard phantoms using a validated Monte Carlo calculation tool. Calculations were performed to obtain ED values for tube voltages from 80 kV to 140 kV in steps of 20 kV for the following examinations: hip (femur), knee, ankle, and computed tomographic (CT) angiography of the lower extremities. Values of the DLP were calculated by multiplying measured CT dose index values by the scan length; k values resulted as the quotients of the ED and DLP values. DLP/ED coefficients averaged over the range of voltage values and their standard deviations were determined for the given lower-extremity CT examinations for all age groups and for both sexes. Results Coefficients depend strongly on the phantom age and size, but little on the kilovolt value. In the case of the newborn, for example, k values were 0.0612, 0.0046, 0.0014, and 0.047 for hip, knee, ankle, and CT angiography, respectively, while in the case of the adult, these respective values were 0.0110, 0.0004, 0.0002, and 0.0062. A substantial difference up to 20% between coefficients in male and female phantoms was observed for CT angiographic examination. Conclusion DLP/ED conversion coefficients are provided for lower extremities and allow estimation of ED for commonly used clinical musculoskeletal CT and CT angiographic protocols. © RSNA, 2014.
    Radiology 06/2014; · 6.34 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Dynamic contrast-enhanced (DCE) micro-computed tomography (micro-CT) has emerged as a valuable imaging tool to noninvasively obtain quantitative physiological biomarkers of drug effect in preclinical studies of antiangiogenic compounds. In this study, we explored the ability of DCE micro-CT to assess the antiangiogenic treatment response in breast cancer xenografts and correlated the results to the structural vessel response obtained from 3-dimensional (3D) fluorescence ultramicroscopy (UM). Two groups of tumor-bearing mice (KPL-4) underwent DCE micro-CT imaging using a fast preclinical dual-source micro-CT system (TomoScope Synergy Twin, CT Imaging GmbH, Erlangen, Germany). Mice were treated with either a monoclonal antibody against the vascular endothelial growth factor or an unspecific control antibody. Changes in vascular physiology were assessed measuring the mean value of the relative blood volume (rBV) and the permeability-surface area product (PS) in different tumor regions of interest (tumor center, tumor periphery, and total tumor tissue). Parametric maps of rBV were calculated of the tumor volume to assess the intratumoral vascular heterogeneity. Isotropic 3D UM vessel scans were performed from excised tumor tissue, and automated 3D segmentation algorithms were used to determine the microvessel density (MVD), relative vessel volume, and vessel diameters. In addition, the accumulation of coinjected fluorescence-labeled trastuzumab was quantified in the UM tissue scans to obtain an indirect measure of vessel permeability. Results of the DCE micro-CT were compared with corresponding results obtained by ex vivo UM. For validation, DCE micro-CT and UM parameters were compared with conventional histology and tumor volume. Examination of the parametric rBV maps revealed significantly different patterns of intratumoral blood supply between treated and control tumors. Whereas control tumors showed a characteristic vascular rim pattern with considerably elevated rBV values in the tumor periphery, treated tumors showed a widely homogeneous blood supply. Compared with UM, the physiological rBV maps showed excellent agreement with the spatial morphology of the intratumoral vascular architecture. Regional assessment of mean physiological values exhibited a significant decrease in rBV (P < 0.01) and PS (P < 0.05) in the tumor periphery after anti-vascular endothelial growth factor treatment. Structural validation with UM showed a significant reduction in reduction of relative vessel volume (rVV) (P < 0.01) and MVD (P < 0.01) in the corresponding tumor region. The reduction in rBV correlated well with the rVV (R = 0.73 for single values and R = 0.95 for mean values). Spatial maps of antibody penetration showed a significantly reduced antibody accumulation (P < 0.01) in the tumor tissue after treatment and agreed well with the physiological change of PS. Examination of vessel diameters revealed a size-dependent antiangiogenic treatment effect, which showed a significant reduction in MVD (P < 0.001) for vessels with diameters smaller than 25 μm. No treatment effect was observed by tumor volume. Noninvasive DCE micro-CT provides valuable physiological information of antiangiogenic drug effect in the intact animal and correlates with ex vivo structural analysis of 3D UM. The combined use of DCE micro-CT with UM constitutes a complementary imaging toolset that can help to enhance our understanding of antiangiogenic drug mechanisms of action in preclinical drug research.
    Investigative radiology 03/2014; · 4.85 Impact Factor
  • Christian Steiding, Daniel Kolditz, Willi A Kalender
    [Show abstract] [Hide abstract]
    ABSTRACT: Thousands of cone-beam computed tomography (CBCT) scanners for vascular, maxillofacial, neurological, and body imaging are in clinical use today, but there is no consensus on uniform acceptance and constancy testing for image quality (IQ) and dose yet. The authors developed a quality assurance (QA) framework for fully automated and time-efficient performance evaluation of these systems. In addition, the dependence of objective Fourier-based IQ metrics on direction and position in 3D volumes was investigated for CBCT. The authors designed a dedicated QA phantom 10 cm in length consisting of five compartments, each with a diameter of 10 cm, and an optional extension ring 16 cm in diameter. A homogeneous section of water-equivalent material allows measuring CT value accuracy, image noise and uniformity, and multidimensional global and local noise power spectra (NPS). For the quantitative determination of 3D high-contrast spatial resolution, the modulation transfer function (MTF) of centrally and peripherally positioned aluminum spheres was computed from edge profiles. Additional in-plane and axial resolution patterns were used to assess resolution qualitatively. The characterization of low-contrast detectability as well as CT value linearity and artifact behavior was tested by utilizing sections with soft-tissue-equivalent and metallic inserts. For an automated QA procedure, a phantom detection algorithm was implemented. All tests used in the dedicated QA program were initially verified in simulation studies and experimentally confirmed on a clinical dental CBCT system. The automated IQ evaluation of volume data sets of the dental CBCT system was achieved with the proposed phantom requiring only one scan for the determination of all desired parameters. Typically, less than 5 min were needed for phantom set-up, scanning, and data analysis. Quantitative evaluation of system performance over time by comparison to previous examinations was also verified. The maximum percentage interscan variation of repeated measurements was less than 4% and 1.7% on average for all investigated quality criteria. The NPS-based image noise differed by less than 5% from the conventional standard deviation approach and spatially selective 10% MTF values were well comparable to subjective results obtained with 3D resolution pattern. Determining only transverse spatial resolution and global noise behavior in the central field of measurement turned out to be insufficient. The proposed framework transfers QA routines employed in conventional CT in an advanced version to CBCT for fully automated and time-efficient evaluation of technical equipment. With the modular phantom design, a routine as well as an expert version for assessing IQ is provided. The QA program can be used for arbitrary CT units to evaluate 3D imaging characteristics automatically.
    Medical Physics 03/2014; 41(3):031901. · 2.91 Impact Factor
  • Jounal of Medical Engeneering. 01/2014; Vol 2014(Article ID 946574).
  • [Show abstract] [Hide abstract]
    ABSTRACT: PURPOSE Intrascan Motion-artifact-correction in C-arm-based flat-detector CT (FD-CT) is an important issue in interventional imaging because of longer scan times as compared to Multi-Slice CT. Our aim was the development and evaluation of an online image-content-based motion-correction technique without using any kind of markers or external motion knowledge. METHOD AND MATERIALS The correction method is based on a gradient descent method, minimizing a gray-value entropy criterion optimizing the underlying acquistion trajectory parameters. It is formed as a multistep approach, including a global, local and projection wise optimization. We are using a locally rigid variation of the systems trajectory parameters like detector- or source-translation or a detector rotation to compensate patient motion. The retrospective evaluation of 30 arbitrary (with weak and strong motion, without motion artifacts) patient head scans included 5s 3D angiography and 20s soft-tissue protocols. All scans were performed on an Artis Q System (Siemens AG). For each dataset three volumes were computed: 1) original reconstruction using the system’s geometry calibration (OR), 2) motion corrected reconstruction without any system information (MCR) and 3) motion corrected reconstruction using the system’s geometry calibration as initialization (MCR+). Two neuroradiology experts performed a visual evaluation according to a 5-point grading scale with respect to general image quality, motion-artifact-content and spatial resolution of the structures of interest, e.g. 3D vessels. RESULTS The average scores for OR, MCR and MCR+ were 2.75, 3.0 and 3.15, respectively. The combined compensation of unknown trajectories and unknown patient motion (MCR) can lead to comparable results to OR. Both experts confirmed a distinct reduction of artifacts by the motion correction algorithm (MCR+) , e.g. blurring and streaks. Especially for 3D angiography even small distal vessels were depicted clearly. MCR+ application on soft-tissue protocols illustrated a constantly better delineation of bone und soft-tissue in the border zones. CONCLUSION Image-based motion correction is possible without a-priori knowledge of the motion pattern and can improve interventional FD-CT imaging. CLINICAL RELEVANCE/APPLICATION Using the proposed algorithm enables good image quality even for unsteady patients and can be helpful for longer FD-CT acquisitions in cases where anaesthesia is contraindicated.
    Radiological Society of North America 2013 Scientific Assembly and Annual Meeting; 12/2013
  • [Show abstract] [Hide abstract]
    ABSTRACT: Purpose Objective evaluation of image quality based on known phantom structures fixed in known positions is hard to do. We designed phantoms which allow arranging tumor-like masses, microcalcifications (µCa) and fibers freely with and without superpositioned structures to enable observer studies with the observer blinded to the ground truth. Material and Methods Phantoms were made of breast-equivalent material as a 40 mm slab and as a 14 cm diameter cylinder simulating the compressed breast for digital mammography (DM) and breast tomosynthesis (BT) and the pendant breast for breast CT (bCT), respectively. Both phantoms provide cavities (45x40² mm³) to insert structures such as cubes (10³ mm³) and plates with different thicknesses simulating normal breast tissue or containing spherical soft-tissue masses (2-8 mm), µCa (0.1-0.4 mm) or connective tissue (fibers, 1 mm) which can be placed in different locations and superimpose each other. Measurements were done on clinical DM and BT systems and on a prototype high-resolution bCT system with and without superimposed structures. Exposure parameters for all systems were kept consistent with the standard clinical settings; i.e., spatial resolution was better than 100 µm and dose was kept below 5 mGy. Images were acquired using random structure arrangements; ROC curves were generated based on 5 observers. Results For DM and BT, low-contrast masses down to 4 mm were recognized without and down to 6 mm with added superimposing structures. All µCa objects were detected down to 200 µm with and without superpositioned structures. Fibers were not seen in either arrangement. For bCT, low-contrast masses down to a size of 2 mm diameter, all µCa objects and all fibers were recognized by the observers. Standard ROC analysis revealed higher sensitivity and specificity for bCT in low-contrast detectability than for DM and BT. Conclusion Objective evaluation of low-contrast detectability with the reader blinded to structure type and position is feasible. In this test, high-resolution bCT showed performance superior to DM and DT, especially when confounding structures were superimposed. Clinical Relevance/Application The proposed phantom setup allows conducting receiver operating characteristic (ROC) studies for the objective evaluation of scanners and scan protocols with ground truth clearly defined.
    RSNA 2013, Chicago; 12/2013
  • Conference Paper: The European Perspective
    Willi A. Kalender
    [Show abstract] [Hide abstract]
    ABSTRACT: LEARNING OBJECTIVES 1) Understand that CTDI is merely a technical concept for scanner acceptance and constancy testing, but not a measure for patient dose. 2) Learn about concepts for patient- and scanner-specific patient dose estimates. 3) Learn about the concept of diagnostic reference levels and its strengths and weaknesses. ABSTRACT There is no major debate regarding the validity of the computed tomography dose index (CTDI) in Europe because it is considered as a tool for scanner acceptance and constancy testing. Its use for that purpose is undisputed. Measures for patient dose have been a major topic for decades. There are no common regulations valid for all of Europe, but there are a number of initiatives and concepts in place already which originated here. Among these are primarily the generation of conversion coefficients k for estimating values of the effective dose E from the dose length product (DLP) by E = k×DLP and the concept of dose reference levels (DRL). DRLs for radiological examinations in the European Union were demanded by law already in 2000. Patient dose assessment relies predominantly on pre-tabulated values generated for anthropomorphic and voxel phantoms. Efforts are underway to provide more patient-specific dose estimates (PSDE) independent of CTDI phantom measurements. The lecture will review the above concepts and will point to both strengths and weaknesses.
    Radiological Society of North America 2013 Scientific Assembly and Annual Meeting; 12/2013
  • Natalia Saltybaeva, Daniel Kolditz, Willi A. Kalender
    [Show abstract] [Hide abstract]
    ABSTRACT: PURPOSE Monte Carlo (MC) simulations can be performed on patient CT image data. However, this data is limited to the scanned volume and does not allow estimating dose for organs outside the directly exposed range. Standard mathematical phantoms do not have this limitation, but they do not reflect individual patient anatomy. The aim of this work was to generate whole-body patient-specific voxel models for accurate organ and effective dose estimations. METHOD AND MATERIALS Three anthropomorphic phantoms representing an adult (Alderson Research Laboratories, New York, USA) and 5 and 1 y.o children (CIRS, Norfolk, VA, USA) were considered as patients and scanned with thorax routine protocols (SOMATOM Definition Flash, Siemens, Forchheim, Germany) with tube voltages of 80, 100 and 120 kV. The absorbed dose was measured using 90 calibrated TLD chips. Whole-body voxel models were generated by amending the patient CT volume by size-adapted versions of the ICRP Reference Male phantom in case of the adult and ORNL voxel phantoms in the case of the pediatric patients. Organ definitions provided with these phantoms were transferred to the combined models and adapted interactively. MC simulations were performed using the validated tool ImpactMC (CT Imaging GmbH, Erlangen, Germany) for a) the unamended patient volumes, b) the combined whole-body models. Simulated 3D dose distributions were compared with TLD measurements chip by chip. RESULTS The mean difference between measurements and simulations based on the unamended CT volumes was 19%; using the whole-body model reduced the difference to 6%. In contrast to unamended CT volumes, whole-body models intrinsically also provided the dose values for organs outside the scanned volume. These organs contributed 17%, 34% and 36% to effective dose for the adult, 5y.o. and 1 y.o. phantom, respectively. CONCLUSION Patient-specific whole-body models allow to increase accuracy of dose estimation and to calculate dose for all relevant organs. CLINICAL RELEVANCE/APPLICATION The approach is useful for individual patient dose estimation, especially for clinical studies.
    Radiological Society of North America 2013 Scientific Assembly and Annual Meeting; 12/2013
  • [Show abstract] [Hide abstract]
    ABSTRACT: PURPOSE There is general consensus that computed tomography (CT) can provide good soft-tissue discrimination and dynamic contrast-enhanced studies of the breast, but with insufficient spatial resolution and dose values exceeding the limits set for screening examinations. We re-evaluated if this assumption still holds true for an innovative high-resolution breast CT (bCT) system. METHOD AND MATERIALS We compared the performance of a bCT prototype (CT Imaging GmbH, Erlangen, Germany) to two clinical systems of two different manufacturers for each digital mammography (DM) and breast tomosynthesis (BT) with respect to detectability of the structures presented by the American College of Radiology (ACR) accreditation phantom. bCT examines one breast at a time with the patient lying prone on the patient bed without exposing the body trunk. The prototype employs a new cadmium telluride detector with 100 µm pixel size, single photon counting electronics and close to 100% detection efficiency [Kalender WA et al. Eur Radiol 2012; 22(1):1-8]. The tests focused on the question if fibers down to 0.75 mm, masses down to 0.50 mm, and specks down to 0.24 mm were clearly distinguished as recommended by the ACR. Tests were also performed to determine image quality and dose. We did not add overlaying structures, which would be potentially confounding the ACR structures for DM and BT. RESULTS Acceptance testing for all 5 systems confirmed that they met the requirements for screening mammography; the bCT system provided better than 100 µm spatial resolution at average glandular dose levels below 5 mGy. Measurements of the ACR phantom revealed the following: DM and BT showed fibers, masses and specks as required; bCT went beyond this and revealed even the finest structures presented in the ACR phantom, i.e. fibers of 0.4 mm, masses of 0.25 mm and specks of 0.16 mm. CONCLUSION Fully 3D high-resolution breast CT showed performance superior to DM and BT, even in the benevolent situation with no confounding structures superimposed. Smaller structures may have to be introduced in test phantoms to provide adequate tests for finer details. CLINICAL RELEVANCE/APPLICATION High-resolution breast CT appears to offer potential for superposition-free fully 3D imaging of the breast at improved detail resolution and dose levels accepted for screening procedures.
    Radiological Society of North America 2013 Scientific Assembly and Annual Meeting; 12/2013
  • [Show abstract] [Hide abstract]
    ABSTRACT: PURPOSE Tube current modulation (TCM) and automatic exposure control (AEC) are widely used in modern CT. The aim of this work was to include the effects of TCM and AEC in a software package for fast and easy organ and effective dose estimates. METHOD AND MATERIALS Measurements were carried out for a SOMATOM Definition Flash scanner (Siemens AG, Forchheim, Germany); the manufacturer provided all necessary information on their CARE Dose4D TCM/AEC product. TCM and AEC curves were derived for anthropomorphic phantoms by generating complete CT projection data sets by means of ray-tracing and predicting the flux at the detector. For all phantoms and parameter combinations studied, Monte Carlo (MC) calculations w&w/o CARE Dose4D were performed to provide tabulated dose values. These tables were included in the software package ImpactDose (CT Imaging GmbH, Erlangen, Germany) which estimates organ and effective dose depending on patient size, scan region and scan protocol. It is based on pre-tabulated dose values calculated by means of MC calculations for the ORNL family of anthropomorphic phantoms. Validation measurements were performed using thermoluminescence dosimeters (TLDs) for each of three different anthropomorphic phantoms (Rando adult, 5-y.o. and 1-y.o. CIRS) w&w/o CARE Dose4D. RESULTS Measured dose values were compared to MC results on a chip-by-chip basis. The mean differences for all TLD chips were 5%, 7%, and 6% for the adult, the 5-year old, and the 1-year old phantom, respectively. This deviation is in the range of the uncertainty associated with TLD measurements and indicates that TCM/AEC were correctly implemented. The derived dose values w&w/o TCM/AEC allowed for assessment of their effect on dose for different patients without the need for measurements or repeated MC calculations. CONCLUSION Dose estimates based on tabulated MC-derived dose distributions can provide accurate information on the effect of TCM and AEC in clinical CT if information about their implementation is provided by the manufacturer. CLINICAL RELEVANCE/APPLICATION The software package allows to obtain fast and accurate dose estimates when TCM/AEC is used and furthermore may serve as a learning tool.
    Radiological Society of North America 2013 Scientific Assembly and Annual Meeting; 12/2013
  • Ronny Hendrych, Marcel Beister, Willi A. Kalender
    [Show abstract] [Hide abstract]
    ABSTRACT: PURPOSE In clinical breast CT it is of interest to calculate low-noise CT volumes for soft-tissue lesion (STL) detection from noisy high-resolution (HR) images for micro-calcification diagnosis to avoid multiple reconstructions. A viewer for multi planar reformatting (MPR) was developed and evaluated to offer a continuous adjustment of spatial resolution, to reduce the time necessary for the diagnostic procedure and to improve the workflow. METHOD AND MATERIALS Simulations of mathematical breast phantoms were performed (ImpactSim, CT Imaging GmbH, Erlangen, Germany) with average glandular dose levels varied from 1.5 up to 6 mGy. Furthermore, a breast CT prototype (CT Imaging GmbH, Erlangen, Germany) was used to scan an ACR Phantom (CIRS, Norfolk, VA, USA). Volumetric images were reconstructed in HR mode and subsequently 3D filtered using the following techniques: Gaussian, median and box filters and an iterative impulse detector using a weighted median filter. The visibility of lesions was assessed by calculating the effective contrast-to-noise ratio (CNReff), combining the usual CNR with the diameter of the lesion in question. RESULTS The MPR viewer allowed for continuous interactive real-time filtering of large HR volumes in an interactive fashion. In the simulated breast CT volumes the applied filters improved the CNReff for lesions of 2 mm from 1.4 unfiltered in the HR volumes up to 39.2, 23.7, 40.5 and 8.5 for box filter, median, Gaussian and impulse detector. For the ACR phantom the Gaussian filter achieved the best results with an increased CNReff from 7.6 to 51.4 for the smallest lesion. Thereby all filters help to surpass the Rose criterion which states that values of 5 or higher are necessary to distinguish objects from the surrounding area. CONCLUSION The MPR viewer eliminates the need for multiple reconstructions in breast CT. It allows adjusting interactively the spatial resolution and thereby changing the effective CNR continuously. CLINICAL RELEVANCE/APPLICATION MPR viewers may help to avoid multiple image reconstructions, increase the effective CNR of lesions and improve the workflow for breast CT exams.
    Radiological Society of North America 2013 Scientific Assembly and Annual Meeting; 12/2013
  • [Show abstract] [Hide abstract]
    ABSTRACT: PURPOSE To validate the use of a single shaped filter with variable source-to-filter distance (SFD) for dedicated breast CT (bCT) and arbitrary breast sizes. METHOD AND MATERIALS The shaped filter was designed using simulations of a dedicated bCT system with the goal to achieve noise homogeneity and dose reduction for breast diameters of 80 to 180 mm. This was accomplished with a filter design method that aims to achieve a homogeneous detector noise but considering a correction factor for the filtered back projection process. According to the simulations a single shaped filter designed for the largest breast diameter works for all breast diameters if SFD can be adjusted. To validate these results the filter was manufactured of an aluminum alloy. The measurements were performed on a bCT prototype with breast phantoms (80% adipose, 20% glandular tissue) of diameters from 80 to 180 mm. The filter was positioned at SFDs from 54 to 112 mm according to the phantom diameter. Image quality was evaluated for the reconstructed volume by assessing CT value accuracy, noise homogeneity and spatial resolution. Furthermore, scatter distribution was determined with the use of a beam-stop phantom with and without shaped filter. Dose reduction was measured using a calibrated ionization chamber in the center and in the periphery of the phantom. RESULTS The results with a single shaped filter at variable SFD resulted in improved noise homogeneity and dose reduction for all breast diameters: noise homogeneity was improved from 15% down to 5% and the overall dose was reduced by about 30 to 40% for all breast diameters. Furthermore, scatter reduction of about 60% was achieved, which reduced cupping artifacts and improved the CT value accuracy. Spatial resolution was not affected by the shaped filter. CONCLUSION By means of shaped filters designed for bCT, significant dose reduction can be achieved and image quality can be improved by reducing noise inhomogeneity as well as scatter-induced artifacts. A single shaped filter designed for the largest breast diameter used with variable SFD appears to be a good solution for bCT. CLINICAL RELEVANCE/APPLICATION The use of a shaped filter for bCT appears essential to keep patient dose as low as reasonably achievable.
    Radiological Society of North America 2013 Scientific Assembly and Annual Meeting; 12/2013
  • [Show abstract] [Hide abstract]
    ABSTRACT: Cadaver and phantom measurements and simulations confirmed that radiation exposure in 3D QCT of the spine can be reduced if 80 kV instead of 120 kV protocols are used; 120 mAs and slice thicknesses of 1-1.3 mm should be usable but obese patient will require higher milliampere-second settings. To develop a low-radiation exposure CT acquisition protocol for 3D QCT of the thoracolumbar spine. Twenty-six cadavers were scanned with a standard protocol of 120 kV, 100 mAs and with a low-dose protocol using 90 kV, 150 mAs. The scan range included the vertebrae T6 to L4. Each vertebra was segmented and the integral volume and BMD of the total vertebral body were determined. Effective dose values were estimated. The impact of milliampere-second reduction on image quality was simulated by adding noise. One hundred ninety-six vertebrae were analyzed. Integral volume as well as integral BMD correlated significantly (p < 0.001) between standard and low-dose protocols (volume, r (2) = 0.991, residual root mean square (RMS) error, 0.77 cm(3); BMD, r (2) = 0.985, RMS error, 4.21 mg/cm(3)). The slope significantly differed from 1 for integral BMD but not for volume hinting at residual field inhomogeneity differences between the two voltage settings that could be corrected by cross-calibration. Compared to the standard protocol, effective dose was reduced by over 50 % in the low-dose protocol. Adding noise in the 90 kV images to simulate a reduction from 150 to 100 mAs did not affect the results for integral volume or BMD. For 3D QCT of the spine, depending on scanner type, 80 or 90 kV instead of 120 kV protocols may be considered as an important option to reduce radiation exposure; 120 mAs and slice thicknesses of 1-1.5 mm are usable if segmentation is robust to noise. In obese patients, higher milliampere-second settings will be required.
    Osteoporosis International 10/2013; · 4.04 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Purpose: With recently introduced technical innovations for CT systems, the dose of CT scan acquisitions has been substantially reduced; even effective dose values below 1 mSv have been reported. Due to this development, dose of the localizer radiograph may contribute substantially to dose of the whole CT examination. Since there are only limited data in the literature regarding patient dose for the different types of localizer radiographs, patient dose values were estimated in our study by measurements and Monte Carlo simulations and compared to dose values of typical CT examinations.Methods: First, dose distributions were measured in anthropomorphic phantoms for three different body regions (head, thorax, abdomen-pelvic) and three positions of the x-ray tube (AP, PA, and lateral views); measured values were compared to simulated data using Monte Carlo techniques for validation purposes. Second, organ and effective dose values for the various investigated localizer radiograph scenarios were calculated and compared with published dose values for standard CT and low-dose CT examinations.Results: For the anthropomorphic phantom, deviations of the dose values between measured and calculated results were in the range of 15%. Organ and effective dose values showed a strong dependence on the tube position. The largest differences were observed for chest localizer radiographs in the female phantom for the dose to the breast (AP: 1.01 mGy vs PA: 0.24 mGy). Overall effective dose values were in the range of 0.04-0.42 mSv per localizer radiograph acquisition.Conclusions: In view of the technical dose-reducing innovations in CT, localizer radiographs may substantially contribute to the total dose of the whole CT examination, particularly in the case of dedicated low-dose scans used, e.g., for young patients or screening purposes. Optimization of dose in localizer radiographs should be pursued further in the same way as it was done in CT.
    Medical Physics 08/2013; 40(8):084301. · 2.91 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Treatment of cardiac diseases via minimally invasive procedures is of major interest in the clinics. An angiographic C-arm CT system is state-of-the-art in an interventional cardiac laboratory. It opens up the possibility of 3D reconstruction during the procedure. Due to the long acquisition time of several seconds of the C-arm, imaging of dynamic structures is a challenging problem. Therefore, motion correction for cardiac applications is an issue for this imaging device. New minimally invasive procedures like the recently introduced TAVI (transcatheter aortic valve implantation) suffer from cardiac motion. The 3D image of the aorta is acquired during rapid pacing of the patient to minimize the cardiac motion and to reduce the blood flow. We present a new algorithmic approach for motion compensation of the aortic root for TAVI procedures under sinus rythm to make rapid pacing unnecessary. Our optimization routine was tested on three clinical datasets of the aortic root, wherein all three show promising results.
    The 12th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, Lake Tahoe, California; 06/2013

Publication Stats

9k Citations
866.04 Total Impact Points

Institutions

  • 1995–2014
    • Universitätsklinikum Erlangen
      • Department of Obstetrics and Gynaecology
      Erlangen, Bavaria, Germany
  • 2012
    • Autonomous University of Queretaro
      Ciudad Queretaro, Querétaro, Mexico
  • 1996–2012
    • Friedrich-Alexander Universität Erlangen-Nürnberg
      • • Institute of Physics
      • • Department of Diagnostic Radiology
      Erlangen, Bavaria, Germany
    • Universität Ulm
      Ulm, Baden-Württemberg, Germany
  • 2008
    • University of Chicago
      • Department of Radiology
      Chicago, IL, United States
  • 2007
    • Mayo Foundation for Medical Education and Research
      • Department of Radiology
      Scottsdale, AZ, United States
  • 2001
    • Phoenix Zoo
      Phoenix, Arizona, United States
  • 1997–1999
    • IST Austria
      Klosterneuberg, Lower Austria, Austria
  • 1994
    • University of Washington Seattle
      • Department of Otolaryngology/Head and Neck Surgery
      Seattle, WA, United States
  • 1988–1991
    • Free University of Brussels
      • Department of Radiology
      Brussels, BRU, Belgium
    • Siemens
      München, Bavaria, Germany