-
[show abstract]
[hide abstract]
ABSTRACT: Assessment of image quality for digital x-ray mammography systems used in European screening programs relies mainly on contrast-detail CDMAM phantom scoring and requires the acquisition and analysis of many images in order to reduce variability in threshold detectability. Part II of this study proposes an alternative method based on the detectability index (d') calculated for a non-prewhitened model observer with an eye filter (NPWE). The detectability index was calculated from the normalized noise power spectrum and image contrast, both measured from an image of a 5 cm poly(methyl methacrylate) phantom containing a 0.2 mm thick aluminium square, and the pre-sampling modulation transfer function. This was performed as a function of air kerma at the detector for 11 different digital mammography systems. These calculated d' values were compared against threshold gold thickness (T) results measured with the CDMAM test object and against derived theoretical relationships. A simple relationship was found between T and d', as a function of detector air kerma; a linear relationship was found between d' and contrast-to-noise ratio. The values of threshold thickness used to specify acceptable performance in the European Guidelines for 0.10 and 0.25 mm diameter discs were equivalent to threshold calculated detectability indices of 1.05 and 6.30, respectively. The NPWE method is a validated alternative to CDMAM scoring for use in the image quality specification, quality control and optimization of digital x-ray systems for screening mammography.
Physics in Medicine and Biology 06/2011; 56(14):4221-38. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: In many European countries, image quality for digital x-ray systems used in screening mammography is currently specified using a threshold-detail detectability method. This is a two-part study that proposes an alternative method based on calculated detectability for a model observer: the first part of the work presents a characterization of the systems. Eleven digital mammography systems were included in the study; four computed radiography (CR) systems, and a group of seven digital radiography (DR) detectors, composed of three amorphous selenium-based detectors, three caesium iodide scintillator systems and a silicon wafer-based photon counting system. The technical parameters assessed included the system response curve, detector uniformity error, pre-sampling modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE). Approximate quantum noise limited exposure range was examined using a separation of noise sources based upon standard deviation. Noise separation showed that electronic noise was the dominant noise at low detector air kerma for three systems; the remaining systems showed quantum noise limited behaviour between 12.5 and 380 µGy. Greater variation in detector MTF was found for the DR group compared to the CR systems; MTF at 5 mm(-1) varied from 0.08 to 0.23 for the CR detectors against a range of 0.16-0.64 for the DR units. The needle CR detector had a higher MTF, lower NNPS and higher DQE at 5 mm(-1) than the powder CR phosphors. DQE at 5 mm(-1) ranged from 0.02 to 0.20 for the CR systems, while DQE at 5 mm(-1) for the DR group ranged from 0.04 to 0.41, indicating higher DQE for the DR detectors and needle CR system than for the powder CR phosphor systems. The technical evaluation section of the study showed that the digital mammography systems were well set up and exhibiting typical performance for the detector technology employed in the respective systems.
Physics in Medicine and Biology 06/2011; 56(14):4201-20. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: A wide variation in patient exposure has been observed in interventional radiology and cardiology. The purpose of this study was to investigate the patient dose from fluoroscopy-guided procedures performed in non-academic centres when compared with academic centres. Four procedures (coronary angiography, percutaneous coronary intervention, angiography of the lower limbs and percutaneous transluminal angioplasty of the lower limbs) were evaluated. Data on the dose-area product, fluoroscopy time and number of images for 1000 procedures were obtained from 23 non-academic centres and compared with data from 5 academic centres. No differences were found for cardiology procedures performed in non-academic centres versus academic ones. However, significantly lower doses were delivered to patients for procedures of the lower limbs when they were performed in non-academic centres. This may be due to more complex procedures performed in the academic centres. Comparison between the centres showed a great variation in the patient dose for these lower limb procedures.
Radiation Protection Dosimetry 02/2011; 148(1):74-82. · 0.82 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: Diagnostic reference levels (DRLs) were established for 21 indication-based CT examinations for adults in Switzerland. One hundred and seventy-nine of 225 computed tomography (CT) scanners operated in hospitals and private radiology institutes were audited on-site and patient doses were collected. For each CT scanner, a correction factor was calculated expressing the deviation of the measured weighted computed tomography dose index (CTDI) to the nominal weighted CTDI as displayed on the workstation. Patient doses were corrected by this factor providing a realistic basis for establishing national DRLs. Results showed large variations in doses between different radiology departments in Switzerland, especially for examinations of the petrous bone, pelvis, lower limbs and heart. This indicates that the concept of DRLs has not yet been correctly applied for CT examinations in clinical routine. A close collaboration of all stakeholders is mandatory to assure an effective radiation protection of patients. On-site audits will be intensified to further establish the concept of DRLs in Switzerland.
Radiation Protection Dosimetry 10/2010; 142(2-4):244-54. · 0.82 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: The aim of this study was to evaluate and compare organ doses delivered to patients in wrist and petrous bone examinations using a multislice spiral computed tomography (CT) and a C-arm cone-beam CT equipped with a flat-panel detector (XperCT). For this purpose, doses to the target organ, i.e. wrist or petrous bone, together with those to the most radiosensitive nearby organs, i.e. thyroid and eye lens, were measured and compared. Furthermore, image quality was compared for both imaging systems and different acquisition modes using a Catphan phantom. Results show that both systems guarantee adequate accuracy for diagnostic purposes for wrist and petrous bone examinations. Compared with the CT scanner, the XperCT system slightly reduces the dose to target organs and shortens the overall duration of the wrist examination. In addition, using the XperCT enables a reduction of the dose to the eye lens during head scans (skull base and ear examinations).
Radiation Protection Dosimetry 03/2010; 139(1-3):164-8. · 0.82 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: The aim of this work is to compare two methods used for determining the proper shielding of computed tomography (CT) rooms while considering recent technological advances in CT scanners. The approaches of the German Institute for Standardisation and the US National Council on Radiation Protection and Measurements were compared and a series of radiation measurements were performed in several CT rooms at the Lausanne University Hospital. The following three-step procedure is proposed for assuring sufficient shielding of rooms hosting new CT units with spiral mode acquisition and various X-ray beam collimation widths: (1) calculate the ambient equivalent dose for a representative average weekly dose length product at the position where shielding is required; (2) from the maximum permissible weekly dose at the location of interest, calculate the transmission factor F that must be taken to ensure proper shielding and (3) convert the transmission factor into a thickness of lead shielding. A similar approach could be adopted to use when designing shielding for fluoroscopy rooms, where the basic quantity would be the dose area product instead of the load of current (milliampere-minute).
Radiation Protection Dosimetry 03/2010; 139(1-3):403-9. · 0.82 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: The purpose of this study was to assess the spatial resolution of a computed tomography (CT) scanner with an automatic approach developed for routine quality controls when varying CT parameters. The methods available to assess the modulation transfer functions (MTF) with the automatic approach were Droege's and the bead point source (BPS) methods. These MTFs were compared with presampled ones obtained using Boone's method. The results show that Droege's method is not accurate in the low-frequency range, whereas the BPS method is highly sensitive to image noise. While both methods are well adapted to routine stability controls, it was shown that they are not able to provide absolute measurements. On the other hand, Boone's method, which is robust with respect to aliasing, more resilient to noise and provides absolute measurements, satisfies the commissioning requirements perfectly. Thus, Boone's method combined with a modified Catphan 600 phantom could be a good solution to assess CT spatial resolution in the different CT planes.
Radiation Protection Dosimetry 02/2010; 139(1-3):443-8. · 0.82 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: The purpose of this paper is to present a strategy to define diagnostic reference levels DRL for fluoroscopic, dose-intensive
examinations in cardiology and interventional radiology. This work is part of the project of the Federal Office of Public
Health of Switzerland to translate the guidelines of the ICRP and the EU into action. After the 2002 survey in all University
Hospitals in Switzerland this work will present the results of the 2006-2007 survey performed in small and medium sized hospitals.
The data of the small and medium sized hospitals are analyzed to establish DRL. They are corrected to patient size and analyzed
in respect to the difficulty of the examination, the experience of the operator and the type of image detection system. The
results of the study will be compared to those of the former study in the University Hospitals.
KeywordsDiagnostic Reference Level DRL-Cardiology-Interventional Radiology
12/2009: pages 136-139;
-
[show abstract]
[hide abstract]
ABSTRACT: Surveying the frequency of medical x-ray examinations is of prime importance in the assessment of the collective detriment
due to diagnostic and interventional radiology. In the past, this was performed using paper questionnaires, but today there
is an increasing interest in the automatic collection of the frequency data for reasons related to reducing the work load
and increasing the accuracy of the results. This paper present the work performed in Switzerland to explore the use of the
Tarmed coding system for this purpose. The preliminary investigation covering a sample of examinations indicates that Tarmed
coding can easily be used for the collection of radiography and CT examinations, but presents some difficulties for fluoroscopy,
mainly in the case of complex angiography and interventional examinations.
Keywordsdiagnostics-x-ray examinations-frequency surveys-coding systems-population exposure
12/2009: pages 277-279;
-
[show abstract]
[hide abstract]
ABSTRACT: Over the past few years the frequency of computed tomography (CT) examinations has dramatically increased. Simultaneously,
there has been also a significant increase in CT patient dose due to high-resolution imaging and application of more complex
scan techniques. Since no dose limit exists for patients, the International Commission on Radiological Protection introduced
the concept of diagnostic reference levels (DRL) as a means of dose optimization. The aim of this project is to collect patient
doses for the most frequently applied CT protocols and to provide a realistic basis for establishing DRL in CT in Switzerland.
Starting in 2007, patient doses of every Swiss radiological institute operating a CT scanner were going to be collected. Volume
computed tomography dose index (CTDIvol) and dose-length product (DLP) for standard patients was collected for selected clinical CT protocols. The 75th percentile of the CTDIvol and DLP distribution was calculated and compared to the proposed DRL which is partly based on the Swiss survey in 1998 and
recommendations of the European Union. For standard examination of the skull/brain the 75th percentiles are higher than the proposed DRL (72 mGy vs. 60 mGy; 1180 mGy∙cm vs. 1000 mGy∙cm). For examination of thorax
and abdomen/pelvis the 75th percentiles are close to the proposed DRL (thorax: 15 mGy vs. 15 mGy; 511 mGy∙cm vs. 450 mGy∙cm;
abdomen/ pelvis: 16 mGy vs. 15 mGy; 701 mGy∙cm vs. 700 mGy∙cm). In conclusion, there is always a trade-off between dose reduction
and diagnostic image quality. However, especially for skull/brain examinations, optimization is still feasible. The concept
of DRL provides a valuable means for practitioners and manufacturers in optimizing CT protocols.
KeywordsDiagnostic Reference Level-Computed Tomography-Switzerland-CTDI-DLP
12/2009: pages 146-149;
-
[show abstract]
[hide abstract]
ABSTRACT: We studied the influence of signal variability on human and model observers for detection tasks with realistic simulated masses superimposed on real patient mammographic backgrounds and synthesized mammographic backgrounds (clustered lumpy backgrounds, CLB). Results under the signal-known-exactly (SKE) paradigm were compared with signal-known-statistically (SKS) tasks for which the observers did not have prior knowledge of the shape or size of the signal. Human observers' performance did not vary significantly when benign masses were superimposed on real images or on CLB. Uncertainty and variability in signal shape did not degrade human performance significantly compared with the SKE task, while variability in signal size did. Implementation of appropriate internal noise components allowed the fit of model observers to human performance.
Journal of the Optical Society of America A 03/2009; 26(2):425-36. · 1.56 Impact Factor
-
Radiographics 01/2008; 28(7):1807-1816. · 2.85 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: A nationwide survey was launched to investigate the use of fluoroscopy and establish national reference levels (RL) for dose-intensive procedures. The 2-year investigation covered five radiology and nine cardiology departments in public hospitals and private clinics, and focused on 12 examination types: 6 diagnostic and 6 interventional. A total of 1,000 examinations was registered. Information including the fluoroscopy time (T), the number of frames (N) and the dose-area product (DAP) was provided. The data set was used to establish the distributions of T, N and the DAP and the associated RL values. The examinations were pooled to improve the statistics. A wide variation in dose and image quality in fixed geometry was observed. As an example, the skin dose rate for abdominal examinations varied in the range of 10 to 45 mGy/min for comparable image quality. A wide variability was found for several types of examinations, mainly complex ones. DAP RLs of 210, 125, 80, 240, 440 and 110 Gy cm2 were established for lower limb and iliac angiography, cerebral angiography, coronary angiography, biliary drainage and stenting, cerebral embolization and PTCA, respectively. The RL values established are compared to the data published in the literature.
European Radiology 07/2007; 17(6):1621-33. · 3.22 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: A nationwide investigation was conducted in Switzerland to establish the exposure of the population by medical x rays and update the results of the 1998 survey. Both the frequency and the dose variations were studied in order to determine the change in the collective dose. The frequency study addressed 206 general practitioners (GPs), 30 hospitals, and 10 private radiology institutes. Except for the latter, the response rate was very satisfactory. The dose study relied on the assessment of the speed class of the screen-film combinations used by the GPs as well as the results of two separate studies dedicated to fluoroscopy and CT. The investigation showed that the total number of all medical x-ray examinations performed by GPs registered a 1% decrease between 1998 and 2003, and that the sensitivities of the film-screen combinations registered a shift towards higher values, leading to a reduction of the dose delivered by a GP of the order of 20%. The study indicated also that the total number of all x-ray examinations performed in hospitals increased by 4%, with a slight increase of radiographies by 1% but significant decrease of examinations involving fluoroscopy (39%), and a 70% increase for CT examinations. Concerning the doses, the investigation of a selection of examinations involving fluoroscopy showed a significant increase of the kerma-area product (KAP) per procedure. For CT the study showed an increase of the dose-length product (DLP) per procedure for skull and abdomen examinations, and a decrease for chest examination. Both changes in the frequency and the effective dose per examination led to a 20% increase in the total collective dose.
Health Physics 06/2007; 92(5):442-8. · 1.68 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: An objective analysis of image quality parameters was performed for six digital mammography systems. The presampled modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) for the systems were determined at different doses, for 28 kVp with a Mo/Mo or W/Al target/filter combination and 2 mm of additional aluminium filtration. The flat-panel units have higher MTF and DQE in the mid to high frequency range than standard CR systems. The highest DQE, over the whole dose range, is for the slit-scanning direct photon counting system. Dual-side read CR can overcome the inherent x-ray absorption and signal collection limitations of standard CR mammography, improving the low-frequency DQE by 40%, to the same level as full-field systems, but it does not improve the poor spatial resolution of phosphor.
Medical Physics 04/2007; 34(3):906-14. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: The age of the patient is of prime importance when assessing the radiological risk to patients due to medical X-ray exposures and the total detriment to the population due to radiodiagnostics. In order to take into account the age-specific radiosensitivity, three age groups are considered: children, adults and the elderly. In this work, the relative number of examinations carried out on paediatric and geriatric patients is established, compared with adult patients, for radiodiagnostics as a whole, for dental and medical radiology, for 8 radiological modalities as well as for 40 types of X-ray examinations. The relative numbers of X-ray examinations are determined based on the corresponding age distributions of patients and that of the general population. Two broad groups of X-ray examinations may be defined. Group A comprises conventional radiography, fluoroscopy and computed tomography; for this group a paediatric patient undergoes half the number of examinations as that of an adult, and a geriatric patient undergoes 2.5 times more. Group B comprises angiography and interventional procedures; for this group a paediatric patient undergoes a one-fourth of the number of examinations carried out on an adult, and a geriatric patient undergoes five times more.
Radiation Protection Dosimetry 02/2007; 123(3):402-8. · 0.82 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: In our recent paper by Monnin et al. [Med. Phys. 33, 411-420 (2006)], an objective analysis of the relative performance of a computed radiography (CR) system using both standard single-side (ST-VI) and prototype dual-side read (ST-BD) plates was reported. The presampled modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for the systems were determined at three different beam qualities representative of paediatric chest radiography, at an entrance detector air kerma of 5 microGy. Experiments demonstrated that, compared to the standard single-side read system, the MTF for the dual-side read system was slightly reduced, but a significant decrease in image noise resulted in a marked increase in DQE (+40%) in the low spatial frequency range. However, the DQE improvement for the ST-BD plate decreased with increasing spatial frequency, and, at spatial frequencies above 2.2 mm(-1), the DQE of the dual-side read system was lower than that of the single-side one.
Medical Physics 11/2006; 33(10):3637-9. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: This study analyzes CT examinations in Switzerland.
Using different sources (administrative data on the equipment, a 1998 nationwide inquiry into practices, and data provided by the Swiss University Hospitals of Basel, Zurich, and Lausanne), we determined the frequency of CT examinations (hospitals and private radiologists) in 1998 according to different descriptive variables and studied the progression in CT use over time.
CT scanners increased by 7% between 1998 and 2004. The average annual number of CT examinations in 1998 was 46.3/1000 population, 3.4% of all radiological examinations in Switzerland in 1997-1998. The most frequent examination was CT of the skull (24%), while private radiology institutes perform more CTs of the spine. More CT examinations were performed for men than for women (sex ratio M/F=1.17). The average annual increase in CT in Swiss hospitals varied from 8% for Basel to 18% for Lausanne. Finally, the proportion of pediatric examinations was 5%; their numbers appear to be stabilizing.
There is a significant increase in CT examinations. It is hoped that our study will heighten awareness among doctors of CT examinations in order to optimize their use.
Journal de Radiologie 10/2006; 87(9):1057-62. · 0.42 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: The development of CT applications might become a public health problem if no effort is made on the justification and the optimisation of the examinations. This paper presents some hints to assure that the risk-benefit compromise remains in favour of the patient, especially when one deals with the examinations of young patients. In this context a particular attention has to be made on the justification of the examination. When performing the acquisition one needs to optimise the extension of the volume investigated together with the number of acquisition sequences used. Finally, the use of automatic exposure systems, now available on all the units, and the use of the Diagnostic Reference Levels (DRL) should allow help radiologists to control the exposure of their patients.
Revue médicale suisse 08/2006; 2(73):1752-7.
-
[show abstract]
[hide abstract]
ABSTRACT: An objective analysis of image quality parameters was performed for a computed radiography (CR) system using both standard single-side and prototype dual-side read plates. The pre-sampled modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for the systems were determined at three different beam qualities representative of pediatric chest radiography, at an entrance detector air kerma of 5 microGy. The NPS and DQE measurements were realized under clinically relevant x-ray spectra for pediatric radiology, including x-ray scatter radiations. Compared to the standard single-side read system, the MTF for the dual-side read system is reduced, but this is offset by a significant decrease in image noise, resulting in a marked increase in DQE (+40%) in the low spatial frequency range. Thus, for the same image quality, the new technology permits the CR system to be used at a reduced dose level.
Medical Physics 03/2006; 33(2):411-20. · 2.83 Impact Factor