-
[show abstract]
[hide abstract]
ABSTRACT: Reliable post-implant evaluation of prostate seed implants requires optimal seed identification and accurate delineation of anatomical structures. In this study the GEC-ESTRO groups BRAPHYQS and PROBATE investigated the interobserver variability in post-implant prostate contouring, seed reconstruction and image fusion and its impact on the dose-volume parameters.
Post-implant T2-TSE, T1-GE and CT images were acquired for three patients, in order to evaluate four post-plan techniques: (a) CT, (b) T1+T2, (c) CT+T2, (d) CT+T1(int)+T2. Three interobserver studies were set up. (1) Contouring: the CTV-prostate was delineated on CT and T2 by eight physicians. Additionally one reference contour was defined on both image modalities for each patient. (2) Seed reconstruction: seven physicists localised the seeds on T1 and CT, manually and with CT seed finder tools. A reference seed geometry was defined on CT and T1. (3) Fusion: six physicists registered the image sets for technique (b)-(d), using seeds (if visible) and anatomical landmarks. A reference fusion was determined for each combined technique.
(1) The SD(ref) for contouring (1 SD with respect to the reference volume) was largest for CT (23%), but also surprisingly large for MRI (17%). This resulted in large SD(ref) values for D90 for all techniques (17-23%). The surprisingly large SD(ref) for MRI was partly due to variations in interpretation of what to include in the prostate contour. (2) The SD(ref) in D90 for seed reconstruction was small (2%) for all techniques, except for T1+T2 (7%). (3) The SD(ref) in D90 due to image fusion was quite large, especially for direct fusion of CT+T2 (16%) where clearly corresponding landmarks were missing (seeds hardly visible on T2). In general, we observed large differences in D90 depending on the technique used.
The dosimetric parameters for prostate post-implant evaluation showed large technique-dependent interobserver variabilities. Contouring and image fusion are the 'weak links' in the procedure. Guidelines and training in contouring together with incorporation of automated fusion software need to be implemented.
Radiotherapy and Oncology 07/2012; 104(2):192-8. · 5.58 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: PURPOSE: To quantify the whole-body-dose delivered during the application of new techniques and compare them to the results obtained by treatment planning systems. The ultimate goal being the use of planning data in combination with complication data to assess the impact of low doses of ionizing radiation. METHODS: A film technique using gafchromic films to assess low doses was used on simplified phantoms and compared to data from treatment planning systems as well as a simplified whole body dose calculation system (Peridose). The types of treatment include open fields, intensity modulated radiation therapy (IMRT) and volumetric arc treatments. The film measurements were confirmed using TLDs in Alderson phantoms. In addition neutron contributions were measured as these are not taken into account in the current modern treatment planning systems, but can add significantly to the patient's whole body dose. RESULTS: Dose outside of the treatment plane diminished to 1% of the prescribed dose, this for open fields, IMRT and rotational treatments alike. Noteworthy was an increase at about 20cm from the central plane in IMRT, and in a more limited fashion for volumetric modulated arc treatment. In open fields this was not observed. Treatment planning systems were good at determining the out-of-field doses of single field treatments. In complex plans the TPS underestimated the dose to the patient. At distances greater than 20cm from the field edge, these systems did not predict any dose. The Peridose program performed well in the case of classical treatments. In the case of IMRT treatments, the overall evolution of the dose as a function of the distance to the field was well-modeled. However, an over estimation of the order of 60-80% was observed, leaving the possibility for a corrective factor based on a point measurement. Dose levels over the whole body were of the order 100mGy or higher over a complete treatment for the more complex treatments. Neutron dose levels were of the order single digit mSv for 10MV treatments. For 18MV the level of neutron contribution was in agreement with recent publications, corroborating reports that the dose from neutrons is lower than previously reported.
Radiotherapy and Oncology 01/2012; · 5.58 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: To introduce a methodology to perform dose measurements using Gafchromic films which can span several decades of dose levels.
The technique is based on a rescaling approach using different films irradiated at different dose levels. This is combined with a registration protocol correcting positioning and scaling factors for each film. The methodology is validated using TLD's for out-of-field doses. Furthermore, two examples are provided using the technique to characterize small sized radiosurgery cones and compared with measurements made with a pinpoint chamber.
Excellent agreement with TLD, planning systems and measurement was found. The superior resolution of the film technique was apparent.
The authors have introduced a new technique allowing users to quantify very low doses in conjunction with commissioning measurements. The use of film also provides 2D information on beam characteristics in high resolution measurements.
Medical Physics 12/2011; 38(12):6443-8. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: We propose a methodology to perform dose measurements using gafchromic films
which can span several decades of dose levels. The technique is based on a
rescaling approach using different films irradiated at different dose levels.
This is combined with a registration protocol correcting positioning and
scaling factors for each film.
07/2011;
-
[show abstract]
[hide abstract]
ABSTRACT: Purpose : To relate the physical transmittance parameters of the water
equivalent Gafchromic EBT 2 Film with the delivered dose in a transparent
absolute calibration protocol. The protocol should be easy to understand, easy
to perform, and should be able to predict the residual dose error.
Conclussions : The gafchromic EBT2 Films are properly calibrated with an
accessible robust calibration protocol. The protocol largely deals with the
uniformity problems of the Film. The proposed method allowed to relate the dose
with the red channel transmittance using only T0, T_inf, and a dose scaling
factor. Based on the local and global uniformity the red channel dose errors
could be predicted to be smaller than 5%.
07/2011;
-
[show abstract]
[hide abstract]
ABSTRACT: To study the impact of clinical predisposing factors on rectal normal tissue complication probability modeling using the updated results of the Dutch prostate dose-escalation trial.
Toxicity data of 512 patients (conformally treated to 68 Gy [n = 284] and 78 Gy [n = 228]) with complete follow-up at 3 years after radiotherapy were studied. Scored end points were rectal bleeding, high stool frequency, and fecal incontinence. Two traditional dose-based models (Lyman-Kutcher-Burman (LKB) and Relative Seriality (RS) and a logistic model were fitted using a maximum likelihood approach. Furthermore, these model fits were improved by including the most significant clinical factors. The area under the receiver operating characteristic curve (AUC) was used to compare the discriminating ability of all fits.
Including clinical factors significantly increased the predictive power of the models for all end points. In the optimal LKB, RS, and logistic models for rectal bleeding and fecal incontinence, the first significant (p = 0.011-0.013) clinical factor was "previous abdominal surgery." As second significant (p = 0.012-0.016) factor, "cardiac history" was included in all three rectal bleeding fits, whereas including "diabetes" was significant (p = 0.039-0.048) in fecal incontinence modeling but only in the LKB and logistic models. High stool frequency fits only benefitted significantly (p = 0.003-0.006) from the inclusion of the baseline toxicity score. For all models rectal bleeding fits had the highest AUC (0.77) where it was 0.63 and 0.68 for high stool frequency and fecal incontinence, respectively. LKB and logistic model fits resulted in similar values for the volume parameter. The steepness parameter was somewhat higher in the logistic model, also resulting in a slightly lower D(50). Anal wall DVHs were used for fecal incontinence, whereas anorectal wall dose best described the other two endpoints.
Comparable prediction models were obtained with LKB, RS, and logistic NTCP models. Including clinical factors improved the predictive power of all models significantly.
International journal of radiation oncology, biology, physics 06/2011; 82(3):1233-42. · 4.59 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: Intrafractional motion consists of two components: (1) the movement between the on-line repositioning procedure and the treatment start and (2) the movement during the treatment delivery. The goal of this study is to estimate this intrafractional movement of the prostate during prostate cancer radiotherapy.
Twenty-seven patients with prostate cancer and implanted fiducials underwent a marker match procedure before a five-field IMRT treatment. For all fields, in-treatment images were obtained and then processed to enable automatic marker detection. Combining the subsequent projection images, five positions of each marker were determined using the shortest path approach. The residual set-up error (RSE) after kV-MV based prostate localization, the prostate position as a function of time during a radiotherapy session and the required margins to account for intrafractional motion were determined.
The mean RSE and standard deviation in the antero-posterior, cranio-caudal and left-right direction were 2.3±1.5 mm, 0.2±1.1 mm and -0.1±1.1 mm, respectively. Almost all motions occurred in the posterior direction before the first treatment beam as the percentage of excursions>5 mm was reduced significantly when the RSE was not accounted for. The required margins for intrafractional motion increased with prolongation of the treatment. Application of a repositioning protocol after every beam could decrease the 1cm margin from CTV to PTV by 2 mm.
The RSE is the main contributor to intrafractional motion. This RSE after on-line prostate localization and patient repositioning in the posterior direction emphasizes the need to speed up the marker match procedure. Also, a prostate IMRT treatment should be administered as fast as possible, to ensure that the pre-treatment repositioning efforts are not erased by intrafractional prostate motion. This warrants an optimized workflow with the use of faster treatment techniques.
Radiotherapy and Oncology 02/2011; 98(2):181-6. · 5.58 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: We evaluated an isocentric technique for conformal irradiation of the breast, internal mammary, and medial supra-clavicular lymph nodes (IM-MS LN) using the oblique parasternal photon (OPP) technique. For 20 breast cancer patients, the OPP technique was compared with a conventional mixed-beam technique (2D) and a conformal partly wide tangential (PWT) technique, using dose-volume histogram analysis and normal tissue complication probabilities (NTCPs). The 3D techniques resulted in a better target coverage and homogeneity than did the 2D technique. The homogeneity index for the IM-MS PTV increased from 0.57 for 2D to 0.90 for PWT and 0.91 for OPP (both p < 0.001). The OPP technique was able to reduce the volume of heart receiving more than 30 Gy (V(30)), the cardiac NTCP, and the volume of contralateral breast receiving 5 Gy (V(5)) compared with the PWT plans (all p < 0.05). There is no significant difference in mean lung dose or lung NTCP between both 3D techniques. Compared with the PWT technique, the volume of lung receiving more than 20 Gy (V(20)) was increased with the OPP technique, whereas the volume of lung receiving more than 40 Gy (V(40)) was decreased (both p < 0.05). Compared with the PWT technique, the OPP technique can reduce doses to the contralateral breast and heart at the expense of an increased lung V(20).
Medical dosimetry: official journal of the American Association of Medical Dosimetrists 01/2011; 36(1):28-34. · 1.26 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: To evaluate the potential for dose escalation with intensity-modulated radiotherapy (IMRT) in positron emission tomography-based radiotherapy planning for locally advanced non-small-cell lung cancer (LA-NSCLC).
For 35 LA-NSCLC patients, three-dimensional conformal radiotherapy and IMRT plans were made to a prescription dose (PD) of 66 Gy in 2-Gy fractions. Dose escalation was performed toward the maximal PD using secondary endpoint constraints for the lung, spinal cord, and heart, with de-escalation according to defined esophageal tolerance. Dose calculation was performed using the Eclipse pencil beam algorithm, and all plans were recalculated using a collapsed cone algorithm. The normal tissue complication probabilities were calculated for the lung (Grade 2 pneumonitis) and esophagus (acute toxicity, grade 2 or greater, and late toxicity).
IMRT resulted in statistically significant decreases in the mean lung (p <.0001) and maximal spinal cord (p = .002 and 0005) doses, allowing an average increase in the PD of 8.6-14.2 Gy (p ≤.0001). This advantage was lost after de-escalation within the defined esophageal dose limits. The lung normal tissue complication probabilities were significantly lower for IMRT (p <.0001), even after dose escalation. For esophageal toxicity, IMRT significantly decreased the acute NTCP values at the low dose levels (p = .0009 and p <.0001). After maximal dose escalation, late esophageal tolerance became critical (p <.0001), especially when using IMRT, owing to the parallel increases in the esophageal dose and PD.
In LA-NSCLC, IMRT offers the potential to significantly escalate the PD, dependent on the lung and spinal cord tolerance. However, parallel increases in the esophageal dose abolished the advantage, even when using collapsed cone algorithms. This is important to consider in the context of concomitant chemoradiotherapy schedules using IMRT.
International journal of radiation oncology, biology, physics 09/2010; 80(1):306-13. · 4.59 Impact Factor
-
Medical Physics 04/2010; 37(4):1379-81. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: A robust and accurate method that allows the automatic detection of fiducial markers in MV and kV projection image pairs is proposed. The method allows to automatically correct for inter or intrafraction motion.
Intratreatment MV projection images are acquired during each of five treatment beams of prostate cancer patients with four implanted fiducial markers. The projection images are first preprocessed using a series of marker enhancing filters. 2D candidate marker locations are generated for each of the filtered projection images and 3D candidate marker locations are reconstructed by pairing candidates in subsequent projection images. The correct marker positions are retrieved in 3D by the minimization of a cost function that combines 2D image intensity and 3D geometric or shape information for the entire marker configuration simultaneously. This optimization problem is solved using dynamic programming such that the globally optimal configuration for all markers is always found. Translational interfraction and intrafraction prostate motion and the required patient repositioning is assessed from the position of the centroid of the detected markers in different MV image pairs. The method was validated on a phantom using CT as ground-truth and on clinical data sets of 16 patients using manual marker annotations as ground-truth.
The entire setup was confirmed to be accurate to around 1 mm by the phantom measurements. The reproducibility of the manual marker selection was less than 3.5 pixels in the MV images. In patient images, markers were correctly identified in at least 99% of the cases for anterior projection images and 96% of the cases for oblique projection images. The average marker detection accuracy was 1.4 +/- 1.8 pixels in the projection images. The centroid of all four reconstructed marker positions in 3D was positioned within 2 mm of the ground-truth position in 99.73% of all cases. Detecting four markers in a pair of MV images takes a little less than a second where most time is spent on the image preprocessing.
The authors have developed a method to automatically detect multiple markers in a pair of projection images that is robust, accurate, and sufficiently fast for clinical use. It can be used for kV, MV, or mixed image pairs and can cope with limited motion between the projection images.
Medical Physics 04/2010; 37(4):1554-64. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: This paper reports on an evaluation of 5 RapidArc optimization approaches vs IMRT. This study includes 11 patients with adenocarcinoma of the prostate. Rectal Normal Tissue Complication Probability is used as a constraint in a dose escalation. RapidArc rectal NTCP's are lower than those of IMRT (p = 0.007). This allows a mean dose escalation of 2.1 Gy([0.7 Gy,3.5 Gy]).
Radiotherapy and Oncology 02/2010; 95(2):149-52. · 5.58 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: To determine the dependence of celecoxib on the tumour micro-environment in vitro and in vivo and to compare the use of (18)F-Fluorodeoxyglucose ((18)F-FDG) and (18)F- 3'-deoxy-3-fluorothymidine ((18)F-FLT) to measure tumour response.
In vitro, colony assays were performed on a cyclo-oxygenase 2 (COX-2) negative (HCT116) and a COX-2 positive cell line (HCA7). Xenograft models of these cell lines were treated with celecoxib and/or radiotherapy. Micro Positron Emission Tomography (microPET) scans with (18)F-FDG and (18)F-FLT were performed at different time-points.
In vitro, no radiosensitising effect was seen in either of the cell lines. In vivo results showed a significant effect of celecoxib in the COX-2 negative tumours (HCT116) (enhancement ratio 1.5, p = 0.02) while no significant effect was observed in the COX-2 positive model (HCA7). A good correlation between (18)F-FDG and (18)F-FLT uptake was seen in both tumour models (r = 0.48, p = 0.002; r = 0.41, p = 0.005). After irradiation, a decrease in the uptake of both tracers was observed in both tumour models, which was more pronounced in the combination group, confirming the growth delay data.
The contradicting in vitro and in vivo results suggest a major role of the tumour micro-environment. (18)F-FLT seems a good alternative for (18)F-FDG to follow tumour growth after radiation treatment.
International Journal of Radiation Biology 10/2009; 85(9):763-71. · 2.28 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: Currently, most available patient alignment tools based on implanted markers use manual marker matching and rigid registration transformations to measure the needed translational shifts. To quantify the particular effect of prostate gland shrinkage, implanted gold markers were tracked during a course of radiotherapy including an isotropic scaling factor to model prostate shrinkage.
Eight patients with prostate cancer had gold markers implanted transrectally and seven were treated with (neo) adjuvant androgen deprivation therapy. After patient alignment to skin tattoos, orthogonal electronic portal images (EPIs) were taken. A semi-automated 2D/3D marker-based registration was performed to calculate the necessary couch shifts. The registration consists of a rigid transformation combined with an isotropic scaling to model prostate shrinkage.
The inclusion of an isotropic shrinkage model in the registration algorithm cancelled the corresponding increase in registration error. The mean scaling factor was 0.89+/-0.09. For all but two patients, a decrease of the isotropic scaling factor during treatment was observed. However, there was almost no difference in the translation offset between the manual matching of the EPIs to the digitally reconstructed radiographs and the semi-automated 2D/3D registration. A decrease in the intermarker distance was found correlating with prostate shrinkage rather than with random marker migration.
Inclusion of shrinkage in the registration process reduces registration errors during a course of radiotherapy. Nevertheless, this did not lead to a clinically significant change in the proposed table translations when compared to translations obtained with manual marker matching without a scaling correction.
Radiotherapy and Oncology 12/2008; 90(3):331-6. · 5.58 Impact Factor
-
Medical Physics 11/2008; 35(10):4285-8. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: This article provides a quantitative evaluation of Varian Medical Systems' beam matching procedure. A one-dimensional y analysis is employed to investigate the level of agreement of matched beams. A customized concept of one-dimensional gamma evaluation was designed. Our algorithm first performs a "local" fit of the reference and the evaluated datasets. For a particular point on the fitted evaluated curve, the y is derived as the shortest distance between the point and the fitted reference curve. This approach removes variations of the obtained y value related to the discrete character and noise in the original datasets. Criteria of 1 mm distance-to-agreement and 1% dose difference were used to evaluate the level of agreement of according profiles. Relative point and profile measurements were performed for all photon and electron beams of two Varian Clinacs 2100C/D. Matched beams show a good level of agreement. 70% of profiles completely pass the chosen criteria. The analysis of remaining 30% of the profiles demonstrates that measurement error becomes a limiting factor in achieving a better score. The highest obtained y value was 1.70. The quality of beam matching allowed us to treat according beams of both treatment units as "identical" and to use the reference beam data for the new unit. Nevertheless, the vendor's acceptance criteria of beam matching are much more benevolent. It might happen that the acceptance criteria are met, however, resulting quality of beam matching does not allow full interchangeability of beams.
Medical Physics 08/2007; 34(7):2917-27. · 2.83 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: Postimplant dosimetry of prostate seed implants is usually performed by seed localisation on transversal CT or MR images. In order to obtain reliable dosimetric evaluation data, it is important that seeds are reconstructed accurately. Currently, there is no comparative data available on seed localisation accuracy of CT-and MRI-based reconstructions, mainly due to the lack of a suitable QA tool. In this study, we developed a CT-and MRI compatible prostate phantom to investigate the intrinsic accuracy of seed detection for both imaging modalities.
A 60 seed geometry was created according to a clinically meaningful plan, including rotated and shifted seeds. After implantation of the seeds in the phantom, CT and MRI scans with 3, 4 and 5mm slice thickness were performed. The seed locations were reconstructed in the treatment planning system and compared with the known reference positions.
Due to the comparable density and relaxation times of the phantom material to prostate tissue, the seeds are visualised similarly as on real patient images. The observed mean reconstruction uncertainties were in general smaller for CT (0.9+/-0.6, 0.9+/-0.6, 2.1+/-0.8 mm on 3, 4 and 5mm scans, respectively), than for MRI (Philips 1.5 T: 2.1+/-1.4, 1.6+/-1.2, 1.9+/-0.9 mm on 3, 4 and 5 mm scans, respectively, and Siemens 1.5 T: 2.3+/-0.8, 2.0+/-1.6, 1.6+/-0.8 mm on 3, 4 and 5mm scans, respectively).
For our clinical sequences of both CT and MRI, the mean deviation of the reconstructed seed positions were all within acceptable limits for clinical use (<2.3 mm). The phantom was found to be a suitable quality assurance tool to assess the reliability and accuracy of the seed reconstruction procedure. Moreover, as the phantom material has the same imaging characteristics as real prostate tissue, it is a useful device to define proper MRI sequences.
Radiotherapy and Oncology 05/2006; 79(2):190-7. · 5.58 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: To implement an on-line correction scheme based on implanted markers to reduce treatment margins in external beam radiation therapy (EBRT) of carcinoma of the prostate. In turn reduction in treatment margins reduces irradiated volumes and offers the possibility of reduced normal tissue complications or escalated target dose.
Five or six gold markers were implanted in 10 patients treated for prostate carcinoma using EBRT. All patients were enlisted in an IRB-approved protocol. Before each fraction two portal images were obtained using a low dose (2MU). Positions of the markers were calculated from these images using an in-house developed program. Corrections were applied with a threshold of 2mm displacement. After correction the procedure was repeated.
Overall systematic errors were reduced from 7.45, 1.29, and 5.12 mm to 0.65, 0.11, and 0.46 mm in, respectively, the antero-posterior, lateral, and cranio-caudal directions. Likewise, the overall SD were reduced from 5.99, 5.34, and 4.44 mm to 2.82, 2.64, and 2.22 mm, respectively. All reductions were highly significant (P < 0.01) using a t-test for systematic and an F-test for random errors. On an individual level all but three patients showed significant improvements in all directions for the random errors. All patients improved in at least one direction. Systematic errors were significantly lower in all patients. Simulated correction schemes using this data suggest that margin reduction using off-line reduction does not benefit substantially from on-line corrections in the first few fractions.
Use of marker-based correction improves the patient position. Factors influencing the accuracy were: (1) number of seeds usable for correction, (2) distribution of markers throughout the volume of interest, and (3) objective instructions for patient realignment.
Radiotherapy and Oncology 05/2006; 79(1):94-100. · 5.58 Impact Factor
-
[show abstract]
[hide abstract]
ABSTRACT: The aim of this paper is to study the correction of prostate motion and position during external beam therapy. The correction was performed using a commercially available ultrasound-based repositioning tool. Electronic portal imaging with the use of fiducial markers was used to assess efficacy and accuracy. Patients undergoing radiation treatment for adenocarcinoma of the prostate were enrolled in a positioning study. Fifteen patients had five to six gold fiducial markers implanted in their prostate. These patients were positioned daily in a standard manner and then were repositioned every other day using an ultrasound-based correction system. Every fraction of a patients' treatment was imaged. This yielded 156 image pairs with and 119 pairs without repositioning available for analysis. This group of patients with markers had the following residual positions measured after the use of ultrasound repositioning. A mean error of -0.4 mm (LL), -2.6 mm (CC), and +2.5 mm (AP) with a standard deviation of 4.3, 5.4, and 5.7 mm. In two directions the improvements of treatment using the ultrasound correction were smaller than the precision of this experiment. They were no larger than 0.81 mm (LAT), and 0.95 mm (CC). In the AP direction a significant improvement was found of 1.6 mm. A highly significant correlation (p < 0.001) was found between the residual errors in the cranio-caudal direction and the shifts performed on the basis of the ultrasound measurements (Spearman ranking R = 0.53). We presented a method to objectively estimate improvements by a correction scheme. This method applied to ultrasound-based adjustment showed significant improvement in one direction and no measurable improvement in two other directions.
Medical Physics 11/2003; 30(11):2878-87. · 2.83 Impact Factor