Article

Corneal and epithelial thickness changes after 4 weeks of overnight corneal refractive therapy lens wear, measured with optical coherence tomography.

Centre for Contact Lens Research, School of Optometry, University of Waterloo, Waterloo, Ontario, N2L 3G1 Canada.
Eye & Contact Lens Science & Clinical Practice (Impact Factor: 1.68). 11/2004; 30(4):189-93; discussion 205-6. DOI: 10.1097/01.ICL.0000140223.60892.16
Source: PubMed

ABSTRACT To investigate thickness changes of the total cornea and epithelium across the horizontal corneal meridian after 4 weeks of overnight corneal refractive therapy (CRT) rigid contact lens (Paragon Vision Sciences, Mesa, AZ) wear.
Thirty subjects were fitted with CRT contact lenses (Dk/t = 67), which were worn overnight for 4 weeks. Corneal thickness was measured at nine locations along the horizontal meridian by using optical coherence tomography (OCT) before lens insertion in the evening. Corneal thickness was measured the next morning immediately after lens removal and 1, 3, 7, and 14 hours later. This was repeated on days 4, 10, and 28 of the study and then 3 days after discontinuing lens wear.
Twenty-three subjects completed the study. At lens removal on day 1, the central and paracentral cornea swelled by 4.9% and 6.2%, respectively (both P = 0.000). The central epithelium thinned by 7.3%, and the mid peripheral epithelium thickened by 13% (both P = 0.000). Corneal swelling recovered throughout the day, with most of the deswelling taking place within the first 3 hours after lens removal. Maximal central epithelial thinning reached 13.5% by day 4. Three days after the study completion, corneal and epithelial thickness had recovered to baseline values.
This study shows that CRT lenses induce differential overnight swelling across the cornea, with rapid deswelling during the day. Central epithelial thinning and paracentral thickening occurs, with recovery 3 days after discontinuation of lens wear.

0 Followers
 · 
66 Views
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: La medida del edema corneal es un parámetro ampliamente utilizado para valorar el impacto del uso de lentes de contacto sobre la superficie ocular con diferentes técnicas (paquimetría ultrasónica, ópticas, microscopía confocal, análisis de sección óptica o tomografía óptica de coherencia). Existe controversia respecto a diferencias en el grado de edema central o periférico con trabajos que proponen una edematización uniformemente, otros un edema menor en el centro, pero la opinión más extendida es que el edema es mayor en el centro corneal. Esta tesis ha validado técnicas de medida y monitorización no invasiva del edema corneal inducido por el uso continuado de lentes de contacto, demostrando que no existen diferencias significativas entre el grado de edema central o periférico. Esta información permitiría una mejor selección del material de la lente, mejorar sus recomendaciones de uso, readaptar a usuarios con complicaciones incipientes y minimizar la incidencia de las complicaciones.
  • [Show abstract] [Hide abstract]
    ABSTRACT: PurposeTo calibrate the new ZEISS VisanteTM anterior segment optical coherence tomographer (OCT) using references with known physical thickness and refractive index equal to the human cornea and to compare the Visante measures to those from a previous generation OCT (Zeiss-Humphrey OCT II).Methods Twenty two semi-rigid lenses of specified thicknesses were manufactured using a material with refractive index of 1.376. Central thickness of these lenses was measured using VisanteTM and Zeiss-Humphrey OCT II OCT's (Zeiss, Germany). Two data sets consisting of nominal measures (with a standard pachymeter) of the lenses and one obtained using a digital micrometer was used as references. Regression equations between the new physical and optical (OCT) measures were derived to calibrate the devices.ResultsBefore calibration, repeated measures ANOVA showed that there were significant differences between mean lens thicknesses from each of the measurement methods (p < 0.01), where Visante measurements were significantly different from the other three (OCT II, MG and OP) methods (p < 0.001). Visante thickness was significantly higher than the microgauge measures (453 ± 37.6 compared to 445.1 ± 38.2) and the OCT II was significantly lower (424.5 ± 36.1 both, p < 0.001). After calibration using the regressions between physical and optical measurements, there were no differences between OCT II and Visante (p < 0.05).Conclusions Using references lenses with refractive index of the cornea (1.376) allows rapid and simple calibration and cross calibration of instruments for measuring the corneal thickness. The Visante and OCT II do not produce measurements that are equal to physical references with refractive index equal to the human cornea.ResumenObjetivoCalibrar el nuevo tomógrafo de coherencia óptica (OCT) del segmento anterior ZEISS VisanteTM utilizando referencias con índice de refracción y espesor físico conocido equivalentes a la córnea humana y comparar las medidas del Visante con las del OCT de la generación anterior (Zeiss-Humphrey OCT II).MétodosSe fabricaron 22 lentes semirrígidas de espesores específicos utilizando un material con un índice de refracción de 1,376. El espesor central de estas lentes se midió con los OCT VisanteTM y Zeiss-Humphrey OCT II (Zeiss, Alemania). Como referencias se utilizaron dos conjuntos de datos compuestos de medidas nominales (con un paquímetro estándar) de las lentes, uno de ellos obtenido mediante un micrómetro digital. Para calibrar los dispositivos se derivaron las ecuaciones de regresión entre las nuevas medidas físicas y ópticas (OCT).ResultadosAntes de la calibración, las mediciones repetidas con el ANOVA mostraron que había diferencias significativas entre las medias de espesor de la lente a partir de cada método de medición (p < 0,01), en los cuales las mediciones con Visante fueron significativamente diferentes de los otros tres métodos (OCT II, MG y OP) (p < 0,001). El espesor con Visante fue significativamente mayor que en las mediciones con micrómetro mecánico (453 ± 37,6 en comparación con 445,1 ± 38,2) y con el OCT II fue significativamente inferior (424,5 ± 36,1 ambos, p < 0,001). Después de la calibración utilizando las regresiones entre las mediciones físicas y ópticas, no hubo diferencias entre el OCT II y el Visante (p < 0,05).ConclusiónEl uso de lentes de referencia con índice de refracción equivalente al de la córnea (1,376) permite calibrar e intercalibrar rápida y fácilmente los instrumentos para medir el espesor corneal. El Visante y el OCT II no proporcionan mediciones equivalentes a las referencias físicas con un índice de refracción equivalente al de la córnea humana.
    Journal of Optometry 10/2011; 4(4):147–155. DOI:10.1016/S1888-4296(11)70057-7
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Purpose. Quantifying adaptation to light distortion of subjects undergoing orthokeratology (OK) for myopia during the first month of treatment. Methods. Twenty-nine healthy volunteers (age: 22.34 ± 8.08 years) with mean spherical equivalent refractive error -2.10 ± 0.93D were evaluated at baseline and days 1, 7, 15, and 30 of OK treatment. Light distortion was determined using an experimental prototype. Corneal aberrations were derived from corneal topography for different pupil sizes. Contrast sensitivity function (CSF) was analyzed for frequencies of 1.50, 2.12, 3.00, 4.24, 6.00, 8.49, 12.00, 16.97, and 24.00 cpd under photopic conditions. Results. Average monocular values of all light distortion parameters measured increased significantly on day 1, returning to baseline after 1 week (P < 0.05 in all cases). Spherical-like aberration stabilized on day 7 for all pupil diameters, while coma-like for smaller pupils only. CSF was significantly reduced on day 1 for all spatial frequencies except for 1.5 cpd, returning to baseline afterwards. Significant correlation was found between light distortion and contrast sensitivity for middle and high frequencies (P < 0.05) after 15 days. Conclusion. Despite consistently increased levels of corneal aberrations, light distortion tends to return to baseline after one week of treatment, suggesting that neural adaptation is capable of overcoming optical quality degradation.
    BioMed Research International 01/2015; 2015:278425. DOI:10.1155/2015/278425 · 2.71 Impact Factor