The use of live three-dimensional Doppler echocardiography in the measurement of cardiac output: an in vivo animal study.

Clinical Care Center for Congenital Heart Disease, Oregon Health and Science University, 3181 SW Sam Jackson Park Road, Portland, OR 97239, USA.
Journal of the American College of Cardiology (Impact Factor: 14.09). 03/2005; 45(3):433-8. DOI: 10.1016/j.jacc.2004.10.046
Source: PubMed

ABSTRACT The purpose of this study was to investigate whether cardiac output (CO) could be accurately computed from live three-dimensional (3-D) Doppler echocardiographic data in an acute open-chested animal preparation.
The accurate measurement of CO is important in both patient management and research. Current methods use invasive pulmonary artery catheters or two-dimensional (2-D) echocardiography or esophageal aortic Doppler measures, with the inherent risks and inaccuracies of these techniques.
Seventeen juvenile, open-chested pigs were studied before undergoing a separate cardiopulmonary bypass procedure. Live 3-D Doppler echocardiography images of the left ventricular outflow tract and aortic valve were obtained by epicardial scanning, using a Philips Medical Systems (Andover, Massachusetts) Sonos 7500 Live 3-D Echo system with a 2.5-MHz probe. Simultaneous CO measurements were obtained from an ultrasonic flow probe placed around the aortic root. Subsequent offline processing using custom software computed the CO from the digital 3-D Doppler DICOM data, and this was compared to the gold standard of the aortic flow probe measurements.
One hundred forty-three individual CO measurements were taken from 16 pigs, one being excluded because of severe aortic regurgitation. There was good correlation between the 3-D Doppler and flow probe methods of CO measurement (y = 1.1x - 9.82, R(2) = 0.93).
In this acute animal preparation, live 3-D Doppler echocardiographic data allowed for accurate assessment of CO as compared to the ultrasonic flow probe measurement.

0 Bookmarks
 · 
79 Views
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Objectives- The aim of this study was to assess the accuracy, feasibility, and reproducibility of determining stroke volume from a novel 3-dimensional (3D) color Doppler flow quantification method for mitral valve (MV) inflow and left ventricular outflow tract (LVOT) outflow at different stroke volumes when compared with the actual flow rate in a pumped porcine cardiac model. Methods- Thirteen freshly harvested pig hearts were studied in a water tank. We inserted a latex balloon into each left ventricle from the MV annulus to the LVOT, which were passively pumped at different stroke volumes (30-80 mL) using a calibrated piston pump at increments of 10 mL. Four-dimensional flow volumes were obtained without electrocardiographic gating. The digital imaging data were analyzed offline using prototype software. Two hemispheric flow-sampling planes for color Doppler velocity measurements were placed at the MV annulus and LVOT. The software computed the flow volumes at the MV annulus and LVOT within the user-defined volume and cardiac cycle. Results- This novel 3D Doppler flow quantification method detected incremental increases in MV inflow and LVOT outflow in close agreement with pumped stroke volumes (MV inflow, r = 0.96; LVOT outflow, r = 0.96; P < .01). Bland-Altman analysis demonstrated overestimation of both (MV inflow, 5.42 mL; LVOT outflow, 4.46 mL) with 95% of points within 95% limits of agreement. Interobserver variability values showed good agreement for all stroke volumes at both the MV annulus and LVOT. Conclusions- This study has shown that the 3D color Doppler flow quantification method we used is able to compute stroke volumes accurately at the MV annulus and LVOT in the same cardiac cycle without electrocardiographic gating. This method may be valuable for assessment of cardiac output in clinical studies.
    Journal of ultrasound in medicine: official journal of the American Institute of Ultrasound in Medicine 02/2014; 33(2):265-71. · 1.40 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: The pig has recently become popular as a large animal experimental model in many fields of biomedical research. The aim of this study is to evaluate the basic anatomical structures in the head region of the pig to lay the groundwork for its practical clinical usage or pre-clinical research in the future. We used three different diagnostic imaging methods: radiography, computed tomography (CT) and magnetic resonance imaging (MRI). The analysis showed that radiographic imaging is suitable only for general evaluation of the facial area of the pig skull. CT images showed excellent spatial definition of bony structures of the whole craniofacial area, and MRI images revealed fine soft tissue details. Radiography is preferentially suited to general assessment of bone structures of the facial skeleton; however, the thick layer of adipose tissue in the craniofacial region of the pig makes the imaging of some parts difficult or even impossible. CT is useful for revealing morphological details of mineralized tissues, whereas MRI is more suitable for soft tissue analysis and the detection of subtle pathologic changes in both bone and soft tissues. Therefore, before using pigs as an experimental model in craniofacial research, it is necessary to evaluate the suitability and disadvantages of potential imaging methods and how appropriate they are for accurate visualization of desired structures.
    Anantomia Histologia Embryologia 11/2013; · 0.88 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Accurate assessment of mitral regurgitation (MR) severity is crucial for clinical decision-making and optimizing patient outcomes. Recent advances in real-time three dimensional (3D) echocardiography provide the option of real-time full volume color Doppler echocardiography (FVCD) measurements. This makes it practical to quantify MR by subtracting aortic stroke volume from the volume of mitral inflow in an automated manner. Thirty-two patients with more than a moderate degree of MR assessed by transthoracic echocardiography (TTE) were consecutively enrolled during this study. MR volume was measured by 1) two dimensional (2D) Doppler TTE, using the proximal isovelocity surface area (PISA) and the volumetric quantification methods (VM). Then, 2) real time 3D-FVCD was subsequently obtained, and dedicated software was used to quantify the MR volume. MR volume was also measured using 3) phase contrast cardiac magnetic resonance imaging (PC-CMR). In each patient, all these measurements were obtained within the same day. Automated MR quantification was feasible in 30 of 32 patients. The mean regurgitant volume quantified by 2D-PISA, 2D-VM, 3D-FVCD, and PC-CMR was 72.1 ± 27.7, 79.9 ± 36.9, 69.9 ± 31.5, and 64.2 ± 30.7 mL, respectively (p = 0.304). There was an excellent correlation between the MR volume measured by PC-CMR and 3D-FVCD (r = 0.85, 95% CI 0.70-0.93, p < 0.001). Compared with PC-CMR, Bland-Altman analysis for 3D-FVCD showed a good agreement (2 standard deviations: 34.3 mL) than did 2D-PISA or 2D-VM (60.0 and 62.8 mL, respectively). Automated quantification of MR with 3D-FVCD is feasible and accurate. It is a promising tool for the real-time 3D echocardiographic assessment of patients with MR.
    Journal of cardiovascular ultrasound 06/2013; 21(2):81-9.