Analysis of skeletal movements in mandibular distraction osteogenesis

Harvard University, Cambridge, Massachusetts, United States
Journal of Oral and Maxillofacial Surgery (Impact Factor: 1.28). 04/2005; 63(3):335-40. DOI: 10.1016/j.joms.2004.06.057
Source: PubMed

ABSTRACT The purpose of this study was to use geometric parameters of movement, calculated from 3-dimensional computed tomography (CT) data, to determine the curvilinear distractor dimensions required to correct mandibular deformities in a series of patients.
Preoperative CT scans from 15 patients with symmetric (n = 5) and asymmetric (n = 10) deformities were imported into a CT-based software program (Osteoplan; an open-source visualization application developed by Gering et al at the Surgical Planning Laboratory [SPL, Brigham and Womens Hospital, Boston, MA]). The software was used to reconstruct virtual 3-dimensional models from these scans. Two experienced surgeons, working with a computer scientist, then used Osteoplan to create an ideal treatment plan for each patient. In each case, the 3-dimensional curvilinear movement was quantified using 4 "parameters of movement" (POMs). These parameters were then used to prescribe a distraction device capable of executing the planned skeletal correction. Curvilinear distractor dimensions calculated by Osteoplan included the radius of curvature of the prescribed device, and the distractor elongation, pitch, and handedness.
Treatment plans including POMs were developed for each patient. The radii of curvature for the prescribed distractors ranged from 2.3 to 14.1 cm, the distractor elongation dimensions ranged from 0.7 to 3.2 cm, and the pitch (horizontal plane) dimensions ranged from 0.005 to 0.8 cm. Handedness was either a left (n = 12) or right (n = 8) turning helix.
The results of this study indicate that, using geometric parameters of movement calculated from 3-dimensional CT scans, curvilinear devices could be prescribed for correction of the range of skeletal deformities in this group of patients.

Download full-text

Full-text

Available from: Ron Kikinis, Mar 01, 2015
2 Followers
 · 
172 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: A next generation compact, Ku-band, pulse compression radar altimeter (RA) measures sea surface topography to a few centimeters accuracy and provides scientific data for assessing ocean circulation, mean sea-level change and other ocean features. The integrated RA is comprised of a basic altimeter, a water vapor radiometer (WVR) for correcting RA range path delay errors, and a shared antenna system. The RA operates at 13.5 GHz with pulsed 7 watt chirp bandwidth of 320 MHz. The system characteristics, integrated RA/WVR design, and support equipment are discussed. The RA was launched on the US Navy GEOSAT follow-on spacecraft
    Southeastcon '96. 'Bringing Together Education, Science and Technology'., Proceedings of the IEEE; 05/1996
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: The purpose of this study was to determine the range of fixed trajectory curvilinear distraction devices required to correct a variety of severe mandibular deformities. Preoperative computed tomography (CT) scans from 18 patients with mandibular deformities were imported into a CT-based software program (Osteoplan). Three-dimensional virtual models of the individual skulls were made with landmarks to track movements. An ideal treatment plan was created for each patient. Upper and lower boundaries for the dimensions of curvilinear distractors were established based on manufacturing and geometric constraints. Then, anatomically acceptable distractor attachment points were identified on the models using proximal and distal grids. Treatment plans were simulated for a series of distractors with varying radii of curvature, elongations (arc-length of device), and placements along the grids. The outcomes using these distractors were compared with the ideal treatment plans. Discrepancies were quantified in millimeters by comparing landmarks in the simulated versus ideal movements. Approximately 400,000 simulated 3-dimensional movements, based on the distractor parameters and variations in placement were computationally evaluated for the 18 cases. It was determined that, by varying distractor placement, a family of 5 distractors, with 3, 5, 7, and 10 cm radii of curvature and a straight-line device, could be used to treat all 18 cases to within 1.8 mm of error. The results of this study indicate that a family of 5 curvilinear distractors may suffice to treat a broad range of mandibular deformities.
    Journal of Oral and Maxillofacial Surgery 03/2006; 64(2):259-64. DOI:10.1016/j.joms.2005.10.015 · 1.28 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: The purpose of this study was to retrospectively evaluate the accuracy and reliability of a computer simulation system that predicts the results of an internal curvilinear mandibular distractor in the treatment of hemifacial microsomia and severe mandibular retrognathia. A two-dimensional computer analysis and simulation system was developed based on cephalometric radiographs and patient photographs. The software was used to generate and superimpose digital templates, corresponding to the actual distractor and logarithmic spiral, onto the lateral cephalometric tracings. The digital distractor's template was then adjusted to fit the radiographic projection of the actual distractor on the 1-week postoperative lateral cephalograph as precisely as possible. It was then activated along the spiral growth curve template until the ideal position of the mandible was reached. This computational surgical simulation tracing was then superimposed onto the actual postdistraction lateral cephalometric radiograph and tracing. Thus, the predicted and actual mandibular distraction movements were compared. In all of the cases, the computational distraction (skeletal and soft tissues) tracing fit the actual postdistraction tracing with an accuracy to within 1 mm. The current study demonstrated that a specific curvilinear distraction can (1) closely reproduce the natural logarithmic spiral movement of the mandibular growth, (2) be closely approximated and thus predicted by a computer simulation system, and (3) be used for correction of mandibular deformities such as hemifacial microsomia and severe mandibular retrognathia.
    Plastic and Reconstructive Surgery 01/2008; 120(7):1975-80. DOI:10.1097/01.prs.0000287327.66824.2f · 3.33 Impact Factor