In vitro assessment of image-guided otologic surgery: submillimeter accuracy within the region of the temporal bone.

Department of Otolaryngology-Head and Neck Surgery, Vanderbilt University, Nashville, TN, USA.
Otolaryngology Head and Neck Surgery (Impact Factor: 1.72). 04/2005; 132(3):435-42. DOI: 10.1016/j.otohns.2004.09.141
Source: PubMed

ABSTRACT Application of image-guided surgery to otology has been limited by the need for submillimeter accuracy via a fiducial system that is easily usable (noninvasive and nonobstructive).
A dental bite-block was fitted with a rigid frame with 7 fiducial markers surrounding each external ear. The temporal bones of 3 cadaveric skulls were removed and replaced with surgical targets arranged in a bull's-eye pattern about the centroid of each temporal bone. The surgical targets were identified both within CT scans and in physical space using an infrared optical tracking system. The difference between positions in CT space versus physical space was calculated as target registration error.
A total of 234 independent target registration errors were calculated. Mean +/- standard deviation = 0.73 mm +/- 0.25 mm.
These findings show that image-guided otologic surgery with submillimeter accuracy is achievable with a minimally invasive fiducial frame. Significance In vivo validation of the system is ongoing. With such validation, this system may facilitate clinically applicable image-guided otologic surgery. EBM rating: A.

0 Bookmarks
 · 
115 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: Due to the importance of surgery in the medical field, a large amount of research has been conducted in this area. Imaging and robotics technologies provide surgeons with the advanced eye and hand to perform their surgeries in a safer and more accurate manner. Recently medical images have been utilized in the operating room as well as in the diagnostic stage. If the image to patient registration is done with sufficient accuracy, medical images can be used as “a map” for guidance to the target lesion. However, the accuracy and reliability of the surgical navigation system should be sufficiently verified before applying it to the patient. Along with the development of medical imaging, various medical robots have also been developed. In particular, surgical robots have been researched in order to reach the goal of minimal invasiveness. The most important factors to consider are determining the demand, the strategy for their use in operating procedures, and how it aids patients. In addition to the above considerations, medical doctors and researchers should always think from the patient's point of view. In this article, the latest medical imaging and robotic technologies focusing on surgical applications are reviewed based upon the factors described in the above.
    International Journal of Optomechatronics 10/2011; 5(4):299-321. · 0.57 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: sion 2 planning station. The route from the surface of the mastoid to the scala tympani of the cochlea was planned as a direct channel passing the facial recess without injuring the facial nerve and other functionally 1 Klinik und Poliklinik für Hals- Nasen-Ohrenheilkunde, important anatomical structures of the temporal bone. During surgery the preoperatively defined trajectory was followed from the entry point Medizinische Hochschule to the target point using a navigated and hand-held surgical drill. MSCT
    01/2007;
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Image-guided surgery systems are limited by registration error, so practical and effective methods to improve accuracy are necessary. A projection point-based method for reducing the surface registration error in image-guided surgery was developed and tested.
    International Journal of Computer Assisted Radiology and Surgery 05/2014; · 1.36 Impact Factor

Full-text (2 Sources)

Download
21 Downloads
Available from
May 27, 2014