[Value of fetal scalp lactate sampling during labour: a comparative study with scalp pH].

Service de gynécologie-obstétrique, CHR Saint-Jacques, avenue du 8-Mai-1945, 25030 Besançon cedex, France.
Gynécologie Obstétrique & Fertilité (impact factor: 0.52). 03/2005; 33(3):107-12. DOI:10.1016/j.gyobfe.2005.01.004
Source: PubMed

ABSTRACT To evaluate the validity and the feasibility of fetal scalp lactate sampling during labour in the assessment of non reassuring fetal status.
A prospective observational study was conducted on 129 fetal scalp blood samplings. Scalp lactate measurements were compared to scalp pH (gold standard procedure), neonatal cord blood gas (pH and lactates) and Apgar score. Pathological values taken were for scalp lactates>5 mmol/l, scalp pH<7.20, cord arterial pH<7.10, cord arterial lactates>6,35 mmol/l, and Apgar score<7 at one and five minutes. All measurements were performed using the Rapid Lab 860 device from Bayer.
Scalp lactate correlated significantly with scalp pH (R=-0.54, P=0.001), with umbilical artery pH (R=-0.46, P=0,01), with umbilical artery lactate (P=0.49, P=0.01), but with neither Apgar score at one minute (R=-0.21, ns) nor at 5 minutes (R=-0.11, ns). Scalp lactate at a cut-off value of 5 mmol/l had the same predictive values than scalp pH at 7.20 to predict neonatal acidosis. Sampling failure with scalp lactate was inferior to 1 vs 18% for scalp pH.
The measurement of lactate in fetal blood scalp seems correlated to the fetal scalp pH. It may be an attractive alternative to pH analysis and a useful tool for monitoring fetal asphyxia, especially with the advent of handheld devices requiring small sample volumes.

0 0
 · 
0 Bookmarks
 · 
29 Views

Keywords

129 fetal scalp blood samplings
 
Apgar score
 
cord arterial lactates>6,35 mmol/l
 
fetal blood scalp
 
fetal scalp lactate sampling
 
fetal scalp pH
 
monitoring fetal asphyxia
 
neonatal acidosis
 
Pathological values
 
predictive values
 
prospective observational study
 
scalp lactate
 
Scalp lactate correlated
 
Scalp lactate measurements
 
scalp lactates>5 mmol/l
 
scalp pH
 
small sample volumes
 
umbilical artery lactate
 
umbilical artery pH
 
useful tool
 