Article

In vivo-in vitro toxicogenomic comparison of TCDD-elicited gene expression in Hepa1c1c7 mouse hepatoma cells and C57BL/6 hepatic tissue.

Department of Biochemistry & Molecular Biology, Michigan State University, East Lansing MI 48824-1319, USA.
BMC Genomics (impact factor: 4.07). 02/2006; 7:80. DOI:10.1186/1471-2164-7-80 pp.80
Source: PubMed

ABSTRACT In vitro systems have inherent limitations in their ability to model whole organism gene responses, which must be identified and appropriately considered when developing predictive biomarkers of in vivo toxicity. Systematic comparison of in vitro and in vivo temporal gene expression profiles were conducted to assess the ability of Hepa1c1c7 mouse hepatoma cells to model hepatic responses in C57BL/6 mice following treatment with 2,3,7,8-tetrachlorodibenzo-p-dioxin (TCDD).
Gene expression analysis and functional gene annotation indicate that Hepa1c1c7 cells appropriately modeled the induction of xenobiotic metabolism genes in vivo. However, responses associated with cell cycle progression and proliferation were unique to Hepa1c1c7 cells, consistent with the cell cycle arrest effects of TCDD on rapidly dividing cells. In contrast, lipid metabolism and immune responses, representative of whole organism effects in vivo, were not replicated in Hepa1c1c7 cells.
These results identified inherent differences in TCDD-mediated gene expression responses between these models and highlighted the limitations of in vitro systems in modeling whole organism responses, and additionally identified potential predictive biomarkers of toxicity.

0 0
 · 
0 Bookmarks
 · 
29 Views
  • Source
    Article: Correcting log ratios for signal saturation in cDNA microarrays.
    [show abstract] [hide abstract]
    ABSTRACT: MOTIVATION: Pixel saturation occurs when the pixel intensity exceeds a threshold and the recorded pixel intensity is truncated. Microarray experiments are commonly afflicted with saturated pixels. As a result, estimators of gene expression are biased, with the amount of bias increasing as a function of the proportion of pixels saturated. Saturation is directly related to the photomultiplier tube (PMT) voltage settings and RNA abundance and is not necessarily associated with poor array or poor spot quality. When choosing PMT settings, higher PMT settings are desired because of improved signal-to-noise ratios of low-intensity spots. This improved signal is somewhat offset by saturation of high-intensity spots. In practice, spots with saturated pixels are discarded or the biased value is used. Neither of these approaches is appealing, particularly the former approach when a highly expressed gene is discarded because of saturation. RESULTS: We present a method to correct for saturation using pixel-level data. The method is based on a censored regression model. Evaluations on several arrays indicate that the method performs well. Simulation studies suggest that the method is robust under certain model violations.
    Bioinformatics 12/2004; 20(16):2685-93. · 5.47 Impact Factor
  • Article: Correcting for signal saturation errors in the analysis of microarray data.
    [show abstract] [hide abstract]
    ABSTRACT: A variety of technical errors have arisen in data analysis when using cDNA or oligonucleotide microarrays. One of the most insidious problems is the saturation of the hybridization signal of high-abundant transcripts. This problem arises from the truncation of the laser fluorescence signal. When the hybridization signal on the microarray is very strong, this truncation can result in serious consequences that may not be readily apparent to the user. As an illustration of this problem, two subclasses of normal human tissue samples (six liver and six lung samples) were analyzed with GeneChip probe arrays to evaluate the patterns of expression for approximately 7000 human genes. Five of these data sets were found to suffer from signal truncation. This caused several tissues to be incorrectly classified using hierarchical clustering. To rectify this problem so that the gene expression data could be properly compared and clustered, we developed a "filtering" procedure that identifies a subset of genes least affected by the signal saturation. This filtering procedure can be obtained at www.hugeindex.org.
    BioTechniques 03/2002; 32(2):330-2, 334, 336. · 2.67 Impact Factor
  • Article: Accuracy and calibration of commercial oligonucleotide and custom cDNA microarrays
    [show abstract] [hide abstract]
    ABSTRACT: We compared the accuracy of microarray measurements obtained with oligonucleotide arrays (GeneChip, Affymetrix) with a laboratory-developed cDNA array by assaying test RNA samples from an experiment using a paradigm known to regulate many genes measured on both arrays. We selected 47 genes represented on both arrays, including both known regulated and unregulated transcripts, and established reference relative expression measurements for these genes in the test RNA samples using quantitative reverse transcriptase realtime PCR (QRTPCR) assays. The validity of the reproducible (average coefficient of variation = 11.8%) QRTPCR measurements were established through application of a new mathematical model. The performance of both array platforms in identifying regulated and non-regulated genes was identical. With either platform, 16 of 17 definitely regulated genes were correctly identified, and no definitely unregulated transcript was falsely identified as regulated. Accuracy of the fold-change measurements obtained with each platform was assessed by determining measurement bias. Both platforms consistently underestimate the relative changes in mRNA expression between experimental and control samples. The bias observed with cDNA arrays was predictable for fold-changes <250-fold by QRTPCR and could be corrected by the calibration function F c = F a(cDNA) ,whereF a(cDNA) is the microarray-determined foldchange comparing experimental with control samples, q is the correction factor and F c is the calibrated value. The bias observed with the commercial oligonucleotide arrays was less predictable and calibration was unfeasible. Following calibration, fold-change measurements generated by custom cDNA arrays were more accurate than those obtained by commercial oligonucleotide ar...
    06/2002;

Full-text (2 Sources)

View
0 Downloads
Available from

Keywords

cell cycle arrest effects
 
cell cycle progression
 
functional gene annotation
 
Gene expression analysis
 
Hepa1c1c7 cells
 
Hepa1c1c7 cells appropriately modeled
 
Hepa1c1c7 mouse hepatoma cells
 
immune responses
 
lipid metabolism
 
model hepatic responses
 
model whole organism gene responses
 
modeling whole organism responses
 
models
 
potential predictive biomarkers
 
predictive biomarkers
 
TCDD-mediated gene expression responses
 
vitro systems
 
vivo temporal gene expression profiles
 
whole organism effects
 
xenobiotic metabolism genes
 

Edward Dere