Article

Comparison of CPAP titration at home or the sleep laboratory in the sleep apnea hypopnea syndrome.

Department of Sleep Medicine, Royal Infirmary of Edinburgh, Scotland, UK.
Sleep (Impact Factor: 5.06). 12/2006; 29(11):1451-5.
Source: PubMed

ABSTRACT Continuous positive airway pressure (CPAP) for the treatment of obstructive sleep apnea hypopnea syndrome (OSAHS) is conventionally started after in-laboratory overnight titration. This use of sleep laboratory space is both costly and limits access for diagnostic studies. This study aimed to evaluate whether automated CPAP titration in the home produced patient outcomes equal to those following laboratory-based automated CPAP titration. The main outcomes were Epworth Sleepiness Scale score, objective daytime sleepiness (Oxford SLEep Resistance test or OSLER test), and CPAP use; we also performed quality-of-life questionnaires: Functional Outcomes of Sleep Questionnaire and SF-36.
Prospective, randomized, single-blind, parallel-group, controlled trial
Regional sleep center and patients' homes.
Two hundred CPAP-naïve patients with OSAHS requiring CPAP treatment.
One hundred patients were randomly assigned to a standard 1-night in-hospital CPAP titration and 100 to 3 nights' home CPAP titration and then issued with fixed pressure CPAP. Data were analyzed on an intention-to-treat basis.
The patient groups did not differ at baseline. The CPAP pressures defined at titration (mean+/- SEM: 10.6+/-0.2, 10.4+/-0.2 cm H20, p = .19), number of mask leaks, and initial acceptance rates were similar in the sleep-laboratory and home-titrated groups. At 3-month follow-up, there was no significant difference in CPAP use (mean+/-SEM: 4.39+/-0.25, 4.38+/-0.25 h/night; p > .9), Epworth Sleepiness Scale score (9.5+/-0.5, 8.5+/-0.5, p = .14), OSLER, Functional Outcomes of Sleep Questionnaire, or SF-36 between the sleep-laboratory and home-titrated groups.
Home-based automated CPAP titration is as effective as automatic in-laboratory titrations in initiating treatment for OSAHS.

0 Followers
 · 
244 Views
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Study objectives: To investigate the need for pressure change in a larger group of patients with sleep disordered breathing (SDB) several weeks after initial CPAP (continuous positive airway pressure)/Bilevel-CPAP and auto-CPAP titration. Design: Prospective cohort study of consecutive patients with sleep disordered breathing and CPAP/Bilevel-CPAP/auto-CPAP therapy. Setting: Sleep laboratory in a regional hospital. Patients: 905 patients (740 m, 165 f) with SDB confirmed by polysomnography (PSG) and therapeutic intervention with CPAP/Bilevel-CPAP. Measurements and interventions: After baseline PSG, patients underwent in laboratory PSG for titration of CPAP, Bilevel-CPAP or auto-CPAP. Several weeks (mean 11) after CPAP initiation in a single night PSG patients were restudied and the differences in CPAP pressure between initial treatment night and re-titration night recorded. Anthropometric data, PSG data and Epworth Sleepiness scale and Berlin Questionnnaire scores were correlated to changes in CPAP pressure of first and control titration. Results: Pressure change was needed in the majority of patients (58,2 %, n 511). Pressure increase was more frequent than pressure recuction [411 (46,8 %) vs. 100 (11,4 %)]. Mean pressure increase in CPAP was 1,3 mbar, mean decrease 1,6 mbar, in Bilevel-CPAP mean increase in inspiratory pressure was 1,3 mbar, in exspiratory pressure 0,8 mbar, decrease in inspiratory pressure 1,9 mbar and 1,4 mbar in exspiratory pressure. No correlation was found between anthropometric data of patients, sleep efficacy or questionnaire scores and pressure change. Conclusion: Pressure changes are necessary in the majority of patients several weeks after CPAP initiation. Therefore re-evaluation of CPAP patients in a certain time frame after CPAP is useful.
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: La CPAP es el tratamiento de elección en pacientes con SAHS. Los pacientes con un IAH > 5 y síntomas, son susceptibles de tratamiento con CPAP nasal. Si no tienen síntomas, con patología cardiovascular, se debe considerar el tratamiento si el IAH >30, especialmente en pacientes con edad inferior a 70 años. Puesto que no es un tratamiento curativo, debe usarse de forma continuada, y para esto debemos hacer un ajuste de la presión óptima, para la que es imprescindible una adecuada información y preparación del enfermo, pudiendo realizarse diversas técnicas para el ajuste. La PSG completa en el laboratorio de sueño es la que mejor asegura la corrección de todos los fenómenos patológicos (respiratorios y neurológicos). Pero debido a lo costoso y largo del procedimiento y a las largas listas de espera se han propuesto otras opciones como son la PSG en noche partida para pacientes con IAH>20, el ajuste empírico mediante fórmula matemática, que puede usarse de manera provisional, a la espera de titulación, y un sistema de Auto-CPAP validado y con análisis visual (Normativa SEPAR 2010). Uno de los procedimientos más extendidos, para titular la presión de CPAP es el uso de equipos de CPAP automática. Existe una gran diversidad de equipos, con comportamientos diferentes, ya que utilizan un algoritmo de presión diferente, y variables distintas. Sin embargo, en otros estudios clínicos no se han encontrado diferencias. Estos, miden varias variables respiratorias, para detectar anormalidades respiratorias. Básicamente existen 3 variables de medida: flujo, ronquido e impedancia (valorada mediante oscilación forzada). Aunque inicialmente no todos los equipos utilizaban todas las variables, hoy la mayoría utilizan todas ellas con diferentes algoritmos.
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: To improve adherence to continuous positive airway pressure (CPAP) treatment in participants with obstructive sleep apnea (OSA) using a cognitive behavioral therapy (CBT) intervention. A randomized controlled trial. A major teaching hospital in Sydney (2005). One hundred individuals (96 men), ranging in age from 32 to 81 years, diagnosed with OSA. Two 1-hour CBT interventions (including a video of real CPAP users) plus treatment as usual (mask fitting and information) or treatment as usual only. Hours of CPAP usage was assessed at 7 nights and 28 nights. Adherence was defined as usage at least 4 hours per night. Questionnaires measuring self-efficacy, social support, and expectancy (mediators of adherence) were given after intervention or after usual treatment. A higher adherence to CPAP therapy was found in the CBT group (2.9 hours difference) relative to treatment as usual (P < 0.001) at 28 days. Only 4 participants in the CBT group did not initiate treatments after their titration study, compared with 15 in the treatment as usual group (P < 0.02). The CBT group had significantly higher scores for self-efficacy (P < 0.001) and social support P < 0.008) but not for expectancy. The CBT intervention resulted in both increased adherence and "uptake" of CPAP and therefore would be expected to reduce the social, economic, and health-related consequences of untreated OSA.
    Sleep 05/2007; 30(5):635-40. · 5.06 Impact Factor