System for intraoperative evaluation of soft-tissue-generated forces during total hip arthroplasty by measurement of the pressure distribution in artificial joints.

Institute for High Dimensional Medical Imaging, Jikei University School of Medicine, Tokyo, Japan.
Computer Aided Surgery (impact factor: 0.3). 02/2007; 12(1):53-9. DOI:10.3109/10929080701210881 pp.53-9
Source: PubMed

ABSTRACT A system for evaluating the soft-tissue-generated forces at the hip joint was developed. The system enabled measurement of contact pressure distribution at hip joint surfaces, as well as evaluation of the artificial hip joint condition during total hip arthroplasty (THA). First, a pressure sensor module that forms part of the artificial joint was constructed. Eight small pressure sensors were installed in the spherical head component of the ball-and-socket joint. Next, software for recording and visualizing the detected pressures at 1-millisecond intervals was developed. The pressure distribution was displayed in real time via 3D computer graphics on a monitor. The system enabled intuitive recognition of the direction of soft-tissue-generated forces and pressure distribution in three dimensions. Accuracy tests were conducted using a high-accuracy 6-degree-of-freedom positioning device and digital force gauge. The error between the applied loads and measured forces was 3.42 +/- 3.26 N (mean +/- standard deviation) for each coordinate in 10 trials involving load application from 10 different directions. Next, a clinical evaluation was conducted during THA. The relative positions of the cup and stem component were measured using a surgical navigation system simultaneously with the pressure measurement. The system allowed real-time acquisition of information regarding the artificial hip joint, as well as comparison of the differences in the hip condition when several types of neck were used. Further improvements to the calibration method should enable more accurate measurements. We believe this system will be a useful tool for selecting an appropriate implant that fits a patient's hip joint or for estimating the risk of complications following surgery.

0 0
 · 
0 Bookmarks
 · 
24 Views

Keywords

+/- standard deviation
 
1-millisecond intervals
 
artificial hip joint
 
artificial hip joint condition
 
artificial joint
 
ball-and-socket joint
 
contact pressure distribution
 
forms part
 
hip condition
 
hip joint
 
hip joint surfaces
 
load application
 
patient's hip joint
 
pressure distribution
 
pressure sensor module
 
real time
 
real-time acquisition
 
small pressure sensors
 
spherical head component
 
surgical navigation system