Median frequency revisited - An approach to improve a classic spectral electroencephalographic parameter for the separation of consciousness from unconsciousness

University of Duisburg-Essen, Essen, North Rhine-Westphalia, Germany
Anesthesiology (Impact Factor: 6.17). 10/2007; 107(3):397-405. DOI: 10.1097/01.anes.0000278904.63884.4c
Source: PubMed

ABSTRACT In the past, several electroencephalographic parameters have been presented and discussed with regard to their reliability in discerning consciousness from unconsciousness. Some of them, such as the median frequency and spectral edge frequency, are based on classic spectral analysis, and it has been demonstrated that they are of limited capacity in differing consciousness and unconsciousness.
A generalized approach based on the Fourier transform is presented to improve the performance of electroencephalographic parameters with respect to the separation of consciousness from unconsciousness. Electroencephalographic data from two similar clinical studies (for parameter development and evaluation) in adult patients undergoing general anesthesia with sevoflurane or propofol are used. The study period was from induction of anesthesia until patients followed command after surgery and includes a reduction of the hypnotic agent after tracheal intubation until patients followed command. Prediction probability was calculated to assess the ability of the parameters to separate consciousness from unconsciousness.
On the basis of the training set of 40 patients, a new spectral parameter called weighted spectral median frequency was designed, achieving a prediction probability of 0.82 on the basis of the "classic" electroencephalographic frequency range up to 30 Hz. Next, in the evaluation data set, the prediction probability was 0.79, which is higher than the prediction probability of median frequency (0.58) or spectral edge frequency (0.59) and the Bispectral Index (0.68) as calculated from the same data set.
A more general approach of the design of spectral parameters leads to a new electroencephalographic spectral parameter that separates consciousness from unconsciousness significantly better than the Bispectral Index.

Download full-text

Full-text

Available from: Denis Jordan, Jun 21, 2015
0 Followers
 · 
81 Views
  • [Show abstract] [Hide abstract]
    ABSTRACT: For decades, monitoring depth of anesthesia was mainly based on unspecific effects of anesthetics, for example, blood pressure, heart rate, or drug concentrations. Today, electroencephalogram-based monitors promise a more specific assessment of the brain function. To date, most approaches were focused on a "head-to-head" comparison of either electroencephalogram- or standard parameter-based monitoring. In the current study, a multimodal indicator based on a combination of both electro encephalographic and standard anesthesia monitoring parameters is defined for quantification of "anesthesia depth." Two hundred sixty-three adult patients from six European centers undergoing surgery with general anesthesia were assigned to 1 of 10 anesthetic combinations according to standards of the enrolling hospital. The anesthesia multimodal index of consciousness was developed using a data-driven approach, which maps standard monitoring and electroencephalographic parameters into an output indicator that separates different levels of anesthesia from awake to electroencephalographic burst suppression. Obtained results were compared with either a combination of standard monitoring parameters or the electroencephalogram-based bispectral index. The anesthesia multimodal index of consciousness showed prediction probability (PK) of 0.96 (95% CI, 0.95 to 0.97) to separate different levels of anesthesia (wakefulness to burst suppression), whereas the bispectral index had significantly lower PK of 0.80 (0.76 to 0.81) at corrected threshold P value of less than 0.05. At the transition between consciousness and unconsciousness, anesthesia multimodal index of consciousness yielded a PK of 0.88 (0.85 to 0.91). A multimodal integration of both standard monitoring and electroencephalographic parameters may more precisely reflect the level of anesthesia compared with monitoring based on one of these aspects alone.
    Anesthesiology 04/2014; 120(4):819-28. DOI:10.1097/ALN.0000000000000151 · 6.17 Impact Factor
  • Source
    [Show abstract] [Hide abstract]
    ABSTRACT: Awareness during general anesthesia with subsequent explicit recall is a serious and frequently preventable problem that is gaining attention from clinicians and patients alike. Cost-effective interventions that increase vigilance should be implemented to decrease the likelihood of this complication.
    F1000 Medicine Reports 01/2009; 1. DOI:10.3410/M1-9
  • [Show abstract] [Hide abstract]
    ABSTRACT: Automatic sleep staging on an online basis has recently emerged as a research topic motivated by fundamental sleep research. The aim of this paper is to find optimal signal processing methods and machine learning algorithms to achieve online sleep staging on the basis of a single EEG signal. The classification performance obtained using six different EEG signals and various signal processing feature sets is compared using the kappa statistic which has very recently become popular in sleep staging research. A variable duration of the EEG segment (or epoch) to decide on the sleep stage is also analyzed. Spectral-domain, time-domain, linear, and nonlinear features are compared in terms of performance and two types of machine learning approaches (random forests and support vector machines) are assessed. We have determined that frontal EEG signals, with spectral linear features, epoch durations between 18 and 30 seconds, and a random forest classifier lead to optimal classification performance while ensuring real-time online operation.