Article

Calculation of MRI-induced heating of an implanted medical lead wire with an electric field transfer function

School of Electrical and Computer Engineering, Purdue University, West Lafayette, Indiana 47907, USA.
Journal of Magnetic Resonance Imaging (Impact Factor: 2.79). 11/2007; 26(5):1278-85. DOI: 10.1002/jmri.21159
Source: PubMed

ABSTRACT To develop and demonstrate a method to calculate the temperature rise that is induced by the radio frequency (RF) field in MRI at the electrode of an implanted medical lead.
The electric field near the electrode is calculated by integrating the product of the tangential electric field and a transfer function along the length of the lead. The transfer function is numerically calculated with the method of moments. Transfer functions were calculated at 64 MHz for different lengths of model implants in the form of bare wires and insulated wires with 1 cm of wire exposed at one or both ends.
Heating at the electrode depends on the magnitude and the phase distribution of the transfer function and the incident electric field along the length of the lead. For a uniform electric field, the electrode heating is maximized for a lead length of approximately one-half a wavelength when the lead is terminated open. The heating can be greater for a worst-case phase distribution of the incident field.
The transfer function is proposed as an efficient method to calculate MRI-induced heating at an electrode of a medical lead. Measured temperature rises of a model implant in a phantom were in good agreement with the rises predicted by the transfer function. The transfer function could be numerically or experimentally determined.

6 Followers
 · 
138 Views
  • Article: Reply.
    Pacing and Clinical Electrophysiology 03/2012; 35(3):372. DOI:10.1111/j.1540-8159.2011.03293.x · 1.25 Impact Factor
  • [Show abstract] [Hide abstract]
    ABSTRACT: Purpose: To provide a rapid method to reduce the radiofrequency (RF) E-field coupling and consequent heating in long conductors in an interventional MRI (iMRI) setup. Methods: A driving function for device heating (W) was defined as the integration of the E-field along the direction of the wire and calculated through a quasistatic approximation. Based on this function, the phases of four independently controlled transmit channels were dynamically changed in a 1.5 T MRI scanner. During the different excitation configurations, the RF induced heating in a nitinol wire immersed in a saline phantom was measured by fiber-optic temperature sensing. Additionally, a minimization of W as a function of phase and amplitude values of the different channels and constrained by the homogeneity of the RF excitation field (B 1) over a region of interest was proposed and its results tested on the benchtop. To analyze the validity of the proposed method, using a model of the array and phantom setup tested in the scanner, RF fields and SAR maps were calculated through finite-difference time-domain (FDTD) simulations. In addition to phantom experiments, RF induced heating of an active guidewire inserted in a swine was also evaluated. Results: In the phantom experiment, heating at the tip of the device was reduced by 92% when replacing the body coil by an optimized parallel transmit excitation with same nominal flip angle. In the benchtop, up to 90% heating reduction was measured when implementing the constrained minimization algorithm with the additional degree of freedom given by independent amplitude control. The computation of the optimum phase and amplitude values was executed in just 12 s using a standard CPU. The results of the FDTD simulations showed similar trend of the local SAR at the tip of the wire and measured temperature as well as to a quadratic function of W, confirming the validity of the quasistatic approach for the presented problem at 64 MHz. Imaging and heating reduction of the guidewire were successfully performed in vivo with the proposed hardware and phase control. Conclusions: Phantom and in vivo data demonstrated that additional degrees of freedom in a parallel transmission system can be used to control RF induced heating in long conductors. A novel constrained optimization approach to reduce device heating was also presented that can be run in just few seconds and therefore could be added to an iMRI protocol to improve RF safety
    Medical Physics 01/2015; 42(1):359-371. DOI:10.1118/1.4903894 · 3.01 Impact Factor
  • Circulation Arrhythmia and Electrophysiology 04/2013; 6(2):419-428. DOI:10.1161/CIRCEP.113.000116 · 5.42 Impact Factor