Comparison of two column characterisation systems based on pharmaceutical applications.

Katholieke Universiteit Leuven, Laboratorium voor Farmaceutische Analyse, O&N 2, PB 923, Herestraat 49, B-3000 Leuven, Belgium.
Journal of Chromatography (impact factor: 4.53). 06/2008; 1189(1-2):59-71. DOI:10.1016/j.chroma.2008.02.012 pp.59-71
Source: PubMed

ABSTRACT A useful column characterisation system should help chromatographers to select the most appropriate column to use, e.g. when a particular chromatographic column is not available or when facing the dilemma of selecting a suitable column for analysis according to an official monograph. Official monographs of the European Pharmacopoeia and the United States Pharmacopeia are not allowed to mention the brand name of the stationary phase used for the method development. Also given the overwhelming offer of several hundreds of commercially available reversed-phase liquid chromatographic columns, the choice of a suitable column could be difficult sometimes. To support rational column selection, a column characterisation study was started in our laboratory in 2000. In the same period, Euerby et al. also developed a column characterisation system, which is now released as Column Selector by ACD/Labs. The aim of this project was to compare the two existing column characterisation systems, i.e. the KUL system and the Euerby system. Other research groups active in this field will not be discussed here. Euerby et al. developed a column characterisation system based on 6 test parameters, while the KUL system is based on 4 chromatographic parameters. Comparison was done using a set of 63 columns. For 7 different pharmaceutical separations (fluoxetine, gemcitabine, erythromycin, tetracycline, tetracaine, amlodipine and bisacodyl), a ranking was built based on an F-value (KUL method) or Column Difference Factor value (Euerby method) versus a (virtual) reference column. Both methods showed a similar ranking. The KUL and Euerby methods do not perfectly match, but they yield very similar results, allowing with a relatively high certainty, the selection of similar or dissimilar columns as compared to a reference column. An analyst that uses either of the two methods, will end up with a similar ranking. From a practical point of view, it must be noted that the KUL method only includes 4 parameters and 3 chromatographic methods compared to 6 parameters and 4 methods for the Euerby method. Hence, the time needed to determine the chromatographic properties of a column is shorter for the KUL approach. Access to the KUL method also requires no download procedures.

0 0
 · 
0 Bookmarks
 · 
35 Views
  • Source
    Article: Miniaturized protein separation using a liquid chromatography column on a flexible substrate
    [show abstract] [hide abstract]
    ABSTRACT: We report a prototype protein separator that successfully miniaturizes existing technology for potential use in biocompatible health monitoring implants. The prototype is a liquid chromatography (LC) column (LC mini-column) fabricated on an inexpensive, flexible, biocompatible polydimethylsiloxane (PDMS) enclosure. The LC mini-column separates a mixture of proteins using size exclusion chromatography (SEC) with polydivinylbenzene beads (5–20 µm in diameter with 10 nm pore size). The LC mini-column is smaller than any commercially available LC column by a factor of ~11 000 and successfully separates denatured and native protein mixtures at ~71 psi of the applied fluidic pressure. Separated proteins are analyzed using NuPAGE-gel electrophoresis, high-performance liquid chromatography (HPLC) and an automated electrophoresis system. Quantitative HPLC results demonstrate successful separation based on intensity change: within 12 min, the intensity between large and small protein peaks changed by a factor of ~20. In further evaluation using the automated electrophoresis system, the plate height of the LC mini-column is between 36 µm and 100 µm. The prototype LC mini-column shows the potential for real-time health monitoring in applications that require inexpensive, flexible implant technology that can function effectively under non-laboratory conditions.
    Journal of Micromechanics and Microengineering 10/2008; 18(12):125010. · 2.11 Impact Factor

Keywords

3 chromatographic methods
 
4 chromatographic parameters
 
4 methods
 
4 parameters
 
6 parameters
 
6 test parameters
 
63 columns
 
appropriate column
 
brand name
 
Column Selector
 
dissimilar columns
 
download procedures
 
Euerby methods
 
KUL approach
 
particular chromatographic column
 
reference column
 
stationary phase
 
suitable column
 
support rational column selection
 
two methods